Significant decrease in lipid core burden index following balloon dilation was associated with the leakage of cholesterol crystals in a patient: a case report

Soshi Moriya 1, Kikuo Isoda 1,, Tomotaka Dohi 1, Shinya Okazaki 1
Editors: Riccardo Liga, Romain Didier, Andre Dias, Christian Fielder Camm, Peregrine Green1
PMCID: PMC7780494  PMID: 34151139

Abstract

Background

Near-infrared spectroscopy (NIRS) has been used for analysis the composition of the atherosclerotic plaque in coronary arteries. However, meaning of significant decrease in max lipid core burden index at 4 mm (max LCBI4mm) during percutaneous coronary intervention (PCI) is poorly understood.

Case summary

A 64-year-old male with unstable angina underwent coronary angiography, which demonstrated a hazy tight culprit lesion in the mid-right coronary artery. Pre-intervention NIRS–intravascular ultrasound (NIRS–IVUS) and chemogram showed plaque with high lipid burden at the culprit lesion. Then, we used a distal protection device before PCI because of high max LCBI4mm in the lesion. After pre-dilation with a scoring balloon, repeat NIRS–IVUS interrogation revealed an almost complete disappearance of the yellow signal and decrease in max LCBI4mm (from 537 to 44) significantly, suggesting decrease in the lipid content of the plaque. Finally, a drug-eluting stent deployment followed by inflation of a non-compliant balloon led to an excellent result. After PCI, we detected trapped large amounts of debris on retrieval of the filter. Pathological diagnosis confirmed that trapped material was lipid-rich plaque including cholesterol crystals.

Discussion

This is the first report directly demonstrated that significant decrease in max LCBI4mm at culprit lesion should be associated with the leakage of cholesterol crystals from lipid-rich plaque during PCI in the clinical patient.

Keywords: Case report, Percutaneous coronary intervention, Near-infrared spectroscopy, Histopathological study


Learning points

  • Near-infrared spectroscopy (NIRS) can be employed to detect extent and distribution of lipid within native coronary arteries.

  • High lipid core burden index (LCBI) measured using NIRS may be associated with distal embolization during percutaneous coronary intervention (PCI).

  • Significant decrease in max LCBI should be associated with the leakage of cholesterol crystals from lipid-rich plaque during PCI.

Introduction

Distal embolization during percutaneous coronary intervention (PCI) is a crucial cause of periprocedural myocardial infarction.1–3 Coronary angiography (CAG) has limited capacity to detect the risk for distal embolization. The development of intracoronary imaging techniques provides an opportunity for detail analysis of plaque composition, which could identify lesions at high risk of causing distal embolization.4 Recently, near-infrared spectroscopy (NIRS) has been used for analysis the composition of the atherosclerotic plaque in coronary arteries.5,6 However, meaning of significant decrease in max lipid core burden index at 4 mm (max LCBI4mm) during PCI is poorly understood.

Herein, we demonstrated that significant decrease in max LCBI4mm during PCI should be associated with the leakage of cholesterol crystals from lipid-rich plaque in the clinical patient using the histopathological study.

Timeline

Time Events
Initial event Patient presented with concerns for progressive exertional chest pain. Diagnosis of unstable angina.
Hospital Day 1 Coronary angiography showed a hazy tight culprit lesion in the mid-right coronary artery. Pre-intervention near-infrared spectroscopy–intravascular ultrasound showed plaque with high lipid burden at the culprit lesion. Then, we used a distal protection device during percutaneous coronary intervention (PCI) and we detected trapped large amounts of debris on retrieval of the filter.
Hospital Day 2 Pathological diagnosis confirmed that trapped material was lipid-rich plaque including cholesterol crystals.
Hospital Day 3 Patient felt no chest pain and was discharged from hospital.
6 months after discharge Patient has been free of chest pain since the PCI.

Case presentation

A 64-year-old male with hypertension and dyslipidaemia was referred for CAG because of progressive exertional angina. He had been medicated for 2 years with amlodipine (5 mg/day) for hypertension and rosuvastatin (2.5 mg/day) daily for dyslipidaemia. He had previously smoked cigarettes but had quit 40 years ago. Twenty days before admission, he experienced new exertional chest pain, the frequency of which increased within a few days. A local doctor diagnosed unstable angina and referred him to our hospital. He had a blood pressure of 138/84 mmHg, a heart rate of 78 b.p.m., respiratory rate of 14 breaths/min, and oxygen saturation of 100% on room air at arrival. On physical examination, he had regular cardiac rhythm with no heart murmurs. Initial laboratory evaluation revealed total, HDL, and LDL cholesterol values of 208 (normal range: 200–219), 38 (>40) and 106 (120–139) mg/dL, respectively.

The CAG findings revealed a hazy, tight culprit lesion in the mid-right coronary artery (RCA) (Figure 1A, yellow arrow, Video 1). Pre-intervention near-infrared spectroscopy–intravascular ultrasound (NIRS–IVUS) (Makoto Intravascular Imaging System, Infraredx) and chemogram showed plaque with high lipid burden [max lipid core burden index at 4 mm (max LCBI4mm), 537] at the culprit lesion (Figure 1A). We then applied a distal protection device (Filtrap II 5 mm, Nipro) before percutaneous coronary intervention (PCI) because of the high max LCBI4mm in the lesion. After crossing another guidewire across the right ventricular branch as a protection wire, pre-dilation was performed with a scoring balloon (AngioSculpt 3.5 × 10 mm, AngioScore) (Figure 1B) because NIRS–IVUS also revealed that almost the entire circumference of the calcified lesion was located at the proximal side of the culprit lesion (Supplementary material online, Figure S1). Repeat NIRS–IVUS interrogation revealed disappearance of the yellow signal and a significant decrease in max LCBI4mm from 537 to 44, suggesting decreased lipid content in the plaque (Figure 1C). Finally, a drug-eluting stent (Xience Sierra 4.0 × 33 mm, Abbott) was deployed followed by inflation of a non-compliant balloon (Powered3 NC 4.5 × 8 mm, Nipro), which led to excellent results at the lesion site (Figure 1D, Video 2). Final angiography confirmed a well-expanded stent without distal embolization (Video 3). After PCI, the retrieved filter contained large amounts of trapped white debris (Figure 2A). Pathological findings confirmed that this material was lipid-rich plaque including cholesterol crystals and inflammatory cells (Figure 2B).

Figure 1.

Figure 1

Large lipid-core plaque detected by near-infrared spectroscopy at culprit site and lipid core burden index was significantly decreased during percutaneous coronary intervention. (A) Pre-intervention near-infrared spectroscopy–intravascular ultrasound and chemogram showed plaque with high lipid burden (max lipid core burden index at 4 mm, 537) at the culprit lesion. (B) Predilatation was performed using a scoring balloon (AngioSculpt 3.5 × 10 mm). (C) Repeat near-infrared spectroscopy–intravascular ultrasound interrogation revealed an almost complete disappearance of the yellow signal and decrease in max lipid core burden index at 4 mm (from 537 to 44) significantly. (D) A drug-eluting stent deployment followed by inflation of a non-compliant balloon led to an excellent result at the lesion site.

Figure 2.

Figure 2

After percutaneous coronary intervention, we detected trapped large amounts of lipid-rich plaque on retrieval of the filter. (A) Large amounts of debris on retrieval of the filter were detected. (B) Pathological study demonstrated that the debris was lipid-rich plaque including cholesterol crystals and inflammatory cells (haematoxylin–eosin stain).

This patient developed acute coronary syndrome (ACS) and had an LDL cholesterol value of 106 mg/dL despite 2 years of therapy with rosuvastatin (2.5 mg/day). We therefore doubled the dose of rosuvastatin to 5 mg to achieve target LDL cholesterol levels < 70 mg/dL. He was discharged on post-PCI Day 2. One month after discharge, laboratory findings showed total, HDL, and LDL cholesterol values of 142, 52, and 62 mg/dL, respectively. The patient has remained free of chest pain since the PCI (for 6 months).

Discussion

Occurrences of no-reflow and periprocedural myocardial infarction can be associated with major adverse events.1–3 Near-infrared spectroscopy was developed for the determination of the chemical composition of plaques with the intention of identifying lipid cores within the atherosclerotic lesions.5,7,8 Recently, NIRS and IVUS were combined in a single catheter to provide the user with information on both the composition and structure of plaques simultaneously.5,9,10 Because the NIRS–IVUS catheter is a little thicker and less flexible than conventional IVUS catheters, we selected a short Amplatz left 0.75 (6 Fr) guiding catheter, which provided excellent backup, and the NIRS–IVUS catheter was passed through the thigh lesion without pre-dilation.

Because the reference diameter of the right coronary artery was very large (∼4.0 mm), we considered that the culprit lesion might include a large lipid core and examined the lesion using NIRS–IVUS. In our patient, NIRS–IVUS revealed the culprit lesion was attenuated plaque with a large lipid core (max LCBI4mm = 537), suggesting high risk of distal embolization during PCI. Indeed, previous study demonstrated that the incidence post-PCI myocardial infarction was significant higher at 50% among patients with max LCBI4mm > 500 vs. 4.2% with LCBI4mm < 500 (P = 0.0002).11 We therefore deployed a filter wire in the distal RCA before ballooning, which resulted in trapping large amounts of debris on the retrieved filter. Finally, the filter wire prevented distal embolization during PCI and angiography confirmed optimal results without delayed coronary flow. These findings suggested that NIRS–IVUS should be actively applied for patients with ACS and large plaque to avoid the no-reflow phenomenon occurring during PCI.

Several reports have described a relationship between a decreased level of LCBI during PCI and no reflow.9,11,12 In addition, autopsy studies of patients who died due to fatal no reflow, have found extensive cholesterol crystal emboli plugs in distal coronary arteries.11,12 These findings indirectly suggested that a decreased LCBI is associated with cholesterol crystal leakage from atherosclerotic plaque during PCI. Here, we detected a significant decrease in max LCBI4mm at a culprit lesion during PCI and pathological findings confirmed that material trapped on the filter wire after PCI comprised lipid-rich plaque including cholesterol crystals and inflammatory cells.

Conclusion

These findings are the first report of evidence of a direct association between a significantly decreased max LCBI4mm at a culprit lesion during PCI and cholesterol crystal leakage from lipid-rich plaque in a clinical patient.

Lead author biography

graphic file with name ytaa148f3.jpg

Soshi Moriya received the MD degree from Kitazato University, Kanagawa, Japan in 2016 and subsequently underwent his specialty training in Internal Medicine and Cardiology at Juntendo University, Tokyo, Japan. Currently, his research interest is mainly focused on multimodal cardiac imaging, in particular in ischaemic heart diseases and heart failure.

Supplementary material

Supplementary material is available at European Heart Journal - Case Reports online.

Funding

This report was supported by Grant-in-Aid for Scientific Research (C) (JSPS KAKENHI Grant Number 16K09522 (K.I.), and 20K08433 (K.I.))

Slide sets: A fully edited slide set detailing this case and suitable for local presentation is available online as Supplementary data.

Consent: The authors confirm that written consent for submission and publication of this case report including images and associated text has been obtained from the patient in line with COPE guidance.

Conflict of interest: none declared.

Supplementary Material

ytaa148_Supplementary_Data

References

  • 1. Cantor WJ, Newby LK, Christenson RH, Tuttle RH, Hasselblad V, Armstrong PW. et al. ; SYMPHONY and 2nd SYMPHONY Cardiac Markers Substudy Investigators. Prognostic significance of elevated troponin I after percutaneous coronary intervention. J Am Coll Cardiol 2002;39:1738–1744. [DOI] [PubMed] [Google Scholar]
  • 2. Kizer JR, Muttrej MR, Matthai WH, McConnell J, Nardone H, Sonel AF, et al. Role of cardiac troponin T in the long-term risk stratification of patients undergoing percutaneous coronary intervention. Eur Heart J 2003;24:1314–1322. [DOI] [PubMed] [Google Scholar]
  • 3. Nageh T, Sherwood RA, Harris BM, Byrne JA, Thomas MR.. Cardiac troponin T and I and creatine kinase-MB as markers of myocardial injury and predictors of outcome following percutaneous coronary intervention. Int J Cardiol 2003;92:285–293. [DOI] [PubMed] [Google Scholar]
  • 4. Patel VG, Brayton KM, Mintz GS, Maehara A, Banerjee S, Brilakis ES.. Intracoronary and noninvasive imaging for prediction of distal embolization and periprocedural myocardial infarction during native coronary artery percutaneous intervention. Circ Cardiovasc Imaging 2013;6:1102–1114. [DOI] [PubMed] [Google Scholar]
  • 5. Negi SI, Didier R, Ota H, Magalhaes MA, Popma CJ, Kollmer MR. et al. Role of near-infrared spectroscopy in intravascular coronary imaging. Cardiovasc Revasc Med 2015;16:299–305. [DOI] [PubMed] [Google Scholar]
  • 6. Horvath M, Hajek P, Stechovsky C, Honek J, Spacek M, Veselka J.. The role of near-infrared spectroscopy in the detection of vulnerable atherosclerotic plaques. Arch Med Sci 2016;6:1308–1316. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7. Sharif F, Murphy RT.. Current status of vulnerable plaque detection. Cathet Cardiovasc Intervent 2009;75:135–144. [DOI] [PubMed] [Google Scholar]
  • 8. Sanon S, Dao T, Sanon VP, Chilton R.. Imaging of vulnerable plaques using near-infrared spectroscopy for risk stratification of atherosclerosis. Curr Atheroscler Rep 2013;15:304. [DOI] [PubMed] [Google Scholar]
  • 9. Schultz CJ, Serruys PW, van der Ent M, Ligthart J, Mastik F, Garg S, et al. First-in-man clinical use of combined near-infrared spectroscopy and intravascular ultrasound: a potential key to predict distal embolization and no-reflow? J Am Coll Cardiol 2010;56:314. [DOI] [PubMed] [Google Scholar]
  • 10. Madder RD, Steinberg DH, Anderson RD.. Multimodality direct coronary imaging with combined near-infrared spectroscopy and intravascular ultrasound: initial US experience. Cathet Cardiovasc Intervent 2013;81:551–557. [DOI] [PubMed] [Google Scholar]
  • 11. Goldstein JA, Maini B, Dixon SR, Brilakis ES, Grines CL, Rizik DG. et al. Detection of lipid-core plaques by intracoronary near-infrared spectroscopy identifies high risk of periprocedural myocardial infarction. Circ Cardiovasc Interv 2011;4:429–437. [DOI] [PubMed] [Google Scholar]
  • 12. Goldstein JA, Grines C, Fischell T, Virmani R, Rizik D, Muller J, et al. Coronary embolization following balloon dilation of lipid-core plaques. JACC Cardiovasc Imaging 2009;2:1420–1424. [DOI] [PubMed] [Google Scholar]