Abstract
Background
Late gadolinium enhancement (LGE) cardiovascular magnetic resonance (CMR) represents the gold standard for assessment of myocardial viability. The purpose of this study was to investigate the clinical potential of Compressed SENSE (factor 5) accelerated free-breathing three-dimensional (3D) whole heart LGE with high isotropic spatial resolution (1.4 mm3 acquired voxel size) compared to standard breath-hold LGE imaging.
Methods
This was a retrospective, single-center study of 70 consecutive patients (45.8 ± 18.1 years, 27 females; February–November 2019), who were referred for assessment of left ventricular myocardial viability and received free-breathing and breath-hold LGE sequences at 1.5 T in clinical routine. Two radiologists independently evaluated global and segmental LGE in terms of localization and transmural extent. Readers scored scans regarding image quality (IQ), artifacts, and diagnostic confidence (DC) using 5-point scales (1 non-diagnostic—5 excellent/none). Effects of heart rate and body mass index (BMI) on IQ, artifacts, and DC were evaluated with ordinal logistic regression analysis.
Results
Global LGE (n = 33) was identical for both techniques. Using free-breathing LGE (average scan time: 04:33 ± 01:17 min), readers detected more hyperenhanced lesions (28.2% vs. 23.5%, P < .05) compared to breath-hold LGE (05:15 ± 01:23 min, P = .0104), pronounced at subepicardial localization and for 1–50% of transmural extent. For free-breathing LGE, readers graded scans with good/excellent IQ in 80.0%, with low-impact/no artifacts in 78.6%, and with good/high DC in 82.1% of cases. Elevated BMI was associated with increased artifacts (P = .0012) and decreased IQ (P = .0237). Increased heart rate negatively influenced artifacts (P = .0013) and DC (P = .0479) whereas IQ (P = .3025) was unimpaired.
Conclusions
In a clinical setting, free-breathing Compressed SENSE accelerated 3D high isotropic spatial resolution whole heart LGE provides good to excellent image quality in 80% of scans independent of heart rate while enabling improved depiction of small and particularly non-ischemic hyperenhanced lesions in a shorter scan time than standard breath-hold LGE.
Keywords: Late gadolinium enhancement, Cardiovascular magnetic resonance, Cardiomyopathy
Background
Cardiovascular magnetic resonance (CMR) late gadolinium enhancement (LGE) imaging has become the non-invasive standard of reference for assessment of myocardial viability in different cardiac diseases [1–3]. In ischemic cardiomyopathy, accurate assessment of the extent of infarct transmurality by LGE as a predictive indicator for prognosis and recovery of contractile function guides revascularisation therapy [4–7]. For non-ischemic cardiomyopathies, LGE enables differentiation among diseases based on different patterns of hyperenhancement [1, 8–10].
Traditionally, a two-dimensional (2D) inversion-recovery fast spoiled gradient-echo sequence is applied for LGE imaging with manually chosen inversion time (TI) to null the signal of healthy myocardium [1, 11]. However, 2D acquisition requires multiple breath-holds in pre-defined orientations with associated disadvantages such as non-isotropic resolution with thick slices, incomplete coverage due to slice gaps, low signal-to-noise ratio (SNR), and slice misregistration [11, 12].
The recent development of three-dimensional (3D) imaging allows data acquisition of the entire heart in a single scan without interslice gaps [3, 13]. In this context, accelerated and extended breath-hold approaches have been proposed [14, 15]. However, these techniques are associated with similar constraints on SNR and spatial resolution as 2D LGE since they can only provide highly anisotropic readouts [14, 15]. Further, they are hampered by long breath-hold duration, hence widely unsuitable for clinical routine [15, 16]. When applying different techniques for navigator-gating such as respiratory bellows signal [17], respiratory self-navigation [18–20], and diaphragmatic pencil-beam navigation [3] to reduce respiratory artifacts, 3D LGE can be performed during free-breathing enabling 3D assessment of the whole heart with higher spatial resolution [21–24]. Nevertheless, free-breathing 3D LGE techniques suffer from long imaging time, especially when aiming to provide readouts with submillimetre high isotropic spatial resolution [24–26]. Consequently, drawbacks like susceptibility to varying heart rate and respiratory motion as well as changing accumulation of contrast agent in injured myocardium with inadequate nulling of healthy tissue question their feasibility and usefulness in clinical routine [27, 28].
Recently, Compressed SENSE was introduced, which combines compressed sensing and parallel imaging using SENSitivity Encoding (SENSE) [29–31]. Compressed SENSE enables image-acquisition acceleration currently not achievable by compressed sensing or parallel imaging alone and has shown promising results in musculoskeletal and cardiovascular imaging [29–36].
The purpose of this study was to investigate the clinical potential of Compressed SENSE accelerated free-breathing 3D whole heart LGE with high isotropic spatial resolution compared to standard breath-hold LGE imaging for assessment of left ventricular (LV) viability.
Methods
Ethics
The institutional review board approved this single-center study (Ethikkommission, Medizinische Fakultät der Universität zu Köln; reference number: 19-1619). Given its retrospective design, written informed consent was waived for the patient cohort.
Patient population
We retrospectively reviewed our internal database at our tertiary care medical center for CMR studies performed between February and November 2019. Patients were included if they received a standardized protocol in clinical routine for assessment of LV myocardial viability with known/suspected ischemic or non-ischemic cardiomyopathy including both, breath-hold and free-breathing LGE sequences. Patients were excluded in case of (I) inadequate nulling of breath-hold LGE or (II) uncertainty in assessment of global LGE in breath-hold sequences as determined by a board certified cardiovascular radiologist with eight years of experience in CMR (J.D.), and (III) a very low navigator efficiency of free-breathing LGE.
The following data were obtained from the medical charts or observed during examination: Patient age, gender, body mass index (BMI) at examination date, indication for viability study, heart rate during LGE imaging, LV ejection fraction (LVEF) as assessed by short-axis (SAx) cine, and scan time for LGE sequences. Final diagnosis of cardiomyopathies was based on multidisciplinary evaluation of clinical data, laboratory examination, biopsy results if available, echocardiographic findings, and CMR examination (breath-hold LGE) by the treating cardiology consultant and a board certified cardiovascular radiologist.
Imaging
All CMR examinations were performed on a commercially available whole body 1.5T CMR system (Philips Ingenia, Philips Healthcare, Best, The Netherlands) equipped with a dedicated 28-channel coil for cardiac imaging, which consists of standard system-provided 12-channel posterior and 16-channel anterior coils. The clinical imaging protocol comprised 2D balanced steady-state free precession (bSSFP) breath-hold cine imaging in standard orientations (4-chamber (4Ch), 2-chamber (2Ch), 3-chamber (3Ch), and SAx) followed by admission of Gadobutrol (Gadovist, Bayer HealthCare Pharmaceuticals, Berlin, Germany; 0.2 mmol/kg), which was automatically injected into an antecubital vein at a flow-rate of 2 ml/s. Ten minutes after application of contrast agent, 2D gradient-echo T1-weighted sequence (Look-Locker) was used to determine TI adjusted to null LV myocardium.
Breath-hold LGE
As clinical standard, electrocardiogram (ECG)-triggered (end-diastolic) 3D Cartesian T1-weighted inversion-recovery fast spoiled gradient-echo multislice breath-hold LGE sequences covering the left ventricle in standard orientations (4Ch, SAx, 2Ch, and 3Ch) were acquired. The following imaging parameters were applied: Slice thickness 10 mm, shot duration 170 ms, repetition time 3.5 ms, echo time 1.7 ms, flip angle 15°, field of view (FOV) 300 × 239 × 50 mm3 (4Ch, SAx)/280 × 204 × 50 mm3 (3Ch, 2Ch), matrix 172 × 119 (4Ch, SAx)/176 × 133 (3Ch, 2Ch), acquired voxel size 1.64 × 1.72 × 10 mm3, and reconstructed voxel size 1.59 × 1.59 × 10 mm3. Spectral fat suppression was used to avoid high-intensity signals from epicardial fat. Scan time was 18 s per breath-hold; to cover the SAx, two stacks were acquired.
Free-breathing LGE
Free-breathing whole heart 3D Cartesian LGE was acquired immediately after breath-hold LGE. Figure 1 shows a schematic pulse sequence diagram of the free-breathing LGE sequence. To compensate for cardiac and respiratory motion, ECG-triggering (end-diastolic) in combination with diaphragmatic pencil-beam navigation was applied. The navigator was placed on the dome of the right hemidiaphragm and a 6 mm gating window during end-expiration was employed. TI as estimated for breath-hold LGE sequences was increased by 50 ms. Similar to breath-hold LGE, spectral fat suppression was employed. bSSFP was used for data acquisition as it allows for a very short repetition time, which is beneficial given the time constraints in high-resolution isotropic LGE imaging and since less signal is lost due to signal spoiling compared to gradient-echo imaging. Imaging parameters were as follows: Repetition time 4.4 ms, echo time 2.2 ms, flip angle 45°, FOV 298 × 265 × 120 mm3, matrix 188 × 212 × 171, acquired voxel size 1.40 × 1.40 × 1.40 mm3, and reconstructed voxel size 0.70 × 0.69 × 0.69 mm3. In every cardiac cycle, 35 k-space lines were acquired resulting in a shot length of 150 ms in end-diastolic cardiac phase.
For acceleration of image acquisition, Compressed SENSE (Philips Healthcare ) was used. Data was acquired using a balanced variable density incoherent sampling patter within high-density in the k-space center and continuously decreasing sampling density towards the k-space periphery. For image reconstruction, an iterative L1 norm minimization ensuring data consistency and sparsity in the wavelet domain in combination with regularization by coil sensitivity distribution and SENSE parallel imaging was used. Image reconstruction, which was performed with standard hardware provided by the manufacturer of the CMR system, took about 30 s. An acceleration factor of 5 was employed, resulting in a nominal scan time of 02:08 min or 04:16 min when considering 50% of navigator efficiency, and a heart rate of 60 beats per minute. Data were acquired in the transaxial plane.
Image analysis
For multiplanar reformatting of free-breathing LGE source images, a radiologist with four years of experience in CMR (M.S.) used the Multiplanar-Reconstruction-(MPR) tool in a commercially available image viewer (IMPAX EE, release 20; Agfa HealthCare N.V., Mortsel, Belgium) to create reconstructions in standard orientations (4Ch, SAx, 2Ch, and 3Ch) with identical slice thickness as breath-hold LGE scans. Scans were than evaluated by two radiologists with four (A.W., R1) and three (L.P., R2) years of experience in CMR using the same IMPAX EE (release 20, Agfa HealthCare N.V) workstation during separate reading sessions. Readers were blinded to clinical and patient data. The reading took place in two sessions: In the first session, the readers evaluated breath-hold LGE scans whereas in the second session, they assessed reconstructed free-breathing LGE images and its source images with the latter being free to visualize in all three directions of space using the MPR-tool. A period of four weeks between sessions was used to minimize recall bias. Readers were free to adjust window leveling.
Assessment of LGE
Readers assessed scans for global and segmental LGE of the LV applying the American Heart Association’s 17-segment model [37]. Hyperenhanced lesions were scored for each segment based on the area of scar per segment (0 = no hyperenhancement, 1 = 1–25%, 2 = 26–50%, 3 = 51–75%, and 4 = 76–100% hyperenhancement) and on the location of scar within the segment (1 = subendocardial, 2 = mid-myocardial, 3 = subepicardial, and 4 = transmural). Further, readers analyzed scans for potential pericardial hyperenhancement (0 = no hyperenhancement, 1 = left ventricular, 2 = right ventricular, and 3 circumferential) and intracardiac thrombi.
Assessment of image quality, artifacts, and diagnostic confidence
Using 5-point scales, readers scored LGE sequences in terms of image quality (1: non-diagnostic, image quality inadequate for diagnosis; 2: poor, suboptimal image quality for diagnosis; 3: fair, mediocre image quality acceptable for diagnosis; 4: good, image quality suitable for confident diagnosis; and 5: excellent, image quality providing highly confident diagnosis) and artifacts considering blurring artifacts, banding artifacts, respiratory artifacts, cardiac motion artifacts, and parallel imaging reconstruction artifacts (1: non-diagnostic; 2: high impact, 3: moderate impact, 4: low impact, and 5: none) as well as diagnostic confidence (1: non-diagnostic, 2: low, 3: fair, 4: good, and 5: high).
Statistical analysis
Statistical analysis was performed using JMP (release 14.1.0, SAS Institute, Cary, North Carolina, USA) with statistical significance being set to P < 0.05. Data are expressed as mean ± standard deviation, unless noted otherwise. Comparison of paired nonparametric variables and scan times was performed with Wilcoxon signed-rank tests. To compare the proportion of detected hyperenhanced lesions between LGE techniques, the McNemar test was applied. Cohen’s Kappa was used to evaluate interobserver agreement of detection of hyperenhanced lesions using free-breathing and breath-hold LGE sequences. The interpretation of an agreement was as follows: 0.01–0.2 slight, 0.21–0.4 fair, 0.41–0.6 moderate, 0.61–0.8 substantial, and 0.81–0.99 almost perfect agreement [38]. Influence of confounders (heart rate and BMI) on subjective evaluation of image quality, artifacts, and diagnostic confidence of LGE sequences was assessed using ordinal logistic regression analysis.
Results
Study population and baseline characteristics
Seventy-nine patients met the inclusion criteria. Thereof, four patients were excluded given inadequate nulling of breath-hold LGE, three due to uncertainty in assessment of global LGE in breath-hold sequences, and two given very low navigator efficiency of free-breathing LGE.
Consequently, 70 patients (27 females) were included in this study, yielding an age of 45.8 ± 18.1 years (range 16–80 years) and BMI of 25.1 ± 4.2 (17.6–37.4). The heart rate was 70 ± 12 (48–109) beats per minute and a LVEF of 60 ± 17% (19%–85%) was observed. Scans were acquired for viability assessment of known/suspected ischemic cardiomyopathy in 19 (27.1%) and non-ischemic cardiomyopathies in 51 (72.9%) patients. Thereof, 18 (35.3%) patients were examined for evaluation of potential (peri-)myocarditis, nine (17.6%) of structural heart disease, seven (13.7%) of sarcoidosis, six (11.8%) of dilated cardiomyopathy (DCM), five (9.8%) of hypertrophic cardiomyopathy (HCM), three (5.8%) of amyloidosis, two (3.9%) of chemotherapy-induced cardiomyopathy, and one patient (2.0%) of Anderson-Fabry disease.
Imaging
All included imaging studies were executed successfully. Depending on patient’s heart rate and breathing frequency, free-breathing LGE took 04:33 ± 01:17 min to acquire (46.9% navigator efficiency). Breath-hold LGE sequences lasted 05:15 ± 01:23 min (P = .0104).
Image analysis
Assessment of LGE
Global LV myocardial LGE (n = 33) was identical for both techniques and readers, present in 13 patients (39.4%) with ischemic and 20 patients (60.6%) with non-ischemic cardiomyopathy (P = 1.0). Thereof, findings consistent with (peri-) myocarditis were present in 11 (55.0%) patients, sarcoidosis in three (15.0%), DCM in two (10.0%), HCM in two (10.0%), amyloidosis in one (5.0%), and chemotherapy-induced cardiomyopathy in one (5.0%) patient.
Table 1 provides an overview of indications for viability studies and final LGE findings.
Table 1.
Indication for viability study | n = 70 | Final LGE findings | n = 33 |
---|---|---|---|
Ischemic cardiomyopathy | 19 (27.1%) | Ischemic cardiomyopathy | 13 (39.4%) |
Non-ischemic cardiomyopathies | 51 (72.9%) | Non-ischemic cardiomyopathies | 20 (60.6%) |
(Peri-) Myocarditis | 18 (35.3%) | (Peri-) Myocarditis | 11 (55.0%) |
Structural heart disease | 9 (17.6%) | Sarcoidosis | 3 (15.0%) |
Sarcoidosis | 7 (13.7%) | Dilated cardiomyopathy | 2 (10.0%) |
Dilated cardiomyopathy | 6 (11.8%) | Hypertrophic cardiomyopathy | 2 (10.0%) |
Hypertrophic cardiomyopathy | 5 (9.8%) | Amyloidosis | 1 (5.0%) |
Amyloidosis | 3 (5.8%) | Chemotherapy-induced cardiomyopathy | 1 (5.0%) |
Chemotherapy-induced cardiomyopathy | 2 (3.9%) | ||
Anderson-Fabry disease | 1 (2.0%) |
Individual and averaged segmental LGE assessment of readers is given in Table 2. Readers detected 132 hyperenhanced lesions in breath-hold (23.5%; R1: 126 (22.5%), R2: 138 (24.6%)) sequences whereas 158 lesions were noted using free-breathing LGE (28.2%; R1: 153 (27.3%), R2: 163 (29.1%); P < 0.0001).
Table 2.
READER 1 | ||||
Overall detected hyperenhanced lesions in patients showing global LGE (n = 33) | ||||
Breath-hold | n = 126/561 (22.5%) detected hyperenhanced lesions | P < .0001 | ||
Free-breathing | n = 153/561 (27.3%) detected hyperenhanced lesions | |||
LGE localization | Subendocardial | Mid-myocardial | Subepicardial | Transmural |
Breath-hold | 24/126 (19.1%) | 25/126 (19.8%) | 21/126 (16.7%) | 56/126 (44.4%) |
Free-breathing | 31/153 (20.3%) | 29/153 (19.0%) | 32/153 (20.9%) | 61/153 (39.9%) |
LGE transmural extent | 1–25% | 26–50% | 51–75% | 76–100% |
Breath-hold | 16/126 (12.7%) | 25/126 (19.8%) | 25/126 (19.8%) | 60/126 (47.6%) |
Free-breathing | 29/153 (18.9%) | 32/153 (20.9%) | 27/153 (17.6%) | 65/153 (42.5%) |
READER 2 | ||||
Overall detected hyperenhanced lesions in patients showing global LGE (n = 33) | ||||
Breath-hold | n = 138/561 (24.6%) detected hyperenhanced lesions | P < .0001 | ||
Free-breathing | n = 163/561 (29.1%) detected hyperenhanced lesions | |||
LGE localization | Subendocardial | Mid-myocardial | Subepicardial | Transmural |
Breath-hold | 33/138 (23.9%) | 24/138 (17.4%) | 27/138 (19.6%) | 54/138 (39.1%) |
Free-breathing | 39/163 (23.9%) | 31/163 (19.0%) | 38/163 (23.3%) | 55/163 (33.7%) |
LGE transmural extent | 1–25% | 26–50% | 51–75% | 76–100% |
Breath-hold | 23/138 (16.7%) | 28/138 (20.3%) | 31/138 (22.5%) | 56/138 (40.6%) |
Free-breathing | 35/163 (21.5%) | 38/163 (23.3%) | 31/163 (19.0%) | 59/163 (36.2%) |
AVERAGED for both readers | ||||
Overall detected hyperenhanced lesions in patients showing global LGE (n = 33) | ||||
Breath-hold | n = 132/561 (23.5%) detected hyperenhanced lesions | P < .0001 | ||
Free-breathing | n = 158/561 (28.2%) detected hyperenhanced lesions | |||
LGE localization | Subendocardial | Mid-myocardial | Subepicardial | Transmural |
Breath-hold | 28.5/132 (21.6%) | 24.5/132 (18.6%) | 24/132 (18.2%) | 55/132 (41.7%) |
Free-breathing | 35/158 (22.2%) | 30/158 (19.0%) | 35/158 (22.2%) | 58/158 (36.7%) |
LGE transmural extent | 1–25% | 26–50% | 51–75% | 76–100% |
Breath-hold | 19.5/132 (14.8%) | 26.5/132 (20.1%) | 28/132 (21.2%) | 58/132 (43.9%) |
Free-breathing | 32/158 (20.2%) | 35/158 (22.2%) | 29/158 (18.4%) | 62/158 (39.2%) |
Averaged for both readers, depiction of transmural scar was comparable between both techniques (free-breathing: 58, breath-hold: 55) whereas readers identified more lesions in mid-myocardial (free-breathing: 30, breath-hold: 24.5), subendocardial (free-breathing: 35, breath-hold: 28.5), and subepicardial localization (free-breathing: 35, breath-hold: 24) in free-breathing LGE. Further, more small lesions were depicted using free-breathing LGE than with breath-hold LGE (1–25%: 32 vs. 19.5, 26–50%: 35 vs. 26.5) while identification of scars of 51–75% (free-breathing: 29, breath-hold: 28) and of 76–100% (free-breathing: 62, breath-hold: 58) of transmural extent was comparable between techniques. Figure 2 depicts distribution of lesions in terms of localization and transmural extent for free-breathing and breath-hold LGE.
For both techniques, an almost perfect interobserver agreement for detection of hyperenhanced lesions was observed [free-breathing LGE: Cohen’s Kappa of 0.94 (0.91–0.97), breath-hold LGE: Cohen’s Kappa of 0.93 (0.89–0.96)].
Concomitant pericardial LGE was identical for both techniques and readers (n = 9) with six patients presenting circular (66.7%), two LV (22.2%), and one patient right ventricular (11.1%) hyperenhancement (P = 1). Both readers detected intracardiac thrombi in three patients in both techniques (P = 1).
Assessment of image quality, artifacts, and diagnostic confidence
For both readers, free-breathing LGE achieved a score of 4.1 ± 0.8 for image quality, of 4.0 ± 0.9 for artifacts, and of 4.4 ± 0.8 for diagnostic confidence. Breath-hold LGE yielded scores of 4.4 ± 0.8 for image quality (P = .0028), of 4.3 ± 0.9 for artifacts (P = .0098), and of 4.5 ± 0.8 for diagnostic confidence (P = .3125).
Considering the overall distribution of scores of both readers combined, Table 3 gives detailed results. Breath-hold LGE provided good/excellent image quality in 87.1%, low-impact/no artifacts in 81.4%, and good/high diagnostic confidence in 90.7% of examinations. For free-breathing LGE, readers graded scans with good/excellent image quality in 80.0%, with low-impact/no artifacts in 78.6%, and with good/high diagnostic confidence in 82.1% of cases.
Table 3.
Criterion | Technique | Non-diagnostic | Poor | Fair | Good | Excellent | Good + excellent |
---|---|---|---|---|---|---|---|
Image | Breath-hold | – | 7 (5.0%) | 11 (7.9%) | 45 (32.1%) | 77 (55.0%) | 122 (87.1%) |
quality | Free-breathing | – | 7 (5.0%) | 21 (15.0%) | 61 (43.6%) | 51 (36.4%) | 112 (80.0%) |
Criterion | Technique | Non-diagnostic | High impact | Moderate impact | Low impact | None | Low impact + none |
---|---|---|---|---|---|---|---|
Artifacts | Breath-hold | – | 5 (3.6%) | 21 (15.0%) | 42 (30.0%) | 72 (51.4%) | 114 (81.4%) |
Free-breathing | – | 11 (7.9%) | 19 (13.6%) | 63 (45.0%) | 47 (33.6%) | 110 (78.6%) |
Criterion | Technique | Non-diagnostic | Low | Fair | Good | High | Good + high |
---|---|---|---|---|---|---|---|
Diagnostic | Breath-hold | – | 6 (4.3%) | 7 (5.0%) | 40 (28.6%) | 87 (62.1%) | 127 (90.7%) |
confidence | Free-breathing | – | 4 (2.9%) | 21 (15.0%) | 33 (23.6%) | 82 (58.6%) | 115 (82.1%) |
Table 4 displays influence of confounders (heart rate and body mass index) on subjective assessment of breath-hold and free-breathing LGE assessed by ordinal logistic regression analysis. Regarding breath-hold LGE, increased heart rate negatively influenced image quality (P = .0362) and artifacts (P = .0453). BMI did not yield significant influence on subjective assessment of breath-hold sequences. For free-breathing LGE, elevated BMI was associated with increased artifacts (P = .0012) and reduced image quality (P = .0237). Increased heart rate negatively influenced artifacts (P = .0013) and diagnostic confidence (P = .0479) whereas image quality (P = .3025) was unimpaired.
Table 4.
Confounder | Criterion | Breath-hold | Free-breathing |
---|---|---|---|
Heart rate | Image quality | .0362 | .3025 |
Artifacts | .0453 | .0013 | |
Diagnostic confidence | .1722 | .0479 | |
Body mass index | Image quality | .2520 | .0237 |
Artifacts | .8481 | .0012 | |
Diagnostic confidence | .1822 | .3523 |
Exemplary comparisons of free-breathing and breath-hold LGE are given in Figs. 3 and 4.
Discussion
In this study, we investigated the clinical potential of Compressed SENSE accelerated free-breathing 3D whole heart LGE with high isotropic spatial resolution compared to standard breath-hold LGE for assessment of LV viability. The major findings of the study are the following: 1. Free-breathing LGE enabled improved depiction of small hyperenhanced lesions of the LV myocardium, in particular of non-ischemic localization, and required a shorter scan time compared to breath-hold LGE. 2. Free-breathing LGE provided good to excellent results for diagnostic confidence, artifacts, and image quality in about 80% of examinations with the latter being independent of heart rate.
Given aforementioned limitations of breath-hold LGE, free-breathing LGE techniques are of particular interest in research and clinical practice. Previous studies have proposed different free-breathing approaches enabling high spatial resolution at 1.5 T and 3 T [18, 21–23, 25, 26]. However, long acquisition time remains one of the main limitations for the majority of techniques. In this context, Andreu et al. reported scan times of 16.4 ± 7.2 min (for 1.4 mm3 reconstructed resolution) [25], Bizino et al. of 09:34 ± 03:04 min (0.9 mm3 reconstructed resolution) [26], and Lintingre et al. of 8–10 min (1.25 × 1.25 × 2.5 mm acquired resolution) [24]. Bratis et al. introduced respiratory self-navigation for whole heart LGE in 03:52 min with 2.0 mm3 acquired and 1.0 mm3 reconstructed resolution [18] whereas Kino et al. demonstrated the use of phase-sensitive inversion-recovery (PSIR) for free-breathing LGE with 2.0 mm3 acquired resolution in 06:10 ± 02:56 min [23]. Akcakaya et al. [21] and Basha et al. [22], respectively, proposed another promising technique for free-breathing LGE providing the highest isotropic acquisition to date (varying between 1.0 and 1.7 mm3 depending on patient’s size and heart rate) in approximately 4 to 6 min (depending on applied acceleration factor) using LOw-dimensional-structure Self-learning and Thresholding (LOST). However, required reconstruction of acquired data using an offline central processing unit cluster, lasting about an hour, limits its immediate application in clinical routine and feasibility for other centers [21, 22].
In the present study, proposed Compressed SENSE accelerated free-breathing 3D whole heart LGE sequence enables high isotropic resolution in acquired (fixed 1.4 mm3, no adaptation to patient’s specifics required) and reconstructed (0.7 mm3) voxel size in an acceptable scan time (04:33 ± 01:17 min) without extensive postprocessing. Image reconstruction took about 30 s and was performed with standard hardware provided by the manufacturer of the CMR system.
As previously shown when comparing free-breathing to breath-hold LGE, readers detected more smaller hyperenhanced lesions using free-breathing LGE, mostly because of its contiguous, high-resolution acquisition in through-plane direction and because breath-hold LGE suffers from partial volume effect due to its lower voxel size [22–24, 39]. Although a voxel partially contains spins with high signal intensity, an overall elimination of high signal intensity of the voxel may occur if it is also occupied by spins with low signal intensity [39]. Therefore, small hyperenhanced lesions are more difficult to detect and to distinguish from another as two separate findings in breath-hold LGE. In breath-hold sequences, the partial volume effect may also lead to impaired detection of patchy mid-myocardial and, in particular, subepicardial lesions. In contrast, higher resolution of free-breathing LGE with possible 3D assessment facilitates their detection since it enables sufficient differentiation from adjoining structures.
Visually, the signal appearance of free-breathing LGE was lower compared to breath-hold LGE, which might have different reasons except for the distinct sequence details with free-breathing LGE being acquired at a later time point after administration of contrast agent and data acquisition comprising a comparable large timeframe. However, no quantitative SNR comparison was carried out in the current work because of the iterative and wavelet denoising based reconstruction used for free-breathing LGE, which is dependent on the actual image content and results in different numbers of iterations or denoising steps for different images interfering with absolute SNR estimations. The visually reduced signal of the blood pool improved detection of subendocardial lesions in this study, mostly due to visually moderately increased differentiation between blood and injured myocardium, facilitating depiction of scars at this localization, comparable to dark-blood LGE [40, 41].
Free-breathing LGE enabled sufficient detection of intracardiac thrombi and pericardial hyperenhancement. Given the low number of patients in the present study presenting respective pathologies, these findings may carry a bias and should warrant future investigations, including assessment of atrial and right ventricular LGE.
Free-breathing LGE did not fully yield the same high scores for image quality, artifacts, and diagnostic confidence as breath-hold LGE. These differences may be due to acquiring the free-breathing sequence at the end of examination when potential contrast washout leads to a change in optimal TI and patients tend to be exhausted and may present irregular breathing patterns. However, there was no statistical difference regarding diagnostic confidence between both techniques and free-breathing LGE achieved good/excellent image quality in 80% of cases. An earlier acquisition of free-breathing LGE may lead to an improvement of image quality and reduction of artifacts. Further, application of PSIR might reduce effects of dynamic changes of TI during examination [19, 23].
In free-breathing LGE, elevated BMI led to impaired image quality and more pronounced artifacts, while in breath-hold LGE, BMI had no impact on image quality and artifacts. It is of note, however, that diagnostic confidence of free-breathing was unimpaired by BMI. A possible explanation may be that the loss in image quality is counterbalanced by the improved post-processing options for free-breathing LGE such as multiplanar reformation. Further, increased heart rate led to decreased diagnostic confidence and increased artifacts for free-breathing LGE and affected artifacts as well as image quality of breath-hold LGE. However, increased heart rate did not lead to impaired image quality for free-breathing LGE. These findings differ to the study by Basha et al. investigating free-breathing LGE using LOST, for which increased heart rate was associated with significant decreased image quality [22]. The shorter shot duration of free-breathing LGE compared to breath-hold LGE (150 vs. 170 ms) may explain the findings of the present study.
Besides being acquired without any acceleration technique, the longer scan time of breath-hold LGE in this study is mostly due to patient recovery time between breath-holds and repetition of scans if the patient was unable to perform the breath-hold or if planning was not accurately. In contrast, free-breathing LGE can be performed in patients with poor compliance unable to tolerate repetitive breath-holds, proves to be widely user-independent, and enables post-acquisition reformatting in any arbitrary orientation with identical image quality as original transaxial images using multiplanar tools.
Limitations
The following limitations need to be considered: First, there is no standard of reference for detection of LGE. In this study, breath-hold LGE was used for comparison since this technique is standard in clinical routine for assessment of LV viability. Histopathological staining or repeated acquisition of the same patient at the following day could have been conducted to confirm additional findings of free-breathing LGE but where not possible or desired given the retrospective setting and potential peri-interventional complications. Second, given the distinct appearance of LGE techniques, readers were not blinded to the type of sequence potentially influencing the results. Third, no direct comparison to other free-breathing (an) isotropic LGE techniques with different undersampling approaches was performed in the current study, which could nurture additional investigations. Finally, the possible impact of proposed free-breathing technique on final diagnosis of cardiomyopathies and further clinical decision making should be investigated in a prospective, multi-center setting, in which the order of acquisition of free-breathing and breath-hold LGE sequences is randomized to provide an unbiased evaluation.
Conclusions
Free-breathing Compressed SENSE accelerated 3D whole heart LGE with high isotropic spatial resolution provides good to excellent image quality in 80% of scans independent of heart rate while enabling improved depiction of small hyperenhanced lesions with in particular non-ischemic localization of the LV myocardium in a shorter scan time than standard breath-hold LGE. Given its fast acquisition, wider user-independence, and omission of breath‐holds, Compressed SENSE accelerated free-breathing LGE has the potential to facilitate clinical workflow.
Acknowledgements
Not applicable.
Abbreviations
- 2D
Two-dimensional
- 2Ch
2-Chamber
- 3Ch
3-Chamber
- 3D
Three-dimensional
- 4CH
4-Chamber
- BMI
Body mass index
- bSSFP
Balanced steady-state free precession
- CMR
Cardiovascular magnetic resonance
- DCM
Dilated cardiomyopathy
- ECG
Electrocardiogram
- FOV
Field of view
- HCM
Hypertrophic cardiomyopathy
- LGE
Late gadolinium enhancement
- LOST
LOw-dimensional-structure Self-learning and Thresholding (LOST)
- LV
Left ventricle/left ventricular
- LVEF
Left ventricular ejection fraction
- MPR
Multiplanar-reconstruction
- PSIR
Phase-sensitive inversion-recovery
- SAx
Short-axis
- SENSE
SENSitivity Encoding
- SNR
Signal-to-noise ratio
- TI
Inversion time
Authors’ contributions
Conceptualization: JD, LP, KW, MG, CPN, DM, SN. Methodology: LP, JD, AW, KW, SN, DM, CPN, ACB, SL, KRL. Formal analysis and investigation: LP, AW, MS, SL, JD, SN. Writing—original draft preparation: LP, SL, AW, KW. Writing—review and editing: JD, CPN, DM, TP, SN, KRL, SL, ACB. Resources: JD, MG, CPN, DM, KW, TP, ACB. Supervision: JD, CPN, DM, TP, ACB. All authors read and approved the final manuscript.
Funding
Not applicable.
Availability of data and materials
The datasets generated and/or analyzed during the current study are not publicly available due to data protection but are available from the corresponding author upon reasonable request. The imaging protocol of proposed free-breathing LGE sequence is available from the corresponding author upon reasonable request.
Ethics approval and consent to participate
The local ethics committee approved this study. Due to the retrospective design of the study, written informed consent requirement was waived. Ethics comittee: Ethikkommission, Medizinische Fakultät der Universität zu Köln. Committee’s reference number: 19-1619.
Consent for publication
Regarding the images attached to the manuscript, written informed consent for publication was obtained from the patients.
Competing interests
Lenhard Pennig: Study leave related to this project as part of a research contract between Philips Healthcare and the University Hospital Cologne. Simon Lennartz: Received research support from Philips Healthcare. David Maintz: Speaker’s bureau, Philips Healthcare. Kilian Weiss: Employee, Philips Healthcare.
Footnotes
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
References
- 1.Ordovas KG, Higgins CB. Delayed contrast enhancement on MR images of myocardium: past, present, future. Radiology. 2011;261:358–374. doi: 10.1148/radiol.11091882. [DOI] [PubMed] [Google Scholar]
- 2.West AM, Kramer CM. Cardiovascular magnetic resonance imaging of myocardial infarction, viability, and cardiomyopathies. Curr Probl Cardiol. 2010;35:176–220. doi: 10.1016/j.cpcardiol.2009.12.002. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3.Kellman P, Arai AE. Cardiac imaging techniques for physicians: late enhancement. J Magn Reson Imaging. 2012;36:529–542. doi: 10.1002/jmri.23605. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4.Kim HW, Farzaneh-Far A, Kim RJ. Cardiovascular Magnetic resonance in patients with myocardial infarction. Current and emerging applications. J Am Coll Cardiol. 2009;55:1–16. doi: 10.1016/j.jacc.2009.06.059. [DOI] [PubMed] [Google Scholar]
- 5.Kim RJ, Wu E, Rafael A, Chen EL, Parker MA, Simonetti O, et al. The use of contrast-enhanced magnetic resonance imaging to identify reversible myocardial dysfunction. N Engl J Med. 2000;343:1445–1453. doi: 10.1056/NEJM200011163432003. [DOI] [PubMed] [Google Scholar]
- 6.Selvanayagam JB, Kardos A, Francis JM, Wiesmann F, Petersen SE, Taggart DP, et al. Value of delayed-enhancement cardiovascular magnetic resonance imaging in predicting myocardial viability after surgical revascularization. Circulation. 2004;110:1535–1541. doi: 10.1161/01.CIR.0000142045.22628.74. [DOI] [PubMed] [Google Scholar]
- 7.Yan AT, Shayne AJ, Brown KA, Gupta SN, Chan CW, Luu TM, et al. Characterization of the peri-infarct zone by contrast-enhanced cardiac magnetic resonance imaging is a powerful predictor of post-myocardial infarction mortality. Circulation. 2006;114:32–39. doi: 10.1161/CIRCULATIONAHA.106.613414. [DOI] [PubMed] [Google Scholar]
- 8.Mahrholdt H, Wagner A, Judd RM, Sechtem U, Kim RJ. Delayed enhancement cardiovascular magnetic resonance assessment of non-ischaemic cardiomyopathies. Eur Heart J. 2005;26:1461–1474. doi: 10.1093/eurheartj/ehi258. [DOI] [PubMed] [Google Scholar]
- 9.Wu KC, Weiss RG, Thiemann DR, Kitagawa K, Schmidt A, Dalal D, et al. Late gadolinium enhancement by cardiovascular magnetic resonance heralds an adverse prognosis in nonischemic cardiomyopathy. J Am Coll Cardiol. 2008;51:2414–2421. doi: 10.1016/j.jacc.2008.03.018. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10.Shehata ML, Turkbey EB, Vogel-Claussen J, Bluemke DA. Role of cardiac magnetic resonance imaging in assessment of nonischemic cardiomyopathies. Top Magn Reson Imaging. 2008;19:43–57. doi: 10.1097/RMR.0b013e31816fcb22. [DOI] [PubMed] [Google Scholar]
- 11.Kim RJ, Shah DJ, Judd RM. How we perform delayed enhancement imaging. J Cardiovasc Magn Reson. 2003;5:505–514. doi: 10.1081/JCMR-120022267. [DOI] [PubMed] [Google Scholar]
- 12.Peukert D, Laule M, Taupitz M, Kaufels N, Hamm B, Dewey M. 3D and 2D Delayed-enhancement magnetic resonance imaging for detection of myocardial infarction: preclinical and clinical results. Acad Radiol. 2007;14:788–794. doi: 10.1016/j.acra.2007.03.006. [DOI] [PubMed] [Google Scholar]
- 13.Kido T, Kido T, Nakamura M, Kawaguchi N, Nishiyama Y, Ogimoto A, et al. Three-dimensional phase-sensitive inversion recovery sequencing in the evaluation of left ventricular myocardial scars in ischemic and non-ischemic cardiomyopathy: comparison to three-dimensional inversion recovery sequencing. Eur J Radiol. 2014;83:2159–2166. doi: 10.1016/j.ejrad.2014.09.014. [DOI] [PubMed] [Google Scholar]
- 14.Roujol S, Basha TA, Akçakaya M, Foppa M, Chan RH, Kissinger KV, et al. 3D late gadolinium enhancement in a single prolonged breath-hold using supplemental oxygenation and hyperventilation. Magn Reson Med. 2014;72:850–857. doi: 10.1002/mrm.24969. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 15.Goetti R, Kozerke S, Donati OF, Sürder D, Stolzmann P, Kaufmann PA, et al. Acute, subacute, and chronic myocardial infarction: quantitative comparison of 2D and 3D late gadolinium enhancement MR imaging. Radiology. 2011;259:704–711. doi: 10.1148/radiol.11102216. [DOI] [PubMed] [Google Scholar]
- 16.Piehler KM, Wong TC, Puntil KS, Zareba KM, Lin K, Harris DM, et al. Free-breathing, motion-corrected late gadolinium enhancement is robust and extends risk stratification to vulnerable patients. Circ Cardiovasc Imaging. 2013;6:423–432. doi: 10.1161/CIRCIMAGING.112.000022. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 17.Peters DC, Shaw JL, Knowles BR, Moghari MH, Manning WJ. Respiratory bellows-gated late gadolinium enhancement of the left atrium. J Magn Reson Imaging. 2013;38:1210–1214. doi: 10.1002/jmri.23954. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 18.Bratis K, Henningsson M, Grigoratos C, Dell’Omodarme M, Chasapides K, Botnar R, et al. Image-navigated 3-dimensional late gadolinium enhancement cardiovascular magnetic resonance imaging: Feasibility and initial clinical results. J Cardiovasc Magn Reson. 2017;19:97. doi: 10.1186/s12968-017-0418-7. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 19.Ginami G, Neji R, Rashid I, Chiribiri A, Ismail TF, Botnar RM, et al. 3D whole-heart phase sensitive inversion recovery CMR for simultaneous black-blood late gadolinium enhancement and bright-blood coronary CMR angiography. J Cardiovasc Magn Reson. 2017;19:94. doi: 10.1186/s12968-017-0405-z. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 20.Munoz C, Bustin A, Neji R, Kunze KP, Forman C, Schmidt M, Hajhosseiny R, Masci PG, Zeilinger M, Wuest W, Botnar RMPC. Motion-corrected 3D whole-heart water-fat high-resolution late gadolinium enhancement cardiovascular magnetic resonance imaging. J Cardiovasc Magn Reson. 2020;22:53. doi: 10.1186/s12968-020-00649-5. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 21.Akçakaya M, Rayatzadeh H, Basha TA, Hong SN, Chan RH, Kissinger KV, et al. Accelerated late gadolinium enhancement cardiac MR imaging with isotropic spatial resolution using compressed sensing: initial experience. Radiology. 2012;264:691–699. doi: 10.1148/radiol.12112489. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 22.Basha TA, Akçakaya M, Liew C, Tsao CW, Delling FN, Addae G, et al. Clinical performance of high-resolution late gadolinium enhancement imaging with compressed sensing. J Magn Reson Imaging. 2017;46:1829–1838. doi: 10.1002/jmri.25695. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 23.Kino A, Zuehlsdorff S, Sheehan JJ, Weale PJ, Carroll TJ, Jerecic R, et al. Three-dimensional phase-sensitive inversion-recovery turbo FLASH sequence for the evaluation of left ventricular myocardial scar. Am J Roentgenol. 2009;193:381–388. doi: 10.2214/AJR.08.1952. [DOI] [PubMed] [Google Scholar]
- 24.Lintingre P-F, Nivet H, Clément-Guinaudeau S, Camaioni C, Sridi S, Corneloup O, et al. High-Resolution late gadolinium enhancement magnetic resonance for the diagnosis of myocardial infarction with nonobstructed coronary arteries. JACC Cardiovasc Imaging. 2020;13:1135–1148. doi: 10.1016/j.jcmg.2019.11.020. [DOI] [PubMed] [Google Scholar]
- 25.Andreu D, Ortiz-Pérez JT, Fernández-Armenta J, Guiu E, Acosta J, Prat-González S, et al. 3D delayed-enhanced magnetic resonance sequences improve conducting channel delineation prior to ventricular tachycardia ablation. Europace. 2015;17:938–945. doi: 10.1093/europace/euu310. [DOI] [PubMed] [Google Scholar]
- 26.Bizino MB, Tao Q, Amersfoort J, Siebelink HMJ, van den Bogaard PJ, van der Geest RJ, et al. High spatial resolution free-breathing 3D late gadolinium enhancement cardiac magnetic resonance imaging in ischaemic and non-ischaemic cardiomyopathy: quantitative assessment of scar mass and image quality. Eur Radiol. 2018;28:4027–4035. doi: 10.1007/s00330-018-5361-y. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 27.Klein C, Nekolla SG, Balbach T, Schnackenburg B, Nagel E, Fleck E, et al. The influence of myocardial blood flow and volume of distribution on late Gd-DTPA kinetics in ischemic heart failure. J Magn Reson Imaging. 2004;20:588–593. doi: 10.1002/jmri.20164. [DOI] [PubMed] [Google Scholar]
- 28.Goldfarb JW, Mathew ST, Reichek N. Quantitative breath-hold monitoring of myocardial gadolinium enhancement using inversion recovery TrueFISP. Magn Reson Med. 2005;53:367–371. doi: 10.1002/mrm.20369. [DOI] [PubMed] [Google Scholar]
- 29.Pruessmann KP, Weiger M, Scheidegger MB, Boesiger P. SENSE: sensitivity encoding for fast MRI. Magn Reson Med. 1999;42:952–962. doi: 10.1002/(SICI)1522-2594(199911)42:5<952::AID-MRM16>3.0.CO;2-S. [DOI] [PubMed] [Google Scholar]
- 30.Liang D, Liu B, Wang J, Ying L. Accelerating SENSE using compressed sensing. Magn Reson Med. 2009;62:1574–1584. doi: 10.1002/mrm.22161. [DOI] [PubMed] [Google Scholar]
- 31.Lustig M, Donoho D, Pauly JM. Sparse MRI: The application of compressed sensing for rapid MR imaging. Magn Reson Med. 2007;58:1182–1195. doi: 10.1002/mrm.21391. [DOI] [PubMed] [Google Scholar]
- 32.Pennig L, Wagner A, Weiss K, Lennartz S, Grunz JP, Maintz D, et al. Imaging of the pulmonary vasculature in congenital heart disease without gadolinium contrast: Intraindividual comparison of a novel Compressed SENSE accelerated 3D modified REACT with 4D contrast-enhanced magnetic resonance angiography. J Cardiovasc Magn Reson. 2020;22:8. doi: 10.1186/s12968-019-0591-y. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 33.Bratke G, Rau R, Weiss K, Kabbasch C, Sircar K, Morelli JN, et al. Accelerated MRI of the lumbar spine using compressed sensing: quality and efficiency. J Magn Reson Imaging. 2019;49:164–175. doi: 10.1002/jmri.26526. [DOI] [PubMed] [Google Scholar]
- 34.Ma Y, Hou Y, Ma Q, Wang X, Sui S, Wang B. Compressed SENSE single-breath-hold and free-breathing cine imaging for accelerated clinical evaluation of the left ventricle. Clin Radiol. 2019;74:325.e9–e325. doi: 10.1016/j.crad.2018.12.012. [DOI] [PubMed] [Google Scholar]
- 35.Pennig L, Wagner A, Weiss K, Lennartz S, Huntgeburth M, Hickethier T, et al. Comparison of a novel Compressed SENSE accelerated 3D modified relaxation-enhanced angiography without contrast and triggering with CE-MRA in imaging of the thoracic aorta. Int J Cardiovasc Imaging. 2020. 10.1007/s10554-020-01979-2. [DOI] [PMC free article] [PubMed]
- 36.Pennig L, Kabbasch C, Hoyer UCI, Lennartz S, Zopfs D, Goertz L, et al. Relaxation-Enhanced Angiography Without Contrast and Triggering (REACT) for Fast Imaging of Extracranial Arteries in Acute Ischemic Stroke at 3 T. Clin Neuroradiol. 2020. 10.1007/s00062-020-00963-6. [DOI] [PMC free article] [PubMed]
- 37.Cerqueira MD, Weissman NJ, Dilsizian V, Jacobs AK, Kaul S, Laskey WK, et al. Standardized myocardial sementation and nomenclature for tomographic imaging of the heart: a statement for healthcare professionals from the cardiac imaging committee of the council on clinical cardiology of the American Heart Association. Circulation. 2002;105:539–542. doi: 10.1161/hc0402.102975. [DOI] [PubMed] [Google Scholar]
- 38.Landis JR, Koch GG. The Measurement of observer agreement for categorical data. Biometrics. 1977;33:159–174. doi: 10.2307/2529310. [DOI] [PubMed] [Google Scholar]
- 39.González Ballester MÁ, Zisserman AP, Brady M. Estimation of the partial volume effect in MRI. Med Image Anal. 2002;6:389–405. doi: 10.1016/S1361-8415(02)00061-0. [DOI] [PubMed] [Google Scholar]
- 40.Kim HW, Rehwald WG, Jenista ER, Wendell DC, Filev P, van Assche L, et al. Dark-Blood delayed enhancement cardiac magnetic resonance of myocardial infarction. JACC Cardiovasc Imaging. 2018;11:1758–1769. doi: 10.1016/j.jcmg.2017.09.021. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 41.Holtackers RJ, Van De Heyning CM, Nazir MS, Rashid I, Ntalas I, et al. Clinical value of dark-blood late gadolinium enhancement cardiovascular magnetic resonance without additional magnetization preparation. J Cardiovasc Magn Reson. 2019;21:44. doi: 10.1186/s12968-019-0556-1. [DOI] [PMC free article] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Data Availability Statement
The datasets generated and/or analyzed during the current study are not publicly available due to data protection but are available from the corresponding author upon reasonable request. The imaging protocol of proposed free-breathing LGE sequence is available from the corresponding author upon reasonable request.