Abstract
Peripheral Nerve Stimulation (PNS) limits the acquisition rate of Magnetic Resonance Imaging data for fast sequences employing powerful gradient systems. The PNS characteristics are currently assessed after the coil design phase in experimental stimulation studies using constructed coil prototypes. This makes it difficult to find design modifications that can reduce PNS. Here, we demonstrate a direct approach for incorporation of PNS effects into the coil optimization process. Knowledge about the interactions between the applied magnetic fields and peripheral nerves allows the optimizer to identify coil solutions that minimize PNS while satisfying the traditional engineering constraints. We compare the simulated thresholds of PNS-optimized body and head gradients to conventional designs, and find an up to 2-fold reduction in PNS propensity with moderate penalties in coil inductance and field linearity, potentially doubling the image encoding performance that can be safely used in humans. The same framework may be useful in designing and operating magneto- and electro-stimulation devices.
Keywords: gradient coil design, MRI safety, peripheral nerve stimulation, electromagnetic field simulation, magneto-stimulation thresholds, neurodynamic nerve model
I. Introduction
Peripheral Nerve Stimulation (PNS) has become a significant limitation in the use of high-performance gradient coils in Magnetic Resonance Imaging (MRI) [1, 2]. Rapid switching of the magnetic field gradient is needed to quickly encode the image but induces electric fields (E-fields) in the conductive body strong enough to generate unwanted action potentials in peripheral nerves. This can lead to bothersome or even painful stimulations [3, 4, 5, 6]. Although most prominent in whole-body gradient systems used in clinical scanners due to the large body areas exposed, the PNS effect is also becoming an acute limitation for state-of-the-art head gradient coils. Head gradients have intrinsically greater PNS thresholds (2-3 fold), but are now also being used at 2-3 fold higher gradient strength and slew rate [7]. Despite its limiting role, PNS effects are not directly incorporated in the gradient winding optimization. The PNS characteristics are assessed after the coil design phase during stimulation experiments which requires construction of coil prototypes and measuring the thresholds in experiments using healthy subjects [8, 9, 10, 11]. These stimulation studies are not only complex, but also provide very limited insights into design changes that might reduce PNS. Instead, indirect scaling relationships between PNS thresholds and gradient coil design parameters such as linear field-of-view (FOV) and coil length are used as surrogates for a PNS metric in the design process [12]. In addition to being possibly sub-optimal, this approach precludes the study of tradeoffs between PNS thresholds and the traditional engineering metrics such as coil inductance, field linearity, wire density, and torque.
The ability to improve a coil design during the iterative coil winding optimization requires a subspace of coil solutions with similar engineering metrics but different PNS thresholds. An explicit PNS metric then allows this sub-space to be explored, possibly yielding improved PNS characteristics with little cost to the other engineering metrics. The PNS metric also provides a principled way to characterize the trade-off between these traditional engineering metrics and PNS.
If the coil design required complete specification of the magnetic field (B-field), Maxwell’s equations would uniquely specify the resulting E-field in a given human body and thus the PNS threshold. However, the designer seeks only a target pattern for the gradient of the Bz field component within the imaging FOV, allowing an infinite number of choices for the transverse field components (Bx and By) and all field components outside of the imaging FOV. These field components constitute degrees-of-freedom that can be chosen to minimize PNS. Hidalgo-Tobon et al. [13] exploited this concept by superimposing an additional Bx or By field to partially cancel the transverse gradient coil’s concomitant fields, resulting in a ~50% improvement in measured PNS thresholds. The simplistic addition of concomitant field coils, however, leads to an exacerbated net torque of the gradient coil in the MRI environment [13]. Thus, in addition to creating a PNS constraint, it is critical to simultaneously consider all other design constraints during coil winding optimization.
The boundary element method stream function (BEM-SF) approach is the most common method for optimizing MRI gradient coils [14, 15, 16]. In the BEM-SF approach, the coil windings are derived from a current density distribution on the coil former surface (usually a cylinder). Each vertex of the cylinder’s surface mesh carries a discrete current density element (the stream function basis). The B-field generated from a candidate design is formed from the weighted superposition of the precomputed fields of these basis elements. Although precomputation of the bases’ fields is time-consuming, the calculation of any candidate design’s field is a simple matrix multiplication. The optimal basis weights are determined by a linear solver that seeks the weights that create a best match to the target field while satisfying a series of engineering constraints (such as coil inductance, wire spacing, and torque), also expressed as linear or quadratic matrix forms. In our work, we provide the ability to include the most important “biological constraint”, PNS thresholds, in exactly the same way.
We have recently shown that activation of peripheral nerves in the body by external electromagnetic fields can be accurately predicted using a coupled electromagnetic-neurodynamic modeling framework. This tool predicts the “forward” PNS response (i.e., generation of PNS by an existing coil) [17, 18, 19, 20]. The pipeline computes the induced E-fields in detailed electromagnetic body models and assesses the field’s capability to induce action potentials in a realistic peripheral nerve system (more than 1900 nerves in our simulations) embedded in the body model. The approach was validated against experimental thresholds measured in healthy volunteers in several MRI gradient coils [21, 22]. As such, this “forward” PNS model is useful for manually assessing a small number of candidate coils prior to construction and can thus guide the designer. This workflow, however, cannot be readily incorporated in the winding optimization due to the time-consuming E-field simulations and nonlinear neurodynamic calculations. The nonlinearity of the neurodynamics arises from the ion channels which themselves are sensitive to the membrane voltage. If the stimulus reaches a certain threshold, the coupled equations for the nerve membrane potential move from a damped (negative) feedback regime to positive-feedback causing the ion channels to open up abruptly. This process alters the nerve membrane potential; a change known as an action potential. Although the coupled differential equations describing action potential generation are well-known, modeling this non-linear process is time-consuming and not amenable to direct incorporation into the gradient winding optimization where the objective function and constraints need to be evaluated thousands of times.
Additionally, expression of BEM-SF design constraints as linear or quadratic forms is an important pre-requisite for preserving the convex nature of the optimization. While the electromagnetic fields naturally obey superposition, it is important to find a linear expression for the nerve responses so that each nerve’s response can be represented as a weighted sum of precomputed responses from each basis function (wire segment). To achieve this, we developed the “PNS oracle”, a generalized linear version of the neurodynamic response function that correlates well with PNS thresholds computed from the full neurodynamic modeling [23, 24]. Armed with a pre-computation of how each basis function contributes to the PNS threshold of each nerve segment, the full response from the candidate winding pattern is obtained from a simple “P-matrix” multiplication with the stream function basis weights.
In this work, we leverage the PNS oracle and perform numerical optimizations of coil windings with – for the first time – explicit PNS constraints [25]. We combine the PNS constraint with traditional engineering constraints (field linearity, coil inductance, wire density, and torque). This encourages the solver to explore the sub-space of coil solutions with similar “engineering metrics” and identify a solution with favorable reduced PNS. It also provides a framework to analyze the optimal tradeoff between the traditional engineering metrics and PNS performance. The convex optimization provides a method to explore the theoretical limits of PNS-optimal coil design. We demonstrate this tool by designing a series of whole-body and head-only gradient coils with typical engineering constraints but up to 2-fold reduction in PNS propensity.
II. Methods
Boundary Element Stream Function Method
The Boundary Element Method Stream Function (BEM-SF) approach is the gold standard for designing efficient gradient coil windings. The goal for each of the three separate windings is a spatially linear Bz field pattern (“gradient”) in the chosen region-of-linearity (ROL) while satisfying a series of engineering constraints such as field linearity, torque balancing, minimum wire spacing (current density), and field magnitude at the magnet surface (controlling the eddy currents) [14, 15, 16]. In the BEM-SF approach, the wires are constrained to the surfaces of a coil former geometry (in most cases two coaxial cylinders). This coil geometry is represented by a surface mesh that carries a continuous current density. Instead of directly optimizing the current density (which may lead to solutions that are not divergence-free), the current density is indirectly described by a scalar stream function whose iso-contours define the paths of current flow. The total stream function on the coil surface is discretized into a set of local stream function bases defined on the former’s mesh vertices (see Fig. 1). Every stream function basis models a divergence-free current circulating around the respective vertex (similar to a small current loop). The full stream function is constructed from a weighted sum of stream function bases described by a vector x. This construction guarantees that the resulting continuous current density is divergence-free, allowing it to be converted to closed discrete wires.
Prior to optimization (finding the optimal weight set x), extensive precomputations are performed to characterize the contributions from each stream function basis to the various design metrics (B-fields, torques, inductance, etc.). For example, the Bz field contribution to each point in the target region is precomputed for each basis and assembled into a matrix C such that Cx describes the total Bz field generated in the ROL. In addition to the imaging ROL, fields are precomputed at the magnet surface to constrain the magnitude of eddy currents. In a similar fashion, we precompute a series of additional matrices linking the stream function bases to relevant physical properties such as the current density (Dx) and the torque (Tx). The pairwise coupling between individual stream function bases is used to generate a matrix L such that the quadratic form xTLx scales with the coil inductance. Note that other metrics have been proposed, such as heating [26, 27], sound pressure levels [28, 29], and secondary effect of eddy currents on the Bz field in the imaging ROL [30, 31].
Finally, the BEM-SF coil design problem can be formulated as a quadratic minimization of the coil inductance subject to linear constraints for the other engineering metrics:
Eq. 1 |
where btar defines the target B-field pattern (linear Bz field gradient within the ROL and zero normal field on the magnet cryostat), ϵ is the tolerated field gradient error, and Tmax / Dmax denote the torque and current density constraints.
PNS model and P-matrix assembly
Addition of an explicit PNS constraint requires construction of a matrix (similarly to the magnetic field or torque matrices) that links the basis weight vector x to the PNS propensity of each nerve segment in the body. Multiplication of this “P-matrix” by the weight vector x superimposes the effects of all current bases and generates a vector describing the PNS propensity for each peripheral nerve segment. This P-matrix is formed using our “PNS oracle” formulation (the inverse of the PNS threshold) rather than the threshold itself since the oracle is linear with respect to the coil current. This allows for expressing the optimization constraint as a simple inequality (Px ≥ Pmax). Of course, the more commonly discussed PNS threshold can be easily calculated via the inverse PNS oracle.
The workflow to obtain the P-matrix starts with computation of the electromagnetic fields created by a winding pattern or current density basis element within a detailed male or female body model [18]. In this work, we study two body positions: head-centered or heart-centered (head first supine). The electromagnetic fields are simulated for a 1 kHz sinusoidal current waveform applied to the current basis element using a low-frequency magneto quasi-static solver (Sim4Life, Zurich MedTech) on a 1 mm3 hexahedral mesh. Given the E-fields induced by a subset of the basis elements (~10%), we compute the electric potential changes V(r) along each of the 1900 nerve segments in the body (each of which is labeled by an estimate of the local axon diameter). The electric potentials for the remaining ~90% of basis elements are obtained using a barycentric interpolation scheme. We then predict the response of each nerve fiber using either a full nonlinear neurodynamic model [32, 33, 34] or our linear PNS oracle [24].
The PNS oracle metric (which is the reciprocal PNS threshold) follows a previous definition of the neural activation function and modified driving function [35]. It estimates the threshold of action potential generation of a nerve segment based on the spatial pattern of the applied electric potential using only linear operations. Note that the PNS oracle does not explicitly model ion dynamics and action potential initiation or propagation and thus ignores the non-linear aspects of nerve stimulation. This provides the needed linear relationship between the electric potential V(r) and the propensity for stimulation. The PNS oracle (PNSO) is defined as
Eq. 2 |
where V(r) is the extracellular potential along the nerve position r, L(D) is the distance between consecutive nodes of Ranvier (which is a function of the axon diameter D), K(D) is a spatial kernel, and m(D) is a calibration factor. The PNS oracle is based on the second spatial difference of the electric potential V(r) across adjacent nodes of Ranvier. The result is convolved with a spatial kernel K(D) that describes cross-talk between adjacent nodes of Ranvier. Finally, the factor m(D) accounts for the different excitability due to the axon-dependent level of myelination. Both the kernel K(D) and the myelination factor m(D) are calibrated for a specific temporal waveform of the driving fields (in this work a 1 kHz sinusoid). We have recently shown that the PNS oracle correlates very well with the inverse thresholds obtained from the full neurodynamic model (R2 > 0.995). The linear nature of the PNS oracle with respect to the electric potentials along the nerve and thus the local E-field and ultimately the coil currents or current density allows it to be assembled into a matrix form (the P-matrix). Note that a P-matrix must be pre-computed for each body model, coil former geometry (including the body position within the coil former) and each driving waveform. This process yields a matrix of size p × n where p is the number of nerve locations and n is the number of stream function bases. Figure 2 illustrates this process for a whole-body cylindrical coil loaded with the female body model. The resulting P-matrix is incorporated in the BEM-SF optimization as an additional linear constraint:
Eq. 3 |
Simulated Body and Head Gradient
We design a series of actively shielded whole-body and head-only gradient coils with constraints similar to commercially available gradient coils. We assess three gradient design geometries: 2 and 3-layer body coils and 3-layer head coils. The symmetric body coil geometry consists of two or three coaxial cylinders (110 cm long) with diameters of d1 = 66 cm and d3 = 82 cm, and an optional intermediate layer for the 3-layer design (d2 = 74 cm). Each cylindrical surface carries 2340 stream function bases. The P-matrix is computed for each layer using the female model positioned with the head at iso-center. The target efficiency for the body coils is set to 0.1 mT/m for 1 A coil current. The head-only gradient coil geometry consists of three concentric cylinders (diameters: d1 = 46 cm, d2 = 62 cm, d3 = 75 cm, lengths: h1 = 65 cm, h2 = 110 cm, h3 = 120 cm). Note that the head coil geometry is asymmetric (as opposed to the symmetric body coil geometry). The head coil layers support 2592, 4400, and 4080 stream function bases, respectively. The respective P-matrices are also computed using the female model. The head gradients’ target efficiency is set to 0.15 mT/m for 1 A coil current. This efficiency is on the lower end of typical head gradient coils, which have efficiencies ranging from 0.15-0.3 mT/m/A [36, 37, 7]. Although at the lower end of this range, the design choice of 0.15 mT/m/A provided enough degrees-of-freedom for PNS mitigation. We design several families of gradient coils with 5%, 10%, 15%, and 20% deviation from the ideal linear gradient field pattern. A coil winding’s linearity deviation ε is computed from the coil’s induced Bz-field in all target points () and the target field () [38, 39] via
Eq. 4 |
The PNS constraint is varied as well (including coil designs without PNS constraints). Additionally, we constrained the maximum torque by Tmax = 0.1 Nm/A, the minimum wire spacing by Dmax = 7 mm, and the active shielding capability in terms of B-field magnitude on the surface of the magnet cryostat by 1 μT/A. All BEM-SF simulations are performed in Matlab R2018b (MathWorks, Natick, MA, USA) using the MOSEK quadratic optimizer (MOSEK Aps, Copenhagen, Denmark). For each stream function solution, we extract discrete coil windings by tracing the iso-contours of the stream function solution and compute the coil inductance using FastHenry2 [40].
Additionally, we analyze the coil solution’s concomitant B-field components (Bx and By) that affect the local Larmor frequency and thus potentially interfere with spin encoding. We assess the magnitude of this effect by computing the fractional deviation of the total B-field from the idealized encoding field without concomitant components:
Eq. 5 |
We report the average and maximum of this positive valued metric over the ROL.
Finally, we evaluate the robustness of the coil solutions with respect to body position (optimize for head at iso-center, simulate PNS characteristics for heart at iso-center) as well as body size and gender (optimize for the female model, assess PNS characteristics for the male model).
III. Results
L-curves
Figure 3 shows L-curves quantifying the tradeoff between PNS and coil inductance for X, Y, and Z-axes body coils (2-layer and 3-layer geometries) and head-coils (3-layer geometry). The L-curves were generated for gradient linearity constraints of 5%, 10%, 15%, and 20%. Every point of the L-curves corresponds to a coil winding solution; the first point (circled x) corresponds to the standard solution without PNS constraints.
For all coils studied, incorporation of PNS constraints yields winding solutions with significantly reduced PNS propensity (decreased PNS oracle, i.e., increased PNS threshold). One traditional approach for PNS reduction is relaxation of the linearity constraint. Our simulations show this effect, for example relaxing the linearity of the Y-body coil from 5% to 10% leads to a 13% PNS oracle reduction (36.6 m/T to 31.7 m/T) while also reducing the coil inductance by 27% (840 μH to 613 μH). In contrast, incorporation of PNS constraints yields a coil with the same inductance (840 μH) and relaxed linearity (10%), but with a 52% reduction in PNS oracle (36.6 m/T to 17.4 m/T). Alternatively, PNS can be reduced without penalty in field linearity: for a 5% linearity Y-axis body coil, a PNS improvement of 20% can be achieved at a modest ~5% penalty in inductance. Accepting a 20% increase in inductance allows reducing PNS by 40% (these coil solutions are marked B1 and B2 in Fig. 3). The PNS oracle improvements where comparable for the body X-axis, but came at higher cost in terms of inductance for the Z-axis. In this case, increasing the inductance by 20% allowed for a 20% PNS reduction. Additional relaxation of the linearity constraint (5% to 10%) leads to a 45% PNS oracle reduction. Figure 3 (left column) also shows results from adding a third cylinder to the body coil design (dash lines). This further improved PNS performance for larger inductance X and Y-axis coil designs.
Figure 3 (right column) shows results for the head gradient coils. In this case, PNS oracle reduction was achieved with a smaller coil inductance penalty. For the X-axis, the PNS oracle is reduced by 40% for only a 15% increase in inductance (the respective coil windings are marked H1 and H2). Tolerating a penalty in both coil inductance and field linearity (from 5% to 10% error) yields a PNS reduction of 60%. Similarly to the body gradient’s Z-axis, improving the PNS performance of the head gradient’s Z-axis led to higher inductance penalties (45% increase for 40% PNS reduction).
Specific winding solutions and stimulation locations
Figure 4 shows the coil windings and PNS oracle maps in the female model for the unoptimized and optimized 2-layer body Y-axis coils (solutions B1 and B2 in Fig. 3) and Z-axis coils (solutions B3 and B4 in Fig. 3). All coils achieve the same gradient efficiency (0.1 mT/m/A), field linearity (5%), and are actively-shielded and torque-balanced. Unfolded winding patterns are shown in supplementary Fig. S1.
The PNS oracle map shows hot-spots greater than 1% of the maximum PNS oracle in the body as colored spheres. The color and size of each sphere corresponds to the PNS oracle value (reciprocal PNS threshold). For example for coil B1 (Fig. 4, left), stimulation of the “yellow-sphere” nerves in the shoulder, biceps, and upper vertebrae requires a gradient amplitude of approx. 27 mT/m (for a 1 kHz sinusoidal gradient waveform), while stimulation of the “red-sphere” nerves (mostly neck and lower vertebrae) requires a higher amplitude of approx. 50 mT/m. The coil B1 without PNS-optimization corresponds to a classical symmetric Y-axis coil with typical stimulation sites in the shoulder, triceps, and close to the vertebrae. The optimized Y-axis coil B2 deviates significantly from the conventional solution B1. This modification leads to a reduced interaction with the peripheral nerves, and thus to a 66% increase in the PNS thresholds (from 27.3 mT/m to 45.5 mT/m), which corresponds to a 40% PNS oracle reduction. The classical symmetric Z-axis coil B3 without PNS-optimization mainly stimulates the intercostal nerves leaving the vertebrae. The optimized coil B4 has an unconventional and asymmetric wire pattern which leads to a 43% PNS threshold increase (from 42.9 mT/m to 61.4 mT/m), corresponding to a 25% PNS oracle reduction.
Figure 5 shows the winding solutions and PNS oracle maps for the unoptimized and optimized 3-layer head X-axis coils H1 and H2 and Z-axis coils H3 and H4 (unfolded winding patterns are shown in supplementary Fig. S2). The unoptimized X-axis coil H1 strongly stimulates the facial nerves (forehead and nose) and mildly stimulates the shoulder nerves. The optimized coil H2 significantly reduces stimulations of the facial nerves while slightly increasing stimulations in the shoulder, thus balancing stimulations in these two body parts. The Z-axis head coil H3 without PNS-optimization mainly stimulates the intercostal nerves (similarly to the body Z-axis coils) as well as the shoulder nerves. The optimized Z-axis windings increase the PNS thresholds by 66% (109 mT/m to 222 mT/m), corresponding to a 40% PNS oracle reduction.
Body Surface B-fields and E-fields
Figure 6 shows the B-field and E-field magnitudes on the surface of the female body model induced by the body Y-axis coils B1 and B2 and the head X-axis coils H1 and H2 (results for the Z-axis body coils B3/B4 and head coils H3/H4 are shown in supplementary Fig. S3). The unoptimized body coil B1 induces a rather uniform B-field in the chest. The resulting E-fields are shaped by the conductive body model, with hot-spots in the neck/shoulders, upper arm, hips and close to the spine. These hotspots roughly coincide with the PNS hot-spots shown in Fig. 4. The optimized body coil B2 reduces the B-field exposure of the chest (thus reducing the induced E-field and PNS), for example, by redirecting some of the B-fields through the head where the E-fields increase mildly.
The unoptimized head coil H1 induces strong B-fields (and thus high E-fields) in the head, with lower fields in the shoulders and thorax. The optimized coil H2 reduces the B-field magnitude in the head and increases the B-field in the chest. This modification shifts the high E-field region from the head to the chest where fewer peripheral nerves are present, thus reducing the total PNS response.
Linearity and image distortion
Figure 7, left, shows maps of the gradient field error for the unoptimized and optimized Y-axis body coils B1 and B2 and the resulting spatial deformations in the center and off-center sagittal planes (transverse planes are shown in supplementary Fig. S4). Every wrap of the colormap corresponds to a 5% increase in linearity error. The unoptimized coil B1 creates typical symmetric field error characteristics with zero error at the isocenter and a steady increase towards the boundary of the FOV (dashed line) where the first 5% error contour occurs. Consequently, the spatial deformations (bottom, black lines) are mostly confined to the edges of the FOV. The PNS-optimized coil B2 creates a less symmetric, but still smooth error map with a baseline error that leads to a mild shift of the image (red lines). The baseline linearity error and resulting image shift are stronger in the off-center slice. Both coils B1 and B2, however, have the same peak linearity error (5%) and thus the same maximum pixel displacement. The unoptimized and optimized head coils H1 and H2 (Fig. 7, right column) achieve more similar linearity error maps; the only significant differences occur at the left/right boundaries of the ROL, while both coils achieve the same 5% peak linearity error. Linearity results for the Z-axis body coils B3 and B4 and Z-axis head coils H3 and H4 are shown in supplementary Figs. S5 and S6.
Concomitant magnetic fields
The concomitant field metric within the ROL was larger for the PNS-optimized body coils. For the Y-axis body coils B1 and B2, PNS-optimization increased the average and maximum of this metric from Δc = 0.044 (maximum Δc = 0.203) to Δc = 0.087 (maximum Δc = 0.434). For the Z-axis body coils B3 and B4, the increase was from Δc = 0.022 (maximum Δc = 0.053) to Δc = 0.0254 (maximum Δc = 0.076). For the head gradient X-axis coils (H1 and H2), PNS-optimization lowered the concomitant field metrics from Δc = 0.067 (maximum Δc = 0.209) to Δc = 0.046 (maximum Δc = 0.163). The concomitant field metrics for the Z-axis head coils H3 and H4 were essentially unaffected by the optimization (average Δc = 0.006 and maximum Δc = 0.015 in both cases).
Robustness to body model type and position
Figure 8 summarizes the results of a preliminary robustness analysis. Panels A and B show L-curves (PNS oracle vs. coil inductance) for the 5% linearity body Y-axis coils optimized for the female model for head imaging. The black curve shows the PNS responses in the female model (same L-curve as in Fig. 3), whereas the dashed blue curve shows PNS responses in the male model for this “female-model-designed” coil. Similarly to our previous results, the female model has a significantly higher threshold (lower PNS oracle) than the male [41], but the relative PNS improvements obtained from the PNS-informed coil optimization are similar for both body models. For example, optimizing and testing on the female model reduces the PNS oracle by 40%, (B1 vs. B2) while optimizing on the female and testing on the male model reduces the PNS oracle by 45%. Similar results are achieved for the head X-axis coils H1 and H2 (Fig. 8, top right) where the PNS oracle is reduced by 40% (7.5 m/T to 4.5 m/T) and 30% (9.6 m/T to 6.7 m/T) in the female and male models, respectively.
Figure 8C shows the effect of changing the body model position within the 5% linearity, head-at-isocenter PNS-optimized body Y-axis coils. The black L-curve shows the PNS responses for the female model placed with the head at isocenter; the dashed blue L-curve shows the PNS responses when the female model is shifted by 30 cm along the z-direction (mimicking cardiac imaging). Using these coils for cardiac imaging instead of head imaging reduces the PNS improvement for coils B1/B2 from 40% to 18%. Alternatively, the coil windings can be optimized simultaneously for both scan positions by stacking the respective P-matrices in the BEM-SF optimization. These jointly optimized windings (red L-curves in Fig. 8C) achieve similar PNS oracle values for both cardiac and head imaging positions (31 m/T and 30 m/T for the 1060 μH coil solution, corresponding to 28% and 19% PNS oracle reductions).
IV. Discussion
In this work, we describe an approach to incorporate PNS effects into the design process of MRI gradient coils and demonstrate its potential for addressing this growing biological limit to MRI performance. The representation of the PNS model in matrix form (the P-matrix) allows the winding optimizer to quickly assess and compare thousands of candidate coil designs and iteratively approach an optimal design. The convexity of the optimization guarantees convergence to an optimal solution and stems from the linearity of the PNS oracle [23, 24] with respect to the current applied to individual coil segments (basis elements). Although formulation of the P-matrix for a given coil basis set requires considerable precomputation, it can be applied to multiple design problems for that geometry. The approach also incorporates traditional engineering constraints (such as field linearity, coil inductance, wire spacing, and torque) and assesses optimal tradeoffs between these constraints and PNS improvements. We demonstrated that this framework allows designing body and head gradient coils with more than 2-fold reduction in PNS propensity with moderate penalties in coil inductance or field linearity.
The L-curve analysis (Fig. 3) showed that significant PNS improvements can be achieved, albeit at the cost of some additional coil inductance or degradation of field linearity. For example, tolerating a 10% increase in inductance allows reducing PNS for the X and Y-axes of body and head gradients by 25% and 40%, respectively. For the body Y-axis coil, relaxing both the inductance and linearity constraints (from 5% to 10% linearity error) yields a 2-fold PNS reduction. Improving the performance of the Z-axes in either body or head gradients comes at higher inductance penalties. This is most likely due to the fact that Z-axis coils mostly create Bz field components. Reshaping of these field components for PNS reduction must be confined to outside of the FOV to avoid interference with spin encoding. In contrast, X and Y-axes mostly create Bx and By components outside of the FOV. These field components only mildly affect spin encoding, and thus can be modified both inside and outside of the FOV to reduce PNS.
Analysis of the concomitant field terms in the linear volume showed that the winding optimizer utilizes this degree-of-freedom for PNS reduction. For the optimized body Y-axis coil B2, this led to 2-fold higher average concomitant field metric (compared to the unoptimized coil B1). This places increased burden on field characterization [42, 43] and correction schemes, such as gradient pre-emphasis [44] and frequency modulation of the RF coils [45, 46]. These concepts have been successfully used in various high-performance scanners, including the Connectome system [1] and a compact head gradient 3T scanner [47].
The algorithm can also trade the field linearity constraint for PNS improvements but even if this constraint is held constant, undesirable effects can be introduced. For example, both body Y-axis coils B1 and B2 achieve the same 5% maximum linearity error (leading to the same maximum pixel displacement), but the image distortion pattern for the PNS-optimized coil B2 is more complex, i.e., less symmetric and less uniform. Similar effects were observed for the body Z-axis coils B3 and B4, while being less pronounced for the head X-axis and Z-axis coils H1-H4. This will potentially require more advanced correction techniques [48, 49, 50].
The inductance penalty for the PNS-optimized coils arises from the need to reshape the coil’s B-fields in an attempt to reduce B-field exposure to the body’s high cross-sectional area regions (which reduces the E-fields). This B-field shaping requires additional wire paths, leading to higher coil inductance. It is important to note that an increase in coil inductance results in a reduced slew rate for the user assuming a given maximum voltage available for the gradient amplifier. Additionally, higher inductance requires increased heat removal (from increased resistance due to longer wire lengths) and partial discharge concerns due to large voltages between neighboring wires. While these are negatives, the importance of the two parameters (inductance and PNS) underscores the need to rigorously assess this tradeoff between optimal designs.
The gradient coil windings shown in this work were PNS-optimized for an 8 ms sinusoidal 1 kHz coil waveform (the parameters of the PNS oracle were calibrated for this waveform). Analyzing a single frequency is sufficient for our needs since the main goal of our winding optimization is to enforce relative PNS improvements. This means that a PNS-optimized coil that achieves a 20% threshold improvement for sinusoidal waveforms will achieve a very similar relative improvement for trapezoidal pulses or for other frequency sinusoids. While it is not fully studied that optimizing a gradient coil for one waveform guarantees optimality for other waveforms, the existence of the simple monotonic relationship between the common waveform types (such as sinusoidal and trapezoid) is encouraging [10]. Finally, while the PNS oracle utilized in this work is computed for a specific 1 kHz frequency readout waveform, the PNS oracle and the neurodynamic modeling approach have been shown to be accurate for a range of frequencies up to about 10 kHz [18].
The PNS oracle plays an important role in informing the winding optimizer about the PNS characteristics of different coil solutions. In fact, constraining the E-field magnitude at the nerve locations (rather than the PNS oracle) disregards several important aspects of nerve excitability. First, many nerve tracks in the body do not contain sensitive motor nerves but only less-sensitive sensory nerves that can tolerate higher E-fields. Second, the nerve’s excitability strongly depends on the relative orientation between the nerve and the imposed E-field. For example, a kinked nerve segment in a moderate E-field region can be much more susceptible to stimulation, than a straight nerve segment in a high E-field region. Third, a nerve that is stimulated moderately at two segments a few millimeter apart is likely more susceptible than a nerve being stimulated more strongly at a single nerve segment (due to cross-talk between neighboring nerve segments). Correctly accounting for these aspects provides a more accurate PNS metric and offer degrees-of-freedom that can be harnessed by the optimizer for PNS reduction. Most importantly, however, simplified metrics (such as E-field at the location of nerves) still require the relatively hard-to-come-by electromagnetic model with a nerve atlas. Given this necessary model, the PNS oracle then captures these aspects in an easily computed formulation and preserves linearity in the E-field (and thus the current density basis functions).
The level of achievable PNS reduction is expected to be determined by the number of degrees-of-freedom in the BEM-SF optimization problem, and different strategies can be used to create these needed degrees-of-freedom. One strategy is to relax some of the traditional engineering constraints such as gradient field linearity. This tradeoff between linearity and PNS performance has been previously noted [12], but our optimization strategy can maximally exploit it. For example, using only the traditional coil design approach without PNS metric achieves a PNS improvement of only 15%. This traditional approach also reduces the coil inductance by 25%, however, this may not be useful if PNS limits the usable slew rate.
Adding coil layers appears to be another viable strategy for providing more degrees-of-freedom to the optimization problem. For the body coils, two coil layers are sufficient to achieve significant PNS improvements. Addition of a third layer (dashed L-curves in Fig. 3, left column) reduces the inductance “cost” at constant PNS improvements. If additional inductance is acceptable, the 3-layer approach can take advantage of this to improve PNS performance beyond ~40%. For the head coils, we show only 3-layer coil layouts since the BEM-SF optimization for the 2-layer setup lacks solutions for levels of PNS reduction beyond ~15% (especially for coils with 5% target linearity). The winding optimizer used the intermediate coil layer mainly for field generation (rather than shielding). For the PNS-optimized coils, the intermediate layer also played an important role in bending away the X-axis coil’s B-fields from the facial area (Fig. 6) or the Z-axis coil’s B-fields from the shoulder area (Fig. S3) in an attempt to reduce the induced E-fields and thus PNS. Although the PNS informed coil design/operation approach is demonstrated on cylindrical geometry coils, the method can be easily applied to unconventional coil geometries [51, 52, 29, 53], coils with 3D winding distributions [54, 55], and matrix gradient coils [56, 57, 58].
Although we have previously shown that our male and female body models have differing PNS thresholds [41], it appears that an optimal winding pattern for one gender model is near optimal for the other (Fig. 8). Most likely, this robustness stems from the fact that the BEM-SF basis set does not provide enough degrees-of-freedom to reduce PNS on a microscopic scale. In fact, adjacent stream function bases create very similar smooth magnetic fields, and thus similar E-fields and PNS oracles (see Fig. 2C). This similarity, together with the relatively large distance between the coil windings and the body, does not allow the optimizer to create local PNS cancelation for the studied coil geometries. Instead, the winding optimizer uses macroscopic strategies for PNS reduction, such as bending the B-field away from the torso and redirecting it to smaller body parts (such as the extremities or the head) where they lead to smaller E-field amplitudes, and therefore reduced PNS propensity. In reality, there are of course many sources of PNS threshold variability among subjects, including body shapes, tissue composition (such as body fat percentage) [19, 59], varying dielectric tissue properties, and anatomical and physiological details of the nerve fibers (such as level of myelination, node of Ranvier spacing, etc) [19]. We hypothesize that this variability does not significantly impair the PNS improvements of our optimized windings, as we are aiming for relative threshold improvements (i.e., increasing the PNS threshold averaged over a large population). That being said, it is desirable to test this hypothesis in as many models as possible, and ultimately in threshold experiments on individual subjects whose anatomy differs from our models.
Aside from purely reducing the E-fields in the body, the winding optimizer explored a different strategy: redistribution of the E-fields in the body, by either moving them to body parts without sensitive nerves or balancing E-fields (and thus PNS) in different body parts. This is especially noticeable for the unoptimized and optimized head X-axis coils H1 and H2, where the E-field hotspot moved from the facial area to the chest area (Fig. 6). This led to higher PNS thresholds in the head, but slightly lower PNS thresholds in the shoulder, thus balancing stimulation in these two body parts. Note that this is a valid strategy, as the coil’s PNS thresholds are determined only by the most sensitive nerve, while secondary stimulation sites are less relevant in practice. Even though the E-fields increase in the chest area (reaching ~0.6 V/m at 100 T/m/s), cardiac stimulation requires substantially higher E-field magnitudes (experimental work in mammals indicates E-field thresholds above approx. 40 V/m) [8]. We have recently extended our PNS modeling tool towards electromagnetically induced cardiac stimulation to study this concern in more detail [60].
As shown in Fig. 8C, the optimized coils are most effective in reducing PNS for the scan position used in the optimization. For example, when the body gradients are optimized for head imaging and used with a cardiac imaging body position (30 cm shift), the PNS oracle improvement is reduced from 40% to 15%. For greater shifts of the body in the coil (e.g., corresponding to kidney imaging), the PNS thresholds may even become slightly worse (lower) compared to the unoptimized coil. In this case, it would likely be beneficiary to use a feet-first patient position. The optimization tends to de-symmetrize the coil, placing a high-field-region near the service end. This ensures that the torso is located in the region of reduced B-field magnitude (see Fig. 6, top) for head-first imaging. Alternatively, the coil windings can be jointly optimized for multiple scan positions. For example, for the head and cardiac scan positions this yielded a smaller PNS improvement which was valid for both scan positions (similar PNS oracle values for cardiac and head imaging, see Fig. 8C, red curves). For improved PNS performance at multiple scan positions, it is likely that providing more degrees-of-freedom to the BEM-SF winding optimizer, for example by adding a third coil layer, would be helpful. Finally, while we did not assess feet-first imaging, clinical use cases for the feet-first position such as foot, knee, or wrist imaging would be relatively unconstrained by PNS since a low cross-section body part is present in the high field area at the service end.
The manufacturability of the PNS-optimized winding patterns is likely going to be more complex, compared to conventional designs. The impact on the winding complexity is most prominent for the body coils, where the introduced twist in the windings breaks most of the typical coil symmetries, except the symmetry along the spatial x-axis. The main challenge for manufacturing of the proposed head gradients is likely the additional intermediate winding layer that was needed to provide degrees-of-freedom for PNS reduction (although 3-layer coil designs have been realized before [61]). As with every engineering constraint, the tolerated winding complexity, if quantified, can be freely traded against PNS improvements. We are currently assessing strategies to further constrain the winding optimization to improve manufacturability.
Although only required once, formation of the P-matrix is computationally expensive. Simulation of the PNS response for a single stream function basis (one column in the P-matrix) takes up to 10 minutes, resulting in 2-3 days for the P-matrix computation of a single coil layer. Modification of the coil former shape, changing the body model type (male/female) or its position within the coil requires re-computation of the P-matrix. Fortunately, this lengthy step can be avoided by simply relating a pre-computed P-matrix defined on a Huygens’ surface closely fitting the body model to the coil geometry-specific P-matrix [62]. This approach avoids re-computation of a “new” P-matrix for each new design geometry or body position. The Huygens P-matrix is obtained by populating the Huygens’ surface with magnetic basis functions (such as magnetic dipoles or small coils) and computing the induced E-fields and PNS oracles along all nerves for each basis. The Huygens’ P-matrix is then assembled from the PNS oracle values induced by all basis functions (columns) along all nerve segments (rows). Due to the linearity of the PNS oracle with respect to the coil currents, the P-matrix of a coil geometry (i.e., its current density basis set) can be represented by a weighted superposition of the Huygens’ basis PNS oracles. Linking the Huygens’ P-matrix to a coil geometry specific P-matrix only requires a mapping matrix which can be obtained very efficiently using Biot-Savart simulations of the B-fields generated by both the Huygens’ basis set and coil former geometry and solving an inverse problem.
The P-matrix formulation involves a few important approximations. First, the response of a nerve to the induced E-field is modeled by a linear predictor (the PNS oracle), that has a PNS threshold accuracy of ~7% compared to the reference thresholds obtained using full non-linear neurodynamic models [23, 24]. Other factors of uncertainty arise from typical limitations of the electromagnetic body models itself (such as the topology of conductive tissues and dielectric tissue properties) and the limited number of nerve segments included in our nerve atlases (~1900 segments per model). The body models and the processing pipelines including the electromagnetic simulations that we (and others) use for neurodynamic modeling are constantly being refined [63, 64]. Additional work is also required on a more in-depth analysis of the robustness of the optimized coil layouts with respect to other parameters, such as subject variability, manufacturability and tolerances in coil fabrication. Moreover, there are a number of other engineering metrics such as secondary effects from the cryostat eddy currents [30, 31], acoustic properties [28, 29], and the exact properties of heat dissipation [26, 27] that could be incorporated. Finally, it will be important to validate that the optimized designs achieve the predicted benefits. This will require directly comparing the predicted PNS thresholds to measured thresholds in prototype coils. Fortunately, much of this can likely be done on single-layer coils outside of a magnet (with reduced force and torque concerns). After validating the success of PNS mitigation through B-field reshaping in simplified coils, more realistic gradient prototypes need to be tested.
V. Conclusion
In this work, we demonstrate an approach for incorporation of PNS effects into the numeric optimization of MRI gradient coils. Knowledge of the interaction between the applied magnetic fields and peripheral nerves allows the coil winding optimizer to iteratively refine the windings to minimize PNS at the cost of coil inductance or gradient linearity. Representation of the PNS model in matrix forms maintains the convexity and thus simplifies the optimization process. Our design approach simultaneously considers PNS and traditional engineering parameters (such as field linearity, torque, and coil inductance), and thus allows studying optimal tradeoffs between these metrics and PNS improvements. A comparison of our PNS-optimized coils with conventional gradient coils showed a more than 2-fold PNS improvement with moderate penalties in coil inductance and field linearity, potentially doubling the usable image acquisition performance of state-of-the-art MRI systems in humans.
Supplementary Material
Acknowledgment
The authors would like to acknowledge the help of past and present members of the gradient coil group at Siemens Healthineers, including Peter Dietz, Axel vom Endt, Ralph Kimmlingen, Eva Eberlein, and Franz Hebrank.
Manuscript received May 16, 2020; Research reported in this publication was supported by the NIBIB and the NIMH of the NIH under award numbers R24MH106053, U01EB026996, U01EB025162, U01EB025121, R01EB028250. The content is solely the responsibility of the authors and does not necessarily represent the official views of the National Institutes of Health.
Contributor Information
Mathias Davids, A. A. Martinos Center for Biomedical Imaging, Department of Radiology, Massachusetts General Hospital, Charlestown, MA, United States; Wald are with the Harvard Medical School, Boston, MA, United States; Computer Assisted Clinical Medicine, Medical Faculty Mannheim, Heidelberg University, Heidelberg, Germany.
Bastien Guérin, A. A. Martinos Center for Biomedical Imaging, Department of Radiology, Massachusetts General Hospital, Charlestown, MA, United States; Wald are with the Harvard Medical School, Boston, MA, United States.
Valerie Klein, A. A. Martinos Center for Biomedical Imaging, Department of Radiology, Massachusetts General Hospital, Charlestown, MA, United States; Computer Assisted Clinical Medicine, Medical Faculty Mannheim, Heidelberg University, Heidelberg, Germany.
Lawrence L. Wald, A. A. Martinos Center for Biomedical Imaging, Department of Radiology, Massachusetts General Hospital, Charlestown, MA, United States; Wald are with the Harvard Medical School, Boston, MA, United States; Harvard-MIT Division of Health Sciences and Technology, Cambridge, MA, United States
References
- [1].Setsompop K et al. , “Pushing the limits of in vivo diffusion MRI for the human connectome project,” NeuroImage, vol. 80, pp. 220–233, 2013 [DOI] [PMC free article] [PubMed] [Google Scholar]
- [2].McNab JA et al. , “The human connectome project and beyond: Initial applications of 300 mT/m gradients,” NeuroImage, vol. 80, pp. 234–245, 2013 [DOI] [PMC free article] [PubMed] [Google Scholar]
- [3].Budinger TF, “Thresholds for physiological effects due to RF and magnetic fields used in NMR imaging,” IEEE Trans Nucl Sci, vol. 26, no. 2, pp. 2821–2825, 1979. [Google Scholar]
- [4].Cohen M, Weisskoff R, and Kantor H, “Evidence of peripheral stimulation by time-varying magnetic fields,” in Proceedings of the 75th anniversary scientific assembly and annual meeting Radiological Society of North America (Abstracts), 1989. [Google Scholar]
- [5].Cohen MS, Weisskoff RM, Rzedzian RR, and Kantor HL, “Sensory stimulation by time-varying magnetic fields,” Magn Reson Med, vol. 14, no. 2, pp. 409–414, 1990. [DOI] [PubMed] [Google Scholar]
- [6].Irnich W and Schmitt F, “Magnetostimulation in MRI,” Magn Reson Med, vol. 33, no. 5, pp. 619–623, 1995. [DOI] [PubMed] [Google Scholar]
- [7].Tan ET et al. , “Peripheral nerve stimulation limits of a high amplitude and slew rate magnetic field gradient coil for neuroimaging,” Magn Reson Med, vol. 83, no. 1, pp. 352–366, 2020. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [8].Irnich W, “Electrostimulation by time-varying magnetic fields,” Magn Reson Mater Phy, vol. 2, no. 1, pp. 43–49, 1994. [Google Scholar]
- [9].Den Boer JA et al. , “Comparison of the threshold for peripheral nerve stimulation during gradient switching in whole body MR systems,” J Magn Reson Imaging, vol. 15, no. 5, pp. 520–525, 2002. [DOI] [PubMed] [Google Scholar]
- [10].Irnich W and Hebrank FX, “Stimulation threshold comparison of time-varying magnetic pulses with different waveforms,” J Magn Reson Imaging, vol. 29, no. 1, pp. 229–236, 2009. [DOI] [PubMed] [Google Scholar]
- [11].Feldman RE, Hardy CJ, Aksel B, Schenck J, and Chronik BA, “Experimental determination of human peripheral nerve stimulation thresholds in a 3-axis planar gradient system,” Magn Reson Med, vol. 62, no. 3, pp. 763–770, 2009. [DOI] [PubMed] [Google Scholar]
- [12].Zhang B, Yen Y-F, Chronik BA, McKinnon GC, Schaefer DJ, and Rutt BK, “Peripheral nerve stimulation properties of head and body gradient coils of various sizes,” Magn Reson Med, vol. 50, no. 1, pp. 50–58, 2003. [DOI] [PubMed] [Google Scholar]
- [13].Hidalgo-Tobon SS, Bencsik M, and Bowtell R, “Reducing peripheral nerve stimulation due to gradient switching using an additional uniform field coil,” Magn Reson Med, vol. 66, no. 5, pp. 1498–1509, 2011. [DOI] [PubMed] [Google Scholar]
- [14].Peeren G, “Stream function approach for determining optimal surface currents,” J Comput Phys, vol. 191, no. 1, pp. 305–321, October 2003. [Google Scholar]
- [15].Lemdiasov RA and Ludwig R, “A stream function method for gradient coil design,” Concepts Magn Reson, vol. 26B, no. 1, pp. 67–80, August 2005. [Google Scholar]
- [16].Poole M and Bowtell R, “Novel gradient coils designed using a boundary element method,” Concepts Magn Reson, vol. 31B, no. 3, pp. 162–175, August 2007. [Google Scholar]
- [17].Davids M, Guérin B, Schad LR, and Wald LL, “Realistic simulation of peripheral nerve stimulations in magnetic particle imaging for improved MPI molecular imaging in humans,” Proceedings of the World Molecular Imaging Congress, New York, NY, USA, 2016. [Google Scholar]
- [18].Davids M, Guérin B, Schad LR, and Wald LL, “Predicting magnetostimulation thresholds in the peripheral nervous system using realistic body models,” Sci Rep, vol. 7, no. 1, p. 5316, 2017. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [19].Klein V, Davids M, Wald LL, Schad LR, and Guérin B, “Sensitivity analysis of neurodynamic and electromagnetic simulation parameters for robust prediction of peripheral nerve stimulation,” Phys Med Biol, 2018. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [20].Davids M, Guérin B, Schad LR, and Wald LL, Peripheral Nerve Stimulation Modeling for MRI. Wiley, 2019, pp. 87–102. [Google Scholar]
- [21].Davids M, Guérin B, Klein V, Schad LR, and Wald LL, “Simulation of peripheral nerve stimulation thresholds of MRI gradient coils,” Proceedings of the 26th Annual Meeting of ISMRM, Paris, France, 2018. [Google Scholar]
- [22].Davids M, Guérin B, Schad LR, and Wald LL, “Prediction of peripheral nerve stimulation thresholds of MRI gradient coils using coupled electromagnetic and neurodynamic simulations,” Proceedings of the 27th Annual Meeting of ISMRM, Montreal, Canada, 2019. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [23].Davids M, Guérin B, Klein V, Schad LR and Wald LL, “The PNS oracle: a modified neural activation function metric for rapid assessment of peripheral nerve stimulation (PNS),” Proceedings of the 27th Annual Meeting of ISMRM, Montreal, Canada, 2019. [Google Scholar]
- [24].Davids M, Guérin B, Klein V, Schmelz M, Schad LR, and Wald LL, “Optimizing selective stimulation of peripheral nerves with arrays of coils or surface electrodes using a linear peripheral nerve stimulation metric,” J Neural Eng, vol. 17, no. 1, p. 016029, 2020. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [25].Davids M, Guérin B, Schad LR, and Wald LL, “Peripheral nerve stimulation (PNS) constrained gradient coil design within a boundary element method stream function (BEM-SF) optimization,” Proceedings of the 27th Annual Meeting of ISMRM, Montreal, Canada, 2019. [Google Scholar]
- [26].Poole MS, While PT, Lopez HS, and Crozier S, “Minimax current density gradient coils: Analysis of coil performance and heating,” Magn Reson Med, vol. 68, no. 2, pp. 639–648, 2012. [DOI] [PubMed] [Google Scholar]
- [27].While PT, Poole MS, Forbes LK, and Crozier S, “Minimum maximum temperature gradient coil design,” Magn Reson Med, vol. 70, no. 2, pp. 584–594, 2013. [DOI] [PubMed] [Google Scholar]
- [28].Edelstein WA, Hedeen RA, Mallozzi RP, El-Hamamsy S-A, Ackermann RA, and Havens TJ, “Making MRI quieter,” Magn Reson Imaging, vol. 20, no. 2, pp. 155–163, 2002. [DOI] [PubMed] [Google Scholar]
- [29].Winkler SA, Alejski A, Wade T, McKenzie CA, and Rutt BK, “On the accurate analysis of vibroacoustics in head insert gradient coils,” Magn Reson Med, vol. 78, no. 4, pp. 1635–1645, 2017. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [30].Trakic A, Liu F, Lopez HS, Wang H, and Crozier S, “Longitudinal gradient coil optimization in the presence of transient eddy currents,” Magn Reson Med, vol. 57, no. 6, pp. 1119–1130, 2007. [DOI] [PubMed] [Google Scholar]
- [31].Sanchez Lopez H et al. , “Multilayer integral method for simulation of eddy currents in thin volumes of arbitrary geometry produced by MRI gradient coils,” Magn Reson Med, vol. 71, no. 5, pp. 1912–1922, 2014. [DOI] [PubMed] [Google Scholar]
- [32].McIntyre CC and Grill MW, “Sensitivity analysis of a model of mammalian neural membrane,” Biol Cybern, vol. 79, no. 1, pp. 29–37, 1998. [DOI] [PubMed] [Google Scholar]
- [33].Richardson AG, McIntyre CC, and Grill WM, “Modelling the effects of electric fields on nerve fibres: Influence of the myelin sheath,” Med Biol Eng Comput, vol. 38, no. 4, pp. 438–446, 2000. [DOI] [PubMed] [Google Scholar]
- [34].McIntyre CC, Richardson AG, and Grill WM, “Modeling the excitability of mammalian nerve fibers: Influence of afterpotentials on the recovery cycle,” J Neurophysiol, vol. 87, no. 2, pp. 995–1006, 2002. [DOI] [PubMed] [Google Scholar]
- [35].Rattay F, “Analysis of models for external stimulation of axons,” IEEE Trans Bio-Med Eng, vol. BME-33, no. 10, pp. 974–977, October 1986. [DOI] [PubMed] [Google Scholar]
- [36].Handler WB et al. , “New head gradient coil design and construction techniques,” J Magn Reson Imaging, vol. 39, no. 5, pp. 1088–1095, 2014. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [37].Kimmlingen R et al. , “An easy to exchange high performance head gradient insert for a 3T whole body MRI system: First results,” Proceedings of the 12th Annual Meeting of ISMRM, Kyoto, Japan, 2004. [Google Scholar]
- [38].Hidalgo-Tobon S, “Theory of gradient coil design methods for Magnetic Resonance Imaging,” Concepts Magn Reson, vol. 36A, no. 4, pp. 223–242, 2010. [Google Scholar]
- [39].Sanchez C, Pantoja M, Poole M, and Bretones A, “Gradient-coil design: A multi-objective problem,” IEEE Trans Magn, vol. 48, no. 6, pp. 1967–1975, June 2012. [Google Scholar]
- [40].Silveira LM, Kamon M, and White J, “Efficient reduced-order modeling of frequency-dependent coupling inductances associated with 3D interconnect structures,” IEEE Trans Compon Packaging Manuf Technol, vol. 19, no. 2, pp. 283–288, 1996. [Google Scholar]
- [41].Davids M, Guérin B, vom Endt A, Schad LR, and Wald LL, “Prediction of peripheral nerve stimulation thresholds of MRI gradient coils using coupled electromagnetic and neurodynamic simulations,” Magn Reson Med, vol. 81, no. 1, pp. 686–701, 2019. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [42].Bernstein MA et al. , “Concomitant gradient terms in phase contrast MR: Analysis and correction,” Magn Reson Med, vol. 39, no. 2, pp. 300–308, 1998. [DOI] [PubMed] [Google Scholar]
- [43].Meier C, Zwanger M, Feiweier T, and Porter D, “Concomitant field terms for asymmetric gradient coils: Consequences for diffusion, flow, and echo-planar imaging,” Magn Reson Med, vol. 60, no. 1, pp. 128–134, 2008. [DOI] [PubMed] [Google Scholar]
- [44].Tao S et al. , “Gradient pre-emphasis to counteract first-order concomitant fields on asymmetric MRI gradient systems,” Magn Reson Med, vol. 77, no. 6, pp. 2250–2262, 2017. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [45].Weavers PT et al. , “B0 concomitant field compensation for MRI systems employing asymmetric transverse gradient coils,” Magn Reson Med, vol. 79, no. 3, pp. 1538–1544, 2018. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [46].Tao S et al. , “The effect of concomitant fields in fast spin echo acquisition on asymmetric MRI gradient systems,” Magn Reson Med, vol. 79, no. 3, pp. 1354–1364, 2018. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [47].Foo TKF et al. , “Lightweight, compact, and high-performance 3T MR system for imaging the brain and extremities,” Magn Reson Med, vol. 80, no. 5, pp. 2232–2245, 2018. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [48].Glover GH and Pelc NJ, “Method for correcting image distortion due to gradient nonuniformity,” May 27 1986, uS Patent 4,591,789.
- [49].Tao S et al. , “Noncartesian MR image reconstruction with integrated gradient nonlinearity correction,” Med Phys, vol. 42, no. 12, pp. 7190–7201, 2015. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [50].Tao S et al. , “Partial fourier and parallel MR image reconstruction with integrated gradient nonlinearity correction,” Magn Reson Med, vol. 75, no. 6, pp. 2534–2544, 2016. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [51].Mansfield P and Haywood B, “Controlled E-field gradient coils for MRI,” Phys Med Biol, vol. 53, no. 7, p. 1811, 2008. [DOI] [PubMed] [Google Scholar]
- [52].Wade TP, Alejski A, McKenzie CA, and Rutt BK, “Peripheral nerve stimulation thresholds of a high performance insertable head gradient coil,” Proceedings of the 24th Annual Meeting of ISMRM, Singapore, 2016. [Google Scholar]
- [53].Weiger M et al. , “A high-performance gradient insert for rapid and short-T2 imaging at full duty cycle,” Magn Reson Med, vol. 79, no. 6, pp. 3256–3266, 2018. [DOI] [PubMed] [Google Scholar]
- [54].While PT, Forbes LK, and Crozier S, “3D gradient coil design: toroidal surfaces,” J Magn Reson, vol. 198, no. 1, pp. 31–40, 2009. [DOI] [PubMed] [Google Scholar]
- [55].While PT, Forbes LK, and Crozier S, “3D gradient coil design for open MRI systems,” J Magn Reson, vol. 207, no. 1, pp. 124–133, 2010. [DOI] [PubMed] [Google Scholar]
- [56].Kroboth S et al. , “Optimization of a switching circuit for a matrix gradient coil,” Proceedings of the 24th Annual Meeting of ISMRM, Singapore, 2016. [Google Scholar]
- [57].Littin S et al. , “Development and implementation of an 84-channel matrix gradient coil,” Magn Reson Med, vol. 79, no. 2, pp. 1181–1191, 2018. [DOI] [PubMed] [Google Scholar]
- [58].Kroboth S et al. , “Switching circuit optimization for matrix gradient coils,” Tomography, vol. 5, no. 2, p. 248, 2019. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [59].Demirel O, Kilic T, Cukur T, and Saritas E, “Simple anatomical measures correlate with individual PNS thresholds for kHz-range homogeneous magnetic fields,” Proceedings of the 28th Annual Meeting of ISMRM, 2020. [DOI] [PubMed] [Google Scholar]
- [60].Klein V, Davids M, Schad LR, Wald LL, and Guérin B, “Investigating cardiac stimulation limits of MRI gradient coils using electromagnetic and electrophysiological simulations in human and canine body models,” Magn Reson Med, (in press) 2020. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [61].Foo TKF et al. , “Highly efficient head-only magnetic field insert gradient coil for achieving simultaneous high gradient amplitude and slew rate at 3.0T (MAGNUS) for brain microstructure imaging,” Magn Reson Med, vol. 83, no. 6, pp. 2356–2369, 2020. [DOI] [PubMed] [Google Scholar]
- [62].Davids M, Guérin B, and Wald LL, “Assembly of a PNS predicting “P-matrix” on a Huygens’ surface for rapid PNS assessment of 2D or 3D gradient coil windings,” Proceedings of the 28th Annual Meeting of ISMRM, Sydney, Australia, 2020. [Google Scholar]
- [63].Neufeld E, Cassará AM, Montanaro H, Kuster N, and Kainz W, “Functionalized anatomical models for EM-neuron interaction modeling,” Phys Med Biol, vol. 61, no. 12, pp. 4390–4401, 2016. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [64].Cassara A, Lloyd B, Farcito S, Neufeld E, and Hagberg KN, Gisela, “Neurostimulation safety investigations using reference, posable anatomical models,” Proceedings of the 27th Annual Meeting of ISMRM, Montreal, Canada, 2019. [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.