Abstract.
Significance: Uncontrolled cryoablation of tissues is a strong reason limiting the wide application of cryosurgery and cryotherapy due to the certain risks of unpredicted damaging of healthy tissues. The existing guiding techniques are unable to be applied in situ or provide insufficient spatial resolution. Terahertz (THz) pulsed spectroscopy (TPS) based on sensitivity of THz time-domain signal to changes of tissue properties caused by freezing could form the basis of an instrument for observation of the ice ball formation.
Aim: The ability of TPS for in situ monitoring of tissue freezing depth is studied experimentally.
Approach: A THz pulsed spectrometer operated in reflection mode and equipped with a cooled sample holder and ex vivo samples of bovine visceral adipose tissue is applied. Signal spectrograms are used to analyze the changes of THz time-domain signals caused by the interface between frozen and unfrozen tissue parts.
Results: Experimental observation of TPS signals reflected from freezing tissue demonstrates the feasibility of TPS to detect ice ball formation up to depth.
Conclusions: TPS could become the promising instrument for in situ control of cryoablation, enabling observation of the freezing front propagation, which could find applications in various fields of oncology, regenerative medicine, and THz biophotonics.
Keywords: cryoablation, cryotherapy, terahertz biophotonics, tetrahertz pulsed spectroscopy, ice ball formation
1. Introduction
Terahertz (THz) pulsed spectroscopy (TPS) is known as a promising tool of biophotonics and medical diagnosis.1–4 It demonstrates strong potential of study and differentiate various conditions of tissues, both ex vivo and in vivo,5–9 and imaging and characterization of neoplasms in different localizations,10–14 by means of endogenous markers of pathology, such as water content. However, high concentration of free and bound water in tissues restricts the penetration depth of THz waves by only several hundreds of microns.1 To overcome this limitation, different approaches can be applied. Among them are such techniques as tissue dehydration,15 paraffin-embedding,16 lyophilization,17 compression,18 immersion optical clearing,19,20 and freezing.21–24 It was demonstrated that the latter one helps to increase the THz-wave penetration up to 1 mm, changing at the same time dielectric properties of tissues in THz range.23 Along with reduction of absorption, low temperatures alter the refractive index within certain limits, thus, yielding the dielectric contrast between tissues in different states. This opens further horizons of TPS applications, in particular, for monitoring of freezing depth in tissues during their cryosurgery.
Uncontrolled cryoablation of living tissues is a severe problem of modern cryosurgery, i.e., application of extreme cold aimed at rapid freezing and thus destroying or renewal of pathological tissues.25–29 Unless cryoablation possesses several advantages, such as relative painlessness, hemostatic effect, short recovering of patients, and immunostimulating effect,25,30,31 it is associated with certain risks of damaging healthy tissues surrounding the pathology and with possibility of incomplete cell death.28 In particular, the use of cryosurgery for the treatment of brain tumors and epilepsy is currently limited by the possible risks for the patient.32–34 Such guiding techniques as ultrasonography,35,36 magnetic resonance, and computer imaging37–42 do not enable the monitoring of the tissue freezing in situ with high spatial resolution near the cryoprobe. Alternative approach is based on measuring of tissue temperature using thermocouples (TCs) combined with cryoprobe.43 However, it allows controlling of the tissue condition only at the cryoprobe–tissue interface. Therefore, cryosurgery still needs guiding approaches for the efficient control of ice ball formation.
In this work, we study the ability of TPS to solve this problem. We experimentally demonstrate the changes of THz time-domain signal during tissue freezing, using ex vivo bovine adipose specimen and laboratory spectrometer equipped with special sample holder. Processing of THz pulsed signal within 45 s demonstrates the ability to detect the movement of freezing front up to the depth of . The results of this work reveal the potential of TPS to control the initial stage of ice ball formation during tissue cryoablation.
2. Experimental Setup
To experimentally test the feasibility of TPS to detect the freezing depth in biological tissues, we use a laboratory THz pulsed spectrometer, which was described in detail in Ref. 44. It operates in a reflection mode and has a maximum spectral operating range from about 0.05 to 4.0 THz and a maximum spectral resolution of about 0.002 THz. The THz wave features s-polarization. To avoid the impact of water vapors on the measurements, the spectrometer is purged with nitrogen gas. It is also equipped with a sample holder (Fig. 1). The latter consists of a reference window made of sapphire, which on the one hand is placed on the THz beam path in contact with the tissue sample [see Fig. 1(a)]. On the other hand, it is inserted in the metal frame connected with the reservoir filled with liquid nitrogen prior to the measurements [Fig. 1(b)]. Thus, due to high thermal conductivity of metal and sapphire, the window possesses low and stable temperature during the measurements.
As a tissue sample, we use ex vivo bovine visceral adipose tissue because of its relatively homogeneous structure. The sample thickness is near 30 mm enabling prolonged duration of the freezing process. By two off-axis parabolic mirrors, the incident THz beam is focused on the interface between the reference window and the tissue; the reflected beam is collimated to the detector. Generation and detection of THz pulses are provided by two photoconductive antennas (PCAs). A typical form of the reflected THz time-domain signal is shown in Fig. 1(c). The sequence of pulses includes signal (A) reflected from free space–window interface and signal (B) reflected from the window–sample interface. Since we are interested in the latter one, the scanning range of the spectrometer is adjusted at this pulse (marked with red rectangle).
The measurements are performed at room temperature 24°C. However, the sample is preliminary heated to the temperature 37°C aimed at the slowing down the initial stage of freezing. After the contact between the tissue and the cooled window, THz pulsed signal is constantly recorded with the period of 15 s until the sample is completely frozen. For additional control of sample freezing, its temperature is measured by the inserted TCs.
3. THz Pulsed Spectroscopy of a Freezing Tissue
3.1. THz Dielectric Properties of a Sample
The sample refractive index was characterized before and after freezing. For this purpose, the complex amplitude of the incident THz field is considered to have the form , where and are the amplitude and phase of the THz field, respectively. Assuming the negligible losses in the reference window and the normal incidence of THz beam, the signal amplitudes and of pulses reflected from the first and second interfaces [see wavelets (A) and (B) in Fig. 1(c)] are described by the Fresnel coefficients
(1) |
(2) |
where and are the Fresnel coefficients of amplitude reflection and transmission between ’th and ’th media, which for s-polarized wave are determined by the corresponding refractive indices and
(3) |
(4) |
In this work, refractive indices for free space and reference window are and , respectively, at 1-THz frequency.45 Estimating the relation from the measured TPS waveforms and accounting the dispersion relation, the values of the sample refractive index at 1-THz frequency were determined for two conditions of the tissue, i.e., at room temperature and after freezing. Thus, freezing of the adipose tissue leads to significant changes of its refractive index. The value is used further for estimation of the freezing depth according to the relation , where is the temporal position of the detected freezing depth, is speed of light in free space.
3.2. Analysis of THz Signal Spectrograms
According to the measurement of sample temperature, the freezing depth of 4 mm was reached in . This depth significantly exceeds the typical values of THz penetration depth in frozen biological tissues.23 Thus, only several THz waveforms can be recorded during this stage of the analyzed process. To increase the number of detected signals, the accumulation time and discretization of the spectrometer should be reduced.
Figure 2 shows the sequence of THz waveforms detected at (I) 0 s, (II) 15 s, (III) 30 s, and (IV) 45 s from the beginning of the sample cooling. At the initial moment of the ice front propagation, the temporal width of first THz pulse, caused by the window–tissue interface, is about 5 ps, which corresponds to of the sample depth. Within this tissue layer, the second pulse caused by the existence of the ice front interface, is probably overlapped by the first one. Its amplitude is expected to be significantly smaller, since the dielectric contrast between frozen and unfrozen tissue is lower than that for the tissue and the reference window. Thus, it is difficult to estimate directly from the THz waveforms the moving position of the second pulse.
To overcome this problem, we analyze the signal intensity spectrograms in addition to their temporal waveforms. Figures 3(a)–3(d) show the spectrograms of the measured signals shown in the insets [Figs. 3(e)–3(h)]. Comparing them between each other, it is possible to detect the temporal position of the pulse reflected from the ice front. Figures 3(i)–3(k) show the difference between the normalized spectrogram at , 30, and 45 s and the initial normalized spectrogram at . Obviously, the first pulse is altering during the first 30 s, then its appearance remains stable. We can clearly observe the presence of the second pulse at and 30 s, which indicates the freezing depths of and , respectively; then, its amplitude becomes lower due to absorbtion, which does not allow us to detect it. Therefore, during 30 s starting from the tissue cooling, TPS provides an ability to analyze the ice ball formation and estimate the freezing depth. It should be mentioned that time-domain THz signals feature rather weak changes at the determined moments, not allowing us to evaluate them directly. To demonstrate this, the detected positions of the second pulses at and 30 s are marked with red rectangles on the THz waveforms [see Figs. 3(f) and 3(g)]. Due to the absence of second pulses at and 45 s, there are no marks on Figs. 3(e) and 3(h).
4. Results and Discussion
The estimation of the tissue freezing depth is shown in Fig. 4, where the values obtained by TPS are combined with the direct measurements of the sample temperature. Figure 4(b) demonstrates the temperature change at two depths 4 and 11 mm from the window–tissue interface. The slowdown of the freezing process at the beginning can be explained by the initial cooling of the tissue from 37°C to 0°C. In our feasibility study, the maximal freezing depth observed by TPS is limited by the value . However, we believe that it could be increased using THz time-domain spectrometer with more efficient performance, particularly, scanning rate, which would yield the reduction of the detection period and increasing of the temporal resolution.
Though the penetration depth of THz waves even in frozen tissue remains extremely low, the detection of the freezing depth by TPS could provide rather high precision of monitoring ice front propagation, enabling the observation and control of cryoablation and cooling of cell and artificial tissue structures, and agglomeration of cell spheroids. Thus, it could become a beneficial research instrument in regenerative medicine and microbiology.
THz technologies are still far from clinical application, suffering from a list of problems, such as low performance of THz sources and detectors, and low spatial resolution due to high absorbtion and scattering.46,47 However, the rapid progress of THz sources and optical components in the recent years 48–52 stimulates the further improvements of THz instruments for biophotonics. A strong limitation is caused by the absence of efficient waveguides and fibers.53 Nevertheless, the sapphire shaped crystal waveguides, exploiting both anti-resonant reflecting optical and photonic crystal waveguidance principles, have certain perspectives for delivering THz pulses with relatively low losses and dispersion;44,54 thus, they could form the basis for further developments of THz instruments for biophotonics and medicine. In addition, sapphire features high thermal conductivity at cryogenic temperatures, being more efficient material for cryosurgery than metal (copper, titanium nickelide, and stainless steel), since it enables higher tissue freezing rate.53 Combination of these advantages opens future perspectives for the appearance of sapphire probes for controlled cryoablation.
Cryosurgery demonstrates high potential in oncology, in particular, to remove neoplasms and metastases of different localizations, such as liver,37 breast,55 lungs,40prostate,56 kidney,57 stomach,58 and skin.59 Considering the ability of THz technologies for tumor diagnosis along with currently developing THz instruments and methods for healthy and pathological tissues differentiation,2 the combination of THz preoperative characterization of tissue with further TPS-controlled cryoablation could become an effective tool for oncology.
It is worth mentioning that several techniques enable increasing of low penetration depth of THz waves into living tissues. Among them are methods of immersion optical clearing,19 which could also be combined with THz freezing depth detection. However, a list of hyperosmotic agents applied for tissue clearing act as cryoprotectors (e.g., glycerol, dimethyl sulfoxide, and propylene glycol), since they substitute the interstitial water. Thus, the particular approaches of their application for increasing of THz-wave penetration into tissues during their freezing, such as concentration of aqueous solutions and exposure time, are the subjects of further studies.
5. Conclusion
The performed experimental study demonstrates the feasibility of in situ observation of the ice ball formation in tissues using TPS instrumentation. Applying the spectrogram analysis of THz time-domain signals recorded in the reflection-mode of THz pulsed spectrometer during tissue freezing, it is possible to detect the positions of the freezing depth up to . We believe that the described approach would be helpful for the development of novel approaches and instruments of controlled cryoablation and would be useful for various tasks of oncology, regenerative medicine, microbiology, and further progress of THz biophotonics.
Acknowledgments
This work was supported by the Russian Science Foundation (RSF), Project No. #19-79-10212.
Biography
Biographies of the authors are not available.
Disclosures
The authors declare no conflict of interest.
Contributor Information
Arsen K. Zotov, Email: AKZotov@hotmail.com.
Arsenii A. Gavdush, Email: arsenii.a.gavdush@gmail.com.
Gleb M. Katyba, Email: micalych@mail.ru.
Larisa P. Safonova, Email: larisa.safonova@gmail.com.
Nikita V. Chernomyrdin, Email: chernik-a@yandex.ru.
Irina N. Dolganova, Email: in.dolganova@gmail.com.
Code, Data, and Materials Availability
All data generated or analyzed during this study are included in this article.
References
- 1.Smolyanskaya O. A., et al. , “Terahertz biophotonics as a tool for studies of dielectric and spectral properties of biological tissues and liquids,” Prog. Quantum Electron. 62, 1–77 (2018). 10.1016/j.pquantelec.2018.10.001 [DOI] [Google Scholar]
- 2.Zaytsev K. I., et al. , “The progress and perspectives of terahertz technology for diagnosis of neoplasms: a review,” J. Opt. 22(1), 013001 (2020). 10.1088/2040-8986/ab4dc3 [DOI] [Google Scholar]
- 3.Son J.-H., Oh S. J., Cheon H., “Potential clinical applications of terahertz radiation,” J. Appl. Phys. 125(19), 190901 (2019). 10.1063/1.5080205 [DOI] [Google Scholar]
- 4.Cheon H., Yang H., Son J., “Toward clinical cancer imaging using terahertz spectroscopy,” IEEE J. Sel. Top. Quantum Electron. 23(4), 1–9 (2017). 10.1109/JSTQE.2017.2704905 [DOI] [Google Scholar]
- 5.Wang J., et al. , “In vivo terahertz imaging to evaluate scar treatment strategies: silicone gel sheeting,” Biomed. Opt. Express 10(7), 3584–3590 (2019). 10.1364/BOE.10.003584 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 6.Hernandez-Cardoso G., et al. , “Terahertz imaging for early screening of diabetic foot syndrome: a proof of concept,” Sci. Rep. 7, 42124 (2017). 10.1038/srep42124 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 7.Bajwa N., et al. , “Non-invasive terahertz imaging of tissue water content for flap viability assessment,” Biomed. Opt. Express 8(1), 460–474 (2017). 10.1364/BOE.8.000460 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8.Cao Y., et al. , “Qualitative and quantitative detection of liver injury with terahertz time-domain spectroscopy,” Biomed. Opt. Express 11(2), 982–993 (2020). 10.1364/BOE.381660 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9.Parrott E. P., et al. , “Terahertz pulsed imaging in vivo: measurements and processing methods,” J. Biomed. Opt. 16(10), 106010 (2011). 10.1117/1.3642002 [DOI] [PubMed] [Google Scholar]
- 10.Li D., et al. , “Detecting melanoma with a terahertz spectroscopy imaging technique,” Spectrochim. Acta A 234, 118229 (2020). 10.1016/j.saa.2020.118229 [DOI] [PubMed] [Google Scholar]
- 11.Oh S., et al. , “Study of freshly excised brain tissues using terahertz imaging,” Biomed. Opt. Express 5(8), 2837–2842 (2014). 10.1364/BOE.5.002837 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12.Gavdush A., et al. , “Terahertz spectroscopy of gelatin-embedded human brain gliomas of different grades: a road toward intraoperative THz diagnosis,” J. Biomed. Opt. 24(2), 027001 (2019). 10.1117/1.JBO.24.2.027001 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13.Ashworth P. C., et al. , “Terahertz pulsed spectroscopy of freshly excised human breast cancer,” Opt. Express 17(15), 12444–12454 (2009). 10.1364/OE.17.012444 [DOI] [PubMed] [Google Scholar]
- 14.Zaytsev K. I., et al. , “Highly accurate in vivo terahertz spectroscopy of healthy skin: variation of refractive index and absorption coefficient along the human body,” IEEE Trans. Terahertz Sci. Technol. 5(5), 817–827 (2015). 10.1109/TTHZ.2015.2460677 [DOI] [Google Scholar]
- 15.He Y., et al. , “Determination of terahertz permittivity of dehydrated biological samples,” Phys. Med. Biol. 62(23), 8882–8893 (2017). 10.1088/1361-6560/aa8ebe [DOI] [PubMed] [Google Scholar]
- 16.Meng K., et al. , “Terahertz pulsed spectroscopy of paraffin-embedded brain glioma,” J. Biomed. Opt. 19(7), 077001 (2014). 10.1117/1.JBO.19.7.077001 [DOI] [PubMed] [Google Scholar]
- 17.Png G., et al. , “The impact of hydration changes in fresh bio-tissue on THz spectroscopic measurements,” Phys. Med. Biol. 53(13), 3501–3517 (2008). 10.1088/0031-9155/53/13/007 [DOI] [PubMed] [Google Scholar]
- 18.Wang J., et al. , “THz in vivo measurements: the effects of pressure on skin reflectivity,” Biomed. Opt. Express 9(12), 6467–6476 (2018). 10.1364/BOE.9.006467 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 19.Musina G. R., et al. , “Optimal hyperosmotic agents for tissue immersion optical clearing in terahertz biophotonics,” J. Biophotonics 13(12), e202000297 (2020). 10.1002/jbio.202000297 [DOI] [PubMed] [Google Scholar]
- 20.Kolesnikov A., et al. , “THz monitoring of the dehydration of biological tissues affected by hyperosmotic agents,” Phys. Wave Phenom. 22, 169–176 (2014). 10.3103/S1541308X14030029 [DOI] [Google Scholar]
- 21.Hoshina H., et al. , “Terahertz pulsed imaging of frozen biological tissues,” Appl. Phys. Lett. 94(12), 123901 (2009). 10.1063/1.3106616 [DOI] [Google Scholar]
- 22.Sim Y., et al. , “Temperature-dependent terahertz imaging of excised oral malignant melanoma,” IEEE J. Biomed. Health 17(4), 779–784 (2013). 10.1109/JBHI.2013.2252357 [DOI] [PubMed] [Google Scholar]
- 23.Sim Y. C., et al. , “Terahertz imaging of excised oral cancer at frozen temperature,” Biomed. Opt. Express 4(8), 1413–1421 (2013). 10.1364/BOE.4.001413 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 24.Park J. Y., et al. , “Terahertz imaging of metastatic lymph nodes using spectroscopic integration technique,” Biomed. Opt. Express 8(2), 1122–1129 (2017). 10.1364/BOE.8.001122 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 25.Vansonnenberg E., et al. , Tumor Ablation: Principles and Practice, Springer-Verlag, New York: (2005). [Google Scholar]
- 26.Mazur P., “Cryobiology: the freezing of biological systems,” Science 168(3934), 939–949 (1970). 10.1126/science.168.3934.939 [DOI] [PubMed] [Google Scholar]
- 27.Gage A. A., Baust J., “Mechanisms of tissue injury in cryosurgery,” Cryobiology 37(3), 171–186 (1998). 10.1006/cryo.1998.2115 [DOI] [PubMed] [Google Scholar]
- 28.Baust J., et al. , “Mechanisms of cryoablation: clinical consequences on malignant tumors,” Cryobiology 68(1), 1–11 (2014). 10.1016/j.cryobiol.2013.11.001 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 29.Korpan N., Basics of Cryosurgery, Springer-Verlag, Wien: (2001). [Google Scholar]
- 30.Chu K. F., Dupuy D. E., “Thermal ablation of tumours: biological mechanisms and advances in therapy,” Nat. Rev. Cancer 14(3), 199–208 (2014). 10.1038/nrc3672 [DOI] [PubMed] [Google Scholar]
- 31.Yuan F., Zhao G., Panhwar F., “Enhanced killing of HepG2 during cryosurgery with Fe3O4-nanoparticle improved intracellular ice formation and cell dehydration,” Oncotarget 8(54), 92561–92577 (2017). 10.18632/oncotarget.21499 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 32.Tokiwa T., et al. , “A palm-sized cryoprobe system with a built-in thermocouple and its application in an animal model of epilepsy,” IEEE Trans. Biomed. Eng. 66(11), 3168–3175 (2019). 10.1109/TBME.2019.2901530 [DOI] [PubMed] [Google Scholar]
- 33.Li C., et al. , “MR imaging-guided cryoablation of metastatic brain tumours: initial experience in six patients,” Eur. Radiol. 20(2), 404–409 (2010). 10.1007/s00330-009-1554-8 [DOI] [PubMed] [Google Scholar]
- 34.Martynov B., et al. , “Stereotactic cryodestruction of gliomas,” Prog. Neurol. Surg. 32, 27–38 (2018). 10.1159/000469677 [DOI] [PubMed] [Google Scholar]
- 35.Berim I. G., et al. , “Radial probe ultrasound-guided cryobiopsy,” J. Bronchol. Interventional Pulmonol. 24(2), 170–173 (2017). 10.1097/LBR.0000000000000368 [DOI] [PubMed] [Google Scholar]
- 36.Ahmed M., et al. , “Percutaneous ultrasound-guided cryoablation for symptomatic plantar fibromas,” CardioVasc. Interventional Radiol. 41, 298–304 (2018). 10.1007/s00270-017-1801-3 [DOI] [PubMed] [Google Scholar]
- 37.Zhou K., et al. , “Combined magnetic resonance imaging and optical surgical navigation system guidance of percutaneous liver cryoablation in a porcine model,” Am. J. Transl. Res. 10(5), 1531–1538 (2018). [PMC free article] [PubMed] [Google Scholar]
- 38.Lichter J., et al. , “Real-time magnetic resonance imaging-guided cryoablation of the pulmonary veins with acute freeze-zone and chronic lesion assessment,” EP Europace 21(1), 154–162 (2019). 10.1093/europace/euy089 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 39.Fan W., et al. , “Percutaneous computed tomography-guided cryoablation for recurrent retroperitoneal soft tissue sarcoma: a study of safety and efficacy,” Oncotarget 7(27), 42639–42649 (2016). 10.18632/oncotarget.9476 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 40.Prud’homme C., et al. , “Image-guided lung metastasis ablation: a literature review,” Int. J. Hyperth. 36(2), 37–45 (2019). 10.1080/02656736.2019.1647358 [DOI] [PubMed] [Google Scholar]
- 41.Pohlan J., et al. , “Computed tomography thermography for ablation zone prediction in microwave ablation and cryoablation: advantages and challenges in an ex vivo porcine liver model,” J. Comput. Assist. Tomogr. 44(5), 744–749 (2020). 10.1097/RCT.0000000000001081 [DOI] [PubMed] [Google Scholar]
- 42.Yang Y., et al. , “Retrospective analysis of CT-guided percutaneous cryoablation for treatment of painful osteolytic bone metastasis,” Cryobiology 92, 203–207 (2020). 10.1016/j.cryobiol.2020.01.012 [DOI] [PubMed] [Google Scholar]
- 43.Shurrab M., et al. , “The cooling performance of a cryoprobe: establishing guidelines for the safety margins in cryosurgery,” Int. J. Refrig. 67, 308–318 (2016). 10.1016/j.ijrefrig.2016.03.007 [DOI] [Google Scholar]
- 44.Katyba G. M., et al. , “Sapphire photonic crystal waveguides for terahertz sensing in aggressive environments,” Adv. Opt. Mater. 6(22), 1800573 (2018). 10.1002/adom.201800573 [DOI] [Google Scholar]
- 45.Grischkowsky D., et al. , “Far-infrared time-domain spectroscopy with terahertz beams of dielectrics and semiconductors,” J. Opt. Soc. Am. B 7, 2006–2015 (1990). 10.1364/JOSAB.7.002006 [DOI] [Google Scholar]
- 46.Dolganova I. N., et al. , “The role of scattering in quasi-ordered structures for terahertz imaging: local order can increase an image quality,” IEEE Trans. Terahertz Sci. Technol. 8(4), 403–409 (2018). 10.1109/TTHZ.2018.2844104 [DOI] [Google Scholar]
- 47.Zaytsev K. I., et al. , “Overcoming the Abbe diffraction limit using a bundle of metal-coated high-refractive-index sapphire optical fibers,” Adv. Opt. Mater. 8(18), 2000307 (2020). 10.1002/adom.202000307 [DOI] [Google Scholar]
- 48.Lavrukhin D. V., et al. , “Terahertz photoconductive emitter with dielectric-embedded high-aspect-ratio plasmonic grating for operation with low-power optical pumps,” AIP Adv. 9(1), 015112 (2019). 10.1063/1.5081119 [DOI] [Google Scholar]
- 49.Lavrukhin D. V., et al. , “Shaping the spectrum of terahertz photoconductive antenna by frequency-dependent impedance modulation,” Semicond. Sci. Technol. 34, 034005 (2019). 10.1088/1361-6641/aaff31 [DOI] [Google Scholar]
- 50.Ponomarev D. S., et al. , “Enhanced terahertz emission from strain-induced InGaAs/InAlAs superlattices,” J. Appl. Phys. 125(15), 151605 (2019). 10.1063/1.5079697 [DOI] [Google Scholar]
- 51.Ulitko V. E., et al. , “Nanoporous SiO2 based on annealed artificial opals as a favorable material platform of terahertz optics,” Opt. Mater. Express 10(9), 2100–2113 (2020). 10.1364/OME.402185 [DOI] [Google Scholar]
- 52.Gavdush A. A., et al. , “Proof of concept for continuously-tunable terahertz bandpass filter based on a gradient metal-hole array,” Opt. Express 28(18), 26228–26238 (2020). 10.1364/OE.401608 [DOI] [PubMed] [Google Scholar]
- 53.Katyba G., et al. , “Sapphire shaped crystals for waveguiding, sensing and exposure applications,” Prog. Cryst. Growth Charact. Mater. 64(4), 133–151 (2018). 10.1016/j.pcrysgrow.2018.10.002 [DOI] [Google Scholar]
- 54.Zaytsev K. I., et al. , “Terahertz photonic crystal waveguides based on sapphire shaped crystals,” IEEE Trans. Terahertz Sci. Technol. 6(4), 576–582 (2016). 10.1109/TTHZ.2016.2555981 [DOI] [Google Scholar]
- 55.Habrawi Z., et al. , “Cryoablation: a promising non-operative therapy for low-risk breast cancer,” Am. J. Surg. 221, 127–133 (2021). 10.1016/j.amjsurg.2020.07.028 [DOI] [PubMed] [Google Scholar]
- 56.Kinsman K., et al. , “Whole-gland prostate cancer cryoablation with magnetic resonance imaging guidance: one-year follow-up,” CardioVasc. Interventional Radiol. 41, 344–349 (2018). 10.1007/s00270-017-1799-6 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 57.Hebbadj S., et al. , “Safety considerations and local tumor control following percutaneous image-guided cryoablation of T1b renal tumors,” CardioVasc. Interventional Radiol. 41, 449–458 (2018). 10.1007/s00270-017-1820-0 [DOI] [PubMed] [Google Scholar]
- 58.Zheng W., et al. , “A novel endoscopic visible full-thickness cryoablation device on stomach,” Sci. Rep. 10, 5203 (2020). 10.1038/s41598-020-61595-x [DOI] [PMC free article] [PubMed] [Google Scholar]
- 59.Mpourazanis G., et al. , “The effectiveness of photodynamic therapy and cryotherapy on patients with basal cell carcinoma: a systematic review and meta-analysis,” Dermatologic Therapy 33, e13881 (2020). 10.1111/dth.13881 [DOI] [PubMed] [Google Scholar]