Variation in LOD Across SARS-CoV-2 Assay Systems: Need for Standardization

Youvraj Sohni 1,
PMCID: PMC7798922

Abstract

Multiple SARS-CoV-2 emergency use authorization (EUA) tests are being used for clinical testing across various clinical testing laboratories for meeting the diagnostic challenges of the ongoing pandemic. However, cross-assay variations in performance characteristics need to be recognized. A better understanding is needed of the clinical implications of cross-assay variation in performance characteristics, particularly in the limit of detection (LOD) of the SARS-CoV-2 assays used for clinical testing. Herein, a snapshot of the diversity of SARS-CoV-2 EUA analytical assay systems including methodologies, assay designs, and technology platforms is presented. Factors affecting the variations in LOD are discussed. Potential measures that may standardize across the various assay systems are suggested. Development of international standards and reference materials for the establishment of performance characteristics may substantially alleviate potential clinical decision-making challenges. Finally, cross-assay variation in LODs among the diverse SARS-CoV-2 diagnostic assays impacts clinical decision-making with multiple assay systems in use and lack of standardization across platforms. International standards in parallel with continued cross-platform studies and collaborative efforts across pertinent healthcare entities will help mitigate some of the clinical decision-making challenges.

Keywords: SARS-CoV-2, RT-PCR, LOD, CT, Standards, COVID-19


In response to the ongoing public health emergency for COVID-19, the U.S. Food & Drug Administration (FDA) has issued emergency use authorization (EUA) for several SARS-CoV-2 diagnostic assays developed by IVD manufacturers, academic centers, and commercial reference laboratories. Several of these EUA assays are now part of the SARS-CoV-2 clinical testing workflow. The list of assays that receive EUA continues to increase, and the complete list of these assays can be accessed from the FDA website.1

The need for widespread availability of accurate diagnostic testing for SARS-CoV-2 is critical for appropriate and timely clinical decision-making. Multiple testing options are needed as SARS-CoV-2 testing capacity increases across the various clinical testing laboratories. In the foreseeable future, a limited manufacturing allocation of testing kits and related consumables compels individual clinical laboratories to use multiple assay systems to meet SARS-CoV-2 testing needs.2,3

It is reasonable to expect that cross-assay variation in performance characteristics of the EUA assays currently in use will have a bearing on the results of patient testing and subsequent clinical decision-making. In the context of the ongoing pandemic, ensuring that patient testing is not compromised is essential. Therefore, the clinical implications of cross-assay variation in performance characteristics must be recognized. This is particularly relevant for those patient specimen testing results that are close to the limit of detection (LOD) of the assay systems in use. These high cycle threshold (CT) SARS-CoV-2 results may not meet the 95% confidence interval (CI) detection criteria, and such test results may be discordant when comparatively tested across different assay systems. The potential impact on SARS-CoV-2 transmission and the clinical significance of such patient results is yet to be completely understood.

Clinical Implications

A significant majority of those with SARS-CoV-2 infections present with mild symptoms, are asymptomatic with positive results, or never develop any signs or clinical symptoms of COVID-19.4 Thus a better understanding of the spectrum of the disease is needed because SARS-CoV-2 infection can be detected before the development of clinical symptoms. Although an ideal objective for a test is to eliminate the possibility of presumed false-negative patient results, the challenge is knowing the clinical implications of SARS-CoV-2 testing results that fall within the window of difference in LOD between the assay systems used in the clinical testing workflow.

Currently the EUA assays in use for SARS-CoV-2 diagnostics are all qualitative tests and generally provide a binary test with a positive or negative result. A third presumptive positive result is part of some tests (for example, the Cepheid GeneXpert assay; Cepheid, Sunnyvale, CA). As per the Centers for Disease Control and Prevention (CDC) recommendations, viral testing for SARS-CoV-2 is considered to be diagnostic when conducted among individuals with symptoms consistent with COVID-19. In addition, diagnostic testing may be considered for asymptomatic individuals with known or suspected recent exposure to SARS-CoV-2 to control transmission or to determine the resolution of infection. Viral testing is considered as screening when conducted among asymptomatic individuals without known or suspected exposure to SARS-CoV-2 for early identification and as surveillance when conducted among asymptomatic individuals to detect transmission hot spots or characterize disease trends. The CDC also recommends testing for all close contacts of persons with SARS-CoV-2 infection and for those in some high-risk settings.5

Assays with a relatively low LOD may detect SARS-CoV-2 in specimens with low viral loads that may be missed by assays with higher LODs. Viral loads in throat swab and sputum specimens peak at approximately 5–6 days after symptom onset, but specimens taken from an individual with a recently acquired SARS-CoV-2 infection can have low viral loads. One small study suggests that viral load detected in an asymptomatic patient may be similar to that in symptomatic patients, which in turn suggests the transmission potential of asymptomatic or minimally symptomatic patients.6,7 Ultimately, correlating virus counts and therapeutic measures with outcomes may result in different strategies of care or isolation.

SARS-CoV-2 Shedding and Reverse-Transcription Polymerase Chain Reaction

Viral RNA may be detected from patient specimens weeks after the infectious stage has ended post-onset of symptoms. Sun et al8 reported persistent shedding of viral RNA in nasopharyngeal (NP) swabs and feces specimens, with the estimated time until loss of RNA detection ranging from 45.6 days for NP swab specimens to 46.3 days for fecal specimens in mild cases of infection and from 48.9 days for NP swab specimens to 49.4 days for fecal specimens in severe cases of infection. However, the authors noted that the median time for throat specimens from mild cases of infection was 15.6 days (95% CI, 11.8–20.7 days) and that the 95th percentile was 32.8 days (95% CI, 25.9–42.3 days). Therefore, Sun et al.8 suggested that the detection of viral RNA for mild cases of infection in throat swab specimens at the 50th day after illness onset should be a low-probability event, beyond the 95th percentile limit, as for fecal specimens. Similarly, according to van Kampen et al,9 the probability of detecting infectious virus drops below 5% 15 days post-onset of symptoms. Lan et al10 reported positive reverse-transcription polymerase chain reaction (RT-PCR) results in throat swab specimens from patients who recovered from mild COVID-19 for 50 days at maximum.

Chen et al11 studied viral shedding at multiple timepoints in stool specimens and analyzed its correlation with clinical manifestations and the severity of illness. Those authors found that SARS-CoV-2 RNA in stool specimens was not associated with the presence of gastrointestinal symptoms and the severity of illness. A majority of patients may remain positive for viral RNA in the feces after the pharyngeal swabs are returned negative. The results show the presence of SARS-CoV-2 RNA in the feces of patients with COVID-19 and suggest the possibility of SARS-CoV-2 transmission via the fecal-oral route.9 Gupta et al12 reported that the duration for fecal shedding of viral RNA ranged from 1 to 33 days and in one patient was up to 47 days from symptom onset, longer after the clearance of respiratory specimens. In spite of viral RNA concentration, virus isolation from stool specimens was not successful from patient specimens collected between days 6 and 12. There appears to be active replication of the virus in the throat during the first 5 days after the onset of symptoms with no virus replication in stool specimens.13

An RT-PCR assay may have the analytical sensitivity for detecting a few hundred copies of viral RNA, but that does not indicate the infectious status of the patient. For these reasons, identifying such patients with low levels of detectable viral RNA that straddle the LOD of the assay of choice needs more clarity. It may be challenging to interpret RT-PCR tests for SARS-CoV-2 infection particularly early in the course of infection when these results underlie clinical decision-making to minimize onward transmission.

CT Values and Infectivity

A relatively lower CT value increases the probability of the presence of infectious virus, but the CT value itself is not enough to determine whether the virus can be cultured from the specimens. Thus a limitation of PCR and other nucleic acid-based diagnostics is that they do not establish infectivity. Studies have shown that viral infectivity, as defined by culturability, is decreased when RT-PCR CT values are >24, with the odds ratio for infectivity being decreased by 32% for every 1-unit increase in CT.14 Wölfel et al13 were able to isolate virus during the first week of symptoms particularly from a majority of sputum specimens but could not isolate the virus from specimens taken after day 8. The infectivity of the virus culture does appear to correlate with viral load, and specimens that contain <106 copies/mL do not yield an isolate.13,15

In addition, Huang et al15 reported that the mean CT values of culturable oropharyngeal (OP) and NP specimens were similar whereas culturable sputum specimens had relatively lower CT values. Specimens for a successful viral culture contained higher viral RNA counts, but note that viral genome integrity is an important factor for culturability and infectivity assessment. In their study, Huang et al15 stated that a genome copy number of 5 to 6 log10 genome copies/mL appears to be the minimal viral load necessary for virus isolation. Detection of SARS-CoV-2 by RT-PCR is still the diagnostic gold standard, and the World Health Organization recommends 2 negative tests at least 24 hours apart as 1 of the key criteria for a clinically recovered patient.16 In their study, Bullard et al14 showed a link between in vitro viral growth, CT value, and symptom-to-test. Thus, these authors suggested that a CT value >24 in conjunction with duration of symptoms >8 days could be used together to determine the duration of infectivity in patients.

Nucleic acid amplification kits may sometimes produce false-negative results that do not match clinical features. The false-negative rate is a function of time with virus exposure and subsequent pathogenesis whether infection is presymptomatic, asymptomatic, or paucisymptomatic. In a recent case study, Lv et al17 reported a successively RT-PCR-negative patient with COVID-19. The authors emphasized the importance of clinical signs and symptoms, other laboratory findings, and chest computed tomography images that should be considered in spite of RT-PCR-negative results. They also reported the dynamic change process of SARS-CoV-2 target genes by RT-PCR testing during the course of infection of a patient with COVID-19 from successive negative results to a successive single positive N gene to 2 positive ORF1a/b and N genes.

If clinical suspicion is high, then infection should not be ruled out on the basis of RT-PCR alone and the clinical and epidemiologic situation should be carefully considered.18 Because RT-PCR positivity persists beyond the period of infectivity, clinical criteria that include isolation of hospitalized patients for 14 days from symptom onset or 72 hours symptom-free would be the prudent approach. Early discharge followed by home isolation of patients after day 10 of symptoms appears to have little risk of infectivity.13,14

Factors such as comorbidities and/or age are certainly part of the consideration in addition to other clinical findings that inform clinical decision-making. Children with SARS-CoV-2 infection have milder clinical symptoms and fewer laboratory and radiologic abnormalities. Nonetheless, understanding the perinatal outcomes of infants born to women infected with SARS-CoV-2 during pregnancy is important for early identification of children with SARS-CoV-2 to provide optimal medical care and control the pandemic.19

Finally, the CDC advises clinicians to use their judgment to determine if a patient has signs or symptoms compatible with COVID-19 and whether the patient should be tested. Clinicians are encouraged to consider testing for other causes of respiratory illness in addition to testing for SARS-CoV-2 depending on patient age, season, or clinical setting. Further details and updated recommendations may be accessed on the CDC website.5

SARS-CoV-2 Assays

Table 1 is a representation of some of the molecular SARS-CoV-2 EUA assays that are run on a diverse array of platforms. Several of the assay systems have integrated automated nucleic acid extraction and analytical steps. The automated high-throughput testing platforms such as the Roche (Indianapolis, IN) cobas 6800/8800 feature integrated specimen prep and RT-PCR. Similarly, the Hologic (San Diego, CA) Panther/Fusion systems are high-throughput and automated but employ transcription-mediated amplification (TMA) methodology.

Table 1.

Representative SARS-CoV-2 EUA Assay Systems.

Test/Instrument Methodology Gene Targets Claimed LOD
Roche cobas SARS-CoV-2/cobas 6800/8800 systems RT-PCR ORF1a/b 0.007 TCID50/mL
Thermofisher TaqPath COVID-19 Combo Kit/Applied Biosystems 7500, QuantStudio 5 and 7 systems RT-PCR ORF1a/b, N, S 10 GCE/reaction
Hologic Panther Fusion SARS-CoV-2 Assay/Panther Fusion TMA ORF1 0.01 TCID50/mL
Abbott Real-Time SARS-CoV-2 Assay/m2000sp plus m2000rt systems RT-PCR N, RdRp 3.1 GE/reaction or 100 copies/mL
DiaSorin Molecular Simplexa COVID-19 Direct/Liaison MDX RT-PCR ORF1, S 500 copies/mL
Cepheid Xpert Xpress SARS-CoV-2 Test/GeneXpert Dx or GeneXpert Infinity systems RT-PCR N2 and E 250 copies/mL
BioFire COVID-19 Test/FilmArray 2.0, FilmArray Torch System RT-PCR ORF1a/b, ORF8 0.022 TCID50/mL or 330 copies/mL
Mesa Biotech/Accula SARS-CoV-2 Molecular LFA N 200 copies/mL
Abbott ID NOW COVID-19 Isothermal RdRp 125 GE/mL
Luminex NxTAG CoV Extended Panel Assay/Luminex MAGPIX RT-PCR ORF1a/b, N, E 5 × 103 GCE/mL
NeuMoDx SARS-CoV-2 Assay/288 Molecular System and NeuMoDx 96 Molecular System RT-PCR Nsp2, N 150 copies/mL
QIAstat-Dx Respiratory SARS-CoV-2 Panel RT-PCR Orf1b (Rdrp), E 500 copies/mL
Luminex ARIES SARS-CoV-2 Assay RT-PCR ORF1a/b, N 333 copies/mL
Gnomegen COVID-19 RT-Digital PCR Detection Kit/Gnomegen Real-Time Digital PCR Instrument or QuantStudio 3D Digital PCR System RT-digital PCR N1/N2 8 copies/reaction
InBios Smart Detect SARS-CoV-2 rRT-PCR Kit/7500 Fast Dx Real-Time PCR Instrument (Applied Biosystems) or CFX96 Touch Real-Time PCR Detection System (BioRad) E, N, ORF1b/RdRp 1100 GE/mL or 12.5 GE/reaction
BD SARS-CoV-2 Reagents/BD MAX System RT-PCR N1, N2 40 GE/mL
Atila Biosystems iAMP COVID-19 Detection Kit/BioRad CFX96 Real-Time System Isothermal amplification N, ORF1a/b 4 copies/ μL
Rheonix COVID-19 MDx Assay/Rheonix Encompass MDx Workstation RT-PCR N1 0.025 TCID50/mL 625 GE/mL)
Bio-Rad SARS-CoV-2 ddPCR Test/QX200 and QXDx Droplet Digital PCR systems ddPCR N1, N2 625 copies/mL
BioFire Respiratory Panel 2.1 (RP2.1)/BioFire FilmArray System RT-PCR S, M 500 copies/mL or 0.069 TCID50/mL
Sherlock CRISPR SARS-CoV-2 kit RT-LAMP/CRISPR-based detection N, O 4.5 copies/μL
Abbott Alinity m SARS-CoV-2 assay/Alinity m System RT-PCR RdRp, N 100 copies/mL
HDPCR SARS-CoV-2 Assay/Applied Biosystems QuantStudio 12K system RT-PCR N1, N2 250 copies/mL
Illumina COVIDSeq Test/NovaSeq 6000 Sequencing System NGS Virus genome 1000 copies/mL
BioCode SARS-CoV-2 Assay/BioCode MDx-3000 RT-PCR N 0.0117 TCID50/mL

CRISPR, clustered regularly interspaced short palindromic repeats; ddPCR, digital droplet polymerase chain reaction; EUA, emergency use authorization; FDA, U.S. Food and Drug Administration; GCE, genome copy equivalent; GE, genome equivalents; IFU, instructions for use; LFA, lateral flow assay; LOD, limit of detection; NGS, next-generation sequencing; RT-LAMP, reverse-transcription loop-mediated isothermal amplification; RT-PCR, reverse-transcription polymerase chain reaction; TCID50, median tissue culture infectious dose; TMA, transcription-mediated amplification. Complete assay list may be accessed on the FDA website. The information summarized herein has been derived from the manufacturer’s IFU, also accessible on the FDA website. Test/instrument, gene targets, and claimed LOD units are as reported in the manufacturer’s IFU. Roche Diagnostics, Indianapolis, IN; Thermofisher Scientific, Waltham, MA/Applied Biosystems, Foster City, CA; Hologic, Inc., San Diego, CA; Abbott Diagnostics, Lake Forest, IL; Diasorin Molecular, Cypress, CA; Cepheid, Sunnyvale, CA; Biofire, Salt Lake City, UT; Mesa Biotech, Inc., San Diego, CA; Abbott Diagnostics, Lake Forest, IL; Luminex Corporation, Austin, TX; NeuMoDx Molecular, Inc., Ann Arbor, MI; Qiagen, Germantown, MD; Gnomegen, San Diego, CA; InBios, Seattle, WA; Becton Dickinson and Company, Sparks, MD; Atila Biosystems, Mountain View, CA; Rheonix Inc., Ithaca, NY; Bio-Rad Laboratories, Hercules, CA; Biofire, Salt Lake City, UT; Sherlock Biosciences, Cambridge, MA; Abbott Diagnostics, Lake Forest, IL; Applied Biosystems, Foster City, CA; Illumina, San Diego, CA; Applied BioCode, Santa Fe Springs, CA.

At the other end of the spectrum, designated point-of-care (POC) devices such as the Abbott ID NOW (Lake Forest, IL) and the Mesa Biotech (San Diego, CA) Accula platform exist, with the latter designed to be a molecular lateral flow assay readout system. The Cepheid Xpert Xpress SARS-CoV-2 assay also has the option to be used in a POC format. The POC-type devices include those that are deemed waived to be tested at the point of collection.

Some IVD manufacturers have integrated the SARS-CoV-2 EUA assay to be part of respiratory panel testing. The BioFire (Salt Lake City, UT) Respiratory Panel 2.1 is a multiplexed nucleic acid test intended for the simultaneous qualitative detection and differentiation of nucleic acids from multiple viral and bacterial respiratory organisms including SARS-CoV-2. Similarly, the QIAstat Respiratory panel (Qiagen, Germantown, MD) is designed to allow the discrimination of respiratory pathogens including SARS-CoV-2.

Other novel technologies continue to be applied to develop SARS-CoV-2 assays and have received EUA for SARS-CoV-2 testing. The Rheonix (Ithaca, NY) assay uses a proprietary microfluidic Rheonix CARD cartridge technology that integrates specimen preparation and target detection using RT-PCR. Digital PCR-based assays have received EUA, including the Gnomegen (San Diego, CA) COVID-19 RT-Digital PCR Kit, which is validated on the Gnomegen real-time PCR instrument, and the QuantStudio 3D Digital PCR Systems (Applied Biosystems, Foster City, CA). Similarly, the Bio-Rad (Hercules, CA) SARS-CoV-2 ddPCR test is a droplet digital PCR assay.

The majority of the EUA assays are based on RT-PCR methodology, although isothermal nucleic acid amplification methodology such as TMA along with next-generation sequencing (NGS) and clustered regularly interspaced short palindromic repeats–based methodologies have also received EUA for SARS-CoV-2 clinical diagnostic testing. For example, the Illumina (San Diego, CA) COVIDSeq Test on the NovaSeq 6000 platform is an NGS assay in which the viral genome is sequenced. It is entirely conceivable that novel and cutting-edge technologies would be used in SARS-CoV-2 diagnostic testing. Many such technology platforms would be amenable for use in niche testing areas for SARS-CoV-2 depending on patient testing requirements.

In many instances, commercial IVD manufacturers have developed and validated current EUA assays on their own hardware systems for both nucleic acid extraction and the analytical method. Alternatively, other commercial manufacturers have developed kits that have been validated on generic platforms for either manual or automated nucleic acid extraction followed by analytical runs on, eg, RT-PCR platforms sold by other manufacturers.

As expected, cross-comparison studies have shown that there is variation in analytical sensitivity across these different methodologies and technology platforms.6,20-23

Cross-Platform Comparison Studies

Earlier in the pandemic, Carter et al20 reviewed molecular assay methodologies used for EUA tests. Subsequently, several studies have emerged in which performance characteristics across several SARS-CoV-2 EUA assays were compared and useful cross-comparison data with respect to LOD were provided.22,23

Cradic et al2 evaluated the clinical performance of 3 molecular assays using NP swab specimens and compared the Abbott (Lake Forest, IL) ID NOW COVID-19, the DiaSorin Molecular (Cypress, CA) Simplexa COVID-19 Direct, and the Roche cobas 6800 SARS-CoV-2 assays. The Simplexa and Roche cobas assays reportedly had approximately 10 to 100 times lower LODs than the Abbott ID NOW. Based on these evaluations, Cradic et al2 proposed a multiplatform testing approach in relation to the patient population and the assay performance characteristics including assay sensitivity.

Craney et al24 compared the diagnostic performance of 2 high-throughput assay systems, the Roche cobas 6800 and the Hologic Panther Fusion, using 389 nasopharyngeal specimens. The overall percentage agreement between the platforms was 96.4% (κ = 0.922). Remarkably, as reported, no significant difference existed between the corresponding CT values generated on the 2 systems (P value = .88). The authors suggested that these assay systems can be considered comparable in terms of their clinical performance.

Lieberman et al3 compared the analytical performance of the CDC primer set LDT and the Cepheid, DiaSorin Simplexa, Hologic Panther, and Roche cobas assays using a total of 169 NP swabs. The CDC-LDT and Cepheid Xpert Xpress SARS-CoV-2 assays were the most sensitive assays for SARS-CoV-2, with 100% agreement across specimens. However, as per Lieberman et al,3 the Hologic Panther Fusion, the DiaSorin Simplexa, and the Roche cobas 6800 assays only failed to detect positive specimens near the LOD of the CDC-LDT assay. The authors highlighted the importance of having multiple viral detection testing platforms available in a public health emergency.

Zhen, Manji, et al22 stressed the importance of accurate testing results in patient management and, to that end, the importance of false negative results because they inevitably lead to more exposures that can be potentially devastating. Zhen, Smith, et al23 compared the performance characteristics of 4 PCR methods that yielded comparable results (κ ≥ 0.96); however, they did observe a notable difference when it came to overall LOD, with the GenMark (Carlsbad, CA) having the overall highest LOD of all 4 platforms evaluated and the DiaSorin Simplexa having the lowest LOD (39 ± 23 copies/mL), closely followed by the Hologic Panther Fusion (83 ± 36 copies/mL). Their modified CDC assay showed a final LOD of 779 ± 27 copies/mL. The clinical correlation was also consistent with LOD findings where both the DiaSorin Simplexa and Hologic Panther Fusion assays had 100% positive agreement and detected all specimens deemed positive by the consensus test result, which was 3 of 4 evaluated assays as the gold standard.

As is to be expected, differences exist in LODs across assays, as reported in these studies. More such studies25,26 continue to be published with data that will better define the applicability and optimal use of these different assays in laboratory testing to inform clinical decision-making.

Factors Affecting LOD

Test manufacturers seeking EUA are required to establish performance characteristics such as accuracy, precision, analytical sensitivity, and analytical specificity. The analytical sensitivity of an assay is the ability of an assay to detect very low concentrations of a given analyte in a biological specimen. The LOD is the lowest actual concentration of an analyte that can be consistently detected in typically ≥95% of specimens tested; thus it is an important performance characteristic. Analytical performance at the low concentration limit is critical for infectious disease diagnostics.

As the representative assays presented in Table 1 indicate, there is considerable variability in the manner in which molecular assay LODs are reported in manufacturer IFUs. Per manufacturer IFUs, the claimed LOD is reported as median tissue culture infectious dose (TCID50)/mL, copies/mL, genome copy equivalent/mL, or copies/reaction. The genome equivalent is the amount of DNA necessary to be present in a purified specimen to guarantee that all genes are present.

The calculation for TCID50/mL depends on the cell line being used for culture and is based on the degree of cytopathic effect observed in the tissue culture. Viral counts reported by TCID50 tend to be much lower than RT-quantitative PCR measurements (Bar-On27). The RNA counts can overestimate infectious virions because the presence of viral RNA does not necessarily imply the presence of infectious virions. To assess the concentration of infectious viruses, researchers typically measure TCID50, which involves infecting replicate cultures of susceptible cells with dilutions of the virus and noting the dilution at which half the replicate dishes become infected.

Several key variables define what is ultimately the LOD for a particular assay system because individual assay performance characteristics vary. Although LOD can be determined by calculating the point at which a signal can be discriminated from the background, it is more often determined empirically by testing serial dilutions of target analyte such as viral stocks.

In addition to the physicochemical factors that impact assay performance, some assay technologies involve direct specimen testing but most are methods that have a preanalytical specimen extraction step. The direct specimen assay systems include the Diasorin Simplexa Covid-19 Direct Assay on the Liaison MDX and the Atila Biosystems (Mountain View, CA) iAmp COVID-19 Detection Kit, which is run on the Bio-Rad CFX Real-Time System using an isothermal amplification method.

Among these assay systems, there is a variation in the input specimen volume for a given SARS-CoV-2 molecular assay that also affects what is ultimately determined to be the LOD for the assay.

Gene Targets

As shown in Table 1, assay designs vary across the assays and instruments on which they have been validated. The SARS-CoV-2 RNA transcript encodes multiple genes such as the replicase complex ORF1ab, the spike protein, the viral envelope, the membrane, and the nucleocapsid proteins. Although gene targets are selected for their SARS-CoV-2 specificity, the assay designs have included 1 or more gene sequences among the ORf1a/b, nucleocapsid, spike protein, envelope, membrane, and RNA-dependent RNA polymerase genes.

The specificity of gene targets for SARS-CoV-2 is a critical part of assay design. This specificity is necessary for SAR-CoV-2 detection and discrimination among other related SARS viruses. Coronaviruses have a significantly lower mutation rate than those observed in other RNA viruses, such as the influenza virus, because of the presence of intrinsic proofreading activity during viral replication.28-33 It is possible that successful SARS-CoV-2 replication cycles may accumulate mutations that could ultimately contribute to differences in clinical outcomes between patients with differing viral populations. Gussow et al32 and Artesi et al34 identified a C-to-U transition at position 26,340 of the SARS-CoV-2 genome that spans the envelope gene. Moreover, they stressed the importance of targeting 2 genomic regions to minimize the chance for false negative results. It is thus prudent to monitor circulating SARS-CoV-2 variants that may adversely impact the performance of RT-PCR diagnostic assays.35

Assay Protocol

Currently, the assays with EUA are all qualitative, providing a binary positive or negative answer (or a third answer as presumptive positive or indeterminate in a few assay systems). Most RT-PCR assay systems are set up for a CT value <40 to be clinically reported as PCR-positive. However, this situation does not always occur because some assay systems continue amplification cycles beyond 40. It is likely that a patient test result that may be positive has a late CT (eg, >40 CT) and may remain undetected on a different assay platform. Nevertheless, a PCR-positive result indicates the presence of viral RNA and may not necessarily indicate the presence of viable virus. Thus, it is difficult to correlate the CT values from one assay system to another in an accurate manner.36,37

It is assumed during PCR that each template molecule is duplicated once per cycle. However, for templates expressed at very low levels variance exists in the results from PCR reactions with scattered CT values between replicate PCR reactions.38 As can be envisaged for low copy number templates, comparison across different assay designs, instruments, target genes, and PCR physicochemical characteristics is difficult.

Specimen Types

The SARS-CoV-2 molecular tests discussed are performed after sampling using NP swabs or other upper respiratory tract specimens, including OP swabs or saliva. Aside from the diversity of assay designs, target genes, molecular methodologies, and platforms used to detect SARS-CoV-2, specific specimen types, which include NP and OP swabs, BAL, and sputum impact the LOD of an assay. In addition, alternate specimen types such as saliva and cotton swabs have also been validated for specific assays and intended uses, as has the use of saline as an alternative transport medium vs conventional viral transport medium. Limited allocations of swabs (and reagents) have compelled manufacturers (and clinical laboratories) to validate alternate swabs and transport to be used for patient specimen collection and testing.39

The timeline of testing for PCR positivity may vary with the type of specimen. However, BAL fluid specimens seem to show the highest positive rates, followed by sputum, nasal swabs, fibrobronchoscope brush biopsy, pharyngeal swabs, feces, and blood.40 False-negative results may occur because of inappropriate timing of specimen collection in relation to illness onset and deficiency in sampling technique, especially of NP swabs. Research has shown that PCR positivity declines more slowly in sputum and may still be positive after NP swabs are negative. The persistence of PCR in sputum and stool has been found to be similar as assessed.13

International Standards

There is a dire need for the development of international standards for a SARS-CoV-2 RNA copy number. Development of such standards would permit comparison of the myriad assay systems such that the assay LOD can be reported in a standardized manner. This would be very useful for clinical practitioners and laboratory and other healthcare professionals.

To support its evaluation of diagnostic tests for COVID-19, the FDA has announced that it will provide a SARS-CoV-2 reference panel to ensure the quality of the tests, validation of new assays, test calibration, and monitoring of assay performance. The FDA panel will be available to commercial and laboratory developers who are interacting with the FDA through the pre-EUA process.

The Coronavirus Standards Working Group, bringing together commercial, academic, and public-sector interests, has been formed to develop tools that ensure the reliability and accuracy of this vital COVID-19 testing.41 As stated on its website, this working group will develop an annotated inventory of resources, identify gaps, and conduct international collaborative experiments to evaluate and establish comparability of the diverse control materials and test methods available. A commendable objective is to pool knowledge to collaboratively produce the most-needed reference materials, reference samples, and reference measurement methods to underpin an enduring and reliable COVID testing enterprise.

Bustin and Nolan42 stated in 2020 that current RT-PCR testing for SARS-CoV-2 has several shortcomings because of the shortage in the availability of suitable assays, reagents, equipment, and testing capacity. Interestingly, they recommended centralized RNA extraction procedures in containment level 3 facilities with dissemination of analytical testing of noninfectious RNA to nationwide laboratories. They suggested that these procedures be considered by public health organizations such as the CDC and Public Health England in the United Kingdom for protocols for emergency testing systems that can be rapidly adopted in emergency situations.

Conclusion

Worldwide multifarious assay systems are currently in use to test for SARS-CoV-2. To effectively implement clinical countermeasures, it is imperative to recognize the clinical decision-making that may be influenced by variation in the performance characteristics of the diverse assay systems in use for SARS-CoV-2 diagnostic testing. The assay results, particularly those that fall near the LOD, may not be commutable from one platform to another. As discussed herein, efforts to standardize across the different assays, platforms, and technologies would enhance clinical diagnostic testing to the ultimate benefit of patients. It is anticipated that cross-platform comparison studies will continue to add to our knowledge base. Simultaneous ongoing efforts to learn more about SARS-CoV-2 pathogenesis would lead to improved diagnostic testing such that successful preventive, diagnostic, and prognostic countermeasures may be implemented for controlling the ongoing pandemic. LM

Glossary

Abbreviations

EUA

emergency use authorization

LOD

limit of detection

FDA

U.S. Food & Drug Administration

IVD

in vitro diagnostics

CT

cycle threshold

CI

confidence interval

CDC

Centers for Disease Control and Prevention

NP

nasopharyngeal

RT-PCR

reverse-transcription polymerase chain reaction

OP

oropharyngeal

TMA

transcription-mediated amplification

POC

point of care

NGS

next-generation sequencing

LDT

laboratory-developed test

IFU

instructions for use

TCID50

median tissue culture infectious dose

BAL

bronchoalveolar lavage

References

  • 1. https://www.fda.gov/medical-devices/coronavirus-disease-2019-covid-19-emergency-use-authorizations-medical-devices/vitro-diagnostics-euas#individual-molecular. Individual EUAs for molecular diagnostic tests for SARS-CoV-2. U.S Food & Drug Administration. Last updated November 6, 2020. Accessed November 13, 2020.
  • 2. Cradic K, Lockhart M, Ozbolt P, et al.  Clinical evaluation and utilization of multiple molecular in vitro diagnostic assays for the detection of SARS-CoV-2. Am J Clin Pathol.  2020;154(2):201–207. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3. Lieberman JA, Pepper G, Naccache SN, et al.  Comparison of commercially available and laboratory developed assays for in vitro detection of SARS-CoV-2 in clinical laboratories. J Clin Microbiol.  2020;58:e00821-20. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4. Tan C, Xiao Y, Meng X, et al.  Asymptomatic SARS-CoV-2 infections: What do we need to know?  Infect Control Hosp Epidemiol. Published online ahead of print May 6, 2020. doi: 10.1017/ice.2020.201. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5. https://www.cdc.gov/coronavirus/2019-ncov/hcp/testing-overview.html. Overview of testing for SARS-CoV-2 (COVID-19). Centers for Disease Control and Prevention. Updated October 21, 2020. Accessed November 13, 2020.
  • 6. Zou L, Ruan F, Huang M, et al.  SARS-CoV-2 viral load in upper respiratory specimens of infected patients. N Engl J Med.  2020;382:12. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7. He X, Lau EHY, Wu P, et al.  Temporal dynamics in viral shedding and transmissibility of COVID-19. Nat Med.  2020;26(9):672–675. [DOI] [PubMed] [Google Scholar]
  • 8. Sun J, Xiao J, Sun R, et al.  Prolonged persistence of SARS-CoV-2 RNA in body fluids. Emerg Infect Dis.  2020;26(8):1834–1838. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9. van Kampen JJA, van de Vijver DAMC, Fraaij PLA, et al.  Shedding of infectious virus in hospitalized patients with coronavirus disease-2019 (COVID-19): duration and key determinants. Preprint. Posted online June 9, 2020. medRxiv. doi: 10.1101/2020.06.08.20125310. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10. Lan L, Xu D, Ye G, et al.  Positive RT-PCR test results in patients recovered from COVID-19. JAMA. 2020;323(15):1502–1503. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11. Chen Y, Chen L, Deng Q, et al.  The presence of SARS-CoV-2 RNA in the feces of COVID-19 patients. J Med Virol.  2020;92(7):833–840. [DOI] [PubMed] [Google Scholar]
  • 12. Gupta S, Parker J, Smits S, et al.  Persistent viral shedding of SARS-CoV-2 in faeces—a rapid review. Colorectal Dis.  2020;22(6):611–620. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13. Wölfel R, Corman VM, Guggemos W, et al.  Virological assessment of hospitalized patients with COVID-2019. Nature.  2020;581(7809):465–469. [DOI] [PubMed] [Google Scholar]
  • 14. Bullard J, Durst K, Funk D, et al.  Predicting infectious SARS-CoV-2 from diagnostic samples. Clin Infect Dis. Published online May 22, 2020. doi: 10.1093/cid/ciaa638. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15. Huang C-G, Lee K-M, Hsiao M-J, et al.  Culture-based virus isolation to evaluate potential infectivity of clinical specimens tested for COVID-19. J Clin Microbiol.  2020; 58:e01068-20. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 16. https://www.who.int/publications/i/item/who-2019-nCoV-surveillanceguidance-2020.7. Public health surveillance for COVID-19: interim guidance. World Health Organization. Updated August 7, 2020. Accessed November 13, 2020.
  • 17. Lv DF, Ying QM, Weng YS, et al.  Dynamic change process of target genes by RT-PCR testing of SARS-Cov-2 during the course of a coronavirus disease 2019 patient. Clin Chim Acta.  2020;506:172–175. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18. Kucirka L, Lauer S, Laeyendecker O, Boon D. Variation in false negative rate of RT-PCR based SARS-CoV-2 tests by time since exposure. Ann Intern Med.  2020;173(4):262–267. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19. Zimmermann P, Curtis N. Coronavirus infections in children including COVID-19. An overview of the epidemiology, clinical features, diagnosis, treatment and prevention options in children. Pediatric Infect Dis J.  2020;39(5):355–368. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 20. Carter LJ, Garner LV, Smoot JW, et al.  Assay techniques and test development for COVID-19 diagnosis. ACS Cent Sci.  2020;6(5):591–605. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 21. Gorzalski AJ, Tian H, Laverdure C, et al.  High-throughput transcription-mediated amplification on the Hologic Panther is a highly sensitive method of detection for SARS-CoV-2. J Clin Virol.  2020;129:104501. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 22. Zhen WR, Manji E, Smith GJ, Berry GJ. Comparison of four molecular in vitro diagnostic assays for the detection of SARS-CoV-2 in nasopharyngeal specimens. J Clin Microbiol. 2020;58(8):e00743-20. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 23. Zhen W, Smith E, Manji R, et al.  Clinical evaluation of three sample-to-answer platforms for the detection of SARS-CoV-2. J Clin Microbiol. 2020;58(8):e00783-20. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 24. Craney AR, Velu P, Satlin MJ, et al.  Comparison of two high-throughput reverse transcription-polymerase chain reaction systems for the detection of severe acute respiratory syndrome coronavirus 2. J Clin Microbiol.  2020;58(8):e00890-20. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 25. Rhoads DD, Cherian SS, Roman K, et al.  Comparison of Abbott ID Now, DiaSorin Simplexa, and CDC FDA EUA methods for the detection of SARS-CoV-2 from nasopharyngeal and nasal swabs from individuals diagnosed with COVID-19. J Clin Microbiol.  2020;58(8):e00760-20. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 26. Harrington A, Cox B, Snowden J, et al.  Comparison of Abbott ID Now and Abbott m2000 methods for the detection of SARS-CoV-2 from nasopharyngeal and nasal swabs from symptomatic patients. J Clin Microbiol.  2020;58(8):e00798-20. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 27. Bar-On YM, Flamholz A. SARS-CoV-2 (COVID-19) by the numbers. Elife  2020;9:e57309. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 28. Denison MR, Graham RL, Donaldson EF, Eckerle LD, Baric RS. Coronaviruses: an RNA proofreading machine regulates replication fidelity and diversity. RNA Biol.  2011;8(2):270–279. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 29. Fehr AR, Perlman S. Coronaviruses: an overview of their replication and pathogenesis. Coronaviruses  2020;1282:1–23. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 30. Sanjuán R, Nebot MR, Chirico N, Mansky LM, Belshaw R. Viral mutation rates. J Virol.  2010;84(19):9733–9748. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 31. Minskaia E, Hertzig T, Gorbalenya AE, et al.  Discovery of an RNA virus exoribonuclease that is critically involved in coronavirus RNA synthesis. PNAS  2006;103:5108–5113. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 32. Gussow AB, Auslander N, Faure G, Wolf YI, Zhang F, Koonin EV. Genomic determinants of pathogenicity in SARS-CoV-2 and other human coronaviruses. Proc Natl Acad Sci U S A.  2020;117(26):15193–15199. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 33.Cimons M. SARS-CoV-2 is mutating slowly, and that’s a good thing. https://hub.jhu.edu/2020/06/10/sars-cov-2-dna-suggests-single-vaccine-will-be-effective/. Accessed November 13, 2020.
  • 34. Artesi M, Bontems S, Göbbels P, et al.  A recurrent mutation at position 126340 of SARS-CoV-2 is associated with failure of the E-gene qRT-PCR utilized in a commercial dual-target diagnostic assay. J Clin Microbiol.  2020;58(10):e01598-20. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 35. Hu J, He C-L, Gao Q-Z, et al.  D614G mutation of SARS-CoV-2 spike protein enhances viral infectivity. Preprint. Posted online June 19,  2020. bioRxiv  178509. doi: 10.1101/2020.06.20.161323. [DOI] [Google Scholar]
  • 36. Sethuraman N, Jeremiah SS, Ryo A. Interpreting diagnostic tests for SARS-CoV-2. JAMA.  2020;323(22):2249–2251. [DOI] [PubMed] [Google Scholar]
  • 37. Nalla AK, Casto AM, Huang MW, et al.  Comparative performance of SARS-CoV-2 detection assays using seven different primer/probe sets and one assay kit. J Clin Microbiol.  2020;58(6):e00557-20. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 38. Bustin SA, Nolan T. Data analysis and interpretation. In: Bustin SA, ed. A-Z of Quantitative PCR. La Jolla, CA: IUL Press; 2004: 447–44 8. [Google Scholar]
  • 39. Rodino KG, Espy MG, Buckwalter SP, et al.  Evaluation of saline, phosphate buffered saline and minimum essential medium as potential alternatives to viral transport media for SARS-CoV-2 testing. J Clin Microbiol. 2020;58(6):e00590-20. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 40. Wang W, Xu Y, Gao R, et al.  Detection of SARS-CoV-2 in different types of clinical specimens. JAMA. 2020;323(18):1843–1844. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 41. https://jimb.stanford.edu/covid-19-standards. Coronavirus Standards Working Group: a COVID-19 diagnostic standards development partnership. The Joint Initiative for Metrology in Biology. Accessed November 13, 2020.
  • 42. Bustin SA, Nolan T. RT-qPCR testing of SARS-CoV-2: a primer. Int J Mol Sci.  2020;21:3004. [DOI] [PMC free article] [PubMed] [Google Scholar]

Articles from Laboratory Medicine are provided here courtesy of Oxford University Press

RESOURCES