Abstract
With the rise in public health awareness, research on point-of-care testing (POCT) has significantly advanced. Electrochemical biosensors (ECBs) are one of the most promising candidates for the future of POCT due to their quick and accurate response, ease of operation, and cost effectiveness. This review focuses on the use of metal nanoparticles (MNPs) for fabricating ECBs that has a potential to be used for POCT. The field has expanded remarkably from its initial enzymatic and immunosensor-based setups. This review provides a concise categorization of the ECBs to allow for a better understanding of the development process. The influence of structural aspects of MNPs in biocompatibility and effective sensor design has been explored. The advances in MNP-based ECBs for the detection of some of the most prominent cancer biomarkers (carcinoembryonic antigen (CEA), cancer antigen 125 (CA125), Herceptin-2 (HER2), etc.) and small biomolecules (glucose, dopamine, hydrogen peroxide, etc.) have been discussed in detail. Additionally, the novel coronavirus (2019-nCoV) ECBs have been briefly discussed. Beyond that, the limitations and challenges that ECBs face in clinical applications are examined and possible pathways for overcoming these limitations are discussed.
Keywords: electrochemical biosensors, point-of-care testing, metal nanoparticles, cancer biomarkers, glucose, novel coronavirus
1. Introduction
Biosensors are chemical sensors that utilize biomolecules as the target recognizing component and a transducer that produce an identifiable signal through their interaction [1,2]. In the case of electrochemical biosensors (ECBs), the transducer converts the chemical signal to an electrical signal that allows for qualitative and quantitative identification of the target biomolecules [1,3,4,5]. With the increasing risk of cancer, diabetes, viral infections, and other pathogenic diseases, point-of-care testing (POCT) systems have become essential in the health sector [4,6,7,8]. As a result, research in ECBs has seen an exponential growth because they are inexpensive, provide fast and accurate responses, require almost no sample preparation, and are easy to use [3,8,9,10,11,12,13].
ECBs often take advantage of the unique chemical properties possessed by nanomaterials [14,15,16]. Particularly, metal nanoparticles (MNPs) are mostly used due to their biocompatibility, low toxicity, excellent conductivity, and high surface area [17,18,19,20,21]. Among many, gold (AuNPs), silver (AgNPs), and platinum NPs (PtNPs) are some of the most commonly utilized in ECBs [21,22,23,24,25,26,27,28]. In fabricating biosensors, these MNPs often provide the anchoring site for biorecognition components such as antibodies, enzymes, single stranded RNA (ssRNA) and DNA (ssDNA), aptamers, and affibodies [21,22,23,24,27,29,30,31]. The effectiveness and stability of these biochemical interactions are largely dependent on the physicochemical properties of the MNPs [1,20,32]. This is why material researchers have devised unique strategies for controlling size, shape, and other structural features of these MNPs [32,33]. However, MNPs are often combined with a scaffold for increasing stability and catalytic activity that is usually made of nanostructured material [34,35,36,37]. Of these, the carbon nanostructures are the most popular candidate due to their availability, good conductivity, and stability [9,35,38,39,40]. The various carbon nanostructures used are 0D fullerenes, 1D carbon nanotubes (CNTs), 2D graphene (GR), and 3D graphite materials [41,42,43,44]. In ECBs, the composites of MNPs and these carbon nanostructures are used to anchor the biorecognition components to the transducer that converts the chemical signal to electronic signal. Figure 1 depicts the various MNPs and their composites that are often utilized in ECBs.
ECBs can be manufactured in miniaturized size that can be used as POCT devices for clinical purposes [45,46]. These POCT devices are often fabricated on paper strips or carbon paste electrodes based on lab-on-chip principles and can be used with a portable electroanalytical device [46,47]. Electrochemical glucose biosensors are one of the most successful and promising examples of this technology [48]. However, the practical applications of ECBs for a wide variety of bioanalytes have not been completely realized. Besides the cost of the electrodes, the key technical factors that determine the applicability of ECBs in clinical purposes are: ease of preparation, sensitivity, accuracy, reproducibility, and stability of the modified electrode [49,50,51]. However, the use of biomolecules makes it very challenging to meet all these criteria.
The review is devoted to discussing how using MNPs and carbon composites can help to overcome the limitations of ECBs. The field of ECBs is enormous; hence a general classification of the ECBs is considered for simplification purposes. The progress in the ECB research has also been discussed to provide the reader with a better understanding of the development process over the last decade. The design strategies for tailoring the properties of MNPs and carbon nanostructures that influence the sensing ability has also been explored. The review also discusses the advances in ECBs for sensing small biomolecules and cancer biomarkers. Beyond these, the challenges and perspective course of actions to overcome them have been explored. Hence, the authors hope that the discussions and concepts presented in this review would envisage the fabrication of ECBs that can be applied for POCT.
2. Electrochemical Biosensors
ECBs can achieve high selectivity and accuracy through combining bio-selectivity of biomolecules and sensitivity of the electroanalytical techniques (EATs) [9,52]. Application of ECBs encompasses a wide variety of research area from small biomolecules (dopamine, glucose, xanthine, etc.) to cancer biomarkers and other large biological systems [3,25,26,27,28,29,30,31,32,33,34,35,36,37,42,44]. These biosensors vary from each other based on the use of biorecognition components (BRC) and EAT for the detection process [1,4]. Typically, ECBs comprise three electrodes (working electrode (WE), reference electrode (RE), and counter electrode (CE)) that are placed in contact with the analyte solution in an electrochemical cell [1,6,53]. These three electrodes are connected with an electrochemical workstation that is capable of applying potential and measuring the electrochemical changes due to electron transfer at the interfacial region between the WE (transducer) and solution [46,51]. The circuitry component of the workstation converts such chemical changes at the WE into readable data in terms of current, potential, or conductivity with respect to the WE [51]. That is why understanding the properties of the WE is most important when discussing about ECBs. The WE acts as the transducer which can convert electrochemical reactions into electrical signals [53,54]. The WE is modified with various BRCs and utilizes different EATs to make them sensitive and selective towards a particular type of analyte. In this review, ECBs have been categorized to simplify these wide varieties. Initially, ECBs can be broadly categorized into two classes, based on the type of EATs and BRCs employed. The classification of ECBs based on EATs and BRCs used are represented by a hierarchical list in Figure 2.
2.1. Classification of ECBs Based on BRCs
When considering how the biorecognition systems work, the ECBs can be categorized into two classes:
ECBs modified with biocatalysts
ECBs operating through bioaffinity
2.1.1. ECBs Modified with Biocatalysts
Biocatalysts are BRCs that can produce electroactive species by interacting with biomolecules [55,56]. Enzymes, cells, tissues, and small biomolecules are often used as biocatalysts in the ECBs [55,56]. These ECBs most commonly employ impedimetric, voltammetric, and amperometric methods in their analyses [56,57]. Preparing ECBs with biocatalysts is often cost-effective, simple, and easily scalable to make POCT devices [58]. Of the various biocatalysts used, enzymes are most popular [59]. This is because of enzymes are amino acids that are capable of inducing biochemical catalysis [1,56]. They are capable of interacting with electrochemically inactive bioanalytes and producing electroactive species [26,27,60]. Enzymes often significantly increase the rate of reaction and the kinetic parameters can be readily investigated with simple EATs [1,60]. However, these enzymes are often very sensitive to temperature, pH of the solution, and humidity [61]. That is why it is important to maintain optimal conditions while preparing and using this type of sensors. Researchers are working towards fabricating enzymatic sensors that are more tolerant to these limiting factors. Marzo and coworkers have reported the fabrication of horseradish peroxidase (HRP)-based ECB for highly sensitive H2O2 detection [62]. The sensor uses a composite of AuNP-HRP that is immobilized on 3D screen printed (3D-SP) GR-polylactic (PLA) substrate. The HRP acts as the catalytic enzyme that induces electron transfer from H2O2. The 3D-SP electrode fabrication and its H2O2 detection mechanism are illustrated in Figure 3a. The 3D-SP sensor showed relatively good stability [62]. An enzymatic ECB has been reported, where virus (tobacco mosaic virus (TMV)) was used as a carrier to enhance the sensitivity and stability of glucose oxidase (GOx) enzyme for glucose detection [63]. The TMV carrier containing a glucose sensor showed almost double the current response for the same concentration of glucose compared to the non-TMV containing sensor [63]. Figure 3b shows the glucose sensor chips and their response towards glucose detection. Further research to the advancement of biocatalytic sensors have made them a promising candidate for POCT.
2.1.2. ECBs Operating through Bioaffinity
In these biosensors, a biorecognition element is used that can strongly bind with the target biomolecules to produce a detectable electrical signal [64]. Affinity ECBs utilize antigens to bind with antibodies, various oligomers, ssRNA or ssDNA, and membrane receptors to bind specifically [1,4,65]. These bioaffinity sensors are often used when biocatalytic systems are not applicable. For instance, there are many biomolecules for which there is no enzyme that can selectively catalyze them [1,65,66]. Based on the type of biorecognition molecules used, the bioaffinity-based ECBs can be classified as the following:
Immunosensors
Aptasensors
Affibody-based sensors
Figure 4a shows schematic diagrams of the marker (signal inducer)-labeled sandwich and label-free immunosensor, aptasensor and affisensor.
Immunosensors utilize antibody-antigen (Anb–Ang) interactions for producing detectable electrochemical signals [67]. Immunosensors take advantage of the strong selective chemical affinity between antigen and antibody [68]. As a result, immunosensors are highly selective, very sensitive, and accurate in their detection. Immunosensors can be categorized as traditional label-free/direct, indirect, sandwich, and competitive type based on the mode of operation [69]. Again, based on the change of signal response immunosensors can be categorized as “signal on” and “signal off” [68,69]. In the case of label-free immunosensors, the electrode surface is tailored with an antibody that can bind with the specific antigen [70]. The electrolyte solution usually contains an electroactive redox pair that is responsible for the electrochemical signal. The redox pair interaction with the electrode surface changes based on the concentration of the antigen present on the solution. As a result, the electronic signal varies and allows for quantitative detection of the antigen [71]. The setup can be reversed, where the antigen is immobilized for specific antibody detection. In the direct label-free method, the antigen is immobilized on the substrate and allowed to interact with the antibody [72]. An antibody of hantavirus has been detected from serum solution using an AuNP-hantavirus antigen-modified electrode [73. The sensor used a linear sweep voltammetry (LSV) technique and showed linearity for hantavirus antibody detection for 0.4 to 300 µg/mL. Sensor fabrication and hantavirus detection is shown in Figure 5a. The reported sensor also showed good stability over 21 days [73]. In another work, prostate specific antigen (PSA) detection was carried out with direct label-free method by Camilo et al [72]. The report showed that AuNPs and layer by layer (LBL)-assembled nanostructures can be used for signal amplification in a direct immunosensor detection system while simultaneously lowering the number of biomolecules (antibody) needed. Their proposed sensor required approximately 10 times fewer antibodies compared to traditional PSA immunosensors [72]. Label-free immunosensors are also employed for detecting proteins, hormones, bacteria, etc. [74].
In the case of indirect immunoassays, the quantitative analysis of antigens is carried out by measuring the changes in electrochemical signal due to the conjugation of a labeled secondary antibody with the primary antibody which is already bound to antigen [75]. The design of indirect immunoassay follows two-step binding strategies in which the primary label-free antibody binds to antigen which is first immobilized on the substrate. Later, a labeled secondary antibody is immobilized on it which can recognize the primary antibody and subsequently bind to it. The secondary antibody can be labeled with various electroactive compounds or enzymes which helps in generating the desired signal.
A competitive immunosensing process utilizes the ability of antigen–antibody binding affinity along with the catalytic properties of biocatalysts [76]. Typically, different labeled secondary antibodies compete to bind with inadequately available primary antibody sites. AuNP-modified electrodes have been utilized for the detection of phenolic estrogens through indirect competitive binding processes [75]. The four phenolic estrogens conjugated with the secondary antibody and the binding affinity followed: diethylstilbestrol > dienestrol > bisphenol A > hexestrol. This work utilized a differential pulse voltammetry (DPV) technique [75]. In another work, Hou et al. reported a direct competitive ECB that utilized an electrochemical impedance spectroscopy (EIS) technique for detecting chlorpyrifos [77]. Chlorpyrifos antibodies were initially anchored on a glassy carbon electrode (GCE) surface. Analyte competitor was prepared by combining spherical AuNPs with HRP, bovine serum albumin (BSA), and chlorpyrifos. This analyte competitor then competed with chlorpyrifos to bind with the anchored antibody. This resulted in the formation of insoluble 4-chloro-1(4H)-naphthalenone through biocatalytic process in the presence of H2O2 and 4-chloro-1-naphthol. The proposed electrode linear range from 1.0 × 10−3 ng mL−1 ~10 ng mL−1 [77]. The competitive immunoassay method can be utilized with other immunoassay method for amplifying the signal. This is very effective in lowering the limit of detection (LOD) of ECBs.
Although label-free immunosensors are very selective, they are not adequately sensitive [78]. Hence, sandwich type immunosensors were conceived to overcome this limitation [1,78]. Similar to the label-free system, the antibody (Anb1) is first immobilized on the ECB surface and allowed to interact with the antigen (Ang). However, a second antibody (Anb2) is introduced to the system that interacts only with the Anb1–Ang sites on the electrode surface to produce the sandwich (Anb1–Ang–Anb2) [68,79]. As a result, the change in electronic signal is amplified and the sensitivity is improved. Jampasa et al. developed an ECB for the sensitive detection of LipL32 protein through a “signal on” process [80]. The sensor utilized a graphene oxide (GO) layer for immobilizing the Anb1. The modified electrode was allowed to interact with the antigen. Finally, Au conjugated Anb2 was introduced to the electrode system. The electrode fabrication process is shown in detail in Figure 5b. This interaction process ensured selectivity and high sensitivity. The DPV technique was utilized for the detection process. The sensor showed a stable current response towards LipL32 for over 14 days [80].
Aptasensors were developed to overcome the limitations posed by the immunosensors [81,82]. POCT devices need to be cheap, robust, and easily scalable [65,83]. Using immunosensors it is often difficult to fulfill these criteria. Aptasensors use aptamers (ssRNA, ssDNA, and peptides containing 15–40 bases) with unique binding sites that utilize their nucleic acid arrangements for interacting selectively with target biomolecules [1,84]. Aptamer spatial configurations are changed to enable the best interaction with the target biomolecules [84]. Additionally, aptamers are more stable than antigens, and can easily recover their active spatial configuration after usage thereby allowing for the reuse of the same electrode multiple times [85]. Based on how the immobilized aptamers interact with the target analyte, Han and coworkers proposed the following categories: (a) spatial configuration rearrangement of aptamers based on target interaction [86,87,88]; (b) sandwich type interactions [89,90,91]; (c) dissociation or displacement of aptamers through target interaction [92,93]; and (d) competitive replacement of aptamers [14,94,95,96].
In the case of type-a biosensors, the immobilized aptamers change their configuration with respect to interactions with the target biomolecules [87]. Mazaafrianto and co-workers developed an aptasensor for detecting ochratoxin A (OHA) based on structure switching [88]. The proposed sensor was able to obtain an LOD of 113 pm through the “signal on” method. Figure 6a shows the OHA sensor setup and its interaction process for electrochemical signaling. OHA interaction induced the structure change in the aptamer that then allowed interaction with methylene blue (MB) that resulted in the increased signal [88].
Similar to immunosensor, sandwich type aptasensors (type b) also utilize signal amplification through a double interaction system [69]. Research strategies have been focused on developing label-free sandwich aptasensors for cost-effective and rapid biomolecule detection. Wang et al. proposed an antibody and label-free sandwich sensor for the detection of carcinoembryonic antigen (CEA) cancer biomarker [89]. In this setup, the aptamer was deposited on the Au electrode surface and allowed to interact with CEA. The sensor showed an increased DPV response when Concanavalin A (conA) was allowed to interact with the Au/aptamer–CEA conjugate. However, in the absence of CEA, conA did not show any interaction with the Au/aptamer electrode system shown in Figure 6b. The sensor showed very good selectivity along with a low LOD of 3.4 ng/mL [89].
For type-c ECB systems, the aptamer probe is modified through displacement/dissociation at certain sequences in the presence of the target biomolecules [96]. This allows for high selectivity towards the analyte and amplification of the electrochemical signal. In a work by Li et al. the sgc8 aptamer was used for modifying a hairpin probe (HP2) and detecting protein tyrosine kinase-7 (PTK-7) [93]. HP2 was immobilized on the GCE surface along with HP1. In the presence of PTK-7, the HP2 may undergo structural change exposing the aptamer that hybridized with the HP1. Finally, a redox probe carrying HP3 is introduced that upon interaction produced suitable voltammetric signals. Besides the very low LOD of 1.8 fM, the sensor surface is regenerated through the removal of PTK-7 at the end of each cycle [93]. The ECB fabrication process and PTK-7 detection mechanism are shown in Figure 6c.
For type-d ECBs, the target analyte replaces the aptamer to produce the desired electrical signal [94]. Figure 6d shows the mechanism for such a sensor that was used for detecting hepatocellular carcinoma (HepG2) tumor cells through the “signal off” process [96]. An LBL assembly system was used where AuNPs were initially deposited on ITO (indium tin oxide) electrodes along with a TLS11a aptamer. This was then allowed to interact with the LBL assembly of PtNP-Fc-labeled cDNA (complementary DNA). When no tumor cells were present, the PtNP assembly gave a high current response. However, in the presence of tumor cells the cDNA could no longer effectively bind with the aptamer due to denaturation of the double strand DNA. This resulted in a decreased current signal that was linearly proportional to the logarithm of the HepG2 cell concentration. [96].
Affibody-based sensors are a result of using antibody mimicking bioengineered small protein (6 to 7 kDa) molecules to overcome the limitations of immunosensors [97]. These affibodies are engineered according to the need and have high binding affinity, selectivity, and survivability in high temperature conditions [98]. Antibodies typically contain disulfide bonds that lead to poor heat stability [1]. However, only a small portion of the multidomain protein structure of antibodies is used in antigen detection [1,98]. This is where affibody technology comes into use. The parts of antibodies that are responsible for their affinity and selectivity towards antigens are engineered in vitro. These affibodies are often paired with various metal nanoparticles to further enhance their efficacy [99]. An impedimetric strip ECB for human epidermal growth factor receptor 2 (HER2) biomarker that utilized affibody as the biorecognition element is shown in Figure 7 [100]. AuNPs were used for immobilizing the anti-HER2 affibodies. This resulted in selective interaction with the HER2. Because of that, the impedimetric charge transfer resistance increased linearly with increasing HER2 concentration. Analysis of the experimental results provided an LOD of 6 µg/L for the proposed sensor. Compared to conventional immunosensors, the affibody sensor was more sensitive, provided a more rapid response, and higher specificity [100].
2.2. Classification of ECBs Based on EATs
A wide variety of EATs are currently employed in ECBs [53]. These techniques are sometimes combined to work synergistically to further amplify the electronic signals [53,54]. Therefore, based on the different EATs utilized, ECBs can be categorized as follows:
Amperometric technique
Voltammetric technique
Potentiometric technique
Conductometric technique
Impedimetric technique
Figure 4b shows a schematic representation of ECB sensing process and the different EATs used in them. Below, these techniques are discussed in detail.
2.2.1. Amperometric Technique
This is a sensitive EAT that utilizes an applied potential for inducing oxidation or reduction of the target analyte and the response is observed as a change of current signal with respect to time and the analyte concentration [101]. It is one of the most popular EATs because it offers sensitive detection and is very simple to use [101,102]. The LOD of the amperometric technique is in the range of 10−5 M [103]. The use of specific analyte is advantageous because it allows for limiting interference [104]. At the same time, the charging current is also minimized within a few hundred seconds allowing for a very low limit of detection (LOD) [1,102]. Because of these advantageous properties, ECBs have often utilized amperometric techniques. For instance, a portable ECB was proposed by Dong and coworkers for the sensitive and selective detection of histamine (HA) through an AuNP–chitosan–Prussian blue-modified electrode system [105]. The electrode used an HA antigen that competitively interacted with an HRP-labelled HA antibody for HA detection using an amperometry technique within the 0.01 to 100 µg/L linear range [105]. Figure 8 shows a schematic presentation of the portable sensor fabrication process and its application for HA detection. Combined hydrodynamic and amperometry techniques can further enhance the sensitivity and lower the LOD [106]. A hydrodynamic amperometry technique-based aqueous uranyl ion ECB was reported, which showed a higher sensitivity and lower detection limit compared to steady state amperometric systems that were previously reported [106].
2.2.2. Voltammetric Technique
These are EATs where a certain potential region is scanned, and the signal is displayed in the form of a peak or a plateau [107]. The current response is proportional to the concentration of the analyte present in the system [13,107]. The most commonly used voltammetric techniques include linear sweep voltammetry (LSV), cyclic voltammetry (CV), DPV, and square wave voltammetry (SWV) [108,109,110]. Although the principle is same for all the techniques, they differ in the way that the potential region is scanned. Based on the scanning method, the most sensitive are DPV and SWV [111]. The detection limits for LSV, CV, DPV, and SWV are 10−5, 10−5, 10−7, and 10−8 M, respectively [103]. DPV-based highly sensitive ECB was reported for the detection of Tau-441 protein, which is correlated to cognitive disorder [112]. The proposed ECB utilized an Au electrode that was modified with multi-walled carbon nanotubes-reduced graphene oxide (MWCNT–rGO) and Tau-441 specific antibody. The sensor showed a linear range from 0.5–80 fM with an LOD of 0.46 fM [112]. The voltammetric electrode preparation process is shown in Figure 9. Such low-level detection of bioanalytes with voltametric-based ECBs shows their potential for application in POCT.
2.2.3. Potentiometric Technique
For the potentiometric technique, the change of potential in the electrochemical cell is measured, while the current change is minimal [113]. The potentiometric sensors are also known as ion selective electrodes (ISEs) because they are often designed to generate responses with respect to the change in concentration of a specific ion [113,114]. Their setups are different from the traditional amperometric and voltammetric cells, because they often utilize two reference electrodes that measure the potential change with respect to the target analyte concentration in the cell [113,115]. These ISEs can be converted to ECBs by modifying the electrode with biocatalysts that interact with biomolecules to produce ions that the ISEs can detect [1,113]. Like other ECBs, they can also operate independent of sample volume, have a low LOD, small size, and produce a rapid response. On top of these, potentiometric ISEs are able to provide information regarding the concentration of free ions or ion activity in the cell [114,116]. Manjakkal et al. reported the fabrication of a potentiometric pH sensor that can be used as a wearable device [117]. The sensor showed excellent stability to washing and a good sensitivity of 4 mV/pH in the pH range of 6–9 [117], making it an excellent candidate for POCT for various bioanalytes through the incorporation of proper biorecognition component. The fabricated wearable device is shown in Figure 10.
2.2.4. Conductometric Technique
The change of conductance in the electrochemical setup is measured as a response of interactions between the BRC and analyte [118]. These types of ECBs usually use catalytic biorecognition modifications that result in the change of ionic strength in the cell [118,119]. This change is then measured to determine the biomolecules qualitatively and quantitatively [73,120]. Kolahchi et al. have developed a miniaturized conductometric electrodes for phenol detection [121]. The optical microscopic image and electrochemical setup of the device is shown in Figure 11. In this setup, they used AuNP-immobilized bacteria (Pseudomonas sp. (GSN23)) as the biorecognition component for the detection process. This setup enabled bypassing of the micro-extraction process required for phenol detection with a simple dilution procedure [121]. Conductometric sensors have also been used for the detection of biomolecules from human serum and urine samples, and pathogens from foods for biosecurity purposes [118,122,123].
2.2.5. Impedimetric Technique
In the impedimetric technique, the changes in resistance and capacitance at the interfacial region of the working electrode are measured with respect to analyte concentration [68]. This is also known as electrochemical impedance spectroscopy (EIS) technique. The measurements are carried out through the application of an alternating current as the exciting factor that perturbs the voltage as a function of frequency (1 × 10−5 to 0.7 V) [1,124]. This is one of the most popular techniques used in bioaffinity sensors (immunosensor and aptasensors), because of its high sensitivity to slight changes in impedance [125]. A paper-based impedimetric ECB was developed for the rapid and on-site detection of bacterial contaminations in drinking water [126]. The paper strip was initially carboxylated, and was then used for immobilizing Concanavalin A (conA) lectin. The paper sensor setup process is shown in Figure 12. The impedimetric sensor showed variation of resistance over a wide concentration window with a low LOD of 1.9 × 103 CFU/mL [126]. Impedimetric ECBs are frequently employed for detecting small biomolecules and various biomarkers [127].
3. MNPs and their Composites in ECBs
3.1. Influence of MNP Morphology in Biosensing
The size and shape of metal nanoparticles plays a crucial role in improving the electrocatalytic activity, selectivity in biomolecular binding, metal–electrode/metal–support interactions for electrochemical biosensing applications [63,82,128,129,130,131]. As a result, research into MNP-based ECBs has received considerable attention. Figure 13 shows the rise in research interest in the field of ECBs from 2010–2020. In this section, the extensively studied MNPs are discussed (Au, Ag, Cu, Pt NPs etc.) based on their size, shape, and structure dependent electronic properties that significantly influence their sensing ability [130,131,132,133]. The synthetic procedures that are regularly utilized to create these MNPs and the technological advancement for characterizing these MNPs are also explored in sequence.
3.1.1. AuNPs: Effect of Size and Shape in Biosensing
AuNPs are the most commonly employed MNPs in ECBs due to their outstanding chemical and electrical properties, excellent biocompatibility, and catalytic ability [54,130,131]. These properties resolutely depend on the size and shape of the AuNPs [133,134]. It is well known that the high index facets and edges of the Au surface are more likely to enhance catalytic activity in contrast to flat or spherical surfaces [135]. Depending upon the synthesis protocols, the properties of the AuNPs, particularly the size and shape, can be precisely controlled. Tremendous efforts have been put forward over the decade to synthesize AuNPs with precise control over size (varying from 1–100 nm) and shape [136]. In a typical wet chemical synthesis process (Turkevich and Lee Miesel process), the metal salt is reduced in the presence of a stabilizing agent or an adsorbate or a capping agent which selectively binds to the surface of AuNPs in order to protect them from aggregation and, therefore, imparting greater stability [137,138]. Furthermore, the controlled nucleation and crystal growth mechanism influence the morphology of the prepared AuNPs and thus allow the formation of different shapes, such as Au nanorods, nanocubes, nanowires, nanopyramids/bipyramids nanocages, nanoflowers, etc. [131,133,139].
In one study, 3D-printed tubes were designed for the simultaneous detection of glucose and H2O2 [140]. The 3D-printed tube utilized two working electrodes (WEs). One WE was modified with colloidal PtNPs, and the other one with spherical AuNPs and HRP. The colloidal PtNP was utilized for glucose sensing, while the AuNP–HRP system was used for H2O2 sensing. The sensor showed a broad linear range and low LOD for both analytes, showing the effectiveness of MNP shape control in biosensing ability [140]. The effect of AuNP size on the effectiveness of ECB was investigated by Quintero-Jaime and coworkers [141]. AuNPs were impregnated on functionalized MWCNTs (fMWCNTs) in a ratio of 0.5 and 50. The AuNP–fMWCNT-0.5 ratio system showed AuNPs of 9.5 nm, and the AuNP–fMWCNT-50 ratio system showed AuNPs of 6.6 nm. Based on the size of AuNPs, the prepared ECB showed different linear range and sensitivity for PSA detection. The AuNP–fMWCNT-0.5 ratio and AuNP–fMWCNT-50 ratio systems showed linear ranges from 0–4 ng/mL and 0–6 ng/mL, respectively [141]. The sensor fabrication process is shown in Figure 14.
3.1.2. AgNPs: Effect of Size and Shape in Biosensing
Besides being highly conducting and biocompatible, AgNPs are also one of the most commonly manufactured MNPs [54,142]. They are more easily oxidized compared to AuNPs in an electrochemical setup [19,26,54,143]. This makes them an excellent candidate for ECBs. Despite these advantages, the use of AgNPs is limited by the fact that they are less stable and cannot easily be functionalized [54,144]. Contemporary research in AgNP synthesis for biosensing processes is aimed towards eliminating these limitations. As a result, various methods have been developed for size- and shape-controlled stable AgNPs synthesis [145].
AgNPs of definite size can be produced through the chemical reduction process. The same method that Turkevich and coworkers developed for spherical AuNPs synthesis can also be used for AgNPs [54,137]. AgNO3 is the most commonly used metal salt due to its good solubility in polar solvents. The size of synthesized AgNPs can be controlled through the use of an appropriate reducing agent. Citrate usually produces AgNPs between 50–100 nm diameter, while 5–20 nm AgNPs are obtained when NaBH4 is used [146,147].
Morphological properties of AgNPs can strongly influence its applicability in electrochemical sensing applications. For instance, an MWCNT–AgNPs modifier was shown to be able to detect glucose from 0.025 to 1.0 mM when incorporated on a GCE with GOx [148]. The ECB followed the first-generation mechanism of dissolved oxygen reduction for glucose [9,53]. The average size of AgNPs was determined to be around 5 nm for this system [148]. It has been previously reported that the oxidation potential of AgNPs shifts towards a more negative potential with decreasing size [128]. The thermal scattering is also accelerated in AgNPs smaller than 5 nm [129], suggesting that the small size of the AgNPs played a crucial role in dissolved O2 reduction process.
3.1.3. PtNPs: Effect of Size and Shape in Biosensing
Besides gold and silver NPs, platinum NPs (PtNPs) are also frequently employed in ECB fabrication [149,150]. PtNPs are highly conductive, relatively stable, and have good catalytic activity [150]. Aside from these benefits, PtNPs can catalyze hydrogen peroxide (H2O2) decomposition during an electrochemical process [151,152]. This is an important property because it can work in synergy with enzymatic processes to significantly amplify the electrochemical current response while lowering the overpotential requirement [152]. The electrocatalytic activity of PtNPs is also dependent on structural properties.
As is the case with other MNPs, it is possible to prepare PtNPs that have a definite size and shape. The most commonly employed synthesis strategies include chemical reduction, electrochemical reduction, electrodeposition, and the photochemical reduction of platinum salts (PtCl62− and PtCl42−) [152].
PtNPs with cubic, polygonal, or rod shapes offer better anchoring sites for biorecognition components compared to spherical NPs [151]. Huang et al. reported the development of a highly effective glucose and H2O2 ECB [153]. For this, flower-like AgNPs were decorated with dewdrop-like PtNPs for enhancing the electrocatalytic surface area, selectivity, and stability. Figure 15 shows the synthesis and morphological structures of the ECB. The sensor showed linear range from 1 µM to 2 mM for H2O2 and 1 to 14 mM glucose [152].
3.2. Properties of MNPs Composites
MNPs have excellent electrocatalytic activity. However, when used in biosensing application these properties need to be fine-tuned [153,154]. The most convenient way of tuning the properties of MNPs is through combining MNPs with other conducting nanomaterials (CNMs). Tran and coworkers reported a graphene quantum dot (GQD) and AgNP nanocomposite for detection of glucose [155]. The composite showed a wide linear range of 1–10 mM, although the composite had larger size (~40 nm) compared to previously reported NPs [155]. Because of their ability to work synergistically, CNMs such as fullerenes, GR, rGO, quantum dots (QDs), calixarenes, and carbon nanotubes (CNTs) are frequently employed for preparing MNP composites (MNPCs) [156,157]. The electrocatalytic activity of these MNPCs is largely dependent on the choice of MNPs and CNMs [154]. Hence, it is crucial to have a sound understanding of the physical, chemical, and electrical properties of these CNMs before using them as electrocatalysts in preparing ECBs.
3.2.1. Fullerene-Based MNPCs
Since their discovery in 1985, these sp2 carbon-containing truncated icosahedron-shaped fullerenes have found practical use in a wide range of applications [158]. The C60 and C70 fullerenes can be reduced in a reversible manner for up to six electrons (1e− each step) transfer process in nonaqueous solvents [159]. The electrooxidation of fullerenes is often irreversible in nature [159]. The ease of electrooxidation or reduction is dependent on the size and symmetry of the fullerenes [158,159]. Usually, larger fullerenes are more electroactive [160]. Fullerenes can be easily functionalized through such redox processes. Furthermore, fullerenes can be synthesized in a way that they interact with cations, metal atoms, or small molecules via coordination or adduct formation to form endohedral and exohedral systems [159]. Electron spin resonance (ESR) analysis showed that the metals in endohedral systems are typically in the oxidized form, making the fullerene skeleton negative charge bearers. These endohedral fullerenes act like organic salts, meaning these are capable of interacting with both positively and negatively charged systems. As a result, biorecognition systems carrying opposite charges can be utilized with ease for preparing biosensors [159,161]. Compton et al. first reported fullerene-modified electrodes for sensor application. MNPs can be incorporated with fullerenes through both endohedral and exohedral means for preparing ECBSs [159].
3.2.2. CNT-Based MNPCs
CNTs are a class of nanostructured CNMs that are of tubular shape with few nanometers in diameter, and lengths in the micrometer range [161]. These are either single wall CNTs (SWCNTs) or MWCNTs, based on the number of rolled-up layers [158]. Some of the unique properties of CNTs are excellent electrical and thermal conductivity, good elasticity (~18%), high tensile strength and flexibility, and good biocompatibility [118,152]. It has been shown that synthetic cardiac cells can be grown on CNTs without any significant toxic effects [162]. Besides, both SWCNTs and MWCNTs can be functionalized to facilitate binding with biorecognition entities through electrostatic interactions or covalent linkages for fabricating effective ECBs [141,163].
3.2.3. GR-Based MNPCs
Single sheet GR (graphene) was first prepared through mechanical exfoliation by Geim and coworkers in 2004 [9]. Of all the allotropes of carbon, the electronic properties of GR are the most fascinating. GR can be considered in biosensing applications as single layer GR (SLGR), few layer GR (FLGR), and multilayer GR (MLGR) [164]. SLGR is crystalline in nature, FLGR is crystalline up to 10 layers, and beyond that 3D graphite-like properties are observed [9,156]. SLGR sheets have very high conductivity (~1.6 × 105 S/cm) and low resistance (30 Ω/sq) [9]. The anharmonic stretching and bending vibrational modes of sp2 hybridized GR sheets are responsible for forming the finite “rippled” structures that stabilize the 2D sheets while promoting unique electronic properties that are not observable in other allotropes of carbon [165]. These properties include the absence of a weak localization force, ambipolarity of charge carrier concentration, and ballistic electron transport [166,167]. GR can be derivatized for producing graphene nanoribbons (GRNs), GO, rGO, GR nanowalls (GRWs), and GQDs [9]. Each of these derivatives have their own electronic properties that enables them to be used in diverse biosensing platforms [155]. GR sheets can be considered as the building block for the other allotropes of carbon. Figure 16 shows a schematic representation of how CNTs, fullerenes, and graphite are related to GR.
Table 1 discusses MNPC-based ECBs that have been reported for the detection of various biomolecules based on the size and shape of MNPs. It can be seen from the table that MNPs with sizes between 5 and 20 nm are the most-used individual component in the ECB fabrication process [4,23,26,27,29,41,44]. Although size variation is well studied, the shape of the MNPs used is almost always spherical [4,26,41,44,48,85]. The area of MNP shape control deserves significant attention, because NPs with unique shapes (hollow spheres, cubic, porous, pyramidal, etc.) are likely to offer better catalytic activity, increased surface area, and enhanced stability for the ECBs [23,28,39].
Table 1.
Transducer | Biorecognition Component | MNP | MNP Size | MNP Shape | Biointeraction and EAT | Analyte | References |
---|---|---|---|---|---|---|---|
Ab-N,S-GQDs@AuNP–PANI | anti-HEV antibody | AuNP | 6–14 nm | spherical | Bioaffinity immunosensor, pulse impedance | HEV | [4] |
AuNP/SWCNTs/PDA/gold electrode | probe DNA | AuNP | 15 nm | oval | Bioaffinity aptasensor, LSV | target DNA | [23] |
HRP@PGA-C/AgNP | HRP | AgNP | 5–8 nm | spherical | Enzyme-based Biocatalytic, amperometry | H2O2 | [26] |
PtNPs/MWCNT/PEDOT | glutamate oxidase | PtNP | ~12–20 nm | spherical | Enzyme-based Biocatalytic, amperometry | Glutamate | [27] |
PtNP/GR/SPCE | ortho-phenylenediamine | PtNP | 400 nm | urchin | Non-enzymatic Biocatalytic, CV | Cotinine | [28] |
AuNPs–CS/GR/CPE | single-stranded DNA Aptamer | AuNP | 10–20 nm | spherical | Bioaffinity aptasensor, DPV | Activated protein C | [29] |
Au/AuNP–avidin-HRP | avidin-HRP | AuNP | 40 nm | nanowall | Enzyme-based Biocatalytic, amperometry | DNA Methylation | [39] |
MWCNT–AuNPs/GCE | E-cadherin antibody–QD | AuNP | 5 nm | spherical | Bioaffinity immunosensor, DPV | Epithelial-mesenchymal transition | [41] |
Cu–nanoflower@AuNPs–GO NFs coated Au chip | GOx–HRP–Cu nanoflower | AuNP | 20 nm | spherical | Enzyme-based Biocatalytic, amperometry | Glucose | [44] |
PB–CS–AuNPs/SPCE | HRP-labeled histamine antibody | AuNP | ~50 nm | spherical | Bioaffinity immunosensor, amperometry | HA | [105] |
DMF–EC/AuNPs/HRP | HRP | AuNP | ~20–30 nm | spherical | Enzyme-based Biocatalytic, amperometry | H2O2 | [62] |
BSA/anti-A(1–42)/AuNPs/MPA/Au | monoclonal antibody mAb DE2B4 | AuNP | ~30 nm | spherical | Bioaffinity immunosensor, SWV | amyloid beta 1–42 | [70] |
ds-ATPA/TBA on AuNPs–MoS2/GCE | ds-ATPA and TBA | AuNP | 10 nm | spherical | Bioaffinity aptasensor, SWV | ATP and Thrombin | [85] |
AgNPs@GQDs/CS/GCE | GOx | AgNP | 40 nm | spherical | Enzyme-based Biocatalytic, CV and amperometry | Glucose | [155] |
GOx/PtNP@SnS2/Nafion/GCE | GOx | PtNP | ~20–40 nm | spherical | Enzyme-based Biocatalytic, amperometry | Glucose | [168] |
XO/AuNP/PtNP/MWCNT/GCPE | Xanthine oxidase | AuNP PtNP |
50 nm 5 nm |
Spherical spherical |
Enzyme-based Biocatalytic, CV | xanthine | [169] |
AgNPs–Aβ/PrP95–110/GE | peptide | AgNP | ~15 nm | spherical | Bioaffinity aptasensor, LSV | beta-amyloid | [170] |
HEV: Hepatitis E virus; GQD: graphene quantum dot; PANI: polyaniline; PDA: poly dopamine; PEDOT: Poly(3,4-ethylenedioxythiophene); SPCE: screen-printed carbon electrode; CS: chitosan; DMF: dimethylformamide; PB: prussian blue; BSA: bovine serum; MPA: 3-mercaptopropionic acid; ATPA: ATP aptamer; XO: Xanthine oxidase.
4. MNP-Based ECBs for Biomolecule Detection
The previous section discussed the properties of MNPs and MNPCs that influence their effectiveness in ECBs. This section will discuss about the advances in ECB design strategies for the rapid and effective detection of small biomolecules (SBMs), cancer biomarkers, and the COVID-19 virus.
4.1. MNPs in Small Biomolecule Sensing
SBMs are organic compounds that do not have extended polymerization and are responsible for maintaining chemical balance throughout the body [171,172]. In the case of diabetes, the world’s ninth deadliest disease, the blood glucose level can change drastically (up to 30 mM) from the normal level of 4 to 8 mM [173,174]. Dopamine (DA) is a neurotransmitter that is responsible for controlling the motor and sensory nerves, feelings, and various other body functions [22,175]. Abnormality (normal level 0.01–1 µM) in DA level is responsible for attention deficit hyperactivity disorder, Alzheimer’s disease, and schizophrenia [175]. Furthermore, the irregularity in DA level can be used for determining Parkinson’s disease and HIV [175,176]. Uric acid (UA) is the end product of purine metabolism that is normally present in the body within 100 µM concentration range [176]. It helps to relieve stress caused by oxidative processes. However, in excessive content it can cause gout and hyperuricemia [176]. During the production of antioxidant UA through the xanthine oxidation process, H2O2 is formed as a short-lived oxidizing agent that can show acute toxicity [177]. Usually, only a trace amount of H2O2 (10–100 nM) is found in the circulated blood, which makes detection an arduous task [177]. While glucose detection requires a wide linear range, other SBMs necessitate a highly sensitive approach. ECBs that are used in the detection of SBMs are often biocatalytic in nature and utilize impedimetric, amperometric, and voltammetric EATs [129,154].
GOx is one of the most frequently used enzymes for the glucose detection ECBs, because it is capable of inducing a direct electron transfer through its two flavin adenosine dinucleotide (FAD) coenzymes [9]. The enzyme-based glucose sensors are categorized as: (i) first generation ECBs that utilize O2 molecules as mediators to oxidize FAD to FADH2; (ii) second generation ECBs that use artificial mediators for glucose sensing; (iii) third generation ECBs that induce direct electron transfer between glucose and the immobilized enzymes [9]. The mechanisms of these three types of glucose sensors are shown in Figure 17. MNPs can significantly enhance the performance of enzymatic glucose sensors through providing a high surface area, alternative low energy catalytic pathway, and stability for immobilized enzymes [9]. A PtNP-coated SnS2 enzymatic (GOx) glucose sensor was reported with linear range from 0.1–12 mM [168]. Authors concluded from morphological analysis of the prepared electrochemical glucose biosensor that the use of hydrophilic PtNPs significantly enhanced GOx immobilization. This in turn resulted in the sensitive detection of glucose over the wide linear range [168,178]. Magnetic NiNPs have been used for directly immobilizing GOx [179]. The glucose sensor showed linearity up to 12 mM with an LOD of 0.42 mM. The proposed sensing mechanism for the sensor is shown in Figure 18a. The magnetic NiNP sensor did not need to incorporate any other binding material for GOx immobilization [179].
Although HRP is most commonly employed for H2O2 detection, other redox-inducing biorecognition components such as ferredoxin, cytochrome C, and hemoglobin are also utilized [180]. A myoglobin-based H2O2 sensor was reported which used MoS2 NPs and GO [176]. The myoglobin/MoS2 NP/GO system showed the best current response along with better stability compared to only myoglobin/MoS2 NPs or myoglobin/GO systems [181]. In another study, cytochrome C enzyme was used for fabricating the H2O2 sensor—Au nanocubes were utilized for immobilization of cytochrome C [182]. Figure 18b shows the fabrication process of the reported H2O2 ECB. The sensor showed a linear range from 100–1000 µM for H2O2 detection [182]. The use of cubic NPs enhanced the electroactive surface and incorporated a better electron transfer mechanism for biocatalytic H2O2 reduction. This work shows the importance of shape- and size-controlled MNP fabrication for use in the ECBs.
Unlike H2O2 and glucose, there is no specific biocatalyst that is employed for the detection of DA. Paulraj and coworkers showed that polyaniline (PANI)-coated AgNP-modified electrodes can be used for simultaneously detecting DA and H2O2 [178,183]. The sensor was utilized for oxidizing DA and reducing H2O2. The proposed sensor showed a low LOD of 0.03 and 0.12 µM for H2O2 and DA, respectively [178,183]. However, UA is often detected with the help of uricase (UOx) enzyme [182,184]. A Cu2ZnSnS4 NP-modified ECB was reported for the detection of UA [182,184]. The sensor utilized UOx in combination with the Cu2ZnSnS4 NPs for the detection of UA. The low LOD of 0.066 µM and wide linear range of 50 to 700 µM shows the effectiveness of using UOx enzyme in the modification process [182,184]. In another work, an AuNP and rGO complex was utilized for the immobilization of UOx [183,185]. This biosensor was utilized for the rapid detection of UA from human serum samples. The sensor required lower positive potential (low overpotential) compared to traditional electrochemical sensors. The linear range was from 50 to 800 µM with good selectivity and real sample analysis results [183,185]. These works show the advantage of using ECBs compared to traditional methods in the detection of small biomolecules. The high sensitivity, good selectivity, ease of preparation, rapid detection, and cost-effectiveness are the most attractive aspects of these ECBs. Table 2 discusses the fabrication process and EATs utilized for the detection of small biomolecules.
Table 2.
Transducer | MNP | Biorecognition Component | Analyte | Linear Range | LOD | References |
---|---|---|---|---|---|---|
HRP@PGA-C/AgNP | AgNPs | HRP | H2O2 | 1–3000 µM | 0.35 µM | [26] |
Cu–nanoflower@AuNPs–GO NFs coated Au chip | AuNPs | GOx–HRP–Cu nanoflower | Glucose | 0.001–0.1 mM | 0.018 µM | [44] |
DMF-EC/AuNPs/HRP | AuNPs | HRP | H2O2 | 25–100 µM | 9.1 µM | [62] |
AgNPs@GQDs/CS/GCE | AgNPs | GOx | Glucose | 0.1–10 mM | 0.01 mM | [155] |
GOx/PtNP@SnS2/Nafion/GCE | PtNPs | GOx | Glucose | 0.1–1 mM 1–12 mM |
2.5 µM | [168] |
GOx–NiNP/Ni/Au | NiNPs | GOx | Glucose | 1–12 nM | 0.42 mM | [179] |
Ag-doped PANInanocomposites/GCE | AgNPs | AgNP | DA H2O2 |
10–90 µM 10–90 µM |
0.12 µM 0.03 µM |
[183] |
Ag/MoS2/ITO | AgNPs | Ag/MoS2 | DA | 0.2–50 µM | 0.2 µM | [186] |
GOx/PtNP/acetic acid-treated LIG | PtNPs | GOx | glucose | 0.3 µM–2.1 mM | 0.3 µM | [187] |
GOx/PVA-Fe3O4/Sn | Fe3O4 | GOx | glucose | 1–30 mM | 0.5 mM | [188] |
UOx/EDC:NHS/CZTS/ITO | Cu2ZnSnS4 | UOx | UA | 50–700 μM | 1.3 µM | [184] |
UOx/Au-rGO/ITO | AuNPs | UOx | UA | 50–800 μM | 7.32 µM | [185] |
AuNFs/Fe3O4@ZIF-8-MoS2 | AuNPs | Fe3O4@ZIF-8 | H2O2 | 5 µM–120 mM | 0.9 µM | [45] |
GCE/HUA/HNT/FAD | HNT | HUA and FAD | H2O2 | 1–250 μM | 0.49 µM | [189] |
PGA: poly(glutamic acid); NF: nanoflower; GOx: glucose oxidase; UOx: uricase; HUA: humic acid; HNT: Halloysite nanotube; FAD: flavin adenosine dinucleotide; LIG: laser-induced graphene; ITO: indium tin oxide.
4.2. MNPs in Cancer Biomarker Detection
Carcinogenesis happens at a genetic level in the cell, and follows a complex pathway that ultimately disturbs the homeostatic equilibrium by altering the cell death and cell proliferation rate [185]. Despite tremendous efforts and development, treatment of cancer is still challenging because of the following reasons: (i) proliferation of cancer cells and cell proteins through mutation of the proto-oncogenes; (ii) rejection of growth inhibition signals; (iii) evasion of apoptosis or activating anti-apoptotic genes in cells [69,190,191]. For early-stage diagnoses of cancers, tracking of the disease-specific biomarkers is essential. Biomarkers are characteristics biomolecules overexpressed in the beginning of carcinogenesis, either by the body immune system in response to the disease or by the tumor cell itself [11,71,97,100]. Biomarkers could be utilized to assess the responses from the body towards a specific treatment process for controlling disease [191]. A wide range of biomarkers based on genetic, proteomic, glycomic, etc., are well established for detecting cancers and the prognosis processes [192,193,194]. Evaluation of these different biomarkers in bodily fluids such as serum, blood, urine, saliva, tears, and sputum would require noninvasive and cost-effective methods for cancer screening [191,192,193,194]. As a result, biomarker detection based on electrochemical methods has been perceived as an effective early-stage diagnosis of cancer, even though clinical sampling and analysis is still in its infancy. The type of materials utilized for electrode modification are mostly MNPs and MNPCs. This has been discussed in detail in Section 3. MNPs provide improved biocompatibility, better surficial stability, and binding affinity for biomolecule conjugation [82,105,130]. Such surface immobilization of BRCs (antibody, peptide, or aptamer) for preparing immune/aptasensors depends on the functionality of the biomolecule and the type of nanostructured electrode modifiers used. Both of these need to be compatible with each other [62,155,179,183]. For example, AuNPs allow thiol-functionalized antibody/aptamers to be anchored over the electrode surface via activation of the thiol group (SH)–Au bond, which is one of the fundamental pathways followed in most electrochemical bioaffinity sensor preparations [97,152]. In short, bioaffinity ECB functions are based on such binding capacities of the BRCs via interactions with the nanostructured electrode materials. The BRCs are mainly antibodies (mono/polyclonal), aptamers (single stranded ssDNA sequence/RNA), peptides, etc., which can effectively capture target antigens or biomarkers while constructing bioaffinity ECBs [44,85,98,155]. Table 3 shows the mostly studied bioaffinity ECBs for different biomarkers based on MNPs and nanocomposites assays.
Table 3.
Nanostructure Biorecognition Molecules Modified Transducer (Base Electrode *) |
Type of ECBs | EAT | Linear Range (ng/mL) |
LOD (ng/mL) |
Reference |
---|---|---|---|---|---|
Carcinoembryonic antigen (CEA): Colorectal, pancreatic, breast, and liver cancers | |||||
Au*/COOH–rGO/ PdAuPt/antiCEA | Immunosensor | DPV | 0.012–85 | 0.008 | [195] |
AuNPs/TiO2–GR/HRP–Ab2 and GCE*/AuNPs/Ab1/CEA | HRP-labeled Sandwich immunosensor | DPV | 0.005–200 | 3.33 × 10−6 | [196] |
Fc–AuNPs–Ab2 and Au*/LPA–NHS/Ab1/CEA | Fc-labeled Sandwich immunosensor | SWV | 0.05–20 | 0.01 | [197] |
Ag@CeO2–Au–Ab2 and GCE*/AuNPs/Ab1/CEA | Sandwich immunosensor | CV, EIS | 0.0001–5 | 32 × 10−6 | [198] |
MoS2/g-C3N4/PtCu/Ab2 and GCE*/AuNPs/Ab1/CEA | Sandwich immunosensor | i-t | 0.0001–80 | 3 × 10−5 | [199] |
Fe3O4@AuNPs–DNA(S1)–S2–S3–CEA-Exoll/Hemin | Magnetic aptasensor | DPV | 0.1–200 | 0.0004 | [200] |
Ag–PANI@MnO2/Ab2 and GCE*/AuNPs/Ab1/BSA/CEA | Sandwich immunosensor | DPV | 0.0005–80 | 0.00017 | [201] |
Cu–MOFs–TB/PDA/Ab2 and GCE*/MWCNT/CS/Ab1/CEA | TB-labeled sandwich immunosensor | DPV | 2 × 10−5–200 | 3 × 10−6 | [202] |
Mag–SPCE*/AuNP–MnO2/ Fe3O4@Au/antiCEA |
Immunosensor | LSV EIS |
0.001–100 | 0.0001 (LSV) 0.0003 (EIS) |
[203] |
GCE*/PDA–rGO/Ag–Au/antiCEA | Immunosensor | CV | 0.001–80 | 2.86 × 10−4 | [204] |
GCE*/HNF/AuNP/cMWCNT/antiCEA | immunosensor | EIS | 0.4–125 | 0.09 | [205] |
CSH/Ab2/BSA and GCE*/MoS2–Au/Ab1/CEA | Sandwich immunosensor | DPV | 0.0001–80 | 3 × 10−5 | [206] |
GCE*/Au@PDA@Fe–MOF/NH2–aptamer/BSA/CEA | Aptasensor | DPV | 1 × 10−6–1000 | 3.3 × 10−7 | [207] |
GCE*/CNT@PAMAM/CdSe NP/Ab2/CEA/Ab1/Fe3O4 | Cation-labeled sandwich immunosensor | SWV | 0.005–50 | 0.0017 | [208] |
MWCNTs/CoS2@PANI/HRP and GCE*/Au/Ab1/BSA/CEA | HRP-labeled sandwich immunosensor | i-t | 0.001–40 | 0.0003 | [209] |
GCE*/NCMT@Fe3O4@CuSiO3/ConA/CEA/AuNC-aptamer | Cation-labeled aptasensor | DPV | 0.03–6 | 5.38 × 10−3 | [210] |
CeO2-MoS2/Pb2+/Ab2 and GCE*/Au/Ab1/BSA/CEA | Cation-labeled sandwich immunosensor | SWV | 0.001–80 | 0.0003 | [211] |
MoS2@Cu2O/Fc/Ab2 and GCE*/Au/Ab1/BSA/CEA | Fc-labeled sandwich immunosensor | SWV | 0.001–80 | 3 × 10−5 | [212] |
Au*/Ni-Co-PBA/aptamer/CEA | Aptasensor | EIS | 0.001–5 | 7.4 × 10−7 | [213] |
GCE*/MWCNT-SO3H/Rh@Pd ND/Ab1/BSA/CEA | Immunosensor | DPV | 2.5 × 10−5–100 | 8.3 × 10−6 | [214] |
GCE*/rGO-PtAu NP/antiCEA | Immunosensor | SWV | 1 × 10−5–100 | 7 × 10−6 | [215] |
Ag-MOF/AuNP/Ab2 and GCE*/MWCNT/Ab1/BSA/CEA | Ag(I)-labeled sandwich immunosensor | DPV | 0.05–120 | 8 × 10−6 | [216] |
GCE*/GO-AuNP/antiCEA | Immunosensor | SWV | 1–40 | 0.0158 | [217] |
AgNP@Strp-HRP/Ab2 and GCE*/T–GO/AuNP@Strp/Ab1/CEA | HRP-labeled sandwich immnusensor | DPV | 0.0001–0.005 | 7.5 × 10−5 | [218] |
Fc: ferrocene; LPA-NHS: Lipoic acid N-hydroxysuccinimide ester; rGO: reduced GO; HNF: honey nanofibers; BSA: bovine serum; PAMAM: poly(amidoamine); Strp: streptavidin; | |||||
Prostate specific antigen (PSA): Prostate cancer | |||||
Ab2–HRP and GCE*/C60/PANI@PdNP/Ab1/BSA/PSA | HRP-labeled sandwich immnusensor | CV | 0.00016–38 | 1.95 × 10−5 | [219] |
GCE*/MoS2–GA@AuNP/antiPSA/BSA | Immunosensor | DPV | 1 × 10−5–50 | 3 × 10−6 | [220] |
Cu3(BTC)2/Ab2 and Au*/PG@PDA/Ab1/CAS/PSA | Sandwich immunosensor | i-t | 0.1–10 | 0.025 | [221] |
d-Ti3C2Tx@AuNP/Ab2 and GCE*/ATP–GO@AuNP/Ab1/BSA/PSA | Sandwich immunosensor | DPV | 0.00001–0.001 | 3 × 10−6 | [222] |
SPCE*/AuNP@aptamer/MCH/PSA | Aptasensor | DPV | 0.001–200 | 7.7 × 10−5 | [223] |
GCE*/CeO2–MnO2/antiPSA | Immunosensor | SWV | 0.005–50 | 0.005 | [224] |
SPCE*/GO/antiPSA | Immunosensor | DPV | 0.75–100 | 0.27 | [225] |
Au*/MPA/f-PSA/BSA andAu*/MPA/t-PSA/BSA | Immunosensor | EIS | 0.00002–200 | 3 × 10−6 (f-PSA) 4 × 10−6 (t-PSA) |
[226] |
Ab2-CdNi QDs and GCE*/Fe3O4@TMU-10(MOF)-CS/Ab1/BSA | QD-labeled sandwich immunosensor | DPV | 0.001–100 | 0.00045 | [227] |
Ab2-HRP and GCE*/RC60/CuNP@HQ/Ab1/BSA | HRP-labeled Sandwich immunosensor | DPV | 0.005–20 | 0.002 | [228] |
GCE*/GQDs–CS–Naf-IL/MWCNT–GR–IL/PPY-MoS2–il–AuPt NP/aptamer/PSA | Aptasensor | SWV | 0.0005–0.35 | 0.00014 | [229] |
GCE*/MWCNT@AuNP-GQD/Ab1/BSA/PSA | Immunosensor | EIS | 0.001–10 | 0.00048 | [230] |
CPE*/Gr-Fe3O4 NP/antiPSA | Immunosensor | DPV | 1–150 | 0.38 | [231] |
Au IDE*/16-MUA+EDC/NHS/antiPSA | Immunosensor | f-EIS and nf-EIS | 0.01–100 and 0.5–1000 | 0.01 and 0.5 | [232] |
Au*/Peptide/GO@AgNP | ECB | LSV | 0.005–20 | 0.00033 | [233] |
GCE*/rGO-NH2/AgPtPd-COOH/antiPSA | Immunosensor | DPV | 4 × 10−6–300 | 4 × 10−6 | [234] |
GCE*/AuNP/rGO@AuNP/antiPSA | Immunosensor | SWV EIS |
5.5 × 10−8–0.25; 1–36 (SWV) and 0.0018–41.15 |
0.06; 0.002 (SWV) and 0.006 (EIS) |
[235] |
GCE*/PANI@AuNP/Pep/aptamerPSA/BSA | Aptasensor | DPV | 0.0001–100 | 8.5 × 10−5 | [236] |
Ab2-HRP and CASPAuE-MFD*/Magb-Ab1/BSA/PSA | HRP-labeled Sandwich immunosensor | i-t SWV |
0.001–10 | 0.00084 (i-t) 25.4 fM (SWV) |
[237] |
Ab2/AuNP@cys-AgNP@Cu2+ and GCE*/GS@SnO2/Au@Pt/Ab1/BSA/PSA | Cu2+-labeled sandwich immunosensor | SWV | 0.01–100 | 0.00384 | [238] |
GCE*/fMWCNT@AuNP-0.5/antiPSA and GCE*/fMWCNT@AuNP-50/antiPSA | Immunosensor | CV | 0–4 and 0–6 | 85 and 56 | [141] |
GCE*/AuNP/Ab1/BSA/PSA/Ab2-S0/DNA concatemer(S1-S2)-AgNP Signal probe–DNA concatemer (S1-S2)–AgNP |
Ag-labeled sandwich immunosensor | DPV | 0.0001–75 | 3.3 × 10−5 | [239] |
Ab2–HRP and SPE*/CS/AuNP/Ab1/BSA/PSA | HRP-labeled Sandwich immunosensor | SWV | 1–18 | 0.001 | [240] |
SPCE*/rGO@ thionine–AuNP/DNA aptamer/ | Aptasensor | DPV | 0.05–200 | 0.01 | [241] |
PtNP-Ab2/BSA–CuNP and GCE*/AuNP/Ab1/BSA/PSA | CuNP-labeled sandwich immunosensor | SWV | 0.0005–100 | 14.57 × 10−5 | [242] |
Au@Ag-Cu2O/Ab2 and GCE*/Au@N–GQDs/Ab1/BSA/PSA | Sandwich immunosensor | i-t | 1 × 10−5–100 | 3 × 10−6 | [243] |
CAS: casein; MCH: 6-mercapto-1-hexanol; QD: quantum dot; MPA: 3-mercaptopropionic acid; 16-MUA: 16-[Mercaptoundecanoic acid; EDC: N-(3, Dimethylaminopropyl)-N-ethyl-carbodiimidehydrochloride; NHS: Lipoic acid N-hydroxysuccinimide ester; CASPE-MFD: commercially available screen-printed electrode-based microfluidic devices; AE: gold electrode; | |||||
Cancer Antigen 125 (CA125): Ovarian cancer, breast cancer, lymphoma | |||||
Ab2–Suc–Chi@MNPs–TB and GCE*/PAMAM/AuNP–3DrGO–MWCNT/Ab1/Glu/CA125 | TB-labeled sandwich immunosensor | SWV | 0.0005–10 and 10–75 U/mL | 0.006 mU/mL | [244] |
MB–mAb–HRP–CA125 and Au*/Aptamer | HRP-labeled sandwich aptasensor | CV | 2–100 U/mL. | 0.08 U/mL | [245] |
Ag–PPy–pAb2 and ITO*/MB–mAb1 | Sandwich immunosensor | LSV | 0.001– 300 U/mL |
7.6 mU/mL | [246] |
Ab2–GPTMS–SiNPs and Au*/f-GNS/Ab1/CA125 | Sandwich immunosensor | DPV | 1 × 10−9–1 × 10−15 | 1 × 10−15 | [247] |
TB or Fc–Probe@Au–TiO2@Ab2 and Ta*/BDD/VBG–Au/Ab1–CA125 or CEA | TB/Fc-labeled sandwich immunosensor | DPV | CA125: 0.5–100 mU/mL CEA: 0.0005-0.1 |
CA125: 0.09 mU/mL CEA: 0.00015 |
[248] |
SPCE*/rGO/thionine /AuNPs/antiCA125/BSA | POCT immunosensor | DPV | 0.1–200 U/mL | 0.01 U/mL | [249] |
ITO*/AgNPs–PAN–oxime NFs/aptamer/cDNA–MB/CA125 | MB-labeled aptasensor | DPV | 0.01–350 U/mL | 0.0042 U/mL | [92] |
GCE*/AgNPs–GQD/antiCA125/BSA | Immunosensor | DPV SWV |
0.01–400 U/mL | 0.01 U/mL | [250] |
Hollow MB–PDA–Ab2 and GCE*/Au–rGO/Ab1/BSA/CA125 | MB-labeled sandwich immunosensor | DPV | 0.0001–100 U/mL | 336 nU/mL | [251] |
GCE*/ATA–CNT–MSA; CdS–Ab2/AFP/Ab1; ZnS–Ab2/CEA/Ab1; HgS–Ab2/CA19-9/Ab1; PbS–Ab2/CA125/Ab1 |
CdS, ZnS, HgS and PbS-labeled ratiometric multiplex sandwich immunosensor | DPASV | AFP and CEA: 0.0004–10; CA19-9 and CA125: 0.004–100 U/mL |
AFP: 0.00011; CEA: 0.0023; CA19-9: 0.68 mU/mL; CA125: 1.4 mU/mL | [252] |
SPCE*–AuNP/antiCA125/BSA and SPCE*–PtNP/antiCA125/BSA | Immunosensor | EIS | 450–2916 | AuNP: 419; PtNP: 386 | [253] |
GCE*/Ag–DPA–GQDs/CysA–AuNP/antiCA125/BSA/CA125 | Immunosensor | DPV | 0.001–400 U/mL | 0.001 U/mL | [254] |
SPE*/Au–AgNPs/antiCA125/BSA (ISA) and SPE*/Au–AgNPs/CysA/antiCA125/BSA (ISB) | Immunosensor | EIS | ISA: 1–500 IU/mL; ISB: 1–1000 IU/mL | 1.03 IU/mL | [255] |
GCE/FA@H–PANI@CS–HCl/Ab–Ag@Co3O4/BSA/CA125 | FA-labeled immunosensor | DPV | 0.001–25 | 0.00025 | [256] |
Ab2–AuNPs–LOx and GCE*/GO/MWCNT/AuNPs–CS/Ab1/BSA | Enzymatic immunosensor | i-t | 0.01–0.5 and 0.5–100 U/mL | 0.002 U/mL | [257] |
AuNP–Ab2–Cd2+ and ITO*/GNR/Ab1/CA125 | Cd2+-labeled sandwich immunosensor | DPV | 20–100 U/mL | 3.4 U/ mL | [258] |
GCE*/p(CTAB-CS)-AuNP/antiCA125/BSA | Immunosensor | DPV | 0.001–400 U/mL | 0.001 U/mL | [259] |
GCE*/PDA/ERGO/CysA–AuNPs/antiCA125–HRP/BSA | HRP-labeled immunosensor | SWV | 0.1–400 U/mL | 0.1 U/mL | [260] |
TB: toluidine blue; MB: methylene blue; GNS: gold Nanostructures; GPTMS: glycidyloxypropyl trimethoxysilane; BDD: boron-doped diamond; VBG: vertical boron doped graphene; FA: ferro-cenecarboxylic acid; pCTAB: poly cetyl trimethylammoniumbromide; AE: gold electrode; GNR: gold nanorods; | |||||
Human epidermal growth factor receptor 2 (HER2): Breast cancer | |||||
SPCE*/AuNPs–MPA/NSCeO2/NHS–PEG– Maleimide/antiCA125/BSA | Immunosensor | DPV | 0.001–0.5 and 0.5–20 | 0.0349 | [261] |
CdSe@ZnS QDs–Ab2 and SPCE*/c-MagBs/Ab1/EA/HER2 | Cd2+-labeled sandwich immunosensor | DPASV | 0.50–50 | 0.29 | [262] |
AE*/MnFePBA@AuNP/Aptamer/HER2 and AE*/MnFePBA@AuNP/Aptamer/MCF7 | Aptasensor | EIS | HER2: 0.001–1; MCF7: 500-5 × 104 cell/mL | 0.000247; 36 cell/mL | [263] |
SPCE*/Ab1/HER2–ECD/Ab2/CdSe@ZnS QDs | Cd2+-labeled sandwich immunosensor | DPASV | 10–150 | 2.1 | [264] |
AE*/CDs@ZrHf–MOF/Aptamer/HER2 | Aptasensor | EIS | 0.001–10; 1000–1 × 105 cell/mL | HER2: 19 × 10−6; MCF7: 23 cell/mL | [265] |
Ab2–CDI–PbS QDs and SPCE*–COOH/EDC–NHS/Ab1 | Pb2+-labeled sandwich immunosensor | SWV | 1–100 | 0.28 | [266] |
GCE*/ErGO–SWCNT/AuNP/Aptamer/HER2 | Aptasensor | EIS | 0.0001–1 | 5 × 10−5 | [267] |
GNR–Pd SS–Aptamer–HRP and AE*/DNA tetrahedron/BSA/HER2 | HRP-labeled sandwich aptasensor | DPV | 10–200 | 0.15 | [268] |
ITO*/MoO3@rGO/APTES/antiHER2/BSA | Immunosensor | DPV | 0.001–500 | 0.001 | [269] |
GSPE*/AuNPs/antiHER2 affibody/MCH/HER2 | Affisensor | EIS | 0–4 × 104 | 6000 | [100] |
Ab2–AuNPs–dC20 AE*/peptide/MCH/HER2 | DNA-labeled sandwich immunosensor | SWV | 0.0001–1 | 0.0005 | [270] |
Ab2/Hyd@AuNPs–APTMS–Fe3O4 and GCE*/Fe3O4–APTMS/Ab1/BSA/HER2 | Sandwich immunosensor | DPV | 0.0005–50 | 2 × 10−5 | [271] |
PEG: polyethylene glycol; MagB: magnetic bead; CDI: carbonyldi-imidazole; MCH: 6-mercapto-1-hexanol; APTES: 3-aminopropyltriethoxysilane; APTMS: 3-aminopropyltrimethoxysilane; AE: gold electrode; | |||||
α-fetoprotein (AFP): Gastrointestinal tumor and liver cancer | |||||
GCE*/PGNR/AuNPs/antiAFP/BSA/AFP | Immunosensor | DPV | 5–60 | 1.0 | [272] |
Thiolated LAPS*/MPTES–AuNPs/Aptamer/AFP | Aptasensor | potentiometric | 100–1 × 105 | 92 | [273] |
AuNP–DNA2–MB and AE*/DNA1/MCH/Fc–CP/AFP/RecJf | Ratiometric aptasensor | ACV | 1 × 10−5–100 | 269.4 ag/mL | [274] |
AE*/AlCu MOF2.5,2.5/Aptamer/AFP | Aptasensor | EIS | 0.001–0.5 | 0.00012 | [275] |
Pd@PtNPs–Ab2– thionine and AE*/AuNPs/Ab1/BSA/AFP | Thionine-labeled sandwich immunosensor | DPV | 0.0001–100 | 0.000035 | [276] |
Ab2–AgNP–HRP and GCE*/PANI–AgNP/Ab1/BSA/AFP | HRP-labeled sandwich immunosensor | i-t | 0.01–1 and 1–10 | 0.0047 | [277] |
GCE*/ZnFe2O4–AgNP@rGO/antiAFP/BSA/AFP | Label-free immunosensor | CV | 0.001–200 | 0.00098 | [278] |
GCE*/Cu3Pt NPs/antiAFP/BSA/AFP | Immunosensor | DPV | 0.0001–10 | 0.000033 | [279] |
MO/CNT–AuNP–Ab2 and GCE*/VG–AuPt/Ab1/BSA/AFP | MO-labeled sandwich immunosensor | DPV | 1 × 10−6–100 | 7 × 10−7 | [280] |
MoSe2 NSs–NH2/Au@Pt DNRs–Ab2 and GCE*/GS–NH2/AuNPs/Ab1/BSA/AFP | Sandwich immunosensor | i-t | 1 × 10−5–200 | 3.3 × 10−6 | [281] |
PGNR: porous graphene nanoribbon; MPTES: 3-mercaptopropyltriethoxysilane; AE: gold electrode; BSA: bovine serum; MO: methyl orange; DNR: dendritic nanorods; | |||||
Interleukin-6 (IL-6): Colorectal cancer | |||||
MCH/Apt/AuNPs/pATP/pABA/GCE | Sandwich aptasensor | EIS | 0.005–100 | 0.0016 | [282] |
MCHApt/AuNPs/PPyNPs/SPGE | Structure switching aptasensor | EIS | 0.001–15000 | 0.00033 | [283] |
MCH: 6-mercapto-1-hexanol; Apt: aptamer; pATP: p-aminothiophenol; pABA: p-aminobenzoic acid; PPY: polypyrrole; | |||||
Interleukin-8 (IL-8): Oral cancer | |||||
BSA/Anti-IL 8/β–Ag2MoO4/ITO | Label-free immunosensor | DPV | 1 × 10−6–40 | 0.09 | [284] |
DNA-templated CdTe/CdS QDs/MB | Aptasensor | ASV | 1 × 10−6–0.005 | 3.36 × 10−6 | [285] |
Anti-IL8/AuNPs–rGO/ITO | Label-free immunosensor | DPV | 0.5–4 | 0.072 | [286] |
BSA: bovine serum; MB: methylene blue; ITO: indium tin oxide; |
4.2.1. MNPs in Carcinoma Embryonic Antigen Sensing
Carcinoma embryonic antigen (CEA) is a cell adhesive acidic glycoprotein with properties similar to the human embryonic cell. Normally, the level of CEA is around 5 µg/mL in serum, but in the blood the level is very low (<5 ng/mL) [195,196,197,198]. Meanwhile, serum CEA has been found to elevate up to 20 µg/mL in people with lung cancer [197]. Blood CEA levels above 10 ng/mL are indicative of cancer in the patient [201]. In several other types of carcinomas, such as breast cancer, ovarian cancer, pancreatic cancer, and gastrointestinal cancer, CEA often shows elevation in serum level, which indicates its potentiality as being a tumor marker for clinical cancer diagnosis [195,196,197]. Therefore, quantitative measurements of CEA in biological fluids such as blood and serum are critical for locating, and understanding the prognosis, staging, and recurrence of multiple cancers [197,198,199,200,201]. For electrochemical immunosensor/aptasensor assays, MNP-based probes which have strong biocompatibility and electrical conductivity are of great interest [201,202,203,204,205]. MNPs have superior efficiency as tags or labels for amplifying biomolecular interactions and as the enhancers of electrochemical signals [198,199,200,201,202,203]. Depending upon the complexity and necessities, mono/bi/tri-metallic composites are utilized for constructing different types of ECBs with label-free or labeled, sandwich or non-sandwich strategies for CEA immunosensor/aptasensor [195,196,199,207,210]. In mono-metallic-based ECBs, the primary concern is to enhance electrocatalytic surface area for anchoring biomolecules such as the antibody, and labeling enzymes or conductive dyes or biomolecules for developing labeled immunosensors [197,201]. Normally, direct immobilization of electrical signaling molecules such as HRP, MB, Fc to MNPs is not beneficial because it causes the loss of signaling molecules during electrochemical experiments and hence poor stability and reproducibility of the sensor [197]. To overcome this problem, Gu et al. constructed an Fc-labeled AuNP-based sandwich immunosensor assay for the highly sensitive electrochemical detection of CEA [197]. In this work, they introduced a thiol group (-SH) into an Fc molecule which assisted the stable chemisorption of Fc over AuNPs via initiating an S-Au covalent bond. After immobilization of Fc–SH, the colloidal Au nanoprobes were stabilized with PEG800. Figure 19a shows the transmission electron microscope (TEM) image of the nanostructured Fc–SH/AuNP–Ab2 composite along with the schematics of the fabrication process. Through this Fc-labeled immunosensing assay, they achieved a detection limit for CEA as low as 0.01 ng/mL [197]. Apart from AuNPs as the effective sensing platform and capture antibody, AgNPs also showed great promise in enhancing conductivity for electrocatalytic sensing and constructing immunosensors. Zhao et al. utilized a dual reduction signal amplification strategy based on AgNPs and MnO2 for constructing a sandwich immunosensing assay for CEA [201]. AgNPs and MnO2 together displayed catalytic activity for the reducing of H2O2 into H2O and molecular O2. At the same time, they utilized PANI, which acted as the sacrificial reducing agent for AgNP and as the base material for providing active sites for AgNP and MnO2 immobilization. Figure 19b shows the SEM images of the composites at different fabrication stages. This ultrasensitive dual amplifying sandwich immunosensor showed an LOD for CEA of about 0.17 pg/mL with a broad sensing range of 0.0005–80 ng/mL [201].
In bimetallic-based ECBs, synergistic interactions between multiple MNP components might be able to induce significant signal amplification, when compared with mono metal components [199,215]. In addition, the bimetallic counterpart shows enhanced photo-induced charge transfer properties and biocompatibility [199]. For instance, Song et al. constructed an MoS2/C3N4 (graphitic) composite supported with bimetallic Pt–Cu nano-dendrimers for visible light-induced amperometric sandwich immunosensing assay for CEA detection [199]. Graphitized carbon nitride usually has low conductivity due to its compact conduction band. However, by compositing with 2D MoS2, interactions between the conduction band of C3N4 and valence band of MoS2 enhanced the catalytic and charge transfer capacity of the support for bimetallic Pt–Cu NPs. The synergistic charge transfer/flow efficiency of the bimetallic counterparts along with the conductive support allowed for visible light-induced increment of the amperometric current signal for H2O2 reduction after the formation of the sandwich immunosensing array [199]. Figure 19c shows the synthesis process of the bimetallic ECB. In another work, monodisperse flower morphology-based Pt–Au NPs-supported rGO composite was prepared through gamma irradiation followed by microemulsion strategies for constructing a pulse voltammetric ultrasensitive CEA immunosensor [215]. Gamma irradiation ensured the simultaneous reduction of both GO and Pt–Au NPs, while water droplet-based microemulsion controlled the unique flower-like morphology and the size of the bimetallic Pt–Au NPs (<15 nm). The low LOD for CEA obtained using this immunosensor was only 7 fg/mL [215].
Trimetallic nanoparticles as the electrode probe are even more attractive in a sense of fast electron transfer affinity and great electrode stability when compared to the mono/bi-metallic counterparts [195]. Their finite geometrical orientation of the metallic components and HOMO (Highest occupied molecular orbital) -LUMO (Lowest unoccupied molecular orbital) distribution due to the formation of mixed bonds with the help of ligands enables formidable synergistic electrocatalytic activity. Barman et al. constructed a trimetallic Pd–Au–Pt NP-supported COOH functionalized rGO composite for developing an immunosensor for both CEA and PSA detection [195]. Figure 19d shows a schematic representation of the trimetallic sensor fabrication process. In this report, CV-assisted electrochemical deposition of trimetallic composites over COOH–rGO-modified gold electrode was implemented because the composition and morphology of the MNPs can be tailored by precisely controlling the concentration of the precursor metal salt solution, pH, scan rate, CV cycle, and deposition potential. The composite was successfully utilized to anchor the capture antibody and subsequently the antigen. The sensor delivered an ultrasensitive response toward CEA with an LOD of 8 pg/mL [195]. The authors emphasized that the sensor could be used for the POCT of CEA from human serum. Aside from MNP–carbon material-based composites, metal oxide NPs, core shell MNPs, quantum dot NPs, graphene, MOF, etc., have extensively been studied as the electrode probes for fabricating both labeled or label-free sandwich immunosensing arrays for the ultrasensitive detection of CEA for early-stage cancer diagnoses [196,202,209,211,212,213,214,215,216].
4.2.2. MNPs in Prostate-Specific Antigen Sensing
Prostate cancer (PSC) is one of the most common cancers for males with high mortality [219]. At present, prostate cancer can be treated by removal of the cancerous cells, but only if diagnosed in the early stages [219,220,221,222,223]. However, in the advanced stages, PSC cancer is lethal. Research into prostate cancer has shown that the early stage of the cancer is asymptomatic [223,224,225]. Hence, sensitive and selective detection of PSC biomarkers is most desired for early-stage detection. Prostate-specific antigens (PSAs) are one of the most reliable biomarkers for the early-stage detection of PSC [220,221]. Previous research has shown that a presence of PSA between 4–10 ng/mL is indicative of the possible PSC (>27%) risk [225]. If the PSA level is 10 ng/mL or above, then the risk is increased to greater than 67% [222,225]. As a result, the PSA biomarker is used for monitoring both the progression of the prostate cancer and mediating therapeutics [220]. Hence, significant research effort has been put into the development of ECBs for the detection of PSA.
PSA is a glycoprotein of around 34 kDa in molecular weight (MW) [223]. The low level of PSA makes it very hard to detect. As a result, researchers have explored different strategies for designing effective BRCs for PSA sensing. A highly sensitive biosensor was reported to utilize PdNPs along with conducting PANI and fullerene-C60 for PSA sensing [219]. The PANI–C60 combination worked towards activating the PdNPs through electrostatic interactions. This in turn allowed for the successful immobilization of the PSA on the sandwich type immunosensor. The sensor reported a promising linear range from 0.00016–38 ng/mL with a level of detection of 1.95 × 10−5 ng/mL. The sensor was also tested for the low detection of prostate cancer from serum and urine samples, with a recovery between 95–97% [219]. Another work utilized a screen-printing process to make prostate cancer sensors that could be used for practical applications [225]. The sensor utilized a GO and AgNP composite over screen-printed carbon electrodes (SPCEs) for the immobilization of PSAs. The label-free immunosensor offered a simple preparation electrode process compared to the complicated sandwich sensors. The biosensor utilized a “signal off” methodology for the PSA interaction with the antibody through the DPV technique. Despite the simple preparation process, the sensor showed a wide linear range from 0.75–100 ng/mL [225]. This indicated the feasibility of the screen-printed label-free sensor in POCT. A number of works have reported that bimetallic, MOF, QD, and core@shell structure further enhance the sensitivity and stability of ECBs [220,221,222,224,227,235]. However, as discussed in the previous sections, it is important to be able to prepare ECBs on strips so that they can be used for the on-the-spot testing of biomolecules. To this end, Chen and coworkers reported the fabrication of a PSA biosensor based on the microfluidic principles through screen printing [237]. The ultra-sensitive PSA sensor was prepared through screen printing, making it readily scalable and cost-effective. The sensor fabrication process is shown in Figure 20a. The proposed sensor used printed gold electrodes as the WE and CE, while an Ag electrode was used as the control. MagBs were utilized for anchoring the PSA antibody on the printed gold electrode. The sensor utilized an amperometric technique for the PSA detection from 0.001–10 ng/mL, with a low LOD of 0.00084 ng/mL. Authors claimed that the reported sensor was cheap, easy to fabricate and operate, highly reproducible, and extremely sensitive [237]. Such sensors could be the key to solving the problems associated with utilization of ECBs for the POCT.
4.2.3. MNPs in Cancer Antigen 125 Sensing
Ovarian cancer is one of the most commonly occurring cancers for woman and has high mortality rate [194,247]. The main reason behind the high mortality rate is the fact that early stages of ovarian cancer are usually asymptotic (stage I), and for later stages the symptoms are unspecified (stage II and III) [245]. A promising biomarker for the early-stage detection of ovarian cancer is cancer antigen 125 (CA125) [244,246]. CA125 is the recommended biomarker for ovarian cancer diagnosis by the FDA [194]. CA125 is a mucin-like transmembrane glycoprotein (200 kDa) that is overexpressed even during the early stages of ovarian cancer [245]. For a healthy person, the CA125 level is below 35 U/mL in blood and serum [245]. However, the level of CA125 in the blood and serum increases significantly for patients with ovarian cancer. That is why CA125 is known as a “Gold Standard” biomarker for ovarian cancer diagnoses [287]. CA125 overexpression is observed in about 50% of stage I patients, and about 90% overexpression is observed for stages II, III, and IV patients [244,245]. Therefore, a sensitive, selective, and accurate POCT system for CA125 is essential for the early-stage diagnosis and treatment of ovarian cancer.
The demand for early-stage detection of ovarian cancer has led to the development of a large number of ECBs for CA125. AuNP-modified biosensors have been the most reported for CA125 detection [244,246,248,249,251]. These works have utilized various CNMs for increasing the selectivity and sensitivity. For instance, AuNP, PAMAM, MWCNT, and rGO composites were utilized for immobilizing the antibody (Anb1) [244]. Here, MB was used to label the Anb2. This sandwich (Anb1–Ang–Anb2) system utilized the SWV technique for CA125 detection. The sensor showed a moderate linear range from 10–75 U/mL with a low LOD of 0.006 U/mL [244]. In another study, Huang et al. utilized an AgNP-MagBs-based sandwich immunosensor system for CA125 detection over a wide linear range of 0.001–76 U/mL [246]. The electrochemical cell setup utilized a magnet-controlled microfluidic flow system and the LSV technique for the detection process. The proposed sandwich system showed significant signal enhancement and selectivity towards CA125 detection [246]. Although these sensors have potential, it is often difficult to transition from laboratory to clinical applications. A paper-based strip ECB design was used to address this issue [251]. The proposed sensor utilized AuNP, rGO, and thionine composite as the substrate for immobilization of the CA125 antibody to fabricate the “signal off” ECB. The immunosensor utilized the affinity-based binding between the CA125 Anb–Ang to decrease the current signal produced by the thionine. The fabrication and detection of CA125 is shown in Figure 20b. As a result, the current signal decreased with increasing concentration of the CA125. The sensor showed a linear range from 0.01 to 200 U/mL for CA125 [251]. Development of these strip sensors might solve the problems associated with the clinical transition of ECBs.
4.2.4. MNPs in HER2 Sensing
One of the most prominent candidates for early-stage breast cancer detection is the HER2 [262,263]. This protein expression status is either positive or negative for the cancer [257]. It is often overexpressed in the early stages of breast cancer [262]. Breast cancer is one of the most frequently observed (~34%) cancers in all woman [263]. HER2 concentration in the blood of a healthy human body is about 2–15 ng/mL [288]. However, in a cancer patient the HER2 protein increases to 15–75 ng/mL [263]. The increase in HER2 concentration is significant, which makes it a prominent candidate for the early-stage cancer detection. However, the very low concentration of the biomarker makes it difficult for effective detection using conventional ECBs. This is why researchers have focused on using MNP-incorporated immunosensors, aptasensors, and affisensors for the sensitive and selective detection of HER2 [100,262,263,264,265].
Two ECBs that utilized AuNPs and CdSe@ZnS QDs were reported for the selective detection of HER2 [261,262]. The AuNP-modified electrode used a label-free immunosensor and the DPV technique [261]. The sensor showed a low LOD of 0.035 ng/mL, but the linear range was also very short, 0.001–20 ng/mL [261]. The CdSe@ZnS QD sensor had a longer linear range of 0.5–50 ng/mL, with an LOD of 0.29 ng/mL [262]. This ECB utilized functionalized MagBs for anchoring the BRC. The sensor could detect HER2 from the SK-BR-3 (an HER2-positive cell line) for only 2 cells/mL [262]. The detection process is shown in Figure 21a. Gold nanorods (GNRs) and Pd composite electrodes were proposed by Chen and coworkers for fabricating an HER2 biosensor with wide linear range [268]. The ECB used super structure aptamers for the detection of HER2. The sandwich type sensor mechanism pathway was followed, utilizing the DPV technique. A wide linear range from 10–200 ng/mL was obtained with an LOD of just 0.15 ng/mL [268]. Such an HER2 sensor could be used for POCT and diagnosis of breast cancer.
4.2.5. MNPs in Alpha Fetoprotein Sensing
Alpha fetoprotein is a glycoprotein that can be utilized for the early-stage diagnosis of hepatocellular carcinoma (HCC) [272,273]. HCC is one of the most common types of liver cancer that often leads to the death of the patient [273]. Alpha fetoprotein is about 70 kDa in MW, and is produced in the yolk sack, liver, and gastrointestinal tract [272,277]. During cancerous conditions, the alpha fetoprotein concentration is above 500 ng/mL in the human body, while in a healthy body it is only around 20 ng/mL [273,274]. This large abnormality in concentration makes alpha fetoprotein a significant marker for HCC detection. However, it is a challenging task to prepare biosensors for the detection of a low concentration analyte with a wide linear range.
An ECB was proposed that utilized an AuNP–GRN composite and a DPV technique for alpha fetoprotein detection [272]. The label-free immunosensor used Anb–alpha fetoprotein modification for the detection of alpha fetoprotein. It had a low LOD of 1 ng/mL and a linear range of 5–60 ng/mL. Though the LOD is low, the linear range was very small for the proposed senor for POCT of alpha fetoprotein in cancer patients [272]. To address the challenge of wide linear range, Li et al. developed an ECB utilizing AuNPs that had a wide linear range from 0.1 to 100 µg/mL [273]. However, the LOD was 92 ng/mL, which indicates that this sensor could not be used for the low-level detection of alpha fetoprotein. A label-free immunosensor was developed using AgNPs and an rGO composite along with ZnFe2O4 for the sensitive detection of alpha fetoprotein [278]. The signal amplification of the sensor enabled the CV technique to be used for detecting the pg/mL analyte. The alpha fetoprotein sensor development process and consequent detection mechanism is shown in Figure 21b. The sensor showed a very low LOD of 0.98 pg/mL, with a linear range in the region of 0.001–200 ng/mL [278]. A possible way to improve alpha fetoprotein sensors can be considering the size- and shape-dependent properties of MNPs. At the same time, utilizing different CNMs for preparing MNPCs is a promising way for further improving the biosensing capability.
4.2.6. MNPs in Interleukin Sensing
The human genome has about 50 different small proteins or cytokines that are responsible for maintaining important immunomodulatory responses [289]. Hence, interleukin (IL) is a term that describes this group of cytokines that are capable of being signaling cells. Of the various IL species, in this section the focus will be on IL-6 and IL-8, which are important biomarkers for colorectal and oral cancers, respectively [282,284]. IL-6 contains 184 amino acids that iss about 26 kDa in MW [284]. This glycoprotein is responsible for signaling cancer cells, and overexpression is related to colorectal cancer [282]. On the other hand, the MW of IL-8 is about 8.5 kDa, with its 70 amino acids [284]. Enzyme-linked immunosorbent assay (ELISA) and a few other methods are commercially available for the detection of these biomarkers [285,290]. However, one ELSIA kit costs about USD 320 [291], whereas ECBs can be prepared for much less. Sandwich and structure switching type aptasensors were used for the sensitive and selective detection of IL-6 [282,283], utilizing AuNPs and the impedimetric technique for the detection process. These sensors had an impressive LOD of 0.0016 and 0.00033 ng/mL for IL-6 [282,283]. These reports demonstrate the significance of using MNPs in ECBs. There are various ECB designs for the detection of IL-8. Normally, only 250 pg/mL of IL-8 is present in a healthy person’s saliva [285,286]. IL-8 expression higher than 750 pg/mL is indicative of oral cancer [285,286]. ECBs for IL-8 detection have used Ag2MoO4 and CdTe/CdS QDs, besides AuNPs [285,286]. These multimetallic NPs with a controlled size and unique shapes often perform better than spherical monometallic systems. The multimetallic ECB fabrication is shown in Figure 21c. However, the use of EATs and other supporting ingredients of the biosensors also play an important role in the overall performance of the sensor. The DNA templated CdTe/CdS QD sensor had a very low LOD of 3.36 × 10−6 ng/mL for IL-8 detection [285]. The AuNP sensor with the setup of anti-IL8/AuNPs–rGO/ITO showed an LOD of 0.072 ng/mL [286]. Both of these sensors are well within the range necessary for oral cancer patient identification through IL-8 biomarkers [285,286].
4.3. MNPs in Novel Coronavirus Sensing
The 2019 novel coronavirus (2019-nCoV) is responsible for the COVID-19 pandemic of the year 2020 [292]. It is also known as severe acute respiratory syndrome coronavirus 2 [8]. It is a ribonucleic virus, i.e., an RNA virus that is known to infect and attack various parts of the body, but causes most damage to the respiratory and cardiovascular system [292,293]. Besides economic impact, the long-term lockdown due to COVID-19 will have a severe mental health impact on both individuals and society as a whole [294]. One of the proven and most effective way to control the spread and minimize the 2019-nCoV impact is through testing. Asymptomatic and pre-asymptomatic individuals infected with 2019-nCoV are also highly contagious [292]. As a result, researchers around the globe have invested their time and knowledge in developing POCT systems that can be used for rapid, accurate, and early detection of the virus [8,292].
ECBs can be used to reach all these goals in a cost-effective manner [293]. There are various proposals for the synthesis of ECB strips that can effectively detect 2019-nCoV [294,295]. Smartphone-based ECBs for ultrasensitive detection of 2019-nCoV were reported by Zhao et al [296]. The reported aptasensor utilized a super-sandwich setup for the detection of 2019-nCoV through a “signal on” process with the DPV technique. The sensor fabrication process and its 2019-nCoV detection process is shown in Figure 22a. For the sandwich setup, initially, thiolated capture probes (CPs) were immobilized on the surface of the Au@Fe3O4 NPs (Premix A). Calixarene8 (CX8) was used for enhancing the electrochemical activity of TB through a supramolecular interaction process. Essentially, the host–guest complex utilized AuNPs, CX8, TB, LP (label probe) and an rGO system. Finally, the immobilized host–guest system was combined with the AP (auxiliary probe) to make the final modifications (Premix B). After extraction, 2019-nCoV RNA was first incubated with Premix A for 1 h, followed by 2 h incubation with Premix B. The sensor showed a significant increase in DPV current signal for the 2019-nCoV RNA combined Premix A and B setup compared to just the Premix B setup. The reported sensor was also tested on 2019-nCoV active and recovering patients. The proposed sensor showed higher effectiveness compared to the reverse transcription real-time polymerase chain reaction (RT-qPCR) for 2019-nCoV detection from both active and recovering patients. The sensor can be used with a smartphone, in a plug-and-play system for the effective POCT diagnosis of 2019-nCoV [296]. Once commercialized, these portable ECBs could be highly effective in contact tracing and controlling the spread of 2019-nCoV.
5. Advances in POCT Devices: Prospects and Challenges in the Clinical Transition of ECBs
POCT provides health experts and patients with the opportunity to monitor health conditions and diagnose a disease quickly and accurately. Furthermore, the introduction of personalized heath care would be possible with the large-scale implementation of POCT [6,8,12]. This would allow for early-stage detection of cancer biomarkers, senescent β-cells in type-I diabetes patients, or beta-amyloids in Alzheimer’s disease patients [65,297,298]. Because of this, research in ECBs has focused on developing prototype biosensor devices that can be used with mobile or other convenient electronic systems for the rapid analysis of biomolecules [6,65]. Wearable paper ECBs were reported for the detection of sulfur mustard that used EmStat3, a portable potentiostat [12]. A glucose sensor was developed that could be used with a smartphone for blood glucose level detection [298]. Castro et al. reported the fabrication of label-free immunosensor strips for the sensitive detection of CA125 cancer biomarkers from human serum sample [71]. An ECB device has been reported for the rapid detection of the COVID-19 virus, that could be used as an alternative to the RT-qPCR-based 2019-nCoV test [296]. Figure 23 shows the fabrication and detection mechanism of these sensors.
ECBs fall within of one of the five EATs that have been discussed in the Section 2.1. These techniques utilize Ohm’s law, the Nernst equation, and other well-established theories of electrochemistry [82]. For instance, the commercial glucose sensors utilize a biocatalytic process for blood glucose detection through voltammetry or amperometry techniques. There are ECBs that have been developed to communicate with smartphones through micro-USB ports, audio channels, or even wirelessly [82]. Despite the increasing research and steady development in POCT electrochemical devices, there are almost no ECBs in the market for consumer use other than the glucose sensor for diabetes patients [48]. These devices are often used only in laboratory testing and do not progress towards clinical trials. Long trial times for evaluation and lack of funding are some of the reasons for such shortcomings. ECBs utilize enzymes, antibodies, proteins, peptides, and many other biomolecules as BRCs. All of these have their advantages and limitations. The storage stability, data reproducibility, and complicated sensor fabrication process are some of the key issues that has hindered the clinical transition of ECBs [49,50,51]. Discussion of Table 1 on the MNP-based ECBs shows that despite the use of diverse biomolecules, there still lacks the development of size- and shape-controlled MNPs; there is very little use of affibody molecules. There are more than 353,000 research studies listed on the clinicaltrials.gov website. However, when we searched for the term “electrochemical biosensor”, only three studies were found. Of these, two were ongoing and one study had been completed. The clinical trial identification number for the completed study is NCT00591240, and they published two reports based on their findings [299,300]. The completed study was conducted with the help of 116 patients that utilized ECBs for the detection of pathogens that cause urinary infections [300]. The sandwich mechanism was employed by the sensor for simultaneous detection of multiple bacterial species (E. coli, Enterococcus, etc.). Figure 23c shows the fabrication process and depth effect in the ECB. The researchers compared their ECB with other urine cultures and concluded that the ECB design needed further modification to improve the LOD and specificity [300].
There are reports of ECBs for detecting bacterial systems that showed better performance due to the incorporation of MNPCs for anchoring the biorecognition component [31,301]. In one of the ongoing studies (NCT04053140), researchers were using microneedle array-ECBs for administrating Benzylpenicillin IV 1200 mg. The work was in Phase 1 trial, and had not published any results at time of writing. The other clinical trial was studying the ECB system towards detection of leukocyte esterase biomarkers for periprosthetic joint infection (PJI) (NCT04390607). The studies were aiming to test the ECB on subjects that underwent revision joint surgery. The study was set to begin from November 2020. From Table 2 and Table 3, it becomes clear that using MNPs can significantly enhance the performance, stability, and reproducibility of ECBs, and lower the production cost at the same time. Hence, thorough investigation of the impact of MNP properties on the design and effectiveness of ECBs might help to realize the POCT in the near future [297,302]. The following key points might be inferred based on our analysis of more than 250 articles on ECBs that utilized MNPs for improving their practical applicability:
ECBs that utilize MNPs or MNPCs usually show higher sensitivity, stability, and wider linear range compared to those that do not utilize MNPs. However, it is important to choose which MNPs are most compatible with a specific BRC. Therefore, research into MNP and BRC compatibility might greatly enhance the stability of the fabricated ECBs.
Using bi- or tri-metallic NPs significantly enhances the performance of ECBs. The possible reason for this is that the metal–metal interaction helps in lowering the HOMO-LUMO energy gap. This in turn allows for more active sites on the MNPs. Hence, using multimetallic NPs that interact synergistically with each other will allow for stronger interactions with the BRCs.
MNPs with cubic, pyramidal, oval, and other unique shapes show higher catalytic activity and have an increased surface area compared to the commonly employed spherical MNPs. This would allow for immobilization of a greater amount of BRCs. At the same time, the edge sites of these MNPs show higher activity compared to the basal sites.
MNPs that have QD or core shell-like structure show some unique physical, chemical, and electronic properties. These unique properties usually make them highly desirable for fabrication of ECBs.
Green synthesis of MNPs is becoming ever popular. This not only allows for the preparation of MNPs in an environmentally friendly way, but also introduces various functional groups on the MNP surface. These functional groups, when properly utilized, might help in the robust anchoring of BRCs and enhance the stability and overall activity of the ECBs.
MNPs interact differently with various CNMs. Hence, it is essential to properly choose the MNPs and CNMs before composites can be prepared for fabricating effective ECBs. Future research should focus on understanding the fundamental properties of various MNPCs. This would allow for the intelligent designing of MNPCs for fabricating ECBs.
The screen-printing technique is most commonly used in the fabrication of ECB strips. However, methods such as inkjet printing, doctor blading, and aerosol-assisted chemical vapor deposition should be explored for determining the best approach for the fabrication of ECB strips.
Aside from the above-mentioned topics, ECB researchers should work towards the commercialization of their laboratory models. This would then reveal the limitations of their proposed systems, and make way for future research that would help to overcome these shortcomings.
The authors hope that this review will help researchers to see the importance of the proper utilization of the various properties of MNPs in developing effective ECBs. Furthermore, this would allow the fabrication of cost-effective ECBs with high stability and accuracy for the POCT of small biomolecules, cancer biomarkers, and other pathogenic diseases.
Acknowledgments
We acknowledge the research grant supported by the Lizanell and Colbert Coldwell Foundation.
Abbreviations
ADHD | Attention deficit hyperactivity disorder |
AFP | Alpha-fetoprotein |
Anb | Antibody |
AgNPs | Silver NPs |
Anb | Antibody |
Ang | Antigen |
AP | Auxiliary probe |
APTES | 3-aminopropyltriethoxysilane |
APTMS | 3-aminopropyltrimethoxysilane |
ATPA | ATP aptamer |
ASV | anodic stripping voltammetry |
AuNPs | Gold NPs |
BDD | Boron-doped diamond |
BRC | biorecognition component |
BSA | bovine serum albumin |
CAS | Casein |
CASPE-MFD | Commercially available screen-printed electrode-Based microfluidic devices |
CDI | Carbonyldi-imidazole |
cDNA | Complimentary DNA |
CE | Counter electrode |
CEA | Carcinoembryonic antigen |
CNMs | Conducting nanomaterials |
CNT | Carbon nanotube |
conA | Concanavalin A |
CP | Capture probe |
CPE | Carbon paste electrode |
CS | Chitosan |
CSH | Copper silicate hollow spheres |
CV | cyclic voltammetry |
CysA | Cysteamine |
DA | Dopamine |
DMF | Dimethylformamide |
DNR | Dendritic nanorods |
DPV | differential pulse voltammetry |
DPASV | differential pulse anodic stripping voltammetry |
3D-SP | 3D screen printed |
EATs | electroanalytical techniques |
ECBs | Electrochemical biosensors |
ECD | Extracellular domain |
EDC | N-(3,Dimethylaminopropyl)-N-ethyl-Carbodiimidehydrochloride |
EIS | electrochemical impedance spectroscopy |
ELISA | Enzyme-linked immunosorbent assay |
ESR | Electron spin resonance |
FA | Ferro-cenecarboxylic acid |
FAD | Flavin adenosine dinucleotide |
FC | Ferrocene |
FLGR | few layer GR |
fMWCNT | Functionalized MWCNTs |
Glu | Glutaraldehyde |
GNR | Gold nanorods |
GNS | Gold Nanostructures |
GO | GR oxide |
GOx | glucose oxidase |
GPTMS | Glycidyloxypropyl trimethoxysilane |
GQD | Graphene quantum dot |
GR | Graphene |
GRNs | graphene nanoribbons |
GRWs | GR nanowalls |
HA | Histamine |
HCC | Hepatocellular carcinoma |
HepG2 | Hepatocellular carcinoma |
HER2 | Human epidermal growth factor receptor 2 |
HEV | Hepatitis E virus |
HNF | Honey nanofibers |
HNT | Halloysite nanotube |
HP | Hairpin probe |
HRP | horseradish peroxidase |
HUA | Humic acid |
IL | Interleukin |
ISE | ion selective electrode |
ITO | Indium tin oxide |
LBL | Layer-by-layer |
LIG | laser-induced graphene |
LOD | limit of detection |
LOx | Lactate oxidase |
LP | Label probe |
LPA-NHS | Lipoic acid N-hydroxysuccinimide ester |
LSV | linear sweep voltammetry |
16-MUA | 16-Mercaptoundecanoic acid |
MagB | Magnetic bead |
MB | Methylene blue |
MCH | 6-mercapto-1-hexanol |
MLGR | multilayer GR |
MNPs | Metal nanoparticles |
MNPCs | MNP composites |
MO | Methyl orange |
MOF | Metal organic framework |
MPA | 3-mercaptopropionic acid |
MPTES | 3-mercaptopropyltriethoxysilane |
MSA | Mercaptosuccinic acid |
MWCNT | Multiwall CNT |
2019-nCoV | Novel coronavirus |
NCMTs | Nitrogen-doped magnetic carbon microtubes |
NF | Nanoflower |
OC | Ovarian cancer |
OHA | Ochratoxin A |
pABA | p-aminobenzoic acid |
PAMAM | Poly(amidoamine) |
PAN-oxime | Amidoxime-modified polyacrylonitrile |
PANI | Polyaniline |
pATP | p-aminothiophenol |
PB | Prussian blue |
pCTAB | Poly cetyl trimethylammoniumbromide |
PDA | Poly DA |
PEDOT | Poly(3,4-ethylenedioxythiophene) |
PEG | Polyethylene glycol |
PGNR | porous graphene nanoribbon |
PJI | Periprosthetic joint infection |
PLA | Polylactic |
POCT | Point-of-care testing |
PGA | Poly(glutamic acid) |
PPY | Polypyrrole |
PS | Prostate cancer |
PSA | Prostate specific antigen |
PTK-7 | Protein tyrosine kinase-7 |
PtNPs | Platinum NPs |
PVA | Polyvinyl alcohol |
RE | Reference electrode |
rGO | Reduced GO |
SBMs | Small biomolecules |
SLGR | single layer GR |
SPCE | Screen-printed carbon electrode |
ssDNA | Single stranded DNA |
ssRNA | Single stranded RNA |
Strp | Streptavidin |
SWCNT | Single wall CNT |
SWV | square wave voltammetry |
TB | Toluidine blue |
TBA | Thrombin binding aptamer |
TMV | tobacco mosaic virus |
UA | Uric acid |
UOx | Uricase |
VBG | Vertical boron doped graphene |
WE | Working electrode |
XO | Xanthine oxidase |
Funding
This research received no external funding.
Conflicts of Interest
The authors declare no conflict of interest.
Footnotes
Publisher’s Note: MDPI stays neutral with regard to jurisdictional claims in published maps and institutional affiliations.
References
- 1.Ronkainen N.J., Halsall H.B., Heineman W.R. Electrochemical biosensors. Chem. Soc. Rev. 2010;39:1747–1763. doi: 10.1039/b714449k. [DOI] [PubMed] [Google Scholar]
- 2.Ouyang M., Tu D., Tong L., Sarwar M., Bhimaraj A., Li C., Coté G.L., Di Carlo D. A review of biosensor technologies for blood biomarkers toward monitoring cardiovascular diseases at the point of care. Biosens. Bioelectron. 2020;171:112621. doi: 10.1016/j.bios.2020.112621. [DOI] [PubMed] [Google Scholar]
- 3.Zhang N., Ruan Y.F., Zhang L.B., Zhao W.W., Xu J.J., Chen H.Y. Nanochannels Photoelectrochemical Biosensor. Anal. Chem. 2018;90:2341–2347. doi: 10.1021/acs.analchem.7b04862. [DOI] [PubMed] [Google Scholar]
- 4.Chowdhury A.D., Takemura K., Li T.C., Suzuki T., Park E.Y. Electrical pulse-induced electrochemical biosensor for hepatitis E virus detection. Nat. Commun. 2019;10:4–7. doi: 10.1038/s41467-019-11644-5. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 5.Dai Y., Somoza R.A., Wang L., Welter J.F., Li Y., Caplan A.I., Liu C.C. Exploring the Trans-Cleavage Activity of CRISPR-Cas12a (cpf1) for the Development of a Universal Electrochemical Biosensor. Angew. Chem. 2019;131:17560–17566. doi: 10.1002/ange.201910772. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 6.Chen L.C., Wang E., Tai C.S., Chiu Y.C., Li C.W., Lin Y.R., Lee T.H., Huang C.W., Chen J.C., Chen W.L. Improving the reproducibility, accuracy, and stability of an electrochemical biosensor platform for point-of-care use. Biosens. Bioelectron. 2020;155:112111. doi: 10.1016/j.bios.2020.112111. [DOI] [PubMed] [Google Scholar]
- 7.Gill A.A.S., Singh S., Nate Z., Chauhan R., Thapliyal N.B., Karpoormath R., Maru S.M., Reddy T.M. A novel copper-based 3D porous nanocomposite for electrochemical detection and inactivation of pathogenic bacteria. Sens. Actuators B Chem. 2020;321:128449. doi: 10.1016/j.snb.2020.128449. [DOI] [Google Scholar]
- 8.Khan M.Z.H., Hasan M.R., Hossain S.I., Ahommed M.S., Daizy M. Ultrasensitive detection of pathogenic viruses with electrochemical biosensor: State of the art. Biosens. Bioelectron. 2020;166:112431. doi: 10.1016/j.bios.2020.112431. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9.Ahammad A.J.S., Islam T., Hasan M.M. Biomedical Applications of Graphene and 2D Nanomaterials. Elsevier Inc.; Amsterdam, The Netherlands: 2019. Graphene-Based Electrochemical Sensors for Biomedical Applications; pp. 249–282. [Google Scholar]
- 10.Islam T., Hasan M.M., Akter S.S., Alharthi N.H., Karim M.R., Aziz M.A., Hossain M.D., Ahammad A.J.S. Fabrication of Ni–Co-Based Heterometallo-Supramolecular Polymer Films and the Study of Electron Transfer Kinetics for the Nonenzymatic Electrochemical Detection of Nitrite. ACS Appl. Polym. Mater. 2020;2:273–284. doi: 10.1021/acsapm.9b00797. [DOI] [Google Scholar]
- 11.Chang J., Wang X., Wang J., Li H., Li F. Nucleic Acid-Functionalized Metal-Organic Framework-Based Homogeneous Electrochemical Biosensor for Simultaneous Detection of Multiple Tumor Biomarkers. Anal. Chem. 2019;91:3604–3610. doi: 10.1021/acs.analchem.8b05599. [DOI] [PubMed] [Google Scholar]
- 12.Colozza N., Kehe K., Dionisi G., Popp T., Tsoutsoulopoulos A., Steinritz D., Moscone D., Arduini F. A wearable origami-like paper-based electrochemical biosensor for sulfur mustard detection. Biosens. Bioelectron. 2019;129:15–23. doi: 10.1016/j.bios.2019.01.002. [DOI] [PubMed] [Google Scholar]
- 13.Ahammad A.J.S., Alam M.K., Islam T., Hasan M.M., Karim R., Anju A.N., Mozumder M.N.I. Poly (brilliant cresyl blue)-reduced graphene oxide modified activated GCE for nitrite detection: Analyzing the synergistic interactions through experimental and computational study. Electrochim. Acta. 2020;349:136375. doi: 10.1016/j.electacta.2020.136375. [DOI] [Google Scholar]
- 14.Zhou L., Wang Y., Yang C., Xu H., Luo J., Zhang W., Tang X., Yang S., Fu W., Chang K., et al. A label-free electrochemical biosensor for microRNAs detection based on DNA nanomaterial by coupling with Y-shaped DNA structure and non-linear hybridization chain reaction. Biosens. Bioelectron. 2019;126:657–663. doi: 10.1016/j.bios.2018.11.028. [DOI] [PubMed] [Google Scholar]
- 15.Sun A.C., Hall D.A. Point-of-Care Smartphone-based Electrochemical Biosensing. Electroanalysis. 2019;31:2–16. doi: 10.1002/elan.201800474. [DOI] [Google Scholar]
- 16.Haque M.A., Hasan M.M., Islam T., Razzak M.A., Alharthi N.H., Sindan A., Karim M.R., Basha S.I., Aziz M.A., Ahammad A.J.S. Hollow Reticular Shaped Highly Ordered Rice Husk Carbon for the Simultaneous Determination of Dopamine and Uric Acid. Electroanalysis. 2020:1–15. doi: 10.1002/elan.202060059. [DOI] [Google Scholar]
- 17.Chellapandian C., Ramkumar B., Puja P., Shanmuganathan R., Pugazhendhi A., Kumar P. Gold nanoparticles using red seaweed Gracilaria verrucosa: Green synthesis, characterization and biocompatibility studies. Process Biochem. 2019;80:58–63. doi: 10.1016/j.procbio.2019.02.009. [DOI] [Google Scholar]
- 18.Kang M.S., Lee S.Y., Kim K.S., Han D.W. State of the art biocompatible gold nanoparticles for cancer theragnosis. Pharmaceutics. 2020;12:701. doi: 10.3390/pharmaceutics12080701. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 19.Baygar T., Sarac N., Ugur A., Karaca I.R. Antimicrobial characteristics and biocompatibility of the surgical sutures coated with biosynthesized silver nanoparticles. Bioorg. Chem. 2019;86:254–258. doi: 10.1016/j.bioorg.2018.12.034. [DOI] [PubMed] [Google Scholar]
- 20.Eanest Jebasingh B., Valan Arasu A. A comprehensive review on latent heat and thermal conductivity of nanoparticle dispersed phase change material for low-temperature applications. Energy Storage Mater. 2020;24:52–74. doi: 10.1016/j.ensm.2019.07.031. [DOI] [Google Scholar]
- 21.Agarwal H., Nakara A., Shanmugam V.K. Anti-inflammatory mechanism of various metal and metal oxide nanoparticles synthesized using plant extracts: A review. Biomed. Pharmacother. 2019;109:2561–2572. doi: 10.1016/j.biopha.2018.11.116. [DOI] [PubMed] [Google Scholar]
- 22.Nurunnabi M., Nafiujjaman M., Lee S.J., Park I.K., Huh K.M., Lee Y.K. Preparation of ultra-thin hexagonal boron nitride nanoplates for cancer cell imaging and neurotransmitter sensing. Chem. Commun. 2016;52:6146–6149. doi: 10.1039/C5CC10650H. [DOI] [PubMed] [Google Scholar]
- 23.Han S., Liu W., Zheng M., Wang R. Label-Free and Ultrasensitive Electrochemical DNA Biosensor Based on Urchinlike Carbon Nanotube-Gold Nanoparticle Nanoclusters. Anal. Chem. 2020;92:4780–4787. doi: 10.1021/acs.analchem.9b03520. [DOI] [PubMed] [Google Scholar]
- 24.Lv Q., Wang Y., Su C., Lakshmipriya T., Gopinath S.C.B., Pandian K., Perumal V., Liu Y. Human papilloma virus DNA-biomarker analysis for cervical cancer: Signal enhancement by gold nanoparticle-coupled tetravalent streptavidin-biotin strategy. Int. J. Biol. Macromol. 2019;134:354–360. doi: 10.1016/j.ijbiomac.2019.05.044. [DOI] [PubMed] [Google Scholar]
- 25.Yuan Q., Liu Y., Ye C., Sun H., Dai D., Wei Q., Lai G., Wu T., Yu A., Fu L., et al. Highly stable and regenerative graphene–diamond hybrid electrochemical biosensor for fouling target dopamine detection. Biosens. Bioelectron. 2018;111:117–123. doi: 10.1016/j.bios.2018.04.006. [DOI] [PubMed] [Google Scholar]
- 26.Xu S., Huang X., Chen Y., Liu Y., Zhao W., Sun Z., Zhu Y., Liu X., Wong C.P. Silver Nanoparticle-Enzyme Composite Films for Hydrogen Peroxide Detection. ACS Appl. Nano Mater. 2019;2:5910–5921. doi: 10.1021/acsanm.9b01346. [DOI] [Google Scholar]
- 27.Nguyen T.N.H., Nolan J.K., Park H., Lam S., Fattah M., Page J.C., Joe H.E., Jun M.B.G., Lee H., Kim S.J., et al. Facile fabrication of flexible glutamate biosensor using direct writing of platinum nanoparticle-based nanocomposite ink. Biosens. Bioelectron. 2019;131:257–266. doi: 10.1016/j.bios.2019.01.051. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 28.Parate K., Karunakaran C., Claussen J.C. Electrochemical cotinine sensing with a molecularly imprinted polymer on a graphene-platinum nanoparticle modified carbon electrode towards cigarette smoke exposure monitoring. Sens. Actuators B Chem. 2019;287:165–172. doi: 10.1016/j.snb.2019.02.032. [DOI] [Google Scholar]
- 29.Hosseini G.M., Mirzaei M., Torkzadeh-Mahani M. Electrochemical aptasensor for activated protein C using a gold nanoparticle—Chitosan/graphene paste modified carbon paste electrode. Bioelectrochemistry. 2019;130:107322. doi: 10.1016/j.bioelechem.2019.06.007. [DOI] [PubMed] [Google Scholar]
- 30.Hou L., Huang Y., Hou W., Yan Y., Liu J., Xia N. Modification-free amperometric biosensor for the detection of wild-type p53 protein based on the in situ formation of silver nanoparticle networks for signal amplification. Int. J. Biol. Macromol. 2020;158:580–586. doi: 10.1016/j.ijbiomac.2020.04.271. [DOI] [PubMed] [Google Scholar]
- 31.Wang L., Hao L., Qi W., Huo X., Xue L., Liu Y., Zhang Q., Lin J. A sensitive Salmonella biosensor using platinum nanoparticle loaded manganese dioxide nanoflowers and thin-film pressure detector. Sens. Actuators B Chem. 2020;321:128616. doi: 10.1016/j.snb.2020.128616. [DOI] [Google Scholar]
- 32.Roach K.A., Stefaniak A.B., Roberts J.R. Metal nanomaterials: Immune effects and implications of physicochemical properties on sensitization, elicitation, and exacerbation of allergic disease. J. Immunotoxicol. 2019;16:87–124. doi: 10.1080/1547691X.2019.1605553. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 33.Jamkhande P.G., Ghule N.W., Bamer A.H., Kalaskar M.G. Metal nanoparticles synthesis: An overview on methods of preparation, advantages and disadvantages, and applications. J. Drug Deliv. Sci. Technol. 2019;53:101174. doi: 10.1016/j.jddst.2019.101174. [DOI] [Google Scholar]
- 34.Gherab K., AI-Douri Y., Hashim U., Ameri M., Bouhemadou A., Batoo K.M., Adil S.F., Khan M., Raslan E.H. Fabrication and characterizations of Al nanoparticles doped ZnO nanostructures-based integrated electrochemical biosensor. J. Mater. Res. Technol. 2020;9:857–867. doi: 10.1016/j.jmrt.2019.11.025. [DOI] [Google Scholar]
- 35.Asadian E., Ghalkhani M., Shahrokhian S. Sensors and Actuators B: Chemical Electrochemical sensing based on carbon nanoparticles: A review. Sens. Actuators B. Chem. 2019;293:183–209. doi: 10.1016/j.snb.2019.04.075. [DOI] [Google Scholar]
- 36.Shetti N.P., Bukkitgar S.D., Reddy K.R., Reddy C.V., Aminabhavi T.M. Nanostructured titanium oxide hybrids-based electrochemical biosensors for healthcare applications. Colloids Surf. B Biointerfaces. 2019;178:385–394. doi: 10.1016/j.colsurfb.2019.03.013. [DOI] [PubMed] [Google Scholar]
- 37.Zhu L., Miao M., Shao X., Du Z., Huang K., Luo Y., Xu W. A Universal Electrochemical Biosensor Using Nick-HCR Nanostructure as Molecular Gate of Nanochannel for Detecting Chromium(III) Ions and MicroRNA. Anal. Chem. 2019;91:14992–14999. doi: 10.1021/acs.analchem.9b03489. [DOI] [PubMed] [Google Scholar]
- 38.Dimcheva N. Nanostructures of noble metals as functional materials in biosensors. Curr. Opin. Electrochem. 2020;19:35–41. doi: 10.1016/j.coelec.2019.09.008. [DOI] [Google Scholar]
- 39.Chen X., Huang J., Zhang S., Mo F., Su S., Li Y., Fang L., Deng J., Huang H., Luo Z., et al. Electrochemical Biosensor for DNA Methylation Detection through Hybridization Chain-Amplified Reaction Coupled with a Tetrahedral DNA Nanostructure. ACS Appl. Mater. Interfaces. 2019;11:3745–3752. doi: 10.1021/acsami.8b20144. [DOI] [PubMed] [Google Scholar]
- 40.Li Z., Wang L., Li Y., Feng Y., Feng W. Carbon-based functional nanomaterials: Preparation, properties and applications. Compos. Sci. Technol. 2019;179:10–40. doi: 10.1016/j.compscitech.2019.04.028. [DOI] [Google Scholar]
- 41.Du X., Zhang Z., Zheng X., Zhang H., Dong D., Zhang Z., Liu M., Zhou J. An electrochemical biosensor for the detection of epithelial-mesenchymal transition. Nat. Commun. 2020;11:192. doi: 10.1038/s41467-019-14037-w. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 42.Ayaz S., Karakaya S., Emir G., Dilgin D.G., Dilgin Y. A novel enzyme-free FI-amperometric glucose biosensor at Cu nanoparticles modified graphite pencil electrode. Microchem. J. 2020;154:104586. doi: 10.1016/j.microc.2019.104586. [DOI] [Google Scholar]
- 43.Hondred J.A., Medintz I.L., Claussen J.C. Enhanced electrochemical biosensor and supercapacitor with 3D porous architectured graphene via salt impregnated inkjet maskless lithography. Nanoscale Horiz. 2019;4:735–746. doi: 10.1039/C8NH00377G. [DOI] [Google Scholar]
- 44.Baek S.H., Roh J., Park C.Y., Kim M.W., Shi R., Kailasa S.K., Park T.J. Cu-nanoflower decorated gold nanoparticles-graphene oxide nanofiber as electrochemical biosensor for glucose detection. Mater. Sci. Eng. C. 2020;107:110273. doi: 10.1016/j.msec.2019.110273. [DOI] [PubMed] [Google Scholar]
- 45.Lu J., Hu Y., Wang P., Liu P., Chen Z., Sun D. Electrochemical biosensor based on gold nanoflowers-encapsulated magnetic metal-organic framework nanozymes for drug evaluation with in-situ monitoring of H2O2 released from H9C2 cardiac cells. Sens. Actuators B Chem. 2020;311:127909. doi: 10.1016/j.snb.2020.127909. [DOI] [Google Scholar]
- 46.Gattani A., Singh S.V., Agrawal A., Khan M.H., Singh P. Recent progress in electrochemical biosensors as point of care diagnostics in livestock health. Anal. Biochem. 2019;579:25–34. doi: 10.1016/j.ab.2019.05.014. [DOI] [PubMed] [Google Scholar]
- 47.Mohanraj J., Durgalakshmi D., Rakkesh R.A., Balakumar S., Rajendran S., Karimi-Maleh H. Facile synthesis of paper based graphene electrodes for point of care devices: A double stranded DNA (dsDNA) biosensor. J. Colloid Interface Sci. 2020;566:463–472. doi: 10.1016/j.jcis.2020.01.089. [DOI] [PubMed] [Google Scholar]
- 48.Newman J.D., Turner A.P.F. Home blood glucose biosensors: A commercial perspective. Biosens. Bioelectron. 2005;20:2435–2453. doi: 10.1016/j.bios.2004.11.012. [DOI] [PubMed] [Google Scholar]
- 49.Zhao C., Thuo M.M., Liu X. Erratum: A microfluidic paper-based electrochemical biosensor array for multiplexed detection of metabolic biomarkers (Science and Technology of Advanced Materials (2013) 14 (054402)) Sci. Technol. Adv. Mater. 2015;16:054402. doi: 10.1088/1468-6996/16/4/049501. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 50.Liu J., Siavash Moakhar R., Sudalaiyadum Perumal A., Roman H.N., Mahshid S., Wachsmann-Hogiu S. An AgNP-deposited commercial electrochemistry test strip as a platform for urea detection. Sci. Rep. 2020;10:9527. doi: 10.1038/s41598-020-66422-x. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 51.Loncaric C., Tang Y., Ho C., Parameswaran M.A., Yu H.Z. A USB-based electrochemical biosensor prototype for point-of-care diagnosis. Sens. Actuators B Chem. 2012;161:908–913. doi: 10.1016/j.snb.2011.11.061. [DOI] [Google Scholar]
- 52.Uniyal S., Sharma R.K. Technological advancement in electrochemical biosensor based detection of Organophosphate pesticide chlorpyrifos in the environment: A review of status and prospects. Biosens. Bioelectron. 2018;116:37–50. doi: 10.1016/j.bios.2018.05.039. [DOI] [PubMed] [Google Scholar]
- 53.Shetti N.P., Malode S.J., Roy S., Chandra P., Reddy K.R., Chatterjee S. Nanomaterials in Diagnostic Tools and Devices. Elsevier; Amsterdam, The Netherlands: 2020. Electroanalytical techniques for investigating biofilms: Applications in biosensing and biomolecular interfacing. [Google Scholar]
- 54.Farka Z., Juřík T., Kovář D., Trnková L., Skládal P. Nanoparticle-Based Immunochemical Biosensors and Assays: Recent Advances and Challenges. Chem. Rev. 2017;117:9973–10042. doi: 10.1021/acs.chemrev.7b00037. [DOI] [PubMed] [Google Scholar]
- 55.Hasan A., Nurunnabi M., Morshed M., Paul A., Polini A., Kuila T., Al Hariri M., Lee Y.K., Jaffa A.A. Recent advances in application of biosensors in tissue engineering. Biomed Res. Int. 2014;2014 doi: 10.1155/2014/307519. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 56.Cipolatti E.P., Valério A., Henriques R.O., Moritz D.E., Ninow J.L., Freire D.M.G., Manoel E.A., Fernandez-Lafuente R., De Oliveira D. Nanomaterials for biocatalyst immobilization-state of the art and future trends. RSC Adv. 2016;6:104675–104692. doi: 10.1039/C6RA22047A. [DOI] [Google Scholar]
- 57.Bilal M., Adeel M., Rasheed T., Iqbal H.M.N. Multifunctional metal-organic frameworks-based biocatalytic platforms: Recent developments and future prospects. J. Mater. Res. Technol. 2019;8:2359–2371. doi: 10.1016/j.jmrt.2018.12.001. [DOI] [Google Scholar]
- 58.Hollmann F., Schmid A. Electrochemical regeneration of oxidoreductases for cell-free biocatalytic redox reactions. Biocatal. Biotransform. 2004;22:63–88. doi: 10.1080/10242420410001692778. [DOI] [Google Scholar]
- 59.Bandodkar A.J., Jia W., Ramírez J., Wang J. Biocompatible Enzymatic Roller Pens for Direct Writing of Biocatalytic Materials: “Do-it-Yourself” Electrochemical Biosensors. Adv. Healthc. Mater. 2015;4:1215–1224. doi: 10.1002/adhm.201400808. [DOI] [PubMed] [Google Scholar]
- 60.Campaña A.L., Florez S.L., Noguera M.J., Fuentes O.P., Puentes P.R., Cruz J.C., Osma J.F. Enzyme-based electrochemical biosensors for microfluidic platforms to detect pharmaceutical residues in wastewater. Biosensors. 2019;9:41. doi: 10.3390/bios9010041. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 61.Kucherenko I.S., Soldatkin O.O., Dzyadevych S.V., Soldatkin A.P. Electrochemical biosensors based on multienzyme systems: Main groups, advantages and limitations—A review. Anal. Chim. Acta. 2020;1111:114–131. doi: 10.1016/j.aca.2020.03.034. [DOI] [PubMed] [Google Scholar]
- 62.López Marzo A.M., Mayorga-Martinez C.C., Pumera M. 3D-printed graphene direct electron transfer enzyme biosensors. Biosens. Bioelectron. 2020;151:111980. doi: 10.1016/j.bios.2019.111980. [DOI] [PubMed] [Google Scholar]
- 63.Bäcker M., Koch C., Eiben S., Geiger F., Eber F., Gliemann H., Poghossian A., Wege C., Schöning M.J. Tobacco mosaic virus as enzyme nanocarrier for electrochemical biosensors. Sens. Actuators B Chem. 2017;238:716–722. doi: 10.1016/j.snb.2016.07.096. [DOI] [Google Scholar]
- 64.Alizadeh N., Salimi A. Ultrasensitive Bioaffinity Electrochemical Sensors: Advances and New Perspectives. Electroanalysis. 2018;30:2803–2840. doi: 10.1002/elan.201800598. [DOI] [Google Scholar]
- 65.Tu J., Torrente-Rodríguez R.M., Wang M., Gao W. The Era of Digital Health: A Review of Portable and Wearable Affinity Biosensors. Adv. Funct. Mater. 2020;30:1906713. doi: 10.1002/adfm.201906713. [DOI] [Google Scholar]
- 66.Huang Y., Xu J., Liu J., Wang X., Chen B. Disease-related detection with electrochemical biosensors: A review. Sensors. 2017;17:2375. doi: 10.3390/s17102375. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 67.Abdorahim M., Rabiee M., Alhosseini S.N., Tahriri M., Yazdanpanah S., Alavi S.H., Tayebi L. Nanomaterials-based electrochemical immunosensors for cardiac troponin recognition: An illustrated review. TrAC Trends Anal. Chem. 2016;82:337–347. doi: 10.1016/j.trac.2016.06.015. [DOI] [Google Scholar]
- 68.Cho I.H., Lee J., Kim J., Kang M.S., Paik J.K., Ku S., Cho H.M., Irudayaraj J., Kim D.H. Current technologies of electrochemical immunosensors: Perspective on signal amplification. Sensors. 2018;18:207. doi: 10.3390/s18010207. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 69.Khanmohammadi A., Aghaie A., Vahedi E., Qazvini A., Ghanei M., Afkhami A., Hajian A., Bagheri H. Electrochemical biosensors for the detection of lung cancer biomarkers: A review. Talanta. 2020;206:120251. doi: 10.1016/j.talanta.2019.120251. [DOI] [PubMed] [Google Scholar]
- 70.Carneiro P., Loureiro J., Delerue-Matos C., Morais S., do Carmo Pereira M. Alzheimer’s disease: Development of a sensitive label-free electrochemical immunosensor for detection of amyloid beta peptide. Sens. Actuators B Chem. 2017;239:157–165. doi: 10.1016/j.snb.2016.07.181. [DOI] [Google Scholar]
- 71.de Castro A.C.H., Alves L.M., Siquieroli A.C.S., Madurro J.M., Brito-Madurro A.G. Label-free electrochemical immunosensor for detection of oncomarker CA125 in serum. Microchem. J. 2020;155:104746. doi: 10.1016/j.microc.2020.104746. [DOI] [Google Scholar]
- 72.Camilo D.E., Miyazaki C.M., Shimizu F.M., Ferreira M. Improving direct immunoassay response by layer-by-layer films of gold nanoparticles—Antibody conjugate towards label-free detection. Mater. Sci. Eng. C. 2019;102:315–323. doi: 10.1016/j.msec.2019.04.055. [DOI] [PubMed] [Google Scholar]
- 73.Gogola J.L., Martins G., Caetano F.R., Ricciardi-Jorge T., Duarte dos Santos C.N., Marcolino-Junior L.H., Bergamini M.F. Label-free electrochemical immunosensor for quick detection of anti-hantavirus antibody. J. Electroanal. Chem. 2019;842:140–145. doi: 10.1016/j.jelechem.2019.04.066. [DOI] [Google Scholar]
- 74.Filik H., Avan A.A. Nanostructures for nonlabeled and labeled electrochemical immunosensors: Simultaneous electrochemical detection of cancer markers: A review. Talanta. 2019;205:120153. doi: 10.1016/j.talanta.2019.120153. [DOI] [PubMed] [Google Scholar]
- 75.Pan G., Zhao G., Wei M., Wang Y., Zhao B. Design of nanogold electrochemical immunosensor for detection of four phenolic estrogens. Chem. Phys. Lett. 2019;732:136657. doi: 10.1016/j.cplett.2019.136657. [DOI] [Google Scholar]
- 76.Jampasa S., Pummoree J., Siangproh W., Khongchareonporn N. Sensors and Actuators B: Chemical “ Signal-On ” electrochemical biosensor based on a competitive immunoassay format for the sensitive determination of oxytetracycline. Sens. Actuators B. Chem. 2020;320:128389. doi: 10.1016/j.snb.2020.128389. [DOI] [Google Scholar]
- 77.Hou L., Zhang X., Kong M., Jiang G., Sun Y., Mo W., Lin T., Ye F., Zhao S. A competitive immunoassay for electrochemical impedimetric determination of chlorpyrifos using a nanogold-modified glassy carbon electrode based on enzymatic biocatalytic precipitation. Microchim. Acta. 2020;187:204. doi: 10.1007/s00604-020-4175-1. [DOI] [PubMed] [Google Scholar]
- 78.Pei X., Zhang B., Tang J., Liu B., Lai W., Tang D. Sandwich-type immunosensors and immunoassays exploiting nanostructure labels: A review. Anal. Chim. Acta. 2013;758:1–18. doi: 10.1016/j.aca.2012.10.060. [DOI] [PubMed] [Google Scholar]
- 79.Felix F.S., Angnes L. Electrochemical immunosensors—A powerful tool for analytical applications. Biosens. Bioelectron. 2018;102:470–478. doi: 10.1016/j.bios.2017.11.029. [DOI] [PubMed] [Google Scholar]
- 80.Jampasa S., Lae-ngee P., Patarakul K., Ngamrojanavanich N., Chailapakul O., Rodthongkum N. Electrochemical immunosensor based on gold-labeled monoclonal anti-LipL32 for leptospirosis diagnosis. Biosens. Bioelectron. 2019;142:111539. doi: 10.1016/j.bios.2019.111539. [DOI] [PubMed] [Google Scholar]
- 81.Mishra G.K., Sharma V., Mishra R.K. Electrochemical aptasensors for food and environmental safeguarding: A review. Biosensors. 2018;8:28. doi: 10.3390/bios8020028. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 82.Yoon J.-Y. Basic Principles of Electrochemical Biosensing Using a Smartphone. Elsevier Inc.; Amsterdam, The Netherlands: 2020. [Google Scholar]
- 83.Shrivastava S., Trung T.Q., Lee N.E. Recent progress, challenges, and prospects of fully integrated mobile and wearable point-of-care testing systems for self-testing. Chem. Soc. Rev. 2020;49:1812–1866. doi: 10.1039/C9CS00319C. [DOI] [PubMed] [Google Scholar]
- 84.Yao Y., Jiang C., Ping J. Flexible freestanding graphene paper-based potentiometric enzymatic aptasensor for ultrasensitive wireless detection of kanamycin. Biosens. Bioelectron. 2019;123:178–184. doi: 10.1016/j.bios.2018.08.048. [DOI] [PubMed] [Google Scholar]
- 85.Dunn M.R., Mccloskey C.M., Buckley P., Rhea K., Chaput J.C. Generating Biologically Stable TNA Aptamers that Function with High Affinity and Thermal Stability. J. Am. Chem. Soc. 2020;142:7721–7724. doi: 10.1021/jacs.0c00641. [DOI] [PubMed] [Google Scholar]
- 86.Su S., Sun H., Cao W., Chao J., Peng H., Zuo X., Yuwen L., Fan C., Wang L. Dual-Target Electrochemical Biosensing Based on DNA Structural Switching on Gold Nanoparticle-Decorated MoS2 Nanosheets. ACS Appl. Mater. Interfaces. 2016;8:6826–6833. doi: 10.1021/acsami.5b12833. [DOI] [PubMed] [Google Scholar]
- 87.Wang L., Wen Y., Yang X., Xu L., Liang W., Zhu Y., Wang L., Li Y., Li Y., Ding M., et al. Ultrasensitive Electrochemical DNA Biosensor Based on a Label-Free Assembling Strategy Using a Triblock polyA DNA Probe. Anal. Chem. 2019;91:16002–16009. doi: 10.1021/acs.analchem.9b04757. [DOI] [PubMed] [Google Scholar]
- 88.Mazaafrianto D.N., Ishida A., Maeki M., Tani H., Tokeshi M. An electrochemical sensor based on structure switching of dithiol-modified aptamer for simple detection of Ochratoxin A. Anal. Sci. 2019;35:1221–1226. doi: 10.2116/analsci.19P240. [DOI] [PubMed] [Google Scholar]
- 89.Wang Q.L., Cui H.F., Song X., Fan S.F., Chen L.L., Li M.M., Li Z.Y. A label-free and lectin-based sandwich aptasensor for detection of carcinoembryonic antigen. Sens. Actuators B Chem. 2018;260:48–54. doi: 10.1016/j.snb.2017.12.105. [DOI] [Google Scholar]
- 90.Abnous K., Danesh N.M., Ramezani M., Emrani A.S., Taghdisi S.M. A novel colorimetric sandwich aptasensor based on an indirect competitive enzyme-free method for ultrasensitive detection of chloramphenicol. Biosens. Bioelectron. 2016;78:80–86. doi: 10.1016/j.bios.2015.11.028. [DOI] [PubMed] [Google Scholar]
- 91.Zhong H., Yu C., Gao R., Chen J., Yu Y., Geng Y., Wen Y., He J. A novel sandwich aptasensor for detecting T-2 toxin based on rGO-TEPA-Au@Pt nanorods with a dual signal amplification strategy. Biosens. Bioelectron. 2019;144:111635. doi: 10.1016/j.bios.2019.111635. [DOI] [PubMed] [Google Scholar]
- 92.Farzin L., Sadjadi S., Shamsipur M., Sheibani S., Mousazadeh M. hasan Employing AgNPs doped amidoxime-modified polyacrylonitrile (PAN-oxime) nanofibers for target induced strand displacement-based electrochemical aptasensing of CA125 in ovarian cancer patients. Mater. Sci. Eng. C. 2019;97:679–687. doi: 10.1016/j.msec.2018.12.108. [DOI] [PubMed] [Google Scholar]
- 93.Li Z., Zhou Z., Xue N., Wu S., Miao X. Electrochemical aptamer-based determination of protein tyrosine kinase-7 using toehold-mediated strand displacement amplification on gold nanoparticles and graphene oxide. Microchim. Acta. 2019;186:720. doi: 10.1007/s00604-019-3849-z. [DOI] [PubMed] [Google Scholar]
- 94.Wang Y., Yao L., Ning G., Wu Y., Wu S., Mao S., Liu G.Q. An electrochemical strategy for tetracycline detection coupled triple helix aptamer probe with catalyzed hairpin assembly signal amplification. Biosens. Bioelectron. 2019;143:111613. doi: 10.1016/j.bios.2019.111613. [DOI] [PubMed] [Google Scholar]
- 95.Nameghi M.A., Danesh N.M., Ramezani M., Alibolandi M., Abnous K., Taghdisi S.M. An ultrasensitive electrochemical sensor for 17β-estradiol using split aptamers. Anal. Chim. Acta. 2019;1065:107–112. doi: 10.1016/j.aca.2019.02.062. [DOI] [PubMed] [Google Scholar]
- 96.Jiang Y., Sun D., Liang Z., Chen L., Zhang Y., Chen Z. Label-free and competitive aptamer cytosensor based on layer-by-layer assembly of DNA-platinum nanoparticles for ultrasensitive determination of tumor cells. Sens. Actuators B Chem. 2018;262:35–43. doi: 10.1016/j.snb.2018.01.194. [DOI] [Google Scholar]
- 97.Cui F., Zhou Z., Zhou H.S. Review—Measurement and Analysis of Cancer Biomarkers Based on Electrochemical Biosensors. J. Electrochem. Soc. 2020;167:037525. doi: 10.1149/2.0252003JES. [DOI] [Google Scholar]
- 98.Löfblom J., Feldwisch J., Tolmachev V., Carlsson J., Ståhl S., Frejd F.Y. Affibody molecules: Engineered proteins for therapeutic, diagnostic and biotechnological applications. FEBS Lett. 2010;584:2670–2680. doi: 10.1016/j.febslet.2010.04.014. [DOI] [PubMed] [Google Scholar]
- 99.Frejd F.Y., Kim K.T. Affibody molecules as engineered protein drugs. Exp. Mol. Med. 2017;49:e306. doi: 10.1038/emm.2017.35. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 100.Ravalli A., da Rocha C.G., Yamanaka H., Marrazza G. A label-free electrochemical affisensor for cancer marker detection: The case of HER2. Bioelectrochemistry. 2015;106:268–275. doi: 10.1016/j.bioelechem.2015.07.010. [DOI] [PubMed] [Google Scholar]
- 101.Tang Z., Ma Z. Multiple functional strategies for amplifying sensitivity of amperometric immunoassay for tumor markers: A review. Biosens. Bioelectron. 2017;98:100–112. doi: 10.1016/j.bios.2017.06.041. [DOI] [PubMed] [Google Scholar]
- 102.Ahammad A.J.S., Pal P.R., Shah S.S., Islam T., Hasan M.M., Qasem M.A.A., Odhikari N., Sarker S., Kim D.M., Aziz M.A. Activated jute carbon paste screen-printed FTO electrodes for nonenzymatic amperometric determination of nitrite. J. Electroanal. Chem. 2019;832:368–379. doi: 10.1016/j.jelechem.2018.11.034. [DOI] [Google Scholar]
- 103.Joshi A., Kim K. Biosensors and Bioelectronics Recent advances in nanomaterial-based electrochemical detection of antibiotics: Challenges and future perspectives. Biosens. Bioelectron. 2020;153:112046. doi: 10.1016/j.bios.2020.112046. [DOI] [PubMed] [Google Scholar]
- 104.Sahin B., Kaya T. Electrochemical amperometric biosensor applications of nanostructured metal oxides: A review. Mater. Res. Express. 2019;6:042003. doi: 10.1088/2053-1591/aafa95. [DOI] [Google Scholar]
- 105.Dong X.X., Yang J.Y., Luo L., Zhang Y.F., Mao C., Sun Y.M., Lei H.T., Shen Y.D., Beier R.C., Xu Z.L. Portable amperometric immunosensor for histamine detection using Prussian blue-chitosan-gold nanoparticle nanocomposite films. Biosens. Bioelectron. 2017;98:305–309. doi: 10.1016/j.bios.2017.07.014. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 106.Muñoz J., Montes R., Bastos-Arrieta J., Guardingo M., Busqué F., Ruíz-Molina D., Palet C., García-Orellana J., Baeza M. Carbon nanotube-based nanocomposite sensor tuned with a catechol as novel electrochemical recognition platform of uranyl ion in aqueous samples. Sens. Actuators B Chem. 2018;273:1807–1815. doi: 10.1016/j.snb.2018.07.093. [DOI] [Google Scholar]
- 107.Meirinho S.G., Dias L.G., Peres A.M., Rodrigues L.R. Voltammetric aptasensors for protein disease biomarkers detection: A review. Biotechnol. Adv. 2016;34:941–953. doi: 10.1016/j.biotechadv.2016.05.006. [DOI] [PubMed] [Google Scholar]
- 108.Shrivastava S., Jadon N., Jain R. Next-generation polymer nanocomposite-based electrochemical sensors and biosensors: A review. TrAC Trends Anal. Chem. 2016;82:55–67. doi: 10.1016/j.trac.2016.04.005. [DOI] [Google Scholar]
- 109.Kumar T.H.V., Sundramoorthy A.K. Electrochemical biosensor for methyl parathion based on single-walled carbon nanotube/glutaraldehyde crosslinked acetylcholinesterase-wrapped bovine serum albumin nanocomposites. Anal. Chim. Acta. 2019;1074:131–141. doi: 10.1016/j.aca.2019.05.011. [DOI] [PubMed] [Google Scholar]
- 110.Diouf A., Motia S., El Alami El Hassani N., El Bari N., Bouchikhi B. Development and characterization of an electrochemical biosensor for creatinine detection in human urine based on functional molecularly imprinted polymer. J. Electroanal. Chem. 2017;788:44–53. doi: 10.1016/j.jelechem.2017.01.068. [DOI] [Google Scholar]
- 111.Scholz F. Voltammetric techniques of analysis: The essentials. ChemTexts. 2015;1:17. doi: 10.1007/s40828-015-0016-y. [DOI] [Google Scholar]
- 112.Li X., Jiang M., Cheng J., Ye M., Zhang W., Jaffrezic-Renault N., Guo Z. Signal multi-amplified electrochemical biosensor for voltammetric determination of tau-441 protein in biological samples using carbon nanomaterials and gold nanoparticles to hint dementia. Microchim. Acta. 2020;187:302. doi: 10.1007/s00604-020-04273-z. [DOI] [PubMed] [Google Scholar]
- 113.Ding J., Qin W. Recent advances in potentiometric biosensors. TrAC Trends Anal. Chem. 2020;124:115803. doi: 10.1016/j.trac.2019.115803. [DOI] [Google Scholar]
- 114.Nishitani S., Sakata T. Enhancement of Signal-to-Noise Ratio for Serotonin Detection with Well-Designed Nanofilter-Coated Potentiometric Electrochemical Biosensor. ACS Appl. Mater. Interfaces. 2020;12:14761–14769. doi: 10.1021/acsami.9b19309. [DOI] [PubMed] [Google Scholar]
- 115.Mello H.J.N.P.D., Mulato M. Enzymatically functionalized polyaniline thin films produced with one-step electrochemical immobilization and its application in glucose and urea potentiometric biosensors. Biomed. Microdevices. 2020;22:22. doi: 10.1007/s10544-020-00478-4. [DOI] [PubMed] [Google Scholar]
- 116.Silva N.F.D., Almeida C.M.R., Magalhães J.M.C.S., Gonçalves M.P., Freire C., Delerue-Matos C. Development of a disposable paper-based potentiometric immunosensor for real-time detection of a foodborne pathogen. Biosens. Bioelectron. 2019;141:111317. doi: 10.1016/j.bios.2019.111317. [DOI] [PubMed] [Google Scholar]
- 117.Manjakkal L., Dang W., Yogeswaran N., Dahiya R. Textile-based potentiometric electrochemical PH sensor for wearable applications. Biosensors. 2019;9:14. doi: 10.3390/bios9010014. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 118.Muhammad-Tahir Z., Alocilja E.C. A conductometric biosensor for biosecurity. Biosens. Bioelectron. 2003;18:813–819. doi: 10.1016/S0956-5663(03)00020-4. [DOI] [PubMed] [Google Scholar]
- 119.Cesewski E., Johnson B.N. Electrochemical biosensors for pathogen detection. Biosens. Bioelectron. 2020;159:112214. doi: 10.1016/j.bios.2020.112214. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 120.Kamel S., Khattab T. Recent Advances in Cellulose-Based Biosensors for medical diagnosis. Biosensors. 2020;10:67. doi: 10.3390/bios10060067. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 121.Kolahchi N., Braiek M., Ebrahimipour G., Ranaei-Siadat S.O., Lagarde F., Jaffrezic-Renault N. Direct detection of phenol using a new bacterial strain-based conductometric biosensor. J. Environ. Chem. Eng. 2018;6:478–484. doi: 10.1016/j.jece.2017.12.023. [DOI] [Google Scholar]
- 122.Soldatkin O.O., Stepurska K.V., Arkhypova V.M., Soldatkin A.P., El’skaya A.V., Lagarde F., Dzyadevych S.V. Conductometric enzyme biosensor for patulin determination. Sens. Actuators B Chem. 2017;239:1010–1015. doi: 10.1016/j.snb.2016.08.121. [DOI] [Google Scholar]
- 123.Lavanya N., Leonardi S.G., Marini S., Espro C., Kanagaraj M., Reddy S.L., Sekar C., Neri G. MgNi2O3 nanoparticles as novel and versatile sensing material for non-enzymatic electrochemical sensing of glucose and conductometric determination of acetone. J. Alloy. Compd. 2020;817:152787. doi: 10.1016/j.jallcom.2019.152787. [DOI] [Google Scholar]
- 124.Bahadir E.B., Sezgintürk M.K. A review on impedimetric biosensors. Artif. Cells Nanomed. Biotechnol. 2016;44:248–262. doi: 10.3109/21691401.2014.942456. [DOI] [PubMed] [Google Scholar]
- 125.Ahammad A.J.S., Al Mamun A., Akter T., Mamun M.A., Faraezi S., Monira F.Z. Enzyme-free impedimetric glucose sensor based on gold nanoparticles/polyaniline composite film. J. Solid State Electrochem. 2016;20:1933–1939. doi: 10.1007/s10008-016-3199-2. [DOI] [Google Scholar]
- 126.Rengaraj S., Cruz-Izquierdo Á., Scott J.L., Di Lorenzo M. Impedimetric paper-based biosensor for the detection of bacterial contamination in water. Sens. Actuators B Chem. 2018;265:50–58. doi: 10.1016/j.snb.2018.03.020. [DOI] [Google Scholar]
- 127.Xu M., Yadavalli V.K. Flexible Biosensors for the Impedimetric Detection of Protein Targets Using Silk-Conductive Polymer Biocomposites. ACS Sens. 2019;4:1040–1047. doi: 10.1021/acssensors.9b00230. [DOI] [PubMed] [Google Scholar]
- 128.Ivanova O.S., Zamborini F.P. Size-dependent electrochemical oxidation of silver nanoparticles. J. Am. Chem. Soc. 2010;132:70–72. doi: 10.1021/ja908780g. [DOI] [PubMed] [Google Scholar]
- 129.Voisin C., Christofilos D., Del Fatti N., Vallée F., Prével B., Cottancin E., Lermé J., Pellarin M., Broyer M. Size-dependent electron-electron interactions in metal nanoparticles. Phys. Rev. Lett. 2000;85:2200–2203. doi: 10.1103/PhysRevLett.85.2200. [DOI] [PubMed] [Google Scholar]
- 130.Ma H., Gao P., Qian P., Su Y. Size-Dependent Electrochemical Properties of Pure Metallic Nanoparticles. J. Phys. Chem. C. 2020;124:3403–3409. doi: 10.1021/acs.jpcc.9b10962. [DOI] [Google Scholar]
- 131.Xiao T., Huang J., Wang D., Meng T., Yang X. Au and Au-Based nanomaterials: Synthesis and recent progress in electrochemical sensor applications. Talanta. 2020;206:120210. doi: 10.1016/j.talanta.2019.120210. [DOI] [PubMed] [Google Scholar]
- 132.Tang L., Li X., Cammarata R.C., Friesen C., Sieradzki K. Electrochemical stability of elemental metal nanoparticles. J. Am. Chem. Soc. 2010;132:11722–11726. doi: 10.1021/ja104421t. [DOI] [PubMed] [Google Scholar]
- 133.Elahi N., Kamali M., Baghersad M.H. Recent biomedical applications of gold nanoparticles: A review. Talanta. 2018;184:537–556. doi: 10.1016/j.talanta.2018.02.088. [DOI] [PubMed] [Google Scholar]
- 134.Alex S., Tiwari A. Functionalized gold nanoparticles: Synthesis, properties and applications—A review. J. Nanosci. Nanotechnol. 2015;15:1869–1894. doi: 10.1166/jnn.2015.9718. [DOI] [PubMed] [Google Scholar]
- 135.Chen H., Kou X., Yang Z., Ni W., Wang J. Shape- and size-dependent refractive index sensitivity of gold nanoparticles. Langmuir. 2008;24:5233–5237. doi: 10.1021/la800305j. [DOI] [PubMed] [Google Scholar]
- 136.Eustis S., El-Sayed M.A. Why gold nanoparticles are more precious than pretty gold: Noble metal surface plasmon resonance and its enhancement of the radiative and nonradiative properties of nanocrystals of different shapes. Chem. Soc. Rev. 2006;35:209–217. doi: 10.1039/B514191E. [DOI] [PubMed] [Google Scholar]
- 137.Khatoon U.T., Rao G.V.S.N., Mantravadi K.M., Oztekin Y. Strategies to synthesize various nanostructures of silver and their applications—A review. RSC Adv. 2018;8:19739–19753. doi: 10.1039/C8RA00440D. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 138.Lee P.C., Meisel D. Adsorption and surface-enhanced Raman of dyes on silver and gold sols. J. Phys. Chem. 1982;86:3391–3395. doi: 10.1021/j100214a025. [DOI] [Google Scholar]
- 139.Suchomel P., Kvitek L., Prucek R., Panacek A., Halder A., Vajda S., Zboril R. Simple size-controlled synthesis of Au nanoparticles and their size-dependent catalytic activity. Sci. Rep. 2018;8:4589. doi: 10.1038/s41598-018-22976-5. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 140.Narayanan J.S., Slaughter G. Towards a dual in-line electrochemical biosensor for the determination of glucose and hydrogen peroxide. Bioelectrochemistry. 2019;128:56–65. doi: 10.1016/j.bioelechem.2019.03.005. [DOI] [PubMed] [Google Scholar]
- 141.Quintero-Jaime A.F., Berenguer-Murcia Á., Cazorla-Amorós D., Morallón E. Carbon nanotubes modified with Au for electrochemical detection of prostate specific antigen: Effect of au nanoparticle size distribution. Front. Chem. 2019;7:147. doi: 10.3389/fchem.2019.00147. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 142.Akter M., Sikder M.T., Rahman M.M., Ullah A.K.M.A., Hossain K.F.B., Banik S., Hosokawa T., Saito T., Kurasaki M. A systematic review on silver nanoparticles-induced cytotoxicity: Physicochemical properties and perspectives. J. Adv. Res. 2018;9:1–16. doi: 10.1016/j.jare.2017.10.008. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 143.Wonner K., Evers M.V., Tschulik K. The electrochemical dissolution of single silver nanoparticles enlightened by hyperspectral dark-field microscopy. Electrochim. Acta. 2019;301:458–464. doi: 10.1016/j.electacta.2019.01.129. [DOI] [Google Scholar]
- 144.Pich A., Karak A., Lu Y., Ghosh A.K., Adler H.J.P. Preparation of hybrid microgels functionalized by silver nanoparticles. Macromol. Rapid Commun. 2006;27:344–350. doi: 10.1002/marc.200500761. [DOI] [Google Scholar]
- 145.Natsuki J. A Review of Silver Nanoparticles: Synthesis Methods, Properties and Applications. Int. J. Mater. Sci. Appl. 2015;4:325. doi: 10.11648/j.ijmsa.20150405.17. [DOI] [Google Scholar]
- 146.Fang Y. Optical absorption of nanoscale colloidal silver: Aggregate band and adsorbate-silver surface band. J. Chem. Phys. 1998;108:4315–4318. doi: 10.1063/1.475831. [DOI] [Google Scholar]
- 147.Austin L.A., MacKey M.A., Dreaden E.C., El-Sayed M.A. The optical, photothermal, and facile surface chemical properties of gold and silver nanoparticles in biodiagnostics, therapy, and drug delivery. Arch. Toxicol. 2014;88:1391–1417. doi: 10.1007/s00204-014-1245-3. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 148.Chen L., Xie H., Li J. Electrochemical glucose biosensor based on silver nanoparticles/multiwalled carbon nanotubes modified electrode. J. Solid State Electrochem. 2012;16:3323–3329. doi: 10.1007/s10008-012-1773-9. [DOI] [Google Scholar]
- 149.Gao C., Lyu F., Yin Y. Encapsulated Metal Nanoparticles for Catalysis. Chem. Rev. 2020 doi: 10.1021/acs.chemrev.0c00237. [DOI] [PubMed] [Google Scholar]
- 150.Yao Y., Lan L., Liu X., Ying Y., Ping J. Spontaneous growth and regulation of noble metal nanoparticles on flexible biomimetic MXene paper for bioelectronics. Biosens. Bioelectron. 2020;148:111799. doi: 10.1016/j.bios.2019.111799. [DOI] [PubMed] [Google Scholar]
- 151.García-Cruz L., Montiel V., Solla-Gullón J. Shape-controlled metal nanoparticles for electrocatalytic applications. Phys. Sci. Rev. 2019;4:1–34. doi: 10.1515/psr-2017-0124. [DOI] [Google Scholar]
- 152.Kucherenko I.S., Soldatkin O.O., Kucherenko D.Y., Soldatkina O.V., Dzyadevych S.V. Advances in nanomaterial application in enzyme-based electrochemical biosensors: A review. Nanoscale Adv. 2019;1:4560–4577. doi: 10.1039/C9NA00491B. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 153.Huang Z., Zhang A., Zhang Q., Pan S., Cui D. Electrochemical Biosensor Based on Dewdrop-Like Platinum Nanoparticles-Decorated Silver Nanoflowers Nanocomposites for H2O2 and Glucose Detection. J. Electrochem. Soc. 2019;166:B1138–B1145. doi: 10.1149/2.0471913jes. [DOI] [Google Scholar]
- 154.Lü K., Zhao G.X., Wang X.K. A brief review of graphene-based material synthesis and its application in environmental pollution management. Chin. Sci. Bull. 2012;57:1223–1234. doi: 10.1007/s11434-012-4986-5. [DOI] [Google Scholar]
- 155.Tran H.V., Le T.A., Giang B.L., Piro B., Tran L.D. Silver nanoparticles on graphene quantum dots as nanozyme for efficient H2O2 reduction in a glucose biosensor. Mater. Res. Express. 2019;6:115403. doi: 10.1088/2053-1591/ab46ca. [DOI] [Google Scholar]
- 156.Simfukwe J., Mapasha R.E., Braun A., Diale M. Biopatterning of Keratinocytes in Aqueous Two-Phase Systems as a Potential Tool for Skin Tissue Engineering. MRS Adv. 2017;357:2443–2449. doi: 10.1557/adv.201. [DOI] [Google Scholar]
- 157.You Z., Qiu Q., Chen H., Feng Y., Wang X., Wang Y., Ying Y. Laser-induced noble metal nanoparticle-graphene composites enabled flexible biosensor for pathogen detection. Biosens. Bioelectron. 2020;150:111896. doi: 10.1016/j.bios.2019.111896. [DOI] [PubMed] [Google Scholar]
- 158.Smalley R.E. Discovering the fullerenes. Rev. Mod. Phys. 1997;69:723–730. doi: 10.1103/RevModPhys.69.723. [DOI] [Google Scholar]
- 159.Sherigara B.S., Kutner W., D’Souza F. Electrocatalytic properties and sensor applications of fullerenes and carbon nanotubes. Electroanalysis. 2003;15:753–772. doi: 10.1002/elan.200390094. [DOI] [Google Scholar]
- 160.Li Q., Wudl F., Thilgen C., Whetten R.L., Diederich F. Unusual Electrochemical Properties of the Higher Fullerene, Chiral C76. J. Am. Chem. Soc. 1992;114:3994–3996. doi: 10.1021/ja00036a068. [DOI] [Google Scholar]
- 161.Kuzmany H., Pfeiffer R., Hulman M., Kramberger C. Raman spectroscopy of fullerenes and fullerene-nanotube composites. Philos. Trans. R. Soc. A Math. Phys. Eng. Sci. 2004;362:2375–2406. doi: 10.1098/rsta.2004.1446. [DOI] [PubMed] [Google Scholar]
- 162.Ren J., Xu Q., Chen X., Li W., Guo K., Zhao Y., Wang Q., Zhang Z., Peng H., Li Y.G. Superaligned Carbon Nanotubes Guide Oriented Cell Growth and Promote Electrophysiological Homogeneity for Synthetic Cardiac Tissues. Adv. Mater. 2017;29:1702713. doi: 10.1002/adma.201702713. [DOI] [PubMed] [Google Scholar]
- 163.Hu C.G., Wang W.L., Wang S.X., Zhu W., Li Y. Investigation on electrochemical properties of carbon nanotubes. Diam. Relat. Mater. 2003;12:1295–1299. doi: 10.1016/S0925-9635(03)00071-2. [DOI] [Google Scholar]
- 164.Baig N., Saleh T.A. Electrodes modified with 3D graphene composites: A review on methods for preparation, properties and sensing applications. Microchim. Acta. 2018;185:283. doi: 10.1007/s00604-018-2809-3. [DOI] [PubMed] [Google Scholar]
- 165.Fasolino A., Katsnelson M.I. Intrinsic ripples in graphene. Nat. Mater. 2007;6:6–9. doi: 10.1038/nmat2011. [DOI] [PubMed] [Google Scholar]
- 166.Morozov S.V., Novoselov K.S., Katsnelson M.I., Schedin F., Ponomarenko L.A., Jiang D., Geim A.K. Strong suppression of weak localization in graphene. Phys. Rev. Lett. 2006;97:7–10. doi: 10.1103/PhysRevLett.97.016801. [DOI] [PubMed] [Google Scholar]
- 167.Pacilé D., Meyer J.C., Fraile Rodríguez A., Papagno M., Gómez-Navarro C., Sundaram R.S., Burghard M., Kern K., Carbone C., Kaiser U. Electronic properties and atomic structure of graphene oxide membranes. Carbon N. Y. 2011;49:966–972. doi: 10.1016/j.carbon.2010.09.063. [DOI] [Google Scholar]
- 168.Wang L., Li J., Pan Y., Min L., Zhang Y., Hu X., Yang Z. Platinum nanoparticle-assembled nanoflake-like tin disulfide for enzyme-based amperometric sensing of glucose. Microchim. Acta. 2017;184:2357–2363. doi: 10.1007/s00604-017-2209-0. [DOI] [Google Scholar]
- 169.Çevik S. Xanthine biosensor based on XO/AuNP/PtNP/MWCNT hybrid nanocomposite modified GCPE. Biotechnol. Bioprocess Eng. 2016;21:314–320. doi: 10.1007/s12257-016-0014-y. [DOI] [Google Scholar]
- 170.Xing Y., Feng X.Z., Zhang L., Hou J., Han G.C., Chen Z. A sensitive and selective electrochemical biosensor for the determination of beta-amyloid oligomer by inhibiting the peptide-triggered in situ assembly of silver nanoparticles. Int. J. Nanomed. 2017;12:3171–3179. doi: 10.2147/IJN.S132776. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 171.Gibson M.I., Seyedsayamdost M.R. Small Biomolecules for Big Applications. ACS Cent. Sci. 2018;4:437–439. doi: 10.1021/acscentsci.8b00088. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 172.Zardecki C., Dutta S., Goodsell D.S., Voigt M., Burley S.K. RCSB Protein Data Bank: A Resource for Chemical, Biochemical, and Structural Explorations of Large and Small Biomolecules. J. Chem. Educ. 2016;93:569–575. doi: 10.1021/acs.jchemed.5b00404. [DOI] [Google Scholar]
- 173.Luong J.H.T., Glennon J.D., Gedanken A., Vashist S.K. Achievement and assessment of direct electron transfer of glucose oxidase in electrochemical biosensing using carbon nanotubes, graphene, and their nanocomposites. Microchim. Acta. 2017;184:369–388. doi: 10.1007/s00604-016-2049-3. [DOI] [Google Scholar]
- 174.Mathers C.D., Loncar D. Projections of global mortality and burden of disease from 2002 to 2030. PLoS Med. 2006;3:2011–2030. doi: 10.1371/journal.pmed.0030442. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 175.Hasan M.M., Ehsan M.A., Islam T., Alharthi N.H., Alharbi H.F., Karim M.R., Abdul Aziz M., Saleh Ahammad A.J. Selective detection of dopamine at the AACVD synthesized palladium nanoparticles and understanding the sensing mechanism through electrochemical and computational study. J. Electrochem. Soc. 2019;166:B1528–B1542. doi: 10.1149/2.0631915jes. [DOI] [Google Scholar]
- 176.Ahammad A.J.S., Odhikari N., Shah S.S., Hasan M.M., Islam T., Pal P.R., Ahmed Qasem M.A., Aziz M.A. Porous tal palm carbon nanosheets: Preparation, characterization and application for the simultaneous determination of dopamine and uric acid. Nanoscale Adv. 2019;1:613–626. doi: 10.1039/C8NA00090E. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 177.Ehsan M.A., Hasan M.M., Islam T., Hossain M.D., Aziz M.A., Ahammad A.J.S. Fabrication of Nanostructured Pd Thin Films Using Aerosol-Assisted Chemical Vapor Deposition for the Nonenzymatic Electrochemical Detection of H2O2. ACS Appl. Electron. Mater. 2019;1:417–429. doi: 10.1021/acsaelm.8b00131. [DOI] [Google Scholar]
- 178.Meng W., Wen Y., Dai L., He Z., Wang L. A novel electrochemical sensor for glucose detection based on Ag@ZIF-67 nanocomposite. Sens. Actuators B Chem. 2018;260:852–860. doi: 10.1016/j.snb.2018.01.109. [DOI] [Google Scholar]
- 179.Barsan M.M., Enache T.A., Preda N., Stan G., Apostol N.G., Matei E., Kuncser A., Diculescu V.C. Direct Immobilization of Biomolecules through Magnetic Forces on Ni Electrodes via Ni Nanoparticles: Applications in Electrochemical Biosensors. ACS Appl. Mater. Interfaces. 2019;11:19867–19877. doi: 10.1021/acsami.9b04990. [DOI] [PubMed] [Google Scholar]
- 180.Liu H., Weng L., Yang C. A review on nanomaterial-based electrochemical sensors for H2O2, H2S and NO inside cells or released by cells. Microchim. Acta. 2017;184:1267–1283. doi: 10.1007/s00604-017-2179-2. [DOI] [Google Scholar]
- 181.Yoon J., Lee T., Bapurao B., Jo J., Oh B.K., Choi J.W. Electrochemical H2O2 biosensor composed of myoglobin on MoS2 nanoparticle-graphene oxide hybrid structure. Biosens. Bioelectron. 2017;93:14–20. doi: 10.1016/j.bios.2016.11.064. [DOI] [PubMed] [Google Scholar]
- 182.Manickam P., Vashist A., Madhu S., Sadasivam M., Sakthivel A., Kaushik A., Nair M. Gold nanocubes embedded biocompatible hybrid hydrogels for electrochemical detection of H2O2. Bioelectrochemistry. 2020;131:107373. doi: 10.1016/j.bioelechem.2019.107373. [DOI] [PubMed] [Google Scholar]
- 183.Kumar R., Manikandan E., Pandian K., Mahnashi M.H., Alsaiari M.A., Ibrahim A.A., Bouropoulos N., Baskoutas S. Solid-state synthesis of Ag-doped PANI nanocomposites for theirend-use as an electrochemical sensor for hydrogen peroxide anddopamine. Electrochim. Acta. 2020;363:137158. doi: 10.1016/j.electacta.2020.137158. [DOI] [Google Scholar]
- 184.Jain S., Verma S., Singh S.P., Sharma S.N. An electrochemical biosensor based on novel butylamine capped CZTS nanoparticles immobilized by uricase for uric acid detection. Biosens. Bioelectron. 2019;127:135–141. doi: 10.1016/j.bios.2018.12.008. [DOI] [PubMed] [Google Scholar]
- 185.Verma S., Choudhary J., Singh K.P., Chandra P., Singh S.P. Uricase grafted nanoconducting matrix based electrochemical biosensor for ultrafast uric acid detection in human serum samples. Int. J. Biol. Macromol. 2019;130:333–341. doi: 10.1016/j.ijbiomac.2019.02.121. [DOI] [PubMed] [Google Scholar]
- 186.Shin J.W., Yoon J., Shin M., Choi J.W. Electrochemical Dopamine Biosensor Composed of Silver Encapsulated MoS 2 Hybrid Nanoparticle. Biotechnol. Bioprocess Eng. 2019;24:135–144. doi: 10.1007/s12257-018-0350-1. [DOI] [Google Scholar]
- 187.Yoon H., Nah J., Kim H., Ko S., Sharifuzzaman M., Barman S.C., Xuan X., Kim J., Park J.Y. A chemically modified laser-induced porous graphene based flexible and ultrasensitive electrochemical biosensor for sweat glucose detection. Sens. Actuators B Chem. 2020;311:127866. doi: 10.1016/j.snb.2020.127866. [DOI] [Google Scholar]
- 188.Sanaeifar N., Rabiee M., Abdolrahim M., Tahriri M., Vashaee D., Tayebi L. A novel electrochemical biosensor based on Fe3O4 nanoparticles-polyvinyl alcohol composite for sensitive detection of glucose. Anal. Biochem. 2017;519:19–26. doi: 10.1016/j.ab.2016.12.006. [DOI] [PubMed] [Google Scholar]
- 189.Preethika M., Sundramoorthy A.K. Humic acid/halloysite nanotube/flavin adenine dinucleotide nanocomposite based selective electrochemical biosensor for hydrogen peroxide. Appl. Surf. Sci. 2019;488:503–511. doi: 10.1016/j.apsusc.2019.05.255. [DOI] [Google Scholar]
- 190.Vargas A.J., Harris C.C. Biomarker development in the precision medicine era: Lung cancer as a case study. Nat. Rev. Cancer. 2016;16:525–537. doi: 10.1038/nrc.2016.56. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 191.Xiao T., Ying W., Li L., Hu Z., Ma Y., Jiao L., Ma J., Cai Y., Lin D., Guo S., et al. An approach to studying lung cancer-related proteins in human blood. Mol. Cell. Proteom. 2005;4:1480–1486. doi: 10.1074/mcp.M500055-MCP200. [DOI] [PubMed] [Google Scholar]
- 192.Nagata M., Noman A.A., Suzuki K., Kurita H., Ohnishi M., Ohyama T., Kitamura N., Kobayashi T., Uematsu K., Takahashi K., et al. ITGA3 and ITGB4 expression biomarkers estimate the risks of locoregional and hematogenous dissemination of oral squamous cell carcinoma. BMC Cancer. 2013;13:410. doi: 10.1186/1471-2407-13-410. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 193.Shinmura K., Igarashi H., Kato H., Kawanishi Y., Inoue Y., Nakamura S., Ogawa H., Yamashita T., Kawase A., Funai K., et al. CLCA2 as a novel immunohistochemical marker for differential diagnosis of squamous cell carcinoma from adenocarcinoma of the lung. Dis. Markers. 2014;2014:619273. doi: 10.1155/2014/619273. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 194.Bottoni P., Scatena R. The role of CA 125 as tumor marker: Biochemical and clinical aspects. Adv. Exp. Med. Biol. 2015;867:229–244. doi: 10.1007/978-94-017-7215-0_14. [DOI] [PubMed] [Google Scholar]
- 195.Barman S.C., Hossain M.F., Yoon H., Park J.Y. Trimetallic Pd@Au@Pt nanocomposites platform on -COOH terminated reduced graphene oxide for highly sensitive CEA and PSA biomarkers detection. Biosens. Bioelectron. 2018;100:16–22. doi: 10.1016/j.bios.2017.08.045. [DOI] [PubMed] [Google Scholar]
- 196.Chen Y., Li Y., Deng D., He H., Yan X., Wang Z., Fan C., Luo L. Effective immobilization of Au nanoparticles on TiO2 loaded graphene for a novel sandwich-type immunosensor. Biosens. Bioelectron. 2018;102:301–306. doi: 10.1016/j.bios.2017.11.009. [DOI] [PubMed] [Google Scholar]
- 197.Gu X., She Z., Ma T., Tian S., Kraatz H.B. Electrochemical detection of carcinoembryonic antigen. Biosens. Bioelectron. 2018;102:610–616. doi: 10.1016/j.bios.2017.12.014. [DOI] [PubMed] [Google Scholar]
- 198.Chen S., Yang Y., Li W., Song Y., Shi L., Hong C. A sandwich-type electrochemical immunosensor using Ag@CeO2-Au as a lable for sensitive detection of carcinoembryonic antigen. Microchem. J. 2020;159:105415. doi: 10.1016/j.microc.2020.105415. [DOI] [Google Scholar]
- 199.Song Y., Qiao J., Li W., Ma C., Chen S., Li H., Hong C. Bimetallic PtCu nanoparticles supported on molybdenum disulfide-functionalized graphitic carbon nitride for the detection of carcinoembryonic antigen. Microchim. Acta. 2020;187:538. doi: 10.1007/s00604-020-04498-y. [DOI] [PubMed] [Google Scholar]
- 200.Li X., Weng C., Wang J., Yang W., Lu Q., Yan X., Sakran M.A., Hong J., Zhu W., Zhou X. A label-free electrochemical magnetic aptasensor based on exonuclease III–assisted signal amplification for determination of carcinoembryonic antigen. Microchim. Acta. 2020;187:492. doi: 10.1007/s00604-020-04457-7. [DOI] [PubMed] [Google Scholar]
- 201.Zhao C., Ma C., Wu M., Li W., Song Y., Yang Y., Yin B., Hong C., Qiao X. An immunosensor detects carcinoembryonic antigen by a double reduction strategy based on polyphenylamine as a sacrifice reducing agent. Int. J. Hydrog. Energy. 2020;45:5055–5066. doi: 10.1016/j.ijhydene.2019.12.041. [DOI] [Google Scholar]
- 202.Liu J., Shang Y., Xu J., Chen Y., Jia Y., Zheng J. A novel electrochemical immunosensor for carcinoembryonic antigen based on Cu-MOFs-TB/polydopamine nanocarrier. J. Electroanal. Chem. 2020;877:114563. doi: 10.1016/j.jelechem.2020.114563. [DOI] [Google Scholar]
- 203.Butmee P., Tumcharern G., Thouand G., Kalcher K., Samphao A. An ultrasensitive immunosensor based on manganese dioxide-graphene nanoplatelets and core shell Fe3O4@Au nanoparticles for label-free detection of carcinoembryonic antigen. Bioelectrochemistry. 2020;132:107452. doi: 10.1016/j.bioelechem.2019.107452. [DOI] [PubMed] [Google Scholar]
- 204.Yang Y., Jiang M., Cao K., Wu M., Zhao C., Li H., Hong C. An electrochemical immunosensor for CEA detection based on Au-Ag/rGO@PDA nanocomposites as integrated double signal amplification strategy. Microchem. J. 2019;151:104223. doi: 10.1016/j.microc.2019.104223. [DOI] [Google Scholar]
- 205.Paimard G., Shahlaei M., Moradipour P., Akbari H., Jafari M., Arkan E. An Impedimetric Immunosensor modified with electrospun core-shell nanofibers for determination of the carcinoma embryonic antigen. Sens. Actuators B Chem. 2020;311:127928. doi: 10.1016/j.snb.2020.127928. [DOI] [Google Scholar]
- 206.Song Y., Li W., Ma C., Sun Y., Qiao J., Li H., Hong C. First use of inorganic copper silicate-transduced enzyme-free electrochemical immunosensor for carcinoembryonic antigen detection. Sens. Actuators B Chem. 2020;319:128311. doi: 10.1016/j.snb.2020.128311. [DOI] [Google Scholar]
- 207.Li J., Liu L., Ai Y., Liu Y., Sun H., Liang Q. Self-Polymerized Dopamine-Decorated Au NPs and Coordinated with Fe-MOF as a Dual Binding Sites and Dual Signal-Amplifying Electrochemical Aptasensor for the Detection of CEA. ACS Appl. Mater. Interfaces. 2020;12:5500–5510. doi: 10.1021/acsami.9b19161. [DOI] [PubMed] [Google Scholar]
- 208.Zhao Y., Cai X., Zhu C., Yang H., Du D. A novel fluorescent and electrochemical dual-responsive immunosensor for sensitive and reliable detection of biomarkers based on cation-exchange reaction. Anal. Chim. Acta. 2020;1096:61–68. doi: 10.1016/j.aca.2019.10.054. [DOI] [PubMed] [Google Scholar]
- 209.Zhao C., Ma C., Wu M., Li W., Song Y., Hong C., Qiao X. A novel electrochemical immunosensor based on CoS2 for early screening of tumor marker carcinoembryonic antigen. New J. Chem. 2020;44:3524–3532. doi: 10.1039/C9NJ05745E. [DOI] [Google Scholar]
- 210.Zheng J., Wang J., Song D., Xu J., Zhang M. Electrochemical Aptasensor of Carcinoembryonic Antigen Based on Concanavalin A-Functionalized Magnetic Copper Silicate Carbon Microtubes and Gold-Nanocluster-Assisted Signal Amplification. ACS Appl. Nano Mater. 2020;3:3449–3458. doi: 10.1021/acsanm.0c00194. [DOI] [Google Scholar]
- 211.Li W., Qiao X., Hong C., Ma C., Song Y. A sandwich-type electrochemical immunosensor for detecting CEA based on CeO2-MoS2 absorbed Pb2+ Anal. Biochem. 2020;592:113566. doi: 10.1016/j.ab.2019.113566. [DOI] [PubMed] [Google Scholar]
- 212.Song Y., Li W., Ma C., Qiao J., Li H., Hong C. The Synergistic Effect of Ferrocene and Cu2 O to Construct a Sandwich-Type Multi-Signal Amplification Ultra-Sensitive Immunosensor for Carcinoembryonic Antigen Detection. J. Electrochem. Soc. 2020;167:027538. doi: 10.1149/1945-7111/ab6c53. [DOI] [Google Scholar]
- 213.He L., Li Z., Guo C., Hu B., Wang M., Zhang Z., Du M. Sensors and Actuators B: Chemical Bifunctional bioplatform based on NiCo Prussian blue analogue: Label-free impedimetric aptasensor for the early detection of carcino-embryonic antigen and living cancer cells. Sens. Actuators B. Chem. 2019;298:126852. doi: 10.1016/j.snb.2019.126852. [DOI] [Google Scholar]
- 214.Lv H., Li Y., Zhang X., Gao Z., Feng J., Wang P., Dong Y. The label-free immunosensor based on rhodium@palladium nanodendrites/sulfo group functionalized multi-walled carbon nanotubes for the sensitive analysis of carcino embryonic antigen. Anal. Chim. Acta. 2018;1007:61–70. doi: 10.1016/j.aca.2017.12.030. [DOI] [PubMed] [Google Scholar]
- 215.Li X., Liu L., Xu Z., Wang W., Shi J., Liu L., Jing M., Li F., Zhang X. Gamma irradiation and microemulsion assisted synthesis of monodisperse flower-like platinum-gold nanoparticles/reduced graphene oxide nanocomposites for ultrasensitive detection of carcinoembryonic antigen. Sens. Actuators B Chem. 2019;287:267–277. doi: 10.1016/j.snb.2019.02.026. [DOI] [Google Scholar]
- 216.Liu J., Shang Y., Zhu Q., Zhang X., Zheng J. A voltammetric immunoassay for the carcinoembryonic antigen using silver(I)-terephthalate metal-organic frameworks containing gold nanoparticles as a signal probe. Microchim. Acta. 2019;186:509. doi: 10.1007/s00604-019-3638-8. [DOI] [PubMed] [Google Scholar]
- 217.Zhao X., Wang W., Liu L., Hu Y., Xu Z., Liu L., Wu N., Li N. Microstructure evolution of sandwich graphite oxide/interlayer-embedded Au nanoparticles induced from γ-rays for carcinoembryonic antigen biosensor. Nanotechnology. 2019;30:495501. doi: 10.1088/1361-6528/ab3e1e. [DOI] [PubMed] [Google Scholar]
- 218.Akbari Nakhjavani S., Afsharan H., Khalilzadeh B., Ghahremani M.H., Carrara S., Omidi Y. Gold and silver bio/nano-hybrids-based electrochemical immunosensor for ultrasensitive detection of carcinoembryonic antigen. Biosens. Bioelectron. 2019;141:111439. doi: 10.1016/j.bios.2019.111439. [DOI] [PubMed] [Google Scholar]
- 219.Suresh L., Bondili J.S., Brahman P.K. Fabrication of Immunosensor Based on Polyaniline, Fullerene-C60 and Palladium Nanoparticles Nanocomposite: An Electrochemical Detection Tool for Prostate Cancer. Electroanalysis. 2020;32:1439–1448. doi: 10.1002/elan.201900659. [DOI] [Google Scholar]
- 220.Xu Q., Jia H., Duan X., Lu L., Tian Q., Chen S., Xu J., Jiang F. Label-free electrochemical immunosensor for the detection of prostate specific antigen based three-dimensional Au nanoparticles/MoS2-graphene aerogels composite. Inorg. Chem. Commun. 2020;119:108122. doi: 10.1016/j.inoche.2020.108122. [DOI] [Google Scholar]
- 221.Liu X., Yue T., Qi K., Qiu Y., Guo X. Porous graphene based electrochemical immunosensor using Cu3(BTC)2 metal-organic framework as nonenzymatic label. Talanta. 2020;217:121042. doi: 10.1016/j.talanta.2020.121042. [DOI] [PubMed] [Google Scholar]
- 222.Medetalibeyoglu H., Kotan G., Atar N., Yola M.L. A novel and ultrasensitive sandwich-type electrochemical immunosensor based on delaminated MXene@AuNPs as signal amplification for prostate specific antigen (PSA) detection and immunosensor validation. Talanta. 2020;220:121403. doi: 10.1016/j.talanta.2020.121403. [DOI] [PubMed] [Google Scholar]
- 223.Hassani S., Maghsoudi A.S., Akmal M.R., Rahmani S., Sarihi P., Ganjali M.R., Norouzi P., Abdollahi M. A sensitive aptamer-based biosensor for electrochemical quantification of PSA as a specific diagnostic marker of prostate cancer. J. Pharm. Pharm. Sci. 2020;23:243–258. doi: 10.18433/jpps31171. [DOI] [PubMed] [Google Scholar]
- 224.Dai Y., Wang X., Zhu X., Liu H., Wang P., Wang X., Zhang S., Sun Y., Gao D., Han R., et al. Electrochemical assays for determination of H2O2 and prostate-specific antigen based on a nanocomposite consisting of CeO2 nanoparticle-decorated MnO2 nanospheres. Microchim. Acta. 2020;187:428. doi: 10.1007/s00604-020-04403-7. [DOI] [PubMed] [Google Scholar]
- 225.Thunkhamrak C., Chuntib P., Ounnunkad K., Banet P., Aubert P.H., Saianand G., Gopalan A.I., Jakmunee J. Highly sensitive voltammetric immunosensor for the detection of prostate specific antigen based on silver nanoprobe assisted graphene oxide modified screen printed carbon electrode. Talanta. 2020;208:120389. doi: 10.1016/j.talanta.2019.120389. [DOI] [PubMed] [Google Scholar]
- 226.Han L., Wang D., Yan L., Petrenko V.A., Liu A. Specific phages-based electrochemical impedimetric immunosensors for label-free and ultrasensitive detection of dual prostate-specific antigens. Sens. Actuators B Chem. 2019;297:126727. doi: 10.1016/j.snb.2019.126727. [DOI] [Google Scholar]
- 227.Ehzari H., Amiri M., Safari M. Enzyme-free sandwich-type electrochemical immunosensor for highly sensitive prostate specific antigen based on conjugation of quantum dots and antibody on surface of modified glassy carbon electrode with core–shell magnetic metal-organic frameworks. Talanta. 2020;210:120641. doi: 10.1016/j.talanta.2019.120641. [DOI] [PubMed] [Google Scholar]
- 228.Suresh L., Bondili J.S., Brahman P.K. Development of proof of concept for prostate cancer detection: An electrochemical immunosensor based on fullerene-C60 and copper nanoparticles composite film as diagnostic tool. Mater. Today Chem. 2020;16:100257. doi: 10.1016/j.mtchem.2020.100257. [DOI] [Google Scholar]
- 229.Soleimani S., Arkan E., Jalalvand A.R., Goicoechea H.C. Fabrication of a novel electrochemical aptasensor assisted by a novel computerized monitoring system for real-time determination of the prostate specific antigen: A computerized experimental method brought elegancy. Microchem. J. 2020;157:104898. doi: 10.1016/j.microc.2020.104898. [DOI] [Google Scholar]
- 230.Ghanavati M., Tadayon F., Bagheri H. A novel label-free impedimetric immunosensor for sensitive detection of prostate specific antigen using Au nanoparticles/MWCNTs- graphene quantum dots nanocomposite. Microchem. J. 2020;159:105301. doi: 10.1016/j.microc.2020.105301. [DOI] [Google Scholar]
- 231.Suhanto R.N., Harimurti S., Septiani N.L.W., Utari L., Anshori I., Wasisto H.S., Suzuki H., Suyatman, Yuliarto B. Sonochemical synthesis of magnetic Fe3O4/graphene nanocomposites for label-free electrochemical biosensors. J. Mater. Sci. Mater. Electron. 2020;31:15381–15393. doi: 10.1007/s10854-020-04102-2. [DOI] [Google Scholar]
- 232.Ibau C., Arshad M.K.M., Gopinath S.C.B., Nuzaihan M., Fathil M.F.M., Shamsuddin S.A. Immunosensing prostate-specific antigen: Faradaic vs non-Faradaic electrochemical impedance spectroscopy analysis on interdigitated microelectrode device. Int. J. Biol. Macromol. 2020;162:1924–1936. doi: 10.1016/j.ijbiomac.2020.08.125. [DOI] [PubMed] [Google Scholar]
- 233.Meng F., Sun H., Huang Y., Tang Y., Chen Q., Miao P. Peptide cleavage-based electrochemical biosensor coupling graphene oxide and silver nanoparticles. Anal. Chim. Acta. 2019;1047:45–51. doi: 10.1016/j.aca.2018.09.053. [DOI] [PubMed] [Google Scholar]
- 234.Sharifuzzaman M., Barman S.C., Rahman M.T., Zahed M.A., Xuan X., Park J.Y. Green Synthesis and Layer-by-Layer Assembly of Amino-Functionalized Graphene Oxide/Carboxylic Surface Modified Trimetallic Nanoparticles Nanocomposite for Label-Free Electrochemical Biosensing. J. Electrochem. Soc. 2019;166:B983–B993. doi: 10.1149/2.0821912jes. [DOI] [Google Scholar]
- 235.Assari P., Rafati A.A., Feizollahi A., Asadpour Joghani R. An electrochemical immunosensor for the prostate specific antigen based on the use of reduced graphene oxide decorated with gold nanoparticles. Microchim. Acta. 2019;186:484. doi: 10.1007/s00604-019-3565-8. [DOI] [PubMed] [Google Scholar]
- 236.Li Z., Yin J., Gao C., Qiu G., Meng A., Li Q. The construction of electrochemical aptasensor based on coral-like poly-aniline and Au nano-particles for the sensitive detection of prostate specific antigen. Sens. Actuators B Chem. 2019;295:93–100. doi: 10.1016/j.snb.2019.05.070. [DOI] [Google Scholar]
- 237.Chen S., Wang Z., Cui X., Jiang L., Zhi Y., Ding X., Nie Z., Zhou P., Cui D. Microfluidic Device Directly Fabricated on Screen-Printed Electrodes for Ultrasensitive Electrochemical Sensing of PSA. Nanoscale Res. Lett. 2019;14:71. doi: 10.1186/s11671-019-2857-6. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 238.Jiang L., Li Y., Gao Z., Wang P., Li D., Dong Y. Sensitive Detection of Prostate Specific Antigen Based on Copper Ions Doped Ag-Au Nanospheres Labeled Immunosensor. J. Electrochem. Soc. 2019;166:B1637–B1643. doi: 10.1149/2.1341915jes. [DOI] [Google Scholar]
- 239.Chen X., Wang Y., Zhang J., Zhang Y. DNA concatemer-silver nanoparticles as a signal probe for electrochemical prostate-specific antigen detection. Analyst. 2019;144:6313–6320. doi: 10.1039/C9AN01484E. [DOI] [PubMed] [Google Scholar]
- 240.Suresh L., Brahman P.K., Reddy K.R., Bondili J.S. Development of an electrochemical immunosensor based on gold nanoparticles incorporated chitosan biopolymer nanocomposite film for the detection of prostate cancer using PSA as biomarker. Enzym. Microb. Technol. 2018;112:43–51. doi: 10.1016/j.enzmictec.2017.10.009. [DOI] [PubMed] [Google Scholar]
- 241.Wei B., Mao K., Liu N., Zhang M., Yang Z. Graphene nanocomposites modified electrochemical aptamer sensor for rapid and highly sensitive detection of prostate specific antigen. Biosens. Bioelectron. 2018;121:41–46. doi: 10.1016/j.bios.2018.08.067. [DOI] [PubMed] [Google Scholar]
- 242.Zhao L., Ma Z. New immunoprobes based on bovine serum albumin-stabilized copper nanoclusters with triple signal amplification for ultrasensitive electrochemical immunosensing for tumor marker. Sens. Actuators B Chem. 2017;241:849–854. doi: 10.1016/j.snb.2016.11.012. [DOI] [Google Scholar]
- 243.Yang Y., Yan Q., Liu Q., Li Y., Liu H., Wang P., Chen L., Zhang D., Li Y., Dong Y. An ultrasensitive sandwich-type electrochemical immunosensor based on the signal amplification strategy of echinoidea-shaped Au@Ag-Cu2O nanoparticles for prostate specific antigen detection. Biosens. Bioelectron. 2018;99:450–457. doi: 10.1016/j.bios.2017.08.018. [DOI] [PubMed] [Google Scholar]
- 244.Samadi Pakchin P., Fathi M., Ghanbari H., Saber R., Omidi Y. A novel electrochemical immunosensor for ultrasensitive detection of CA125 in ovarian cancer. Biosens. Bioelectron. 2020;153:112029. doi: 10.1016/j.bios.2020.112029. [DOI] [PubMed] [Google Scholar]
- 245.Sadasivam M., Sakthivel A., Nagesh N., Hansda S., Veerapandian M., Alwarappan S., Manickam P. Magnetic bead-amplified voltammetric detection for carbohydrate antigen 125 with enzyme labels using aptamer-antigen-antibody sandwiched assay. Sens. Actuators B Chem. 2020;312:127985. doi: 10.1016/j.snb.2020.127985. [DOI] [Google Scholar]
- 246.Huang J., Huang C., Zhong W., Lin Y. A magneto-controlled microfluidic device for voltammetric immunoassay of carbohydrate antigen-125 with silver–polypyrrole nanotags. Anal. Methods. 2020;12:4211–4219. doi: 10.1039/D0AY01225D. [DOI] [Google Scholar]
- 247.Rafique S., Tabassum S., Akram R. Sensitive competitive label-free electrochemical immunosensor for primal detection of ovarian cancer. Chem. Pap. 2020;74:2591–2603. doi: 10.1007/s11696-020-01100-w. [DOI] [Google Scholar]
- 248.Li H., Qin J., Li M., Li C., Xu S., Qian L., Yang B. Sensors and Actuators B: Chemical Gold-nanoparticle-decorated boron-doped graphene/BDD electrode for tumor marker sensor. Sens. Actuators B. Chem. 2020;302:127209. doi: 10.1016/j.snb.2019.127209. [DOI] [Google Scholar]
- 249.Fan Y., Shi S., Ma J., Guo Y. A paper-based electrochemical immunosensor with reduced graphene oxide/thionine/gold nanoparticles nanocomposites modification for the detection of cancer antigen 125. Biosens. Bioelectron. 2019;135:1–7. doi: 10.1016/j.bios.2019.03.063. [DOI] [PubMed] [Google Scholar]
- 250.Jafari M., Hasanzadeh M., Solhi E., Hassanpour S., Shadjou N., Mokhtarzadeh A., Jouyban A., Mahboob S. Ultrasensitive bioassay of epitope of Mucin-16 protein (CA 125) in human plasma samples using a novel immunoassay based on silver conductive nano-ink: A new platform in early stage diagnosis of ovarian cancer and efficient management. Int. J. Biol. Macromol. 2019;126:1255–1265. doi: 10.1016/j.ijbiomac.2019.01.009. [DOI] [PubMed] [Google Scholar]
- 251.Liang X., Han H., Ma Z. pH responsive amperometric immunoassay for carcinoma antigen 125 based on hollow polydopamine encapsulating methylene blue. Sens. Actuators B Chem. 2019;290:625–630. doi: 10.1016/j.snb.2019.04.030. [DOI] [Google Scholar]
- 252.Yu L., Cui X., Li H., Lu J., Kang Q., Shen D. A ratiometric electrochemical sensor for multiplex detection of cancer biomarkers using bismuth as an internal reference and metal sulfide nanoparticles as signal tags. Analyst. 2019;144:4073–4080. doi: 10.1039/C9AN00775J. [DOI] [PubMed] [Google Scholar]
- 253.Baradoke A., Jose B., Pauliukaite R., Forster R.J. Properties of Anti-CA125 antibody layers on screen-printed carbon electrodes modified by gold and platinum nanostructures. Electrochim. Acta. 2019;306:299–306. doi: 10.1016/j.electacta.2019.03.081. [DOI] [Google Scholar]
- 254.Saadati A., Hassanpour S., Bahavarnia F., Hasanzadeh M. A novel biosensor for the monitoring of ovarian cancer tumor protein CA 125 in untreated human plasma samples using a novel nano-ink: A new platform for efficient diagnosis of cancer using paper based microfluidic technology. Anal. Methods. 2020;12:1639–1649. doi: 10.1039/D0AY00299B. [DOI] [Google Scholar]
- 255.Raghav R., Srivastava S. Core-shell gold-silver nanoparticles based impedimetric immunosensor for cancer antigen CA125. Sens. Actuators B Chem. 2015;220:557–564. doi: 10.1016/j.snb.2015.05.108. [DOI] [Google Scholar]
- 256.Ren X., Wang H., Wu D., Fan D., Zhang Y., Du B., Wei Q. Talanta Ultrasensitive immunoassay for CA125 detection using acid site com- pound as signal and enhancer. Talanta. 2015;144:535–541. doi: 10.1016/j.talanta.2015.06.086. [DOI] [PubMed] [Google Scholar]
- 257.Samadi P., Ghanbari H., Saber R., Omidi Y. Biosensors and Bioelectronics Electrochemical immunosensor based on chitosan-gold nanoparticle/carbon nanotube as a platform and lactate oxidase as a label for detection of CA125 oncomarker. Biosens. Bioelectron. 2018;122:68–74. doi: 10.1016/j.bios.2018.09.016. [DOI] [PubMed] [Google Scholar]
- 258.Kumar N., Sharma S., Nara S. Dual gold nanostructure-based electrochemical immunosensor for CA125 detection. Appl. Nanosci. 2018;8:1843–1853. doi: 10.1007/s13204-018-0857-y. [DOI] [Google Scholar]
- 259.Hasanzadeh M., Mohammadzadeh A., Jafari M., Habibi B. Ultrasensitive immunoassay of glycoprotein 125 (CA 125) in untreated human plasma samples using poly (CTAB-chitosan) doped with silver nanoparticles. Int. J. Biol. Macromol. 2018;120:2048–2064. doi: 10.1016/j.ijbiomac.2018.09.208. [DOI] [PubMed] [Google Scholar]
- 260.Hasanzadeh M., Sahmani R., Solhi E., Mokhtarzadeh A., Shadjou N., Mahboob S. Ultrasensitive immunoassay of carcinoma antigen 125 in untreated human plasma samples using gold nanoparticles with flower like morphology: A new platform in early stage diagnosis of ovarian cancer and efficient management. Int. J. Biol. Macromol. 2018;119:913–925. doi: 10.1016/j.ijbiomac.2018.08.008. [DOI] [PubMed] [Google Scholar]
- 261.Hartati Y.W., Letelay L.K., Gaffar S., Wyantuti S., Bahti H.H. Cerium oxide-monoclonal antibody bioconjugate for electrochemical immunosensing of HER2 as a breast cancer biomarker. Sens. Bio Sens. Res. 2020;27:100316. doi: 10.1016/j.sbsr.2019.100316. [DOI] [Google Scholar]
- 262.Freitas M., Nouws H.P.A., Keating E., Fernandes V.C., Delerue-Matos C. Immunomagnetic bead-based bioassay for the voltammetric analysis of the breast cancer biomarker HER2-ECD and tumour cells using quantum dots as detection labels. Mikrochim. Acta. 2020;187:184. doi: 10.1007/s00604-020-4156-4. [DOI] [PubMed] [Google Scholar]
- 263.Zhou N., Su F., Li Z., Yan X., Zhang C., Hu B., He L., Wang M., Zhang Z. Gold nanoparticles conjugated to bimetallic manganese(II) and iron(II) Prussian Blue analogues for aptamer-based impedimetric determination of the human epidermal growth factor receptor-2 and living MCF-7 cells. Microchim. Acta. 2019;186:3–12. doi: 10.1007/s00604-018-3184-9. [DOI] [PubMed] [Google Scholar]
- 264.Freitas M., Neves M.M.P.S., Nouws H.P.A., Delerue-Matos C. Quantum dots as nanolabels for breast cancer biomarker HER2-ECD analysis in human serum. Talanta. 2020;208:120430. doi: 10.1016/j.talanta.2019.120430. [DOI] [PubMed] [Google Scholar]
- 265.Gu C., Guo C., Li Z., Wang M., Zhou N., He L., Zhang Z., Du M. Bimetallic ZrHf-based metal-organic framework embedded with carbon dots: Ultra-sensitive platform for early diagnosis of HER2 and HER2-overexpressed living cancer cells. Biosens. Bioelectron. 2019;134:8–15. doi: 10.1016/j.bios.2019.03.043. [DOI] [PubMed] [Google Scholar]
- 266.Lah Z.M.A.N.H., Ahmad S.A.A., Zaini M.S., Kamarudin M.A. An Electrochemical Sandwich Immunosensor for the Detection of HER2 using Antibody-Conjugated PbS Quantum Dot as a label. J. Pharm. Biomed. Anal. 2019;174:608–617. doi: 10.1016/j.jpba.2019.06.024. [DOI] [PubMed] [Google Scholar]
- 267.Rostamabadi P.F., Heydari-Bafrooei E. Impedimetric aptasensing of the breast cancer biomarker HER2 using a glassy carbon electrode modified with gold nanoparticles in a composite consisting of electrochemically reduced graphene oxide and single-walled carbon nanotubes. Microchim. Acta. 2019;186:495. doi: 10.1007/s00604-019-3619-y. [DOI] [PubMed] [Google Scholar]
- 268.Chen D., Wang D., Hu X., Long G., Zhang Y., Zhou L. A DNA nanostructured biosensor for electrochemical analysis of HER2 using bioconjugate of GNR@Pd SSs—Apt—HRP. Sens. Actuators B Chem. 2019;296:126650. doi: 10.1016/j.snb.2019.126650. [DOI] [Google Scholar]
- 269.Augustine S., Kumar P., Malhotra B.D. Amine-Functionalized MoO3@RGO Nanohybrid-Based Biosensor for Breast Cancer Detection. ACS Appl. Bio Mater. 2019;2:5366–5378. doi: 10.1021/acsabm.9b00659. [DOI] [PubMed] [Google Scholar]
- 270.Li X., Shen C., Yang M., Rasooly A. Polycytosine DNA Electric-Current-Generated Immunosensor for Electrochemical Detection of Human Epidermal Growth Factor Receptor 2 (HER2) Anal. Chem. 2018;90:4764–4769. doi: 10.1021/acs.analchem.8b00023. [DOI] [PubMed] [Google Scholar]
- 271.Shamsipur M., Emami M., Farzin L., Saber R. Biosensors and Bioelectronics A sandwich-type electrochemical immunosensor based on in situ silver deposition for determination of serum level of HER2 in breast cancer patients. Biosens. Bioelectron. 2018;103:54–61. doi: 10.1016/j.bios.2017.12.022. [DOI] [PubMed] [Google Scholar]
- 272.Jothi L., Jaganathan S.K., Nageswaran G. An electrodeposited Au nanoparticle/porous graphene nanoribbon composite for electrochemical detection of alpha-fetoprotein. Mater. Chem. Phys. 2020;242:122514. doi: 10.1016/j.matchemphys.2019.122514. [DOI] [Google Scholar]
- 273.Li G., Li W., Li S., Shi X., Liang J., Lai J., Zhou Z. Regular article A novel aptasensor based on light-addressable potentiometric sensor for the determination of Alpha-fetoprotein. Biochem. Eng. J. 2020;164:107780. doi: 10.1016/j.bej.2020.107780. [DOI] [Google Scholar]
- 274.Yang X., Zhao C., Zhang C., Wen K., Zhu Y. Bi-directionally amplified ratiometric electrochemical aptasensor for the ultrasensitive detection of alpha-fetoprotein. Sens. Actuators B Chem. 2020;323:128666. doi: 10.1016/j.snb.2020.128666. [DOI] [Google Scholar]
- 275.Gu C., Peng Y., Li J., Liu C.S., Pang H. Controllable synthesis of copper ion guided MIL-96 octadecahedron: Highly sensitive aptasensor toward alpha-fetoprotein. Appl. Mater. Today. 2020;20:100745. doi: 10.1016/j.apmt.2020.100745. [DOI] [Google Scholar]
- 276.Zhang K., Cao Z., Wang S., Chen J., Wei Y., Feng D. A novel sandwich-type electrochemical immunosensor based on the signal amplification strategy of core-shell Pd@Pt nanoparticles for α-fetoprotein detection. Int. J. Electrochem. Sci. 2020;15:2604–2613. doi: 10.20964/2020.03.32. [DOI] [Google Scholar]
- 277.Chellachamy Anbalagan A., Sawant S. Polyaniline Based Electrochemical Biosensor for α-fetoprotein Detection Using Bio-functionalized Nanoparticles as Detection Probe. Electroanalysis. 2020;32:2415–2421. doi: 10.1002/elan.202060219. [DOI] [Google Scholar]
- 278.Wu M., Yang Y., Cao K., Zhao C., Qiao X., Hong C. Microwave-assisted preparation of ZnFe2O4-Ag/rGO nanocomposites for amplification signal detection of alpha-fetoprotein. Bioelectrochemistry. 2020;132:107434. doi: 10.1016/j.bioelechem.2019.107434. [DOI] [PubMed] [Google Scholar]
- 279.Wang A.J., Zhu X.Y., Chen Y., Yuan P.X., Luo X., Feng J.J. A label-free electrochemical immunosensor based on rhombic dodecahedral Cu3Pt nanoframes with advanced oxygen reduction performance for highly sensitive alpha-fetoprotein detection. Sens. Actuators B Chem. 2019;288:721–727. doi: 10.1016/j.snb.2019.03.061. [DOI] [Google Scholar]
- 280.Sun D., Li H., Li M., Li C., Qian L., Yang B. Electrochemical immunosensors with AuPt-vertical graphene/glassy carbon electrode for alpha-fetoprotein detection based on label-free and sandwich-type strategies. Biosens. Bioelectron. 2019;132:68–75. doi: 10.1016/j.bios.2019.02.045. [DOI] [PubMed] [Google Scholar]
- 281.Zhang S., Zhang C., Jia Y., Zhang X., Dong Y., Li X., Liu Q., Li Y., Zhao Z. Sandwich-type electrochemical immunosensor based on Au@Pt DNRs/NH2 -MoSe2 NSs nanocomposite as signal amplifiers for the sensitive detection of alpha-fetoprotein. Bioelectrochemistry. 2019;128:140–147. doi: 10.1016/j.bioelechem.2019.03.012. [DOI] [PubMed] [Google Scholar]
- 282.Tertis M., Leva P.I., Bogdan D., Suciu M., Graur F., Cristea C. Impedimetric aptasensor for the label-free and selective detection of Interleukin-6 for colorectal cancer screening. Biosens. Bioelectron. 2019;137:123–132. doi: 10.1016/j.bios.2019.05.012. [DOI] [PubMed] [Google Scholar]
- 283.Tertiş M., Ciui B., Suciu M., Săndulescu R., Cristea C. Label-free electrochemical aptasensor based on gold and polypyrrole nanoparticles for interleukin 6 detection. Electrochim. Acta. 2017;258:1208–1218. doi: 10.1016/j.electacta.2017.11.176. [DOI] [Google Scholar]
- 284.Pachauri N., Lakshmi G.B.V.S., Sri S., Gupta P.K., Solanki P.R. Silver molybdate nanoparticles based immunosensor for the non-invasive detection of Interleukin-8 biomarker. Mater. Sci. Eng. C. 2020;113:110911. doi: 10.1016/j.msec.2020.110911. [DOI] [PubMed] [Google Scholar]
- 285.Xu J., Yu X., Xie L., Shao M. Facile incorporation of DNA-templated quantum dots for sensitive electrochemical detection of the oral cancer biomarker interleukin-8. Anal. Bioanal. Chem. 2020;412:2599–2606. doi: 10.1007/s00216-020-02487-x. [DOI] [PubMed] [Google Scholar]
- 286.Verma S., Singh A., Shukla A., Kaswan J., Arora K., Ramirez-Vick J., Singh P., Singh S.P. Anti-IL8/AuNPs-rGO/ITO as an Immunosensing Platform for Noninvasive Electrochemical Detection of Oral Cancer. ACS Appl. Mater. Interfaces. 2017;9:27462–27474. doi: 10.1021/acsami.7b06839. [DOI] [PubMed] [Google Scholar]
- 287.Yuan J., Duan R., Yang H., Luo X., Xi M. Detection of serum human epididymis secretory protein 4 in patients with ovarian cancer using a label-free biosensor based on localized surface plasmon resonance. Int. J. Nanomed. 2012;7:2921–2928. doi: 10.2147/IJN.S32641. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 288.Eletxigerra U., Martinez-Perdiguero J., Barderas R., Pingarrón J.M., Campuzano S., Merino S. Surface plasmon resonance immunosensor for ErbB2 breast cancer biomarker determination in human serum and raw cancer cell lysates. Anal. Chim. Acta. 2016;905:156–162. doi: 10.1016/j.aca.2015.12.020. [DOI] [PubMed] [Google Scholar]
- 289.Brocker C., Carpenter C., Nebert D.W., Vasiliou V. Evolutionary divergence and functions of the human acyl-CoA thioesterase gene (ACOT) family. Hum. Genom. 2010;4:411–420. doi: 10.1186/1479-7364-4-6-411. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 290.Khan M.A., Mujahid M. Recent advances in electrochemical and optical biosensors designed for detection of Interleukin 6. Sensors. 2020;20:646. doi: 10.3390/s20030646. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 291.Aydın E.B., Sezgintürk M.K. A disposable and ultrasensitive ITO based biosensor modified by 6-phosphonohexanoic acid for electrochemical sensing of IL-1β in human serum and saliva. Anal. Chim. Acta. 2018;1039:41–50. doi: 10.1016/j.aca.2018.07.055. [DOI] [PubMed] [Google Scholar]
- 292.Ji T., Liu Z., Wang G.Q., Guo X., Akbar khan S., Lai C., Chen H., Huang S., Xia S., Chen B., et al. Detection of COVID-19: A review of the current literature and future perspectives. Biosens. Bioelectron. 2020;166:112455. doi: 10.1016/j.bios.2020.112455. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 293.Guan W.J., Liang W.H., Zhao Y., Liang H.R., Chen Z.S., Li Y.M., Liu X.Q., Chen R.C., Tang C.L., Wang T., et al. Comorbidity and its impact on 1,590 patients with Covid-19 in China: A nationwide analysis. Eur. Respir. J. 2020:55. doi: 10.1183/13993003.00547-2020. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 294.Driggin E., Madhavan M.V., Bikdeli B., Chuich T., Laracy J., Biondi-Zoccai G., Brown T.S., Der Nigoghossian C., Zidar D.A., Haythe J., et al. Cardiovascular Considerations for Patients, Health Care Workers, and Health Systems During the COVID-19 Pandemic. J. Am. Coll. Cardiol. 2020;75:2352–2371. doi: 10.1016/j.jacc.2020.03.031. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 295.Chandra P. Miniaturized label-free smartphone assisted electrochemical sensing approach for personalized COVID-19 diagnosis. Sens. Int. 2020;1:100019. doi: 10.1016/j.sintl.2020.100019. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 296.Zhao H., Liu F., Xie W., Zhou T.C., OuYang J., Jin L., Li H., Zhao C.Y., Zhang L., Wei J., et al. Ultrasensitive supersandwich-type electrochemical sensor for SARS-CoV-2 from the infected COVID-19 patients using a smartphone. Sens. Actuators B Chem. 2021;327:128899. doi: 10.1016/j.snb.2020.128899. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 297.Menon S., Mathew M.R., Sam S., Keerthi K., Kumar K.G. Recent advances and challenges in electrochemical biosensors for emerging and re-emerging infectious diseases. J. Electroanal. Chem. 2020;878:114596. doi: 10.1016/j.jelechem.2020.114596. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 298.Ji D., Liu L., Li S., Chen C., Lu Y., Wu J., Liu Q. Biosensors and Bioelectronics Smartphone-based cyclic voltammetry system with graphene modi fi ed screen printed electrodes for glucose detection. Biosens. Bioelectron. 2017;98:449–456. doi: 10.1016/j.bios.2017.07.027. [DOI] [PubMed] [Google Scholar]
- 299.Chen C.H., Lu Y., Sin M.L.Y., Mach K.E., Zhang D.D., Gau V., Liao J.C., Wong P.K. Antimicrobial susceptibility testing using high surface-to-volume ratio microchannels. Anal. Chem. 2010;82:1012–1019. doi: 10.1021/ac9022764. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 300.Mach K.E., Du C.B., Phull H., Haake D.A., Shih M.C., Baron E.J., Liao J.C. Multiplex Pathogen Identification for Polymicrobial Urinary Tract Infections Using Biosensor Technology: A Prospective Clinical Study. J. Urol. 2009;192:2735–2741. doi: 10.1016/j.juro.2009.08.028. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 301.Feng Y., Zhou D., Gao L., He F. Electrochemical biosensor for rapid detection of bacteria based on facile synthesis of silver wire across electrodes. Biosens. Bioelectron. 2020;168:112527. doi: 10.1016/j.bios.2020.112527. [DOI] [PubMed] [Google Scholar]
- 302.Khan N.I., Song E. Lab-on-a-chip systems for aptamer-based biosensing. Micromachines. 2020;11:220. doi: 10.3390/mi11020220. [DOI] [PMC free article] [PubMed] [Google Scholar]