Performance of SARS-CoV-2 serology tests: Are they good enough?

Isabelle Piec 1,*, Emma English 2, Mary Annette Thomas 3, Samir Dervisevic 4, William D Fraser 1,5, William Garry John 2,5
Editor: Muhammad Adrish6
PMCID: PMC7888604  PMID: 33596236

Abstract

In the emergency of the SARS-CoV-2 pandemic, great efforts were made to quickly provide serology testing to the medical community however, these methods have been introduced into clinical practice without the complete validation usually required by the regulatory organizations. SARS-CoV-2 patient samples (n = 43) were analyzed alongside pre-pandemic control specimen (n = 50), confirmed respiratory infections (n = 50), inflammatory polyarthritis (n = 22) and positive for thyroid stimulating immunoglobulin (n = 30). Imprecision, diagnostic sensitivity and specificity and concordance were evaluated on IgG serologic assays from EuroImmun, Epitope Diagnostics (EDI), Abbott Diagnostics and DiaSorin and a rapid IgG/IgM test from Healgen. EDI and EuroImmun imprecision was 0.02–14.0% CV. Abbott and DiaSorin imprecision (CV) ranged from 5.2%–8.1% and 8.2%–9.6% respectively. Diagnostic sensitivity of the assays was 100% (CI: 80–100%) for Abbott, EDI and EuroImmun and 95% (CI: 73–100%) for DiaSorin at ≥14 days post PCR. Only the Abbott assay had a diagnostic specificity of 100% (CI: 91–100%). EuroImmun cross-reacted in 3 non-SARS-CoV-2 respiratory infections and 2 controls. The DiaSorin displayed more false negative results and cross-reacted in six cases across all conditions tested. EDI had one cross-reactive sample. The Healgen rapid test showed excellent sensitivity and specificity. Overall, concordance of the assays ranged from 76.1% to 97.9%. Serological tests for SARS-CoV-2 showed good analytical performance. The head-to-head analysis of samples revealed differences in results that may be linked to the use of nucleocapsid or spike proteins. The point of care device tested demonstrated adequate performance for antibody detection.

Introduction

The scientific community has had to rapidly develop and manufacture tests for the new SARS-CoV-2 pandemic at unprecedented speed, taking three months to develop assays that would ordinarily take three years. Serology testing, that can identify those who have previously been exposed to the SARS-CoV-2 virus and have mounted an immune response, has been hailed as key to managing the pandemic however controversy remains over both the accuracy and utility of serology testing in disease management. Structural proteins, including the spike (essential for viral infection) and the nucleocapsid (important for viral RNA transcription), are both potential targets for early detection of infection and known to elicit an immune response in the host [1] with antibodies detectable within 20 days of disease onset [24].

Systematic reviews [5, 6] challenged the diagnostic accuracy of serological tests, particularly when using lateral flow immunoassays (LFIAs). Public Health England (PHE) showed only the Siemens and the Roche Diagnostics assays met the minimum UK Medicines and Healthcare products Regulatory Agency Target Product Profile criteria for sensitivity [7] after the threshold of positivity was adjusted to 0.128. Assays from DiaSorin and Abbott Diagnostics [8] also provided acceptable diagnostic results. These evaluations did not address cross-reactivity. To our knowledge little has been done regarding interference from antibodies produced during other viral infection and autoimmune disorders. Additionally, with the focus on diagnostic sensitivity and specificity, little has been done to evaluate the analytical accuracy, which if poor, has the potential to negate all of these study findings. Indeed in the editorial, Duong and colleagues clearly states that there is a need for critical independent evaluations of these tests, using the same specimen panels [9].

This study provides a head-to-head evaluation of the diagnostic and analytical performance of four commercially available IgG based serology assays for SARS-CoV-2 and a diagnostic accuracy study of one point of care LFIA.

Materials and methods

Specimen collection and storage

Patients were not involved in any part of the work. All samples were from archived specimens and were fully anonymized before we accessed them. Therefore, our study is in accordance with the blanket Ethical standards of University of East Anglia on de-identified samples for method development. Moreover, using the UK NHS Research Ethics Committee decision toolkit (http://www.hra-decisiontools.org.uk/ethics/) we confirmed that separate ethical review is not required for this study which is in concordance with the Helsinki Declaration.

All serum samples were collected between April and June 2020, anonymized, aliquoted and stored at –80°C until analyzed. SARS-CoV-2 PCR-positive patients (AusDiagnostics platform, Chesham, UK) were of both genders, age range 66 to 93 and hospitalized at the Norfolk and Norwich University Hospital (NNUH) or Queen Elizabeth Hospital in King Lynn (QEH). Samples were taken 8–44 days after testing positive for SARS-CoV-2. Negative control samples were collected in 2018 from patients with no history of infection or immune disorder. Pre-pandemic samples from patients who had a range of confirmed respiratory infections (including Influenza A, B and seasonal coronaviruses [Table 1]), samples collected from patients with inflammatory polyarthritis positive for anti-cyclic citrullinated peptide antibodies (anti-CCP) along with samples positive for thyroid stimulating immunoglobulin (TSI) were used to test the non-specific binding of non-SARS-CoV-2 antibodies. These groups of samples are referred to as N (negative control), CR (cross-reactivity), RA (Rheumatoid Arthritis), TSI (patients with thyroid stimulating immunoglobulin) and P (SARS-CoV-2 Positive). A total of 195 individual serum samples (43 P, 50 N, 50 CR, 22 RA and 30 TSI) were analyzed for SARS-CoV-2 IgG antibodies. For a subset of patients, samples were available for a series of time-points thus allowing for a time course analysis (43 patients, 142 samples).

Table 1. Respiratory infections tested for cross reactivity in the SARS-CoV-2 IgG immunoassays.

Infection No patients
Epstein-Barr virus 8
Influenza A virus 8
Respiratory syncytial virus 7
Seasonal Coronaviruses 7
Borrelia burgerdorferii 4
Cytomegalovirus 3
Varicellazoster virus 3
Bordella Pertussis 2
Hepatitis B 2
Human immunodeficiency virus 2
Adenovirus 1
Mycoplasma 1
ParaInfluenza 1
Rhinovirus 1

Study design

SARS-CoV-2 IgG immunoassays were from 1) Epitope Diagnostics Inc. (EDI, San Diego, CA, USA) performed using the Agility ELISA automate (Dynex Technologies, Chantilly, VA, USA), 2) EuroImmun UK ITC (UK) performed manually, 3) Abbott Diagnostics (Maidenhead, UK) on the Alinity i analyzer and 4) DiaSorin (London, UK) on the Liaison XL analyzer. A subset of samples was also tested using the point of care testing (POCT) device SARS-CoV-2 IgG/IgM rapid test from Healgen (Houston, TX, USA). Due to a limited number of cassettes available, 49 samples from 27 P were analyzed along with 3 N, 8 CR, 4 RA and 4 TSI. Cross-reactive and negative samples were primarily chosen from patient samples proven positive for seasonal coronaviruses and influenza A or a false positive result in one or more of the immunoassays. We focused on the IgG results in order to compare with the immunoassays.

Assays were performed by trained biomedical scientists using manufacturer’s instructions. The SARS-CoV-2 Abbott assay was performed in the clinical biochemistry department at NNUH and the other SARS-CoV-2 assays were performed at the University of East Anglia. All other non- SARS-CoV-2 related tests were performed at the NNUH virology department. DiaSorin SARS-Cov-2 is a quantitative assay and antibody concentrations are expressed in AU/mL. The Abbott, EDI and EuroImmun are qualitative assays for which the result is calculated using the ratio of the sample optical density (OD) against the negative or calibrator control (S1 Table). EuroImmun and DiaSorin assays detect antibodies to, respectively, recombinant S1 and S1/S2 domains of the SARS-CoV-2 spike protein while both the EDI and Abbott assay detect antibodies to the nucleocapsid. The POCT from Healgen is a solid phase lateral flow immunochromatographic assay (LFIA) for detection of SARS-CoV-2 of IgG and IgM, antigen not specified.

Imprecision

As the results are expressed with a values correlating with the amount of antibody detectable, imprecision was assessed using a Clinical and Laboratory Standards Institute (CLSI) EP-15 based protocol on the automated clinical laboratory analyzers protocol [10]. Positive and negative patient pools and/or controls of different concentrations were prepared and frozen as aliquots and assayed as 5 replicates per day on 5 different days. For the plate-based assays, inter- and intra-assay CVs were calculated. Intra-assay was determined using the CV of the optical density (OD) of duplicated samples. Inter-assay was determined using the CV obtained from the sample pool and the kit positive control across the plates.

Statistics

Using IBM SPSS Statistics 25.0.0.1, Mann-Whitney and Cohen’s Kappa tests were used to compare OD results between groups and to determine the concordance between the assays, respectively. Analysis of EP15 was performed using EP evaluator. Variation was estimated on calculated values (R) or response (Relative Light Unit, RLU) as intra and inter-assay coefficient of variation (CV). Graphical representations were conducted with GraphPad Prism version 8.0 (GraphPad Software, Inc., USA). Throughout the tables, figures, and legends, the following terminology is used to show statistical significance: *P<0.05; **P<0.01 and ***P<0.001.

Results

Imprecision

Abbott

EP15 and was performed on two Alinity analyzers (Table 2). Overall, negative pool imprecision was CV = 8.1% and 6.8% on equipment 1 and 2 respectively. Positive pool imprecisions were CV = 2.3% and 1.1% respectively.

Table 2. EP15 analysis on two Abbott Alinity, DiaSorin Liaison XL and ELISAs imprecision tests.
Sample n Mean Intra-assay imprecision Inter-assay imprecision
SD %CV SD %CV
ABBOTT Alinity 1 (Neg) 25 0.136 (R) 0.011 8.1 0.011 8.1
Alinity 1 (Pos) 25 7.254 (R) 0.167 2.3 0.170 2.3
Alinity 2 (Neg) 25 0.143 (R) 0.007 5.2 0.010 6.8
Alinity 2 (Pos) 25 7.242 (R) 0.081 1.1 0.082 1.1
DIASORIN Kit Negative control 20 2457 (RLU) 1730 70.4 2860 116.4
Level 1 (Neg pool) 25 6945 (RLU) 4003 57.6 5887 84.8
Kit Positive control 25 58662 (RLU) 4815 8.2 5608 9.6
Level 2 (Pool 1) 25 83236 (RLU) 11128 13.4 11128 13.4
Level 3 (Pool 2) 25 410600 (RLU) 56802 13.8 56802 13.8
Level 4 (pool 3) 25 557660 (RLU) 55667 10.0 58092 10.4
EDI Kit Negative control 27 0.074 (OD) - - 0.014 10.9
Kit positive control 9 0.482 (OD) - - 0.068 14.2
Duplicate samples 308 - 3.8 3.3 - -
EURO-IMMUN Kit Negative control 3 0.074 (OD) - - 0.003 3.7
Kit positive control 3 1.169 (OD) - - 0.15 2.9
Calibrator 3 0.277 (OD) - - 0.027 9.5
Duplicate samples 44 - 6.7 6.1 - -

For the DiaSorin, negative samples (QC or pools) results were typically below the limit of detection of 3.8 AU/mL and variation was estimated on the response in relative light units (RLU).

DiaSorin

EP15 imprecision was estimated based on response intensity (RLU). Positive control imprecision was between 8.2% and 13.8% (Table 2). The negative quality control material results were consistently below the lower limit of detection of 3.8AU/mL and the negative pool concentration was consistently below 10AU/mL, the resulting calculated imprecision was therefore expectedly elevated.

EDI and EuroImmun

Intra-assay imprecision on duplicate samples (Table 2) was on average CV = 3.3±3.8% and 6.1±6.7% respectively. Inter-assay imprecision of EDI was CV = 14.2% for the kit positive pool and 16.5% for the negative pool. Baseline OD varied between the plates increasing the inter-assay variations, however, the ratio positive/cut-off was on average 1.43 ±0.16 (CV = 11.1%). Inter-assay of EuroImmun was evaluated using the positive kit QC, the calibrator and the negative kit control. Coefficient of variation were CV = 12.9%, 9.5% and 3.7% respectively.

Specificity and sensitivity

A total of 43 individual P was analyzed for SARS-CoV-2 IgG antibodies. Of these, twenty had samples taken at least 14 days after a positive PCR result (P≥14) and 23 were taken prior (P<14). All P≥14 had detectable antibodies in the EDI, EuroImmun, Abbott and Healgen assays. However, one sample returned a negative result using the DiaSorin assay. These results suggest a true positive rate of 100% with EDI, EuroImmun, Abbott and Healgen assays and 95% for the DiaSorin assay.

Amongst the 23 P<14 samples, antibodies were detected for 65% (Abbott & EuroImmun), 61% (EDI) and 43% (DiaSorin) of the samples. Two samples R were close to the threshold in EDI and Abbott (EDI: 0.8 and Abbott 1.9; EDI: 1.0 and Abbott 0.8) resulting in one being positive in one assay and negative in the other (and vice-versa).

All 50 N were negative on the Abbott and EDI. Two samples were positive and 48 were negative on the EuroImmun (although 2 were equivocal). Two false positive samples were also observed on the DiaSorin, one being positive on both DiaSorin and EuroImmun assays.

The IgG kits showed a very good diagnostic ability to differentiate between P and N (Table 3). Overall, EuroImmun and DiaSorin showed lower sensitivity and specificity than EDI and Abbott. Sensitivity ranged between 81–100% on all time points for EDI, EuroImmun and Abbott. DiaSorin sensitivity was 71% on all time points and 95% for P≥14. Specificity was consistently 100% for the Abbott while it ranged between 92 to 100% for the other assays.

Table 3. Sensitivity of the assays was estimated on all time points and including only samples >14 days post PCR.

Assay Total Tested SARS-CoV-2 IgG Positive SARS-CoV-2 IgG Negative Equivocal result Result (95%CI)
SENSITIVITY SARS-CoV-2 Positive all time points EDI 43 35 8 n/a 81 (66–91)
EuroImmun 43 35 8 0 81 (66–91)
Abbott 43 35 8 n/a 81 (66–91)
DiaSorin 42 30 12 0 71 (55–84)
Healgen 27 27 0 n/a 100 (84–100)
SARS-CoV-2 Positive ≥14 days post PCR EDI 20 20 0 n/a 100 (80–100)
EuroImmun 20 20 0 0 100 (80–100)
Abbott 20 20 0 n/a 100 (80–100)
DiaSorin 20 19 1 0 95 (73–100)
Healgen 20 20 0 n/a 100 (80–100)
SPECIFICITY Pre-pandemic controls (N) EDI 50 0 50 n/a 100 (91–100)
EuroImmun 50 2 46 2 92 (79–97)
Abbott 50 0 50 n/a 100 (91–100)
DiaSorin 50 2 48 0 96 (85–99)
Healgen 4 0 4 n/a -
Other Respiratory Infection (CR) EDI 50 0 50 n/a 100 (91–100)
EuroImmun 50 3 47 0 94 (82–98)
Abbott 50 0 50 n/a 100 (91–100)
DiaSorin 50 2 48 0 96 (85–99)
Healgen 9 1 8 n/a -
Rheumatoid Arthritis (RA) EDI 22 0 22 n/a 92 (72–99)
EuroImmun 22 0 22 0 92 (72–99)
Abbott 22 0 22 n/a 100 (82–100)
DiaSorin 22 1 21 0 95 (75–100)
Healgen 4 0 4 n/a -
Thyroid Disorder (TSI) EDI 30 1 29 n/a 97 (81–100)
EuroImmun 30 0 28 2 93 (76–99)
Abbott 30 0 30 n/a 100 (85–100)
DiaSorin 30 1 29 0 97 (81–100)
Healgen 4 1 3 n/a -

Specificity was estimated on pre-2020 samples (N) from healthy individuals and patients with disorders that induce the production of potentially interfering substances. n/a = no equivocal range available.

Cross-reactivity

There were no SARS-CoV-2 IgG positive results from patients with non-SARS-CoV-2 infection (CR, n = 50, including seasonal flu (n = 7)), anti-CCP positive (RA, n = 22) nor TSI positive (n = 30) using the Abbott and the EDI assays. Overall, DiaSorin showed the highest (4%) cross-reactivity (2CR, 1 RA and 1 TSI), followed by EuroImmun (3%–3CR) and EDI (1%–1 TSI). The Mann-Whitney test showed that on the EDI only, the R value of samples used to test cross-reactivity (RA and TSI) was significantly elevated, however only one sample was falsely positive for SARS-CoV-2 (Fig 1).

Fig 1. Dot-plots of R values for each condition (N, P, CR, RA and TSI) for the (A) EDI, (B) Abbott, (C) EuroImmun and (D) DiaSorin tests.

Fig 1

Mann-Whitney analysis demonstrated a significant increase in the R value for the positive samples. Mann-Whitney statistical significance *p<0.05; **p<0.01 and ***p<0.001. Dotted line represents the positive cut-off for each assay.

Any sample that gave a false positive result in any of the immunoassays was also tested on the Healgen POCT and none were IgG positive. However, a very weak signal could be detected on one TSI sample and one sample from a patient with seasonal flu. Because of the very small number of samples tested; specificity calculation was not performed for the rapid test.

Time course analysis

We analyzed 1 to 13 data points for 43 P. We observed an increase of the signal for presence of IgG over time going from negativity to positivity and reaching a plateau (Fig 2). Sigmoid curve-fitting indicated a time from PCR to seroconversion at 9.8 days (95% CI 10.7–13.7), 10.2 (95% CI 8.5–11.8), 12.2 days (95% CI 10.7–13.7) and 10.4 days (95% CI 7.9–12.9) for EDI, Abbott, DiaSorin and EuroImmun assays respectively. Note that due to a limited number of EuroImmun tests available, we only had measurements for 56 (of 142) data points. One data point was missing for Abbott and 4 were missing for DiaSorin due to insufficient sample volume.

Fig 2. Seropositivity in specimen with PCR positive relative to day of PCR.

Fig 2

Dashed line represents the cut-off ratio for each assay. Solid black line and dotted lines represent the 4-parameter logistic curve-fit of the points with confidence interval. Time to PCR onset is calculated as curve inflection point.

We tested 48 samples from 27 P patients using the Healgen rapid test. Ninety four percent (n = 45) displayed a positive test for IgG. Samples showed positive results with POCT from day 7 post PCR although these were still negative in the other immunoassays (SARS-CoV-2 positive at day 12).

Assay concordance

Abbott and EDI had the greatest concordance with Cohen’s Kappa of 0.957 and 97.9% agreement between the all results (Table 4). DiaSorin was the most different, with agreements below 95%. The Healgen POCT concordance with the other assays was low (below 90%) but reflect a limited number of samples and may not be representative. Modifying the threshold to 0.8 for EDI would allow the detection of 2 more P<14 without increasing the rate of false positive. No change in threshold in the other assay would reclassify any results without dramatically affecting the specificity to either have a high rate of false positive or false negative.

Table 4. Cohen’s Kappa concordance analysis of the assays and overall (all samples included) agreement of results given as %.

Abbott Euroimmun DiaSorin Healgen Cohen’s Kappa (±SD) % agreement
0.950 (0.019) 97.9% 0.907 (0.034) 96.6% 0.892 (0.027) 94.8% 0.745 (0.083) 88.2% EDI Inline graphic
0.892 (0.037) 96.1% 0.891 (0.027) 94.8% 0.777 (0.078) 89.6% Abbott
0.852 (0.043) 94.7% 0.631 (0.111) 82.7% EuroImmun
0.485 (0.109) 76.1% DiaSorin

Equivocal results were considered negative.

Discussion

Statement of principal findings

In this head to head study we demonstrated the good performance of four commercially available serologic assays for SARS-CoV-2 and one POCT. Abbott, Epitope Diagnostics Ltd and EuroImmun demonstrated higher sensitivity and specificity than the DiaSorin assay on the same specimens. The Abbott assay showed no cross-reactivity to any other potential interfering substances tested while EDI, EuroImmun and DiaSorin cross-reacted in 1%, 3% and 4% of the sample tested. However, no assay cross-reacted with Influenza A and B or other coronaviruses. The analytical performance was deemed acceptable although it varied considerably between the different methods.

It is estimated that there are nearly 300 different SARS-CoV-2 antibody tests in development globally ranging from POCT through to assays on large clinical laboratory analyzers. Whilst data is accruing on the sensitivity and specificity of a number of these assays [5, 6] there are still many with little or no published, independent performance evaluations. Whilst there is a focus on the diagnostic accuracy of these tests, much less is understood about the analytical performance of these devices such as imprecision and cross reactivity with common respiratory illnesses or immunoassay interferences. Without this knowledge the sensitivity and specificity data is brought into question and it is important that the limitations of assay are fully understood before applying the results in clinical practice. The Food and Drug Administration and European Medicines Agency acceptance criteria for biological assays typically define the required between-run and within-run precision as CV≤15% for positive samples and ≤20% for samples at the lower limit of quantification [11, 12]. All immunoassays passed the criteria for positive samples.

Published median seroconversion time for IgG is around 14 days post symptoms [1315]. As we did not have access to symptom onset for most patients, we used PCR day to date the samples, before and after day 14. We included in the positive group only one sample per patient, thus limiting our sample size. However, our results are not biased by repeat measurements. All samples post day 14 were positive in all assay except DiaSorin, which returned one false negative (day 39). Positivity prior to day 14 was consistent between EDI, EuroImmun and Abbott. These results are differing from those published by PHE who observed more false negative results in the Abbott than the DiaSorin (92.7% sensitivity vs 95% sensitivity, respectively) [8]. We estimated seroconversion post PCR positivity to be between 9 and 12 days on these assays. Although we couldn’t do a full comparison of the POCT with the immunoassays, 100% of the P≥14 samples were IgG positive. More samples were also positive with POCT prior day 14 than in the other assays.

In regard to the POCT, our study showed excellent sensitivity and specificity. We observed no false negative results on P≥14 after a positive SARS-CoV-2 PCR and more samples were IgG positive P<14 than the other immunoassays. Two potential false positive were detected (including seasonal flu) but the signal was very weak and confirmation would be necessary. The results of systematic reviews on point-of care serological tests for SARS-CoV-2 suggest discontinuing the use of the devices due to low sensitivity [5]. Our results tend to reveal a different pattern however we only performed a limited number of tests.

We analyzed 50 samples collected in 2018 from patients with no known infection as negative controls. Both the EDI and the Abbott showed 100% specificity. However, EuroImmun and DiaSorin produced false positives (n = 4 and 2, respectively). Only one of these samples was common between both assays. PHE also showed lower specificity of the DiaSorin assay (vs Abbott). We analyzed 50 samples from patients (pre-pandemic) presenting with respiratory infection. Among those 7 had the seasonal flu, 8 had influenza A., other viruses included EBV, Varicellazoster virus, parainfluenza, Adenovirus. EDI and Abbott showed 100% specificity with no false positive; however, we observed 3 positive results with the EuroImmun, two of these also being positive with the DiaSorin. These samples were from patients with EBV (n = 1) and RSV (n = 2). Our results on EuroImmun differ slightly from a previous evaluation [16], where specificity of the assay was excellent as early as 4 days after positive PCR and only 2 of 28 samples showed borderline cross-reactivity to common human coronaviruses. None of the assays showed cross-reactivity either to the seasonal CoV flu or to Influenza A. Although it is based on a small number of sample (n = 7 for each), it brings confidence that assays will be able to discriminate SARS-CoV-2 antibodies during the next seasonal flu. Tang et al., showed similar results on 5 patients using EuroImmun and Abbott Assay [17]. A great variety of endogenous substances such as polyreactive antibodies or autoantibodies, can interfere with the reaction between analyte and reagent antibodies in immunoassays. Assays for SARS-CoV-2 are no exception. Manufacturers, and evaluation studies to date, offer a limited insight into cross-reactivity of other antibodies in particular to other SARS-CoV antibodies [1822]. A small independent study showed no cross-reactivity was seen for patients with Influenza A (n = 2), Influenza B (n = 2) and other coronaviruses (n = 5) [17]. Samples with potentially interfering antibodies did not cross-react in the Abbott Diagnostics assay, and a limited number cross-reacted in the other assays. Using the EDI assay, the signal obtained for both RA and TSI samples is significantly higher than the negative controls however all but one TSI sample remain below the cut-off of positivity. The cut-off is therefore appropriate for use with the assay with potentially cross-reactive substances. None of these samples was common between the different assays and modification of the various threshold would not improve performance of any assay.

Successful attempts to treat SARS-CoV-2 patients with blood from convalescent individuals suggest antibodies against SARS-CoV-2 may have the ability to confer protective immunity to the disease [2328]. Spike proteins are the most likely target for neutralizing antibodies are displayed on the surface of the virus whereas the nucleocapsid is contained within the viral envelope [29, 30]. Antibodies against the nucleocapsid have been shown to appear first [31, 32], followed by the production of antibodies against the spike protein [13, 14]. Therefore, assays based on the nucleocapsid detection appear to be more sensitive early on in the disease recovery but presence of anti-S1/S2 antibodies may indicate presence of neutralizing antibodies. Both the EuroImmun and the DiaSorin are targeted the spike protein of SARS-Cov-2 while the EDI and Abbott are targeted to the nucleocapsid protein of the virus. We observe a highest specificity of both nucleocapsid assays (EDI and Abbott, 100% (91–100%)) compared to the two spike assays (DiaSorin (96% (85–99%)) and EuroImmun (92% (79–97%)). Although the EuroImmun assay had the same sensitivity (all time points to PCR) as the EDI and Abbott, the DiaSorin assay was less sensitive (71% (73–100%) vs (81% (66–91%)), potentially supporting this hypothesis.

Overall, the assays had high concordance, DiaSorin being the least identical to the others, with higher false negative and false positive, and lower performance. This is in accordance with the high false positive rate observed by Boukli et al. [33] with the DiaSorin Liaison SARS-CoV-2 IgG assay on patients with non-SARS-CoV-2 acute infections. In April 2020 both the DiaSorin assay and the Abbott assays were authorized by Public Health England for emergency use in the clinical setting; 15 sites using the Abbott method reported to the WEQAS scheme in October while 84 (Abbott Architect and Alinity) reported to the UKNEQAS scheme in November. However, 3 sites using the DiaSorin method reported to WEQAS and 9 to UKNEQAS for the same period. The same samples were analyzed on the different platforms and therefore the direct comparison is possible. However, one needs to consider the potential variance in antigen as the Wuhan strain has evolved as geographic spread has occurred between the different regions of the globe (GISAID) [34] and it is possible that these differences will not be seen on a different set of samples. Harmonization of the assays is necessary but will be near impossible with such variation between assay designs (spike vs nucleocapsid). The Wales External Quality Assessment Scheme (WEQAS, UK, https://www.weqas.com/) and the UK NEQAS (https://ukneqas.org.uk/) are now offering a SARS-CoV-2 antibody external quality assessment (EQA) program for laboratories which will reduce uncertainty associated with different methods.

Strengths and limitations of this study

The main strength of our study is the direct comparison (same specimens) of five SARS-CoV-2 assays and the analysis of potentially cross-reactive substances produced during other respiratory infections and disorders such as rheumatoid arthritis and thyroid imbalance which are known to affect immunoassays.

Limitations include the limited number of positive samples due to the UK East Anglian region’s low prevalence and unavailability of onset date of SARS-CoV-2 symptoms. PCR may have been done from symptom onset day to several days post symptom; therefore, we based our seroconversion on PCR-positive date. The severity of symptoms was not available for all patients however these patients were hospitalized and we cannot comment on whether the production of antibodies correlate with the severity of symptoms.

Conclusion and policy implications

The role of serology testing in the management of people with SARS-CoV-2 infection will remain controversial until we have clear data that enables an understanding of how production of IgG relates to immunity over time and whether or not the presence or absence of antibodies can inform risk of future infection. Whilst the clinical utility of serology tested is debated, it is important that the diagnostic and analytical performance of these tests is understood and adequate for need so that there can be confidence in the results when a meaningful clinical use is determined. Without high quality analytical testing the clinical application of serology testing in the future is not viable.

This study examines the performance of four commercially available serologic assays for SARS-CoV-2 in a head to head study. Our study demonstrated good analytical performance for all of the assays, however we observed Abbott, EDI and EuroImmun demonstrated higher sensitivity and specificity than the DiaSorin assay in this study. Whilst a full evaluation was not possible the P14+ samples from the main study were used in a sub analysis using the Healgen POCT device which showed 100% specificity, this contradicts earlier studies [5, 6] and indicates that the evolution of the quality of POC devices has been rapid and some may now demonstrate adequate performance for antibody detection.

Assays showed 0–4% cross-reactivity, however none with Influenza viruses. This may give increase confidence of the test during the seasonal flu period. We observed differences between the assay responses with DiaSorin being the most different from the other three. We hypothesize that these differences may be linked to the design of the assay themselves (spike glycoprotein or nucleocapsid) and the timeline of production of antibodies for either antigen. We also suggested the possibility that the antigen plasticity and the antigen used when the manufacturer set up the test may influence the sensitivity of the CoV-2 assays. These findings highlight the importance of following the evolution of the antibody production and evolution of the virus over time. But it also highlights how harmonization of the assays will be complex.

Supporting information

S1 Table. Characteristics of the immunoassays evaluated, as provided by the manufacturers.

(DOCX)

Acknowledgments

The authors would like to thank the staff at the Norfolk and Norwich University Hospital and the Queen Elizabeth Hospital, Kings Lynn who collected the samples from SARS-CoV-2 patients, in particular Christopher Jeanes, and the Norfolk Arthritis Register (NOAR) for kindly providing historical samples of patients with inflammatory polyarthritis. We would also like to express our gratitude to Myra Del Rosario and Simon Clements who analyzed the samples on the Abbott Alinity, and Christopher McDonnell and Reenesh Prakash who procured EuroImmun kits.

Abbreviations

anti-CCP

Cyclic citrullinated peptide antibodies

CLSI

Clinical and Laboratory Standards Institute

CR

Cross-reactivity

CV

Coefficient of variation expressed as percentage

EBV

Epstein Barr Virus

EDI

Epitope Diagnostics Ltd

EQA

External quality assessment

IgG

Immunoglobulin G

LFIAs

Lateral flow immunoassays

N

Negative control

NNUH

Norwich and Norfolk University Hospital

NOAR

Norfolk Arthritis Register

OD

Optical density

OD

Optical density/absorbance

P

SARS-CoV-2 Positive

PHE

Public Health England

POCT

Point of Care Testing

QEH

Queen Elizabeth Hospital in King Lynn

R

Threshold of positivity

RA

Rheumatoid Arthritis

RLU

Relative Light Unit

S1/S2

Spike protein 1 and 2

TSI

Thyroid stimulating immunoglobulin

WEQAS

Wales External Quality Assessment Scheme (UK)

Data Availability

All relevant data are within the manuscript and its Supporting information files.

Funding Statement

The authors received no specific financial support for this study. They acknowledge the Norfolk, Suffolk, Essex and Bedfordshire Freemasons for their generous material support in providing funding for some equipment used in this study.

References

  • 1.Chang MS, Lu YT, Ho ST, Wu CC, Wei TY, Chen CJ, et al. Antibody detection of SARS-CoV spike and nucleocapsid protein. Biochem Biophys Res Commun. 2004;314(4):931–6. 10.1016/j.bbrc.2003.12.195 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 2.Woelfel R, Corman VM, Guggemos W, Seilmaier M, Zange S, Mueller MA, et al. Clinical presentation and virological assessment of hospitalized cases of coronavirus disease 2019 in a travel-associated transmission cluster. medRxiv. 2020;2020.03.05.20030502. [Google Scholar]
  • 3.Long QX, Liu BZ, Deng HJ, Wu GC, Deng K, Chen YK, et al. Antibody responses to SARS-CoV-2 in patients with COVID-19. Nat Med. 2020;26(6):845–8. 10.1038/s41591-020-0897-1 [DOI] [PubMed] [Google Scholar]
  • 4.Okba NMA, Müller MA, Li W, Wang C, GeurtsvanKessel CH, Corman VM, et al. Severe Acute Respiratory Syndrome Coronavirus 2-Specific Antibody Responses in Coronavirus Disease Patients. Emerg Infect Dis. 2020;26(7):1478–88. 10.3201/eid2607.200841 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5.Deeks JJ, Dinnes J, Takwoingi Y, Davenport C, Spijker R, Taylor-Phillips S, et al. Antibody tests for identification of current and past infection with SARS-CoV-2. Cochrane Database Syst Rev. 2020;2 10.1002/14651858.CD013652 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6.Lisboa Bastos M, Tavaziva G, Abidi SK, Campbell JR, Haraoui L-PP, Johnston JC, et al. Diagnostic accuracy of serological tests for covid-19: systematic review and meta-analysis. Bmj [Internet]. 2020;370:m2516. Available from: http://www.bmj.com/lookup/doi/10.1136/bmj.m2516 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7.UK Medicines and Healthcare products Regulatory Agency (MHRA). Target Product Profile: enzyme Immunoassay (EIA) Antibody tests to help determine if people have antibodies to SARS-CoV-2 [Internet]. [cited 2020 Aug 10]. https://www.gov.uk/government/publications/how-tests-and-testing-kits-for-coronavirus-covid-19-work/target-product-profile-enzyme-immunoassay-eia-antibody-tests-to-help-determine-if-people-have-antibodies-to-sars-cov-2
  • 8.Public Health England. Evaluation of sensitivity and specificity of four commercially available SARS-CoV-2 antibody immunoassays About Public Health England. 2020;(July).
  • 9.Duong YT, Wright CG, Justman J. Antibody testing for coronavirus disease 2019: Not ready for prime time. Vol. 370, The BMJ. 2020. 10.1136/bmj.m2655 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10.Neil Carey R, Paul Durham A, Hauck WW, Kallner A, Kondratovich MV., Guy Middle J, et al. EP15-A3 User Verification of Precision and Estimation of Bias; Approved Guideline—Third Edition. Clin Lab Stand Inst. 2014;34(12):1–81. [Google Scholar]
  • 11.Kaza M, Karaźniewicz-Łada M, Kosicka K, Siemiątkowska A, Rudzki PJ. Bioanalytical method validation: new FDA guidance vs. EMA guideline. Better or worse? J Pharm Biomed Anal. 2019;165:381–5. 10.1016/j.jpba.2018.12.030 [DOI] [PubMed] [Google Scholar]
  • 12.Center for Drug Evaluation and Research. Bioanalytical Method Validation Guidance for Industry Bioanalytical Method Validation. FDA Guid Ind [Internet]. 2018;(May):1–22. http://www.fda.gov/Drugs/GuidanceComplianceRegulatoryInformation/Guidances/default.htmand/orhttp://www.fda.gov/AnimalVeterinary/GuidanceComplianceEnforcement/GuidanceforIndustry/default.htm
  • 13.Guo L, Ren L, Yang S, Xiao M, Chang D, Yang F, et al. Profiling Early Humoral Response to Diagnose Novel Coronavirus Disease (COVID-19). Clin Infect Dis. 2020;(Xx Xxxx):1–8. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 14.Zhao J, Yuan Q, Wang H, Liu W, Liao X, Su Y, et al. Antibody responses to SARS-CoV-2 in patients of novel coronavirus disease 2019. Clin Infect Dis. 2020; 10.1093/cid/ciaa344 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15.Burbelo PD, Riedo FX, Morishima C, Rawlings S, Smith D, Das S, et al. Sensitivity in Detection of Antibodies to Nucleocapsid and Spike Proteins of Severe Acute Respiratory Syndrome Coronavirus 2 in Patients With Coronavirus Disease 2019. J Infect Dis. 2020; 10.1093/infdis/jiaa273 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 16.Beavis KG, Matushek SM, Abeleda APF, Bethel C, Hunt C, Gillen S, et al. Evaluation of the EUROIMMUN Anti-SARS-CoV-2 ELISA Assay for detection of IgA and IgG antibodies. J Clin Virol. 2020;129 10.1016/j.jcv.2020.104468 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 17.Tang MS, Hock KG, Logsdon NM, Hayes JE, Gronowski AM, Anderson NW, et al. Clinical Performance of Two SARS-CoV-2 Serologic Assays. Clin Chem. 2020; 10.1093/clinchem/hvaa120 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18.Zheng Z, Monteil VM, Maurer-Stroh S, Yew CW, Leong C, Mohd-Ismail NK, et al. Monoclonal antibodies for the S2 subunit of spike of SARS-CoV-1 cross-react with the newly-emerged SARS-CoV-2. Euro Surveill. 2020; 10.2807/1560-7917.ES.2020.25.28.2000291 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19.Patrick DM, Petric M, Skowronski DM, Guasparini R, Booth TF, Krajden M, et al. An outbreak of human coronavirus OC43 infection and serological cross-reactivity with SARS coronavirus. Can J Infect Dis Med Microbiol. 2006;17(6):330–6. 10.1155/2006/152612 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 20.Lv H, Wu NC, Tsang OTY, Yuan M, Perera RAPM, Leung WS, et al. Cross-reactive Antibody Response between SARS-CoV-2 and SARS-CoV Infections. Cell Rep. 2020; [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 21.Che XY, Qiu LW, Liao ZY, Di Wang Y, Wen K, Pan YX, et al. Antigenic cross-reactivity between severe acute respiratory syndrome-associated coronavirus and human coronaviruses 229E and OC43. J Infect Dis. 2005;191(12):2033–7. 10.1086/430355 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 22.Chan KH, Cheng VCC, Woo PCY, Lau SKP, Poon LLM, Guan Y, et al. Serological responses in patients with severe acute respiratory syndrome coronavirus infection and cross-reactivity with human coronaviruses 229E, OC43, and NL63. Clin Diagn Lab Immunol. 2005;12(11):1317–21. 10.1128/CDLI.12.11.1317-1321.2005 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 23.Hegerova L, Gooley T, Sweerus KA, Maree CL, Bailey N, Bailey M, et al. Use of Convalescent Plasma in Hospitalized Patients with Covid-19—Case Series. Blood. 2020; 10.1182/blood.2020006964 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 24.Duan K, Liu B, Li C, Zhang H, Yu T, Qu J, et al. Effectiveness of convalescent plasma therapy in severe COVID-19 patients. Proc Natl Acad Sci U S A. 2020;117(17):9490–6. 10.1073/pnas.2004168117 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 25.Shen C, Wang Z, Zhao F, Yang Y, Li J, Yuan J, et al. Treatment of 5 Critically Ill Patients with COVID-19 with Convalescent Plasma. JAMA—J Am Med Assoc. 2020;323(16):1582–9. 10.1001/jama.2020.4783 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 26.Zhang B, Liu S, Tan T, Huang W, Dong Y, Chen L, et al. Treatment With Convalescent Plasma for Critically Ill Patients With Severe Acute Respiratory Syndrome Coronavirus 2 Infection. Chest. 2020;158(1):e9–13. 10.1016/j.chest.2020.03.039 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 27.Ahn JY, Sohn Y, Lee SH, Cho Y, Hyun JH, Baek YJ, et al. Use of convalescent plasma therapy in two covid-19 patients with acute respiratory distress syndrome in Korea. J Korean Med Sci. 2020;35(14). 10.3346/jkms.2020.35.e149 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 28.Ye M, Fu D, Ren Y, Wang F, Wang D, Zhang F, et al. Treatment with convalescent plasma for COVID-19 patients in Wuhan, China. J Med Virol. 2020; 10.1002/jmv.25882 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 29.Wang C, Li W, Drabek D, Okba NMA, van Haperen R, Osterhaus ADME, et al. A human monoclonal antibody blocking SARS-CoV-2 infection. Nat Commun. 2020; [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 30.Ni L, Ye F, Cheng ML, Feng Y, Deng YQ, Zhao H, et al. Detection of SARS-CoV-2-Specific Humoral and Cellular Immunity in COVID-19 Convalescent Individuals. Immunity. 2020; 10.1016/j.immuni.2020.04.023 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 31.To KKW, Tsang OTY, Leung WS, Tam AR, Wu TC, Lung DC, et al. Temporal profiles of viral load in posterior oropharyngeal saliva samples and serum antibody responses during infection by SARS-CoV-2: an observational cohort study. Lancet Infect Dis. 2020; 10.1016/S1473-3099(20)30196-1 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 32.Zou L, Ruan F, Huang M, Liang L, Huang H, Hong Z, et al. SARS-CoV-2 viral load in upper respiratory specimens of infected patients. New England Journal of Medicine. 2020. 10.1056/NEJMc2001737 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 33.Boukli N, Le Mene M, Schnuriger A, Cuervo NS, Laroche C, Morand-Joubert L, et al. High incidence of false positive results in patients with other acute infections, using the LIAISON® SARS-CoV-2 commercial chemiluminescent micro-particle immunoassay for detection of IgG anti SARS-CoV-2 antibodies. J Clin Microbiol. 2020; [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 34.GISAID. Genomic epidemiology of hCoV-19 [Internet]. 2020 [cited 2020 Aug 18]. https://www.gisaid.org/epiflu-applications/next-hcov-19-app/

Decision Letter 0

Muhammad Adrish

9 Dec 2020

PONE-D-20-31021

Performance of SARS-CoV-2 Serology tests: Are they good enough?

PLOS ONE

Dear Dr. Piec,

Thank you for submitting your manuscript to PLOS ONE. After careful consideration, we feel that it has merit but does not fully meet PLOS ONE’s publication criteria as it currently stands. Therefore, we invite you to submit a revised version of the manuscript that addresses the points raised during the review process.

==============================

ACADEMIC EDITOR: Please review comments made by the reviewers and provide point by point response in your revised manuscript.

==============================

Please submit your revised manuscript by due date. If you will need more time than this to complete your revisions, please reply to this message or contact the journal office at plosone@plos.org. When you're ready to submit your revision, log on to https://www.editorialmanager.com/pone/ and select the 'Submissions Needing Revision' folder to locate your manuscript file.

Please include the following items when submitting your revised manuscript:

  • A rebuttal letter that responds to each point raised by the academic editor and reviewer(s). You should upload this letter as a separate file labeled 'Response to Reviewers'.

  • A marked-up copy of your manuscript that highlights changes made to the original version. You should upload this as a separate file labeled 'Revised Manuscript with Track Changes'.

  • An unmarked version of your revised paper without tracked changes. You should upload this as a separate file labeled 'Manuscript'.

If you would like to make changes to your financial disclosure, please include your updated statement in your cover letter. Guidelines for resubmitting your figure files are available below the reviewer comments at the end of this letter.

If applicable, we recommend that you deposit your laboratory protocols in protocols.io to enhance the reproducibility of your results. Protocols.io assigns your protocol its own identifier (DOI) so that it can be cited independently in the future. For instructions see: http://journals.plos.org/plosone/s/submission-guidelines#loc-laboratory-protocols

We look forward to receiving your revised manuscript.

Kind regards,

Muhammad Adrish

Academic Editor

PLOS ONE

Journal Requirements:

When submitting your revision, we need you to address these additional requirements.

1) Please ensure that your manuscript meets PLOS ONE's style requirements, including those for file naming. The PLOS ONE style templates can be found at

https://journals.plos.org/plosone/s/file?id=wjVg/PLOSOne_formatting_sample_main_body.pdf and

https://journals.plos.org/plosone/s/file?id=ba62/PLOSOne_formatting_sample_title_authors_affiliations.pdf

2)  Thank you for stating the following in the Acknowledgments Section of your manuscript:

[We finally wish to thank the Norfolk, Suffolk, Essex and Hertfordshire Freemasons for their generous financial support to purchase the Agility system.]

We note that you have provided funding information that is not currently declared in your Funding Statement. However, funding information should not appear in the Acknowledgments section or other areas of your manuscript. We will only publish funding information present in the Funding Statement section of the online submission form.

Please remove any funding-related text from the manuscript and let us know how you would like to update your Funding Statement. Currently, your Funding Statement reads as follows:

 [The funders had no role in study design, data collection and analysis, decision to

publish, or preparation of the manuscript.]

Please include your amended statements within your cover letter; we will change the online submission form on your behalf.

3) Please include captions for your Supporting Information files at the end of your manuscript, and update any in-text citations to match accordingly. Please see our Supporting Information guidelines for more information: http://journals.plos.org/plosone/s/supporting-information.

[Note: HTML markup is below. Please do not edit.]

Reviewers' comments:

Reviewer's Responses to Questions

Comments to the Author

1. Is the manuscript technically sound, and do the data support the conclusions?

The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented.

Reviewer #1: Yes

Reviewer #2: Yes

Reviewer #3: Yes

**********

2. Has the statistical analysis been performed appropriately and rigorously?

Reviewer #1: Yes

Reviewer #2: Yes

Reviewer #3: Yes

**********

3. Have the authors made all data underlying the findings in their manuscript fully available?

The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified.

Reviewer #1: Yes

Reviewer #2: Yes

Reviewer #3: Yes

**********

4. Is the manuscript presented in an intelligible fashion and written in standard English?

PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here.

Reviewer #1: Yes

Reviewer #2: Yes

Reviewer #3: Yes

**********

5. Review Comments to the Author

Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters)

Reviewer #1: Major comment: The manuscript# PONE-D-20-31021entitled “Performance of SARS-CoV-2 Serology tests: Are they good enough?” assess the performance of SARS-CoV-2 serologic tests available in clinical practices.

These methods were introduced into clinical practice quickly without an extensive validation usually required by the regulatory authorities. The IgG serologic Assays from EuroImmun, Epitope Diagnostics (EDI), Abbott Diagnostics and DiaSorin and a rapid IgG/IgM test from Healgen were evaluated for the imprecision, diagnostic sensitivity and specificity and concordance. The SARS-CoV-2 patient samples were analyzed as pre-pandemic control specimen with confirmed respiratory infections, inflammatory polyarthritis and several samples positive for thyroid stimulating immunoglobulin and post pandemic samples. Authors showed that the diagnostic sensitivity of the assays were 100% (CI: 80-100%) for Abbott, EDI and EuroImmun and 95% (CI: 73-100%) for DiaSorin at ≥14 days post infection. Only the Abbott assay had a diagnostic specificity of 100% (CI: 91-100%). The DiaSorin displayed more false negative results across all conditions tested. The imprecision for EDI and EuroImmun, was 0.02-14.0% CV while for Abbott and DiaSorin imprecision (CV) ranged from 5.2% - 8.1% and 8.2% - 9.6% respectively. Overall, concordance of the assays ranged from 76.1% to 97.9%. The point of care rapid test Healgen, showed excellent sensitivity and specificity. Authors attributed the differences in results to the use of nucleocapsid and spike proteins. The point of care device tested demonstrated adequate performance for antibody detection.

Limited sample size should be included in the limitation of the study.

The Figure legend describes Figure 1 as Box-plot even though data shown as a dot plot.

The Figure 1A reveled very little to no cross reactivity. The statistical analysis among groups show significant differences between N to RA and N to TSI. None of the values in RA and only one value in TSI groups is above the positive cutoff value. May be author should comment on meaning of such statistical significance.

In this study the DiaSorin displayed the most-false negatives and showed cross-reactivity when comparing with other serologic assays. It will important to add in the discussion the frequency of use of these tests within UK or globally specially if that data is available through health care systems or through the company.

Reviewer #2: This is a well written manuscript. COVID -19 is also a huge demanding topic. I have some minor comments.

Overall: The number of positive samples is very low (43). Though the present analysis showed a good agreement. But it may alter if the sample size increased. There are so many published paper with the information of different type of serological testing of SARS-CoV-2. Presently we need more specific decision for specific serological test.

Please add some explanation in discussion section, why the specificity of DiaSorin assay showed low.

Line 111: Please add the time period of the SARS-CoV-2 positivity; like March to May.

Reviewer #3: The study is well designed and results are clear presented.

Concerning the results higher sensitivity for point of care devices was found, compared with other publications. The absence of cross-reactivity to seasonal coronavirus and influenza A is a remarkable result. Evolution of the virus and antibodies production turns serology assays a complex and hard work.

Although Authors did not consider strengths or weaknesses of the work.

To date the samples for serology PCR day was used, but it would be better to date from onset of symptoms.

The study is small and results probably will need to be reproduced in larger studies, in different places and in systematic reviews.

Line 268 – we chose – please correct.

**********

6. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files.

If you choose “no”, your identity will remain anonymous but your review may still be made public.

Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy.

Reviewer #1: No

Reviewer #2: No

Reviewer #3: No

[NOTE: If reviewer comments were submitted as an attachment file, they will be attached to this email and accessible via the submission site. Please log into your account, locate the manuscript record, and check for the action link "View Attachments". If this link does not appear, there are no attachment files.]

While revising your submission, please upload your figure files to the Preflight Analysis and Conversion Engine (PACE) digital diagnostic tool, https://pacev2.apexcovantage.com/. PACE helps ensure that figures meet PLOS requirements. To use PACE, you must first register as a user. Registration is free. Then, login and navigate to the UPLOAD tab, where you will find detailed instructions on how to use the tool. If you encounter any issues or have any questions when using PACE, please email PLOS at figures@plos.org. Please note that Supporting Information files do not need this step.

Attachment

Submitted filename: Review PONE-D-20-31021-11-6-20.docx

PLoS One. 2021 Feb 17;16(2):e0245914. doi: 10.1371/journal.pone.0245914.r002

Author response to Decision Letter 0


22 Dec 2020

Ethics changes had been requested previous to the decision letter and author stated the following:

The University of East Anglia Ethics Committee was consulted and confirmed no extra ethics was required as all samples were anonymized and this is covered by the UEA blanket Ethics for method development.

We also consulted the UK National Heath System Health Research Authority online tool kit which confirmed no supplemental Ethics or patient consent were required.

Attachment

Submitted filename: Response to Reviewers.docx

Decision Letter 1

Muhammad Adrish

11 Jan 2021

Performance of SARS-CoV-2 Serology tests: Are they good enough?

PONE-D-20-31021R1

Dear Dr. Piec,

We’re pleased to inform you that your manuscript has been judged scientifically suitable for publication and will be formally accepted for publication once it meets all outstanding technical requirements.

Within one week, you’ll receive an e-mail detailing the required amendments. When these have been addressed, you’ll receive a formal acceptance letter and your manuscript will be scheduled for publication.

An invoice for payment will follow shortly after the formal acceptance. To ensure an efficient process, please log into Editorial Manager at http://www.editorialmanager.com/pone/, click the 'Update My Information' link at the top of the page, and double check that your user information is up-to-date. If you have any billing related questions, please contact our Author Billing department directly at authorbilling@plos.org.

If your institution or institutions have a press office, please notify them about your upcoming paper to help maximize its impact. If they’ll be preparing press materials, please inform our press team as soon as possible -- no later than 48 hours after receiving the formal acceptance. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information, please contact onepress@plos.org.

Kind regards,

Muhammad Adrish

Academic Editor

PLOS ONE

Additional Editor Comments (optional):

All comments have been addressed

Reviewers' comments:

Reviewer's Responses to Questions

Comments to the Author

1. If the authors have adequately addressed your comments raised in a previous round of review and you feel that this manuscript is now acceptable for publication, you may indicate that here to bypass the “Comments to the Author” section, enter your conflict of interest statement in the “Confidential to Editor” section, and submit your "Accept" recommendation.

Reviewer #1: All comments have been addressed

Reviewer #2: All comments have been addressed

Reviewer #3: All comments have been addressed

**********

2. Is the manuscript technically sound, and do the data support the conclusions?

The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented.

Reviewer #1: Yes

Reviewer #2: Yes

Reviewer #3: (No Response)

**********

3. Has the statistical analysis been performed appropriately and rigorously?

Reviewer #1: Yes

Reviewer #2: Yes

Reviewer #3: (No Response)

**********

4. Have the authors made all data underlying the findings in their manuscript fully available?

The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified.

Reviewer #1: Yes

Reviewer #2: Yes

Reviewer #3: (No Response)

**********

5. Is the manuscript presented in an intelligible fashion and written in standard English?

PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here.

Reviewer #1: Yes

Reviewer #2: Yes

Reviewer #3: (No Response)

**********

6. Review Comments to the Author

Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters)

Reviewer #1: Comment: The manuscript# PONE-D-20-31021R1 entitled “Performance of SARS-CoV-2

Serology tests: Are they good enough?” assessed the performance of SARS-CoV-2 serologic tests available in clinical practices. The topic is highly relevant in the current settings. In this revised manuscript authors have addressed the comments and suggestions provided by each of the reviewers.

Reviewer #2: Author addressed all the comments properly. It is a great opportunity for me to be a reviewer of this manuscript which is related to COVID-1.

Reviewer #3: (No Response)

**********

7. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files.

If you choose “no”, your identity will remain anonymous but your review may still be made public.

Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy.

Reviewer #1: No

Reviewer #2: No

Reviewer #3: No

Acceptance letter

Muhammad Adrish

2 Feb 2021

PONE-D-20-31021R1

Performance of SARS-CoV-2 Serology tests: Are they good enough?

Dear Dr. Piec:

I'm pleased to inform you that your manuscript has been deemed suitable for publication in PLOS ONE. Congratulations! Your manuscript is now with our production department.

If your institution or institutions have a press office, please let them know about your upcoming paper now to help maximize its impact. If they'll be preparing press materials, please inform our press team within the next 48 hours. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information please contact onepress@plos.org.

If we can help with anything else, please email us at plosone@plos.org.

Thank you for submitting your work to PLOS ONE and supporting open access.

Kind regards,

PLOS ONE Editorial Office Staff

on behalf of

Dr. Muhammad Adrish

Academic Editor

PLOS ONE

Associated Data

    This section collects any data citations, data availability statements, or supplementary materials included in this article.

    Supplementary Materials

    S1 Table. Characteristics of the immunoassays evaluated, as provided by the manufacturers.

    (DOCX)

    Attachment

    Submitted filename: Review PONE-D-20-31021-11-6-20.docx

    Attachment

    Submitted filename: Response to Reviewers.docx

    Data Availability Statement

    All relevant data are within the manuscript and its Supporting information files.


    Articles from PLoS ONE are provided here courtesy of PLOS

    RESOURCES