. 2020 Nov 3;34(6):3601–3610. doi: 10.21873/invivo.12205

Comparison of the Outcome of Patients Protected by the Wearable Cardioverter Defibrillator (WCD) for <90 Wear Days versus ≥90 Wear Days

TOBIAS C DREHER 1, IBRAHIM EL-BATTRAWY 1,2, SUSANNE RÖGER 1,2, STEPHANIE L ROSENKAIMER 1,2, STEFAN GERHARDS 1,2, JÜRGEN KUSCHYK 1,2, MARTIN BORGGREFE 1,2, IBRAHIM AKIN 1,2
PMCID: PMC7811610  PMID: 33144474

Abstract

Background/Aim: The wearable cardioverter/defibrillator (WCD) is recommended to prevent sudden cardiac death (SCD). Guidelines suggest a 90 days’ period, but prolongation of WCD wear time until increasing the ejection fraction (≥35%) might be suggested. Patients and Methods: A cohort of 153 patients with prescribed WCD were divided into two groups: A <90 wear days’ group (n=112) vs. ≥90 wear days’ group (n=41) and followed. Results: In the first group, WCD shock occurred in 3.6% of patients, 47.3% improved in left ventricular ejection fraction (LVEF) after 3 months, and 37.5% had a cardiac implantable electronic device (CIED) implantation with appropriate implantable cardioverter defibrillator (ICD) shock events occurring in 6 patients. Two of these patients already received WCD shock therapy due to ventricular fibrillation. A 20.5% improved in LVEF after 6-12 months, but 73% were already implanted with ICD. In the second group, 4.9% received WCD shock, 34.1% improved in LVEF after 3 months, 48.8% were implanted with ICD, and 2 had ICD shocks during follow up time. LVEF improvement after 6-12 months occurred in 26.8%. ICD implantation was prevented in 7.3% of patients due to LVEF recovery. Conclusion: Prolonging wearing days of WCD may reduce the number of inappropriate ICD implantation.

Keywords: Wearable cardioverter defibrillator (WCD), heart insufficiency, sudden cardiac death


The Wearable cardioverter defibrillator (WCD) may be considered for patients with low (<35%) left ventricular ejection fraction (LVEF) who are at risk of sudden cardiac death (SCD), but implantable cardioverter defibrillator (ICD) implantation is not yet indicated (1). These can be patients with transient or unknown risk for ventricular tachyarrhythmia and SCD (1,2).

Because many patients recover after diagnosis of heart failure with severely low LVEF <35% and high risk of SCD, WCD is most commonly prescribed for 90 days (3-5). During this time, physicians should evaluate heart failure therapy and treat patients with optimal medical therapy (OMT) according to current European Society of Cardiology (ESC) guidelines for heart failure (6). Different studies have shown that early ICD implantation might not improve the outcome of patients with ischemic (ICM) and non-ischemic cardiomyopathy (NICM) (7,8).

In patients with infected cardiac implantable electronic device (CIED) who are still at risk of arrhythmic events, WCD has demonstrated high efficacy and can be used as a bridge until ICD can be re-implanted (9).

A prophylactic ICD implantation among patients after acute myocardial infarction (MI) and high risk of SCD did not reduce overall mortality (7,10). Results of the recent VEST trial showed that WCD did not lead to a significantly lower rate of arrhythmic deaths among patients with a recent MI and LVEF <35% (11). ICD implantation for primary prevention is therefore still a difficult topic especially in patients with newly diagnosed NICM (12).

The goal of this study was to compare and analyze differences between less and equal or more than 90 wear days of WCD by looking at (a) baseline characteristics and disease etiology, (b) ICD implantation vs. no ICD implantation and (c) improvement of LVEF.

Patients and Methods

We included all patients (n=153) who were prescribed with a WCD between April 2012 and March 2019 at Mannheim University Hospital (Mannheim, Germany). The study was approved by the local ethics committee and conforms to the 1975 Declaration of Helsinki.

WCD is an easy to handle, not invasive and reversible medical device that can detect and terminate ventricular tachycardia (VT) or ventricular fibrillation (VF) (13,14). If an arrhythmic event such as VT or VF occurs, WCD first alarms the patient with vibration and acoustic and optical signals and then treats the patient automatically with up to 5 electric shocks using 150 Joules, if not inhibited by the patient pressing two buttons on the WCD (15). Indications for WCD use were patients at risk of SCD like myocarditis, ICD or cardiac resynchronization therapy (CRT) defibrillator explantation due to infection, newly diagnosed ICM with low LVEF (<35%), newly diagnosed non-ischemic cardiomyopathy (NICM) with LVEF <35% or congenital diseases (HCM/channelopathies) (Figure 1). Patients received a ZOLL Life Vest TM system (LifeVest, ZOLL, Pittsburgh, PA, USA) and OMT according to current heart failure guidelines (16). In general, WCD was prescribed for 3 months irrespective of the underlying WCD indication described above. Data were prospectively collected. All patients provided consent for collection and analysis of standard clinical data.

Figure 1. Indications for wearable cardioverter defibrillator (WCD) use.

Figure 1

Baseline characteristics were obtained at time of heart failure diagnosis and WCD implementation including clinical parameters, medical history and cardiovascular risk factors. We also obtained patient’s medication record. Each patient was followed up after 3 months (short-term follow up) and after 6-12 month (long-term follow up) or when clinically indicated. At each examination patient received an electrocardiogram and also an echocardiogram. The biplane Simpson’s method was used to calculate LVEF. Improvement was defined as an increase to LVEF over 35%. Reasons for stopping WCD use were either improved LVEF, device implantation, incompliance, death or others.

WCD data were also taken from the manufacturer’s online platform (LifeVest Network, ZOLL) including alert records, wear days and daily wear time.

We divided patients into two groups: patients with less than 90 wear days of WCD (n=112, <90 wear days’ group) and patients with equal or more than 90 wear days (n=41, ≥90 wear days’ group). After that, we analyzed if patients had ICD implantation or LVEF improvement and compared disease etiology. We used a method similar to the PROLONG study and related a prolongation of WCD wearing days to a) a relevant increase in LVEF with potential to more improvement, especially in cases with LVEF between 30% to 35%; and b) lack of OMT of heart failure (4).

Each individual alert record, patient-initiated recording or treatment was reviewed and classified into the following categories: appropriate shock, inappropriate shock, inhibition of shock. An arrhythmic episode was divided into sustained VT lasting 30 s or longer with WCD shock therapy, VF with WCD shock therapy, others with no WCD therapy or none. There were no bradyarrhythmic or asystole events and therefore, they were not categorized in our data table.

Statistical analysis. All statistical analyses were performed using IBM SPSS Statistics for Macintosh (Version 25.0. Armonk, NY, USA: IBM Corp.). Baseline characteristics were compared between patients that used WCD for less than 90 days and ≥90 days. Data are presented as mean±standard deviation or median (interquartile range) for continuous variables or as a number of cases for categorical variables expressed as frequencies and percentages. A paired t-test was performed for comparison of continuous variables and a chi-squared test was used for comparison of categorical variables. A two-tailed p-value of <0.05 was considered statistically significant.

Results

WCD Data. A total of 153 patients were prescribed WCD use. Forty-one of these patients (27%) used this device for exactly or more than 90 days. Thirty patients (73% of those prescribed WCD ≥90 days) wore the WCD for 90-119 days and 11 patients (27%) from 120 to a maximum of 201 days. Median in the <90 wear days’ group was 48.5 days [interquartile range (IQR)=24.25-64] and 104 (IQR=94-139.5) days in the other one. The average daily wear time in the ≥90 wear days’ group was greater than in the <90 wear days’ group (22.38±2 vs. 21.11±4 h per day, p<0.007).

Baseline characteristics (Table I). Looking at the indications for WCD use, most of the patients in both groups had a newly diagnosed NICM (43.8% in the <90 wear day’s group vs. 51.2% in the ≥90 wear day’s group, p<0.411). Second most frequent indication was newly diagnosed ICM (37.5% vs. 34.1%, p<0.974) followed by ICD/cardiac resynchronization therapy defibrillator (CRT- D) explantation (11.6% vs. 7.3%, p<0.442) and myocarditis (4.5% vs. 7.3%, p<0.483). There were three patients with channelopathies but only in the <90 wear days’ group. None of them received an CIED implantation.

Table I. Baseline characteristics of 153 patients with wearable cardioverter defibrillator (WCD) in comparison of less and equal to or more than 90 wear days.

graphic file with name in_vivo-34-3603-i0001.jpg

ICD: Implantable cardioverter defibrillator; CIED: cardiac implantable electronic device; CRT: cardiac resynchronization therapy; LVEF: Left ventricular ejection fraction; ICM: ischemic cardiomyopathy; NICM: non-ischemic cardiomyopathy; VT: ventricular tachycardia; VF: ventricular fibrillation; CAD: coronary artery disease; CABG: coronary artery bypass graft; CKD: chronic kidney disease; CHF: congestive heart failure; TIA: transient ischemic attack; COPD: chronic obstructive pulmonary disease; BMI: body-mass-index; CVD: cardiovascular disease; CRT-D: cardiac resynchronization therapy-defibrillator; SD: standard deviation

Follow up clinical data (Table II). There were four patients in the <90 wear days’ group who had a pacemaker rhythm at baseline (3.6% vs. 0% in the ≥90 wear days’ group, p<0.045). After 6-12 months, elevation of Pro-BNP occurred numerically more often in the ≥90 wear days’ group than in the other group (41.5% vs. 23.2%, p<0.041) with a greater median of BNP value (720 vs. 334.5, p<0.179).

Table II. Comparison of follow up clinical data of patients with <90 wear days of the wearable cardioverter defibrillator (WCD) versus ≥90 wear days.

graphic file with name in_vivo-34-3604-i0001.jpg

LVEF: Left ventricular ejection fraction; ECG: electrocardiography; BNP: brain natriuretic peptide; AV-Block: atrioventricular-block. 1number of patients if not mentioned otherwise; 2p-value for comparison of <90 wear days versus ≥90 wear days

WCD arrhythmia events (Table III). We could not find any statistically relevant differences between the two groups looking at arrhythmic events during WCD use. Most of the patients had no arrhythmic events (93.8% and 92.7%, p<0.813). There were 4 appropriate shocks in the <90 wear days’ group and 2 shocks in the ≥90 wear days’ group (3.6% vs. 4.9%, p<0.712). Only one of these shocks happened after 90 wear days. This patient was prescribed WCD due to recurrent VTs, but his health status became worse when his knee replacement surgery resulted in infection with decompensated global heart failure and pleural effusions. LVEF improved from 20% to only 26% in the first 3 months, but remained at this level during follow up. It took 159 WCD wear days until ICD implantation was possible. WCD shock occurred after 151 days.

Table III. Follow up data of patients after wearable cardioverter defibrillator (WCD) use in comparison of <90 wear days versus ≥90 wear days.

graphic file with name in_vivo-34-3605-i0001.jpg

LVEF: Left ventricular ejection fraction; LBBB: left bundle branch block; RBBB: right bundle branch block; ICD: implantable cardioverter defibrillator; CRT: cardiac resynchronization therapy; VT: ventricular tachycardia; VF: ventricular fibrillation; SD: standard deviation; IQR: interquartile range; CRT-D: cardiac resynchronization therapy-defibrillator; AV-Block: atrioventricular-block. 1p-value in comparison of <90 wear days versus ≥90 wear days; 22/87 patients died due to septic shock and terminal heart failure, 2/87 denied implantation, 4/153 patients lost to follow up; 3over mean follow up of 36.2±15.6 months, more than one reason possible

None of the patients in our cohort experienced bradyarrhythmic or asystole events, but one patient was falsely treated with an inappropriate shock. In this event, WCD misinterpreted artefactual voltage fluctuations as VT and therefore, activated its tactile and audible alarms, but the patient was not able to push the response button and therefore, could not save himself from the electric shock.

Figure 2 shows shock events in both groups.

Figure 2. Comparison of wear days and wearable cardioverter defibrillator (WCD) shock therapy.

Figure 2

Echocardiographic data. While both groups improved significantly in LVEF during follow up from mean 28.6±10.15% at baseline to 36.94±11.34% after 3 months (p<0.001) and to 40.96±12.68% after 6-12 months (p<0.001), there were no significant differences between the groups at baseline (<90 wear days: 29.73±11% vs. 27.10±10% ≥90 wear days, p<0.170), after 3 months follow up (<90 wear days: 38.43±11% vs. 35.33±12% ≥90 wear days, p<0.150) and after 6-12 months (<90 wear days: 41.97±12% vs. 41.35±12% ≥90 wear days, p<0.789).

In the <90 wear days’ group LVEF did not change during follow up in 34 patients vs. 15 patients in the other group (30.4% vs. 36.6%, p<0.465), declined in 2 patients (1.8%) vs. 1 patient in the ≥90 wear days’ group (2.4%, p<0.796), but improved in 53 patients (47.3%) vs. 14 (34.1%, p<0.146) after 3 months and 24 patients (21.4%) vs. 11 (26.8%, p<0.407) after 6-12 months.

WCD prolongation outcome (Figure 3). Forty-one out of 153 patients (26.8%) had a prolonged WCD use for ≥90 wear days and improved in mean±SD LVEF from 27.10±10% at baseline to 41.34±12% after 6-12 months. Almost half of these patients were diagnosed with NICM (n=20, 48.8%). Twenty-seven patients did not improve in LVEF after 3 months follow up (65.9%), but 11 patients did so after 6-12 months (26.8%).

Figure 3. Comparison of wear days and device implantation.

Figure 3

Eight patients received ICD implantation because LVEF did not improve during the first 3 months, but improved after 6-12 months when ICD was already implanted. All of them had NICM. In fact, from all 20 NICM patients with extended WCD use, 75% improved in LVEF either after 3 months (n=5, 25%) or after 6-12 months (n=10, 50%).

Three patients with prolonged WCD wear time improved in LVEF after 6-12 months and did not receive an ICD.

Device implantation. As shown in Figure 3, there is no significant difference in LVEF improvement if the wear days’ groups are further divided into no device implantation and device implantation, however, looking at an etiology, there is a higher number of patients with device implantation in the ≥90 WCD wear days’ group that were prescribed WCD use because of newly diagnosed NICM than in the device implantation and <90 wear days’ group (60% vs. 28.5%, p<0.018).

ICD events. Six patients from the <90 wear days’ group with ICD implantation were treated by their device due to VT/VF, and two of them had already received WCD shocks. In the other group, there were two documented VT/VF events during mean follow up, also successfully treated by ICD (14.3% <90 wear days’ group vs. 10% ≥90 wear days’ group, p=0.638).

Death during follow up. Mean follow up time was 36.2±15.6 months (median: 34, IQR=24-46 months). Fourteen patients with less than 90 wear days and 1 patient with more than 90 wear days died during this time (12.5% vs. 2.4%, p<0.064). In detail, in the first group 7 patients died without and 7 with ICD implantation. In the ≥90 wear days’ group, 1 patient died with device implantation. Reason for death is not known in most of the patients.

Comparison by disease etiology NICM. In detail, there were 69 patients with newly diagnosed non-ischemic cardiomyopathy.

Forty-nine patients with <90 WCD wear days (71%). Seven patients (10%) did not improve in LVEF and either received ICD (n=4) or denied implantation (n=3). Forty-two patients (61%) improved in LVEF after 3 months (n=29, 42%) or after 6-12 months (n=13, 19%). Nine of the 13 patients (29%) received device implantation, leaving 4 patients with late LVEF improvement, but still without CIED. Four/49 patients from this group died during follow up. One patient with device implantation and one reported successful ICD shock. Reason for death is not known.

We documented 20 WCD patients with ≥90 WCD wear days (29%). Five of them showed little LVEF improvement, but never reached more than 35%. Four of them were implanted with CIED and 1 patient was lost to long-term follow up. Six patients improved in LVEF after 3 months. One patient with LVEF of 36% received CRT-D implantation with one reported show during follow up. Nine patients improved in LVEF after 6-12 months (13%) with 8 of 9 ICD implantations.

From the second group, no patient died during follow up (1 patient lost to follow up). There was 1 patient who received an inappropriate ICD shock due to t-wave oversensing (falsely detected as VT) and 1 patient with documented non-sustained VT during WCD wear time.

Discussion

Our study shows that a median LVEF improvement occurred in 66% of patients prescribed WCD. If worn for more than 90 days, more than every fourth patient (26.8%) improved in LVEF. Therefore, a prolongation of WCD use may prevent patients from inappropriate ICD implantation, but, nevertheless, a longer WCD wear time could imply that patients start to become impatient or bored with a decrease in wear time (hours per day), which could then influence the patient’s outcome.

The VEST study showed a mean wear time of 14.0±9.3 h in the device group and therefore, WCD did not show a significantly lower rate of arrhythmic deaths in comparison to only medical therapy (11). Subanalysis confirmed that when patients were wearing the WCD, a significantly lower percentage died than when patients did not wear the WCD. This emphasizes the importance of patient’s compliance and wear time and that it has direct influence on his or her outcome. A recently published study showed predictors for WCD compliance such as cardiac arrest during myocardial infarction, diabetes or prior heart failure. Whereas factors like divorce, Asian race, high body mass index or WCD shock lead to worse WCD compliance (17). These predictors may be useful in evaluating whether or not to extend WCD use.

Although patients in our cohort with more than 90 wear days had a statistically significant better daily wear time than patients with less than 90 wear days, the rate of arrhythmic deaths in comparison of both groups is not quite statistically significant (p=0.064), but overall survival is high during long-term follow up.

In our cohort, we described patients with newly diagnosed NICM in more detail due to the fact, that prolongation of WCD wear time in NICM patients is still debatable. Twelve of 69 NICM patients (17%) did not improve in LVEF during our follow up study. This means that 83% (n=57) did improve over time. Nevertheless, 27 patients (39%) already received ICD implantation. All NICM patients with prolonged WCD wear time and ICD implantation had no documented appropriate arrhythmic episodes during follow up.

Two patients that had worn WCD for more than 90 days, without WCD shocks, but subsequently had ICD implantation, showed adequate ICD shock events during follow up and were therefore prevented from SCD. Therefore, WCD might be helpful for risk stratification in patients with an impaired EF.

Every patient who received WCD shock therapy in our cohort was implanted with ICD. During the mean follow up period (36.2±15.6 months), two of them showed arrhythmic events, such as VT/VF after WCD wear time. Because of ICD implantation, they were successfully treated through defibrillation. Nevertheless, to evaluate whether ICD implantation or a prolonged WCD use is more secure, further randomized and double blinded studies are necessary. Of note, WCD shocks occurred in two patients of the ≥90 wear days’ group (1 shock after 151 days), both with ICD implantation after WCD use but without further arrhythmic events during follow up. In addition, three patients with myocarditis and prolonged WCD wear time improved in LVEF and were not implanted with ICD. All 3 survived during mean follow up.

Looking at cost effectiveness, studies have already shown that WCD likely is cost-effective in protecting patients from SCD after removal of infected ICD, while waiting for reimplantation (18,19). In our cohort, further studies need to show if a prolongation of WCD wear time for more than 90 days has a major impact for insurance companies, as long as patients can be at home instead of in a hospital and if a prevention of ICD implantation can result in fewer primary and secondary costs for the health care system.

In addition to a prolongation of WCD wear days there could also be an improvement in OMT. A study has shown that patients with reverse remodeling improved in LVEF due to optimization of heart failure medication (4). Further studies are needed to show whether time, i.e. more than 90 days, or dosage lead to LVEF improvement. Of note, sacubitril-valsartan has shown optimistic effects in heart failure patients (20). In the PARADIGM trial, data have shown that in addition to the lower mortality rate and re-hospitalization rate the risk of SCD was decreased. Therefore, testing the concomitant effect of sacubitril/valsartan with WCD as compared to the historical treatment with ACE-inhibitor and/or AT II antagonists might be a good approach to reduce the number of electronic device implantation, its complications and subsequent costs.

Limitations

This study is only a single center study and the number of patients is too small to find significant predictors for LVEF recovery or arrhythmias. Patients were not randomized and there was no control group.

Conclusion

While two third of our patients recovered from low LVEF to >35%, still 40% were implanted with ICD. If WCD was worn for equal to or more than 90 days, every fourth patient recovered from low LVEF. Nevertheless, patients are still at risk of arrhythmic events and should therefore decide carefully whether to prolong WCD wear time or get ICD implantation to prevent them from sudden cardiac death. Overall, prolonging wearing days of WCD for more than 90 days (3 months) does not statistically reduce the number of untimely implanted ICDs.

Conflicts of Interest

All Authors declare that there are no conflicts of interest regarding this study.

Authors’ Contributions

TD and IE wrote the manuscript and were involved in statistical analysis and design of the study. IE, SR, SLR, SG and JK collected data. IA and MB corrected and revised the manuscript critically. All Authors read and approved the final manuscript.

References

  • 1.Priori SG, Blomstrom-Lundqvist C, Mazzanti A, Blom N, Borggrefe M, Camm J, Elliott PM, Fitzsimons D, Hatala R, Hindricks G, Kirchhof P, Kjeldsen K, Kuck KH, Hernandez-Madrid A, Nikolaou N, Norekval TM, Spaulding C, Van Veldhuisen DJ, Group ESCSD. 2015 esc guidelines for the management of patients with ventricular arrhythmias and the prevention of sudden cardiac death: The task force for the management of patients with ventricular arrhythmias and the prevention of sudden cardiac death of the european society of cardiology (esc). Endorsed by: Association for european paediatric and congenital cardiology (aepc) Eur Heart J. 2015;36(41):2793–2867. doi: 10.1093/eurheartj/ehv316. [DOI] [PubMed] [Google Scholar]
  • 2.Piccini JP Sr., Allen LA, Kudenchuk PJ, Page RL, Patel MR, Turakhia MP, American Heart Association Electrocardiography and Arrhythmias Committee of the Council on Clinical Cardiology and Council on Cardiovascular and Stroke Nursing Wearable cardioverter-defibrillator therapy for the prevention of sudden cardiac death: A science advisory from the american heart association. Circulation. 2016;133(17):1715–1727. doi: 10.1161/CIR.0000000000000394. [DOI] [PubMed] [Google Scholar]
  • 3.Brooks GC, Lee BK, Rao R, Lin F, Morin DP, Zweibel SL, Buxton AE, Pletcher MJ, Vittinghoff E, Olgin JE, PREDICTS Investigators Predicting persistent left ventricular dysfunction following myocardial infarction: The predicts study. J Am Coll Cardiol. 2016;67(10):1186–1196. doi: 10.1016/j.jacc.2015.12.042. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4.Duncker D, Konig T, Hohmann S, Bauersachs J, Veltmann C. Avoiding untimely implantable cardioverter/defibrillator implantation by intensified heart failure therapy optimization supported by the wearable cardioverter/defibrillator-the prolong study. J Am Heart Assoc. 2017;6(1) doi: 10.1161/JAHA.116.004512. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5.Kutyifa V, Moss AJ, Klein H, Biton Y, McNitt S, MacKecknie B, Zareba W, Goldenberg I. Use of the wearable cardioverter defibrillator in high-risk cardiac patients: Data from the prospective registry of patients using the wearable cardioverter defibrillator (wearit-ii registry) Circulation. 2015;132(17):1613–1619. doi: 10.1161/CIRCULATIONAHA.115.015677. [DOI] [PubMed] [Google Scholar]
  • 6.McMurray JJ, Adamopoulos S, Anker SD, Auricchio A, Bohm M, Dickstein K, Falk V, Filippatos G, Fonseca C, Gomez-Sanchez MA, Jaarsma T, Kober L, Lip GY, Maggioni AP, Parkhomenko A, Pieske BM, Popescu BA, Ronnevik PK, Rutten FH, Schwitter J, Seferovic P, Stepinska J, Trindade PT, Voors AA, Zannad F, Zeiher A, Task Force for the D, Treatment of A, Chronic Heart Failure of the European Society of C, Bax JJ, Baumgartner H, Ceconi C, Dean V, Deaton C, Fagard R, Funck-Brentano C, Hasdai D, Hoes A, Kirchhof P, Knuuti J, Kolh P, McDonagh T, Moulin C, Popescu BA, Reiner Z, Sechtem U, Sirnes PA, Tendera M, Torbicki A, Vahanian A, Windecker S, McDonagh T, Sechtem U, Bonet LA, Avraamides P, Ben Lamin HA, Brignole M, Coca A, Cowburn P, Dargie H, Elliott P, Flachskampf FA, Guida GF, Hardman S, Iung B, Merkely B, Mueller C, Nanas JN, Nielsen OW, Orn S, Parissis JT, Ponikowski P, ESC Committee for Practice Guidelines ESC guidelines for the diagnosis and treatment of acute and chronic heart failure 2012: The task force for the diagnosis and treatment of acute and chronic heart failure 2012 of the european society of cardiology. Developed in collaboration with the heart failure association (hfa) of the esc. Eur J Heart Fail. 2012;14(8):803–869. doi: 10.1093/eurjhf/hfs105. [DOI] [PubMed] [Google Scholar]
  • 7.Steinbeck G, Andresen D, Seidl K, Brachmann J, Hoffmann E, Wojciechowski D, Kornacewicz-Jach Z, Sredniawa B, Lupkovics G, Hofgartner F, Lubinski A, Rosenqvist M, Habets A, Wegscheider K, Senges J, Investigators I. Defibrillator implantation early after myocardial infarction. N Engl J Med. 2009;361(15):1427–1436. doi: 10.1056/NEJMoa0901889. [DOI] [PubMed] [Google Scholar]
  • 8.Sheppard R, Mather PJ, Alexis JD, Starling RC, Boehmer JP, Thohan V, Pauly DF, Markham DW, Zucker M, Kip KE, McNamara DM, Investigators I. Implantable cardiac defibrillators and sudden death in recent onset nonischemic cardiomyopathy: Results from imac2. J Card Fail. 2012;18(9):675–681. doi: 10.1016/j.cardfail.2012.07.004. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9.Ellenbogen KA, Koneru JN, Sharma PS, Deshpande S, Wan C, Szymkiewicz SJ. Benefit of the wearable cardioverter-defibrillator in protecting patients after implantable-cardioverter defibrillator explant: Results from the national registry. JACC Clin Electrophysiol. 2017;3(3):243–250. doi: 10.1016/j.jacep.2016.09.002. [DOI] [PubMed] [Google Scholar]
  • 10.Hohnloser SH, Kuck KH, Dorian P, Roberts RS, Hampton JR, Hatala R, Fain E, Gent M, Connolly SJ, Investigators D. Prophylactic use of an implantable cardioverter-defibrillator after acute myocardial infarction. N Engl J Med. 2004;351(24):2481–2488. doi: 10.1056/NEJMoa041489. [DOI] [PubMed] [Google Scholar]
  • 11.Olgin JE, Pletcher MJ, Vittinghoff E, Wranicz J, Malik R, Morin DP, Zweibel S, Buxton AE, Elayi CS, Chung EH, Rashba E, Borggrefe M, Hue TF, Maguire C, Lin F, Simon JA, Hulley S, Lee BK, Investigators V. Wearable cardioverter-defibrillator after myocardial infarction. N Engl J Med. 2018;379(13):1205–1215. doi: 10.1056/NEJMoa1800781. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12.Haugaa KH, Dan GA, Iliodromitis K, Lenarczyk R, Marinskis G, Osca J, Scherr D, Dagres N. Management of patients with ventricular arrhythmias and prevention of sudden cardiac death-translating guidelines into practice: Results of the european heart rhythm association survey. Europace. 2018;20(FI2):f249–f253. doi: 10.1093/europace/euy112. [DOI] [PubMed] [Google Scholar]
  • 13.Barraud J, Cautela J, Orabona M, Pinto J, Missenard O, Laine M, Thuny F, Paganelli F, Bonello L, Peyrol M. Wearable cardioverter defibrillator: Bridge or alternative to implantation. World J Cardiol. 2017;9(6):531–538. doi: 10.4330/wjc.v9.i6.531. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 14.Rosenkaimer SL, El-Battrawy I, Dreher TC, Gerhards S, Roger S, Kuschyk J, Borggrefe M, Akin I. The wearable cardioverter-defibrillator: Experience in 153 patients and a longterm follow-up. J Clin Med. 2020;9(3) doi: 10.3390/jcm9030893. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15.Duncker D, Veltmann C. The wearable cardioverter/ defibrillator - toy or tool? J Atr Fibrillation. 2016;8(6):1367. doi: 10.4022/jafib.1367. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 16.Priori SG, Blomstrom-Lundqvist C, Mazzanti A, Blom N, Borggrefe M, Camm J, Elliott PM, Fitzsimons D, Hatala R, Hindricks G, Kirchhof P, Kjeldsen K, Kuck KH, Hernandez-Madrid A, Nikolaou N, Norekval TM, Spaulding C, Van Veldhuisen DJ. 2015 ESC guidelines for the management of patients with ventricular arrhythmias and the prevention of sudden cardiac death. Rev Esp Cardiol (Engl Ed) 2016;69(2):176. doi: 10.1016/j.rec.2016.01.001. [DOI] [PubMed] [Google Scholar]
  • 17.Olgin JE, Lee BK, Vittinghoff E, Morin DP, Zweibel S, Rashba E, Chung EH, Borggrefe M, Hulley S, Lin F, Hue TF, Pletcher MJ. Impact of wearable cardioverter-defibrillator compliance on outcomes in the vest trial: As-treated and per-protocol analyses. J Cardiovasc Electrophysiol. 2020;31(5):1009–1018. doi: 10.1111/jce.14404. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18.Jiang X, Ming WK, You JHS. Potential cost-effectiveness of wearable cardioverter-defibrillator for patients with implantable cardioverter-defibrillator explant in a high-income city of china. J Cardiovasc Electrophysiol. 2019;30(11):2387–2396. doi: 10.1111/jce.14153. [DOI] [PubMed] [Google Scholar]
  • 19.Healy CA, Carrillo RG. Wearable cardioverter-defibrillator for prevention of sudden cardiac death after infected implantable cardioverter-defibrillator removal: A cost-effectiveness evaluation. Heart Rhythm. 2015;12(7):1565–1573. doi: 10.1016/j.hrthm.2015.03.061. [DOI] [PubMed] [Google Scholar]
  • 20.McMurray JJ, Packer M, Desai AS, Gong J, Lefkowitz MP, Rizkala AR, Rouleau JL, Shi VC, Solomon SD, Swedberg K, Zile MR, PARADIGM-HF Investigators and Committees Angiotensin-neprilysin inhibition versus enalapril in heart failure. N Engl J Med. 2014;371(11):993–1004. doi: 10.1056/NEJMoa1409077. [DOI] [PubMed] [Google Scholar]