Noninvasive Continuous Monitoring of Vital Signs With Wearables: Fit for Medical Use?

Malte Jacobsen 1,, Till A Dembek 2, Guido Kobbe 3, Peter W Gaidzik 4, Lutz Heinemann 5
PMCID: PMC7783016  PMID: 32063034

Abstract

Background:

Wearables (= wearable computer) enable continuous and noninvasive monitoring of a range of vital signs. Mobile and cost-effective devices, combined with powerful data analysis tools, open new dimensions in assessing body functions (“digital biomarkers”).

Methods:

To answer the question whether wearables are ready for use in the medical context, a PubMed literature search and analysis for their clinical-scientific use using publications from the years 2008 to 2018 was performed.

Results:

A total of 79 out of 314 search hits were publications on clinical trials with wearables, of which 16 were randomized controlled trials. Motion sensors were most frequently used to measure defined movements, movement disorders, or general physical activity. Approximately 20% of the studies used sensors to detect cardiovascular parameters. As for the sensor location, the wrist was chosen in most studies (22.8%).

Conclusion:

Wearables can be used in a precisely defined medical context, when taking into account complex influencing factors.

Keywords: wearables, vital signs, clinical trials, digital biomarkers, noninvasive

Introduction

Vital parameters such as heart rate, blood pressure, oxygen saturation, and temperature are essential for the clinical evaluation of patients. With new technological developments, such vital parameters can be recorded wirelessly and continuously, noninvasively and cost-effectively using the so-called wearables. The combination of wearables with powerful data analysis tools opens up new dimensions in the assessability of body functions (“digital biomarkers”), with the possibility of remote monitoring further enhancing the relevance of such approaches. Although many of the technologies used in wearables are not really innovative, technological advances mean that they can now be integrated into portable battery-operated devices in a compact and cost-effective form. They make it possible to use them in patients’ everyday lives and allow important functional aspects to be quantified and qualified, such as physical activity and sleep behavior.

The aim of this review is to present the status quo of the application of wearables for the measurement of vital parameters in the medical context. Additionally, the results of a PubMed search about clinical trials with wearables will be presented.

Background

Nowadays, many healthy people also monitor their vital parameters independently and regularly, but no published data on the number of vital parameters in Germany are yet available. In relation to the total population, it can be assumed that this proportion is relatively small. However, the proportion of those who collect information about their body functions under everyday conditions through wearables, driven by the “Quantified Self” movement, is increasing significantly.1 Smartphones now routinely record a number of parameters and provide the user with suitable evaluations.

To date, vital parameters have primarily been recorded analogously in the medical context, eg, by blood pressure protocols or movement protocols for Parkinson’s patients. These are used by the attending physicians as “orientation” and usually not systematically evaluated. The keeping of such protocols means a high additional effort for patients, which often results in only inconsistent protocol keeping.

At the outpatient care level, vital parameters are usually randomly collected because of the limitations of analog measurement methods and the lack of data storage. The established methods for continuous recording of vital parameters under everyday conditions (eg, long-term ECG or blood pressure measurements) are carried out over a comparatively short period of a maximum of a few days.

On an inpatient level, a set of vital parameters is routinely recorded by hospital staff at several points in time per day; this is a manual serial measurement with different recording frequencies. There are no published data available on the regularity and measurement quality of vital parameters in hospitals, although there are corresponding guidelines from the specialist societies.2 Although various vital parameters are continuously recorded on monitoring stations, they are usually only analyzed as “snapshots” or in simple trend tables. In addition, many of the parameters need to be measured invasively and are associated with considerable restrictions on patient mobility. A transferability of the vital parameters measured under inpatient conditions to the everyday situation of the patients is only possible to a limited extent.

Overall, the recording of vital parameters currently requires considerable personnel and economic effort; it is often unclear to what extent the information obtained is reliable and what actual medical added value it offers.

Monitoring of Vital Signs With Wearables in the Medical Field

The rapidly increasing number of different wearables is primarily based on their use in the private sector (eg, fitness tracking); for medical use, there is only a limited number of systems.3 The recording of various vital parameters measurable with wearables should enable a clinically relevant assessment of intra- and interindividual physiological and pathophysiological processes over time. They are to be used for monitoring vital parameters in outpatient, inpatient, institutionalized (eg, nursing home), and scientific settings.

Medical interventions based on vital parameters recorded with wearables require high data reliability, ie, the measured parameters must be validated in a precisely defined context of use (eg, age and pre-existing conditions).4 Despite the rapid technical development, this validation still turns out to be a complex requirement; a number of influencing factors have to be considered systematically and structured before the use of wearables (Figure 1):

Figure 1.

Figure 1.

Noninvasive continuous measurement of vital parameters with wearables. Steps from the signals to the analysis and interpretation of the measured values. PPG, photoplethysmography. Source: Graphs from www.visionom.de.

  • Environmental factors: Ambient light, temperature, humidity, height above sea level, and the exact location of the patient must be taken into account.5

  • Legal framework conditions: The medical purpose (diagnostic or therapeutic) of wearables crosses the line from lifestyle/fitness to medical devices, so that manufacturers and users in Europe are subject to the Medical Devices Act (MPG) and the (medical) product used must be CE-certified. In Europe, the regulatory authorities have so far not made any recommendations for the use of wearables in the medical context; in the United States, on the other hand, they have done so for clinical application, albeit not for use in clinical research. There are only a few initial statements on this subject.6,7

  • Body: The target population should be clearly defined and has to be carefully considered which parameters are relevant in the given disease-related context. The majority of the clinical studies identified with the PubMed research mentioned above were conducted with a young and healthy study population. Such volunteers differ considerably from the patients in the medical use of wearables and represent an inadequate starting point. Even when wearables are used in an adequate target population, pronounced interindividual differences must be taken into account: Thus, factors of body constitution such as skin color, fluid status, and body fat content can vary significantly between patients in a target population. Especially in photometric procedures (eg, photoplethysmography [PPG]), body temperature has a negative influence on the measurement quality via local skin perfusion or substances applied or applied to the skin (eg, creams and tattoos).5

  • Movements cause artifacts in many measurement methods and are one of the biggest problems in data evaluation. For example, heart rate analysis has shown that measurement accuracy decreases with increasing movement intensity.8 Furthermore, the heart rhythm can influence the accuracy of the measurement, so that the heart rate can be determined more reliably in patients with sinus rhythm than in patients with atrial fibrillation.9 Therefore, influential disturbances such as movements should be addressed and compensated for as well as possible.

  • Sensors: The sensors built into wearables usually convert a physiological signal into an electrical signal, which is then usually digitized and stored or transmitted in this form. In most wearables, combinations of different sensors are used. Many of the measurement methods are not based on traceable measurement quantities in the metrological sense, they are measured with arbitrary methods (ie, without calibration). This means, for example, in the case of PPG that the light measured by the photon detector is usually not specified with an SI unit. However, since the absolute measured values are in most cases not clinically relevant, but relative changes, interindividually comparable values only result after further processing (eg, determination of the heart rate from pulse wave analysis).

  • The attachment of the sensors to the body can have a negative influence on the quality of the measurement of some parameters, eg, the accuracy of the pulse measurement at the wrist is inferior to that at other locations.10 The ear, for example, represents a low artifact measuring point for many parameters. The reasons for this are the lack of ambient light in photometric measurements in the auditory canal and comparatively few movement artifacts.11 It nonetheless remains to be seen how compliance with in-ear wearables will be in clinical studies.

  • Depending on the body localization at which the wearable is to be measured, different fixation techniques are considered, eg, with a plaster or an elastic band. The contact pressure can influence the local blood flow in such a way that the quality of the measurement is negatively affected.12 If the wearable is worn in the same place for a longer period of time, skin reactions may occur which may even lead to the development of relevant inflammations. Such side effects can massively impair the use of wearables by patients.

  • Processor: The wearables in themselves often represent a “black box,” ie, the raw signal detected by the sensors is processed by proprietary algorithms of the manufacturers; artifacts created shall be eliminated, the signal-to-noise ratio shall be improved, etc. The algorithms are usually permanently further developed, since an improvement in the performance of the wearables is possible without major technical effort (in terms of hardware). Therefore, the most current algorithm version should always be used for practical application and the version used should be documented.

  • In everyday use, user-friendliness and intuitive handling of wearables are decisive factors for long-term use by patients. Factors such as battery life, capacity of the data memory, and easy attachment to the body are of great importance in this context. Such characteristics vary significantly between the available wearables which means the spectrum of battery life goes from a few hours up to 18 months (Spire Health Tag, Spire Inc., United States).

  • Data transfer: The processed data are not only stored in different formats (ISHNE, CSV, etc.), but depending on the underlying operating system, relevant differences may arise for further data transmission and processing. If the data from wearables are to be merged with data from other signal sources, the necessary interoperability is often not given. The integration of wearable data into the existing clinical information systems (CIS) or practice systems can thus be made considerably more difficult. There are individual technical solutions in which vital parameters are automatically transferred into such information systems (Connex Spot Monitor, Welch Allyn Inc., United States; No Wearable), but as a rule this possibility does not exist for vital parameters recorded with wearables.

  • Vital sign data can be transferred from the given wearable to a device via Bluetooth or similar methods using different transmission techniques. In practice, the transmission technique used has a considerable effect on the reliability of the transmission per se and on the transmission speed. In many cases, the wearables are linked to a smartphone and the data are automatically stored in cloud databases. There are studies that have shown that the data transmission of medical devices using such techniques is not sufficiently secure.13 Even the act of data transmission as such must be covered by the express consent of the patient previously informed of his rights. In the European Union, this is controlled by the General Data Protection Regulation and in Germany by the Federal Data Protection Act.

  • Devices: These devices receive the data and represent the interface to the user. The vital parameters are visualized on the display of the devices: this ranges from the representation of a numerical value to complex graphics, eg, for glucose day profiles. It is crucial for the user to be able to get a comprehensive picture of the data immediately. Furthermore, key figures can be calculated and displayed in the terminal device: for example, the average glucose concentration, sleep duration, or heart rate.

  • Analysis and interpretation: The progress made in data processing and analysis in recent years as well as over the past decades due to the broad availability of smartphones has made a decisive contribution to the pronounced use of wearables in the private sphere. The amount of data to be analyzed is considerable, if, for example, a range of vital parameters with a frequency of, eg, 1 Hz (per parameter 86 400 data records/day) is recorded, several million values result over weeks, months, and years. In most clinical studies with wearables, data processing was “manual”, ie, with the help of suitable software (such as Matlab). With methods from the field of artificial intelligence and “deep learning,” a considerably faster and “automated” processing is possible. An increasing use of such approaches in data processing in connection with vital parameters can be observed, however, these methods have so far only been used in smaller experimental investigations.

  • The potential of new evaluation methods for large amounts of data (ie, “Big Data”) can hardly be overestimated.14 It is, however, important to consider the benefits in a differentiated way, as it is possible, with the help of artificial intelligence, to differentiate images of the fundus of the eye according to gender without it being comprehensible on the basis of which criteria the algorithm used makes this allocation.15 As a result, this circumstance is per se goal-oriented, but unsatisfactory in terms of knowledge gain; similar problematic results are also conceivable in the area of vital parameters.

This list of factors makes it clear that there are no wearables that can be used ubiquitously (“one fits all”), but that the specific selection of a suitable wearable must be made for the respective question.

Self-Monitoring by the Patients

The measurement of vital parameters can support patients especially in dealing with diseases of civilization such as obesity-related, neurodegenerative, and cardiovascular diseases. Self-monitoring enables patients to continuously track their own physical condition based on objective data in everyday life at any time. Such vital parameters recorded under everyday conditions support approaches to individualized medicine. Defined key figures (ie, digital biomarkers) can be determined through the evaluation of longitudinally recorded “vital parameter profiles”; their value can be further increased through the use of further information (eg, genomics).16

A possible negative consequence of self-monitoring is that hypochondriasis may worsen and patients are constantly concerned with their vital parameters and the possible significance of these parameters on their health. The American “mobile health” expert Eric Topol sees the predominant benefit of digital medicine in the larger context, but also points to the problem of such “side effects.17

Usage of Wearables in Research and Clinic

The clinical studies to date provide indications of the potential of the use of wearables for clinical research. Thus, wearables provide important additional information for the interpretation of the questions posed by the studies (“surrogate parameters”). The combination of noninvasively recorded vital parameters with subjective patient experiences (electronic patient reported outcomes, ePRO) under real conditions represents a further option for clinical research. Such correlated data could be used ubiquitously in healthcare: From clinical research to outpatient care to inpatient monitoring, practitioners are de facto “live” informed about serious incidents and changes in the monitored patients.

In the Apple Heart Study, data from approximately 420 000 study participants were analyzed for cardiac arrhythmias using Apple Watch (a photoplethysmographic method) and immediate automated analysis. Recruitment for the study was via an app (“ResearchKit”), with the first >10 000 study participants enrolling within 24 hours.18 Whether clinical studies with wearables with large study cohorts (ie, “virtual studies”) will be conducted more frequently in the future and how principles of Good Clinical Practice will be implemented in such studies remains to be seen.

The “dynamic” life cycle of wearables and their software represents a considerable obstacle to the use of wearables in research: Software updates come in monthly intervals and new device generations in annual intervals. A context-related validation for a vital parameter only applies to a defined wearable generation with a given software version. With regard to a study duration that regularly takes years, this problem must be adequately considered: A wearable that was selected during study planning may differ significantly in its properties from the wearable that is used at the start of the study and, above all, at the end of the study.

The use of wearables in hospitals enables a completely new monitoring option, the diverse application potential of which has not yet been fully exploited.

Possible use cases are the following:

  • New procedures for the collection of vital signs. After appropriate validation, wearables can be used to replace the previously used methods for vital signs collection. For a broad use of wearables in the medical environment, it would be supportive if they could measure two additional vital signs:
    • - Noninvasive and cuff-less real-time blood pressure measurement: In various approaches, PPG is mostly used in combination with ECG recording.19 A clinical trial from 2016 showed that such a measurement of systolic blood pressure is possible; it even meets the requirements of the Society for Measurement Accuracy for new blood pressure monitors.20 Since the end of 2018, the first wristwatch wearable for measuring blood pressure has had FDA approval, which measures blood pressure noninvasively several times a day.21
    • - Noninvasive continuous blood glucose measurement as a replacement for conventional capillary blood glucose measurement and continuous interstitial glucose monitoring, which is still a burdening factor for millions of patients with diabetes. Although various approaches have been researched in recent decades, no marketable product is yet available.22,23
  • Outpatient/inpatient monitoring of patients during critical treatment phases: The automated analysis of vital signs for the collection of clinical scores is an important application example, eg, the “Early Warning Score” or the “quick Sepsis-related organ failure assessment” (qSOFA) score. Such scores are clinical tools that take vital parameters and state of consciousness into account and allow early detection of disease worsening or sepsis.24 Through consistent further development of such scores, clinically relevant events can be automatically detected by wearables.

  • Treatment support for chronic diseases or for therapies involving a high risk that lead to frequent hospitalization: An example is the assessment of volume status on the basis of vital signs in patients with heart failure. In addition, the potential of novel therapies can be better detected by wearables, eg, by the option of monitoring for specific undesired side effects, such as those occurring with CAR-T cell or antibody therapies in the form of cytokine release syndrome.25

  • Secondary prevention: One example is the early detection of heart rhythm disorders in large collectives of previously healthy patients. As with all screening examinations, a careful balance must be struck between the expected incidence and the false-positive rates.

Health-Economic Aspects

The direct costs or acquisition costs for wearables are comparatively low compared to other medical equipment. If, however, clinical validation studies are required for the use of wearables in accordance with the MPG, it can be assumed that the manufacturers will pass on the costs incurred to users in the form of higher acquisition costs.

Due to the abnormalities in the vital parameters of patients detected by wearables, there may initially be increased use of medical services, ie, the health systems have to bear higher costs.26 In the long run, costs can be reduced by avoiding hospitalization and achieving a better state of health.27 Overall, it is not yet possible to estimate whether wearables have a positive cost-benefit ratio.

Methods

To describe the status quo, a PubMed literature search was carried out and publications on clinical studies with wearables in the last 10 years were sought. The search terms “wearable” and “clinical trial” or “clinical study” were used in the years 2008 to 2018 (accessed February 20, 2019).

Results

The search yielded 314 hits, with the number of hits increasing from n = 4 in 2008 to n = 84 in 2017 and n = 49 in 2018. The identified articles were then analyzed to determine whether they described the results of clinical studies with wearables for the noninvasive measurement of vital parameters. This applied to n = 79 (of which n = 16 were randomized controlled trials) of the 314 publications. Relevant data from these 79 publications were collected and evaluated (Table 1; the respective references are given in the last column of the table).

Table 1.

In PubMed Literature Research Identified Clinical Studies Published in the Years 2008-2018 in Which a Noninvasive Continuous Measurement of Vital Parameters With Wearables Was Performed.

Parameter Sensors Calculated parameter Localization of sensor Indication Examples Clinical studies
Heart rate PPG RR and HRV pulse wave distances pulse wave amplitude Wrist and ear Cardiac arrhythmia, monitoring, stress, and menstrual cycle Apple Watch Series 3
FitBit Blaze
Gillinov et al8; Koshy et al9; Conroy et al28; Gaggioli et al29; and Shilaih et al30
ECG RR and HRV
R-wave-distances
Chest Cardiac arrhythmia, monitoring, and sleep ZioXT Schoot et al20; Steinhubl et al26; Mlakar et al31; Breteler et al32; Razjouyan et al33; Jeffs et al34; Frias et al35; Yu et al36; and Aktaruzzaman et al37
Blood pressure (without cuff) ECG and PPG Pulse transit time Handheld device Systolic blood pressure Checkme Health Monitor Schoot et al20
Respiratory rate ECG RR Chest Monitoring HealthPatch MD Breteler et al32
Oxygen saturation Photometric Pulse oxymetric oxygen saturation Upper arm, wrist, and finger Monitoring Hidalgo EQ02, Fitbit, and Checkme Health Monitor Jeffs et al34 and Kroll et al38
Electromyography (surfaces) Electrical surface activity Muscle contraction Upper arm, abdomen, thigh, and lower leg Generalized tonic-clonic seizures and fatigue symptoms Multi Bio Sensors Yu et al36 and Halford et al39
Physical activity Accelerometer, gyroscope, and magnetometers General activity sleep ingestion of food Upper arm, wrist, and back Health promotion, weight loss, and rehabilitation Misfit, Fit Core, and Bite Counter Frias et al35; McDermott et al40; Mendoza et al41; Kim et al42; Moyle et al43; Bade et al44; Patel et al45; Fernandez-Luque et al46, Brakenridge et al47; Jakicic et al48; Takami et al49; Jayaraman et al50; Bentley et al51; Martin et al52; Shuger et al53; Knight and Bidargaddi54; Ryu et al55; Onoue et al56; Najafi et al57; Whelan et al58; Leonardi et al59; Yang et al60; Kwasnicki et al61,62; Schwenk et al63; Sleep: Aktaruzzaman et al37; Moyle et al43; Maneyapanda et al64; and Rondanelli et al65
Movement, gait and fall analysis Motion sequences Up to 17 localizations Neurodegenerative diseases and rehabilitation LEGSys, Opal sensors, and ActiGraph Schwenk et al63; Kobsar and Ferber66; Silva de Lima et al67; Muchna et al68; Paul et al69; King et al70; Turner-McGrievy et al71; Brogioli et al72; Ashton et al73; Whelan et al74; Jagos et al75; Nguyen et al76; Weathers et al77; Carpinella et al78; Farooq et al79; Gorsic et al80; Conradsson et al81; Andrzejewski et al82; Grewal et al83; Menz et al84; Lee et al85; Schwenk et al86-88; and Zampieri et al89
Movement disorder Tremor and hypokinetic phases Chest, abdomen, finger, and lower leg Neurodegenerative diseases SNUMAP device and Mimamori-Gait system Silva de Lima et al67; Kim et al90; Uchino et al91; Bhidayasiri et al92,93; and Volpe et al94
Bed position Turns Back, hip, knee, and sole Immobilized patients Leaf sensor Pickham et al95,96
Electrodermal activity Conductivity Vegetative activity (sympathetic nervous system) Chest and wrist Stress Chiron wearable sensor platform, and Empathica Gaggioli et al29; Mlakar et al31; and Azevedo et al97
Temperature Temperature-dependent electrical resistance Vegetative activity Chest and upper arm wrist Monitoring, fatigue symptoms, and weight loss VitalPatch, Fit Core, and Sense Wear Mlakar et al31; Breteler et al32; and Jeffs et al34
Brain current EEG Cognitive activity Head Rett syndrome (with eye movement analysis) STARLAB Fabio et al98
Position and contact determination GPS and radiofrequency proximity sensors Surrounding contacts Head Rett syndrome and spread of infections STARLAB Signal et al99 and Voirin et al100

Abbreviation: ECG, Elctrocardiogramm; GPS, Global positioning system; HRV, Heart rate variability; PPG, photoplethysmography; RR, Respiratory rate.

Parameter and Sensors

In 60% of the studies, sensors were used that can measure linear (accelerometer/magnetometer) or angular acceleration (gyrometer) motion to record.40,41 Motion sensors were used for various applications:

  • In 31 studies, defined motion sequences (eg, rehabilitation exercises) or disorders (eg, neurodegenerative diseases) were examined with the aid of motion sensors.66,70

  • In 20 other studies, the focus was on general physical activity (eg, weight loss, sleep, or health promotion).40,41,42-53

In 20% of the identified studies, wearables were used to measure cardiovascular parameters (eg, heart rate, heart rate variability, and respiratory rate).8,9,20,26,31-38 A smaller number of sensors were used to measure (surface) skin temperature, electrodermal activity, oxygen saturation, and muscle activity.31,32,34,35

Calculated Parameters

At least with some wearables, the measurement method used does not directly measure the parameters, but estimate these from other signals, eg, if the heart rate variability is derived from PPG. Telemonitoring of vital signs with conventional medical products is not appropriate for capturing real-world data continuously.

Sensor Location and Areas of Usage

In 22.8% of the study goods, the sensors were attached to the wrist, which was the most frequently chosen localization.

In 25% of the 79 studies, the included patients were healthy or had frequent diseases of civilization such as obesity or arterial hypertension, with wearable activities (eg, physical activity and food intake) as endpoints.35,42 Patients with neurodegenerative diseases were the second-most studied cohort with 17.7%. On average, the studies had 135 participants; 25 of the 79 studies had >100 participants. The largest number of participants with 2659 participants was the randomized controlled clinical study of Steinhubl et al for the detection of atrial fibrillation with a wearable (adhesive ECG patch) in a risk population.26 The average age of the patients in the study cohorts was 54.5 years.

Additional Aspects

The evaluation of the publications regarding the sensors used shows a considerable problem, the nomenclature is used—at least partially—inconsistently. For example, the terms “electrodermal activity” and “galvanic skin reaction” are used synonymously.

In 81% of the studies, the authors assess the use of wearables technology as predominantly positive with regard to the specific question of the study and the authors’ prior considerations for wearable usage. While the other studies found the wearable not consistently beneficial due to, eg, optimizable wearable features for the studied question, lack of interpretation options because of poor reference method or technical issues.39,46,67,73

In two studies, the evaluation was based on machine learning approaches.

One general problem with wearables that occurred in the studies was false-negative or false-positive alarms. These can lead to a low acceptance by patients if, for example, a false-positive alarm rate of 2.5 in 24 hours occurs during the detection of generalized tonic-clonic seizures by a wearable.39

Discussion

In view of the rapid technical developments and the procedures already used by many patients as “self-monitoring,” a discussion of this topic is necessary. Wearables have an enormous potential for medical care, both for the care of individual patients and for prevention in the sense of public health approaches. The analysis of recent studies conducted with wearables shows that they can be successfully used in the medical context; however, certain aspects have to be taken into consideration:

  • Adequate selection of wearables taking into account the influencing factors with a precisely defined application context.

  • (Pilot) testing of the selected wearable in a small cohort of the target population and comparison of the results with those from an established reference method.

  • Clear definition of the practical procedures for the use of wearables for complete data collection and handling.

  • Selection of a suitable evaluation model of the determined data in relation to the research question.

  • Use of validated digital biomarkers (their development has only just begun, see “real-world walking speed” of the MOBILISED-D collaboration).

Vital parameters can be measured reliably and routinely with wearables. Ethical questions and data protection aspects need to be discussed transparently.

Footnotes

Declaration of Conflicting Interests: The author(s) declared no potential conflicts of interest with respect to the research, authorship, and/or publication of this article.

Funding: The author(s) received no financial support for the research, authorship, and/or publication of this article.

ORCID iD: Lutz Heinemann Inline graphic https://orcid.org/0000-0003-2493-1304

References

  • 1. Sim I. Mobile devices and health. N Engl J Med. 2019;381(10):956-968. [DOI] [PubMed] [Google Scholar]
  • 2. Williams B, Mancia G, Spiering W, et al. 2018. Practice guidelines for the management of arterial hypertension of the European Society of Cardiology and the European Society of Hypertension. Blood Press. 2018;27(6):314-340. [DOI] [PubMed] [Google Scholar]
  • 3. Piwek L, Ellis DA, Andrews S, Joinson A. The rise of consumer health wearables: promises and barriers. PLoS Med. 2016;13(2):e1001953. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4. Izmailova ES, Wagner JA, Perakslis ED. Wearable devices in clinical trials: hype and hypothesis. Clin Pharmacol Ther. 2018;104(1):42-52. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5. Allen J. Photoplethysmography and its application in clinical physiological measurement. Physiol Meas. 2007;28(3):R1-R39. [DOI] [PubMed] [Google Scholar]
  • 6. Evenson KR, Goto MM, Furberg RD. Systematic review of the validity and reliability of consumer-wearable activity trackers. Int J Behav Nutr Phys Act. 2015;12(1):159-181. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7. Byrom B, Watson C, Doll H, et al. Selection of and evidentiary considerations for wearable devices and their measurements for use in regulatory decision making: recommendations from the ePRO consortium. Value Health. 2018;21(6):631-639. [DOI] [PubMed] [Google Scholar]
  • 8. Gillinov S, Etiwy M, Wang R, et al. Variable accuracy of wearable heart rate monitors during aerobic exercise. Med Sci Sports Exerc. 2017;49(8):1697-1703. [DOI] [PubMed] [Google Scholar]
  • 9. Koshy AN, Sajeev JK, Nerlekar N, et al. Smart watches for heart rate assessment in atrial arrhythmias. Int J Cardiol. 2018;266:124-127. [DOI] [PubMed] [Google Scholar]
  • 10. Wang R, Blackburn G, Desai M, et al. Accuracy of wrist-worn heart rate monitors. JAMA Cardiol. 2017;2(1):104-106. [DOI] [PubMed] [Google Scholar]
  • 11. Goverdovsky V, von Rosenberg W, Nakamura T, et al. Hearables: multimodal physiological in-ear sensing. Sci Rep. 2017;7(1):6948. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12. Liu C, Correia R, Ballaji HK, Korposh S, Hayes-Gill BR, Morgan SP. Optical fibre-based pulse oximetry sensor with contact force detection. Sensors (Basel). 2018;18(11):pii:E3632. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13. Koppel R, Kuziemsky C. Healthcare data are remarkably vulnerable to hacking: connected healthcare delivery increases the risks. Stud Health Technol Inform. 2019;257:218-222. [PubMed] [Google Scholar]
  • 14. Ferrari FA, Viana RL, Lopes SR, Stoop R. Phase synchronization of coupled bursting neurons and the generalized Kuramoto model. Neural Netw. 2015;66:107-118. [DOI] [PubMed] [Google Scholar]
  • 15. Poplin RV, Varadarajan AV, Blumer K, et al. Prediction of cardiovascular risk factors from retinal fundus photographs via deep learning. Nat Biomed Eng. 2018;2(3):158-164. [DOI] [PubMed] [Google Scholar]
  • 16. Chen R, Snyder M. Promise of personalized omics to precision medicine. Wiley Interdiscip Rev Syst Biol Med. 2013;5(1):73-82. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 17. Topol EJ. The Creative Destruction of Medicine: How the Digital Revolution Will Create Better Health Care. New York: Basic Books; 2013. [Google Scholar]
  • 18. Perez MV, Mahaffey KW, Hedlin H, et al. Large-scale assessment of a smartwatch to identify atrial fibrillation. N Engl J Med. 2019;381(20):1909-1917. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19. Steinhubl SR, Muse ED, Barrett PM, Topol EJ. Off the cuff: rebooting blood pressure treatment. Lancet. 2016;388(10046):749. [DOI] [PubMed] [Google Scholar]
  • 20. Schoot TS, Weenk M, van de Belt TH, Engelen LJ, van Goor H, Bredie SJ. A new cuffless device for measuring blood pressure: a real-life validation study. J Med Internet Res. 2016;18(5):e85. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 21. Topouchian J, Agnoletti D, Blacher J, et al. Validation of four automatic devices for self-measurement of blood pressure according to the international protocol of the European Society of Hypertension. Vasc Health Risk Manag. 2011;7:709-717. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 22. Hertzberg O, Bauer A, Kuderle A, Pleitez MA, Mantele W. Depth-selective photothermal IR spectroscopy of skin: potential application for non-invasive glucose measurement. Analyst. 2017;142(3):495-502. [DOI] [PubMed] [Google Scholar]
  • 23. Caduff A, Zanon M, Mueller M, et al. The effect of a global, subject, and device-specific model on a noninvasive glucose monitoring multisensor system. J Diabetes Sci Technol. 2015;9(4):865-872. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 24. Weenk M, Koeneman M, van de Belt TH, Engelen LJLPG, van Goor H, Bredie SJH. Wireless and continuous monitoring of vital signs in patients at the general ward. Resuscitation. 2019;136:47-53. [DOI] [PubMed] [Google Scholar]
  • 25. Hay KA, Hanafi LA, Li D, et al. Kinetics and biomarkers of severe cytokine release syndrome after CD19 chimeric antigen receptor-modified T-cell therapy. Blood. 2017;130(21):2295-2306. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 26. Steinhubl SR, Waalen J, Edwards AM, et al. Effect of a home-based wearable continuous ECG monitoring patch on detection of undiagnosed atrial fibrillation: the mSToPS randomized clinical trial. JAMA. 2018;320(2):146-155. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 27. Koehler F, Koehler K, Deckwart O, et al. Efficacy of telemedical interventional management in patients with heart failure (TIM-HF2): a randomised, controlled, parallel-group, unmasked trial. Lancet. 2018;392(10152):1047-1057. [DOI] [PubMed] [Google Scholar]
  • 28. Conroy T, Guzman JH, Hall B, Tsouri G, Couderc JP. Detection of atrial fibrillation using an earlobe photoplethysmographic sensor. Physiol Meas. 2017;38(10):1906-1918. [DOI] [PubMed] [Google Scholar]
  • 29. Gaggioli A, Pallavicini F, Morganti L, et al. Experiential virtual scenarios with real-time monitoring (interreality) for the management of psychological stress: a block randomized controlled trial. J Med Internet Res. 2014;16(7):e167. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 30. Shilaih M, Clerck V, Falco L, Kubler F, Leeners B. Pulse rate measurement during sleep using wearable sensors, and its correlation with the menstrual cycle phases, a prospective observational study. Sci Rep. 2017;7(1):1294. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 31. Mlakar M, Puddu PE, Somrak M, et al. Mining telemonitored physiological data and patient-reported outcomes of congestive heart failure patients. PLoS One. 2018;13(3):e0190323. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 32. Breteler MJM, Huizinga E, van Loon K, et al. Reliability of wireless monitoring using a wearable patch sensor in high-risk surgical patients at a step-down unit in the Netherlands: a clinical validation study. BMJ Open. 2018;8(2):e020162. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 33. Razjouyan J, Grewal GS, Talal TK, Armstrong DG, Mills JL, Najafi B. Does physiological stress slow down wound healing in patients with diabetes? J Diabetes Sci Technol. 2017;11(4):685-692. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 34. Jeffs E, Vollam S, Young JD, Horsington L, Lynch B, Watkinson PJ. Wearable monitors for patients following discharge from an intensive care unit: practical lessons learnt from an observational study. J Adv Nurs. 2016;72(8):1851-1862. [DOI] [PubMed] [Google Scholar]
  • 35. Frias J, Virdi N, Raja P, Kim Y, Savage G, Osterberg L. Effectiveness of digital medicines to improve clinical outcomes in patients with uncontrolled hypertension and type 2 diabetes: prospective, open-label, cluster-randomized pilot clinical trial. J Med Internet Res. 2017;19(7):e246. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 36. Yu F, Bilberg A, Stenager E, Rabotti C, Zhang B, Mischi M. A wireless body measurement system to study fatigue in multiple sclerosis. Physiol Meas. 2012;33(12):2033-2048. [DOI] [PubMed] [Google Scholar]
  • 37. Aktaruzzaman M, Rivolta MW, Karmacharya R, et al. Performance comparison between wrist and chest actigraphy in combination with heart rate variability for sleep classification. Comput Biol Med. 2017;89:212-221. [DOI] [PubMed] [Google Scholar]
  • 38. Kroll RR, Boyd JG, Maslove DM. Accuracy of a wrist-worn wearable device for monitoring heart rates in hospital inpatients: a prospective observational study. J Med Internet Res. 2016;18(9):e253. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 39. Halford JJ, Sperling MR, Nair DR, et al. Detection of generalized tonic-clonic seizures using surface electromyographic monitoring. Epilepsia. 2017;58(11):1861-1869. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 40. McDermott MM, Spring B, Berger JS, et al. Effect of a home-based exercise intervention of wearable technology and telephone coaching on walking performance in peripheral artery disease: the HONOR randomized clinical trial. JAMA. 2018;319(16):1665-1676. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 41. Mendoza JA, Baker KS, Moreno MA, et al. A Fitbit and Facebook mHealth intervention for promoting physical activity among adolescent and young adult childhood cancer survivors: a pilot study. Pediatr Blood Cancer. 2017;64(15):e26660. [DOI] [PubMed] [Google Scholar]
  • 42. Kim Y, Lumpkin A, Lochbaum M, Stegemeier S, Kitten K. Promoting physical activity using a wearable activity tracker in college students: a cluster randomized controlled trial. J Sports Sci. 2018;36(16):1889-1896. [DOI] [PubMed] [Google Scholar]
  • 43. Moyle W, Jones C, Murfield J, et al. Effect of a robotic seal on the motor activity and sleep patterns of older people with dementia, as measured by wearable technology: a cluster-randomised controlled trial. Maturitas. 2018;110:10-17. [DOI] [PubMed] [Google Scholar]
  • 44. Bade BC, Brooks MC, Nietert SB, et al. Assessing the correlation between physical activity and quality of life in advanced lung cancer. Integr Cancer Ther. 2018;17(1):73-79. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 45. Patel MS, Benjamin EJ, Volpp KG, et al. Effect of a game-based intervention designed to enhance social incentives to increase physical activity among families: the BE FIT randomized clinical trial. JAMA Intern Med. 2017;177(11):1586-1593. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 46. Fernandez-Luque L, Singh M, Ofli F, et al. Implementing 360° quantified self for childhood obesity: feasibility study and experiences from a weight loss camp in Qatar. BMC Med Inform Decis Mak. 2017;17(1):37. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 47. Brakenridge CL, Fjeldsoe BS, Young DC, et al. Evaluating the effectiveness of organisational-level strategies with or without an activity tracker to reduce office workers’ sitting time: a cluster-randomised trial. Int J Behav Nutr Phys Act. 2016;13(1):115. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 48. Jakicic JM, Davis KK, Rogers RJ, et al. Effect of wearable technology combined with a lifestyle intervention on long-term weight loss: the IDEA randomized clinical trial. JAMA. 2016;316(11):1161-1171. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 49. Takami E, Sanchez JJ, Province R, Torres AJ. Implantable gastric electric stimulator with automatic daily activity compliance reporting. Conf Proc IEEE Eng Med Biol Soc. 2016;2016:5254–5257. [DOI] [PubMed] [Google Scholar]
  • 50. Jayaraman C, Mummidisetty CK, Jayaraman A. Effect of wearable sensor dynamics on physical activity estimates: a comparison between SCI vs. healthy individuals. Conf Proc IEEE Eng Med Biol Soc. 2016;2016:3282-3285. [DOI] [PubMed] [Google Scholar]
  • 51. Bentley CL, Otesile O, Bacigalupo R, et al. Feasibility study of portable technology for weight loss and HbA1c control in type 2 diabetes. BMC Med Inform Decis Mak. 2016;16(1):92. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 52. Martin SS, Feldman DI, Blumenthal RS, et al. mActive: a randomized clinical trial of an automated mhealth intervention for physical activity promotion. J Am Heart Assoc. 2015;4(11):pii:e002239. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 53. Shuger SL, Barry VW, Sui X, et al. Electronic feedback in a diet- and physical activity-based lifestyle intervention for weight loss: a randomized controlled trial. Int J Behav Nutr Phys Act. 2011;8(1):41. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 54. Knight A, Bidargaddi N. Commonly available activity tracker apps and wearables as a mental health outcome indicator: a prospective observational cohort study among young adults with psychological distress. J Affect Disord. 2018;236:31-36. [DOI] [PubMed] [Google Scholar]
  • 55. Ryu B, Kim N, Heo E, et al. Impact of an electronic health record-integrated personal health record on patient participation in health care: development and randomized controlled trial of MyHealthKeeper. J Med Internet Res. 2017;19(12):e401. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 56. Onoue T, Goto M, Kobayashi T, et al. Randomized controlled trial for assessment of Internet of things system to guide intensive glucose control in diabetes outpatients: Nagoya health navigator study protocol. Nagoya J Med Sci. 2017;79:323-329. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 57. Najafi B, Grewal GS, Bharara M, Menzies R, Talal TK, Armstrong DG. Can’t stand the pressure: the association between unprotected standing, walking, and wound healing in people with diabetes. J Diabetes Sci Technol. 2017;11(4):657-667. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 58. Whelan DF, O’Reilly MA, Ward TE, Delahunt E, Caulfield B. Technology in rehabilitation: evaluating the single leg squat exercise with wearable inertial measurement units. Methods Inf Med. 2017;56(2):88-94. [DOI] [PubMed] [Google Scholar]
  • 59. Leonardi L, Aceto MG, Marcotulli C, et al. A wearable proprioceptive stabilizer for rehabilitation of limb and gait ataxia in hereditary cerebellar ataxias: a pilot open-labeled study. Neurol Sci. 2017;38(3):459-463. [DOI] [PubMed] [Google Scholar]
  • 60. Yang YP, Wang CJ, Wang JJ, et al. The effects of an activity promotion system on active living in overweight subjects with metabolic abnormalities. Obes Res Clin Pract. 2017;11(6):718-727. [DOI] [PubMed] [Google Scholar]
  • 61. Kwasnicki RM, Ali R, Jordan SJ, et al. A wearable mobility assessment device for total knee replacement: a longitudinal feasibility study. Int J Surg. 2015;18:14-20. [DOI] [PubMed] [Google Scholar]
  • 62. Kwasnicki RM, Hettiaratchy S, Jarchi D, et al. Assessing functional mobility after lower limb reconstruction: a psychometric evaluation of a sensor-based mobility score. Ann Surg. 2015;261(4):800-806. [DOI] [PubMed] [Google Scholar]
  • 63. Schwenk M, Mohler J, Wendel C, et al. Wearable sensor-based in-home assessment of gait, balance, and physical activity for discrimination of frailty status: baseline results of the Arizona frailty cohort study. Gerontology. 2015;61(3):258-267. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 64. Maneyapanda MB, Stork R, Ingraham B, et al. Association of sleep with neurobehavioral impairments during inpatient rehabilitation after traumatic brain injury. Neuro Rehabilitation. 2018;43(3):319-325. [DOI] [PubMed] [Google Scholar]
  • 65. Rondanelli M, Opizzi A, Monteferrario F, Antoniello N, Manni R, Klersy C. The effect of melatonin, magnesium, and zinc on primary insomnia in long-term care facility residents in Italy: a double-blind, placebo-controlled clinical trial. J Am Geriatr Soc. 2011;59(1):82-90. [DOI] [PubMed] [Google Scholar]
  • 66. Kobsar D, Ferber R. Wearable sensor data to track subject-specific movement patterns related to clinical outcomes using a machine learning approach. Sensors (Basel). 2018;18(9):pii: E2828. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 67. Silva de Lima AL, Evers LJW, Hahn T, et al. Impact of motor fluctuations on real-life gait in Parkinson’s patients. Gait Posture. 2018;62:388-394. [DOI] [PubMed] [Google Scholar]
  • 68. Muchna A, Najafi B, Wendel CS, Schwenk M, Armstrong DG, Mohler J. Foot problems in older adults associations with incident falls, frailty syndrome, and sensor-derived gait, balance, and physical activity measures. J Am Podiatr Med Assoc. 2018;108(2):126-139. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 69. Paul SS, Dibble LE, Walther RG, Shelton C, Gurgel RK, Lester ME. Characterization of head-trunk coordination deficits after unilateral vestibular hypofunction using wearable sensors. JAMA Otolaryngol Head Neck Surg. 2017;143(10):1008-1014. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 70. King LA, Mancini M, Fino PC, et al. Sensor-based balance measures outperform modified balance error scoring system in identifying acute concussion. Ann Biomed Eng. 2017;45(9):2135-2145. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 71. Turner-McGrievy GM, Wilcox S, Boutte A, et al. The Dietary Intervention to Enhance Tracking with Mobile Devices (DIET Mobile) study: a 6-month randomized weight loss trial. Obesity (Silver Spring). 2017;25(8):1336-1342. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 72. Brogioli M, Popp WL, Schneider S, et al. Multi-day recordings of wearable sensors are valid and sensitive measures of function and independence in human spinal cord injury. J Neurotrauma. 2017;34(6):1141-1148. [DOI] [PubMed] [Google Scholar]
  • 73. Ashton LM, Morgan PJ, Hutchesson MJ, Rollo ME, Collins CE. Feasibility and preliminary efficacy of the ‘HEYMAN’ healthy lifestyle program for young men: a pilot randomised controlled trial. Nutr J. 2017;16(1):2. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 74. Whelan DF, O’Reilly MA, Ward TE, Delahunt E, Caulfield B. Technology in rehabilitation: comparing personalised and global classification methodologies in evaluating the squat exercise with wearable IMUs. Methods Inf Med. 2017;56(5):361-369. [DOI] [PubMed] [Google Scholar]
  • 75. Jagos H, Pils K, Haller M, et al. Mobile gait analysis via eSHOEs instrumented shoe insoles: a pilot study for validation against the gold standard GAITRite®. J Med Eng Technol. 2017;41(5):375-386. [DOI] [PubMed] [Google Scholar]
  • 76. Nguyen H, Lebel K, Boissy P, Bogard S, Goubault E, Duval C. Auto detection and segmentation of daily living activities during a Timed Up and Go task in people with Parkinson’s disease using multiple inertial sensors. J Neuroeng Rehabil. 2017;14(1):26. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 77. Weathers D, Siemens JC, Kopp SW. Tracking food intake as bites: effects on cognitive resources, eating enjoyment, and self-control. Appetite. 2017;111:23-31. [DOI] [PubMed] [Google Scholar]
  • 78. Carpinella I, Cattaneo D, Bonora G, et al. Wearable sensor-based biofeedback training for balance and gait in Parkinson disease: a pilot randomized controlled trial. Arch Phys Med Rehabil. 2017;98(4):622-630. [DOI] [PubMed] [Google Scholar]
  • 79. Farooq M, McCrory MA, Sazonov E. Reduction of energy intake using just-in-time feedback from a wearable sensor system. Obesity (Silver Spring). 2017;25(4):676-681. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 80. Gorsic M, Cikajlo I, Novak D. Competitive and cooperative arm rehabilitation games played by a patient and unimpaired person: effects on motivation and exercise intensity. J Neuroeng Rehabil. 2017;14(1):23. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 81. Conradsson D, Nero H, Lofgren N, Hagstromer M, Franzen E. Monitoring training activity during gait-related balance exercise in individuals with Parkinson’s disease: a proof-of-concept-study. BMC Neurol. 2017;17(1):19. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 82. Andrzejewski KL, Dowling AV, Stamler D, et al. Wearable sensors in Huntington disease: a pilot study. J Huntingtons Dis. 2016;5(2):199-206. [DOI] [PubMed] [Google Scholar]
  • 83. Grewal GS, Schwenk M, Lee-Eng J, et al. Sensor-based interactive balance training with visual joint movement feedback for improving postural stability in diabetics with peripheral neuropathy: a randomized controlled trial. Gerontology. 2015;61(6):567-574. [DOI] [PubMed] [Google Scholar]
  • 84. Menz HB, Auhl M, Tan JM, Levinger P, Roddy E, Munteanu SE. Biomechanical effects of prefabricated foot orthoses and rocker-sole footwear in individuals with first metatarsophalangeal joint osteoarthritis. Arthritis Care Res (Hoboken). 2016;68(5):603-611. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 85. Lee SH, Yeh SC, Chan RC, Chen S, Yang G, Zheng LR. Motor ingredients derived from a wearable sensor-based virtual reality system for frozen shoulder rehabilitation. Biomed Res Int. 2016;2016:7075464. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 86. Schwenk M, Grewal GS, Holloway D, Muchna A, Garland L, Najafi B. Interactive sensor-based balance training in older cancer patients with chemotherapy-induced peripheral neuropathy: a randomized controlled trial. Gerontology. 2016;62(5):553-563. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 87. Schwenk M, Sabbagh M, Lin I, et al. Sensor-based balance training with motion feedback in people with mild cognitive impairment. J Rehabil Res Dev. 2016;53(3):945-958. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 88. Schwenk M, Grewal GS, Honarvar B, et al. Interactive balance training integrating sensor-based visual feedback of movement performance: a pilot study in older adults. J Neuroeng Rehabil. 2014;11(1):164. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 89. Zampieri C, Salarian A, Carlson-Kuhta P, Nutt JG, Horak FB. Assessing mobility at home in people with early Parkinson’s disease using an instrumented Timed Up and Go test. Parkinsonism Relat Disord. 2011;17(4):277-280. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 90. Kim HB, Lee WW, Kim A, et al. Wrist sensor-based tremor severity quantification in Parkinson’s disease using convolutional neural network. Comput Biol Med. 2018;95:140-146. [DOI] [PubMed] [Google Scholar]
  • 91. Uchino K, Shiraishi M, Tanaka K, Akamatsu M, Hasegawa Y. Impact of inability to turn in bed assessed by a wearable three-axis accelerometer on patients with Parkinson’s disease. PLoS One. 2017;12(11):e0187616. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 92. Bhidayasiri R, Sringean J, Chaiwong S, et al. Rotigotine for nocturnal hypokinesia in Parkinson’s disease: quantitative analysis of efficacy from a randomized, placebo-controlled trial using an axial inertial sensor. Parkinsonism Relat Disord. 2017;44:124-128. [DOI] [PubMed] [Google Scholar]
  • 93. Bhidayasiri R, Sringean J, Anan C, Boonpang K, Thanawattano C, Ray Chaudhuri K. Quantitative demonstration of the efficacy of night-time apomorphine infusion to treat nocturnal hypokinesia in Parkinson’s disease using wearable sensors. Parkinsonism Relat Disord. 2016;33(suppl 1):s36-s41. [DOI] [PubMed] [Google Scholar]
  • 94. Volpe D, Giantin MG, Fasano A. A wearable proprioceptive stabilizer (Equistasi®) for rehabilitation of postural instability in Parkinson’s disease: a phase II randomized double-blind, double-dummy, controlled study. PLoS One. 2014;9(11):e112065. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 95. Pickham D, Pihulic M, Valdez A, Mayer B, Duhon P, Larson B. Pressure injury prevention practices in the intensive care unit: real-world data captured by a wearable patient sensor. Wounds. 2018;30(8):229-234. [PubMed] [Google Scholar]
  • 96. Pickham D, Berte N, Pihulic M, Valdez A, Mayer B, Desai M. Effect of a wearable patient sensor on care delivery for preventing pressure injuries in acutely ill adults: a pragmatic randomized clinical trial (LS-HAPI study). Int J Nurs Stud. 2018;80:12-19. [DOI] [PubMed] [Google Scholar]
  • 97. Azevedo RT, Bennett N, Bilicki A, Hooper J, Markopoulou F, Tsakiris M. The calming effect of a new wearable device during the anticipation of public speech. Sci Rep. 2017;7(1):2285. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 98. Fabio RA, Billeci L, Crifaci G, Troise E, Tortorella G, Pioggia G. Cognitive training modifies frequency EEG bands and neuropsychological measures in Rett syndrome. Res Dev Disabil. 2016;53-54:73-85. [DOI] [PubMed] [Google Scholar]
  • 99. Signal LN, Smith MB, Barr M, et al. Kids’Cam: an objective methodology to study the world in which children live. Am J Prev Med. 2017;53(3):e89-e95. [DOI] [PubMed] [Google Scholar]
  • 100. Voirin N, Payet C, Barrat A, et al. Combining high-resolution contact data with virological data to investigate influenza transmission in a tertiary care hospital. Infect Control Hosp Epidemiol. 2015;36(3):254-260. [DOI] [PubMed] [Google Scholar]

Articles from Journal of Diabetes Science and Technology are provided here courtesy of Diabetes Technology Society

RESOURCES