To the Editor:
Tissue Doppler imaging (TDI) is a robust ultrasound technique used in cardiology to quantify myocardial motion velocity, but the validity and clinical applications of diaphragm TDI are as yet uncertain. We read with great interest the work of Soilemezi and colleagues (1) on TDI to describe diaphragm motion properties of critically ill patients during ventilator weaning. The authors mention that diaphragm pulsed-wave (PW)-TDI is a straightforward method with a fast learning curve. Although we agree that these are essential features for any ultrasound method, we would like to address some important challenges that should be taken into account before implementing this technique in clinical practice.
First, PW-TDI results are very dependent on ultrasound settings and probe position. Doppler signals within a region of interest (“sample volume,” or “gate”) are converted into an average velocity signal. However, increasing the gain broadens the velocity spectrum and results in higher peak velocities (2, 3). Furthermore, as TDI measures the motion vector that is parallel to the ultrasound beam, the insonation angle must be kept as low as possible and changes in angle will affect both pulse length and spectral width. Angle correction can be applied but only for angles <60° because this correction is nonlinear. In addition, reverberation artifacts may occur, resulting in a clutter band along the baseline. This is presented in Figure E1B in the online supplement of Reference 1, where the concept of “smoothing” is introduced. Although smoothing improves visualization of the contour of the velocity signal, peaks are directly affected by the gain and too much smoothing results in signal loss. The authors mentioned that different filters and gains were set to obtain the best velocity images according to the speed of the diaphragmatic motion and the subject under examination. It would be great if they could address how this method was standardized (i.e., quantify “best”), as PW-TDI results can be manipulated easily by adjusting settings on the ultrasound machine.
Furthermore, despite high reliability of results reported in healthy volunteers, it is not entirely clear if PW-TDI results represent what they are intended to. As the velocity–time integral (VTI) reflects diaphragm displacement, VTI should match M-mode displacement (less sensitive to measurement errors compared with PW-TDI), but a large discrepancy was reported in patients (mean 1.27 vs. 0.78 cm for M-mode displacement vs. VTI, respectively; see Table 2 of Reference 1). Also, inspiratory and expiratory VTI should be similar over a large number of breaths, as end-expiratory diaphragm position should not change. Based on our own experience with PW-TDI, and in line with the presented examples (1), VTI inspiration is often larger than VTI expiration. Therefore, reporting a direct comparison of M-mode displacement versus VTI, and VTI inspiration versus VTI expiration, would be valuable to address the validity of PW-TDI results. If there is a systematic underestimation of VTI (and thus also of velocity results), it should be explored whether differences can be minimized sufficiently by adapting ultrasound settings or by performing offline correction.
Noninvasive measures to quantify diaphragm mechanics are highly needed, and correlations between transdiaphragmatic pressure (Pdi) and PW-TDI were evaluated. No relationship between VTI and diaphragm pressure–time product (PTPdi) was found. The authors assumed that VTI could possibly represent diaphragmatic work, because it is defined as the area under the inspiratory PW-TDI curve, similarly to PTPdi being the area under the Pdi–time waveform. This lack of relationship is not surprising given the poor correlation between diaphragm displacement and breathing effort (4, 5). A correlation was found between TDI-maximal relaxation rate (MRR) and Pdi-MRR. TDI-MRR was defined as the slope of the steepest part of the PW-TDI signal during expiration. From a physiological perspective, however, TDI-MRR represents diaphragm deceleration as the signal is already a derivative of displacement. Hence, peak relaxation velocity better reflects MRR and it would be of clinical interest to evaluate its relationship with Pdi-MRR.
Undoubtedly, diaphragm TDI is an exciting approach. However, standardization of the method and further understanding of the capabilities and limitations is important before using this technique for clinical decision-making.
Supplementary Material
Footnotes
Originally Published in Press as DOI: 10.1164/rccm.202007-2958LE on September 22, 2020
Author disclosures are available with the text of this letter at www.atsjournals.org.
References
- 1.Soilemezi E, Savvidou S, Sotiriou P, Smyrniotis D, Tsagourias M, Matamis D. Tissue Doppler imaging of the diaphragm in healthy subjects and critically ill patients. Am J Respir Crit Care Med. 2020;202:1005–1012. doi: 10.1164/rccm.201912-2341OC. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2.Manouras A, Shahgaldi K, Winter R, Nowak J, Brodin L-A. Comparison between colour-coded and spectral tissue Doppler measurements of systolic and diastolic myocardial velocities: effect of temporal filtering and offline gain setting. Eur J Echocardiogr. 2009;10:406–413. doi: 10.1093/ejechocard/jen298. [DOI] [PubMed] [Google Scholar]
- 3.Nestaas E, Schubert U, de Boode WP, El-Khuffash A European Special Interest Group ‘Neonatologist Performed Echocardiography’ (NPE) Tissue Doppler velocity imaging and event timings in neonates: a guide to image acquisition, measurement, interpretation, and reference values. Pediatr Res. 2018;84:18–29. doi: 10.1038/s41390-018-0079-8. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4.Soilemezi E, Tsagourias M, Talias MA, Soteriades ES, Makrakis V, Zakynthinos E, et al. Sonographic assessment of changes in diaphragmatic kinetics induced by inspiratory resistive loading. Respirology. 2013;18:468–473. doi: 10.1111/resp.12011. [DOI] [PubMed] [Google Scholar]
- 5.Spiesshoefer J, Herkenrath S, Henke C, Langenbruch L, Schneppe M, Randerath W, et al. Evaluation of respiratory muscle strength and diaphragm ultrasound: normative values, theoretical considerations, and practical recommendations. Respiration. 2020;99:369–381. doi: 10.1159/000506016. [DOI] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.