Abstract
Multidetector row computed tomography (CT) scanners perform dynamic scanning and have a wide scan range. Time-resolved three-dimensional CT (i.e., 4D CT) has recently enabled visualization of flow in neurovascular vessels. We hypothesized that 4D CT technology would be a useful and non-invasive method for visualizing the flow dynamics of the portal circulation. The aim of this study was to evaluate the technical feasibility of 4D CT for visualizing flow dynamics in the portal circulation using 320-detector-row CT. 4D CT images of 18 consecutive patients with portal circulation including gastrorenal shunt were retrospectively evaluated for their ability to generate flow dynamics of the portal circulation. Flow dynamics could be visualized by 4D CT in 68 of the 72 vessels in the portal vein, splenic vein, superior mesenteric vein, and gastrorenal shunt. Flow direction could not be identified in four vessels, all of them being superior mesenteric veins. Flow direction was recognized on 4D CT in the 68 vessels of the portal circulation. A preliminary validation study revealed that flow direction of all 19 vessels in the portal circulation had concordance between 4D CT and color Doppler ultrasound. 4D CT could visualize flow dynamics of the portal circulation.
Supplementary Information
The online version contains supplementary material available at 10.1186/s41747-020-00197-8.
Keywords: Four-dimensional computed tomography, Hypertension (portal), Multidetector computed tomography, Portal vein, Portasystemic shunt
Key points
Four-dimensional computed tomography (4D CT) could visualize flow dynamics in the portal circulation.
Radiologists could recognize flow direction from the flow dynamics of the portal circulation on 4D CT.
Flow direction in the portal circulation demonstrated concordance between 4D CT and color Doppler ultrasound.
Background
Portal hypertension leads to serious complications, including bleeding from gastroesophageal varices, ascites, and portosystemic encephalopathy, in patients with hepatic cirrhosis. Hepatic venous pressure gradient is the gold standard method for the diagnosis of portal hypertension [1]; however, as it is measured by catheter, it is highly invasive. Endoscopic evaluation of varices is not only an indirect method of assessing portal hypertension, but it is also an invasive technique.
Noninvasive techniques proposed in recent decades for the measurement of portal hypertension include flow imaging using ultrasound (US) or magnetic resonance imaging (MRI) [1–6]. Doppler US and two-dimensional phase-contrast MRI are widely performed and well validated for measuring velocity and flow direction [7–9]. Doppler US is a simple technique used to evaluate portal hemodynamics without radiation exposure; however, this technique has relatively poor reproducibility and accuracy characterized by intraobserver and interobserver variation, and factors such as gas-filled bowels, obesity, and complex anatomy hinder observation of vessels in the portal circulation [3, 4, 10–12]. Although two-dimensional phase-contrast MRI can overcome these obstacles, elaborate preparation is necessary to define the plane in which all portal vessel flows are measured. Four-dimensional (4D) flow MRI is a recently developed technique that generates three-dimensional visualization of vessels, time-resolved flow images, and quantification of hemodynamics [6, 13]. Although several 4D flow studies have been conducted, multicenter validation has not yet been performed [13]. A disadvantage of 4D flow MRI imaging is the longer scan time required to obtain flow data and anatomical data compared with Doppler US.
Although conventional computed tomography (CT) technology cannot produce time-resolved flow images of the portal circulation in an efficient manner, CT studies are generally performed in patients with portal-hypertension-related diseases to determine the portal anatomy prior to interventional radiology (IR) procedures such as balloon-occluded retrograde transvenous obliteration, transjugular intrahepatic portosystemic shunt, and percutaneous transhepatic sclerotherapy. Despite the radiation burden to the patient, CT has significant advantages over US and MRI in terms of planning operative procedures, understanding the complicated anatomy of the portosystemic shunt, determining whether or not to catheterize, and deciding which vein from multiple portosystemic shunts should be embolized prior to therapy [14]. CT is also superior to US and MRI for detecting the tiny multiple drainage veins connected to the portosystemic shunt [15]. After IR treatment, CT is the modality of choice for evaluating thrombosis of varices or portosystemic shunts after embolization [16, 17].
Regarding portal-hypertension-related diseases, IR treatments raise serious post-therapeutic concerns that include the emergence of ascites, deterioration of other varices, portal thrombosis, and liver failure [16, 17]. Although each modality has drawbacks, recent US and 4D flow MRI studies have revealed the relationship between portal flow and the deterioration of gastroenteric varices and liver function [2, 18–20]. IR treatments can alter the flow dynamics of the portal circulation between pre- and post-therapy, resulting in changed liver function [18, 20]. If flow dynamic imaging could be performed by 4D flow CT in addition to routine CT scans, it would simplify the preoperative examination for patients and would indicate portosystemic shunts that should and should not be embolized. The addition of 4D flow CT images of the portal circulation to routine CT images could potentially improve assessment of the liver and could also be used to evaluate treatment efficacy.
Multidetector row CT scanners perform dynamic scanning and have a wide scan range. Time-resolved three-dimensional CT (i.e., 4D CT) has recently enabled visualization of flow in neurovascular vessels [21]. We hypothesized that 4D CT would be a useful and non-invasive method for visualizing the flow dynamics of the portal circulation. The aim of this study was to evaluate the technical feasibility of 4D CT with 320-detector row CT for visualizing flow dynamics in the portal circulation.
Methods
This retrospective study was approved by our Institutional Committee on Human Investigation. Informed consent for 4D CT in addition to routine CT imaging was obtained from all patients prior to 4D CT scanning. From March 2018 to September 2019, 66 patients underwent IR treatment for such as balloon-occluded retrograde transvenous obliteration and percutaneous transhepatic sclerotherapy at our department. After excluding 27 patients with diagnoses other than gastrorenal shunt, 39 patients with gastric varices and/or portosystemic encephalopathy including gastrorenal shunt met the criteria for 4D CT prior to IR treatment. Of these, the following were also excluded: 10 patients who were scanned less than 3 months prior to IR treatment, 6 who had a history of varix rupture, and 5 with renal failure. A final total of 18 consecutive patients (9 men, 9 women; mean age 65.3 years; age range 41–83 years; mean weight 61.7 kg, weight range 46–83 kg) who had undergone 4D CT before IR procedures were enrolled in the study.
4D CT protocol
4D CT was acquired by a 320-detector-row CT scanner (Aquilion one, Canon Medical Systems, Otawara, Japan) equipped with 320 × 0.5-mm-wide detector rows covering a 16-cm-long volume per rotation. A bolus of contrast medium was injected by an automatic injector via the cubital vein, followed by a saline flush. To visualize the aorta at the level of the celiac artery bifurcation, a bolus-tracking scan was performed after intravenous injection of 10 mL of contrast medium (Oypalomin 300, Fuji Pharma, Tokyo, Japan) into the cubital vein to confirm enhancement of the aorta and celiac artery. After the bolus-tracking scan, 600 mgI/kg of contrast medium was injected at a rate of 3.7–4.8 mL/s over 20 s, followed by 20 mL of saline, during 4D CT scanning. When contrast enhancement of the aorta at the level of the celiac artery bifurcation reached > 100 HU, patients were asked to inhale and hold their breath and 4D CT scanning was initiated. Scan parameters were as follows: detector collimation 320 × 0.5 mm; matrix 512 × 512; slice thickness 0.5 mm; tube voltage 120 kVp; mean tube current 58.8 mA (range, 40–80 mA); gantry rotation time 0.5 s; interval 1 s; number of imaging volumes 18; total scan time 26 s. Temporal resolution was 1.0 s, corresponding to the interval time. The current was determined based on twice the CT dose index for commonly performed helical CT examinations such as portal phase imaging. In this protocol, a series of 18 dynamic volumes was obtained, consisting of 320 slices covering the portal circulation. To make it easier for patients to hold their breath during scanning, they were given oxygen via a mask beforehand to increase their oxygen saturation [22].
After 4D CT scanning, portal, venous, and late phase imaging was performed from the diaphragm to the pelvis. Portal and venous phase acquisitions were initiated 35 and 55 s after the start of 4D scanning. The late phase scan was initiated 180 s after injection of contrast medium. Scan parameters were as follows: detector collimation 320 × 0.5 mm; matrix 512 × 512; table feed 65.04 mm/s; slice thickness 0.5 mm; tube voltage 120 kVp; mean tube current 392.8 mA (range, 277.5–562.5 mA); gantry rotation time 0.5 s; interval 0.5 mm.
Post-processing of 4D CT images
The 18 phases of the 4D CT images were transferred to a workstation (PhyZiodynamics; Ziosoft, Tokyo, Japan) for post-processing. This commercially available software creates smooth video images using a motion coherence function that fills in interphase motion with interpolated data. Using this software, we generated two additional phases between the original phases to create interphase motion, thus creating partially artificially constructed smooth and clear 4D motion [23]. To display only the portal circulation, the abdominal arteries were removed from the cine images, and only portal vessels with enhancement > 70 HU were displayed. Finally, a total of 52 phases were reconstructed into three-dimensional volume-rendering and maximum intensity projection images. Multiple-frame dynamic cine loops from the anterior and inferior aspects were generated for evaluation. A radiographer generated dynamic cine loops of the maximum intensity projection and volume-rendering images of all patients in multiple views, which took 3–5 h to create, depending on the experience of the radiographer.
Evaluation of 4D CT images
Two radiologists, each of them with more than 15 years of experience in diagnostic and interventional radiology, independently evaluated all 4D CT examinations. A 2-point scale (visualized or not visualized) was used to assess visualization of flow in the portal vein, splenic vein, superior mesenteric vein, and gastrorenal shunt. They watched videos of the flow dynamics in the portal vein, splenic vein, and superior mesenteric vein and assessed flow direction of these vessels using a 2-point scale (hepatopetal flow, toward the liver or hepatofugal flow, away from the liver) by consensus.
Radiation doses of CT scans
Radiation doses were recorded in all 18 cases for the 4D CT scans and all other scans.
Validation study with US
In the first 7 patients, color Doppler US was performed as a preliminary examination to assess the direction of hepatopetal or hepatofugal flow in the portal, splenic, and superior mesenteric veins, using a LOGIQ S7 Expert machine (GE Healthcare, Tokyo, Japan) equipped with a C1-5-D convex probe. US was performed with the patient in the supine position and was scheduled on the same day as 4D CT.
Statistical analysis
No statistical analysis was performed due to complete concordance observed in the validation study.
Results
Flow dynamics could be visualized by 4D CT in 68 of the 72 vessels (portal vein, splenic vein, superior mesenteric vein, and gastrorenal shunt) in the 18 patients. Flow direction could not be identified by 4D CT in four superior mesenteric veins. The following scans performed after 4D CT visualized superior mesenteric veins in the 4 cases. In the 68 vessels that could be visualized by 4D CT, both readers recognized flow direction from the flow dynamics of the portal circulation on 4D CT. All patients had hepatopetal flow in the portal vein (Figs. 1 and 2, Supplemental data 1, 2, 3, 4). The splenic vein had hepatofugal flow in 6 patients (Fig. 2, Supplemental data 3 and 4, Table 1) and the superior mesenteric vein had hepatofugal flow in 1 patient (Table 1).
Table 1.
Patient characteristic | 4D CT scan parameter | Portal phase parameter | Venous phase parameter | Late phase parameter | Total | Flow direction on CT | Flow direction on US | |||||||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Case number | Sex | Age (years) | Height (cm) | Weight (kg) | Voltage (kV) | Current (mA) | CTDI (mGy) | DLP(mGy·cm) | Effective dose(mSv) | CTDI (mGy) | DLP(mGy·cm) | Effective dose(mSv) | CTDI (mGy) | DLP(mGy·cm) | Effective dose(mSv) | CTDI (mGy) | DLP(mGy·cm) | Effective dose(mSv) | CTDI (mGy) | DLP(mGy·cm) | Effective dose(mSv) | Portal vein | Splenic vein | SMV | Gastrorenal shunt | Portal vein | Splenic vein | SMV |
1 | F | 80 | 151 | 58 | 120 | 70 | 33.8 | 540.0 | 8.1 | 15.1 | 787.9 | 11.8 | 15.1 | 787.9 | 11.8 | 15.9 | 709.7 | 10.6 | 79.9 | 2,825.5 | 42.3 | Pe | Fu | Pe | Fu | Pe | Fu | Pe |
2 | F | 47 | 150 | 83 | 120 | 80 | 38.5 | 615.6 | 9.2 | 19.9 | 1,079.6 | 16.1 | 20.1 | 716.9 | 10.7 | 20.1 | 716.9 | 10.7 | 98.6 | 3,129 | 46.7 | Pe | Fu | Pe | Fu | Pe | Fu | N |
3 | M | 64 | 172 | 72 | 120 | 70 | 33.8 | 540.0 | 8.1 | 15.6 | 971 | 14.5 | 15.3 | 524.3 | 7.8 | 16.4 | 560.6 | 8.4 | 81.1 | 2,595.9 | 38.8 | Pe | Pe | N | Fu | Pe | Pe | Fu |
4 | M | 65 | 161 | 64 | 120 | 60 | 28.9 | 462.6 | 6.9 | 15.4 | 820.3 | 12.3 | 14.6 | 527.6 | 7.9 | 15.4 | 820.3 | 12.3 | 74.3 | 2,630.8 | 39.4 | Pe | Pe | Pe | Fu | Pe | Pe | Pe |
5 | F | 69 | 151 | 70 | 120 | 70 | 33.8 | 540.0 | 8.1 | 17 | 1,056.5 | 15.8 | 17.7 | 700.8 | 10.5 | 17.1 | 979.5 | 14.6 | 85.6 | 3,276.8 | 49 | Pe | Pe | Fu | Fu | Pe | Pe | Fu |
6 | F | 87 | 144 | 51 | 120 | 50 | 24.1 | 385.2 | 5.7 | 10.1 | 522.5 | 7.8 | 10.1 | 351.8 | 5.2 | 10.8 | 478.3 | 7.1 | 55.1 | 1,737.8 | 25.8 | Pe | Pe | Pe | Fu | Pe | Pe | Pe |
7 | M | 63 | 183 | 61 | 120 | 50 | 24.1 | 385.2 | 5.7 | 12.3 | 823.2 | 12.3 | 12.9 | 537.3 | 8.0 | 12.8 | 775.6 | 11.6 | 62.1 | 2,521.3 | 37.6 | Pe | Fu | Pe | Fu | Pe | Fu | Pe |
8 | F | 61 | 151 | 57 | 120 | 70 | 33.8 | 540.0 | 8.1 | 15.6 | 837.9 | 12.5 | 15.6 | 564.4 | 8.4 | 15.8 | 769.8 | 11.5 | 80.8 | 2,712.1 | 40.5 | Pe | Fu | Pe | Fu | NA | NA | NA |
9 | F | 41 | 149 | 71 | 120 | 70 | 33.8 | 540.0 | 8.1 | 16.8 | 1,030.1 | 15.4 | 17.3 | 710.5 | 10.6 | 16.9 | 965.9 | 14.4 | 84.8 | 3,246.5 | 48.5 | Pe | Pe | N | Fu | NA | NA | NA |
10 | M | 78 | 161 | 53 | 120 | 50 | 24.1 | 385.2 | 5.7 | 11 | 655.0 | 9.8 | 10.7 | 429.3 | 6.4 | 11.7 | 569.2 | 8.5 | 57.5 | 2,038.7 | 30.4 | Pe | Pe | N | Fu | NA | NA | NA |
11 | F | 83 | 155 | 46 | 120 | 40 | 19.2 | 307.8 | 4.6 | 10.5 | 568.1 | 8.5 | 10.2 | 375.2 | 5.6 | 11.2 | 544.8 | 8.1 | 51.1 | 1,795.9 | 26.8 | Pe | Fu | Pe | Fu | NA | NA | NA |
12 | M | 56 | 173 | 61 | 120 | 60 | 28.9 | 462.6 | 6.9 | 14.5 | 943.8 | 14.1 | 16.1 | 662 | 9.9 | 14.6 | 1,094.3 | 16.4 | 74.1 | 3,162.7 | 47.3 | Pe | Pe | Pe | Fu | NA | NA | NA |
13 | M | 66 | 151 | 50 | 120 | 40 | 19.2 | 307.1 | 4.6 | 10.6 | 611.8 | 9.1 | 9.9 | 373.8 | 5.6 | 10.9 | 547.7 | 8.2 | 50.6 | 1,840.4 | 27.5 | Pe | Fu | Pe | Fu | NA | NA | NA |
14 | M | 58 | 177 | 67 | 120 | 60 | 28.9 | 462.6 | 6.9 | 13 | 901.2 | 13.5 | 12.6 | 545 | 8.1 | 13.1 | 1,092.1 | 16.3 | 67.6 | 3,000.9 | 44.8 | Pe | Pe | N | Fu | NA | NA | NA |
15 | M | 67 | 172 | 68 | 120 | 60 | 28.9 | 462.6 | 6.9 | 14.4 | 1,147.9 | 17.2 | 14.4 | 1,147.9 | 17.2 | 15.6 | 858.7 | 12.8 | 73.3 | 3,617.1 | 54.1 | Pe | Pe | Pe | Fu | NA | NA | NA |
16 | M | 55 | 163 | 65 | 120 | 60 | 28.9 | 462.6 | 6.9 | 14.5 | 878.8 | 13.1 | 14.2 | 536.4 | 8.0 | 15.2 | 795 | 11.9 | 72.8 | 2,672.8 | 39.9 | Pe | Pe | Pe | Fu | NA | NA | NA |
17 | F | 69 | 159 | 57 | 120 | 50 | 24.1 | 385.2 | 5.7 | 12.2 | 882.1 | 13.2 | 12 | 522.7 | 7.8 | 13.6 | 730 | 10.9 | 61.9 | 2,520 | 37.6 | Pe | Pe | Pe | Fu | NA | NA | NA |
18 | F | 67 | 167 | 56 | 120 | 60 | 28.9 | 462.6 | 6.9 | 11 | 937.8 | 14.0 | 11 | 1,039.1 | 15.5 | 11.5 | 884.6 | 13.2 | 62.4 | 3,324.1 | 49.6 | Pe | Pe | Pe | Fu | NA | NA | NA |
4D CT Four-dimensional computed tomography, CT Computed tomography, CTDI CT dose index, DLP Dose-length product, F Female, Fu Fepatofugal flow away from the liver, M Male, N Not depicted, NA Not applicable, Pe Hepatopetal flow into the liver, SMV Superior mesenteric vein, US Ultrasound
The mean CT dose index, dose-length product, and effective dose of all 4D CT scans were 28.6 mGy (range 19.2–38.5), 458.2 mGy × cm (range 307.1–615.6), and 6.9 mSv (range 4.6–9.2), respectively (Table 1). The doses for the entire CT examination are reported in Table 1.
In the preliminary validation study, color Doppler US detected 20/21 vessels in the first seven patients, and 4D CT in these patients also detected 20/21 vessels (one vessel was not depicted on 4D CT in one patient, and one vessel was not depicted on US in one patient) (Table 1). After excluding these two vessels, there was a concordance of flow direction between 4D CT and US in 19/19 vessels.
Discussion
This is the first study to visualize flow dynamics in the portal circulation on 4D CT with 320-detector-row computed tomography. As CT technology has evolved, the number of detectors has continued to increase, from 4-row CT scanners to the 320-detector-row CT scanners in current use. The most striking feature of multidetector CT is its broad scanning length in the z-axis. The 160-mm detector width enables acquisition of the portal circulation in a single rotation of the gantry and without couch movement. The established 4D flow MRI technique has a wider scan range and higher spatial resolution compared with our 4D-flow CT, in which the scan range is limited to 16 cm [13]. However, the advantages of the 4D CT method include short scan time, the ability to visualize the portal circulation at the same time as regular diagnostic imaging for a liver lesion or a gastric varix, and safety for patients with contraindication to MRI. The advantages of color Doppler US include the small size and portability of the US unit, non-invasiveness, and the short examination time with no radiation exposure. However, detection of portal vessels is dependent on operator skill and is made more difficult in the case of bowel gas and obesity [3, 4]. In the seven patients in the present preliminary study, one splenic vein could not be depicted on US due to obesity. The present results demonstrated that 4D CT might play a role as an additional imaging tool for understanding the portal circulation.
Breath-holding during 4D CT has important effects in terms of the patient’s oxygenation, portal flow dynamics, and detection of the portal circulation, but requires a high degree of patient cooperation. A previous report that compared the durations of prolonged breath-holding between preoxygenation and breathing room air [22] found that preoxygenation avoids hypoxemia and makes it easier for patients to hold their breath. Another study found that breath-holding after inspiration decreased portal blood flow [24]. In our study, we speculated that the Valsalva effect would lead to decreased portal flow during 4D CT scanning compared with breath-holding after expiration. As the breath-hold for 4D CT scanning should be as short as possible to reduce patient discomfort, we aimed for a scan time of less than 30 s. This is the reason why superior mesenteric venous flow was not detected in four of the present patients. The time difference in the arrival of the contrast medium happened between visualization of superior mesenteric venous flow and splenic venous flow. We prioritized visualization of splenic venous flow in order to visualize the short, retro, and left gastric veins as afferent feeders to a gastrorenal shunt. In addition, when the scan timing is adjusted to enable visualization of contrasted superior mesenteric venous flow, splenic venous flow cannot be seen because the splenic vein is already contrasted.
Previous studies have revealed changes in blood flow between pre-IR and post-IR treatment for gastrointestinal varices and hepatic encephalopathy using US and MRI [18, 20], increasing our understanding of changes in liver function due to changes in portal flow following IR therapies. In most patients of our study, the portal circulation had normal hepatopetal flow into the liver, except some patients with reversed hepatofugal flow in the splenic veins. These findings of flow direction are similar to the results of an US study of patients with gastric varices and hepatic encephalopathy [20]. US studies have also demonstrated that hepatofugal flow indicates the severity of the condition of the liver [2, 20]. Our preliminarily validation study revealed concordance of flow direction between 4D CT and US. By indicating flow direction, 4D CT may also predict the condition of the liver, although 4D CT provides limited detail regarding quantification of flow velocity and flow volume in the portal circulation compared with 4D MRI and US. Therefore, 4D flow CT imaging has potential as a diagnostic tool that could predict liver condition, emergence of complications, and therapeutic effects prior to commencement of IR treatment that can cause changes to dynamic flow in the portal circulation.
Compared with US and MRI, 4D CT is an emerging imaging method. As yet, the minimum radiation dose for 4D CT scanning is unknown, and therefore we are unable to compare the radiation exposure in our study and similar studies in terms of minimizing exposure. In this regard, we are able only to compare 4D CT scanning with conventional multiphasic CT as a reference. Effective radiation dose of 4D CT was lower than those of typical dynamic enhanced helical CT scans in this study. Helical scans usually covered whole abdominal cavity, whereas 4D CT scans covered only 16 cm. The radiation dose per unit length is higher for 4D CT than for typical helical CT [25]. However, the mean dose–length product in our series was 458.2 mGy × cm, which is similar to those reported previously for typical helical imaging of the upper abdomen with contrast medium (400–740 mGy × cm) [26, 27]. In further work, we will refine the examination protocol to keep radiation exposure as low as possible.
This study has some limitations. The first drawback is the limited breath-hold. Theoretically, it is possible to visualize superior mesenteric venous flow after visualization of splenic venous flow, but only if the patient can maintain a long enough breath-hold during 4D CT scanning. The second drawback is the small number of cases. The third drawback is lack of a control group consisting of healthy persons with normal portal circulation. The fourth drawback is that we did not evaluate 4D CT image quality, as there is no established imaging standard. Therefore, further investigations are required to enhance the validity of our results.
In conclusion, 4D CT could visualize flow dynamics in the portal circulation and might be a useful technique for capturing flow direction of the portal circulation.
Abbreviations
- 4D
Four-dimensional
- CT
Computed tomography
- IR
Interventional radiology
- MRI
Magnetic resonance imaging
- US
Ultrasound
Authors’ contributions
KK, AY, and AJ conceived and designed the study. KK, SI, DH, and AK acquired and processed the data. KK, AY, AJ, ES, and YM analyzed and interpreted the data. KK, AY, AJ, and SI drafted the manuscript. All authors critically revised the manuscript. All authors read and approved final manuscript.
Funding
This research did not receive any grant from public or commercial funding resources.
Availability of data and materials
The relevant data have been included in the manuscript. The datasets used and analyzed during the current study are available from the corresponding author on reasonable request.
Ethics approval and consent to participate
This retrospective study was approved by our Institutional Committee on Human Investigation (Reference number, 2019-027).
Consent for publication
Written informed consent was waived by the Institutional Review Board.
Competing interests
The authors declare that they have no competing interests.
Footnotes
Publisher’s Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
References
- 1.Thabut D, Moreau R, Lebrec D. Noninvasive assessment of portal hypertension in patients with cirrhosis. Hepatology. 2011;53:683–694. doi: 10.1002/hep.24129. [DOI] [PubMed] [Google Scholar]
- 2.Kondo T, Maruyama H, Sekimoto T, Shimada T, Takahashi M, Yokosuka O. Reversed portal flow: Clinical influence on the long-term outcomes in cirrhosis. World J Gastroenterol. 2015;21:8894–8902. doi: 10.3748/wjg.v21.i29.8894. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3.Baik SK. Haemodyanmic evaluation by doppler ultrasonography in patients with portal hypertension: a review. Liver Int. 2010;30:1403–1413. doi: 10.1111/j.1478-3231.2010.02326.x. [DOI] [PubMed] [Google Scholar]
- 4.Kim G, Cho YZ, Baik SK, Kim MY, Hong WK, Kwon SO (2015) The accuracy of ultrasonography for the evaluation of portal hypertension in patients with cirrhosis: a systemic review. Korean J Radiol 16:314–324. 10.3348/kjr.2015.16.2.314 [DOI] [PMC free article] [PubMed]
- 5.Stankovic Z, Rössle M, Euringer W, et al. Effect of TIPS placement on portal and splanchnic arterial blood flow in 4-dimensional flow MRI. Eur Radiol. 2015;25:2634–2640. doi: 10.1007/s00330-015-3663-x. [DOI] [PubMed] [Google Scholar]
- 6.Landgraf BR, Johnson KM, Roldán-Alzate A, Francois CJ, Wieben O, Reeder SB. Effect of temporal resolution on 4D flow MRI in the portal circulation. J Magn Reson Imaging. 2014;39:819–826. doi: 10.1002/jmri.24233. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 7.Gouya H, Vignaux O, Sogni P, et al. Chronic liver disease: systemic and splanchnic venous flow mapping with optimized cine phase-contrast MR imaging validated in a phantom model and prospectively evaluated in patients. Radiology. 2011;261:144–155. doi: 10.1148/radiol.11101541. [DOI] [PubMed] [Google Scholar]
- 8.Zhang L, Yin J, Duan Y, Yang Y, Yuan L, Cao T. Assessment of intrahepatic blood flow by Doppler ultrasonography: relationship between the hepatic vein, portal vein, hepatic artery and portal pressure measured intraoperatively in patients with portal hypertension. BMC Gastroenterol. 2011;11:84. doi: 10.1186/1471-230X-11-84. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9.Chawla Y, Santa N, Dhiman RK, Dilawari JB. Portal hemodynamics by duplex Doppler sonography in different grades of cirrhosis. Dig Dis Sci. 1998;43:354–357. doi: 10.1023/a:1018814624307. [DOI] [PubMed] [Google Scholar]
- 10.Finn JP, Kane RA, Edelman RR, et al. Imaging of the portal venous system in patients with cirrhosis: MR angiography vs duplex Doppler sonography. AJR Am J Roentgenol. 1993;161:989–994. doi: 10.2214/ajr.161.5.8273643. [DOI] [PubMed] [Google Scholar]
- 11.de Vries PJ, van Hattum J, Hoekstra JB, de Hooge P. Duplex Doppler measurements of portal venous flow in normal subjects. Inter- and intra-observer variability. J Hepatol. 1991;13:358–363. doi: 10.1016/0168-8278(91)90081-l. [DOI] [PubMed] [Google Scholar]
- 12.Lim AK, Patel N, Eckersley RJ, et al. Can Doppler sonography grade the severity of hepatitis C-related liver disease? AJR Am J Roentgenol. 2005;184:1848–1853. doi: 10.2214/ajr.184.6.01841848. [DOI] [PubMed] [Google Scholar]
- 13.Soulat G, McCarthy P, Markl M. 4D flow with MRI. Annu Rev Biomed Eng. 2020;22:103–126. doi: 10.1146/annurev-bioeng-100219-110055. [DOI] [PubMed] [Google Scholar]
- 14.Matsumoto A, Kitamoto M, Imamura M, et al. Three-dimensional portography using multislice helical CT is clinically useful for management of gastric fundic varices. AJR Am J Roentgenol. 2001;176:899–905. doi: 10.2214/ajr.176.4.1760899. [DOI] [PubMed] [Google Scholar]
- 15.Kageyama K, Nishida N, Matsui H, Yamamoto A, Nakamura K, Miki Y. Successful balloon-occluded retrograde transvenous obliteration for gastric varix mainly draining into the pericardiophrenic vein. Cardiovasc Intervent Radiol. 2012;35:180–183. doi: 10.1007/s00270-011-0233-8. [DOI] [PubMed] [Google Scholar]
- 16.Jogo A, Nishida N, Yamamoto A, et al. Factors associated with aggravation of esophageal varices after B-RTO for gastric varices. Cardiovasc Intervent Radiol. 2014;37:1243–1250. doi: 10.1007/s00270-013-0809-6. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 17.Jogo A, Yamamoto A, Kaminoh T, et al. Utility of low-dose gelatin sponge particles and 5% ethanolamine oleate iopamidol mixture in retrograde transvenous obliteration (GERTO) for gastric varices. Br J Radiol. 2020;93:20190751. doi: 10.1259/bjr.20190751. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 18.Bannas P, Roldán-Alzate A, Johnson K, et al. Longitudinal monitoring of hepatic blood flow before and after TIPS by using 4D-flow MR imaging. Radiology. 2016;281:574–582. doi: 10.1148/radiol.2016152247. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 19.Motosugi U, Roldán-Alzate A, Bannas P, et al. Four-dimensional flow MRI as a marker for risk stratification of gastroesophageal varices in patients with liver cirrhosis. Radiology. 2019;290:101–107. doi: 10.1148/radiol.2018180230. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 20.Ishikawa T, Sasaki R, Nishimura T, et al. Comparison of patients with hepatic encephalopathy and those with gastric varices before and after balloon-occluded retrograde transvenous obliteration. Hepatol Res. 2018;48:1020–1030. doi: 10.1111/hepr.13199. [DOI] [PubMed] [Google Scholar]
- 21.Yamaguchi S, Takeda M, Mitsuhara T, et al. Application of 4D-CTA using 320-row are detector computed tomography on spinal arteriovenous fistulae: initial experience. Neurosurg Rev. 2013;36:289–296. doi: 10.1007/s10143-012-0440-z. [DOI] [PubMed] [Google Scholar]
- 22.Parkes MJ, Green S, Stevens AM, Clutton-Brock TH. Assessing and ensuring patient safety during breath-holding for radiotherapy. Br J Radiol. 2014;87:20140454. doi: 10.1259/bjr.20140454. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 23.Mori S, Yashiro H, Inoue M, Takahara K, Kusaka M, Shiroki R. Urodynamic 4D-CT evaluation: 320-row area detector CT scanner combined with PhyZiodynamics software analysis provides an innovative system to evaluate urinary flow and outlet obstructions. Acta Radiol. 2019;61:558–567. doi: 10.1177/0284185119868909. [DOI] [PubMed] [Google Scholar]
- 24.Sugano S, Yamamoto K, Sasao K, Watanabe M (1999) Portal venous blood flow while breath-holding after inspiration or expiration and during normal respiration in controls and cirrhotics. J Gastroenterol 34:613–618. 10.1007/s005350050381 [DOI] [PubMed]
- 25.Kanal KM, Butler PF, Sengupta D, Bhargavan-Chatfield M, Coombs LP, Morin RL. U.S. diagnostic reference levels and achievable doses for 10 adult CT examinations. Radiology. 2017;284:120–133. doi: 10.1148/radiol.2017161911. [DOI] [PubMed] [Google Scholar]
- 26.Japanese Network for Research and Information on Medical Exposure (2015) Diagnostic reference levels based on latest surveys in Japan – Japan DRLs 2015. http://www.radher.jp/J-RIME/report/DRLhoukokusyoEng.pdf. Accessed 26 Aug 2020
- 27.European Commission . Radiation Protection No. 180 – Diagnostic reference levels in thirty-six European countries (Part 2/2) 2014. [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Data Availability Statement
The relevant data have been included in the manuscript. The datasets used and analyzed during the current study are available from the corresponding author on reasonable request.