Abstract
Background:
The microscope high-power field (HPF) is the cornerstone for histopathology diagnostic evaluation such as the quantification of mitotic figures, lymphocytes, and tumor grading. With traditional light microscopy, HPFs are typically evaluated by quantifying histologic events in 10 fields of view at × 400 magnification. In the era of digital pathology, new variables are introduced that may affect HPF evaluation. The aim of this study was to determine the parameters that influence HPF in whole slide images (WSIs).
Materials and Methods:
Glass slides scanned on various devices (Leica's Aperio GT450, AT2, and ScanScope XT; Philips UltraFast Scanner; Hamamatsu's Nanozoomer 2.0HT; and 3DHistech's P1000) were compared to acquired digital slides reviewed on each vendor's respective WSI viewer software (e.g., Aperio ImageScope, ImageScope DX, Philips IMS, 3DHistech CaseViewer, and Hamamatsu NDP.view) and an in-house developed vendor-agnostic viewer. WSIs were reviewed at “×40” equivalent HPF on different sized monitors with varying display resolutions (1900 × 1080–4500 × 3000) and aspect ratios (e.g., Food and Drug Administration [FDA]-cleared 27” Philips PS27QHDCR, FDA-cleared 24” Dell MR2416, 24” Hewlett Packard Z24n G2, and 28” Microsoft Surface Studio). Digital and microscopic HPF areas were calculated and compared.
Results:
A significant variation of HPF area occurred between differing monitor size and display resolutions with minor differences between WSI viewers. No differences were identified by scanner or WSIs scanned at different resolutions (e.g., 0.5, 0.25, 0.24, and 0.12 μm/pixel).
Conclusion:
Glass slide HPF at × 400 magnification with conventional light microscopy was not equivalent to “×40” digital HPF areas. Digital HPF quantification may vary due to differences in the tissue area displayed by monitor sizes, display resolutions, and WSI viewers but not by scanner or scanning resolution. These findings will need to be further clinically validated with potentially new digital metrics for evaluation.
Keywords: Digital pathology, grade, high-power field, mitosis, whole slide imaging
INTRODUCTION
The microscope high-power field (HPF) is routinely used as a tool in the reporting of certain pathology diagnoses. Pathologists using HPF for standardized quantification of mitosis for tumor grading are the current standard of care.[1] In addition, pathology evaluation using HPF counts aids in determining patient prognosis or management. For example, the number of mitotic figures in an HPF area is a standard reporting metric in patient pathology reports for many tumors. Criteria have been established and clinically validated, with specific quantification ranges, for a variety of tumors in which a mitotic count is essential for determination of tumor grade or for cellular quantification of eosinophils, lymphocytes, and other histopathologic features. Most tumors that require a mitotic count have a three-tiered grading system based on a defined HPF area. For instance, gastrointestinal stromal tumors necessitate counting of 50 HPF overall comprising a 5 mm2 area for evaluation of risk stratification.[2] In comparison, gastrointestinal neuroendocrine tumors, carcinoid tumors of the lung, and invasive ductal carcinoma of the breast require 10 HPFs consisting of a 2.0 mm2 area.[3,4] These metrics are largely based on predigital era publications reporting outcome differences based on histopathology events in selected HPF areas.[5,6,7,8,9,10,11,12,13] In current practice, variables affecting the microscope HPF include objective and ocular magnification differences. While such variables are few, these differences would alter the HPF area presented to the pathologist when performing such tasks.[14]
Recent advances in pathology have established a new standard of care using digital pathology systems. In 2013, the College of American Pathologists (CAP) acknowledged that the usage of digital slides for primary diagnosis was emerging and recommended guidelines for validating a digital pathology system that generated whole slide images (WSIs).[15] In addition, the Food and Drug Administration (FDA) has cleared two WSI platforms (Aperio AT2DX and Philips IntelliSite Pathology Solution) for primary diagnostic purposes.[16,17] Both systems have been extensively validated for a digital primary pathologic evaluation and diagnosis.[18,19,20] However, there is a paucity of literature providing insight into how novel digital workflows affect standardized pathology reporting, specifically related to reporting elements based on the HPF. With a digital workflow, there are differences between glass slide visualization on a traditional light microscope versus a digital representation of the WSI. In a digital pathology system comprising both hardware (e.g., WSI scanner) and software (e.g., image viewer and computer monitor) components, there are several parameters (e.g., scanner optics, camera sensors, and digital resolution) that can contribute to variations affecting representation of the HPF area.[21,22] These parameters can, in turn, affect the evaluation of histopathologic tissue, specifically the area that is represented in a digital HPF compared to an HPF on a bright-field microscope.
The aim of this study was to pilot and evaluate variables by comparing conventional light microscope HPF (at × 400 magnification) to a digital HPF (at “×40” magnification). Potential variables evaluated included the whole slide scanner, scan resolution, image viewer, monitor size, aspect ratio, and display resolution.
MATERIALS AND METHODS
This study evaluated different variables deemed to play a critical role in a digital workflow when reporting patient pathology. These variables included the whole slide scanner device, scan resolution, software image viewer, monitor display resolution, monitor size, and aspect ratio. To evaluate these variables, a precision stage WSI slide calibrator with 0.005 mm and 0.01 mm markers [Figure 1] was scanned at × 40 equivalent resolution on the following commercially available whole slide scanners: (i) Aperio AT2DX (0.25 μm/pixel), Aperio AT2 (0.25 μm/pixel), Aperio ScanScope XT (0.25 μm/pixel), and Aperio AT2 GT450 (0.26 μm/pixel) scanners (Leica Biosystems, Buffalo Grove, IL), (ii) Hamamatsu Nanozoomer 2.0HT (0.23 μm/pixel) scanner (Hamamatsu City, Japan), (iii) P1000 (0.12 μm/pixel) scanner (3DHistech, Budapest, Hungary), and (iv) ultra-fast scanner (UFS) (0.25 μm/pixel) (Philips, Amsterdam, The Netherlands). In addition, this calibrator slide was scanned at a “×20” equivalent resolution (0.5 and 0.46 μm/pixel) on the Aperio AT2, Aperio AT2DX, Aperio ScanScope XT, and Hamamatsu Nanozoomer. The Aperio GT450 and Philips UFS can only scan at “×40” equivalent resolutions.
All WSI file types were recursively analyzed using six different slide viewers (ImageScope v12.4 and ImageScope DX v1.0 from Leica, Philips IMS v3.2, CaseViewer v2.3 from 3DHistech, NDP.view v2.7 from Hamamatsu, and a homegrown MSK Slide Viewer v1.3). Four monitors were used for this study and consisted of an FDA-cleared 24” Dell monitor MR2416 (aspect ratio, 16:10; pixel pitch, 0.27 mm), consumer-grade 24” Hewlett Packard Z24n G2 monitor (aspect ratio, 16:10; pixel pitch, 0.27 mm), FDA-cleared 27” Philips monitor PS27QHDCR (aspect ratio, 16:9; pixel pitch, 0.2331 mm), and a 28” Microsoft Surface Studio monitor (aspect ratio, 3:2). Screen resolutions of 1920 × 1080 and 1920 × 1200 pixels were analyzed for all monitors and WSI viewers. Of note, the Microsoft Surface Studio's monitor is part of an all-in-one computer that is unique compared to the other monitors. Other monitors were reviewed on a 64-bit Hewlett Packard computer with an i5 3.2 GHz processor. For those monitors with high resolution (Philips monitor and Microsoft Surface Studio), additional higher resolutions were tested: 2560 × 1440 and 4500 × 3000 pixels, respectively. For the Microsoft Surface Studio monitor, application and text scaling were evaluated at 4500 × 3000 pixels native resolution, with the WSI viewer evaluated at both 100% and 200% (employing Windows 10 operating system). Display resolutions were modified using the Windows operating system Control Panel.
In addition, to the monitors described previously, the workstation included a standard keyboard and mouse to interact with the slides. The monitors were connected to a Hewlett Packard EliteDesk 800 with an Intel core i5, 3.2 GHz processor, and 8 GB random access memory.
For each combination of variables, the horizontal and vertical screen area for a digital × 40 field was measured using respective WSI software measurement tools. The “X” and “Y” axis measurements on the image viewer for the × 40 designated field were recorded and the total area calculated (Area digital HPF = Height × Width). Corresponding light microscope (Olympus BX43) ×40 HPFs were also measured, calculated (Area microscope HPF = π × radius2), and recorded. Further, comparisons were made between the × 40 digital HPF for each platform and WSI viewer with a × 10 ocular 20 mm and 22 mm diameter lens. Digital and microscopic HPF areas were then compared and quantified to calculate the equivalent number of digital versus microscope HPF.
RESULTS
The glass slides were successfully scanned on each whole slide scanner at scan resolutions of 0.5, 0.46, 0.26, 0.25, 0.23, and 0.12 μm/pixel. Each respective WSI, WSI viewer, whole slide scanner, monitor, and screen resolution were assessed with all available permutations permitted based on vendor interoperability. These digital HPF area combinations showed divergent results described below. A summary of the results for each variable is shown in Table 1.
Table 1.
Resolution | Monitor Size and Resolution | ||||||
---|---|---|---|---|---|---|---|
24” HP (aspect ratio 16:10) | 24” Dell (aspect ratio 16:10) | 27” Philips (aspect ratio 16:9) | |||||
1920×1080 (mm2) | 1920×1200 (mm2) | 1920×1080 (mm2) | 1920×1200 (mm2) | 1920×1080 (mm2) | |||
Whole slide image viewer (X and Y dimensions=area) | ImageScope | 0.245×0.478=0.117 | 0.275×0.478=0.131 | 0.245×0.478=0.117 | 0.275×0.478=0.131 | 0.235×0.477=0.112 | |
Philips IMS | 0.236×0.479=0.113 | 0.265×0.479=0.127 | 0.236×0.479=0.113 | 0.265×0.479=0.127 | 0.237×0.479=0.114 | ||
Case Viewer | 0.248×0.479=0.119 | 0.277×0.479=0.133 | 0.248×0.479=0.119 | 0.277×0.479=0.133 | 0.248×0.479=0.119 | ||
NDP.view | 0.245×0.437=0.107 | 0.274×0.437=0.120 | 0.245×0.437=0.107 | 0.274×0.437=0.120 | 0.246×0.438=0.108 | ||
MSK Slide Viewer | 0.23×0.48=0.110 | 0.27×0.48=0.130 | 0.23×0.48=0.11 | 0.27×0.48=0.130 | 0.24×0.48=0.115 | ||
ImageScope DX | NA | NA | 0.2345×0.4731=0.111 | 0.2641×0.4731=0.125 | NA | ||
Monitor Size and Resolution | |||||||
27” Philips (aspect ratio 16:9) | 28” Microsoft studio (aspect ratio 3:2) | ||||||
1920×1200 (mm2) | 2560×1440 (mm2) | 1920×1080 (mm2) | 1920×1200 (mm2) | 4500×3000 (mm2) | 4500×3000* 200% (mm2) | ||
Whole slide image viewer (X and Y dimensions=area) | ImageScope | 0.265×0.477=0.126 | 0.335×0.638=0.214 | 0.246×0.479=0.118 | 0.275×0.479=0.132 | 1.469×2.249=3.3 | 0.715×1.12=0.801 |
Philips IMS | 0.267×0.479=0.128 | 0.326×0.639=0.208 | 0.236×0.479=0.113 | 0.266×0.479=0.127 | 0.716×1.12=0.802 | 0.344×0.561=0.193!error | |
Case Viewer | 0.278×0.479=0.133 | 0.338×0.639=0.216 | 0.248×0.479=0.119 | 0.278×0.479=0.133 | 1.47×2.247=3.3 | 0.353×0.562=0.198 | |
NSP.view | 0.274×0.436=0.119 | 0.328×0.584=0.192 | 0.224×0.436=0.098 | 0.273×0.436=0.119 | 0.684×1.03=0.705 | 0.684×1.03=0.705 | |
MSK Slide Viewer | 0.27×0.48=0.130 | 0.33×0.64=0.211 | 0.23×0.48=0.110 | 0.27×0.48=0.130 | 0.72×1.12=0.806 | 0.34×0.56=0.190 | |
ImageScope DX | NA | NA | NA | NA | NA | NA |
As the resolution and screen size increases so does the×40 digital HPF area. *One web-based viewer was not compatible with the high-resolution monitor at 200% scaling, shown in Figure 5. (HPF measurements differ, however, the represented tissue area is not reflected accurately). HPF: High-power field
Whole slide image viewer software
When controlling each of the four different monitors at the same screen resolution (e.g., 1920 × 1080 or 1920 × 1200), the visualized HPF area of the WSI on each respective WSI viewer at a digital × 40 magnification differed by a factor of up to 21%. A digital × 40 HPF using Leica's Aperio ImageScope and ImageScope DX, Philips IMS, 3DHistech's CaseViewer, Hamamatsu's NDP.view, and the MSK Slide Viewer was viewed on the 24” Dell monitor at a resolution of 1920 × 1080 resulting in areas of 0.117 mm2, 0.111 mm2, 0.113 mm2, 0.119 mm2, 0.107 mm2, and 0.110 mm2, respectively. Of note, the ImageScope DX viewer can only be used on the respective Dell monitor; when attempting to launch ImageScope DX on other monitors, an error message displays and does not launch the WSI. Table 1 shows the HPF area presented in each of the tested WSI viewers on each respective monitor at 1920 × 1080 and 1920 × 1200 pixel resolutions.
Monitor size and display resolution
Higher monitor resolutions were tested on the 27” Philips monitor as well as the 28” Microsoft Studio monitor. With default display resolution set at a 2560 × 1440 on the Philips monitor, the HPF areas calculated are shown in Table 1. Of note, these values were captured at the default (recommended) 200% scaling. Scaling is an operating system setting that adjusts the sizes of windows, images, icons, and text relative to the monitor resolution or pixel density. Scaling showed differences for desktop application and native text (only tested on the Microsoft Surface Studio). Scaling was modified at a 4500 × 3000 resolution to 100% and 200%. Scaling at 100% resulted in a larger area as the monitor scaled the applications and text halved their size. At 100% scaling, a digital × 40 HPF area was 3.30 mm2 (ImageScope), 0.802 mm2 (Philips IMS), 3.30 mm2 (CaseViewer), 0.705 mm2 (NDP.view), and 0.806 mm2 (MSK Slide Viewer).
Scanning resolution
The different whole slide scanners tested had varying scanning resolutions. The scanning resolutions for each scanner are 0.25 and 0.5 μm/pixel at “×20” and “×40” scan resolution for the Aperio ScanScope XT, Aperio AT2, and for the Aperio AT2DX; 0.26 μm/pixel at “×40” scan resolution for the Aperio GT450; 0.23 and 0.46 μm/pixel at “×20” and “×40” scan resolution for the Hamamatsu Nanozoomer 2.0HT; 0.12 μm/pixel at “×40” scan resolution for the 3DHistech P1000; and 0.25 μm/pixel at “×40” scan resolution for the Philips UFS. It is worth mentioning that the reported scanning resolution of each platform is rounded, and small differences may not be accounted for as such. Each respective WSI was evaluated at a × 40 digital HPF on all monitors and using all WSI viewer software. There was no change in the presented HPF area for WSIs scanned at the different μm/pixel resolutions. Controlling for WSI scan resolution, a single WSI was viewed on each monitor, respectively. The presented HPF area was the same on each respective monitor when the other variables remained constant (e.g., screen size, aspect ratio, and display resolution), regardless of the scanning μm/pixel resolution. Of note, ImageScope and ImageScope DX did not have the ability to display a × 40 magnification in the viewport when the WSI was scanned at a “×20” equivalent resolution (i.e., digital zoom).
Comparison of digital HPF to glass HPF
The various digital HPF areas at × 40 were compared to the area encompassed in 10 HPFs, the value typically required for many histopathologic quantifications (e.g., mitotic figure counts) using a light microscope within a 2 mm2 area. The number of equivalent HPFs for each respective monitor and display resolution is shown in Table 2. These represent the number of digital × 40 HPFs to equal 10 HPFs on a × 40 light microscope. An overlay of the area seen in one 27” Philips monitor at 1920 × 1200 display resolution HPF over a × 40 light microscope HPF using a × 10 ocular is shown in Figure 2. The largest monitor and highest resolution of monitors evaluated was the 28” Microsoft Surface Studio, and with 4500 × 3000 resolution at 100% scaling, the number of areas to achieve the same area as 10 HPFs on the light microscope was 0.6 (ImageScope), 2.5 (Philips IMS), 0.6 (CaseViewer), 2.8 (NDP.view), and 2.5 (MSK Slide Viewer) with each viewer. At 200% scaling for the Microsoft Studio monitor, the number of areas required was 2.5 (ImageScope), 10.4 (Philips IMS), 10.1 (CaseViewer), 2.8 (NDP.view), and 10.5 (MSK Slide Viewer).
Table 2.
Monitor | Resolution | Whole slide image viewer (number of digital HPF to achieve 10 standard microscopic HPF [2 mm2]) | |||||
---|---|---|---|---|---|---|---|
ImageScope | Philips IMS | Case Viewer | NDP view | MSK slide viewer | ImageScope DX | ||
24” HP (aspect ratio 16:10) | 1920×1080 | 17.1 | 17.7 | 16.8 | 18.7 | 18.2 | 18.0 |
1920×1200 | 15.3 | 15.7 | 15.0 | 16.7 | 15.4 | 16.0 | |
24” Dell (aspect Ratio 16:10) | 1920×1080 | 17.1 | 17.7 | 16.8 | 18.7 | 18.2 | 18.0 |
1920×1200 | 15.3 | 15.7 | 15.0 | 16.7 | 15.4 | 16.0 | |
27” Philips (aspect ratio 16:9) | 1920×1080 | 17.9 | 17.5 | 16.8 | 18.5 | 17.4 | NA |
1920×1200 | 15.9 | 15.6 | 15.0 | 16.8 | 15.4 | NA | |
2560×1440 | 9.3 | 9.6 | 9.3 | 10.4 | 9.5 | NA | |
28” Microsoft Studio (aspect ratio 3:2) | 1920×1080 | 16.9 | 17.7 | 16.8 | 20.4 | 18.2 | NA |
1920×1200 | 15.2 | 15.7 | 15.0 | 16.8 | 15.4 | NA | |
4500×3000 | 0.6 | 2.5 | 0.6 | 2.8 | 2.5 | NA | |
4500×3000* | 2.5 | 10.4 | 10.1 | 2.8 | 10.5 | NA |
This was extrapolated from values in Table 1 showing various 40×HPF areas for each monitor and combination of display resolution and image viewer. *At 200% scaling. HPF: High-power field
To compare the digital HPF to standard × 40 objectives of most light microscopes, field areas for a × 40 digital HPF were compared to that of a × 40 field seen on a conventional bright-field microscope (e.g., Olympus BX43). These results are reported in Table 3. The difference in area ranged between -50% and 1586% comparing display resolution to a × 10 ocular lens with 20 mm and 22 mm diameter on a light microscope.
Table 3.
Monitor | Slide viewer | “×40” HPF standard resolution | Microscope (HPF) area (mm2) | |||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|
Aperio AT2 (SVS) | Aperio GT450 (SVS) | Aperio AT2DX (SVS) | Philips (iSyntax) | 3DHistech (MRXS) | Hamamatsu (NDPI) | WSI Viewer; “×40” HPF area | ×10 ocular/20 mm diameter; ×40 objective | ×10 ocular/22 mm diameter; ×40 objective | Delta compared to ×10 ocular, 20 mm diameter, 40× objective (%) | Delta compared to ×10 ocular, 22 mm diameter, 40× objective (%) | ||
24” HP (Aspect ratio 16:9) | ImageScope | 0.275×0.478 | 0.275×0.478 | 0.275×0.478 | 0.275×0.478 | 0.275×0.478 | 0.131 | 0.196 | 0.238 | −33 | −45 | |
Philips IMS | 0.265×0.479 | 0.127 | 0.196 | 0.238 | −35 | −47 | ||||||
CaseViewer | 0.277×0.479 | 0.277×0.479 | 0.277×0.479 | 0.277×0.479 | 0.277×0.479 | 0.133 | 0.196 | 0.238 | −32 | −44 | ||
NDP.view | 0.274×0.437 | 0.120 | 0.196 | 0.238 | −39 | −50 | ||||||
MSK Slide Viewer | 0.27×0.48 | 0.27×0.48 | 0.27×0.48 | 0.27×0.48 | 0.27×0.48 | 0.27×0.48 | 0.130 | 0.196 | 0.238 | −34 | −46 | |
ImageScope DX | 0.125 | 0.196 | 0.238 | −36 | −47 | |||||||
24” Dell (Aspect ratio 16:10) | ImageScope | 0.275×0.478 | 0.275×0.478 | 0.275×0.478 | 0.275×0.478 | 0.275×0.478 | 0.131 | 0.196 | 0.238 | −33 | −45 | |
Philips IMS | 0.265×0.479 | 0.127 | 0.196 | 0.238 | −35 | −47 | ||||||
CaseViewer | 0.277×0.479 | 0.277×0.479 | 0.277×0.479 | 0.277×0.479 | 0.277×0.479 | 0.133 | 0.196 | 0.238 | −32 | −44 | ||
NDP.view | 0.274×0.437 | 0.120 | 0.196 | 0.238 | −39 | −50 | ||||||
MSK Slide Viewer | 0.27×0.48 | 0.27×0.48 | 0.27×0.48 | 0.27×0.48 | 0.27×0.48 | 0.27×0.48 | 0.130 | 0.196 | 0.238 | −34 | −46 | |
ImageScope DX | 0.264×0.473 | 0.125 | 0.196 | 0.238 | −36 | −47 | ||||||
27” Philips (Aspect ratio 16:9) | ImageScope | 0.335×0.638 | 0.335×0.638 | 0.335×0.638 | 0.335×0.638 | 0.335×0.638 | 0.214 | 0.196 | 0.238 | 9 | −10 | |
Philips IMS | 0.326×0.639 | 0.208 | 0.196 | 0.238 | 6 | −12 | ||||||
CaseViewer | 0.338×0.639 | 0.338×0.639 | 0.338×0.639 | 0.338×0.639 | 0.338×0.639 | 0.216 | 0.196 | 0.238 | 10 | −9 | ||
NDP.view | 0.328×0.584 | 0.192 | 0.196 | 0.238 | −2 | −20 | ||||||
MSK Slide Viewer | 0.33×0.64 | 0.33×0.64 | 0.33×0.64 | 0.33×0.64 | 0.33×0.64 | 0.33×0.64 | 0.211 | 0.196 | 0.238 | 8 | −11 | |
ImageScope DX | 0.196 | 0.238 | ||||||||||
28” Microsoft Studio (Aspect ratio 3:2) | ImageScope | 1.469×2.249 | 1.469×2.249 | 1.469×2.249 | 1.469×2.249 | 1.469×2.249 | 3.304 | 0.196 | 0.238 | 1586 | 1288 | |
Philips IMS | 0.716×1.12 | 0.802 | 0.196 | 0.238 | 309 | 237 | ||||||
CaseViewer | 1.47×2.247 | 1.47×2.247 | 1.47×2.247 | 1.47×2.247 | 1.47×2.247 | 3.303 | 0.196 | 0.238 | 1585 | 1288 | ||
NDP.view | 0.684×1.03 | 0.705 | 0.196 | 0.238 | 259 | 196 | ||||||
MSK Slide Viewer | 0.72×1.12 | 0.72×1.12 | 0.72×1.12 | 0.72×1.12 | 0.72×1.12 | 0.72×1.12 | 0.806 | 0.196 | 0.238 | 311 | 239 | |
ImageScope DX | 0.196 | 0.238 |
This dimension was then compared to the diameter of a 10×/20 mm and 10×/22 mm, 40×objective by calculating the delta as a percentage change. HPF: High-power field
DISCUSSION
Counting HPFs using microscopy in pathology is essential in the evaluation of certain diseases (e.g., eosinophil and lymphocyte counts) and tumor type grading (e.g., mitotic figure counting). Specific histologic quantifications, such as counting of mitotic figures, are highly dependent on the precise area that is presented to the pathologist in a defined HPF.[14] Most conventional bright-field microscopes use a customary ocular lens (e.g., 20 mm or 22 mm) and magnification objectives that allow for a standardized and translatable definition for microscopic HPF. Using the tissue area viewed through a bright-field microscope, the mitotic count relies on the number of mitotic figures found based on a set number of HPFs. A mitotic count is quantified as the number of mitoses per the number of HPFs evaluated. Higher mitotic rates generally equate to a higher grade and more aggressive tumor biology.
Recent advances in care delivery methods have allowed pathologists to render pathologic diagnoses based on patient tissue samples such as biopsies and resections using a digital workflow.[23,24,25,26] Digital pathology using whole slide imaging to generate a digitized version of a glass slide with high resolution has been proven to be noninferior to its glass counterpart.[27,28,29,30] Digital pathology is a tremendous asset to the pathologist by providing novel tools for histopathology review, annotation and objective quantification for diagnostic evaluation, and computer-assisted image algorithms for analysis. However, from a regulatory perspective, vendor hardware and software thus far have been marketed to only use a specific display device (i.e., computer monitor) with accompanying vendor-specific WSI viewing software. Nonetheless, several pathology departments around the world have initiated primary diagnosis in pathology using digital means.[18,28,29] Therefore, a comparison of diagnostic tools that were formally performed using a traditional light microscope must now be re-evaluated in this emerging digital environment. One factor that warrants critical reappraisal is the HPF and its role in the histologic assessment of disease.
Our study sought to investigate if any differences between conventional light microscope and digital modalities exist, to provide guidance in reporting of pathology parameters using the HPF. The differences we uncovered suggest a need for either additional software development to ensure comparable tissue presentation of a single HPF to a pathologist using WSI or that formal pathologic criteria/scoring systems be re-evaluated to accommodate digital pathology. Pathologists may reference the results in the tables published in this study for identification of HPF area, especially for comparison between varied workstation settings between pathology departments. Investigating possible permutations of potential variables that may impact the HPF, we first identified the different components of a digital pathology system that could affect the area viewed in a × 400 microscopic field (×40 objective). Using digital pathology, the ocular and objective lenses of a light microscope are substituted by different digital components, such as the whole slide scanner, digital camera used to acquire an image (e.g., sensor resolution), image scanning resolution (pixels/micron), digital image file (e.g., image compression and file type), WSI viewing software, and display monitor (i.e., differing display resolutions, monitor size, and aspect ratio). The aim of this study was to evaluate how these differences may affect the pathologic evaluation of tissue features reliant on HPFs. The results overall showed that there is wide variability in the HPF area using different hardware and software and that such differences in tissue area presented to a pathologist are divergent from those seen using a light microscope [Figure 3]. Using standard monitor configurations commonly used in clinical settings, the screen resolutions studied were between 1920 × 1080 and 4500 × 3000. The lowest display resolution (e.g., 1920 × 1080) resulted in a digital HPF that was nearly 50% smaller than that seen with a light microscope HPF. This would hypothetically require almost twice the amount of HPF tissue to be microscopically evaluated to be comparable to an equivalent evaluation on a light microscope. For example, if 10 HPFs are required to adequately evaluate a mitotic count on a light microscope, a 1920 × 1080 resolution display would require around 20 HPF to be assessed. In contrast, the highest display resolution (e.g., Microsoft Surface Studio; 4500 × 3000 at 100% scaling) evaluated could require less than a single HPF (e.g., 0.6 HPF) to display all of the pixels necessary for evaluating an area comprising 10 conventional light microscopic HPFs. Trying to maximize the tissue area shown across each WSI viewer, this study reports small differences identified between the various software viewers due to user interface variances (e.g., toolbar placements and design layout). However, in a given WSI viewer, there may be significant differences in tissue area in the viewport based on the user interface [Figure 4]. Therefore, depending on the display resolution, monitor size, and image viewer that a pathologist uses when viewing a WSI, there may be differences in the size of the tissue area a pathologist evaluates when engaged in a digital workflow. Understanding these differences and validating these alterations are necessary to ensure that the tasks pathologists are required to perform (e.g., tumor grading) are appropriate and accurate. Pathologists who are unaware of the impact different variables may have on the HPF can lead to an over- or underestimation of mitotic counts, eosinophil quantification, lymphocyte quantitation, and so on when rendering pathology reports.
An additional variable inherent with higher resolution displays is scaling. Scaling is a variable dependent on the operating system and application utilized for image viewing. The purpose of scaling exists for ultra-high-resolution monitors, as text and icons can be very hard to read at the native high resolution, and they are also often displayed at a much smaller size (and less readable) than at a lower resolution. Therefore, certain elements are “scaled” up to be more user friendly and thereby improve the readable viewing experience. Hence, with higher resolution monitors, the pixel density can change the visualization of items presented on the screen. Scaling is dependent and inherent to the application development and may affect both desktop and web-based viewers. Of the desktop viewers tested on multiple monitors: ImageScope, CaseViewer, and NDP.view, scaling affected ImageScope and CaseViewer but not NDP.view. The Microsoft Surface Studio monitor used in this study showed a significant difference between 100% and 200% scaling at the same resolution and using the same image viewer software compared to other displays in all except NDP.view WSI viewer. The Microsoft Surface Studio with scaling at 100% showed 3.30 mm2, and at 200%, scaling showed 0.801 mm2 HPF areas using the ImageScope viewer but retained the same HPF area for both 100% and 200% scaling (e.g., 0.705 mm2) with the NDP.view software, meaning the NDP.view WSI viewer was not affected by scaling. Scaling also affected web-based viewers such as the Philips IMS and MSK Viewer, as the browser application interface changes. Similarly, web-based viewers have additional inherent web browser setting controls that may change HPF tissue images which are presented to the pathologist. Based on compatibility, some WSI viewers may not be programmed to open on other monitors other than those that were FDA-cleared for, such as the ImageScope DX viewer. In addition, proprietary file format compatibility with the WSI viewer may not allow for viewing of certain vendor WSI on other WSI viewers, such as Philips viewing of non-Philips WSI on the Philips IMS viewer. Of note, the MSK Slide Viewer was the only WSI viewer to have the functionality to view and navigate all WSIs from every vendor tested. This is critical for institutions with multiple vendor whole slide scanners to have a consistent user interface for the various WSI from each vendor.
Not all light microscopes have similar components to visualize the exact HPF area. This is due to differences in their ocular lenses and objectives. For this reason, for certain cancer checklist protocols (e.g., invasive breast carcinoma), the CAP offers a tabular guide for how mitotic counts vary using various microscope configurations.[31] As each light microscope may vary, the HPF evaluated that will, in turn, affect the mitotic count, and the CAP cancer protocols accordingly offer different mitotic cutoffs based on the light microscope field diameter and area. In addition, these CAP cancer protocols give guidance on how to determine the field diameter and area. Without such guidance in a digital workflow, diagnoses that rely on microscopic assessments using the HPF may result in wide variations as a result of differences with the hardware and software used. Lower resolution screens in this study had a smaller area compared to the microscope HPF with × 40 objective in both a 20 mm and 22 mm × 10 lens. In our study, as the display resolution increased, changes were observed up to 1586 % with the Microsoft Surface Studio monitor, compared to the HPF area of a microscope. A single HPF with the highest resolution monitor evaluated in this study was equivalent to a 16.9 HPF area of a conventional microscope (×10/20 mm ocular, ×40 objective).
This study also identified minor differences in the HPF area as a result of image viewer software. This difference can be attributed to variations with each viewer application interface (e.g., toolbars and navigation panes). Further, each viewer has different elements presented on the screen that may obscure tissue being viewed [Figure 5]. With a conventional light microscope, there is no toolbar or overlay superimposed on the image field. For the purposes of this study, the user interface in each software viewer and operating system (e.g., Windows taskbar) were minimized as much as possible to maximally display the WSI viewport tissue area. Ideally, an image viewer could standardize a digital HPF by inserting bounding boxes of a fixed input size (i.e., 2 mm2) over an image. However, no such feature yet exists in any of the image viewers we tested. Fortunately, the different image viewers had a minor impact on HPF area in comparison to the differences seen with display resolution and monitor size.
Currently, in the United States, there are only two FDA-cleared commercial platforms available for primary diagnosis by WSI.[16,17] These platforms are the Philips IntelliSite Pathology Solution and the Aperio AT2Dx system. This study compared the Philips system to other imaging platforms as well as a standard light microscope. Interestingly, without hiding the Windows taskbar (as most users have configured), the Philips 27” monitor with 2560 × 1440 display resolution when combined with Leica's Aperio ImageScope software created the most similar field area to that of a light microscope with 10.15 HPF required to equal 10 HPFs on a light microscope; using the FDA-cleared Philips IMS viewer with the same Philips 27” monitor 9.71 digital HPF was found to be equivalent to 10 HPFs of a light microscope (not shown in the table). In addition, the FDA-cleared Dell MR2416 monitor has the exact same specifications (i.e., monitor size, resolution, and aspect ratio) as the consumer-grade standard hospital configuration monitor tested (HP z24n G2) and similarly showed identical HPF tissue areas. Lower resolutions and smaller monitor sizes revealed smaller HPF areas compared to the FDA-cleared Philips system, while larger size monitors as well as higher resolutions revealed larger HPF areas. Therefore, the greatest variation in HPF was seen with changes in display resolution and to a lesser extent monitor size. There was minimal variation with the various image viewers and no observed difference due to WSI scanners and at varying scanning resolutions. To illustrate the variation and effect of monitor sizes and display resolution on HPF when using the same WSI scanned on a Philips UFS, the WSI viewed on a 24” HP monitor at 1920 × 1080 required 17.7 HPF to equal 10 light microscope HPFs while on a 28” Microsoft Studio monitor at 4500 × 3000 with 100% scaling required only 2.5 HPF fields. When comparing the same slide scanned on Philips UFS viewed on the 27” Philips monitor using different image viewer software, there was only a minimal difference ranging from 9.3 to 10.4 HPF required to equal 10 light microscope HPFs.
For an FDA-cleared digital pathology system intended for primary diagnosis use, current approvals include a packaged end-to-end digital closed system that includes a whole slide scanner, specific monitor, and proprietary WSI viewer software. The data from this study suggest that as long as display resolution and monitor size remain equivalent when substituting components, little variation in HPF size is likely to be caused by different types of whole slide scanner or WSI viewer. Therefore, pathologists can be assured that the HPF area is relatively unchanged when a digital slide generated by the FDA-cleared Philips UFS is viewed on a non-FDA-cleared monitor of similar specifications to the Philips recommended monitor (27”, 2560 × 1400) or if examined with a non-Philips image viewer. These data thus suggest that current regulatory restrictions that prevent interchanging the components of a digital pathology system may not be warranted. As long as a digital pathology system is appropriately validated,[15,32] regardless of the type of slide scanner, monitor, or WSI viewer used, it is probable that a pathologist's ability to render an accurate diagnosis will not be hindered.
In summary, as digital pathology becomes increasingly adopted and pathologists turn to primary diagnosis using a digital workflow, ensuring clinical reproducibility in regard to HPF area measurements is necessary. This study demonstrates critical components of the digital workflow and respective effects of each component on HPF area representation for clinical reporting of patient pathology. The future role of digital and computational pathology in areas such as HPF assessment of pathology parameters needs to be considered, analyzed, and standardized.[33,34] It is plausible the ground truth and outcomes used today may no longer be applicable in a digital workflow where novel metrics using artificial intelligent tools could be instituted (i.e., automated mitotic figure quantification and ratio of mitoses per tissue area or per number of tumor cells).[35] Notably, this gives an opportunity to validate these tools against prospective clinical cohorts and comparatively evaluate the data to current pathology grading systems and outcomes.[36]
CONCLUSION
As the pathology community transitions into a fully digitized workflow for primary diagnosis, pathologists must be aware of technical differences between available diagnostic methodologies. While digital slides provide the pathologist with innovative diagnostic tools, pathologists should evaluate and assess the current standard of care with that of the digital era. This study showed that glass slide HPF at × 400 magnification on a light microscope was not equivalent to “×40” digital HPF areas. Tumor grading and other features requiring quantification dependent on HPFs may differ when using WSI. Digital HPF quantification may vary due to differences in the tissue area displayed by monitor sizes, display resolutions, and WSI viewers but not by scanner or scanning resolution. Furthermore, clinical validation of quantification of pathology parameters in HPF areas needs to be performed to ensure appropriate diagnostic standardization and reproducibility for diagnoses performed today.
Financial support and sponsorship
This research was funded in part through the NIH/NCI Cancer Center Support Grant P30 CA008748.
Conflicts of interest
Liron Pantanowitz is a consultant for Hamamatsu. Matthew Hanna is a consultant for Paige and advisor for PathPresenter.
Footnotes
Available FREE in open access from: http://www.jpathinformatics.org/text.asp?2020/11/1/33/297572
REFERENCES
- 1.Facchetti F. A proposal for the adoption of a uniform metrical system for mitosis counting. Int J Surg Pathol. 2005;13:157–9. doi: 10.1177/106689690501300205. [DOI] [PubMed] [Google Scholar]
- 2.Rubin BP, Blanke CD, Demetri GD, Dematteo RP, Fletcher CD, Goldblum JR, et al. Protocol for the examination of specimens from patients with gastrointestinal stromal tumor. Arch Pathol Lab Med. 2010;134:165–70. doi: 10.5858/134.2.165. [DOI] [PubMed] [Google Scholar]
- 3.Lakhani SR, Ellis IO, Schnitt SJ, Tan PH, van de Vijver MJ. WHO Classification of Tumours of the Breast. 4th ed (IARC, ed) Lyon, France: IARC Press; 2012. [[Last accessed on 2020 Apr 14]]. Available from: http://publicationsiarcfr/Book-And-Report-Series/Who-Iarc-Classification-Of-Tumours/Who-Classification-Of-Tumours-Of-The-Breast-2012 . [Google Scholar]
- 4.Bosman FT, Carneiro F, Hruban RH, Theise ND. WHO Classification of Tumours of the Digestive System? 4th ed Lyon. France: IARC Press: WHO-OMS; 2010. [[Last accessed on 2020 Apr 14]]. Available from: http://appswhoint/bookorders/anglais/detart1jspsesslan=1&codlan=1&codcol=70&codcch=4003 . [Google Scholar]
- 5.Perry A, Stafford SL, Scheithauer BW, Suman VJ, Lohse CM. Meningioma grading: An analysis of histologic parameters. Am J Surg Pathol. 1997;21:1455–65. doi: 10.1097/00000478-199712000-00008. [DOI] [PubMed] [Google Scholar]
- 6.CAP. Protocol for the Examination of Specimens from Patients with Neuroendocrine Tumors (Carcinoid Tumors) of the Stomach CAP. 2012. [[Last accessed on 2020 Apr 14]]. Available from: https://webappscaporg/apps/docs/committees/cancer/cancer_protocols/2012/StomachNET_12protocol_3200pdf .
- 7.Robbins P, Pinder S, de Klerk N, Dawkins H, Harvey J, Sterrett G, et al. Histological grading of breast carcinomas: A study of interobserver agreement. Hum Pathol. 1995;26:873–9. doi: 10.1016/0046-8177(95)90010-1. [DOI] [PubMed] [Google Scholar]
- 8.Fletcher CD, Berman JJ, Corless C, Gorstein F, Lasota J, Longley BJ, et al. Diagnosis of gastrointestinal stromal tumors: A consensus approach. Hum Pathol. 2002;33:459–65. doi: 10.1053/hupa.2002.123545. [DOI] [PubMed] [Google Scholar]
- 9.Metter GE, Nathwani BN, Burke JS, Winberg CD, Mann RB, Barcos M, et al. Morphological subclassification of follicular lymphoma: Variability of diagnoses among hematopathologists, a collaborative study between the repository center and pathology panel for lymphoma clinical studies. J Clin Oncol. 1985;3:25–38. doi: 10.1200/JCO.1985.3.1.25. [DOI] [PubMed] [Google Scholar]
- 10.Furuta GT, Liacouras CA, Collins MH, Gupta SK, Justinich C, Putnam PE, et al. Eosinophilic esophagitis in children and adults: A systematic review and consensus recommendations for diagnosis and treatment. Gastroenterology. 2007;133:1342–63. doi: 10.1053/j.gastro.2007.08.017. [DOI] [PubMed] [Google Scholar]
- 11.Greenson JK, Huang SC, Herron C, Moreno V, Bonner JD, Tomsho LP, et al. Pathologic predictors of microsatellite instability in colorectal cancer. Am J Surg Pathol. 2009;33:126–33. doi: 10.1097/PAS.0b013e31817ec2b1. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12.Guillou L, Coindre JM, Bonichon F, Nguyen BB, Terrier P, Collin F, et al. Comparative study of the National Cancer Institute and French Federation of Cancer Centers Sarcoma Group grading systems in a population of 410 adult patients with soft tissue sarcoma. J Clin Oncol. 1997;15:350–62. doi: 10.1200/JCO.1997.15.1.350. [DOI] [PubMed] [Google Scholar]
- 13.Travis WD, Rush W, Flieder DB, Falk R, Fleming MV, Gal AA, et al. Survival analysis of 200 pulmonary neuroendocrine tumors with clarification of criteria for atypical carcinoid and its separation from typical carcinoid. Am J Surg Pathol. 1998;22:934–44. doi: 10.1097/00000478-199808000-00003. [DOI] [PubMed] [Google Scholar]
- 14.Yigit N, Gunal A, Kucukodaci Z, Karslioglu Y, Onguru O, Ozcan A. Are we counting mitoses correctly? Ann Diagn Pathol. 2013;17:536–9. doi: 10.1016/j.anndiagpath.2013.05.005. [DOI] [PubMed] [Google Scholar]
- 15.Pantanowitz L, Sinard JH, Henricks WH, Fatheree LA, Carter AB, Contis L, et al. Validating whole slide imaging for diagnostic purposes in pathology: Guideline from the College of American Pathologists Pathology and Laboratory Quality Center. Arch Pathol Lab Med. 2013;137:1710–22. doi: 10.5858/arpa.2013-0093-CP. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16.US Food and Drug Administration. Intellisite Pathology Solution (PIPS, Philips Medical Systems) US Food and Drug Administration. 2017. [[Last accessed on 2020 Apr 14]]. Available from: https://wwwfdagov/drugs/resources-information-approved-drugs/intellisite-pathology-solution-pips-philips-medical-systems .
- 17.CAP Today. FDA Clears Aperio AT2 DX for Clinical Diagnosis Today. 2019. [[Last accessed on 2020 Apr 14]]. Available from: https://wwwcaptodayonlinecom/fda-clears-aperio-at2-dx-for-clinical-diagnosis/#:~:text=June 11%2C 2019–Leica Biosystems, diagnosis in the United States .
- 18.Mukhopadhyay S, Feldman MD, Abels E, Ashfaq R, Beltaifa S, Cacciabeve NG, et al. Whole Slide Imaging Versus Microscopy for Primary Diagnosis in Surgical Pathology: A Multicenter Blinded Randomized Noninferiority Study of 1992 Cases (Pivotal Study) Am J Surg Pathol. 2018;42:39–52. doi: 10.1097/PAS.0000000000000948. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 19.Borowsky AD, Glassy EF, Wallace WD, Kallichanda NS, Behling CA, Miller DV, et al. Digital whole slide imaging compared with light microscopy for primary diagnosis in surgical pathology: A multicenter, double-blinded, randomized study of 2045 cases? Arch Pathol Lab Med. 2020 Feb 14; doi: 10.5858/arpa.2019-0569-OA. doi: 105858/arpa2019-0569-OA Epub ahead of print PMID: 32057275. [DOI] [PubMed] [Google Scholar]
- 20.Evans AJ, Bauer TW, Bui MM, Cornish TC, Duncan H, Glassy EF, et al. US Food and Drug Administration approval of whole slide imaging for primary diagnosis: A key milestone is reached and new questions are raised. Arch Pathol Lab Med. 2018;142:1383–7. doi: 10.5858/arpa.2017-0496-CP. [DOI] [PubMed] [Google Scholar]
- 21.Hanna M, Pantanowitz L. Redefining the high power field when counting mitoses using digital pathology. Mod Pathol. 2017;30:394–403. [Google Scholar]
- 22.Sellaro TL, Filkins R, Hoffman C, Fine JL, Ho J, Parwani AV, et al. Relationship between magnification and resolution in digital pathology systems. J Pathol Inform. 2013;4:21. doi: 10.4103/2153-3539.116866. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 23.Stathonikos N, Nguyen TQ, Spoto CP, Verdaasdonk MA, van Diest PJ. Being fully digital: Perspective of a Dutch academic pathology laboratory. Histopathology. 2019;75:621–35. doi: 10.1111/his.13953. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 24.Retamero JA, Aneiros-Fernandez J, Del Moral RG. Complete digital pathology for routine histopathology diagnosis in a multicenter hospital network. Arch Pathol Lab Med. 2020;144:221–8. doi: 10.5858/arpa.2018-0541-OA. [DOI] [PubMed] [Google Scholar]
- 25.Fraggetta F, Garozzo S, Zannoni GF, Pantanowitz L, Rossi ED. Routine digital pathology workflow: The Catania experience. J Pathol Inform. 2017;8:51. doi: 10.4103/jpi.jpi_58_17. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 26.Cheng CL, Tan PH. Digital pathology in the diagnostic setting: Beyond technology into best practice and service management. J Clin Pathol. 2017;70:454–7. doi: 10.1136/jclinpath-2016-204272. [DOI] [PubMed] [Google Scholar]
- 27.Goacher E, Randell R, Williams B, Treanor D. The diagnostic concordance of whole slide imaging and light microscopy: A systematic review. Arch Pathol Lab Med. 2017;141:151–61. doi: 10.5858/arpa.2016-0025-RA. [DOI] [PubMed] [Google Scholar]
- 28.Hanna MG, Reuter VE, Hameed MR, Tan LK, Chiang S, Sigel C, et al. Whole slide imaging equivalency and efficiency study: Experience at a large academic center. Mod Pathol. 2019;32:916–28. doi: 10.1038/s41379-019-0205-0. [DOI] [PubMed] [Google Scholar]
- 29.Tabata K, Mori I, Sasaki T, Itoh T, Shiraishi T, Yoshimi N, et al. Whole-slide imaging at primary pathological diagnosis: Validation of whole-slide imaging-based primary pathological diagnosis at twelve Japanese academic institutes. Pathol Int. 2017;67:547–54. doi: 10.1111/pin.12590. [DOI] [PubMed] [Google Scholar]
- 30.Krishnamurthy S, Mathews K, McClure S, Murray M, Gilcrease M, Albarracin C, et al. Multi-institutional comparison of whole slide digital imaging and optical microscopy for interpretation of hematoxylin-eosin-stained breast tissue sections. Arch Pathol Lab Med. 2013;137:1733–9. doi: 10.5858/arpa.2012-0437-OA. [DOI] [PubMed] [Google Scholar]
- 31.CAP. Protocol for the Examination of Specimens From Patients With Invasive Carcinoma of the Breast. Northfield, Illinois, USA: CAP; 2018. [[Last accessed on 2020 Apr 14]]. Available from: http://wwwcaporg/apps/docs/committees/cancer/cancer_protocols/2012/BreastInvasive_12protocol_3100pdf . [Google Scholar]
- 32.CAP. Validating Whole Slide Imaging (WSI) for Diagnostic Purposes in Pathology Guideline CAP. 2015. [[Last accessed on 2020 Apr 14]]. Available from: https://documentscaporg/documents/wsi-summary-recommendations-2013pdf .
- 33.Tabata K, Uraoka N, Benhamida J, Hanna MG, Sirintrapun SJ, Gallas BD, et al. Validation of mitotic cell quantification via microscopy and multiple whole-slide scanners. Diagn Pathol. 2019;14:65. doi: 10.1186/s13000-019-0839-8. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 34.Wei BR, Halsey CH, Hoover SB, Puri M, Yang HH, Gallas BD, et al. Agreement in histological assessment of mitotic activity between microscopy and digital whole slide images informs conversion for clinical diagnosis. Acad Pathol. 2019;6:237428951985984. doi: 10.1177/2374289519859841. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 35.Bonert M, Tate AJ. Mitotissc counts in breast cancer should be standardized with a uniform sample area. Biomed Eng Online. 2017;16:28. doi: 10.1186/s12938-016-0301-z. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 36.Hanna M, Xing J, Monaco SE, Hartman D, Pantanowitz L. Evaluation of diagnostic concordance between manual mitotic figure counting on glass slides versus whole slide images. J Pathol Inform. 2017;8:26. [Google Scholar]