Published in final edited form as: Phys Med Biol. 2020 Dec 23;65(23):235028. doi: 10.1088/1361-6560/abc5a8

Evaluating attenuation correction strategies in a dedicated, single-gantry breast PET-Tomosynthesis scanner

Srilalan Krishnamoorthy 1, Trevor Vent 2, Bruno Barufaldi 1, Andrew DA Maidment 1, Joel S Karp 1,3, Suleman Surti 1
PMCID: PMC7870546  NIHMSID: NIHMS1642263  PMID: 33113520

Abstract

We are developing a dedicated, combined breast PET-tomosynthesis scanner. Both the PET and digital breast tomosynthesis (DBT) scanners are integrated in a single gantry to provide spatially co-registered 3D PET-tomosynthesis images. The DBT image will be used to identify the breast boundary and breast density to improve the quantitative accuracy of the PET image. This paper explores PET attenuation correction (AC) strategies that can be performed with the combined breast PET-DBT scanner to obtain more accurate, quantitative high-resolution 3D PET images. The PET detector is comprised of a 32 x 32 array of 1.5x1.5x15 mm3 LYSO crystals. The PET scanner utilizes two detector heads separated by either 9 or 11 cm, with each detector head having a 4x2 arrangement of PET detectors. GATE simulations were performed using an anthropomorphic breast phantom with heterogeneous attenuation under clinical DBT-compression. FDG-avid lesions, each 5 mm in diameter with 8:1 uptake, were simulated at four locations within the breast. Simulations were performed with a scan time of 2 min. PET AC was performed using the actual breast simulation model as well as DBT reconstructed volumetric images to derive the breast outline. In addition to using the known breast density as defined by the breast model, we also modeled it as uniform patient-independent soft-tissue, and as a uniform patient-specific material derived from breast tissue composition. Measured absolute lesion uptake was used to evaluate the quantitative accuracy of performing AC using the various strategies. This study demonstrates that AC is necessary to obtain a closer estimate of the true lesion uptake and background activity in the breast. The DBT image dataset assists in measuring lesion uptake with low bias by facilitating accurate breast delineation as well as providing accurate information related to the breast tissue composition. While both the uniform soft-tissue and patient-specific material approaches provides a close estimate to the ground truth, <5% bias can be achieved by using a uniform patient-specific material to define the attenuation map.

1. Introduction

Whole-body (WB) positron emission tomography (PET) imaging is now routinely used in cancer diagnosis and treatment. While time-of-flight (TOF) resolution has steadily improved since its re-emergence 15 years ago, spatial resolution for most WB PET scanners continues to be in the 4–5 mm range (Jakoby et al., 2011, Bettinardi et al., 2011, Miller et al., 2015, Vandendriessche et al., 2019). The only exception is the recently introduced Siemens Biograph Vision that has a reconstructed spatial resolution of 3.6 mm. It therefore remains challenging for WB PET to accurately quantify lesion uptake in smaller lesions (<5 mm) that are prevalent in early-stage breast cancer. The ability of PET to provide quantitative information by accurately measuring radiotracer uptake is equally important for applications involving tumor characterization and treatment monitoring (Dehdashti et al., 2009, Kurland et al., 2012, Eary et al., 2012, Jones et al., 2019). The availability of breast-cancer specific radiotracers (e.g. 18FLT, 18FMISO, 18FES) further motivates the need for a dedicated breast scanner with the ability to accurately measure radiotracer uptake in small tumors.

Over a period of more than two decades, a number of dedicated breast PET scanners have been developed to overcome some of the limitations of WB PET in breast imaging. Besides improving spatial resolution, dedicated breast PET scanners also increase photon sensitivity by using a compact scanner geometry placed close to the breast, thereby reducing the number of photons that would normally get attenuated by the chest or thorax when imaged in a WB-PET scanner. While academic institutions led the effort (Thompson et al., 1994, Thompson et al., 1995, Freifelder and Karp, 1997, Baghaei et al., 2000, Doshi et al., 2001, Freifelder et al., 2001, Abreu et al., 2006, Wang et al., 2006, Peng and Levin, 2010, Ravindranath et al., 2011, Iima et al., 2012, Surti, 2013, Murthy et al., 2000), initial studies demonstrating clinical benefits also led to the commercial development of dedicated breast PET scanners (MacDonald et al., 2009, Moliner et al., 2012, Miyake et al., 2014).

PET performs an important role in cancer detection and its quantitative imaging capability can play an equally important role by providing an accurate measure of radiotracer uptake. Just as in WB PET, accurate data corrections are necessary in dedicated breast PET scanners in order to achieve quantitative images. Random coincidences are generally easy to correct using a delayed window coincidence technique to estimate the random coincidences in the electronics as routinely done in commercial whole-body PET scanners. This reduces bias in the resultant image due to random coincidences but can impact image noise properties. Considering the absence of other organs within the scanner FOV, the fraction of scattered photons is related primarily to the size of the breast being imaged. In previous work investigating scatter in a dedicated breast PET scanner with design similar to the one presented in this paper, Qi et al., (Qi and Huesman, 2002) determined a scatter fraction of 32% that is also consistent with measurements performed by Bowen et al., (Bowen et al., 2009). Bowen et al., (Bowen et al., 2012) used a Monte Carlo based scatter correction technique in their dedicated breast scanner design and report a residual error of 3% after scatter correction. In WB PET scatter correction is routinely performed using the single scatter simulation (Watson et al., 1996, Werner et al., 2006, Watson, 2007) and similar residual errors have been previously reported (Zaidi, 2000, Accorsi et al., 2004). While a dedicated breast PET scanner design eliminates the loss of 511 keV photons that traverse through the chest or thorax, AC is still necessary to reduce bias that can be as high as 50% as shown previously by Bowen et al., (Bowen et al., 2012). Thus, attenuation represents the largest residual source of bias in dedicated breast PET, and careful AC is necessary in order to accurately quantify radiotracer uptake.

Clinical WB PET AC is performed using the CT image to map the volumetric object outline as well as identify the material density distribution within the object. In recent years dedicated breast PET scanners (Bowen et al., 2009, Raylman et al., 2018) have been developed that image uncompressed breasts with patients lying in the prone position. These two systems also have an integrated CT scanner that provides anatomic localization and AC similar to clinical PET/CT systems. Dedicated breast PET scanners without a companion CT, either exclude AC, or, perform AC by assuming the breast to be comprised of uniform soft-tissue, and use the emission image to define the breast boundaries. Many of these dedicated breast PET scanners also image the breast compressed and use planar detectors in partial-ring scanner design. In these systems (Murthy et al., 2000, MacDonald et al., 2009), image artifacts arising from the limited-angle coverage introduce an additional challenge in accurately delineating the breast boundary, which is necessary for accurate AC.

The breast is primarily composed of glandular and adipose tissue and thus heterogenous. Based on the distribution of the two tissue types, the American College of Radiology (ACR) Breast Imaging Reporting and Data System (BI-RADS) (www.acr.org) classification system categorizes breast tissue into four categories – fatty, scattered fibroglandular, heterogeneously dense, and extremely dense. While the linear attenuation coefficients of adipose (0.0914 cm−1) and fibroglandular (0.0974 cm−1) tissues differ by only 6.5% at 511 keV, the fraction by volume and spatial distribution of the two tissue types can vary substantially from patient to patient. Thus, for breast imaging it is important to delineate the volumetric breast outline and also identify breast composition to perform AC accurately.

Our group at the University of Pennsylvania is building a dedicated high spatial resolution, time-of-flight breast PET scanner. In comparison with conventional dual-headed PET scanners that suffer from image artifacts and reduced quantitation accuracy due to limited angle effects, time-of-flight (TOF) information is expected to reduce these effects and improve lesion quantification (Surti and Karp, 2008, Lee et al., 2013). The breast scanner will be combined with a digital breast tomosynthesis (DBT) unit in an integrated gantry providing spatially co-registered 3D PET-DBT images. In addition to providing an anatomical image co-registered with the PET image, the same DBT image can be used to perform accurate AC of the PET images, which is necessary to produce accurate quantitative information.

This paper explores PET AC strategies that can be performed with the combined breast PET-DBT scanner to obtain fully quantitative high-resolution 3D PET images. Monte Carlo simulations of a compressed anthropomorphic breast with heterogeneous attenuation due to varying tissue composition and density, and containing FDG-avid lesions, were performed. PET image reconstruction was performed without AC, and with AC using attenuation maps derived from different estimation strategies. The accuracy of PET quantification was evaluated by comparing the absolute lesion uptake measured from varying AC strategies.

1.1. Anthropomorphic breast phantom

The open-source Virtual Clinical Trial (OpenVCT) framework was used to generate an anthropomorphic breast phantom (Pokrajac et al., 2012, Barufaldi et al., 2018). The anthropomorphic breast phantom is a voxelized 3D model, which consists of predefined tissue types. While a variety of breast tissue components can be defined (Bakic et al., 2018), for this study we only modeled the primary components – adipose tissue, fibroglandular tissue, and skin. Finer structures like Cooper’s ligaments and microcalcifications were not simulated. A breast model with volume and density representative of the average U.S. female population was generated (Yaffe et al., 2009). To simulate DBT acquisition conditions, a 50% physical compression to the breast phantom using a GPU-accelerated 3D mapping technique with pre-calculated meshes (Lago et al., 2013) was also performed. While the 50% compression is slightly larger than normal, these choices were driven by the availability of compression meshes that matched the simulated breast phantom. The simulated breast contained 18.96% glandular tissue, and after compression measured 18.9 × 9.5 × 5.25 cm3. Figure 1(a) shows a volumetric rendering of the breast phantom. To optimize computational resources and PET Monte Carlo simulation time, the OpenVCT phantom was down-sampled from a voxel size of 100 μm3 to 500 μm3 using nearest-neighbor interpolation with a winner-take-all scheme for categorical data (Matlab imresize3). The down-sampled and original breast phantoms were compared, to ensure that down-sampling did not alter the simulated breast phantom density.

Figure 1.

Figure 1.

(a) Volumetric rendering of the breast phantom generated by the VCT pipeline. The different breast tissue types simulated are highlighted, and distribution of the two most abundant breast tissue types - adipose (pink) and glandular (yellow) can be visualized. (b) GATE simulation snapshots of the breast PET scanner along with the outline of the anthropomorphic breast phantom placed within the scanner field of view. (c) and (d) depict the front and top views of the simulation. Each of the two PET detector panels are 20 cm (Y) x 10 cm (Z) and separated by 9 cm (X). To accommodate the compression paddle and breast support, the scanner has a 20 (Y) x 10 (Z) x 8 (X) cm3 FOV. The 50% compressed breast is 18.9 (Y) x 9.5 (Z) x 5.25 cm3 (X).

1.2. DBT imaging simulation

DBT projections of the phantom acquisition were simulated using GPU-enabled x-ray tracing software (Kuo et al., 2011). A custom DBT acquisition geometry was used for this study. Conventional DBT utilizes a linear acquisition geometry, where the x-ray source rotates about the patient’s breast along a one-dimensional arc in the mediolateral (ML) direction. Typically, fifteen projections over an angular range of ±7.5° are acquired. As a screening modality, conventional DBT has shown to benefit the detection of masses, but not calcifications (Rafferty et al., 2013). To improve this, the next-generation (NG) DBT system (Eben et al., 2018) is being developed to investigate alternative acquisition geometries for DBT. NG acquisition geometries are achieved with the introduction of posteroanterior (PA) x-ray source motion and cranio-caudal (CC) detector motion during the tomosynthesis acquisition. While images acquired with NG-DBT acquisition geometries have improved spatial resolution, lower noise and higher contrast, they have also been shown to have superior volume estimation in comparison with conventional DBT (Acciavatti et al., 2019). A bowtie x-ray source motion that also contained fifteen projection, eight of which extend into the PA direction (Vent et al., 2019a, Vent et al., 2019b) was simulated for this study. All DBT image reconstructions were performed using commercial reconstruction software (Briona – version 3.6, Real Time Tomography, Villanova, PA). This software uses the background signal (air) information from each DBT projection image to segment each projection image automatically before performing filtered back-projection reconstruction. DBT reconstruction generates images with ~0.1 mm voxel size and 0.5 mm slice thickness.

1.3. PET simulations and image reconstruction

PET Monte Carlo simulations were performed using the GEANT4 Application for Tomographic Emission (GATE) v.8.0 toolkit (Jan et al., 2004) and the anthropomorphic breast phantom described above. The PET detector consists of a 32 × 32 array of 1.5 × 1.5 × 15 mm3 LYSO crystals (Krishnamoorthy et al., 2014). The PET scanner is comprised of two detector heads, each having a 4×2 arrangement of the PET detectors separated by 9 or 11 cm. The scanner has a 20 (Y) x 10 (Z) x 8 or 10 (X) cm3 field of view (FOV), to accommodate the breast compression paddle and support. Figure 1b1d includes pictures from the GATE simulation depicting the scanner geometry as well as the compressed breast phantom placed within the scanner FOV. The breast phantom was translated along the ‘Z’ direction to ensure that the edge of the scanner was ~3 mm away from the edge of the breast phantom (breast chest wall). The breast support plate and the compression paddle (blue in Fig. 1c) were not simulated as they are expected to have fixed and minimal attenuation. The heterogeneous breast phantom was uniformly filled with an activity concentration of 0.1 μCi/cc to emulate background activity from a clinical 18FDG scan. Four 18FDG-avid 5 mm diameter lesions with 8:1 uptake were simulated individually within the breast phantom (lesion-present data). All four lesions were located within the central axial plane of the breast phantom. A separate set of simulations was also run for the breast phantom without any lesions (lesion-absent data). Table. 1 specifies locations of the four lesions relative to the center of the breast.

Table 1.

Centers for the four 5 mm diameter FDG-avid lesions simulated within the breast. Since attenuation is largest at the center of the breast, all four lesions were located in the central axial plane (Z) of the breast phantom.

Lesion # Location relative to breast center
X (cm) Y (cm)
1 0 0
2 0 1.5
3 5 0
4 5 1.5

The volume of the breast model simulated in this paper is representative of the average U.S. female population. Yaffe et al., (Yaffe et al., 2009) report a mean compressed breast thickness of 5.9 cm, and thus the 9 cm detector separation is expected to accommodate a large fraction of the U.S. population. However, it is not straightforward to model the exact DBT-compression and compressed three-dimensional shape, and it is possible that the scanner will be unable to accommodate a small portion of the population. Since the PET scanner is not primarily intended for breast cancer screening, a locoregional breast scan would nevertheless be helpful for such patients. This might, however, necessitate relaxing the detector separation to accommodate a larger than average breast. Relaxing the detector separation might also be necessary under conditions where the DBT-compression needs to relaxed (i.e. milder) for patient comfort during the PET scan. To explore this possibility, GATE simulations were also performed with an increased detector separation of 11 cm. Increasing detector separation to accommodate a breast with relaxed compression decreases scanner angular coverage and could affect scanner performance (Lee et al., 2013). For ease of comparison, we did not reduce the amount of breast compression in this study. For both detector separations (9 cm and 11 cm), ten independent simulations were performed for a scan time of 2 min each. Coincidence simulation data output from GATE was converted to list mode data, and images with 1 mm3 voxels were reconstructed using a blob-based iterative OSEM algorithm utilizing a Gaussian 300 ps FWHM TOF kernel (Popescu, 2004). As this study is focused on evaluating AC strategies that introduce minimal bias, the simulation only modeled true coincidences, and PET data corrections for scatter and randoms corrections were not performed.

1.4. Attenuation correction strategies

Baseline PET AC was first performed using an ideal AC map which uses the exact knowledge of the modeled breast outline as well as the voxel-by-voxel (heterogeneous) breast density (ideal AC). For our AC strategy, there were then two main areas of evaluation: heterogeneous breast density and breast boundary delineation.

In order to investigate the importance of knowing a voxel-by-voxel breast density that can be challenging if not impossible to obtain, we treated the breast as having a uniform composition, where the tissue was either: (i) a patient-independent soft tissue as is typically performed in breast PET (avg unif AC), or (ii) a weighted-average of all tissue types within the simulated model and hence is patient specific (pat-spec unif AC). In practice, the reconstructed DBT image is a gray-scale image that reflects the material density of underlying breast tissue and can be used to estimate the volumetric breast density necessary to ascertain a patient-specific uniform AC. However, as the DBT image generation process in this work did not model scatter, the DBT-based pat-spec unif AC technique evaluated in this study used a breast density estimate directly from the modeled phantom.

We also evaluated the use of the DBT image to derive the breast outline and repeated AC using the two strategies for defining the uniform breast density (DBT avg unif AC and DBT pat-spec unif AC). To match the PET data processing, the DBT reconstructions were down sampled to 0.5 mm3 voxels using 3D linear interpolation (Matlab Imresize3). As described in section 2.2, the DBT reconstruction not only provides a 3D volumetric reconstruction, it also uses the projection information to discern between breast and air intelligently. All non-air elements from the DBT image were used to delineate the reconstructed breast volume.

For each of the AC strategies, the breast composition (i.e. density) information was translated to corresponding linear attenuation coefficient at 511 keV (ICRU, 1989, Berger and Hubbell, 1987) and used by the PET image reconstruction software for AC. Figure 2 shows a single central cranio-caudal slice of the breast phantom and the material definitions used to generate each of the corresponding attenuation maps. Table 2 summarizes the different AC strategies evaluated, and figure 3 outlines the workflow used to evaluate the different AC strategies.

Figure 2.

Figure 2.

A single central cranio-caudal slice of the breast phantom highlighting the different tissue types (a), and corresponding attenuation map derived from using the phantom outline along with – uniform soft tissue i.e. avg unif AC (b), volumetric weighted-average over the entire breast volume i.e. pat-spec unif AC (c), and the true voxel-by-voxel (i.e. ideal AC) based definition as described in Table. 1 (d).

Table 2.

Summary of the different strategies evaluated for AC. The different techniques were compared to the ideal AC performed using the actual breast model as defined on a voxel-by-voxel basis to perform AC.

Breast delineation Breast Density (material definition for attenuation map)
Breast model (ideal) voxel-by-voxel (ideal)
uniform soft-tissue (avg unif)
uniform volumetric weighted-average (pat-spec unif)
DBT-based (practical) uniform soft-tissue (DBT avg unif)
uniform volumetric weighted-average (DBT pat-spec unif)

Figure 3.

Figure 3.

Flowchart highlighting the different approaches investigated for performing attenuation correction, as well as, the workflow used to evaluate their performance.

1.5. PET image analysis

Absolute lesion activity concentration was used to evaluate the quantitative accuracy of the different AC strategies. Lesion activity concentration was measured by evaluating the mean counts within a 5 mm diameter volume of interest (VOI) drawn in the images with lesions. The VOI center for each dataset was determined by maximizing the signal through iterating over a volume surrounding the true center of each lesion. Background activity concentration was measured in the set of lesion-absent images using 10 mm VOIs placed at the same locations as in the lesion-present images. Uptake was measured for each of the four lesions simulated in the scanner with 9 cm and 11 cm detector separation. PET image uptake calibration for both scanner geometries was performed by imaging a large uniform phantom filling the entire FOV. For each set of lesions and reconstructions, mean and standard deviation of absolute measured activity concentration was calculated across the ten 2 mins scan data sets. Quantitation accuracy (bias) for each lesion was also calculated by comparing with the ideal AC performed using the attenuation map defined on a voxel-by-voxel basis and an outline derived from the model breast.

2. Results

2.1. DBT image reconstruction

Figure 4 shows a single central cranio-caudal slice from the DBT reconstruction of the anthropomorphic breast phantom. In comparison with mammography, DBT provides a volumetric reconstruction of the breast which should translate to better delineation of the breast outline (volume), which is important for accurate AC.

Figure 4.

Figure 4.

A single central cranio-caudal slice from the DBT reconstruction of the anthropomorphic breast phantom.

2.2. PET image reconstructions and attenuation correction evaluation

Figure 5 shows a single central slice of the reconstructed PET image (along the three imaging planes) from using the ideal AC generated from the breast model and with the lesion at the center of the breast. PET detectors were separated by 9 cm and the phantom was imaged for 2 min. Also shown are the reconstructed central transverse slices (figure 5e5g) with the lesion located off-center along either or both X and Y directions. Image analysis was performed after reconstructing for 10 iterations as counts in the lesion VOI had converged. An average activity concentration of 0.41 μCi/cc was measured for the four lesions when using the ideal AC. Figure 6 summarizes the relative bias in uptake calculated for the background and each of the four lesions when compared to image generated with the ideal AC performed using a voxel-by-voxel attenuation and breast outline derived from the model. The results demonstrate that AC is necessary to quantify radiotracer uptake accurately. A large negative bias (~45%) relative to using the ideal AC is observed in absence of AC (Fig. 6, top figure). Using a homogeneous model of soft-tissue (patterned green bar in bottom figure) reduces this bias, albeit overestimating lesion uptake, and using a patient-specific uniform definition (dashed green bar) slightly lowered this overestimation. Similar improvement in bias (red bars) is also observed when using the DBT images to delineate the breast boundary. Quantitative accuracy with <5% bias can be obtained when using the DBT-images for performing AC.

Figure 5.

Figure 5.

PET images were reconstructed with 1 mm3 voxels. Central slices (a-c) of the reconstructed image with the lesion at the center of the breast. The PET scanner detector panels were separated by 9 cm and phantom was imaged for 2 min. Ideal AC material definition and breast delineation from the breast model was used for AC, and average activity concentration of 0.41 μCi/cc was measured over the four lesions. Also shown are the central X-Y slices (d-e) where the lesions are located off-center along either or both directions.

Figure 6.

Figure 6.

Summary of bias measurements with detector separation of 9 cm: The percent bias in uptake calculated relative to the image using the ideal attenuation map, for both background and the four lesions, are summarized for the no-AC (top figure) and for the varying AC strategies evaluated (bottom figure) in this study. As summarized in Table 2, these include AC using the breast model (avg unif – uniform soft-tissue; pat-spec unif – uniform volumetric weighted-average), as well as using the DBT images for breast boundary delineation (DBT avg unif and DBT pat-spec unif). Error bars for each of the measurements are calculated as the standard deviation of bias measured across the ten 2 min scans.

2.3. Effect of increasing detector separation

Figure 7a shows a single central transverse slice of the reconstructed PET image from scanner with PET detectors separated by 11 cm and the phantom imaged for 2 mins. Also included are central transverse slices from reconstructions with the lesion located off-center along either or both X & Y directions (figure 7b7d). In comparison with images having 9 cm detector separation, the decrease in angular coverage from increasing detector separation to 11 cm has no visual impact on image quality. Figure 8 summarizes the relative bias measurements for background and all four lesions imaged with the scanner using 11 cm detector separation. Increasing detector separation in a two-detector design decreases angular coverage and can affect scanner performance. When comparing with the ideal AC data from scanner with 9 cm detector separation, the decrease in angular coverage resulted in ~5% reduction (averaged over the 4 lesions) in the measured absolute lesion uptake. There is also a slight increase in bias when using the DBT image for AC (red bars in Fig. 8 bottom figure). Since the same DBT image from measurements with 9 cm detector separation were used for AC, the small change in bias can be attributed to changes in the limited-angle effect affecting the PET image reconstruction. Overall, consistent with previous measurements in this paper (section 3.2), quantification with low bias can be maintained with using the DBT images for AC even after increasing the detector separation to 11 cm.

Figure 7.

Figure 7.

Central X-Y slices of the reconstructed images from PET scanner with 11 cm separation and scan time of 2 mins. All four source positions listed in Table. 2 are shown from left to right. As described in figure 5, the ideal AC map along with the breast outline from the breast model was used to perform AC.

Figure 8.

Figure 8.

Figure 8.

Summary of the relative bias in uptake measurements for background and the four lesions with detector separation of 11 cm and scan time of 2 min. The percent bias in uptake calculated relative to the image using the ideal (voxel) attenuation map, is summarized for no AC (top figure) and the varying AC techniques (bottom figure) evaluated in this paper. As summarized in Table. 2, the varying AC strategies include using the breast model (avg unif – uniform soft-tissue; pat-spec avg – uniform volumetric weighted-average) as well as using the DBT images for breast boundary delineation (DBT avg unif and DBT pat-spec unif). Error bars for each of the measurements are calculated as the standard deviation of bias measured across the ten 2 min scans.

3. Discussion

This paper evaluates AC strategies that can be performed with a dedicated, combined breast PET-DBT scanner. Some dedicated breast PET scanners without a companion CT have previously used the PET image to define the breast boundary necessary for AC. However, image artifacts accompanying a limited-angle PET scanner design limits the accuracy in breast boundary delineation for these scanners. This accuracy is also impacted when imaging with breast-cancer specific radiotracers, which unlike FDG, have high focal uptake and thus are inappropriate for delineating the breast boundary. Hence, the use of a DBT image acquired in the same gantry as the breast PET can be utilized for AC. In addition, DBT can also aid in determining breast density by estimating breast tissue composition.

This study modeled an anthropomorphic breast with heterogeneous attenuation, and used simulations to evaluate various AC strategies. With a focus on clinical impact, evaluations in this paper were performed for a scan time of 2 min, which could be an upper limit for imaging breasts under reduced (mild) or DBT-compression (full). When compared with the ground truth, the study demonstrates that no AC results in a large underestimation (~45% negative relative bias) of radiotracer uptake for both lesion and background (figure 6). Our results demonstrate that AC can be performed fairly accurately and that for all AC strategies studied here the mean relative bias was < 5% over the two detector separations and all lesions and background. For reference, in a prior simulation study, Kinahan et al., (Kinahan et al., 2013) and MacDonald et al., (MacDonald et al., 2020) evaluated AC correction using a single cranio-caudal view mammogram for the dedicated PET/X scanner (MacDonald et al., 2018) currently under development. The mammogram provides a 2-D image of a breast under compression. With noise-free PET emission data, and using uniform soft-tissue definition along with the breast boundary from the 2-D mammogram, Kinahan et al., report an error of ±6% in measuring peak SUV. Thus, the <5% bias measured in this study shows an improvement likely due to the use of a DBT image to define the breast boundary.

While the patient-specific uniform AC (pat-spec unif AC) provides the lowest mean relative bias (~1% positive bias), the statistical significance of the results summarized in figures 6 & 8 was assessed by performing t-tests. Two groups were compared – (a) significance of using the DBT image to delineate the breast boundary versus the exact breast boundary as defined in the breast mode i.e. comparison between patterned red and patterned green bars, or, between dashed red and dashed green bars for each lesion, and (b) significance of using uniform patient-specific AC versus uniform patient-independent AC i.e. comparison between patterned and dashed red bars, or, between patterned and dashed green bars for each lesion. The t-tests were separately performed for each of the 4-lesion measurements from PET scanner with 9 and 11 cm detector separation. For both detector separations, the t-tests (p-value ranging from 0.4 to 0.9) indicated the difference in residual bias when using the DBT image to mark the breast boundary versus the simulated model boundary was not statistically significant. This indicates that the DBT provides good breast boundary delineation necessary for accurate PET AC. The t-tests also suggested no statistically significant difference (p-values ranging from 0.07 to 0.4 for the ideal and DBT-based measurements) in the bias when comparing the uniform patient-specific AC with uniform patient-independent AC. Thus, using an average uniform material definition and outline from the DBT image might be the most practical approach for performing fairly accurate AC with low bias.

The results shown here used the bowtie NG-DBT acquisition that may have an impact on the quality (accuracy) of the DBT image. However, upon comparison with a conventional DBT acquisition, we did not observe any change in the PET quantitation bias. This suggests that the conventional DBT image is sufficient for performing accurate PET AC. However, the bowtie acquisition tested here is only one of many acquisition geometries supported by the NG-DBT system. Acciavatti et al., (Acciavatti et al., 2018) have demonstrated that NG-DBT improves breast volume estimation and better boundary delineation in the posterior-anterior direction. Thus, we could still expect additional benefits under conditions that have not been evaluated in this paper.

While we did not explicitly investigate techniques to estimate the patient-specific breast density from the DBT images, this feasibility has been demonstrated previously. Pertuz et al., (Pertuz et al., 2016) have demonstrated that volumetric breast density obtained from DBT images were significantly correlated with breast density estimates obtained from full-field digital mammography and MR imaging-based techniques. The ability to automatically estimate breast density from DBT images is also supported by commercially available (Ekpo et al., 2016), and research or open-source software (Kontos et al., 2009, Keller et al., 2012, Qin et al., 2014), some of which have also been validated using patient data (Gastounioti et al., 2018). It is also relevant to remember that the “patient-specific uniform AC” evaluated in this paper does not require information about breast tissue composition on a voxel-by-voxel basis. Thus, the limited depth-resolution of the DBT system is not a limiting factor for accurate PET AC, and findings from this study can be easily extended to clinical measurements with a combined PET-DBT scanner.

The < 5% residual error in lesion uptake measurement after AC is also similar to the 3% residual error expected after scatter correction for events scattered with the FOV (Bowen et al., 2012). In a dedicated breast scanner, scatter from out-of-FOV activity could also affect quantitation. In previous work Qi et al., (Qi and Huesman, 2002) estimated that the scatter contribution from out-of-FOV activity is 7% relative to the within FOV scatter, leading to a total scatter fraction of ~34% (as opposed to 32% with only in-FOV scatter) with the lower energy threshold set at 270 keV. Thus, the overall contribution from out-of-FOV scatter is expected to be small. With a higher setting for the lower energy threshold in the range of 400 to 440 keV in our system due to good energy resolution (~13%) (Krishnamoorthy et al., 2014), we expect the overall scatter fraction to be lower than this 34% value (closer to 18% as shown by Freifelder et al., (Freifelder et al., 2001)) and the residual error from scatter correction to not be higher than that from AC.

For both scanner geometries, true lesion activity concentration is not fully recovered even after AC due to partial volume effects arising due to finite detector spatial resolution and variations in spatial resolution from limited-angle and parallax effects. We have investigated this in prior work (Matej et al., 2016) and have also developed a working reconstruction algorithm making use of a spatially-variant resolution model to reduce any image degrading effects due to limited angular coverage (Gravel et al., 2019). The algorithm not only improves quantitation, but also improves uniformity over the FOV, thereby decreasing the variability in uptake measured over the different lesions in the FOV.

4. Conclusion

In this work we have evaluated PET attenuation correction strategies that can be performed in a combined breast PET-Tomosynthesis scanner. As the primary goal of the scanner is to help with breast cancer treatment, it is important to maintain quantitative accuracy of the scanner. The study demonstrates that attenuation correction is necessary to obtain a more accurate estimation of both lesion uptake and background activity in the breast. The DBT scanner assists in measuring lesion uptake with low bias by facilitating accurate breast delineation as well as providing information related to the breast tissue composition. While both uniform soft-tissue or uniform patient-specific material definition provides a close estimate to the ground truth, <5% bias can be achieved by using a uniform patient-specific material definition to define the attenuation map.

Acknowledgements

The work was supported by the National Institutes of Health Grants No. R01-CA196528 and R01-CA113941, Komen Foundation (IIR 13264610) and Burroughs Wellcome Foundation (IRSA 1016451) grants.

References

  1. ABREU MC, AGUIAR JD, ALMEIDA FG, ALMEIDA P, BENTO P, CARRIÇO B, FERREIRA M, FERREIRA NC, GONÇALVES F & LEONG C 2006. Design and evaluation of the Clear-PEM scanner for positron emission mammography. Nuclear Science, IEEE Transactions on, 53, 71–77. [Google Scholar]
  2. ACCIAVATTI RJ, BARUFALDI B, VENT TL, WILYETO EP & MAIDMENT ADA Personalization of x-ray tube motion in digital breast tomosynthesis using virtual Defrise phantoms. Proceedings Volume 10948, Medical Imaging 2019: Physics of Medical Imaging, 2019. [DOI] [PMC free article] [PubMed] [Google Scholar]
  3. ACCIAVATTI RJ, RODRIGUEZ-RUIZ A, VENT TL, BAKIC PR, REISER I, SECHOPOULOS I & MAIDMENT ADA 2018. Analysis of volume overestimation artifacts in the breast outline segmentation in tomosynthesis. Proc. SPIE 10573, Medical Imaging 2018: Physics of Medical Imaging. [DOI] [PMC free article] [PubMed] [Google Scholar]
  4. ACCORSI R, ADAM LE, WERNER ME & KARP JS 2004. Optimization of a fully 3D single scatter simulation algorithm for 3D PET. Phys Med Biol, 49, 2577–2598. [DOI] [PubMed] [Google Scholar]
  5. BAGHAEI H, WONG WH, URIBE J, LI HD, ZHANG N & WANG Y 2000. Breast cancer imaging studies with a variable field of view PET cameral. Ieee Transactions on Nuclear Science, 47, 1080–1084. [Google Scholar]
  6. BAKIC PR, BARUFALDI B, POKRAJAC DD, WEINSTEIN SP & MAIDMENT ADA Optimized simulation of breast anatomy for virtual clinical trials. The Fourteenth International Workshop on Breast Imaging (IWBI 2018), 2018. Atlanta. [Google Scholar]
  7. BARUFALDI B, HIGGINBOTHAM D, BAKIC PR & MAIDMENT ADA OpenVCT: a GPU-accelerated virtual clinical trial pipeline for mammography and digital breast tomosynthesis. Proc. SPIE 10573, Medical Imaging 2018: Physics of Medical Imaging, 2018. [Google Scholar]
  8. BERGER MJ & HUBBELL JH 1987. XCOM: Photon Cross Sections on a Personal Computer NBSIR 87–3597. NIST, PML, Radiation Physics Division. [Google Scholar]
  9. BETTINARDI V, PRESOTTO L, RAPISARDA E, PICCHIO M, GIANOLLI L & GILARDI MC 2011. Physical Performance of the new hybrid PET/CT Discovery-690. Med. Phys, 38, 5394–5411. [DOI] [PubMed] [Google Scholar]
  10. BOWEN SL, FERRERO A & BADAWI RD 2012. Quantification with a dedicated breast PET/CT scanner. Med. Phys, 39, 2694–2707. [DOI] [PMC free article] [PubMed] [Google Scholar]
  11. BOWEN SL, WU Y, CHAUDHARI AJ, FU L, PACKARD NJ, BURKETT GW, YANG K, LINDFORS KK, SHELTON DK & HAGGE R 2009. Initial characterization of a dedicated breast PET/CT scanner during human imaging. Journal of Nuclear Medicine, 50, 1401–1408. [DOI] [PMC free article] [PubMed] [Google Scholar]
  12. DEHDASHTI F, MORTIMER JE, TRINKAUS K, NAUGHTON MJ, ELLIS M, KATZENELLENBOGEN JA, WELCH MJ & SIEGEL BA 2009. PET-based estradiol challenge as a predictive biomarker of response to endocrine therapy in women with estrogen-receptor-positive breast cancer. Breast Cancer Res Treat, 113, 509–517. [DOI] [PMC free article] [PubMed] [Google Scholar]
  13. DOSHI NK, SILVERMAN RW, SHAO Y & CHERRY SR 2001. maxPET: A dedicated mammary and axillary region PET imaging system for breast cancer. Ieee Transactions on Nuclear Science, 48, 811–815. [Google Scholar]
  14. EARY J, LINK JM, MANKOFF DA, MUZI M, O’SULLIVAN F, FINK J, ROCKHILL J, LINDEN H & KROHN KA Novel PET imaging in the clinic: Selecting patient cohorts and measuring early response Proc. Am. Assoc. Cancer Res, 2012. Chicago, IL. Philadelphia (PA): AACR; 2012. Abstract nr SY42–02. [Google Scholar]
  15. EBEN JE, VENT TL, CHOI CJ, YARRABOTHULA S, CHAI L, NOLAN M, KOBE E, ACCIAVATTI RJ & MAIDMENT ADA Development of a next generation tomosynthesis system. Medical Imaging 2018: Physics of Medical Imaging, 2018. Published in SPIE Proceedings. [Google Scholar]
  16. EKPO EU, MELLO-THOMAS C, RICKARD M, BRENNAN PC & MCENTEE MF 2016. Breast density (BD) assessment with digital breast tomosynthesis (DBT): Agreement between Quantra and 5th edition BI-RADS. Breast, 30, 185–190. [DOI] [PubMed] [Google Scholar]
  17. FREIFELDER R, CARDI C, GRIGORAS I, SAFFER JR & KARP JS First results of a dedicated breast PET imager (B-PET) using NaI(Tl) curve-plate detectors. In: SIEBERT JA, ed. 2001 IEEE Nuclear Science Symposium and Medical Imaging Conference, 2001. San Diego, CA IEEE Nuclear & Plasma Sciences Society. [Google Scholar]
  18. FREIFELDER R & KARP JS 1997. Dedicated PET scanners for breast imaging. Phys. Med. Biol, 42, 2463–2480. [DOI] [PubMed] [Google Scholar]
  19. GASTOUNIOTI A, HSIEH M-K, PANTALONE L, CONANT EF & KONTOS D Volumetric versus area-based density assessment: comparisons using automated quantitative measurements from a large screening cohort Proc. SPIE 10574, Medical Imaging 2018: Image Processing, 105742H, 2018. Houston, Texas, United States. [Google Scholar]
  20. GRAVEL P, SURTI S, KRISHNAMOORTHY S, KARP JS & MATEJ S 2019. Spatially-variant image-based modeling of PSF deformations with application to a limited angle geometry from a dual-panel breast-PET imager. Phys Med Biol, 64, 225015. [DOI] [PMC free article] [PubMed] [Google Scholar]
  21. ICRU 1989. Tissue Substitutes in Radiation Dosimetry and Measurement Report 44 of the International Commission on Radiation Units and Measurements. Bethesda, MD. [Google Scholar]
  22. IIMA M, NAKAMOTO Y, KANAO S, SUGIE T, UENO T, KAWADA M, MIKAMI Y, TOI M & TOGASHI K 2012. Clinical Performance of 2 Dedicated PET Scanners for Breast Imaging: Initial Evaluation. J Nucl Med, 53, 1534–1542. [DOI] [PubMed] [Google Scholar]
  23. JAKOBY BW, BERCIER Y, CONTI M, CASEY ME, BENDRIEM B & TOWNSEND DW 2011. Physical and clinical performance of the mCT time-of-flight PET/CT scanner. Phys. Med. Biol, 56, 2375–2389. [DOI] [PubMed] [Google Scholar]
  24. JAN S, SANTIN G, STRUL D, STAELENS S, ASSIE K, AUTRET D, AVNER S, BARBIER R, BARDIES M & BLOOMFIELD PM 2004. GATE: a simulation toolkit for PET and SPECT. Phys Med Biol, 49, 4543. [DOI] [PMC free article] [PubMed] [Google Scholar]
  25. JONES EF, RAY KM, LI W, CHIEN AJ, MUKHTAR RA, ESSERMAN LJ, FRANC BL, SEO Y, PAMPALONI MH, JOE B & HYLTON NM 2019. Initial experience of dedicated breast PET imaging of ER+ breast cancers using [F-18]fluoroestradiol. NPJ Breast Cancer, 5. [DOI] [PMC free article] [PubMed] [Google Scholar]
  26. KELLER BM, NATHAN DL, WANG Y, ZHENG Y, GEE JC, CONANT EF & KONTOS D 2012. Estimation of breast percent density in raw and processed full field digital mammography images via adaptive fuzzy c-means clustering and support vector machine segmentation. Medical Physics, 39, 4903–17. [DOI] [PMC free article] [PubMed] [Google Scholar]
  27. KINAHAN PE, ZENG C, PIERCE II LA, KANAL KM & MACDONALD L Attenuation Correction Using a Single-View Mammogram for a 3D PET Scanner. The 12th International Meeting on Fully Three-Dimensional Image Reconstruction in Radiology and Nuclear Medicine, 2013. Lake Taho, CA. [Google Scholar]
  28. KONTOS D, BERGER R, BAKIC PR & MAIDMENT ADA Breast tissue classification in digital breast tomosynthesis images using texture features: a feasibility study. Proceedings Volume 7260, Medical Imaging 2009: Computer-Aided Diagnosis, 2009. [Google Scholar]
  29. KRISHNAMOORTHY S, LEGEYT B, WERNER ME, KAUL M, NEWCOMER FM, KARP JS & SURTI S 2014. Design and performance of a high spatial-resolution, time-of-flight PET detector. IEEE Trans. Nucl. Sci, 61, 1092–1098. [DOI] [PMC free article] [PubMed] [Google Scholar]
  30. KUO J, RINGER PA, FALLOWS SG, BAKIC PR, MAIDMENT ADA & NG S Dynamic reconstruction and rendering of 3D tomosynthesis images In: PELC NJ, SAMEI E & NISHIKAWA RM, eds. Proc. SPIE, Medical Imaging 2011: Physics of Medical Imaging, 2011. San Diego, CA. [Google Scholar]
  31. KURLAND BF, GADI VK, SPECHT JM, ALLISON KH, LIVINGSTON RB, RODLER ET, PETERSON LM, SCHUBERT EK, CHAI X & MANKOFF DA 2012. Feasibility study of FDG PET as an indicator of early response to aromatase inhibitors and trastuzumab in a heterogeneous group of breast cancer patients. EJNMMI research, 2, 1–9. [DOI] [PMC free article] [PubMed] [Google Scholar]
  32. LAGO MA, MAIDMENT ADA & BAKIC PR Modelling of mammographic compression of anthropomorphic software breast phantom using FEBio Int’l Symposium on Computer Methods in Biomechanics and Biomedical Engineering, 2013. Salt Lake City, UT. [Google Scholar]
  33. LEE E, WERNER ME, KARP JS & SURTI S 2013. Design optimization of a time-of-flight, breast PET scanner. IEEE Trans. Nucl. Sci, 60, 1645–1652. [DOI] [PMC free article] [PubMed] [Google Scholar]
  34. MACDONALD L, EDWARDS J, LEWELLEN T, HASELEY D, ROGERS J & KINAHAN P 2009. Clinical Imaging Characteristics of the Positron Emission Mammography Camera: PEM Flex Solo II. J Nucl Med, 50, 1666–1675. [DOI] [PMC free article] [PubMed] [Google Scholar]
  35. MACDONALD LR, HUNTER WCJ, ZENG C, PIERCE LA, DOLINSKI S, DEWITT D, MIYAOKA R & KINAHAN PE 2018. The PET/X dedicated breast-PET scanner for optimizing cancer therapy. Proc. SPIE 10718, 2018. [Google Scholar]
  36. MACDONALD LR, LO JY, STURGEON GM, ZENG C, HARRISON RL & KINAHAN PE 2020. Impact of Using Uniform Attenuation Coefficients for Heterogeneously Dense Breasts in a Dedicated Breast PET/X-Ray Scanner. IEEE Transactions on Radiation and Plasma Medical Sciences, 4, 585–593. [DOI] [PMC free article] [PubMed] [Google Scholar]
  37. MATEJ S, LI Y, PANETTA J, KARP JS & SURTI S 2016. Image-based modeling of PSF deformation with application to limited angle PET data IEEE Transactions on Nuclear Science, 63, 2599–2606. [DOI] [PMC free article] [PubMed] [Google Scholar]
  38. MILLER M, ZHANG J, BINZEL K, GRIESMER J, LAURENCE T, NARAYANAN M, NATARAJAMANI D, WANG S & KNOPP M 2015. Characterization of th Vereos Digital Photon Counting PET system. J Nucl Med, 56, 434S. [Google Scholar]
  39. MIYAKE KK, MATSUMOTO K, INOUE M, NAKAMOTO Y, KANAO S, OISHI T, KAWASE S, KITAMURA K, YAMAKAWA Y, AKAZAWA A, KOBAYASHI T, OHI J & TOGASHI K 2014. Performance Evaluation of a New Dedicated Breast PET Scanner Using NEMA NU4–2008 Standards. Journal of Nuclear Medicine, 55, 1198–1203. [DOI] [PubMed] [Google Scholar]
  40. MOLINER L, GONZAzLEZ AJ, SORIANO A, SAzNCHEZ F, CORRECHER C, ORERO A, CARLES M, VIDAL LF, BARBERAz J & CABALLERO L 2012. Design and evaluation of the MAMMI dedicated breast PET. Med Phys, 39, 5393. [DOI] [PubMed] [Google Scholar]
  41. MURTHY K, AZNAR M, THOMPSON CJ, LOUTFI A, LISBONA R & GAGNON JH 2000. Results of preliminary clinical trials of the positron emission mammography system PEM-I: a dedicated breast imaging system producing glucose metabolic images using FDG. Journal of Nuclear Medicine, 41, 1851–1858. [PubMed] [Google Scholar]
  42. PENG H & LEVIN CS 2010. Design study of a high-resolution breast-dedicated PET system built from cadmium zinc telluride detectors. Phys Med Biol, 55, 2761–2788. [DOI] [PMC free article] [PubMed] [Google Scholar]
  43. PERTUZ S, MCDONALD ES, WEINSTEIN SP, CONANT EF & KONTOS D 2016. Fully Automated Quantitative Estimation of Volumetric Breast Density from Digital Breast Tomosynthesis Images: Preliminary Results and Comparison with Digital Mammography and MR Imaging. Radiology, 279, 65–74. [DOI] [PMC free article] [PubMed] [Google Scholar]
  44. POKRAJAC DD, MAIDMENT AD & BAKIC PR 2012. Optimized generation of high-resolution breast anthropomorphic software phantoms. Medical Physics, 39, 2290–302. [DOI] [PMC free article] [PubMed] [Google Scholar]
  45. POPESCU LM Iterative image reconstruction using geometrically ordered subsets with list-mode data. IEEE Nuclear Science Symposium and Medical Imaging Conference, 2004. Rome, Italy 3536–3540. [Google Scholar]
  46. QI J & HUESMAN RH 2002. Scatter correction for positron emission mammography. Phys. Med. Biol, 47, 2759–2771. [DOI] [PubMed] [Google Scholar]
  47. QIN X, LU G, SECHOPOULOS I & FEI B Breast tissue classification in digital tomosynthesis images based on global gradient minimization and texture features. Proceedings Volume 9034, Medical Imaging 2014: Image Processing, 2014. [DOI] [PMC free article] [PubMed] [Google Scholar]
  48. RAFFERTY EA, PARK JM, PHILPOTTS LE, POPLACK SP, SUMKIN JH, HALPERN EF & NIKLASON LT 2013. Assessing Radiologist Performance Using Combined Digital Mammography and Breast Tomosynthesis Compared with Digital Mammography Alone: Results of a Multicenter, Multireader Trial. Radiology, 266, 104–113. [DOI] [PMC free article] [PubMed] [Google Scholar]
  49. RAVINDRANATH B, JUNNARKAR S, PURSCHKE M, STOLL S, HONG X, HUANG P-J, MARAMRAJU SH, VASKA P, WOODY C & SCHLYER D 2011. Results from a simultaneous PET-MRI breast scanner. J. Nucl. Med. Meeting Abstract, 52, 432. [Google Scholar]
  50. RAYLMAN RR, KAMPEN WV, STOLIN AV, GONG W, JALIPARTHI G, MARTONE PF, SMITH MF, SARMENT D, CLINTHORNE NH & PERNA M 2018. A dedicated breast-PET/CT scanner: Evaluation of basic performance characteristics. Medical Physics, 45, 1603–13. [DOI] [PMC free article] [PubMed] [Google Scholar]
  51. SURTI S 2013. Radionuclide Methods and Instrumentation for Breast Cancer Detection and Diagnosis. Sem. Nucl. Med, 43, 271–280. [DOI] [PMC free article] [PubMed] [Google Scholar]
  52. SURTI S & KARP JS 2008. Design considerations for a limited angle, dedicated breast, TOF PET scanner. Phys. Med. Biol, 53, 2911–2921. [DOI] [PMC free article] [PubMed] [Google Scholar]
  53. THOMPSON CJ, MURTHY K, PICARD Y, WEINBERG IN & MAKO R 1995. Positron Emission Mammography (PEM) - a Promising Technique for Detecting Breast-Cancer. IEEE Trans. Nucl. Sci, 42, 1012–1017. [Google Scholar]
  54. THOMPSON CJ, MURTHY K, WEINBERG IN & MAKO F 1994. Feasibility Study for Positron Emission Mammography. Med. Phys, 21, 529–538. [DOI] [PubMed] [Google Scholar]
  55. VANDENDRIESSCHE D, URIBE J, BERTIN H & GEETER FH 2019. Performance characteristics of silicon photomultiplier based 15-cm AFOV TOF PET/CT. EJNMMI Physics, 6. [DOI] [PMC free article] [PubMed] [Google Scholar]
  56. VENT TL, BARUFALDI B & MAIDMENT ADA Simulation and experimental validation of high-resolution test objects for evaluating a next-generation digital breast tomosynthesis prototype. Proceedings Volume 10948, Medical Imaging 2019: Physics of Medical Imaging, 2019a. [Google Scholar]
  57. VENT TL, LEPORE B & MAIDMENT ADA Evaluating the imaging performance of a next-generation digital breast tomosynthesis prototype. Proceedings Volume 10948, Medical Imaging 2019: Physics of Medical Imaging, 2019b. [Google Scholar]
  58. WANG GC, HUBER JS, MOSES WW, QI J & CHOONG WS 2006. Characterization of the LBNL PEM camera. Nuclear Science, IEEE Transactions on, 53, 1129–1135. [Google Scholar]
  59. WATSON CC 2007. Extension of Single Scatter Simulation to Scatter Correction of Time-of-Flight PET. IEEE Transactions on Nuclear Science, 54, 1679–1686. [Google Scholar]
  60. WATSON CC, NEWPORT D & CASEY ME 1996. A Single Scatter Simulation Technique for Scatter Correction in 3D PET. Three-Dimensional Image Reconstruction in Radiology and Nuclear Medicine, 4, 255–268. [Google Scholar]
  61. WERNER ME, SURTI S & KARP JS Implementation and Evaluation of a 3D PET Single Scatter Simulation with TOF Modeling. IEEE Nuclear Science Symposium and Medical Imaging Conference, 2006. San Diego, CA 1768–1773. [Google Scholar]
  62. YAFFE MJ, BOONE JM, PACKARD N, ALONZO-PROULX O, HUANG SY, PERESSOTTI CL, AL-MAYAH A & BROCK K 2009. The myth of the 50–50 breast. Medical Physics, 36, 5437–43. [DOI] [PMC free article] [PubMed] [Google Scholar]
  63. ZAIDI H 2000. Comparitive evaluation of scatter correction techniques in 3D positron emission tomography. Eur J Nucl Med, 27, 1813–1826. [DOI] [PubMed] [Google Scholar]

RESOURCES