Skip to main content
PLOS One logoLink to PLOS One
. 2021 Mar 5;16(3):e0247356. doi: 10.1371/journal.pone.0247356

Development and validation of an LC-MS/MS method for determination of hydroxychloroquine, its two metabolites, and azithromycin in EDTA-treated human plasma

Vong Sok 1, Florence Marzan 1, David Gingrich 1, Francesca Aweeka 1, Liusheng Huang 1,*
Editor: Pasquale Avino2
PMCID: PMC7935301  PMID: 33667247

Abstract

Background

Hydroxychloroquine (HCQ) and azithromycin (AZM) are antimalarial drugs recently reported to be active against severe acute respiratory syndrome coronavirus- 2 (SARS-CoV-2), which is causing the global COVID-19 pandemic. In an emergency response to the pandemic, we aimed to develop a quantitation method for HCQ, its metabolites desethylhydroxychloroquine (DHCQ) and bisdesethylchloroquine (BDCQ), and AZM in human plasma.

Methods

Liquid chromatography tandem mass spectrometry was used to develop the method. Samples (20 μL) are extracted by solid-phase extraction and injected onto the LC-MS/MS system equipped with a PFP column (2.0 × 50 mm, 3 μm). ESI+ and MRM are used for detection. Ion pairs m/z 336.1→247.1 for HCQ, 308.1→179.1 for DHCQ, 264.1→179.1 for BDCQ, and 749.6→591.6 for AZM are selected for quantification. The ion pairs m/z 342.1→253.1, 314.1→181.1, 270.1→181.1, and 754.6→596.6 are selected for the corresponding deuterated internal standards (IS) HCQ-d4, DHCQ-d4, BDCQ-d4, and AZM-d5. The less abundant IS ions from 37Cl were used to overcome the interference from the analytes.

Results

Under optimized conditions, retention times are 0.78 min for BDCQ, 0.79 min for DHCQ, 0.92 min for HCQ and 1.87 min for AZM. Total run time is 3.5 min per sample. The calibration ranges are 2–1000 ng/mL for HCQ and AZM, 1–500 ng/mL for DHCQ and 0.5–250 ng/mL for BDCQ; samples above the range are validated for up to 10-fold dilution. Recoveries of the method ranged from 88.9–94.4% for HCQ, 88.6–92.9% for DHCQ, 88.7–90.9% for BDCQ, and 98.6%-102% for AZM. The IS normalized matrix effect were within (100±10) % for all 4 analytes. Blood samples are stable for at least 6 hr at room temperature. Plasma samples are stable for at least 66 hr at room temperature, 38 days at -70°C, and 4 freeze-thaw cycles.

Conclusions

An LC-MS/MS method for simultaneous quantitation of HCQ, DHCQ, BDCQ, and AZM in human plasma was developed and validated for clinical studies requiring fast turnaround time and small samples volume.

1. Introduction

The new coronavirus disease 2019 (COVID-19), caused by severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2), has evolved into a world pandemic since the first 4 cases were reported in December 29, 2019, in Wuhan, China [1]. As of August 25, 2020, there are 23.5 million COVID-19 cases worldwide with 810,492 deaths [2], among which 5.77 million cases with178,129 deaths are in USA [3]. There is an urgent need for effective drugs to treat COVID-19. Initial studies found hydroxychloroquine (HCQ) active against SARS-CoV-2 [4] and potentially useful for the treatment of COVID-19 illness clinically [5] and azithromycin (AZM) was evaluated in combination with HCQ to treat COVID-19 [5]. However recent clinical trials, lacking critical pharmacology evaluations to inform optimal dosing and requiring drug quantitation methods, have reported substantial toxicities and contradicted results. Some studies reported benefits [57] while others reported no benefits [810].

HCQ, primarily used previously as an antimalarial drug, has also been used for autoimmune diseases such as rheumatoid arthritis for several decades [11, 12]. HCQ is 50% bound to plasma proteins, absorbed completely and rapidly(70–80% in the gastrointestinal tract) [13] and is characterized by a long half-life(up to 40 days). Its peak concentration(Cmax) in the context of multiple dosing may reach up to 1000 ng/mL [14, 15]. Hepatically, HCQ is metabolized by cytochrome p450 (CYP) 2D6 to desethyl-chloroquine (DCQ) and desethyl-hydroxychloroquine (DHCQ)—both of which exhibiti activity for rheumatoid arthritis [16]; as well as bis-desethyl-hydroxychloroquine (BDCQ), a metabolite implicated in HCQ toxicity [17, 18]. At steady state, DHCQ in blood reaches approximately the same concentration as HCQ while BDCQ exhibits ~1/10 of HCQ concentrations [18].

AZM is a 2nd generation macrolide antibacterial that inhibits bacterial protein synthesis [19]. It also exhibits moderate activity against malaria and is used in combination with chloroquine for malaria chemoprevention [20] and treatment [21]. AZM Cmax has been reported to be ~400 ng/mL following a 500 mg single dose [22], with higher Cmax expected following multiple doses due to its long half-life of ~70 hrs. AZM is characterized by both low oral bioavailability(17–37%) and low plasma protein binding (~30%) [19]. It accumulates in tissues and blood leukocytes.

As our group is a leading pharmacology laboratory for HIV and malaria and as part of the wide-spread international emergency response to the outbreak of COVID-19, our laboratory rapidly developed, validated and received approval from a NIH Division of AIDS (DAIDS) supported quality assurance program, for a liquid chromatography tandem mass spectrometry (LC-MS/MS) method. This method was developed to support clinical trials and to assess the pharmacokinetics (PK) and pharmacodynamics (PD) of HCQ, DHCQ, BDCQ and AZM. LC-MS/MS is the preferred technique for drug analysis due to its high sensitivity and selectivity. While LC-MS/MS methods to quantitate HCQ [13, 23, 24] and its metabolites [17, 18, 25] have been reported, they are mainly for analyzing whole blood samples. A number of LC-MS/MS methods have also been published to measure AZM in human plasma [26, 27]. To the best of our knowledge, this is the first method for the simultaneous quantitation of HCQ, its metabolites and AZM in human plasma. Previous studies reported plasma/serum HCQ ranges from 1.0 to 2440 ng/mL with the majority of samples being between 50.0–1700 ng/mL [24] while AZM Cmax in plasma ranges from 200 ng/mL to 1500 ng/mL depending on the dosage [19]. Therefore, the assay calibration curve ranges were tailored for 2–1000 ng/mL for both AZM and HCQ, 1–500 ng/mL for DHCQ, and 0.5–250 ng/mL for BDCQ. This assay requires only 20 μL plasma sample volume.

2. Materials and methods

2.1. Materials

Azithromycin and hydroxychloroquine (Fig 1) were USP reference standards purchased from Sigma-Aldrich. Desethyl-hydroxychloroquine, bisdesethyl-chloroquine, the internal standards azithromycin-d5, hydroxychloroquine-d4, Desethyl-hydroxychloroquine-d4, and bisdesethyl-chloroquine-d4 were purchased from Toronto Research Chemicals. Trifluoroacetic acid, acetonitrile, methanol and water were purchased from Thermo-fisher (Optima LC/MS grade). Blank human plasma and blood (K2 or K3 EDTA added as anticoagulants) was obtained from Biological Specialty Co (Comar, PA, USA).

Fig 1. Chemical structures of azithromycin, hydroxychloroquine, desethylhydroxychoroquine, bisdesethylchloroquine and the internal standards.

Fig 1

2.2. Instrumentation

Sciex API5000 tandem mass spectrometer was coupled with a Shimadzu Prominence 20ADXR LC pumps and SIL-20ACXR autosampler. The LC column was Pursuit pentafluorophenyl (PFP) (50×2.0 mm, 3μm) fitted with a guard column (10×2.0 mm, 3μm) (Agilent Tech., Santa Clara, CA, USA) and eluted with water (A) and acetonitrile (B) both containing 0.05% trifluoroacetic acid (TFA) at a flow rate of 0.5 mL/min in a gradient mode: 20% solvent B (0–0.2 min), 20 to 50% B (0.2–1.5 min), 50–90% B (1.5–1.6 min), 90 to 100% B (1.6–2.0 min), 100% B (2.0–2.5 min), 100 to 20% B (2.5–2.6 min), and 20% B (2.6–3.5 min). Electro Spray ionization in positive mode (ESI+) was used as the ion source with multiple reaction monitoring (MRM) of m/z 749.6→591.6 for AZM, m/z 336.1→247.1 for HCQ, m/z 308.1→179.1 for DHCQ, and m/z 264.1→179.1 for BDCQ. The ion pairs for their corresponding internal standards (IS) were m/z 754.6→596.6 for AZM-d5,342.1→253.1 for HCQ-d4, 314.1→181.1 for DHCQ-d4, and 270.1→181.1 for BDCQ-d4. Samples were diverted into MS source between 0.5–2.3 min.

2.3. Preparation of stock, calibration standards, and quality control samples

HCQ stock solution was prepared in water. DHCQ and BDCQ stock solutions were prepared in methanol. AZM stock solution and all working solutions were prepared in methanol-water (1:1, v/v). Nine combined calibration standard samples at concentrations of 2, 5, 10, 20, 50, 100, 200, 500 and 1000 ng/mL for AZM and HCQ; 1, 2.5, 5, 10, 25, 50, 100, 250, and 500 ng/mL for DHCQ; 0.5, 1.25, 2.5, 5, 12.5, 25, 50, 125, 250 ng/mL for BDCQ were prepared in blank EDTA-treated human plasma by serial dilution from a combined working solution of AZM/HCQ/DHCQ/BDCQ (40/40/20/10 μg/mL). Quality control (QC) samples QC-L (6.00/6.00/3.00/1.50 ng/mL for AZM/HCQ/DHCQ/BDCQ), QC-M(60.0/60.0/30.0/15.0 ng/mL for AZM/HCQ/DHCQ/BDCQ), and QC-H (800/800/400/200 ng/mL for AZM/HCQ/DHCQ/BDCQ) were prepared in blank plasma from different stock solutions or the same verified stock solutions as those used for calibrators. All solutions and plasma samples were stored at -70°C before use.

2.4. Sample preparation

Hydrophilic lipophilic balance (HLB) solid phase extraction micro-elution 96-well plates were preconditioned with 200μL MeOH and 200μL water sequentially. Plasma samples (20 μL) were added in the wells containing 40 μL 0.1N NaOH, Then 20 μL combined IS solution(100 ng/mL AZM-d5, 40 ng/mL HCQ-d4, 20 ng/mL DHCQ-d4, and 20 ng/mL BDCQ-d4 in 50% MeOH) was added and mixed briefly. The wells were washed with 200 μL water, followed by 200 μL 10% MeOH under vacuum, and eluted with 25 μL MeOH containing 0.5%FA under gradually increased vacuum. Elution was repeated for the second time with another 25 μL eluent. To the collection plate, 150 μL water was added with a 12-channel pipette to constitute a final volume of 200 μL and mixed by pipetting up and down three times. Sample was then injected into the LC-MS/MS at 1 μL.

2.5. Validation

The method was validated in accordance with guidelines outlined by both the NIH-sponsored Clinical Pharmacology Quality Assurance Program (CPQA) [28] and the FDA [29]. A full validation includes precision and accuracy, dilution integrity, selectivity, matrix effect and recovery, and stability. Dilution integrity was evaluated by diluting the extra-high QC sample (4000/4000/2000/1000 ng/mL AZM/HCQ/DHCQ/BDCQ) by 10-fold with blank plasma. Stability in plasma was evaluated at -70°C, room temperature and after 4 freeze-thaw cycles by comparing the treated samples with untreated samples in plasticmicrocentrifuge tubes. To evaluate autosampler stability, the processed low and high QC samples were first tested on the same day of processing (as control) and 3 days after having been in the autosampler. Solution stability was evaluated by diluting the solutions to within the calibration range with methanol-water (1:1, v/v). To test stability in blood, blank blood was spiked with analytes at high QC levels and mixed gently on a rotator for 5 min before centrifuging at 2000 g for 10 min to obtain plasma, which was analyzed along with freshly spiked calibrators and QCs. The remaining blood was rotated briefly and left on benchtop. A series of plasma samples was then prepared from this blood at 1hr, 2hr, 4hr, and 6hr. Each timed sample was processed and analyzed immediately following preparation. All measurements were performed in triplicates at the minimum. Selectivity was evaluated with 6 different lots of human plasma with K3EDTA as the anticoagulant.

Matrix effect (ME), recovery (RE) and process efficiency (PE) were evaluated with three sets of samples: Set 1 samples were prepared by spiking 20 μL analytes in 50% MeOH solution at QC-L, QC-M, QC-H concentrations and 20 μL IS solution (100/40/20/20ng/mL AZM-d5, HCQ-d4, DHCQ-d4, and BDCQ-d4) into 160 μL 50% MeOH and analyzed in triplicates. Set 2 samples were spiked at the same concentration as Set 1 in extracted solutions from blank plasma in triplicate. Set 3 samples were prepared by spiking analytes in blank plasma with final concentrations of 6/6/3/1.5 ng/mL (QC-L), 60/60/30/15 ng/mL (QC-M), and 800/800/400/200 ng/mL (QC-H) for AZM/HCQ/DHCQ/BDCQ. These plasma samples were then processed in triplicate using protocols as described above.

To test impact of hemolysis on plasma sample analysis, a 1 mL aliquot of whole blood underwent 3 freeze-thaw cycles to lyse the blood cells. Fifty microliter of the treated blood was spiked into 950 μL plasma in triplicates to give 2–3% hemolyzed plasma, which were spiked with AZM/HCQ/DHCQ/BDCQ at QC-L and QC-H concentrations. The prepared QC-L and QC-H were processed and analyzed along with freshly spiked calibrators and QCs.

Clinical samples are likely to be collected in tubes with K2EDTA instead of K3EDTA as the anticoagulant. To test the impact on quantification, QC-L and QC-H were prepared in two lots of K2EDTA plasma and one lot of K3EDTA plasma as the control. Triplicates of these samples were processed and analyzed along with a set of calibrators.

3. Results and discussion

3.1. Method development

3.1.1. LC-MS/MS optimization

The LC-MS/MS system was optimized in both APCI+ and ESI+ modes. Initially APCI+ was chosen for its less matrix effect and lower baseline signal. However, APCI+ limited linearity to a range narrower than that desired (S1 Fig). ESI+ ion source was finally chosen for this assay. The optimized MS/MS parameters are shown in Table 1. LC separation parameters were adopted from a previous assay we had developed for piperaquine [30]. It was further discovered that mobile phases 0.05% TFA in water and 0.05% TFA in acetonitrile gave similar peak shapes and retention to those obtained using 20 mM NH4FA 0.14% TFA in water and 0.1% TFA in acetonitrile. Therefore, the former solvent combination was used due to its simplicity. The retention times are 0.78 min for BDCQ, 0.79 min for DHCQ, 0.92 min for HCQ and 1.87 min for AZM. Total run time is 3.5 min per sample.

Table 1. Optimized MS/MS parameters.
Source parameters TEM, IS, CAD, CUR, Gas1, Gas2,
°C v psi psi psi psi
  500 1250 12 25 50 40
Compound parameters DP, EP, CE, CXP, Dwell time,
v v v V ms
749.6/591.6 (AZM) 50 10 40 39 50
754.6/596.6 (AZM-d5) 50 10 40 39 50
336.1/247.1 (HCQ) 50 10 29 16 50
342.1/253.1 (HCQ-d4) 50 10 29 16 50
308.1/179.1 (DHCQ) 50 10 31 16 50
314.1/181.1 (DHCQ-d4) 50 10 31 16 50
264.1/179.1 (BDCQ) 50 10 30 16 50
270.1/181.1 (BDCQ-d4) 50 10 30 16 50

TEM, source temperature; IS, ionspray voltage; CUR, curtain gas, Gas1, nebulizer gas; gas2, auxiliary gas; CAD, collision-activated dissociation; DP, declustering potential; EP, entrance potential; CE, collision energy; CXP, collision cell exit potential.

3.1.2. IS selection

The ideal IS for LC-MS/MS methods are stable isotopically-labelled analytes. In this method, we were able to obtain deuterated IS. To avoid cross talks from analytes, the less abundant ion pairs from the 37Cl isotope (i.e. the most abundant ion plus 2) were selected for HCQ-d4, DHCQ-d4 and BDCQ-d4. For AZM-d5, a higher concentration was used to avoid cross talk from AZM especially at ULOQ level.

Selection of the appropriate concentrations for stable isotopically labelled internal standards is based on both the number of stable isotopic atoms in the IS and the mass abundance of analytes at the IS mass levels. The IS signal originated from analyte at ULOQ should not be more than 5%. The naturally occurring isotopic masses of the 4 analytes are calculated using an online calculator and shown in Table 2 [31]. The percentage is based on the most abundant mass(EM).

Table 2. Natural abundances of isotopic mass of analytes and minimal IS concentration.
  AZM HCQ DHCQ BDCQ
Exact Mass (EM) 748.51 335.18 307.15 263.12
EM+4 0.740% 0.601% 0.408%
EM+5 0.052%
EM+6 0.003% 0.002% 0.001%
CIS,min, ng/mL 10.4 148 120 81.6
*CIS,min, ng/mL 0.6 0.4 0.2

*, 37Cl isotope used for HCQ-d4, DHCQ-d4, and BDCQ-d4.

To avoid interference of cross talk signal from the analyte, the formula to calculate the minimum concentration of stable-isotope labelled internal standard is as follows [32]:

CIS,min=m%×ULOQ5% (1)

Where m is the cross-signal percentage from analyte to IS.

To avoid interference of cross talk signal from the IS to the analyte, the formula to calculate the maximum IS concentration is as follows:

CIS,max=20%×LLOQn% (2)

Where n is the cross-signal percentage from IS to analyte.

According to the equations, IS concentrations higher than 100 ng/mL are needed for HCQ, DHCQ and BDCQ if the most abundant IS ions (EM+4) are selected. However, in order to prevent the spiking of signals at LLOQ levels by highly concentrated IS, we chose the less abundant 37Cl isotope signal (EM+6) for HCQ-d4, DHCQ-d4, and BDCQ-d4. The final IS concentrations were 100ng/mL AZM-d5, 40 ng/mL HCQ-d4, 20 ng/mL DHCQ-d4, and 20 ng/mL BDCQ-d4.

3.1.3. Sample preparation

Solid phase extraction (SPE) was used in this assay as it yielded a cleaner extract than protein precipitation method did. HLB microelution plate was selected in consideration ofthe small sample volumes. While the PRiME HLB plate is more user-friendly for its omission of well preconditioning step, samples loaded into its wells would gradually drain away, leaving little time for the user to homogenize sample with IS and reagents. As a result, we chose the traditional HLB plate as it allows plasma samples to be homogenized with IS and reagents in the wells prior to passing the mixture onto the stationary phase. Methods in literature utilized liquid-liquid extraction with alkalized organic solvents for AZM extraction [26, 33] and protein precipitation with acidified organic solvents for extraction of HCQ and its metabolites [25]. Considering these analytes are weak bases, we alkalized plasma samples with NaOH to help retain analytes on SPE absorbent and acidified elution solvent to help elute analytes. The combined effect improved recovery greatly.

3.2. Method validation

3.2.1. Calibration curves

Nine combined calibration standards, prepared in K3EDTA human plasma, consisting of 2, 5, 10, 20, 50, 100, 200, 500 and 1000 ng/mL for AZM and HCQ; 1, 2.5, 5, 10, 25, 50, 100, 250, and 500 ng/mL for DHCQ; 0.5, 1.25, 2.5, 5, 12.5, 25, 50, 125, 250 ng/mL for BDCQ were used to establish the calibration curves. At the lower limit of quantitation (LLOQ) (2/2/1/0.5 ng/mL AZM/HCQ/DHCQ/BDCQ) the S/N ratios were 47 for AZM, 75 for HCQ, 31 for DHCQ, and 18 for BDCQ. The calibration curves were constructed using concentration vs. peak area ratio fitted with least square linear regression weighted by 1/x for HCQ, DHCQ, and BDCQ while quadratic fitting weighted by 1/x2 was needed for AZM for better accuracy at lower concentrations. The relative error(%RE) sum, defined as the sum of absolute %RE values, was used to evaluate the goodness of fit when using different weighting factors for calibration curve [34]. The %RE sum of calibrators for the 4 intra/inter-day A&P runs were the lowest for quadratic 1/x2 weighted curve compared to those of linear regression. To compare regression models with different parameters, an effective way is to use Akaikes information criterion (AIC) [35]. This criterion not only takes into account the sum of squares of relative errors (SSR), it also includes a term proportional to the number of parameters used. AIC is calculated via the formula:

AIC=n×ln(SSR)+2M (3)

Where n is the number of calibrators and M is the number of parameters. The model producing the smallest AIC is preferred. Quadratic regression weighted by 1/x2 gave the least AIC values (Table 3).

Table 3. Comparison of regression models.
%RE sum AIC
Run ID 2 8 12 17 2 8 12 17
Linear, 1/x weighted 46 96 82 31 60 72 67 51
Linear, 1/x2 weighted 32 78 41 25 49 64 56 45
Quadratic, 1/x weighted 22 82 36 20 48 69 54 45
Quadratic, 1/x2 weighted 17 60 32 22 43 43 50 45

The correlation coefficient(r) was typically >0.995. Representative chromatograms for blank plasma, the lower limit of quantification (LLOQ) and its IS, and double blank plasma injected after ULOQ and its IS are shown in Fig 2.

Fig 2. Representative chromatograms of blank (red dash line), the LLOQ (blue solid line) and double blank samples following ULOQ (grey solid line).

Fig 2

Blank sample was processed with IS, double blank sample was processed without IS.

Fig 2 (red dash line) demonstrates that there are no significant analyte signal and cross-talk from the IS (<20% of LLOQ) for all analytes. Compared to calibrator#1 [LLOQ, where peak area = 2530 (AZM), 29900 (HCQ), 6240 (DHCQ), and 8490 (BDCQ)], the peak areas in blank samples are minimal: 94.3 (3.7%) for AZM, 839 (2.8%) for HCQ, 364 (5.8%) for DHCQ, and 775 (9.1%) for BDCQ.

Fig 2 (grey line) displays each analyte’s and IS’s carryover signals relative to their signal intensities in LLOQ. The peak areas at the retention times of analytes in double blank samples injected after ULOQ for AZM, HCQ, DHCQ, and BDCQ are 217 (8.6%), 3000 (10%), 1050 (16.8%), and 1170 (13.9%), respectively, all within 20% of LLOQ signal (Left panel), and no peaks were found at the retention time of ISs in the chromatograms of double blank samples after ULOQ for AZM-d5, DHCQ-d4, BDCQ-d4 (right panel). The peak for HCQ-d4 in the double blank following ULOQ was not significant: peak area = 1250, representing 1.5% IS signal (83800). The results suggest carryover for all analytes and ISs are not significant. Furthermore, the IS signals from ULOQ of all analytes are no more than the IS signals from the LLOQ, suggesting the cross talk from analytes are negligible (S2 Fig).

3.2.2. Intra-/inter-day precision and accuracy

Precision is the degree of reproducibility; it characterizes the degree of agreement among a series of individual measurements. Precision is calculated as the coefficient of variation (%CV). Accuracy is the degree of correctness and is expressed as the percent deviation from the true concentration value. Precision and accuracy (P&A) of method should be no more than 15% except for the lower limit of quantitation (LLOQ), where ≤20% is acceptable. For inter-assay precision and accuracy, at least 3 runs with at least 5 replicates of LLOQ, low, medium, and high concentration validation samples in each run should be performed. These samples are designated as LLOQ, QC-L, QC-M, and QC-H with AZM/HCQ/DHCQ/BDCQ concentrations of 2/2/1/0.5 ng/mL, 6/6/3/1.5 ng/mL, 60/60/30/15 ng/mL, and 800/800/400/200 ng/mL, respectively.

During the validation, we found the IS solutions were not stable in glass vial especially for hydroxychloroquine and its metabolites. This resulted in unacceptable data in two inter-day P&A runs. Further stability test between storage in glass and in plastic Eppendorf tubes revealed all ISs lost signals in a few hours in glass vial but remained stable in Eppendorf tube for at least overnight. AZM-d5 was reduced by ~20% depending on the container size. HCQ-d4, DHCQ-d4 and BDCQ-d4 in 50% MeOH were reduced much more significantly (over 50%) due to adsorption on glass surface (S1 Table).

Among the 4 runs of intra-/inter-day P&A in this report, the 1st run was performed with freshly prepared calibrators and QCs from separately weighed stocks. In summary, intra- day P&A meet the criteria except for one of four runs at the LLOQ for AZM (dev +25%) using 1/x weighted linear regression calibration curve. When we reanalyzed the intra-/inter-day P&A data for AZM using quadratic regression of 1/x2 weighted calibration curve, all 4 runs passed the acceptance criteria.

The inter-assay precisions(%CV) of the method at low, medium, and high concentrations are 7.5%, 6.9%, 4.9% for AZM; 7.3%, 5.3%, 6.2% for HCQ; 7.5%, 6.3%, 5.1% for DHCQ; 13%, 7.4%, 8.0% for BDCQ. The overall accuracy (%dev) from nominal low, medium and high concentrations are 8.0%, 5.3%, 2.2% for AZM; 0.0%, -0.4%, -1.3% for HCQ; -6.3%, -0.4%, -2.1% for DHCQ; and -4.9%, —5.3%, -5.6% for BDCQ (Table 4).

Table 4. Precision and accuracy.
AZM Intra-day Inter-day
Nominal, ng/mL 2.00 6.00 60.0 800 2.00 6.00 60.0 800
%RSD 7.3–12% 5.6–11% 4.1–7.2% 1.9–3.7% 13 7.4 6.5 5.4
%dev -19-6.1% 6.0–9.3% 0.6–13% -3.1–8.0% -6.2 5.8 2.5 8.9
N 6 6 6 6 24 24 24 24
HCQ Intra-day Inter-day
Nominal, ng/mL 2.00 6.00 60.0 800 2.00 6.00 60.0 800
%RSD 5.7–12% 3.6–8.9% 3.1–6.6% 3.5–6.7% 12 7.3 5.3 6.2
%dev -3.8–19% -4.0–8.2% -2.5–4.1% -7.5–2.7% 6.4 0.0 -0.4 -1.3
N 6 6 6 6 24 24 24 24
DHCQ Intra-day Inter-day
Nominal, ng/mL 1.00 3.00 30.0 400 1.00 3.00 30.0 400
%RSD 9.2–13% 4.7–6.9% 3.8–8.5% 3.9–6.7% 14 7.5 6.3 5.1
%dev -12-13% -13-(0.3%) -5.9–2.3 -4.5–0.5% 0.8 -6.3 -0.4 -2.1
N 6 6 6 6 24 24 24 24
BDCQ Intra-day Inter-day
Nominal, ng/mL 0.500 1.50 15.0 200 0.500 1.50 15.0 200
%RSD 9.0–14% 7.6–13% 4.7–6.0% 4.1–12% 15 13 7.4 8.0
%dev -11-12% -13-8.1% -112.4% -7.8-(3.3)% 0.7 -4.9 -5.3 -5.6
N 6 6 6 6 24 24 24 24

Intra-assay precision and accuracy were calculated from 6 replicate samples of low, medium, and high concentrations analyzed on the same day on 4 unique days. The intra-day precision (%CV) of this method at low, medium, and high concentrations ranges from 5.5–11%, 4.1–7.2%, 2.0–4.2% for AZM; 3.6–8.9%, 3.1–6.6%, 3.5–6.7% for HCQ; 4.7–6.9%, 3.8–8.5%, 3.9–6.7% for DHCQ; and 7.6–13%, 4.7–6.0%, 4.1–12% for BDCQ. Accuracy (dev%) for low, medium and high levels ranges from 6.0–9.3%, 0.6–13%, -3.1–8.0% for AZM; -4.0–8.2%, -2.5–4.1%, -7.5–2.7% for HCQ; -13-(-0.3)%, -5.9–2.3%, -4.5–0.5% for DHCQ; and -13-8.1%, -11-2.4%, -7.8-(-3.3)% for BDCQ (Table 4).

LLOQ. Six replicates of validation samples at the lowest calibration concentration (2/2/1/0.5 ng/mL for AZM/HCQ/DHCQ/BDCQ) were analyzed on 4 different days to determine the inter- and intra- assay precision and accuracy of the lowest point on the calibration curve. The inter-assay precision (%CV) is 14% for AZM, 12% for HCQ, 14% for DHCQ, and 15% for BDCQ. The inter- assay percent deviation is -7.9% for AZM, 6.4% for HCQ, 0.8% for DHCQ, and 0.7% for BDCQ. The intra-assay %CV for the mean of these 4 replicate days ranges from 7.3–12% for AZM, 5.7–12% for HCQ, 9.2–13% for DHCQ, and 9.0–14% for BDCQ. The mean accuracy (%dev) ranges from -19-6.1% for AZM, -3.8–19% for HCQ, -12-13% for DHCQ, and -11-12% for BDCQ (Table 4).

3.2.3. Dilution integrity

An extra-high QC plasma samples at a nominal concentration of 4000 ng/mL for AZM and HCQ, 2000ng/mL for DHCQ, and 1000ng/mL for BDCQ (4 times the ULOQ concentration) were diluted with blank plasma by 10-fold. Five replicates of the diluted samples processed and analyzed. The mean values (n = 5) were within 15% of the nominal concentration for all analytes. The %CV was 5.3%, 9.4%, 5.5% and 3.9%; and %dev was 14%, -7.0%, -11% and -10% for AZM, HCQ, DHCQ, and BDCQ, respectively, suggesting the samples can be diluted by up to 10-fold without compromising sample integrity.

3.2.4. Stability

Freeze/thaw stability. QC-L and QC-H samples undergone 4 freeze-thaw cycles were processed and analyzed along with freshly spiked calibrators and QCs. The percent differences from freshly made controls at QC-L and QC-H concentrations are -3.7% and -10% for AZM, -8.8% and -1.1% for HCQ, -12% and -5.3% for DHCQ, -7.1% and -4.2% for BDCQ (S2 Table). The percent remaining analytes compared to nominal concentrations are all within 100 (±15) % (Table 5). The results reveal that AZM, HCQ, DHCQ and BDCQ in plasma are stable after 4 freeze-thaw cycles.

Table 5. Stability of AZM, HCQ, DHCQ, and BDCQ (n = 3).
  AZM HCQ DHCQ BDCQ
Treated conditions conc., ng/mL % remaining conc., ng/mL % remaining conc., ng/mL % remaining conc., ng/mL % remaining
38 days, -70°C                
Low 6.13±0.26 102 5.20±0.04 86.7 2.58±0.03 86.0 1.32±0.13 88.0
High 859±33 107 790±32 98.8 367±25 91.8 178±16 89.0
66hr, 22±3°C (RT)
Low 5.98±0.13 99.7 5.46±0.32 91.0 2.74±0.12 91.3 1.30±0.07 86.7
high 801±33 100 786±36 98.3 363±15 90.8 177±7 88.5
Reinjection, 3 days
Low 5.67±0.35 94.4 5.81±0.13 96.9 3.16±0.08 105 1.50±0.14 100
high 845±31 106 718±64 89.8 410±2 102 187±12 93.5
4-freeze-thaw cycles
low 5.14±0.36 85.7 5.78±0.02 96.3 2.92±0.12 97.3 1.49±0.12 99.3
high 765±23 95.6 784±17 98.0 373±11 93.3 189±2 94.5
Blood, 6hr
high 878±34 104 687±21 112 303±10 106 135±11 103
Combined working solution, 50%MeOH          
RT, 8 days   96.8   96.6   97.1   94.5

*% remaining was calculated by comparing to the nominal values (100%) for plasma and reinjection stability, comparing to the spiked blood samples at 22min for blood stability, and comparing to the same solution frozen at -70°C for solution room temperature (RT) stability.

Room temperature stability in plasma. After standing on the bench for 66 hours, QC-L and QC-H samples were processed and analyzed along with freshly prepared QC-L and QC-H samples and the freshly prepared calibrators. The % change from controls at QC-L and QC-H concentrations was 4.3% and 6.1% for AZM, -2.3% and 0.2% for HCQ, -7.6 and -6.8% for DHCQ, -12% and -6.3% for BDCQ, all of which were within ±15% (S2 Table). When compared to spiked nominal concentrations, the remaining concentration are all within 100 (±15) %. The results suggest that plasma samples are stable for at least 66 hours at room temperature (Table 5).

Reinjection reproducibility/autosampler stability. To test autosampler stability, the analyzed samples were left in the autosampler and re-injected 3 days after (71hr). The percent remaining drug concentrations from nominal values at QC-L and QC-H were 94.4 and 106% for AZM, 96.9% and 89.8% for HCQ, 105% and 102% for DHCQ, 100% and 93.5% for BDCQ, suggesting the processed samples are stable in autosampler for at least 3 days (Table 5).

Long-term stability at -70°C of plasma samples. To test long-term stability at -70°C, triplicates of the QC-L and QC-H plasma samples stored at -70°C for 38 days were analyzed along with freshly spiked calibrators and QC samples as controls in triplicate. The treated samples were all within ±15% difference from controls. The percent differences from control QC-L and QC-H samples are 3.7% and -3.0% for AZM, 1.3% and 2.7% for HCQ, -1.9% and -0.9% for DHCQ, and -2.0% and -5.7% for BDCQ (S2 Table). When compared to nominal values, the treated samples were also within 100 (±15) % (Table 5). The data demonstrate plasma samples are stable at -70°C for at least 38 days. Previous study reported AZM in plasma is stable at -70° C for at least 92 days [26].Stability of blood samples at room temperature. When compared to plasma samples separated from blood at 22 min, less than 15% difference over 6 hr was found for all drugs, suggesting AZM, HCQ, DHCQ and BDCQ are stable in blood for at least 6 hr at room temperature (Table 5). When compared to nominal concentration, less than 15% change was found for AZM over 6 hr at room temperature, suggesting AZM is equally distributed in blood cells and plasma. Whereas, for HCQ and its metabolites, over 20% (-20% for HCQ, -30% for DHCQ, -35% for BDCQ) lower than nominal concentration was found at 22 min after the drugs had been spiked into blood (S2 Table), suggesting these analytes concentrated in blood cells. Previous studies reported that AZM concentrated in blood leukocytes and inflamed tissues and its concentration in blood was double of that in plasma on day 3 and 4-fold higher after day 30 [19]–Likely because AZM is slowly distributed into blood cells. HCQ level is also higher in blood than in plasma [36], consistent with our results.

Solution stability. AZM stock solution (2 mg/mL in 50% MeOH) was stable for at least 23 hr at room temperature and for at least 45 days at -70°C, the % difference from the untreated fresh controls are -1.3% and -0.1% respectively. HCQ stock (2 mg/mL in water) was stable at -70°C for at least 63 days and at room temperature (19–22°C) for at least 5 days, with the %difference from fresh controls at 6.6% and 1.5%, respectively (S2 Table). The combined working solution (40/40/20/10 μg/mL AZM/HCQ/DHCQ/BDCQ in 50% MeOH) was stable at room temperature for at least 8 days (Table 5). Previous studies reported that AZM stock in methanol is stable for 92 days at -10°C [26]. HCQ and DHCQ stocks (0.2 mg/mL) in water are stable at -80°C for 12 months [18].

The working solutions for all deuterated ISs were stable at room temperature for at least 23 hr in plastic tube (S1 Table). However, due to adsorption on glass surface, the IS solution should be prepared in plastic container. Similarly, the stock and working solutions of analytes—especially those of HCQ, DHCQ and BDCQ should be prepared and stored in plastic containers—even though the impact of adsorption on glass surface may diminish at higher drug concentration. For example, AZM stock solution at 0.5 mg/mL in glass vial is comparable to those in plastic vial (<5% difference) (S2 Table).

3.2.5. Matrix effect, recovery, and process efficiency

Three sets of samples each at three concentration levels (low, medium, and high validation concentration levels) were prepared and analyzed to determine ME, RE, and PE. The mean peak area and peak area ratio (analyte/IS) were calculated for each level in each set of samples and comparisons are presented in Table 6.

Table 6. Matrix effect, recovery and process efficiency.
Analytes Conc. Matrix Effect Recovery PE
(ng/ml) analyte IS analyte IS analyte IS
Low (6) 100 102 102 101 102 103
AZM Med (60) 98.8 96.4 99.4 100 98.3 96.8
  High (800) 97.9 95.7 98.6 102 96.5 97.8
Low (6) 110 106 90.5 96.1 99.8 101
HCQ Med (60) 98.0 101 94.4 97.2 92.4 97.7
  High (800) 103 100 88.9 94.1 91.9 94.1
Low (3) 102 103 88.6 99.8 90.4 103
DHCQ Med (30) 101 98.9 90.3 99.3 91.4 98.2
  High (400) 106 101 92.9 96.7 98.0 97.8
Low (1.5) 97.1 104 90.9 91.8 88.3 95.6
BDCQ Med (15) 103 110 88.7 94.0 91.6 104
  High (200) 101 101 89.0 87.7 90.2 88.4
ME=100×peakareaofpostextractionspikedsample(set2)peakareaofcleansample(set1) (4)
RE=100×peakareaofpreextractionspikedsample(set3)peakareaofpostextractionspikedsample(set2) (5)
PE=100×peakareaofpreextractionspikedsample(set3)peakareaofcleansample(set1) (6)

Recovery (RE). The recovery of analytes from plasma following sample preparation was assessed by comparing the peak areas from set 3 and set 2. The recoveries for AZM were 102, 99.4 and 98.6% at low, medium, and high concentration, respectively, and the recovery for the IS ranged from 100–102%. The recoveries ranged from 88.9–94.4% for HCQ, 88.6–92.9% for DHCQ, and 88.7–90.9% for BDCQ. The %CV of peak areas for recovery experiment was all within 12%. These results suggest the assay is highly reproducible across the concentration range with consistent and high recovery.

Matrix Effect (ME). The matrix effect of analytes from plasma following sample preparation was assessed by comparing the peak areas from set 2 and set 1. The ME for AZM were 100, 98.8 and 97.9% at low, medium, and high concentration, respectively. ME for the IS ranged from 95.7–102%. The ME for HCQ, DHCQ and BDCQ ranged from 98.0–110%, 101–106%, 97.1–103%, respectively. The IS normalized ME were within (100±10) % for all 4 analytes.

Process Efficiency (PE). The PE of analytes from plasma following sample preparation was assessed by comparing the peak areas from set 3 and set 1. The PE for AZM were 102, 98.3 and 96.5% at low, medium, and high concentration, respectively. ME for the IS ranged from 96.8–103%. The PE for HCQ, DHCQ and BDCQ ranged from 91.9–99.8%, 90.4–98.0%, 88.3–91.6%, respectively.

3.2.6. Selectivity

To test selectivity, 6 lots of blank plasma were processed without adding ISs and analyzed along with a LLOQ sample. The results are shown in Fig 3. significant signals were found at the retention times of both analyte and IS for each analyte. The data suggest the method is highly selective.

Fig 3. Chromatograms of six lots of blank plasma and LLOQ sample.

Fig 3

3.2.7. Impact of hemolysis

Compared to controls, the differences of hemolyzed plasma samples at QC-L and QC-H concentrations were within 15% for all analytes (10.8 and -1.4% for AZM, 3.0 and 0.94% for HCQ, 6.3% and -8.1% for DHCQ, -4.3 and -3.0 for BDCQ) (S2 Table). The results suggest hemolysis does not impact quantitation of the analytes in this assay. However, since AZM and HCQ are known to present higher in blood than plasma, it is recommended to avoid hemolysis when processing clinical samples.

3.2.8. Impact of anticoagulant counter ions

The differences of K2EDTA plasma samples from K3EDTA plasma samples (controls) were all within ±15% at QC-L and QC-H concentrations. For AZM, the differences from the controls at QC-L and QC-H were -1.6% and-8.0%, -4.2% and -5.0% in the two lots of K2EDTA plasma. For HCQ, the differences at QC-L and QC-H were 4.6% and -6.3%, 7.3% and 2.1% in the two lots of K2EDTA plasma. For DHCQ, they are -3.9% and -13%, 9.1% and 0.1%. For BDCQ, they are-3.1% and -11%, 11% and -0.8%, respectively (S2 Table). The data suggest different counter ions in anticoagulant won’t interfere assay performance, and K2EDTA collection tubes could be used as an alternative for sample collection.

4. Conclusions

A high-throughput method for simultaneous quantitation of HCQ, DHCQ, BDCQ and AZM in plasma was developed and validated based on guidelines from FDA and NIH-sponsored CPQA, and suitable for clinical studies of those drugs. The method required only 20 μL plasma sample and 3.5 min run time and carryover is negligible. Although clinical data for HCQ and AZM use in COVID-19 have resulted in poor outcomes, understanding the pharmacological basis for toxicity is still of interest to inform any potential future use of these compounds. Therefore, methods to quantitate these drugs reliably still remain relevant. Furthermore, these compounds continue to be used or evaluated for other illnesses besides COVID-19 such as malaria, inflammatory diseases, and other viral and bacterial infections [37].Such evaluations will require state of the art analytical methods such as one described here. For it requires only a small sample volume, our method can be used for pediatric studies where sample volume is limited. and it can be coupled with capillary tube sampling by a finger prick or more advanced microsampling techniques such as Seventh Sense Tap™ to facilitate clinical studies. With the highly sensitive LC-MS/MS system, our method may also be modified for dried blood spot samples.

Supporting information

S1 Fig. Linearity of AZM calibration curve in ESI versus APCI.

(PDF)

S2 Fig. LLOQ versus ULOQ chromatograms.

(PDF)

S1 Table. Adsorption of internal standards on container surface.

(DOCX)

S2 Table. Stability data.

(PDF)

S1 File. Assay standard operating procedure.

(PDF)

Acknowledgments

We wish to thank supporting staff at University of California San Francisco (UCSF). We also want to thank Difrancesco Robin and Andrew Ocque from University of New York at Buffalo, and Lane Bushman from University of Corolado for their constructive discussion on the method validation and thank supporting staff and reviewers for CPQA.

Data Availability

All relevant data are within the manuscript and its Supporting Information files.

Funding Statement

The APC was funded by UCSF library. No additional external funding was received for this study.

References

  • 1.Li Q, Guan X, Wu P, Wang X, Zhou L, Tong Y, et al. Early Transmission Dynamics in Wuhan, China, of Novel Coronavirus-Infected Pneumonia. N Engl J Med. 2020;382(13):1199–207. 10.1056/NEJMoa2001316 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 2.WHO. Coronavirus disease (COVID-19): Situation Report–WHO COVID-19 Dashboard https://www.who.int/emergencies/diseases/novel-coronavirus-2019/situation-reports: World Health Organization; 2020. [cited 2020 August 25]. [Google Scholar]
  • 3.Johns Hopkins CSSE. Tracking COVID-19: COVID-19 Dashboard by the Center for Systems Science and Engineering (CSSE) at Johns Hopkins University (JHU) https://systems.jhu.edu/: John Hopkins University Center for Systems Science and Engineering; 2020. [cited 2020 August 25]. Available from: https://www.arcgis.com/apps/opsdashboard/index.html#/bda7594740fd40299423467b48e9ecf6. [Google Scholar]
  • 4.Liu J, Cao R, Xu M, Wang X, Zhang H, Hu H, et al. Hydroxychloroquine, a less toxic derivative of chloroquine, is effective in inhibiting SARS-CoV-2 infection in vitro. Cell Discov. 2020;6:16. 10.1038/s41421-020-0156-0 PubMed Central PMCID: PMC7078228. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5.Gautret P, Lagier JC, Parola P, Hoang VT, Meddeb L, Mailhe M, et al. Hydroxychloroquine and azithromycin as a treatment of COVID-19: results of an open-label non-randomized clinical trial. Int J Antimicrob Agents. 2020:105949. 10.1016/j.ijantimicag.2020.105949 [DOI] [PMC free article] [PubMed] [Google Scholar] [Retracted]
  • 6.Arshad S, Kilgore P, Chaudhry ZS, Jacobsen G, Wang DD, Huitsing K, et al. Treatment with hydroxychloroquine, azithromycin, and combination in patients hospitalized with COVID-19. Int J Infect Dis. 2020;97:396–403. 10.1016/j.ijid.2020.06.099 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7.Mikami T, Miyashita H, Yamada T, Harrington M, Steinberg D, Dunn A, et al. Risk Factors for Mortality in Patients with COVID-19 in New York City. J Gen Intern Med. 2020. 10.1007/s11606-020-05983-z [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8.Geleris J, Sun Y, Platt J, Zucker J, Baldwin M, Hripcsak G, et al. Observational Study of Hydroxychloroquine in Hospitalized Patients with Covid-19. N Engl J Med. 2020;382(25):2411–8. 10.1056/NEJMoa2012410 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9.Boulware DR, Pullen MF, Bangdiwala AS, Pastick KA, Lofgren SM, Okafor EC, et al. A Randomized Trial of Hydroxychloroquine as Postexposure Prophylaxis for Covid-19. N Engl J Med. 2020;383(6):517–25. 10.1056/NEJMoa2016638 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10.Cavalcanti AB, Zampieri FG, Rosa RG, Azevedo LCP, Veiga VC, Avezum A, et al. Hydroxychloroquine with or without Azithromycin in Mild-to-Moderate Covid-19. N Engl J Med. 2020. 10.1056/NEJMoa2019014 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11.Sperber K, Hom C, Chao CP, Shapiro D, Ash J. Systematic review of hydroxychloroquine use in pregnant patients with autoimmune diseases. Pediatr Rheumatol Online J. 2009;7:9. 10.1186/1546-0096-7-9 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12.Rempenault C, Combe B, Barnetche T, Gaujoux-Viala C, Lukas C, Morel J, et al. Clinical and Structural Efficacy of Hydroxychloroquine in Rheumatoid Arthritis: A Systematic Review. Arthritis Care Res (Hoboken). 2020;72(1):36–40. 10.1002/acr.23826 . [DOI] [PubMed] [Google Scholar]
  • 13.Fan HW, Ma ZX, Chen J, Yang XY, Cheng JL, Li YB. Pharmacokinetics and Bioequivalence Study of Hydroxychloroquine Sulfate Tablets in Chinese Healthy Volunteers by LC-MS/MS. Rheumatol Ther. 2015;2(2):183–95. 10.1007/s40744-015-0012-0 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 14.Colson P, Rolain JM, Lagier JC, Brouqui P, Raoult D. Chloroquine and hydroxychloroquine as available weapons to fight COVID-19. Int J Antimicrob Agents. 2020:105932. 10.1016/j.ijantimicag.2020.105932 . [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15.Mehnert JM, Kaveney AD, Malhotra J, Spencer K, Portal D, Goodin S, et al. A phase I trial of MK-2206 and hydroxychloroquine in patients with advanced solid tumors. Cancer Chemother Pharmacol. 2019;84(4):899–907. 10.1007/s00280-019-03919-x . [DOI] [PubMed] [Google Scholar]
  • 16.Munster T, Gibbs JP, Shen D, Baethge BA, Botstein GR, Caldwell J, et al. Hydroxychloroquine concentration-response relationships in patients with rheumatoid arthritis. Arthritis Rheum-Us. 2002;46(6):1460–9. 10.1002/art.10307 WOS:000176199200006. [DOI] [PubMed] [Google Scholar]
  • 17.Soichot M, Megarbane B, Houze P, Chevillard L, Fonsart J, Baud FJ, et al. Development, validation and clinical application of a LC-MS/MS method for the simultaneous quantification of hydroxychloroquine and its active metabolites in human whole blood. J Pharm Biomed Anal. 2014;100:131–7. 10.1016/j.jpba.2014.07.009 . [DOI] [PubMed] [Google Scholar]
  • 18.Qu Y, Noe G, Breaud AR, Vidal M, Clarke WA, Zahr N, et al. Development and validation of a clinical HPLC method for the quantification of hydroxychloroquine and its metabolites in whole blood. Future Sci OA. 2015;1(3):FSO26. 10.4155/fso.15.24 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19.Parnham MJ, Erakovic Haber V, Giamarellos-Bourboulis EJ, Perletti G, Verleden GM, Vos R. Azithromycin: mechanisms of action and their relevance for clinical applications. Pharmacol Ther. 2014;143(2):225–45. 10.1016/j.pharmthera.2014.03.003 . [DOI] [PubMed] [Google Scholar]
  • 20.Chico RM, Chandramohan D. Azithromycin plus chloroquine: combination therapy for protection against malaria and sexually transmitted infections in pregnancy. Expert Opin Drug Metab Toxicol. 2011;7(9):1153–67. 10.1517/17425255.2011.598506 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 21.Kshirsagar NA, Gogtay NJ, Moran D, Utz G, Sethia A, Sarkar S, et al. Treatment of adults with acute uncomplicated malaria with azithromycin and chloroquine in India, Colombia, and Suriname. Res Rep Trop Med. 2017;8:85–104. 10.2147/RRTM.S129741 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 22.Supattanapong S, Konsil J. Solid phase extraction and high performance liquid chromatography for the determination of azithromycin in human plasma. Southeast Asian J Trop Med Public Health. 2008;39(6):978–87. . [PubMed] [Google Scholar]
  • 23.Wang LZ, Ong RYL, Chin TM, Thuya WL, Wan SC, Wong ALA, et al. Method development and validation for rapid quantification of hydroxychloroquine in human blood using liquid chromatography-tandem mass Spectrometry. J Pharmaceut Biomed. 2012;61:86–92. 10.1016/j.jpba.2011.11.034 WOS:000299861700013. [DOI] [PubMed] [Google Scholar]
  • 24.Fuzery AK, Breaud AR, Emezienna N, Schools S, Clarke WA. A rapid and reliable method for the quantitation of hydroxychloroquine in serum using turbulent flow liquid chromatography-tandem mass spectrometry. Clin Chim Acta. 2013;421:79–84. 10.1016/j.cca.2013.02.018 . [DOI] [PubMed] [Google Scholar]
  • 25.Chhonker YS, Sleightholm RL, Li J, Oupicky D, Murry DJ. Simultaneous quantitation of hydroxychloroquine and its metabolites in mouse blood and tissues using LC-ESI-MS/MS: An application for pharmacokinetic studies. J Chromatogr B. 2018;1072:320–7. 10.1016/j.jchromb.2017.11.026 WOS:000423637600041. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 26.Filist M, Bus-Kwasnik K, Ksycinska H, Rudzki PJ. Simplified LC-MS/MS method enabling the determination of azithromycin in human plasma after a low 100mg dose administration. J Pharm Biomed Anal. 2014;100:184–9. 10.1016/j.jpba.2014.07.015 . [DOI] [PubMed] [Google Scholar]
  • 27.Nirogi RVS, Kandikere VN, Shukla M, Mudigonda K, Maurya S, Boosi R, et al. Sensitive and selective liquid chromatography-tandem mass spectrometry method for the quantification of azithromycin in human plasma. Anal Chim Acta. 2005;553(1–2):1–8. 10.1016/j.aca.2005.08.007 WOS:000233077800001. [DOI] [Google Scholar]
  • 28.CPQA. Clinical Pharmacology Quality Assurance (CPQA) Guidelines for Bioanalytical Chromatographic Method, Development, Validation, and Application, v1.0. Effective April 1, 2017. 2017. Epub April 1, 2017.
  • 29.FDA. US Food and Drug Administration (FDA) Guidance for Industry–Bioanalytical Method Validation (BMV), May 2018. 2018.
  • 30.Kjellin LL, Dorsey G, Rosenthal PJ, Aweeka F, Huang L. Determination of the antimalarial drug piperaquine in small volume pediatric plasma samples by LC-MS/MS. Bioanalysis. 2014;6(23):3081–9. 10.4155/bio.14.254 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 31.Christoph Gohlke. https://www.lfd.uci.edu/~gohlke/molmass/?q=C14H18ClN3. Available from: https://www.lfd.uci.edu/~gohlke/molmass/?q=C14H18ClN3.
  • 32.Tan AM, Awaiye K. Use of Internal Standards in Lc-Ms Bioanalysis. Handbook of Lc-Ms Bioanalysis: Best Practices, Experimental Protocols, and Regulations. 2013:217–27. Book_Doi 10.1002/9781118671276 WOS:000332641500018. [DOI] [Google Scholar]
  • 33.Ben-Eltriki M, Somayaji V, Padwal RS, Brocks DR. A liquid chromatography-mass spectrometric method for the quantification of azithromycin in human plasma. Biomed Chromatogr. 2013;27(8):1012–7. 10.1002/bmc.2896 . [DOI] [PubMed] [Google Scholar]
  • 34.Almeida AM, Castel-Branco MM, Falcao AC. Linear regression for calibration lines revisited: weighting schemes for bioanalytical methods. J Chromatogr B. 2002;774(2):215–22. Pii S1570-0232(02)00244-1 10.1016/s1570-0232(02)00244-1 WOS:000176996300010. [DOI] [PubMed] [Google Scholar]
  • 35.Kirkup L, Mulholland M. Comparison of linear and non-linear equations for univariate calibration. J Chromatogr A. 2004;1029(1–2):1–11. 10.1016/j.chroma.2003.12.013 . [DOI] [PubMed] [Google Scholar]
  • 36.Brocks DR, Skeith KJ, Johnston C, Emamibafrani J, Davis P, Russell AS, et al. Hematologic disposition of hydroxychloroquine enantiomers. J Clin Pharmacol. 1994;34(11):1088–97. 10.1002/j.1552-4604.1994.tb01986.x . [DOI] [PubMed] [Google Scholar]
  • 37.Sinha N, Balayla G. Hydroxychloroquine and COVID-19. Postgrad Med J. 2020;96(1139):550–5. 10.1136/postgradmedj-2020-137785 . [DOI] [PMC free article] [PubMed] [Google Scholar]

Decision Letter 0

Pasquale Avino

23 Dec 2020

PONE-D-20-26989

A validated LC-MS/MS method for simultaneous determination of azithromycin, hydroxychloroquine, and its metabolites desethylhydroxychloroquine and bisdesethylchoroquine in human plasma

PLOS ONE

Dear Dr. Huang,

Thank you for submitting your manuscript to PLOS ONE. After careful consideration, we feel that it has merit but does not fully meet PLOS ONE’s publication criteria as it currently stands. Therefore, we invite you to submit a revised version of the manuscript that addresses the points raised during the review process.

Please submit your revised manuscript by Jan 15 2021 11:59PM. If you will need more time than this to complete your revisions, please reply to this message or contact the journal office at plosone@plos.org. When you're ready to submit your revision, log on to https://www.editorialmanager.com/pone/ and select the 'Submissions Needing Revision' folder to locate your manuscript file.

Please include the following items when submitting your revised manuscript:

  • A rebuttal letter that responds to each point raised by the academic editor and reviewer(s). You should upload this letter as a separate file labeled 'Response to Reviewers'.

  • A marked-up copy of your manuscript that highlights changes made to the original version. You should upload this as a separate file labeled 'Revised Manuscript with Track Changes'.

  • An unmarked version of your revised paper without tracked changes. You should upload this as a separate file labeled 'Manuscript'.

If you would like to make changes to your financial disclosure, please include your updated statement in your cover letter. Guidelines for resubmitting your figure files are available below the reviewer comments at the end of this letter.

If applicable, we recommend that you deposit your laboratory protocols in protocols.io to enhance the reproducibility of your results. Protocols.io assigns your protocol its own identifier (DOI) so that it can be cited independently in the future. For instructions see: http://journals.plos.org/plosone/s/submission-guidelines#loc-laboratory-protocols

We look forward to receiving your revised manuscript.

Kind regards,

Pasquale Avino, Ph.D.

Academic Editor

PLOS ONE

Journal requirements:

When submitting your revision, we need you to address these additional requirements.

1. Please ensure that your manuscript meets PLOS ONE's style requirements, including those for file naming. The PLOS ONE style templates can be found at

https://journals.plos.org/plosone/s/file?id=wjVg/PLOSOne_formatting_sample_main_body.pdf and

https://journals.plos.org/plosone/s/file?id=ba62/PLOSOne_formatting_sample_title_authors_affiliations.pdf

2. For reproducibility purposes please clearly specify in your methods section the source of the blood and plasma used in your study (eg. brand, product number)

3. Thank you for stating in your Funding Statement:

"This work was partially supported by the National Institutes of Health (NIH) through AIDS clinical Trials Group (ACTG), grant number 1UM1 AI068636 (F.A.). URL: www.NIH.gov. The funder plays no role in the study or prepataion of the manuscript."

Please provide an amended statement that declares *all* the funding or sources of support (whether external or internal to your organization) received during this study, as detailed online in our guide for authors at http://journals.plos.org/plosone/s/submit-now.  Please also include the statement “There was no additional external funding received for this study.” in your updated Funding Statement.

Please include your amended Funding Statement within your cover letter. We will change the online submission form on your behalf.

4. We note that you have included the phrase “data not shown” in your manuscript. Unfortunately, this does not meet our data sharing requirements. PLOS does not permit references to inaccessible data. We require that authors provide all relevant data within the paper, Supporting Information files, or in an acceptable, public repository. Please add a citation to support this phrase or upload the data that corresponds with these findings to a stable repository (such as Figshare or Dryad) and provide and URLs, DOIs, or accession numbers that may be used to access these data. Or, if the data are not a core part of the research being presented in your study, we ask that you remove the phrase that refers to these data.

__________________

Reviewers' comments:

Reviewer #1: Comments to Author:

A validated LC-MS/MS method for simultaneous determination of azithromycin, hydroxychloroquine, and its metabolites desethylhydroxychloroquine and

bisdesethylchoroquine in human plasma

The paper presented in this study is a validated LC-MS/MS method for simultaneous determination of Hydroxychloroquine (HCQ) and azithromycin (AZM) are antimalarial drugs recently reported to be active against severe acute respiratory syndrome coronavirus-2 (SARS-CoV-2) and quantitation method to assess the pharmacokinetics of AZM, HCQ, and its metabolites desethylhydroxychloroquine (DHCQ) and bisdesethylchloroquine (BDCQ) in patients’ plasma. The work is well written and organized and the results are satisfactorily supported by the reported data. Figures and tables are comprehensive and helpful. However, some minor changes are required as discussed in the following:

1. Abstract should be limited to 200 words

2. Keywords should not be the repetitions of the title words, please find such words which are not in the title, this way search engines of the web will find your manuscript with higher probability.

3. Suggestions for improvements in the Title

4. The structure of scientific publication should include the general chapters (Introduction, Materials and Methods, Results, Discussion, Conclusion). Please follow instructions on journal webpage. Conclusion chapter is missing.

5. Recommendations for future studies are needed in the conclusion section. Kindly provide strong recommendations for future researches.

6. Some citations are missing from the References section.

7. Some references are not cited in the text.

8. Formatting does not match journal criteria. (e.g. References section; section title spacing; paragraph indents)

The paper generates the following kinds of data.

1. Only 20 µL plasma sample by volume is needed for simultaneous quantitation of AZM, HCQ, DHCQ, and BDCQ.

2. The run time is 3.5 min which is fast turnaround time.

3. Statistical measurements are missing in whole of the data which should be added in the manuscript.

However, before I can recommend its publication, the authors should address the following questions

Some questions to author

1. The %differences from controls are 3.8% and -2.9% for AZM, 1.3% and 2.7% for HCQ, -1.9% and -0.9% for DHCQ, and -2.0% and -5.7% for BDCQ. Can author explain what would be the structural changes?

2. Why the study is important, what benefits readers can get from this study?

Reviewer #2: This is an interesting manuscript describing a method to analyze the compounds of interest. The authors follow the FDA and CPQA methodology. The quality of the data and results are outstanding. There are several minor points that the authors should address prior to acceptance of the manuscript. These points are listed below in order of appearance in the paper.

1. Page 5, line 3: There are two periods after [21].

2. Page 6, Figure 1: I recommend having the same relative geometry of the rings in the structure for the last 6 compounds to aid in visualizing the differences between structures.

3. Page 6, Section 2.2: PFP should be defined for the more general reader.

4. Page 7, section 2.4: HLB should be defined for the more general reader.

5. Page 5 and 22-23: These equations should be reformatted and numbered using the recommended standards of the journal.

6. Page 19, end of Room Temperature Stability in Plasma section: (Table] should be replaced with (Table 5).

While revising your submission, please upload your figure files to the Preflight Analysis and Conversion Engine (PACE) digital diagnostic tool, https://pacev2.apexcovantage.com/. PACE helps ensure that figures meet PLOS requirements. To use PACE, you must first register as a user. Registration is free. Then, login and navigate to the UPLOAD tab, where you will find detailed instructions on how to use the tool. If you encounter any issues or have any questions when using PACE, please email PLOS at figures@plos.org. Please note that Supporting Information files do not need this step.

PLoS One. 2021 Mar 5;16(3):e0247356. doi: 10.1371/journal.pone.0247356.r002

Author response to Decision Letter 0


16 Jan 2021

Comments from the Editor’s office.

1. Please ensure that your manuscript meets PLOS ONE's style requirements, including those for file naming.

Response: The manuscript was revised with updated format and style.

2. For reproducibility purposes please clearly specify in your methods section the source of the blood and plasma used in your study (eg. brand, product number)

Response: blank blood and plasma sources are now added.

3. Thank you for stating in your Funding Statement:

"This work was partially supported by the National Institutes of Health (NIH) through AIDS clinical Trials Group (ACTG), grant number 1UM1 AI068636 (F.A.). URL: www.NIH.gov. The funder plays no role in the study or prepataion of the manuscript."

Please provide an amended statement that declares *all* the funding or sources of support (whether external or internal to your organization) received during this study, as detailed online in our guide for authors at http://journals.plos.org/plosone/s/submit-now. Please also include the statement “There was no additional external funding received for this study.” in your updated Funding Statement.

Please include your amended Funding Statement within your cover letter. We will change the online submission form on your behalf.

Response: amended funding statement was included in this letter.

4. We note that you have included the phrase “data not shown” in your manuscript. Unfortunately, this does not meet our data sharing requirements. PLOS does not permit references to inaccessible data. We require that authors provide all relevant data within the paper, Supporting Information files, or in an acceptable, public repository. Please add a citation to support this phrase or upload the data that corresponds with these findings to a stable repository (such as Figshare or Dryad) and provide and URLs, DOIs, or accession numbers that may be used to access these data. Or, if the data are not a core part of the research being presented in your study, we ask that you remove the phrase that refers to these data.

Response: We addressed the “data not shown” issues by providing the data in Supporting S2 Figure and S2 Tables.

__________________

Reviewers' comments:

Reviewer #1: Comments to Author:

1. Abstract should be limited to 200 words

Response: The abstract was truncated but still over 300 words limit. We deem exception should be granted in this case as this is a comprehensive assay with 4 analytes.

2. Keywords should not be the repetitions of the title words, please find such words which are not in the title, this way search engines of the web will find your manuscript with higher probability.

Response: additional key words are included: malaria; solid-phase extraction;

3. Suggestions for improvements in the Title

Response: The title is updated as follows: Development and validation of an LC-MS/MS method for determination of hydroxychloroquine, its two metabolites, and azithromycin in EDTA-treated human plasma.

4. The structure of scientific publication should include the general chapters (Introduction, Materials and Methods, Results, Discussion, Conclusion). Please follow instructions on journal webpage. Conclusion chapter is missing.

Response: we have revised the conclusion section (4), and discussion was incorporated in the results section.

5. Recommendations for future studies are needed in the conclusion section. Kindly provide strong recommendations for future researches.

Response: In Conclusion section, we discussed the potential use of the method for studies other than Covid-19. We also added the following sentences:

“For it requires only a small sample volume, our method can be used for pediatric studies where sample volume is limited, and it can be coupled with capillary tube sampling by a finger prick or more advanced microsampling techniques such as Seventh Sense Tap� to facilitate clinical studies. With the highly sensitive LC-MS/MS system, our method may be modified for dried blood spot samples.”

6. Some citations are missing from the References section.

Response: The citations in text can all be found in references section

7. Some references are not cited in the text.

Response: All 37 references are now cited in the text using ENDNote program.

8. Formatting does not match journal criteria. (e.g. References section; section title spacing; paragraph indents)

The paper generates the following kinds of data.

1. Only 20 µL plasma sample by volume is needed for simultaneous quantitation of AZM, HCQ, DHCQ, and BDCQ.

2. The run time is 3.5 min which is fast turnaround time.

3. Statistical measurements are missing in whole of the data which should be added in the manuscript.

However, before I can recommend its publication, the authors should address the following questions

Some questions to author

1. The %differences from controls are 3.8% and -2.9% for AZM, 1.3% and 2.7% for HCQ, -1.9% and -0.9% for DHCQ, and -2.0% and -5.7% for BDCQ. Can author explain what would be the structural changes?

2. Why the study is important, what benefits readers can get from this study?

Response: Formatting was updated. The % differences were calculated from the measured control QC-L and QC-H concentrations. The data to calculate the % differences are now provided in Supporting S2 tables.

The method was developed to support potential clinical studies. Readers in the field of COVID-19, malaria, and inflammatory diseases using these drugs may be interested in the method and drug stability data. The small sample volume used in our method will benefit pediatric studies as stated in the conclusion section now. Furthermore, the approaches for method development are exemplary for readers in bioanalysis field,.

Reviewer #2:

1. Page 5, line 3: There are two periods after [21].

Response: The extra period is removed now.

2. Page 6, Figure 1: I recommend having the same relative geometry of the rings in the structure for the last 6 compounds to aid in visualizing the differences between structures.

Response: The Figure 1 is updated.

3. Page 6, Section 2.2: PFP should be defined for the more general reader.

Response: It is now defined as follows: pentafluorophenyl (PFP)

4. Page 7, section 2.4: HLB should be defined for the more general reader.

Response: It is now defined as follows: Hydrophilic lipophilic balance (HLB) solid phase extraction.

5. Page 5 and 22-23: These equations should be reformatted and numbered using the recommended standards of the journal.

Response: The equations are reformatted and numbered now.

6. Page 19, end of Room Temperature Stability in Plasma section: (Table] should be replaced with (Table 5).

Response: It is now corrected.

Attachment

Submitted filename: ResponseToReviewers.docx

Decision Letter 1

Pasquale Avino

8 Feb 2021

Development and validation of an LC-MS/MS method for determination of hydroxychloroquine, its two metabolites, and azithromycin in EDTA-treated human plasma

PONE-D-20-26989R1

Dear Dr. Huang,

We’re pleased to inform you that your manuscript has been judged scientifically suitable for publication and will be formally accepted for publication once it meets all outstanding technical requirements.

Within one week, you’ll receive an e-mail detailing the required amendments. When these have been addressed, you’ll receive a formal acceptance letter and your manuscript will be scheduled for publication.

An invoice for payment will follow shortly after the formal acceptance. To ensure an efficient process, please log into Editorial Manager at http://www.editorialmanager.com/pone/, click the 'Update My Information' link at the top of the page, and double check that your user information is up-to-date. If you have any billing related questions, please contact our Author Billing department directly at authorbilling@plos.org.

If your institution or institutions have a press office, please notify them about your upcoming paper to help maximize its impact. If they’ll be preparing press materials, please inform our press team as soon as possible -- no later than 48 hours after receiving the formal acceptance. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information, please contact onepress@plos.org.

Kind regards,

Pasquale Avino, Ph.D.

Academic Editor

PLOS ONE

Acceptance letter

Pasquale Avino

19 Feb 2021

PONE-D-20-26989R1

Development and validation of an LC-MS/MS method for determination of hydroxychloroquine, its two metabolites, and azithromycin in EDTA-treated human plasma

Dear Dr. Huang:

I'm pleased to inform you that your manuscript has been deemed suitable for publication in PLOS ONE. Congratulations! Your manuscript is now with our production department.

If your institution or institutions have a press office, please let them know about your upcoming paper now to help maximize its impact. If they'll be preparing press materials, please inform our press team within the next 48 hours. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information please contact onepress@plos.org.

If we can help with anything else, please email us at plosone@plos.org.

Thank you for submitting your work to PLOS ONE and supporting open access.

Kind regards,

PLOS ONE Editorial Office Staff

on behalf of

Professor Pasquale Avino

Academic Editor

PLOS ONE

Associated Data

    This section collects any data citations, data availability statements, or supplementary materials included in this article.

    Supplementary Materials

    S1 Fig. Linearity of AZM calibration curve in ESI versus APCI.

    (PDF)

    S2 Fig. LLOQ versus ULOQ chromatograms.

    (PDF)

    S1 Table. Adsorption of internal standards on container surface.

    (DOCX)

    S2 Table. Stability data.

    (PDF)

    S1 File. Assay standard operating procedure.

    (PDF)

    Attachment

    Submitted filename: ResponseToReviewers.docx

    Data Availability Statement

    All relevant data are within the manuscript and its Supporting Information files.


    Articles from PLoS ONE are provided here courtesy of PLOS

    RESOURCES