Abstract
High-resolution imaging is essential for analysis of the steps and way stations of cargo transport in in vitro models of the endothelium. In this study, we demonstrate a microfluidic system consisting of two channels horizontally separated by a cell-growth-promoting membrane. Its design allows for high-resolution (down to single-molecule level) imaging using a high numerical aperture objective with a short working distance. To reduce optical aberrations and enable single-molecule-sensitive imaging, an observation window was constructed in the membrane via laser cutting with subsequent structuring using 3D multiphoton lithography for improved cell growth. The upper channel was loaded with endothelial cells under flow conditions, which showed polarization and junction formation. A coculture of human vascular endothelial cells with pericytes was developed that mimics the blood–brain barrier. Finally, this dual channel microfluidics system enabled 3D localization microscopy of the cytoskeleton and 3D single-molecule-sensitive tracing of lipoprotein particles.
Keywords: microfluidics, 3D multiphoton lithography, single-molecule imaging, blood-brain barrier, endothelial cells, 3D particle tracking
The development of reliable in vitro models for human blood barriers, which allow for monitoring of drug delivery processes or processes occurring in tissue disorders, is highly desirable. Additionally, faithful mimicking of these interfaces provides us the opportunity to study transport without the use of human/animal in vivo models, which are limited by ethical concerns. Accessibility to in vitro models facilitates tuning of microenvironmental conditions and understanding the biological, biochemical, and biophysical effects of chemo-physical treatment.1−3 The majority of these in vitro models are based on a microfluidic platform, which introduces flow dynamics into the biological system. Cells either form a growing layer on top of a 2D surface or are used as a barrier between two microfluidic channels.4−10 These cellular constructs facilitate the formation of vascularization networks directly on the microfluidic chip which can be generated either via endothelial-lined patterned channels or via self-assembled networks.11 Fluorescence imaging is frequently used as the primary measurement technique for standard microfluidic systems that are used to observe cellular dynamics or the transport of biomolecules across barriers.12,13 Microfluidic systems have been used to characterize cell reactions after treatment with exosomes, using time-lapse confocal microscopy or to quantify mitochondrial dynamics, morphology, and nanoscale mitochondrial protein distributions using live cell and super-resolution microscopy.12
Multiphoton lithography (MPL) has recently been used to structure arbitrary nano/microscopic features into microfluidic systems.14 MPL is suitable for 3D structuring with lateral and axial resolutions of ∼200 and 500 nm, respectively.15 The high flexibility in feature sizes, material properties, and 3D writing capability makes MPL a well-suited method for structuring mechanical barriers, which can also be used for cell support.
However, a microfluidic chip for mimicking different kinds of blood barriers with the feature of providing for single-molecule-sensitive imaging in 3D is still lacking. Most of the current blood–brain barrier (BBB) models are based on the transwell system,16,17 where live optical imaging is not possible at all or likely only with low NA (i.e., high working distance) objectives. Under such conditions, neither single-molecule-sensitive nor 3D imaging is possible at the desired resolution. In the existing microfluidics,18,19 3D tracking is challenging due to the thickness of the channels and the membrane’s porosity. Moreover, thick membranes reduce the contact between cells and introduce optical aberrations, consequently reducing achievable imaging resolution. The platform proposed in this paper allows for 3D imaging and tracking of single molecules that move within or along cells forming the blood barrier. Our microfluidic system consists of two channels, which are horizontally separated by a membrane, enabling cocultivation. To enable transport across the membrane, we have introduced a window produced via laser cutting that was subsequently filled with a 3D polymer into which a complex, defined geometry of holes and support structures was engineered. The combination of subtractive and additive manufacturing allows creation of complex structures and enhancement of the functionality of the final device.20−22 We were able to show that when endothelial cells (ECs) were introduced into the system they became polarized on the membrane and developed interendothelial junctions. By coculture of ECs with pericytes, a model of the BBB23 was demonstrated. In summary, our microfluidic chip enables 3D single-molecule-sensitive imaging using a high numerical aperture (NA) objective with a short working distance. In a showcase experiment using CD34+-ECs in the microfluidics device we were able to perform 3D localization microscopy of the cytoskeleton and 3D single molecule tracing of high-density lipoprotein (HDL) particles.
Results and Discussion
The primary goal of this study was to develop a microfluidic system that mimics cellular blood vessel barriers (Figure 1a) providing a means to study the transport of biomolecules (e.g., lipoprotein particles, extracellular vesicles, but also lipids and hormones) across the barrier.24 The layout of this dual channel microfluidics system allows for cocultivation of various cell-types and imaging of their cellular processes (e.g., extracellular vesicles uptake, cell–cell interactions) via single-molecule-sensitive fluorescence microscopy. In order to introduce a cell barrier between the microfluidic channels, a MPL structure was introduced which enables an application-specific optimization of the barrier (i.e., observation window) (see Figure 1b,c and Figure S1).
Structure of the Microfluidic Device
Our aim was to design a microfluidic channel with two flow channel compartments, separated by a membrane suited for cell cultivation. The channel was designed to provide for nanoscopic imaging of single molecule transport to living cells. In order to use objective lenses with a NA > 1.4 and a working distance of 300 μm, the height of the microfluidics compartment above the imaging window (i.e., cell membrane interface) was adjusted to ∼250 μm. A cross-section of the microfluidic chamber is depicted in Figure 1b. The microfluidic chamber consists of a coverslip substrate, a spacer that forms the bottom channel, a 30 μm thick polyethylene terephthalate (PET)-foil that separates the lower and upper channels, and spacers which define the upper channel including inlets. The top was covered with hard acrylic glass which provided mechanical stability and rigidity for the whole microfluidic device. A detailed description of the layout can be found in Figure S1a. The PET-foil has proven to be well suited as a cell substrate.5,25,26 However, although transparency is provided, distortion of the fluorescence signal was observed; moreover, such impermeable PET-foil does not allow the study of diffusion of single molecules and ions that move across the foil. To permit diffusion and reduce distortions in optical measurements, custom-defined, and shaped holes were cut into the PET-foil by femtosecond (fs)-long laser pulses. Here, without loss of generality, a star-shaped geometry has been chosen (Figure 2b, Figure S1b depicts alternative forms). Typically, three to six starlike holes with a characteristic length of 170 μm (tip-to-tip distance) were cut in one microfluidic device.
Thereafter, a 3D grating was created that acts as a barrier between the two compartments. The 3D grating inside the star-shaped holes was structured using multiphoton lithography, allowing for high flexibility in the geometry and polymer composition of the structure. Furthermore, writing of a grating-shaped cell support reduces the contact area of the cells with the substrate (i.e., minimizes the artificial contact area) and simultaneously increases the direct interaction area between cells. For 3D printing, we used a biocompatible BisSR-photoresist,27 which allows for structuring with sub-200 nm feature sizes. The photoresist was placed on top of the hole, and MPL in dip-in configuration was performed (Figure 2c). In the dip-in configuration, the photoresist is used as the immersion medium. The final structure consisted of a 100 μm × 100 μm grid with a customizable periodicity. The critical steps for production of the imaging window are shown in Figure 2b,c: a laser cut star-shaped hole in the PET-foil and the same hole filled with a suspended 3D polymeric grating. The flexibility of MPL structuring allows facile tuning of grid periodicity and shape at both sides of the hole and enables cell type dependent customization. In the transwell based BBB model, it was shown that increase of the membrane pore size from 0.4 to 3 μm substantially increases the contact between cells and resulted in greater barrier strength.16 Here, we employ a grating with an upper layer periodicity of 6 μm and a lower layer periodicity of 12 μm, filling the star-shaped hole (Figure 2d). The layers are interconnected by 30 μm high polymeric structures. The larger mesh size of the bottom grid facilitates penetration of cells. In contrast, the upper mesh is sufficiently dense to inhibit ECs invasion.
Additionally, flexibility in terms of functionalization of the polymeric structures is shown in Figures S1d,e. The figure shows grids with a 12 μm periodicity, made of protein-repellent pentaerythritoltriacrylate/ethylene glycoldiacrylate (PETA–PEGDA) polymer28−30 and ring substructures made of biotinylated PETA (Figure S1d) or pentaerythritol triacrylate/2-carboxyethyl acrylate (PETA-CEA) monomers (Figure S1e). Both polymers, employed for the ring substructures, can be used to immobilize the desired proteins.29,30 To show protein binding selectivity, we incubated the entire structure with Alexa647-streptavidin. As shown in the fluorescence image of the polymer structure (Figure S1f), streptavidin efficiently bound to the biotinylated PETA containing the ring substructure and did not bind to the protein repellent rectangular grid consisting of PETA–PEGDA. Such selective polymer functionalization28,30−32 enables local immobilization of biomolecules inside the imaging window and could potentially be used for cell-type-specific stimulation.29,33,34
The optical properties of the polymer were optimized to minimize autofluorescence and maximize transparency.27 This enabled fluorescence microscopy measurements, even at a single molecule level (Table S1). The biocompatibility of the foil and polymers used for the grid structuring facilitates cell seeding and culturing within the channels. We used a syringe pump to introduce a constant flow of 0.2 mL/h in the upper channel. In the lower channel, medium was replaced every 2 h or if the pH change (i.e., color change of phenol red) was visually observed. We used a two-cell-type system wherein we seeded primary endothelial cells (CD34+-ECs), or alternatively human umbilical vein endothelial cells (HUVECs) to form an endothelial cell layer in the top channel. The flow-speed was adjusted to 70 μm/s, which is in the same order of magnitude as natural flow-speeds (300 μm/s for capillaries35).
Optimization of cultivation and imaging. In the next step, cell cultivation inside the microfluidic channel was optimized. Cells were grown for 3–5 days under a constant, directional nutrient flow of 0.2 mL/h. This produced cells that were partially aligned with the direction of flow (arrow in Figure 3a) and which formed a confluent monolayer complete with interendothelial junctions on both the foil and the grid structures. To characterize the cell-layer, fluorescence microscopy was used. Cell positions were visualized using a Hoechst 33342 stain for the nuclei (blue). We visualized the cellular network by labeling cell–cell junctions with an Alexa647-anti-CD31/PECAM-1 antibody (Figure 3b). A two-color fluorescence image in the right panel of Figure 3b demonstrates a dense cell layer, indicated by the cell–cell junctions and the labeled nuclei. Overall, the cell nuclei showed no signs of apoptosis, and cells on the polymer grid formed a dense monolayer, with interdigitated contacts to neighboring cells. Generally, morphology of the cells on the grid was indistinguishable from cells on the PET substrate, demonstrating once again the excellent biocompatibility of the coated polymer grid, which readily supports the growth of primary endothelial cells for longer time periods.
ECs exposed to flow typically exhibit a cell polarization36 resulting in a change in the protein distributions between the apical and basal membranes. Here, we used CD146/MCAM and CD31/PECAM-1 proteins as polarization markers.37 It is expected that, upon polarization, CD31 will have a higher apical density38,39 and that CD146 will show a more homogeneous distribution.40,41 For imaging, we stained ECs inside the microfluidic channel with Alexa647-conjugated anti-CD146 (red) antibodies and Alexa488-conjugated anti-CD31 (blue) antibodies (Figure 3c). To demonstrate the polarization of ECs, confocal imaging of the endothelial cells on top of the grid was performed. As expected the CD146 signal was homogeneously distributed throughout the cell membrane, while the CD31 signal was preferentially localized on the apical side of the ECs, highlighting the interendothelial junctions. XZ-axis-cross sections with their respective profiles are shown in Figure 3c. Similar results, in terms of confluence and polarization, were also obtained for HUVECs and are presented in the Supporting Information (Figure S2).
Establishment of a Coculture System
The designed microfluidic device is well-suited for cocultures of two cell types based on our ability to tune the periodicity and geometry of the 3D grid as well as the hole size for tailoring the cell–cell interaction area. We used ECs and bovine pericytes cultured on the upper and lower sides of the grid, respectively (Figure 3d), to mimic the BBB;23 however, any other desired cell combination can be cocultured. The endothelial cells were cultured for 3 days under flow conditions in the upper channel and then cocultured with bovine pericytes42 in the lower channel. Direct interactions between the cocultivated ECs and the pericytes form a BBB like structure23 (see Figure 1a). For visualization, the pericytes were stained with Nuclear Green dye (green). To visualize the EC network, the nuclei of fixed cells were stained with Hoechst 33342 (blue), and the cell junctions were stained with Alexa647-conjugated anti-CD31/PECAM-1 antibody (red). The right panel in Figure 3d shows a fluorescence image of the ECs and bovine pericytes cocultured in the microfluidic device. Human vascular ECs were captured on the top of the grid, and the pericytes were imaged on a focal plane below the grid. The images of the separate color channels are depicted in Figure S3. This result demonstrates the applicability of the microfluidic device for imaging of cocultivated cells.
Single-Molecule Imaging within the Dual-Channel Microfluidics
Here, we show that the transport of molecules across the mimicked cellular barrier can be followed with single-molecule sensitivity. In Figure 4a, we show a bright-field image of a scaffold populated with ECs. To prove the microfluidic device’s feasibility for localization microscopy, we stained the cytoskeleton of these CD34+-ECs with Alexa Fluor 647 conjugated phalloidin. The astigmatism of the signal was used to determine the axial arrangement of the cytoskeleton.43,44 The refractive index mismatch between the polymer structure and the cells was adjusted using a glucose buffer (n ∼ 1.42). Figure 4b shows a super-resolved (3D-dSTORM45) image of the cytoskeleton. Detailed analyses of the super-resolved images revealed that the signal-to-noise-ratio (SNR) in the suspended areas (SNR ∼ 13) is higher when compared to areas where the cells lie directly on top of the grid (SNR ∼ 10). This difference was minimized due to the low autofluorescence of the chosen polymer; nevertheless, the advantage of intentional polymer-free volumes areas in the imaging area is notable.
In a second experiment, we traced fluorescently labeled HDL particles across human vascular ECs cultivated under flow conditions. Particles were transported along the cell membrane. The observed movement was consistent with the compartmentation of HDL-like particles between liquor and blood circulation.46,47 We were able to trace the movement of individual HDL particles in 3D and determine their diffusion constants. Projections of the trajectories of two particles are shown in Figure 4c. Two traces are explicitly depicted in Figure 4d,e, the black line represents the 3D trace, and the green, blue, and red lines represent the projections in xz, yz, and xy planes, respectively. The data indicate that, consistent with previous results,48 an average diffusion constant of 3 × 10–3 ± 2 × 10–3 μm2/s (assuming a Brownian motion model, 12 cells) was obtained. The analysis was performed using Trackpy49 and msdanalysis software.50 For comparison, cells were also incubated with extracellular vesicles (EVs) originating from HEK293T cells. For fluorescent imaging, CD63 proteins in the EVs were labeled with green fluorescent proteins. These additional diffusion analyses are shown in Figure S4. The data indicate that EVs are taken up by the cells.
Conclusion
The platform presented in this paper provides for 3D imaging and tracking of single HDL particles moving across cells grown on an artificial membrane. We show a microfluidic system consisting of two channels, which are horizontally separated by a cell growth-promoting membrane. High-resolution imaging is enabled by observation windows introduced into the membrane (we were not able to image single molecules through the PET membrane, the higher background induced by the membranes is clearly visible in Figure S2). These windows were produced via laser cutting and subsequently filled with 3D polymer structures. As proof of concept, we seeded our microfluidic chambers with ECs and observed cellular polarization and interendothelial junction formation under flow conditions. A coculture of ECs with pericytes imitating blood–brain barrier formation was demonstrated. In contrast to other available systems,16,17 our microfluidic model allows coculture, appliance of flow, live high-resolution microscopy as well as analysis of transport processes at a single molecule level. In a showcase demonstration, 3D localization microscopy of the cytoskeleton and 3D single molecule tracking of HDL particles was performed which showed that HDL particles do not pass through the cell layer. A diffusion analysis of the HDL particles revealed that on average, the particles’ positions varied by x = 3.4 ± 2.8 μm, y = 2.4 ± 1.4 μm, and z = 0.26 ± 0.11 μm, indicating that the HDL particles were not entering cells but rather diffusing along the membrane curvature. In comparison, we were able to show that EVs were taken up by the cells. The average x,y,z-displacements for EVs in ECs were 0.8 ± 0.4 μm, 0.8 ± 0.3 μm, and 0.5 ± 0.1 μm, respectively. These data indicate that EVs move across the membrane and inside the cell. Thus, we conclude that EVs were taken up by the cells in contrast to HDL particles. This result shows the selectivity of the barrier via transferring some molecules while blocking others. Taken together, HDL particles were moving only in two dimensions (on the surface of the cells), indicating that these cells formed a valid blood–brain barrier model. Note that it is well-described that lipoproteins do not pass the blood–brain barrier; instead, the brain–liquor has a unique set of HDL like lipoproteins.51
Materials and Methods
Fabrication
Lithography System
A scheme of the Workshop of Photonics (WOP, Lithuania) multiphoton lithography instrument is shown in Figure 2a. The instrument is equipped with an ultrashort pulsed laser (CARBIDE, 1 MHz repetition rate, 290 fs pulse duration, Light Conversion, Lithuania) with two available wavelengths. The first harmonic of the laser output (1030 nm) is used for direct laser drilling and the second harmonic output (515 nm) for multiphoton lithography. All stated intensities are peak intensities in the focal plane, calculated from the power measured in front of the objective, taking the transmission of the objective and a diffraction-limited spot area into account. Writing was performed by a combination of sandwich and dip-in configuration.52 A 3-axis stage (AEROTECH Nanopositioner, USA) was used for sample motion. Optionally, faster lateral writing was performed by a galvanometric mirror scanner (AEROTECH, USA).
Direct Laser Cutting
For laser ablation, the first harmonic (1030 nm) of the WOP system laser was used as shown in Figure 2b, left. The laser beam was focused with a 50× magnification air objective (Mitutoyo, Japan) onto the biocompatible PET foil (biaxially oriented polyethylene terephthalate foil, Bleher). Cutting was performed by moving the focal spot with a velocity of 80 μm/s and executed in two steps: first a peak intensity of 1.65 TW/cm2 was used for the first cutting, and then the intensity was increased to 1.84 TW/cm2 and the was beam shifted closer to the center by 5 μm. After a completed run, the foil was lifted by 2.5 μm and the cutting procedure was repeated until the inset was cut out. If necessary, the inner discarded part was removed mechanically by means of an ultrasonic bath. To protect the optic from the evaporating material, the PET-foil was sandwiched in-between two glass slides.
Multiphoton Lithography
In order to close the hole with a cell-supporting scaffold, the hole was covered with a droplet of resin (see Figure 2b, right-hand side). Writing was carried out by sandwich writing (two spacers with 170 μm and a glass lid with 170 μm) using the 50× air objective as described in the previous section and/or in dip-in configuration (Zeiss, 63×, NA 1.25). Due to the lower NA, sandwich writing enables a higher throughput, while dip-in writing has the benefit of small feature sizes and a higher intensity. The photoinitiator of the resin was excited by using the second harmonic (515 nm) of the WOP system. For 2D or 3D structuring, the focus of the laser was traced along a path by 3D stage and/or moving of a galvanometric mirror. To improve attachment between the structure and the foil, a frame (35 μm wide and 30 μm thick) is written into the hole, usually by the aid of sandwich writing. The hole was closed optionally by a 2D grid or 3D scaffold. The 2D grid consists of quadratic squares with a grid constant of 6 μm and a commissioned height of 2 μm. Each square was exposed twice, making the resulting bar thickness dependent on the used writing speed of 200 μm/s and writing peak intensity of 1.05 TW/cm2. The 3D scaffold was generated by the use of a STL-file. The 3D model consisted of an upper grid (grid constant of 6 μm, bar thickness and height of 1.5 μm) separated to a lower grid by 30 μm long spacers with the same bar dimension but a higher grid constant of 12 μm. For structuring, a writing speed 1 mm/s and writing peak intensity of 1.58 TW/cm2 was used.
Photoresist Formulations
The biocompatible photoresist BisSR was a mixture of 29% of Bis-GMA (Esschem Europe, England) and 71% ethoxylated bisphenol A dimethacrylate (SR348C) (Sartomer, France), and prepared as published.27 CEA/PETA was composed of 10 wt.% 2-carboxyethyl acrylate (Sigma-Aldrich, Germany) mixed with PETA. The biotinylated resin consisted of 1 wt.% biotin arylate (Nanocs, Inc., USA) and 99% PETA. The protein repelling resins were made of 20% PEGDA (Sigma-Aldrich, Germany) and 80% PETA as recently reported.29 A photo initiator 1 wt.% of Bis(2,4,6–trimethylbenzoyl)–phenylphosphineoxide (IC 819) was used for all resins. All monomers were stirred with a magnetic stirrer until homogeneously mixed.
Assembly of the Microfluidic Chip
Two microfluidic channels were sandwiched between a microscope coverslip (Menzel 24 mm × 50 mm #1 SPEZIAL, Thermo Scientific) and an acrylic glass piece. While the microscope coverslip enabled observation of the sample area with high magnification/high NA objectives, the acrylic glass, with openings for tube connectors, made the microfluidic chip rigid and easy to handle. An industrial laser cutting device was used to cut the dimensions and openings of the acrylic glass out of a 3 mm thick sheet. The building blocks with the microfluidic channel structure were cut using a craft cutter (Silhouette Portrait 2).53 The individual layers of the channel structure were made from a Bleher 175 μm thick PET foil (Optimont 501, Bleher Folientechnik GmbH, Germany) and two kinds of double-sided adhesive tape, Adhesive Research Arcare 90445 (thickness: 80 μm) and 3 M 467MP (thickness: 50 μm) (purchased from adhesives research, Ireland, and local supplier, respectively). The lower channel consisted of a 5 mm wide gap with a circular opening in the middle (radius: 7.5 mm). The upper channel consisted of a rectangular opening of 18 mm by 22 mm. An impermeable 30 μm thick PET foil (Bleher Folientechnik GmbH, Germany) separated the two channels. The same PET foil was used to form a lid on the upper microfluidic channel, which was fixed to the acrylic glass piece by a double-sided adhesive tape. Likewise, four tube connectors (EV Group, Austria) were fixed to the double-sided adhesive tape. The gap between the acrylic glass cavity and the tube connector was filled with a liquid two-component glue (pico dent twinsil extra hard, Henry Schein, Austria), to prevent leakage from both channels. To assemble the microfluidic chip, each layer was stacked on top of the previous one. As a precaution against leakage, surrounding edges of the microfluidic chip were covered with a thin layer of a UV light curing polymer mixture SR 499 (Sartormer, France) with 2 wt.% IC784 (BASF AG, Switzerland) which solidified during the sterilization process. For sterilization, tube connectors were closed by male Luer stoppers (Carl Roth, Austria) and exposed to UV light for 2 min (Dymax ECE, USA). For the last step, the microfluidic chip was baked at 60 °C for 1 h (Memmert GmbH, Germany). If desired, tubing, tube connectors, male Luer stoppers, and acrylic glass pieces can be reused after proper washing with 2-propanol and deionized water.
Cell Cultures
Human Vascular Endothelial Cells
Primary ECs were differentiated from CD34+ cells isolated from human cord blood as described23 and were provided in frozen aliquots of 106 cells at passage five by Prof. Fabien Gosselet, Université d’Artois, France. After thawing, cells were seeded onto gelatin (0.1% in PBS)-coated 10 cm-dishes in ECM-5 medium (ECM from Sciencell) supplemented with 1% endothelial cell growth supplement (ECGS), gentamycin (50 μg/mL, BiochromAG, ref A-2712), and 5% of preselected, heat-inactivated FBS and cultivated at 37 °C, 5% CO2. After reaching confluency, cells were washed once with PBS, detached with a Trypsin/EDTA solution, counted, and reseeded at approximately 2 × 105 cells/cm2. Expression of the EC marker CD31/PECAM-1 was confirmed by flow cytometry and immunofluorescence.
Brain Pericytes
Bovine brain pericytes were purchased from Prof. Gosselet’s group and cultured in DMEM (Dulbecco’s Medium Eagle Modified, Life Technology) supplemented with 2 g/L NaHCO3 (Merck, Austria), 20% heat-inactivated FCS (Sigma-Aldrich, Austria), l-glutamine, and 50 μg/mL gentamycin (Biochrom AG, Germany). Cocultures of human ECs and pericytes are an established and well-characterized blood–brain model.23,47 Co-cultivation protocols were adapted for use in the microfluidics system.
HUVECtert2 Line
Immortalized human umbilical vein endothelial cells (HUVECtert2) were purchased from Evercyte (Vienna, Austria) and cultured in phenol red-free M199 medium (Carl Roth, Karlsruhe, Germany) supplemented with 10% BCS (Sigma-Aldrich, Austria), LVES (Thermo Fisher Scientific, USA), GlutaMAX (100×, Thermo Fisher Scientific, USA), penicillin-streptomycin (100 IU/mL penicillin, 100 μg/mL streptomycin), and 20 μg/mL G418 (Gemini, Sussex, United Kingdom) at 37 °C and 5% CO2.
Cell Cultivation in the Dual-Channel Microfluidics System
Human Vascular Endothelial Cells
The UV-sterilized microfluidic system was thoroughly rinsed with PBS, followed by coating of the flow chambers with 0.1% gelatin solution. Next, the flow chambers were rinsed with prewarmed ECM-5 medium, and ECs were seeded into the channel (105 cells/cm2). Cells adhered for 1–2 h at 37 °C. Then the tubing was filled with prewarmed ECM-5 medium and connected to the MF system and the syringe pump. Nonadherent cells were removed by rinsing the flow chamber with 1 mL of ECM-5 at a flow rate of 10 mL/h. Next, the flow rate was decreased to 0.2 mL/h, and cells were cultivated for up to 6 days under constant flow conditions. In the lower channel, medium was replaced every 2 h or if the pH change (i.e., color change of phenol red) was visually observed. For immunostaining at the indicated time points, the MF system was rinsed with prewarmed HBSS containing Ca2+ and Mg2+, followed by fixation with 4% paraformaldehyde in HBSS for 15 min at room temperature.
HUVECtert2 Line
The position of the polymeric cell scaffolds was marked with a solvent-resistant lab marker. Subsequently, the chip was rinsed with 0.02% SDS in distilled water in order to reduce the chip’s surface energy (first the lower than the upper channel). The lower channel (LC) was flushed with PBS, whereas the upper channel (UC) was coated with 100 μg/mL fibrinogen for 30 min at 37 °C and 5% CO2. The UC was washed with PBS, and the LC was filled with phenol-free medium. In the UC, HUVECtert2 were seeded at a density of 105 cells/cm2. Cells were allowed to settle for 3 h and were then subjected to shear stress via flow of phenol red-free M199 medium applied by a syringe pump connected to the chip providing fresh medium every 3 h (100 μL/min, 400 μL total volume).
Imaging
Confocal Fluorescence Microscopy
Confocal images were acquired using an inverted laser scanning microscope (Zeiss LSM 700, Medical University of Vienna, Imaging Core Facility) with an oil-immersion objective (Plan-Apochromat, 63×, NA 1.4) or Zeiss LSM 800 (Johannes Kepler University Linz) with an oil-immersion objective (Plan-Apochromat, 40×, NA 1.2). For recording of 3D stacks, the samples were illuminated using 4 solid-state laser lines (405 nm, 488 nm, 555 and 639 nm). The signals were detected using two photomultiplier tubes (PMT). For PMT 1, the following filter sets were used: SP 490, SP 555, SP 640 nm and for PMT 2 long-pass filters LP 490, LP 560, LP 640 nm. Images were recorded with the ZEN 2009 black software and evaluations were performed using ImageJ (version 1.52n)54 and Gwyddion (version 2.5).55
Actin Staining for dSTORM Imaging
Cells were washed with prewarmed PBS, fixed using 4% paraformaldehyde in PBS for 10 min and washed again. The actin cytoskeleton was visualized using Alexa Fluor 647 conjugated phalloidin (Cell Signaling Technology, Leiden, Netherlands). Actin staining was performed in cytoskeleton buffer with sucrose (CBS) containing 10 mM MES pH 6.1, 138 mM KCl, 3 mM MgCl2, 2 mM EGTA and 0.32 M sucrose according to the protocol from Louise Cramer (MRC Laboratory for Molecular Cell Biology, UCL). Briefly, cells were permeabilized in 0.5% Triton X-100 with CBS, blocked in 10% albumin from chicken egg white (Sigma-Aldrich, Vienna, Austria) and stained for 20 min with 66 nM Alexa Fluor 647 conjugated phalloidin.
Fluorescence Microscopy and 3D dSTORM Image Acquisition
Images were acquired using a modified Olympus IX81 inverted epifluorescent microscope with an oil-immersion objective (PlanApo N, 60×, NA 1.42, Olympus, Vienna, Austria). Samples were mounted on a XYZ piezo stage (PI Mars; P-562, Physical Instruments) which has nanometer accuracy, combined with a coarse mechanical stage with a travel range of 1 cm × 1 cm (Hybrid, JPK Instruments, Berlin, Germany). A tube lens with an additional magnification of 1.6 was used to achieve a final imaging magnification of 96 (corresponding to a pixel size of 167 nm). ECs were illuminated with a 642 nm laser diode (Omicron-laserage Laserprodukte GmbH, Phoxx 642, Rodgau-Dudenhofen, Germany) and a 488 nm laser (Toptica Photonics, Germany). Signals were collected using an Andor iXonEM+ 897 (back-illuminated) EMCCD camera (16 μm pixel size). The following filter sets were used: dichroic filter (ZT405/488/561/640rpc, Chroma, Germany), emission filter (446/523/600/677 nm BrightLine quad-band band-pass filter, Semrock, Rochester, NY, USA), and additional emission filters: ET 700/75 M, Chroma Technology GmbH, Olching, Germany; ET 525/50 M, Chroma Technology GmbH, Olching, Germany. For 3D measurements, a cylindrical lens (f = 500 mm; Thorlabs, Newton, USA) was placed into the detection path of the microscope.
3D dSTORM Imaging Protocol
Single-molecule photoswitching of the rhodamine dye Alexa Fluor 488 and the cyanine dye Alexa Fluor 647 was performed in a buffer optimized for both dyes (no need for buffer exchange). This buffer contained OxEA as described by Nahidiazar et al.(56) and was applied to the cells immediately prior to measurements. For imaging, samples were illuminated for 40 ms with a 1.4 kW/cm2 excitation intensity (647 nm) and for 20 ms at a 3.3 kW/cm2 excitation intensity (488 nm) on the sample plane (both frame rates: 25 images/s). For image reconstruction, a sequence of 20000 images was recorded. All illumination protocols were performed using custom-written acquisition software.
Drift Correction
The drift correction method was applied according to Han et al.(57) The concept of drift correction is similar to a cross correlation analysis and is based on the Parzen-window density estimation.58,59 For more details, see ref (44).
Background Correction for Grid Artifacts
Polymers used in lithography can introduce additional background signals in fluorescent images. In our case, cells on the BisSR 3D polymer scaffold had a slightly higher autofluorescence, and super-resolution analyses of the cells’ cytoskeletons were more challenging. Closer inspection of the recorded super-resolution image stack showed that the autofluorescence of the polymer bleaches slowly over the acquisition time and is visible in all frames. Since, stochastic blinking fluorophores are only visible over a short-range of frames, an averaged image of the stack would result in a constant background image. Thus, a subtraction of this background from each frame in the stack can be used to remove the unwanted autofluorescence. The standard procedure of averaging of all frames and subsequent subtraction of the resulting background from each frame was impossible to apply due to the slowly decreasing autofluorescent signal (resulting from bleaching). Therefore, an advanced software tool for background subtraction was written, where the following procedures were applied. For calculation of the background (averaged) image, each frame was normalized to the prior summation step, which resulted in a more accurate background image. In the second step, prior to subtraction, the obtained background image was scaled in accordance with the intensity of each frame. The parameters for scaling and offset were calculated by linear regression between the background image and corresponding frame.
High Density Lipoprotein (HDL) Preparation
Plasma was obtained from normolipidemic healthy volunteers. Informed consent was obtained from all subjects. This was approved by the Ethics Committee of the Medical University of Vienna (license no. 1414/2016). Afterward, HDL particles were isolated by serial ultracentrifugation60 (for a detailed video, see ref (61)). The apolipoprotein part of the HDL particle was covalently labeled with Alexa647-succinimidyl ester according to the protocol (Thermo Fisher) (for details, see ref (62)).
Extracellular Vesicle (EV) Preparation
HEK293T cells (ATCC no. CRL-3216) were cultured in Dulbecco’s modified eagles medium (DMEM, Thermo-Fisher, Waltham, MA), high glucose, supplemented with 10% fetal bovine serum (FBS, Thermo-Fisher), and 1% penicillin/streptomycin (complete medium) at 37 °C in a 5% CO2 atmosphere. Transfection of HEK293T with the GFP-CD63 plasmid (Addgene no. 62964) for overexpression of the GFP-labeled EV marker CD63 was performed with 1 μg of plasmid DNA and 2 μL of Endofectin Max reagent in 6-well plates. After expansion of the transfected cells for several weeks, GFP+ cells were sorted by fluorescence-activated cell sorting (FACS), single-cell derived clones were isolated and characterized by flow cytometry and microscopy. For EV isolation, parental or transfected HEK293T cells were cultured to a subconfluent state in DMEM supplemented with 10% FCS and 1% penicillin/streptomycin. A 20% FCS solution in OptiMEM was centrifuged at 110000g for 12 h to deplete EVs from bovine origin. For collection of the conditioned media, the media was changed to OptiMEM (Gibco, Carlsbad, CA) containing 2% EV-depleted FCS, and supernatants of approximately 130–150 × 106 cells were collected after 24 h. The cell number and viability of EV producing cells was verified with a CASY cell counter (OMNI Life Science, Bremen, Germany). Cellular debris, apoptotic bodies, and larger EVs were removed from the supernatants by centrifugation at 200g (5 min), 2000g (10 min and 10000g (30 min) at 4 °C. Smaller EVs were obtained by ultracentrifugation at 110000g (1 h 10 min) at 4 °C (Sorvall MX150, Thermo Fisher Scientific, Waltham, MA) in a fixed angle rotor (Hitachi S58A-0095, Chiyoda, Japan). The pellet was washed in PBS, concentrated by ultracentrifugation, and subsequently resuspended in 30 μL of PBS. Samples were stored at −80 °C.
Acknowledgments
We thank Heidi Piglmayer-Brezina for taking the SEM images, Andreas Horner for help with confocal measurements, Ilse Kammerhofer for support with the administration organization, Fabien Gosselet and Julien Saint-Pol for helping with the establishment of the BBB cellular model, and EV-Group GmbH for providing microfluidic parts. This work was funded by the Interreg project ATCZ14 and the Austria Science Fund (FWF) project P 31827-B21.
Supporting Information Available
The Supporting Information is available free of charge at https://pubs.acs.org/doi/10.1021/acsnano.0c09263.
Movie S1: eGFP-labeled EVs diffusing across a 3D structure without cells (AVI)
Movie S2: Static, pseudo-3D reconstruction of diffusing particles in the lower channel (AVI)
Detailed description of the microfluidic chip layout; alternative forms of holes in PET-foil; functionalization of the polymeric structures; autofluorescence of the polymer grid for different excitation wavelength; cultivation of HUVECs in the microfluidic chip; color-separated images of cocultured ECs and bovine pericytes; 3D trajectories of extracellular vesicles diffusion in the ECs (PDF)
The authors declare no competing financial interest.
Supplementary Material
References
- Hwa A. J.; Fry R. C.; Sivaraman A.; So P. T.; Samson L. D.; Stolz D. B.; Griffith L. G. Rat Liver Sinusoidal Endothelial Cells Survive without Exogenous VEGF in 3D Perfused Co-Cultures with Hepatocytes. FASEB J. 2007, 21 (10), 2564–2579. 10.1096/fj.06-7473com. [DOI] [PubMed] [Google Scholar]
- Twardowski R. L.; Black L. D. Cardiac Fibroblasts Support Endothelial Cell Proliferation and Sprout Formation but Not the Development of Multicellular Sprouts in a Fibrin Gel Co-Culture Model. Ann. Biomed. Eng. 2014, 42 (5), 1074–1084. 10.1007/s10439-014-0971-2. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Rothbauer M.; Rosser J. M.; Zirath H.; Ertl P. Tomorrow Today: Organ-on-a-Chip Advances towards Clinically Relevant Pharmaceutical and Medical in Vitro Models. Curr. Opin. Biotechnol. 2019, 55, 81–86. 10.1016/j.copbio.2018.08.009. [DOI] [PubMed] [Google Scholar]
- Huang R.; Zheng W.; Liu W.; Zhang W.; Long Y.; Jiang X. Investigation of Tumor Cell Behaviors on a Vascular Microenvironment-Mimicking Microfluidic Chip. Sci. Rep. 2015, 5 (1), 17768. 10.1038/srep17768. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Walter F. R.; Valkai S.; Kincses A.; Petneházi A.; Czeller T.; Veszelka S.; Ormos P.; Deli M. A.; Dér A. A Versatile Lab-on-a-Chip Tool for Modeling Biological Barriers. Sens. Actuators, B 2016, 222, 1209–1219. 10.1016/j.snb.2015.07.110. [DOI] [Google Scholar]
- Chen L.-J.; Ito S.; Kai H.; Nagamine K.; Nagai N.; Nishizawa M.; Abe T.; Kaji H. Microfluidic Co-Cultures of Retinal Pigment Epithelial Cells and Vascular Endothelial Cells to Investigate Choroidal Angiogenesis. Sci. Rep. 2017, 7 (1), 3538. 10.1038/s41598-017-03788-5. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Sleeboom J. J. F.; Amirabadi H. E.; Nair P.; Sahlgren C. M.; Toonder J. M. J. den Metastasis in Context: Modeling the Tumor Microenvironment with Cancer-on-a-Chip Approaches. Disease Models & Mechanisms 2018, 10.1242/dmm.033100. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Mandt D.; Gruber P.; Markovic M.; Tromayer M.; Rothbauer M.; Krayz S. R. A.; Ali F.; Hoorick J.; van Holnthoner W.; Mühleder S.; Dubruel P.; Vlierberghe S. V.; Ertl P.; Liska R.; Ovsianikov A. Fabrication of Placental Barrier Structures within a Microfluidic Device Utilizing Two-Photon Polymerization. Int. J. Bioprinting 2018, 10.18063/ijb.v4i2.144. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Hudecz D.; Khire T.; Chung H. L.; Adumeau L.; Glavin D.; Luke E.; Nielsen M. S.; Dawson K. A.; McGrath J. L.; Yan Y. Ultrathin Silicon Membranes for in Situ Optical Analysis of Nanoparticle Translocation across a Human Blood-Brain Barrier Model. ACS Nano 2020, 14 (1), 1111–1122. 10.1021/acsnano.9b08870. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Kincses A.; Santa-Maria A. R.; Walter F. R.; Dér L.; Horányi N.; Lipka D. V.; Valkai S.; Deli M. A.; Dér A. A Chip Device to Determine Surface Charge Properties of Confluent Cell Monolayers by Measuring Streaming Potential. Lab Chip 2020, 20 (20), 3792–3805. 10.1039/D0LC00558D. [DOI] [PubMed] [Google Scholar]
- Haase K.; Kamm R. D. Advances in on-Chip Vascularization. Regener. Med. 2017, 12 (3), 285–302. 10.2217/rme-2016-0152. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Tam J.; Cordier G. A.; Bálint Š.; Álvarez Á. S.; Borbely J. S.; Lakadamyali M. A Microfluidic Platform for Correlative Live-Cell and Super-Resolution Microscopy. PLoS One 2014, 9 (12), e115512 10.1371/journal.pone.0115512. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Sharghi-Namini S.; Tan E.; Ong L.-L. S.; Ge R.; Asada H. H. Dll4-Containing Exosomes Induce Capillary Sprout Retraction in a 3D Microenvironment. Sci. Rep. 2015, 4 (1), 4031. 10.1038/srep04031. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Xu B.-B.; Zhang Y.-L.; Xia H.; Dong W.-F.; Ding H.; Sun H.-B. Fabrication and Multifunction Integration of Microfluidic Chips by Femtosecond Laser Direct Writing. Lab Chip 2013, 13 (9), 1677–1690. 10.1039/c3lc50160d. [DOI] [PubMed] [Google Scholar]
- Fischer J.; Wegener M. Three-Dimensional Optical Laser Lithography beyond the Diffraction Limit. Laser & Photonics Reviews 2013, 7 (1), 22–44. 10.1002/lpor.201100046. [DOI] [Google Scholar]
- Stone N. L.; England T. J.; O’Sullivan S. E. A Novel Transwell Blood Brain Barrier Model Using Primary Human Cells. Front. Cell. Neurosci. 2019, 13, 13. 10.3389/fncel.2019.00230. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Elbakary B.; Badhan R. K. S. A Dynamic Perfusion Based Blood-Brain Barrier Model for Cytotoxicity Testing and Drug Permeation. Sci. Rep. 2020, 10 (1), 3788. 10.1038/s41598-020-60689-w. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Yeon J. H.; Na D.; Choi K.; Ryu S.-W.; Choi C.; Park J.-K. Reliable Permeability Assay System in a Microfluidic Device Mimicking Cerebral Vasculatures. Biomed. Microdevices 2012, 14 (6), 1141–1148. 10.1007/s10544-012-9680-5. [DOI] [PubMed] [Google Scholar]
- Ahn S. I.; Sei Y. J.; Park H.-J.; Kim J.; Ryu Y.; Choi J. J.; Sung H.-J.; MacDonald T. J.; Levey A. I.; Kim Y. Microengineered Human Blood-Brain Barrier Platform for Understanding Nanoparticle Transport Mechanisms. Nat. Commun. 2020, 11 (1), 175. 10.1038/s41467-019-13896-7. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Tičku̅nas T.; Perrenoud M.; Butkus S.; Gadonas R.; Rekstyte S.; Malinauskas M.; Paipulas D.; Bellouard Y.; Sirutkaitis V. Combination of Additive and Subtractive Laser 3D Microprocessing in Hybrid Glass/Polymer Microsystems for Chemical Sensing Applications. Opt. Express 2017, 25 (21), 26280–26288. 10.1364/OE.25.026280. [DOI] [PubMed] [Google Scholar]
- Jonusauskas L.; Rekstyte S.; Buividas R.; Butkus S.; Gadonas R.; Juodkazis S.; Malinauskas M. Hybrid Subtractive-Additive-Welding Microfabrication for Lab-on-Chip Applications via Single Amplified Femtosecond Laser Source. Opt. Eng. 2017, 56 (9), 094108. 10.1117/1.OE.56.9.094108. [DOI] [Google Scholar]
- Sugioka K. Hybrid Femtosecond Laser Three-Dimensional Micro-and Nanoprocessing: A Review. Int. J. Extrem. Manuf. 2019, 1 (1), 012003. 10.1088/2631-7990/ab0eda. [DOI] [Google Scholar]
- Cecchelli R.; Aday S.; Sevin E.; Almeida C.; Culot M.; Dehouck L.; Coisne C.; Engelhardt B.; Dehouck M.-P.; Ferreira L. A Stable and Reproducible Human Blood-Brain Barrier Model Derived from Hematopoietic Stem Cells. PLoS One 2014, 9 (6), e99733 10.1371/journal.pone.0099733. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Hudecz D.; Rocks L.; Fitzpatrick L. W.; Herda L.-M.; Dawson K. A. Reproducibility in Biological Models of the Blood-Brain Barrier. Eur. J. Nanomed. 2014, 6 (3), 185–193. 10.1515/ejnm-2014-0021. [DOI] [Google Scholar]
- Kuo Y.-C.; Chung C.-Y. Transcytosis of CRM197-Grafted Polybutylcyanoacrylate Nanoparticles for Delivering Zidovudine across Human Brain-Microvascular Endothelial Cells. Colloids Surf., B 2012, 91, 242–249. 10.1016/j.colsurfb.2011.11.007. [DOI] [PubMed] [Google Scholar]
- Benam K. H.; Villenave R.; Lucchesi C.; Varone A.; Hubeau C.; Lee H.-H.; Alves S. E.; Salmon M.; Ferrante T. C.; Weaver J. C.; Bahinski A.; Hamilton G. A.; Ingber D. E. Small Airway-on-a-Chip Enables Analysis of Human Lung Inflammation and Drug Responses in Vitro. Nat. Methods 2016, 13 (2), 151–157. 10.1038/nmeth.3697. [DOI] [PubMed] [Google Scholar]
- Buchroithner B.; Hartmann D.; Mayr S.; Jin Oh Y.; Sivun D.; Karner A.; Buchegger B.; Griesser T.; Hinterdorfer P. A.; Klar T.; Jacak J. 3D Multiphoton Lithography Using Biocompatible Polymers with Specific Mechanical Properties. Nanoscale Adv. 2020, 2 (6), 2422–2428. 10.1039/D0NA00154F. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Wollhofen R.; Buchegger B.; Eder C.; Jacak J.; Kreutzer J.; Klar T. A. Functional Photoresists for Sub-Diffraction Stimulated Emission Depletion Lithography. Opt. Mater. Express 2017, 7 (7), 2538–2559. 10.1364/OME.7.002538. [DOI] [Google Scholar]
- Wollhofen R.; Axmann M.; Freudenthaler P.; Gabriel C.; Röhrl C.; Stangl H.; Klar T. A.; Jacak J. Multiphoton-Polymerized 3D Protein Assay. ACS Appl. Mater. Interfaces 2018, 10 (2), 1474–1479. 10.1021/acsami.7b13183. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Buchegger B.; Kreutzer J.; Axmann M.; Mayr S.; Wollhofen R.; Plochberger B.; Jacak J.; Klar T. A. Proteins on Supported Lipid Bilayers Diffusing around Proteins Fixed on Acrylate Anchors. Anal. Chem. 2018, 90 (21), 12372–12376. 10.1021/acs.analchem.8b02588. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Buchegger B.; Kreutzer J.; Plochberger B.; Wollhofen R.; Sivun D.; Jacak J.; Schütz G. J.; Schubert U.; Klar T. A. Stimulated Emission Depletion Lithography with Mercapto-Functional Polymers. ACS Nano 2016, 10 (2), 1954–1959. 10.1021/acsnano.5b05863. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Buchegger B.; Vidal C.; Neuwirth J.; Buchroithner B.; Karner A.; Hochreiner A.; Klar T. A.; Jacak J. Gold Nanoislands Grown on Multiphoton Polymerized Structures as Substrate for Enzymatic Reactions. ACS Materials Lett. 2019, 1 (4), 399–403. 10.1021/acsmaterialslett.9b00182. [DOI] [Google Scholar]
- Aekbote B. L.; Jacak J.; Schütz G. J.; Csányi E.; Szegletes Z.; Ormos P.; Kelemen L. Aminosilane-Based Functionalization of Two-Photon Polymerized 3D SU-8 Microstructures. Eur. Polym. J. 2012, 48 (10), 1745–1754. 10.1016/j.eurpolymj.2012.06.011. [DOI] [Google Scholar]
- Wolfesberger C.; Wollhofen R.; Buchegger B.; Jacak J.; Klar T. A. Streptavidin Functionalized Polymer Nanodots Fabricated by Visible Light Lithography. J. Nanobiotechnol. 2015, 13 (1), 27. 10.1186/s12951-015-0084-6. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Marieb E. N.; Hoehn K.. The Cardiovascular System: Blood Vessels. In Human Anatomy & Physiology, 9th ed.; Pearson: Boston, 2013; p 1264. [Google Scholar]
- Dejana E.; Orsenigo F.; Molendini C.; Baluk P.; McDonald D. M. Organization and Signaling of Endothelial Cell-to-Cell Junctions in Various Regions of the Blood and Lymphatic Vascular Trees. Cell Tissue Res. 2009, 335 (1), 17–25. 10.1007/s00441-008-0694-5. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Conway D. E.; Breckenridge M. T.; Hinde E.; Gratton E.; Chen C. S.; Schwartz M. A. Fluid Shear Stress on Endothelial Cells Modulates Mechanical Tension across VE-Cadherin and PECAM-1. Curr. Biol. 2013, 23 (11), 1024–1030. 10.1016/j.cub.2013.04.049. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Liu G.; Place A. T.; Chen Z.; Brovkovych V. M.; Vogel S. M.; Muller W. A.; Skidgel R. A.; Malik A. B.; Minshall R. D. ICAM-1-Activated Src and ENOS Signaling Increase Endothelial Cell Surface PECAM-1 Adhesivity and Neutrophil Transmigration. Blood 2012, 120 (9), 1942–1952. 10.1182/blood-2011-12-397430. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Reglero-Real N.; Colom B.; Bodkin J. V.; Nourshargh S. Endothelial Cell Junctional Adhesion Molecules: Role and Regulation of Expression in Inflammation. Arterioscler., Thromb., Vasc. Biol. 2016, 36 (10), 2048–2057. 10.1161/ATVBAHA.116.307610. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Solovey A. N.; Gui L.; Chang L.; Enenstein J.; Browne P. V.; Hebbel R. P. Identification and Functional Assessment of Endothelial P1H12. J. Lab. Clin. Med. 2001, 138 (5), 322–331. 10.1067/mlc.2001.118519. [DOI] [PubMed] [Google Scholar]
- Bardin N.; Anfosso F.; Massé J. M.; Cramer E.; Sabatier F.; Le Bivic A.; Sampol J.; Dignat-George F. Identification of CD146 as a Component of the Endothelial Junction Involved in the Control of Cell-Cell Cohesion. Blood 2001, 98 (13), 3677–3684. 10.1182/blood.V98.13.3677. [DOI] [PubMed] [Google Scholar]
- Nakagawa S.; Deli M. A.; Kawaguchi H.; Shimizudani T.; Shimono T.; Kittel A.; Tanaka K.; Niwa M. A New Blood-Brain Barrier Model Using Primary Rat Brain Endothelial Cells, Pericytes and Astrocytes. Neurochem. Int. 2009, 54 (3–4), 253–263. 10.1016/j.neuint.2008.12.002. [DOI] [PubMed] [Google Scholar]
- Mayr S.; Hauser F.; Peterbauer A.; Tauscher A.; Naderer C.; Axmann M.; Plochberger B.; Jacak J. Localization Microscopy of Actin Cytoskeleton in Human Platelets. Int. J. Mol. Sci. 2018, 19 (4), 1150. 10.3390/ijms19041150. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Mayr S.; Hauser F.; Puthukodan S.; Axmann M.; Göhring J.; Jacak J. Statistical Analysis of 3D Localisation Microscopy Images for Quantification of Membrane Protein Distributions in a Platelet Clot Model. PLoS Comput. Biol. 2020, 16 (6), e1007902 10.1371/journal.pcbi.1007902. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Rust M. J.; Bates M.; Zhuang X. Sub-Diffraction-Limit Imaging by Stochastic Optical Reconstruction Microscopy (STORM). Nat. Methods 2006, 3 (10), 793–796. 10.1038/nmeth929. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Coisne C.; Hallier-Vanuxeem D.; Boucau M.-C.; Hachani J.; Tilloy S.; Bricout H.; Monflier E.; Wils D.; Serpelloni M.; Parissaux X.; Fenart L.; Gosselet F. β-Cyclodextrins Decrease Cholesterol Release and ABC-Associated Transporter Expression in Smooth Muscle Cells and Aortic Endothelial Cells. Front. Physiol. 2016, 7, 7. 10.3389/fphys.2016.00185. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Kuntz M.; Candela P.; Saint-Pol J.; Lamartinière Y.; Boucau M.-C.; Sevin E.; Fenart L.; Gosselet F. Bexarotene Promotes Cholesterol Efflux and Restricts Apical-to-Basolateral Transport of Amyloid-β Peptides in an in Vitro Model of the Human Blood-Brain Barrier. J. Alzheimer's Dis. 2015, 48 (3), 849–862. 10.3233/JAD-150469. [DOI] [PubMed] [Google Scholar]
- Rhode S.; Breuer A.; Hesse J.; Sonnleitner M.; Pagler T. A.; Doringer M.; Schütz G. J.; Stangl H. Visualization of the Uptake of Individual HDL Particles in Living Cells via the Scavenger Receptor Class B Type I. Cell Biochem. Biophys. 2004, 41 (3), 343–356. 10.1385/CBB:41:3:343. [DOI] [PubMed] [Google Scholar]
- Allan D.; van der Wel C.; Keim N.; Caswell T. A.; Wieker D.; Verweij R.; Reid T. C.; Grueter L.; Ramos K.; Zoeith A.; Perry R. W.; Boulogne F.; Sinha P.; Pfigliozzi N. B.; Uieda L.; Katins J.; Hadrien M.; Ahmadia A.. Soft-Matter/Trackpy: Trackpy, v0.4.2; 2019, 10.5281/zenodo.3492186. [DOI]
- Tarantino N.; Tinevez J.-Y.; Crowell E. F.; Boisson B.; Henriques R.; Mhlanga M.; Agou F.; Israël A.; Laplantine E. TNF and IL-1 Exhibit Distinct Ubiquitin Requirements for Inducing NEMO-IKK Supramolecular Structures. J. Cell Biol. 2014, 204 (2), 231–245. 10.1083/jcb.201307172. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Vitali C.; Wellington C. L.; Calabresi L. HDL and Cholesterol Handling in the Brain. Cardiovasc. Res. 2014, 103 (3), 405–413. 10.1093/cvr/cvu148. [DOI] [PubMed] [Google Scholar]
- Bückmann T.; Stenger N.; Kadic M.; Kaschke J.; Frölich A.; Kennerknecht T.; Eberl C.; Thiel M.; Wegener M. Tailored 3D Mechanical Metamaterials Made by Dip-in Direct-Laser-Writing Optical Lithography. Adv. Mater. 2012, 24 (20), 2710–2714. 10.1002/adma.201200584. [DOI] [PubMed] [Google Scholar]
- Kokalj T.; Park Y.; Vencelj M.; Jenko M.; Lee L. P. Self-Powered Imbibing Microfluidic Pump by Liquid Encapsulation: SIMPLE. Lab Chip 2014, 14 (22), 4329–4333. 10.1039/C4LC00920G. [DOI] [PubMed] [Google Scholar]
- Schneider C. A.; Rasband W. S.; Eliceiri K. W. NIH Image to ImageJ: 25 Years of Image Analysis. Nat. Methods 2012, 9 (7), 671–675. 10.1038/nmeth.2089. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Nečas D.; Klapetek P. Gwyddion: An Open-Source Software for SPM Data Analysis. Open Physics 2012, 10 (1), 181–188. 10.2478/s11534-011-0096-2. [DOI] [Google Scholar]
- Nahidiazar L.; Agronskaia A. V.; Broertjes J.; Broek B.; van den Jalink K. Optimizing Imaging Conditions for Demanding Multi-Color Super Resolution Localization Microscopy. PLoS One 2016, 11 (7), e0158884 10.1371/journal.pone.0158884. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Han R.; Wang L.; Xu F.; Zhang Y.; Zhang M.; Liu Z.; Ren F.; Zhang F. Drift Correction for Single-Molecule Imaging by Molecular Constraint Field, a Distance Minimum Metric. BMC Biophys. 2015, 8 (1), 1. 10.1186/s13628-014-0015-1. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Rosenblatt M. Remarks on Some Nonparametric Estimates of a Density Function. Ann. Math. Stat. 1956, 27 (3), 832–837. 10.1214/aoms/1177728190. [DOI] [Google Scholar]
- Parzen E. On Estimation of a Probability Density Function and Mode. Ann. Math. Stat. 1962, 33 (3), 1065–1076. 10.1214/aoms/1177704472. [DOI] [Google Scholar]
- Schumaker V. N.; Puppione D. L. Sequential Flotation Ultracentrifugation. Methods Enzymol. 1986, 128, 155–170. 10.1016/0076-6879(86)28066-0. [DOI] [PubMed] [Google Scholar]
- Axmann M.; Karner A.; Meier S. M.; Stangl H.; Plochberger B. Enrichment of Native Lipoprotein Particles with MicroRNA and Subsequent Determination of Their Absolute/Relative MicroRNA Content and Their Cellular Transfer Rate. J. Visualized Exp. 2019, 147. 10.3791/59573. [DOI] [PubMed] [Google Scholar]
- Plochberger B.; Röhrl C.; Preiner J.; Rankl C.; Brameshuber M.; Madl J.; Bittman R.; Ros R.; Sezgin E.; Eggeling C.; Hinterdorfer P.; Stangl H.; Schütz G. J. HDL Particles Incorporate into Lipid Bilayers - A Combined AFM and Single Molecule Fluorescence Microscopy Study. Sci. Rep. 2017, 7 (1), 15886. 10.1038/s41598-017-15949-7. [DOI] [PMC free article] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.