Abstract
Objectives:
Modern reconstruction and post-processing software aims at reducing image noise in CT images, potentially allowing for a reduction of the employed radiation exposure. This study aimed at assessing the influence of a novel deep-learning based software on the subjective and objective image quality compared to two traditional methods [filtered back-projection (FBP), iterative reconstruction (IR)].
Methods:
In this institutional review board-approved retrospective study, abdominal low-dose CT images of 27 patients (mean age 38 ± 12 years, volumetric CT dose index 2.9 ± 1.8 mGy) were reconstructed with IR, FBP and, furthermore, post-processed using a novel software. For the three reconstructions, qualitative and quantitative image quality was evaluated by means of CT numbers, noise, signal-to-noise ratio (SNR), and contrast-to-noise ratio (CNR) in six different ROIs. Additionally, the reconstructions were compared using SNR, peak SNR, root mean square error and mean absolute error to assess structural differences.
Results:
On average, CT numbers varied within 1 Hounsfield unit (HU) for the three assessed methods in the assessed ROIs. In soft tissue, image noise was up to 42% lower compared to FBP and up to 27% lower to IR when applying the novel software. Consequently, SNR and CNR were highest with the novel software. For both IR and the novel software, subjective image quality was equal but higher than the image quality of FBP-images.
Conclusion:
The assessed software reduces image noise while maintaining image information, even in comparison to IR, allowing for a potential dose reduction of approximately 20% in abdominal CT imaging.
Advances in knowledge:
The assessed software reduces image noise by up to 27% compared to IR and 48% compared to FBP while maintaining the image information.
The reduced image noise allows for a potential dose reduction of approximately 20% in abdominal imaging.
Introduction
The radiation exposure in CT with its concomitant risk of inducing cancerous diseases motivates the need for dose reduction and optimization of scan protocols.1,2 Advancements for dose reduction in CT are based on hardware (i.e. increasing detector sensitivity), on scan protocol settings (reduction of tube potential and/or tube current–time product) and on software (automatic exposure control, reconstruction techniques and post-processing methods).3 For scan protocol settings with reduced radiation exposure to the patient, the image quality deteriorates since the image detector receives less signal due to the decreased radiation exposure. Depending on the underlying image reconstruction method, the effect of dose reduction on the processed image varies.
Traditional filtered back-projection (FBP) is a computationally simple technique, where the measured raw data from projection space are transformed to processed image space. FBP is a fast and robust technique, requires limited computational power and is applicable in clinical routine.4 However, the reconstructed images can appear noisy if low dose techniques are employed or only insufficient projection data are available, causing a lower visibility of low-contrast structures and image artifacts are often significant.4–6
Iterative reconstruction (IR) techniques provide a lower amount of image noise and improve visibility of low-contrast structures.7,8 Due to the reduced image noise, IR allows to decrease the applied radiation dose in CT while maintaining image quality.6,9 However, reconstruction time is considerably longer than FBP. Another side-effect of IR is the altered image appearance: Images often look artificially smooth (plastic-like) due to altered signal values (pixel data).4,5,10,11
In the last years, more and more software has been developed for reduction of image noise in CT. In literature, some of the software has been described previously.12–15 The commercial product PixelShine (AlgoMedica, Inc., Sunnyvale, CA) is a novel software based on an artificial neural network technique (deep learning) that reduces image noise in CT.16 Ultra-large amounts of high-dose and low-dose data have been used for training of the software.17 After training, the software makes use of filtered back-projected CT images for post-processing. In contrast, the techniques FBP or IR use projection data (raw data) as input for image reconstruction instead if image data. The vendor claims that the software decreases image noise without altering the image information or rendering the image appearance to blotchy or plastic-like impression. Up to now, only one full paper covers the application and value of PixelShine in human imaging (low-dose pelvic arterial phase images),17 whereas just two others cover phantom studies.16,18 Further studies need to be performed to evaluate the software in other clinical settings.
When performing low-dose CT imaging of the abdomen, considerable image noise occurs especially for patients with high body mass indices or large effective diameters due to increased absorption of X-rays in the abdominal (fat) tissue. In the institution, where this study is performed, low-dose abdominal imaging is performed in cases of urinary stone diagnostics or in case of suspected drug body packing as part of a criminal investigation. These indications allow for low-dose imaging since urinary stones or body packs are usually high-contrast materials, however, only if IR for noise suppression is applied.19 The drawback of IR images is the increased blotchiness of images with decreasing noise level. Therefore, the aim of the study was to evaluate whether PixelShine might be an alternative to IR or even lead to a superior image quality than IR in human low-dose abdominal imaging. For this purpose, the study evaluates and compares the subjective and objective image quality of images reconstructed with FBP, IR and PixelShine in low-dose abdominal CT examinations.
Methods and materials
In this IRB-approved retrospective study, 27 patients (19 male, 8 female) receiving a low-dose abdominal CT scan between November 2014 and February 2016 were included. The examined patients were suspected of body packing, therefore, a low-dose protocol was employed. A full-dose abdominal CT scan was not performed in these patients.
Image acquisition and post-processing
CT protocol parameters were as follows: tube potential 80 kVp, reference tube current–time product 160 mAs, 0.5 s rotation time, 0.6 mm collimation, pitch 0.6. Images were acquired on a Somatom Definition Flash CT scanner (Siemens Healthineers, Forchheim, Germany). The scan range covered the diaphragm to the ischium.
Raw CT data were reconstructed and post-processed as follows:
FBP: 2.0 mm slice thickness, kernel B30f, increment 1.7 mm.
PS: FBP (see reconstruction 1) with PixelShine (PS), settings: P214A8S.
IR: 2.0 mm slice thickness, kernel I30f, level 3, increment 1.7 mm (SAFIRE, Siemens Healthineers, Fochheim, Germany).
SAFIRE (sinogram-affirmed iterative reconstruction) utilizes both raw data (projection data) and image data iterations with up to five strength levels to control for image impression and noise reduction.4 In SAFIRE, the initial reconstruction is performed using FBP images, after which two different correction loops are implemented, during which noise is reduced.
PixelShine v. 1.2.104’s adaptive processing mode was employed with the following reconstruction parameters for abdominal imaging: sharpening level P2 (a linear filter applied before the noise reduction process), target noise level 14 (based on Hounsfield unit, HU), maximum processing strength A8 (55% noise reduction, equivalent dose or thickness x4.9), and smoothing type S (soft kernel). In the adaptive processing mode, the software estimates noise levels in the original images and adaptively applies appropriate processing strength to obtain the target noise level images. The vendor (AlgoMedica) specified the parameters for the combination of the CT scanner type and abdominal imaging (soft tissue). Since PixelShine is a commercial product, we were not able to obtain information related to the deep-learning structure, training and operating principle of the software.
Objective image analysis
For all patients and data sets, circular regions-of-interest (ROIs) (r ≈ 10 mm) were drawn in the liver, spleen, aorta, paravertebral muscle, fat and in the background (air) using the software ImageJ, v. 1.52 n (National Institutes of Health, USA). For each patient, ROIs were drawn in exactly the same position for each reconstruction method. Mean CT number and standard deviation (noise) were determined in each ROI.
Contrast-to-noise ratio (CNR) and signal-to-noise (SNRROI) ratio were calculated using the following formulas:
for the muscle, spleen, liver and the aorta and
for all ROIs with CTtissue or CTROI as CT number of a certain ROI and SDfat or SDROI for the standard value of a certain ROI.
Furthermore, for the total image data sets, the SNR (in decibel, dB), peak SNR (PSNR, in dB), root mean square error (RMSE) and mean absolute error (MAE) were calculated with the FBP image data set as reference images, using the plugin “SNR” for the software ImageJ, v. 1.52 n (National Institutes of Health, USA).20 The calculation for each quantity is performed image-wise. The measures were then averaged over the all images for each patient and reconstruction.
Noise-images were acquired using the software ImageJ, v. 1.52 n (National Institutes of Health, USA). ImageJ offers the possibility to subtract images pixelwise. This was performed to calculate subtraction images from the three processed image data sets.
Subjective image analysis
Two radiologists (10 and 2 years of experience in radiology) independently rated image quality using a 5-point Likert scale [(1) excellent image quality, (2) good image quality, (3) moderate image quality, (4) poor image quality, (5) non-diagnostic image quality].
Statistical analysis
For both the objective and the subjective image quality, the following comparisons were obtained:
Comparison PS with FBP
Comparison PS with IR
Comparison IR with FBP
For all 6 ROIs, the CT numbers and image noise (12 data sets per reconstruction type) were compared and evaluated.
Commercially available software (Microsoft Excel 2016, Redmond, WA; SPSS, 26.0, Inc., Chicago, IL) was used to perform statistical analysis. Continuous data are expressed as mean ± SD. Differences in objective and subjective image quality among the reconstruction methods were calculated by means of a Bonferroni-corrected Wilcoxon-test using a significance level dependent on the number of evaluated tests (ROIs and comparisons). Inter-reader reliability was determined using a Fleiss κ with chance-correction according to Brennan and Prediger.21
Results
Patient collective and effective dose
In this study, 27 patients (19 male, 8 female) with a mean age of 38 ± 12 years (range 20–58) and an average effective abdominal diameter of 29.3 cm (measured on central reconstructed CT slice () were included (c.f. Table 1). A CTDIvol of 2.9 ± 1.8 mGy and a DLP of 137.6 ± 88.6 mGy∙cm with an effective tube current–time product of 148.0 ± 93.9 mAs at 80 kVp resulted in an average effective dose of 2.24 ± 1.44 mSv (conversion factor of 0.0163 mSv/(mGy∙cm) for the abdomen22).
Table 1.
Parameter | Mean | ± | SD |
---|---|---|---|
Age (years) | 37.6 | ± | 11.5 |
Effective TCTP (mAs) | 148.0 | ± | 93.9 |
CTDIvol (mGy) | 2.9 | ± | 1.8 |
DLP (mGy∙cm) | 137.6 | ± | 88.6 |
Effective dose (mSv) | 2.24 | ± | 1.44 |
Scan length (cm) | 46.5 | ± | 4.1 |
Diameter lateral (mm) | 349.6 | ± | 57.5 |
Diameter ap (mm) | 246.1 | ± | 55.6 |
Effective diameter (mm) | 293.0 | ± | 55.6 |
AP, anteroposterior; CTDIvol, volumetric computed tomography dose index; DLP, dose–length product; SD, standard deviation;effective TCTP, effective tube current–time product.
Objective image quality
CT number
The average CT numbers obtained by the three reconstruction methods for the evaluated ROI positions were within ±1 HU (absolute difference) or 3% (relative difference), see Tables 2 and 3. There were significant differences between CT numbers in the muscle and aorta (PS vs FBP and PS vs IR) and in the air (IR vs FBP), although the maximum difference in CT numbers was only 3 HU, see Tables 2–4.
Table 2.
Liver | Muscle | Spleen | Air | Fat | Aorta | |
---|---|---|---|---|---|---|
CT number (HU) | ||||||
FBP | 49.5 ± 14.7 | 52.9 ± 10.7 | 36.7 ± 10.1 | −992.1 ± 10.8 | −126.8 ± 12.6 | 45.3 ± 9.4 |
PS | 49.4 ± 14.8 | 52.2 ± 10.6 | 36.4 ± 10.2 | −992.5 ± 10.6 | −126.6 ± 12.3 | 44.2 ± 9.1 |
IR | 49.6 ± 14.8 | 52.9 ± 10.5 | 36.8 ± 10.1 | −992.8 ± 11.2 | −126.7 ± 12.6 | 45.4 ± 9.1 |
Noise (HU) | ||||||
FBP | 45.9 ± 9.6 | 42.7 ± 7.6 | 43.9 ± 9.3 | 22.2 ± 12.4 | 35.3 ± 5.6 | 36.6 ± 6.7 |
PS | 26.8 ± 7.5 | 24.0 ± 6.2 | 25.1 ± 6.6 | 10.0 ± 13.4 | 18.0 ± 4.9 | 24.8 ± 5.1 |
IR | 31.7 ± 6.7 | 29.6 ± 5.4 | 30.5 ± 6.3 | 16.9 ± 13.0 | 24.9 ± 4.2 | 26.6 ± 5.0 |
FBP: Reconstruction with filtered back-projection, HU: Hounsfield unit, IR: Iterative reconstructed images; PS: PixelShine employed on FBP-images;SD, standard deviation.
Table 3.
Liver | Muscle | Spleen | Air | Fat | Aorta | |
---|---|---|---|---|---|---|
CT number | ||||||
Absolute difference (HU) | ||||||
PS vs FBP | −0.1 ± 0.5 | −0.7 ± 0.7 | −0.3 ± 0.7 | −0.4 ± 0.7 | 0.2 ± 0.7 | −1.1 ± 0.9 |
PS vs IR | −0.2 ± 0.5 | −0.7 ± 0.5 | −0.3 ± 0.6 | 0.4 ± 1.1 | 0.1 ± 0.7 | −1.2 ± 1.1 |
IR vs FBP | 0.0 ± 0.4 | 0.0 ± 0.4 | 0.1 ± 0.4 | −0.7 ± 0.8 | 0.1 ± 0.6 | 0.1 ± 0.7 |
Relative difference (%) | ||||||
PS vs FBP | −0.4 ± 1.8 | −1.2 ± 1.4 | −1.1 ± 4.3 | 0.0 ± 0.1 | −0.1 ± 0.6 | −2.4 ± 1.9 |
PS vs IR | −0.2 ± 1.8 | −1.4 ± 1.2 | −1.3 ± 4.3 | 0.0 ± 0.1 | −0.1 ± 0.5 | 2.8 ± 2.5 |
IR vs FBP | −0.1 ± 2.2 | 0.2 ± 1.1 | 0.2 ± 1.4 | 0.1 ± 0.1 | −0.1 ± 0.5 | 0.4 ± 1.8 |
Noise | ||||||
Absolute difference (HU) | ||||||
PS vs FBP | −19.1 ± 7.2 | −18.7 ± 7.0 | −18.8 ± 7.1 | −12.2 ± 3.8 | −17.3 ± 6.4 | −11.7 ± 5.1 |
PS vs IR | −4.9 ± 5.0 | −5.6 ± 5.3 | −5.4 ± 4.8 | −6.9 ± 2.3 | −6.9 ± 4.9 | −1.8 ± 3.4 |
IR vs FBP | −14.2 ± 3.3 | −13.1 ± 2.6 | −13.4 ± 3.4 | −5.3 ± 1.7 | −10.4 ± 2.1 | −9.9 ± 2.4 |
Relative difference (%) | ||||||
PS vs FBP | −41.2 ± 12.5 | −43.3 ± 13.0 | −42.4 ± 11.9 | −62.5 ± 18.8 | −48.2 ± 14.8 | −31.4 ± 12.0 |
PS vs IR | −15.3 ± 16.0 | −18.4 ± 17.4 | −17.4 ± 15.5 | −49.7 ± 18.4 | −27.1 ± 18.7 | −6.1 ± 14.3 |
IR vs FBP | −30.9 ± 2.7 | −30.7 ± 2.5 | −30.4 ± 2.9 | −27.6 ± 8.4 | −29.4 ± 3.9 | −27.1 ± 4.1 |
FBP: reconstruction with filtered back-projection, HU: Hounsfield unit, IR: iterative reconstructed images; PS: PixelShine employed on FBP-images;SD, standard deviation.
Table 4.
Liver | Muscle | Spleen | Air | Fat | Aorta | |
---|---|---|---|---|---|---|
CT number | ||||||
PS vs FBP | 0.179 | 0.000 | 0.032 | 0.007 | 0.178 | 0.000 |
PS vs IR | 0.088 | 0.000 | 0.007 | 0.058 | 0.588 | 0.000 |
IR vs FBP | 0.492 | 0.879 | 0.568 | 0.000 | 0.525 | 0.486 |
Noise | ||||||
PS vs FBP | 0.000 | 0.000 | 0.000 | 0.000 | 0.000 | 0.000 |
PS vs IR | 0.000 | 0.000 | 0.000 | 0.000 | 0.000 | 0.000 |
IR vs FBP | 0.000 | 0.000 | 0.000 | 0.000 | 0.000 | 0.000 |
FBP: reconstruction with filtered back-projection, IR: iterative reconstructed images; PS: PixelShine employed on FBP-images.
A p-value of 0.0028 was considered statistically significant (3 × 6 comparisons → p = 0.5/18 = 0.0028).
Image noise
Image noise was lowest when employing PixelShine, and highest when using FBP, see Tables 2 and 3. See Figure 1 for an abdominal slice and Figure 2 for zoomed sections of the slice, reconstructed with the three evaluated methods. Between PS and FBP, the noise in the evaluated ROIs is reduced by 12–19 HU (air: average reduction of 12 HU or 63%, tissues: reduction of approximately 18 HU or 40–48%, aorta: reduction of 12 HU or 31%)), see Tables 2–4. Comparing PS and IR, the image noise was lower in PS reconstructions by 2–7 HU (muscle, liver, spleen −15–18%, air-50%, fat −27%, aorta −6%). Image noise was approximately 30% lower in all evaluated ROIs when employing IR compared to FBP. Significant differences were visible for all evaluated ROI positions and reconstruction methods, see Table 4.
SNR in ROIs
The absolute value of the SNR was highest (Table 5) for images reconstructed with PS and lowest for FBP images. Differences among the reconstruction methods were statistically significant for all evaluated ROIs (p < 0.001).
Table 5.
Liver | Muscle | Spleen | Air | Fat | Aorta | |
---|---|---|---|---|---|---|
SNR | ||||||
FBP | 1.2 ± 0.5 | 1.3 ± 0.4 | 0.9 ± 0.4 | −51.3 ± 13.9 | −3.7 ± 0.7 | 1.3 ± 0.4 |
PS | 2.0 ± 1.0 | 2.3 ± 0.9 | 1.6 ± 0.7 | −168.7 ± 74.7 | −7.7 ± 2.8 | 1.9 ± 0.8 |
IR | 1.7 ± 0.7 | 1.9 ± 0.6 | 1.3 ± 0.5 | −73.0 ± 22.2 | −5.2 ± 1.0 | 1.8 ± 0.6 |
CNR | ||||||
FBP | 5.1 ± 1.1 | 5.2 ± 1.0 | 4.7 ± 0.9 | 5.0 ± 0.9 | ||
PS | 10.6 ± 3.6 | 10.7 ± 3.3 | 9.8 ± 3.0 | 10.4 ± 3.9 | ||
IR | 7.3 ± 1.5 | 7.4 ± 1.4 | 6.7 ± 1.2 | 7.1 ± 1.3 |
CNR: contrast-to-noise ratio, FBP: reconstruction with filtered back-projection, IR: iterative reconstructed images; PS: PixelShine employed on FBP-images, ROI: region of interest,SD, standard deviation; SNR: signal-to-noise ratio.
Differences between reconstruction methods for SNR were significant (p < 0.001) for all ROIs except for the aorta (differences between IR and PS: p = 0.155). Differences between reconstruction methods for CNR were significant (p < 0.001) for all ROIs.
CNR in ROIs
Between the four evaluated ROIs, CNR was highest for the liver and muscle (Table 5). Between the three evaluated reconstruction techniques, CNR was highest when using PS, due to the low image noise (see last section), increasing by about 40% compared to IR images and 90% compared to FBP. Differences among the reconstruction methods were statistically significant for all evaluated ROIs (p < 0.001).
SNR, PSNR, RMSE, MAE in image data sets
Comparing PS and IR image sets to the corresponding FBP image sets, SNR and PSNR were higher when comparing IR and FBP data sets with PS and FBP images (SNRIR-FBP 35.4 ± 1.4 dB vs SNRPS-FBP 33.6 ± 2.7 dB and PSNRIR-FBP to 43.2 ± 1.5 dB vs PSNRPS-FBP 41.4 ± 2.8 dB). The RMSE and MAE were lower when comparing IR and FBP data sets with PS and FBP data sets (RMSEIR-FBP 11.7 ± 1.4 vs RMSEPS-FBP 15.1 ± 4.0 and MAEIR-FBP 8.3 ± 1.2 vs MAEPS-FBP 11.8 ± 3.0).
Subjective image quality
The image quality of the abdominal CT reconstructions was moderate to poor, since a low-dose protocol was employed (Figures 1 and 3). However, the image quality was sufficient for the detection of foreign materials. For the two raters, inter-reader reliability was fair to moderate (κ = 0.40–0.49). Applying solely FBP, the image quality was poor (4.0 ± 0.8), whereas it was moderate when applying PS (3.3 ± 0.7) or IR (3.3 ± 0.8). There was a significant difference in image quality between PS- and FBP-reconstructions and between IR- and FBP-reconstructions (p < 0.001, Figure 3). No significant difference in image quality was noted between PS- and IR-reconstructions although the visual impression differed (p = 0.876, Figures 1 and 2). As visible in Figures 1 and 2, the liver appears considerably noisier in FBP images compared to IR- or PS-processed images. The PS-processed images seem to be more homogeneous and smoother in each structure, removing little coarseness. With the increased homogeneity in the organs, IR- and PS-processed images appear less contrasty, compared to the FBP images. The difference in visual impression gets obvious when subtracting the identical axial slice for the three reconstruction techniques (Figure 4). The stronger image noise is obvious when comparing PS- or IR-images with FBP-images, especially in the liver or spleen.
Discussion
In this retrospective study, the influence of the post-processing algorithm PixelShine (AlgoMedica, Sunnyvale, CA) on the image characteristics (CT number and noise) and image quality in comparison to two standard CT reconstruction algorithms (FBP and IR) was investigated. For the purpose of this study, a low-dose protocol of the abdomen was assessed. Image noise is often pronounced in low-dose acquisitions so that differences in image noise can become obvious when altering the reconstruction process.
Objective image quality
IR is known to alter the textural features compared to FBP by means of a blotchy appearance or induced/increased artefacts.11 PS claims to provide for CT images with exactly the same information as with FBP, however, while still reducing the image noise. Although significant differences in the CT numbers for the evaluated reconstruction methods were notable for the muscle and aorta, those differences were within 3 HU.
On average, CT numbers between the methods varied by less than 1 HU (maximum absolute difference of 3 HU for all evaluated patients and ROIs in the muscle). CT numbers provide a measure of a materials density and therefore link to its composition. In case of urinary stones, a shifted CT number might result in the assumption of the wrong genesis and affect the clinical diagnosis. Regarding the image noise, there were statistical significant differences between the reconstruction methods notable: employing PS on the FBP images reduced the image noise by up to 63% in the air and 48% in the soft tissues. Compared to IR, image noise was reduced by 15–27% in the soft tissues, 50% in the air and 6% in the aorta when using PS. Among the three reconstruction techniques, image noise was lowest for PS although the average CT numbers in the evaluated ROIs were not affected. Hence, calculated CNR and SNR values were highest for PS- and lowest for FBP-images throughout all ROIs. The reduction in image noise might allow for the reduction of the employed radiation exposure by 15–20%, while holding the image noise to the same level. The performed analysis on CT numbers did not show any notable differences in CT numbers, whereas the analysis on the image noise did. However, these traditional quantitative metrics do not represent the visual differences between difference reconstruction techniques.11
SNR, PSNR, RMSE and MAE describe how close a processed image is compared to a reference image. Since there is no ground-truth image available (such as a full-dose image), we chose to use the FBP images as reference. The analysis of the SNR, PSNR, RMSE and MAE has shown that IR images are closer to FBP images than PS processed images, since signal measures (SNR, PSNR) were higher and error measures (RMSE and MAE) were lower for IRs.
Subjective image quality
The image quality as rated by the two radiologists ranked between poor and moderate due to the low radiation exposure (average effective dose of 2.2 ± 1.4 mSv), c.f. Figure 3. Interrater-reliability was fair to moderate (0.40–0.49), due to the difference in experience (10 vs 2 years) and the stronger familiarization of rater 1 to images processed with PS. FBP images resulted in the lowest image quality (4.0 ± 0.8) for both raters, as they looked noisier than IR- or PS-processed images, especially in the liver (Figures 1 and 2). No differences in rated image quality were notable between IR-images and PS-processed images (Figure 3), although the image texture differed between the reconstruction methods (Figure 2).
Noise reduction techniques such as IR or PS cause a shift of the central frequency to a lower frequency, causing a texture change that is different for different technique.16 Pan et al16 compared PS to the reconstruction techniques ASiR (adaptive statistical iterative reconstruction) and ASiR-V from GE Halthcare (Waukeha, Wisconsin) and SAFIRE and ADMIRE of Siemens Healthineers (Forchheim, Germany). They concluded in their study that PS processed images lead to the least central frequency shift for the same amount of noise reduction or the most noise reduction for the same amount of central frequency shift.16 These results show, that for the different evaluated reconstruction techniques, the smallest textural change is expected from PS for the same noise reduction. Other publications have not yet treated the structure sharpness and textural appearance in detail. Figure 4 shows image subtractions for an axial slice. Although there was a significant difference between CT numbers in the air when comparing IR and FBP images, the visual impression of the air in both cases looks similar. As was noted earlier, CT numbers and noise do not provide enough detail on the visual impression. Therefore, further analysis of the frequency distribution, noise power spectrum, SNR, PSNR, RMSE and MAE for the three reconstruction techniques is required.
Until now, there are only few publications on the software PS. Tian et al17 performed Pixelshine on IR images (ASiR-VEO-FBP blending, GE Healthcare) of pelvic arterial phase CT. They concluded in their study that PS can improve the image quality, significantly reduce noise and improve the SNR and CNR. Our study confirms the positive influence on image noise, SNR and CNR for native phase abdominal CT examinations. Both mentioned research groups employ PS on IR images, which is contrary to our information of using PS with FBP images. The impact of the underlying reconstruction method for the in PS processed images needs to be assessed in a future study. Wisselink et al18 employed PS on a COPDGene phantom images to assess the potential for dose reduction in lung emphysema densitometry. They used a variety of scan parameters and reconstruction techniques in combination with PS and demonstrated that PS can reduce image noise in low-density inserts of the phantom, especially in air and emphysema. Our study supports their findings with the highest noise reduction being determined in the air (up to 63% compared to FBP).
Costs and benefits
For this study, the evaluated software required FBP images from the CT scanner. On modern CT scanner installations, IR techniques are available and employed for most of the acquisitions. Using PS as a complement to IR increases the reconstruction time and image storage by a factor of approximately 3 since FBP images and further processed images need to be created from raw data, in addition to the IR images. Furthermore, the number of reconstructed images included in the generation of a clinical diagnosis also increases. Therefore, PS might be implemented as reconstruction tool for individual CT protocols or as an alternative to IR techniques, ones the software is evaluated for more CT protocols and vendors. For existing CT scanner installations where no IR is available, PS could be implemented to either increase the image quality or reduce the employed radiation exposure.
Limitations
There was only a limited number of patients included in this study. This study needs to be extended to a larger study cohort and other body regions, such as the chest, for more thorough claims regarding the limitations and benefits of the evaluated software. Furthermore, only native acquisitions without contrast enhancement were included in this study. The influence of PS on the visibility of low-contrast lesions or small subtle-enhancing nodules has not been evaluated in detail yet. The CT examinations in this study were performed to visualize body packs with a large contrast, allowing for a non-contrast and low-dose protocol. Low-contrast lesions are difficult to visualize and evaluate in this setting. Here, all included patients had normal parenchymatous upper abdominal organs as far as visible from the low-dose non-enhanced images.
We did not give a comparison of the reader’s detection performance on body packs for the included patients, since the packs were clearly identifiable on all images due to their size and density, independent of the reconstruction algorithm. This retrospective study was performed in this specific patient collective due to the employed low-dose CT protocol rather than the analysis whether foreign bodies are detectable. More methods exist to assess the image quality with regard to texture properties (i.e. blotchy or plastic-like impression) between the three evaluated techniques. However, we were primarily interested how the three reconstruction techniques influence the visual image quality and diagnostic performance.
Furthermore, this study did not evaluate the influence on the reconstruction type on image artifacts in particular. However, when judging the image quality by the radiologists, there was no difference on the strength or visibility of artifacts (i.e. streak artifacts, beam hardening) notable among the three evaluated reconstruction techniques. The defined settings PS is using for image reconstruction are not specified to reduce or influence the visibility of image artifacts. Therefore, a difference between the methods was not expected.
In this study, the standard PS software settings were used as suggested by the vendor for the specific CT scanner, reconstruction kernel and body region. More research on the impact of the settings on the image quality is necessary to validate the chosen parameter. Furthermore, since the number of DL methods for noise reduction is increasing, it would be of interest to compare the assessed DL method to other published DL methods.
Contributor Information
Andrea Steuwe, Email: andrea.steuwe@med.uni-duesseldorf.de.
Marie Weber, Email: marie.weber@med.uni-duesseldorf.de.
Oliver Thomas Bethge, Email: oliver.t.bethge@med.hhu.de.
Christin Rademacher, Email: christin.rademacher@med.uni-duesseldorf.de.
Matthias Boschheidgen, Email: matthias.boschheidgen@med.uni-duesseldorf.de.
Lino Morris Sawicki, Email: linomorris.sawicki@med.uni-duesseldorf.de.
Gerald Antoch, Email: antoch@med.uni-duesseldorf.de.
Joel Aissa, Email: joel.aissa@med.uni-duesseldorf.de.
REFERENCES
- 1.Mangold S, Wichmann JL, Schoepf UJ, Poole ZB, Canstein C, Varga-Szemes A, et al. . Automated tube voltage selection for radiation dose and contrast medium reduction at coronary CT angiography using 3(rd) generation dual-source CT. Eur Radiol 2016; 26: 3608–16. doi: 10.1007/s00330-015-4191-4 [DOI] [PubMed] [Google Scholar]
- 2.Greffier J, Frandon J, Pereira F, Hamard A, Beregi JP, Larbi A, et al. . Optimization of radiation dose for CT detection of lytic and sclerotic bone lesions: a phantom study. Eur Radiol 2020; 30: 1075–8. doi: 10.1007/s00330-019-06425-z [DOI] [PubMed] [Google Scholar]
- 3.Lell MM, Kachelrieß M. Recent and upcoming technological developments in computed tomography. Invest Radiol 2020; 55: 8–19. doi: 10.1097/RLI.0000000000000601 [DOI] [PubMed] [Google Scholar]
- 4.Geyer LL, Schoepf UJ, Meinel FG, Nance JW, Bastarrika G, Leipsic JA, et al. . State of the art: iterative CT reconstruction techniques. Radiology 2015; 276: 339–57. doi: 10.1148/radiol.2015132766 [DOI] [PubMed] [Google Scholar]
- 5.Stiller W. Basics of iterative reconstruction methods in computed tomography: a vendor-independent overview. Eur J Radiol 2018; 109: 147–54. doi: 10.1016/j.ejrad.2018.10.025 [DOI] [PubMed] [Google Scholar]
- 6.May MS, Wüst W, Brand M, Stahl C, Allmendinger T, Schmidt B, et al. . Dose reduction in abdominal computed tomography. Invest Radiol 2011; 46: 465–70. doi: 10.1097/RLI.0b013e31821690a1 [DOI] [PubMed] [Google Scholar]
- 7.den Harder AM, Willemink MJ, de Ruiter QMB, Schilham AMR, Krestin GP, Leiner T, et al. . Achievable dose reduction using iterative reconstruction for chest computed tomography: a systematic review. Eur J Radiol 2015; 84: 2307–13. doi: 10.1016/j.ejrad.2015.07.011 [DOI] [PubMed] [Google Scholar]
- 8.Greffier J, Frandon J, Larbi A, Beregi JP, Pereira F. Ct iterative reconstruction algorithms: a task-based image quality assessment. Eur Radiol 2020; 30: 487–500. doi: 10.1007/s00330-019-06359-6 [DOI] [PubMed] [Google Scholar]
- 9.Pontana F, Moureau D, Schmidt B, Duhamel A, Faivre J-B, Yasunaga K, et al. . CT pulmonary angiogram with 60% dose reduction: Influence of iterative reconstructions on image quality. Diagn Interv Imaging 2015; 96: 487–93. doi: 10.1016/j.diii.2014.08.006 [DOI] [PubMed] [Google Scholar]
- 10.van Osch JAC, Mouden M, van Dalen JA, Timmer JR, Reiffers S, Knollema S, et al. . Influence of iterative image reconstruction on CT-based calcium score measurements. Int J Cardiovasc Imaging 2014; 30: 961–7. doi: 10.1007/s10554-014-0409-9 [DOI] [PubMed] [Google Scholar]
- 11.Fletcher JG, Leng S, Yu L, McCollough CH. Dealing with uncertainty in CT images. Radiology 2016; 279: 5–10. doi: 10.1148/radiol.2016152771 [DOI] [PubMed] [Google Scholar]
- 12.Kim JH, Yoon HJ, Lee E, Kim I, Cha YK, Bak SH. Validation of Deep-Learning image reconstruction for low-dose chest computed tomography scan: emphasis on image quality and noise. Korean J Radiol 2020; 21: S5. doi: 10.3348/kjr.2020.0116 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13.Akagi M, Nakamura Y, Higaki T, Narita K, Honda Y, Zhou J, et al. . Deep learning reconstruction improves image quality of abdominal ultra-high-resolution CT. Eur Radiol 2019; 29: 6163–71. doi: 10.1007/s00330-019-06170-3 [DOI] [PubMed] [Google Scholar]
- 14.Benz DC, Benetos G, Rampidis G, von Felten E, Bakula A, Sustar A, et al. . Validation of deep-learning image reconstruction for coronary computed tomography angiography: impact on noise, image quality and diagnostic accuracy. J Cardiovasc Comput Tomogr 2020; 14: 444–51. doi: 10.1016/j.jcct.2020.01.002 [DOI] [PubMed] [Google Scholar]
- 15.Shin YJ, Chang W, Ye JC, Kang E, Oh DY, Lee YJ, et al. . Low-Dose abdominal CT using a deep Learning-Based denoising algorithm: a comparison with CT reconstructed with filtered back projection or iterative reconstruction algorithm. Korean J Radiol 2020; 21: 356. doi: 10.3348/kjr.2019.0413 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16.Pan T, Hasegawa A, Luo D, Wu CC, Vikram R. Technical note: impact on central frequency and noise magnitude ratios by advanced CT image reconstruction techniques. Med Phys 2020; 47: 480–7. doi: 10.1002/mp.13937 [DOI] [PubMed] [Google Scholar]
- 17.Tian S-F, Liu A-L, Liu J-H, Liu Y-J, Pan J-D. Potential value of the PixelShine deep learning algorithm for increasing quality of 70 kVp+ASiR-V reconstruction pelvic arterial phase CT images. Jpn J Radiol 2019; 37: 186–90. doi: 10.1007/s11604-018-0798-0 [DOI] [PubMed] [Google Scholar]
- 18.Wisselink HJ, Pelgrim GJ, Rook M, van den Berge M, Slump K, Nagaraj Y, et al. . Potential for dose reduction in CT emphysema densitometry with post-scan noise reduction: a phantom study. Br J Radiol 2020; 93: 20181019. doi: 10.1259/bjr.20181019 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 19.den Harder AM, Willemink MJ, van Doormaal PJ, Wessels FJ, Lock MTWT, Schilham AMR, et al. . Radiation dose reduction for CT assessment of urolithiasis using iterative reconstruction: a prospective intra-individual study. Eur Radiol 2018; 28: 143–50. doi: 10.1007/s00330-017-4929-2 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 20.Sage D, Unser M. Teaching image-processing programming in Java. IEEE Signal Process Mag 2003; 20: 43–52. doi: 10.1109/MSP.2003.1253553 [DOI] [Google Scholar]
- 21.Brennan RL, Prediger DJ. Coefficient kappa: some uses, misuses, and alternatives. Educ Psychol Meas 1981; 41: 687–99. doi: 10.1177/001316448104100307 [DOI] [Google Scholar]
- 22.Huda W, Magill D, He W. Ct effective dose per dose length product using ICRP 103 weighting factors. Med Phys 2011; 38: 1261–5. doi: 10.1118/1.3544350 [DOI] [PubMed] [Google Scholar]