Introduction
Catheter-based radiofrequency (RF) ablation utilizes alternating electrical current to create a tissue lesion through the production of resistive heat and conductive heat, this latter spreading away from the point of contact. Several parameters affect lesion size: catheter tip temperature, power output, catheter diameter and orientation, the duration of ablation, and the degree of contact between the catheter and the tissue. Recently introduced catheters that can monitor the catheter–tissue contact force have been found to improve the efficiency of RF ablation in patients undergoing ablation for atrial fibrillation.1, 2, 3, 4, 5 However, only little data are available on how to monitor lesion formation during RF delivery. Recently, a new technology, the DirectSense™ tool, has been proposed that assesses the real-time effectiveness of ablation6 by means of a novel measurement of local impedance (LI) through mini-electrodes on the catheter tip of a dedicated ablation catheter (IntellaNav Mifi OI; Boston Scientific, Marlborough, MA). This system uses highly localized impedance measurements to provide insight into tissue characteristics, and may provide more useful information regarding catheter stability and lesion formation than the use of generator impedance alone.7 Here, we describe how localized impedance information can be leveraged in order to guide the successful ablation of complex arrhythmias.
Key Teaching Points.
-
•
Evaluation of local impedance through dedicated technology is useful not only in assessing effective lesion formation but also in directing radiofrequency (RF) applications to proper sites.
-
•
Local impedance could be more helpful than ablation signals in confirming the presence of the gap and in guiding the ablation strategy. Particularly, the highly sensitive detection of increased local impedance in the presence of gaps can be valuable in redo ablation procedures.
-
•
A multiparametric approach can better define the structural, electrical, and contact properties of the ablation target, such as the combined use of localized electrical and impedance information, and it is mandatory in order to avoid unnecessary RF lesions.
-
•
Ablation lesions in diseased tissue, unlike in healthy myocardium, are not always predicted by the amount of power and duration of the application.
Case report
Case 1: Spotting a gap in a case of redo pulmonary vein isolation
A 50-year-old man with a history of persistent atrial fibrillation underwent a redo procedure 2 years after his first ablation procedure. An activation map was acquired by means of the Orion™ catheter and the Rhythmia™ system (Boston Scientific) and revealed that all 4 veins had become reconnected. The propagation map (Supplemental Video 1) showed that a single gap in a wide antral circumferential ablation line reconnected both left veins. When the ablation catheter (IntellaNav Mifi OI) was positioned at the corresponding location, the fragmented signals due to gap conduction into the veins were barely visible on ablation traces and were completely different from the Orion signals previously recorded at the same location (Figure 1, left panel), suggesting either that the catheter tip was not exactly on the gap or that a large cancellation effect had occurred because of the wider field of view of the ablation electrodes than that of the Orion catheter one.8 LI values at the same location showed an impedance value greater than 120 ohms (Figure 1, right panel), which was more compatible with the presence of surviving myocardial fibers in a gap region than with nearly unexcitable tissue, as suggested by signals on the ablation traces (LI values were around 88 ohms in the blood and around 95 ohms at lines of block in this patient). We applied RF energy and the veins became isolated when the LI drop reached 20 ohms (Supplemental Video 2).
Figure 1.
Left: The propagation map acquired with the Orion™ catheter (Boston Scientific, Marlborough, MA) revealed a single gap (dark red front) reconnecting both left veins. The electrograms detected by the Orion catheter at the gap location were highly fragmented and well represented. Right: The tip of the ablation catheter is at the gap location, as indicated by the white halo on the activation map. The real-time signal graph on the left shows both Orion signals (with far-field and pulmonary vein components) and ablation traces. Conventional ablation traces show no signals (white traces in the red rectangle) while mini-electrode tracings (yellow traces in the red rectangle) show very tiny and barely visible signals that are completely different from those previously detected by the Orion catheter at the same spot. The red rectangle on the right of the panel shows the real-time local impedance (LI) value when the catheter tip was at the location indicated by the white halo. This impedance value is more compatible with a conduction gap than with unexcitable tissue.
Case 2: Choosing the best ablation approach to tricuspid annular premature ventricular contractions
A 19-year-old man with premature ventricular contractions (PVC) was referred to our electrophysiology lab after a previous failed ablation procedure in another center. High-resolution mapping with the Orion catheter revealed that the PVC arose slightly anterior to the His bundle region. Figure 2A and 2B shows the Rhythmia activation map of this tricuspid annular PVC; the red spot indicates the ablation target (earliest ventricular signal during PVC) and yellow tags indicate the position of the His bundle. On approaching the parahisian region with the ablation catheter, we were never able to obtain a baseline LI value greater than 88 ohms, which was quite close to the blood LI value in this patient (around 80 ohms). Nevertheless, we decided to apply RF energy, since we were “visually” on the red spot of the map, and local signals were very early. We obtained a very low impedance drop, and the PVC did not disappear (Supplemental Video 3 and Figure 2A). Considering the young age of the patient and the high-risk location of the ablation target, we decided to create additional RF lesions only if the baseline LI value was greater than around 100 ohms, since these values were easily achieved with good contact in the adjacent right ventricle in this patient. Despite adopting a superior approach to the parahisian region or using a steerable sheath, we could not obtain a higher baseline LI value. Only by looping both the ablation catheter and the steerable sheath into the right ventricle apex and approaching the tricuspid valve by means of a “reverse” curve did we finally achieve a baseline LI value of around 115 ohms. RF applications at the same red spot as in the previous application immediately led to PVC disappearance (Figure 2C and Supplemental Video 4).
Figure 2.
In each panel, the red spot on the activation map indicates the ablation target (earliest ventricular signal during premature ventricular contraction [PVC] mapping) while yellow tags indicate the position of the His bundle. A: A lower baseline local impedance (LI) value predicts a low impedance drop during unsuccessful ablation lesion. The orange value in the red rectangle (-0.8 ohm) represents the real-time LI drop calculated from start of ablation. B: On using a different approach (“reverse curve”), the ablation tip comes into contact with the ablation target, yielding a higher LI value, which is suggestive of better electrical coupling with the tissue and hence greater potential for resistive heating. With this approach, the baseline LI value is 120 ohms, as indicated in the red rectangle. C: Radiofrequency application with the reverse curve approach yielded a greater impedance drop (orange value, -11 ohms, in the red rectangle), causing PVC to disappear.
Case 3: High-power ablation in diseased left atrial tissue
A 60-year-old woman with rheumatic mitral valve disease was referred to our lab for RF ablation of atypical atrial flutter of 2 different morphologies. High-resolution mapping revealed that the 2 arrhythmias were clockwise and counterclockwise perimitral flutters with a critical isthmus at a very diseased anterior wall. Since it was difficult to achieve good LI drops by applying standard power, high-power (50 watt) RF lesions were created along the anterior line connecting the right superior vein to the anterior mitral annulus. Once the counterclockwise arrhythmia had been interrupted, the clockwise form was spontaneously induced when 1 application remained to complete the ablation line. Although the baseline LI was low at the remaining gap, signals on the mini-electrodes confirmed contact with a diseased tissue rather than with blood6; we therefore continued to apply 50 W. Although flutter interruption (Figure 3, left panel, and Supplemental Video 5) suggested a good lesion, a very low LI drop (only 2 ohms) did not. A validation map of the anterior line during pacing from the left appendage revealed a gap exactly where the previous ablation had failed to achieve a good LI drop (Figure 3, right panel). We decided to close the gap at an adjacent location, where an LI drop of at least 15 ohms could be achieved.
Figure 3.
Left: The last remaining gap in the anterior ablation line is closed, while the clockwise form of the perimitral flutter is spontaneously induced. The presence of atrial signals detected by mini-electrodes (yellow traces in the red rectangle on the left) in combination with a low local impedance value (75 ohms in the red rectangle on the right) indicates that the catheter tip is in contact with diseased tissue. Despite applying 50 W, a very low impedance drop is achieved (orange value, -1 ohm, in the red rectangle on the right). Right: The validation map, acquired with the Orion™ catheter (Boston Scientific, Marlborough, MA) while pacing with the ablation catheter from the left atrial appendage, indicates a conduction gap at the exact location where previous high-power application had achieved a low impedance drop.
Discussion
With regard to emerging technologies, several knowledge gaps have been identified, including how to determine the best approach to safely achieving successful ablation in complex procedures, and a large amount of research is being focused on developing new tools that allow the creation of permanent transmural lesions.9 LI is a direct measure of the resistive load on the ablation catheter and is a surrogate for the surface area of myocardium covered by the distal electrode10; its real-time evaluation is helpful in order to precisely assess the electrical contact of the catheter and tip stability. It also helps to distinguish catheter–tissue contact from non-contact10 and might be an adjunct to fluoroscopy, local electrograms, and 3D maps in assessing tissue contact in low-voltage areas.7 The cases presented are examples of how DirectSense capabilities are useful not only in assessing effective lesion formation but also in directing RF applications to proper sites. The first case showed that information derived from LI was more helpful than ablation signals in confirming the presence of the gap and in guiding the ablation strategy. Fractionated electrograms at gap locations may be completely filtered out by conventional ablation catheters (20% of gaps could be missed by a 3.5-mm-tip ablation catheter)11 and, although signal resolution is improved by adding mini-electrodes at the tip, the best mini-electrode mapping resolution can still filter out some elements that would be detected by ultra-high-resolution catheters.12 For this reason, the highly sensitive detection of increased LI in the presence of gaps can be valuable in redo ablation procedures. In the second case described, LI information was important in achieving complete and durable success, as the high LI values at the resulting ablation spots indicated a close catheter–tissue interface, a large resistive load, and therefore the ability to create significant resistive heating. This example demonstrates that multiparametric approaches that can better define the structural, electrical, and contact properties of the ablation target, such as the combined use of localized electrical and impedance information, are mandatory in order to avoid unnecessary RF lesions. Our last case perfectly highlights a puzzling conundrum that has recently emerged in the literature: ablation lesions in diseased tissue, unlike in healthy myocardium, are not always predicted by the amount of power and duration of the application.13,14 In this case, LI revealed that no lesion was being created in the diseased tissue during the high-power application, thus prompting us to choose another ablation strategy. As contact-force sensing technology with impedance sensing was not available, we could not speculate the role of contact force in obtaining safe and effective lesions beyond the role of LI in our cases series.
Conclusion
LI is a helpful new parameter that can monitor RF lesions and guide ablation strategies in complex procedures.
Footnotes
Disclosures: F. Maddaluno and M. Malacrida are employees of Boston Scientific. No other conflicts of interest to declare. Funding Sources: No extramural funding was used to support this study.
Supplementary data associated with this article can be found in the online version at https://doi.org/10.1016/j.hrcr.2020.11.004.
Appendix. Supplementary data
Case 1 - Spotting a gap in a case of redo pulmonary vein isolation. The propagation map showed that a single gap in a wide antral circumferential ablation line reconnected both left veins map.
Case 1 - Spotting a gap in a case of redo pulmonary vein isolation. Application of RF energy and the veins became isolated when the LI drop reached 20 ohm.
Case 2 - Choosing the best ablation approach to tricuspid annular PVC. Application of RF energy at the red spot of the map, obtaining a very low impedance drop, without PVC disappearance.
Case 2 - Choosing the best ablation approach to tricuspid annular PVC. RF applications at the same red spot as in the previous application immediately led to PVC disappearance.
Case 3 - High-power ablation in diseased left atrial tissue. Although the baseline LI was low at the remaining gap, signals on the mini-electrodes confirmed contact with a diseased tissue rather than with blood; we therefore continued to apply 50 W. Although flutter interruption suggested a good lesion, a very low LI drop (only 2 ohms) did not.
References
- 1.Reddy V.Y., Shah D., Kautzner J. The relationship between contact force and clinical outcome during radiofrequency catheter ablation of atrial fibrillation in the TOCCATA study. Heart Rhythm. 2012;9:1789–1795. doi: 10.1016/j.hrthm.2012.07.016. [DOI] [PubMed] [Google Scholar]
- 2.Marijon E., Fazaa S., Narayanan K. Real-time contact force sensing for pulmonary vein isolation in the setting of paroxysmal atrial fibrillation: procedural and 1-year results. J Cardiovasc Electrophysiol. 2014;25:130–137. doi: 10.1111/jce.12303. [DOI] [PubMed] [Google Scholar]
- 3.Stabile G., Solimene F., Calò L. Catheter-tissue contact force for pulmonary veins isolation: a pilot multicentre study on effect on procedure and fluoroscopy time. Europace. 2014;16:335–340. doi: 10.1093/europace/eut262. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4.Sigmund E., Puererfellner H., Derndorfer M. Optimizing radiofrequency ablation of paroxysmal and persistent atrial fibrillation by direct catheter force measurement-a case-matched comparison in 198 patients. Pacing Clin Electrophysiol. 2015;38:201–208. doi: 10.1111/pace.12549. [DOI] [PubMed] [Google Scholar]
- 5.Lee G., Hunter R.J., Lovell M.J. Use of a contact force-sensing ablation catheter with advanced catheter location significantly reduces fluoroscopy time and radiation dose in catheter ablation of atrial fibrillation. Europace. 2016;18:211–218. doi: 10.1093/europace/euv186. [DOI] [PubMed] [Google Scholar]
- 6.Martin C.A., Martin R., Gajendragadkar P.R. First clinical use of novel ablation catheter incorporating local impedance data. J Cardiovasc Electrophysiol. 2018;29:1197–1206. doi: 10.1111/jce.13654. [DOI] [PubMed] [Google Scholar]
- 7.Segreti L, De Simone A, Schillaci V, et al. A novel local impedance algorithm to guide effective pulmonary vein isolation in AF patients: preliminary experience across different ablation sites from the CHARISMA Pilot Study. J Cardiovasc Electrophysiol. https://doi.org/10.1111/jce.14647. 2020 Jul 1. [DOI] [PubMed]
- 8.Münkler P., Gunawardene M.A., Jungen C. Local impedance guides catheter ablation in patients with ventricular tachycardia. J Cardiovasc Electrophysiol. 2020;31:61–69. doi: 10.1111/jce.14269. [DOI] [PubMed] [Google Scholar]
- 9.Al-Khatib S.M., Benjamin E.J., Buxton A.E. Research needs and priorities for catheter ablation of atrial fibrillation: A report from a National Heart, Lung, and Blood Institute virtual workshop. Circulation. 2020;141:482–492. doi: 10.1161/CIRCULATIONAHA.119.042706. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10.Sulkin M.S., Laughner J.I., Hilbert S. Novel measure of local impedance predicts catheter-tissue contact and lesion formation. Circ Arrhythm Electrophysiol. 2018;11 doi: 10.1161/CIRCEP.117.005831. [DOI] [PubMed] [Google Scholar]
- 11.Masuda M., Fujita M., Iida O. The identification of conduction gaps after pulmonary vein isolation using a new electroanatomic mapping system. Heart Rhythm. 2017;14:1606–1614. doi: 10.1016/j.hrthm.2017.08.016. [DOI] [PubMed] [Google Scholar]
- 12.Mantovan R., Corò L., Allocca G., Sitta N., Rivetti L., Ricarda Marinigh R. How small could a detectable reentrant circuit be in a localized microreentrant tachycardia? HeartRhythm Case Rep. 2020;6:222–225. doi: 10.1016/j.hrcr.2020.01.003. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13.Barkagan M., Leshem E., Shapira-Daniels A. Histopathological characterization of radiofrequency ablation in ventricular scar tissue. JACC Clin Electrophysiol. 2019;5:920–931. doi: 10.1016/j.jacep.2019.05.011. [DOI] [PubMed] [Google Scholar]
- 14.Tofig B.J., Lukac P., Nielsen J.M. Radiofrequency ablation lesions in low-, intermediate-, and normal-voltage myocardium: an in vivo study in a porcine heart model. Europace. 2019 doi: 10.1093/europace/euz247. pii:euz247. [DOI] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Supplementary Materials
Case 1 - Spotting a gap in a case of redo pulmonary vein isolation. The propagation map showed that a single gap in a wide antral circumferential ablation line reconnected both left veins map.
Case 1 - Spotting a gap in a case of redo pulmonary vein isolation. Application of RF energy and the veins became isolated when the LI drop reached 20 ohm.
Case 2 - Choosing the best ablation approach to tricuspid annular PVC. Application of RF energy at the red spot of the map, obtaining a very low impedance drop, without PVC disappearance.
Case 2 - Choosing the best ablation approach to tricuspid annular PVC. RF applications at the same red spot as in the previous application immediately led to PVC disappearance.
Case 3 - High-power ablation in diseased left atrial tissue. Although the baseline LI was low at the remaining gap, signals on the mini-electrodes confirmed contact with a diseased tissue rather than with blood; we therefore continued to apply 50 W. Although flutter interruption suggested a good lesion, a very low LI drop (only 2 ohms) did not.



