Skip to main content
Diagnostics logoLink to Diagnostics
. 2021 Feb 19;11(2):349. doi: 10.3390/diagnostics11020349

Evaluation of MicroScan Bacterial Identification Panels for Low-Resource Settings

Sien Ombelet 1,2,*, Alessandra Natale 3, Jean-Baptiste Ronat 3,4,5, Olivier Vandenberg 6,7,8, Liselotte Hardy 1, Jan Jacobs 1,2
Editor: Raul Colodner
PMCID: PMC7922174  PMID: 33669829

Abstract

Bacterial identification is challenging in low-resource settings (LRS). We evaluated the MicroScan identification panels (Beckman Coulter, Brea, CA, USA) as part of Médecins Sans Frontières’ Mini-lab Project. The MicroScan Dried Overnight Positive ID Type 3 (PID3) panels for Gram-positive organisms and Dried Overnight Negative ID Type 2 (NID2) panels for Gram-negative organisms were assessed with 367 clinical isolates from LRS. Robustness was studied by inoculating Gram-negative species on the Gram-positive panel and vice versa. The ease of use of the panels and readability of the instructions for use (IFU) were evaluated. Of species represented in the MicroScan database, 94.6% (185/195) of Gram-negative and 85.9% (110/128) of Gram-positive isolates were correctly identified up to species level. Of species not represented in the database (e.g., Streptococcus suis and Bacillus spp.), 53.1% out of 49 isolates were incorrectly identified as non-related bacterial species. Testing of Gram-positive isolates on Gram-negative panels and vice versa (n = 144) resulted in incorrect identifications for 38.2% of tested isolates. The readability level of the IFU was considered too high for LRS. Inoculation of the panels was favorably evaluated, whereas the visual reading of the panels was considered error-prone. In conclusion, the accuracy of the MicroScan identification panels was excellent for Gram-negative species and good for Gram-positive species. Improvements in stability, robustness, and ease of use have been identified to assure adaptation to LRS constraints.

Keywords: clinical bacteriology, low-resource settings, bacterial identification, blood cultures, MicroScan identification system

1. Introduction

Bacterial Identification in Low-Resource Settings

Clinical bacteriology laboratories equipped to diagnose bacterial infections and perform antibiotic susceptibility testing (AST) are scarce in low-resource settings (LRS) [1,2]. Correct identification of bacteria in clinical samples is a cornerstone for proper management of bacterial infections because both empirical and directed antibiotic treatments depend on the identified organism [3,4].

Current “gold standard” identification methods in high-resource settings, such as matrix-assisted laser desorption/ionization time-of-flight mass spectrometry (MALDI-TOF MS) and molecular techniques are not adapted to diagnostic laboratories in LRS [1,5]. Most laboratories in LRS still rely on “conventional” phenotypic identification techniques, in which isolates are inoculated on different culture media containing different carbohydrates and enzyme substrates, and interpretation of test results is carried out using dichotomous decision trees [6,7,8]. Commercialized panels consisting of phenotypic tests, such as the Analytical Profile Index (API®) panels (bioMérieux, Marcy l’Etoile, France) offer probability-based identification; combined test results are compared to databases containing the profiles of thousands of bacterial isolates [9].

In response to the global rise in antimicrobial resistance and the need to implement context-tailored antibiotic stewardship programs, the non-governmental humanitarian organization Médecins Sans Frontières (MSF) committed to developing an all-in-one, affordable, and transportable clinical bacteriology laboratory, the Mini-lab [10]. Because the Mini-lab is intended for use in LRS by non-expert laboratory technicians, a system for bacterial identification is required that is easy to inoculate, easy to read, inexpensive, and with long shelf life. After defining technical specifications (target product profile) and performing a market analysis, the Dried Overnight MicroScan ID panels by Beckman Coulter (Brea, CA, USA) were chosen because they have a long shelf life and can be read without the use of automated instruments.

To meet Mini-lab specifications, Beckman Coulter developed a customized “Research Use Only” ID panel containing all test wells for the identification of both Gram-negative and Gram-positive organisms on one single panel, coded MSFNPID1. The Gram-negative and Gram-positive test wells were assembled in separate groups on the panel, and therefore, knowledge of the Gram stain result is still essential to read the panel correctly (Figure 1). The primary objective of this study was to assess the diagnostic accuracy of these identification panels with clinical isolates originating from LRS. A secondary objective was to assess the ease of use of the system.

Figure 1.

Figure 1

Example of an inoculated customized Médecins Sans Frontières (MSF) ID panel (MSFNPID1): Gram-negative test wells on the top side and Gram-positive test wells on the opposite side. Gram-positive and Gram-negative test panels can thus be inoculated simultaneously; for reading, one of the two sides must be picked.

2. Materials and Methods

2.1. Identification Panels

We evaluated the Dried Overnight Positive ID Type 3 (PID3) and Dried Overnight Negative ID Type 2 (NID2) panels and the customized version of the MSF Negative Positive ID Type 1 (MSFNPID1) panel. We will further use the terms “Gram-positive panels” and “Gram-negative panels” when referring to the composition of test wells. When dealing with the specific panels, we will use the terms PID3, NID2, and MSF ID panels, respectively. Table 1 demonstrates the different tests performed on each panel and reagents to be added. The Beckman Coulter LabPro software version 4.43 and the MicroScan autoSCAN-4 automated reader were used for this evaluation. Lot numbers of panels are listed in Supplementary Table S1.

Table 1.

Test wells on the MicroScan Gram-positive and Gram-negative panels (Dried Overnight Positive ID Type 3 (PID3) and Dried Overnight Negative ID Type 2 (NID2), respectively) or embedded in the customized MSF panel (MSFNPID1) and reagents to be added. Underlined substrates are wells for which coverage with mineral oil before incubation is needed after inoculation. Substrates in bold are tests for which additional reagents need to be added after incubation and before reading.

Substrate. Abbreviation Panel Substrate Abbreviation Panel
Glucose GLU Gram-negative Lactose LAC Gram-positive
Sucrose SUC Gram-negative Trehalose TRE Gram-positive
Inositol INO Gram-negative Mannose MNS Gram-positive
Adonitol ADO Gram-negative Ribose RBS Gram-positive
Rhamnose RHA Gram-negative Inulin INU Gram-positive
Melibiose MEL Gram-negative Mannitol MAN Gram-positive
Penicillin G 4 µg/mL P4 Gram-negative PNP-β-d-Glucuronide PGR Gram-positive
Kanamycin 4 µg/mL K4 Gram-negative PNP-β-d-Galactopyranoside PGT Gram-positive
Colistin 4 µg/mL Cl4 Gram-negative Indoxyl Phosphatase IDX Gram-positive
Cephalothin 8 µg/mL Cf8 Gram-negative Phosphatase PHO Gram-positive
Nitrofurantoin 64 g/mL Fd64 Gram-negative Pyrrolidonyl-B-naphtylamide PYR Gram-positive
Tobramycin 4 g/mL To4 Gram-negative Bile-Esculin BE Gram-positive
Cetrimide CET Gram-negative Pyruvate PRV Gram-positive
Lysine LYS Gram-negative Bacitracin BAC Gram-positive
Ornithine ORN Gram-negative Crystal Violet CV Gram-positive
Tryptophan Deaminase TDA Gram-negative Bacitracin 0.05 g/mL MS Gram-positive
Esculin ESC Gram-negative Novobiocin 1.6 g/mL NOV Gram-positive
o-Nitrophenyl-d-Galactopyranoside ONPG Gram-negative Optochin OPT Gram-positive
Citrate CIT Gram-negative 6.5% NaCl NACL Gram-positive
Malonate MAL Gram-negative Sorbitol SOR Both panels
Acetamide ACE Gram-negative Raffinose RAF Both panels
Tartrate TAR Gram-negative Arabinose ARA Both panels
Oxidation Base Control OF/B Gram-negative Arginine ARG Both panels
Oxidation of glucose OF/G Gram-negative Nitrate NIT Both panels
Hydrogen Sulfide H2S Gram-negative Voges-Proskauer VP Both panels
Indole IND Gram-negative Urea URE Both panels
Decarboxylase Base Control DCB Gram-negative

2.2. Clinical Isolates and Reference Strains

A total of 367 isolates were tested (Table 2 and Table 3), including 332 anonymized clinical isolates and 35 reference strains corresponding to the most common bloodstream pathogens or contaminants in LRS. Of the clinical isolates, 66.9% originated from sub-Saharan Africa, 26.8% from Asia, 4.2% from South America, and 2.1% from Europe (See Supplementary Table S2 for details per species). Reference strains were obtained from the Belgian Sciensano (Scientific Institute of Public Health) (https://www.wiv-isp.be/QML/activities/external_quality/rapports/_nl/rapports_annee.htm; accessed 18 February 2021) as part of external quality assessment programs. The collection included 323 isolates (195 Gram-negative and 128 Gram-positive) of species represented in the MicroScan database. The remaining 44 isolates, belonging to species that were not present in the MicroScan database, were used to look for anticipated incorrect identifications.

Table 2.

Overall performance of the MicroScan Gram-negative, Gram-positive, and MSF ID panels for species represented in the MicroScan database. N = total number of isolates tested. Details of incorrect identifications with their probability scores are listed in Table 3. * Formerly part of Streptococcus bovis group.

Species N Correct Identification Incorrect Identification Identification Not Possible
High Probability Score Low Probability Score High Probability Score Low Probability Score
Enterobacterales
Escherichia coli 28 24 3 - 1 -
Klebsiella pneumoniae 13 13 - - - -
Klebsiella oxytoca 3 3 - - - -
Enterobacter cloacae complex 15 14 - - 1 -
Citrobacter freundii complex 9 9 - - - -
Kluyvera ascorbata 1 1 - - - -
Salmonella Typhi 10 10 - - - -
Salmonella Paratyphi A 10 9 - 1 - -
Salmonella Typhimurium 10 10 - - - -
Salmonella Choleraesuis 9 8 - - 1 -
Shigella species 10 9 - - 1 -
Morganella morganii 4 4 - - - -
Proteus mirabilis 7 7 - - - -
Providencia rettgerii 1 1 - - - -
Enterobacterales 130 122 (93.8%) 3 (2.3%) 1 (0.8%) 4 (3.1%) 0 (0%)
Non-fermenting Gram-negative organisms and Aeromonas/Vibrio species
Pseudomonas aeruginosa 11 11 - - - -
Acinetobacter baumannii 9 8 - 1 - -
Achromobacter xylosoxidans 10 8 - 1 1 -
Burkholderia cepacia 15 13 1 - 1 -
Stenotrophomonas maltophilia 15 14 1 - - -
Aeromonas species 3 2 - - 1 -
Vibrio alginolyticus 2 2 - - - -
Non-fermenters 65 58 (89.2%) 2 (3.1%) 2 (3.1%) 3 (4.6%) 0 (0%)
Total Gram-negative isolates
Total Gram-negative 195 180 (92.3%) 5 (2.6%) 3 (1.5%) 7 (3.6%) -
Species N Correct Identification Incorrect Identification Identification Not Possible
High Probability Score Low Probability Score High Probability Score Low Probability Score
Micrococcaceae
Staphylococcus aureus 14 14 - - - -
Staphylococcus hominis 3 2 1 - - -
Staphylococcus epidermidis 11 10 - 1 - -
Staphylococcus lugdunensis 3 3 - - - -
Staphylococcus lentus 1 - - 1 - -
Staphylococcus haemolyticus 6 3 - 2 1 -
Listeria monocytogenes 2 2 - - - -
Micrococcus species 4 4 - - - -
Total Staphylococcus/Micrococcus 44 38 (86.4%) 1 (2.3%) 4 (9.1%) 1 (2.3%) 0 (0%)
Streptococcaceae
Enterococcus faecalis 7 7 - - - -
Enterococcus faecium 7 4 - 2 - 1
Enterococcus casseliflavus 1 - - 1 - -
Streptococcus agalactiae 13 11 1 - 1 -
Streptococcus anginosus 8 2 2 3 - 1
Streptococcus pneumoniae 12 9 - 2 1 -
Streptococcus gallolyticus * 10 9 1 - - -
Streptococcus lutetiensis * 1 - - - 1 -
Streptococcus pyogenes 13 13 - - - -
Streptococcus dysgalactiae (group C & G) 6 6 - - - -
Streptococcus mitis 5 2 3 - - -
Streptococcus salivarius 1 1 - - - -
Total Streptococcus/Enterococcus species 84 64 (76.2%) 7 (8.3%) 8 (9.5%) 3 (3.6%) 2 (2.4%)
Total Gram-positive isolates
Total Gram-positive 128 102 (79.7%) 8 (6.3%) 12 (9.4%) 4 (3.9%) 2 (1.6%)
Overall total 323 282 (87.3%) 13 (4.0%) 15 (4.6%) 11 (3.4%) 2 (0.6%)

Table 3.

Overall performance of the MicroScan Gram-negative and Gram-positive panels for micro-organisms not represented in the MicroScan database. All identifications by the MicroScan system are by definition incorrect because the species are not represented in the database. The expected result for these species is “identification not possible”. N = total number of isolates tested. Overall, * 44 different isolates were tested, five of which on both the Gram-positive as the Gram-negative panel.

Species N Identification Not Possible Low Probability Identification High Probability Identification
Gram-positive bacilli (Bacillus & Corynebacterium species) 16 1 8 7
Streptococcus suis 11 - 5 6
Burkholderia species (non-cepacia, non-pseudomallei) 8 2 2 4
Yeast species (on Gram-positive panel) 9 - - 9
Yeast species (on Gram-negative panel) 5 - 5 -
Total (species not represented in MicroScan database) 49 * 3 (6.1%) 20 (40.8%) 26 (53.1%)

2.3. Reference Identification of Clinical Isolates

Identification of clinical isolates was confirmed by MALDI-TOF MS, using Microflex™ device (Bruker Daltonics, Billerica, MA, USA) with MALDI Biotyper® (MBT) 7854 MSP Library. Serotyping of Salmonella was performed with the Salmonella antisera of Pro-lab diagnostics (Richmond Hill, ON, Canada). Optochin disks (Rosco Diagnostica, Taastrup, Denmark) were used to distinguish between Streptococcus pneumoniae and other viridans Streptococcus species, after identification by MALDI-TOF as Streptococcus pneumoniae/mitis/oralis/parapneumoniae.

2.4. Preparation of Suspension

Frozen isolates (stored at −80 °C) underwent two passages on sheep blood agar. Inoculation from single colonies was conducted using the Prompt™ Inoculation System (Beckman Coulter, Brea, CA, USA), consisting of a wand designed to hold a specific quantity of bacterial material and a 30 mL Prompt™ Inoculation Water bottle (Beckman Coulter, Brea, CA, USA). As per the manufacturer’s instructions, some slow-growing Streptococcus species were inoculated using the turbidity standard method, in which a 0.5 McFarland solution was prepared in 3 mL of Inoculum Water, of which 100 µL was pipetted into 25 mL of Inoculum Water.

2.5. Inoculation of Panels

The bacterial suspension was transferred to the panels using MicroScan Renok (inoculating device by Beckman Coulter, Brea, CA, USA), which delivers 115 µL of broth suspension to each well. Mineral oil was added to the indicated wells (Table 1). Panels were incubated in an atmospheric incubator for 16–18 h at 35 °C.

2.6. Reading of the Panels

Reagents were added and panels were read visually using transmitted light and magnifier with interchangeable white or black backgrounds according to the manufacturer’s instructions for use (IFU). We used a microplate viewer prototype adapted for this purpose by JP Selecta (Barcelona, Spain) (Figure 2). The results were recorded on PID3 and NID2 worksheets (supplied by the manufacturer) and transferred to the LabPro software (version 4.43). The LabPro software calculated the “biotype”, a code containing the biochemical profile, and generated the corresponding identification for each biotype with probability scores. Results with high probability scores (≥85%) were considered reliable, while results with low probability scores (<85%) were considered “unconfirmed”, as defined by the IFU. No additional tests for confirmation of low probability identifications were performed in our study. If the biochemical profile did not match any identification in LabPro’s database, the result generated was “very rare biotype”.

Figure 2.

Figure 2

Prototype of the microplate viewer by JP Selecta, used for the visual reading. The background at the bottom can be changed to white.

2.7. Robustness Testing

Incorrect panel: To anticipate possible errors in Gram stain, we inoculated 64 Gram-negative isolates on Gram-positive panels and vice versa (80 isolates).

Incorrect algorithm: For Gram-positive organisms, the LabPro software asked to choose between the algorithms for Streptococcaceae (Streptococcus species and Enterococcus species (catalase negative) versus Micrococcaceae (Micrococcus species, Staphylococcus species, and Listeria monocytogenes (catalase positive). To assess the impact of errors in microscopy or catalase results, we compared the results of Gram-positive bacteria with both the Streptococcaceae and Micrococcaceae algorithm.

Hemolysis: For the Streptococcaceae, visual assessment of beta-hemolysis (complete hemolysis) on blood agar was part of the biotype. In many LRS, sheep blood agar is hard to come by, hence hemolysis patterns may be less reliable in these settings [11,12]. To assess the impact of hemolysis reading, we compared results for beta-hemolytic streptococci with hemolysis read as positive versus negative.

Oxidase: For Gram-negative organisms, the software required the result for oxidase testing. We used OxiSticks™ Oxidase Swabs (Hardy Diagnostics, Santa Maria, CA, USA) for this purpose. For a subset of Stenotrophomonas maltophilia isolates, oxidase testing was repeated with Bactident® oxidase (Merck Millipore, Burlington, MA, USA).

2.8. Comparison between Automated and Visual Reading, Repeatability and Inter-Observer Agreement

The panels of 190 isolates were read both by visual and automated reading. To assess the “repeatability” of the panels, 31 isolates were tested in triplicate. Panels were inoculated from a single blood agar plate and processed and read simultaneously and by the same operator. “Inter-observer agreement” of visual reading was assessed for 50 isolates, subsequently read by two operators blinded to each other’s results. For the above comparisons, species agreement was defined as an identical species identification; biotype agreement was defined as an identical biotype, implying identical results for each separate test well.

2.9. Definitions of Correct and Incorrect Identifications

Correct identification was defined as concordance between the reference identification method and the identification given by MicroScan software up to the level provided by the MicroScan system (species level for most organisms, serotype level for certain Salmonella serotypes and genus level for Micrococcus species, for instance). We assessed the potential impact of incorrect identifications in the light of diagnostic and therapeutic consequences, as well as epidemiological significance.

For species not represented in the database or when panels were incorrectly used (e.g., Gram-positive isolate on Gram-negative panel), the expected “correct” identification by the MicroScan system was “very rare biotype”, indicating the system did not find sufficient agreement with expected biotypes of species available in the database. We will further refer to this result as “no identification possible”.

2.10. Management of Incorrect Identifications

For incorrect identifications with high probability scores, MALDI-TOF was repeated. If initial reference identification was confirmed, the MicroScan panel was repeated or included in the repeatability testing panel: if the isolate was correctly identified upon repeat MicroScan testing, we assumed technical error on first testing and accepted the second result. For repeatability testing, if at least two of three repetitions yielded again an incorrect result, the isolate was considered incorrectly identified. For Streptococcus species, all incorrect identifications were repeated using the turbidity standard method instead of the Prompt™ method according to the MicroScan panel IFU.

2.11. Ease of Use

Ease of use was assessed by surveying the operators for feedback on each of the components of the system. The readability level of the IFU was assessed using Flesch–Kincaid Grade Levels (https://www.online-utility.org/english/readability_test_and_improve.jsp; accessed 18 February 2021) [13].

3. Results

3.1. Results for Species Included in the MicroScan Database

Gram-negative isolates: Gram-negative isolates represented in the MicroScan database were correctly identified with a high probability score in 92.3% of 195 isolates (Table 2). An additional 2.6% of isolates were correctly identified but with a low probability score. Enterobacterales were more accurately identified than non-fermenters, with correct identifications in 96.1% and 90.0%, respectively. Table 4 shows an overview of the incorrectly identified isolates. Most errors for Enterobacterales (6/9) were at the species level and had limited clinical relevance. Six out of 15 Stenotrophomonas maltophilia isolates tested oxidase positive with the OxiSticks™; this led to incorrect identifications as other non-fermenters in four isolates. Repeat oxidase testing with Bactident® oxidase tests gave negative results for all isolates; the corrected biotypes all produced correct identifications with high probability scores.

Table 4.

Details of incorrectly identified isolates. Potential clinical relevance is also indicated; limited clinical relevance means that the two pathogens are similar in clinical, epidemiological, and infection control aspects. Probability scores are reported by the MicroScan LabPro software with two decimals. A probability score of ≥85.00% is considered as “high probability”. N = total number of isolates tested. * As identified by matrix-assisted laser desorption/ionization time-of-flight mass spectrometry (MALDI-TOF) (and serotyping when necessary).

Species/Serotype * Nr. of Isolates with Incorrect Identification/N Incorrect Result by MicroScan Probability Score Clinical Relevance of Error
Gram-negative species
Escherichia coli 1/28 Escherichia fergusonii 67.88% Limited
Enterobacter cloacae 1/15 Enterobacter amnigenus 58.26% Limited
Salmonella Paratyphi A 1/10 Salmonella Typhi 98.87% Moderate: pathogens are clinically similar, but distinction has epidemiological relevance
Salmonella Choleraesuis 1/9 Salmonella Typhi 57.06% High: pathogens are both clinically and epidemiologically distinct
Shigella species 1/10 Klebsiella ozaenae 72.69% High: pathogens have different clinical presentations; Shigella infection is of high epidemiological importance (outbreak potential)
Acinetobacter baumannii 1/9 Acinetobacter lwoffii 92.88% Moderate: pathogens may have different clinical presentation
Achromobacter xylosoxidans 2/10 Rhizobactrum radiobacter
Burkholderia cepacia
49.88%
94.19%
Limited
Aeromonas caviae 1/3 Aeromonas hydrophila 61.71% Limited
Gram-positive species
Staphylococcus epidermidis 1/11 Staphylococcus hyicus 91.14% Limited
Staphylococcus lentus 1/1 Staphylococcus sciuri 96.73% Limited
Staphylococcus haemolyticus 3/6 Staphylococcus lugdunensis
Staphylococcus lugdunensis
53.67%
93.66%
Moderate: pathogens may have different clinical presentation
Enterococcus faecium 2/7 Enterococcus raffinosus
Enterococcus durans
95.39%
99.68%
Moderate: pathogens have different clinical significance
Enterococcus casseliflavus 1/1 Enterococcus gallinarum 59.68% Limited: both are associated with increased vancomycin resistance
Streptococcus anginosus 3/8 Streptococcus mitis/oralis
Streptococcus salivarius
Streptococcus parasanguinis
86.99%
88.56%
99.99%
High: Streptococcus anginosus is associated with purulent infections; other viridans streptococci are more likely contaminants or implicated in infectious endocarditis
Streptococcus pneumoniae 3/12 Rhodococcus equi
Gemella species
Gemella species
99.99%
51.48%
95.79%
High: Streptococcus pneumoniae is a more common pathogen than Rhodococcus equi or Gemella and clinical presentation is very different
Streptococcus agalactiae 1/13 Streptococcus dysgalactiae 99.99% Limited; except for screening in pregnancy

Gram-positive isolates: Of 128 Gram-positive isolates, 79.7% and 6.3% were correctly identified with high and low probability scores, respectively (Table 2). Low probability identifications were more frequent for Streptococcaceae than for Micrococcaceae (Staphylococcus, Micrococcus, and Listeria species), i.e., 8.3% of isolates versus 2.6%, respectively. Incorrect identifications were especially common—and considered as highly clinically relevant—for Streptococcus pneumoniae (3/12) and Streptococcus anginosus (4/8) (Figure 3 and Figure 4). All S. aureus isolates (n = 14) were identified correctly.

Figure 3.

Figure 3

Streptococcus pneumoniae clinical isolate incorrectly identified as Gemella species; PNP-β-d-Glucuronide test (PGT) was considered negative (any shade of yellow = positive). A positive result for PGT would have led to correct identification of S. pneumoniae with a probability score of 94.79%. Only indoxyl phosphatase (IDX) test was positive in this panel.

Figure 4.

Figure 4

Streptococcus anginosus reference isolate incorrectly identified as S. mitis/oralis. Note that both PHO and PGT wells were considered positive when this panel was read; “any shade of yellow” leads to a positive result according to the instructions for use (IFU). A negative reading would have resulted in the correct identification of S. anginosus. Note also that Voges–Proskauer (VP) test was judged negative because “pale pink” is to be considered negative according to the IFU. A positive result of VP would have led to an identification of S. parasanguinis.

3.2. Isolates Not Represented in the Database

Testing of bacteria not represented in the MicroScan database led to incorrect high probability identification in 53.1% of 49 isolates tested, and only for 6.1% of isolates, the expected result “identification not possible” was generated (Table 4). The yeast species were consistently misidentified by the MicroScan Gram-positive panel as either Staphylococcus cohnii (Micrococcus/Staphylococcus algorithm) or Rhodococcus equi (Streptococcus algorithm) (Table 5).

Table 5.

Most frequent incorrect identifications with high probability score for species not represented in the database or “incorrect” use of the panels (Gram-positive isolates inoculated on Gram-negative panels and vice versa, Streptococcus using Micrococcus/Staphylococcus algorithm and vice versa). N = total number of isolates tested.

Species. Algorithm Used (for Gram-Positive Panels) N Number of Isolates with this Result Result
Species not represented in the database
Yeast species Micrococcus 9 9 Staphylococcus cohnii
Streptococcus 9 Rhodococcus equi
Corynebacterium species Micrococcus 6 1 Micrococcus species
Streptococcus 6 1 Rhodococcus equi
Bacillus species Micrococcus 10 4 Staphylococcus auricularis
Gram-positive species on Gram-negative panel
Streptococcus dysgalactiae 2 2 Providencia rustigiannii
Streptococcus suis 10 9 Providencia rustigiannii
Staphylococcus lentus 1 1 Providencia rustigiannii
Streptococcus agalactiae 5 1 Providencia rustigiannii
Gram-negative species on Gram-positive panel
Escherichia coli/paracoli Streptococcus 10 3 Rhodococcus equi
Acinetobacter baumannii Micrococcus 4 3 Staphylococcus cohnii
Micrococcus 1 Staphylococcus auricularis
Streptococcus 1 Rhodococcus equi
Stenotrophomonas maltophilia Streptococcus 4 2 Rhodococcus equi
Burkholderia cepacia Streptococcus 4 3 Rhodococcus equi
Burkholderia thailandensis Streptococcus 4 1 Rhodococcus equi
Burkholderia ubonensis Streptococcus 1 1 Rhodococcus equi
Gram-positive species using incorrect algorithm on Gram-positive panel
Staphylococcus aureus Streptococcus 14 3 Rhodococcus equi
Micrococcus species Streptococcus 4 2 Rhodococcus equi
Staphylococcus lentus Streptococcus 1 1 Rhodococcus equi
Streptococcus gallolyticus Micrococcus 10 9 Listeria monocytogenes
Streptococcus pneumoniae Micrococcus 12 9 Micrococcus species
2 Staphylococcus auricularis
Streptococcus pyogenes Micrococcus 13 2 Staphylococcus auricularis
Streptococcus agalactiae Micrococcus 13 1 Staphylococcus auricularis
Streptococcus anginosus Micrococcus 8 1 Staphylococcus auricularis
Streptococcus mitis/oralis Micrococcus 5 1 Staphylococcus auricularis

3.3. Robustness Testing

Incorrect panel: Inoculating a Gram-positive isolate on the Gram-negative panel or vice versa led to an incorrect identification with a high probability score in 32.6% of 144 isolates tested (Table 6). This risk was higher for isolates wrongly inoculated on the Gram-negative panel (43.8%) compared to the Gram-positive panel (18.8%). A total of 38.2% isolates inoculated on the wrong panel generated the expected result “no identification possible”. Rhodococcus equi was a common incorrect identification with the Gram-positive panel (n = 27); when the Gram-negative panel was used for Gram-positive isolates, identification of Providencia rustigiannii recurred (n = 13) (Table 5).

Table 6.

Results when isolates were tested on an incorrect panel or using an incorrect algorithm, i.e., inoculation of Gram-negative species on Gram-positive panels or vice versa, or of Gram-positive species incorrectly assigned as “Streptococcus” or “Micrococcus” based on faulty catalase or microscopy results. The expected result for this type of incorrect panel use is “identification not possible”. For Gram-negative species inoculated on Gram-positive panel, we assumed Micrococcus interpretation for the results in this table, as most of them would be catalase positive. Note that all results other than “identification not possible” are wrong; detailed results for each species can be found in Supplementary Table S3. Incorrect identifications with high probability are listed in Table 5. N = total number of isolates tested.

Species. N Panel/Algorithm Used Number “Identification not Possible” (%) Number Low Probability Identification (%) Number High Probability Identification (%)
Gram-negative species tested on Gram-positive panels and vice versa
Enterobacterales 41 Gram-positive 27 9 5
Non-fermenters 23 10 6 7
Total Gram-negatives on Gram-positive panel 64 37 (57.8%) 15 (23.4%) 12 (18.8%)
Staphylococcus species 16 Gram-negative 1 11 4
Streptococcus species 53 15 12 26
Gram-positive bacilli (Bacillus & Corynebacterium species) 11 2 4 5
Total Gram-positive on Gram-negative panel 80 18 (22.5%) 27 (33.8%) 35 (43.8%)
Staphylococcus/Micrococcus species using Streptococcus species algorithm and vice versa
Staphylococcus/Micrococcus species and Gram-positive rods 60 Streptococcus algorithm 35 11 14
Streptococcus/Enterococcus species 95 Micrococcus algorithm 27 24 44
Total Gram-positive with incorrect algorithm 155 62 (40.0%) 35 (22.6%) 58 (37.4%)
Total tested on incorrect panel/incorrect algorithm 299 117 (39.1%) 77 (25.8%) 105 (35.1%)

Incorrect algorithm: When Staphylococcus species were interpreted erroneously using the Streptococcus algorithm, this led to incorrect high probability identification for 23.3% of isolates tested (Table 6). Vice versa, interpreting the results of Streptococcus species with the Micrococcus algorithm led to incorrect high probability identifications in 46.3% of isolates tested.

Hemolysis: Impact of hemolysis on identification of beta-hemolytic Streptococcus species (S. agalactiae, S. pyogenes, S. dysgalactiae) was limited; beta-hemolysis results led to different identifications in only 2 of 32 isolates (one S. agalactiae isolate and one S. pyogenes isolate).

3.4. Comparison between Automated and Visual Reading, Repeatability and Inter-Observer Agreement

Automated versus visual reading: Visual reading was slightly more accurate than automated reading (90.6% correct versus 87.5% correct), except for Streptococcus species. For species included in the MicroScan database, species agreement between visual and automated reading was 91.9%, and biotype agreement 58.1% (Supplementary Table S4).

Repeatability was better for automated reading than visual reading (89.7% versus 88.7% for species agreement; 87.5% versus 79.0% for biotype agreement). Especially for Staphylococcus species and Enterobacterales, repeatability was excellent with 100% biotype agreement for the automated reading. The lowest repeatability was seen for Streptococcus species (70.8% biotype agreement for automated reading versus 73.1% for visual reading).

Inter-observer agreement: On 50 isolates read by two operators, species and biotype agreements were 98.0% and 66.0%, respectively. VP (Voges-Proskauer test) was the test most often read differently by different observers (inter-observer and visual-automated) (Supplementary Table S5).

3.5. Ease of Use

Instructions for use: The instructions for interpretation (e.g., color) of positive or negative results were often equivocal (Table 7). Flesch–Kincaid Grade Levels of the Gram-negative, Gram-positive, and MSF panel IFU were 9, 11, and 10, respectively. Flesch–Kincaid Grade Levels refer to US grade levels (i.e., years of schooling) necessary to understand the text.

Table 7.

Ease-of-use assessment with a focus on use in low resource settings of different aspects of the MicroScan identification panels. Abbreviations: IFU = instructions for use; PEP = peptidase; PYR = pyrrolidonyl-b-naphtylamide; VP = Voges–Proskauer test.

Favorable Observations Unfavorable Observations
Storage of panels and reagents
Long shelf life of the panels and of most reagents VP2 reagent must be reconstituted with 95% ethanol (not provided); after reconstitution, vial is stable for only 2 weeks
Panels and some reagents can be stored at room temperature Many reagents (nitrate, VP, ferric chloride, Kovac’s) require storage at 2–8 °C
Instructions for use, labeling, and packaging
Panels are packaged separately with desiccant, (StripPax®, Multisorb Technologies, New York, NY, USA) Reconstitution of VP2 reagent not explained in IFU, only on box of reagent
Sturdy outer packaging
IFU were delivered at request by the distributor IFU are not delivered in the box with the products; they are available online but not easy to find
Flesch Kincaid Grade level of IFU was 9.14 for the Gram-negative IFU, 11.06 for the Gram-positive IFU and 10.30 for the MSF panel IFU; these levels point to a high level of education needed to comprehend the text
Instructions in IFU for interpretation of negative/positive wells are often ambiguous.e.g., yellow to orange color for carbohydrate tests must be considered positive, whereas orange to red must be considered negative
Reagent bottles are labeled with the test name: easy to add right reagent to right test well Use of reagent names is inconsistent between IFU and reagent bottles (e.g., 5% alpha naphtol in IFU, VP2 on bottle)
The PYR test reagent has “PEP” on the bottle and not “PYR”
Limited space for writing patient identification and sample number on the panel
Inoculation and reading
Most panels could be inoculated using the Prompt™ system Prompt™ is not advised for use in slow-growing Streptococcus species; we repeated testing of 14 Streptococcus isolates with turbidity standard method due to incorrect results with Prompt™; 50% of these had a correct result with turbidity standard method
Prompt ™ was easy to use and time-efficient
RENOK system was easy to use and time-efficient Large amounts of plastic waste associated with the RENOK system
Difficulties in disposing of waste without spillage
Difficulties in incubation of panels without spillage
Favorable observations Unfavorable observations
IFU recognizes that interpretation of test results requires trained clinical staff Visual reading of panels was considered difficult as positivity of test wells was often doubtful
Automated reading was more user-friendly as it removed subjectivity and eliminated clerical errors autoSCAN-4 does not indicate automatically that panels must be re-incubated
autoSCAN-4 reader dimensions are very large: 48.5 cm × 24 cm × 58 cm
Prototype microplate viewer (JP Selecta) with magnifier and interchangeable white/black background considerably facilitated the reading Both white and black background needed for reading of wells
Reagents can be added only after visual assessment of certain wells indicating the need for re-incubation; panels must be re-incubated if specific combinations of wells are negative—there are 6 of these combinations that consist each of 1 to 3 wells unevenly distributed on the panels
Different reading times for different reagents
Re-incubation was required for 15.3% of panels in our study; 13.3% (27/203) of Gram-negative isolates (mostly non-fermenters) and 17.7% (29/164) of Gram-positive isolates (mainly Streptococcus isolates).
Software and Biotype Lookup Tool
LabPro software and reader easy to use Beckman Coulter worksheets not delivered with the panels; available online but hard to find
Calibration and QC of reader are alerted by the reader automatically and do not require user input Test position on worksheet different from test position on the panel (for Gram-negative panel) (Figure 5).
Biotype Lookup Tools freely available online
Both the online Biotype Lookup Tool and LabPro software propose additional tests to confirm low-probability identification Many of the confirmatory tests proposed by the software are unlikely to be easily available in LRS Examples of recommended confirmatory tests:
  1. E. coli: cellobiose fermentation, growth in potassium cyanide and Christensen urea agar

  2. Salmonella Typhi: trehalose fermentation, serotyping

  3. S. pneumoniae: leucine aminopeptidase and bile solubility

  4. S. anginosus: beta-d-acetylgalactosaminidase, alpha-glucosidase, beta-glucosidase, hyaluronidase, neuraminidase

Inoculation and reading: Both the Prompt™ and the RENOK system (Beckman Coulter, Brea, CA, USA) were considered user-friendly and time-efficient, especially compared to other inoculation methods using 0.5 McFarland standards and pipettes. Reading of the panels was considered rather difficult. Visual assessment of positivity of wells was frequently doubtful. A harmonized workflow for reading multiple panels was hindered by the re-incubation requirements; reagents could only be added to the panels when certain combinations of tests were positive—if this was not the case, re-incubation for another 24 h was needed. Furthermore, waiting times between adding reagents and visual reading were variable, ranging from immediate reaction to 20 min. Automated reading was more user-friendly but a visual check of the required positive wells was still necessary. Re-incubation of panels was needed for 15.3% of panels.

Obtaining identifications: The worksheets provided with the PID3 and NID2 panels had some tests in different positions compared to the real-life panel or the software picture (Figure 5). To obtain identification results from the biotype, the LabPro software and a freely available online Biotype Lookup Tool could be used.

Figure 5.

Figure 5

Worksheet for the NID2 panel. Tests marked in orange are in a different order on the panel than they are on the worksheet (see Figure 1).

Customized panels: Because Gram stain results were still necessary to interpret the results of the inoculated panels, advantages in the usability of the customized MSF ID panels were mainly logistic (simplicity, stock management). On the other hand, care should be taken when reading the panels to avoid reading the wrong side.

3.6. Stability

The shelf life of the panels was within the limits predefined by the Mini-lab target product profile (minimally 12 months); storage conditions for the panels (2–25 °C) were not entirely within the Mini-Lab specifications (2–40 °C). Packaging of individual panels was of good quality, providing air-sealed individual aluminum-plastic pouches with a humidity indicator. Other components were provided within sturdy cardboard boxes. The shelf life of the VP2 reagent (Voges Proskauer test) was too short; it could be used for just two weeks after reconstitution of the reagent.

4. Discussion

4.1. Performance of the MicroScan Panels

Our results showed excellent performance of the Gram-negative panel, with 94.9% of isolates correctly identified to the species level, which was in line with several previous studies evaluating the MicroScan panels [14,15,16]. The performance of MicroScan for Gram-negatives was similar or better than this of API® (95.2%), Phoenix™ system from Becton–Dickinson (Franklin Lakes, NJ, USA) (92.5%), or Vitek® 2 from bioMérieux (Marcy-l’Etoile, France) (88.6%) as reported by other studies [17,18].

The Gram-positive panel had a lower performance with 85.9% of tested isolates correctly identified up to species level. This was again in line with other published studies for Staphylococcus and Enterococcus species. There were no previous publications that reported the accuracy of MicroScan panels for Streptococcus species. Performance of MicroScan for Gram-positive isolates was lower than the reported performance of API® (92.6%), Phoenix™ (93.6% accuracy), or Vitek® 2 (90.2%) [17,19]. All Staphylococcus aureus isolates were correctly identified in our study, but Streptococcus species were identified less accurately, generating clinically relevant misidentifications for Streptococcus pneumoniae and, to a lesser extent, Streptococcus anginosus. Overall, 7.4% of the isolates tested would need additional testing; this was needed more often for Gram-positive than Gram-negative isolates.

Compared to conventional phenotypic testing as it is still conducted in most LRS, MicroScan performed considerably better—a study in Nigeria on 145 Enterobacterales isolates, comparing MALDI-TOF to conventional testing, found only 57.2% accuracy on genus level and 33.1% on species level [20].

4.2. Testing Isolates Not Represented in the Database, Robustness Testing

When inoculated on the MicroScan panels, isolates not represented in the database were misidentified with high probability scores in half of the cases, and only in a minority of them (6.1%), the user was alerted about the impossibility of identification. Likewise, testing on inappropriate panels (Gram-positive isolates on the Gram-negative panel and vice versa) or using an incorrect algorithm regularly led to incorrect identification with high probabilities instead of the expected “no identification possible” result. By contrast, errors in interpretation of beta-hemolysis did not impact the identification of hemolytic Streptococcus species.

The experiences of misidentifications at robustness testing highlight the importance of accurate performance and interpretation of orienting tests such as Gram stain or “spot tests” such as catalase and oxidase [21,22]. This poses challenges in LRS, as laboratory staff is frequently not familiar with clinical bacteriology [5] and the Gram stain is error-prone in inexperienced hands [23,24,25,26,27]. Apart from training, supervision, and guidance by the laboratory information support system, careful selection and quality control of orienting tests is needed, as was shown by the present experience with oxidase brands. Based on the present study (Table 5), identification results such as Rhodococcus equi and Providencia rustigiannii (both very rare in clinical bacteriology) should raise alerts to incorrect identifications.

4.3. Comparison between Automated and Visual Reading, Repeatability and Inter-Observer Agreement

Visual reading was slightly more accurate than reading by the autoSCAN-4 reader, except for Streptococcus species. The species agreement of 87.5%, was, however, lower than the 95% reported for the previous version of the reader (autoSCAN-3) [28]. No studies reporting agreement between visual reading and autoSCAN-4 reader have been published, although some studies showed excellent accuracy of autoSCAN-4 for Gram-negative bacteria [29,30]. Agreement between repetitions was higher with automated reading, which was not unexpected given the subjectivity in the reading of certain wells. Repeatability was poor for Streptococcus species.

4.4. Adaptation to LRS: Stability, Ease-of-Use

The temperature stability of the MicroScan panels, currently assured up to 25 °C, does not fulfill the requirements for tropical settings because cool storage (<30 °C) is not always feasible [31]. The shelf life of the reconstituted VP2 is too short, particularly for laboratories processing small amounts of samples. In addition, the 95% ethanol needed for reconstitution is often not easy to find in LRS.

As to ease of use, strong points are the MicroScan inoculation system. On the downside, visual reading of panels was considered complicated, poorly amenable to a swift work-flow, and error-prone. In the absence of the automated reader, the dedicated prototype microplate viewer (allowing quick change from white to the black background) considerably facilitated the reading process. The customized panels provided all tests on one panel; however, Gram stain results are still needed for reading and interpretation. Added value was therefore considered limited.

The Flesch–Kincaid Grade Level scores of the IFU indicated that a high level of schooling is required to understand the IFU. For patient leaflets in LRS, Flesch–Kincaid levels below 6 are desirable [13,32]. Although the language used in professional documents may be more complex, it should be reminded that the IFU will often not be available in the user’s native language, therefore low complexity is still preferable.

To obtain identification results, the user must purchase the LabPro software or use the (freely available) online Biotype Lookup Tool. For results with low probability scores, confirmatory tests are proposed by the software; however, many of these tests are unlikely to be available in LRS (Table 7).

4.5. Recommendations for Use and Further Development of the MicroScan System

Within the Mini-Lab project, the current limitations of the panels and IFU will be tackled by training, adapted laboratory procedures, bench aids, and the microplate viewer for visual reading. The Mini-lab project will further use the present results to create an expert decision support system that will anticipate potential errors such as those related to Gram stain interpretation. These expert rules will be made accessible outside MSF. In addition, we encourage the manufacturer to mitigate to the best possible extent the risks described above. To build in an extra fail-safe for Gram stain errors, the development of control wells for Gram-positive, Gram-negative, and yeast growth (consisting of vancomycin, colistin, and amphotericin B, respectively) has been proposed to the manufacturer. We further suggest simplifying the IFU and harmonize it with packaging. Bench aids and improved worksheets could be included with the product. Accuracy can further be improved over time by adding more LRS pathogens and contaminants to the database. Lastly, extended shelf-life testing and stability testing in tropical environments are necessary to assure product quality in LRS.

4.6. Strengths and Limitations of the Study

To our knowledge, this is the first study evaluating the MicroScan system on clinical isolates from LRS and the largest evaluating both Gram-positive and Gram-negative organisms, among which typical LRS pathogens [33,34]. Furthermore, we assessed robustness and ease of use. Robustness testing of in vitro diagnostics is encouraged to anticipate end users’ challenges, particularly in LRS [35,36]. To our knowledge, other commercial phenotypic identification systems have not yet been tested for robustness.

Because we did not dispose of Biosafety Level-3 (BSL-3) facilities, we could not test pathogens such as B. pseudomallei and Brucella species. Laboratory technicians were not blinded to the isolate’s identification when interpreting results; however, we do not believe this led to bias in the interpretation of test wells because they were unaware of which tests were supposed to be positive for which species. Lastly, this evaluation was performed in a reference laboratory in optimal conditions, by experienced laboratory technicians, and results can therefore not be extrapolated to a field setting. Performance and usability should be evaluated prospectively in LRS settings by the intended end user.

5. Conclusions

Challenged with clinical isolates from LRS, MicroScan Dried Overnight Identification panels had excellent performance for Gram-negative organisms. For important Gram-positive pathogens, accuracy ranged from excellent (S. aureus) to fair (S. pneumoniae and S. anginosus). The study further identified potential improvements in stability, robustness, and ease of use to assure adaptation of the MicroScan system to the constraints of LRS, particularly outside the MSF Mini-lab setting.

Acknowledgments

We thank Kaat Eggermont, Tine Vermoesen, and Marta Oller Sala for the laboratory work involved in this study, and the diagnostic laboratory of the ITM travel clinic, and our partners of the surveillance studies for providing the clinical isolates. Furthermore, we are indebted to Birgit De Smet for her valuable feedback on the use of the panels in the field. Last but not least, we would like to thank the Mini-lab team and the Mini-lab scientific committee for their continued and valued scientific and practical input.

Supplementary Materials

The following are available online at https://www.mdpi.com/2075-4418/11/2/349/s1, Table S1: Lot numbers of tested panels, Table S2: Geographical origin of the isolates, Table S3: Detailed results for incorrect use of the panels, Table S4: Agreement between visual and automated (autoSCAN-4) reading, Table S5: Results for inter-observer agreement between two independent readers who read the panels visually.

Author Contributions

Conceptualization, S.O., A.N., J.-B.R., O.V., L.H., and J.J.; data curation, S.O.; formal analysis, S.O.; funding acquisition, J.-B.R. and J.J.; methodology, S.O., A.N., L.H., and J.J.; project administration, S.O., A.N., and J.-B.R.; supervision, L.H. and J.J.; visualization, S.O.; writing—original draft preparation, S.O., L.H., and J.J.; writing—review and editing, A.N., J.-B.R., and O.V. All authors have read and agreed to the published version of the manuscript.

Funding

This study was funded by Médecins Sans Frontières (Mini-lab project).

Institutional Review Board Statement

Because the study did not involve research on human or animal subjects, data, or material, ethical clearance was waived. Bacterial isolates were obtained from ITM travel clinic and microbiological surveillance studies (surveillance of antimicrobial resistance among consecutive blood culture isolates in tropical settings, Institutional Review Board number 613/08, Ethical Committee Antwerp University Hospital 8/20/96).

Informed Consent Statement

Not applicable (study did not involve human subjects).

Data Availability Statement

The database for this manuscript will be made open access. Access requests for ITM research data can be made to ITM’s central point for research data access by means of submitting the completed Data Access Request Form. These requests will be reviewed for approval by ITMs Data Access Committee (https://www.itg.be/E/data-sharing-open-access).

Conflicts of Interest

The authors declare no conflict of interest. The manufacturer of the panels had no role in the design of the study, in the collection, analyses, or interpretation of data, in the writing of the manuscript, or in the decision to publish the results.

Footnotes

Publisher’s Note: MDPI stays neutral with regard to jurisdictional claims in published maps and institutional affiliations.

References

  • 1.Ombelet S., Ronat J.B., Walsh T., Yansouni C.P., Cox J., Vlieghe E., Martiny D., Semret M., Vandenberg O., Jacobs J., et al. Clinical bacteriology in low-resource settings: Today’s solutions. Lancet Infect. Dis. 2018;18:e248–e258. doi: 10.1016/S1473-3099(18)30093-8. [DOI] [PubMed] [Google Scholar]
  • 2.Petti C.A., Polage C.R., Quinn T.C., Ronald A.R., Sande M.A. Laboratory Medicine in Africa: A Barrier to Effective Health Care. Clin. Infect. Dis. 2006;42:377–382. doi: 10.1086/499363. [DOI] [PubMed] [Google Scholar]
  • 3.The European Committee on Antimicrobial Susceptibility Testing Breakpoint Tables for Interpretation of MICs and Zone Diameters. Version 10.0. [(accessed on 18 February 2021)];2020 Available online: http://www.eucast.org.
  • 4.CLSI . M100: Performance Standards for Antimicrobial Susceptibility Testing. 30th ed. Clinical and Laboratory Standards Institute; Wayne, NJ, USA: 2020. [Google Scholar]
  • 5.Jacobs J., Hardy L., Semret M., Lunguya O., Phe T., Affolabi D., Yansouni C., Vandenberg O. Diagnostic Bacteriology in District Hospitals in Sub-Saharan Africa: At the Forefront of the Containment of Antimicrobial Resistance. Front. Med. 2019;6 doi: 10.3389/fmed.2019.00205. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6.Franco-Duarte R., Černáková L., Kadam S., Kaushik K.S., Salehi B., Bevilacqua A., Corbo M.R., Antolak H., Dybka-Stępień K., Leszczewicz M., et al. Advances in chemical and biological methods to identify microorganisms—From past to present. Microorganisms. 2019;7:130. doi: 10.3390/microorganisms7050130. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7.Leber A.L. Clinical Microbiology Procedures Handbook. ASM Press; Washington, DC, USA: 2016. [Google Scholar]
  • 8.Jorgensen J.H. In: Manual of Clinical Microbiology. 11th ed. Jorgensen J.H., Carroll K.C., Funke G., Pfaller M.A., editors. ASM Press; Washington, DC, USA: 2015. [Google Scholar]
  • 9.Sandle T. Pharmaceutical Microbiology. Elsevier Inc.; Amsterdam, The Netherlands: 2016. Microbial identification. [Google Scholar]
  • 10.Natale A., Ronat J.-B., Mazoyer A., Rochard A., Boillot B., Hubert J., Baillet B., Ducasse M., Mantelet F., Oueslati S., et al. The Mini-Lab: Accessible clinical bacteriology for low-resource settings. Lancet Microbe. 2020;1:56–58. doi: 10.1016/S2666-5247(20)30012-4. [DOI] [PubMed] [Google Scholar]
  • 11.Russell F.M., Biribo S.S.N., Selvaraj G., Oppedisano F., Warren S., Seduadua A., Mulholland E.K., Carapetis J.R. As a bacterial culture medium, citrated sheep blood agar is a practical alternative to citrated human blood agar in laboratories of developing countries. J. Clin. Microbiol. 2006;44:3346–3351. doi: 10.1128/JCM.02631-05. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12.Zomorodian K., Javad M., Safaei A., Bazargani A. Analysis of beta-hemolysis in human blood agars by Streptococcus pyogenes Analysis of beta-hemolysis in human blood agars by Streptococcus pyogenes. J. Microbiol. Methods. 2011;85:233–234. doi: 10.1016/j.mimet.2011.03.009. [DOI] [PubMed] [Google Scholar]
  • 13.WHO . Technical Guidance Series (TGS) for WHO Prequalification—Diagnostic Assessment Guidance on Test Guidance on Test Method Validation for In Vitro Diagnostic Medical Devices TGS–4. WHO; Geneva, Switzerland: 2017. [(accessed on 18 February 2021)]. pp. 1–23. Tech. Guid. Ser. WHO Prequalification—Diagnostic Assess. Available online: http://apps.who.int/bookorders.%0Ahttp://apps.who.int/iris/bitstream/10665/258971/1/WHO-EMP-RHT-PQT-TGS4-2017.04-eng.pdf?ua=1. [Google Scholar]
  • 14.Jin W.Y., Jang S.J., Lee M.J., Park G., Kim M.J., Kook J.K., Kim D.M., Moon D.S., Park Y.J. Evaluation of VITEK 2, MicroScan, and Phoenix for identification of clinical isolates and reference strains. Diagn. Microbiol. Infect. Dis. 2011;70:442–447. doi: 10.1016/j.diagmicrobio.2011.04.013. [DOI] [PubMed] [Google Scholar]
  • 15.Rhoads S., Marinelli L., Imperatrice C.A., Nachamkin I. Comparison of MicroScan WalkAway system and Vitek system for identification of gram-negative bacteria. J. Clin. Microbiol. 1995;33:3044–3046. doi: 10.1128/JCM.33.11.3044-3046.1995. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 16.Snyder J.W., Munier G.K., Johnson C.L. Direct comparison of the BD phoenix system with the MicroScan WalkAway system for identification and antimicrobial susceptibility testing of Enterobacteriaceae and nonfermentative gram-negative organisms. J. Clin. Microbiol. 2008;46:2327–2333. doi: 10.1128/JCM.00075-08. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 17.Chatzigeorgiou K.S., Sergentanis T.N., Tsiodras S., Hamodrakas S.J., Bagos P.G. Phoenix 100 versus Vitek 2 in the identification of gram-positive and gram-negative bacteria: A comprehensive meta-analysis. J. Clin. Microbiol. 2011;49:3284–3291. doi: 10.1128/JCM.00182-11. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18.O’Hara C.M. Manual and Automated Instrumentation for Identication of Enterobacteriaceae and Other Aerobic Gram-Negative Bacilli. Clin Microbiol Rev. 2005;18:147–162. doi: 10.1128/CMR.18.1.147-162.2005. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19.Von Baum H., Klemme F.R., Geiss H.K., Sonntag H.G. Comparative evaluation of a commercial system for identification of gram-positive cocci. Eur. J. Clin. Microbiol. Infect. Dis. 1998;17:849–852. doi: 10.1007/s100960050205. [DOI] [PubMed] [Google Scholar]
  • 20.Jesumirhewe C., Ogunlowo P.O., Olley M., Springer B., Allerberger F., Ruppitsch W. Accuracy of conventional identification methods used for Enterobacteriaceae isolates in three Nigerian hospitals. PeerJ. 2016;2016:1–12. doi: 10.7717/peerj.2511. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 21.Baron E.J. Rapid Identification of Bacteria and Yeast: Summary of a National Committee for Clinical Laboratory Standards Proposed Guideline. Clin. Infect. Dis. 2001;33:220–225. doi: 10.1086/321816. [DOI] [PubMed] [Google Scholar]
  • 22.Baron E.J., York M.K., Ferraro M.J., Rex J.H., Body B.A., Forbes B.A., Poole F.M., Sahm D.F., Tenover F.C., Turnidge J.D., et al. M35-A2 Abbreviated Identification of Bacteria and. 2nd ed. CLSI; Wayne, NJ, USA: 2008. [Google Scholar]
  • 23.Rand K.H., Tillan M. Errors in interpretation of gram stains from positive blood cultures. Am. J. Clin. Pathol. 2006;126:686–690. doi: 10.1309/V4KE2FPM5T8V4552. [DOI] [PubMed] [Google Scholar]
  • 24.Samuel L.P., Balada-Llasat J.M., Harrington A., Cavagnolo R. Multicenter assessment of gram stain error rates. J. Clin. Microbiol. 2016;54:1442–1447. doi: 10.1128/JCM.03066-15. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 25.Munson E., Block T., Basile J., Hryciuk J.E., Schell R.F. Mechanisms to assess Gram stain interpretation proficiency of technologists at satellite laboratories. J. Clin. Microbiol. 2007;45:3754–3758. doi: 10.1128/JCM.01632-07. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 26.Chandler, L. Challenges in Clinical Microbiology Testing. 1st ed. Elsevier Inc.; Amsterdam, The Netherlands: 2013. [Google Scholar]
  • 27.Yuan S., Astion M.L., Schapiro J., Limaye A.P. Clinical impact associated with corrected results in clinical microbiology testing. J. Clin. Microbiol. 2005;43:2188–2193. doi: 10.1128/JCM.43.5.2188-2193.2005. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 28.Ellner P.D., Myers D.A. Preliminary evaluation of the autoSCAN-3, an instrument for automated reading and interpretation of microdilution trays: Identification of aerobic gram-negative bacilli. J. Clin. Microbiol. 1981;14:326–328. doi: 10.1128/JCM.14.3.326-328.1981. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 29.Rhoden D.L., Smith P.B., Baker C.N., Schable B. autoSCAN-4 system for identification of gram-negative bacilli. J. Clin. Microbiol. 1985;22:915–918. doi: 10.1128/JCM.22.6.915-918.1985. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 30.Gavini F., Husson M.O., Izard D., Bernigaud A., Quiviger B. Evaluation of Autoscan-4 for identification of members of the family Enterobacteriaceae. J. Clin. Microbiol. 1988;26:1586–1588. doi: 10.1128/JCM.26.8.1586-1588.1988. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 31.World Health Organization—R&D Blueprint . COVID-19 Target Product Profiles for Priority Diagnostics to Support Response to the COVID-19 Pandemic v.0.1. WHO; Geneva, Switzerland: 2020. [Google Scholar]
  • 32.Gillet P., Maltha J., Hermans V., Ravinetto R., Bruggeman C., Jacobs J. Malaria rapid diagnostic kits: Quality of packaging, design and labelling of boxes and components and readability and accuracy of information inserts. Malar. J. 2011;10:39. doi: 10.1186/1475-2875-10-39. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 33.Reddy E.A., Shaw A.V., Crump J.A. Community-acquired bloodstream infections in Africa: A systematic review and meta-analysis. Lancet Infect. Dis. 2010;10:417–432. doi: 10.1016/S1473-3099(10)70072-4. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 34.Deen J., von Seidlein L., Andersen F., Elle N., White N.J., Lubell Y. Community-acquired bacterial bloodstream infections in developing countries in south and southeast Asia: A systematic review. Lancet Infect. Dis. 2012;12:480–487. doi: 10.1016/S1473-3099(12)70028-2. [DOI] [PubMed] [Google Scholar]
  • 35.WHO Technical Specification Series. [(accessed on 18 February 2021)];2020 Available online: https://www.who.int/diagnostics_laboratory/guidance/technical-specifications-series/en/
  • 36.Gerth-Guyette E., Malacad C.C., Demonteverde M.P., Faulx D., Lochhead M.J., Lupisan S.P., Leader B.T., Tallo V.L. Understanding user requirements to improve adoption of influenza diagnostics in clinical care within Metro Manila. Health Sci. Rep. 2018;1:1–8. doi: 10.1002/hsr2.75. [DOI] [PMC free article] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

Data Availability Statement

The database for this manuscript will be made open access. Access requests for ITM research data can be made to ITM’s central point for research data access by means of submitting the completed Data Access Request Form. These requests will be reviewed for approval by ITMs Data Access Committee (https://www.itg.be/E/data-sharing-open-access).


Articles from Diagnostics are provided here courtesy of Multidisciplinary Digital Publishing Institute (MDPI)

RESOURCES