Skip to main content
PLOS One logoLink to PLOS One
. 2021 Jan 11;16(1):e0245189. doi: 10.1371/journal.pone.0245189

HIV false positive screening serology due to sample contamination reduced by a dedicated sample and platform in a high prevalence environment

Michael A Linström 1,2,*, Wolfgang Preiser 1,2, Nokwazi N Nkosi 1,2, Helena W Vreede 3,4, Stephen N J Korsman 3,5, Annalise E Zemlin 1,6, Gert U van Zyl 1,2
Editor: Jean-Luc EPH Darlix7
PMCID: PMC7799780  PMID: 33428663

Abstract

Automated testing of HIV serology on clinical chemistry analysers has become common. High sample throughput, high HIV prevalence and instrument design could all contribute to sample cross-contamination by microscopic droplet carry-over from seropositive samples to seronegative samples resulting in false positive low-reactive results. Following installation of an automated shared platform at our public health laboratory, we noted an increase in low reactive and false positive results. Subsequently, we investigated HIV serology screening test results for a period of 21 months. Of 485 initially low positive or equivocal samples 411 (85%) tested negative when retested using an independently collected sample. As creatinine is commonly requested with HIV screening, we used it as a proxy for concomitant clinical chemistry testing, indicating that a sample had likely been tested on a shared high-throughput instrument. The contamination risk was stratified between samples passing the clinical chemistry module first versus samples bypassing it. The odds ratio for a false positive HIV serology result was 4.1 (95% CI: 1.69–9.97) when creatinine level was determined first, versus not, on the same sample, suggesting contamination on the chemistry analyser. We subsequently issued a notice to obtain dedicated samples for HIV serology and added a suffix to the specimen identifier which restricted testing to a dedicated instrument. Low positive and false positive rates were determined before and after these interventions. Based on measured rates in low positive samples we estimate that before the intervention, of 44 117 HIV screening serology samples, 753 (1.71%) were false positive, declining to 48 of 7 072 samples (0.68%) post-intervention (p<0.01). Our findings showed that automated high throughput shared diagnostic platforms are at risk of generating false-positive HIV test results, due to sample contamination and that measures are required to address this. Restricting HIV serology samples to a dedicated platform resolved this problem.

Introduction

Laboratory-based diagnosis of Human Immunodeficiency Virus (HIV) infection in adults relies on 4th generation serological testing. In high-prevalence settings like South Africa, this requires a single screening test followed by one or two confirmatory tests for samples with reactive screening results [1]. Although screening is mostly done by point of care testing using rapid lateral flow chromatographic assays, samples with discordant rapid test results or from hospitalised patients are referred for laboratory-based testing. Due to high testing volumes, the public sector Tygerberg Hospital (TBH) National Health Laboratory Service (NHLS) laboratory has implemented automated 4th generation HIV serology on a high-throughput platform. Platforms that allow the analysis of different diagnostic tests across disciplines are often preferred due to the perceived economy of scale and efficient utilisation of space and personnel. These solutions are nevertheless not without challenges.

In a high prevalence setting, prior research by Hardie et al. (2017) found that repeatedly processing known HIV negative samples through automated platforms increased the probability of the sample subsequently testing positive [2]. A similar platform and setting exist at TBH NHLS where the Roche® Modular Pre-Analytic System (MPA) is used to process samples received for Clinical Chemistry, Immunology and Virology. Once samples are loaded, processing commences automatically, which entails centrifugation of the sample, sample aspiration into aliquots, directing samples to the appropriate testing module, and eventually sample storage.

TBH NHLS utilises an HIV serology screening assay which claims a 100% sensitivity and a 99.81% specificity for routine diagnostic samples [3]. The assay is said to detect p24 antigen on average five days after detectable viraemia and low-level HIV antibodies early after seroconversion, making it valuable for the early diagnosis of HIV in high incidence settings [4]. The high analytical sensitivity however poses a challenge for interpreting low positive (LP) results that may occur due to cross contamination of negative samples. Considering the high concentration of HIV antibodies and antigen in some patients, carry-over of microscopically small droplets may be sufficient to cause false positive (FP) results as HIV-positive patient samples remain positive on this platform even if diluted up to 10−5 [5]. Such FP results are determined by negative retesting of an independently obtained subsequent sample from a patient. Prompted by published evidence and investigation of individual cases that were FP, we suspected that sample contamination associated with pre-analytical processing on the MPA and testing on a high-throughput clinical chemistry module may be responsible for a large proportion of FP results.

Our study aims were

  1. To assess the factors that were associated with low positive and equivocal (LP&E) and FP HIV serology testing.

  2. To investigate the impact of the dedicated sample on the overall prevalence of LP&E and FP results.

Methods

Ethics approval was obtained from the Stellenbosch University’s Health Research Ethics Committee (HREC) # N19/04/044. The data was fully anonymised and de-identified before analysis. The HREC waived the requirement for informed consent on the grounds of the research involved retrospective electronic laboratory test data and records.

Study population

TBH NHLS Virology processes samples directly received from clinics and hospitals and referred from other laboratories in the drainage area which includes parts of the Cape Town Metropole, the Western Cape and the Northern Cape provinces.

Samples included were those sent to TBH NHLS for HIV screening serology (HIV-S) between 1 September 2017 (when the MPA was installed) and 31 May 2019. As creatinine (CRT) testing is a common clinical chemistry test it was used as proxy for a sample that had concurrent clinical chemistry testing and was therefore at risk of pre-analytical or clinical chemistry module-associated carry-over contamination.

Laboratory tests

The Roche® Cobas® 6000 E601 analyser with the Roche® Elecsys® HIV combi PT fourth generation assay was used as the screening and the Abbott® ARCHITECT® HIV Ag/Ab Combo® as the confirmatory test for HIV serology. Both are 4th generation HIV serology assays, combining antibody with antigen detection. Both tests are sandwich microparticle immunoassays, where Abbott® uses chemiluminescence and Roche® uses electrochemiluminescence.

The Roche® Cobas AmpliPrep®/Cobas TaqMan® HIV-1 Qualitative Test version 2 (qualitative total nucleic acid PCR) and the Roche® Cobas AmpliPrep®/Cobas TaqMan® HIV-1 Quantitative Test version 2 (quantitative RNA PCR) were used as tie-breaker assays to determine infection status for patients with indeterminate HIV serology using a subsequent fresh new sample. These assays employ the principle of reverse transcriptase real-time polymerase chain reaction (PCR) to detect HIV nucleic acids.

Dedicated sample implementation

Subsequent to our realization that samples may become cross-contaminated on the high throughput chemistry module we attempted to assure that samples for HIV-1 serology are tested on the virology module only. This intervention included a notice to clinicians to send a separate blood sample for HIV serology and once received in the laboratory appending a “Q” suffix to all HIV-1 serology sample labels, which prevents these samples from being recognized and tested on any other module than the dedicated virology module. We were able to extract data and compare rates before and after this intervention.

Data retrieval and analysis

As molecular testing is often required in the work-up of discordant serology, the proportion of LP&E results that had a fresh independently collected sample for subsequent serological or molecular HIV test was assessed. The combined outcome of the subsequent independently collected sample’s test results was then classified as negative, positive, and indeterminate. The median time between the original and subsequent fresh independently collected test result was used to indicate clinician response to LP&E results.

When a subsequent independently collected serology test or molecular test gave a negative result, initial results were classified as FP. When a subsequent independently collected follow-up serology result was high positive and confirmed on the confirmatory assay or the molecular test result was positive, the sample was classified as true positive. If subsequent testing produced an LP&E result like the original, with no additional molecular or serology testing results, samples were classified as indeterminate.

The FP rate was defined as the percentage of LP&E samples with a negative definitive result, as clinicians are required to send independently collected follow-up samples if an LP&E result is generated. It is expected that not all LP&E results have testing of subsequent independently collected follow-up samples due to the reliance on clinicians to provide such samples. Therefore, the false positive rate was extrapolated based on the LP&E results with known subsequent independently collected follow sample testing results. This rate was then used to calculate the prevalence of true positives, the true positive predictive value (PPV) and the specificity. The PPV calculated from the prevalence and the manufacturer’s claimed specificity was used as comparison.

The outcome of confirmatory testing was described as overall proportion of FP results that generated a discordant confirmatory test result on a separate platform, bearing in mind that the sample could have been contaminated before confirmatory testing.

Results for HIV serology including HIV-S and HIV confirmatory (HIV-C) testing and subsequent tests linked to the same individual were retrieved from the NHLS TrakCare laboratory information system. As a single sample identifier (ID) is generated when different sample tubes linked to the same request form are registered, the same sample-ID for different tests indicates concurrent testing. However, it does not necessarily indicate testing of the same tube on a different instrument. Therefore, in order to identify samples at risk of contamination, we first used CRT registered on the same sample ID of HIV-S samples as a screen for being at risk of clinical chemistry module processing and contamination. Second, we used samples being tested on the clinical chemistry module first, rather than second to further stratify contamination risk. We could be confident that samples with CRT results released after HIV-S results did not pass through the clinical chemistry module first and would not have been a source of contamination risk.

LP&E and FP results were stratified into samples with concurrent CRT testing vs. those without, CRT results released first (CRT1st) vs. CRT results released second (CRT2nd), and before and after the dedicated sample intervention. These comparisons included prevalences and the calculation of odds ratios (OR) to determine the risk of generating a LP&E and FP result within these strata. Fischer’s Exact Test was used to assess whether there were significant associations with LP&E or FP results, respectively before and after dedicated sample intervention, whereas the Mantel-Haenszel test was used to assess whether these associations were independent of the dedicated sample intervention.

Results

A total of 51189 results of samples sent for HIV-S were included over the 21 month period as seen in Table 1. Of these, 41156 (80.4%) were resulted as negative, 9085 (17.7%) as positive and 946 (1.8%) as LP&E. Table 1 separates these values into totals found before and after the dedicated sample intervention and illustrates the proportional changes seen.

Table 1. Summary of Low Positive and Equivocal (LP&E) HIV antibody rates in samples tested before and after introduction of a dedicated sample tube for HIV serology.

Before dedicated sample implementation After dedicated sample implementation
  Number Proportion Number Proportion p-Value
Original Sample HIV-S Total 44117   7072    
Negative 35141 79.66% 6015 85.05% NA
Positive 8091 18.34% 994 14.06% NA
LP&E Total 883 2.00% 63 0.89% <0.001
HIV-S Only Total 26986   3481    
HIV-S Only LP&E 354 1.31% 42 1.21% 0.602
HIV-S/CRT Total 17130   3591    
HIV-S/CRT LP&E 529 3.09% 21 0.58% <0.001
CRT1st Total 15505   2563    
CRT1st LP&E 520 3.35% 21 0.82% <0.001
CRT2nd Total 1625   1028    
CRT2nd LP&E 9 0.52% 0 0.00% 0.202
Follow-up Sample LP&E Samples with a follow-up Test 449   36    
Final Diagnosis Positive 27 6.03% 4 11.11% 0.306
Final Diagnosis Negative* 386 86.16% 25 69.44% 0.014
Final Diagnosis Indeterminate 36 8.04% 7 19.44% 0.031

HIV-S: HIV Screening Serology Results, CRT: Creatinine Testing, LP&E: Low Positive and Equivocal Results, HIV-S Only Total: Total Number of HIV Screening Serology Results without Concurrent Creatinine Testing.

HIV-S Only LP&E: Total Number of Low Positive and Equivocal HIV Screening Serology Results without Concurrent Creatinine Testing HIV-S/CRT total: Total Number of HIV Screening Serology Result with Concurrent Creatinine Testing, HIV-S/CRT LP&E: Low Positive and Equivocal Results Generated by HIV Screening Serology with Concurrent Creatinine Testing CRT1st Total: Total number of HIV Screening Serology Results with Concurrent Creatinine Result Released Before HIV Result, CRT1st LP&E: Number of Low Positive and Equivocal HIV Screening Serology Results with Concurrent Creatinine Result Released Before HIV Result CRT2nd Total: Total number of HIV Screening Serology Results with Concurrent Creatinine Testing with Creatinine Result Released after HIV Result, CRT2nd LP&E: Number of Low Positive and Equivocal HIV Screening Serology Results with Concurrent Creatinine Testing with Creatinine Result Released after HIV Result.

*False positive samples.

A summary of LP&E and FP results stratified by concurrent CRT and CRT1st vs CRT2nd is provided in Table 1. The largest proportion of LP&E results were in the equivocal range: 553 (59%) of the 946 overall non-stratified LP&E results (LP&E Total). This was not significantly different before and after the dedicated sample intervention. Of the 946 LP&E results, 485 (51%) had an independently collected follow-up sample for serological or molecular testing (Table 1). Of these 485 subsequent independently collected follow-up sample results, 411 (85%) were negative (making the initial result FP), 31 (6%) were positive (making the initial result true positive), and 43 (9%) remained indeterminate. 179 of the 485 independently collected follow-up samples (37%) had molecular testing performed on the follow-up sample instead of serological testing. The subsequent independently collected follow-up sample was submitted a median of 5 [IQR 2–22] and maximum of 497 days after the initial sample. Over the study period, which included a series of quality improvement interventions (as mentioned previously), there was a declining trend in the prevalence of LP&E samples. The dedicated sample intervention brought about the most significant change in this rate (Fig 1). Similarly, the prevalence of FP results declined over the study period with the lowest levels reached after the dedicated sample intervention (Fig 2). Moreover, based on measured rates in low positive samples we estimate that before the intervention, of 44 117 HIV screening serology samples, 753 (1.71%) were false positive, declining to 48 of 7 072 samples (0.68%) post-intervention (p<0.01). Of the 235 FP results with concurrent CRT testing, 227 (96%) were CRT1st samples. The overall true positive proportion of the results, reflecting the positive prevalence in the tested population, was 17.9% (9228 of 51189).

Fig 1. Low positive and equivocal results monthly proportions displayed over the 21 month study period.

Fig 1

A downward trend in percentage of LP&E results was observed with serial introduction of measures to reduce sample contamination on the analyser.

Fig 2. False positive proportions in groups with the last group being after the dedicated sample intervention.

Fig 2

Bar chart shows the percentage of LP&E results over 4 month intervals from September 2017 to May 2019.

Before the dedicated sample intervention, the odds ratio (OR) for obtaining a LP&E HIV-S result vs a negative result was 2.40 (95% CI: 2.09–2.75) in samples with concurrent CRT. Furthermore, the OR for producing FP vs. negative results in samples exposed to concurrent CRT was 2.20 (95% CI: 1.79–2.69). Following the dedicated sample intervention, these odds ratios declined to 0.48 (95% CI: 0.26–0.85) for LP&E and 0.89 (95%CI: 0.41–2.0) for FP results, respectively. This association of LP&E HIV-S results with concurrent CRT decreased significantly as a result of the dedicated sample intervention (p<0.001). In addition, before the dedicated sample intervention in samples with CRT1st vs CRT2nd, the OR was 7.04 (95% CI: 3.33–14.87) for LP&E HIV-S results and 4.1 (95% CI: 1.69–9.97) for FP HIV-S result. Taken together this provides strong support that this intervention reduced or prevented contamination.

The manufacturer’s assay estimated PPV using this study’s seroprevalence is 99.1% whereas the actual PPV in our setting was 91.9%. The manufacturer’s claimed specificity was 99.81% (4), compared to the actual specificity of 98.1%. Of 411 results shown to be FP (by testing a different sample or by testing on a different platform), 123 (30%) had same sample concurrent confirmatory serology testing on a separate platform. Of these only 52 (42%) had discordant results indicating that confirmatory testing on the same sample would not identify the majority of false positives.

Positive sample signals showed a bimodal distribution as indicated in Fig 3. The majority of positive results are expected above a S/CO of 200 but the first peak in the low signal range is not expected from an assay with a high reported specificity.

Fig 3. The frequency of HIV-S results’ signal distribution over the positive range.

Fig 3

The chart depicts the bimodal distribution of sample signals peaking in the low positive and equivocal as well as the high positive range.

Discussion

The demand on clinical laboratories to produce hundreds to thousands of results every hour has led to the utilisation of automated high throughput analysers. As evidenced in our study, this can increase the risk of sample cross-contamination. A sample had a significant risk of producing a FP result if previously exposed to the clinical chemistry (CRT) analyser (OR of 2.20 95% CI: 1.79–2.69). The most significant association pointing to sample contamination was the increased risk seen in samples with CRT performed before HIV-S. Samples had an OR of 7.04 (95% CI: 3.33–14.87) for testing LP&E vs negative when comparing CRT1st vs. CRT2nd. If CRT was tested second it could not have been contaminated by the clinical chemistry analyser before being tested for HIV. The success of the “dedicated sample implementation” intervention is demonstrated by the subsequent reduction of LP&E results by 55% and FP results by 60%. A significant reduction of 82% in the LP&E results with concurrent CRT testing after the dedicated sample intervention was observed. Of the LP&E samples that had subsequent tests, 85% were negative, supporting that the large majority of LP&E results are FP. Moreover, same sample confirmatory testing on a different platform averted only 42% of false positives.

The high FP rate, especially before the interventions, leading to a much lower PPV than predicted by the manufacturer's specificity claim, are concerning. Only 51% of clinicians responded by providing a follow-up sample in response to the comment issued with LP&E results; even though most of these responded early (after a median of 5 days), most FP cases were not rectified.

FP results usually generate LP&E S/CO signals as seen by other centres [2, 6, 7]. Using a combination of independent sample testing and different platform testing, for confirmation, we found that 85% of LP&E results were FP and therefore, LP&E result prevalence can be used as an indicator of FP rate. In other settings, the PPV for this assay approached 100% when using a higher positive threshold than recommended by the manufacturer (S/CO over 30) in European and Asian populations [79]. Given the high proportion of FP results among LP&E (S/CO <30) in our setting, the use of a higher threshold for positivity helped to identify cases likely to be false positive in our context, whereas the high proportion that tested negative when testing an independent sample highlights the importance of confirming HIV serology on a separately collected sample. Blaich et al (2017) and Uetwille-Geiger et al (2018) both found a lower than expected specificity of this assay [8, 10]. Patient factors were the major factor associated with FP results in their research. However, in our context instrument sample contamination was the primary cause.

A large sample size (n = 51189) over an extended period of 21 months allowed us to estimate rates and odds ratios with high precision. However, extrapolation of the FP proportion of samples with no subsequent independently collected follow-up testing was limited by uncertainty, minimal patient information and the possibility that patients may be elite controllers or on antiretroviral therapy. Clinicians may have relied on rapid point of care tests at clinics to repeat HIV testing on patients with LP&E results instead of laboratory testing. However this approach is not without risk as the performance of HIV rapid immunochromatographic tests is operator dependent [11, 12]. Rarely, FP results have a high S/CO [13]. However, our focus was to reduce on-platform contamination, which usually results in low S/CO values.

An HIV serology screening assay with such a low PPV is likely to result in an inordinately high number of HIV uninfected patients being diagnosed as positive. Although subsequent testing of an independent sample would reduce this risk, a contamination rate of around 2% at the study start would have resulted in some patients randomly testing positive on the subsequent independently collected follow-up sample and therefore inappropriate ‘confirmed diagnosis’ and initiation of antiretroviral therapy in HIV uninfected individuals may have happened. The accuracy of HIV serological diagnosis was improved through implementation of multiple measures at TBH NHLS, which resulted in a gradual decline in FP and LP&E prevalence. It may however not be feasible to implement all these measures in all high burden HIV testing settings. The national and international HIV testing guidelines [1, 14] recommend same sample confirmatory testing on a different platform, and confirmation of reactivity on a follow-up sample. However, less than half of the FP results produced discordant confirmation results when the original sample was tested on a separate confirmatory assay as per guidelines. Therefore, if sample contamination occurs, confirmatory testing of the same sample has limited value. Our quality improvement intervention highlights the importance of a dedicated sample for HIV testing, which reduces the risk of contamination and which should be emphasised in national and international HIV programs.

In summary to reduce the risk of contamination and FP results we employed the following strategies:

  1. Setting-appropriate and conservative cut-offs: The manufacturer regards a signal/sample cut-off (S/CO) > 1 as positive (4). However, based on local verification and data from other sites, a S/CO from 0.9–3) is regarded as equivocal (E), 3–30 as low positive (LP) and ≥ 30 as positive (P) [6].

  2. Dedicated sample requirement for HIV testing: We issued a notice to clinicians requiring them to send separate blood samples when HIV testing is requested, to reduce the risk of contamination on pre-analytical instrumentation and high throughput chemistry analysers [15].

  3. Training of staff to ensure that samples for HIV serology are processed only on the dedicated Virology module.

  4. Dedicated sample labelling: To enforce dedicated sample testing for HIV serology, a suffix (referred to hereafter as “Q” suffix) is added to the barcode label on all dedicated HIV serology samples [16]. This suffix ensures that samples for HIV serology can only be processed on a dedicated virology module. Other modules (clinical chemistry, haematology, immunology, etc.) in the laboratory are unable to process a sample with a suffix on its label; isolating samples from suspected areas of cross-contamination risk.

  5. After requests for HIV serology are no longer permitted on samples already processed on the shared interdisciplinary module.

The fact that nearly 800 people included in this study interval had FP results and more than 500 people may still not know their true HIV status as there was no subsequent testing, despite having contact with health care, is a concern and action should be taken to prevent this.

Conclusions

This investigation highlights the risk of using shared platforms, which may be prone to sample contamination affecting HIV serological testing, and emphasizes the need for proper evaluation, environment adapted thresholds for positive vs. indeterminate (LP&E) serology, the importance of expert oversight and quality improvement to ensure that HIV serology is accurate. The implementation of a dedicated sample for HIV serology testing ultimately improved the overall validity of the test results. Further research to understand the properties and mechanisms of instrument-sample contamination to improve platform design and workflow may facilitate the prevention of false positive HIV serology. These instruments are also used to test other viral infections such as Hepatitis B. As shown in our study with regards to HIV, it is conceivable that micro-contamination between positive and negative Hepatitis B surface antigen samples could occur. This too warrants further investigation.

Supporting information

S1 Dataset. Dataset of all test results and corresponding values over the study interval.

(XLSX)

Data Availability

All relevant data are within the paper and its Supporting Information files.

Funding Statement

The author(s) received no specific funding for this work.

References

Decision Letter 0

Jean-Luc EPH Darlix

5 Nov 2020

PONE-D-20-03947

HIV false positive screening serology due to sample contamination reduced by a dedicated sample and platform in a high prevalence environment

PLOS ONE

Dear Dr. Linström,

Thank you for submitting your manuscript to PLOS ONE. After careful consideration, we feel that it has merit but does not fully meet PLOS ONE’s publication criteria as it currently stands. Therefore, we invite you to submit a revised version of the manuscript that addresses the points raised during the review process.

In accordance with reviewer's comments, this manuscript is really in need of clarification (the aim and the style). To that end an extensive english editing is required to render the manuscript readable by non-experts. In addition  sample contaminations are a real diagnostic issue and should be solved using simple means, but which ones? This is not clear as 'proposed' in the conclusion section  < Automated high throughput shared diagnostic platforms risk generating false-positive HIV test results, due to sample contamination. "Measures" are required to address this. Combining separate samples for HIV serology with sample labels restricting testing to dedicated modules successfully reduced false positive HIV rates>. The authors should list what should be the measures, please specify in detail the measures to avoid contamination.

Please submit your revised manuscript by Dec 06 2020 11:59PM. If you will need more time than this to complete your revisions, please reply to this message or contact the journal office at plosone@plos.org. When you're ready to submit your revision, log on to https://www.editorialmanager.com/pone/ and select the 'Submissions Needing Revision' folder to locate your manuscript file.

Please include the following items when submitting your revised manuscript:

  • A rebuttal letter that responds to each point raised by the academic editor and reviewer(s). You should upload this letter as a separate file labeled 'Response to Reviewers'.

  • A marked-up copy of your manuscript that highlights changes made to the original version. You should upload this as a separate file labeled 'Revised Manuscript with Track Changes'.

  • An unmarked version of your revised paper without tracked changes. You should upload this as a separate file labeled 'Manuscript'.

If you would like to make changes to your financial disclosure, please include your updated statement in your cover letter. Guidelines for resubmitting your figure files are available below the reviewer comments at the end of this letter.

If applicable, we recommend that you deposit your laboratory protocols in protocols.io to enhance the reproducibility of your results. Protocols.io assigns your protocol its own identifier (DOI) so that it can be cited independently in the future. For instructions see: http://journals.plos.org/plosone/s/submission-guidelines#loc-laboratory-protocols

We look forward to receiving your revised manuscript.

Kind regards,

Jean-Luc EPH Darlix, MG, Ph.D.

Academic Editor

PLOS ONE

Journal Requirements:

When submitting your revision, we need you to address these additional requirements.

1. Please ensure that your manuscript meets PLOS ONE's style requirements, including those for file naming. The PLOS ONE style templates can be found at

https://journals.plos.org/plosone/s/file?id=wjVg/PLOSOne_formatting_sample_main_body.pdf and

https://journals.plos.org/plosone/s/file?id=ba62/PLOSOne_formatting_sample_title_authors_affiliations.pdf

2. In the ethics statement in the manuscript and in the online submission form, please provide additional information about the patient records/samples used in your retrospective study.

Specifically, please ensure that you have discussed whether all data/samples were fully anonymized before you accessed them and/or whether the IRB or ethics committee waived the requirement for informed consent.

If patients provided informed written consent to have data/samples from their medical records used in research, please include this information.

[Note: HTML markup is below. Please do not edit.]

Reviewers' comments:

Reviewer's Responses to Questions

Comments to the Author

1. Is the manuscript technically sound, and do the data support the conclusions?

The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented.

Reviewer #1: Yes

**********

2. Has the statistical analysis been performed appropriately and rigorously?

Reviewer #1: Yes

**********

3. Have the authors made all data underlying the findings in their manuscript fully available?

The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified.

Reviewer #1: Yes

**********

4. Is the manuscript presented in an intelligible fashion and written in standard English?

PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here.

Reviewer #1: Yes

**********

5. Review Comments to the Author

Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters)

Reviewer #1: This paper provides good evidence for how false positive HIV serology results can be generated on automated analysers that perform a combination of chemistry, immunology and serology tests frequently used in clinical pathology laboratories. The key finding was the 4-fold increase in likelihood that a sample would test HIV positive if it had first passed through the chemistry module of the analyser (HIV serology tested after creatinine was tested). The authors also demonstrate the effectiveness of measures to reduce sample contamination, in particular use of a dedicated sample tube for HIV serology testing. Recognition of this problem is very important, especially where these instruments are used in areas of high HIV sero-prevalence. I recommend publication, but there are a number of aspects of the manuscript that require fixing. I found the write up of the results particularly difficult to follow

Abstract:

Line 24: Alternative lead in sentence: Automated testing of HIV serology on clinical chemistry analysers has become common.

Line 24: “high throughput, high HIV….”

Line 38: “repeat independent testing on follow up samples of 485…”

Lines 40-42: “The odds ratio for generating a false positive HIV serology result was 4.1 (95%CI:1.69-9.97) when creatinine was tested before HIV serology on the same sample…”

Introduction:

P3, line 51; full stop after “serological testing”

P3, line 80: delete “in order…recently”. Replace with “We issued a notice…”

P3 line 86: Dedicated sample labelling: “ to enforce dedicated sample testing for HIV serology a suffix (referred to hereafter as “Q”suffix) is added to the barcode sample label on all…”

I would prefer the authors to refer to the use of a dedicated sample tube, rather than “Q suffix” (which was the technical means by which this end was achieved) throughout the manuscript, but I don’t feel too strongly about it.

Remove references 11 and 12 – these are not documents available to journal readers.

Line 92: “After requests for HIV serology are not permitted on the…”

P4:

Study aims: (switch around)

1) To assess factors associated with LP&E serology

2) To investigate the impact of measures to ensure the use of a dedicated sample on the overall…”

Methods:

Under data retrieval and analysis:

After the first lead-in sentence, switch the order of paragraphs as follows

section from line 136 to 152 should come before lines 121-135.

Line 128: “…would not have been at risk of contamination from this source”

Line 140 : replace “these samples” with “initial results” and “negative” with “false positive”

Line 142: “the sample was classified as true positive”

Line 145: “It is expected that not all LP&E results have…”

Results:

158: “…before and after the dedicated sample tube intervention…”

162-175: this section is very hard to follow. Please rewrite focussing on the main findings you want to highlight.

I am confused by some of these results. In table 1 and in the text it is stated that a total of 3591 samples were concurrently tested for creatinine “after “Q suffix” implementation (in 2563 creatinine was tested first and in 1028 it was tested second). How can this be? The Q suffix was supposed to ensure that the sample was only tested for HIV. These numbers should all be “0”.

It seems to me that the important figures to compare are the rates of LP&E between the samples where creatinine was tested first, vs where it was tested 2nd . i.e. 3.32% vs 0.52%

Lines 182-191: again very difficult to follow.

Table 1

Title should read: “Summary of low positive and equivocal HIV antibody rates in samples tested before and after introduction of a dedicated sample tube for HIV serology testing.

The figures (numbers) for LP&E results: please amend or make it clear what the numbers in columns 3 and 4 (rows 7-12) refer to.

Middle section of the table:

Column 1 row one: LP&E samples with a repeat test (or positive samples with a repeat test)

Bottom section: (extrapolation data) adds no value. I would remove

Figure 1:

No legend is given. I would suggest “ A downward trend in percentage of LP&E samples was observed with serial introduction of measures to reduce sample contamination on the analyser.”

Figure 2:

Legend: Bar chart shows percentage of LP&E results over 4 month intervals from September 2017 to May 2019.

Discussion:

Rewrite the first paragraph with a lead in sentence contextualising the problem. Highlight the main findings. Don’t restate all the results in the first paragraph.

References:

The following references are not available to journal readers. Either provide a URL or remove from the list: 7, 8, 10, 11, 12,

Provide URL for 4 and 20

**********

6. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files.

If you choose “no”, your identity will remain anonymous but your review may still be made public.

Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy.

Reviewer #1: Yes: Diana Hardie

[NOTE: If reviewer comments were submitted as an attachment file, they will be attached to this email and accessible via the submission site. Please log into your account, locate the manuscript record, and check for the action link "View Attachments". If this link does not appear, there are no attachment files.]

While revising your submission, please upload your figure files to the Preflight Analysis and Conversion Engine (PACE) digital diagnostic tool, https://pacev2.apexcovantage.com/. PACE helps ensure that figures meet PLOS requirements. To use PACE, you must first register as a user. Registration is free. Then, login and navigate to the UPLOAD tab, where you will find detailed instructions on how to use the tool. If you encounter any issues or have any questions when using PACE, please email PLOS at figures@plos.org. Please note that Supporting Information files do not need this step.

PLoS One. 2021 Jan 11;16(1):e0245189. doi: 10.1371/journal.pone.0245189.r002

Author response to Decision Letter 0


7 Dec 2020

Thank you for considering our manuscript and taking time to review it. Herewith, find our response to the reviewer:

“This paper provides good evidence for how false positive HIV serology results can be generated on automated analysers that perform a combination of chemistry, immunology and serology tests frequently used in clinical pathology laboratories. The key finding was the 4-fold increase in likelihood that a sample would test HIV positive if it had first passed through the chemistry module of the analyser (HIV serology tested after creatinine was tested). The authors also demonstrate the effectiveness of measures to reduce sample contamination, in particular use of a dedicated sample tube for HIV serology testing. Recognition of this problem is very important, especially where these instruments are used in areas of high HIV sero-prevalence. I recommend publication, but there are a number of aspects of the manuscript that require fixing. I found the write up of the results particularly difficult to follow”

• We clarified many aspects of the results write up.

‘Abstract:

Line 24: Alternative lead in sentence: Automated testing of HIV serology on clinical chemistry analysers has become common.

Line 24: “high throughput, high HIV….”

Line 38: “repeat independent testing on follow up samples of 485…”

Lines 40-42: “The odds ratio for generating a false positive HIV serology result was 4.1 (95%CI:1.69-9.97) when creatinine was tested before HIV serology on the same sample…”’

• These changes have been made.

• “Automated testing of HIV serology on clinical chemistry analysers has become common. High sample throughput, high HIV prevalence and instrument design could all contribute to sample cross-contamination by microscopic droplet carry-over from seropositive samples to seronegative samples resulting in false positive low-reactive results.”

• “Of 485 initially low positive or equivocal samples 411 (85%) tested negative when retested using an independently collected sample.”

• “The odds ratio for a false positive HIV serology result was 4.1 (95% CI: 1.69-9.97) when creatinine level was determined first, versus not, on the same sample, suggesting contamination on the chemistry analyser.”

‘Introduction:

P3, line 51; full stop after “serological testing”

P3, line 80: delete “in order…recently”. Replace with “We issued a notice…”’

• These changes have been made.

• “Laboratory-based diagnosis of Human Immunodeficiency Virus (HIV) infection in adults relies on 4th generation serological testing.”

• “Dedicated sample requirement for HIV testing: We issued a notice to clinicians requiring them to send separate blood samples when HIV testing is requested, to reduce the risk of contamination on pre-analytical instrumentation and high throughput chemistry analysers[15].”

“P3 line 86: Dedicated sample labelling: “ to enforce dedicated sample testing for HIV serology a suffix (referred to hereafter as “Q”suffix) is added to the barcode sample label on all…”

I would prefer the authors to refer to the use of a dedicated sample tube, rather than “Q suffix” (which was the technical means by which this end was achieved) throughout the manuscript, but I don’t feel too strongly about it.

Remove references 11 and 12 – these are not documents available to journal readers.”

• We concur and have removed reference to “Q” suffix. We have replaced it with reference to a dedicated separate sample or dedicated sample intervention.

‘Line 92: “After requests for HIV serology are not permitted on the…”’

• These changes have been made.

• “After requests for HIV serology are no longer permitted on samples already processed on the shared interdisciplinary module.”

“P4:

Study aims: (switch around)

1) To assess factors associated with LP&E serology

2) To investigate the impact of measures to ensure the use of a dedicated sample on the overall…”’

• These changes have been made.

• “1) To assess the factors that were associated with low positive and equivocal (LP&E) and FP HIV serology testing.

2) To investigate the impact of the dedicated sample on the overall prevalence of LP&E and FP results.”

“Methods:

Under data retrieval and analysis:

After the first lead-in sentence, switch the order of paragraphs as follows

section from line 136 to 152 should come before lines 121-135.

Line 128: “…would not have been at risk of contamination from this source”

Line 140 : replace “these samples” with “initial results” and “negative” with “false positive”

Line 142: “the sample was classified as true positive”

Line 145: “It is expected that not all LP&E results have…”’

• These changes have been made.

• “As molecular testing is often required in the work-up of discordant serology, the proportion of LP&E results that had a fresh independently collected sample for subsequent serological or molecular HIV test was assessed. The combined outcome of the subsequent independently collected sample’s test results was then classified as negative, positive, and indeterminate. The median time between the original and subsequent fresh independently collected test result was used to indicate clinician response to LP&E results.

When a subsequent independently collected serology test or molecular test gave a negative result, initial results were classified as FP. When a subsequent independently collected follow-up serology result was high positive and confirmed on the confirmatory assay or the molecular test result was positive, the sample was classified as true positive. If subsequent testing produced an LP&E result like the original, with no additional molecular or serology testing results, samples were classified as indeterminate.

The FP rate was defined as the percentage of LP&E samples with a negative definitive result, as clinicians are required to send independently collected follow-up samples if an LP&E result is generated. It is expected that not all LP&E results have testing of subsequent independently collected follow-up samples due to the reliance on clinicians to provide such samples. Therefore, the false positive rate was extrapolated based on the LP&E results with known subsequent independently collected follow sample testing results. This rate was then used to calculate the prevalence of true positives, the true positive predictive value (PPV) and the specificity. The PPV calculated from the prevalence and the manufacturer’s claimed specificity was used as comparison.

The outcome of confirmatory testing was described as overall proportion of FP results that generated a discordant confirmatory test result on a separate platform, bearing in mind that the sample could have been contaminated before confirmatory testing.

Results for HIV serology including HIV-S and HIV confirmatory (HIV-C) testing and subsequent tests linked to the same individual were retrieved from the NHLS TrakCare laboratory information system. As a single sample identifier (ID) is generated when different sample tubes linked to the same request form are registered, the same sample-ID for different tests indicates concurrent testing. However, it does not necessarily indicate testing of the same tube on a different instrument. Therefore, in order to identify samples at risk of contamination, we first used CRT registered on the same sample ID of HIV-S samples as a screen for being at risk of clinical chemistry module processing and contamination. Second, we used samples being tested on the clinical chemistry module first, rather than second to further stratify contamination risk. We could be confident that samples with CRT results released after HIV-S results did not pass through the clinical chemistry module first and would not have been a source of contamination risk.

LP&E and FP results were stratified into samples with concurrent CRT testing vs. those without, CRT results released first (CRT1st) vs. CRT results released second (CRT2nd), and before and after the “Q” suffix implementation. These comparisons included prevalences and the calculation of odds ratios (OR) to determine the risk of generating a LP&E and FP result within these strata. Fischer’s Exact Test was used to assess whether there were significant associations with LP&E or FP results, respectively before and after dedicated sample intervention, whereas the Mantel-Haenszel test was used to assess whether these associations were independent of the dedicated sample intervention.”

‘Results:

158: “…before and after the dedicated sample tube intervention…”’

• These changes have been made.

• “Table 1 separates these values into totals found before and after the dedicated sample intervention and illustrates the proportional changes seen.”

“162-175: this section is very hard to follow. Please rewrite focussing on the main findings you want to highlight.

• We have made extensive changes to the article and have attempted to more clearly highlight the main aims of the study.

I am confused by some of these results. In table 1 and in the text it is stated that a total of 3591 samples were concurrently tested for creatinine “after “Q suffix” implementation (in 2563 creatinine was tested first and in 1028 it was tested second). How can this be? The Q suffix was supposed to ensure that the sample was only tested for HIV. These numbers should all be “0”.

• We have clarified that these samples had creatinine testing requested on them, but due to the intervention a dedicated separate sample was processed. These results illustrate, despite concurrent creatinine testing being requested, there was a significant decline in the number of false positive results after the intervention.

“It seems to me that the important figures to compare are the rates of LP&E between the samples where creatinine was tested first, vs where it was tested 2nd . i.e. 3.32% vs 0.52%”

• We agree and have emphasised these findings.

‘Lines 182-191: again very difficult to follow”’

• We have made significant changes to the text here.

“Table 1

Title should read: “Summary of low positive and equivocal HIV antibody rates in samples tested before and after introduction of a dedicated sample tube for HIV serology testing.

The figures (numbers) for LP&E results: please amend or make it clear what the numbers in columns 3 and 4 (rows 7-12) refer to.

Middle section of the table:

Column 1 row one: LP&E samples with a repeat test (or positive samples with a repeat test)

Bottom section: (extrapolation data) adds no value. I would remove

Figure 1:

No legend is given. I would suggest “ A downward trend in percentage of LP&E samples was observed with serial introduction of measures to reduce sample contamination on the analyser.”

Figure 2:

Legend: Bar chart shows percentage of LP&E results over 4 month intervals from September 2017 to May 2019.”

• Suggestions have been accepted and necessary changes to the table and text have been made.

• “Table 1. Summary of low positive and equivocal (LP&E) HIV antibody rates in samples tested before and after introduction of a dedicated sample tube for HIV serology.

HIV-S: HIV Screening Serology Results, CRT: Creatinine Testing, LP&E: Low Positive and Equivocal Results, HIV-S Only Total: Total Number of HIV Screening Serology Results without Concurrent Creatinine Testing

HIV-S Only LP&E: Total Number of Low Positive and Equivocal HIV Screening Serology Results without Concurrent Creatinine Testing HIV-S/CRT total: Total Number of HIV Screening Serology Result with Concurrent Creatinine Testing, HIV-S/CRT LP&E: Low Positive and Equivocal Results Generated by HIV Screening Serology with Concurrent Creatinine Testing CRT1st Total: Total number of HIV Screening Serology Results with Concurrent Creatinine Result Released Before HIV Result, CRT1st LP&E: Number of Low Positive and Equivocal HIV Screening Serology Results with Concurrent Creatinine Result Released Before HIV Result CRT2nd Total: Total number of HIV Screening Serology Results with Concurrent Creatinine Testing with Creatinine Result Released after HIV Result, CRT2nd LP&E: Number of Low Positive and Equivocal HIV Screening Serology Results with Concurrent Creatinine Testing with Creatinine Result Released after HIV Result

*False positive samples”

• “Fig1. Low positive and equivocal results monthly proportions displayed over the 21 month study period. A downward trend in percentage of LP&E results was observed with serial introduction of measures to reduce sample contamination on the analyser.”

• “Fig2. False positive proportions in groups with the last group being after the dedicated sample intervention. Bar chart shows the percentage of LP&E results over 4 month intervals from September 2017 to May 2019.

“Discussion:

Rewrite the first paragraph with a lead in sentence contextualising the problem. Highlight the main findings. Don’t restate all the results in the first paragraph.”

• We have rewritten the first paragraph and improved the flow of the text.

• “The demand on clinical laboratories to produce hundreds to thousands of results every hour has led to the utilisation of automated high throughput analysers. As evidenced in our study, this can increase the risk of sample cross-contamination. A sample had a significant risk of producing a FP result if previously exposed to the clinical chemistry (CRT) analyser (OR of 2.20 95% CI: 1.79-2.69). The most significant association pointing to sample contamination was the increased risk seen in samples with CRT performed before HIV-S. Samples had an OR of 7.04 (95% CI: 3.33-14.87) for testing LP&E vs negative when comparing CRT1st vs. CRT2nd. If CRT was tested second it could not have been contaminated by the clinical chemistry analyser before being tested for HIV. The success of the dedicated sample intervention is demonstrated by the subsequent reduction of LP&E results by 55% and FP results by 60%. A significant reduction of 82% in the LP&E results with concurrent CRT testing after the dedicated sample intervention was observed. Of the LP&E samples that had subsequent tests, 85% were negative, supporting that the large majority of LP&E results are FP. Moreover, same sample confirmatory testing on a different platform averted only 42% of false positives.”

“References:

The following references are not available to journal readers. Either provide a URL or remove from the list: 7, 8, 10, 11, 12,

Provide URL for 4 and 20”

• Changes were made.

Attachment

Submitted filename: Response to Reviewers.docx

Decision Letter 1

Jean-Luc EPH Darlix

28 Dec 2020

HIV false positive screening serology due to sample contamination reduced by a dedicated sample and platform in a high prevalence environment

PONE-D-20-03947R1

Dear Dr. Linström,

We’re pleased to inform you that your manuscript has been judged scientifically suitable for publication and will be formally accepted for publication once it meets all outstanding technical requirements.

Within one week, you’ll receive an e-mail detailing the required amendments. When these have been addressed, you’ll receive a formal acceptance letter and your manuscript will be scheduled for publication.

An invoice for payment will follow shortly after the formal acceptance. To ensure an efficient process, please log into Editorial Manager at http://www.editorialmanager.com/pone/, click the 'Update My Information' link at the top of the page, and double check that your user information is up-to-date. If you have any billing related questions, please contact our Author Billing department directly at authorbilling@plos.org.

If your institution or institutions have a press office, please notify them about your upcoming paper to help maximize its impact. If they’ll be preparing press materials, please inform our press team as soon as possible -- no later than 48 hours after receiving the formal acceptance. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information, please contact onepress@plos.org.

Kind regards,

Jean-Luc EPH Darlix, MG, Ph.D.

Academic Editor

PLOS ONE

Additional Editor Comments (optional):

Reviewers' comments:

Acceptance letter

Jean-Luc EPH Darlix

2 Jan 2021

PONE-D-20-03947R1

HIV false positive screening serology due to sample contamination reduced by a dedicated sample and platform in a high prevalence environment

Dear Dr. Linström:

I'm pleased to inform you that your manuscript has been deemed suitable for publication in PLOS ONE. Congratulations! Your manuscript is now with our production department.

If your institution or institutions have a press office, please let them know about your upcoming paper now to help maximize its impact. If they'll be preparing press materials, please inform our press team within the next 48 hours. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information please contact onepress@plos.org.

If we can help with anything else, please email us at plosone@plos.org.

Thank you for submitting your work to PLOS ONE and supporting open access.

Kind regards,

PLOS ONE Editorial Office Staff

on behalf of

Professor Jean-Luc EPH Darlix

Academic Editor

PLOS ONE

Associated Data

    This section collects any data citations, data availability statements, or supplementary materials included in this article.

    Supplementary Materials

    S1 Dataset. Dataset of all test results and corresponding values over the study interval.

    (XLSX)

    Attachment

    Submitted filename: Response to Reviewers.docx

    Data Availability Statement

    All relevant data are within the paper and its Supporting Information files.


    Articles from PLoS ONE are provided here courtesy of PLOS

    RESOURCES