Closed-loop circuit for reduce oxygen waste on hollow-fiber oxygenators during extracorporeal technologies

Ignazio Condello 1,, Flavio Rimmaudo 2, Giuseppe Speziale 1
PMCID: PMC7908748  PMID: 33632260

Closed-loop systems are designed to dynamically regulate a given variable around a desired set point. Examples thereof surround our everyday lives, from cruise control maintaining the correct speed on the highway, to auto-pilot flying modern airplanes safely [1]. Currently, there are two groups of hollow-fiber membrane oxygenator used in practice. The first types are diffusion, plasma-resistant oxygenators that have been increasingly used for extracorporeal life support or extracorporeal membrane oxygenation for patients who can no longer be supported by mechanical ventilation. The second types are hollow-fiber membranes made of microporous poly-propylene that have been widely used for standard cardiopulmonary bypass (CPB) [2]. Microporous hollow-fiber membranes are primarily used for short-term cardiopulmonary bypass application, whereas non-microporous hollow-fiber membranes are primarily used for extracorporeal membrane oxygenation application (ECMO) [3]. During the use of the oxygenator, the lung membrane uses about 30% of the medical oxygen delivered by the gas mixer, and the remainder comes out of the gas exhaust and is dispersed into the environment. In this context, we present a closed system aimed at recovering oxygen from the gas outlet of the oxygenator. First, the system recovers through a disposable polymer tube the gas flow from the oxygenator outlet; second, the condensate and water vapor will be removed with a water separator; third, the CO2 will be removed through a soda lime container; fourth, an electric control unit will decide whether to enrich the% of oxygen recovered through an oxygen source, in relation to the percentage of FiO2 set on the gas mixer (Fig. 1). The closed-loop circuit for oxygenators is a “concept development,” and no data are provided on feasibility in this context. The system should be implemented with continuous monitoring to avoid:

  • the water separator saturation, through excess water evacuation and appropriate filters

  • the risk of increasing pressure in the gas outlet through a safety valve

  • the risk for “rebreathing” and hence CO2 accumulation.

Fig. 1.

Fig. 1

Closed-loop circuit for reduce oxygen waste on hollow-fiber oxygenators during extracorporeal technologies

Our proposal has the objective of reducing waste and optimizing the use of medical oxygen; at the same time, this closed system is crucial during the transport phases (mostly airplane), in particular for patients on ECMO, to maximize the use of oxygen, guaranteeing greater autonomy.

Acknowledgements

Not applicable.

Authors’ contributions

IC designed the work and the conceptualized the idea. RF gave support for the realization of the manuscript. GS gave support for manuscript language. All authors read and approved the final manuscript.

Funding

None.

Availability of data and materials

Not applicable.

Ethical approval and consent to participate

Not applicable.

Consent for publication

Not applicable.

Competing interests

None.

Footnotes

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

References

  • 1.Platen, PV, Pomprapa A., Lachmann B, et al. The dawn of physiological closed-loop ventilation—a review. Crit Care 24, 121 (2020). https://doi.org/10.1186/s13054-020-2810-1 [DOI] [PMC free article] [PubMed]
  • 2.Berdajs DA, de Stefano E, Delay D, et al. The new advanced membrane gas exchanger. Interact Cardiovasc Thorac Surg. 2011;13(6):591–596. doi: 10.1510/icvts.2011.276873. [DOI] [PubMed] [Google Scholar]
  • 3.Iwahashi H, Yuri K, Nosé Y. Development of the oxygenator: past, present, and future. J Artif Organs. 2004;7(3):111–120. doi: 10.1007/s10047-004-0268-6. [DOI] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Data Availability Statement

Not applicable.


Articles from Critical Care are provided here courtesy of BMC

RESOURCES