Estimation and correction of instrument artefacts in dynamic impedance spectra

Collins Erinmwingbovo 1, Fabio La Mantia 1,
PMCID: PMC7809267  PMID: 33446729

Abstract

Dynamic impedance spectroscopy is one of the most powerful techniques in the qualitative and quantitative mechanistic studies of electrochemical systems, as it allows for time-resolved investigation and dissection of various physicochemical processes occurring at different time scales. However, due to high-frequency artefacts connected to the non-ideal behaviour of the instrumental setup, dynamic impedance spectra can lead to wrong interpretation and/or extraction of wrong kinetic parameters. These artefacts arise from the non-ideal behaviour of the voltage and current amplifier (I/E converters) and stray capacitance. In this paper, a method for the estimation and correction of high-frequency artefacts arising from non-ideal behaviour of instrumental setup will be discussed. Using resistors, [Fe(CN)6]3-/4- redox couple and nickel hexacyanoferrate nanoparticles, the effect of high-frequency artefacts will be investigated and the extraction of the impedance of the system from the measured dynamic impedance is proposed. It is shown that the correction allows acquiring proper dynamic impedance spectra at frequencies higher than the bandwidth of the potentiostat, and simultaneously acquire high precision cyclic voltammetry.

Subject terms: Analytical chemistry, Electrochemistry, Physical chemistry

Introduction

Electrochemical impedance spectroscopy (EIS) is a highly sensitive technique used for the elucidation of reaction mechanisms and determination of kinetic parameters13. The acquisition of qualitative and quantitative mechanistic information relies on the fitting of high-quality impedance data, acquired over a broad range of frequencies and potentials using physicochemical models2. However, cell and instrument setups have been reported as major sources of artefacts in impedance spectra46. These artefacts can result in the wrong interpretation of the reaction mechanism or the extraction of wrong kinetic parameters. Several methods for identifying artefacts arising from cell setup and procedures for avoiding and/or correcting these artefacts have been reported in literature4,5,713.

Following the works of Fletcher7 and Sadkowski et al.8, Battistel et al. proposed solutions to circumvent the artefacts occurring in a three-electrode cell setup4. It comprises of using a three-electrode cell in a coaxial geometry and a capacitive bridge between the working electrode and reference electrode4. Tran et al. with the help of simulations explained the use of low impedance reference electrode fabricated by connecting a platinum wire to the reference electrode through a capacitor, which has been used in circumventing high-frequency artefacts1416. Low impedance reference electrodes were used to circumvent high-frequency artefacts during the acquisition of dynamic impedance spectra of cathode materials for aqueous batteries such as nickel hexacyanoferrate thin films and LiMn2O4 films17,18.

A zero-gap cell combined with by-pass modified and sensing electrodes was proposed by Stojadinovic and co-workers to minimize distortions arising from artefacts in a four-electrode cell setup10,19. The use of a four-probe setup to measure high impedance values (up to 10 GΩ) to circumvent artefacts arising from four-electrode cell setup has been demonstrated by Fafilek et al.5. Finite element method (FEM) simulations have also been used as a means of identifying limiting conditions for reliable impedance measurement in three-electrode cell setup for lithium-ion batteries20,21. To avoid artefacts arising from geometric and electrochemical asymmetries, Klink et al. suggested the use of modified Swagelok cells with precisely aligned electrodes and a coaxially located reference electrode22.

The aforementioned methods reduce and/or eliminate artefacts arising from cell setups, but they do not address artefacts arising from the instrument setup. The instrument setup used for the data acquisition in EIS contains parts (cables, voltage and current amplifiers), which deviates from ideal behaviour at high frequencies4. Manufacturers use calibration to reduce such artefacts at high frequencies23. In classic EIS, data acquisition parameters, among which the current amplifier, are automatically optimized for each frequency by the software, thereby reducing instrumental artefacts. This is not possible in dynamic impedance spectroscopy, in which the perturbation signal contains all the frequencies at once. Dynamic impedance spectroscopy has been reported as a better alternative for investigating the kinetics of unstable electrochemical systems and electrochemical systems under non-stationary conditions17,18,2437. However, instrumental artefacts are particularly troublesome in dynamic impedance spectroscopy due to two contrasting requirements: on one side the acquisition of good dc data (cyclic voltammetry) requires the selection of a low-current amplifier; on the other side, to reach high frequencies a high-current amplifier with high bandwidth has to be chosen.

In this paper, we show how, by estimating the transimpedance response of the instrument, it is possible to correct the instrumental artefacts at high frequencies and increase the bandwidth of the current amplifier. In particular, we will illustrate as practical examples, the measurement of dynamic impedance spectra of redox couples in solution and potassium (de-)insertion in nickel hexacyanoferrate.

Sources of instrumental artefacts

Instrument setups used for acquisition of impedance spectra such as potentiostats contain components (voltage and current amplifiers), which deviate from their ideal behaviour at high frequencies thereby introducing artefacts to the measured impedance. The current amplifier of the potentiostat measures the current as a voltage drop through a resistor Rm. Switchable resistors of different magnitude are used to cover different current ranges, thus allowing for high precision measurements38. The value of Rm determines also the voltage per full current range of the current amplifier, which in the instrument used in this work is 1 V. As an example, this translates to Rm equal to 1 kΩ when the current range of 1 mA is selected. The achievable gain of the current amplifier depends on the input impedance (measured impedance) and the transimpedance of the I/E converter. The bandwidth of the I/E converter depends on the gain such that the greater the gain (lower current range), the lower the bandwidth of the I/E converter. Measuring impedance at frequencies approaching or exceeding the bandwidth of the I/E converter could result in the introduction of artefacts arising from the instrumentation. Therefore, it is important to investigate the bandwidth of the I/E converter at different current ranges to properly choose it in connection to the perturbation signal.

The voltage signal applied to the cell at high-frequency by the control loop in the potentiostat may be shifted in phase due to its response bandwidth39. The use of larger inputs and monitoring of the feedback signal to the cell has been suggested for circumventing this problem2,39. Another source of artefacts in impedance measurements connected to both instrument and cell setups is the stray capacitance. Stray capacitances have been reported to occur between CE–RE, RE—ground, RE–WE and WE—ground23. Despite the calibration done by manufacturers to eliminate these stray capacitance, it has been reported that electrochemical cells affect the value of the stray capacitances23. Here, as a low-impedance RE is used in the experiments, it is possible to ignore the stray capacitances between CE–RE and RE—ground, as explained in literature23.

Theoretical description of the potentiostat

Figure 1 shows the electric circuit of a potentiostat in a two-electrode cell configuration. Zs in Fig. 1 is the impedance of the system being measured while Cst is the stray capacitance between the WE–CE/RE. This is the stray capacitance of interest as the effect of the other stray capacitance in the potentiostat (stray capacitance from RE—ground and from RE–CE) becomes negligible when a low impedance RE is used23. The measured impedance (Zm) can be described as23:

Zm=V2Im=V2V1Rm 1

where V2 is the voltage drop across Zs and Cst while Im is the measured current. V2 can be described as:

V2=IsZsZV2=IstjωCstZV2 2

where Is and Ist are the current flowing through Zs and Cst respectively. ZV2 is the transimpedance of the electrometer (V2) which is unknown. V1, which measures the current as a voltage drop across Rm is given by:

V1=ImZV1Rm=ICm1jωCmZV1 3

where ZV1 is the impedance of V1 and ICm is the current flowing through Cm. Equation (1) can then be rewritten as:

Zm=IsZsZV2ImZV1 4

Figure 1.

Figure 1

Simplified schematic diagram of a potentiostat in a two-electrode cell setup used for measuring the transimpedance and stray capacitance of the potentiostat used in this work. Zs denotes the impedance of the system being measured, Is is the current flowing through Zs, Cst is the stray capacitance and Ist is the current flowing through Cst. Rm is the current measuring resistor, V2 denotes the voltage drop across Zs and Cst, while V1 denotes the voltage drop across Rm and Cm. Icm and Im represent the current flowing through Cm and the measured current respectively.

The total current (IT) is given by IT=Is+Ist=Im+ICm, subsequently IT=Im(1+jωCmRm) and Im=IT/(1+jωCmRm). Defining ZV1=ZV1/(1+jωCmRm), Eq. (4) can be rewritten as:

Zm=IsZsZV2ITZV1 5
Zm=IsZsZV2(Ist+Is)ZV1 6
1Zm=ZV1ZV2Zs+IstZV1IsZsZV2 7
ZsZm=ZV1ZV2+IstZV1IsZV2 8

Redefining Is and Ist from Eq. (2) will lead to Is=V2/ZV2Zs and Ist=V2jωCst/ZV2. Substituting the term Ist/Is in Eq. (8) with jωCstZs leads to:

ZsZm=ZV1ZV2+jωCstZsZV1ZV2 9

The transimpedance of the potentiostat Ztr can be defined as Ztr=ZV1/ZV2. Subsequently, Eq. (9) results to:

ZsZm=Ztr(1+jωCstZs) 10

Equation (10) indicates a linear relationship between the relative admittance of the system being measured (Zs/Zm) and Zs. Ztr is frequency-dependent and extrapolated at each frequency as explained in S2 of the supporting information, while Cst reported in this paper is an average of the Cst at different current ranges.

Results and discussions

Extracting the transimpedance and stray capacitance

To extract the transimpedance of the potentiostat (Ztr) and stray capacitance (Cst), the impedance of high precision resistors at different current ranges of the potentiostat from ten times to one-tenth of the corresponding Rm of the current range was measured. Ztr and Cst were extracted from the intercept and slope of the first-order polynomial fit of the relative admittance (Zs/Zm) versus Zs according to Eq. (10). The stray capacitance due to the potentiostat and cell setup Cst is obtained from the slope of the plot of the high-frequency data point of jωCst versus ω (Fig. S2 in the supporting information). The stray capacitance (Cst) was estimated to be 240 ± 5 pF.

Bandwidth of the potentiostat

The bandwidth of the potentiostat in terms of the transimpedance (Ztr) at different current ranges is shown in Fig. 2. Ideally, Ztr should be unitary at all frequencies. However, the result shows that the bandwidth (frequency where the modulus of Ztr deviates from the ideal unitary value of − 30%), decreases from 800 kHz to 8 kHz when current range decreases from 1 mA to 1 μA as expected.

Figure 2.

Figure 2

The plot of the magnitude of the transimpedance of the potentiostat (Ztr) at different current range versus frequency. The plot shows the 3 dB line, which allows for the bandwidth of the potentiostat at different current ranges to be read off.

Extracting impedance of the system from measured impedance

The impedance of the various electrochemical systems (resistors at different current ranges, [Fe(CN)6]3-/4- redox couple and nickel hexacyanoferrate) devoid of the instrumental artefacts were extracted from the measured impedance using Eq. (11) and the estimated Cst of the potentiostat and Ztr of the current range of interest.

Zcorr=ZtrZm1-jωCstZmZtr 11

Resistor

The use of Eq. (11) to obtain the real impedance from the measured impedance of various resistors is illustrated in this section of the paper. The magnitude and phase of the impedance of 2 kΩ at 1 mA and 20 kΩ at 100 μA current range is shown in Fig. 3.

Figure 3.

Figure 3

Correction method applied to 2 kΩ at 1 mA and 20 kΩ at 100 μA current range. circle represents the measured impedance, while dash line represents the corrected impedance using the proposed method.

Ideally, the magnitude of the impedance of a resistor is constant and the phase is zero at all frequencies. The measured impedance (circle in Fig. 3) was observed to deviate from its constant magnitude and zero phase at ca. 100 kHz in both cases. The deviation can be attributed to instrument artefacts. Using the correction method, the real impedance (unitary magnitude and zero phase at all frequencies) can be extracted (continuous lines in Fig. 3). This illustrates that the proposed correction method can be used for extending digitally the bandwidth of the potentiostat. The limit of the correction method for the different current ranges is shown in section S3 of the supporting information.

Redox couple

The effect of instrument artefacts and application of the proposed correction method was also investigated using a [Fe(CN)6]3-/4- redox couple on a 250 μm Pt electrode in a three-electrode cell configuration. The dynamic impedance spectra were recorded during quasi-cyclic voltammetry, as explained in31,32,35. In this case, as discussed previously, the current range has to be selected to obtain undistorted spectra at high-frequency and simultaneously have good quality dc data (quasi-cyclic voltammetry). The uncompensated cell resistance was 275 Ω, in agreement with the conductivity of the solution and the dimension of the working electrode. For measuring such resistance properly at 1 MHz, a current range of 10 mA ought to be used. However, as seen in Fig. 4a, the use of 10 mA current range would strongly reduce the quality of the extracted voltammogram. It is important to highlight that the reduced data quality of the voltammogram at 10 mA does not originate from interactions between the ac and dc signal as shown in section S4 of the supporting information but from the precision of the current amplifier.

Figure 4.

Figure 4

(a) Quasi Voltammogram obtained at the different current range for 250 μm Pt electrode in 5 mM K3[Fe(CN)6] solution with 1 M KCl as supporting electrolyte. (b) Nyquist plot of impedance spectra obtained for 250 μm Pt electrode in 5 mM K3[Fe(CN)6] solution with 1 M KCl as supporting electrolyte using 1 mA and 10 mA current range with an insert showing the high-frequency region. (c) and (d) Magnitude (Z) and Phase of impedance spectra obtained for 250 μm Pt electrode in 5 mM K3[Fe(CN)6] solution with 1 M KCl as supporting electrolyte using 1 mA and 10 mA current range.

Repeating the same experiment with a 1 mA current range improves noticeably the data quality. However, it introduces a distortion connected to the transimpedance of the potentiostat in the form of an inductive component, as observable in the inset in Fig. 4b and the magnitude and phase of the impedance shown in Fig. 4c,d respectively. It is assumed that the potentiostat is at virtual ground and the current is then measured as the voltage drop across Rm. However, this is not the case as the frequency increases, rather the characteristic of the current amplifier at high frequencies resembles that of an inductor, hence the potentiostat, in reality, controls the current via an RLC element6. The inductive component in the measured impedance, which is an artefact from the current amplifier due to the usage of an inappropriate current range (1 mA), can be compensated by ZV1, which is one of the components of Ztr. Using the proposed correction method (Eq. (11)) and the extracted Ztr for 1 mA current range, the real impedance can be extracted from the measured impedance as shown in the inset of Fig. 4b (green curve), and Fig. 4c,d. The result indicates that the proposed correction method allows choosing an appropriate current range, without compromising the high-frequency data in the impedance spectra, thus enabling the measurement of dynamic impedance at high-frequency. This is particularly important in the analysis of adsorption phenomena and the study of the double layer at microelectrodes.

Nickel hexacyanoferrate nanoparticles

In this section, we illustrate furthermore the effect of instrumental artefacts on the data evaluation of more complex systems. The system of choice is nickel hexacyanoferrate (NiHCF) nanoparticles, which have been reported as a promising cathode material for aqueous and non-aqueous battery systems40,41 and lithium recovery4148.

For intercalation materials such as NiHCF nanoparticles, the impedance is normally acquired over a wide range of frequencies as it allows for the study of the various physicochemical processes occurring during the intercalation process such as (de)adsorption, interfacial charge transfer and mass transport in the solid. The Nyquist plot of the impedance spectra in Fig. 5a shows two RC time constants followed by a straight line attributed to mass transport in the solid. A comparison of the high-frequency data points between the measured and corrected impedance indicates that the former is shifted with respect to the latter (Fig. 5a,b), which can be attributed to the transimpedance of the potentiostat. Although the measured impedance can still be fitted using the same equivalent circuit (Fig. 5c)17, the extracted kinetic parameters are different, as shown in Table 1. A 5.6 % error was observed for the extracted capacitance of the double layer (Cdl). The error introduced by instrumental artefacts exceeds 10 % for other parameters in the high-frequency region such as adsorption resistance (Rad), charge transfer resistance (Rct) and the resistance of the pores (Rp). As expected, kinetic parameters of physicochemical processes occurring in the low frequency were unaffected.

Figure 5.

Figure 5

(a) and (b) Nyquist plot and Bode plot of NiHCF nanoparticles in 0.5 M K2SO4 within the frequency range of 350 kHz to 2.8 Hz at 0.50 V during the cathodic scan respectively. (c) Equivalent circuit of the porous electrode including the equivalent circuit obtained from modelling the reversible insertion as two step process treating the capacitance of adsorption (Cad ) as a short circuit and describing the mass transport in the solid with a capacitor17.

Table 1.

Extracted parameters from fitting the measured and corrected impedance of NiHCF nanoparticles in 0.5 M K2SO4 with the equivalent circuit (Fig. 5c. Ru denotes the uncompensated cell resistance, Cdl is the capacitance of the double layer, Rad is the adsorption resistance, σad is the Warburg coefficient of the cation in the liquid, Rct is the charge transfer resistance, Cct is the capacitor describing the mass transport in the solid and Rp is the pore resistance49.

Parameters Measured impedance Corrected impedance
Ru (Ω) 18.24 18.67
Cdl (F) 5.64 · 10-6 5.98 · 10-6
Rad (Ω) 1.39 1.64
σad (Ω·s0.5) 57.24 57.92
Rct (Ω) 0.86 0.72
Cct (F) 6.54 · 10-3 6.56 · 10-3
Rp (Ω) 4.44 5.56

Conclusion

The results shown in this paper indicate that the artefacts arising from transimpedance of the potentiostat and stray capacitance between WE–CE can be properly corrected by taking into account the equivalent circuit of the potentiostat and the frequency response of the voltage and current amplifiers. This is particularly important in dynamic impedance spectroscopy when a broadband signal is used to measure the impedance in a large range of frequencies. When current ranges above 1 mA are used, the impedance can be acquired with an error less than 1% in magnitude and 1 in phase. The bandwidth of the potentiostat decreases with decreasing current range. The possibility to estimate the correct impedance from the measured impedance was illustrated using 2 kΩ and 20 kΩ resistors. The results indicate that it is possible to acquire the proper impedance values of the resistors up to 1 MHz despite the use of current ranges with bandwidth as low as 100 kHz, thus extending digitally the bandwidth of the potentiostat. Using the proposed correction method, the high-frequency artefacts (inductive behaviour) observed in the measured impedance of a redox couple acquired using a current range of 1 mA were properly corrected. This highlights that the proposed correction method allows choosing properly the current range not only for obtaining reliable high-frequency impedance data, but also for measuring appropriately the cyclic voltammetry. The need to correct measured impedance data was illustrated using NiHCF nanoparticles, where the quantification of kinetic parameters can be severely compromised by the presence of the high-frequency artefacts.

Methods

The stray capacitance CE–RE and the transimpedance of the potentiostat were measured using high precision resistors polarized at 0 V in a two-electrode cell configuration. A multisine wave consisting of 45 frequencies covering 5 decades was used as the perturbation signal. The base frequency of the multisine (fac) was 1 Hz corresponding to measuring the impedance from 1 MHz to 8 Hz. A summary of the parameters used for the acquisition of the impedance spectra can be found in Table 2.

Table 2.

Summary of DMFA parameters used in the acquisition of dynamic impedance.

Parameters Resistors [Fe(CN)6]3-/4- NiHCF
Eh (V) 0 0.25 0.1
fdc (mHz) 50 5
ΔUdc (mVpp) 250 400
fac (Hz) 1 1 0.35
ΔUac (mVpp) 100 50 50
No of Msample 20 200 400
sample rate (μs) 0.2 0.2 1
current range (mA) 10 1 100
bw (Hz) 2 2 1

*The amplitude of the multisine was scaled with the measured resistors to avoid errors due to current-voltage amplification. 100 mVpp was used for the equivalent resistor (Rm) of the current range and the amplitude used for other resistors was scaled by the ratio of the resistor to Rm of the current range.

**A current range of 1 mA was also used to investigate the effect of selecting an inappropriate current range.

NiHCF nanoparticles were synthesized using the co-precipitation method reported in42. The procedure consists of the simultaneous addition of 120 ml of 50 mM of K3Fe(CN)6 and 120 ml of 100 mM of Ni(NO3)2 using a flow-rate of 1 ml per minute to 60 ml of distilled water held at 70C in a water bath while stirring constantly42. The resulting suspension was sonicated for 30 minutes at 70C and left overnight to settle. The precipitates were washed with distilled water and dried for 12 hours at 60C . To prepare NiHCF electrodes, 3 mm glassy carbon electrodes (GCE) were polished using consecutively a 2000 grit sandpaper rotating at 400 rpm, diamond suspension (Struers) of 0.250 μm and 0.1 μm rotating at 300 rpm. The polished electrodes were rinsed with distilled water and sonicated for three minutes to remove any diamond suspension left on the electrode surface. The cleaned GCE surface was coated using the doctor blade method (200 μm) with NiHCF slurries made with a weight ratio of 80:10:10 of the synthesized NiHCF, carbon C65 (Timcal, Bodio Switzerland) and polyvinylidene fluoride (PVDF) solution in NMP (Solef S5130, Solvay) respectively.

The electrochemical measurements of the NiHCF nanoparticles and the redox couple were carried out in a three-electrode cell setup in coaxial cell geometry to reduce artefacts arising from electrode configuration4. The electrolytes used for NiHCF nanoparticles and redox couple measurement were 0.5 M K2SO4 and 20 mM K3[Fe(CN)6] in 1 M KCl respectively. The working electrode (WE) was the 3 mm modified NiHCF nanoparticle electrode or a 250 μm platinum electrode depending on the system under investigation, while the counter electrode (CE) was a platinum mesh (Labor Platina). The reference electrode (RE) was a capacitively-coupled low-impedance (20 Ω) Ag/AgCl (3M KCl) fabricated by coiling a 0.1 mm platinum wire around the tip of a Ag/AgCl reference electrode and connecting it through a 100 nF capacitor to the Ag wire15. The use of this low impedance reference electrode reduces high-frequency artefacts associated with the cell setup4,15,18.

Dynamic impedance spectra were acquired by perturbing the WE electrode held at predetermined potential (Eh), with a combined quasi-triangular waveform (QTW) and a multisine wave created by a waveform generator. The frequency of the QTW (fdc) and the amplitude (ΔUdc) used for the [Fe(CN)6]3-/4- redox couple and NiHCF nanoparticles are shown in Table 2, which results in an equivalent scan rate of 50 mVs-1 and 8 mVs-1 respectively. Dynamic impedance spectra were acquired between 1 MHz and 8 Hz for the [Fe(CN)6]3-/4- redox couple and 350 kHz to 2.8 Hz for the NiHCF nanoparticle using the base frequency of the multisine (fac) in Table 2. The amplitude of the multisine (ΔUac) used for both the [Fe(CN)6]3-/4- redox couple and NiHCF nanoparticles was 50 mVpp resulting in negligible error from the nonlinear components as shown in Fig. 6. Details of the effect of the intensity on the multisine can be found in the supporting information. The input (voltage perturbation) and output (current response) signals were measured using a two-channel oscilloscope 4262 PicoScope (Pico technology). The measured impedance was extracted with a home-made MATLAB scripts using one of the heuristic description of dynamic impedance32,35:

Z=iFT[ΔU(ω).g(ω-ω),bw]iFT[ΔI(ω).g(ω-ω),bw] 12

where Δ U and Δ I are the Fourier transform of the potential and current signals respectively, ω is the angular frequency, g is a quadrature filter function, and bw is the bandwidth of the quadrature filter (Table 2). More details about the application of the quasi-triangular waveform, dynamic multi-frequency analysis and the data extraction and evaluation can be found in31,32,35.

Figure 6.

Figure 6

Plot of the estimated error due to the nonlinear component of the fundamental frequencies of dynamic impedance acquired using a multisine intensity of 50 mVpp in (a) [Fe(CN)6]3-/4- redox couple (b) nickel hexacyanoferrate.

Supplementary Information

Acknowledgements

The authors gratefully acknowledge the support of the European Research Council (ERC) under the European Union’s Horizon 2020 research and innovation programme (Grant Agreement Number 772579).

Author contributions

C.E. wrote the main manuscript text, prepared the figures, performed the experiments; F.L.M. developed the theory. All authors revised the manuscript.

Funding

Open Access funding enabled and organized by Projekt DEAL.

Data availability

The datasets generated during and/or analysed during the current study are available from the corresponding author on reasonable request.

Competing interests

The authors declare no competing interests.

Footnotes

Publisher's note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Supplementary Information

The online version contains supplementary material available at 10.1038/s41598-020-80468-x.

References

  • 1.Bard, A. J. & Faulkner, L. R. Electrochemical Methods: Fundamentals and Applications (eds. Bard, A. J., Faulkner, L. R., 2nd ed. edn (Wiley, New York, 2001).
  • 2.Orazem ME, Tribollet B. Electrochemical Impedance Spectroscopy. New York: Wiley; 2008. [Google Scholar]
  • 3.Chan CK, et al. Advanced and in situ analytical methods for solar fuel materials. Top. Curr. Chem. 2016;371:253–324. doi: 10.1007/128_2015_650. [DOI] [PubMed] [Google Scholar]
  • 4.Battistel A, Fan M, Stojadinović J, La Mantia F. Analysis and mitigation of the artefacts in electrochemical impedance spectroscopy due to three-electrode geometry. Electrochim. Acta. 2014;135:133–138. doi: 10.1016/J.ELECTACTA.2014.05.011. [DOI] [Google Scholar]
  • 5.Fafilek G, Breiter M. Instrumentation for ac four-probe measurements of large impedances. J. Electroanal. Chem. 1997;430:269–278. doi: 10.1016/S0022-0728(97)00263-5. [DOI] [Google Scholar]
  • 6.Davis JE, Toren EC. Instability of current followers in potentiostat circuits. Anal. Chem. 1974;46:647–650. doi: 10.1021/ac60342a006. [DOI] [Google Scholar]
  • 7.Fletcher S. The two-terminal equivalent network of a three-terminal electrochemical cell. Electrochem. commun. 2001;3:692–696. doi: 10.1016/S1388-2481(01)00233-8. [DOI] [Google Scholar]
  • 8.Sadkowski A, Diard J-P. On the Fletcher’s two-terminal equivalent network of a three-terminal electrochemical cell. Electrochim. Acta. 2010;55:1907–1911. doi: 10.1016/J.ELECTACTA.2009.11.008. [DOI] [Google Scholar]
  • 9.Balabajew M, Roling B. Minimizing artifacts in three-electrode double layer capacitance measurements caused by stray capacitances. Electrochim. Acta. 2015;176:907–918. doi: 10.1016/J.ELECTACTA.2015.07.074. [DOI] [Google Scholar]
  • 10.Fan M, Stojadinović J, Battistel A, LaMantia F. Analysis and mitigation of electrochemical impedance spectroscopy artefacts in four-electrode cells: model and simulations. ChemElectroChem. 2015;2:970–975. doi: 10.1002/celc.201500121. [DOI] [Google Scholar]
  • 11.Khandpur RS. Handbook of Analytical Instruments. London: McGraw-Hill; 2007. [Google Scholar]
  • 12.Anderson EL, Bühlmann P. Electrochemical impedance spectroscopy of ion-selective membranes: artifacts in two-, three-, and four-electrode measurements. Anal. Chem. 2016;88:9738–9745. doi: 10.1021/acs.analchem.6b02641. [DOI] [PubMed] [Google Scholar]
  • 13.Veal BW, Baldo PM, Paulikas AP, Eastman JA. Understanding artifacts in impedance spectroscopy. J. Electrochem. Soc. 2015;162:H47–H57. doi: 10.1149/2.0791501jes. [DOI] [Google Scholar]
  • 14.Herrmann CC, Perrault GG, Pilla AA. Dual reference electrode for electrochemical pulse studies. Anal. Chem. 1968;40:1173–1174. doi: 10.1021/ac60263a011. [DOI] [Google Scholar]
  • 15.Mansfeld F. Minimization of high-frequency phase shifts in impedance measurements. J. Electrochem. Soc. 1988;135:906. doi: 10.1149/1.2095825. [DOI] [Google Scholar]
  • 16.Tran AT, Huet F, Ngo K, Rousseau P. Artefacts in electrochemical impedance measurement in electrolytic solutions due to the reference electrode. Electrochim. Acta. 2011;56:8034–8039. doi: 10.1016/j.electacta.2010.12.088. [DOI] [Google Scholar]
  • 17.Erinmwingbovo C, Koster D, Brogioli D, LaMantia F. Dynamic impedance spectroscopy of nickel hexacyanoferrate thin films. ChemElectroChem. 2019 doi: 10.1002/celc.201900805. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18.Erinmwingbovo C, et al. Dynamic impedance spectroscopy of LiMn2O4 thin films made by multi-layer pulsed laser deposition. Electrochim. Acta. 2019 doi: 10.1016/j.electacta.2019.135385. [DOI] [Google Scholar]
  • 19.Stojadinovic J, Fan M, Battistel A, La Mantia F. Analysis and mitigation of electrochemical impedance spectroscopy artefacts in four-electrode cells: experimental aspects. ChemElectroChem. 2015;2:1031–1035. doi: 10.1002/celc.201500029. [DOI] [Google Scholar]
  • 20.Klink S, Höche D, La Mantia F, Schuhmann W. FEM modelling of a coaxial three-electrode test cell for electrochemical impedance spectroscopy in lithium ion batteries. J. Power Sources. 2013;240:273–280. doi: 10.1016/j.jpowsour.2013.03.186. [DOI] [Google Scholar]
  • 21.Ender M, Weber A, Ivers-Tiffe E. Analysis of three-electrode setups for AC-impedance measurements on lithium-ion cells by FEM simulations. J. Electrochem. Soc. 2012;159:A128. doi: 10.1149/2.100202jes. [DOI] [Google Scholar]
  • 22.Klink S, et al. The importance of cell geometry for electrochemical impedance spectroscopy in three-electrode lithium ion battery test cells. Electrochem. Commun. 2012;22:120–123. doi: 10.1016/j.elecom.2012.06.010. [DOI] [Google Scholar]
  • 23.Petrescu B, Diard J. Equivalent Model of an Electrochemical Cell Including the Reference Electrode Impedance and the Potentiostat Parasitics. Paris: Claix; 2013. [Google Scholar]
  • 24.Bond AM, Schwall RJ, Smith DE. On-line FFT faradaic admittance measurements application to A.C. cyclic voltammetry. J. Electroanal. Chem. 1977;85:231–247. doi: 10.1016/S0022-0728(77)80290-8. [DOI] [Google Scholar]
  • 25.Stoynov Z, Savova-Stoynov B, Kossev T. Non-stationary impedance analysis of lead/acid batteries. J. Power Sources. 1990;30:275–285. doi: 10.1016/0378-7753(93)80085-4. [DOI] [Google Scholar]
  • 26.Van Gheem E, et al. Electrochemical impedance spectroscopy in the presence of non-linear distortions and non-stationary behaviour Part I: Theory and validation. Electrochim. Acta. 2004;49:4753–4762. doi: 10.1016/j.electacta.2004.05.039. [DOI] [Google Scholar]
  • 27.Darowicki K, Ślepski P, Szociński M. Application of the dynamic EIS to investigation of transport within organic coatings. Prog. Org. Coat. 2005;52:306–310. doi: 10.1016/j.porgcoat.2004.06.007. [DOI] [Google Scholar]
  • 28.Van Gheem E, et al. Electrochemical impedance spectroscopy in the presence of non-linear distortions and non-stationary behaviour: Part II. Application to crystallographic pitting corrosion of aluminium. Electrochim. Acta. 2006;51:1443–1452. doi: 10.1016/j.electacta.2005.02.096. [DOI] [Google Scholar]
  • 29.Sacci RL, Harrington DA. Dynamic electrochemical impedance spectroscopy. ECS Trans. 2009;19:31–42. doi: 10.1149/1.3247564. [DOI] [Google Scholar]
  • 30.Breugelmans T, et al. Odd random phase multisine electrochemical impedance spectroscopy to quantify a non-stationary behaviour: Theory and validation by calculating an instantaneous impedance value. Electrochim. Acta. 2012;76:375–382. doi: 10.1016/j.electacta.2012.05.051. [DOI] [Google Scholar]
  • 31.Battistel A, Du G, La Mantia F. On the analysis of non-stationary impedance spectra. Electroanalysis. 2016;28:2346–2353. doi: 10.1002/elan.201600260. [DOI] [Google Scholar]
  • 32.Koster D, Du G, Battistel A, La Mantia F. Dynamic impedance spectroscopy using dynamic multi-frequency analysis: a theoretical and experimental investigation. Electrochim. Acta. 2017;246:553–563. doi: 10.1016/j.electacta.2017.06.060. [DOI] [Google Scholar]
  • 33.Koster D, et al. Measurement and analysis of dynamic impedance spectra acquired during the oscillatory electrodissolution of p-type silicon in fluoride-containing electrolytes. ChemElectroChem. 2018;5:1548–1551. doi: 10.1002/celc.201800252. [DOI] [Google Scholar]
  • 34.Koster D, Zeradjanin AR, Battistel A, La Mantia F. Extracting the kinetic parameters of the hydrogen evolution reaction at Pt in acidic media by means of dynamic multi-frequency analysis. Electrochim. Acta. 2019 doi: 10.1016/J.ELECTACTA.2019.04.013. [DOI] [Google Scholar]
  • 35.Battistel A, La Mantia F. On the physical definition of dynamic impedance: how to design an optimal strategy for data extraction. Electrochim. Acta. 2019;304:513–520. doi: 10.1016/J.ELECTACTA.2019.03.033. [DOI] [Google Scholar]
  • 36.Pajkossy T. Dynamic electrochemical impedance spectroscopy of quasi-reversible redox systems. Properties of the Faradaic impedance, and relations to those of voltammograms. Electrochim. Acta. 2019;308:410–417. doi: 10.1016/j.electacta.2019.03.197. [DOI] [Google Scholar]
  • 37.Pajkossy T, Mészáros G. Connection of CVs and impedance spectra of reversible redox systems, as used for the validation of a dynamic electrochemical impedance spectrum measurement system. J. Solid State Electrochem. 2020;24:2883–2889. doi: 10.1007/s10008-020-04661-8. [DOI] [Google Scholar]
  • 38.Unwin PR. Introduction to electroanalytical techniques and instrumentation. Encycl. Electrochem. 2007 doi: 10.1002/9783527610426.bard030101. [DOI] [Google Scholar]
  • 39.Battistel, A. Development of the Intermodulated Differential Immittance Spectroscopy for Electrochemical Analysis (2014).
  • 40.Wessells CD, Peddada SV, Huggins RA, Cui Y. Nickel hexacyanoferrate nanoparticle electrodes for aqueous sodium and potassium ion batteries. Nano Lett. 2011;11:5421–5425. doi: 10.1021/nl203193q. [DOI] [PubMed] [Google Scholar]
  • 41.Erinmwingbovo C, Palagonia M, Brogioli D, La Mantia F. Intercalation into a prussian blue derivative from solutions containing two species of cations. ChemPhysChem. 2017 doi: 10.1002/cphc.201700020. [DOI] [PubMed] [Google Scholar]
  • 42.Trócoli R, Battistel A, LaMantia F. Nickel hexacyanoferrate as suitable alternative to ag for electrochemical lithium recovery. ChemSusChem. 2015;8:2514–2519. doi: 10.1002/cssc.201500368. [DOI] [PubMed] [Google Scholar]
  • 43.Trócoli R, Erinmwingbovo C, La Mantia F. Optimized lithium recovery from brines by using an electrochemical ion-pumping process based on λ-MnO2 and nickel hexacyanoferrate. ChemElectroChem. 2017 doi: 10.1002/celc.201600509. [DOI] [Google Scholar]
  • 44.Palagonia MS, Brogioli D, Mantia FL. Influence of hydrodynamics on the lithium recovery efficiency in an electrochemical ion pumping separation process. J. Electrochem. Soc. 2017;164:E586–E595. doi: 10.1149/2.1531714jes. [DOI] [Google Scholar]
  • 45.Palagonia MS, Brogioli D, La Mantia F. Effect of current density and mass loading on the performance of a flow-through electrodes cell for lithium recovery. J. Electrochem. Soc. 2019;166:E286–E292. doi: 10.1149/2.0221910jes. [DOI] [Google Scholar]
  • 46.Palagonia MS, Brogioli D, La Mantia F. Lithium recovery from diluted brine by means of electrochemical ion exchange in a flow-through-electrodes cell. Desalination. 2020;475:114192. doi: 10.1016/j.desal.2019.114192. [DOI] [Google Scholar]
  • 47.Trocoli R, Bidhendi GK, La Mantia F. Lithium recovery by means of electrochemical ion pumping: a comparison between salt capturing and selective exchange. J. Phys. Condens. Matter. 2016;28:114005. doi: 10.1088/0953-8984/28/11/114005. [DOI] [PubMed] [Google Scholar]
  • 48.Xie N, Li Y, Lu Y, Gong J, Hu X. Electrochemically controlled reversible lithium capture and release enabled by LiMn2O4. ChemElectroChem. 2020;7:105–111. doi: 10.1002/celc.201901728. [DOI] [Google Scholar]
  • 49.Erinmwingbovo, C. Kinetics of the reversible insertion of cations in positive electrode materials under dynamic conditions, 10.26092/elib/153 (2020).

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

Data Availability Statement

The datasets generated during and/or analysed during the current study are available from the corresponding author on reasonable request.