Abstract
Inertial measurement unit (IMU)-based gait analysis systems have become popular in clinical environments because of their low cost and quantitative measurement capability. When a shank is selected as the IMU mounting position, an inverted pendulum model (IPM) can accurately estimate its spatial gait parameters. However, the stride-by-stride estimation of gait parameters using one IMU on each shank and the IPMs has not been validated. This study validated a spatial gait parameter estimation method using a shank-based IMU system. Spatial parameters were estimated via the double integration of the linear acceleration transformed by the IMU orientation information. To reduce the integral drift error, an IPM, applied with a linear error model, was introduced at the mid-stance to estimate the update velocity. the gait data of 16 healthy participants that walked normally and slowly were used. The results were validated by comparison with those extracted from an optical motion-capture system; the results showed strong correlation () and good agreement with the gait metrics (stride length, stride velocity, and shank vertical displacement). In addition, the biases of the stride length and stride velocity extracted using the motion capture system were smaller in the IPM than those in the previous method using the zero-velocity-update. The error variabilities of the gait metrics were smaller in the IPM than those in the previous method. These results indicated that the reconstructed shank trajectory achieved a greater accuracy and precision than that of previous methods. This was attributed to the IPM, which demonstrates that shank-based IMU systems with IPMs can accurately reflect many spatial gait parameters including stride velocity.
Subject terms: Human behaviour, Engineering, Physiology
Introduction
In recent years, many studies have focused on the development of an inertial measurement unit (IMU) equipped with an accelerometer, gyroscope sensor, and magnetometer for a gait analysis system that can provide quantitative gait parameters such as stride length, velocity, gait cycle. Compared to the golden standard for gait analysis—a motion capture system and instrumented walkways—the IMU-based system is cost effective, lightweight, and versatile, which are suitable characteristics for clinical and residential applications1. In particular, the inverted pendulum model (IPM) is considered useful for estimating spatial gait parameters such as stride length from acceleration and angular velocities measured by the IMU2,3. However, the step-by-step accuracy of the spatial parameters estimated using the IMU data and IPM model has not been validated thus far.
Kinematics information in the gait cycle is a component of gait considered in clinical gait analyses4. For example, variability in stride length can be used to assess the progression of Parkinson’s disease (PD)5, and stride velocity can be used to predict the risk of adverse events in the elderly6. To estimate spatial gait parameters from acceleration and angular velocities measured by the IMU, the previous studies implemented a segmentation algorithm7–10. Integral computations for spatial parameter estimation are reset at each segmentation point, which reduces errors caused by the measurement noise. The implementation details of these algorithms depend on the attachment position of the IMU because different positions produce different signal characteristics and different assumptions need to be considered to improve estimation accuracy. A wide selection of attachment positions for the IMUs involved bilateral shanks7,11, bilateral insteps12–15, and bilateral heels16,17. In terms of ensuring attachment stability, the shank position seems to be a good choice because the soft tissue of the lower shank has less movement than that at the other positions18. Further, when the IMU is fixed to the heel or instep position on footwear, the moving artifact of the footwear affects the estimation accuracy. However, compared to the heel or instep attachment positions closer to the ground, the shank position has room for improvement in terms of estimating spatial parameters such as stride length and stride velocity. The shank-based method can be used to obtain reliable gait parameters and visualise the 3D-trajectory of each stride19,20. In a previous study, zero-velocity-update (ZUPT) was introduced, which re-initialises the integrated velocity to zero at the segmentation point (mid-stance). However, the shank possesses a small velocity at the mid-stance2, which results in the error being included in the estimated spatial gait parameters. Instead of assuming the velocity at the mid-stance to be zero, the inverted pendulum model (IPM)2,3, which considers the IMU movement at the mid-stance as a circular motion in the sagittal plane and estimates the velocity using angular velocity has the potential to provide a more accurate result. In fact, Wu et al. showed that the total error of the walking distance estimated using shank-mounted IMUs and an IPM was smaller than that estimated using the same IMUs and a ZUPT19.
However, it remains unclear if stride-by-stride estimation using shank-mounted IMUs and the IPM is superior to that using ZUPT. As mentioned above, some variability in the spatial gait parameters is important for the gait assessment of people with gait disorders, and to investigate such variability in gait parameters, stride-by-stride estimation of gait parameters is necessary. This study aimed to validate the stride-by-stride estimation method for spatial gait parameters using shank-mounted IMUs and the IPM. In addition, we investigated whether the stride-by-stride estimation of the spatial gait parameters using the IPM was superior to that using the ZUPT. A modified IPM that considered the posture of the IMU for shank trajectory estimation was implemented. For evaluating the proposed method, we performed a concurrent validation experiment using an optical motion capture system with a high spatial resolution. Next, we compared the proposed method to a ZUPT method to investigate whether the IPM improved the accuracy of the stride-by-stride estimation. We used a previously reported method20 as the ZUPT method because this method had been evaluated for stride-by-stride spatial gait parameters.
Results
All 16 participants completed both tasks; the data for the normal speed task of one of the participants could not be analysed because of synchronisation issues (the motion capture system did not capture the stamp event), and therefore, it was excluded. The motion capture system suffered from loss of data because of the reflective markers; these invalid data were excluded, and a total of 695 strides were extracted from the motion capture system, with 283 strides for the normal speed task and 412 strides for the slower speed task.
First, we validated the proposed method compared to the results from the motion capture system. Figure 1 shows scatter plots for the stride length, stride velocity, and shank vertical displacement between the data extracted from the proposed method and those obtained from the motion capture system. The stride length and velocity decreased when the participants walked slowly. Pearson’s correlation coefficient and the error distribution are summarised in Table 1. All parameters achieve high correlation with Pearson correlation coefficient Thus, the proposed method estimated the parameters appropriately.
Table 1.
Parameters | r | E | |E| | |
---|---|---|---|---|
Stride length (m) | ||||
Normal | 0.98 | 0.012 (0.026) | 0.024 (0.016) | 1.7% (1.1%) |
Slow | 0.99 | 0.002 (0.025) | 0.020 (0.015) | 1.6% (1.2%) |
Overall | 1.00 | 0.007 (0.025) | 0.020 (0.016) | 1.9% (1.4%) |
Shank vertical displacement (m) | ||||
Normal | 0.93 | − 0.011 (0.007) | 0.011 (0.006) | 6.9% (3.9%) |
Slow | 0.91 | − 0.009 (0.007) | 0.009 (0.006) | 6.3% (4.0%) |
Overall | 0.92 | − 0.010 (0.007) | 0.010 (0.006) | 6.5% (4.0%) |
Stride velocity (m/s) | ||||
Normal | 0.98 | 0.014 (0.028) | 0.026 (0.018) | 2.0% (1.4%) |
Slow | 0.99 | 0.002 (0.021) | 0.016 (0.013) | 1.8% (1.4%) |
Overall | 1.00 | 0.007 (0.025) | 0.020 (0.016) | 1.9% (1.4%) |
Second, we investigated whether the proposed method estimated the parameters more accurately than the previous method. The Bland–Altman plots for the parameters estimated using the proposed method and a previous method are shown in Fig. 2. The biases of stride length, stride velocity, and shank vertical displacement for the proposed and previous methods were 0.006 m and m, 0.007 m/s and m/s, and m and m, respectively. The stride length and stride velocity estimated using the proposed method achieved a very small bias compared with those estimated using the previous method. The 95% confidence intervals of stride length, stride velocity, and shank vertical displacement for the proposed and previous methods were 0.099 m and 0.115 m, 0.095 m/s and 0.137 m/s, and 0.026 m and 0.030 m, respectively. Thus, the error variability for all parameters estimated using the proposed method was smaller than that using the previous method.
Discussion
This study aimed to validate stride-by-stride gait estimation using IMUs attached to the shank and an IPM. The parameters were extracted from the walking data of 16 participants, and they were compared with those obtained using a motion capture system as reference. The results showed high correlation and good agreement between the two systems for stride length, stride velocity, and shank vertical displacement, which verifies the technical effectiveness of the proposed method.
In terms of the Bland–Altman analysis, the proposed method achieved a to 0.057 m limit of agreement for the stride length, which is comparable to that achieved using the heel- or instep-based method14,15,17,21. The relative absolute error of the shank vertical displacement is slightly worse than other parameters with an overall value of 6.5% (Table 1). When comparing the spatial parameters in the forward (stride length) and vertical directions (shank vertical displacement), the magnitude of the error tended to be different. The value of movement in the vertical direction was smaller than that in the forward direction, and therefore, the shank vertical displacement would be susceptible to noise. However, the shank vertical displacement still achieved good agreement with a mean of -0.010 m and the limit of agreement of m and 0.003 m.
Compared to the previous method20 as shown in Fig. 2, the proposed method achieved a considerably smaller bias with a stride length of 0.005 m and a stride velocity of 0.007 m/s. This is because the modified IPM used in this study compensates the velocity estimation error at each segmentation point. Further, this result demonstrated that the IPM achieves good performance in terms of 3D-trajectory estimation. The previous IPM required an event assumption that the shank tilt angle in the sagittal plane is zero; however, this was difficult to guarantee22. The the proposed method combines orientation estimation that computes the update velocity in three dimensions and does not require an event assumption. The estimated shank vertical displacement from both methods still have a close bias, which indicates that the vertical direction has a very small velocity at the segmentation point and does not gain considerable benefit from the IPM. The bias of the shank vertical displacement that appears in both methods can be affected by IMU calibration where the linear acceleration in the vertical direction is calculated by subtracting the gravity component23,24.
The above results indicates that the IPM contributes to spatial gait parameter estimation with sufficient accuracy compared to the motion-capture system. The limitation of the current study is that it does not use patient data or elderly data. However, although measurement targets can walk such that the sole of their foot contacts the ground and the feasibility of the pendulum model for abnormal gait assessed in a prior research2 indicates that the proposed method can be applied to patients or the elderly, this still needs to be proved. In particular, the IPM may be strict in estimating the gait parameters of people with severe gait disorders. For example, the IPM may work well when people limp or shuffle because their gait trajectories would not fit to the pendulum model. Future research needs to address the gait events detection problem for abnormal gait because it is the prerequisite for most segmentation-algorithm-based gait trajectory estimation methods. However, the IPM would be useful in distinguishing people with slight gait disorders from healthy people. For example, some studies have attempted to classify early PD patients and healthy elders using gait parameters to develop an early diagnosis method for PD patients25,26. Because the gait of early PD patients is similar to that of healthy people, it is necessary to estimate the walking trajectory as accurately as possible. Therefore, the IPM would work well to classify early PD patients and healthy elders.
For data processing, we did not use high- or low-pass filters for the IMU data because we initially considered that the proposed method could almost eliminate the drift. In fact, the proposed method showed high accuracy in trajectory estimation. However, such filters would theoretically improve the estimation accuracy. Some studies developed a filtering method for IMU data in gait trajectory estimation23,24. In the future, such filters would be applied or added to the proposed method.
Although we could not discuss gender-related difference for the limitation number of the participants, it is possible to increase the accuracy of trajectory estimation using gender differences in the gait. In fact, some studies investigated gender differences in the gait28–30. For example, women’s foot progression angle tends to be more internally rotated than those of men in the stance phase28. This difference can affect the trajectory and estimation accuracy of the trajectory.
Conclusions
In this study, an IMU-based stride-by-stride gait analysis method was validated with an optical motion capture system. The method explored the possibility of using the IPM to analyse the shank trajectory that estimates the stride length, stride velocity, and shank vertical displacement. The accuracy of each parameter estimated by the IPM method was compared with that estimated using a previous method based on the ZUPT. Using the proposed method, the average error of the extracted stride length achieved a small bias of 0.006 m. Overall, this study is a step toward the development of a shank-based gait analysis system that will serve as a convenient objective measurement tool for future clinical diagnoses.
Methods
System setups
In this study, two IMUs (accelerometer and gyroscope) were mounted on the shanks on both sides, immediately above the malleolus at a distance of r. The attachment position and coordinate system of the IMUs are shown in Fig. 3. The x, y, and z axes represented inferior/superior, posterior/anterior, and medial/lateral directions, respectively.
Gait event detection and data segmentation via angular velocity
Heel-strike (HS) and toe-off (TO) events were first detected based on the angular velocity in the sagittal plane 7 (Fig. 4). The search regions for HS and TO events were defined based on the shank tilt angle in the sagittal plane 20 calculated using the integration of . Then, each HS was defined as the first peak that appears after the local maximum of (shank max forward; SMF) and each TO was defined as the minimum of that appears before the local minimum of (shank max backward; SMB). After the HS and TO in each gait cycle were detected, the data were segmented by the mid-stance (MS), which is defined as the maximum point of between HS and TO11.
Orientation estimation from acceleration and angular velocity
The quaternion system was introduced to describe orientation information in this study. We use superscripts to represent the coordinate frame wherein a variable is located and the rotation direction of a quaternion. Superscripts S and E represent the IMU coordinate frame and the laboratory coordinate frame, respectively. Further, we use (k) to represent the kth sample of an instantaneous variable and (i) to represent a variable in the ith gait cycle. For example, represents the instantaneous quaternion that can convert a vector from frame S to frame E. In addition, ms(i) is the mid-stance event in the ith gait cycle.
At each ms(i), we assumed that the IMU is reset and the accelerometer only detects gravity. Therefore, the rotation quaternion at ms(i) is computed using the accelerometer output and gravity as
1 |
with
2 |
3 |
4 |
where ∥ ∥ represents the L2-norm, is the dot product, and is the cross product.
The remaining rotation quaternions in each segment were computed via the integration of angular velocity for each sample as
5 |
with
6 |
where denotes the quaternion product and represents the sampling interval.
Using we can transform the measured acceleration into the laboratory coordinate frame as
7 |
where is the conjugate of
Finally, the linear acceleration is obtained by removing the gravitational component from .
8 |
Velocity estimation and drift removal via inverted pendulum model
After linear acceleration is determined, the velocity is computed using the trapezoidal integral with the sampling interval as
9 |
However, this integral result contains significant drift. This drift occurs from the accumulation of sensor errors such as bias in the gyroscope and jitter in the accelerometer. To solve this problem, a modified IPM was introduced in this phase. At each ms(i), we assume the movement of the shank to be rotational motion with the malleolus as a fulcrum (Fig. 5). Thus, the velocity at ms(i) can be calculated as the cross product between the angular velocity in the laboratory coordinate frame and the position of the IMU
10 |
with
11 |
We assume that contains the gravitational component only at ms(i), and therefore, is defined as the vector in the direction with a magnitude of the distance between the IMU and the malleolus r as
12 |
If the IMU sensor is attached at a certain distance from the ankle, r is known in advance and there is no need to measure it each time.
Then, we model the drift error as a linear variation31 that occurs over time with a constant slope and intercept in each segment as
13 |
where and are the slope and the intercept of the error model in the ith gait cycle, respectively.
Using the velocity estimated from the IPM, and are computed as
14 |
15 |
Finally, the drift error is removed from velocity by subtracting the modelled error, and the correction result is obtained as
16 |
Trajectory estimation and coordinate frame transformation
First, the trajectory is estimated by the direct integration of the corrected velocity via a trapezoidal rule first as
17 |
Then, a new coordinate frame P is introduced for visualising the stride trajectory and computing the spatial parameters. In frame P, the y axis direction is aligned to the stride forward direction and the x axis is aligned along the vertical direction. The rotation matrix between the frames P and E is computed by solving the equation
18 |
where
19 |
and each column of matrix is defined as
20 |
21 |
22 |
The final estimate trajectory in each gait cycle is transformed using the rotation matrix applied to :
23 |
The example of the stride before and after the transformation is shown in Fig. 6.
Spatial gait parameter computation
The definitions of stride length and shank vertical displacement are shown in Fig. 6. Stride length during one gait cycle is computed as the displacement in the y direction between two successive mid-stance events.
24 |
Shank vertical displacement is calculated as the maximum displacement value in the x direction .
25 |
Stride velocity is defined as the division of the stride length and stride duration.
26 |
with
27 |
Evaluation experiment
In this study, six females and ten males (age: ; height: ) participated in the concurrent validation experiment. This experiment was conducted in accordance with the Declaration of Helsinki and approved by the ethics committee of the Tokyo Institute of Technology. Written informed consent was obtained from all participants.
The participants completed a walking tasks at two different self-selected walking speeds: normal and a slower speed, in a fixed order. In the future, we assume elders or patients with gait disorder such as PD patients as the target for our system. Aging and disorders decrease people’s gait velocity32,33, and therefore, we employed the slow walking condition. The participants conducted one trial under each condition. Two IMUs (TSND121, ATR-Promotions, Japan) with features of an accelerometer () and a gyroscope () were used to implement the proposed method. The size of the TSND121 is and the weight is approximately . Each IMU was placed into a housing pocket with an elastic band, and it was attached to the shank in the position above the malleolus.
A motion capture system with 12 cameras (VENUS3D, NOBBYTECH, Japan) and optical motion capture software (Motive:Tracker, NaturalPoint, Inc.) was used as the reference system. The motion capture system was calibrated well to ensure that the overall displacement error was under . The sampling frequencies were set to for both the IMU and the motion capture system. Before the start of the experiment, participants were asked to stamp their left and right foot once to synchronise events between the IMUs and the motion-capture system. We used Python (Python Software Foundation) for data processing and analysis.
For the statistical analysis, we compute the difference in the stride-by-stride gait parameter results extracted using the IMU and the motion-capture system as the error for the proposed method. We computed the mean and standard deviation of the error, the absolute error, and the relative absolute error for each parameter. Further, Bland–Altman analysis34 was introduced to assess the agreement between the IMU and the motion-capture measurement system for the previous20 and proposed methods.
One participants’ data in normal speed was excluded because the motion capture data cannot be synchronized with the IMU data. Finally, 15 normal speed data and 16 slower speed data were analyzed. After removing invalid data resulting from the reflective markers loss in the invisible area, a total of 722 strides were extracted from the motion capture system, with 289 strides (19.27 strides per one participant) in the normal speed task and 433 strides (27.06 strides per one participant) from the slower speed task. The number of strides analyzed is not less than that measured in previous studies13,14,20. All strides can be found in the IMU estimation results without missing detection. Thus, we analyzed these data to validate the proposed method.
Acknowledgements
This work was supported by JST CREST Grant Number JPMJCR1433.
Author contributions
Y.M. and H.O. developed the gait measurement system. Y.M. and N.T. conducted the evaluation experiment. Y.M., T.O., H.O., N.T., and Y.M. discussed and interpreted the results. Y.M. and T.O. drafted the manuscript.
Competing interests
The authors declare no competing interests.
Footnotes
Publisher's note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
References
- 1.Chen S, Lach J, Lo B, Yang G-Z. Toward pervasive gait analysis with wearable sensors: A systematic review. IEEE J. Biomed. Health Inform. 2016;20:1521–1537. doi: 10.1109/JBHI.2016.2608720. [DOI] [PubMed] [Google Scholar]
- 2.Yang S, Zhang JT, Novak AC, Brouwer B, Li Q. Estimation of spatio-temporal parameters for post-stroke hemiparetic gait using inertial sensors. Gait Posture. 2013;37:354–358. doi: 10.1016/j.gaitpost.2012.07.032. [DOI] [PubMed] [Google Scholar]
- 3.Wu, X., Wang, Y. & Pottier, G. A non-zupt gait reconstruction method for ankle sensors. in 36th Annual International Conference of the IEEE Engineering in Medicine and Biology Society, 5884–5887. 10.1109/EMBC.2014.6944967 (2014). [DOI] [PubMed]
- 4.Sutherland DH. The evolution of clinical gait analysis. Part II kinematics. Gait Posture. 2002;16:159–79. doi: 10.1016/s0966-6362(02)00004-8. [DOI] [PubMed] [Google Scholar]
- 5.Morris ME, Iansek R, Matyas TA, Summers JJ. Stride length regulation in Parkinson’s disease. Brain. 1996;119:551–568. doi: 10.1093/brain/119.2.551. [DOI] [PubMed] [Google Scholar]
- 6.Montero-Odasso M, et al. Gait velocity as a single predictor of adverse events in healthy seniors aged 75 years and older. J. Gerontol. Series A Biol. Sci. Med. Sci. 2005;60:1304–1309. doi: 10.1093/gerona/60.10.1304. [DOI] [PubMed] [Google Scholar]
- 7.Salarian A, et al. Gait assessment in Parkinson’s disease: Toward an ambulatory system for long-term monitoring. IEEE Trans. Biomed. Eng. 2004;51:1434–1443. doi: 10.1109/TBME.2004.827933. [DOI] [PubMed] [Google Scholar]
- 8.Sabatini AM. Quaternion-based strap-down integration method for applications of inertial sensing to gait analysis. Med. Biol. Eng. Comput. 2005;43:94–101. doi: 10.1007/BF02345128. [DOI] [PubMed] [Google Scholar]
- 9.Barth J, et al. Stride segmentation during free walk movements using multi-dimensional subsequence dynamic time warping on inertial sensor data. Sensors. 2015;15:6419–6440. doi: 10.3390/s150306419. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10.Hannink J, et al. Benchmarking foot trajectory estimation methods for mobile gait analysis. Sensors. 2017;17:1940. doi: 10.3390/s17091940. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 11.Li Q, Young M, Naing V, Donelan JM. Walking speed estimation using a shank-mounted inertial measurement unit. J. Biomech. 2010;43:1640–1643. doi: 10.1016/j.jbiomech.2010.01.031. [DOI] [PubMed] [Google Scholar]
- 12.Mariani B, Rochat S, Büla CJ, Aminian K. Heel and toe clearance estimation for gait analysis using wireless inertial sensors. IEEE Trans. Biomed. Eng. 2012;59:3162–3168. doi: 10.1109/TBME.2012.2216263. [DOI] [PubMed] [Google Scholar]
- 13.Kitagawa N, Ogihara N. Estimation of foot trajectory during human walking by a wearable inertial measurement unit mounted to the foot. Gait Posture. 2016;45:110–114. doi: 10.1016/j.gaitpost.2016.01.014. [DOI] [PubMed] [Google Scholar]
- 14.Tunca C, et al. Inertial sensor-based robust gait analysis in non-hospital settings for neurological disorders. Sensors. 2017;17:825. doi: 10.3390/s17040825. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 15.Zhou, L. et al. Validation of an IMU gait analysis algorithm for gait monitoring in daily life situations. in 2020 42nd Annual International Conference of the IEEE Engineering in Medicine Biology Society (EMBC), 4229–4232. 10.1109/EMBC44109.2020.9176827 (2020). [DOI] [PubMed]
- 16.Bamberg SJ, Benbasat AY, Scarborough DM, Krebs DE, Paradiso JA. Gait analysis using a shoe-integrated wireless sensor system. IEEE Trans. Inf. Technol. Biomed. 2008;12:413–423. doi: 10.1109/TITB.2007.899493. [DOI] [PubMed] [Google Scholar]
- 17.Kluge F, et al. Towards mobile gait analysis: Concurrent validity and test-retest reliability of an inertial measurement system for the assessment of spatio-temporal gait parameters. Sensors. 2017;17:1522. doi: 10.3390/s17071522. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 18.Catalfamo P, Ghoussayni S, Ewins D. Gait event detection on level ground and incline walking using a rate gyroscope. Sensors. 2010;10:5683–5702. doi: 10.3390/s100605683. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 19.Washabaugh EP, Kalyanaraman T, Adamczyk PG, Claflin ES. Validity and repeatability of inertial measurement units for measuring gait parameters. Gait Posture. 2017;55:87–93. doi: 10.1016/j.gaitpost.2017.04.013. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 20.Ono Y, et al. Inertial measurement unit-based estimation of foot trajectory for clinical gait analysis. Front. Physiol. 2020;10:1530. doi: 10.3389/fphys.2019.01530. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 21.Mariani B, et al. 3D gait assessment in young and elderly subjects using foot-worn inertial sensors. J. Biomech. 2010;43:2999–3006. doi: 10.1016/j.jbiomech.2010.07.003. [DOI] [PubMed] [Google Scholar]
- 22.Yang S, Laudanski A, Li Q. Inertial sensors in estimating walking speed and inclination: An evaluation of sensor error models. Med. Biol. Eng. Comput. 2012;50:383–393. doi: 10.1007/s11517-012-0887-7. [DOI] [PubMed] [Google Scholar]
- 23.Benoussaad M, Sijobert B, Mombaur K, Coste CA. Robust foot clearance estimation based on the integration of foot-mounted IMU acceleration data. Sensors. 2015;16:12. doi: 10.3390/s16010012. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 24.Fan B, Li Q, Liu T. Accurate foot clearance estimation during level and uneven ground walking using inertial sensors. Meas. Sci. Technol. 2020;31:055106. doi: 10.1088/1361-6501/ab6917. [DOI] [Google Scholar]
- 25.Klucken J, et al. Unbiased and mobile gait analysis detects motor impairment in Parkinson’s disease. PLOS ONE. 2013;8:e56956. doi: 10.1371/journal.pone.0056956. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 26.Cuzzolin F, et al. Metric learning for Parkinsonian identification from IMU gait measurements. Gait Posture. 2017;54:127–132. doi: 10.1016/j.gaitpost.2017.02.012. [DOI] [PubMed] [Google Scholar]
- 27.Giavarina D. Understanding bland Altman analysis. Biochem. Med. 2015;25:141–51. doi: 10.11613/BM.2015.015. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 28.Røislien J, et al. Simultaneous estimation of effects of gender, age and walking speed on kinematic gait data. Gait Posture. 2009;30:441–445. doi: 10.1016/j.gaitpost.2009.07.002. [DOI] [PubMed] [Google Scholar]
- 29.Nigg BM, Tecante GKE, Federolf P, Landry SC. Gender differences in lower extremity gait biomechanics during walking using an unstable shoe. Clin. Biomech. 2010;25:1047–1052. doi: 10.1016/j.clinbiomech.2010.07.010. [DOI] [PubMed] [Google Scholar]
- 30.Chiu M-C, Wu H-C, Chang L-Y. Gait speed and gender effects on center of pressure progression during normal walking. Gait Posture. 2013;37:43–48. doi: 10.1016/j.gaitpost.2012.05.030. [DOI] [PubMed] [Google Scholar]
- 31.Rampp A, et al. Inertial sensor-based stride parameter calculation from gait sequences in geriatric patients. IEEE Trans. Biomed. Eng. 2015;62:1089–1097. doi: 10.1109/TBME.2014.2368211. [DOI] [PubMed] [Google Scholar]
- 32.Bohannon RW. Comfortable and maximum walking speed of adults aged 20–79 years: Reference values and determinants. Age Ageing. 1997;26:15–19. doi: 10.1093/ageing/26.1.15. [DOI] [PubMed] [Google Scholar]
- 33.Hass CJ, et al. Quantitative normative gait data in a large cohort of ambulatory persons with Parkinson’s disease. PLoS One. 2012;7:e42337. doi: 10.1371/journal.pone.0042337. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 34.Martin Bland J, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986;327:307–310. doi: 10.1016/S0140-6736(86)90837-8. [DOI] [PubMed] [Google Scholar]