. 2020 Dec 1;5(56):2881–2888. doi: 10.1557/adv.2020.378

Strap Performance of N95 Filtering Facepiece Respirators After Multiple Decontamination Cycles

Aaron W Richardson 1,, Kent C Hofacre 1, Patrick H Keyes 1, Rachel M Thurston 1, John D Clay 1
PMCID: PMC7790017  PMID: 33437533

Abstract

The Battelle Critical Care Decontamination System™ (CCDS™) decontaminates N95 filtering facepiece respirators (FFRs) using vapor phase hydrogen peroxide (VPHP) for reuse when there is a critical supply shortage. The Battelle CCDS received an Emergency Use Authorization (EUA) from the US Food and Drug Administration (FDA) in March 2020. This research focused on evaluating the mechanical properties of the straps as an indicator of respirator fit. The objective was to characterize the load generated by the straps following up to 20 don/doff and decontamination cycles in Battelle’s CCDS. In general, the measured loads at 50 and 100% strains after 20 cycles were similar (±15%) to the as-received controls. Qualitatively, reductions in the load may be associated with loss of elasticity in the straps, potentially reducing the ability to obtain a proper fit. However, small changes in strap elasticity may not affect the ability to obtain a proper fit given the potential for variation in strap length and positioning on the head. Regardless, prior to reusing a N95 respirator, it is important to complete a visual inspection to ensure it is not damaged, malformed, or soiled. If so, it is recommended to discard the respirator and use a different one. Similarly, the respirator should be discarded if the wearer cannot obtain a proper fit during the user seal check.

Footnotes

At the date this manuscript was written, URLs or links referenced herein were deemed to be useful supplementary material to this manuscript. Neither the author nor the Materials Research Society warrants or assumes liability for the content or availability of URLs referenced in this manuscript.

References

  • [1].Centers for Disease ControlPrevention . Implementing Filtering Facepiece Respirator (FFR) Reuse, Including Reuse after Decontamination, When There Are Known Shortages of N95 Respirators. 2020. [Google Scholar]
  • [2].U.S. FoodDrug Administration . Emergency User Authorizations. 2020. [Google Scholar]
  • [3].Viscusi, D., M. Bergman, B. Eimer, and R. Shaffer, “Evaluation of Five Decontamination Methods for Filtering Facepiece Respirators,” Annal of Occupational Hygiene, Vol. 53(8), pp. 815-827. [DOI] [PMC free article] [PubMed]
  • [4].Occupational SafetyHealth Administration . Enforcement Guidance on Decontamination of Filtering Facepiece Respirators in Healthcare During the Coronavirus Disease (COVID-19) Pandemic. 2020. [Google Scholar]
  • [5].National Institute for Occupational SafetyHealth NIOSH, NIOSH Test Plan . Assessment of Filter Penetration Performance and Fit for Decontaminated N95 Respirators. 2010. [Google Scholar]
  • [6].Roberge R, Niezgoda G, Benson S. Analysis of Forces Generated by N95 Filtering Facepiece Respirator Tethering Devices: A Pilot Study. Journal of Occupational and Environmental Hygiene. 2012;9:517–523. doi: 10.1080/15459624.2012.695962. [DOI] [PubMed] [Google Scholar]
  • [7].Rottach DR. Stockpiled N95 Filtering Facepiece Respirators Polyisoprene Strap Performance. Journal of the International Society for Respiratory Protection. 2017;34(2):69–80. [PMC free article] [PubMed] [Google Scholar]
  • [8].Roberge R. Effect of Upper Strap Downward Displacement of N95 Filtering Facepiece Respirator Fit Factors: A Pilot Study. Journal of Occupational and Environmental Hygiene. 2014;11(5):338–341. doi: 10.1080/15459624.2013.866716. [DOI] [PMC free article] [PubMed] [Google Scholar]