Abstract
G nerve agents are colorless, odorless, and lethal chemical warfare agents (CWAs). The threat of CWAs, which cause critical damage to humans, continues to exist, e.g., in warfare or terrorist attacks. Therefore, it is important to be able to detect these agents rapidly and with a high degree of sensitivity. In this study, a surface acoustic wave (SAW) array device with three SAW sensors coated with different sensing materials and one uncoated sensor was tested to determine the most suitable material for the detection of nerve agents and related simulants. The three materials used were polyhedral oligomeric silsesquioxane (POSS), 1-benzyl-3-phenylthiourea (TU-1), and 1-ethyl-3-(4-fluorobenzyl) thiourea (TU-2). The SAW sensor coated with the POSS-based polymer showed the highest sensitivity and the fastest response time at concentrations below the median lethal concentration (LCt50) for tabun (GA) and sarin (GB). Also, it maintained good performance over the 180 days of exposure tests for dimethyl methylphosphonate (DMMP). A comparison of the sensitivities of analyte vapors also confirmed that the sensitivity for DMMP was similar to that for GB. Considering that DMMP is a simulant which physically and chemically resembles GB, the sensitivity to a real agent of the sensor coated with POSS could be predicted. Therefore, POSS, which has strong hydrogen bond acid properties and which showed similar reaction characteristics between the simulant and the nerve agent, can be considered a suitable material for nerve agent detection.
Keywords: SAW array sensor, chemical warfare agent, surface acoustic wave, polyhedral oligomeric silsesquioxane, dimethyl methylphosphonate
1. Introduction
Chemical Warfare Agents (CWAs) are classified into nerve, blood, choking, and blister agents according to the effects they have on the body and their intended use. Nerve agents such as sarin (GA), tabun (GB), soman (GD), and VX are among the most toxic, and usually contain organophosphates [1,2,3]. The toxicity of these nerve agents can be quantitatively expressed using the median lethal concentration (LCt50) expressed in . The LCt50 of GA is 400 , and that of GB is 100 ; exposure at these concentrations can cause serious damage to humans [1]. The production and use of CWAs are limited by the Organization for the Prohibition of Chemical Weapons (OPCW), but these compounds could potentially still be used in warfare and terrorism, posing a threat not only to military personnel, but also to civilians [1]. Therefore, early detection is key. Generally, nerve agents can be detected using various techniques such as photoionization [4], ion mobility spectroscopy [5,6], an electric nose [7,8], flame photometry [9,10,11], and IR spectroscopy [12,13], but most detectors using these methods are heavy, bulky, and consume lots of power.
Surface acoustic wave (SAW) sensors, a kind of the chemical detector, are sensitive, small, and require extremely low power to operate [14,15,16,17,18]. As such, they have great potential for use as handheld detectors. The sensing material on the surface of the delay line of the SAW sensor is one of the most important components to selectively detect CWA vapors, because it has a large impact on the mass loading effect or conductivity changes, which are the prime sensing principles of SAW sensors [19,20]. Hydrogen bond acidic polymers such as fluoropolyol, fluoroalcohol polysiloxane, and fluoroalcoholic linear polysiloxane have been utilized as sensing materials for the detection of nerve agents in security and defense applications because they can selectively interact with the target analytes by hydrogen bonding [21]. These hydrogen-bond acidic polymers contain hydroxyl derivatives including alcohol, carboxylic acid, or phenolic groups, and the addition of electron withdrawing halogens in close proximity to hydroxyl groups yields a dramatic increase in hydrogen bond acidity with a concomitant reduction in hydrogen bond basicity [21,22,23,24].
In this paper, SAW sensors coated with polyhedral oligomeric silsesquioxane (POSS), 1-benzyl-3-phenylthiourea (TU-1), and 1-ethyl-3-(4-fluorobenzyl)thiourea (TU-2), synthesized by Jang [25] and Song [26] in a laboratory in Korea, were tested to investigate their sensitivity to GA, GB, and dimethyl methylphosphonate (DMMP). These sensors were tested simultaneously using a portable SAW array device which included one uncoated SAW sensor to reduce the effect of moisture and three coated SAW sensors under the same environmental conditions.
2. Experiment
2.1. Design of SAW Array Sensor
The SAW sensor module is shown in Figure 1a. The SAW sensors are based on piezoelectric materials, influencing their characteristics. The most commonly used piezoelectric materials for SAW sensors are quartz, lithium niobate (LiNbO3), and lithium tantalate (LiTaO3) [27]. Different wave types are obtained depending on the piezoelectric material, crystal cut, and temperature dependence [28]. While LiNbO3 and LiTaO3 show higher frequency change by temperature variation, ST-cut quartz has the advantage of thermal stability. The sensitivity of the sensor based on the choice of piezoelectric material is affected by , which is the center frequency of the SAW sensor, where , , and are wavelength, SAW velocity, and center frequency, respectively [29,30,31]. Figure 1b,c show optical microscope images of the SAW delay line. The SAW sensors were fabricated on a ST-cut quartz substrate having an aluminum interdigit transducer (Al-IDT) structure designed for a 250-MHz center frequency. Al electrodes were deposited on the substrate and titanium was used as an adhesion metal for the deposition of the IDTs. The center frequency was calculated as 250 MHz because the SAW velocity on the ST-cut quartz substrate was 3158 and the wavelength of the Al electrode was 12.632 .
Figure 1.
Optical microscope image of the IDT on a SAW sensor module: (a) SAW sensor module; (b) IDT pattern; (c) Design parameter of IDT.
The SAW array device was constructed in a four-channel configuration consisting of reference and sensing channels, as shown in Figure 2a. The uncoated SAW sensor improved the performance of the SAW array device by compensating for the frequency shift caused by environmental effects. Each of the three sensors was coated with different sensing materials on the delay lines. Each channel to output the frequency signal was made up of a low pass filter (LPF), an amplifier, and a SAW sensor. The coated SAW sensors from channels 1–3 are for detecting vapors, while the uncoated sensor on channel 4 is a reference sensor. The frequency signals of the coated SAW sensors are cross-coupled with the signal of the reference sensor to transmit the difference between the two signals [32]. Then, real-time measurement signals generated by each channel are transmitted to the display via the multiplexer (MUX).
Figure 2.
Four-channel SAW monitoring system: (a) Signal processing diagram; (b) Structure of the SAW array device with a portable battery. The insertion numbers indicate four different SAW sensors.
Figure 2b shows the relevant circuit-board design with the portable battery, which is divided into two parts: the oscillator board and the digital-signal processing board. The fourth sensor is the reference sensor and the others are coated sensors. Each sensor is mounted on each channel on the oscillator board. The board with mounted SAW sensors is connected to the digital-signal processing board. This SAW array device is powered by connecting the portable battery.
2.2. Materials Synthesis and Coating Process
POSS and thiourea (TU) -based polymers are used as sensing materials to detect GA, GB, and D MMP (Sigma Aldrich, MO, USA), as shown in Figure 3. GA and GB were supplied by the Netherlands Organization for applied scientific research (TNO, Rijswijk, Netherland). These were obtained from previously synthesized stocks of TNO. The POSS-based polymer was synthesized using 4-(Trifluoromethyl) phenol and PSS-Octakis (dimethylsilyloxy) substituted through Karstedt’s catalyst [25]. TU-1 was synthesized using a TU-based polymer and N-benzyl groups attached to both sides of the TU base. The N-benzyl groups successfully reduced the self-aggregation of TU molecules, and the additional CH − π interaction strengthened the binding of TU and organophosphonate. TU-2 was synthesized using a TU-based polymer and 4-fluoro-benzyl, as this enhances the hydro bonding by altering the electron density of the N-benzyl group via electron-withdrawing [26].
Figure 3.
Chemical structure of the sensing materials: (a) POSS; (b) TU-1; and (c) TU-2 [32].
The POSS and TU-based polymers were diluted to a 1:1 concentration using ethanol and dimethylformamide as solvents, respectively. These three solutions were dispersed with a sonicator for 1 h, and then 0.5 of these solutions was coated on the surface of the delay line of the SAW sensors by a drop coating process. Finally, these sensors were heated to 60 °C for 2 h in a convection oven [32].
2.3. Experimental Apparatus
This experiment was conducted in two ways, depending on the detection of nerve agents or the simulant, because nerve agents cannot be handled in a typical laboratory. The experiment detecting nerve agents took place at the High Tox laboratory of TNO, while experiments detecting the simulant were conducted in our laboratory in South Korea. The SAW array device was evaluated in a test facility at TNO that has a dynamic vapor generation system with a canister, as shown in Figure 4. The needle of the syringe filled with the agents was inserted into the heated injector (140 °C for GB and 175 °C for GA) of a heat chamber. The agents were evaporated in the injector and mixed with nitrogen () gas delivered at a flow rate or 1 in the heat chamber. A set of valves controlled the exposure condition of the SAW array device. Figure 4a shows the main mechanism of the experiment when the device was exposed to CWAs. The vapor concentration of the CWAs was controlled by changing the volume flow rate of the syringe driver. The flow rate of GA was changed from 10 to 100 , and that of GB from 5 to 500 . After the exposure time, the agent was released to the outside through the canister, and pure gas was injected into the SAW array device as shown Figure 4b.
Figure 4.
Schematic diagram of the experimental setup for testing the SAW array device with nerve agents (TNO, Netherland): (a) exposure to G nerve agents; (b) exposure to gas for purging.
A gas feeding system was used to test the SAW array device shown in Figure 5. It consisted of stainless-steel tubes through which the gases flowed, check valves that allowed vapor to flow in only one direction, two mass-flow controllers (KOFLOG, Japan) to control the concentration of the gases, and a readout unit. A bubbler was mounted on the carrier gas line, and the target material was maintained at 25 °C in a two-necked-flask using a constant-temperature water bath. gas was used as both the dilution and carrier gas. The gas entered the two-necked flask at a constant flow rate maintained by the mass-flow controllers, and DMMP, the target material, was released into the vapor by the vapor pressure [30,32].
Figure 5.
Schematic diagram of the in house experimental setup for testing the SAW array device with a simulant, DMMP (Inha University, South Korea).
3. Results
3.1. Sensitivity to GA Vapor
The response of the SAW array device to GA in the range of 50–100 (7.5–15 ppm) is shown in Figure 6. Upon exposure to GA vapor, the frequency response immediately changed, showing a saturated signal after a constant frequency shift according to the vapor concentration. The frequency shift of the SAW-1 showed a change of 11.0 kHz for GA at a concentration of 50 . There was some data loss, shown in Figure 6 as a dash line, in the initial response at 50 , but this did not affect the determination of an equilibrium state of adsorption. The 80% response time, T80, was defined as the time interval for the frequency shift reaching 80% of the maximum frequency to reduce the effect of noise from the equilibrium state of the sensors. T80 was about 50 s for GA at 100 , and finally achieved a maximum frequency of 18.9 kHz. The frequency shifts of the SAW-2 and SAW-3 changed to 1.7 kHz and 2.8 kHz at a concentration of 50 , respectively, which represented frequency shifts which were lower than the SAW-1. The SAW-2 or SAW-3 showed unstable signals in the equilibrium state upon exposure to a concentration of 100 due to their small frequency shifts relative to noise.
Figure 6.
Dynamic response of the SAW sensors coated with different sensing materials: (a) POSS; (b) TU-1; and (c) TU-2 upon exposure to GA concentrations ranging from 50 to 100 . The inserted numbers are the different concentrations of GA.
3.2. Sensitivity to GB Vapor
Figure 7 describes the response of each SAW sensor upon exposure to GB in the range of 5.5–550 (0.95–96 ppm). The SAW-1 showed a frequency shift of 0.495 kHz at the lowest concentration of 5.5 which increased with increasing concentration. The frequency shift was represented as 13.7 kHz at 550 , and the , an average of , was 45 s in the range of concentration. The SAW-2 indicated no significant frequency shift up to 55 . The frequency shift of the sensor was 0.59 kHz at a concentration of 110 , which was the limit of detection of SAW-2, i.e., 4.3 kHz at 550 . The SAW-3 showed a frequency shift of 0.39 kHz from the lowest concentration but could not distinguish the signals at concentrations below 110 . The observed frequency shifts were 0.64 kHz for 110 and 6.1 kHz for 550 . It could be assumed that the reaction patterns of SAW-2 and SAW-3 would be similar upon exposure to GB because the sensors were made using the same TU-based polymer. The response patterns presented a spike because the sensors responded slowly to GB, i.e., they needed about 3 min to reach an equilibrium state. The baseline shifts of each sensor were caused by slower desorption than sorption in the concentration range over 110 .
Figure 7.
Dynamic response of the SAW sensors coated with different sensing materials: (a) POSS; (b) TU-1; and (c) TU-2 upon exposure to GB at concentrations from 5.5 to 550 . The inserted numbers are the different concentrations of GB.
3.3. Sensitivity to DMMP Vapor
The dynamic responses of three SAW sensors coated with POSS, TU-1, and TU-2 to DMMP in a concentration range of 63 to 507 (12.5–100 ppm) are shown in Figure 8. One test cycle consisted of an exposure time of 3 min to DMMP vapor and a purging time of 5 min using 99.9% gas. The same test cycle was carried out once again to confirm the reproducibility of the sensors to exposure to DMMP vapor. The SAW-1 presented a frequency shift of 12.6 kHz at 507 and a stable frequency shift of 4.1 kHz at 63 . The was 22 s at high concentration (507 and 381 ), but slowed to about 51 s at low concentration (254, 127, and 63 ). The frequency shifts of the SAW-2 and the SAW-3 were 14.6 kHz and 2.5 kHz at 507 , respectively, and were unstable at 63 . The frequency shift of the SAW-2 was greater than that of SAW-1 at the highest concentration, but lower than that of SAW-1 with decreasing concentrations of DMMP. The of SAW-2 and SAW-3 was approximately 83 s and 88 s, respectively.
Figure 8.
Dynamic response of the SAW sensors coated with different sensing materials: (a) POSS; (b) TU-1; and (c) TU-2 under the exposure to DMMP concentrations ranging from 63 to 507 . The inserted numbers are the different concentrations of DMMP.
3.4. Regression to Frequency Response of the Sensors
The difference of frequency, , in kHz before and after exposure to vapors can be expressed as a function of concentration, , in of the vapor types, as described in Figure 9. Table 1 shows the slope, intercept, and R-square of these functions. All correlation coefficients were above 0.9. The slopes of GB and DMMP similarly appeared for SAW-1, which had high sensitivity with low concentrations of GA. The reactions to GB and DMMP presented a different trend, even though SAW-2 and SAW-3 were made by coating the sensing materials with the same TU-based polymer. The slopes of GB were smaller for SAW-2 and larger for SAW-3 than the slope of DMMP. Also, SAW-2 had the lowest sensitivity to the vapors except for DMMP within the overall concentration range.
Figure 9.
Response of the SAW sensors with different sensing materials of (a) POSS; (b) TU-1; and (c) TU-2 depending on the vapor type.
Table 1.
Linear parameters of the function : a, b, and R2 are slope, intercept, and coefficient of determination, respectively.
a | b | R 2 | |
---|---|---|---|
SAW-1 (POSS) | |||
GA | 0.157 | 3.118 | - |
GB | 0.024 | 0.803 | 0.984 |
DMMP | 0.018 | 3.906 | 0.957 |
SAW-2 (TU-1) | |||
GA | 0.046 | −0.563 | - |
GB | 0.008 | −0.157 | 0.991 |
DMMP | 0.031 | −1.138 | 0.999 |
SAW-3 (TU-2) | |||
GA | 0.079 | −1.527 | - |
GB | 0.011 | 0.048 | 0.972 |
DMMP | 0.004 | 0.485 | 0.935 |
The SAW sensor detected the G nerve agents and their simulants by changing the hydrogen bond acidity of the polymer to maximize the mass loading effect [33]. A particularly effective means of enhancing the hydrogen-bond acidity of the polymer is to incorporate fluorinated alcohol or fluorinated phenolic functional groups as substituents in the polymer structure [33,34,35]. It seems that the fluorinated phenolic functional groups of the POSS-based polymer reacted with the phosphonate oxygen group in GA, GB, or DMMP to form hydrogen bonds [25]. In contrast, TU-based polymers form hydrogen bonds between the NH groups of the polymers and the phosphonate oxygen groups to detect G nerve agents and their simulants [26]. Although all three sensing materials formed hydrogen bonds with the analytes, it was expected that the POSS-based polymer would show higher sensitivity than the other sensing materials because its fluorinated phenolic group maximizes the hydrogen bond acidity of the polymer by the electron-withdrawing effect of the fluorine atoms [33].
From the comparison of these sensitivities to the analyte vapors, we also confirmed that the frequency response to DMMP was similar to that for GB. DMMP is a simulant which physically and chemically resembles GB [36].
3.5. Long-Term SAW Response: Reliability of the POSS as a Sensing Material
Maintaining the performance of sensors for a long period is related to the reliability of the sensor. The stability of the SAW sensor coated with POSS was evaluated by assessing the increase of concentration at 0, 90, and 180 days after fabrication, as shown in Figure 10. One microliter of a mixture of POSS and ethanol at a ratio of 1:1 (mg:mL) was coated onto the 250 MHz SAW sensors. The response was measured using a vector network analyzer (MS46122A, Anritsu). The sensors were tested in a DMMP concentration range of 127–507 (25–100 ppm), and later retested in the same conditions. According to the results, it is hypothesized that the vapor concentration affects the performance of the sensors, making them more stable at higher concentrations. When comparing the performance of the sensor between the first and 180th days after fabrication, the level of reaction was maintained at 79% at the highest concentration, i.e., 507 , and 64% at the lowest, i.e., 127 . Performance degradation was mainly observed 0 to 90 days after fabrication, with only 10% or less occurring over days 90 to 180. This suggests that a sensor coated with POSS could maintain good long-term performance if initial performance degradation is prevented.
Figure 10.
Performance over time of the SAW sensor coated with POSS.
4. Conclusions
We designed and fabricated four 250 MHz SAW sensors coated with POSS (SAW-1), TU-1(SAW-2), and TU-2 (SAW-3), and evaluated their working performance by measuring the frequency shift upon exposure to GA, GB, and DMMP. In a response test to GA exposure within a concentration range of 50–100 mg/m3, all SAW sensors could detect the vapor at 50 . For GB in the range of 5.5–550 , SAW-1 and SAW-3 were effective at 5.5 , the lowest concentration of GB, while the limit of detection of SAW-2 was shown to be 110 . The response pattern of SAW-1 is a rectangular shape, but those of SAW-2 and SAW-3 presented spikes, indicating that the response to GB did not reach equilibrium during the exposure time. Also, SAW-2 and SAW-3 showed similar response patterns for the GB exposure because their sensing materials were synthesized with the same TU-based polymer. A response experiment for DMMP, a simulant of G nerve agents, was implemented in the range of 63–507 . The SAW-1 showed high sensitivity to low concentrations of DMMP, while SAW-2 showed high sensitivity under the highest concentration, which gradually decreased as the concentration declined, ending with an unstable signal at the lowest concentration. The sensitivity of SAW-3 was the lowest among the three sensors for DMMP vapor. The results of the exposure test confirmed that among the three sensing materials, POSS showed the highest sensitivity, the fastest response time to G nerve agents, and the largest frequency response at concentrations of GA and GB below LCt50. The sensitivity of SAW-1 to a real agent could be predicted using the simulant, i.e., DMMP, a compound which physically and chemically resembles GB.
Author Contributions
J.K. (Jinuk Kim) and H.P. performed the experiments, analyzed the data, and wrote the draft; J.K. (Jihyun Kim) and B.-I.S. contributed to the review and editing the manuscript. J.-H.K. contributed to the overall supervision of the experiment and writing the paper. All authors have read and agreed to the published version of the manuscript.
Funding
This work was supported by the Institute of Civil Military Technology Cooperation Center Funded by the Defense Acquisition Program Administration and Ministry of Trade, Industry and Energy, and of Korean government under grant No. 20-CM-BR-05 and supported by the Institute of Civil Military Technology Cooperation through the Civil Military Technology Development Project (No. 18-CM-SS-17).
Conflicts of Interest
The authors declare no conflict of interest.
Footnotes
Publisher’s Note: MDPI stays neutral with regard to jurisdictional claims in published maps and institutional affiliations.
References
- 1.Salem H., Sidell F.R. Nerve Agents. Encycl. Toxicol. 2005:201–206. doi: 10.1016/B0-12-369400-0/10031-6. [DOI] [Google Scholar]
- 2.Chauhan S., Chauhan S., D’Cruz R., Faruqi S., Singh K.K., Varma S., Singh M., Karthik V. Chemical Warfare Agents. Environ. Toxicol. Pharmacol. 2008;26:113–122. doi: 10.1016/j.etap.2008.03.003. [DOI] [PubMed] [Google Scholar]
- 3.Abdollahi M., Mostafalou S. G-Series Nerve Agents. 3rd ed. Volume 2. Elsevier; New York, NY, USA: 2014. [DOI] [Google Scholar]
- 4.Smith P.A., Lepage C.J., Harrer K.L., Brochu P.J. Hand-Held Photoionization Instruments for Quantitative Detection of Sarin Vapor and for Rapid Qualitative Screening of Contaminated Objects. J. Occup. Environ. Hyg. 2007;4:729–738. doi: 10.1080/15459620701547233. [DOI] [PubMed] [Google Scholar]
- 5.Satoh T., Kishi S., Nagashima H., Tachikawa M., Kanamori-Kataoka M., Nakagawa T., Kitagawa N., Tokita K., Yamamoto S., Seto Y. Ion Mobility Spectrometric Analysis of Vaporous Chemical Warfare Agents by the Instrument with Corona Discharge Ionization Ammonia Dopant Ambient Temperature Operation. Anal. Chim. Acta. 2015;865:39–52. doi: 10.1016/j.aca.2015.02.004. [DOI] [PubMed] [Google Scholar]
- 6.Puton J., Namieśnik J. Ion Mobility Spectrometry: Current Status and Application for Chemical Warfare Agents Detection. TrAC Trends Anal. Chem. 2016;85:10–20. doi: 10.1016/j.trac.2016.06.002. [DOI] [Google Scholar]
- 7.Pascual L., Campos I., Bataller R., Olguín C., García-Breijo E., Martínez-Mañez R., Soto J. A “Humid Electronic Nose” for the Detection of Nerve Agent Mimics; A Case of Selective Sensing of DCNP (a Tabun Mimic) Sens. Actuators B Chem. 2014;192:134–142. doi: 10.1016/j.snb.2013.10.089. [DOI] [Google Scholar]
- 8.Olguín C., Laguarda-Miró N., Pascual L., García-Breijo E., Martínez-Mañez R., Soto J. An Electronic Nose for the Detection of Sarin, Soman and Tabun Mimics and Interfering Agents. Sens. Actuators B Chem. 2014;202:31–37. doi: 10.1016/j.snb.2014.05.060. [DOI] [Google Scholar]
- 9.Logan T., Allen E., Way M., Swift A., Soni S.D., Koplovitz I. A Method for the Analysis of Tabun in Multisol Using Gas Chromatographic Flame Photometric Detection. Toxicol. Mech. Methods. 2006;16:359–363. doi: 10.1080/15376520600620083. [DOI] [PubMed] [Google Scholar]
- 10.Seto Y., Kanamori-Kataoka M., Tsuge K., Ohsawa I., Maruko H., Sekiguchi H., Sano Y., Yamashiro S., Matsushita K., Sekiguchi H., et al. Development of an On-Site Detection Method for Chemical and Biological Warfare Agents. Toxin Rev. 2007;26:299–312. doi: 10.1080/15569540701506756. [DOI] [Google Scholar]
- 11.Pacsial-Ong E.J., Aguilar Z.P. Chemical Warfare Agent Detection: A Review of Current Trends and Future Perspective. Front. Biosci. 2013;5:516–543. doi: 10.2741/S387. [DOI] [PubMed] [Google Scholar]
- 12.Holdren S., Tsyshevsky R., Fears K., Owrutsky J., Wu T., Wang X., Eichhorn B.W., Kuklja M.M., Zachariah M.R. Adsorption and Destruction of the G-Series Nerve Agent Simulant Dimethyl Methylphosphonate on Zinc Oxide. ACS Catal. 2019;9:902–911. doi: 10.1021/acscatal.8b02999. [DOI] [Google Scholar]
- 13.Wilmsmeyer A.R., Gordon W.O., Davis E.D., Troya D., Mantooth B.A., Lalain T.A., Morris J.R. Infrared Spectra and Binding Energies of Chemical Warfare Nerve Agent Simulants on the Surface of Amorphous Silica. J. Phys. Chem. C. 2013;117:15685–15697. doi: 10.1021/jp404265s. [DOI] [Google Scholar]
- 14.Pan Y., Zhang G., Guo T., Liu X., Zhang C., Yang J., Cao B., Zhang C., Wang W. Environmental Characteristics of Surface Acoustic Wave Devices for Sensing Organophosphorus Vapor. Sens. Actuators B Chem. 2020;315:127986. doi: 10.1016/j.snb.2020.127986. [DOI] [Google Scholar]
- 15.Nimal A.T., Mittal U., Singh M., Khaneja M., Kannan G.K., Kapoor J.C., Dubey V., Gutch P.K., Lal G., Vyas K.D., et al. Development of Handheld SAW Vapor Sensors for Explosives and CW Agents. Sens. Actuators B Chem. 2009;135:399–410. doi: 10.1016/j.snb.2008.08.040. [DOI] [Google Scholar]
- 16.Drafts B. Acoustic Wave Technology Sensors—Microwave Theory and Techniques. IEEE Trans. Microw. Theory Tech. 2001;49:795–802. doi: 10.1109/22.915466. [DOI] [Google Scholar]
- 17.Kirschner J. Surface Acoustic Wave Sensors (SAWS): Design for Application. Microelectromech. Syst. 2010;12:1–11. [Google Scholar]
- 18.Wang W., Lee K., Kim T., Park I., Yang S. A Novel Wireless, Passive CO2 Sensor Incorporating a Surface Acoustic Wave Reflective Delay Line. Smart Mater. Struct. 2007;16:1382–1389. doi: 10.1088/0964-1726/16/4/053. [DOI] [Google Scholar]
- 19.Ramakrishnan N., Nemade H.B., Palathinkal R.P. Resonant Frequency Characteristics of a SAW Device Attached to Resonating Micropillars. Sensors. 2012;12:3789–3797. doi: 10.3390/s120403789. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 20.Ricco A.J., Martin S.J., Zipperian T.E. Surface Acoustic Wave Gas Sensor Based on Film Conductivity Changes. Sens. Actuators. 1985;8:319–333. doi: 10.1016/0250-6874(85)80031-7. [DOI] [Google Scholar]
- 21.Wang Y., Du X., Long Y., Tang X., Tai H., Jiang Y. The Response Comparison of a Hydrogen-Bond Acidic Polymer to Sarin, Soman and Dimethyl Methyl Phosphonate Based on a Surface Acoustic Wave Sensor. Anal. Methods. 2014;6:1951–1955. doi: 10.1039/C3AY42214C. [DOI] [Google Scholar]
- 22.Bielecki M., Witkiewicz Z., Rogala P. Sensors to Detect Sarin Simulant. Crit. Rev. Anal. Chem. 2020:1–13. doi: 10.1080/10408347.2020.1723401. [DOI] [PubMed] [Google Scholar]
- 23.Yang Z., Peng H., Wang W., Liu T. Crystallization Behavior of Poly(ε-Caprolactone)/Layered Double Hydroxide Nanocomposites. J. Appl. Polym. Sci. 2010;116:2658–2667. doi: 10.1002/app.31787. [DOI] [Google Scholar]
- 24.Higgins B.A., Simonson D.L., Houser E.J., Kohl J.G., Andrew Mcgill R. Synthesis and Characterization of a Hyperbranched Hydrogen Bond Acidic Carbosilane Sorbent Polymer. J. Polym. Sci. Part A Polym. Chem. 2010;48:3000–3009. doi: 10.1002/pola.24078. [DOI] [Google Scholar]
- 25.Lee Y.-J., Kim J.-G., Kim J.-H., Yun J., Jang W.J. Detection of Dimethyl Methylphosphonate (DMMP) Using Polyhedral Oligomeric Silsesquioxane (POSS) J. Nanosci. Nanotechnol. 2018;18:6565–6569. doi: 10.1166/jnn.2018.15698. [DOI] [PubMed] [Google Scholar]
- 26.Ha S., Lee M., Seo H.O., Song S.G., Kim K.S., Park C.H., Kim I.H., Kim Y.D., Song C. Structural Effect of Thioureas on the Detection of Chemical Warfare Agent Simulants. ACS Sens. 2017;2:1146–1151. doi: 10.1021/acssensors.7b00256. [DOI] [PubMed] [Google Scholar]
- 27.Devkota J., Ohodnicki P.R., Greve D.W. SAW Sensors for Chemical Vapors and Gases. Sensors. 2017;17:801. doi: 10.3390/s17040801. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 28.Länge K. Bulk and Surface Acoustic Wave Sensor Arrays for Multi-Analyte Detection: A Review. Sensors. 2019;19:5382. doi: 10.3390/s19245382. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 29.Kim Y.H., Chang D.H., Yoon Y.S., Lee T.J., Kang S.J. Study on the Optimization of a Temperature Sensor Based on SAW Delay Line. J. Kor. Phys. Soc. 2004;45:1366–1371. [Google Scholar]
- 30.Kim E., Kim J., Ha S., Song C., Kim J.-H. Improved Performance of Surface Acoustic Wave Sensors by Plasma Treatments for Chemical Warfare Agents Monitoring. J. Nanosci. Nanotechnol. 2020;20:7145–7150. doi: 10.1166/jnn.2020.18850. [DOI] [PubMed] [Google Scholar]
- 31.Wohltjen H. Mechanism of Operation and Design Considerations for Surface Acoustic Wave Device Vapour Sensors. Sens. Actuators. 1984;5:307–325. doi: 10.1016/0250-6874(84)85014-3. [DOI] [Google Scholar]
- 32.Kim J., Kim E., Kim J., Kim J.-H., Ha S., Song C., Jang W.J., Yun J. Four-Channel Monitoring System with Surface Acoustic Wave Sensors for Detection of Chemical Warfare Agents. J. Nanosci. Nanotechnol. 2020;20:7151–7157. doi: 10.1166/jnn.2020.18851. [DOI] [PubMed] [Google Scholar]
- 33.Grate J.W. Hydrogen-Bond Acidic Polymers for Chemical Vapor Sensing. Chem. Rev. 2008;108:726–745. doi: 10.1021/cr068109y. [DOI] [PubMed] [Google Scholar]
- 34.Hartmann-Thompson C., Hu J., Kaganove S.N., Keinath S.E., Keeley D.L., Dvornic P.R. Hydrogen-Bond Acidic Hyperbranched Polymers for Surface Acoustic Wave (SAW) Sensors. Chem. Mater. 2004;16:5357–5364. doi: 10.1021/cm040346z. [DOI] [Google Scholar]
- 35.Kuo S.W., Chang F.C. POSS Related Polymer Nanocomposites. Prog. Polym. Sci. 2011;36:1649–1696. doi: 10.1016/j.progpolymsci.2011.05.002. [DOI] [Google Scholar]
- 36.Lavoie J., Srinivasan S., Nagarajan R. Using Cheminformatics to Find Simulants for Chemical Warfare Agents. J. Hazard. Mater. 2011;194:85–91. doi: 10.1016/j.jhazmat.2011.07.077. [DOI] [PubMed] [Google Scholar]