Graphical Abstract
Point-of-care diagnostics for infectious diseases.

Keywords: Point-of-care, Infectious disease, Device, Biosensor, COVID-19
Abstract
The current widespread of COVID-19 all over the world, which is caused by SARS-CoV-2 virus, has again emphasized the importance of development of point-of-care (POC) diagnostics for timely prevention and control of the pandemic. Compared with labor- and time-consuming traditional diagnostic methods, POC diagnostics exhibit several advantages such as faster diagnostic speed, better sensitivity and specificity, lower cost, higher efficiency and ability of on-site detection. To achieve POC diagnostics, developing POC detection methods and correlated POC devices is the key and should be given top priority. The fast development of microfluidics, micro electro-mechanical systems (MEMS) technology, nanotechnology and materials science, have benefited the production of a series of portable, miniaturized, low cost and highly integrated POC devices for POC diagnostics of various infectious diseases. In this review, various POC detection methods for the diagnosis of infectious diseases, including electrochemical biosensors, fluorescence biosensors, surface-enhanced Raman scattering (SERS)-based biosensors, colorimetric biosensors, chemiluminiscence biosensors, surface plasmon resonance (SPR)-based biosensors, and magnetic biosensors, were first summarized. Then, recent progresses in the development of POC devices including lab-on-a-chip (LOC) devices, lab-on-a-disc (LOAD) devices, microfluidic paper-based analytical devices (μPADs), lateral flow devices, miniaturized PCR devices, and isothermal nucleic acid amplification (INAA) devices, were systematically discussed. Finally, the challenges and future perspectives for the design and development of POC detection methods and correlated devices were presented. The ultimate goal of this review is to provide new insights and directions for the future development of POC diagnostics for the management of infectious diseases and contribute to the prevention and control of infectious pandemics like COVID-19.
Introduction
Since the new coronavirus (SARS-CoV-2) caused pneumonia (COVID-19) outbreak occurred in December 2019, it has rapidly spread all over the world and caused a substantial threat to global human health [1], [2], [3]. Up to 22nd January 2021, there have been 96,012,792 confirmed cases of COVID-19 and 2,075,870 deaths all over the world, as reported by the World Health Organization (WHO) [4]. Looking back on history, none of the infectious diseases has not brought great disasters to mankind, especially for diseases caused by viruses, which have been one of the greatest existential threats to human beings [5], [6], [7], [8]. For example, the 1918 influenza pandemic [9], [10], the pandemic of human immunodeficiency virus (HIV) infection [11], [12], the pandemic of severe acute respiratory syndrome (SARS) in 2003 [13], [14], [15], the pandemic of Middle East respiratory syndrome (MERS) in 2012 [16], [17], [18], and the current pandemic of COVID-19 were all deprived of thousands of lives [19], [20], [21], [22]. All those appalling infectious diseases had significantly affected the global health, economy, social stability, and eventually the civilization process of human society, and as a result emphasized the importance of developing new novel point-of-care (POC) diagnostics to enhance timely recognition and intervention of those diseases.
Unlike other conventional diseases, infectious disease is a kind of disease caused by live pathogens such as bacteria, viruses and parasites, which is capable of rapid transmission and infection among human or animal vectors by inoculation, airborne or waterborne transmission [7], [23], [24], [25], [26]. For the management of infectious diseases, there are three strategies: 1) control of the infection source; 2) cutting off transmission routes; and 3) protection of susceptible people. Among those strategies, control of the infection source is recommended to be the most important, which requires early detection, early isolation, and early treatment of correlated patients and calls for the development of rapid, sensitive and accurate detection methods and kits [27], [28], [29], [30], [31], [32], [33]. The common strategies for the clinical diagnosis of infectious diseases are majorly relied on the pathogen culture and identification or the detection of specific antigens, antibodies or nucleic acids of infectious pathogens [34], [35], [36], [37]. However, the current conventional diagnostic methods, such as identification of microorganisms by observing characteristic features of cultures, enzyme linked immunosorbent assay (ELISA) and nucleic acid testing (NAT), are usually performed in medical laboratories, whose applications are always limited due to their shared disadvantages such as time-consuming, labor-consuming, high cost, expensive machine dependent, and inability to achieve rapid on-site detection. To achieve fast and highly sensitive NAT, various methods including photothermal nanomaterials-based PCR, isothermal nucleic acid amplification and biosensors [38], [39], [40], [41], have been developed to increase the detection speed and sensitivity of NAT. Besides, though real-time quantitative polymerase chain reaction (qPCR) has been widely employed for NAT of infectious diseases in the past years due to its high sensitivity and accuracy, the qPCR machines and correlated reagents are always very expensive and skilled technicians are essential [42], [43], [44]. Those downsides of those above mentioned diagnostic methods make them inadequate to achieve fast, accurate and on-site diagnosis during the pandemic outbreak of infectious diseases, especially in resource-poor regions where infectious diseases are always the most widespread and serious.
POC diagnostic, refer to diagnosis done in the proximity of patient care, is a new intelligent strategy for real-time, rapid, accurate, and on-site detection at the patient's point-of-need [45], [46], [47], [48]. In the past years, with high demand for on-site detection in clinical medicine, especially in resource-poor regions, POC diagnostics have attracted increasing attention since their emergences and there has been a huge market of POC devices. As coined by the WHO, the fundamental criterions of POC diagnostics are denoted to be “ASSURED”, that are affordable, sensitive, specific, user-friendly, rapid and robust, equipment-free, and deliverable to end user [34], [49]. Firstly, to develop POC diagnostics, design and establishment of novel sensitive and fast detection methods are among the most important. Up to date, there have been a variety of novel detection methods employed for the development of POC diagnostics, including electrochemical biosensors, fluorescence biosensors, surface-enhanced Raman scattering (SERS)-based biosensors, colorimetric biosensors, chemiluminiscence biosensors, surface plasmon resonance (SPR)-based biosensors, and magnetic biosensors [50], [51], [52], [53]. Moreover, empowered by the booming development of microfluidics, micro electro-mechanical systems (MEMS) technology, nanotechnology and materials science, all flowers are in bloom in the development of POC devices. Those POC devices including lab-on-a-chip (LOC) devices [54], [55], [56], lab-on-a-disc (LOAD) devices [57], [58], microfluidic paper-based analytical devices (μPADs) [59], lateral flow devices [60], miniaturized PCR devices [61], and isothermal nucleic acid amplification devices (INAA) devices [62], [63], which substantially promoted the development of POC diagnostics. In particular, the seamless integration of those novel biosensors, advanced devices and recent artificial intelligence (AI) techniques would greatly accelerate the pace of progress in the development POC diagnostics toward labor-free and intelligent diagnosis.
Over the past decades, thanks to the efforts by us and others, there have been a large number of reports about POC diagnostic systems and some of them have been commercialized for clinical disease diagnostics either in hospital or resource-poor regions [64], [65], [66], [67], [68], [69], [70], [71], [72], [73], [74], [75], [76], [77], [78], [79], [80]. Hence, under the background of COVID-19 outbreak and the high demand for rapid, sensitive and accurate on-site diagnostics, it is necessary to review the recent progresses of the development of POC diagnostics for on-site diagnosis of infectious diseases, attempting to provide new approaches and research directions for the future development of POC diagnostics. Although there have been a couple of reviews about the applications of POC diagnostics for infectious diseases, most of them are mainly about the detection methods, nanomaterials employed and target detections, and are comparably one-sided [34], [48], [52], [81], [82], [83], [84]. Herein, both POC detection methods and correlated POC devices for the POC diagnostics of various infectious diseases were systematically summarized and discussed ( Scheme 1). Firstly, various detection methods potentially for POC diagnostic applications were systematically presented. Then, the current progresses in the development of various POC devices and their applications in the detections of various infectious diseases were particularly reviewed. Finally, the challenges and future perspectives in the development of POC diagnostics were discussed in depth.
Scheme 1.
Schematic illustration of point-of-care diagnostics for infectious diseases.
Detection methods for POC diagnostic applications
Detection methods are among the most important component in the development of POC diagnostic systems, which determine the efficiency, sensitivity and accuracy of the whole system. Therefore, developing novel detection methods for rapid and highly sensitive detections of infectious diseases is the foundation of design and fabrication of POC devices. Herein, various detection methods potential for POC diagnostic adaptation were summarized, including electrochemical biosensors, fluorescence biosensors, SERS-based biosensors, colorimetric biosensors, chemiluminiscence biosensors, SPR-based biosensors and magnetic biosensors.
Electrochemical biosensors
Electrochemical biosensors are one kind of the most popular detection methods for the detection of all kinds diseases, due to their advantages of short detection time, instrumental simplicity, low cost and portability, holding a great potential for POC diagnostic applications [85], [86], [87], [88], [89]. As a result, various electrochemical biosensors have been explored for the applications in the development of POC diagnostic platforms, including voltammetric/amperometric biosensors [90], [91], [92], [93], [94], impedimetric biosensors [95], [96], potentiometric biosensors [97], and field-effect transistor (FET) based biosensors [98], [99].
Voltammetric/amperometric biosensors
Voltammetric/amperometric biosensors are one of the widest employed electrochemical biosensors in the analyst detection due to their high sensitivity caused by the enzymes used [100]. Up to date, there have been various methods established for the development of voltammetry biosensors, including cyclic voltammetry, linear sweep voltammetry, square wave voltammetry, and differential pulse voltammetry [91], [101]. Although both voltammetric and amperometric biosensors can achieve the detection of targets by measuring the current arising from electrolysis caused by an electrochemical oxidation or reduction at the working electrode upon being applied a potential to the indicator electrode versus the reference electrode, whose signals are determined by the mass transport rate of the reactant molecules from the solution to the electrode interface, the potential applied in a voltammetric biosensor is ramped at a given rate, while in an amperometric biosensor the applied potential is constant [98]. As a paradigm of POC diagnostic adaptation, Singhal et al. developed an electrochemical voltammetric biosensor based on molybdenum disulfide nanosheets (MoS2 NSs) for POC detection of chikungunya virus DNA ( Fig. 1A) [102]. Briefly, MoS2 NSs were synthesized and subjected to physical adsorption onto the screen printed gold electrodes for immobilizing the probe DNA. Then, the magnitude of the voltammetric signal produced by the different interaction of methylene blue with the guanine bases of the single and double-stranded DNA was detected. The proposed electrochemical voltammetric biosensor exhibited a low detection limit of 3.4 nM [102]. In another work, Hiraiwa et al. developed an amperometric biosensor by a microtip immunoassay for rapid and POC detection of Mycobacterium tuberculosis (MTB) in sputum [103]. Briefly, a capture antibodies coated microtip was used to capture and concentrate the target bacteria by a coffee-ring effect to form immunocomplex on the microtip surface, which could be detected by electric current. The amperometric biosensor could achieve a highly sensitive detection of MTB in spiked human sputum with a detection limit of 100 CFU mL−1, which was simple, portable, and very promising for POC detection in the resource-poor regions.
Fig. 1.
(A) Schematic illustration of MoS2 NSs based biosensor for the POC diagnosis of chikungunya virus (a) Screen printed gold electrode (SPGE). (b) Stepwise representation of the fabrication of SPGE with MoS2 nanosheets, probe DNA and target DNA. (c) Interaction of working electrode (Au) of SPGE and MoS2 shows strong affinity between Au (SPGE) and S (MoS2). (d) MoS2 interacts with probe DNA via Vander Waals forces. (e) Principle for the detection of hybridization of target DNA via redox hybridization indicator, i.e., methylene blue. Reproduced with permission from Ref. [102] Copyright 2018, Springer Nature. (B) Schematic construction representation of the impedimetric label-free immunosensor based on an anti-NS1 modified gold electrode for POC dengue diagnosis. Reproduced with permission from Ref. [108] Copyright 2015, Elsevier. (C) Schematic of the potentiometric sensing device, consisting of two parts: a sensor part for signal generation and a transducer part for signal amplification and readout. Reproduced with permission from Ref. [114] Copyright 2016, Elsevier. (D) Schematic of graphene field-effect transistors (GFETs) nano-biosensor conjugated with single-chain variable fragment (scFv) antibodies for Lyme disease diagnosis. (a) Photograph of a sensor array with 100 GFET devices. (b) Optical micrograph of a GFET channel with dimension of 100 μm × 10 μm, connected to source and drain electrodes highlighted in yellow. The light purple region is the 285 nm thick SiO2 substrate, while the darker purple region is the graphene channel. (c) Schematic of a GFET functionalized with Lyme disease scFv through a 1-pyrenebutyric acid N-hydroxysuccinimide ester linker, with the bound antigen. Reproduced with permission from Ref. [121] Copyright 2020, IOP Publishing.
Impedimetric biosensors
Impedimetric biosensors are another kind of commonly used electrochemical biosensors due to its advantages of label-free and small amplitude perturbation, which employ electrochemical impedance spectroscopy (EIS) to realize target analysis [91], [104], [105]. Electrochemical impedance is the ratio of an incremental voltage change to the resulting current change, and the resistive and capacitive components of the circuit could be determined by EIS through a frequency changed small amplitude sinusoidal AC excitation signal. When the frequency is considerably high, there will be a rate limiting upon the migration of the redox species to the electrode surface, which can be blocked by the targets and generates a frequency-dependent phase lag between the AC voltage and current [98], [106]. The EIS can be performed in Faradaic and non-Faradaic modes. Faradaic EIS involves charge transfer between the electrodes and requires the addition of a redox couple, while non-Faradaic detection requires no additional reagents, in which the capacitive behavior is produced through the separation of charges at the electrode-electrolyte interface [106], [107]. As an example of applications of impedimetric biosensors for POC detection of infectious disease, Cecchetto et al. designed an impedimetric label-free immunosensor based on an anti-NS1, a non-structural dengue protein antibody, modified gold electrode for POC diagnosis of dengue by testing neat serum (Fig. 1B) [108]. Briefly, an impedimetric biosensor was fabricated by covalently immobilizing anti-NS1 to the gold electrode surface for NS1 binding. Then, the impedance spectrum was collected in the presence of a redox probe ([Fe(CN)6]3−/4−) to monitor the charge-transfer resistance changes caused by the target binding. The impedimetric label-free immunosensor could achieve high sensitive detection of NS1 with a detection limit of 30 ng mL−1 in neat serum, suggesting its potential application in POC diagnostics of dengue [108].
Potentiometric biosensors
Potentiometric biosensors apply an electrochemical cell with two reference electrodes, largely ion-selective electrodes (ISEs), to measure the accumulation of a charge at an electrode while drawing negligible current flow [109], [110]. For biological detection, enzymes are usually employed in the potentiometric sensors to catalyze chemical reaction and produce ions in the vicinity of the sensing ISE [111], [112]. Potentiometric biosensors have several advantages of small size, rapid response, ease of use, low cost, resistant to color, turbid interferences, and is independent of the sample volume [113], making it promising for POC diagnostic applications. As a paradigm, Tarasov et al. developed a novel potentiometric biosensor based on extended-gate field-effect transistors in a dual-chip configuration for on-site serological diagnosis of Bovine Herpes Virus-1 (BHV-1) (Fig. 1C) [114]. The potentiometric biosensor was then demonstrated to be highly sensitive and selective to detect BHV-1 in real serum samples from cattle, which could be easily integrated to POC devices for the POC diagnosis of a wide range of human and animal diseases [114].
Field-effect transistor (FET)-based biosensors
FET-based biosensors are a class of electrochemical biosensors that detect the change in source-drain channel conductivity in the electric field caused by the accumulation of charged target species at the sensor surfaces [115], [116], [117]. FET-based biosensors have attractive advantages of label-free, miniaturization, ease of mass production, high versatility and low cost, and thus have become an ideal candidate for the development of POC diagnostics [117], [118], [119], [120]. For instance, Gao et al. developed a highly sensitive graphene FET-based nanobiosensor coupled with Lyme single-chain variable fragment antibodies for multiplexed detection of Lyme disease antigens of Borrelia burgdorferi bacterium, which exhibited a low detection limit of 2 × 10−3 ng mL−1 (Fig. 1D) [121].
Fluorescence biosensors
Fluorescence biosensors are a type of biosensors that have been widely employed for POC diagnostics of infectious diseases due to their appealing advantages of high sensitivity, low cost, easily operation and rapid analysis [122], [123], [124], [125], [126], [127]. Labeling of fluorophores, either fluorescent dyes or fluorescent nanomaterials, is always essential for the establishment of a fluorescence biosensor. For the design of a fluorescence biosensor, one of the most typical strategies is using fluorescence resonance energy transfer (FRET), which refers to the transfer of energy from a donor fluorophore to an acceptor fluorophore [87], [128], [129], [130]. In addition, some nanomaterials can also produce fluorescence signals due to their unique physical, chemical, and electronic transport properties, and thus have attracted great attention for the development of fluorescence biosensors [131], [132]. Based on the fluorescence generation or quencher in the detection, fluorescence biosensors can be classified into direct fluorescence labeling biosensors, “signal-on” fluorescence biosensors and “signal-off” fluorescence biosensors. For direct fluorescence labeling biosensors, the fluorescent tag is labeled to the specific ligand that selectively binding to the targets and generate corresponding proportional fluorescence signal. For “signal-on” fluorescence biosensors, the fluorescence signal is in proportional to the target quantity, and only negligible fluorescence intensity exists in the absence of the target. On the contrary, the fluorescence signal in a “signal-off” fluorescence biosensor is inversely proportional to the target quantity, and the fluorescence intensity is enhanced in the absence of the target. Herein, those potential fluorescence biosensors for POC diagnostics of infectious diseases were systematically discussed.
Direct fluorescence labeling biosensors
In direct fluorescence labeling biosensors, fluorescent dyes are labeled to the binding ligands that specifically recognize the targets and serve as the reporter signals. The fluorescence signals are proportional to the target amounts. Except for traditional fluorescent dyes, some nanomaterials have also been widely employed for developing fluorescence biosensors due to their excellent fluorescence properties, such as quantum dots (QDs) [133], carbon dots (CDs) [134], [135], upconversion nanoparticles (UCNPs) [136], and metal nanoclusters (NCs) [137], [138], [139], which have been widely applied for the fabrication of fluorescence biosensors. Herein, various direct fluorescence labeling biosensors for POC diagnostics of infectious diseases were summarized.
Quantum dots-based fluorescence biosensors
QDs are a kind of nanoparticles with particle sizes less than or equal to their own exciton Bohr radius in three space dimensions, usually less than 10 nm, and have strong fluorescence properties [140], [141]. Compared with other fluorescent materials, QDs exhibit several merits such as high quantum yield, tunable emission spectrum, and excellent photostability [141], [142], [143], and thus have attracted much attention for the development of POC diagnostic platforms for the detection of infectious diseases. As a paradigm, Hu et al. [144] encapsulated fluorescent QDs and Au nanoparticles into antibody and streptavidin modified RNs@Au multifunctional nanospheres (antibody-RNs@Au-streptavidin) and developed a sensitive and quantitative lateral flow assay for the POC detection of Ebola virus glycoprotein. In this strategy, fluorescent QDs were used for quantitative detection and Au nanoparticles were used for visual detection, which was able to achieve naked-eye detection of 2 ng mL−1 glycoprotein within 20 min and a quantitative detection limit of 0.18 ng mL−1. Takemura et al. [125] developed a new QDs-based immunofluorescence biosensor for the POC detection of nonstructural protein 1 (NS1) of Zika virus. In this biosensor, a localized surface plasmon resonance (LSPR) signal from plasmonic gold nanoparticles (AuNPs) was used to amplify the fluorescence intensity signal of QDs to achieve ultrasensitive, rapid, and quantitative detection of NS1 of Zika virus. The biosensor was able to detect NS1 of Zika virus in a wide concentration ranging from 10 fg mL−1 to 1 ng mL−1, and the detection limit was as low as 1.28 fg mL−1.
Carbon dots-based fluorescence biosensors
CDs are a kind of well-known fluorescent zero-dimensional carbon-based nanomaterials with diameters less than 10 nm [145]. Due to their preferable attributes, such as excellent fluorescence properties and biocompatibility, CDs have been widely applied for various biomedical applications, including biosensing, imaging, drug delivery and therapy [146], [147], [148], [149]. Among all those biomedical applications, fluorescence-based biosensing by leveraging their unique fluorescence properties has been extensively investigated for POC diagnostics of infectious diseases. For example, Xu et al. [150] developed a fluorescent lateral flow assay using CDs/SiO2 nanospheres (CSNs) as reporters for ultrasensitive POC detection of severe fever with thrombocytopenia syndrome virus (SFTSV). The fluorescent lateral flow assay owned good reproducibility, high specificity and sensitivity due to the excellent fluorescent properties of CSNs, and was able to achieve a low visual detection limit of as low as 10 pg mL−1 SFTSV nucleoprotein which was 2 orders of magnitude higher than that of the colloidal gold-based lateral flow test strip. Most recently, Wang et al. [151] developed a Eu(III) functionalized carbon dots (CDs-Eu)-based ratiometric fluorescent test paper for sensitive POC detection of dipicolinic acid (DPA), a biomarker of anthrax. This ratiometric fluorescent test paper exhibited excellent detection performance with high sensitivity which could achieve a linearity range from 0.5 nM to 5 μM and a detection limit of 0.8 nM. To make it a more convenient approach for DPA detection, the ratiometric fluorescent test paper was further being combined with a smartphone and was able to achieve visualized POC detection of DPA with a detection limit of 67 nM, which was lower than other reported test paper for DPA detection, holding a great potential for the practical POC diagnostics of anthrax.
Upconversion nanoparticles-based fluorescence biosensors
Lanthanide-doped UCNPs are fluorescent inorganic nanomaterials that can absorb two or more low-energy photons light to produce higher energy photons and emit fluorescence signals [152], [153], [154]. Compared with conventional fluorescent materials, UCNPs have a variety of advantages, such as low toxicity, long luminescence lifetimes, high photostability, high signal-to-noise ratio, and high resistance to photobleaching [155], [156], [157], and thus have been regarded as a kind of very promising fluorescent probes for POC detection of infectious diseases. As a paradigm, Zhao et al. [158] developed an electro-driven immunochromatography assay using UCNPs for high-sensitivity detection of Yersinia pestis EV76. In this assay, UCNPs were used as the fluorescent tags and conjugated with monoclonal antibodies against Yersinia pestis EV76, and electroosmotic flow was applied to increase the assay speed. The electro-driven immunochromatography assay showed high specificity and sensitivity for the detection of Yersinia pestis EV76, and the whole process could be completed within 5 min. More importantly, the assay displayed good repeatability and high reliability in various conditions, and could achieve a low detection limit of 1.2 × 104 CFU mL−1 in real soil samples, showing a great potential for on-site pathogen detection and other POC applications. In another study, Jin et al. [159] used aptamers functionalized UCNPs as probes and developed a lateral flow aptamer assay integrated smartphone-based portable device for highly sensitive and precise detection of Salmonella. Detailedly, aptamers were conjugated with the UCNPs to specifically bind with the targets and produced fluorescence signal at the test line, while a smartphone-based device was used to read the results, making the detection process rapid and portable. After optimization, the assay was demonstrated to be able to provide rapid and sensitive detection of Salmonella with a wide range of 150–2000 CFU mL−1 and a detection limit of 85 CFU mL−1, holding a great potential to be used as a sensitive, specific, convenient and stable platform for POC detection of Salmonella in various fields.
Metal nanoclusters-based fluorescent biosensors
Metal NCs are a type of ultrasmall nanomaterials consisting of several to tens of atoms, which have attracted a great attention for using as a new type of fluorescent probes for biosensing applications [160], [161], [162]. Compared with other fluorescent nanomaterials, metal NCs exhibit outstanding physical and chemical properties, well biocompatibility and strong fluorescence emission [162], [163], [164]. Over the past years, a variety of metal NCs, including AuNCs [165], [166], AgNCs [167], [168], and so on, have been synthesized and employed for the POC detection of infectious diseases. For example, Ji et al. [169] synthesized a protein-AuNC-based fluorescence sensor array for rapid identification of bacteria, which was able to realize discrimination of six kinds of bacteria, including two kinds of drug-resistant bacteria. Moreover, the sensor array could achieve 100% classification accuracy by using two protein-AuNCs probes with outstanding advantages of easy synthesis and convenient to use, demonstrating its promise for facile diagnosis of bacterial infection in resource-poor regions [169]. In another study, a label-free fluorescence biosensor based on DNA template AgNCs fluorescent probe was developed for the detection of clinical significant DNA fragments from human immunodeficiency virus type 1 (HIV-1). The proposed fluorescence sensor exhibited excellent selectivity and high sensitivity against mismatched target DNA and could achieve a low detection limit of 2.9 × 10−10 mol L−1 HIV-1 DNA [170].
“Signal-on” fluorescence biosensors
In the past years, there have been a series of reports about the application of “signal-on” fluorescence biosensors for POC detection of infectious disease. For example, Lee et al. developed a fluorescent probe-based “signal-on” biosensor using the cooperation of quantum dot-aptamer beacons and light guides in a 3D photonic crystal for POC detection of influenza A (H1N1) virus [171]. Firstly, quantum dot-aptamer beacons and dark quencher-labeled guard DNA (G-DNA) were annealed for hybridization, which quenched the fluorescence signals of quantum dot to “OFF” state. In the presence of the H1N1 virus, the specific aptamers would preferentially bind to the virus and release G-DNA, resulting in recovery of fluorescence signal to “ON” state ( Fig. 2A). The fluorescence intensity of the biosensor was quantitatively proportional to the concentration of the target H1N1 virus. The fluorescence biosensor was user-friendly, highly sensitive and selective towards H1N1 virus detection with an ultra-low detection limit of 0.138 ng mL−1. Furthermore, the fluorescence biosensor could also be adapted to a low cost portable home-made chamber for quantitative analysis of H1N1 virus with a smartphone camera [171]. While in another study, He et al. combined a powerful clustered regularly interspaced short palindromic repeats (CRISPR)-Cas assay with a “signal-on” fluorescence biosensor for rapid and POC detection of the nucleic acid of African Swine Fever Virus (ASFV) [172]. Specifically, this fluorescence detection system applied CRISPR-Cas12a programmed with a CRISPR RNA (crRNA) to specifically bind with ASFV target DNA to form activated Cas12a/crRNA/ASFV DNA complex, which then degraded the fluorescent single stranded DNA reporter probe that linked a fluorophore and a quencher, and thus generated fluorescence signal. Finally, the fluorescence signal was detected by a fluorescence-sensing unit aligned with an 80x disposable cartridge. This fluorescence biosensor could achieve rapid, accurate, and sensitive POC detection of ASFV DNA without nucleic acid amplification. The whole detection processes could be completed within 2 h and could achieve a low detection limit of 1 pM [172].
Fig. 2.
(A) Schematic illustration of “Signal-on” fluorescence detection principle based on Qdot-aptamer conjugates and 3D nanoporous photonic crystal. (a) Loading the prepared quenched Qdot-aptamer conjugates (i.e., “OFF” state) for H1N1 virus detection in aerosol spray. (b) Capturing airborne H1N1 virus by aerosol spray and recovering Qdot signals by releasing G-DNA and preferred aptamer-virus binding (i.e., “ON” state). (c) Visualization of virus detection by enhanced Qdot signal owing to the emission light guide in 3D photonic crystals. Reproduced with permission from Ref. [171] Copyright 2018, the Royal Society of Chemistry. (B) Schematic illustration of “Signal-off” fluorescence immunoassay for the detection of AIV. Reproduced with permission from Ref. [173] Copyright 2012, Elsevier.
“Signal-off” fluorescence biosensors
As a paradigm, Li et al. integrated lateral flow test strip technique with a novel fluorescence immunoassay method for fast and ultrasensitive POC detection of avian influenza virus (AIV), in which label-free and high luminescent quantum dots were applied as fluorescence signal output and gold ions released from gold nanoparticles were used as the fluorescence quencher [173]. In the presence of AIV, gold ions would be released from the gold nanoparticles and were captured in the test zone through the sandwich immunoreaction performed on the lateral flow test strip, which then quenched the fluorescence signals of quantum dots, resulting in “signal-off”. Whereas, the fluorescence signals of quantum dots would recover in the absence of AIV (Fig. 2B). The proposed method could directly detect clinical samples without any pretreatment, and showed higher efficiency (90.0%), sensitivity (100.0%) and specificity (88.2%) against AIV compared with virus isolation (gold method), which was able to realize a linear range of 0.27–12 ng mL−1 and a detection limit of 0.09 ng mL−1, showing great promise for rapid, sensitive, and quantitative detection of AIV at POC [173]. While in another study, a rapid, low cost, and sensitive antibody-functionalized MoS2-based “signal-off” fluorescence biosensor was developed for the detection of infectious bronchitis virus (IBV, an avian coronavirus), in which the FRET was utilized between the MoS2 and fluorescence dye during the antibody-antigen interaction. The biosensor was performed on a low cost cotton thread-based microfluidic platform and could achieve a remarkable sensitivity of 0.023 ng mL−1 for IBV standard solutions, which is very potential for POC detection [174].
Surface-enhanced Raman scattering (SERS)-based biosensors
Despite the outstanding attributes and extensive applications of the above mentioned fluorescence biosensors, they always have high background signals, limited sensitivity and inability to achieve multiplex detection [175]. SERS, which is a commonly used powerful sensing technique using enhanced inelastic light scattering by molecules adsorbed onto corrugated metal surfaces, can overcome those disadvantages of fluorescence biosensors due to its attributes of high sensitivity, well multiplexing capabilities, no photobleaching and low background [175], [176], [177], [178]. More importantly, SERS-based biosensors can achieve single molecule detection by means of coinage-metal nanostructures [179]. Therefore, SERS has been extensively implemented for the development of biosensors in a wide variety of fields [180], [181], [182], [183]. Meanwhile, SERS-based biosensors have also been widely investigated for the pathogen detection of infectious diseases at the POC, mainly including viruses and bacteria [175], [184], [185]. For example, Pang et al. developed a magnetically assisted SERS-based biosensor using aptamer and antibiotic molecules for dual-recognition of bacterial cells at the POC [186]. In this biosensor, aptamer-Fe3O4@Au magnetic nanoparticles were synthesized as magnetic and SERS activated substrate for specific bacteria enrichment, while vancomycin-SERS tags were prepared for the sensitive quantification of the pathogenic bacteria ( Fig. 3A). The biosensor showed high specificity and sensitivity towards the target pathogenic bacteria with an ultra-low detection limit of 3 cells mL−1 and a wide dynamic linear range from 10 to 107 cells mL−1, which is more sensitive than most of the current reported biosensors. Moreover, the whole process of the biosensor could be completed within 50 min and exhibited recoveries from 95.0% to 106.4% with relative standard derivation less than 5.3% in real samples [186]. Shi et al. designed a rapid SERS-based lateral flow immunoassay biosensor for the sensitive and quantitative POC detection of Escherichia coli (E. coli) O157:H7 in biological samples [187]. Firstly, novel gold-shell silica-core nanospheres (SiO2/Au) with monodispersity, good stability, and excellent SERS activity were prepared and utilized as high performance tags. The SiO2/Au SERS tags were then modified with two layers of Raman reporter molecules and monoclonal antibodies for specifically and effectively binding with E. coli O157:H7 and forming sandwich immune complexes on the test lines. Under optimal conditions, this biosensor showed a detection limit of 50 cells mL−1 in PBS solution and 100 cells mL−1 in real samples, which are 2000 times more sensitive than colloidal Au-based lateral flow immunoassay strips [187]. Most recently, Lee et al. developed a PCR-coupled paper-based SERS biosensor composed of silver-nanowires (AgNWs) for rapid and sensitive determination of Mycoplasma pneumoniae DNA at the POC (Fig. 3B) [188]. In this system, a DNA intercalating molecule named EvaGreen dye was introduced to the low-cycled PCR product, whose difference in the intercalation of DNA structure could be quantified by Raman spectroscopy. The PCR-coupled SERS biosensor showed enhanced detection capability with a DNA detection limit of 3.12 pg μL−1 with 10 cycles of target DNA amplification. Finally, the SERS substrate was prepared as a rapid kit containing a test line and negative and positive control lines, which could achieve specific discrimination of the target DNA and non-target DNA with statistically relevant after 10 cycles of amplification. Notably, the total assay time using the prototype SERS biosensor was only 30 min, including 20 min of thermal amplification and 10 min of SERS measurement, which is very promising to be used as a new detection method for the POC diagnostics of various bacteria and viruses [188].
Fig. 3.
(A) Schematic illustration of (a) the synthesis of Au-Van SERS tags, (b) the synthesis of aptamer-modified Fe3O4@Au MNPs, and (c) the operating procedure for S. aureus detection via the dual-recognition SERS biosensor. Reproduced with permission from Ref. [186] Copyright 2019, Elsevier. (B) Schematic illustrations of diagnostic process for respiratory bacterial DNA using paper-based SERS substrate. In the presence of target, EvaGreen dye intercalation is dominant in DNA structure while in the presence of non-target or no-target, majority of EvaGreen is adhered on the hot-spots area of the AgNWs. The diagnosis can be made by comparing the Raman intensity between the test line and the control lines. Reproduced with permission from Ref. [188] Copyright 2021, Elsevier.
Colorimetric biosensors
Colorimetric biosensors have long been investigated for POC diagnostic applications due to their superiorities such as easy preparation, easy readout, low cost, and portability [189], [190], [191], [192], [193]. Colorimetric detection can be achieved by the oxidation of peroxidase or peroxidase-like nanomaterials, aggregation of nanomaterials, or addition of dye indicators [193], [194], [195], [196]. In the past years, peroxidase-like nanomaterials and aggregation of nanomaterials have been widely applied for colorimetric biosensor establishment such as gold nanoparticles (Au NPs), platinum nanoparticles, silver nanoparticles, magnetic nanoparticles (MNPs), and so on [196], [197], [198], [199], [200], among which Au NPs have attracted increasing interests for developing colorimetric POC diagnostic platforms for on-site detection of infectious diseases due to their simple preparation, easy surface modification, and interparticle surface plasmon resonance properties that can cause fast significant color changes through the nanoparticle aggregations [201], [202], [203], [204].
As a paradigm, Zheng et al. developed a novel colorimetric biosensor based on AuNPs aggregation and smart phone imaging for the POC detection of E. coli O157:H7 ( Fig. 4A) [205]. Specifically, MNPs modified with the capture antibodies against E. coli O157:H7, polystyrene microspheres (PSs) modified with the detection antibodies against E. coli O157:H7 and catalases were mixed with the sample containing the target E. coli O157:H7 cells in the first mixing channel of the microfluidic chip to form MNP-bacteria-PS complexes, which were then captured in the separation chamber by an external magnetic field. Then, the catalysate from the catalytic reaction of injected hydrogen peroxide by the catalases on the MNP-bacteria-PS complexes were mixed with the AuNPs and crosslinking agents in the second mixing channel, and incubated in the detection chamber. Finally, the aggregation of the AuNPs was triggered through the crosslinking agents, resulting in the color changing from blue to red, which was detected by a smartphone imaging APP to determine the amount of bacteria. The proposed biosensor exhibited rapid, high specific and sensitive detection of E. coli O157:H7 in chicken samples within 1 h, and the detection limit was as low as 50 CFU mL−1 under optimal conditions, which owned several merits such as simple device for data processing, high integration for sample and reagent handling, low cost for microfluidic chip fabrication, and small size for POC detection applications [205]. However, the immune reaction efficiency was not very high with its detection limit and detection time needed to be further improved. To address this problem, a rapid, sensitive, and reliable stochastic DNA dual-walker-based colorimetric biosensor was developed for POC detection of bacteria (Fig. 4B) [206]. This colorimetric biosensor was composed of two kinds of well-dispersed thiol-tagged oligonucleotide-functionalized AuNPs probes and an aptamer specifically binding to the target bacteria. In the presence of the target bacteria, the aptamer would preferably bind to the target bacteria and concomitantly release two kinds of walking strands. Then, the released walking strands started multiple cycles of hybridization with the thiol-tagged oligonucleotides and allowed autonomous walk on the AuNP-based 3D track as propelled by Exo III through consumption of the thiol-tagged oligonucleotides, which eventually resulted in destabilized aggregation of AuNP-based probes and fast color change. Due to the operation of DNA dual-walkers in one system and multiple walking strands concurrently released from one polymer, the enzyme digestion reaction and reaction kinetics were significantly improved, thus shortening the response time and speeding the analytical procedure. The colorimetric biosensor was demonstrated to be very sensitive and specific for bacteria detection, which could be finished within 15 min with a linear response ranging from 100 to 1 × 105 CFU mL−1 and a much lower detection limit of 1 CFU mL−1. All these merits of this colorimetric biosensor such as rapid, sensitive, specific, and reliable analysis made it very promising for POC detection of bacteria [206].
Fig. 4.
(A) The principle of the proposed colorimetric biosensor for rapid detection of E. coli O157:H7 based on gold nanoparticle aggregation and smart phone imaging. Reproduced with permission from Ref. [205] Copyright 2019, Elsevier. (B) Schematic illustration of stochastic DNA dual-walkers for colorimetric detection of bacteria. Reproduced with permission from Ref. [206] Copyright 2020, American Chemical Society.
Chemiluminiscence biosensors
Chemiluminiscence biosensors are very common sensors for disease diagnosis due to their high sensitivity, high signal-to-noise ratio, wide dynamic range, and simpler reader instruments, which have long been widely studied for POC diagnostic adaptations [207], [208], [209], [210], [211], [212], [213]. As early as in 2002, an accurate and sensitive chemiluminiscence POC diagnostics, called the ZstatFlu®-II test, for the detection of influenza virus infections was developed, which has already been commercialized [214], [215]. Besides, other commercialized chemiluminiscence biosensors based POC devices, such as FastPack® IP system, PATHFAST®, OPTIGEN® systems, and so on, have also been developed [212], [216], [217]. However, those devices did not meet the “ASSURED” standard absolutely, which still required complicated operation and complex detector devices. Besides, multiplex detection has been another important challenge for POC detection of infectious diseases, especially for the targeted detection of emerging infectious diseases. Therefore, how to simplify the operations of chemiluminiscence detection and signal acquisition, and increase detection flux has been the top priority in the development of POC devices based on chemiluminiscence detection. In our previous work, we had established various chemiluminiscence biosensors for the detection of infectious diseases, with expectations for advanced POC diagnostic applications [218], [219], [220], [221], [222], [223], [224], [225], [226], [227], [228]. As a paradigm, by combining magnetic separation technology for nucleic acid purification with a chemiluminescence technique for post-amplification detection, a DNA hybridization based chemiluminescence biosensor was developed for simultaneously multiple detection of hepatitis B virus (HBV), hepatitis C virus (HCV), and human immunodeficiency virus (HIV) ( Fig. 5) [228]. First, the viral (HBV, HCV, and HIV) nucleic acids were extracted by silica modified MNPs, followed by one step multiplex reverse transcription PCR (RT-PCR) for amplification. Then, different amino modified capture probes for HBV, HCV, and HIV were attached to carboxyl coated MNPs, and incubated with biotin labeled RT-PCR products in different tubes for capturing HBV, HCV or HIV sequences, respectively. Finally, the chemiluminescence signal was recorded after the addition of streptavidin-modified alkaline phosphatase and substrate. The chemiluminescence biosensor was very specific, efficient and sensitive with a detection limit of 5 × 10−5 ng mL−1 in serum samples in the monoplex detection of HBV, HCV, and HIV, and low detection limits of 5 × 10−2 ng mL−1 for HBV, 5 × 10−2 ng mL−1 for HCV, and 50 ng mL−1 for HIV in serum samples in the multiplex detection assay. This method could be modified into a high throughput automated set up for the simultaneous detection of multiple infections, which would dramatically decrease the cost of infectious diseases detection and showed a great potential for POC diagnostic applications [228].
Fig. 5.
Schematic illustration of one step multiplex RT-PCR and chemiluminescence detection of multiple viruses. Reproduced with permission from Ref. [228] Copyright 2017, the Royal Society of Chemistry.
Surface plasmon resonance (SPR)-based biosensors
SPR, which measures the changes in the refractive index caused by molecular interactions on the metal surface via a surface plasmon wave, has become one of the most powerful detection methods for the POC diagnostic applications due to its merits of real-time, label-free, high cost-effective, noninvasive nature, excellent reutilization performance and outstanding reproducibility [229], [230], [231], [232], [233]. And as a result, various SPR-based biosensors have been developed for POC detection of infectious diseases [82], [168], [234], [235], [236], [237], [238], among which some have already been adapted into POC devices [239], [240], [241]. However, the liquid sample volumes and power consumption of a SPR-based biosensor are still bulky, and the sensitivity and resolution of those SPR-based biosensors or devices need to be further improved, which have been the main bottlenecks in the fabrication of SPR-based biosensors for biomedical applications. To circumvent these drawbacks, nanomaterials, microfluidic devices, and compact as well as power-free pump systems have been attempted to integrate SPR-based biosensor systems [230], [241], [242], [243], [244], [245], [246], [247]. In one example, an innovative SPR-based biosensor was developed for highly sensitive detection of HIV-related DNA based on entropy-driven strand displacement reactions and double-layer DNA tetrahedrons for SPR signal enhancement ( Fig. 6A). In this biosensor system, hairpin probes were firstly immobilized on the surface of the SPR-based sensor chip in advance to bind with the target DNA via the terminal toehold region of three-stranded complexes, which then initiated the toehold-mediated strand displacement reactions and produced plentiful double stranded complex by the reuse of the target DNA. Subsequently, the generated double-stranded complex specifically hybridized with hairpin probes and captured double-layer DNA tetrahedrons nanostructure, forming three-decker composites on SPR-based sensor chip, which eventually amplified the SPR response signal significantly. The proposed SPR-based biosensor could detect HIV-related DNA sensitively and specifically with a wide linear range from 1 pM to 150 nM and a detection limit of 48 fM. More importantly, compared with the conventional methods and other sensing strategies, this SPR-based biosensor system have several advantages such as faster, simpler, lower-cost, more practical, and better stability, which is very promising for the POC detection of HIV infections [248]. In another example, a LSPR biosensor was fabricated for high sensitive detection and identification of dengue virus serotypes with CdSeTeS QDs and AuNPs regulation (Fig. 6B). Specifically, four different CdSeTeS QDs-conjugated hairpin nanoprobes were designed to exactly complement with the target analytes of four different DENV serotypes. Meanwhile, thiolated poly-C functionalized AuNPs were synthesized for complementary binding with the free poly-G region of the hairpin nanoprobe. In the presence of the complementary RNA of DENV, the hairpin structure of the nanoprobe would be opened due to the better hybridization of the nanoprobe ssDNA and complementary RNA of DENV in its loop region, and poly-C-functionalized AuNPs would then complementary bind with the poly-G region of the nanoprobe, resulting in the LSPR enhancement (at around 14 nm) or quenching (at around 3 nm) effect on the fluorescence of the QDs, which was determined by the distance between the QDs and the AuNPs. After optimization, the SPR-based biosensor could achieve the detection of standard dilutions of the target virus DNA from 1 × 10−9 to 1 × 10−4 μM with a detection limit of femtomolar range, demonstrating a great potential for POC detection of dengue virus [249]. Moreover, a portable, multiplex, inexpensive microfluidic-integrated SPR platform was developed for rapidly POC detection and quantification of E. coli (Fig. 6C). Firstly, the disposable microfluidic chips with gold coated surfaces were functionalized with antibodies for efficient, selective and specific capture of E. coli. Then, the capture event happened in the microchannel induced a change in the local refractive index, which produced a signature on the reflected light and was captured by the SPR-based sensor. The platform presented reliable capture and detection of E. coli at concentrations ranging from 1 × 105 to 3.2 × 107 CFU mL−1 in phosphate buffered saline and peritoneal dialysis fluid [250]. Furthermore, Trzaskowski et al. developed a portable SPR-based device for rapid detection of tuberculosis bacteria in sputum and cultures without bacteria lysis and purification [251]. Detailedly, miniature SPR sensor Spreeta 2000 (S2k) chips with gold surfaces were functionalized with anti-MTB secretory protein (Ag85) for specific Ag85 capturing, which was then connected to a SensiQ Discovery two channel manual SPR platform or self-built prototype SPR device. The device was able to detect MTB Ag85 in concentration of 10 ng mL−1 and tuberculosis bacteria cultures in concentration of 1 × 104 CFU mL−1 with no significant interfering response from two other bacteria species [251].
Fig. 6.
(A) Schematic representation of SPR biosensor for POC detection of HIV. (a) Schematic illustration of the preparation of DDTs nanostructure; (b) Schematic representation of SPR biosensing strategy for HIV-related DNA detection based on ESDRs and DDTs nanostructure. Reproduced with permission from Ref. [248] Copyright 2018, Elsevier. (B) Schematic representation of the detection mechanism of dengue virus serotypes by AuNPs and hairpin ssDNA-CdSeTeS quantum dots. (a) Probe design for DProbe 1 and 2 and functionalized AuNPs. (b) Identification and fluorometric enhancement for dengue virus 1/3 serotype and quenching for dengue virus 2/4 serotype depending on the distance between the nanocomposite of AuNP-dsDNA-CdSeTeS. Reproduced with permission from Ref. [249] Copyright 2020, the Royal Society of Chemistry. (C) Schematic of Portable plasmonic platform for pathogen detection and quantification. Reproduced with permission from Ref. [250] Copyright 2015, Springer Nature.
Magnetic biosensors
Magnetic biosensors, which employ the magnetic phenomena of magnetic materials to achieve sensitive detections of the analytes, have received surging interests for POC detection applications in the past decades [53], [123], [252]. Compared with other biosensors, magnetic biosensors have a couple of advantages. Firstly, magnetic biosensors are of low cost and improved detection efficiency due to the elimination of expensive optics components and decreased sample preparation time by utilizing magnetic field [253], [254], [255]. Besides, magnetic biosensors exhibit high specificity, sensitivity and high signal-to-noise ratio due to the negligible magnetic background signal of biological samples [256], [257], [258]. After several decades of development, there have been several classes of magnetic biosensors developed, such as giant magnetoresistance (GMR) biosensors, magnetic tunnel junction biosensors, magnetic particle spectroscopy (MPS) biosensors, and the nuclear magnetic resonance biosensors and so on [252], [259]. Among them, GMR biosensors and MPS biosensors have been widely investigated for POC detection of infectious diseases in the past years.
GMR effect, which was discovered in 1980s, usually happens in a multi-layer structure in which ferromagnetic (FM) and non-magnetic (NM) thin films are deposited alternately [260], [261], [262]. However, subsequently studies demonstrated that GMR effect could also happen in inhomogeneous media, granular films for example, which expanded the biomedical applications of GMR biosensors [263], [264]. Moreover, GMR effect can sense very low magnetic fields, making it very attractive for developing magnetic biosensors for highly sensitive POC detection of infectious diseases. For example, Wang at al. established a portable handheld GMR biosensor named Z-Lab for the POC detection of influenza A virus (IAV) ( Fig. 7A) [265]. This GMR biosensor displayed high specificity and sensitivity towards IAV nucleoprotein and purified H3N2v with quantitative results in less than 10 min and low detection limits of 15 ng mL−1 and 6.25 × 10−3 ng mL−1 for IAV nucleoprotein and purified H3N2v, respectively. However, the detection processes of Z-Lab involved multiple washing and incubating steps, which were time-consuming and tedious, and thus the detection efficiency was limited. To solve this problem, Wang et al. further developed a wash-free magnetic bioassay to integrate with the Z-Lab platform for POC detection of both purified IAV nucleoprotein and IAV in nasal swab samples (Fig. 7B) [266]. The wash-free magnetic bioassay greatly simplified the testing process and significantly improved the detection efficiency, which was able to achieve high sensitive detection of purified IAV nucleoprotein and IAV with limits of detection of 0.017 ng mL−1 for IAV nucleoprotein and 1.25 × 10−2 ng mL−1 for IAV-spiked nasal swab samples. Although GMR biosensors can achieve POC detection of infectious diseases, the fabrication of GMR biosensors is very complex and the test is expensive. The further development of GMR biosensors may focus on their simplification and cost reduction.
Fig. 7.
(A) Schematic of Z-Lab diagnosis platform and magnetic sandwich assay mechanism. (a) Real-time data can be collected and transmitted to a smartphone, a tablet, a laptop and a desktop computer. (b) The Z-Lab platform consists of a plastic cartridge, an electrical interface connecting the electrodes from GMR chip to the circuit board, and a handheld device. (c) Schematic of the GMR chip. (d) Schematic of the magnetic sandwich assay. (e) Real-time binding curves of targeted binding. Reproduced with permission from Ref. [265] Copyright 2017, American Chemical Society. (B) Schematic of a wash-free magnetic bioassay based handheld GMR biosensors for POC detection (a) One-step wash-free magnetic bioassay based on a sandwich assay structure. (b) Photograph of one Z-Lab diagnosis platform. (i) disposable plastic cartridge; (ii) cartridge shell; (iii) Helmholtz coil with ferrite core; (iv) card edge connector; (v) microcontroller; (vi) UART to Bluetooth and USB; (vii) power supply; (viii) current source Helmholtz coil driver. (c) The circuit schematic of Z-Lab. Reproduced with permission from Ref. [266] Copyright 2019, Frontiers. (C) Schematic of the MPS biosensor for POC diagnostics (a) Schematic view of MPS system setups. (b) and (c) are the third and the fifth harmonics along varying driving field frequencies (only samples I, VIII, and IX are plotted) collected by the MPS system. (d) Boxplots of the harmonic ratios (R35) collected from samples I, VIII, and IX. Reproduced with permission from Ref. [271] Copyright 2020, American Chemical Society.
MPS is a novel flourishingly researched detection method closely related to magnetic particle imaging, which can also be interpreted as a zero-dimensional magnetic particle imaging scanner that conducts spectroscopic studies on superparamagnetic iron oxide nanoparticles (SPIONs) [267], [268]. In MPS, SPIONs are periodically driven into magnetically saturated regions in a sinusoidal magnetic field with sufficiently large amplitude and produce dynamic magnetic responses, which contain unique higher odd harmonics. Those harmonics are then obtained and extracted by filtering and fast Fourier transform, which contain important information about the SPIONs solution for analysis [269], [270]. Over the past years, MPS has been extensively explored as a portable, low cost and highly sensitive detection method for various biomedical detection applications, including POC diagnostics of infectious diseases. As a paradigm, a MPS biosensor was investigated for the POC detection of IAV subtype H1N1 nucleoprotein [271]. In this biosensing system, the oscillating harmonics of MNPs were measured by MPS, which was a metric of the freedom of rotational process that indicated the bound states of MNPs, and could be readily collected from nanogram quantities of MNPs within 10 s (Fig. 7C). In the presence of H1N1 nucleoprotein, IgG polyclonal antibodies anchored MNPs would specifically bind with H1N1 nucleoprotein molecules and trigger cross-linking between MNPs and H1N1 nucleoprotein, forming MNP self-assemblies. The MPS biosensor was finally demonstrated to be able to achieve rapid, sensitive, and wash-free detection of H1N1 nucleoprotein and the detection limit was as low as 2.49 ng mL−1, which could be further developed into a portable MPS-based device for the POC detection of H1N1 virus [271].
Despite the well performances of the above summarized biosensors, there are still many shortcomings of them. The detailed comparisons among them are illustrated in Table 1 . Among the above biosensors, electrochemical biosensors are among the most widely investigated methods and show the most potential for POC diagnostic adaptation due to their high sensitivity, fast response and low cost. However, an electrochemical biosensor always displays weak stability and is susceptible to interference from environment [272], [273], [274]. Fluorescence biosensors are widely applied for POC detection because of their merits of rapid response and high sensitivity, flexibility and experimental simplicity. However, they are limited by high fluorescence background and limited fluorescence lifetime [275], [276], [277]. SERS-based biosensors can achieve single molecule detection of targets with high sensitivity, well multiplexing capabilities, no photobleaching and low background, the SERS-based biosensors still displayed limited field of view [177], [178], [278]. Colorimetric biosensors show the advantages of simplicity and detection by naked eyes without expensive instruments, which is the pioneered and most successful method for POC diagnostic applications, while it is difficult to achieve quantitative detection and the sensitivity is limited [279], [280], [281]. For a chemiluminescent biosensor, it has advantages of high sensitivity, wide linear dynamic ranges, simple instrumentation and has been widely employed in clinical detection, but it is limited by tedious washing steps, time-consuming, and is enzyme-dependent [87], [282], [283], [284]. SPR-based biosensors exhibit extreme sensitivity, ability of detection in complex matrices, label-free and real-time detection, distinctive fingerprint capability and ability to achieve high-throughput analysis. However, the SPR equipment is too bulky and expensive [285], [286]. With the booming development of nanotechnology and microfluidics in recent years, SPR-based biosensors have been attempted to be integrated with various POC devices to provide accurate, rapid, low cost and highly sensitive POC detection. Finally, a magnetic biosensor is able to achieve low cost, high-efficient and high signal-to-noise ratio detection without expensive optics components, there is a shortage of miniaturized magnetic signal readout systems. Thus, fabrication of miniaturized magnetic signal readout systems would contribute greatly to the development of portable devices to integrate with magnetic biosensors for POC detection [123], [258], [259]. Notwithstanding the merits and shortcomings of those detection methods, how to simplify those methods and adapt them into POC diagnostics for infectious diseases with increased sensitivity, specificity, detection speed and easy-to-use remains a great challenge and the industry is still in its infancy. The fast development of nanotechnology, materials science, microelectronic techniques and microfluidic systems has made the integration of the above detection methods into portable POC devices to achieve specific and low cost POC detection with low sample volumes, increased speed and sensitivity, and ability of multiplex detection possible.
Table 1.
Comparisons of the biosensors for POC diagnostics of infectious diseases.
| Methods | Pros | Cons | Refs. |
|---|---|---|---|
| Electrochemical biosensors | High sensitivity, fast response and low cost | Weak stability and susceptibility to interference | [272], [273], [274] |
| Fluorescence biosensors | Rapid response, flexibility and experimental simplicity | High fluorescence background | [275], [276], [277] |
| SERS-based biosensors | High sensitivity, well multiplexing capabilities, no photobleaching and low background | Limited field of view | [177], [178], [278] |
| Colorimetric biosensors | Simplicity and visualized detection | Inability to achieve quantitative detection and limited sensitivity | [279], [280], [281] |
| Chemiluminiscence biosensors | High sensitivity, wide linear dynamic ranges, simple instrument | Time-consuming, and enzyme-dependent | [87], [282], [283], [284] |
| SPR-based biosensors | Label-free and real-time detection | Bulky equipment and expensive | [285], [286] |
| Magnetic biosensors | low cost, and high signal-to-noise ratio detection | shortage of miniaturized magnetic readout systems | [123], [258], [259] |
POC devices for infectious disease detection
Since the first POC device GeneXpert endorsed by the WHO in 2010 for the initial detection of HIV-associated tuberculosis, which integrated sample preparation, nucleic acid amplification, and detection into a single device that was able to complete nucleic acid detection with 2 h [287], there have been a series of POC devices fabricated for POC diagnostics of infectious diseases, such as the LIAT analyzer (IQuum) for HIV detection [288], Truelab Uno® for malaria detection [289], GenePoC for Influenza A and B virus detection [290], and so on. Recently, due to the frequently spread of infectious diseases, especially at the circumstance of the COVID-19 outbreak, a research heat wave has been set off in the development of POC devices. As a result, there have been a series of POC devices developed for the detection of SARS-CoV-2 [291], [292], [293]. Part of current commercial available POC devices for infectious diseases can be found in Table 2 [48], [294], [295], [296], [297]. However, most of those established POC devices are just simple transplantation of conventional steps and most of them are not highly integrated and automated despite their high-cost and bulky instruments, there is still a huge space to develop portable, highly-integrated, rapid, sensitive, low cost, and fully-automated POC devices and the POC diagnostic industry is still in its true infancy stage. There are various problems remaining unaddressed, which call for more novel design and fabrication efforts to contribute to the development of POC devices. The most important challenge in developing such practicable POC devices is how to seamlessly integrate all the processes of detection methods into a complete automated instrument and achieve rapid, sensitive, and sample-to-result detection. Recent booming development of microfluidics, nanotechnology and MEMS technologies has greatly facilitated the development of POC devices. Thus, a series of POC devices have been designed and fabricated for the detection of infectious diseases, including LOC devices, LOAD devices, μPADs, lateral flow devices, miniaturized PCR devices, INAA devices, and so on.
Table 2.
Current commercial POC devices for infectious diseases.
| Devices | Types | Manufacturers | Pathogens | Time | Refs. |
|---|---|---|---|---|---|
| GeneXpert | Miniaturized PCR device | Cepheid | Tuberculosis and HIV | 120 min | [287] |
| LIAT analyzer | Miniaturized PCR device | IQuum | HIV | 60 min | [288] |
| FilmArray | Miniaturized PCR device | BioFire Diagnostics | Respiratory viruses and bacteria | 60 min | [298] |
| mL FluAB-RSV assay | Miniaturized PCR device | Enigma | Influenza A and B viruses | 95 min | [299] |
| ALERE i | INAA device | Alere | Influenza A and B viruses | 15 min | [300] |
| Revogene® | LOAD device | Meridian Bioscience | Respiratory viruses and bacteria | 70 min | [301] |
| Simplexa™ | LOAD device | Focus Diagnostics | Influenza A and B viruses, respiratory syncytial virus | 150 min | [302] |
| GenePoC | Miniaturized PCR device | GenePoC | Influenza A and B viruses | 70 min | [290] |
| Truelab Uno® | Miniaturized PCR device | Bigtec Labs | Plasmodium falciparum and Plasmodium vivax | 60 min | [289] |
| Elecsys® Anti-SARS-CoV-2 | Lateral flow device | Roche | SARS-CoV-2 | 12 min | [291] |
| COVID-19 IgG/IgM Rapid Test Cassette | Lateral flow device | Zhejiang Orient Gene Biotech | SARS-CoV-2 | 10 min | [292] |
| COVID-19 IgM/IgG™ | Lateral flow device | Truvian Sciences | SARS-CoV-2 | 10 min | [293] |
| Abbott Panbio™ COVID-19 IgG/IgM Rapid Test | Lateral flow device | Abbott Diagnostics | SARS-CoV-2 | 10–20 min | [303] |
| Alere™ Malaria Ag P.f RDT | Lateral flow device | Abbott Diagnostics | Plasmodium falciparum | 15 min | [304] |
| SD BIOLINE | Lateral flow device | Abbott Diagnostics | Plasmodium vivax | 15 min | [305] |
| RapidChek Select™ | Lateral flow device | Romer Labs | Salmonella | 22–30 h | [306] |
| Reveal® Salmonella 2.0 | Lateral flow device | Reveal | Salmonella | 24 h | [307] |
| Tubex® TF | Lateral flow device | IDL Biotech | Salmonella | 22 h | [308] |
| Atlas® Salmonella SEN Detection Assay | Miniaturized PCR device | Roka Bioscience | Salmonella | 24 h | [309] |
In addition, thanks to the rapid development of hardware and software, smartphone has actually been a handheld miniature computer, which owns ubiquitous availability, extensive connectivity and appealing price. By using a smartphone as the primary user interface for detection, the total cost and overall time of detection are dramatically decreased owing to the elimination of centralized laboratory, specialized equipment and complex operation, making POC diagnostics more accessible and user-friendly [310], [311], [312]. Therefore, smartphone is always explored either as a detector or an instrumental interface to couple with other POC devices for the detection of various diseases due to its portability, wide accessibility, low cost and easily operation [313], [314], [315], [316], [317], [318]. Herein, we will summarize the design and fabrication of various POC devices toward detections of various infectious diseases, which would contribute significantly to the applications of POC devices for urgent prevention and control of emerging public health events like COVID-19.
Lab-on-a-chip (LOC) devices
LOC devices are highly integrated and automated manipulation microfluidic devices that can rapidly conduct sample inlet and preparation, reagent manipulation, detection as well as result analysis in a unique platform, which have significantly facilitated the transformation processes of POC devices towards miniaturization, portability, integration, and automation of multiple assay functions into a single instrument [319], [320], [321], [322], [323]. Compared with conventional detection methods, highly integrated LOC devices exhibited several considerable merits such as rapid detection, ease of multiplexing and integration, small reagent volume consumption, cost effectiveness, high accuracy, and high portability [321], [322], [324], [325]. As a result, LOC devices have attracted a good many attentions for POC diagnostic applications, especially for on-site detection of infectious diseases in resource-poor regions. Herein, fabrication and application of LOC devices for various infectious diseases detection were exhaustively discussed, including infections caused by parasites, bacteria, and viruses.
Parasite detection
As a typical and severe infectious disease caused by Plasmodium parasites continuously multiplying in human red blood cells, malaria has long been one of the severe worldwide health threats and caused hundreds of thousands of deaths every year [326], [327], [328], [329]. However, the fast and accurate detection techniques for malaria are in quite shortage, requiring rapid, sensitive and accurate detection of Plasmodium parasites to improve the diagnosis and effective treatment of malaria. Under this background, on-site detection instruments such as POC devices are in greatly and urgent need for malaria diagnosis especially in resource-poor regions, where malaria outbreaks are most severe [328]. In recent years, LOC devices have attracted significant attention to be applied for POC detection of malaria caused by Plasmodium parasites. As a paradigm, Kadimisetty et al. developed a simple, inexpensive, disposable, fully 3D-printed microfluidic reactor array for rapid POC detection of Plasmodium falciparum gDNA ( Fig. 8A) [330]. The 3D-printed microfluidic reactor integrated the extraction, concentration and isothermal amplification of nucleic acids in a variety of body fluids, yielding a multifunctional diagnostic platform, which was able to achieve both end-point colorimetric qualitative detection and real-time fluorescence quantitative detection. After being evaluated, the microfluidic reactor could detect Plasmodium falciparum in plasma samples by loop-mediated amplification (LAMP) within 50 min, and a low detection limit of 5.85 × 10−3 ng mL−1 could be achieved [330]. While in another study, a complementary metal-oxide-semiconductor LOC platform integrated Ion-Sensitive Field-Effect Transistors (ISFETs) to facilitate direct chemical-to-electronic sensing and LAMP was developed for specific, quantitative detection of kelch 13 gene of Plasmodium falciparum and C580Y single-nucleotide polymorphism (SNP) associated with artemisinin-resistant malaria (Fig. 8B) [331]. The LOC platform demonstrated for the first time to realize DNA quantification on-chip by using DNA samples derived from clinical isolates of Plasmodium falciparum. The detection limits for kelch 13 gene and C580Y SNP were 1 × 102 copies mL−1 and 1 × 103 copies mL−1, respectively, which held significant potential for the development of a rapid, fully portable and sensitive POC device for quantitative malaria diagnostic and identification of mutations related to drug-resistance. Although those nucleic acid detection methods based on LAMP can provide highly accurate and sensitive diagnosis of malaria, the LAMP reagents are always very expensive and trained personnel are necessary, which causes difficulties in on-site diagnosis. Therefore, a film-based immunochromatographic LOC device was fabricated for malaria diagnosis [332]. Firstly, a microfluidic channel was patterned on a polyethylene terephthalate double-sided adhesive film followed by being assembled with a polycarbonate film for the immobilization of capture antibodies against Plasmodium falciparum lactate dehydrogenase (LDH) or Plasmodium vivax LDH. Then, fluorescent labeled Pan-type monoclonal antibodies (mAbs) conjugates were prepared as the signal indicator. In the presence of Plasmodium falciparum LDH, three detection dots including test dot 1 (T1), test dot 2 (T2) and control dot (Ctrl) revealed fluorescence signals where Plasmodium falciparum mAb, Pan-type pLDH mAb and goat anti-mouse IgG were immobilized, respectively. In the presence of Plasmodium vivax LDH, T2 and Ctrl dots showed fluorescence signals while no signal was detected with the negative control (Fig. 8C). This fluorescence-based immunochromatographic LOC device was able to detect Plasmodium falciparum LDH and Plasmodium vivax LDH successfully with detection limits of 50 ng mL−1 and 100 ng mL−1, respectively [332].
Fig. 8.
(A) Schematic illustration of assembly of multifunctional 3D printed reactor array for rapid POC detection of Plasmodium falciparum gDNA. Reproduced with permission from Ref. [330] Copyright 2018, Elsevier. (B) Schematic of the metal-oxide-semiconductor LOC platform integrated ISFETs to facilitate direct chemical-to-electronic sensing and LAMP for malaria detection. (a) Cross-section of an ISFET fabricated in unmodified CMOS technology and equivalent circuit macromodel. (b) Schematic of the pixel circuit implemented as a source follower configuration where changes in Vout reflect changes in pH. (c) Setup showing the LOC platform including a motherboard printed circuit-board that facilitates data readout, a cartridge PCB hosting the microchip and microfluidic chamber, the microchip including an array of 4096 ISFET sensors and an external thermal controller. (d) Cross-section illustration of the LOC platform showing the reaction interface. Reproduced with permission from Ref. [331] Copyright 2019, Elsevier. (C) Schematic illustration of film-based immunochromatographic microfluidic device for malaria diagnosis. (a) An exploded view of the immunochromatographic microfluidic device. (b) Schematic illustration of the malaria parasite capture antibody immobilization procedure on the polycarbonate film. (c) The structure of the immunochromatographic microfluidic device. (d) Photograph of the IMD with a paper case. Reproduced with permission from Ref. [332] Copyright 2019, Springer Nature.
Taking advantages of the high-resolution of the digital camera of a smartphone, a novel smartphone-coupled LOC device was designed to detect HRP-II antigen of Plasmodium falciparum from diluted (10%) human whole blood [333]. In this system, the white LED flash and digital camera of the smartphone were particularly utilized as an incident light source and a detector; anti-HRP-II-conjugated submicrobeads were applied for capturing HRP-II in blood samples in the LOC device; Mie scatter from the sample in the optofluidic channels was employed as the detection signal; and finally a handheld device fabricated by optical mirrors and lenses was attached to the single smartphone ( Fig. 9A). After optimization, the device showed considerably high sensitivity to HRP-II antigen with a detection limit of 1 × 10−3 ng mL−1 in 10% blood and a linear range from 1 × 10−3 ng mL−1 to 1 × 10−2 ng mL−1, and the total assay time of the device was approximately only 10 min [333]. Although the digital camera of a smartphone can be used as a detector in a LOC device, it is not efficient enough for ultra-high sensitive optical detection required in immunodiagnostics. Therefore, more sensitive optical detectors are required to be integrated with smartphone-coupled LOC devices for immunodiagnostics, while the smartphone can serve as the power source, data transfer, storage and analysis, and display. As a paradigm, Ghosh et al. developed a smartphone-coupled LOC device using a microchannel capillary flow assay (MCFA) platform to perform ultra-high sensitive chemiluminescence based ELISA with lyophilized chemiluminescence reagents for the detection of malaria biomarker HRP-II [334]. The whole device was composed of a MCFA lab chip, a high-sensitive optical detector and a smartphone analyzer (Fig. 9B). Firstly, the MCFA platform with functionally microchannels was fabricated by thermoplastic COC for performing high sensitive chemiluminescence assay, which eliminated the liquid reagent handling, control of assay sequence and user intervention in a conventional chemiluminescent detection. Then, a custom designed high-sensitive optical detector, which was applied for detecting the chemiluminescent signal from the chip, was coupled with the smartphone via USB on-the-go port for real-time communication. Finally, the smartphone was mainly used for display, data transfer, data storage and analysis, and also worked as the host and power source for the optical detector. This LOC device was able to detect HRP-II in artificial serum with a detection limit of 8 ng mL−1, which is sensitive enough to detect active malarial infection [334].
Fig. 9.
(A) Schematic of the handheld smartphone-coupled LOC device for the detection of HRP-II antigen. (a) Isometric view of a rendering of the handheld device, less the case. (b) Top view of the handheld device. (c) Exploded view of the entire device. (d) Photograph of actual device with an iPhone 4 installed. Reproduced with permission from Ref. [333] Copyright 2014, SAGE Publishing. (B) Overview of the smartphone-based POC testing analyzer. (a) Functional concept of the MCFA lab chip with on-chip reservoirs in operation sequence to implement chemiluminescence based sandwich ELISA, (b) Schematic of the designed MCFA chip with detailed labeling of the microfluidic components and (c) Schematic of the developed smartphone-based POC testing analyzer. Reproduced with permission from Ref. [334] Copyright 2020, Springer Nature.
Bacteria detection
Bacterial infection has long been one of the most common infectious diseases that threats human health. Among various bacterial infectious diseases, tuberculosis has been one of the leading causes of death worldwide. What is worse, the multidrug-resistant tuberculosis (MDR-TB) always exhibits poor response to antibiotic therapy and bad clinical prognosis, and requires expensive systematical treatment [335], [336], [337]. Therefore, there has been an urgent requirement for fast, sensitive, and specific diagnosis of tuberculosis and MDR-TB. POC diagnostics using LOC devices show considerable advantages for the molecular detection of MDR-TB both in laboratory settings and resource-poor regions. In one study, Muhlig et al. developed a closed droplet based LOC device for the differentiation of six species of mycobacteria using SERS [338]. The platform consisted of a bead-beating module for disruption of bacterial cell and a LOC-SERS device for bacteria discrimination, which was able to differentiate mycobacteria in 1 h without extraction and further treatment of samples via the obtained SERS spectra by the LOC-SERS device. While in another work, Kukhtin et al. fabricated a Lab-on-a-Film disposable device for the detection of MDR-TB from sputum extracts [339]. The Lab-on-a-Film device was composed of 203 gel elements printed on a flexible film, including DNA sequences for 37 mutations, deletions, or insertion elements across 5 genes. Then, the film with printed gel elements was laminated to additional rollable films and formed a microfluidic flow cell ( Fig. 10A). This device was an entirely closed-amplicon system that integrated multiplex amplification, hybridization and post hybridization wash steps into a single microfluidic chamber without buffer exchanges or other manipulations. The Lab-on-a-Film device was finally demonstrated to be capable of highly sensitive detection of MDR-TB from sputum extracts with a detection limit of 32 CFU mL−1, and was able to genotype MDR-TB by computing ratios of mutant to wild-type hybridization signals from gel elements printed on unmodified and untreated film substrates [339]. However, the Lab-on-a-Film device was not able to achieve automatic and highly precise sample-to-result detection of MDR-TB without sample pretreatment. To achieve this goal, there are several challenges: 1) hard sample pretreatment due to the highly viscous and heterogeneous of the sputum; 2) difficulties in the lysis of acid-fast MTB bacilli; 3) diversity of MTB mutations; and 4) high-cost [340]. Therefore, to address those challenges in the POC diagnostic of MDR-TB and further make the Lab-on-a-Film device an automatic workstation, Kukhtin et al, then integrated and automated MagVor homogenization, TruTip purification, and Lab-on-a-Film assembly (LFA) amplification in a multi-sample, sputum-to-genotype system for automated genotyping of MDR-TB directly from raw sputum (Fig. 10B) [340]. Briefly, an automated MagVor using magnetic rotation of an external magnet to induce chaotic mixing of glass beads by the rotation of a disposable stir disc was used for sample homogenization and cell lysis. Then, a TruTip using a pipet tip with an embedded matrix was employed for nucleic acid isolation and purification. Finally, a Lab-on-a-Film device with 203 porous three-dimensional gel elements was utilized to genotype multiple mutations and cut the cost of MDR-TB detection. Noteworthy, this LFA served as a platform for amplification, hybridization, washing, and fluorescent imaging in a closed format and thus prevented amplicon contamination of the workspace. The system was able to detect MDR-TB sensitively with a low detection limit of 43 CFU mL−1 from raw sputum [340].
Fig. 10.
(A) Schematic illustration of the Lab-on-a-Film disposable device for the POC detection of MDR-TB from sputum extracts. Reproduced with permission from Ref. [339] Copyright 2019, the Royal Society of Chemistry. (B) Schematic illustration of the benchtop automated sputum-to-genotype system using the LFA for the detection of MDR-TB. (a) Photograph of the interior of the analyzer depicting the MagVor for lysis and homogenization, the TruTip for purification, the deepwell plate so that all liquids are part of the consumable, pipetting station, the LFA, the thermal cycler, and the imager. (b) Layout of the consumable for six samples in which the following steps occur. (c) Illustration of the LFA on the system with the thermal cycler in the (1) “up” (disengaged) position, (2) in the “down” (engaged) position, and (3) with the imager capable of individually imaging and analyzing each array of gel elements. Reproduced with permission from Ref. [340] Copyright 2020, American Chemical Society.
Besides, E. coli, Staphylococcus aureus, and Salmonella enterica are also among the most common bacterial infectious diseases. In the past years, various LOC devices were developed for their rapid and sensitive detection [341], [342], [343]. As a paradigm, Ha et al. implemented a TEM-1 β-lactamase system engineered zinc finger protein (ZFP) arrays for the direct detection of pathogen-specific DNA sequences of E. coli O157:H7 using a cyclic olefin copolymer (COC) based LOC device ( Fig. 11A) [344]. The engineered ZFPs were immobilized on the COC chip, which could function as a non-PCR based molecular diagnostic device, to directly detect double-stranded DNA of E. coli O157:H7 with high specificity and sensitivity. The system could achieve a low detection limit of 1 nM target DNA, which can be further developed into a simple and novel LOC device for the detection of various pathogens. In another study, for the purpose of simplifying the entire procedures from sample input to quantitative nucleic acid output without external power sources, Yeh et al. developed a self-powered, highly integrated LOC diagnostic device for direct POC quantitative detection of methicillin-resistant Staphylococcus aureus (MRSA) DNA from blood without separate sample preparation steps (Fig. 11B) [345]. The device integrated three technologies into one chip: reagent microfluidic patterning, digital plasma separation for digital nucleic acid amplification, and an integrated vacuum battery on the chip. Briefly, an amplification initiator [magnesium acetate (MgOAc)] was prepatterned on the chip to enable digital nucleic acid amplification, followed by autonomous separation of the plasma into 224 microwells without any hemolysis via a simplified sample preparation step. Finally, an integrated vacuum battery was added to the chip to achieve self-powered microfluidic pumping without any external pumps, controllers, or power sources. This simple, autonomous, portable, LOC device allowed rapid quantitative detection of MRSA DNA directly from human blood samples in 30 min by isothermal recombinant polymerase amplification (RPA). The device was able to achieve a wide detection range from 1 × 104 to 1 × 108 copies mL−1, providing promising foundations for future low cost molecular diagnostic assays [345].
Fig. 11.
(A) Schematic of a COC based LOC device using a ZFP array for POC detection of E. coli O157:H7. (a) An image of the COC array spot chip. (b) A schematic diagram of the ZFP array on the COC chip. Reproduced with permission from Ref. [344] Copyright 2018, the Royal Society of Chemistry. (B) Schematic illustration of the SIMPLE chip for low cost, quantitative, and portable nucleic acid testing. (a) Samples can be simply dropped into the inlet for automatic sample preparation with minimal handling and no external pumps or power sources. (b) The end-point isothermal digital amplification of nucleic acid was done (RPA) by incubating the chip on a reusable heat pack. Scale bar, 2 mm. (c) Dye-loaded chip for visualization of microchannels. (d) Digital microfluidic patterning enables reagent patterning with common laboratory equipment. (e) Side view of the digital plasma separation design. (f) The vacuum battery system frees the chip from external pumps or power sources for pumping. Reproduced with permission from Ref. [345] Copyright 2017, AAAS.
Virus detection
Viruses always cause very serious fulminating infectious diseases, such as Ebola, HIV, SARS, COVID-19 and so on. Therefore, rapid, sensitive and accurate detection of viral infectious diseases at POC is of significant importance for the prevention and treatment of those diseases. LOC device is one of the most popular alternatives to achieve such a goal. As one of the sever hemorrhagic fever viruses, Ebola causes a disease characterized by systemic viral replication, immune suppression, abnormal inflammatory responses, major fluid and electrolyte losses, and high mortality, which always leads to major epidemics in resource-poor regions [346]. Thus, portable POC detection is urgently needed for its control and treatment. Du et al. developed an automated microfluidic sample preparation multiplexer device combined with a sensitive liquid-core antiresonant reflecting optical waveguide biosensor chip for the detection of Ebola RNA ( Fig. 12A) [347]. In this system, thermally stable 4-formyl benzamide functionalized magnetic beads with a high density of capture probes and metered air bubbles controlled by microvalves were used to enhance the hybridization of the target Ebola virus RNA with the capture probes bound to the beads. The system was able to achieve a wide dynamic range that covered the whole clinically applicable concentration range and the detection limit was as low as 1.05 × 10−6 ng mL−1 for clinical samples without target amplification. The whole detection process could be completed within 2 h [347]. However, this system is too complex and requires long sample preparation time, which is time-consuming and cannot meet the requirement of rapid detection. To simplify the approaches of the device and endow it with higher specificity and sensitivity, the authors then improved the system by using a sequence specific barcode fluorescence reporter and a photocleavable capture probe. The new device and protocol could detect Ebola virus RNA in raw blood with clinically relevant sensitivity by the sensitive and compact fluorometer (Fig. 12B). Exploiting a pneumatic multiplexing architecture, the device was able to run up to 80 assays in parallel samples on a single chip and achieved a detection limit of 3.37 × 10−3 ng mL−1 for Ebola RNA in raw blood within 90 min without target amplification. Importantly, this protocol did not require time-consuming and problematic off-chip probe conjugation and washing, which significantly simplified the sample preparation and detection procedures for POC detection Ebola virus [348]. Nevertheless, the fluorometer in this system was not integrated with the microfluidic systems, which cannot achieve automated detection without any external manipulation. Therefore, to further produce a simple, rapid, sensitive, and integrated LOC device for automated POC detection of Ebola virus, the authors combined an automated and multiplexing CRISPR microfluidic chip with a custom designed benchtop fluorometer for rapid detection of Ebola virus with very low volume (∼ 10 μL) (Fig. 12C). The device was capable of automated microfluidic mixing and hybridization, after which the nonspecific cleavage products of Cas13a were immediately measured by a small custom integrated fluorometer. Exploiting the collateral cleavage of CRISPR technology, the integrated device was demonstrated to be able to detect total Ebola RNA in 5 min without solid phase extraction, and a considerably low detection limit of ∼ 1 × 10−3 ng mL−1 was achieved [349].
Fig. 12.
(A) Schematic of an automated microfluidic sample preparation multiplexer (SPM) device combined with a sensitive liquid-core antiresonant reflecting optical waveguide biosensor chip for Ebola detection. (a) Design of the SPM with six incubation reservoirs for target preparation. (b) Photograph of the SPM. (c) Schematic of the solid-phase extraction process and assay. Reproduced with permission from Ref. [347] Copyright 2017, Elsevier. (B) Schematic of (a) Design of the SPM for on-chip solid-phase extraction with 80 incubation reservoirs. (b) Photograph of the SPM. (c) Schematic of the on-chip solid phase extraction experiment. Reproduced with permission from Ref. [348] Copyright 2017, American Chemical Society. (C) Schematic of automated CRISPR microfluidic chip for Ebola virus detection. (a) Design of the automated CRISPR microfluidic chip. (b) Blow up of design of the fluidic layer. (c) Open (left) and closed (right) states of a microvalve. (d) Design of a benchtop fluorometer system integrated with a microfluidic device for in situ virus sensing. (e) Photograph of chip and detection system. Reproduced with permission from Ref. [349] Copyright 2019, American Chemical Society. (D) LOC design for the RT-PCR: (a) design of the polymer RT-PCR LOC; (b) photograph of fabricated RT-PCR LOC including 5 different chambers; and (c) embedded micro pinch valves using silicone tubes. Reproduced with permission from Ref. [352] Copyright 2008, the Royal Society of Chemistry. (E) Schematic illustration of the assembled nucleic acid cassette. Reproduced with permission from Ref. [353] Copyright 2010, Springer. (F) Self-digitization chip for HIV detection. (a) Schematic of the self-digitization chip. 16 × 64 array of 6.5 nL microwells connected by channels is embedded in PDMS and covered with a PDMS-coated glass slide. (b) Photograph of the self-digitization chip. Reproduced with permission from Ref. [355] Copyright 2018, the Royal Society of Chemistry.
In the past years, HIV had become one of the leading causes of human death from a single infectious agent, emphasizing the requirement of POC diagnostics for HIV infection [81]. Although rapid detection of HIV infection can be achieved using anti-HIV antibodies by lateral flow immunoassays, they cannot detect the disease during seroconversion with sufficient sensitivity [81], [350], [351]. LOC device shows the potential for enhancing the sensitivity and reducing the HIV acute phase detection window. For instance, Lee et al. designed an innovative disposable, polymer based LOC device for clinical POC diagnostics of HIV by RT-PCR based chemiluminescence assay (Fig. 12D) [352]. The device integrated on-chip RT-PCR with a portable analyzer that consisted of a non-contact infrared based temperature control system for reverse transcription process and an optical detection system for on-chip detection as well as monitoring the RT-PCR LOC. The developed polymer LOC device was demonstrated to be capable of achieving POC detection of HIV in shorter than 1 h with minimized cross-contamination. Besides, Chen et al. developed a self-contained, integrated, disposable and sample-to-result polycarbonate microfluidic cassette for nucleic acid detection of HIV at POC [353]. The cassette comprised on-chip sample lysis, nucleic acid isolation, enzymatic amplification, amplicon labeling, and detection, which was automatically controlled by an analyzer that provided pouch and valve actuation with electrical motors and heated for the thermal cycling (Fig. 12E). The device was finally demonstrated to detect HIV virus in saliva samples with a detection limit of 0.05 ng mL−1. Most recently, Wang and Chiu et al. took advantages of fast, isothermal property and similar sensitive to qPCR of nucleic acid sequence based amplification (NASBA) and developed a digital format of nucleic acid sequence-based amplification using a self-digitization chip platform, a stationary water-in-oil nanolitre reactor generation chip with the ability to allow an incoming aqueous fluid spontaneously divides itself into an array of chambers that have been primed with an immiscible fluid [354], for the quantification of HIV-1 RNA (Fig. 12F) [355]. The self-digitization chip was demonstrated to be more sensitive and accurate than the real-time quantitative NASBA for quantifying HIV-1 RNA in plasma samples, and a low detection limit of 1 copy per μL of HIV-1 RNA was achieved, which is very promising for utilization in resource-poor regions.
Besides, LOC devices applied for other viral infectious diseases detection have also been widely investigated. In one study, an integrated LOC device using a magnetic nanoparticle immunoassay was developed for rapid detection of Hendra virus antibodies in dilute horse serum samples [356]. The device was characterized by the chaotic fluid mixing and the automatic liquid handling system, which was capable of detecting Hendra virus antibodies within 15 min and the detection limit was as low as 0.48 ng mL−1. In another example, Zhu et al. designed a LOC device that integrated DNA extraction, solid-phase PCR and genotyping detection for high-risk HPV [357]. In this device, the chip size was significantly reduced by applying the pneumatic microvalves for organically combination of the fluid mixing and reagent storage. Moreover, to allow the spatial separation of the primers and reduce undesirable interactions in multiplex amplification, the solid oligonucleotide array were incorporated into the chip. After being tested using positive quality control samples and HPV patient cervical swab specimens, the LOC device was capable of detection of HPV virus within 1 h, whose detection limit was approximately 50 copies of HPV virus per reaction, providing great utility for the screening and monitoring of HPV genotypes [357]. Furthermore, for the rapid and POC detection of H1N1 influenza virus, a self-driven microfluidic device integrated with a reverse transcription LAMP assay was developed [358]. The device comprised virus isolation via H1N1-specific aptamers conjugated magnetic beads, virus lysis, isothermal nucleic acid amplification, and colorimetric detection of the virus, which was capable of performing sample pretreatment, reverse transcription LAMP, and H1N1 virus detection on a single chip. The entire detection could be finished within 40 min and a low detection limit of 0.08 ng mL−1 was achieved, which was sensitive enough for clinical applications [358]. Most recently, to circumvent the elaborate and labor intensive sample preparation of PCR-based methods for the identification and quantification of viruses, Hügle et al. fabricated a LOC device for free-flow electrophoretic preconcentration of viruses and gel electrophoretic DNA extraction [359]. The device integrated free-flow electrophoretic preconcentration of viral particles with thermal lysis and gel-electrophoretic nucleic acid extraction on a single chip, which was able to achieve high sensitive detection of bacteriophage PhiX174 with a low detection limit of 0.5 × 10−4 ng mL−1. Further, to simplify the operation of the device, a custom-made chip holder together with a compact peristaltic pump and power supply were introduced to the device, making it user-friendly, low risk of cross-contamination and high potential for automation [359].
In the past years, barcodes have been widely investigated for disease diagnostics due to their ability to allow simultaneous detection of multiple targets from patient samples with increased analysis speed and improved the precision and accuracy. However, their clinical applications were limited by the lack of sensitivity and expensive read-out device. To address this problem, Ming et al. engineered a simple and low cost chip-based wireless multiplex diagnostic device by combining quantum dot barcoding technology with a smartphone reader and isothermal amplification with expectation to detect multiple pathogens in patients with multiple blood-borne infections ( Fig. 13) [360]. The results of analytical performances of this device demonstrated that the device was capable of rapid and simultaneous detection of HIV and HBV gene in a single test with a detection limit down to 1 × 103 copies mL−1 in less than 1 h, which could also be extended to develop molecular testing panels for other important pathogens by simply modifying the barcode recognition molecules to the target of choice [360].
Fig. 13.
Schematic illustration of the smartphone-coupled LOC device utilizing quantum dot barcodes. (A) Assay involves the addition of patient samples to a chip coated with microbeads, which are optically barcoded by quantum dots and are coated with molecules that recognize a target analyte. (B) Typical microwell chip containing different barcodes in each well. (C) Smartphone camera captures the image of four different quantum dot barcodes arrayed on the surface of the chip. (D) Two excitation sources excite the quantum dot barcoded chip independently. (E) Image of the smartphone device. Reproduced with permission from Ref. [360] Copyright 2015, American Chemical Society.
Lab-on-a-disc (LOAD) devices
LOAD device is a kind of centrifugal microfluidics, which integrates sample preparation, solution translocation, reagent mixing, separation and detection in a single device [57], [58]. Compared with LOC devices, LOAD devices exhibit advantages of automatic fluid actuation without tubing, external pumping, actuators or active valves, a high degree of parallelization, and a huge capacity for total integration [361], [362], [363], [364], [365], [366], [367]. In a LOAD device, the movement of the solution is controlled by the synthetic action of centrifugal forces, capillary forces, Euler forces and Coriolis forces, which can be easily manipulated by modulating the spinning speed, spinning direction, valves or rotation actuated barriers, making the device be able to work smoothly and efficiently with a simple spindle motor and a sophisticated microfluidic design [57], [58], [367], [368], [369], [370]. In the past years, LOAD devices have experienced great prosperity and development, which have been regarded as one of the most promising alternatives for the development of automated POC devices. As a result, LOAD devices have attracted considerable attention in the field of biomedical detection and diagnostics. Herein, recent progresses of LOAD devices for POC detection of infectious diseases were comprehensively provided.
Parasite detection
As described above, centrifuge-based LOAD device has shown great promising for integrated POC diagnostic application, but the centrifugal force is non-linear and energy-intensive due to the extremely high rotational frequencies upon the spinning [361]. Therefore, to circumvent those drawbacks, Choi et al. developed a real-time fluorescence LOAD device using energy efficient non-centrifuge based magnetic field interactions for field detection of Plasmodium falciparum and Plasmodium vivax with automated and scalable sample preparation capability [371]. The device was composed of a compact analyzer and a disposable microfluidic reagent compact disc, which seamlessly integrated parasite DNA sample preparation and subsequent real-time LAMP detection on a single microfluidic compact disc, and was driven by energy efficient non-centrifuge based magnetic field interactions ( Fig. 14). There were four parallel testing units in each disc, which were able to automatically conduct the parasite DNA binding, washing, elution and immediate real-time isothermal amplification and fluorescence detection independently. Thus, the disc could be configured either as four identical tests or as four species-specific tests, increasing the testing throughput. After systematical evaluation, the LOAD device was demonstrated to be capable of rapidly processing four samples simultaneously within 50 min turnaround time and a detection limit of ~ 500 parasites mL−1 for whole blood was achieved, which was sufficient for the detection of asymptomatic parasite carriers [371].
Fig. 14.
Schematic illustration of real-time fluorescence LOAD nucleic acid testing device for malaria detection. (A) Exploded view of the device, showing the assembly of various components. (B) Schematic of the assembled device and the quadplex microfluidic reagent compact disc. The form factor of the analyzer is palm-sized. The reagent compact disc is secured to the spindle platter. A real-time fluorescence sensing scheme is integrated on the analyzer. (C) Workflow of the device. Reproduced with permission from Ref. [371] Copyright 2018, Elsevier.
Bacteria detection
Kim et al. developed a LOAD device, which integrated the DNA extraction, isothermal RPA, and detection onto a single disc, for the detection of Salmonella ( Fig. 15A) [372]. The device applied a single laser diode for wireless control of valve actuation, cell lysis, and noncontact heating for the isothermal amplification, yielding a compact and miniaturized system. Meanwhile, antibody-coated magnetic beads were used in the detection system to enrich rare cells in large volumes of samples for the enhancement of sensitivity. Finally, lateral flow strips were utilized for the detection via direct visual observation. The device was capable of completing the entire detection procedures within 30 min automatically and rapidly, and low detection limits of 10 CFU mL−1 and 1 × 102 CFU mL−1 in PBS and milk were achieved respectively, indicating its great potential for molecular diagnostics of Salmonella [372]. In addition, Law et al. devised a robust LOAD device using RPA for automated real-time detection of MDR-TB [373]. In this platform, real-time RPA was integrated on a LOAD with on-board power to maintain temperature for DNA amplification, and a fluorescence detector was applied for the final detection. After being tested using spiked sputa samples collected from healthy volunteers, the device was demonstrated to be able to detect MDR-TB fast and sensitively with a detection limit of 1 × 102 CFU mL−1 within 15 min. However, the sensors used in this RPA-based LOAD device are complicated. LAMP is another isothermal amplification method, which requires easier method than RPA to be integrated in a LOAD platform for sensitive target detection [364], [374]. For example, Loo et al. developed a LOAD microfluidic platform using a real-time LAMP with SYTO-9 as the signal reporter to provide a sample-to-result solution for POC detection of MTB and Acinetobacter baumanii [375]. The microfluidic channels designed in this LOAD device was able to mimic sequential analytical steps in bench-top environment, in which the actuation of samples and reagents were automatically controlled by the centrifugation force. Notably, the LOAD system was self-contained, which seamlessly integrated DNA extraction, isothermal DNA amplification and real-time signal detection in a predefined sequence, thus making the LOAD device be able to achieve rapid and automated identification of MTB and Acinetobacter baumanii in 2 h with high sensitivity. The detection limits of MTB in sputum and Acinetobacter baumanii in blood were 1 × 103 CFU mL−1 and 1 × 102 CFU mL−1, respectively [375]. In another study, Seo et al. designed an integrated LOAD device for fully automated and colorimetric detection of E. coli O157:H7, Salmonella typhimurium, Vibrio parahaemolyticus and Listeria monocytogenes (Fig. 15B) [376]. The device integrated DNA extraction and purification, DNA amplification, and amplicon detection on a single disc closely. In this system, silica microbeads were incorporated in the disc for the extraction and purification of bacterial genomic DNA; LAMP was performed to amplify the specific genes of the targets; and a metal indicator (Eriochrome Black T) was used for the colorimetric detection via the color change of the LAMP mixtures from purple to sky blue. The whole detection processes could be rapidly completed in an automated manner by the LOAD device in 65 min with a low detection limit of 10 cells mL−1. However, the samples used in this strategy are in small volumes. Therefore, to increase the amount of a sample, the authors further developed a fully integrated LOAD device based platform with a large-volume sample for POC genetic analysis for multiplex bacteria including E. coli O157:H7, Salmonella typhimurium, and Vibrio parahaemolyticus (Fig. 15C) [365]. The platform consisted of a portable genetic analyzer and a fully integrated LOAD device. The integrated LOAD was combined with a 3D-printed solution-loading cartridge for automated operation including the bead-based DNA extraction, isothermal amplification by an Eriochrome Black T-mediated LAMP, and amplicon detection by a colorimetric and UV–vis detector. Finally, the LOAD-based system was demonstrated to be capable of automatically finishing the whole processes in 1 h using a large sample of 1 mL and its detection limit was as low as 1 × 102 cells mL−1. Most recently, inspired by the fidget spinner toy, Michael et al. engineered a diagnostic fidget spinner (Dx-FS) based on an instrument-free LOAD platform for rapid, low cost, electricity-free and colorimetric detection of urinary tract infection (UTI) ( Fig. 16) [377]. At the time of detection, just spun the Dx-FS, bacteria suspended in the sample would be enriched on a nitrocellulose membrane, which could be detected by the naked eye via colorimetric microbial detection. Moreover, a pressure equalization technique was used to circumvent the influence of operators because the Dx-FS was hand-powered. Finally, the Dx-FS was evaluated via a field test performed in India. The results showed that the Dx-FS was able to inform the presence of bacteria in urine within 50 min in contrast to turnaround times of several days, indicating the Dx-FS could be used in resource-poor regions as an inexpensive handheld POC device for the rapid concentration and detection of pathogens in urine samples [377].
Fig. 15.
(A) Schematic illustration of the LOAD device for POC detection of Salmonella. (a) Expanded view of the LOAD showing top and bottom plates made of polycarbonate, strip sensors, adhesive layer, and the metal heater. (b) Top view of a section of the disc featuring the chambers for cell lysis, isothermal amplification, metering, dilution, and detection. (c) Schematic illustration of the experimental setup. Reproduced with permission from Ref. [372] Copyright 2014, American Chemical Society. (B) Schematic illustration of the LOAD device. (a) The function components of the LOAD. (b) Assembly of the LOAD. (c) Photograph of the assembled LOAD. Reproduced with permission from Ref. [376] Copyright 2016, the Royal Society of Chemistry. (C) Schematic illustration of the integrated centrifugal disc. (a) The function components of the integrated centrifugal disc. (b) A digital image of the disc. (c) Components of the centrifugal microdevice. Reproduced with permission from Ref. [365] Copyright 2019, Elsevier.
Fig. 16.
The Dx-FS as a POCT device for low-resource settings. (A) Fidget spinner (left) and Dx-FS (right). (B) Sample enrichment (urine sample preparation) and UTI diagnostics performed by a Dx-FS. (C) Design of the Dx-FS with labeled parts. (D) Images from a high-speed video showing the operation of a Dx-FS: the device was (i) hand-spun, (ii) spun for 1 min and (iii) stopped. The areas colored in red denote the liquid flow during the process. (E) Angular rotational frequency of the Dx-FS versus time. Reproduced with permission from Ref. [377] Copyright 2020, Springer Nature.
Virus detection
As a paradigm, Liu et al. developed a novel LOAD platform integrated membrane-resistance (MembR) valves for fully automated sample-to-result subtyping of highly pathogenic avian influenza viruses (HPAIVs) ( Fig. 17A) [378]. In the LOAD platform, there were six MembR valves utilized, which enabled reagent pre-storage and all the processes including sample lysis, RNA extraction and purification, and specific RNA detection by real-time reverse transcription LAMP to be seamlessly integrated on a single disc. After being evaluated using three HPAIVs H7N3, H7N9, and H9N2 and two other influenza A subtypes H1N1 and H3N2, the LOAD device could achieve automated subtyping of those HPAIVs within 70 min [378]. In another study, Li et al. fabricated an automated sample-to-result disc for rapid HBV detection from whole blood based on a double rotation axes centrifugal microfluidic platform (DRA-CMP) (Fig. 17B) [379]. The disc was capable of conducting the pre-storage of reagents and integrating serum separation from whole blood, magnetic bead-based DNA extraction, aliquot of nucleic acid, and qPCR into the disc. The device was then demonstrated to be able to successfully execute sample-to-result detection of HBV in a 500 μL whole blood sample with a low detection limit of 1 × 102 copies mL−1. Besides, the whole detection process could be finished within about 48 min. All those properties of this device indicated its great potential for POC molecular diagnosis HBV in resource-poor regions [379].
Fig. 17.
(A) Schematic illustration of the sample-to-result LOAD platform. (a) Image of the platform appearance. (b) Photograph of the internal modules. (c) Schematic diagram of the main modules in the platform. (d) Schematic diagram of the non-contact heating and the optical module with the dichroic beam splitter. (e) Photograph of the assembled disc mounted on the rotary axis. (f) Assembly of the disc. (g) Design of the diagnostic disc. Reproduced with permission from Ref. [378] Copyright 2018, Elsevier. (B) Schematic illustration of the DRA-CMP for HBV detection. (a) Exploded view of the chip contains four layers. (b) Schematic illustration of the flow control module consisting of a DRA-CMP and a disc. (c) Schematic illustration of the waste chamber of the disc. (d) Picture of the disc. (e) Schematic of the microfluidic layout for sample-to-result HBV DNA detection from whole blood. Reproduced with permission from Ref. [379] Copyright 2019, American Chemical Society.
Microfluidic paper-based analytical devices (μPADs)
Since firstly introduced by Whitesides et al. in 2007, μPADs have enjoyed booming development for POC diagnostic applications due to their inherent advantages provided by the paper-based substrates, such as hydrophilic and porous structure for microfluidic channels fabrication, low cost, fold ability, disposability, favorable biocompatibility, no need of external driven force for sample transportation, and equipment independence [380], [381], [382], [383], [384], [385], [386]. Typical μPADs were composed of an arrangement of hydrophilic/hydrophobic microstructures patterned on paper substrates that integrated reagent storage, sample transportation, sorting, mixing, flow control and multiplex detection, making them ideal alternatives for automated field testing and POC diagnostic applications [385], [386], [387]. In particular, μPADs can be applied for target analysis in various aqueous solutions or biological fluids such as saliva, urine, serum, or even whole blood with simple operation and no external sources [382], [388], [389], [390]. As a result, μPADs have attracted increasing attention for POC diagnostic applications in the fields of biological detection, environmental protection, clinical diagnosis, and food safety, among which rapid POC detection of infectious diseases especially in resource-poor regions is one of the hottest fields.
Based on the fabrication methods of μPADs, μPADs can be classified into 2D μPADs, and 3D μPADs. 2D μPADs can be fabricated by patterning physical or chemical hydrophobic boundaries on the paper to produce liquid channels [380]. Besides, other strategies, including photolithography, laser treatment, plotting, cutting, inkjet etching, plasma etching, wax printing, screen printing and so on [391], [392], [393], [394], have been attempted for 2D μPADs development. Compared to 2D μPADs, the fabrication of 3D μPADs is more complicated to achieve more complicated reaction process and multiple functionalities, which are usually obtained by stacking the layers of 2D microfluidic paper [59]. Meanwhile, some newly developed methods can also be employed for the development of 3D μPADs, such as origami, slip techniques, open-channel, 3D-printing, and so on [59], [393], [394], [395], [396]. In the past decade, there have been a large number of novel μPADs reported for the detection of various infectious disease, which will certainly contributed to the prevention and control of the disease spreads. Herein, the development of μPADs for POC detection of infectious diseases was discussed.
Parasite detection
Taking malaria detection for an example, Chakma et al. developed a novel label-free spectrophotometric detection method following an indicator displacement assay for the detection of malarial biomarker histidine-rich protein II (HRP-II) [397]. Then, they incorporated this assay into a μPAD for simple POC detection applications in resource-poor regions. The μPAD was fabricated by printing hydrophobic alkyl ketene dimer on a chromatographic paper to create hydrophilic microchannels, test zone, and sample application zone. The device was able to achieve rapid and favorable detection of HRP-II within 5 min with a detection limit of 30 ± 9.6 nM. In particular, the sample used was as low as 20 ± 0.06 μL. These properties of this μPAD made it a promising POC device for rapid, selective, simple, portable, and inexpensive POC diagnostic of malaria [397]. In another study, Xu et al. developed a μPAD using the paper-folding technique of origami via wax printing to integrate the sequential steps of DNA extraction, LAMP, and array-based fluorescence detection for multiplexed malaria diagnostics from whole blood ( Fig. 18A) [398]. The device was demonstrated to be capable of rapid and sensitive detection of Plasmodium falciparum, Plasmodium vivax, and Plasmodium pan directly from a finger-prick volume of whole blood in 45 min with minimal user intervention.
Fig. 18.
(A) Schematic of paper-based multiplexed LAMP detection of malaria in blood. (a) Foldable paper devices: dark areas are printed with hydrophobic wax. The device consists of five panels (1–5) folding onto each other, and a plastic cover for LAMP processing to avoid evaporation. (b) The design incorporates alignment marks on two corners to assign the results. (c) Illustration of the extraction process. (d) By folding the device, the sample is transferred to the LAMP spots where the reaction is carried out. (e) The signal is read out using a UV flashlight (365 nm). Reproduced with permission from Ref. [398] Copyright 2016, Wiley. (B) Illustration of the main microfluidic technologies. The complexity/cost decreases from left to right. Reproduced with permission from Ref. [399] Copyright 2014, IOP Publishing. (C) Schematic illustration of laser-patterned paper-based device. (a) Schematic of the laser-based direct-write setup. (b) Schematic of the laser direct-write fabricated three-layer paper device for bacteria identification and antibiotic-resistance testing. Reproduced with permission from Ref. [403] Copyright 2020, Elsevier.
Bacteria detection
As a paradigm, Costa et al. fabricated a μPAD by an eco-friendly wax printing technology for colorimetric detection of MTB nucleic acids through their hybridization with species specific gold labeled oligonucleotide probes (Fig. 18B) [399]. Firstly, the wax printed 384 well plate was impregnated with a predetermined concentration of salt (MgCl2) to induce Au nanoprobe aggregation. In the presence of the complementary MTB DNA sequence, Au nanoprobe aggregation would be prevented and the paper surface remains red; in the absence of complementary MTB target, Au nanoprobe aggregation would be induced by the presented salt, resulting in a visible color change from red to blue in the paper surface. Besides, by using wax printing, Srisa-Art et al. developed a colorimetric μPAD coupled with immunomagnetic separation (IMS) for the detection of Salmonella typhimurium [400]. In this device, IMS anti-Salmonella coated magnetic beads were applied to capture and separate bacteria from the sample matrix for preconcentration and sample loading. Then, a sandwich immunoassay was conducted using β-galactosidase (β-gal) as the detection enzyme for bacteria quantification. After being tested using culturing solution and spiked bird fecal samples and whole milk, the device was demonstrated to be able to detect Salmonella typhimurium sensitively with detection limits of 1 × 102 CFU mL−1 for culturing solution, 1 × 105 CFU g−1 for starling bird fecal samples, and 1 × 103 CFU mL−1 for whole milk. In the past years, antimicrobial resistance (AMR) has been one of the global threats to human health, which has resulted in hundreds of thousands death all over the world every year [401], [402]. Therefore, to control infections caused by AMR bacteria better, He et al. developed a laser-patterned μPAD for the detection and susceptibility testing of E. coli via a simple visualized color change in resource-poor regions (Fig. 18C) [403]. The results indicated the μPAD a promising device for POC detection of bacterial infections, providing a novel tool for tackling the global challenge of AMR.
Virus detection
In the past decade, there have been a number of articles reported about development of μPADs for POC detection of virus infections, including influenza virus [404], chikungunya virus [405], HCV [406], West Nile virus [407], HBV [408], HPV [409], HIV [410], and so on. For example, Srisomwat et al. developed a pop-up DNA based 3D μPAD platform for label-free electrochemical detection of HBV DNA [408]. The 3D μPAD platform integrated multi-step operation of detection into a single device with facile sample introduction, avoiding contamination and minimizing biofluids exposure. The fluidic path, incubation time and electrical connectivity of the 3D μPAD platform were precisely controlled by the pop-up design through simply folding. Moreover, to achieve electrochemical detection, a pyrrolidinyl peptide nucleic acid, acpcPNA, was covalently immobilized onto the modified cellulose on the working electrode for target DNA capture. After systematically optimization, the device was demonstrated to be capable of sensitive detection of HBV DNA with high specificity and a wide linear range of 50 pM–100 nM, whose detection limit was as low as 1.45 pM. In another study, to avoid the fussy washing steps of conventional heterogeneous immunoassays, Tenda et al. developed an integrated “sample-to-result” 3D μPADs based on bioluminescence resonance energy transfer (BRET) switches for multiplex colorimetric detection of anti-HIV1, anti-HA, and anti-DEN1 antibodies in whole blood [410]. The device was composed of multiple paper layers that vertically arranged through lamination. The first layer served as the sample pad for cellular components separation. The second layer was impregnated with furimazine, which would be dissolved into the vertically flowing sample liquid and transported to the third layer immobilized with the BRET-switching protein (LUMABS). The design of the device enabled an automatic detection with sample volume independent and fully reagent-free operation. The only operation for users was limited to the introduction of a single drop of sample and acquisition of a photograph after sample introduction. All the procedures could be completed within 20 min without requirement for precise pipetting, liquid handling, or analytical equipment except for a camera, making it ideally suit for user-friendly POC diagnostics in resource-poor regions [410]. Recently, detection of influenza with antiviral treatment effectiveness indication is of great significance in reducing the treatment decision time, shortening the impact of symptoms, and minimizing transmission. However, current technologies cannot achieve this goal very well. To address this problem, Murdock et al. developed a novel, POC μPAD by wax printing for influenza detection with the ability to determine antiviral susceptibility of the strain for treatment decision [411]. The μPAD consisted of two wax-patterned filter paper layers and two outside layers of lamination. For specific and sensitive detection, the enzymatic activity of surface proteins present on all influenza strains were employed in this platform. Upon the sample introduction to the device, it would be distributed to 4 different reagent zones, and the results could be read either by eyes or through a colorimetric image analysis smartphone due to the development of the enzymatic substrate under different buffer conditions took place on the bottom of the device [411].
Lateral flow devices
Lateral flow devices are one of the most widely used POC diagnostic tools for the detection of infectious diseases owing to their advantages of simplicity, easy operation, rapidness, durable stability, cost-effectiveness, and minimum user intervention [412], [413], [414]. Since the first lateral flow device developed in the 1980s for pregnancy self-testing, there have been various lateral flow devices developed for all kinds of applications, including biomarker detection, disease diagnosis, food safety, environment monitoring and so on [415], [416], [417], [418], [419], [420]. A typical lateral flow device consists of a nitrocellulose membrane, sample pad, conjugate pad, and absorbent pad with all the required chemicals and reagents pre-stored in the test strip [393], [421]. The sample pad is designed for sample fluid loading and adsorption; the conjugated pad is impregnated with specific molecule-conjugated biolabels for binding to the analyte and signal generation; the nitrocellulose membrane immobilized with capture molecules is the core component for testing; and the absorbent pad provides the driven capillary forces and collects excess sample fluid [422].
Commonly, there are two formats of lateral flow devices, sandwich and competitive lateral flow devices. In a sandwich lateral flow device, a small volume of sample is loaded to the sample pad, which then flows across the conjugate pad, nitrocellulose membrane, and absorbent pad through capillary forces. The target in the fluid will specifically recognize the specific molecule-conjugated biolabels pre-stored on the conjugate pad and form target-molecule-conjugated-biolabels complexes, which are then captured by the capture molecules immobilized on the test zone and generate test and control lines, which can be directly read by the naked eye [412], [422]. Therefore, in a sandwich lateral flow device, the target concentration is proportional to the test line signal. In a competitive lateral flow device, the analyte target will compete with the capture molecules immobilized on the test zone to bind to the specific molecule-conjugated-biolabels pre-stored on the conjugation pad. Thus, in the presence of the analyte target, the test line will not display the signal, while in the absence of the analyte target, there will be signals at both the test line and control line [421], [423]. Thus, the target concentration is inversely proportional to the test line signal. Both sandwich and competitive lateral flow devices can be used for qualitative and quantitative detection of targets. The difference is that a sandwich lateral flow device is usually utilized for the detection of targets with multiple antigen epitopes, while a competitive lateral flow device is more suitable for the detection of a single antigen epitope targets [393].
Parasite detection
The past years have witnessed a booming development of lateral flow devices for POC detection of infectious diseases. Taking malaria detection as a paradigm, Pereira et al. developed a lateral flow device that integrated the concentration and detection steps into a single step for POC detection of a malaria biomarker named Plasmodium lactate dehydrogenase (pLDH) [424]. In this lateral flow device, a micellar aqueous two-phase system was applied in the 3D paper design to reduce the macroscopic phase separation time, which allowed simultaneously concentrate and detect pLDH within 20 min at room temperature. After systematic optimization, the device was capable of sensitive detection of pLDH with a low detection limit of 1.0 ng mL−1, which is 10-fold lower than a conventional lateral flow device. While, dos Santos et al. developed a simple, sensitive and low cost lateral flow device for the early diagnosis of Plasmodium falciparum malaria via an enzyme-based colorimetric detection of HRP-II, a specific biomarker for identifying the Plasmodium falciparum malaria contamination [425]. In this device, the colorimetric detection was achieved by the catalytic reaction of TMB substrate using a detection antibody labeled with the peroxidase enzyme, which could generate a colorimetric signal by the production of an insoluble product. The lateral flow device was able to detect HRP-II with a very low detection limit of 5 ng mL−1.
Bacteria detection
Roskos et al. developed a compact and inexpensive lateral flow device for the detection of MTB genomic DNA through the execution of isothermal DNA amplification in a mesofluidic cartridge that attached to a portable instrument [426]. In the device, one-way passive valves, flexible pouches, and electrolysis-driven pumps were used for the fluid handling inside the closed-system disposable cartridge, and the isothermal DNA amplification occurred in a two-layer pouch that permitted high-efficient heat transfer. Notably, the lateral flow device performed the detection in a closed-system cartridge also prevented workspace amplicon contamination. Finally, LAMP and Exponential Amplification Reaction (EXPAR) assays were coupled in this device for MTB genomic DNA detection, which could be completed within 20 min for LAMP based detection and 70 min for EXPAR based detection with detection limits of 30 copies mL−1 and 2 × 104 copies mL−1, respectively [426]. However, the sample preparation step was not integrated in the device, limiting its application for automatically sample-to-result detection. To develop a fully integrated device for POC nucleic acid detection of infectious diseases, Park et al. designed an integrated lateral flow device on a rotary microfluidic system for POC colorimetric detection of Salmonella Typhimurium and Vibrio parahaemolyticus [427]. The device was a highly integrated detection platform, which was able to conduct glass microbead based DNA extraction, LAMP, and colorimetric strip based detection in a sequential manner under a rotational speed control ( Fig. 19). Then, the device was applied for the multiplex detection of Salmonella Typhimurium and Vibrio parahaemolyticus in contaminated water or milk. The results showed that all the processes could be completed within 80 min and a low detection limit of 1 × 104 CFU mL−1 was achieved [427]. However, the above device requires off-chip reagent storage, complex operation steps and equipment-dependent nucleic acid amplification. To overcome these challenges, Tang et al. designed a fully disposable and integrated sample-to-result lateral flow device for POC nucleic acid detection of Salmonella typhimurium ( Fig. 20) [428]. This established device was composed of a sponge-based reservoir and a paper-based valve for nucleic acid extraction, an integrated battery, a positive temperature coefficient ultrathin heater, temperature control switch and on-chip dried enzyme mix storage for isothermal amplification, and a lateral flow test strip for colorimetric detection, which was capable of integrating nucleic acid extraction, helicase-dependent isothermal amplification and lateral flow assay detection into one paper device and allowing on-chip dried reagent storage and equipment-free nucleic acid amplification with simple operation steps. After being tested using spiked samples, the device was demonstrated to be able to sensitively detect Salmonella typhimurium with a detection limit of as low as 1 × 102 CFU mL−1 in waste water and egg, and 1 × 103 CFU mL−1 in milk and juice in about 1 h, indicating its great potential for future POC detection applications in resource-poor regions [428].
Fig. 19.
Schematic of the integrated lateral flow device on a rotary microfluidic system for POC colorimetric detection of Salmonella Typhimurium and Vibrio parahaemolyticus. (A) A digital image of the integrated rotary microdevice (B) Schematic illustration of the integrated rotary microdevice for the DNA extraction, the LAMP reaction, and the lateral flow strip detection. (C) Schematic illustration of the solid phase DNA extraction unit and the fluorescence images of the FAM-labeled DNA adsorbed glass microbeads (a), the LAMP amplification of target DNA (b), and (c) the lateral flow strip detection. Reproduced with permission from Ref. [427] Copyright 2017, Elsevier.
Fig. 20.
Schematic of a disposable and integrated lateral flow device for nucleic acid extraction, amplification and detection. (A) The different functional modules of the device, including nucleic acid extraction, amplification by the battery heat blocker and lateral flow detection. (B) The substrate and a bridge that was used to separate the upper layer copper sheet from the bottom layer to prevent the lysis buffer and washing buffer from flowing to the dry powder paper. (C) The top cover of the integrated lateral flow device. (D) The integration platform of the different functional modules and the substrate. (E) The model of this device. Reproduced with permission from Ref. [428] Copyright 2017, the Royal Society of Chemistry.
To make lateral flow device more user-friendly, convenient and intelligent, a smartphone-coupled dual lateral flow immunoassay device was developed by Cheng et al. using mesoporous core-shell palladium@platinum (Pd@Pt) nanoparticles as a signal amplifier for simultaneous detection of Salmonella Enteritidis and E. coli O157:H7 [429]. The system exhibited several attractive features: 1) using nanozyme (mesoporous Pd@Pt nanoparticles) for improving the sensitivity and stability; 2) using dual lateral flow immunoassay for simultaneous real-time detection of targets; and 3) a smartphone for imaging and results recording. After optimization, the device was capable of achieving simultaneous real-time detection of Salmonella Enteritidis and E. coli O157:H7 with greatly improved sensitivity and the detection limits were calculated to be ∼ 20 CFU mL−1 for Salmonella Enteritidis and ∼ 34 CFU mL−1 for E. coli O157:H7, respectively [429].
Virus detection
There has been a long history in the development of lateral flow devices for POC detection of viruses, which have also been applied for rapid and cost-effective detection of SARS-CoV-2-specific antibodies. For example, a lateral flow immunochromatography assay was employed for the detection of SARS-CoV-2 IgG/IgM in symptomatic patients presenting to the emergency department. The results displayed a satisfactory analytical performance in clinical practice, indicating the device a potential supplemental serology testing tool for the rapid diagnostic of COVID-19 [430]. Nucleic acid detection is the most commonly used and effective method for specific detection of viruses, which has been widely incorporated in the lateral flow devices to achieve simple, rapid, sensitive and user-friendly POC detection. For example, simple lateral flow devices based on reverse transcription LAMP had been developed and applied for rapid clinical detection of H7N9 virus and foot-and-mouth virus [431], [432]. However, conventional lateral flow devices have limited sensitivity and quantitative analysis, which can only identify targets of high cut-off values via colorimetric detection. To address those problems, nanomaterials were used to enhance its sensitivity and quantitative ability. Fu et al. developed a SERS-based lateral flow device for highly sensitive detection of HIV-1 DNA in the low concentration range ( Fig. 21) [433]. Raman reporter-labeled AuNPs were employed as SERS nano tags for targeting and detecting the HIV-1 DNA marker in the device. After optimization, the device showed favorable ability for the detection of HIV-1 DNA with a considerably low detection limit of 0.24 pg mL−1. Besides, Gong et al. developed a miniaturized and portable upconversion nanoparticle (UCNP)-based lateral flow device for POC detection of HBV DNA, which was composed of a lateral flow assay detection system, an UCNP-lateral flow assay reader and a smartphone-assisted UCNP-lateral flow assay analyzer [434]. The device displayed highly sensitive and quantitative detection capability of HBV DNA with a detection limit of lower by approximately 10 folds than the clinical cut-off value. Moreover, the precision and accuracy of the UCNP-based lateral flow device were also comparable to the gold-standard method with correlation coefficients of 0.922 and 0.995, respectively [434].
Fig. 21.
(A) Schematic illustration of the configuration and (B) the measurement principle of the SERS-based lateral flow device for quantification of HIV-1 DNA. (C is the control line and T is the test line). Reproduced with permission from Ref. [433] Copyright 2016, Elsevier.
For the purpose of realizing more accurate, faster and networked on-site monitoring and detection of avian influenza viruses (AIVs) with high sensitivity, Yeo and Park et al. developed a coumarin-derived dendrimer-based fluorescent lateral flow immunoassay device coupled with an efficient reflective light collection module and a smartphone ( Fig. 22A) [435]. The lateral flow immunoassay device exhibited a two-fold higher detectability as compared to that of the table-top fluorescence strip reader for H5N3, H7N1 and H9N2 subtypes, and could achieve rapid and highly sensitive detection of H5N1 virus within 15 min with a sensitivity of 96.55% (28/29) [95% confidence interval (CI): 82.24–99.91] and a specificity of 98.55% (68/69) (95% CI: 92.19–99.96) in a clinical study of H5N1-confirmed patients. Furthermore, the collected results measured by the smartphone in this system were able to be wirelessly transmitted to a centralized database system for further information processing and data mining via short messaging service. All those properties of the device made it a powerful tool for agile identification and efficient control of infectious diseases dissemination [435]. However, this fluorescent lateral flow immunoassay device was single channel designed and could only detect one pathogen at once, thus limiting its application for simultaneous multiplex analysis. To solve this problem, the authors then improved the lateral flow immunoassay device to a multiplex AIVs detection toolkit and readout system, which was equipped with a portable smartphone-based rapid dual fluorescent diagnostic system (SRDFDS) using two emission filters (Fig. 22B) [436]. The system was designed for simultaneously detection and quantification of both influenza A and its H5 subtype on a single strip using two QD signals (580 nm and 650 nm). The results demonstrated that the system was able to achieve a high sensitivity of 92.86% (13/14) and 78.57% (11/14), and a specificity of 100% (38/38, P < 0.0001) and 97.37% (37/38) for the detection of influenza A and H5 subtype respectively, indicating the system could be used for accurate and early diagnosis of influenza viral infections [436].
Fig. 22.
(A) Schematic of smartphone-coupled fluorescent lateral flow immunoassay device. (a) Schematic description of a smartphone-based fluorescence detector with a reflective light concentrator module. (b) Schematic of a fluorescence detector flow strip for the detection of AIVs, whose fluorescence signal is measured by the smartphone-based diagnostic device. (c) The detection processes of this device. Reproduced with permission from Ref. [435] Copyright 2016, Ivyspring International Publisher. (B) Overview of the smartphone-coupled rapid dual fluorescent lateral flow immunoassay based diagnostic system. (a) Schematic presentation of the multiplexed lateral flow strip with two simultaneous conjugates. (b) Schematic representation of the smartphone-based fluorescence detector with a reflective light concentrator module and two emission filters. Reproduced with permission from Ref. [436] Copyright 2018, Ivyspring International Publisher.
Miniaturized PCR devices
PCR is one of the most popular detection methods and gold standard of molecular diagnosis for infectious diseases, which enables the exponential amplification of one or a few copies of a particular sequence of DNA or RNA, making it very powerful, specific and sensitive [437], [438], [439]. Typical PCR processes were always conducted in an airtight tube with a small volume of no more than 100 μL, and the reaction were performed at temperatures cycled among around 60 °C, 72 °C and 95 °C. However, the common PCR machines are usually complex, bulky and expensive, which can only be used in resource-intensive centralized laboratories, thus precluding their applications for POC detection. The advent of microfluidics in combination with micro electro-mechanical systems (MEMS) technology, have enabled the miniaturization of PCR processes in a chip device with potential merits of ultra-fast speed, low cost and low sample consumption, portability, high throughput, high-integration and automation to be possible [440], [441], which would empower PCR-based diagnostics to be used for POC diagnostic applications in resource-poor regions. However, conventional PCR processes integrate precise heating and cooling control modules to perform thermal cycling increase the challenges and cost for miniaturization, which calls for more sophisticated and careful design to optimize thermal properties [442], [443]. The emergence of qPCR have realized nucleic acid amplification and quantitative real-time detection in a single step without electrophoresis or hybridization steps, which has been the pioneer of PCR-based integrated NAT despite its complex and precise heating and cooling control [444], [445], [446]. To realize rapid or ultrafast NAT, simplifying and shortening the conventional PCR processes are the two key challenges to be addressed. Due to the fact that the heating/cooling rates of conventional PCR systems are only a few degrees centigrade per second and a long time of 1–2 h is needed to complete the whole PCR processes, it is quite slow and the speed is urgently needed to be improved to achieve fast detection. Strategies including increasing the rate of heat transfer, using more efficient heaters, and shrinking of sample volume are the most commonly employed approaches [446], [447], [448], among which a small volume of sample is the most attractive option due to its cost-effective and increased heating/cooling rates, and the application of microfluidics enables very small volume samples used in PCR to be possible. In this part, the design and fabrication of miniaturized PCR devices either based on conventional PCR or qPCR for the POC detection of infectious disease were discussed.
Parasite detection
In a typical example, a handheld, battery operated, chip-based Truelab Uno® microPCR device was developed by bigtec Labs, Bangalore, India, for differential identification of Plasmodium falciparum and Plasmodium vivax parasites. The device was developed based on qPCR method, which was able to detect Plasmodium falciparum and Plasmodium vivax with similar sensitivity and specificity to the WHO nested PCR protocol based on the evaluation of 100 samples, and the detection limit was of< 5 × 103 parasites mL−1 for both Plasmodium falciparum and Plasmodium vivax [289]. In another study, Zaky et al. developed a novel backpack PCR methodology for molecular detection of Brugia malayi in mosquitoes, which integrated a rapid NaOH-based DNA extraction methodology, a portable battery powered PCR platform and a test strip-based DNA detection assay [449]. This backpack PCR diagnostic platform eliminated reliance on expensive and bulky instrumentation without compromising sensitivity or specificity of detection, providing an alternative to cost-prohibitive column-dependent DNA extractions that are typically coupled to detection methodologies requiring advanced laboratory infrastructure, and thus holding a great potential for POC detection of various pathogenic parasites.
Bacteria detection
In the past years, multiplex nucleic acid detection of pathogens has attracted special attention due to its benefits of rapid identification of pathogens and high clinical sensitivity [450], [451]. Manage et al. developed a multiplex cassette PCR device for the detection of enterohemorrhagic E. coli [452]. The cassette PCR contained 10 trenches with a total of 50 capillaries with microliter volumes of LCGreen dye loaded desiccated acrylamide gels to pre-store all the reagents required for the PCR. Finally, the detection was performed by taking CCD images of the entire cassette at each PCR cycle and melt curve analysis at each temperature. The device was capable of simultaneously detecting up to 10 targets of 8 samples within 75 min [452]. Moreover, Salman et al. developed a shunting PCR microfluidic device for nucleic acid detection of E. coli [453]. The device comprised a polycarbonate microfluidic PCR chip, a shunting thermal cycler and a highly sensitive lock-in fluorescence detector, which enabled rapid nucleic acid amplification with heating and cooling rates of 1.8 °C/s and 2 °C/s, respectively. The developed shunting PCR microfluidic device allowed performing low cost PCR amplification with annealing temperatures ranging from 54° to 68°C and the detection limit was 700 ng mL−1, which showed great potential to quantify PCR products in real-time without performing slab gel or capillary electrophoresis [453].
Up to date, the diagnosis of tuberculosis is still a challenge in the clinical medicine by using available conventional methods and it is more difficult to detect extra pulmonary tuberculosis (EPTB) due to the low yield of bacilli in the clinical specimens. For the purpose of addressing this challenge, a chip based TrueNAT device using RT-PCR was introduced for the diagnosis of pulmonary tuberculosis (PTB) and EPTB [454]. A systematic evaluation of the device was conducted using a total of 145 clinical samples including both PTB (80) and EPTB (65). All the results demonstrated that the sensitivity and specificity of TrueNAT device is 93.1% and 72.5% for the diagnosis of PTB and 96.77% and 76.4% for EPTB, respectively. Besides, the device was able to achieve fast detection of PTB and EPTB as well as rifampicin resistance within 2 h, which is portable and compatible for POC detection of Mycobacterium tuberculosis [454]. Moreover, an integrated continuous flow PCR (CF-PCR) and electrophoresis biochip based all-in-one microfluidic device was recently developed by Li and his coworkers for the detection of Porphyromonas gingivalis, Treponema denticola and Tannerella forsythia [455]. The all-in-one device consisted of an innovative pumping unit, two aluminum heaters and an integrated CF-PCR and electrophoresis microfluidic chip ( Fig. 23), which was able to achieve automated sample injection, rapid DNA amplification, and on-site PCR product detection in sequence without costly external precision syringe pump. The results showed that the target DNA amplification could be finished as short as 2′31″ and the detection of PCR products could be completed within 3′43″ with the minimum number of bacteria to be amplified as low as 1.25 × 105 CFU mL−1, indicating its great potential for POC NAT of pathogens [455].
Fig. 23.
The schematic for the (A) integrated CF-PCR-electrophoresis microfluidic chip and the all-in-one device. The photos of the (B) microfluidic chip, (C) prototype system, and (D) all-in-one device. Reproduced with permission from Ref. [455] Copyright 2019, the Royal Society of Chemistry.
Virus detection
As a paradigm, Zhou et al. developed a home-made microfluidic chip system for sensitive detection of SARS coronavirus [456]. This system was composed of a laser-induced fluorescence microfluidic chip analyzer, a glass microchip for both PCR and capillary electrophoresis, a chip thermal cycler based on dual Peltier thermoelectric elements, a RT-PCR SARS diagnostic kit, and a DNA electrophoretic sizing kit. This system could allow efficient cDNA amplification of SARS-CoV and followed electrophoresis on the same a glass microchip. Subsequently, the results of electrophoresis on the microchip were collected and analyzed via the laser-induced fluorescence microfluidic chip analyzer. The test results of this system showed that 17 positive samples were obtained from 18 samples of nasopharyngeal swabs from clinically diagnosed SARS patients, compared to only 12 positive results from the same 18 samples by the conventional RT-PCR with agarose gel electrophoresis detection, indicating the high positive rate, accurate and rapidity of the microfluidic chip system for SARS diagnosis [456]. Furthermore, due to the significant advantages of qPCR and its gold standard for molecular detection, miniaturizing qPCR devices would contribute to the POC detection of infectious diseases much. Ahrberg et al. developed a handheld qPCR device with an approximate size of 100 mm × 60 mm × 33 mm for on-site detection of H7N9 virus [457]. The device was capable of simultaneously performing four reactions in virtual reaction chamber form on a glass cover slip with a sample volume of ≈ 200 nL. The qPCR device was capable of detecting a single DNA copy of H7N9 virus, and the standard curve slope of the device was −3.02 ± 0.16 cycles at threshold per decade corresponding to an amplification efficiency of 0.91 ± 0.05 per cycle, suggesting its potential for rapid detection of infectious diseases in small clinics [457]. Moreover, in one of our previous work, a qPCR based POC device coupled with magnetic nanoparticle nucleic acid extraction was developed for automated sample-to-result detection of adenovirus [458]. The device comprised a 3D-printed closed cartridge and an automated instrument communicated with a computer via Bluetooth, which weighed 6.5 kg with dimensions of 27 × 27.5 × 25 cm (length × width × height). All the liquids and reagents needed were able to be pre-packaged in the cartridge to achieve fully automated sample-to-result detection. The magnetic separation based nucleic acid extraction and qPCR detection were conducted in the cartridge to avoid contamination. Due to the fact that there are five main steps for nucleic acid extraction based on magnetic separation (lysising, binding, multiple washing and elution), the first five centrifuge tubes (4 × 1.5 mL, 1 × 0.2 mL) of the cartridge were used to hold the nucleic acid extraction reagents, while the next three 0.2 mL centrifuge tubes were employed for the qPCR detection. After systematic evaluation, the device showed similar performance to commercialized qPCR device and manual method, indicating its promising for on-site detection of pathogens [458].
Isothermal nucleic acid amplification (INAA) devices
To circumvent the complex heating and cooling control in conventional PCR devices and increase the detection speed, a series of isothermal nucleic acid amplification methods have been developed for POC diagnostic applications, such as rolling circle amplification (RCA) [459], [460], [461], LAMP [354], [462], [463], [464], RPA [465], [466], strand displacement amplification (SDA) [467], [468], helicase dependent amplification (HDA) [469], NASBA [355], multiple cross displacement amplification [470], [471], [472] and so on. These isothermal nucleic acid amplification methods were conducted at a fixed temperature without complex thermal cycling equipment, making them more applicable for POC diagnostics [294], [442], [473], [474]. To achieve POC diagnostics, those INAA devices must integrate sample preparation, nucleic acid amplification, detection and even result analysis into one to realized sample-to-result detection, which are portable and can be friendly used by any unskilled personnel. Up to date, there have been a number of reports about INAA devices designed for POC nucleic acid detection of infectious diseases, and some INAA devices have already become commercially available [475], [476], [477], [478], [479], [480]. Herein, POC devices using isothermal nucleic acid amplification methods were systematically discussed.
Parasite detection
NAT of parasites is currently one of the most precise and accurate methods for the diagnostics of correlated infectious diseases. However, to achieve POC detection, various INAA methods have been coupled with microfluidic techniques to develop INAA devices due to their outstanding merits. However, manipulation of droplets in the channels of traditional microfluidic devices always needs external valves, pumps or pressure supplies, making the devices complicated. Digital microfluidics (DMF) can address this problem via an electrowetting force to manipulate an individual microdroplet on an array of electrodes and has been employed in the development of INAA devices. As a paradigm, Wan et al. developed a LAMP based handheld, automated and detection system free thermal DMF device called LampPort for the DNA detection of blood parasite Trypanosoma brucei ( Fig. 24A) [481]. Firstly, droplet manipulation and real-time temperature control systems were integrated into the handheld device. Then, the control software was installed on an external control system to communicate with the device via Bluetooth. In the experimentation, samples were loaded by an electrowetting force into sandwich-structured DMF chips for an on-chip LAMP reaction, followed by addition of a highly concentrated SYBR Green I droplet to mix with the reaction droplet for visualized detection. Notably, the device was able to prevent aerosol contamination due to the absence of air inside the reaction chamber in the DMF chip. This low cost and compact device can achieve sensitive detection of Trypanosoma brucei DNA without bulky optical system, and a similar detection limit of 4 × 103 copies mL−1 to a commercial thermal cycler was achieved [481].
Fig. 24.
(A) Schematic illustration of the handheld DMF LAMP device for the DNA detection of blood parasite Trypanosoma brucei. (a) 3D schematic of the components of the device. DC: direct current. (b) Schematic of the LAMP reaction and endpoint SYBR Green I denotation of positive LAMP products. (c) Side view of the DMF chip. (d) Droplet actuation and heating control electronics. (e) Schematic of the naked-eyed visualization of the LAMP results. Reproduced with permission from Ref. [481] Copyright 2019, Springer Nature. (B) Schematic illustration of the water-activated, self-heating, non-instrumented cartridge based LAMP device. (a) Exploded view and a photograph (b) of the water-activated, self-heating, non-instrumented cartridge for isothermal amplification of nucleic acids. Reproduced with permission from Ref. [482] Copyright 2011, the Royal Society of Chemistry.
Bacteria detection
For the NAT of bacteria, nucleic acid extraction and amplification are very important, which have not been seamlessly integrated in the INAA devices. How to seamlessly integrated the two procedures and achieve automated, portable, highly-sensitive, fast and easy-to-use POC detection remains a great challenge in this field. In the past years, the advent of microfluidic cartridge has presented an effective solution to address this problem. As a paradigm, Liu et al. developed a LAMP based self-heating cartridge for visual fluorescent detection of E. coli DNA [482]. The device was fully self-contained without any external instruments, which integrated water-activated exothermic heating with temperature regulation facilitated with a phase change material (PCM) and LAMP (Fig. 24B). Porous paper was used to control the rate of exothermic reaction by modulate the water supply to the exothermic reactor, whose temperature was maintained by the PCM material to range from 20 °C to 40 °C, supplying the power for LAMP. Finally, the visual fluorescent detection was achieved by using a reusable, keychain-mounted UV light. The easy-to-use device was able to detecting E. coli DNA consistently with a detection limit of as few as 2 × 103 copies mL−1, which was suitable for use at home, in the field, and in resource poor regions [482]. In another work, Kaur et al. developed a modular fluorescent paper-based LAMP device for the POC detection of MTB gDNA [483]. The device was very user-friendly and cheap with a 12-test-zone device cost of $0.88 and one reaction reagent cost of $0.43, which could be operated by an untrained user. Moreover, all reagents could be dry-stored in the device, facilitating storage and transportation without cold chains. Finally, the device was coupled to an inexpensive imaging platform, which enabled filter-free fluorescence detection of amplified DNA via a cell-phone camera. The device exhibited high specificity to MTB in various complex samples with a low detection limit of 10 copies of MTB gDNA. More importantly, the clinical sensitivity and specificity of the device in clinical sputum samples were 100% (zero false negatives) and 68.75% (five false positives), respectively (N = 30), when utilized Xpert MTB/RIF assay as the reference standard. All those results demonstrate that the device has the potential to provide affordable and accessible molecular diagnostics of MTB at the POC [483].
Virus detection
Rapid detection of virus by NAT is the good standard of the diagnostics of virus caused infections. INAA exhibits many superiorities compared with traditional PCR and qPCR, which has been widely investigated to develop INAA devices for POC diagnostics of infectious diseases caused by viruses. Recently, digital INAA has been one of the hottest research fields, which can be highly integrated with microfluidic devices to achieve miniaturization of the devices, and highly sensitive and precise detection of target nucleic acids. In one previous work, Yu et al. designed a self-partitioning SlipChip (sp-SlipChip) microfluidic device that could generate droplets by slip-induced flow to perform digital LAMP for POC detection and quantification of human papillomavirus (HPV) DNA [484]. Firstly, the sp-SlipChip containing a designed “chain-of-pearls” continuous microfluidic channel was designed to establish the fluidic path for liquids to be introduced into the device. When one simple manual slipping step was added, the aqueous solution could robustly self-partition into individual droplets by capillary pressure-driven flow. Then, digital LAMP was applied by this sp-SlipChip to perform quantitative analysis of HPV-16/18 nucleic acids extracted from clinical swab samples without cross-contamination observed between adjacent droplets, indicating the digital LAMP based device to be a promising diagnostic tool for POC quantitative analysis of viral load, especially in resource-poor regions [484]. Song et al. developed a microfluidic cassette based device using reverse-transcription LAMP for visualized detection of ZIKV virus [485]. The device was powered a chemically heated cup and achieved visualized detection via the leuco crystal violet dye without external instruments. After evaluation, the device was demonstrated to be able to complete rapid and highly sensitive detection of ZIKV in oral samples with a detection limit of 2.5 × 10−3 ng mL−1 within 40 min. Moreover, a portable, smartphone-controlled, automated LAMP based self-driven microfluidic device was developed for rapid colorimetric detection of H1N1 virus ( Fig. 25) [486]. The device integrated magnetic beads based sample purification, pathogen lysis, isothermal nucleic acid amplification, and colorimetric detection in one, which could be completed within 40 min and controlled by a smartphone. Finally, the device was demonstrated to achieve accurate and sensitive molecular diagnosis of H1N1 virus, whose detection limit was 0.8 ng mL−1, which is sensitive enough for clinical adoption.
Fig. 25.
(A) Exploded view of the self-driven microfluidic chip consisting of a PDMS microfluidic structure layer, a hydrophilic film, a PDMS hydrophobic layer, and a glass substrate. (B) A photograph of the chip. (C) Schematic illustration of the chip design. The area enclosed by the red dotted line represents the LAMP reaction module (module B), with the other area being used for sample treatment (module A). (D) Illustration of the portable system composed of two stepping motors, two sets of gears, two photo-interrupters, a punching-press mechanism, an Arduino-based microcontroller circuit, a thermal control module, a color sensor, and a microfluidic chip. (E) Experimental workflow of a sample-to-answer, potable platform for rapid pathogen detection. Reproduced with permission from Ref. [486] Copyright 2019, the Royal Society of Chemistry.
In addition, taking advantages of LAMP, we then developed a portable multi-channel turbidity device was for rapid nucleic acid detection of H7 subtype avian influenza virus [487]. The device was composed of a temperature control unit, photoelectric detection unit, turbidity calibration unit, power management unit, human machine unit, communication unit and ARM-based microcontroller with dimensions of 21 cm in length, 15.5 cm in width and 11.5 cm in depth. The testing results demonstrated that the LAMP-based POC device was capable of rapid and highly sensitive nucleic acid detection of H7 subtype of H7N9 avian influenza virus within 60 min, and a low detection limit of 10 copies mL−1 could be achieved. Owing to the current PCR devices require stable electricity and are not portable, Snodgrass et al. developed a portable LAMP based device called TINY for isothermal nucleic acid quantification with power from electricity, sunlight or a flame and the ability to store intermittent energy sources for operation when electrical power is not available or reliable, enabling the device to be used either in a centralized laboratory with electricity supply or in the field when electricity is unavailable [488]. The TINY was able to quantify target nucleic acid in a handheld package with both weight and volume approximately to be an order of magnitude smaller than commercial qPCR machines ( Fig. 26). Besides, the latent heat inside TINY was able to keep isothermal amplification for over an hour in case of power outages. Then, the TINY device was tested in two Ugandan health clinics to evaluate the system against commercial machines performing both qPCR and LAMP using the detection of Kaposi’s sarcoma-associated herpesvirus DNA as a model. The results showed that TINY device was of equivalent performance either powered by sunlight or electricity and an agreement of 94% with commercial qPCR machines for detection of Kaposi’s sarcoma-associated herpesvirus DNA using samples from 71 Ugandan patients was achieved [488]. Most recently, Sun et al. developed a portable LAMP device coupled with a smartphone to perform POC detection of live virus from nasal swab media, using a panel of equine respiratory infectious diseases as a model system for corresponding human diseases such as COVID-19 [489]. In this system, LAMP on a microfluidic chip was applied for target nucleic acid amplification and the smartphone was used as a detector for detecting at the end of reactions. The total assay could be completed in 30 min with high accuracy and sensitivity, whose detection limit was comparable to that of traditional lab-based test and PCR, indicating its great potential for clinical POC detection of infectious diseases such as COVID-19 [489].
Fig. 26.
Schematic illustration of the portable LAMP based device called TINY for isothermal nucleic acid quantification with power from electricity, sunlight or a flame. (A) TINY is portable and easily carried in one hand, which is far smaller than the GeneXpert IV by Cepheid (footprint outlined by the dark purple box), or the ViiA 7 Real-Time PCR System by Thermo Fisher Scientific (light purple box) (B). (C) TINY heated by a Bunsen burner through an opening in the bottom of the system. (D) TINY heated via electricity, using an integrated cartridge heater. (E) TINY heated via concentrated sunlight at the infectious diseases institute in Uganda. (F) A photograph of the measurement unit separated from the temperature-regulation unit. (G) A cross-section of the measurement unit. (H) LEDs are placed on the bottom side of the top PCB. (I) Looking down into the TINY system with the solar absorption plate removed. (J) A cross-section of the temperature-regulation unit. Reproduced with permission from Ref. [488] Copyright 2018, Springer Nature.
In the field of POC devices, development of portable easy-to-use devices that allow sample-to-result detection is the ultimate goal. Thanks to the fast development of nanotechnology, microfluidics and MEMS technology, a couple of portable devices integrated with various POC detection methods have been investigated for the detection of infectious diseases. In this part of this review, fabrications and applications of POC devices including LOC devices, LOAD devices, μPADs, lateral flow devices, miniaturized PCR devices, and INAA devices for the POC diagnostics of various infectious diseases were systematically presented and discussed. Despite all the distinct merits of those POC devices, they all failed to fully satisfy the ASSURED criterions, leaving many challenges in developing POC devices toward clinical applications. As illustrated in Table 3, LOC devices have advantages of faster, high-throughput and multiplex analysis, smaller sample volume, less power consumption and more efficient control and manipulation, despite the existing challenges in fabricating, sealing and interfacing LOC devices, and difficulty of multiplexing [56], [490], [491], [492]. LOAD devices displayed particular advantages of: 1) the liquids can be easily handled by centrifugal force via a low cost and controllable spinning motor; 2) can be easily applied for large-scale parallelization and multiplexing. But they are still limited by their relatively short shelf life, difficult on-board reagent storage, and are hard to deal with samples of very small volumes [492], [493], [494], [495]. μPADs are attractive POC devices that are very potential toward commercialization due to their low cost, easy fabrication and operation, no need of external pumps or power sources, and well compatibility. However, they are currently limited by their inherent shortcomings such as lack of flow control, difficulty of multiplexing and multi-step assays, limited sensitivity, inability of instant response and repeatable use [386], [496], [497]. Lateral flow devices have been widely deployed POC diagnostics due to their superiorities of user-friendly, low cost, easy miniaturization and development, and independent on equipment, they have limited capability to achieve quantitative detection, and the multiplexing capability and throughput need to be further improved [421], [498], [499]. For miniaturized PCR devices, they own common merits of traditional PCR devices of accurate quantification, small sample and high sensitivity, the deficiencies are high production cost, expensive complex device, and trained professional person are needed [294], [441], [500]. Finally, INAA devices are simple and high sensitive devices that can achieve fast POC detection. However, current POC diagnostics based on INAA devices are limited by high requirements for the primers and expensive reagents [501], [502], [503]. Among all those challenges, the most important challenge in the fabrication of POC devices for the detection of infectious diseases is how to integrate all the detection steps into a compact single device, including sample pretreatment, reaction, signal amplification and acquisition, and data analysis and display. The other important challenge is how to miniaturize the device and keep all its functions as well, which requires the cooperation work of several research areas such as nanotechnology, MEMS, microfluidics, 3D-printing, and materials science. Moreover, guaranteeing the performance and clinical utility of the device comparable to the gold standard methods used in the laboratory settings is also very crucial, which needs very precise design and fabrication. Besides, reducing the fabrication cost and making it an individual affordable device would guarantee the device to be very popular and own a broad market. Finally, in consideration of the complex situation of clinical diseases and mixed infection, achieving high throughput and simultaneously detection without any interference and minimal human intervention is very attractive and fascinating, and efforts have been made to achieve this goal.
Table 3.
Comparisons of the POC devices for the diagnostics of infectious diseases.
| Devices | Pros | Cons | Refs. |
|---|---|---|---|
| LOC devices | High-throughput, multiplex and faster analysis, small sample volume, less power consumption and more efficient control and manipulation | Difficulty of fabricating, packaging and interfacing, and difficulty of multiplexing | [56], [490], [491], [492] |
| LOAD devices | Low cost and easy control by centrifugal force, simple and independent liquid handling, large-scale parallelization and multiplexing | Short shelf life, difficult on-board reagent storage, and hard to work with very small volumes | [492], [493], [494], [495] |
| μPADs devices | Low cost, easy fabrication and operation, no need of external pumps or power sources, and well compatibility | Lack of flow control, difficulty of multiplexing and multi-step assays, limited sensitivity, inability of instant response and repeatable use | [386], [496], [497] |
| Lateral flow devices | Low cost, user-friendly, independent on equipment easy miniaturization and development | Difficulty of achieving quantitative detection, limited multiplexing capability and throughput | [421], [498], [499] |
| Miniaturized PCR devices | Accurate quantification, small sample and high sensitivity | High production cost, expensive complex device | [294], [441], [500] |
| INAA devices | Fast, simple device, and high sensitivity | High requirements for the primers and expensive reagents | [501], [502], [503] |
Conclusion and future perspectives
The current outbreak of COVID-19 worldwide has exposed our deficiencies in the prevention and control of infectious diseases, especially in the aspects of rapid detection and early screening, which urgently require portable POC diagnostics and correlated devices to contribute to rapid discovery of infections in time, especially in resource-poor regions where the medical conditions are always limited. In this review paper, we systematically summarized recent progresses in the development of POC detection methods and correlated POC devices for the diagnostics of various pathogens caused infectious diseases including parasites, bacteria and viruses. In the first part, detection methods for POC detection of infectious diseases were summarized, including electrochemical biosensors, fluorescence biosensors, SERS-based biosensors, colorimetric biosensors, chemiluminiscence biosensors, SPR-based biosensors, and magnetic biosensors. Then, the progresses in the development of various POC devices and their applications in the detections of various infectious diseases, including LOC devices, LOAD devices, μPADs, lateral flow devices, miniaturized PCR devices, and INAA devices, were particularly reviewed.
Despite the fast development of POC detection methods and correlated POC devices for infectious diseases, achieving fast, highly sensitive and accurate POC diagnostics of infectious diseases remains a big challenge in this field. The performances of a good practicable POC device should meet the “ASSURED” standard, which requires the device to be of fast detection, high accuracy, low detection limit, low cost, good repeatability and stability, and to be user-friendly. To achieve those goals, a highly integrated POC device able to achieve sample-to-result detection that can conduct sample pretreatment, reaction, signal amplification, detection, data analysis and reader automatically is the most essential [353], [504]. However, to achieve detection without sample pretreatment is always very hard especially for infectious diseases due to their complex components with human body fluids, which can interfere the detection results. Therefore, the future development of POC diagnostics for infectious diseases may focus on designing assays using original clinical samples with no need of sample pretreatment. The second challenge is that the detection speed and sensitivity of current POC devices should be improved, which mainly depends on the design of novel detection methods [505], [506]. The third challenge of the development of POC devices for infectious diseases is that the costs need to be decreased. Only when the costs of the devices are widely affordable can the POC devices realize their social values and benefit human health [507], [508].
The fast development of mobile networks, internet of things (IoT), machine learning and AI has facilitated the development of POC devices towards remote assay operation and interpretation, more user-friendly and convenient operation, fully automated and intelligent detection, and high-efficient result readout, enabling personalized diagnosis and therapy in a fast, low cost and high-efficient way [509], [510], [511], [512], [513], [514]. By using IoT, POC devices would be able to communicate with centralized location remotely, provide important information to the operators, and speed up the data transmission for big data analysis, which can contribute greatly to the prevention and control of pandemic such as COVID-19 [515], [516]. The current commercialized fifth generation of mobile network will provide a massively increased data rate and accelerate the application of IoT for POC detection. Besides, due to the mushrooming of machine learning in recent years, it has also been harnessed with POC devices to improve the medicine practice by integrating clinical detection and disease diagnostics based on big data and intensive evidences [511], [517], [518]. Finally, the deep learning methods in AI would advance the fabrication of POC devices towards AI-based devices, which will revolutionize medical diagnostics and endow them the ability to achieve automated and intelligent medical practice with precise diagnostics in complex biological environments at POC and provide personalized therapeutics in time. Those approaches will contribute to realize earlier identification of infectious diseases, which is more effective than that at the advanced stages, thus helping the prevention of pandemic outbreak [519], [520], [521], [522].
Finally, the data processing of POC devices is also very important. In the past years, the fast growth in cloud computing has made the data transmission, storage and analysis easier and more efficient and convenient. POC devices combined with cloud computing would accelerate the detection speed, enhance diagnosis accuracy and efficiency, which is of great significance to the control of infectious diseases and enables precise personalized medicines [314], [523], [524]. However, by using cloud computing, all the healthcare data concerning personal privacy will be uploaded to the cloud storage, which can be downloaded by different platforms, parties and even individuals. Thus, how to keep data safety of cloud computing is a great challenge. To address this problem, on the one hand, internet laws are essential to guide the data storage by cloud computing servers. On the other hand, new sophisticated encryption methods should be developed for cloud computing to guarantee the data safety.
CRediT authorship contribution statement
Chao Wang: Conceptualization, Writing-original draft. Mei Liu: Conceptualization, Writing-original draft, Review & Editing, Supervision. Zhifei Wang: Review & Editing, Supervision. Song Li: Review & Editing. Yan Deng: Review & Editing. Nongyue He: Conceptualization, Writing, Review & Editing, Supervision, Funding acquisition.
Declaration of Competing Interest
The authors declare that they have no known competing financial interests or personal relationships that could have appeared to influence the work reported in this paper.
Acknowledgments
This work was financially supported by the National Natural Science Foundation of China (Nos. 62071119, 61527806 and 61871180), the National Key Research and Development Program of China (Nos. 2018YFC1602905 and 2017YFA0205301), and Jiangsu Provincial Key Research and Development Program (BE 2018695).
Biographies

Chao Wang is received his B.S. degree in biomedical engineering in 2014 from Nanjing Medical University and M.S. degree in biomedical engineering in 2017 from Southeast University under the supervision of Prof. Nongyue He. Now, he is a Ph.D. student in the research group of Prof. Nongyue He at Southeast University in the Department Biomedical Engineering. His research interests focus on biosensors and electronic devices for the detection of tumor biomarkers and infectious diseases.

Mei Liu received his B.S. degree in Veterinary Medicine in 2014 from Northwest A&F University and M.S. degree in Biophysics in 2017 from Southeast University. He is currently a joint Ph.D. student in the research group of Prof. Zhifei Wang in the Department of Materials Physics and Chemistry, Southeast University and Prof. Nongyue He in the Department of Biomedical Engineering, Southeast University. His research interests are focused on biosensors and targeted drug delivery. He has published more than 10 scientific papers and 2 patents.

Dr. Zhifei Wang received his B.S. in Applied Chemistry from JiangSu Normal University in 1999, M.S. in Physical Chemistry from Lanzhou Institute of Chemical Physics, Chinese Academy of Science, in 2002, and his Ph.D. in Biomedical Engineering from Southeast University in 2006. From Oct. 2010 to Apr. 2011, he pursued his post-doctoral work in the group of Prof. Min-hui Li at Cure Institute in France. His research interest is focused on the preparation of multifunctional composite nanostructures, and their applications in bio-sensing, bio-imaging, and cancer theranostics.

Prof. Song Li obtained his Ph.D. in Biomedical Engineering from State Key Laboratory of Bioelectronics, Southeast University in 2008. Then, he entered Hudsonalpha Institute for Biotechnology as a Postdoctor and became a senior scientist in Hudsonalpha Institute for Biotechnology, iCubate Inc. and iRepertoire Inc since 2012. Currently, he is a full professor in School of Life Science and Chemistry, Hunan University of Technology. His research interests are focused on molecular diagnostic technology, nanobiotechnology, and biosensors. He has published about 65 papers and 6 patents.

Prof. Yan Deng obtained her M.S. degree from Hunan Normal University, Changsha, China, and her Ph.D. degree from Southeast University, China. She is currently a full professor in School of Life Science and Chemistry, Hunan University of Technology. Her research interest is focused on biosensors and controlled drug delivery. She has published more than 70 papers.

Nongyue He completed his Ph.D. degree from Nanjing University, China. Initially, he joined Southeast University, Nanjing and Nanjing Medical University, Nanjing as a postdoctoral researcher. But since 2000, he emerged as a professor in School of Biomedical Science and Medical Engineering, Southeast University, Nanjing. His research interests are focused on biochips and biosensors, functional nanomaterials, and electronic devices for point-of-care testing. He has published more than 300 scientific journal articles, 6 books, and more than 55 patents.
References
- 1.Wu Y.C., Chen C.S., Chan Y.J. The outbreak of COVID-19: an overview. J. Chin. Med. Assoc. 2020;83:217–220. doi: 10.1097/JCMA.0000000000000270. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2.Remuzzi A., Remuzzi G. COVID-19 and Italy: what next. Lancet. 2020;395:1225–1228. doi: 10.1016/S0140-6736(20)30627-9. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3.Sohrabi C., Alsafi Z., O'Neill N., Khan M., Kerwan A., Al-Jabir A., Iosifidis C., Agha R. World Health Organization declares global emergency: a review of the 2019 novel coronavirus (COVID-19) Int. J. Surg. 2020;76:71–76. doi: 10.1016/j.ijsu.2020.02.034. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4.W.H. Organization, Geneva, 2021.
- 5.Cowie B.C., Dore G.J. The perpetual challenge of infectious diseases. N. Engl. J. Med. 2012;367 doi: 10.1056/NEJMc1204960. 89-89. [DOI] [PubMed] [Google Scholar]
- 6.Daszak P., Cunningham A.A., Hyatt A.D. Emerging infectious diseases of wildlife--threats to biodiversity and human health. Science. 2000;287:443–449. doi: 10.1126/science.287.5452.443. [DOI] [PubMed] [Google Scholar]
- 7.Johnson P.T.J., De Roode J.C., Fenton A. Why infectious disease research needs community ecology. Science. 2015;349 doi: 10.1126/science.1259504. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8.Jones K.E., Patel N.G., Levy M.A., Storeygard A., Balk D., Gittleman J.L., Daszak P. Global trends in emerging infectious diseases. Nature. 2008;451:990–993. doi: 10.1038/nature06536. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9.Morens D.M., Fauci A.S. The 1918 influenza pandemic: insights for the 21st century. J. Infect. Dis. 2007;195:1018–1028. doi: 10.1086/511989. [DOI] [PubMed] [Google Scholar]
- 10.Gram F.C. The influenza epidemic and its after-effects in the city of Buffalo: a detailed survey. JAMA. 1919;73:886–891. [Google Scholar]
- 11.Traeger M.W., Schroeder S.E., Wright E.J., Hellard M.E., Cornelisse V.J., Doyle J.S., Stoove M.A. Effects of pre-exposure prophylaxis for the prevention of human immunodeficiency virus infection on sexual risk behavior in men who have sex with men: a systematic review and meta-analysis. Clin. Infect. Dis. 2018;67:676–686. doi: 10.1093/cid/ciy182. [DOI] [PubMed] [Google Scholar]
- 12.Chin H.B., Sipe T.A., Elder R., Mercer S.L., Chattopadhyay S.K., Jacob V., Wethington H.R., Kirby D., Elliston D.B., Griffith M., Chuke S.O., Briss S.C., Ericksen I., Galbraith J.S., Herbst J.H., Johnson R.L., Kraft J.M., Noar S.M., Romero L.M., Santelli J., Task C.P.S. The effectiveness of group-based comprehensive risk-reduction and abstinence education interventions to prevent or reduce the risk of adolescent pregnancy, human immunodeficiency virus, and sexually transmitted infections: two systematic reviews for the guide to community preventive services. Am. J. Prev. Med. 2012;42:272–294. doi: 10.1016/j.amepre.2011.11.006. [DOI] [PubMed] [Google Scholar]
- 13.Ksiazek T.G., Erdman D., Goldsmith C.S., Zaki S.R., Peret T., Emery S., Tong S.X., Urbani C., Comer J.A., Lim W., Rollin P.E., Dowell S.F., Ling A.E., Humphrey C.D., Shieh W.J., Guarner J., Paddock C.D., Rota P., Fields B., DeRisi J., Yang J.Y., Cox N., Hughes J.M., LeDuc J.W., Bellini W.J., Anderson L.J., Grp S.W. A novel coronavirus associated with severe acute respiratory syndrome. N. Engl. J. Med. 2003;348:1953–1966. doi: 10.1056/NEJMoa030781. [DOI] [PubMed] [Google Scholar]
- 14.Rota P.A., Oberste M.S., Monroe S.S., Nix W.A., Campagnoli R., Icenogle J.P., Penaranda S., Bankamp B., Maher K., Chen M.H., Tong S.X., Tamin A., Lowe L., Frace M., DeRisi J.L., Chen Q., Wang D., Erdman D.D., Peret T.C.T., Burns C., Ksiazek T.G., Rollin P.E., Sanchez A., Liffick S., Holloway B., Limor J., McCaustland K., Olsen-Rasmussen M., Fouchier R., Gunther S., Osterhaus A.D.M.E., Drosten C., Pallansch M.A., Anderson L.J., Bellini W.J. Characterization of a novel coronavirus associated with severe acute respiratory syndrome. Science. 2003;300:1394–1399. doi: 10.1126/science.1085952. [DOI] [PubMed] [Google Scholar]
- 15.Drosten C., Gunther S., Preiser W., van der Werf S., Brodt H.R., Becker S., Rabenau H., Panning M., Kolesnikova L., Fouchier R.A.M., Berger A., Burguiere A.M., Cinatl J., Eickmann M., Escriou N., Grywna K., Kramme S., Manuguerra J.C., Muller S., Rickerts V., Sturmer M., Vieth S., Klenk H.D., Osterhaus A.D.M.E., Schmitz H., Doerr H.W. Identification of a novel coronavirus in patients with severe acute respiratory syndrome. N. Engl. J. Med. 2003;348:1967–1976. doi: 10.1056/NEJMoa030747. [DOI] [PubMed] [Google Scholar]
- 16.Zaki A.M., van Boheemen S., Bestebroer T.M., Osterhaus A.D.M.E., Fouchier R.A.M. Isolation of a novel coronavirus from a man with pneumonia in Saudi Arabia. N. Engl. J. Med. 2012;367:1814–1820. doi: 10.1056/NEJMoa1211721. [DOI] [PubMed] [Google Scholar]
- 17.Assiri A., McGeer A., Perl T.M., Price C.S., Al Rabeeah A.A., Cummings D.A.T., Alabdullatif Z.N., Assad M., Almulhim A., Makhdoom H., Madani H., Alhakeem R., Al-Tawfiq J.A., Cotten M., Watson S.J., Kellam P., Zumla A.I., Memish Z.A., Team K.M.-C.I. Hospital outbreak of Middle East respiratory syndrome coronavirus. N. Engl. J. Med. 2013;369:407–416. doi: 10.1056/NEJMoa1306742. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 18.Assiri A., Al-Tawfiq J.A., Al-Rabeeah A.A., Al-Rabiah F.A., Al-Hajjar S., Al-Barrak A., Flemban H., Al-Nassir W.N., Balkhy H.H., Al-Hakeem R.F., Makhdoom H.Q., Zumla A.I., Memish Z.A. Epidemiological, demographic, and clinical characteristics of 47 cases of Middle East respiratory syndrome coronavirus disease from Saudi Arabia: a descriptive study. Lancet Infect. Dis. 2013;13:752–761. doi: 10.1016/S1473-3099(13)70204-4. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 19.Chan J.F.W., Yuan S.F., Kok K.H., To K.K.W., Chu H., Yang J., Xing F.F., Liu J.L., Yip C.C.Y., Poon R.W.S., Tsoi H.W., Lo S.K.F., Chan K.H., Poon V.K.M., Chan W.M., Ip J.D., Cai J.P., Cheng V.C.C., Chen H.L., Hui C.K.M., Yuen K.Y. A familial cluster of pneumonia associated with the 2019 novel coronavirus indicating person-to-person transmission: a study of a family cluster. Lancet. 2020;395:514–523. doi: 10.1016/S0140-6736(20)30154-9. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 20.Holshue M.L., DeBolt C., Lindquist S., Lofy K.H., Wiesman J., Bruce H., Spitters C., Ericson K., Wilkerson S., Tural A., Diaz G., Cohn A., Fox L., Patel A., Gerber S.I., Kim L., Tong S.X., Lu X.Y., Lindstrom S., Pallansch M.A., Weldon W.C., Biggs H.M., Uyeki T.M., Pillai S.K., Washington State 2019-nCoV Case Investigation Team First case of 2019 novel coronavirus in the United States. N. Engl. J. Med. 2020;382:929–936. doi: 10.1056/NEJMoa2001191. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 21.Yang W.J., Cao Q.Q., Qin L., Wang X.Y., Cheng Z.H., Pan A.S., Dai J.Y., Sun Q.F., Zhao F.Q., Qu J.M., Yan F.H. Clinical characteristics and imaging manifestations of the 2019 novel coronavirus disease (COVID-19): a multi-center study in Wenzhou city, Zhejiang, China. J. Infect. 2020;80:388–393. doi: 10.1016/j.jinf.2020.02.016. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 22.Guan W.J., Ni Z.Y., Hu Y., Liang W.H., Ou C.Q., He J.X., Liu L., Shan H., Lei C.L., Hui D.S.C., Du B., Li L.J., Zeng G., Yuen K.Y., Chen R.C., Tang C.L., Wang T., Chen P.Y., Xiang J., Li S.Y., Wang J.L., Liang Z.J., Peng Y.X., Wei L., Liu Y., Hu Y.H., Peng P., Wang J.M., Liu J.Y., Chen Z., Li G., Zheng Z.J., Qiu S.Q., Luo J., Ye C.J., Zhu S.Y., Zhong N.S., China Medical Treatment Expert Group Clinical characteristics of coronavirus disease 2019 in China. N. Engl. J. Med. 2020;382:1708–1720. doi: 10.1056/NEJMoa2002032. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 23.Boucher H.W., Talbot G.H., Bradley J.S., Edwards J.E., Gilbert D., Rice L.B., Scheld M., Spellberg B. Bad bugs, no drugs: no ESKAPE! an update from the Infectious Diseases Society of America. Clin. Infect. Dis. 2009;48:1–12. doi: 10.1086/595011. [DOI] [PubMed] [Google Scholar]
- 24.McDonald L.C., Gerding D.N., Johnson S., Bakken J.S., Carroll K.C., Coffin S.E., Dubberke E.R., Garey K.W., Gould C.V., Kelly C., Loo V., Sammons J.S., Sandora T.J., Wilcox M.H. Clinical practice guidelines for Clostridium difficile infection in adults and children: 2017 update by the Infectious Diseases Society of America (IDSA) and Society for Healthcare Epidemiology of America (SHEA) Clin. Infect. Dis. 2018;66:e1–e48. doi: 10.1093/cid/cix1085. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 25.Zumla A., Rao M., Wallis R.S., Kaufmann S.H.E., Rustomjee R., Mwaba P., Vilaplana C., Yeboah-Manu D., Chakaya J., Ippolito G., Azhar E., Hoelscher M., Maeurer M., Network H.-D.T. Host-directed therapies for infectious diseases: current status, recent progress, and future prospects. Lancet Infect. Dis. 2016;16:e47–e63. doi: 10.1016/S1473-3099(16)00078-5. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 26.Libertucci J., Young V.B. The role of the microbiota in infectious diseases. Nat. Microbiol. 2019;4:35–45. doi: 10.1038/s41564-018-0278-4. [DOI] [PubMed] [Google Scholar]
- 27.Weigl J. Challenges in infectious disease control and the current pandemic by skewed distributions. Pravent. Gesundheit. 2020;15:97–101. [Google Scholar]
- 28.Dong Y.H., Wang L.P., Burgner D.P., Miller J.E., Song Y., Ren X., Li Z.J., Xing Y., Ma J., Sawyer S.M., Patton G.C. Infectious diseases in children and adolescents in China: analysis of national surveillance data from 2008 to 2017. BMJ-Brit. Med. J. 2020;369:m1043. doi: 10.1136/bmj.m1043. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 29.Nie K.X., Wang C., Li X.W. Success of big infectious disease reimbursement policy in china. Inq. J. Health Care. 2020;57 doi: 10.1177/0046958020907788. 0046958020907788. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 30.Nahid P., Dorman S.E., Alipanah N., Barry P.M., Brozek J.L., Cattamanchi A., Chaisson L.H., Chaisson R.E., Daley C.L., Grzemska M., Higashi J.M., Ho C.S., Hopewell P.C., Keshavjee S.A., Lienhardt C., Menzies R., Merrifield C., Narita M., O'Brien R., Peloquin C.A., Raftery A., Saukkonen J., Schaaf H.S., Sotgiu G., Starke J.R., Migliori G.B., Vernon A. Executive summary: Official American Thoracic Society/Centers for Disease Control and Prevention/Infectious Diseases Society of America clinical practice guidelines: treatment of drug-susceptible tuberculosis. Clin. Infect. Dis. 2016;63:853–867. doi: 10.1093/cid/ciw566. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 31.Wu Z., McGoogan J.M. Characteristics of and important lessons from the coronavirus disease 2019 (COVID-19) outbreak in China: summary of a report of 72 314 cases from the Chinese Center for Disease Control and Prevention. JAMA. 2020;323:1239–1242. doi: 10.1001/jama.2020.2648. [DOI] [PubMed] [Google Scholar]
- 32.Deng S.Q., Peng H.J. Characteristics of and public health responses to the coronavirus disease 2019 outbreak in China. J. Clin. Med. 2020;9:575. doi: 10.3390/jcm9020575. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 33.Zhang L., Liu Y.H. Potential interventions for novel coronavirus in China: a systematic review. J. Med. Virol. 2020;92:479–490. doi: 10.1002/jmv.25707. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 34.Tram D.T.N., Wang H., Sugiarto S., Li T., Ang W.H., Lee C., Pastorin G. Advances in nanomaterials and their applications in point of care (POC) devices for the diagnosis of infectious diseases. Biotechnol. Adv. 2016;34:1275–1288. doi: 10.1016/j.biotechadv.2016.09.003. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 35.Gou L.X., Sheng Y.R., Peng Q.L., Ling J.J., Yue H., Chen F.J., Tang H. Ternary nanocube-based “off-on” blinking-type electrochemiluminescence towards enzyme-free detection of hepatitis B virus (HBV)-related DNA. Sens. Actuators B Chem. 2020;312 [Google Scholar]
- 36.Ramanan P., Bryson A.L., Binnicker M.J., Pritt B.S., Patel R. Syndromic panel-based testing in clinical microbiology. Clin. Microbiol. Rev. 2018;31 doi: 10.1128/CMR.00024-17. (UNSP) [DOI] [PMC free article] [PubMed] [Google Scholar]
- 37.Tang Y.J., Ii B., Dai J.G., Dai J.F., Wang X.H., Si J., Ali Z.S., Li T.T., He N.Y. Genotyping of Pseudomonas aeruginosa type III secretion system using magnetic enrichment multiplex polymerase chain reaction and chemiluminescence. J. Biomed. Nanotechnol. 2016;12:762–769. doi: 10.1166/jbn.2016.2222. [DOI] [PubMed] [Google Scholar]
- 38.You M., Li Z., Feng S., Gao B., Yao C., Hu J., Xu F. Ultrafast photonic PCR based on photothermal nanomaterials. Trends Biotechnol. 2020;38:637–649. doi: 10.1016/j.tibtech.2019.12.006. [DOI] [PubMed] [Google Scholar]
- 39.Yang B., Kong J., Fang X. Bandage-like wearable flexible microfluidic recombinase polymerase amplification sensor for the rapid visual detection of nucleic acids. Talanta. 2019;204:685–692. doi: 10.1016/j.talanta.2019.06.031. [DOI] [PubMed] [Google Scholar]
- 40.Liu Z., Shang C., Ma H., You M. An upconversion nanoparticle-based photostable FRET system for long-chain DNA sequence detection. Nanotechnology. 2020;31 doi: 10.1088/1361-6528/ab776d. [DOI] [PubMed] [Google Scholar]
- 41.Wang T., Liu Y., Sun H.H., Yin B.C., Ye B.C. An RNA‐guided Cas9 nickase‐based method for universal isothermal DNA amplification. Angew. Chem. Int. Ed. 2019;58:5382–5386. doi: 10.1002/anie.201901292. [DOI] [PubMed] [Google Scholar]
- 42.Liu R., Han H., Liu F., Lv Z.H., Wu K.L., Liu Y.L., Feng Y., Zhu C.L. Positive rate of RT-PCR detection of SARS-CoV-2 infection in 4880 cases from one hospital in Wuhan, China, from Jan to Feb 2020. Clin. Chim. Acta. 2020;505:172–175. doi: 10.1016/j.cca.2020.03.009. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 43.Santiago G.A., Vazquez J., Courtney S., Matias K.Y., Andersen L.E., Colon C., Butler A.E., Roulo R., Bowzard J., Villanueva J.M., Munoz-Jordan J.L. Performance of the trioplex real-time RT-PCR assay for detection of Zika, dengue, and chikungunya viruses. Nat. Commun. 2018;9:1391. doi: 10.1038/s41467-018-03772-1. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 44.Corman V.M., Rasche A., Baronti C., Aldabbagh S., Cadar D., Reusken C.B.E.M., Pas S.D., Goorhuis A., Schinkel J., Molenkamp R., Kummerer B.M., Bleicker T., Brunink S., Eschbach-Bludau M., Eis-Hubinger A.M., Koopmans M.P., Schmidt-Chanasit J., Grobusch M.P., de Lamballerie X., Drosten C., Drexler J.F. Assay optimization for molecular detection of Zika virus. Bull. World Health Organ. 2016;94:880–892. doi: 10.2471/BLT.16.175950. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 45.Kiechle F.L., Malinski H., Dandurand D.M., Mcgill J.B. The effect of amino-acids, monoamines and polyamines on pyruvate-dehydrogenase activity in mitochondria from rat adipocytes. Mol. Cell. Biochem. 1990;93:195–206. doi: 10.1007/BF00226192. [DOI] [PubMed] [Google Scholar]
- 46.Chan H.N., Tan M.J.A., Wu H. Point-of-care testing: applications of 3D printing. Lab Chip. 2017;17:2713–2739. doi: 10.1039/c7lc00397h. [DOI] [PubMed] [Google Scholar]
- 47.He W.H., You M.L., Wan W.T., Xu F., Li F., Li A. Point-of-care periodontitis testing: biomarkers, current technologies, and perspectives. Trends Biotechnol. 2018;36:1127–1144. doi: 10.1016/j.tibtech.2018.05.013. [DOI] [PubMed] [Google Scholar]
- 48.Niemz A., Ferguson T.M., Boyle D.S. Point-of-care nucleic acid testing for infectious diseases. Trends Biotechnol. 2011;29:240–250. doi: 10.1016/j.tibtech.2011.01.007. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 49.Sista R., Hua Z.S., Thwar P., Sudarsan A., Srinivasan V., Eckhardt A., Pollack M., Pamula V. Development of a digital microfluidic platform for point of care testing. Lab Chip. 2008;8:2091–2104. doi: 10.1039/b814922d. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 50.Citartan M., Tang T.H. Recent developments of aptasensors expedient for point-of-care (POC) diagnostics. Talanta. 2019;199:556–566. doi: 10.1016/j.talanta.2019.02.066. [DOI] [PubMed] [Google Scholar]
- 51.Nicolini A.M., McCracken K.E., Yoon J.Y. Future developments in biosensors for field-ready Zika virus diagnostics. J. Biol. Eng. 2017;11:7. doi: 10.1186/s13036-016-0046-z. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 52.Tepeli Y., Ülkü A. Electrochemical biosensors for influenza virus a detection: the potential of adaptation of these devices to POC systems. Sens. Actuators B Chem. 2018;254:377–384. [Google Scholar]
- 53.Pai A., Khachaturian A., Chapman S., Hu A., Wang H., Hajimiri A. A handheld magnetic sensing platform for antigen and nucleic acid detection. Analyst. 2014;139:1403–1411. doi: 10.1039/c3an01947k. [DOI] [PubMed] [Google Scholar]
- 54.Shrivastava S., Trung T.Q., Lee N.E. Recent progress, challenges, and prospects of fully integrated mobile and wearable point-of-care testing systems for self-testing. Chem. Soc. Rev. 2020;49:1812–1866. doi: 10.1039/c9cs00319c. [DOI] [PubMed] [Google Scholar]
- 55.Varongchayakul N., Hersey J.S., Squires A., Meller A., Grinstaff M.W. A solid-state hard microfluidic nanopore biosensor with multilayer fluidics and on-chip bioassay/purification chamber. Adv. Funct. Mater. 2018;28 doi: 10.1002/adfm.201804182. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 56.Temiz Y., Lovchik R.D., Kaigala G.V., Delamarche E. Lab-on-a-chip devices: how to close and plug the lab. Microelectron. Eng. 2015;132:156–175. [Google Scholar]
- 57.Nguyen H.V., Nguyen V.D., Nguyen H.Q., Chau T.H.T., Lee E.Y., Seo T.S. Nucleic acid diagnostics on the total integrated lab-on-a-disc for point-of-care testing. Biosens. Bioelectron. 2019;141 doi: 10.1016/j.bios.2019.111466. [DOI] [PubMed] [Google Scholar]
- 58.Ho H.P., Lau P.M., Kwok H.C., Wu S.Y., Gao M.H., Cheung A.K.L., Chen Q.L., Wang G.H., Kwan Y.W., Wong C.K., Kong S.K. Allergen screening bioassays: recent developments in lab-on-a-chip and lab-on-a-disc systems. Bioanalysis. 2014;6:2005–2018. doi: 10.4155/bio.14.153. [DOI] [PubMed] [Google Scholar]
- 59.Li F., You M., Li S.X., Hu J., Liu C., Gong Y., Yang H.Y., Xu F. Paper-based point-of-care immunoassays: recent advances and emerging trends. Biotechnol. Adv. 2020;39 doi: 10.1016/j.biotechadv.2019.107442. [DOI] [PubMed] [Google Scholar]
- 60.Mahmoudi T., de la Guardia M., Baradaran B. Lateral flow assays towards point-of-care cancer detection: a review of current progress and future trends. TrAC Trends Anal. Chem. 2020;125 [Google Scholar]
- 61.Shin D.J., Trick A.Y., Hsieh Y.H., Thomas D.L., Wang T.H. Sample-to-answer droplet magnetofluidic platform for point-of-care hepatitis C viral load quantitation. Sci. Rep. 2018;8:9793. doi: 10.1038/s41598-018-28124-3. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 62.Kumar S., Gallagher R., Bishop J., Kline E., Buser J., Lafleur L., Shah K.M., Lutz B., Yager P. Long-term dry storage of enzyme-based reagents for isothermal nucleic acid amplification in a porous matrix for use in point-of-care diagnostic devices. Analyst. 2020;145:6875–6886. doi: 10.1039/d0an01098g. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 63.Craw P., Balachandran W. Isothermal nucleic acid amplification technologies for point-of-care diagnostics: a critical review. Lab Chip. 2012;12:2469–2486. doi: 10.1039/c2lc40100b. [DOI] [PubMed] [Google Scholar]
- 64.Meyer-Rath G., Schnippel K., Long L., MacLeod W., Sanne I., Stevens W., Pillay S., Pillay Y., Rosen S. The impact and cost of scaling up GeneXpert MTB/RIF in South Africa. Plos One. 2012;7 doi: 10.1371/journal.pone.0036966. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 65.Rong Z., Bai Z.K., Li J.N., Tang H., Shen T.Y., Wang Q., Wang C.W., Xiao R., Wang S.Q. Dual-color magnetic-quantum dot nanobeads as versatile fluorescent probes in test strip for simultaneous point-of-care detection of free and complexed prostate-specific antigen. Biosens. Bioelectron. 2019;145 doi: 10.1016/j.bios.2019.111719. (UNSP) [DOI] [PubMed] [Google Scholar]
- 66.Yun K., Choi J., Song I.U., Chung Y.A. Smartphone-based point-of-care cholesterol blood test performance evaluation compared with a clinical diagnostic laboratory method. Appl. Sci. 2019;9:3334. [Google Scholar]
- 67.Chin C.D., Linder V., Sia S.K. Commercialization of microfluidic point-of-care diagnostic devices. Lab Chip. 2012;12:2118–2134. doi: 10.1039/c2lc21204h. [DOI] [PubMed] [Google Scholar]
- 68.Darwish N.T., Sekaran S.D., Khor S.M. Point-of-care tests: a review of advances in the emerging diagnostic tools for dengue virus infection. Sens. Actuators B Chem. 2018;255:3316–3331. [Google Scholar]
- 69.Hsieh K., Ferguson B.S., Eisenstein M., Plaxco K.W., Soh H.T. Integrated electrochemical microsystems for genetic detection of pathogens at the point of care. Acc. Chem. Res. 2015;48:911–920. doi: 10.1021/ar500456w. [DOI] [PubMed] [Google Scholar]
- 70.Huang L., Zhang D., Jiao L.R., Su E.B., He N.Y. A new quality control method for lateral flow assay. Chin. Chem. Lett. 2018;29:1853–1856. [Google Scholar]
- 71.Khan H., Khan A., Liu Y.F., Wang S., Bibi S., Xu H.P., Liu Y., Durrani S., Jin L., He N.Y., Xiong T. CRISPR-Cas13a mediated nanosystem for attomolar detection of canine parvovirus type 2. Chin. Chem. Lett. 2019;30:2201–2204. doi: 10.1016/j.cclet.2019.10.032. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 72.Boehme C.C., Nabeta P., Hillemann D., Nicol M.P., Shenai S., Krapp F., Allen J., Tahirli R., Blakemore R., Rustomjee R., Milovic A., Jones M., O'Brien S.M., Persing D.H., Ruesch-Gerdes S., Gotuzzo E., Rodrigues C., Alland D., Perkins M.D. Rapid molecular detection of tuberculosis and rifampin resistance. N. Engl. J. Med. 2010;363:1005–1015. doi: 10.1056/NEJMoa0907847. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 73.Chen Z., Wu Y.Q., Kang M., He N.Y., Wan S.R., Su E.B., Wang L.J. Research on automated nucleic acid extraction instrument based on magnetic nanoparticles separation. Nanosci. Nanotechnol. Lett. 2018;10:60–68. [Google Scholar]
- 74.Hussain M., Lv M., Dong X.H., Shen H., Wang W., Li S., Chen Z., Jin L., He N.Y., Li Z.Y., Liu B. Design of rapid bacterial identification system based on scattering of laser light and classification of binned plots. J. Nanosci. Nanotechnol. 2020;20:4047–4056. doi: 10.1166/jnn.2020.17491. [DOI] [PubMed] [Google Scholar]
- 75.Wang C., Guo Z.K., Zhu H., He N.Y., Chen Z., Wang Z.L. Mechanical gripper design and force analysis of microplates for automated high-throughput nucleic acid detection system. J. Nanosci. Nanotechnol. 2020;20:1401–1408. doi: 10.1166/jnn.2020.17355. [DOI] [PubMed] [Google Scholar]
- 76.Chen H., Wu Y.Q., Fang Y.L., Liao P., Zhao K.C., Deng Y., He N.Y. Integrated and automated, sample-in to result-out, system for nanotechnology-based detection of infectious pathogens. Nanosci. Nanotechnol. Lett. 2018;10:1423–1428. [Google Scholar]
- 77.Chen Z., Yang H.W., Peng J., Fang Y.L., Dong H.M., Chen H., Yu J., Nie L.B., Deng Y., Li S., Liu H.N., Wang L.J., He N.Y. Highly precise and fast polymerase chain reaction thermal cycling module with a novel integrated temperature control strategy. Nanosci. Nanotechnol. Lett. 2018;10:23–31. [Google Scholar]
- 78.He H.R., Wu Y.Q., He N.Y., Deng Y. The latest progress of on-site pathogens detection techniques and instruments based on nucleic acid. J. Nanosci. Nanotechnol. 2015;15:6342–6356. doi: 10.1166/jnn.2015.10710. [DOI] [PubMed] [Google Scholar]
- 79.Fang Y., Liao P., Chen Z., Chen H., Wu Y., Li S., Deng Y., He N. Improvement and application of qPCR (real-time quantitative polymerase chain reaction) data processing method for home-made integrated nucleic acid detection system. J. Nanosci. Nanotechnol. 2020;20:7369–7375. doi: 10.1166/jnn.2020.18714. [DOI] [PubMed] [Google Scholar]
- 80.Wang C., Huang Y.P., Zhu H., Wang Z.L., Chen Z., Chen H., Li S., He N.Y., Liu T.H. Design and implementation of high-throughput magnetic separation module for automated nucleic acid detection system based on magnetic nano-beads. J. Nanosci. Nanotechnol. 2020;20:2138–2143. doi: 10.1166/jnn.2020.17327. [DOI] [PubMed] [Google Scholar]
- 81.Mauk M., Song J., Bau H.H., Gross R., Bushman F.D., Collman R.G., Liu C. Miniaturized devices for point of care molecular detection of HIV. Lab Chip. 2017;17:382–394. doi: 10.1039/c6lc01239f. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 82.Li Z.H., Leustean L., Inci F., Zheng M., Demirci U., Wang S.Q. Plasmonic-based platforms for diagnosis of infectious diseases at the point-of-care. Biotechnol. Adv. 2019;37 doi: 10.1016/j.biotechadv.2019.107440. [DOI] [PubMed] [Google Scholar]
- 83.Chen H., Liu K.K., Li Z., Wang P. Point of care testing for infectious diseases. Clin. Chim. Acta. 2019;493:138–147. doi: 10.1016/j.cca.2019.03.008. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 84.Tu J.B., Torrente-Rodriguez R.M., Wang M.Q., Gao W. The era of digital health: a review of portable and wearable affinity biosensors. Adv. Funct. Mater. 2019 [Google Scholar]
- 85.Maduraiveeran G., Sasidharan M., Ganesan V. Electrochemical sensor and biosensor platforms based on advanced nanomaterials for biological and biomedical applications. Biosens. Bioelectron. 2018;103:113–129. doi: 10.1016/j.bios.2017.12.031. [DOI] [PubMed] [Google Scholar]
- 86.Chen L.C., Wang E., Tai C.S., Chiu Y.C., Li C.W., Lin Y.R., Lee T.H., Huang C.W., Chen J.C., Chen W.L. Improving the reproducibility, accuracy, and stability of an electrochemical biosensor platform for point-of-care use. Biosens. Bioelectron. 2020;155 doi: 10.1016/j.bios.2020.112111. [DOI] [PubMed] [Google Scholar]
- 87.Liu M., Khan A., Wang Z.F., Liu Y., Yang G.J., Deng Y., He N.Y. Aptasensors for pesticide detection. Biosens. Bioelectron. 2019;130:174–184. doi: 10.1016/j.bios.2019.01.006. [DOI] [PubMed] [Google Scholar]
- 88.Nishitani S., Sakata T. Enhancement of signal-to-noise ratio for serotonin detection with well-designed nanofilter-coated potentiometric electrochemical biosensor. ACS Appl. Mater. Interfaces. 2020;12:14761–14769. doi: 10.1021/acsami.9b19309. [DOI] [PubMed] [Google Scholar]
- 89.Dai Y.F., Liu C.C. Recent advances on electrochemical biosensing strategies toward universal point‐of‐care systems. Angew. Chem. Int. Ed. 2019;58:12355–12368. doi: 10.1002/anie.201901879. [DOI] [PubMed] [Google Scholar]
- 90.Prasad K.S., Cao X.Y., Gao N., Jin Q.J., Sanjay S.T., Henao-Pabon G., Li X.J. A low-cost nanomaterial-based electrochemical immunosensor on paper for high-sensitivity early detection of pancreatic cancer. Sens. Actuators B Chem. 2020;305 doi: 10.1016/j.snb.2019.127516. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 91.Golichenari B., Nosrati R., Farokhi-Fard A., Maleki M.F., Hayat S.M.G., Ghazvini K., Vaziri F., Behravan J. Electrochemical-based biosensors for detection of Mycobacterium tuberculosis and tuberculosis biomarkers. Crit. Rev. Biotechnol. 2019;39:1056–1077. doi: 10.1080/07388551.2019.1668348. [DOI] [PubMed] [Google Scholar]
- 92.Chand R., Ramalingam S., Neethirajan S. A 2D transition-metal dichalcogenide MoS2 based novel nanocomposite and nanocarrier for multiplex miRNA detection. Nanoscale. 2018;10:8217–8225. doi: 10.1039/c8nr00697k. [DOI] [PubMed] [Google Scholar]
- 93.Martins G.V., Marques A.C., Fortunato E., Sales M.G.F. Wax-printed paper-based device for direct electrochemical detection of 3-nitrotyrosine. Electrochim. Acta. 2018;284:60–68. [Google Scholar]
- 94.Low S.S., Chia J.S.Y., Tan M.T.T., Loh H.S., Khiew P.S., Singh A., Chiu W.S. A proof of concept: detection of avian influenza H5 gene by a graphene-enhanced electrochemical genosensor. J. Nanosci. Nanotechnol. 2016;16:2438–2446. doi: 10.1166/jnn.2016.11714. [DOI] [PubMed] [Google Scholar]
- 95.Reich P., Preuss J.A., Bahner N., Bahnemann J. Impedimetric aptamer-based biosensors: principles and techniques. Adv. Biochem. Eng. Biotechnol. 2020;174:17–41. doi: 10.1007/10_2019_113. [DOI] [PubMed] [Google Scholar]
- 96.Preuss J.A., Reich P., Bahner N., Bahnemann J. Impedimetric aptamer-based biosensors: applications. Adv. Biochem. Eng. Biotechnol. 2020;174:43–91. doi: 10.1007/10_2020_125. [DOI] [PubMed] [Google Scholar]
- 97.Mousavi M.P.S., Ainla A., Tan E.K.W., Abd El-Rahman M.K., Yoshida Y., Yuan L., Sigurslid H.H., Arkan N., Yip M.C., Abrahamsson C.K., Homer-Vanniasinkam S., Whitesides G.M. Ion sensing with thread-based potentiometric electrodes. Lab Chip. 2018;18:2279–2290. doi: 10.1039/c8lc00352a. [DOI] [PubMed] [Google Scholar]
- 98.Labib M., Sargent E.H., Kelley S.O. Electrochemical methods for the analysis of clinically relevant biomolecules. Chem. Rev. 2016;116:9001–9090. doi: 10.1021/acs.chemrev.6b00220. [DOI] [PubMed] [Google Scholar]
- 99.Li Q., Lu N., Wang L.H., Fan C.H. Advances in nanowire transistor-based biosensors. Small Methods. 2018;2 [Google Scholar]
- 100.Bollella P., Gorton L. Enzyme based amperometric biosensors. Curr. Opin. Electrochem. 2018;10:157–173. [Google Scholar]
- 101.Chen A.C., Chatterjee S. Nanomaterials based electrochemical sensors for biomedical applications. Chem. Soc. Rev. 2013;42:5425–5438. doi: 10.1039/c3cs35518g. [DOI] [PubMed] [Google Scholar]
- 102.Singhal C., Khanuja M., Chaudhary N., Pundir C.S., Narang J. Detection of chikungunya virus DNA using two-dimensional MoS2 nanosheets based disposable biosensor. Sci. Rep. 2018;8:7734. doi: 10.1038/s41598-018-25824-8. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 103.Hiraiwa M., Kim J.H., Lee H.B., Inoue S., Becker A.L., Weigel K.M., Cangelosi G.A., Lee K.H., Chung J.H. Amperometric immunosensor for rapid detection of Mycobacterium tuberculosis. J. Micromech. Microeng. 2015;25 doi: 10.1088/0960-1317/25/5/055013. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 104.Brosel-Oliu S., Mergel O., Uria N., Abramova N., van Rijn P., Bratov A. 3D impedimetric sensors as a tool for monitoring bacterial response to antibiotics. Lab Chip. 2019;19:1436–1447. doi: 10.1039/c8lc01220b. [DOI] [PubMed] [Google Scholar]
- 105.Elshafey R., Tlili C., Abulrob A., Tavares A.C., Zourob M. Label-free impedimetric immunosensor for ultrasensitive detection of cancer marker Murine double minute 2 in brain tissue. Biosens. Bioelectron. 2013;39:220–225. doi: 10.1016/j.bios.2012.07.049. [DOI] [PubMed] [Google Scholar]
- 106.Felix F.S., Angnes L. Electrochemical immunosensors – a powerful tool for analytical applications. Biosens. Bioelectron. 2018;102:470–478. doi: 10.1016/j.bios.2017.11.029. [DOI] [PubMed] [Google Scholar]
- 107.Daniels J.S., Pourmand N. Label-free impedance biosensors: opportunities and challenges. Electroanalysis. 2007;19:1239–1257. doi: 10.1002/elan.200603855. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 108.Cecchetto J., Carvalho F.C., Santos A., Fernandes F.C.B., Bueno P.R. An impedimetric biosensor to test neat serum for dengue diagnosis. Sens. Actuators B Chem. 2015;213:150–154. [Google Scholar]
- 109.Kimmel D.W., LeBlanc G., Meschievitz M.E., Cliffel D.E. Electrochemical sensors and biosensors. Anal. Chem. 2012;84:685–707. doi: 10.1021/ac202878q. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 110.Luo X.L., Davis J.J. Electrical biosensors and the label free detection of protein disease biomarkers. Chem. Soc. Rev. 2013;42:5944–5962. doi: 10.1039/c3cs60077g. [DOI] [PubMed] [Google Scholar]
- 111.Bratov A., Abramova N., Ipatov A. Recent trends in potentiometric sensor arrays—a review. Anal. Chim. Acta. 2010;678:149–159. doi: 10.1016/j.aca.2010.08.035. [DOI] [PubMed] [Google Scholar]
- 112.Parrilla M., Canovas R., Andrade F.J. Paper-based enzymatic electrode with enhanced potentiometric response for monitoring glucose in biological fluids. Biosens. Bioelectron. 2017;90:110–116. doi: 10.1016/j.bios.2016.11.034. [DOI] [PubMed] [Google Scholar]
- 113.Ding J.W., Qin W. Recent advances in potentiometric biosensors. TrAC Trends Anal. Chem. 2020;124 [Google Scholar]
- 114.Tarasov A., Gray D.W., Tsai M.Y., Shields N., Montrose A., Creedon N., Lovera P., O'Riordan A., Mooney M.H., Vogel E.M. A potentiometric biosensor for rapid on-site disease diagnostics. Biosens. Bioelectron. 2016;79:669–678. doi: 10.1016/j.bios.2015.12.086. [DOI] [PubMed] [Google Scholar]
- 115.Lee M.Y., Lee H.R., Park C.H., Han S.G., Oh J.H. Organic transistor-based chemical sensors for wearable bioelectronics. Acc. Chem. Res. 2018;51:2829–2838. doi: 10.1021/acs.accounts.8b00465. [DOI] [PubMed] [Google Scholar]
- 116.Matsumoto A., Miyahara Y. Current and emerging challenges of field effect transistor based bio-sensing. Nanoscale. 2013;5:10702–10718. doi: 10.1039/c3nr02703a. [DOI] [PubMed] [Google Scholar]
- 117.Gutierrez-Sanz O., Andoy N.M., Filipiak M.S., Haustein N., Tarasov A. Direct, label-free, and rapid transistor-based immunodetection in whole serum. ACS Sens. 2017;2:1278–1286. doi: 10.1021/acssensors.7b00187. [DOI] [PubMed] [Google Scholar]
- 118.Syedmoradi L., Ahmadi A., Norton M.L., Omidfar K. A review on nanomaterial-based field effect transistor technology for biomarker detection. Microchim. Acta. 2019;186:739. doi: 10.1007/s00604-019-3850-6. [DOI] [PubMed] [Google Scholar]
- 119.Ganguli A., Faramarzi V., Mostafa A., Hwang M.T., You S., Bashir R. High sensitivity graphene field effect transistor‐based detection of DNA amplification. Adv. Funct. Mater. 2020;30 [Google Scholar]
- 120.Picca R.A., Manoli K., Macchia E., Sarcina L., Di Franco C., Cioffi N., Blasi D., Österbacka R., Torricelli F., Scamarcio G., Torsi L. Ultimately sensitive organic bioelectronic transistor sensors by materials and device structure design. Adv. Funct. Mater. 2020;30 [Google Scholar]
- 121.Gao Z.L., Ducos P., Ye H.C., Zauberman J., Sriram A., Yang X.P., Wang Z.Y., Mitchell M.W., Lekkas D., Brisson D., Johnson A.T.C. Graphene transistor arrays functionalized with genetically engineered antibody fragments for Lyme disease diagnosis. 2D Mater. 2020;7 [Google Scholar]
- 122.Han Y.X., Chen J., Li Z., Chen H.L., Qiu H.D. Recent progress and prospects of alkaline phosphatase biosensor based on fluorescence strategy. Biosens. Bioelectron. 2020;148 doi: 10.1016/j.bios.2019.111811. [DOI] [PubMed] [Google Scholar]
- 123.Pashchenko O., Shelby T., Banerjee T., Santra S. A comparison of optical, electrochemical, magnetic, and colorimetric point-of-care biosensors for infectious disease diagnosis. ACS Infect. Dis. 2018;4:1162–1178. doi: 10.1021/acsinfecdis.8b00023. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 124.Zhang R.Q., Liu S.L., Zhao W., Zhang W.P., Yu X., Li Y., Li A.J., Pang D.W., Zhang Z.L. A simple point-of-care microfluidic immunomagnetic fluorescence assay for pathogens. Anal. Chem. 2013;85:2645–2651. doi: 10.1021/ac302903p. [DOI] [PubMed] [Google Scholar]
- 125.Takemura K., Adegoke O., Suzuki T., Park E.Y. A localized surface plasmon resonance-amplified immunofluorescence biosensor for ultrasensitive and rapid detection of nonstructural protein 1 of Zika virus. Plos One. 2019;14 doi: 10.1371/journal.pone.0211517. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 126.Guerreiro M.R., Freitas D.F., Alves P.M., Coroadinha A.S. Detection and quantification of label-free infectious adenovirus using a switch-on cell-based fluorescent biosensor. ACS Sens. 2019;4:1654–1661. doi: 10.1021/acssensors.9b00489. [DOI] [PubMed] [Google Scholar]
- 127.Huang R.R., He L., Li S., Liu H.N., Jin L., Chen Z., Zhao Y.X., Li Z.Y., Deng Y., He N.Y. A simple fluorescence aptasensor for gastric cancer exosome detection based on branched rolling circle amplification. Nanoscale. 2020;12:2445–2451. doi: 10.1039/c9nr08747h. [DOI] [PubMed] [Google Scholar]
- 128.Wang Y., Li Z.H., Wang J., Li J.H., Lin Y.H. Graphene and graphene oxide: biofunctionalization and applications in biotechnology. Trends Biotechnol. 2011;29:205–212. doi: 10.1016/j.tibtech.2011.01.008. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 129.Saha K., Agasti S.S., Kim C., Li X.N., Rotello V.M. Gold nanoparticles in chemical and biological sensing. Chem. Rev. 2012;112:2739–2779. doi: 10.1021/cr2001178. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 130.Yang D.D., Liu M., Xu J., Yang C., Wang X.X., Lou Y.B., He N.Y., Wang Z.F. Carbon nanosphere-based fluorescence aptasensor for targeted detection of breast cancer cell MCF-7. Talanta. 2018;185:113–117. doi: 10.1016/j.talanta.2018.03.045. [DOI] [PubMed] [Google Scholar]
- 131.Li B.M., Yu Q.L., Duan Y.X. Fluorescent labels in biosensors for pathogen detection. Crit. Rev. Biotechnol. 2015;35:82–93. doi: 10.3109/07388551.2013.804487. [DOI] [PubMed] [Google Scholar]
- 132.Chandan H.R., Venkataramana M., Kurkuri M.D., Balakrishna R.G. Simple quantum dot bioprobe/label for sensitive detection of Staphylococcus aureus TNase. Sens. Actuators B Chem. 2016;222:1201–1208. [Google Scholar]
- 133.Tan X.L., Wang X.Q., Hao A.Y., Liu Y.F., Wang X.P., Chu T., Jiang L., Yang Y.Q., Ming D.M. Aptamer-based ratiometric fluorescent nanoprobe for specific and visual detection of zearalenone. Microchem. J. 2020;157 [Google Scholar]
- 134.Cui F.C., Ye Y.L., Ping J.F., Sun X.L. Carbon dots: current advances in pathogenic bacteria monitoring and prospect applications. Biosens. Bioelectron. 2020;156 doi: 10.1016/j.bios.2020.112085. [DOI] [PubMed] [Google Scholar]
- 135.Li S.J., Wang J.P., Sheng W., Wen W.J., Gu Y., Wang S. Fluorometric lateral flow immunochromatographic zearalenone assay by exploiting a quencher system composed of carbon dots and silver nanoparticles. Microchim. Acta. 2018;185:388. doi: 10.1007/s00604-018-2916-1. [DOI] [PubMed] [Google Scholar]
- 136.You M.L., Lin M., Gong Y., Wang S.R., Li A., Ji L.Y., Zhao H.X., Ling K., Wen T., Huang Y., Gao D.F., Ma Q., Wang T.Z., Ma A.Q., Li X.L., Xu F. Household fluorescent lateral flow strip platform for sensitive and quantitative prognosis of heart failure using dual-color upconversion nanoparticles. ACS Nano. 2017;11:6261–6270. doi: 10.1021/acsnano.7b02466. [DOI] [PubMed] [Google Scholar]
- 137.Chatterjee K., Kuo C.W., Chen A., Chen P.L. Detection of residual rifampicin in urine via fluorescence quenching of gold nanoclusters on paper. J. Nanobiotechnol. 2015;13:46. doi: 10.1186/s12951-015-0105-5. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 138.Li D., Kumari B., Makabenta J.M., Gupta A., Rotello V. Effective detection of bacteria using metal nanoclusters. Nanoscale. 2019;11:22172–22181. doi: 10.1039/c9nr08510f. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 139.Zhang Z.K., Liu Y., Liu P.F., Yang L., Jiang X.Y., Luo D., Yang D.Y. Non-invasive detection of gastric cancer relevantd-amino acids with luminescent DNA/silver nanoclusters. Nanoscale. 2017;9:19367–19373. doi: 10.1039/c7nr07337b. [DOI] [PubMed] [Google Scholar]
- 140.Wang L.D., Liu M., Chen Y. Carbon dots and terbium co-enhanced fluorescence of europium nanoparticles for cell imaging. J. Biomed. Nanotechnol. 2018;14:1898–1905. doi: 10.1166/jbn.2018.2630. [DOI] [PubMed] [Google Scholar]
- 141.Wang L.D., Xu D., Gao J., Chen X., Duo Y.H., Zhang H. Semiconducting quantum dots: modification and applications in biomedical science. Sci. China Mater. 2020;63:1631–1650. [Google Scholar]
- 142.Diaz-Gonzalez M., de la Escosura-Muniz A., Fernandez-Arguelles M.T., Alonso F.J.G., Costa-Fernandez J.M. Quantum dot bioconjugates for diagnostic applications. Top. Curr. Chem. 2020;378:35. doi: 10.1007/s41061-020-0296-6. [DOI] [PubMed] [Google Scholar]
- 143.Zhu S.J., Song Y.B., Wang J., Wan H., Zhang Y., Ning Y., Yang B. Photoluminescence mechanism in graphene quantum dots: quantum confinement effect and surface/edge state. Nano Today. 2017;13:10–14. [Google Scholar]
- 144.Hu J., Jiang Y.Z., Wu L.L., Wu Z., Bi Y.H., Wong G., Qiu X.G., Chen J.J., Pang D.W., Zhang Z.L. Dual-signal readout nanospheres for rapid point-of-care detection of Ebola virus glycoprotein. Anal. Chem. 2017;89:13105–13111. doi: 10.1021/acs.analchem.7b02222. [DOI] [PubMed] [Google Scholar]
- 145.Ji C.Y., Zhou Y.Q., Leblanc R.M., Peng Z.L. Recent developments of carbon dots in biosensing: a review. ACS Sens. 2020;5:2724–2741. doi: 10.1021/acssensors.0c01556. [DOI] [PubMed] [Google Scholar]
- 146.Zhu Z.J., Zhai Y.L., Li Z.H., Zhu P.Y., Mao S., Zhu C.Z., Du D., Belfiore L.A., Tang J.G., Lin Y.H. Red carbon dots: optical property regulations and applications. Mater. Today. 2019;30:52–79. [Google Scholar]
- 147.Dhenadhayalan N., Lin K.C., Saleh T.A. Recent advances in functionalized carbon dots toward the design of efficient materials for sensing and catalysis applications. Small. 2020;16 doi: 10.1002/smll.201905767. [DOI] [PubMed] [Google Scholar]
- 148.Qu J.H., Wei Q.Y., Sun D.W. Carbon dots: principles and their applications in food quality and safety detection. Crit. Rev. Food Sci. Nutr. 2018;58:2466–2475. doi: 10.1080/10408398.2018.1437712. [DOI] [PubMed] [Google Scholar]
- 149.Sharma A., Das J. Small molecules derived carbon dots: synthesis and applications in sensing, catalysis, imaging, and biomedicine. J. Nanobiotechnol. 2019;17:92. doi: 10.1186/s12951-019-0525-8. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 150.Xu L.D., Zhang Q., Ding S.N., Xu J.J., Chen H.Y. Ultrasensitive detection of severe fever with thrombocytopenia syndrome virus based on immunofluorescent carbon Dots/SiO2 nanosphere-based lateral flow assay. ACS Omega. 2019;4:21431–21438. doi: 10.1021/acsomega.9b03130. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 151.Wang J., Li D.Q., Qiu Y., Liu X.Y., Huang L., Wen H.M., Hu J. An europium functionalized carbon dot-based fluorescence test paper for visual and quantitative point-of-care testing of anthrax biomarker. Talanta. 2020;220 doi: 10.1016/j.talanta.2020.121377. [DOI] [PubMed] [Google Scholar]
- 152.Mao L.L., Lu Z.X., He N.Y., Zhang L.M., Deng Y., Duan D.M. A new method for improving the accuracy of miRNA detection with NaYF4:Yb,Er upconversion nanoparticles. Sci. China Chem. 2017;60:157–162. [Google Scholar]
- 153.Li X.M., Zhang F., Zhao D.Y. Highly efficient lanthanide upconverting nanomaterials: progresses and challenges. Nano Today. 2013;8:643–676. [Google Scholar]
- 154.Zhang L.M., Mao L.L., Lu Z.X., Deng Y., He N.Y. One-step hydrothermal synthesis of butanetetracarboxylic acid-coated NaYF4:Yb3+, Er3+ upconversion phosphors with enhancement upconversion luminescence. J. Nanosci. Nanotechnol. 2016;16:1220–1224. doi: 10.1166/jnn.2016.11679. [DOI] [PubMed] [Google Scholar]
- 155.Sun L.N., Wei R.Y., Feng J., Zhang H.J. Tailored lanthanide-doped upconversion nanoparticles and their promising bioapplication prospects. Coord. Chem. Rev. 2018;364:10–32. [Google Scholar]
- 156.Fan Y., Liu L., Zhang F. Exploiting lanthanide-doped upconversion nanoparticles with core/shell structures. Nano Today. 2019;25:68–84. [Google Scholar]
- 157.Wang Y.H., Song S.Y., Zhang S.T., Zhang H.J. Stimuli-responsive nanotheranostics based on lanthanide-doped upconversion nanoparticles for cancer imaging and therapy: current advances and future challenges. Nano Today. 2019;25:38–67. [Google Scholar]
- 158.Zhao Y., Zhang P.P., Wang J., Zhou L., Yang R.F. A novel electro-driven immunochromatography assay based on upconversion nanoparticles for rapid pathogen detection. Biosens. Bioelectron. 2020;152 doi: 10.1016/j.bios.2020.112037. [DOI] [PubMed] [Google Scholar]
- 159.Jin B.R., Yang Y.X., He R.Y., Park Y.I., Lee A., Bai D., Li F., Lu T.J., Xu F., Lin M. Lateral flow aptamer assay integrated smartphone-based portable device for simultaneous detection of multiple targets using upconversion nanoparticles. Sens. Actuators B Chem. 2018;276:48–56. [Google Scholar]
- 160.Shang L., Xu J., Nienhaus G.U. Recent advances in synthesizing metal nanocluster-based nanocomposites for application in sensing, imaging and catalysis. Nano Today. 2019;28 [Google Scholar]
- 161.Li T.T., Yi H., Liu Y., Wang Z.L., Liu S.Q., He N.Y., Liu H.N., Deng Y. One-step synthesis of DNA templated water-soluble Au–Ag bimetallic nanoclusters for ratiometric fluorescence detection of DNA. J. Biomed. Nanotechnol. 2018;14:150–160. doi: 10.1166/jbn.2018.2491. [DOI] [PubMed] [Google Scholar]
- 162.Li T.T., Yang J.J., Ali Z.S., Wang Z.L., Mou X.B., He N.Y., Wang Z.F. Synthesis of aptamer-functionalized Ag nanoclusters for MCF-7 breast cancer cells imaging. Sci. China Chem. 2017;60:370–376. [Google Scholar]
- 163.Li T.T., He N.Y., Wang J.H., Li S., Deng Y., Wang Z.L. Effects of the i-motif DNA loop on the fluorescence of silver nanoclusters. RSC Adv. 2016;6:22839–22844. [Google Scholar]
- 164.Guo S.J., Wang E.K. Noble metal nanomaterials: controllable synthesis and application in fuel cells and analytical sensors. Nano Today. 2011;6:240–264. [Google Scholar]
- 165.Yan R., Shou Z.X., Chen J., Wu H., Zhao Y., Qiu L., Jiang P.J., Mou X.Z., Wang J.H., Li Y.Q. On–off–on gold nanocluster-based fluorescent probe for rapid Escherichia coli differentiation, detection and bactericide screening. ACS Sustain. Chem. Eng. 2018;6:4504–4509. [Google Scholar]
- 166.Li W.T., Sun X.M., Zhao X., Wang W., Xu S.H., Luo X.L. Rapid pattern recognition of different types of sulphur-containing species as well as serum and bacteria discrimination using Au NCs-Cu2+ complexes. Chin. Chem. Lett. 2020;31:2473–2477. [Google Scholar]
- 167.Ye Y.D., Xia L., Xu D.D., Xing X.J., Pang D.W., Tang H.W. DNA-stabilized silver nanoclusters and carbon nanoparticles oxide: a sensitive platform for label-free fluorescence turn-on detection of HIV-DNA sequences. Biosens. Bioelectron. 2016;85:837–843. doi: 10.1016/j.bios.2016.06.001. [DOI] [PubMed] [Google Scholar]
- 168.Farzin L., Shamsipur M., Samandari L., Sheibani S. HIV biosensors for early diagnosis of infection: the intertwine of nanotechnology with sensing strategies. Talanta. 2020;206 doi: 10.1016/j.talanta.2019.120201. [DOI] [PubMed] [Google Scholar]
- 169.Ji H.W., Wu L., Pu F., Ren J.S., Qu X.G. Point-of-care identification of bacteria using protein-encapsulated gold nanoclusters. Adv. Healthc. Mater. 2018;7 doi: 10.1002/adhm.201701370. [DOI] [PubMed] [Google Scholar]
- 170.Zhao H.Y., Wang L., Zhu J., Wei H.P., Jiang W. Label-free nucleic acids detection based on DNA templated silver nanoclusters fluorescent probe. Talanta. 2015;138:163–168. doi: 10.1016/j.talanta.2015.02.021. [DOI] [PubMed] [Google Scholar]
- 171.Lee N., Wang C., Park J. User-friendly point-of-care detection of influenza A (H1N1) virus using light guide in three-dimensional photonic crystal. RSC Adv. 2018;8:22991–22997. doi: 10.1039/c8ra02596g. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 172.He Q., Yu D.M., Bao M.D., Korensky G., Chen J.H., Shin M.Y., Kim J.W., Park M., Qin P.W., Du K. High-throughput and all-solution phase African Swine Fever Virus (ASFV) detection using CRISPR-Cas12a and fluorescence based point-of-care system. Biosens. Bioelectron. 2020;154 doi: 10.1016/j.bios.2020.112068. [DOI] [PubMed] [Google Scholar]
- 173.Li X.P., Lu D.L., Sheng Z.H., Chen K., Guo X.B., Jin M.L., Han H.Y. A fast and sensitive immunoassay of avian influenza virus based on label-free quantum dot probe and lateral flow test strip. Talanta. 2012;100:1–6. doi: 10.1016/j.talanta.2012.08.041. [DOI] [PubMed] [Google Scholar]
- 174.Weng X., Neethirajan S. Immunosensor based on antibody-functionalized MoS2 for rapid detection of avian coronavirus on cotton thread. IEEE Sens. J. 2018;18:4358–4363. doi: 10.1109/JSEN.2018.2829084. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 175.Chen H., Das A., Bi L., Choi N., Moon J.I., Wu Y., Park S. Recent advances in surface-enhanced Raman scattering-based microdevices for point-of-care diagnosis of viruses and bacteria. Nanoscale. 2020;12:21560–21570. doi: 10.1039/d0nr06340a. [DOI] [PubMed] [Google Scholar]
- 176.Granger J.H., Schlotter N.E., Crawford A.C., Porter M.D. Prospects for point-of-care pathogen diagnostics using surface-enhanced Raman scattering (SERS) Chem. Soc. Rev. 2016;45:3865–3882. doi: 10.1039/c5cs00828j. [DOI] [PubMed] [Google Scholar]
- 177.Vendrell M., Maiti K.K., Dhaliwal K., Chang Y.T. Surface-enhanced Raman scattering in cancer detection and imaging. Trends Biotechnol. 2013;31:249–257. doi: 10.1016/j.tibtech.2013.01.013. [DOI] [PubMed] [Google Scholar]
- 178.Langer J., de Aberasturi D.J., Aizpurua J., Alvarez-Puebla R.A., Auguie B., Baumberg J.J., Bazan G.C., Bell S.E.J., Boisen A., Brolo A.G., Choo J., Cialla-May D., Deckert V., Fabris L., Faulds K., de Abajo F.J.G., Goodacre R., Graham D., Haes A.J., Haynes C.L., Huck C., Itoh T., Ka M., Kneipp J., Kotov N.A., Kuang H., Le E.C., Lee Ru, H.K., Li J.F., Ling X.Y., Maier S.A., Mayerhofer T., Moskovits M., Murakoshi K., Nam J.M., Nie S., Ozaki Y., Pastoriza-Santos I., Perez-Juste J., Popp J., Pucci A., Reich S., Ren B., Schatz G.C., Shegai T., Schlucker S., Tay L.L., Thomas K.G., Tian Z.Q., Van Duyne R.P., Vo-Dinh T., Wang Y., Willets K.A., Xu C., Xu H., Xu Y., Yamamoto Y.S., Zhao B., Liz-Marzan L.M. Present and future of surface-enhanced Raman scattering. ACS Nano. 2020;14:28–117. doi: 10.1021/acsnano.9b04224. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 179.Zrimsek A.B., Chiang N.H., Mattei M., Zaleski S., McAnally M.O., Chapman C.T., Henry A.I., Schatz G.C., Van Duyne R.P. Single-molecule chemistry with surface- and tip-enhanced Raman spectroscopy. Chem. Rev. 2017;117:7583–7613. doi: 10.1021/acs.chemrev.6b00552. [DOI] [PubMed] [Google Scholar]
- 180.Xu K.C., Zhou R., Takei K., Hong M.H. Toward flexible surface‐enhanced Raman scattering (SERS) sensors for point‐of‐care diagnostics. Adv. Sci. 2019;6 doi: 10.1002/advs.201900925. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 181.Gao X.F., Zheng P., Kasani S., Wu S., Yang F., Lewis S., Nayeem S., Engler-Chiurazzi E.B., Wigginton J.G., Simpkins J.W., Wu N.Q. Paper-based surface-enhanced raman scattering lateral flow strip for detection of neuron-specific enolase in blood plasma. Anal. Chem. 2017;89:10104–10110. doi: 10.1021/acs.analchem.7b03015. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 182.Huang C.C., Cheng C.Y., Lai Y.S. Paper-based flexible surface enhanced Raman scattering platforms and their applications to food safety. Trends Food Sci. Technol. 2020;100:349–358. [Google Scholar]
- 183.Fan M.K., Andrade G.F.S., Brolo A.G. A review on recent advances in the applications of surface-enhanced Raman scattering in analytical chemistry. Anal. Chim. Acta. 2020;1097:1–29. doi: 10.1016/j.aca.2019.11.049. [DOI] [PubMed] [Google Scholar]
- 184.Shanmukh S., Jones L., Driskell J., Zhao Y.P., Dluhy R., Tripp R.A. Rapid and sensitive detection of respiratory virus molecular signatures using a silver nanorod array SERS substrate. Nano Lett. 2006;6:2630–2636. doi: 10.1021/nl061666f. [DOI] [PubMed] [Google Scholar]
- 185.Maneeprakorn W., Bamrungsap S., Apiwat C., Wiriyachaiporn N. Surface-enhanced Raman scattering based lateral flow immunochromatographic assay for sensitive influenza detection. RSC Adv. 2016;6:112079–112085. [Google Scholar]
- 186.Pang Y.F., Wan N., Shi L.L., Wang C.W., Sun Z.W., Xiao R., Wang S.Q. Dual-recognition surface-enhanced Raman scattering (SERS) biosensor for pathogenic bacteria detection by using vancomycin-SERS tags and aptamer-Fe3O4@Au. Anal. Chim. Acta. 2019;1077:288–296. doi: 10.1016/j.aca.2019.05.059. [DOI] [PubMed] [Google Scholar]
- 187.Shi L.L., Xu L., Xiao R., Zhou Z.H., Wang C.W., Wang S.Q., Gu B. Rapid, quantitative, high-sensitive detection of Escherichia coli O157:H7 by gold-shell silica-core nanospheres-based surface-enhanced Raman scattering lateral flow immunoassay. Front. Microbiol. 2020;11 doi: 10.3389/fmicb.2020.596005. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 188.Lee H.G., Choi W., Yang S.Y., Kim D.H., Park S.G., Lee M.Y., Jung H.S. PCR-coupled paper-based surface-enhanced Raman scattering (SERS) sensor for rapid and sensitive detection of respiratory bacterial DNA. Sens. Actuators B Chem. 2021;326 [Google Scholar]
- 189.Kim H.N., Ren W.X., Kim J.S., Yoon J. Fluorescent and colorimetric sensors for detection of lead, cadmium, and mercury ions. Chem. Soc. Rev. 2012;41:3210–3244. doi: 10.1039/c1cs15245a. [DOI] [PubMed] [Google Scholar]
- 190.Priyadarshini E., Pradhan N. Gold nanoparticles as efficient sensors in colorimetric detection of toxic metal ions: a review. Sens. Actuators B Chem. 2017;238:888–902. [Google Scholar]
- 191.Morbioli G.G., Mazzu-Nascimento T., Stockton A.M., Carrilho E. Technical aspects and challenges of colorimetric detection with microfluidic paper-based analytical devices (μPADs) – a review. Anal. Chim. Acta. 2017;970:1–22. doi: 10.1016/j.aca.2017.03.037. [DOI] [PubMed] [Google Scholar]
- 192.Sriram G., Bhat M.P., Patil P., Uthappa U.T., Jung H.Y., Altalhi T., Kumeria T., Aminabhavi T.M., Pai R.K., Madhuprasad, Kurkuri M.D. Paper-based microfluidic analytical devices for colorimetric detection of toxic ions: a review. TrAC Trends Anal. Chem. 2017;93:212–227. [Google Scholar]
- 193.Choi Y., Hwang J.H., Lee S.Y. Recent trends in nanomaterials-based colorimetric detection of pathogenic bacteria and viruses. Small Methods. 2018;2 doi: 10.1002/smtd.201700351. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 194.Calvert A.E., Biggerstaff B.J., Tanner N.A., Lauterbach M., Lanciotti R.S. Rapid colorimetric detection of Zika virus from serum and urine specimens by reverse transcription loop-mediated isothermal amplification (RT-LAMP) Plos One. 2017;12 doi: 10.1371/journal.pone.0185340. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 195.Roy S., Mohd-Naim N.F., Safavieh M., Ahmed M.U. Colorimetric nucleic acid detection on paper microchip using loop mediated isothermal amplification and crystal violet dye. ACS Sens. 2017;2:1713–1720. doi: 10.1021/acssensors.7b00671. [DOI] [PubMed] [Google Scholar]
- 196.Wu J.J.X., Wang X.Y., Wang Q., Lou Z.P., Li S.R., Zhu Y.Y., Qin L., Wei H. Nanomaterials with enzyme-like characteristics (nanozymes): next-generation artificial enzymes (II) Chem. Soc. Rev. 2019;48:1004–1076. doi: 10.1039/c8cs00457a. [DOI] [PubMed] [Google Scholar]
- 197.Wang Z.F., Yang X., Yang J.J., Jiang Y.Y., He N.Y. Peroxidase-like activity of mesoporous silica encapsulated Pt nanoparticle and its application in colorimetric immunoassay. Anal. Chim. Acta. 2015;862:53–63. doi: 10.1016/j.aca.2014.12.046. [DOI] [PubMed] [Google Scholar]
- 198.Wang Z.F., Yang X., Feng J., Tang Y.J., Jiang Y.Y., He N.Y. Label-free detection of DNA by combining gated mesoporous silica and catalytic signal amplification of platinum nanoparticles. Analyst. 2014;139:6088–6091. doi: 10.1039/c4an01539h. [DOI] [PubMed] [Google Scholar]
- 199.Vilela D., Gonzalez M.C., Escarpa A. Sensing colorimetric approaches based on gold and silver nanoparticles aggregation: chemical creativity behind the assay. A review. Anal. Chim. Acta. 2012;751:24–43. doi: 10.1016/j.aca.2012.08.043. [DOI] [PubMed] [Google Scholar]
- 200.Alhogail S., Suaifan G.A.R.Y., Bikker F.J., Kaman W.E., Weber K., Cialla-May D., Popp J., Zourob M.M. Rapid colorimetric detection of Pseudomonas aeruginosa in clinical isolates using a magnetic nanoparticle biosensor. ACS Omega. 2019;4:21684–21688. doi: 10.1021/acsomega.9b02080. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 201.Guo L.H., Xu Y., Ferhan A.R., Chen G.N., Kim D.H. Oriented gold nanoparticle aggregation for colorimetric sensors with surprisingly high analytical figures of merit. J. Am. Chem. Soc. 2013;135:12338–12345. doi: 10.1021/ja405371g. [DOI] [PubMed] [Google Scholar]
- 202.Aldewachi H., Chalati T., Woodroofe M.N., Bricklebank N., Sharrack B., Gardiner P. Gold nanoparticle-based colorimetric biosensors. Nanoscale. 2018;10:18–33. doi: 10.1039/c7nr06367a. [DOI] [PubMed] [Google Scholar]
- 203.Sun J.S., Xianyu Y.L., Jiang X.Y. Point-of-care biochemical assays using gold nanoparticle-implemented microfluidics. Chem. Soc. Rev. 2014;43:6239–6253. doi: 10.1039/c4cs00125g. [DOI] [PubMed] [Google Scholar]
- 204.Godakhindi V.S., Kang P.Y., Serre M., Revuru N.A., Zou J.M.H., Roner M.R., Levitz R., Kahn J.S., Randrianalisoa J., Qin Z.P. Tuning the gold nanoparticle colorimetric assay by nanoparticle size, concentration, and size combinations for oligonucleotide detection. ACS Sens. 2017;2:1627–1636. doi: 10.1021/acssensors.7b00482. [DOI] [PubMed] [Google Scholar]
- 205.Zheng L.Y., Cai G.Z., Wang S.Y., Liao M., Li Y.B., Lin J.H. A microfluidic colorimetric biosensor for rapid detection of Escherichia coli O157:H7 using gold nanoparticle aggregation and smart phone imaging. Biosens. Bioelectron. 2019;124–125:143–149. doi: 10.1016/j.bios.2018.10.006. [DOI] [PubMed] [Google Scholar]
- 206.Yang H.H., Xiao M.S., Lai W., Wan Y., Li L., Pei H. Stochastic DNA dual-walkers for ultrafast colorimetric bacteria detection. Anal. Chem. 2020;92:4990–4995. doi: 10.1021/acs.analchem.9b05149. [DOI] [PubMed] [Google Scholar]
- 207.Deng J.Q., Yang M.Z., Wu J., Zhang W., Jiang X.Y. A self-contained chemiluminescent lateral flow assay for point-of-care testing. Anal. Chem. 2018;90:9132–9137. doi: 10.1021/acs.analchem.8b01543. [DOI] [PubMed] [Google Scholar]
- 208.Joung H.A., Oh Y.K., Kim M.G. An automatic enzyme immunoassay based on a chemiluminescent lateral flow immunosensor. Biosens. Bioelectron. 2014;53:330–335. doi: 10.1016/j.bios.2013.10.004. [DOI] [PubMed] [Google Scholar]
- 209.Ibarra-Silva E., Raff A.B., Cardenas A., Franco W. Point‐of‐care detection of neutrophils in live skin microsamples using chemiluminescence. J. Biophotonics. 2020;13 doi: 10.1002/jbio.201960170. [DOI] [PubMed] [Google Scholar]
- 210.Liu M., Wang Z.Y., Zhang C.Y. Recent advance in chemiluminescence assay and its biochemical applications. Chin. J. Anal. Chem. 2016;44:1934–1941. [Google Scholar]
- 211.Maffert P., Reverchon S., Nasser W., Rozand C., Abaibou H. New nucleic acid testing devices to diagnose infectious diseases in resource-limited settings. Eur. J. Clin. Microbiol. Infect. Dis. 2017;36:1717–1731. doi: 10.1007/s10096-017-3013-9. [DOI] [PubMed] [Google Scholar]
- 212.Park J.Y., Kricka L.J. Prospects for the commercialization of chemiluminescence-based point-of-care and on-site testing devices. Anal. Bioanal. Chem. 2014;406:5631–5637. doi: 10.1007/s00216-014-7697-8. [DOI] [PubMed] [Google Scholar]
- 213.Wang C.G., Xiao R., Dong P.T., Wu X.Z., Rong Z., Xin L., Tang J., Wang S.Q. Ultra-sensitive, high-throughput detection of infectious diarrheal diseases by portable chemiluminescence imaging. Biosens. Bioelectron. 2014;57:36–40. doi: 10.1016/j.bios.2014.01.016. [DOI] [PubMed] [Google Scholar]
- 214.Hamilton M.S., Abel D.M., Ballam Y.J., Otto M.K., Nickell A.F., Pence L.M., Appleman J.R., Shimasaki C.D., Achyuthan K.E. Clinical evaluation of the ZstatFlu-II test: a chemiluminescent rapid diagnostic test for influenza virus. J. Clin. Microbiol. 2002;40:2331–2334. doi: 10.1128/JCM.40.7.2331-2334.2002. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 215.Achyuthan K.E., Pence L.M., Appleman J.R., Shimasaki C.D. ZstatFlu (R)-II test: a chemiluminescent neuraminidase assay for influenza viral diagnostics. Luminescence. 2003;18:131–139. doi: 10.1002/bio.714. [DOI] [PubMed] [Google Scholar]
- 216.Kim J.K., Yoon Y.M., Jang W.J., Choi Y.J., Hong S.C., Cho J.H. Comparison study between MAST CLA and OPTIGEN. Am. J. Rhinol. Allergy. 2011;25:e156–e159. doi: 10.2500/ajra.2011.25.3646. [DOI] [PubMed] [Google Scholar]
- 217.Kurihara T., Yanagida A., Yokoi H., Koyata A., Matsuya T., Ogawa J., Okamura Y., Miyamoto D. Evaluation of cardiac assays on a benchtop chemiluminescent enzyme immunoassay analyzer, PATHFAST. Anal. Biochem. 2008;375:144–146. doi: 10.1016/j.ab.2007.12.030. [DOI] [PubMed] [Google Scholar]
- 218.He L., Yang H.W., Xiao P.F., Singh R., He N.Y., Liu B., Li Z.Y. Highly selective, sensitive and rapid detection of Escherichia coli O157:H7 using duplex PCR and magnetic nanoparticle-based chemiluminescence assay. J. Biomed. Nanotechnol. 2017;13:1243–1252. [Google Scholar]
- 219.Wang J.H., Lu P., Yan J.N., Zhang Y.F., Huang L.Y., Ali Z.S., Liu B., Li Z.Y., He N.Y. Rapid and sensitive detection of RNA viruses based on reverse transcription loop-mediated isothermal amplification, magnetic nanoparticles, and chemiluminescence. J. Biomed. Nanotechnol. 2016;12:710–716. doi: 10.1166/jbn.2016.2244. [DOI] [PubMed] [Google Scholar]
- 220.Liu M., Hu P., Zhang G., Zeng Y., Yang H.W., Fan J., Jin L., Liu H.N., Deng Y., Li S., Zeng X., Elingarami S., He N.Y. Copy number variation analysis by ligation-dependent PCR based on magnetic nanoparticles and chemiluminescence. Theranostics. 2015;5:71–85. doi: 10.7150/thno.10117. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 221.Yang H.W., Liang W.B., He N.Y., Deng Y., Li Z.Y. Chemiluminescent labels released from long spacer arm-functionalized magnetic particles: a novel strategy for ultrasensitive and highly selective detection of pathogen infections. ACS Appl. Mater. Interfaces. 2015;7:774–781. doi: 10.1021/am507203s. [DOI] [PubMed] [Google Scholar]
- 222.Yang H.W., Liang W.B., Si J., Li Z.Y., He N.Y. Long spacer arm-functionalized magnetic nanoparticle platform for enhanced chemiluminescent detection of hepatitis B virus. J. Biomed. Nanotechnol. 2014;10:3610–3619. doi: 10.1166/jbn.2014.2047. [DOI] [PubMed] [Google Scholar]
- 223.Ma N.N., Ma C., Wang N.Y., Li C.Y., Elingarami S., Mou X.B., Tang Y.J., Zheng S., He N.Y. Application of functional microsphere in human hepatitis B virus surface antigen detection. J. Nanosci. Nanotechnol. 2014;14:3348–3355. doi: 10.1166/jnn.2014.8236. [DOI] [PubMed] [Google Scholar]
- 224.Ali Z., Liang W.B., Jin L., Tang Y.J., Mou X.B., Shah M.A.A., Yang H.W., Deng Y., He N.Y., Li Z.Y. Development of magnetic nanoparticles based nucleic acid extraction method and application in hepatitis C virus chemiluminescent detection. Sci. Adv. Mater. 2015;7:1233–1240. [Google Scholar]
- 225.He N.Y., Wang F., Ma C., Li C.Y., Zeng X., Deng Y., Zhang L.M., Li Z.Y. Chemiluminescence analysis for HBV-DNA hybridization detection with magnetic nanoparticles based DNA extraction from positive whole blood samples. J. Biomed. Nanotechnol. 2013;9:267–273. doi: 10.1166/jbn.2013.1478. [DOI] [PubMed] [Google Scholar]
- 226.Xi Z.J., Huang R.R., Li Z.Y., He N.Y., Wang T., Su E.B., Deng Y. Selection of HBsAg-specific DNA aptamers based on carboxylated magnetic nanoparticles and their application in the rapid and simple detection of hepatitis B virus infection. ACS Appl. Mater. Interfaces. 2015;7:11215–11223. doi: 10.1021/acsami.5b01180. [DOI] [PubMed] [Google Scholar]
- 227.Li Z.Y., Yang H.W., He N.Y., Liang W.B., Ma C., Shah M.A.A., Tang Y.J., Li S., Liu H.N., Jiang H.S., Guo Y.F. Solid-phase hybridization efficiency improvement on the magnetic nanoparticle surface by using dextran as molecular arms. J. Biomed. Nanotechnol. 2013;9:1945–1949. doi: 10.1166/jbn.2013.1694. [DOI] [PubMed] [Google Scholar]
- 228.Ali Z., Wang J.H., Tang Y.J., Liu B., He N.Y., Li Z.Y. Simultaneous detection of multiple viruses based on chemiluminescence and magnetic separation. Biomater. Sci. 2017;5:57–66. doi: 10.1039/c6bm00527f. [DOI] [PubMed] [Google Scholar]
- 229.Firdous S., Anwar S., Rafya R. Development of surface plasmon resonance (SPR) biosensors for use in the diagnostics of malignant and infectious diseases. Laser Phys. Lett. 2018;15 [Google Scholar]
- 230.Patil P.O., Pandey G.R., Patil A.G., Borse V.B., Deshmukh P.K., Patil D.R., Tade R.S., Nangare S.N., Khan Z.G., Patil A.M., More M.P., Veerapandian M., Bari S.B. Graphene-based nanocomposites for sensitivity enhancement of surface plasmon resonance sensor for biological and chemical sensing: a review. Biosens. Bioelectron. 2019;139 doi: 10.1016/j.bios.2019.111324. [DOI] [PubMed] [Google Scholar]
- 231.Csaki A., Stranik O., Fritzsche W. Localized surface plasmon resonance based biosensing. Expert Rev. Mol. Diagn. 2018;18:279–296. doi: 10.1080/14737159.2018.1440208. [DOI] [PubMed] [Google Scholar]
- 232.Malmqvist M. Biospecific interaction analysis using biosensor technology. Nature. 1993;361:186–187. doi: 10.1038/361186a0. [DOI] [PubMed] [Google Scholar]
- 233.Chen C., Wang J.S. Optical biosensors: an exhaustive and comprehensive review. Analyst. 2020;145:1605–1628. doi: 10.1039/c9an01998g. [DOI] [PubMed] [Google Scholar]
- 234.Rich R.L., Myszka D.G. Spying on HIV with SPR. Trends Microbiol. 2003;11:124–133. doi: 10.1016/s0966-842x(03)00025-8. [DOI] [PubMed] [Google Scholar]
- 235.Trilling A.K., Hesselink T., van Houwelingen A., Cordewener J.H.G., Jongsma M.A., Schoffelen S., van Hest J.C.M., Zuilhof H., Beekwilder J. Orientation of llama antibodies strongly increases sensitivity of biosensors. Biosens. Bioelectron. 2014;60:130–136. doi: 10.1016/j.bios.2014.04.017. [DOI] [PubMed] [Google Scholar]
- 236.Singh V.V., Gupta G., Batra A., Nigam A.K., Boopathi M., Gutch P.K., Tripathi B.K., Srivastava A., Samuel M., Agarwal G.S., Singh B., Vijayaraghavan R. Greener electrochemical synthesis of high quality graphene nanosheets directly from pencil and its SPR sensing application. Adv. Funct. Mater. 2012;22:2352–2362. [Google Scholar]
- 237.Singh P. SPR biosensors: historical perspectives and current challenges. Sens. Actuators B Chem. 2016;229:110–130. [Google Scholar]
- 238.Masson J.F. Surface plasmon resonance clinical biosensors for medical diagnostics. ACS Sens. 2017;2:16–30. doi: 10.1021/acssensors.6b00763. [DOI] [PubMed] [Google Scholar]
- 239.Chinowsky T.M., Soelberg S.D., Baker P., Swanson N.R., Kauffman P., Mactutis A., Grow M.S., Atmar R., Yee S.S., Furlong C.E. Portable 24-analyte surface plasmon resonance instruments for rapid, versatile biodetection. Biosens. Bioelectron. 2007;22:2268–2275. doi: 10.1016/j.bios.2006.11.026. [DOI] [PubMed] [Google Scholar]
- 240.Trzaskowski M., Ciach T. SPR system for on-site detection of biological warfare. Curr. Anal. Chem. 2017;13:144–149. [Google Scholar]
- 241.Na W., Kang M., Jang D., Shin S. Miniaturized surface plasmon resonance biosensor with vacuum-driven hydrodynamic focusing. Sens. Actuators B Chem. 2018;254:64–71. [Google Scholar]
- 242.Chen H.X., Liu F., Koh K., Lee J., Ye Z.H., Yin T.T., Sun L.Z. Sensitive detection of tuberculosis using nanoparticle-enhanced surface plasmon resonance. Microchim. Acta. 2013;180:431–436. [Google Scholar]
- 243.Lynn N.S., Dandy D.S. Passive microfluidic pumping using coupled capillary/evaporation effects. Lab Chip. 2009;9:3422–3429. doi: 10.1039/b912213c. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 244.Bhardwaj H., Sumana G., Marquette C.A. A label-free ultrasensitive microfluidic surface plasmon resonance biosensor for aflatoxin B1 detection using nanoparticles integrated gold chip. Food Chem. 2020;307 doi: 10.1016/j.foodchem.2019.125530. (UNSP) [DOI] [PubMed] [Google Scholar]
- 245.Nguyen T.T., Trinh K.T.L., Yoon W.J., Lee N.Y., Ju H. Integration of a microfluidic polymerase chain reaction device and surface plasmon resonance fiber sensor into an inline all-in-one platform for pathogenic bacteria detection. Sens. Actuators B Chem. 2017;242:1–8. [Google Scholar]
- 246.Marchesini G.R., Koopal K., Meulenberg E., Haasnoot W., Irth H. Spreeta-based biosensor assays for endocrine disruptors. Biosens. Bioelectron. 2007;22:1908–1915. doi: 10.1016/j.bios.2006.08.005. [DOI] [PubMed] [Google Scholar]
- 247.Zhao S.S., Bukar N., Toulouse J.L., Pelechacz D., Robitaille R., Pelletier J.N., Masson J.F. Miniature multi-channel SPR instrument for methotrexate monitoring in clinical samples. Biosens. Bioelectron. 2015;64:664–670. doi: 10.1016/j.bios.2014.09.082. [DOI] [PubMed] [Google Scholar]
- 248.Diao W., Tang M., Ding S.J., Li X.M., Cheng W.B., Mo F., Yan X.Y., Ma H.M., Yan Y.R. Highly sensitive surface plasmon resonance biosensor for the detection of HIV-related DNA based on dynamic and structural DNA nanodevices. Biosens. Bioelectron. 2018;100:228–234. doi: 10.1016/j.bios.2017.08.042. [DOI] [PubMed] [Google Scholar]
- 249.Chowdhury A.D., Takemura K., Khorish I.M., Nasrin F., Tun M.M.N., Morita K., Park E.Y. The detection and identification of dengue virus serotypes with quantum dot and AuNP regulated localized surface plasmon resonance. Nanoscale Adv. 2020;2:699–709. doi: 10.1039/c9na00763f. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 250.Tokel O., Yildiz U.H., Inci F., Durmus N.G., Ekiz O.O., Turker B., Cetin C., Rao S., Sridhar K., Natarajan N., Shafiee H., Dana A., Demirci U. Portable microfluidic integrated plasmonic platform for pathogen detection. Sci. Rep. 2015;5:9152. doi: 10.1038/srep09152. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 251.Trzaskowski M., Napiorkowska A., Augustynowicz-Kopec E., Ciach T. Detection of tuberculosis in patients with the use of portable SPR device. Sens. Actuators B Chem. 2018;260:786–792. [Google Scholar]
- 252.Kim S.E., Tieu M.V., Hwang S.Y., Lee M.H. Magnetic particles: their applications from sample preparations to biosensing platforms. Micromachines. 2020;11:302. doi: 10.3390/mi11030302. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 253.Schotter J., Kamp P.B., Becker A., Puhler A., Reiss G., Bruckl H. Comparison of a prototype magnetoresistive biosensor to standard fluorescent DNA detection. Biosens. Bioelectron. 2004;19:1149–1156. doi: 10.1016/j.bios.2003.11.007. [DOI] [PubMed] [Google Scholar]
- 254.Santiesteban O.J., Kaittanis C., Perez J.M. Identification of toxin inhibitors using a magnetic nanosensor-based assay. Small. 2014;10:1202–1211. doi: 10.1002/smll.201301824. [DOI] [PubMed] [Google Scholar]
- 255.Hong H.B., Krause H.J., Song K.B., Choi C.J., Chung M.A., Son S.W., Offenhausser A. Detection of two different influenza A viruses using a nitrocellulose membrane and a magnetic biosensor. J. Immunol. Methods. 2011;365:95–100. doi: 10.1016/j.jim.2010.12.005. [DOI] [PubMed] [Google Scholar]
- 256.Murzin D., Mapps D.J., Levada K., Belyaev V., Omelyanchik A., Panina L., Rodionova V. Ultrasensitive magnetic field sensors for biomedical applications. Sensors. 2020;20:1569. doi: 10.3390/s20061569. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 257.Sideris C., Khial P.P., Hajimiri A. Design and implementation of reference-free drift-cancelling CMOS magnetic sensors for biosensing applications. IEEE J. Solid-State Circuits. 2018;53:3065–3075. [Google Scholar]
- 258.Denmark D.J., Bustos-Perez X., Swain A., Phan M.H., Mohapatra S., Mohapatra S.S. Readiness of magnetic nanobiosensors for point-of-care commercialization. J. Electron. Mater. 2019;48:4749–4761. [Google Scholar]
- 259.Xianyu Y.L., Wang Q.L., Chen Y.P. Magnetic particles-enabled biosensors for point-of-care testing. TrAC Trends Anal. Chem. 2018;106:213–224. [Google Scholar]
- 260.Nabaei V., Chandrawati R., Heidari H. Magnetic biosensors: modelling and simulation. Biosens. Bioelectron. 2018;103:69–86. doi: 10.1016/j.bios.2017.12.023. [DOI] [PubMed] [Google Scholar]
- 261.Binasch G., Grunberg P., Saurenbach F., Zinn W. Enhanced magnetoresistance in layered magnetic structures with antiferromagnetic interlayer exchange. Phys. Rev. B. 1989;39:4828–4830. doi: 10.1103/physrevb.39.4828. [DOI] [PubMed] [Google Scholar]
- 262.Lin G.G., Makarov D., Schmidt O.G. Magnetic sensing platform technologies for biomedical applications. Lab Chip. 2017;17:1884–1912. doi: 10.1039/c7lc00026j. [DOI] [PubMed] [Google Scholar]
- 263.Su D.Q., Wu K., Saha R., Peng C.Y., Wang J.P. Advances in magnetoresistive biosensors. Micromachines. 2020;11:34. doi: 10.3390/mi11010034. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 264.Ren C.H., Bayin Q.G., Feng S.L., Fu Y.S., Ma X., Guo J.H. Biomarkers detection with magnetoresistance-based sensors. Biosens. Bioelectron. 2020;165 doi: 10.1016/j.bios.2020.112340. [DOI] [PubMed] [Google Scholar]
- 265.Wu K., Klein T., Krishna V.D., Su D.Q., Perez A.M., Wang J.P. Portable GMR handheld platform for the detection of influenza A virus. ACS Sens. 2017;2:1594–1601. doi: 10.1021/acssensors.7b00432. [DOI] [PubMed] [Google Scholar]
- 266.Su D.Q., Wu K., Krishna V.D., Klein T., Liu J.M., Feng Y.L., Perez A.M., Cheeran M.C.J., Wang J.P. Detection of influenza A virus in swine nasal swab samples with a wash-free magnetic bioassay and a handheld giant magnetoresistance sensing system. Front. Microbiol. 2019;10:1077. doi: 10.3389/fmicb.2019.01077. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 267.Wu K., Su D.Q., Saha R., Wong D., Wang J.P. Magnetic particle spectroscopy-based bioassays: methods, applications, advances, and future opportunities. J. Phys. D Appl. Phys. 2019;52 [Google Scholar]
- 268.Poller W.C., Lowa N., Wiekhorst F., Taupitz M., Wagner S., Moller K., Baumann G., Stangl V., Trahms L., Ludwig A. Magnetic particle spectroscopy reveals dynamic changes in the magnetic behavior of very small superparamagnetic iron oxide nanoparticles during cellular uptake and enables determination of cell-labeling efficacy. J. Biomed. Nanotechnol. 2016;12:337–346. doi: 10.1166/jbn.2016.2204. [DOI] [PubMed] [Google Scholar]
- 269.Wu K., Liu J.M., Su D.Q., Saha R., Wang J.P. Magnetic nanoparticle relaxation dynamics-based magnetic particle spectroscopy for rapid and wash-free molecular sensing. ACS Appl. Mater. Interfaces. 2019;11:22979–22986. doi: 10.1021/acsami.9b05233. [DOI] [PubMed] [Google Scholar]
- 270.Wu K., Su D.Q., Saha R., Liu J.M., Chugh V.K., Wang J.P. Magnetic particle spectroscopy: a short review of applications using magnetic nanoparticles. ACS Appl. Nano Mater. 2020;3:4972–4989. [Google Scholar]
- 271.Wu K., Liu J.M., Saha R., Su D.Q., Krishna V.D., Cheeran M.C.J., Wang J.P. Magnetic particle spectroscopy for detection of influenza A virus subtype H1N1. ACS Appl. Mater. Interfaces. 2020;12:13686–13697. doi: 10.1021/acsami.0c00815. [DOI] [PubMed] [Google Scholar]
- 272.Wongkaew N., Simsek M., Griesche C., Baeumner A.J. Functional nanomaterials and nanostructures enhancing electrochemical biosensors and lab-on-a-chip performances: recent progress, applications, and future perspective. Chem. Rev. 2019;119:120–194. doi: 10.1021/acs.chemrev.8b00172. [DOI] [PubMed] [Google Scholar]
- 273.Cesewski E., Johnson B.N. Electrochemical biosensors for pathogen detection. Biosens. Bioelectron. 2020;159 doi: 10.1016/j.bios.2020.112214. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 274.Reta N., Saint C.P., Michelmore A., Prieto-Simon B., Voelcker N.H. Nanostructured electrochemical biosensors for label-free detection of water- and food-borne pathogens. ACS Appl. Mater. Interfaces. 2018;10:6055–6072. doi: 10.1021/acsami.7b13943. [DOI] [PubMed] [Google Scholar]
- 275.Jiang P.J., Guo Z.J. Fluorescent detection of zinc in biological systems: recent development on the design of chemosensors and biosensors. Coord. Chem. Rev. 2004;248:205–229. [Google Scholar]
- 276.VanEngelenburg S.B., Palmer A.E. Fluorescent biosensors of protein function. Curr. Opin. Chem. Biol. 2008;12:60–65. doi: 10.1016/j.cbpa.2008.01.020. [DOI] [PubMed] [Google Scholar]
- 277.Pazos E., Vazquez O., Mascarenas J.L., Vazquez M.E. Peptide-based fluorescent biosensors. Chem. Soc. Rev. 2009;38:3348–3359. doi: 10.1039/b908546g. [DOI] [PubMed] [Google Scholar]
- 278.Tu Q., Chang C. Diagnostic applications of Raman spectroscopy. Nanomed. Nanotechnol. Biol. Med. 2012;8:545–558. doi: 10.1016/j.nano.2011.09.013. [DOI] [PubMed] [Google Scholar]
- 279.Tang L.H., Li J.H. Plasmon-based colorimetric nanosensors for ultrasensitive molecular diagnostics. ACS Sens. 2017;2:857–875. doi: 10.1021/acssensors.7b00282. [DOI] [PubMed] [Google Scholar]
- 280.Yu L., Li N. Noble metal nanoparticles-based colorimetric biosensor for visual quantification: a mini review. Chemosensors. 2019;7:53. [Google Scholar]
- 281.Shende P., Prabhakar B., Patil A. Color changing sensors: a multimodal system for integrated screening. TrAC Trends Anal. Chem. 2019;121 [Google Scholar]
- 282.Sun Y., Lu J.Z. Chemiluminescence-based aptasensors for various target analytes. Luminescence. 2018;33:1298–1305. doi: 10.1002/bio.3557. [DOI] [PubMed] [Google Scholar]
- 283.Xiao Q., Xu C.X. Research progress on chemiluminescence immunoassay combined with novel technologies. TrAC Trends Anal. Chem. 2020;124 [Google Scholar]
- 284.Al Mughairy B., Al-Lawati H.A.J. Recent analytical advancements in microfluidics using chemiluminescence detection systems for food analysis. TrAC Trends Anal. Chem. 2020;124 [Google Scholar]
- 285.Zhou J.R., Qi Q.Q., Wang C., Qian Y.F., Liu G.M., Wang Y.B., Fu L.L. Surface plasmon resonance (SPR) biosensors for food allergen detection in food matrices. Biosens. Bioelectron. 2019;142 doi: 10.1016/j.bios.2019.111449. [DOI] [PubMed] [Google Scholar]
- 286.Mahmoudpour M., Dolatabadi J.E.N., Torbati M., Tazehkand A.P., Homayouni-Rad A., de la Guardia M. Nanomaterials and new biorecognition molecules based surface plasmon resonance biosensors for mycotoxin detection. Biosens. Bioelectron. 2019;143 doi: 10.1016/j.bios.2019.111603. [DOI] [PubMed] [Google Scholar]
- 287.Stevens W.S., Scott L., Noble L., Gous N., Dheda K. Impact of the GeneXpert MTB/RIF technology on tuberculosis control. Microbiol. Spectr. 2017;5 doi: 10.1128/microbiolspec.tbtb2-0040-2016. UNSP TBTB2-0040-2016. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 288.Tanriverdi S., Chen L.J., Chen S.Q. A rapid and automated sample‐to‐result HIV load test for near‐patient application. J. Infect. Dis. 2010;201:S52–S58. doi: 10.1086/650387. [DOI] [PubMed] [Google Scholar]
- 289.Nair C.B., Manjula J., Subramani P.A., Nagendrappa P.B., Manoj M.N., Malpani S., Pullela P.K., Subbarao P.V., Ramamoorthy S., Ghosh S.K. Differential diagnosis of malaria on Truelab Uno®, a portable, real-time, microPCR device for point-of-care applications. Plos One. 2016;11 doi: 10.1371/journal.pone.0146961. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 290.Bissonnette L., Bergeron M.G. The GenePOC platform, a rational solution for extreme point-of-care testing. Micromachines. 2016;7:94. doi: 10.3390/mi7060094. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 291.Chan C.W., Parker K., Tesic V., Baldwin A., Tang N.Y., van Wijk X.M.R., Yeo K.J. Analytical and clinical evaluation of the automated elecsys anti–SARS-CoV-2 antibody assay on the Roche cobas e602 analyzer. Am. J. Clin. Pathol. 2020;154:620–626. doi: 10.1093/ajcp/aqaa155. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 292.Delliere S., Salmona M., Minier M., Gabassi A., Alanio A., Le Goff J., Delaugerre C., Chaix M.L., Grp S.-L.C.C.R. Evaluation of the COVID-19 IgG/IgM rapid test from orient gene biotech. J. Clin. Microbiol. 2020;58 doi: 10.1128/JCM.01233-20. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 293.Chan C.W., Shahul S., Coleman C., Tesic V., Parker K., Yeo K.J. Evaluation of the truvian easy check COVID-19 IgM/IgG lateral flow device for rapid anti-SARS-CoV-2 antibody detection. Am. J. Clin. Pathol. 2020;73:267–272. doi: 10.1093/ajcp/aqaa221. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 294.Petralia S., Conoci S. PCR technologies for point of care testing: progress and perspectives. ACS Sens. 2017;2:876–891. doi: 10.1021/acssensors.7b00299. [DOI] [PubMed] [Google Scholar]
- 295.Urusov A.E., Zherdev A.V., Dzantiev B.B. Towards lateral flow quantitative assays: detection approaches. Biosensors-Basel. 2019;9:89. doi: 10.3390/bios9030089. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 296.Magro L., Escadafal C., Garneret P., Jacquelin B., Kwasiborski A., Manuguerra J.C., Monti F., Sakuntabhai A., Vanhomwegen J., Lafaye P., Tabeling P. Paper microfluidics for nucleic acid amplification testing (NAAT) of infectious diseases. Lab Chip. 2017;17:2347–2371. doi: 10.1039/c7lc00013h. [DOI] [PubMed] [Google Scholar]
- 297.Yang H.W., Chen Z., Cao X.B., Li Z.Y., Stavrakis S., Choo J., deMello A.J., Howes P.D., He N. A sample-in-digital-answer-out system for rapid detection and quantitation of infectious pathogens in bodily fluids. Anal. Bioanal. Chem. 2018;410:7019–7030. doi: 10.1007/s00216-018-1335-9. [DOI] [PubMed] [Google Scholar]
- 298.Andersson M.E., Olofsson S., Lindh M. Comparison of the filmarray assay and in-house real-time PCR for detection of respiratory infection. Scand. J. Infect. Dis. 2014;46:897–901. doi: 10.3109/00365548.2014.951681. [DOI] [PubMed] [Google Scholar]
- 299.Goldenberg S.D., Edgeworth J.D. The Enigma mL FluAB-RSV assay: a fully automated molecular test for the rapid detection of influenza A, B and respiratory syncytial viruses in respiratory specimens. Expert Rev. Mol. Diagn. 2015;15:23–32. doi: 10.1586/14737159.2015.983477. [DOI] [PubMed] [Google Scholar]
- 300.Nie S.P., Roth R.B., Stiles J., Mikhlina A., Lu X.D., Tang Y.W., Babady N.E. Evaluation of Alere i influenza A&B for rapid detection of influenza viruses A and B. J. Clin. Microbiol. 2014;52:3339–3344. doi: 10.1128/JCM.01132-14. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 301.Banerjee D., Michael J., Schmitt B., Salimnia H., Mhaissen N., Goldfarb D.M., Lachance P., Faron M.L., Aufderheide T., Ledeboer N., Weissfeld A., Selvarangan R. Multicenter clinical evaluation of the revogene strep A molecular assay for detection of Streptococcus pyogenesfrom throat swab specimens. J. Clin. Microbiol. 2020;58 doi: 10.1128/JCM.01775-19. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 302.Alby K., Popowitch E.B., Miller M.B. Comparative evaluation of the nanosphere verigene RV+ assay and the simplexa flu A/B & RSV kit for detection of influenza and respiratory syncytial viruses. J. Clin. Microbiol. 2013;51:352–353. doi: 10.1128/JCM.02504-12. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 303.Batra R., Olivieri L.G., Rubin D., Vallari A., Pearce S., Olivo A., Prostko J., Nebbia G., Douthwaite S., Rodgers M., Cloherty G. A comparative evaluation between the Abbott Panbio™ COVID-19 IgG/IgM rapid test device and Abbott Architect™ SARS CoV-2 IgG assay. J. Clin. Virol. 2020;132 doi: 10.1016/j.jcv.2020.104645. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 304.Das S., Peck R.B., Barney R., Jang I.K., Kahn M., Zhu M.L., Domingo G.J. Performance of an ultra-sensitive Plasmodium falciparum HRP2-based rapid diagnostic test with recombinant HRP2, culture parasites, and archived whole blood samples. Malar. J. 2018;17:118. doi: 10.1186/s12936-018-2268-7. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 305.Tseroni M., Pervanidou D., Tserkezou P., Rachiotis G., Pinaka O., Baka A., Georgakopoulou T., Vakali A., Dionysopoulou M., Terzaki I., Marka A., Detsis M., Evlampidou Z., Mpimpa A., Vassalou E., Tsiodras S., Tsakris A., Kremastinou J., Hadjichristodoulou C., Project M. Field application of SD bioline malaria Ag Pf/Pan rapid diagnostic test for malaria in Greece. Plos One. 2015;10 doi: 10.1371/journal.pone.0120367. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 306.Torlak E., Akan I.M., Inal M. Evaluation of rapidchek select for the screening of Salmonella in meat and meat products. J. Microbiol. Methods. 2012;90:217–219. doi: 10.1016/j.mimet.2012.05.010. [DOI] [PubMed] [Google Scholar]
- 307.Hoerner R., Feldpausch J., Gray R.L., Curry S., Islam Z., Goldy T., Klein F., Tadese T., Rice J., Mozola M., Hammack T., Marrow T., Fernandez M.C. Reveal®Salmonella 2.0 test for detection of Salmonella spp. in foods and environmental samples. J. AOAC Int. 2011;94:1467–1480. doi: 10.5740/jaoacint.11-051. [DOI] [PubMed] [Google Scholar]
- 308.Ley B., Thriemer K., Ame S.M., Mtove G.M., von Seidlein L., Amos B., Hendriksen I.C.E., Mwambuli A., Shoo A., Kim D.R., Ochiai L.R., Favorov M., Clemens J.D., Wilfing H., Deen J.L., Ali S.M. Assessment and comparative analysis of a rapid diagnostic test (Tubex®) for the diagnosis of typhoid fever among hospitalized children in rural Tanzania. BMC Infect. Dis. 2011;11:147. doi: 10.1186/1471-2334-11-147. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 309.Hu L.J., Deng X.H., Brown E.W., Hammack T.S., Ma L.M., Zhang G.D. Evaluation of Roka Atlas Salmonella method for the detection of Salmonella in egg products in comparison with culture method, real-time PCR and isothermal amplification assays. Food Control. 2018;94:123–131. [Google Scholar]
- 310.Xu D.D., Huang X.W., Guo J.H., Ma X. Automatic smartphone-based microfluidic biosensor system at the point of care. Biosens. Bioelectron. 2018;110:78–88. doi: 10.1016/j.bios.2018.03.018. [DOI] [PubMed] [Google Scholar]
- 311.Liu J.J., Geng Z.X., Fan Z.Y., Liu J., Chen H.D. Point-of-care testing based on smartphone: the current state-of-the-art (2017–2018) Biosens. Bioelectron. 2019;132:17–37. doi: 10.1016/j.bios.2019.01.068. [DOI] [PubMed] [Google Scholar]
- 312.Draz M.S., Kochehbyok K.M., Vasari A., Battalapalli D., Sreeram A., Kanakasabapathy M.K., Kallakuri S., Tsibris A., Kuritzkes D.R., Shafiee H. DNA engineered micromotors powered by metal nanoparticles for motion based cellphone diagnostics. Nat. Commun. 2018;9:4282. doi: 10.1038/s41467-018-06727-8. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 313.Roda A., Michelini E., Zangheri M., Di Fusco M., Calabria D., Simoni P. Smartphone-based biosensors: a critical review and perspectives. TrAC Trends Anal. Chem. 2016;79:317–325. [Google Scholar]
- 314.Hu J., Cui X.Y., Gong Y., Xu X.Y., Gao B., Wen T., Lu T.J., Xu F. Portable microfluidic and smartphone-based devices for monitoring of cardiovascular diseases at the point of care. Biotechnol. Adv. 2016;34:305–320. doi: 10.1016/j.biotechadv.2016.02.008. [DOI] [PubMed] [Google Scholar]
- 315.Xu Z.Z., Liu Z.J., Xiao M., Jiang L.L., Yi C.Q. A smartphone-based quantitative point-of-care testing (POCT) system for simultaneous detection of multiple heavy metal ions. Chem. Eng. J. 2020;394 [Google Scholar]
- 316.Kong J.E., Wei Q., Tseng D., Zhang J., Pan E., Lewinski M., Garner O.B., Ozcan A., Di Carlo D. Highly stable and sensitive nucleic acid amplification and cell-phone-based readout. ACS Nano. 2017;11:2934–2943. doi: 10.1021/acsnano.6b08274. [DOI] [PubMed] [Google Scholar]
- 317.Laksanasopin T., Guo T.W., Nayak S., Sridhara A.A., Xie S., Olowookere O.O., Cadinu P., Meng F.X., Chee N.H., Kim J., Chin C.D., Munyazesa E., Mugwaneza P., Rai A.J., Mugisha V., Castro A.R., Steinmiller D., Linder V., Justman J.E., Nsanzimana S., Sia S.K. A smartphone dongle for diagnosis of infectious diseases at the point of care. Sci. Transl. Med. 2015;7 doi: 10.1126/scitranslmed.aaa0056. [DOI] [PubMed] [Google Scholar]
- 318.Wood C.S., Thomas M.R., Budd J., Mashamba-Thompson T.P., Herbst K., Pillay D., Peeling R.W., Johnson A.M., McKendry R.A., Stevens M.M. Taking connected mobile-health diagnostics of infectious diseases to the field. Nature. 2019;566:467–474. doi: 10.1038/s41586-019-0956-2. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 319.Arshavsky-Graham S., Segal E. Lab-on-a-chip devices for point-of-care medical diagnostics. Adv. Biochem. Eng. Biotechnol. 2020 doi: 10.1007/10_2020_127. [DOI] [PubMed] [Google Scholar]
- 320.Fang X.E., Liu Y.Y., Kong J.L., Jiang X.Y. Loop-mediated isothermal amplification integrated on microfluidic chips for point-of-care quantitative detection of pathogens. Anal. Chem. 2010;82:3002–3006. doi: 10.1021/ac1000652. [DOI] [PubMed] [Google Scholar]
- 321.Foudeh A.M., Didar T.F., Veres T., Tabrizian M. Microfluidic designs and techniques using lab-on-a-chip devices for pathogen detection for point-of-care diagnostics. Lab Chip. 2012;12:3249–3266. doi: 10.1039/c2lc40630f. [DOI] [PubMed] [Google Scholar]
- 322.Nasseri B., Soleimani N., Rabiee N., Kalbasi A., Karimi M., Hamblin M.R. Point-of-care microfluidic devices for pathogen detection. Biosens. Bioelectron. 2018;117:112–128. doi: 10.1016/j.bios.2018.05.050. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 323.Chin C.D., Laksanasopin T., Cheung Y.K., Steinmiller D., Linder V., Parsa H., Wang J., Moore H., Rouse R., Umviligihozo G., Karita E., Mwambarangwe L., Braunstein S.L., van de Wijgert J., Sahabo R., Justman J.E., El-Sadr W., Sia S.K. Microfluidics-based diagnostics of infectious diseases in the developing world. Nat. Med. 2011;17:1015–1019. doi: 10.1038/nm.2408. [DOI] [PubMed] [Google Scholar]
- 324.Wang J.W., Karnaushenko D., Medina-Sanchez M., Yin Y., Ma L.B., Schmidt O.G. Three-dimensional microtubular devices for lab-on-a-chip sensing applications. ACS Sens. 2019;4:1476–1496. doi: 10.1021/acssensors.9b00681. [DOI] [PubMed] [Google Scholar]
- 325.Liao Z.R., Zhang Y., Li Y.R., Miao Y.F., Gao S.M., Lin F.K., Deng Y.L., Geng L.N. Microfluidic chip coupled with optical biosensors for simultaneous detection of multiple analytes: a review. Biosens. Bioelectron. 2019;126:697–706. doi: 10.1016/j.bios.2018.11.032. [DOI] [PubMed] [Google Scholar]
- 326.Rutledge G.G., Bohme U., Sanders M., Rrid A.J., Cotton J.A., Maiga-Ascofare O., Djimde A.A., Apinjoh T.O., Amenga-Etego L., Manske M., Barnwell J.W., Renaud F., Ollomo B., Prugnolle F., Anstey N.M., Auburn S., Price R.N., McCarthy J.S., Kwiatkowski D.P., Newbold C.I., Berriman M., Otto T.D. Plasmodium malariae and P. ovale genomes provide insights into malaria parasite evolution. Nature. 2017;542:101–104. doi: 10.1038/nature21038. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 327.Birnbaum J., Scharf S., Schmidt S., Jonscher E., Hoeijmakers W.A.M., Flemming S., Toenhake C.G., Schmitt M., Sabitzki R., Bergmann B., Frohlke U., Mesen-Ramirez P., Soares A.B., Herrmann H., Bartfai R., Spielmann T. A Kelch13-defined endocytosis pathway mediates artemisinin resistance in malaria parasites. Science. 2020;367:51–59. doi: 10.1126/science.aax4735. [DOI] [PubMed] [Google Scholar]
- 328.Ashley E.A., Pyae Phyo A., Woodrow C.J. Malaria. Lancet. 2018;391:1608–1621. doi: 10.1016/S0140-6736(18)30324-6. [DOI] [PubMed] [Google Scholar]
- 329.W.H. Organization, Geneva, World malaria report, 2019.
- 330.Kadimisetty K., Song J.Z., Doto A.M., Hwang Y., Peng J., Mauk M.G., Bushman F.D., Gross R., Jarvis J.N., Liu C.C. Fully 3D printed integrated reactor array for point-of-care molecular diagnostics. Biosens. Bioelectron. 2018;109:156–163. doi: 10.1016/j.bios.2018.03.009. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 331.Malpartida-Cardenas K., Miscourides N., Rodriguez-Manzano J., Yu L.S., Moser N., Baum J., Georgiou P. Quantitative and rapid Plasmodium falciparum malaria diagnosis and artemisinin-resistance detection using a CMOS lab-on-chip platform. Biosens. Bioelectron. 2019;145 doi: 10.1016/j.bios.2019.111678. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 332.Choi J., Cho S.J., Kim Y.T., Shin H. Development of a film-based immunochromatographic microfluidic device for malaria diagnosis. Biomed. Microdevices. 2019;21:86. doi: 10.1007/s10544-019-0431-8. [DOI] [PubMed] [Google Scholar]
- 333.Stemple C.C., Angus S.V., Park T.S., Yoon J.Y. Smartphone-based optofluidic lab-on-a-chip for detecting pathogens from blood. J. Lab. Autom. 2014;19:35–41. doi: 10.1177/2211068213498241. [DOI] [PubMed] [Google Scholar]
- 334.Ghosh S., Aggarwal K., Vinitha T.U., Nguyen T., Han J., Ahn C.H. A new microchannel capillary flow assay (MCFA) platform with lyophilized chemiluminescence reagents for a smartphone-based POCT detecting malaria. Microsyst. Nanoeng. 2020;6:5. doi: 10.1038/s41378-019-0108-8. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 335.Cabibbe A.M., Miotto P., Moure R., Alcaide F., Feuerriegel S., Pozzi G., Nikolayevskyy V., Drobniewski F., Niemann S., Reither K., Cirillo D.M., Consortium T.-R., Consortium T.-C. Lab-on-chip-based platform for fast molecular diagnosis of multidrug-resistant tuberculosis. J. Clin. Microbiol. 2015;53:3876–3880. doi: 10.1128/JCM.01824-15. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 336.Vinuelas-Bayon J., Vitoria M.A., Samper S. Rapid diagnosis of tuberculosis. Detection of drug resistance mechanisms. Enferm. Infecc. Microbiol. Clin. 2017;35:518–526. doi: 10.1016/j.eimc.2017.01.015. [DOI] [PubMed] [Google Scholar]
- 337.Engstrom A. Fighting an old disease with modern tools: characteristics and molecular detection methods of drug-resistant Mycobacterium tuberculosis. Infect. Dis. 2016;48:1–17. doi: 10.3109/23744235.2015.1061205. [DOI] [PubMed] [Google Scholar]
- 338.Muhlig A., Bocklitz T., Labugger I., Dees S., Henk S., Richter E., Andres S., Merker M., Stockel S., Weber K., Cialla-May D., Popp J. LOC-SERS: a promising closed system for the identification of mycobacteria. Anal. Chem. 2016;88:7998–8004. doi: 10.1021/acs.analchem.6b01152. [DOI] [PubMed] [Google Scholar]
- 339.Kukhtin A.C., Sebastian T., Golova J., Perov A., Knickerbocker C., Linger Y., Bueno A., Qu P., Villanueva M., Holmberg R.C., Chandler D.P., Cooney C.G. Lab-on-a-film disposable for genotyping multidrug-resistant Mycobacterium tuberculosis from sputum extracts. Lab Chip. 2019;19:1217–1225. doi: 10.1039/c8lc01404c. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 340.Kukhtin A.V., Norville R., Bueno A., Qu P., Parrish N., Murray M., Chandler D.P., Holmberg R.C., Cooney C.G. A benchtop automated sputum-to-genotype system using a lab-on-a-film assembly for detection of multidrug-resistant Mycobacterium tuberculosis. Anal. Chem. 2020;92:5311–5318. doi: 10.1021/acs.analchem.9b05853. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 341.Hugle M., Dame G., Behrmann O., Rietzel R., Karthe D., Hufert F.T., Urban G.A. A lab-on-a-chip for preconcentration of bacteria and nucleic acid extraction. RSC Adv. 2018;8:20124–20130. doi: 10.1039/c8ra02177e. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 342.Abeyrathne C.D., Huynh D.H., Mcintire T.W., Nguyen T.C., Nasr B., Zantomio D., Chana G., Abbott I., Choong P., Catton M., Skafidas E. Lab on a chip sensor for rapid detection and antibiotic resistance determination of Staphylococcus aureus. Analyst. 2016;141:1922–1929. doi: 10.1039/c5an02301g. [DOI] [PubMed] [Google Scholar]
- 343.Tsougeni K., Papadakis G., Gianneli M., Grammoustianou A., Constantoudis V., Dupuy B., Petrou P.S., Kakabakos S.E., Tserepi A., Gizeli E., Gogolides E. Plasma nanotextured polymeric lab-on-a-chip for highly efficient bacteria capture and lysis. Lab Chip. 2016;16:120–131. doi: 10.1039/c5lc01217a. [DOI] [PubMed] [Google Scholar]
- 344.Ha D.T., Ghosh S., Ahn C.H., Segal D.J., Kim M.S. Pathogen-specific DNA sensing with engineered zinc finger proteins immobilized on a polymer chip. Analyst. 2018;143:4009–4016. doi: 10.1039/c8an00395e. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 345.Yeh E.C., Fu C.C., Hu L., Thakur R., Feng J., Lee L.P. Self-powered integrated microfluidic point-of-care low-cost enabling (SIMPLE) chip. Sci. Adv. 2017;3 doi: 10.1126/sciadv.1501645. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 346.Malvy D., McElroy A.K., de Clerck H., Gunther S., van Griensven J. Ebola virus disease. Lancet. 2019;393:936–948. doi: 10.1016/S0140-6736(18)33132-5. [DOI] [PubMed] [Google Scholar]
- 347.Du K., Cai H., Park M., Wall T.A., Stott M.A., Alfson K.J., Griffiths A., Carrion R., Patterson J.L., Hawkins A.R., Schmidt H., Mathies R.A. Multiplexed efficient on-chip sample preparation and sensitive amplification-free detection of Ebola virus. Biosens. Bioelectron. 2017;91:489–496. doi: 10.1016/j.bios.2016.12.071. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 348.Du K., Park M., Griffiths A., Carrion R., Patterson J., Schmidt H., Mathies R. Microfluidic system for detection of viral RNA in blood using a barcode fluorescence reporter and a photocleavable capture probe. Anal. Chem. 2017;89:12433–12440. doi: 10.1021/acs.analchem.7b03527. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 349.Qin P.W., Park M., Alfson K.J., Tamhankar M., Carrion R., Patterson J.L., Griffiths A., He Q., Yildiz A., Mathies R., Du K. Rapid and fully microfluidic ebola virus detection with CRISPR-Cas13a. ACS Sens. 2019;4:1048–1054. doi: 10.1021/acssensors.9b00239. [DOI] [PubMed] [Google Scholar]
- 350.Nabatiyan A., Baumann M.A., Parpia Z., Kelso D. A lateral flow-based ultra-sensitive p24 HIV assay utilizing fluorescent microparticles. JAIDS J. Acquir. Immune Defic. Syndr. 2010;53:55–61. doi: 10.1097/QAI.0b013e3181c4b9d5. [DOI] [PubMed] [Google Scholar]
- 351.Loynachan C.N., Thomas M.R., Gray E.R., Richards D.A., Kim J., Miller B.S., Brookes J.C., Agarwal S., Chudasama V., McKendry R.A., Stevens M.M. Platinum nanocatalyst amplification: redefining the gold standard for lateral flow immunoassays with ultrabroad dynamic range. ACS Nano. 2018;12:279–288. doi: 10.1021/acsnano.7b06229. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 352.Lee S.H., Kim S.W., Kang J.Y., Ahn C.H. A polymer lab-on-a-chip for reverse transcription (RT)-PCR based point-of-care clinical diagnostics. Lab Chip. 2008;8:2121–2127. doi: 10.1039/b811131f. [DOI] [PubMed] [Google Scholar]
- 353.Chen D.F., Mauk M., Qiu X.B., Liu C.C., Kim J.T., Ramprasad S., Ongagna S., Abrams W.R., Malamud D., Corstjens P.L.A.M., Bau H.H. An integrated, self-contained microfluidic cassette for isolation, amplification, and detection of nucleic acids. Biomed. Microdevices. 2010;12:705–719. doi: 10.1007/s10544-010-9423-4. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 354.Gansen A., Herrick A.M., Dimov I.K., Lee L.P., Chiu D.T. Digital LAMP in a sample self-digitization (SD) chip. Lab Chip. 2012;12:2247–2254. doi: 10.1039/c2lc21247a. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 355.Wang J.S., Kreutz J.E., Thompson A.M., Qin Y.L., Sheen A.M., Wang J.G., Wu L., Xu S.H., Chang M., Raugi D.N., Smith R.A., Gottlieb G.S., Chiu D.T. SD-chip enabled quantitative detection of HIV RNA using digital nucleic acid sequence-based amplification (dNASBA) Lab Chip. 2018;18:3501–3506. doi: 10.1039/c8lc00956b. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 356.Petkovic K., Metcalfe G., Chen H., Gao Y., Best M., Lester D., Zhu Y. Rapid detection of Hendra virus antibodies: an integrated device with nanoparticle assay and chaotic micromixing. Lab Chip. 2017;17:169–177. doi: 10.1039/c6lc01263a. [DOI] [PubMed] [Google Scholar]
- 357.Zhu C.C., Hu A.Z., Cui J.S., Yang K., Zhu X.C., Liu Y., Deng G.Q., Zhu L. A lab-on-a-chip device integrated DNA extraction and solid phase PCR array for the genotyping of high-risk HPV in clinical samples. Micromachines. 2019;10:537. doi: 10.3390/mi10080537. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 358.Ma Y.D., Chen Y.S., Lee G.B. An integrated self-driven microfluidic device for rapid detection of the influenza A (H1N1) virus by reverse transcription loop-mediated isothermal amplification. Sens. Actuators B Chem. 2019;296 (UNSP) [Google Scholar]
- 359.Hugle M., Behrmann O., Raum M., Hufert F.T., Urban G.A., Dame G. A lab-on-a-chip for free-flow electrophoretic preconcentration of viruses and gel electrophoretic DNA extraction. Analyst. 2020;145:2554–2561. doi: 10.1039/c9an02333j. [DOI] [PubMed] [Google Scholar]
- 360.Ming K., Kim J., Biondi M.J., Syed A., Chen K., Lam A., Ostrowski M., Rebbapragada A., Feld J.J., Chan W.C.W. Integrated quantum dot barcode smartphone optical device for wireless multiplexed diagnosis of infected patients. ACS Nano. 2015;9:3060–3074. doi: 10.1021/nn5072792. [DOI] [PubMed] [Google Scholar]
- 361.Madou M., Zoval J., Jia G.Y., Kido H., Kim J., Kim N. Lab on a CD. Annu. Rev. Biomed. Eng. 2006;8:601–628. doi: 10.1146/annurev.bioeng.8.061505.095758. [DOI] [PubMed] [Google Scholar]
- 362.Delgado S.M.T., Kinahan D.J., Sandoval F.S., Julius L.A.N., Kilcawley N.A., Ducree J., Mager D. Fully automated chemiluminescence detection using an electrified-lab-on-a-disc (eLoaD) platform. Lab Chip. 2016;16:4002–4011. doi: 10.1039/c6lc00973e. [DOI] [PubMed] [Google Scholar]
- 363.Loo J.F.C., But G.W.C., Kwok H.C., Lau P.M., Kong S.K., Ho H.P., Shaw P.C. A rapid sample-to-answer analytical detection of genetically modified papaya using loop-mediated isothermal amplification assay on lab-on-a-disc for field use. Food Chem. 2019;274:822–830. doi: 10.1016/j.foodchem.2018.09.049. [DOI] [PubMed] [Google Scholar]
- 364.Sayad A.A., Ibrahim F., Uddin S.M., Pei K.X., Mohktar M.S., Madou M., Thong K.L. A microfluidic lab-on-a-disc integrated loop mediated isothermal amplification for foodborne pathogen detection. Sens. Actuators B Chem. 2016;227:600–609. [Google Scholar]
- 365.Van Nguyen H., Nguyen V.D., Lee E.Y., Seo T.S. Point-of-care genetic analysis for multiplex pathogenic bacteria on a fully integrated centrifugal microdevice with a large-volume sample. Biosens. Bioelectron. 2019;136:132–139. doi: 10.1016/j.bios.2019.04.035. [DOI] [PubMed] [Google Scholar]
- 366.Morelli L., Serioli L., Centorbi F.A., Jendresen C.B., Matteucci M., Ilchenko O., Demarchi D., Nielsen A.T., Zor K., Boisen A. Injection molded lab-on-a-disc platform for screening of genetically modified E. coli using liquid–liquid extraction and surface enhanced Raman scattering. Lab Chip. 2018;18:869–877. doi: 10.1039/c7lc01217a. [DOI] [PubMed] [Google Scholar]
- 367.Michael I.J., Kim T.H., Sunkara V., Cho Y.K. Challenges and opportunities of centrifugal microfluidics for extreme point-of-care testing. Micromachines. 2016;7:32. doi: 10.3390/mi7020032. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 368.Strohmeier O., Keller M., Schwemmer F., Zehnle S., Mark D., von Stetten F., Zengerle R., Paust N. Centrifugal microfluidic platforms: advanced unit operations and applications. Chem. Soc. Rev. 2015;44:6187–6229. doi: 10.1039/c4cs00371c. [DOI] [PubMed] [Google Scholar]
- 369.Sanger K., Zor K., Jendresen C.B., Heiskanen A., Amato L., Nielsen A.T., Boisen A. Lab-on-a-disc platform for screening of genetically modified E. coli cells via cell-free electrochemical detection of p-Coumaric acid. Sens. Actuators B Chem. 2017;253:999–1005. [Google Scholar]
- 370.Bhamla M.S., Benson B., Chai C., Katsikis G., Johri A., Prakash M. Hand-powered ultralow-cost paper centrifuge. Nat. Biomed. Eng. 2017;1:0009. [Google Scholar]
- 371.Choi G., Prince T., Miao J., Cui L.W., Guan W.H. Sample-to-answer palm-sized nucleic acid testing device towards low-cost malaria mass screening. Biosens. Bioelectron. 2018;115:83–90. doi: 10.1016/j.bios.2018.05.019. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 372.Kim T.H., Park J., Kim C.J., Cho Y.K. Fully integrated lab-on-a-disc for nucleic acid analysis of food-borne pathogens. Anal. Chem. 2014;86:3841–3848. doi: 10.1021/ac403971h. [DOI] [PubMed] [Google Scholar]
- 373.Law I.L.G., Loo J.F.C., Kwok H.C., Yeung H.Y., Leung C.C.H., Hui M., Wu S.Y., Chang H.S., Kwan Y.W., Ho H.P., Kong S.K. Automated real-time detection of drug-resistant Mycobacterium tuberculosis on a lab-on-a-disc by recombinase polymerase amplification. Anal. Biochem. 2018;544:98–107. doi: 10.1016/j.ab.2017.12.031. [DOI] [PubMed] [Google Scholar]
- 374.Zhang L., Tian F., Liu C., Feng Q., Ma T.X., Zhao Z.S., Li T.J., Jiang X.Y., Sun J.S. Hand-powered centrifugal microfluidic platform inspired by the spinning top for sample-to-answer diagnostics of nucleic acids. Lab Chip. 2018;18:610–619. doi: 10.1039/c7lc01234a. [DOI] [PubMed] [Google Scholar]
- 375.Loo J.F.C., Kwok H.C., Leung C.C.H., Wu S.Y., Law I.L.G., Cheung Y.K., Cheung Y.Y., Chin M.L., Kwan P., Hui M., Kong S.K., Ho H.P. Sample-to-answer on molecular diagnosis of bacterial infection using integrated lab-on-a-disc. Biosens. Bioelectron. 2017;93:212–219. doi: 10.1016/j.bios.2016.09.001. [DOI] [PubMed] [Google Scholar]
- 376.Oh S.J., Park B.H., Choi G., Seo J.H., Jung J.H., Choi J.S., Kim do H., Seo T.S. Fully automated and colorimetric foodborne pathogen detection on an integrated centrifugal microfluidic device. Lab Chip. 2016;16:1917–1926. doi: 10.1039/c6lc00326e. [DOI] [PubMed] [Google Scholar]
- 377.Michael I., Kim D., Gulenko O., Kumar S., Kumar S., Clara J., Ki D.Y., Park J., Jeong H.Y., Kim T.S., Kwon S., Cho Y.K. A fidget spinner for the point-of-care diagnosis of urinary tract infection. Nat. Biomed. Eng. 2020;4:591–600. doi: 10.1038/s41551-020-0557-2. [DOI] [PubMed] [Google Scholar]
- 378.Liu Q., Zhang X.L., Chen L.P., Yao Y.H., Ke S.R., Zhao W., Yang Z.F., Sui G.D. A sample-to-answer labdisc platform integrated novel membrane-resistance valves for detection of highly pathogenic avian influenza viruses. Sens. Actuators B Chem. 2018;270:371–381. [Google Scholar]
- 379.Li L., Miao B.G., Li Z., Sun Z.M., Peng N.C. Sample-to-answer hepatitis B virus DNA detection from whole blood on a centrifugal microfluidic platform with double rotation axes. ACS Sens. 2019;4:2738–2745. doi: 10.1021/acssensors.9b01270. [DOI] [PubMed] [Google Scholar]
- 380.Martinez A.W., Phillips S.T., Butte M.J., Whitesides G.M. Patterned paper as a platform for inexpensive, low-volume, portable bioassays. Angew. Chem. Int. Ed. 2007;46:1318–1320. doi: 10.1002/anie.200603817. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 381.Tian T., Bi Y.P., Xu X., Zhu Z., Yang C.Y. Integrated paper-based microfluidic devices for point-of-care testing. Anal. Methods. 2018;10:3567–3581. [Google Scholar]
- 382.Chinnadayyala S.R., Park J., Le H.T.N., Santhosh M., Kadam A.N., Cho S. Recent advances in microfluidic paper-based electrochemiluminescence analytical devices for point-of-care testing applications. Biosens. Bioelectron. 2019;126:68–81. doi: 10.1016/j.bios.2018.10.038. [DOI] [PubMed] [Google Scholar]
- 383.Cate D.M., Adkins J.A., Mettakoonpitak J., Henry C.S. Recent developments in paper-based microfluidic devices. Anal. Chem. 2015;87:19–41. doi: 10.1021/ac503968p. [DOI] [PubMed] [Google Scholar]
- 384.Draz M.S., Moazeni M., Venkataramani M., Lakshminarayanan H., Saygili E., Lakshminaraasimulu N.K., Kochehbyoki K.M., Kanakasabapathy M.K., Shabahang S., Vasan A., Bijarchi M.A., Memic A., Shafiee H. Hybrid paper-plastic microchip for flexible and high-performance point-of-care diagnostics. Adv. Funct. Mater. 2018;28 doi: 10.1002/adfm.201707161. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 385.Fu L.M., Wang Y.N. Detection methods and applications of microfluidic paper-based analytical devices. TrAC Trends Anal. Chem. 2018;107:196–211. [Google Scholar]
- 386.Akyazi T., Basabe-Desmonts L., Benito-Lopez F. Review on microfluidic paper-based analytical devices towards commercialisation. Anal. Chim. Acta. 2018;1001:1–17. doi: 10.1016/j.aca.2017.11.010. [DOI] [PubMed] [Google Scholar]
- 387.Yamada K., Shibata H., Suzuki K., Citterio D. Toward practical application of paper-based microfluidics for medical diagnostics: state-of-the-art and challenges. Lab Chip. 2017;17:1206–1249. doi: 10.1039/c6lc01577h. [DOI] [PubMed] [Google Scholar]
- 388.Nie Z.H., Nijhuis C.A., Gong J.L., Chen X., Kumachev A., Martinez A.W., Narovlyansky M., Whitesides G.M. Electrochemical sensing in paper-based microfluidic devices. Lab Chip. 2010;10:477–483. doi: 10.1039/b917150a. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 389.Park C., Kim H.R., Kim S.K., Jeong I.K., Pyun J.C., Park S. Three-dimensional paper-based microfluidic analytical devices integrated with a plasma separation membrane for the detection of biomarkers in whole blood. ACS Appl. Mater. Interfaces. 2019;11:36428–36434. doi: 10.1021/acsami.9b13644. [DOI] [PubMed] [Google Scholar]
- 390.Bender A.T., Borysiak M.D., Levenson A.M., Lillis L., Boyle D.S., Posner J.D. Semiquantitative nucleic acid test with simultaneous isotachophoretic extraction and amplification. Anal. Chem. 2018;90:7221–7229. doi: 10.1021/acs.analchem.8b00185. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 391.Yu L., Shi Z.Z. Microfluidic paper-based analytical devices fabricated by low-cost photolithography and embossing of Parafilm®. Lab Chip. 2015;15:1642–1645. doi: 10.1039/c5lc00044k. [DOI] [PubMed] [Google Scholar]
- 392.Yamada K., Henares T.G., Suzuki K., Citterio D. Paper-based inkjet-printed microfluidic analytical devices. Angew. Chem. Int. Ed. 2015;54:5294–5310. doi: 10.1002/anie.201411508. [DOI] [PubMed] [Google Scholar]
- 393.Hu J., Wang S.Q., Wang L., Li F., Pingguan-Murphy B., Lu T.J., Xu F. Advances in paper-based point-of-care diagnostics. Biosens. Bioelectron. 2014;54:585–597. doi: 10.1016/j.bios.2013.10.075. [DOI] [PubMed] [Google Scholar]
- 394.Xia Y.Y., Si J., Li Z.Y. Fabrication techniques for microfluidic paper-based analytical devices and their applications for biological testing: a review. Biosens. Bioelectron. 2016;77:774–789. doi: 10.1016/j.bios.2015.10.032. [DOI] [PubMed] [Google Scholar]
- 395.Lim H., Jafry A.T., Lee J. Fabrication, flow control, and applications of microfluidic paper-based analytical devices. Molecules. 2019;24:2869. doi: 10.3390/molecules24162869. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 396.Almeida M.I.G.S., Jayawardane B.M., Kolev S.D., McKelvie I.D. Developments of microfluidic paper-based analytical devices (μPADs) for water analysis: a review. Talanta. 2018;177:176–190. doi: 10.1016/j.talanta.2017.08.072. [DOI] [PubMed] [Google Scholar]
- 397.Chakma B., Jain P., Singh N.K., Goswami P. Development of an indicator displacement based detection of malaria targeting HRP-II as biomarker for application in point-of-care settings. Anal. Chem. 2016;88:10316–10321. doi: 10.1021/acs.analchem.6b03315. [DOI] [PubMed] [Google Scholar]
- 398.Xu G.L., Nolder D., Reboud J., Oguike M.C., van Schalkwyk D.A., Sutherland C.J., Cooper J.M. Paper-origami-based multiplexed malaria diagnostics from whole blood. Angew. Chem. Int. Ed. 2016;55:15250–15253. doi: 10.1002/anie.201606060. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 399.Costa M.N., Veigas B., Jacob J.M., Santos D.S., Gomes J., Baptista P.V., Martins R., Inacio J., Fortunato E. A low cost, safe, disposable, rapid and self-sustainable paper-based platform for diagnostic testing: lab-on-paper. Nanotechnology. 2014;25 doi: 10.1088/0957-4484/25/9/094006. [DOI] [PubMed] [Google Scholar]
- 400.Srisa-Art M., Boehle K.E., Geiss B.J., Henry C.S. Highly sensitive detection of Salmonella typhimurium using a colorimetric paper-based analytical device coupled with immunomagnetic separation. Anal. Chem. 2018;90:1035–1043. doi: 10.1021/acs.analchem.7b04628. [DOI] [PubMed] [Google Scholar]
- 401.Ferri M., Ranucci E., Romagnoli P., Giaccone V. Antimicrobial resistance: a global emerging threat to public health systems. Crit. Rev. Food Sci. Nutr. 2017;57:2857–2876. doi: 10.1080/10408398.2015.1077192. [DOI] [PubMed] [Google Scholar]
- 402.Ayukekbong J.A., Ntemgwa M., Atabe A.N. The threat of antimicrobial resistance in developing countries: causes and control strategies. Antimicrob. Resist. Infect. Control. 2017;6:47. doi: 10.1186/s13756-017-0208-x. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 403.He P.J.W., Katis I.N., Kumar A.J.U., Bryant C.A., Keevil C.W., Somani B.K., Mahobia N., Eason R.W., Sones C.L. Laser-patterned paper-based sensors for rapid point-of-care detection and antibiotic-resistance testing of bacterial infections. Biosens. Bioelectron. 2020;152 doi: 10.1016/j.bios.2020.112008. [DOI] [PubMed] [Google Scholar]
- 404.Wu D., Zhang J.H., Xu F.H., Wen X., Li P.F., Zhang X.L., Qiao S., Ge S.X., Xia N.S., Qian S.Z., Qiu X.B. A paper-based microfluidic Dot-ELISA system with smartphone for the detection of influenza A. Microfluid. Nanofluid. 2017;21:43. [Google Scholar]
- 405.Theillet G., Grard G., Galla M., Maisse C., Enguehard M., Cresson M., Dalbon P., Leparc-Goffart I.L., Bedin F. Detection of chikungunya virus‐specific IgM on laser‐cut paper‐based device using pseudo‐particles as capture antigen. J. Med. Virol. 2019;91:899–910. doi: 10.1002/jmv.25420. [DOI] [PubMed] [Google Scholar]
- 406.Zhao C., Liu X.Y. A portable paper-based microfluidic platform for multiplexed electrochemical detection of human immunodeficiency virus and hepatitis C virus antibodies in serum. Biomicrofluidics. 2016;10 doi: 10.1063/1.4945311. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 407.Channon R.B., Yang Y.Y., Feibelman K.M., Geiss B.J., Dandy D.S., Henry C.S. Development of an electrochemical paper-based analytical device for trace detection of virus particles. Anal. Chem. 2018;90:7777–7783. doi: 10.1021/acs.analchem.8b02042. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 408.Srisomwat C., Teengam P., Chuaypen N., Tangkijvanich P., Vilaivan T., Chailapakul O. Pop-up paper electrochemical device for label-free hepatitis B virus DNA detection. Sens. Actuators B Chem. 2020;316 [Google Scholar]
- 409.Teengam P., Siangproh W., Tuantranont A., Vilaivan T., Chailapakul O., Henry C.S. Multiplex paper-based colorimetric DNA sensor using pyrrolidinyl peptide nucleic acid-induced AgNPs aggregation for detecting MERS-CoV, MTB, and HPV oligonucleotides. Anal. Chem. 2017;89:5428–5435. doi: 10.1021/acs.analchem.7b00255. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 410.Tenda K., van Gerven B., Arts R., Hiruta Y., Merkx M., Citterio D. Paper‐based antibody detection devices using bioluminescent BRET‐switching sensor proteins. Angew. Chem. Int. Ed. 2018;57:15369–15373. doi: 10.1002/anie.201808070. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 411.Murdock R.C., Gallegos K.M., Hagen J.A., Kelley-Loughnane N., Weiss A.A., Papautsky I. Development of a point-of-care diagnostic for influenza detection with antiviral treatment effectiveness indication. Lab Chip. 2017;17:332–340. doi: 10.1039/c6lc01074a. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 412.Ye H.H., Xia X.H. Enhancing the sensitivity of colorimetric lateral flow assay (CLFA) through signal amplification techniques. J. Mater. Chem. B. 2018;6:7102–7111. doi: 10.1039/c8tb01603h. [DOI] [PubMed] [Google Scholar]
- 413.Kim H., Chung D.R., Kang M. A new point-of-care test for the diagnosis of infectious diseases based on multiplex lateral flow immunoassays. Analyst. 2019;144:2460–2466. doi: 10.1039/c8an02295j. [DOI] [PubMed] [Google Scholar]
- 414.Li Z., You M., Bai Y., Gong Y., Xu F. Equipment‐free quantitative readout in paper‐based point‐of‐care testing. Small Methods. 2020;4 [Google Scholar]
- 415.Leuvering J.H., Thal P.J., van der Waart M., Schuurs A.H. Sol particle immunoassay (SPIA) J. Immunoass. 1980;1:77–91. doi: 10.1080/01971528008055777. [DOI] [PubMed] [Google Scholar]
- 416.Huang L., Zhang Y.C., Su E.B., Liu Y., Deng Y., Jin L., Chen Z., Li S., Zhao Y.X., He N.Y. Eight biomarkers on a novel strip for early diagnosis of acute myocardial infarction. Nanoscale Adv. 2020;2:1138–1143. doi: 10.1039/c9na00644c. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 417.Yin R., Sun Y.J., Wang K.Y., Feng N.N., Zhang H.J., Xiao M.Y. Development of a PCR-based lateral flow strip assay for the simple, rapid, and accurate detection of pork in meat and meat products. Food Chem. 2020;318 doi: 10.1016/j.foodchem.2020.126541. [DOI] [PubMed] [Google Scholar]
- 418.Jarvis J.N., Tenforde M.W., Lechiile K., Milton T., Boose A., Leeme T.B., Tawe L., Muthoga C., Rukasha I., Mulenga F., Rulaganyang I., Molefi M., Molloy S.F., Ngidi J., Harrison T.S., Govender N.P., Mine M. Evaluation of a novel semiquantitative cryptococcal antigen lateral flow assay in patients with advanced HIV disease. J. Clin. Microbiol. 2020;58 doi: 10.1128/JCM.00441-20. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 419.Lalremruata A., Nguyen T.T., McCall M.B.B., Mombo-Ngoma G., Agnandji S.T., Adegnika A.A., Lell B., Ramharter M., Hoffman S.L., Kremsner P.G., Mordmuller B. Recombinase polymerase amplification and lateral flow assay for ultrasensitive detection of low-density plasmodium falciparum infection from controlled human malaria infection studies and naturally acquired infections. J. Clin. Microbiol. 2020;58 doi: 10.1128/JCM.01879-19. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 420.Lan J.Q., Sun W.M., Chen L.J., Zhou H.L., Fan Y.M., Diao X.P., Wang B.M., Zhao H.W. Simultaneous and rapid detection of carbofuran and 3-hydroxy-carbofuran in water samples and pesticide preparations using lateral-flow immunochromatographic assay. Food Agric. Immunol. 2020;31:165–175. [Google Scholar]
- 421.Soh J.H., Chan H.M., Ying J.Y. Strategies for developing sensitive and specific nanoparticle-based lateral flow assays as point-of-care diagnostic device. Nano Today. 2020;30 [Google Scholar]
- 422.Meng L.Y., Turner A.P.F., Mak W.C. Soft and flexible material-based affinity sensors. Biotechnol. Adv. 2020;39 doi: 10.1016/j.biotechadv.2019.05.004. [DOI] [PubMed] [Google Scholar]
- 423.Kasetsirikul S., Shiddiky M.J.A., Nguyen N.T. Challenges and perspectives in the development of paper-based lateral flow assays. Microfluid. Nanofluid. 2020;24:17. [Google Scholar]
- 424.Pereira D.Y., Chiu R.Y.T., Zhang S.C.L., Wu B.M., Kamei D.T. Single-step, paper-based concentration and detection of a malaria biomarker. Anal. Chim. Acta. 2015;882:83–89. doi: 10.1016/j.aca.2015.04.040. [DOI] [PubMed] [Google Scholar]
- 425.dos Santos G.P., Correa C.C., Kubota L.T. A simple, sensitive and reduced cost paper-based device with low quantity of chemicals for the early diagnosis of Plasmodium falciparum malaria using an enzyme-based colorimetric assay. Sens. Actuators B Chem. 2018;255:2113–2120. [Google Scholar]
- 426.Roskos K., Hickerson A.I., Lu H.W., Ferguson T.M., Shinde D.N., Klaue Y., Niemz A. Simple system for isothermal DNA amplification coupled to lateral flow detection. Plos One. 2013;8 doi: 10.1371/journal.pone.0069355. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 427.Park B.H., Oh S.J., Jung J.H., Choi G., Seo J.H., Kim D.H., Lee E.Y., Seo T.S. An integrated rotary microfluidic system with DNA extraction, loop-mediated isothermal amplification, and lateral flow strip based detection for point-of-care pathogen diagnostics. Biosens. Bioelectron. 2017;91:334–340. doi: 10.1016/j.bios.2016.11.063. [DOI] [PubMed] [Google Scholar]
- 428.Tang R.H., Yang H., Gong Y., You M.L., Liu Z., Choi J.R., Wen T., Qu Z.G., Mei Q.B., Xu F. A fully disposable and integrated paper-based device for nucleic acid extraction, amplification and detection. Lab Chip. 2017;17:1270–1279. doi: 10.1039/c6lc01586g. [DOI] [PubMed] [Google Scholar]
- 429.Cheng N., Song Y., Zeinhom M.M.A., Chang Y.C., Sheng L., Li H.L., Du D., Li L., Zhu M.J., Luo Y.B., Xu W.T., Lin Y.H. Nanozyme-mediated dual immunoassay integrated with smartphone for use in simultaneous detection of pathogens. ACS Appl. Mater. Interfaces. 2017;9:40671–40680. doi: 10.1021/acsami.7b12734. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 430.Bernasconi L., Oberle M., Gisler V., Ottiger C., Fankhauser H., Schuetz P., Fux C.A., Hammerer-Lercher A. Diagnostic performance of a SARS-CoV-2 IgG/IgM lateral flow immunochromatography assay in symptomatic patients presenting to the emergency department. Clin. Chem. Lab. Med. 2020;58:e159–e161. doi: 10.1515/cclm-2020-0635. [DOI] [PubMed] [Google Scholar]
- 431.Ge Y.Y., Wu B., Qi X., Zhao K.C., Guo X.L., Zhu Y.F., Qi Y.H., Shi Z.Y., Zhou M.H., Wang H., Cui L.B. Rapid and sensitive detection of novel avian-origin influenza A (H7N9) virus by reverse transcription loop-mediated isothermal amplification combined with a lateral-flow device. Plos One. 2013;8 doi: 10.1371/journal.pone.0069941. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 432.Waters R.A., Fowler V.L., Armson B., Nelson N., Gloster J., Paton D.J., King D.P. Preliminary validation of direct detection of foot-and-mouth disease virus within clinical samples using reverse transcription loop-mediated isothermal amplification coupled with a simple lateral flow device for detection. Plos One. 2014;9 doi: 10.1371/journal.pone.0105630. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 433.Fu X., Cheng Z., Yu J., Choo P., Chen L., Choo J. A SERS-based lateral flow assay biosensor for highly sensitive detection of HIV-1 DNA. Biosens. Bioelectron. 2016;78:530–537. doi: 10.1016/j.bios.2015.11.099. [DOI] [PubMed] [Google Scholar]
- 434.Gong Y., Zheng Y.M., Jin B.R., You M.L., Wang J.Y., Li X.J., Lin M., Xu F., Li F. A portable and universal upconversion nanoparticle-based lateral flow assay platform for point-of-care testing. Talanta. 2019;201:126–133. doi: 10.1016/j.talanta.2019.03.105. [DOI] [PubMed] [Google Scholar]
- 435.Yeo S.J., Choi K., Cuc B.T., Hong N.N., Bao D.T., Ngoc N.M., Le M.Q., Hang N.L.K., Thach N.C., Mallik S.K., Kim H.S., Chong C.K., Choi H.S., Sung H.W., Yu K., Park H. Smartphone-based fluorescent diagnostic system for highly pathogenic H5N1 viruses. Theranostics. 2016;6:231–242. doi: 10.7150/thno.14023. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 436.Yeo S.J., Kang H., Dao T.D., Cuc B.T., Nguyen A.T.V., Tien T.T.T., Hang N.L.K., Phuong H.V.M., Thanh L.T., Mai L.Q., Rah Y., Yu K., Shin H.J., Chong C.K., Choi H.S., Park H. Development of a smartphone-based rapid dual fluorescent diagnostic system for the simultaneous detection of influenza A and H5 subtype in avian influenza A-infected patients. Theranostics. 2018;8:6132–6148. doi: 10.7150/thno.28027. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 437.Toze S. PCR and the detection of microbial pathogens in water and wastewater. Water Res. 1999;33:3545–3556. [Google Scholar]
- 438.Furst A.L., Francis M.B. Impedance-based detection of bacteria. Chem. Rev. 2019;119:700–726. doi: 10.1021/acs.chemrev.8b00381. [DOI] [PubMed] [Google Scholar]
- 439.Postollec F., Falentin H., Pavan S., Combrisson J., Sohier D. Recent advances in quantitative PCR (qPCR) applications in food microbiology. Food Microbiol. 2011;28:848–861. doi: 10.1016/j.fm.2011.02.008. [DOI] [PubMed] [Google Scholar]
- 440.Park S., Zhang Y., Lin S., Wang T.H., Yang S. Advances in microfluidic PCR for point-of-care infectious disease diagnostics. Biotechnol. Adv. 2011;29:830–839. doi: 10.1016/j.biotechadv.2011.06.017. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 441.Zhang C.S., Xu J.L., Ma W.L., Zheng W.L. PCR microfluidic devices for DNA amplification. Biotechnol. Adv. 2006;24:243–284. doi: 10.1016/j.biotechadv.2005.10.002. [DOI] [PubMed] [Google Scholar]
- 442.Zhang H., Xu Y., Fohlerova Z., Chang H., Iliescu C., Neuzil P. LAMP-on-a-chip: revising microfluidic platforms for loop-mediated DNA amplification. TrAC Trends Anal. Chem. 2019;113:44–53. doi: 10.1016/j.trac.2019.01.015. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 443.Zhang C.S., Xing D. Miniaturized PCR chips for nucleic acid amplification and analysis: latest advances and future trends. Nucleic Acids Res. 2007;35:4223–4237. doi: 10.1093/nar/gkm389. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 444.Broeders S., Huber I., Grohmann L., Berben G., Taverniers I., Mazzara M., Roosens N., Morisset D. Guidelines for validation of qualitative real-time PCR methods. Trends Food Sci. Technol. 2014;37:115–126. [Google Scholar]
- 445.Sposito A., Hoang V., DeVoe D.L. Rapid real-time PCR and high resolution melt analysis in a self-filling thermoplastic chip. Lab Chip. 2016;16:3524–3531. doi: 10.1039/c6lc00711b. [DOI] [PubMed] [Google Scholar]
- 446.Neuzil P., Zhang C.Y., Pipper J., Oh S., Zhuo L. Ultra fast miniaturized real-time PCR: 40 cycles in less than six minutes. Nucleic Acids Res. 2006;34 doi: 10.1093/nar/gkl416. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 447.You M.L., Cao L., Xu F. Plasmon-driven ultrafast photonic PCR. Trends Biochem. Sci. 2020;45:174–175. doi: 10.1016/j.tibs.2019.11.007. [DOI] [PubMed] [Google Scholar]
- 448.Ullerich L., Campbell S., Krieg-Schneider F., Bursgens F., Stehr J. Ultra-fast PCR technologies for point-of-care testing. LaboratoriumsMedizin. 2017;41:239–244. [Google Scholar]
- 449.Zaky W.I., Tomaino F.R., Pilotte N., Laney S.J., Williams S.A. Backpack PCR: a point-of-collection diagnostic platform for the rapid detection of Brugia parasites in mosquitoes. PLoS Negl. Trop. Dis. 2018;12 doi: 10.1371/journal.pntd.0006962. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 450.Ali Z.S., Wang J.H., Mou X.B., Tang Y.J., Li T.T., Liang W.B., Shah M.A.A., Ahmad R., Li Z.Y., He N.Y. Integration of nucleic acid extraction protocol with automated extractor for multiplex viral detection. J. Nanosci. Nanotechnol. 2017;17:862–870. doi: 10.1166/jnn.2017.12613. [DOI] [PubMed] [Google Scholar]
- 451.Mou X.B., Ali Z., Li B., Li T.T., Yi H., Dong H.M., He N.Y., Deng Y., Zeng X. Multiple genotyping based on multiplex PCR and microarray. Chin. Chem. Lett. 2016;27:1661–1665. [Google Scholar]
- 452.Manage D.P., Lauzon J., McMullen L.M., Pilarski L.M. Application of lab-on-a-chip multiplex cassette PCR for the detection of enterohemorrhagic Escherichia coli. BMC Microbiol. 2019;19:93. doi: 10.1186/s12866-019-1463-1. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 453.Salman A., Carney H., Bateson S., Ali Z. Shunting microfluidic PCR device for rapid bacterial detection. Talanta. 2020;207 doi: 10.1016/j.talanta.2019.120303. [DOI] [PubMed] [Google Scholar]
- 454.Mangayarkarasi V., Sneka P., Sujith R., Jayaprakash Ergonomic diagnostic tool based on chip mini RT-PCR for diagnosis of pulmonary and extra pulmonary tuberculosis. J. Pure Appl. Microbiol. 2019;13:1185–1190. [Google Scholar]
- 455.Li Z.Q., Ju R.X., Sekine S., Zhang D.W., Zhuang S.L., Yamaguchi Y. All-in-one microfluidic device for on-site diagnosis of pathogens based on an integrated continuous flow PCR and electrophoresis biochip. Lab Chip. 2019;19:2663–2668. doi: 10.1039/c9lc00305c. [DOI] [PubMed] [Google Scholar]
- 456.Zhou Z.M., Liu D.Y., Zhong R.T., Dai Z.P., Wu D.P., Wang H., Du Y.G., Xia Z.N., Zhang L.P., Mei X.D., Lin B.C. Determination of SARS-coronavirus by a microfluidic chip system. Electrophoresis. 2004;25:3032–3039. doi: 10.1002/elps.200305966. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 457.Ahrberg C.D., Ilic B.R., Manz A., Neuzil P. Handheld real-time PCR device. Lab Chip. 2016;16:586–592. doi: 10.1039/c5lc01415h. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 458.Chen H., Wu Y.Q., Chen Z., Hu Z.L., Fang Y.L., Liao P., Deng Y., He N.Y. Performance evaluation of a novel sample in–answer out (SIAO) system based on magnetic nanoparticles. J. Biomed. Nanotechnol. 2017;13:1619–1630. doi: 10.1166/jbn.2017.2478. [DOI] [PubMed] [Google Scholar]
- 459.Davari M., van Diepeningen A.D., Babai-Ahari A., Arzanlou M., Najafzadeh M.J., van der Lee T.A.J., de Hoog G.S. Rapid identification of Fusarium graminearum species complex using rolling circle amplification (RCA) J. Microbiol. Methods. 2012;89:63–70. doi: 10.1016/j.mimet.2012.01.017. [DOI] [PubMed] [Google Scholar]
- 460.Yan Y.C., Li J., Li W.H., Wang Y., Song W.L., Bi S. DNA flower-encapsulated horseradish peroxidase with enhanced biocatalytic activity synthesized by an isothermal one-pot method based on rolling circle amplification. Nanoscale. 2018;10:22456–22465. doi: 10.1039/c8nr07294a. [DOI] [PubMed] [Google Scholar]
- 461.Yao Q., Wang Y.Q., Wang J., Chen S.M., Liu H.Y., Jiang Z.R., Zhang X.E., Liu S.M., Yuan Q., Zhou X. An ultrasensitive diagnostic biochip based on biomimetic periodic nanostructure-assisted rolling circle amplification. ACS Nano. 2018;12:6777–6783. doi: 10.1021/acsnano.8b01950. [DOI] [PubMed] [Google Scholar]
- 462.Njiru Z.K. Loop-mediated isothermal amplification technology: towards point of care diagnostics. PLoS Negl. Trop. Dis. 2012;6 doi: 10.1371/journal.pntd.0001572. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 463.Quyen T.L., Ngo T.A., Bang D.D., Madsen M., Wolff A. Classification of multiple DNA dyes based on inhibition effects on real-time loop-mediated isothermal amplification (LAMP): prospect for point of care setting. Front. Microbiol. 2019;10:2234. doi: 10.3389/fmicb.2019.02234. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 464.Nzelu C.O., Kato H., Peters N.C. Loop-mediated isothermal amplification (LAMP): an advanced molecular point-of-care technique for the detection of Leishmania infection. PLoS Negl. Trop. Dis. 2019;13 doi: 10.1371/journal.pntd.0007698. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 465.Lillis L., Lehman D.A., Siverson J.B., Weis J., Cantera J., Parker M., Piepenburg O., Overbaugh J., Boyle D.S. Cross-subtype detection of HIV-1 using reverse transcription and recombinase polymerase amplification. J. Virol. Methods. 2016;230:28–35. doi: 10.1016/j.jviromet.2016.01.010. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 466.Wang J.C., Liu L.B., Han Q.A., Wang J.F., Yuan W.Z. An exo probe-based recombinase polymerase amplification assay for the rapid detection of porcine parvovirus. J. Virol. Methods. 2017;248:145–147. doi: 10.1016/j.jviromet.2017.06.011. [DOI] [PubMed] [Google Scholar]
- 467.Zhou W.H., Hu L., Ying L.M., Zhao Z., Chu P.K., Yu X.F. A CRISPR–Cas9-triggered strand displacement amplification method for ultrasensitive DNA detection. Nat. Commun. 2018;9:5012. doi: 10.1038/s41467-018-07324-5. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 468.Dai Y.F., Furst A., Liu C.C. Strand displacement strategies for biosensor applications. Trends Biotechnol. 2019;37:1367–1382. doi: 10.1016/j.tibtech.2019.10.001. [DOI] [PubMed] [Google Scholar]
- 469.Barreda-Garcia S., Miranda-Castro R., de-los-Santos-Alvarez N., Miranda-Ordieres A., Lobo-Castanon M. Comparison of isothermal helicase-dependent amplification and PCR for the detection of Mycobacterium tuberculosis by an electrochemical genomagnetic assay. Anal. Bioanal. Chem. 2016;408:8603–8610. doi: 10.1007/s00216-016-9514-z. [DOI] [PubMed] [Google Scholar]
- 470.Obande G.A., Singh K.K.B. Current and future perspectives on isothermal nucleic acid amplification technologies for diagnosing infections. Infection and Drug Resistance. 2020;13:455–483. doi: 10.2147/IDR.S217571. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 471.Wang Y., Wang Y., Xu J.G., Ye C.Y. Development of multiple cross displacement amplification label-based gold nanoparticles lateral flow biosensor for detection of shigella spp. Front. Microbiol. 2016;7:1834. doi: 10.3389/fmicb.2016.01834. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 472.Wang Y., Wang Y., Zhang L., Xu J.G., Ye C.Y. Visual and multiplex detection of nucleic acid sequence by multiple cross displacement amplification coupled with gold nanoparticle-based lateral flow biosensor. Sens. Actuators B Chem. 2017;241:1283–1293. [Google Scholar]
- 473.Auroux P.A., Koc Y., deMello A., Manz A., Day P.J.R. Miniaturised nucleic acid analysis. Lab Chip. 2004;4:534–546. doi: 10.1039/b408850f. [DOI] [PubMed] [Google Scholar]
- 474.Lee S.H., Park S.M., Kim B.N., Kwon O.S., Rho W.Y., Jun B.H. Emerging ultrafast nucleic acid amplification technologies for next-generation molecular diagnostics. Biosens. Bioelectron. 2019;141 doi: 10.1016/j.bios.2019.111448. [DOI] [PubMed] [Google Scholar]
- 475.Marx V. PCR heads into the field. Nat. Methods. 2015;12:393–397. doi: 10.1038/nmeth.3369. [DOI] [PubMed] [Google Scholar]
- 476.Kwon H.S., Park H.C., Lee K., An S., Oh Y.L., Ahn E.R., Jung J.Y., Lim S.K. Performance of MiniPCRTMmini8, a portable thermal cycler. Anal. Sci. Technol. 2016;29:79–84. [Google Scholar]
- 477.Guevara E.E., Frankel D.C., Ranaivonasy J., Richard A.F., Ratsirarson J., Lawler R.R., Bradley B.J. A simple, economical protocol for DNA extraction and amplification where there is no lab. Conserv. Genet. Resour. 2018;10:119–125. [Google Scholar]
- 478.Gonzalez-Gonzalez E., Mendoza-Ramos J.L., Pedroza S.C., Cuellar-Monterrubio A.A., Marquez-Ipina A.R., Lira-Serhan D., Trujillo-de Santiago G., Alvarez M.M. Validation of use of the miniPCR thermocycler for Ebola and Zika virus detection. Plos One. 2019;14 doi: 10.1371/journal.pone.0215642. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 479.Kaprou G.D., Papadopoulos V., Loukas C.M., Kokkoris G., Tserepi A. Towards PCB-based miniaturized thermocyclers for DNA amplification. Micromachines. 2020;11:258. doi: 10.3390/mi11030258. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 480.Bae N.H., Lim S.Y., Song Y., Jeong S.W., Shin S.Y., Kim Y.T., Lee T.J., Lee K.G., Lee S.J., Oh Y.J., Park Y.M. A disposable and multi-chamber film-based PCR chip for detection of foodborne pathogen. Sensors. 2018;18:3158. doi: 10.3390/s18093158. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 481.Wan L., Gao J., Chen T.L., Dong C., Li H.R., Wen Y.Z., Lun Z.R., Jia Y.W., Mak P.I., Martins R.P. LampPort: a handheld digital microfluidic device for loop-mediated isothermal amplification (LAMP) Biomed. Microdevices. 2019;21:9. doi: 10.1007/s10544-018-0354-9. [DOI] [PubMed] [Google Scholar]
- 482.Liu C.C., Mauk M.G., Hart R., Qiu X.B., Bau H.H. A self-heating cartridge for molecular diagnostics. Lab Chip. 2011;11:2686–2692. doi: 10.1039/c1lc20345b. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 483.Kaur N., Michael J.S., Toley B.J. A modular paper-and-plastic device for tuberculosis nucleic acid amplification testing in limited-resource settings. Sci. Rep. 2019;9:15367. doi: 10.1038/s41598-019-51873-8. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 484.Yu Z.Q., Lyu W.Y., Yu M.C., Wang Q., Qu H.J., Ismagilov R.F., Han X., Lai D.M., Shen F. Self-partitioning SlipChip for slip-induced droplet formation and human papillomavirus viral load quantification with digital LAMP. Biosens. Bioelectron. 2020;155 doi: 10.1016/j.bios.2020.112107. [DOI] [PubMed] [Google Scholar]
- 485.Song J.Z., Mauk M.G., Hackett B.A., Cherry S., Bau H.H., Liu C.C. Instrument-free point-of-care molecular detection of Zika virus. Anal. Chem. 2016;88:7289–7294. doi: 10.1021/acs.analchem.6b01632. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 486.Ma Y.D., Li K.H., Chen Y.H., Lee Y.M., Chou S.T., Lai Y.Y., Huang P.C., Ma H.P., Lee G.B. A sample-to-answer, portable platform for rapid detection of pathogens with a smartphone interface. Lab Chip. 2019;19:3804–3814. doi: 10.1039/c9lc00797k. [DOI] [PubMed] [Google Scholar]
- 487.Chen Z., Yang T., Yang H.W., Li T.T., Nie L.B., Mou X.B., Deng Y., He N.Y., Li Z.Y., Wang L.J., Li S. A portable multi-channel turbidity system for rapid detection of pathogens by loop-mediated isothermal amplification. J. Biomed. Nanotechnol. 2018;14:198–205. doi: 10.1166/jbn.2018.2524. [DOI] [PubMed] [Google Scholar]
- 488.Snodgrass R., Gardner A., Semeere A., Kopparthy V.L., Duru J., Maurer T., Martin J., Cesarmam E., Erickson D. A portable device for nucleic acid quantification powered by sunlight, a flame or electricity. Nat. Biomed. Eng. 2018;2:657–665. doi: 10.1038/s41551-018-0286-y. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 489.Sun F., Ganguli A., Nguyen J., Brisbin R., Shanmugam K., Hirschberg D.L., Wheeler M.B., Bashir R., Nash D.M., Cunningham B.T. Smartphone-based multiplex 30-minute nucleic acid test of live virus from nasal swab extract. Lab Chip. 2020;20:1621–1627. doi: 10.1039/d0lc00304b. [DOI] [PubMed] [Google Scholar]
- 490.Zhang Y., Ge S.G., Yu J.H. Chemical and biochemical analysis on lab-on-a-chip devices fabricated using three-dimensional printing. TrAC Trends Anal. Chem. 2016;85:166–180. [Google Scholar]
- 491.Li L., Wang C., Nie Y., Yao B., Hu H. Nanofabrication enabled lab-on-a-chip technology for the manipulation and detection of bacteria. TrAC Trends Anal. Chem. 2020;127 [Google Scholar]
- 492.Kong L.X., Perebikovsky A., Moebius J., Kulinsky L., Madou M. Lab-on-a-CD: a fully integrated molecular diagnostic system. J. Lab. Autom. 2016;21:323–355. doi: 10.1177/2211068215588456. [DOI] [PubMed] [Google Scholar]
- 493.Tang M.H., Wang G.H., Kong S.K., Ho H.P. A review of biomedical centrifugal microfluidic platforms. Micromachines. 2016;7:26. doi: 10.3390/mi7020026. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 494.Miyazaki C.M., Carthy E., Kinahan D.J. Biosensing on the centrifugal microfluidic lab-on-a-disc platform. Processes. 2020;8:1360. [Google Scholar]
- 495.Smith S., Mager D., Perebikovsky A., Shamloo E., Kinahan D., Mishra R., Delgado S.M.T., Kido H., Saha S., Ducree J., Madou M., Land K., Korvink J.G. CD-based microfluidics for primary care in extreme point-of-care settings. Micromachines. 2016;7:22. doi: 10.3390/mi7020022. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 496.Baharfar M., Rahbar M., Tajik M., Liu G.Z. Engineering strategies for enhancing the performance of electrochemical paper-based analytical devices. Biosens. Bioelectron. 2020;167 doi: 10.1016/j.bios.2020.112506. [DOI] [PubMed] [Google Scholar]
- 497.Kim T.H., Hahn Y.K., Kim M.S. Recent advances of fluid manipulation technologies in microfluidic paper-based analytical devices (μPADs) toward multi-step assays. Micromachines. 2020;11:269. doi: 10.3390/mi11030269. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 498.Marson B.A., Deshmukh S.R., Grindlay D.J.C., Scammell B.E. Alpha-defensin and the Synovasure lateral flow device for the diagnosis of prosthetic joint infection: a systematic review and meta-analysis. Bone Jt. J. 2018;100B:703–711. doi: 10.1302/0301-620X.100B6.BJJ-2017-1563.R1. [DOI] [PubMed] [Google Scholar]
- 499.Carrell C., Kava A., Nguyen M., Menger R., Munshi Z., Call Z., Nussbaum M., Henry C. Beyond the lateral flow assay: a review of paper-based microfluidics. Microelectron. Eng. 2019;206:45–54. [Google Scholar]
- 500.Ahmad F., Hashsham S.A. Miniaturized nucleic acid amplification systems for rapid and point-of-care diagnostics: a review. Anal. Chim. Acta. 2012;733:1–15. doi: 10.1016/j.aca.2012.04.031. [DOI] [PubMed] [Google Scholar]
- 501.Asiello P.J., Baeumner A.J. Miniaturized isothermal nucleic acid amplification, a review. Lab Chip. 2011;11:1420–1430. doi: 10.1039/c0lc00666a. [DOI] [PubMed] [Google Scholar]
- 502.Chang C.C., Chen C.C., Wei S.C., Lu H.H., Liang Y.H., Lin C.W. Diagnostic devices for isothermal nucleic acid amplification. Sensors. 2012;12:8319–8337. doi: 10.3390/s120608319. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 503.Zhao Y.X., Chen F., Li Q., Wang L.H., Fan C.H. Isothermal amplification of nucleic acids. Chem. Rev. 2015;115:12491–12545. doi: 10.1021/acs.chemrev.5b00428. [DOI] [PubMed] [Google Scholar]
- 504.Liu D., Li X.R., Zhou J.K., Liu S.B., Tian T., Song Y.L., Zhu Z., Zhou L.J., Ji T.H., Yang C.Y. A fully integrated distance readout ELISA-chip for point-of-care testing with sample-in-answer-out capability. Biosens. Bioelectron. 2017;96:332–338. doi: 10.1016/j.bios.2017.04.044. [DOI] [PubMed] [Google Scholar]
- 505.Bacheschi D.T., Polsky W., Kobos Z., Yosinski S., Menze L., Chen J., Reed M.A. Overcoming the sensitivity vs. throughput tradeoff in Coulter counters: a novel side counter design. Biosens. Bioelectron. 2020;168 doi: 10.1016/j.bios.2020.112507. [DOI] [PubMed] [Google Scholar]
- 506.Kiilerich-Pedersen K., Dapra J., Cherre S., Rozlosnik N. High sensitivity point-of-care device for direct virus diagnostics. Biosens. Bioelectron. 2013;49:374–379. doi: 10.1016/j.bios.2013.05.046. [DOI] [PubMed] [Google Scholar]
- 507.Park J., Han D.H., Park J.K. Towards practical sample preparation in point-of-care testing: user-friendly microfluidic devices. Lab Chip. 2020;20:1191–1203. doi: 10.1039/d0lc00047g. [DOI] [PubMed] [Google Scholar]
- 508.Govindarajan A.V., Ramachandran S., Vigil G.D., Yager P., Bohringer K.F. A low cost point-of-care viscous sample preparation device for molecular diagnosis in the developing world; an example of microfluidic origami. Lab Chip. 2012;12:174–181. doi: 10.1039/c1lc20622b. [DOI] [PubMed] [Google Scholar]
- 509.Alonso O., Franch N., Canals J., Arias-Alpizar K., de la Serna E., Baldrich E., Dieguez A. An internet of things-based intensity and time-resolved fluorescence reader for point-of-care testing. Biosens. Bioelectron. 2020;154 doi: 10.1016/j.bios.2020.112074. [DOI] [PubMed] [Google Scholar]
- 510.Attia Z.I., Noseworthy P.A., Lopez-Jimenez F., Asirvatham S.J., Deshmukh A.J., Gersh B.J., Carter R.E., Yao X.X., Rabinstein A.A., Erickson B.J., Kapa S., Friedman P.A. An artificial intelligence-enabled ECG algorithm for the identification of patients with atrial fibrillation during sinus rhythm: a retrospective analysis of outcome prediction. Lancet. 2019;394:861–867. doi: 10.1016/S0140-6736(19)31721-0. [DOI] [PubMed] [Google Scholar]
- 511.McRae M.P., Simmons G., Wong J., McDevitt J.T. Programmable bio-nanochip platform: a point-of-care biosensor system with the capacity to learn. Acc. Chem. Res. 2016;49:1359–1368. doi: 10.1021/acs.accounts.6b00112. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 512.Ballard Z.S., Joung H.A., Goncharov A., Liang J.S., Nugroho K., Di Carlo D., Garner O.B., Ozcan A. Deep learning-enabled point-of-care sensing using multiplexed paper-based sensors. Npj Digit. Med. 2020;3:66. doi: 10.1038/s41746-020-0274-y. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 513.Mejia-Salazar J.R., Cruz K.R., Vasques E.M.M., de Oliveira O.N. Microfluidic point-of-care devices: new trends and future prospects for ehealth diagnostics. Sensors. 2020;20:1951. doi: 10.3390/s20071951. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 514.Mayer M., Baeumner A.J. A megatrend challenging analytical chemistry: biosensor and chemosensor concepts ready for the internet of things. Chem. Rev. 2019;119:7996–8027. doi: 10.1021/acs.chemrev.8b00719. [DOI] [PubMed] [Google Scholar]
- 515.Zhu H.L., Podesva P., Liu X.C., Zhang H.Q., Teply T., Xu Y., Chang H.L., Qian A.R., Lei Y.F., Li Y., Niculescu A., Iliescu C., Neuzil P. IoT PCR for pandemic disease detection and its spread monitoring. Sens. Actuators B Chem. 2020;303 doi: 10.1016/j.snb.2019.127098. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 516.Yang T., Gentile M., Shen C.F., Cheng C.M. Combining point-of-care diagnostics and internet of medical things (IoMT) to combat the COVID-19 pandemic. Diagnostics. 2020;10:224. doi: 10.3390/diagnostics10040224. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 517.Ballard Z.S., Shir D., Bhardwaj A., Bazargan S., Sathianathan S., Ozcan A. Computational sensing using low-cost and mobile plasmonic readers designed by machine learning. ACS Nano. 2017;11:2266–2274. doi: 10.1021/acsnano.7b00105. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 518.Gadalla A.A.H., Friberg I.M., Kift-Morgan A., Zhang J.J., Eberl M., Topley N., Weeks I., Cuff S., Wootton M., Gal M., Parekh G., Davis P., Gregory C., Hood K., Hughes K., Butler C., Francis N.A. Identification of clinical and urine biomarkers for uncomplicated urinary tract infection using machine learning algorithms. Sci. Rep. 2019;9:19694. doi: 10.1038/s41598-019-55523-x. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 519.Xie Q., Faust K., Van Ommeren R., Sheikh A., Djuric U., Diamandis P. Deep learning for image analysis: personalizing medicine closer to the point of care. Crit. Rev. Clin. Lab. Sci. 2019;56:61–73. doi: 10.1080/10408363.2018.1536111. [DOI] [PubMed] [Google Scholar]
- 520.Tittl A., John-Herpin A., Leitis A., Arvelo E.R., Altug H. Metasurface‐based molecular biosensing aided by artificial intelligence. Angew. Chem. Int. Ed. 2019;58:14810–14822. doi: 10.1002/anie.201901443. [DOI] [PubMed] [Google Scholar]
- 521.Joung H.A., Ballard Z.S., Wu J., Tseng D.K., Teshome H., Zhang L., Horn E.J., Arnaboldi P.M., Dattwyler R.J., Garner O.B., Di Carlo D., Ozcan A. Point-of-care serodiagnostic test for early-stage lyme disease using a multiplexed paper-based immunoassay and machine learning. ACS Nano. 2020;14:229–240. doi: 10.1021/acsnano.9b08151. [DOI] [PubMed] [Google Scholar]
- 522.Jo Y., Park S., Jung J., Yoon J., Joo H., Kim M.H., Kang S.J., Choi M.C., Lee S.Y., Park Y. Holographic deep learning for rapid optical screening of anthrax spores. Sci. Adv. 2017;3 doi: 10.1126/sciadv.1700606. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 523.Natesan M., Wu S.W., Chen C.I., Jensen S.M.R., Karlovac N., Dyas B.K., Mudanyali O., Ulrich R.G. A smartphone-based rapid telemonitoring system for Ebola and Marburg disease surveillance. ACS Sens. 2019;4:61–68. doi: 10.1021/acssensors.8b00842. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 524.Lopez-Barbosa N., Gamarra J.D., Osma J.F. The future point-of-care detection of disease and its data capture and handling. Anal. Bioanal. Chem. 2016;408:2827–2837. doi: 10.1007/s00216-015-9249-2. [DOI] [PubMed] [Google Scholar]