Distribution of Paclitaxel Particles After Drug-Coated Balloon Treatment ― Comparison Between Intravascular Ultrasound Images and Electronic High-Resolution Angioscopy ―

Madoka Ihara 1,, Yuhei Nojima 1, Nobuyuki Koh 1, Hidenori Adachi 1, Tetsuya Kurimoto 1, Keita Okayama 2, Yasushi Sakata 2, Shinsuke Nanto 1
PMCID: PMC7932808  PMID: 33693189

We observed the distribution of paclitaxel particles on the vascular wall after treatment with a drug-coated balloon (DCB; IN.PACTTM AdmiralTM DCB; Medtronic) using an electronic high-resolution angioscope (Zemporshe®; OVALIS, Osaka, Japan) and intravascular ultrasound (IVUS; Opticross HD; Boston Scientific).

A 78-year-old woman underwent endovascular treatment for chronic total occlusion of the right mid-superficial femoral artery (SFA). Zemporshe® angioscopy and IVUS were performed at the time of predilatation with a balloon (5 mm×60 mm) and DCB (6 mm×80 mm) deployment at 8 atm for 300 s. Compared with both IVUS images (Figure), a hyperechoic dotted line appeared along the lumen surface after DCB deployment. Matching this to the angioscopic image for this patient, the dotted line is expected to correspond to paclitaxel particles. The IVUS images detected paclitaxel particles in the form of a homogeneous and circumferential line. IVUS should be useful in evaluating drug distribution.

Figure.

Figure.

(A,B) Angiographic images after predilatation with a balloon (A) and drug-coated balloon (DCB) deployment (B). (C,D) Angioscopic images after predilatation with a balloon (C) and drug-coated balloon (DCB) deployment (D). Paclitaxel particles were adhered to the vascular wall. (E,F) Intravascular ultrasound images after predilatation (E), showing dissection (arrows), and after DCB deployment (F), showing a hyperechoic dotted line along the lumen surface (arrowheads).

The distribution of paclitaxel particles after DCB treatment has been reported previously using angioscopic images.1,2 However, to the best of our knowledge, the present study is the first to report that IVUS images can be used to show paclitaxel particles.

Sources of Funding

This study did not receive any specific funding.

Disclosures

None.

References

  • 1. Ihara M, Nojima Y, Adachi H, Kurimoto T, Okayama K, Sakata Y, et al.. Distribution of paclitaxel particles on the vascular wall after drug-coated balloon treatment: Electronic high-resolution angioscopy assessment. Circ Rep 2019; 1: 240. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 2. Ihara M, Nojima Y, Adachi H, Kurimoto T, Okayama K, Sakata Y, et al.. Paclitaxel particle distribution 3 months after drug-coated balloon treatment: Zemporshe® angioscopy assessment. Circ Rep 2020; 2: 135–136. [DOI] [PMC free article] [PubMed] [Google Scholar]

Articles from Circulation Reports are provided here courtesy of The Japanese Circulation Society

RESOURCES