Abstract
X-ray based lung function (XLF) as a planar method uses dramatically less X-ray dose than computed tomography (CT) but so far lacked the ability to relate its parameters to pulmonary air volume. The purpose of this study was to calibrate the functional constituents of XLF that are biomedically decipherable and directly comparable to that of micro-CT and whole-body plethysmography (WBP). Here, we developed a unique set-up for simultaneous assessment of lung function and volume using XLF, micro-CT and WBP on healthy mice. Our results reveal a strong correlation of lung volumes obtained from radiographic XLF and micro-CT and demonstrate that XLF is superior to WBP in sensitivity and precision to assess lung volumes. Importantly, XLF measurement uses only a fraction of the radiation dose and acquisition time required for CT. Therefore, the redefined XLF approach is a promising tool for preclinical longitudinal studies with a substantial potential of clinical translation.
Subject terms: Biological techniques, Medical research
Introduction
Respiratory diseases account for ~ 10% of mortality worldwide, which is set to rise due to persistent smoking, pollution and occupational irritants1. Depending on the underlying disease, the characteristics of lung dysfunction may involve inadequate oxygen exchange, breath shortening, chronic cough, chest pain and dyspnoea2,3. To improve our knowledge about the basic mechanisms regarding lung physiology, pathophysiology and treatment strategies, the use of mouse models of lung disease in translational research is inevitable. The pre-clinical assessment of pulmonary function is a valuable tool not only for investigating the pathology of respiratory and allergic diseases but also enables preclinical evaluation of the response to novel therapeutic strategies.
Changes in lung function can be monitored non-invasively in restrained or unrestrained mice by double chamber plethysmography or barometric plethysmography2,4. Other methods require invasive intubation procedures some of which tend to be terminal5. Of all available techniques, unrestrained whole-body plethysmography (WBP) is extensively used in longitudinal studies due to its ease of use and high data output3,6–8. However, there is also considerable concern over the validity of plethysmography techniques9–12. The main limitation of this technique is that the functional outcomes rely on the plethysmograph pressure changes, and the relationship between chamber pressure and the lung mechanics might be non-linear. Furthermore, the accurate tidal volume estimates are significantly affected by subtle changes in temperature, pulmonary mechanism and fluctuations in barometric pressure3,10,13. Plethysmography may also be inconsistent between mouse strains requiring precise selection of the control groups11.
Recently, the use of lab and synchrotron radiation based micro-computed tomography (micro-CT) has been increasingly used in preclinical studies to assess quantitative parameters including lung volumes, air spaces or lesions and mean lung density14–19. These measurements have been performed in an attempt to reflect the structural and functional competence of the lung during disease progression and in response to therapy in numerous animal models of lung diseases, such as cancer, fibrosis, emphysema and transplantation14. Not only in clinical routine, but also in preclinical research, dose restrictions are major obstacles. Small-animal models, especially mouse models, are often utilized in preclinical research. Imaging in live mice for micro-CT based longitudinal lung studies requires a dramatically higher spatial and temporal resolution due to the small size of the mouse and its fast metabolism. Consequently, this limits the use of CT based high-resolution imaging for longitudinal lung studies and also hampers the clinical application of this approach20–26.
Therefore, the development of novel and sophisticated lung function measurement techniques with improved reliability and minimal radiation is fundamentally required. To this end, we recently established a non-invasive lung function method based on planar cinematic X-ray imaging of the chest, namely X-ray-based lung function (XLF) to measure the lung function in preclinical mouse models of allergic airway inflammation (AAI) over time27,28. Using minimal radiation dose and short exposure time, XLF showed significantly higher sensitivity than WBP for reliable assessment of lung function during AAI and response to dexamethasone treatment27. However, previously the functional parameters of XLF were not exactly relatable to commonly used lung function techniques including WBP and micro-CT and none of the XLF parameters represented the lung volume.
In this study, we compared XLF determined end-inspiratory volume (EIV) with the lung volumes from micro-CT and conventional WBP to establish XLF as a biomedically relevant technique. To achieve this, we first designed a unique experimental set-up for performing correlative lung function measurements using XLF, micro-CT and WBP. Using micro-CT as a gold standard approach, XLF was found sensitive and precise to determine lung volumes, as it exhibited a higher degree of correlation with micro-CT than WBP. By successfully correlating two-dimensional (2D) XLF measurements directly with three-dimensional (3D) micro-CT data in healthy mice, we present the refined XLF algorithm as a sensitive and efficient tool for determining lung volume at minimal radiation dose and exposure time.
Results
Experimental set-up facilitates correlative lung function measurements
To correlate the XLF parameters directly with WBP and micro-CT, we designed a unique experimental set-up that enabled simultaneous measurements by combining either XLF or micro-CT with WBP. As shown in Fig. 1, the custom-made plethysmography chamber was placed into the gantry of the in-vivo micro-CT system. The chamber had multiple connectors, which served as isoflurane inlet/outlet and for connecting a differential pressure sensor (DPS; see “Materials and methods” section). Overall, the designed set-up served as an ideal platform to perform correlative lung function.
Correlating the timing of micro-CT and WBP signal
To perform direct correlation between micro-CT and WBP data we tested and synchronized the timing of signal from each entity. Since the CT-scanner had no signal output to provide information about the timepoint the first image frame was acquired, we had to conceive our own trigger from the CT by installing a piezoelectric sensor to provide an acoustic signal for matching the starting point of CT data acquisition and WBP recording. The peak signals from each method were then synchronized to match the timing and traces of signals recorded. When comparing the peak of the average X-ray attenuation at the region of interest (ROI) over time (Fig. 2a) and the peak of the WBP volume signal we noticed a significant delay of 27.5 ± 20.0 ms (p < 0.01, paired t-test) (Fig. 2b,c). However, taking into consideration the exposure time of the X-ray detector of 33 ms, we assumed that the minimum X-ray attenuation at the lung region and the peak of the WBP lung volume trace coincide as expected (Supp. Fig. S1).
The rise time (trise) of the rXTF signal (55.4 ± 16.1 ms) and the WBP volume signal (52.8 ± 8.7 ms) did not differ (p = 0.0004, paired t-test) (Fig. 2d). This indicates that WBP can predict the inspiratory phase of the respiratory cycle very precisely. However, while comparing tfall and tau (τ) of the expiratory phase, the two methods revealed significantly different results: tfall 187.1 ± 53.8 ms for rXTF versus 497.4 ± 100.8 ms for WBP (p < 0.0421; paired t-test, Fig. 2e) and τ 215.4 ± 65.4 ms for rXTF and 341.3 ± 92.6 ms for WBP (p < 0.0039, Fig. 2f). One factor that may account for the difference is the uncertainty regarding the position of the zero-flow point for the end of each respiratory cycle. Moreover, the slower flow for rXTF during expiration that accounts for this difference, might be a consequence of isoflurane-induced bronchoconstriction29.
Micro-CT based volume measurements
We derived separate volumes () for the lung in inspiration and expiration from micro-CT by performing retrospective gating in combination with a modified Feldkamp (FDK) reconstruction algorithm30. Both steps were implemented in a custom-built software called RetrospeCT. We used the FDK algorithm implemented in the TIGRE toolbox (https://github.com/CERN/TIGRE) which—in contrast to the typically applied version of that algorithm—allows using non-uniformly distributed angular projections. From the angular projections, an ROI line was adjusted to select the representative region of lung expansion near the diaphragm (Fig. 3a). Figure 3b shows an exemplary breathing cycle curve generated by the temporal variations in the X-ray attenuation at this ROI. A moving average filtered curve which presented gradual variations in the trendline was subtracted from the signal to factor out the background. Moreover, the minimum peak distance (expiration points) and the number of frames detected per inhaled phase were adapted to optimize the selection of projections that maintained a consistent data quality for both phases (Fig. 3c). The selection was based on thresholding: here, the highest value above the inspiration threshold (green line) and the lowest value below the expiration threshold (yellow line) were used for defining end-inspiration and end-expiration frames (Fig. 3b).
From the breathing curve, frames at the expiration and inspiration phases were selected resulting in two projection sets with strongly reduced frame numbers (approximately 340 each out of the 8892 frames recorded) and non-equidistant angular spacing (Fig. 3d). Owing to the low number of frames, standard reconstruction algorithms yielded a 3D reconstruction of poor quality. Thus, a modified FDK reconstruction algorithm based on the graphics processing unit (GPU) was used for tomographic reconstruction of the lung for both the expiration and inspiration set of projections (Fig. 3d, Suppl. Fig. S2). Still, the resulting 3D data demonstrated a high noise level and a non-uniform background intensity which hindered the segmentation of the lung and thereby a precise volume measurement. To correct the non-uniform background, a slice below the lung region was subtracted from all other slices. Subsequently, a 3D Gaussian-filter was applied to reduce the noise level (Suppl. Fig. S2). From the resulting filtered volume, the lung at both breathing phases was segmented individually. The lung was first masked to eliminate the area of the surrounding thoracic regions. To select the intensity values that enabled an optimal selection of , histograms were obtained from manually defined regions inside and outside the lung cavity (Fig. 3e). A region growing algorithm was utilized in which the upper and lower threshold values from the histogram were exploited to select voxels consistent with the intensity of the aerated regions. The gaps within the resulting segmented region were filled by applying a closing operator. Lastly, was calculated from the segmented region by simply multiplying the number of segmented voxels with the volume per voxel. Other denser regions such as the mediastinum, consolidation of lung parenchyma and the airway tracts were excluded (Fig. 3f). In principle, did not only account for the air spaces within the lung but also for the micro tissue structures including small blood vessels and connective tissue which could not be segmented due to limited resolution of the reconstruction.
For testing the reliability and precision of the described workflow, measurements were performed in three dead mice of different age, weights and strains by mechanically ventilating them with increasing stroke volumes (Tables S1, S2). To achieve this, the tidal volumes from micro-CT () were determined by calculating the difference between the volume of lung cavity during inspiration and expiration (). The fractional increase in correlated strongly with the applied stroke volumes (Fig. 3g). Unlike the ventilator stroke volume which represents the total ventilation volume, measures the volume of lung aerated regions and a strong correlation between these parameters signifies that indeed reflects the volume of air exchanged. All three mice weighing 30.1 g, 28.2 g and 25.0 g expressed a strong linear positive correlation between both variables with R2 values of 0.95, 0.99 and 0.84, respectively (Fig. 3g). However, the relative change between and the stroke volume did not show an absolute match, which could be attributed to an amount of air leakage, which depends on various factors, including the speed of ventilation, increased airway resistance and reduced lung compliance in dead mouse. Nevertheless, the sensitivity of to the changes in stroke volume warranted the use of this approach as a standard procedure for correlating the lung volumes measured by each method.
Correlation of micro-CT determined volumes with WBP
Using the correlative approach, the in-vivo micro-CT measurements were performed in parallel with WBP. For comparing the data obtained from both methods, was correlated with the tidal volume from WBP (TV[WBP]). The parameters presented only a weak correlation (correlation coefficient R2 = 0.5061, p value = 0.0239, Fig. 4).
Interestingly, both parameters correlated rather poorly (R2 = 0.5061), probably due to the fundamental differences in volumetric constituents measured by each method. While quantifies the aerated regions of the lung, TV[WBP] is derived from pressure changes produced by warming and cooling of the inspired air during every single breath31,32. Moreover, the volume is integrated from the calibrated flow signal (Fig. 2). Although band-pass filtering removed the bias flow, the exact zero-flow point for each respiratory cycle remains uncertain due to change of temperature and humidity inside the WBP chamber.
Quantification of advanced XLF parameters
The functional parameters for XLF were measured from the average X-ray transmission at the chest area over time. Previously, we reported the calibration of X-ray transmission over time via the selection of a ROI at left and right lung lobes at the point of maximum contraction, which was normalized by the averaged background intensity selected at a region outside the mouse26. However, to accomplish adequate background selection, a large FOV is required, which in turn compromises the image resolution at the lung region (Fig. 5a, top panel). Therefore, we now used a modified approach for background correction (Fig. 5a, lower panel). Briefly, an adaptive moving average filter was applied on the X-ray transmission functions XTF(t), which in the first pass computed the moving average ma(t), penalised each time point with its absolute distance d(t) to ma(t) and in a second pass calculated the adaptive moving average ama(t) resulting in a baseline of the function that is less affected by the breathing events.
The exemplary relative X-ray transmission [rXTF(t)] shown for two breathing events in Fig. 5b, was then calculated as follows:
The possibility to select a smaller FOV during XLF acquisition by implementing the modified background correction method led to an improved image resolution at the lung region (Fig. 5a, lower panel). This relatively enhanced macroscopic visualization of the lung structures enabled precise selection of the aerated regions at each lung lobe.
Following background correction, the calibrated X-ray transmission over the chest region was used for quantification of in-vivo XLF. The functional parameters measured by XLF were determined from the breathing curve (Fig. 5b). The local maxima I, which denotes the time points of maximum inspiration (yellow circles, Fig. 5b) were detected. Then the time points of maximum expiration E (red circles, Fig. 5b) were identified as the closest local minima before the maxima I. To further suppress the effect of noise a moving average filter was applied to define the minima E. To split the breathing curve into an active breathing (mainly inspiration) and a passive expiration phase, an intersection point was used at a level function (grey horizontal line, Fig. 5b) resulting in the threshold point rXTF T (cyan circle, Fig. 5b). Moreover, this point also separates the background signal and noise from the breathing peaks.
Since quantitative functional parameters in the rXTF are based on the analysis of every single breath, they were measured as an average over the entire recorded data. The inspiration time (tin) was calculated as the average time between the points E and I for all breathing events recorded. Further, the length of the breathing events (L) was calculated as the average distance between the peaks I of the rXTF. The first part of the area under the curve (AUC) in the interval E to T was recorded as end-inspiration lung volume (EIV[XLF]) while the second part in the interval T to E denoted the relative X-ray transmission at end-expiration (rXTE). Lastly, to characterize the decay rate τ of the expiration phase (represented by the blue dotted line), which indirectly measures changes in the elastic recoil of the lung, an exponential decay function
was fitted at the intervals I to E.
XLF measurements for precise lung volumes
To assess the comparability of XLF based 2D measurements with the lung volumes calculated from 3D projections, we investigated the relationship between the XLF parameters that coincided with the (Tables S2, S3). We therefore compared EIV[XLF] which is described by the average AUC in the X-ray transmission function of a breathing event with . Since EIV[XLF] is mostly affected by the total air volume inhaled at the end of the inspiration phase it should be strongly related to the . To analyse the correlation between the two parameters, we again assessed the lung volumes in dead mice that were ventilated with increasing stroke volumes. Since values have a lower standard error than , they showed an improved correlation with increasing stroke volumes (Supp. Fig. S3) as compared to (Fig. 3g). EIV[XLF] also exhibited a strong positive correlation with for all three mice (30.1 g, R2 = 0,99; 28.2 g, R2 = 0.99; and 25.0 g, R2 = 0,99) (Fig. 6a). Importantly, EIV[XLF] also corresponded to the body weight of the dead mice, unlike which was found to be less sensitive to the body weight of the mice (Supp. Fig. S3). This is because EIV[XLF] is directly influenced by temporal variations in the X-ray transmission function during air exchange. EIV[XLF] hence quantifies the air volume more precisely than which determines the volume of the entire aerated lung regions including miniscule soft tissue structures.
Following the calibration of EIV[XLF] in dead mice, we applied this approach in living mice. To establish a direct correlation coefficient between the 2D (XLF) and 3D (micro-CT) datasets in living male mice, which were age and strain matched, the was normalized with the mouse body mass to the power of 1.5 (Table S3). In living mice, a strong linear positive correlation between EIV[XLF] and normalized was observed with a correlation coefficient of R2 = 0.94 (Fig. 6b). These findings revealed that lung volumes determined from XLF were precise and comparable to those measured from 3D data acquired by micro-CT. It is noteworthy that the XLF data were recorded in 34 s to determine both lung volume and function with an average X-ray dose of only 0.17 Gy. In comparison, CT data were acquired in a considerably longer acquisition time of 4.5 min and with a relatively high X-ray dose of approximately 2.7 Gy. This signifies that XLF can perform both lung volume and function measurements in substantially faster acquisition time and by using only a fraction of the X-ray dose that is necessary for CT.
Because tidal volumes cannot be derived from XLF measurements, a direct comparison of XLF and WBP measurements is not possible. Still, a correlation between EIV[XLF] and TV[WBP] was tested which resulted in a weak correlation as expected (R2 = 0.32, Fig. 6c; Table S3). Using as a standard approach, it can be concluded that EIV[XLF] is sensitive and accurate for determining lung volumes at inspiration. Overall, these findings also highlight the difficulties in deriving lung volumes from WBP measurements as compared to XLF.
Discussion
This study presents a refined XLF approach, which enables the determination of lung volumes. This was achieved by using a novel experimental set-up which facilitated correlative assessment of lung volumes in mice using XLF, micro-CT and conventional WBP. To enable the biomedical interpretation of the XLF-acquired parameters as well as their comparison with WBP and CT, XLF parameters described in Dullin et al.26 were redefined27. Our results demonstrate that lung volume derived from a simple 2D measurement (XLF) strongly correlates with the volume of the lung determined from micro-CT data.
Semi-automated segmentation of CT images is commonly used to define total lung volume in healthy and diseased mice15,19,33–37, hence we used this method in this study as a benchmark for correlation of lung volumes. However, poor resolution owing to low signal-to-noise-ratio and motion induced image blurring could limit the lung volume quantification using retrospectively gated CT38. Likewise, in the present study due to the comparably large voxel size of 40 µm and motion artefacts, it was not possible to calculate the volume of air only, instead, we measured the entire volume of the aerated regions of the lung. Moreover, changes in the morphology and compliance of the lung can occur at the early post-mortem stage, affecting the total ventilation volume reaching the lungs16. The extent of these post-mortem alterations may even be different in each individual mouse. This explains the lack of perfect correlation between and stroke volumes as well as the body weight of the dead mice. Other factors such as the relative fast stroke rate of the ventilator (40 strokes/min), leakage at the tracheal cannula and length and elasticity of the tubes can also contribute to the differences in the total air volume exchanged in mechanically ventilated mice. We used a linear increase in stroke volume which resulted in a perfect correlation with micro-CT lung volume measurements, suggesting a high reliability in measurements. Therefore, was used as a reference for our correlative approach.
For a direct correlation between the lung volumes obtained from 3D micro-CT and planar XLF, was normalized by an empirically estimated factor of body mass to the power of 1.5. In the present in-vivo study we used mice of the same age, sex and strain; hence it cannot be excluded that additional normalization parameters may be required to compensate for these factors. However, as most mouse studies are typically performed on gender and age matched cohorts, our XLF approach is generally applicable.
Most importantly, the XLF data were recorded with a significantly lower X-ray dose and about 87% shorter acquisition time when compared to CT (see “Results”). This signifies that by using only approximately 7% of the X-ray dose used in CT scans, XLF not only measures the lung function more precisely than commonly applied WBP but also determines the lung volume, a parameter that until now could only be reliably assessed by CT. This can serve as a breakthrough for facilitating the use of XLF in longitudinal studies on mouse models of lung disease with repeated lung function measurements without the risk of potential radiation damage to the lung.
XLF-defined lung volume only accounts for air volume at end-inspiration and cannot quantify end-expiration and tidal volumes. This contributed to the poor correlation of EIV[XLF] with TV[WBP] On the other hand, slight breath to breath changes in the humidity and temperature inside the chamber can largely influence the tidal volumes obtained from WBP in small animals10,39. Unfortunately, the humidity and temperature change inside the chamber could not be recorded during correlative lung function measurements as the data logger device can interfere with the X-ray transmission and more importantly changes the posture of the mouse which can affect the breathing pattern. Consequently, we could not determine a correction factor offline (see “Materials and methods”) for the WBP tidal volume which resulted in the rather weak correlation with . Moreover, the standard errors involved in the segmentation of the lung volume at each inspiratory and expiratory phase both contribute to the resulting values, therefore reducing its precision and accuracy.
In addition, WBP is complicated to perform in both pre-clinical and clinical use, and can result in significant inconsistencies due to minor technical differences in equipment, handling and surroundings40,41. Henceforth, sophisticated correlative techniques such as presented in this study are required to further test the validity of plethysmography measurements for determining tidal volumes. On the other hand, WBP for preclinical application can be performed in freely moving animals, while XLF requires isoflurane anaesthesia.
In conclusion, we present XLF as a tool for an efficient and reliable assessment of lung function and also lung volumes by successfully correlating radiographic XLF measurements with micro-CT data. Due to the high sensitivity of XLF and the ability to use standard micro-CT equipment or even a simple X-ray tube to perform these studies, XLF can easily be applied in preclinical studies. Most importantly, XLF is performed with minimal X-ray dose and acquisition time and is therefore suitable for longitudinal studies. This also provides a vital basis for clinical translation. X-ray based lung function analysis—especially for studying lung motion—has already been used, but due to X-ray dose limitations has not been routinely applied in the clinic. Fouras et al.42 developed a similar approach for low dose X-ray velocimetry for patient setup. However, to date, no clinical data from this system has been published. Our data, showing the use of low X-ray dose for lung function and volume measurements, support the notion that this technique should be further developed for patients. Since the functional and volumetric parameters measured by XLF are designed to reflect airway remodelling contributing to the airflow obstruction, XLF could be used for assessing multiple pulmonary diseases and even enable an early-stage diagnosis in case of pulmonary fibrosis and edema43,44. Ventilation settings which are currently adjusted manually by the operator and by interpreting ventilator-derived parameters could also be regularly monitored by XLF and adjusted to personalized, case-by-case settings45. Lastly, XLF may be potentially performed in non-compliant or even unconscious patients, in which e.g., spirometry typically used for lung function measurement, cannot be applied.
Material and methods
Animal experiment
Eight healthy male C57Bl/6 mice (13 weeks old) were used for in-vivo lung function measurements (Table S1). Additional three male mice were used for mechanical ventilation experiments (Table S1). The mice were kept in a temperature and humidity-controlled room with a 12 h light–dark cycle and were fed food and water ad libitum. They were euthanized by cervical dislocation. All animal procedures were performed in compliance with the guidelines of the ARRIVE, European Directive (2010/63/EU) and the German ethical laws and were approved by the administration of Lower Saxony, Germany (approval number G15.1747).
Experimental set-up
The experimental set-up was designed to perform either XLF or micro-CT with WBP in parallel (Fig. 1). The custom-made plethysmography chamber (see below) of approximately 14 cm in length and 5 cm in diameter with a capacity of 200 ml was built to fit into the gantry of the in-vivo micro-CT system (QuantumFX, PerkinElmer) allowing acquisition of a 20 × 20 mm2 FOV for XLF and micro-CT imaging. The mouse was positioned in the WBP chamber, which was then placed on the sample stage inside the micro-CT. The position of mouse and stage was adjusted to have the chest-cavity within the FOV. A piezoelectric sensor (PZT, FT31T-1.3A1-472; KEPO, Ningbo, China) was installed at the door of the micro-CT device to synchronize the data acquisitions for WBP with the acoustic signal from the door interlock system at the beginning and at the end of XLF or micro-CT recording (Supp. Fig. S1).
In-vivo XLF measurement
XLF was performed on unrestrained mice that were isoflurane-anesthetized (~ 2% isoflurane in 1 L oxygen per min) by acquiring planar cinematic X-ray images of the mice chest cavity using a low-dose in-vivo micro-CT as previously described by Dullin et al.21 and Markus et al.22. Briefly, mice were placed inside a plastic chamber (Fig. 1) and the breathing frequency of the mice was adjusted manually by changing the isoflurane concentration between 1.5 and 3% to achieve a breathing frequency of approximately 0.7 Hz. This resulted in a total of 21 breathing cycles recorded in ~ 34 s scans. XLF measurements were recorded using an X-ray tube voltage of 90 kVp and a tube current of 100 µA. The radiographs of the chest movements during the breathing process were sequenced to produce a movie, which later on was parametrized at the lung regions.
Micro-CT imaging
In-vivo 3D micro-CT scans were made immediately after combined XLF and WBP measurements on the isoflurane-anesthetized mice. Mouse lungs were scanned by applying a cardiac gating technique using the following parameters: X-ray tube voltage 90 kVp and tube current 200 µA. The radiograph projections were taken for the 360° gantry rotation for a total scan time of 270s and the entire radiograph was gated retrospectively.
XLF and micro-CT measurements in mechanically ventilated dead mice
Three male mice were euthanized using cervical dislocation technique. Tracheotomy was performed to ventilate the mice with increasing stroke volumes ranging from 0.5 to 3.0 ml and a fixed stroke rate of ~ 40 stokes/min using a MiniVent ventilator (Type 845, Harvard Apparatus, Holliston, USA). However, due to faster breathing in the ventilated mice, the breathing frequency was approximately one single breathing event per 860 ms. The XLF measurements and micro-CT imaging were performed as described for the in-vivo experiments.
Unrestrained whole-body plethysmography
We used a chamber in the so-called flow-through configuration, which utilizes the principle of a pneumotach46,47. In detail, the pressure difference between the recording chamber (220 ml) and a reference chamber (50 ml) was captured by a differential pressure sensor (DPS, Board Mount Pressure Sensor, 0–1" H2O, 20 mV, 16 VDC supply; Mfr. No: INCH-D-MV, Amphenol Cooperation, Wallingford, CT 06492). Digitization was performed with an analog–digital interface (PowerLab) and LabChart-software (ADInstruments LTD, Oxford, United Kingdom). The connectors on the lid of the chamber were used for isoflurane inlet/outlet and for connecting the DPS. A positive bias flow of 0.4–0.7 L/min was introduced using the anesthesia unit (VisualSonics VS 4112; 50% Air + 50% O2) to supply the volatile anesthetic isoflurane. The time constant of the pressure decay from the chamber system was adjusted to approximately 40 ms to match commercial chamber systems that we used in earlier papers47. The decay time constant was tested by manually obstructing the air outlet and measuring the decay of the pressure signal after releasing the obstruction (LabChart Peak Analysis Module).
The raw flow signal was bandpass filtered off-line (0.5–20 Hz), to remove movement artifacts, noise and also bias flow, and then integrated for measuring a tidal volume. We used the standard integral settings of the “Integral Channel Calculation module” of the LabChart-software (use of all data points, reset each cycle, whereby the integral is reset each time the source signal passes through zero to a positive value). Temperature and humidity inside the plethysmography chamber were measured by using a LOG32 data logger (Dostmann electronic GmBH, Wertheim-Reicholzheim, Germany).
WBP temperature and humidity measurements
CT induced temperature and humidity changes inside the plethysmography chamber could not be measured online. We therefore approximated the values for humidity and temperature correction based on separate measurements, that were performed under similar conditions to those used during correlative lung function measurements. In short, the data logger device was placed inside the chamber together with the mouse under isoflurane anesthesia and the temperature and humidity was recorded for ~ 4 min. The relative humidity of the air/O2 mix was below 3%. Relative humidity (rh) inside the chamber dropped during the recording. With the mouse inside the chamber, the rh ranged between 22.5 and 9.4% (mean 16.7%). Considering that the data logger reduces the volume in the chamber, the value inside the chamber without the data logger is expected to be even smaller. The average temperature inside the chamber with the mouse was 27 °C. Since the duration of anesthesia was below 15 min, we did not anticipate any significant change in the body temperature (36.5 °C). With an assumed atmospheric pressure of 760 mmHg (1.01308 bar), the volume adjustment factor was 11.7.
WBP data sampling and analysis
Flow signal from the pressure sensor was sampled at 1 kHz, while the PTC signal was sampled at 40 Hz. The intensity of the lung was detected by using the “Plot Z-axis Profile” function of ImageJ software48. Therefore, an ROI was set to the level of the diaphragm in expiration. Data was exported as ACSII and imported to LabChart 8.0 (ADInstruments). For removal of low-frequency intensity changes resulting from the X-ray tube, “the smoothed” intensity trace (median filter; 101 samples = 100 ms) was subtracted from the raw trace. Processed image data and the WBP data was exported to Axon PCLAMP file format and imported into IGOR Pro 6.37 (Wavemetrics, Lake Oswego, US) for resampling (1 kHz) and reimported into LabChart for final offline analysis. The peak detection module of LabChart was used to identify respiratory parameters including tidal volume (integrated flow), Tstart and Tend (10–90% of peak height) as well as the decay of the signal (τ; 100 to 0% of peak).
Data quantification and statistical analysis
The data of lung volumes were quantified using RetrospeCT (https://gitlab.gwdg.de/thomas.rittmann/retrospective_gating) which includes an iterative GPU-based reconstruction toolbox (tigre, https://github.com/CERN/TIGRE). Quantification of XLF measurements were achieved with a custom-built software xLFinal https://github.com/xPITcoding/xLFinal.git. WBP was performed with an analog–digital interface (PowerLab) and LabChart-software (ADInstruments). For statistical analysis, the paired t-test for same mean implemented in MS Excel 2013 was used with a p value of 0.05 (*) as margin for statistical significance. To measure correlation, the Pearson correlation coefficient implemented in GraphPad Prism Software version 8 was used (GraphPad Software, La Jolla California USA).
Ethical approval
All animal procedures were performed in compliance with the guidelines of the ARRIVE, European Directive (2010/63/EU) and the German ethical laws and were approved by the administration of Lower Saxony, Germany (approval number G15.1747).
Supplementary Information
Acknowledgements
We thank Sarah Garbode and Bärbel Heidrich for assistance during animal experiments.
Author contributions
C.D. F.A. and S.H. designed the research. A.K. A.M. C.D. and S.H. performed the experiments with the help from J.A. T.R. designed the software for micro-CT lung volume measurements (RetrospeCT) with assistance from C.D. and C.D. designed the XLF software (XLFinal). S.H. performed data analysis for plethysmography data. A.K. and C.D. performed data analysis for XLF and micro-CT data. A.K. prepared the figures with the input from C.D. and S.H. All authors discussed the manuscript. A.K. A.M. F.A. S.H. and C.D. wrote the manuscript.
Funding
Open Access funding enabled and organized by Projekt DEAL. AK was supported by a seed Grant of the Physics-to-Medicine Initiative Göttingen.
Data availability
All data generated or analyzed during this study are included in this published article and its supplementary information files.
Competing interests
The authors declare no competing interests.
Footnotes
Publisher's note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
These authors jointly supervised this work: Swen Hülsmann and Christian Dullin.
Supplementary Information
The online version contains supplementary material available at 10.1038/s41598-021-83319-5.
References
- 1.World Health Organization, World Health Statistics 2008. WHO, Geneva, Switzerland; https://www.who.int/whosis/whostat/2008/en/ (2008).
- 2.Hoymann HG. Lung function measurements in rodents in safety pharmacology studies. Front. Pharmacol. 2012;3:20. doi: 10.3389/fphar.2012.00156. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3.Lim R, et al. Measuring respiratory function in mice using unrestrained whole-body plethysmography. JoVE. 2014 doi: 10.3791/51755. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4.Glaab T, Taube C, Braun A, Mitzner W. Invasive and noninvasive methods for studying pulmonary function in mice. Respir. Res. 2007;8:63. doi: 10.1186/1465-9921-8-63. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 5.McGovern TK, Robichaud A, Fereydoonzad L, Schuessler TF, Martin JG. Evaluation of respiratory system mechanics in mice using the forced oscillation technique. J. Vis. Exp. 2013 doi: 10.3791/50172. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 6.Hamelmann E, Schwarze J, Takeda K, Oshiba A. Noninvasive measurement of airway responsiveness in allergic mice using barometric plethysmography. Am. J. Respir. Crit. Care Med. 1997;156:10. doi: 10.1164/ajrccm.156.3.9606031. [DOI] [PubMed] [Google Scholar]
- 7.Vanoirbeek JAJ, et al. Noninvasive and invasive pulmonary function in mouse models of obstructive and restrictive respiratory diseases. Am. J. Respir. Cell Mol. Biol. 2010;42:96–104. doi: 10.1165/rcmb.2008-0487OC. [DOI] [PubMed] [Google Scholar]
- 8.Quindry JC, Ballmann CG, Epstein EE, Selsby JT. Plethysmography measurements of respiratory function in conscious unrestrained mice. J. Physiol. Sci. 2016;66:157–164. doi: 10.1007/s12576-015-0408-1. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9.Mitzner W, Tankersley C. Noninvasive measurement of airway responsiveness in allergic mice using barometric plethysmography. Am. J. Respir. Crit. Care Med. 1998;158:340–342. doi: 10.1164/ajrccm.158.1.let1. [DOI] [PubMed] [Google Scholar]
- 10.Lundblad LKA, Irvin CG, Adler A, Bates JHT. A reevaluation of the validity of unrestrained plethysmography in mice. J. Appl. Physiol. 2002;93:1198–1207. doi: 10.1152/japplphysiol.00080.2002. [DOI] [PubMed] [Google Scholar]
- 11.Adler A, Cieslewicz G, Irvin CG. Unrestrained plethysmography is an unreliable measure of airway responsiveness in BALB/c and C57BL/6 mice. J. Appl. Physiol. 2004;97:286–292. doi: 10.1152/japplphysiol.00821.2003. [DOI] [PubMed] [Google Scholar]
- 12.Zhang Q, Lai K, Xie J, Chen G, Zhong N. Does unrestrained single-chamber plethysmography provide a valid assessment of airway responsiveness in allergic BALB/c mice? Respir. Res. 2009;10:61. doi: 10.1186/1465-9921-10-61. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13.Reynolds JS, Frazer DG. Noninvasive pulmonary function screening in spontaneously breathing rodents: An engineering systems perspective. Pharmacol. Ther. 2011;131:359–368. doi: 10.1016/j.pharmthera.2011.05.003. [DOI] [PubMed] [Google Scholar]
- 14.Vande Velde G, et al. Longitudinal micro-CT provides biomarkers of lung disease that can be used to assess the effect of therapy in preclinical mouse models, and reveal compensatory changes in lung volume. Dis. Model. Mech. 2016;9:91–98. doi: 10.1242/dmm.020321. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 15.Ruscitti F, et al. Longitudinal assessment of bleomycin-induced lung fibrosis by micro-CT correlates with histological evaluation in mice. Multidiscip. Respir. Med. 2017;12:8. doi: 10.1186/s40248-017-0089-0. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16.Lovric G, et al. Tomographic in vivo microscopy for the study of lung physiology at the alveolar level. Sci. Rep. 2017;7:12545. doi: 10.1038/s41598-017-12886-3. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 17.Porra L, et al. Quantitative imaging of regional aerosol deposition, lung ventilation and morphology by synchrotron radiation CT. Sci. Rep. 2018;8:20. doi: 10.1038/s41598-018-20986-x. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 18.Dullin C, et al. MÖNCH detector enables fast and low-dose free-propagation phase-contrast computed tomography of in situ mouse lungs. J. Synchrotron. Rad. 2018;25:565–569. doi: 10.1107/S160057751701668X. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 19.Ruscitti F, et al. Quantification of lung fibrosis in IPF-like mouse model and pharmacological response to treatment by micro-computed tomography. Front. Pharmacol. 2020;11:20. doi: 10.3389/fphar.2020.01117. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 20.Plathow C, et al. Computed tomography monitoring of radiation-induced lung fibrosis in mice. Invest. Radiol. 2004;39:600–609. doi: 10.1097/01.rli.0000138134.89050.a5. [DOI] [PubMed] [Google Scholar]
- 21.Rivina L, Davoren M, Schiestl RH. Radiation-induced lung cancers in murine models. ALC. 2014;03:38–44. doi: 10.4236/alc.2014.32006. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 22.Graves PR, Siddiqui F, Anscher MS, Movsas B. Radiation pulmonary toxicity: From mechanisms to management. Semin. Radiat. Oncol. 2010;20:201–207. doi: 10.1016/j.semradonc.2010.01.010. [DOI] [PubMed] [Google Scholar]
- 23.Iwakawa M, et al. Strain dependent differences in a histological study of CD44 and collagen fibers with an expression analysis of inflammatory response-related genes in irradiated murine lung. JRR. 2004;45:423–433. doi: 10.1269/jrr.45.423. [DOI] [PubMed] [Google Scholar]
- 24.Granton PV, et al. A longitudinal evaluation of partial lung irradiation in mice by using a dedicated image-guided small animal irradiator. Int. J. Radiat. Oncol. Biol. Phys. 2014;90:696–704. doi: 10.1016/j.ijrobp.2014.07.004. [DOI] [PubMed] [Google Scholar]
- 25.Vande Velde G, et al. Longitudinal in vivo microcomputed tomography of mouse lungs: No evidence for radiotoxicity. Am. J. Physiol. Lung Cell Mol. Physiol. 2015;309:L271–L279. doi: 10.1152/ajplung.00098.2015. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 26.Berghen N, et al. Radiosafe micro-computed tomography for longitudinal evaluation of murine disease models. Sci. Rep. 2019;9:17598. doi: 10.1038/s41598-019-53876-x. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 27.Dullin C, et al. X-ray based lung function measurement—a sensitive technique to quantify lung function in allergic airway inflammation mouse models. Sci. Rep. 2016;6:3. doi: 10.1038/srep36297. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 28.Markus MA, et al. X-ray-based lung function measurement reveals persistent loss of lung tissue elasticity in mice recovered from allergic airway inflammation. Am. J. Physiol. Lung Cell. Mol. Physiol. 2017;313:L763–L771. doi: 10.1152/ajplung.00136.2017. [DOI] [PubMed] [Google Scholar]
- 29.Eilers H, et al. Pungent general anesthetics activate transient receptor potential-A1 to produce hyperalgesia and neurogenic bronchoconstriction. Anesthesiology. 2010;112:1452–1463. doi: 10.1097/ALN.0b013e3181d94e00. [DOI] [PubMed] [Google Scholar]
- 30.Feldkamp LA, Davis LC, Kress JW. Practical cone-beam algorithm. J. Opt. Soc. Am. A. 1984;10(1118/1):1759828. [Google Scholar]
- 31.Drorbaugh JE, Fenn WO. A barometric method for measuring ventilation in newborn infants. Pediatrics. 1955;16:81–87. [PubMed] [Google Scholar]
- 32.Bartlett D, Tenney SM. Control of breathing in experimental anemia. Respir. Physiol. 1970;10:384–395. doi: 10.1016/0034-5687(70)90056-3. [DOI] [PubMed] [Google Scholar]
- 33.Murrie RP, et al. Real-time in vivo imaging of regional lung function in a mouse model of cystic fibrosis on a laboratory X-ray source. Sci. Rep. 2020;10:447. doi: 10.1038/s41598-019-57376-w. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 34.Bell RD, Rudmann C, Wood RW, Schwarz EM, Rahimi H. Longitudinal micro-CT as an outcome measure of interstitial lung disease in TNF-transgenic mice. PLoS One. 2018;13:e0190678. doi: 10.1371/journal.pone.0190678. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 35.Gradl R, et al. Visualizing treatment delivery and deposition in mouse lungs using in vivo X-ray imaging. J. Control. Release. 2019;307:282–291. doi: 10.1016/j.jconrel.2019.06.035. [DOI] [PubMed] [Google Scholar]
- 36.Gradl R, et al. Dynamic in vivo chest X-ray dark-field imaging in mice. IEEE Trans. Med. Imaging. 2019;38:649–656. doi: 10.1109/TMI.2018.2868999. [DOI] [PubMed] [Google Scholar]
- 37.Backer JWD, et al. Study of the variability in upper and lower airway morphology in sprague-dawley rats using modern micro-CT scan-based segmentation techniques. Anat. Rec. 2009;292:720–727. doi: 10.1002/ar.20877. [DOI] [PubMed] [Google Scholar]
- 38.Blocker SJ, Holbrook MD, Mowery YM, Sullivan DC, Badea CT. The impact of respiratory gating on improving volume measurement of murine lung tumors in micro-CT imaging. PLoS One. 2020;15:20. doi: 10.1371/journal.pone.0225019. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 39.Enhorning G, van Schaik S, Lundgren C, Vargas I. Whole-body plethysmography, does it measure tidal volume of small animals? Can. J. Physiol. Pharmacol. 1998;76:945–951. doi: 10.1139/y99-002. [DOI] [PubMed] [Google Scholar]
- 40.Mailhot-Larouche S, et al. Assessment of respiratory function in conscious mice by double-chamber plethysmography. J. Vis. Exp. 2018 doi: 10.3791/57778. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 41.Paton J, et al. Discrepancies between pediatric laboratories in pulmonary function results from healthy children. Pediatr. Pulmonol. 2012;47:588–596. doi: 10.1002/ppul.21592. [DOI] [PubMed] [Google Scholar]
- 42.Fouras A, et al. Novel X-ray velocimetry-based quantitative regional lung function analysis to assess risk of radiation-induced pneumonitis and pulmonary fibrosis. Chest. 2019;156:A2264–A2265. doi: 10.1016/j.chest.2019.08.315. [DOI] [Google Scholar]
- 43.Sgalla G, et al. Idiopathic pulmonary fibrosis: Pathogenesis and management. Respir. Res. 2018;19:32. doi: 10.1186/s12931-018-0730-2. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 44.Assaad S. Assessment of pulmonary edema_principles and practice. J. Cardiothorac. Vasc. Anesth. 2018;14:20. doi: 10.1053/j.jvca.2017.08.028. [DOI] [PubMed] [Google Scholar]
- 45.Silva PL, Rocco PRM. The basics of respiratory mechanics: Ventilator-derived parameters. Ann. Transl. Med. 2018;6:20. doi: 10.21037/atm.2017.12.22. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 46.Zhang B, McDonald FB, Cummings KJ, Frappell PB, Wilson RJA. Novel method for conscious airway resistance and ventilation estimation in neonatal rodents using plethysmography and a mechanical lung. Respir. Physiol. Neurobiol. 2014;201:75–83. doi: 10.1016/j.resp.2014.07.004. [DOI] [PubMed] [Google Scholar]
- 47.Hülsmann S, Mesuret G, Dannenberg J, Arnoldt M, Niebert M. GlyT2-dependent preservation of MECP2-expression in inhibitory neurons improves early respiratory symptoms but does not rescue survival in a mouse model of Rett syndrome. Front. Physiol. 2016;7:20. doi: 10.3389/fphys.2016.00385. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 48.Schindelin J, et al. Fiji: An open-source platform for biological-image analysis. Nat. Method. 2012;9:20. doi: 10.1038/nmeth.2019. [DOI] [PMC free article] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Supplementary Materials
Data Availability Statement
All data generated or analyzed during this study are included in this published article and its supplementary information files.