. 2021 Mar 4;11:5162. doi: 10.1038/s41598-021-83986-4

3D Mueller matrix mapping of layered distributions of depolarisation degree for analysis of prostate adenoma and carcinoma diffuse tissues

Volodymyr A Ushenko 1, Benjamin T Hogan 2, Alexander Dubolazov 1, Gennadii Piavchenko 3, Sergey L Kuznetsov 3, Alexander G Ushenko 1, Yuriy O Ushenko 1, Mykhailo Gorsky 1, Alexander Bykov 2, Igor Meglinski 2,3,4,5,
PMCID: PMC7933337  PMID: 33664274

Abstract

Prostate cancer is the second most common cancer globally in men, and in some countries is now the most diagnosed form of cancer. It is necessary to differentiate between benign and malignant prostate conditions to give accurate diagnoses. We aim to demonstrate the use of a 3D Mueller matrix method to allow quick and easy clinical differentiation between prostate adenoma and carcinoma tissues with different grades and Gleason scores. Histological sections of benign and malignant prostate tumours, obtained by radical prostatectomy, were investigated. We map the degree of depolarisation in the different prostate tumour tissues using a Mueller matrix polarimeter set-up, based on the superposition of a reference laser beam with the interference pattern of the sample in the image plane. The depolarisation distributions can be directly related to the morphology of the biological tissues. The dependences of the magnitude of the 1st to 4th order statistical moments of the depolarisation distribution are determined, which characterise the distributions of the depolarisation values. To determine the diagnostic potential of the method three groups of histological sections of prostate tumour biopsies were formed. The first group contained 36 adenoma tissue samples, while the second contained 36 carcinoma tissue samples of a high grade (grade 4: poorly differentiated—4 + 4 Gleason score), and the third group contained 36 carcinoma tissue samples of a low grade (grade 1: moderately differentiated—3 + 3 Gleason score). Using the calculated values of the statistical moments, tumour tissues are categorised as either adenoma or carcinoma. A high level (> 90%) accuracy of differentiation between adenoma and carcinoma samples was achieved for each group. Differentiation between the high-grade and low-grade carcinoma samples was achieved with an accuracy of 87.5%. The results demonstrate that Mueller matrix mapping of the depolarisation distribution of prostate tumour tissues can accurately differentiate between adenoma and carcinoma, and between different grades of carcinoma. This represents a first step towards the implementation of 3D Mueller matrix mapping for clinical analysis and diagnosis of prostate tumours.

Subject terms: Nanoscale biophysics, Polarization microscopy, Biophysical methods, High-throughput screening, Microscopy, Imaging, 3-D reconstruction, Circular dichroism, Structure determination, Cancer imaging, Cancer screening, Cancer imaging, Statistics, Characterization and analytical techniques, Imaging techniques, Applied optics, Optical physics, Nanophotonics and plasmonics, Nonlinear optics, Imaging and sensing, Optical techniques, Circular dichroism, Other photonics, Biophotonics, Biophysics, Nanoscience and technology, Optics and photonics, Applied physics, Biological physics, Techniques and instrumentation, Physics, Optical physics, Nanophotonics and plasmonics

Introduction

Prostate cancer is the second most common cancer globally in men, and in some countries is now the most diagnosed form of cancer1,2. Early diagnosis, intervention, and management can give significantly improved patient outcomes3,4. It is necessary to differentiate between benign (adenoma) and malignant (carcinoma) prostate tissue types, and then to further differentiate between grades of tumour tissues4. Optical analysis of tissue types offers distinct advantages, being non-destructive, requiring limited sample preparation, and being cheap and fast510. Analysis of the polarisation properties- known as polarimetric diagnostics- of different tissues holds great promise. Polarimetric diagnostics of optically anisotropic biological tissues are undergoing active development in biomedical optics1115. Several key strands of investigation have emerged, including: scattering matrices5,16,17; Mueller matrix polarimetry10,1823; polar decomposition of Mueller matrices24,25; and two-dimensional Mueller matrix mapping2629 using various approximations69.

Mueller matrix mapping is particularly promising. Experimental and analytical results obtained using Mueller matrices were traditionally represented as 1D angular dependences (indicatrices) of matrix elements10,13,24. The next step in the development of the Mueller matrix polarimetry techniques was the use of digital cameras to obtain of 2D Mueller matrix images2123, i.e. 2D distributions of the elements of the Mueller matrices. For the 2D distributions, the information obtained about the optical properties of the object being diagnosed is integrally averaged over the entire volume of the sample. However, most biological tissues have complex, spatially heterogeneous, structures and a pronounced depolarising ability. Therefore, the further development of new Mueller matrix polarimetry techniques which can provide 3D information is a critical challenge. This work focuses on the development and experimental testing of a new method of 3D Mueller matrix mapping of the anisotropic polarisation properties. Digital holographic reconstruction of 3D layered distributions of the degree of depolarisation is used for express diagnosis and differentiation of diffuse samples of histological sections of prostate tumour biopsy obtained by radical prostatectomy.

Methods and theory

Biological samples

For determination of the type of tumour, native samples of histological sections of the examined prostate tissue were made. This study was conducted in accordance with the principles of the Declaration of Helsinki, and in compliance with the International Conference on Harmonisation-Good Clinical Practice and local regulatory requirements. Ethical approval was obtained from the Ethics Committee of the Bureau of Forensic Medicine of the Chernivtsi National University and the Bukovinian State Medical University (Chernivtsi, Ukraine), and written informed consent was obtained from all subjects prior to study initiation. The type of prostate tumour was determined by an independent assessment of stained histological samples (Fig. 1). The histological analysis was conducted by the following procedure:

  1. Fixation of prostate tissue with formalin (40% formaldehyde aqueous solution);

  2. Washing of samples in running water for 24 h;

  3. Dehydration with alcohols with increasing concentration (70–100%) within 48 h;

  4. Fixing the material in a xylene-paraffin mixture for 1–2 h, at a temperature of 52°–56°, and cutting out a block with a sample enclosed in it;

  5. Making histological sections on a standard freezing microtome;

  6. Staining of histological sections with haematoxylin and eosin (Fig. 1 shows real colour microscopic images);

  7. Microscopic examination of images of the obtained samples with differentiation of their structure by grades and determination of the position of the prostate tumor sample according to the Gleason Pattern scale.

Figure 1.

Figure 1

Representative real colour microscopic images (magnification ×50, biological microscope Ulab XY-B2T LED, digital camera UCMOSS08000KPB) of: (a) a prostate adenoma sample, (b) a moderately differentiated (3 + 3) prostate carcinoma sample, and (c) a poorly differentiated (4 + 4) prostate carcinoma sample.

Three representative groups of histological biopsy sections, obtained from radical prostatectomy, were formed:

  • Group 1 consisted of n=36 adenoma samples;

  • Group 2 consisted of n=36 moderately differentiated (3 + 3 on Gleason’s pattern scale) carcinoma samples;

  • Group 3 consisted of n=36 poorly differentiated (4 + 4 on Gleason’s Pattern scale) carcinoma samples.

Table 1 presents the optical and geometric parameters of the samples of native histological sections of prostate tumour biopsies from each of the groups. The geometric thickness of the histological sections was determined according to the standard values of the freezing microtome scale. The extinction coefficient was determined by measuring the attenuation of the illuminating beam intensity as it passed through the sample. This was achieved using an integral light-scattering sphere following a well-established method30, 31. The measurement of the integral degree of depolarisation (Λ) was performed using a standard Mueller matrix polarimeter in accordance with similar measurements previously underatken23.

Table 1.

Optical and geometric parameters of histological sections of prostate tissue.

Parameter Group 1 Group 2 Group 3
Geometric thickness h,μm 40 ± 0.95 40 ± 0.95 40 ± 0.95
Attenuation (extinction) coefficient τ,cm-1 2.01 ± 0.041 1.96 ± 0.037 1.99 ± 0.039
Depolarisation degree Λ,% 81 ± 0.78 83 ± 0.81 79 ± 0.73

By comparing the optical and geometric parameters of the different histological section groups, we observe that the depolarisation degree is high for all samples. Additionally, there is little difference in the depolarisation values for the different groups. As discussed below, this significantly limits the diagnostic potential of traditional 2D Mueller matrix polarimetry.

Theoretical background

The diagnostic efficiency of detection of oncological conditions by applying the technique of 2D Mueller matrix polarimetry to the polycrystalline component of optically thin (non-depolarising) histological sections of biological tissues of various human organs has previously been demonstrated3235. These previous studies are devoted to the search for, and subsequent diagnostic application of, a set of diagnostically relevant relationships between Mueller matrix images (Fikx,y) and maps of linear and circular birefringence (LBx.y, CBx,y) and dichroism (LDx.y, CDx.y):

Fx·y=F11F12F13F14F21F22F23F24F31F32F33F34F41F42F43F44x·yLBCBLDCDx·y. 1

As a general rule, samples of biological tissues of human organs are diffuse and thus strongly depolarize optical radiation (as is the case with our prostate tissue samples). This significantly limits the potential for differential diagnostics by traditional 2D Mueller matrix polarimetry. In this situation, almost all off-diagonal (Fik,ik) elements of the Mueller matrix of the diffuse biological layer are significantly reduced5,17,21. However, the diagonal elements (Fik,i=k) are still diagnostically relevant. The superposition of the diagonal elements determines the overall value of depolarisation Λ for the optical radiation11,14,15.

Fik,ik0;Fik,i=k0; 2
F=F110000F220000F330000F44Λ; 3
Λ=1-13F22+F33+F44. 4

The value of the parameter Λ is an integral equivalent of the overall optical properties of the object under investigation.

Analysing (4), one can identify the following range of variation of the parameter Λ:

  • for an optically homogeneous isotropic layer (F22;33;441), and thus Λ0;

  • for an ideal diffuser (F22;33;440), and thus Λ1.

In all other cases, the value of the integral depolarisation parameter is determined by a combination of two components. Firstly, the ‘local’ component resulting from the formation of an orthogonal component of the laser radiation amplitude (i.e. a change in the initial state of polarisation) due to the optical anisotropy of the biological layer. We will call this the “A-mechanism”. Secondly, the ‘diffuse’ component resulting from the statistical averaging of the polarisation state due to the superposition of laser waves, scattered in the volume of the biological layer, with different states of polarisation. We will call this the “B-mechanism”.

For optically thin (τ0.01), singly-scattering, optically anisotropic biological layers, the A-mechanism is dominant. The influence of the optical anisotropy (linear and circular birefringence and dichroism) is seen through a slight decrease in the values of the diagonal elements F22;33;44<1 of the Mueller matrix23. Therefore, in comparison with an optically isotropic layer, there is an increase in the value of depolarisation (Λ>0). The specific value of this parameter is interrelated with the specificity of the polycrystalline structure of biological tissues10,32,35.

As the optical thickness increases (τ>0.01), the multiplicity of the light scattering (i.e. the number of scattering events each photon experiences on average) in the volume of the histological sections increases correspondingly. The value of the integral depolarisation parameter Λ is determined by the superposition of the effects of the A- and B-mechanisms. For small values of the attenuation coefficient (τ0.5), the contribution of the two optical mechanisms is comparable. Averaging the polarisation states of the scattered (differently polarized) wavefronts leads to a further increase in the degree of integral depolarisation. There is hence a decrease in the dependence of integral depolarisation on the specific morphological structure of biological tissue.

For diffuse (optically thick, τ>1.5) biological layers, the polarising effects of the individual features of the polycrystalline structure are effectively smoothed by the B-mechanism. Therefore, the effectiveness of differential 2D diagnostics is extremely low or unsatisfactory in this limit. The 2D polarisation diagnostics of frozen prostate tissue samples is limited by the impossibility of obtaining optically thin (h10-15μm) histological sections with geometric consistency between samples. To obtain consistent samples, a geometric thickness h40μm was required. Hence, the B-mechanism is dominant for all samples, with the diffuse scattering resulting in both high and comparable values of the integral depolarisation for all the types of prostate tissues investigated (Table 1).

Experimental setup

The method of 3D Mueller matrix mapping is based on the use of a reference wave of laser radiation. The reference wave is superimposed on the polarisation-inhomogeneous image of the biological layer using an optical interferometry scheme (Fig. 2).

Figure 2.

Figure 2

Optical scheme of 3D Mueller matrix polarimetry. Here: 1—laser; 2—collimator; 3—beam splitter; 4,5—reflecting mirrors; 6,8,11,13,14—polarisers; 7,12—quarter-wave phase plates; 9—object; 10—strain-free polarisation objective; 15—digital camera; 16—processing unit.

Experimentally herein, the parallel (=2×103μm) beam of a He–Ne (λ=0.6328μm) laser (1, see Fig. 2) (Edmund Optics Lumentum High Performance Helium–Neon Laser, 5mW), formed by passing through a collimator (2, see Fig. 2), is divided by a 50/50 beam splitter (3, see Fig. 2) into separate illuminating and reference beams. The illuminating beam is directed by a mirrors (5, see Fig. 2) through a polarising (11–13, see Fig. 2) consisting of two linear polarisers (Edmund Optics TechSpec High Contrast Glass Linear Polarizer) (11 and 13, see Fig. 2) and a quarter wave plate (Astropribor Achromatic Quarterwave Plate APAW-20) (12, see Fig. 2). The polarised illuminating beam then passes through the sample (9, see Fig. 2). A polarisation-inhomogeneous image of the object is projected through a strain-free microscopic objective(10, see Fig. 2) (Nikon CFI Achromat P, 4x) into the imaging plane of the digital camera (15, see Fig. 2) (The Imaging Source DMK 41AU02.AS, monochrome 1/2" CCD, Sony ICX205AL). The reference beam is directed by a mirror (14, see Fig. 2) through an equivalent polarising filter (6–8, see Fig. 2) (optical components are the same as (11–13, see Fig. 2)) into the plane of the polarisation-inhomogeneous image of the object. The angle between the illuminating and reference beams was selected as Ψ=4. As a result, an interference pattern is formed, the period of which is 10μm. According to the Nyquist–Shannon sampling theorem, this periodicity ensures reliable recording of a single fringe by 5 pixels of the digital camera (15, see Fig. 2).

The error in setting the transmission axes of the polarisers8,11,13,14,36 was Δγ=0.2. The phase shift error of the quarter-wave plates13,37 was Δδ=0.5. The error in setting the fast axis of the quarter-wave plates was Δφ=0.2. The resulting error in the linear polarisation states did not exceed 0.0021, and that in the circular polarisation states did not exceed 0.005638.

Before carrying out measurements on histological sections of the prostate, metrological certification of the experimental setup was conducted using model objects (air, linear polariser, quarter- and half-wave plates). From 50 measurements for each model object, the errors in the determination of the diagonal elements of the Mueller matrix were determined: for F22;33- 0.25% and for F44- 0.5%.

Determination of the layered depolarisation distributions

When obtaining the polarisation interference images, one can determine the Mueller matrix elements by the following procedure:

  1. Forming six distinct polarisation states of both the illuminating and reference beams: four linear polarisations at angles of 0°, 90°, 45°, and 135°, and two circular polarisations—right () and left () circular polarisation respectively.

  2. Collection of six images for each partial interference pattern passed through the polariser-analyser [14] with sequential orientation of its transmission axis at angles Ω0=0;Ω90=90. We then have a series of twelve digital photo of the original field with the applied reference wave.

  3. For each partial interference distribution, we perform a two-dimensional discrete Fourier transform (DFTυ,ν) on the image. The two-dimensional DFTυ,ν of a two-dimensional array I0;90x,y (i.e. the image) f(x,y)is a function of two discrete variables coordinates x,y the DFTυ,ν function is defined by by39:
    DFTΩ=0;900;90;45;135;;υ,ν=1M×Nx=0M-1y=0N-1IΩ=0;90x,yexp-i2πx×υM+y×νN 5
    where I0;90x,y=A0;90x,yA0;90x,y are the coordinate distributions of the intensity of the interference pattern (for each polarisation state of the illuminating and reference beams) filtered by the analyser with the orientation of its transmission axis at Ω0=0 and Ω90=90; A0;90x,y are the orthogonal projections of the complex amplitudes; denotes the complex conjugation operation; υ,ν are the spatial frequencies in the x and y directions respectively; and M,N are the number of pixels of the CCD camera in the x and y directions respectively, such that 0x,υM and 0y,νN.
  4. The results of this transformation should contain three peaks, one central (main) peak and two additional side peaks. The Fourier transform acts like a low-pass filter. It removes interference fringes, which enables extraction of the complex representation of the real field due to the object. Also, since the extracted part has limited size, it acts like a low-pass filter for the object field too.

  5. Either of the additional side peaks (in complex representation) can be used to create a new Fourier spectrum by first extracting the peak and then placing it into centre of a newly generated spectrum DFTΩ=0;900;90;45;135;;υ,ν.

  6. Applying a two-dimensional inverse discrete Fourier transform DFTΩ=0;900;90;45;135;;-1x,y on the obtained spectrum DFTΩ=0;900;90;45;135;;υ,ν, one gets:
    DFTΩ=0;900;90;45;135;;-1x,y=1M×Nx=0M-1y=0N-1DFTΩ=0;900;90;45;135;;υ,νexp-i2πx×υM+y×νN. 6
    Here, DFTΩ=0;900;90;45;135;;-1x,yAΩ=0;900;90;45;135;;x,y.
  7. One subsequently obtains (for each polarisation state) a distribution of complex amplitudes:
    Ω0A0Ω90A90expiδ90-δ0 7
    in different phase planes (which are subplanes of the image). The phase planes are defined by:
    θk=δ90-δ0=2πλΔnz;0zh 8
    where Δn is the birefringence; λ is the wavelength; and h is the sample thickness of the object field, separated by an arbitrary step of Δθ.
  8. In each phase plane θk, the corresponding sets of parameters of the Stokes vector of the object field of the biological layer are calculated:
    VS10;90;45;135;;=A02+A9020;90;45;135;;;VS20;90;45;135;;=A02-A9020;90;45;135;;;VS30;90;45;135;;=2A02A902cosθk0;90;45;135;;;VS40;90;45;135;;=2A02A902sinθk0;90;45;135;;. 9

The obtained relations (9) form the basis of the method for determining layered distributions of the values of the Mueller matrix elements Fikx,y,θk. The elements are obtained from the following relationships:

  1. For Stokes vectors of linearly polarized probe beams, VS00;VS090:
    VS0=F1100=F11+F12F21+F22F31+F32F41+F42;VS90=F1-100=F11-F12F21-F22F31-F32F41-F42θkFikθk=F11F12F21F22F31F32F41F42θk. 10
  2. For Stokes vectors of linearly polarized probe beams VS045;VS0135:
    VS45=F1010=F11+F13F21+F23F31+F33F41+F43;VS135=F10-10=F11-F13F21-F23F31-F33F41-F43θkFikθk=F11F13F21F23F31F33F41F43θk. 11
  3. For Stokes vectors of right- and left-circularly polarized probe beams VS0;VS0:
    VS=F1001=F11+F14F21+F24F31+F34F41+F44;VS=F100-1=F11-F14F21-F24F31-F34F41-F44θkFikθk=F11F14F21F24F31F34F41F44θk. 12

From (9)–(12), working relations for determining the values of the elements of the Mueller matrix in the phase section θk are obtained:

Fθj=0.5VS10+VS190VS10-VS190VS145-VS1135VS1-VS1VS20+VS290VS20-VS290VS245-VS2135VS2-VS2VS30+VS390VS30-VS390VS345-VS3135VS3-VS3VS40+VS490VS40-VS490VS445-VS4135VS4-VS4θj 13

Finally, the distribution of the overall degree of depolarisation in each phase section is determined by the following relation:

Λx,y,θk=1-13VS20-VS290+VS345-VS3135+VS4-VS4 14

Statistical assessment of the polarisation distributions

The distributions of the values Λx,y,θk can be quantitatively assessed by calculating the set of central statistical moments of the first to fourth orders Zi=1;2;3;414,32,34 in each phase plane θk:

Z1=1Kj=1KΛj; 15
Z2=1Kj=1K(Λ-Z1)j2; 16
Z3=1Z231Kj=1K(Λ-Z1)j3; 17
Z4=1Z241Kj=1K(Λ-Z1)j4. 18

Here, K is the total number of pixels of the CCD-camera. These parameters characterize the mean value (Z1), dispersion (Z2), skewness (Z3) and kurtosis (Z4) of the distributions Λx,y,θk.

To determine the statistical significance of a representative sampling of the number of samples by the cross-validation method39, the standard deviation σ2 of each of the calculated values of the central statistical moments Zi=1;2;3;4n (15–18), which characterise the distribution of the degree of depolarisation Λ (4), was determined. The specified number (36 for each group) of samples provided the level σ20.025. This standard deviation corresponds to a confidence interval p<0.05, which demonstrates the statistical reliability of the 3D Mueller matrix mapping method for layered depolarisation maps within a representative sample.

Diagnostic method

Our proposed optical procedure, used herein for the differential diagnosis of the prostate tissue samples, comprised the following steps:

  1. We determined the optimal phase plane for further analysis and diagnostics. A "macro" step of discrete phase scanning was chosen—Δθkmax=0.25rad. Algorithmically, a series of layered coordinate distributions of the depolarisation degree Λx,y,θk corresponding to each Δθkmax=0.25rad was reconstructed.

  2. For each phase section θkmax of the object field of scattered radiation, the set of central statistical moments of the first to fourth orders Zi=1;2;3;4, which characterize the distributions Λx,y,θk, was calculated.

  3. The difference ΔZik=Ziθj+1max-Ziθjmax between the values of each of the statistical moments for the different phase planes was calculated.

  4. The phase interval Δθ=θj+1max-θjmax was determined, within which the monotonic growth of the value ΔZi=Ziθj+1max-Ziθjmax0 stops.

  5. Within the limits of Δθ, a new series of values ΔZi=Ziθq+1min-Ziθqmin was calculated with a discrete phase scanning “micro” step Δθqmin=0.05rad.

  6. The optimal phase plane θ was determined, in which ΔZiθ=max. The specified analytical procedure for one sample of prostate tissue took less than 7 min. A more accurate phase scanning algorithm can be chosen, for example with a step of 0.01 rad. In this case the processing time increased to 11 min.

  7. In the plane θ, the mean ΔZ¯i=1;2;3;4 and standard deviations σΔZi were determined within the representative samplings of histological sections from group 1, group 2 and group 3.

  8. To differentiate benign and malignant tumours, for each of the statistical moments Zi=1;2;3;4, the sensitivity (Se12=a12a12+b12100%; Se13=a13a13+b13100%), specificity (Sp12=c12c12+d12100%; Sp13=c13c13+d13100%) and balanced accuracy (Ac12=Se12+Sp122; Ac13=Se13+Sp132) were calculated. Here, a12 (a13) and b12 (b13) are the number of correct and incorrect diagnoses within group 2 (group 3); and c12 (c13) and d12 (d13) are the same within group 1.

  9. Similarly, to differentiate the grade of cancer, the sensitivity (Se23=a23a23+b23100%), specificity (Sp23=c23c23+d23100%) and balanced accuracy (Ac23=Se23+Sp232) were calculated for each of the statistical moments Zi=1;2;3;4. Here, a23 and b23 the number of correct and incorrect diagnoses within group 3; while c23 and d23 are the same within group 2.

The position of the optimal phase plane θ, determined by the indicated methods for each of the three groups separately, coincided with an accuracy of 0.02 rad. The operating characteristics of the method40 (sensitivity Se, specificity Sp, accuracy Ac) remained practically unchanged.

Results

Figure 3 shows the phase dependences of the integral magnitude (averaged over all the pixels of the CCD-camera) of the degree of depolarisation Λθk of laser radiation passing through prostate tissues from each of the three groups defined previously. The degree of depolarisation was experimentally determined using the 3D Mueller matrix mapping method described above.

Figure 3.

Figure 3

Dynamics of layered changes in the magnitude of the degree of depolarisation Λθk by samples of histological sections of adenoma (dark blue curve), well differentiated (blue curve) and poorly differentiated (red curve) prostate carcinoma.

Analysis of the data obtained (see Fig. 3) revealed that, despite the overall depolarisation through the whole samples being more or less the same (Table 1), there are significant differences observed within the volume of the tissue samples. The overall ranges of Λθk by histological biopsy sections of all samples span 15%-19%forθk=0.25radΛ78%-81%forθk=1.5rad. The rate at which the dependencies Λθk change is very different though. The most rapidly increasing depolarising ability is seen for samples of histological sections of adenoma biopsy (see Fig. 3, dark curve), while the least rapid is for samples of native sections of grade 4 carcinoma (see Fig. 3, red curve), with grade 1 carcinoma (see Fig. 3, blue line).

For small values of the phase (θk0.5rad), the dominant contribution to the depolarisation formation scenario is the “A—mechanism”. In general, soft biological tissues (including the prostate) have a relatively low level of optical anisotropy. The magnitude of the circular and linear birefringence Δn does not exceed 1.5×10-358. Therefore, for small values of the phase, the integral depolarisation is low for histological sections of all types of prostate tumours. With increasing θk, the contribution of the “B-mechanism” to the depolarising ability of the prostate samples increases. The contribution of the “B-mechanism” characterizes the measure of diffraction expansion of partial waves on optical inhomogeneities of biological tissue. The most pronounced effect of this mechanism manifests on small-scale (well differentiated) structures of the prostate tissue. Therefore, the most rapidly growing value of Λθk is found for samples of histological sections of prostate adenoma (see Fig. 3, dark curve), the morphological structure of which is at the smallest scale. Conversely, the values of Λθk for histological sections of biopsy of poorly differentiated prostate carcinoma increase most slowly (see Fig. 3, red curve).

For the largest values of phase shifts (θk1.5rad), the diffraction “B—mechanism”, which characterizes the multiple superposition of laser waves scattered in the volume of the biological layer, becomes dominant. The degree of depolarisation then reaches a maximum level for all types of samples. The maximum differences between the values of integral depolarisation by samples of histological sections of biopsy of prostate tumours are realized for a certain “intermediate” range of phase shifts (0.75rad<Δθ<1.05rad). The maximum difference is found at θ=0.85rad, so we choose this value for further investigations.

Figures 4 and 5 present exemplar depolarisation maps obtained for samples of prostate adenoma (see Fig. 4), and carcinoma with a 3 + 3 Gleason score (see Fig. 5). However, no obvious differences or relation to the tissue structure are immediately visible. For carcinoma with a higher (4 + 4) Gleason score, more obvious differences are visible. By analysing the results obtained, we can see that there is a complex and individual topographic structure of the depolarisation maps Λx,y,θk=0.85rad of native histological sections of prostate tumours obtained during radical prostatectomy (Fig. 6).

Figure 4.

Figure 4

3D surfaces (a) and 2D contours (b) of depolarisation maps Λx,y,θk=0.85rad of a sample of a native histological section of adenoma biopsy during radical prostatectomy.

Figure 5.

Figure 5

3D surfaces (a) and 2D contours (b) of depolarisation maps Λx,y,θk=0.85rad of a sample of a native histological biopsy section of a moderately differentiated (3 + 3 Gleason score) carcinoma during radical prostatectomy.

Figure 6.

Figure 6

3D surfaces (a) and 2D contours (b) of depolarisation maps Λx,y,θk=0.85rad of a sample of a native histological biopsy section of a poorly differentiated (4 + 4 Gleason score) carcinoma during radical prostatectomy.

We can then assess the depolarising ability of histological biopsy sections of benign and malignant prostate tumours within the framework of a statistical approach (15–18) to the analysis of the distributions Λx,y,θk. The largest mean and maximum range of coordinate fluctuations of the magnitude of the degree of depolarisation Λx,y,θk=0.85rad are found for histological sections of benign adenoma biopsies. The smallest mean and minimum range of coordinate fluctuations of the magnitude of the degree of depolarisation Λx,y,θk=0.85rad are found for histological sections of poorly differentiated carcinoma biopsies. The experimentally revealed differences in the series of maps Λx,y in the phase section θ=0.85rad can be associated with the different degrees of differentiation of the structures within the prostate tumour tissues. For well differentiated adenoma tissue, diffraction effects are most pronounced. Therefore, the average value and fluctuations of the degree of depolarisation are maximal. For poorly differentiated (4 + 4) carcinoma tissue, the diffraction broadening of laser waves is smaller. Therefore, both the mean and variance of the individual coordinate values Λx,y,θk=0.85rad decrease.

The results of quantitative statistical analysis and comparison of the distributions Λx,y,θk for the different groups of tissues are presented in Table 2. The data obtained shows that statistical moments of higher orders, which characterize the skewness (Z3θk) and kurtosis (Z4θk) of the distributions of the degree of depolarisation, are the most sensitive to changes in the structure of the depolarisation maps Λx,y,θk. Moreover, the maximum differences between the third and fourth order statistical moments (ΔZi=3;4max) are realised in the phase section θ=0.85max. In general, with increasing depolarisation the statistical moments can be characterised by Λx,y,θkZ1;2Λx,y,θkZ3;4Λx,y,θk.

Table 2.

Intergroup differences in the magnitude of the central statistical moments of the 1st–4th orders ΔZ¯i=1;2;3;4, which characterize the distributions of the magnitude of the degree of depolarisation Λx,y,θk in the layered phase sections corresponding to θk=0.5rad, θk=0.85rad, and θk=1.5rad.

graphic file with name 41598_2021_83986_Tab2_HTML.jpg

Table 3 shows the changes in the magnitude of the sensitivity Se, specificity Sp, and balanced accuracy Ac of the diagnostic power of the 3D Mueller matrix mapping method for native histological sections of benign and malignant prostate tumours.

Table 3.

Selectivity, specificity, and balanced accuracy of the diagnostic power of the 3D Mueller matrix mapping method of native histological sections of benign and malignant prostate tumours in the layered phase sections corresponding to θk=0.5rad, θk=0.85rad, and θk=1.5rad.

graphic file with name 41598_2021_83986_Tab3_HTML.jpg

In the region of small (θk0.5rad) and large (θk1.5rad) phase shifts, the diagnostic efficiency of the method is rather low -69.5%Ac12;13;23θk0.5rad75% and 52.8%Ac12;13;23θk1.5rad58.4%. However, in the diagnostically optimal (ΔZi=3;4max) phase section θ=0.85rad, a high efficiency was revealed not only for differential diagnosis of benign and malignant prostate tumours (90.3%Ac12;13θ=0.85rad93.1%), but also for predicting the severity of the oncological process (Ac23θ=0.85rad=87.5%).

Conclusions

The method of 3D Mueller matrix mapping of diffuse biological layers with digital holographic reconstruction of layered depolarisation maps (1–18) is analytically substantiated. This method, built on the platform of polarisation interferometry, was experimentally tested for the task of express (less than 15 min) differential diagnosis of prostate tumours obtained during radical prostatectomy. A representative sample of diffuse samples of and variously differentiated (Gleason scores 3 + 3 and 4 + 4) carcinoma biopsies was studied. Using the method of phase scanning of the object field of histological sections, the optimal cross section (θk=0.85rad) was determined. In this cross-section, the maximum differences (ΔZi=1;2;3;4max) between the values ​​of the central statistical moments Zi=1;2;3;4, which characterize the depolarisation maps of samples of benign and malignant prostate tumours, were realized. The most sensitive statistical parameters were established to be the skewness and kurtosis of the coordinate distributions Λx,y,θk. The operational characteristics (sensitivity Se, specificity Sp and balanced accuracy Ac) have been determined, which demonstrate the diagnostic power of the 3D Mueller matrix mapping of diffuse biological layers with digital holographic reconstruction of layered depolarisation maps. High accuracy (90.3%Ac12;13θ=0.85rad93.1%) of differentiation of benign and malignant samples of native histological sections of biopsy of prostate tumours was achieved. In addition, the possibility of diagnosing samples of variously differentiated malignant tumours with accuracy Ac23θ=0.85rad=87.5% has been demonstrated.

Author contributions

V.U., A.D., G.P., S.K., M.G., A.U., Y.U., and A.B. wrote the initial draft and prepared Fig. 1, 2 and 3. B.H. and I.M. revised the manuscript, and prepared Figs. 4, 5, and 6. All authors reviewed the manuscript and agreed to its submission.

Funding

This work received funding from: the ATTRACT project funded by the EC under Grant Agreement 777222; Academy of Finland (Grants 314639 and 325097); National Research Foundation of Ukraine, Project 2020.02/0061; and INFOTECH strategic funding. I.M. also acknowledges partial support from MEPhI Academic Excellence Project (Contract No. 02.a03.21.0005), and the National Research Tomsk State University Academic D.I. Mendeleev Fund Program.

Competing interests

The authors declare no competing interests.

Footnotes

Publisher's note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

References

  • 1.Mohler J, et al. Prostate cancer. J. Natl. Compr. Cancer Netw. 2010;8:162–200. doi: 10.6004/jnccn.2010.0012. [DOI] [PubMed] [Google Scholar]
  • 2.Grönberg H. Prostate cancer epidemiology. Lancet. 2003;361:859–864. doi: 10.1016/S0140-6736(03)12713-4. [DOI] [PubMed] [Google Scholar]
  • 3.Ilic D, Neuberger MM, Djulbegovic M, Dahm P. Screening for prostate cancer. Cochrane Database Syst. Rev. 2013;1:1465–1858. doi: 10.1002/14651858.CD004720.pub3. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4.Albertsen PC, Hanley JA, Fine J. 20-Year outcomes following conservative management of clinically localized prostate cancer. J. Am. Med. Assoc. 2005;293:2095–2101. doi: 10.1001/jama.293.17.2095. [DOI] [PubMed] [Google Scholar]
  • 5.Tuchin VV. Tissue optics and photonics: light-tissue interaction. J. Biomed. Photonics Eng. 2015;1:98–134. doi: 10.18287/JBPE-2015-1-2-98. [DOI] [Google Scholar]
  • 6.Vo-Dinh, T. Biomedical Photonics Handbook. Therapeutics and Advanced Biophotonics (CRC Press, 2014).
  • 7.Boas, D. A., Pitris, C. & Ramanujam, N. Handbook of Biomedical Optics. (CRC Press, 2011).
  • 8.Tuchin, V. V. Tissue Optics: Light Scattering Methods and Instruments for Medical Diagnosis: Third Edition (Society of Photo-Optical Instrumentation Engineers (SPIE), 2015). 
  • 9.Wang LV, Wu H-I. Biomedical Optics. Biomedical Optics: Principles and Imaging. Hoboken: Wiley; 2009. [Google Scholar]
  • 10.Ushenko AG, Pishak VP. Laser Polarimetry of Biological Tissues: Principles and Applications. In: Tuchin VV, Valeriĭ V, editors. Handbook of Coherent Domain Optical Methods. New York: Springer US; 2004. pp. 93–138. [Google Scholar]
  • 11.Ghosh N, Vitkin A. Tissue polarimetry: concepts, challenges, applications, and outlook. J. Biomed. Opt. 2011;16:110801. doi: 10.1117/1.3652896. [DOI] [PubMed] [Google Scholar]
  • 12.Meglinski, I. et al.Shedding the Polarized Light on Biological Tissues (Springer Briefs in Applied Science and Technology, Springer, Singapore, 2021). 
  • 13.Bachinskyi, V.T. et al.Polarization Correlometry of Scattering Biological Tissues and Fluids (Springer Briefs in Physics, Springer, Singapore, 2020).
  • 14.Novikova T, Meglinski I, Romella-Roman J, Tuchin VV. Polarized Light for Biomedical Applications. J. Biomed. Opt. 2016;21:071001. doi: 10.1117/1.JBO.21.7.071001. [DOI] [PubMed] [Google Scholar]
  • 15.Das NK, et al. Quantitative assessment of submicron scale anisotropy in tissue multifractality by scattering Mueller matrix in the framework of Born approximation. Opt. Commun. 2018;413:172–178. doi: 10.1016/j.optcom.2017.11.082. [DOI] [Google Scholar]
  • 16.Mishchenko MI, Travis LD, Lacis AA. Scattering, Absorption, and Emission of Light by Small Particles. Cambridge: Cambridge University Press; 2002. [Google Scholar]
  • 17.Tuchin VV. Tissue optics and photonics: biological tissue structures. J. Biomed. Photonics Eng. 2015;1:3–21. doi: 10.18287/JBPE-2015-1-1-3. [DOI] [Google Scholar]
  • 18.Swami MK, Patel HS, Gupta PK. Conversion of 3×3 Mueller matrix to 4×4 Mueller matrix for non-depolarizing samples. Opt. Commun. 2013;286:18–22. doi: 10.1016/j.optcom.2012.08.094. [DOI] [Google Scholar]
  • 19.Izotova VF, Maksimova IL, Nefedov IS, Romanov SV. Investigation of Mueller matrices of anisotropic nonhomogeneous layers in application to an optical model of the cornea. Appl. Opt. 1997;36:164–169. doi: 10.1364/AO.36.000164. [DOI] [PubMed] [Google Scholar]
  • 20.Manhas S, et al. Mueller matrix approach for determination of optical rotation in chiral turbid media in backscattering geometry. Opt. Express. 2006;14:190–202. doi: 10.1364/OPEX.14.000190. [DOI] [PubMed] [Google Scholar]
  • 21.Ushenko VA. Embossed Topographic Depolarisation Maps of Biological Tissues with Different Morphological Structures. Sci. Rep. 2021;11:3871. doi: 10.1038/s41598-021-83017-2. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 22.Angelsky OV, Ushenko A, Ushenko YA, Pishak VP, Peresunko AP. Statistical, correlation, and topological approaches in diagnostics of the structure and physiological state of birefringent biological tissues. In: Tuchin VV, editor. Handbook of Photonics for Biomedical Science. Boca Raton: CRC Press; 2010. pp. 283–322. [Google Scholar]
  • 23.Ushenko YA, Boychuk TM, Bachynsky VT, Mincer OP. Diagnostics of structure and physiological state of birefringent biological tissues: statistical, correlation and topological approaches. In: Tuchin VV, Valeriĭ V, editors. Handbook of Coherent-Domain Optical Methods. New York: Springer; 2013. pp. 107–148. [Google Scholar]
  • 24.Lu S-Y, Chipman RA. Interpretation of Mueller matrices based on polar decomposition. J. Opt. Soc. Am. A. 1996;13:1106–1113. doi: 10.1364/JOSAA.13.001106. [DOI] [Google Scholar]
  • 25.Guo Y, et al. A study on forward scattering Mueller matrix decomposition in anisotropic medium. Opt. Express. 2013;21:18361–18370. doi: 10.1364/OE.21.018361. [DOI] [PubMed] [Google Scholar]
  • 26.Buscemi IC, Guyot S. Near real-time polarimetric imaging system. J. Biomed. Opt. 2013;18:116002. doi: 10.1117/1.JBO.18.11.116002. [DOI] [PubMed] [Google Scholar]
  • 27.DeBoo B, Sasian J, Chipman R. Degree of polarization surfaces and maps for analysis of depolarization. Opt. Express. 2004;12:4941–4958. doi: 10.1364/OPEX.12.004941. [DOI] [PubMed] [Google Scholar]
  • 28.Manhas S, et al. Demonstration of full 4×4 Mueller polarimetry through an optical fiber for endoscopic applications. Opt. Express. 2015;23:3047–3507. doi: 10.1364/OE.23.003047. [DOI] [PubMed] [Google Scholar]
  • 29.Pierangelo A, et al. Multispectral Mueller polarimetric imaging detecting residual cancer and cancer regression after neoadjuvant treatment for colorectal carcinomas. J. Biomed. Opt. 2013;18:046014. doi: 10.1117/1.JBO.18.4.046014. [DOI] [PubMed] [Google Scholar]
  • 30.Marchesini R, et al. Extinction and absorption coefficients and scattering phase functions of human tissues in vitro. Appl. Opt. 1989;28:2318–2324. doi: 10.1364/AO.28.002318. [DOI] [PubMed] [Google Scholar]
  • 31.Edwards DK, et al. Integrating sphere for imperfectly diffuse samples. J. Opt. Soc. Am. 1961;51:1279–1288. doi: 10.1364/JOSA.51.001279. [DOI] [Google Scholar]
  • 32.Borovkova M, et al. Mueller-matrix-based polarization imaging and quantitative assessment of optically anisotropic polycrystalline networks. PLOS One. 2019;14(5):e0214494. doi: 10.1371/journal.pone.0214494. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 33.Borovkova M, et al. Complementary analysis of Muller-matrix images of optically anisotropic highly scattering biological tissues. J. Eur. Opt. Soc. Rapid Publ. 2018;14:20. doi: 10.1186/s41476-018-0085-9. [DOI] [Google Scholar]
  • 34.Ushenko V, et al. 3D Mueller-matrix diffusive tomography of polycrystalline blood films for cancer diagnosis. Opt. Spectrosc. 2012;112:884–892. doi: 10.1134/S0030400X12050232. [DOI] [Google Scholar]
  • 35.Ushenko YA, Dubolazov AV, Balanetskaya VO, Karachevtsev AO, Ushenko VA. Wavelet-analysis of polarization maps of human blood plasma. Opt. Spectrosc. 2012;113:332–343. doi: 10.1134/S0030400X12070260. [DOI] [Google Scholar]
  • 36.Tuchin VV, Wang LV, Zimnyakov DA. Optical polarization in biomedical applications. Optical Polarization in Biomedical Applications. Berlin: Springer; 2006. [Google Scholar]
  • 37.Tuchin VV. Handbook of Photonics for Biomedical Science. Boca Raton: CRC Press; 2010. [Google Scholar]
  • 38.Tyo JS. Design of optimal polarimeters: maximization of signal-to-noise ratio and minimization of systematic error. Appl. Opt. 2002;41:619–630. doi: 10.1364/AO.41.000619. [DOI] [PubMed] [Google Scholar]
  • 39.Goodman JW. Statistical Properties of Laser Speckle Patterns. Berlin: Springer; 1975. [Google Scholar]
  • 40.Robinson SP. Principles of Forensic Medicine. London: Greenwich Medical Media; 1996. [Google Scholar]