Table 1.
Acquisition sequence parameters.
Sequence | Parameters | Scan duration |
---|---|---|
T2w | Fast Spin Echo | 2 × 03:12.0 |
TR=12 s | ||
TE=156 ms | ||
SENSE factor: axial=2.11, sagittal=2.58 | ||
Sagittal and axial slice stacks | ||
In-plane resolution = 0.8 × 0.8 mm2 | ||
1.6 mm slice thickness overlapped by 0.8 mm | ||
290 × 290 × 203 voxels | ||
T1w | Inversion Recovery Fast Spin Echo | 2 × 05:45.2 |
TR=4795 ms | ||
TI=1740 ms | ||
TE=8.7 ms | ||
SENSE factor: axial=2.26, sagittal=2.66 | ||
Resolution matched to T2w | ||
290 × 290 × 203 voxels | ||
Multiband EPI (func) | Single-shot EPI | 15:03.5 |
TE=38 ms | ||
TR=392 ms | ||
Flip angle=34° | ||
MB factor=9x | ||
2300 vol | ||
2.15 mm isotropic | ||
67 × 67 × 45 × 2300 voxels | ||
Single-band EPI (sbref) | Bandwidth matched readout to multiband EPI | 00:19.4 |
TR=3871 ms | ||
TE=38 ms | ||
4 vol | ||
67 × 67 × 51 voxels | ||
Spin-echo EPI (field map) | Bandwidth matched readout to multiband EPI | 01:52.7 |
TR=8040 ms | ||
TE=63 ms | ||
4xAP/PA interleaved phase-encoding directions | ||
67 × 67 × 51 × 8 voxels | ||
Dual-echo-time (fieldmap) | Dual-TE gradient-echo | 00:20.2 |
TR=10 ms | ||
TE1=4.6 ms; TE2=6.9 ms | ||
flip angle=10° | ||
3 mm isotropic in-plane resolution, 6 mm slice thickness | ||
52 × 53 × 18 voxels |