Efficient Regularized Field Map Estimation in 3D MRI

Claire Yilin Lin 1, Jeffrey A Fessler 2
PMCID: PMC7943027  NIHMSID: NIHMS1642345  PMID: 33693053

Abstract

Magnetic field inhomogeneity estimation is important in some types of magnetic resonance imaging (MRI), including field-corrected reconstruction for fast MRI with long readout times, and chemical shift based water-fat imaging. Regularized field map estimation methods that account for phase wrapping and noise involve nonconvex cost functions that require iterative algorithms. Most existing minimization techniques were computationally or memory intensive for 3D datasets, and are designed for single-coil MRI. This paper considers 3D MRI with optional consideration of coil sensitivity, and addresses the multi-echo field map estimation and water-fat imaging problem. Our efficient algorithm uses a preconditioned nonlinear conjugate gradient method based on an incomplete Cholesky factorization of the Hessian of the cost function, along with a monotonic line search. Numerical experiments show the computational advantage of the proposed algorithm over state-of-the-art methods with similar memory requirements.

Keywords: Magnetic field inhomogeneity, field map estimation, water-fat imaging, preconditioned conjugate gradient, monotonic line search, incomplete Cholesky factorization

I. Introduction

In magnetic resonance imaging (MRI), scans with long readout times require correction for magnetic field inhomogeneity during reconstruction to avoid artifacts [1]-[5]. Field inhomogeneity is also a nuisance parameter in chemical shift based water-fat imaging techniques [6]-[11]. Field map estimation is thus crucial to field-corrected MR image reconstruction, and for fat and water image separation.

One field map estimation approach is to acquire MR scans at multiple echo times (usually 2 or 3), where a small echo time difference can help resolve any phase wrapping issues and a large echo time difference can help improve SNR. One can then estimate field inhomogeneity using images reconstructed from these scans [5]. Since field maps tend to be smooth within tissue, estimation methods with smoothness assumptions have been proposed for water-fat separation, including region growing techniques [12]-[17], filtering [18], curve fitting [19]-[21], multiresolution and subspace approaches [21]-[24], and graph cut algorithms [25]. To improve robustness of water and fat separation and reduce ambiguity of assignment, field map preestimation methods such as demodulation [26] and magnetization transfer [27] have been proposed as part of the water-fat imaging framework. Most of these methods, however, use various approximations to account for phase wrapping between different acquisitions. In contrast, regularized estimation methods [5],[8]-[10] have been proposed to account for both phase wrapping and the smoothness of the field map from multiple acquisition images. Because the field map affects image phase, these approaches involve a nonconvex optimization problem that requires iterative methods.

To solve such optimization problems, [5],[9],[28] use a majorization-minimization (MM) approach by introducing a quadratic majorizer for their cost functions. The MM approach decreases the cost monotonically, but is computationally intensive, especially for large-scale datasets. Other regularized field map estimation minimization techniques quantize the solution space [8],[10] and may require a second descent algorithm to produce sufficiently smooth estimates. An alternative minimization technique [29] uses nonlinear conjugate gradient (NCG) with a monotonic line search (MLS), and explored various preconditioners in the 3D single-coil case.

This paper considers the regularized field map estimation problem in the 3D multi-coil MRI setting. In particular, we consider a generalized cost function in the multi-coil case for both multi-echo field map estimation and water-fat imaging. We minimize it by a NCG algorithm with an efficient MLS and an iteration-dependent preconditioner based on an incomplete Cholesky factorization [30] of the Hessian of the cost function. The incomplete Cholesky factorization has been applied to field inhomogeneity estimation using surface fitting [31], and recently to single-coil field map estimation with a similar cost function [29]. In addition to faster convergence, this preconditioner exploits the sparse structure of the Hessian, thus it is memory efficient and scales to 3D datasets. Compared to previous works [9],[28],[29], our new approach unifies the field map correction and the water-fat imaging problems, with a generalized expression that optionally considers multiple coils in MRI. Our efficient algorithm on this problem shows significant computational and storage advantages compared with existing MM and NCG methods.

The rest of this paper is organized as follows. Section II describes the optimization problem for the field map estimation problems for multi-coil MRI. Section III presents the NCG-MLS optimization scheme with the proposed preconditioner. Section IV reports simulated and real experimental results, followed by conclusions in Section V.

II. Problem Formulation

We are given reconstructed images yclCNv for the cth receiver coil of the lth scan, with c = 1, … , Nc, l = 1,…, L, where Nv denotes the total number of voxels in the image, Nc denotes the number of coils, and L ≥ 2 denotes the number of echo times. We model the field inhomogeneity effect as

yclj=eiωjtlscjxlj+ϵclj, (1)

where j = 1, …, Nv is the voxel index, ωRNv is the unknown field map, tlR is the echo time shift of the lth scan, scCNv is the (known) coil sensitivity map for the cth coil, and ϵclCNv denotes the noise. For single-coil MRI, or when the coil images are combined as a preprocessing step, we have Nc = 1 and s = 1 in (1).

The unknown image xlCNv for the lth echo is problem-dependent, where

xlj={mjin field map estimation,mw,j+mf,jp=1Pαpei2πΔf,ptlin water-fat imaging,}

where m, mw, mfCNv are respectively the magnetization, water, and fat components, and Δf,pR denotes the (known) frequency shifts of P fat peaks in the multipeak fat model [32],[33],[10] with relative amplitudes p=1Pαp=1 that can be estimated and averaged over all fat pixels as a preprocessing step by existing methods [34]. The goal of the field map estimation problem is to estimate ω and x given y and s.

Assuming the noise ϵ is zero-mean, white complex Gaussian, the joint maximum-likelihood (ML) estimates of the field map ω and image x are the minimizers of the negative log-likelihood as follows:

argminω,xΦ~(ω,x),whereΦ~(ω,x)=j=1Nvl=1Lc=1Ncycljeiωjtlscjxlj2. (2)

For a given field map ω, the ML estimate of x has a closed-form expression [8],[28] that one can substitute into (2) to give a cost function in terms of ω:

Φ(ω)=minxΦ~(ω,x)=j=1Nvm,n=1Lc,d=1Ncϕcdmnj(ωj), (3)

where

ϕcdmnj(ωj)rcdmnj[1cos(rcdmnj+ωj(tmtn))],rcdmnjΓmnc=1Ncscj2scjsdjycmjydnj,Γγ(γγ)1γ, (4)

where ·* denotes the complex conjugate, and L × L matrix Γ is defined in terms of

γ={1in field map estimation,[1p=1Pαpei2πΔf,pt]in water-fat imaging,} (5)

in which 1 denotes an all one vector of length L, and the exponential is applied element-wise. In the field map estimation case, this simplifies to Γmn = 1/Lm, n.

As B0 field maps tend to be spatially smooth in MRI, we add a regularization term to (3) to form a penalized-likelihood (PL) cost function

Ψ(ω)=Φ(ω)+β2Cω22, (6)

where C is a first or second order finite difference operator with optional spatial weights as in [10]. Such regularization has been used in many other prior works [5],[28],[29].

III. Efficient Algorithm

Several approaches have been proposed to solve the field map estimation problem in the single-coil setting, but are demanding in computation or memory. In particular, a quadratic majorizer with a diagonal Hessian [5] takes many iterations to converge even for 2D images, and a quadratic majorizer with an optimal curvature that inverts a Nv × Nv Hessian matrix [28] is memory-limited to small-scale data. In water-fat imaging, [10],[35] process data in a single-coil manner using the graph cut method. Since graph cut requires discretization, [10] proposes to overcome this limitation by additionally running a descent algorithm such as in [9], which considers a quadratic majorizer with a diagonal Hessian that convergences slowly.

Here, we optimize (6) using NCG with a monotonic line search [29], and consider a preconditioner with efficient computation and memory storage. Our field map estimation procedure is tabulated in the Algorithm below. For NCG, we choose the Polak-Ribiere update to compute a μi that satisfies the conjugacy condition [36].

After estimating the field map ω^, we estimate the water and fat components for each voxel in water-fat imaging by applying the closed-form expression [8] using ω^:

[mw,jmf,j]=((γdiag(eiωjt))sj)yj, (7)

where ⊗ denotes the Kronecker product, (·) denotes the pseudo inverse, and sjCNc denotes the coil sensitivity map for the jth voxel.

Next we present our initialization, choice of preconditioner, and derive our iterative monotone line search algorithm in the multi-coil setting.

A. Initialization

For field map estimation, we initialize ω by a field map computed from the phase of the first two echoes of the coil combined images:

(ωj)0=[(c=1Ncscjyc1j)(d=1Ncsdjyd2j)]/(t2t1). (8)

To initialize ω for water-fat imaging, we follow [9] and sweep through a range of 100 values from −∣Δf/2∣ to ∣Δf/2∣ for each voxel, and choose the value with minimal cost (3), denoted as ω~0. We then run a few CG iterations to minimize a penalized weighted least squares (PWLS) problem

ω0=argminωj=1Nvρj(ωjω~j0)2+β2Cω22, (9)

where the spatial weights

ρj=m,n=1Lc,d=1Ncrcdmnj

are given by (4). We then use ω0 as our initial estimate in the water-fat case.

To reduce ambiguity of water and fat assignment, one can also consider robust initialization schemes such as demodulation [26] or magnetization transfer [27].

B. Preconditioning matrices

To accelerate the NCG-based algorithm, given gradient gi of the cost at the ith NCG iteration, we explore a preconditioner Pi with memory efficient implementation of (Pi)−1gi using an incomplete Cholesky factorization [30]. In particular, the gradient gRNv is given by

g=Ψ(ω)=Φ(ω)+βCTCω, (10)

where

(Φ(ω))j=m,n=1Lc,d=1Ncrcdmnj(tmtn)2sin(rcdmnj+ω(tmtn)).

The Hessian of the cost (6) at the ith iteration is the sum of a diagonal matrix and an (approximately, due to the support mask) block Toeplitz with Toeplitz block (BTTB) matrix:

Hi=Di+βCTCRNv×Nv, (11)

where C is the finite difference operation and Di=diag(dji)¯0, where the Hessian of the negative log-likelihood has diagonal elements given by

dji=m,n=1Lc,d=1Ncκcdmnj(ucdmnj(ωji)), (12)

with

κcdmnj(u)=rcdmnj(tmtn)2sin(u)u,anducdmnj(ω)=(rcdmnj+ω(tmtn))modπ. (13)

Since the terms rcdmnj and tmtn are shared across iterations, we precompute them at the initialization stage to efficiently calculate the gradient and Hessian at each iteration i. Note also that Hi is positive definite as long as at least one value of dji is positive (which is true for any nontrivial problem).

Although Hi is sparse and banded, its inverse is approximately full, so directly computing the inverse would require far too much memory. To reduce memory, we propose to use a preconditioner that approximates the symmetric Hessian with a LU factorization of the form

Pi=Li(Li)THi, (14)

where LiRNv×Nv is sparse lower triangular, enabling efficient computation (via back-substitution) of (Pi)−1gi in the precondition step. Taking advantage of the sparsity and positive definiteness of our Hessian (11), preconditioning with an incomplete Cholesky factorization reduces both computation and memory. A popular form of the incomplete Cholesky factorization matches the matrix H on its nonzero set, thus is at least as sparse as H. Similar preconditioning with incomplete LU factorization has been used for simulating anisotropic diffusion in MRI [37]. In practice, for a better approximation one can control the sparsity of the factors by defining a tolerance on the magnitude of the elements of H (below which entries in the factors are set to zero), with the trade-off between approximation accuracy and memory storage.

Fig. 1 illustrates the memory improvement by a toy problem of image size 20×16×8, where we compute H = D+βC C and its inverse, with randomly chosen diagonal elements dj ∈ (0, 0.1) and β = 0.1. Fig. 1 considers the incomplete Cholesky factorization without tolerance, denoted L0, and with a tolerance of Hmax × 10−3, denoted Lt, where Hmax is the element in H with maximum magnitude. Fig. 1 shows the sparse structure of H, its nonsparse inverse H−1, and the Cholesky factorizations as well as their approximation errors. Table I shows the number of nonzero elements of each matrix, their memory storage, and their errors that affect the convergence rate, using the normalized root mean square error (NRMSE) IL1HLTFNv for each factorization L in our example. Fig. 2 illustrates how the sparsity of Lt changes with respect to its tolerance by showing the percentage of nonzero elements in Lt versus the scaling factor of Hmax in the tolerance.

Fig. 1.

Fig. 1.

Matrix structure of each factorization and the error of its inverse, in a toy problem of size 20 × 16 × 8.

Table I.

Number of nonzero elements, memory usage, and NRMSE of the inverse of each factorization in a toy problem of size 20 × 16 × 8.

H H−1 Lc Lt L0
Number of nonzeros (× 105) 1.67 655 72.5 1.77 0.96
Storage (megabytes) 0.31 100.1 11.9 0.53 0.27
NRMSE 3e-16 4e-3 3e-2

Fig. 2.

Fig. 2.

Change of sparsity of Lt with respect to the scaling factor of Hmax in its tolerance.

For memory storage in this case, the number of nonzero elements in the incomplete Cholesky factor without tolerance L0 is more than 70 times less than that in the (complete) Cholesky factor Lc, with more than 40 times memory saving. In general, we observe (by the banded structures) that the number of nonzero elements of Lc is lower bounded by (NvNxNy) * NxNy, while that of L0 is upper bounded by 4Nv. This leads to the generalization that L0 is at least (NvNxNy)/(4Nz) times more sparse than Lc, which scales significantly with the problem size. The storage of the incomplete Cholesky factor with tolerance Lt depends on the tolerance, and with the choice of tolerance here we observe a 40 times fewer nonzero values, saving memory by a factor of more than 20 compared with Lc.

The trade-off with a sparser factorization, however, is a worse approximation error. This is reflected in the error matrices in Fig. 1 and the NRMSE in Table I. While L0 has lower memory usage than Lt, the inverse is a worse approximation to H−1. In practice, nevertheless, both incomplete factorizations LL are positive definite, so as preconditioners they provide a descent direction in addition to storage advantage, whereas storing Lc is infeasible for realistically sized 3D datasets.

C. Monotonic step size line search

With a search direction given by NCG, the choice of step size is important for convergence of the algorithm. To avoid multiple function evaluations required by backtracking line search algorithms [38], we implement a recursive line search algorithm using a quadratic majorizer with an optimal curvature, which guarantees monotone decrease of the cost function [39].

In the line search step, given a current field map estimate ωi and a search direction ziRNv, we aim to find a step size that minimizes the cost (6):

α^=argminαf(α),wheref(α)=Φ(ωi+αzi)+β2C(ωi+αzi)22, (15)

We iteratively minimize the nonconvex problem (15) using a quadratic majorizer based on Huber’s method [40, p. 184] at the kth inner iteration (dropping outer iteration i for brevity):

qk(α)=Φ(ω+α(k)z)+zTΦ(ω+α(k)z)(αα(k))+12d(k)(αα(k))2+β2C(ω+αz)22,

where the optimal curvature is given by [28]

d(k)=j=1Nvzj2dj(k),where
Algorithm: Preconditioned NCG-MLS
Inputs:y,s,t,C,βIntialization:ω0by(8)or(9)z0=Ψ(ω0)α0=0precomputercdmnjby (4) andtmtnfori=0,1,,N1docompute gradientgi=Ψ(ωi)with (10)preconditionpi=(Pi)1giwith (14)computeμiwith conjugacysearch directionzi+1=pi+μiziRNvfork=0,1,,Ni1doupdate step sizeα(k+1)by (17)endforupdateωi+1=ωi+α(Ni)zi+1endforoutput:ωN
dj(k)=m,n=1Lc,d=1Ncκcdmnj(ucdmnj(ωj+α(k)zj)), (16)

with κcdmnj(·) and ucdmnj(·) defined in (13).

Using one step of Newton’s method on the quadratic majorizer qk(α) gives the step size update

α(k+1)=α(k)αqk(α(k))2α2qk(α(k))=α(k)αf(α(k))d(k)+βCz22. (17)

We implement (17) efficiently by computing Cz22 only once per outer NCG iteration i. Since the majorizer satisfies qk(α) ≥ f(α) for all step size α and inner line search iteration k, the update (17) guarantees monotonic decrease of the cost (15).

IV. Results

We investigated our algorithm and its efficiency with two multi-echo field map estimation experiments and two water-fat imaging experiments. Due to the large data size, memory intensive methods with a direct solver using the full Hessian are excluded from our experiments. In particular, we compare our incomplete Cholesky preconditioner (NCG-MLS-IC) method versus a quadratic majorizer update with diagonal Hessian (QM) [5] and versus the NCG algorithm without any preconditioner (NCG-MLS) and with a diagonal preconditioner (NCG-MLS-D) [28]. In addition, we used the Poblano toolbox [41] to compare the convergence of the quasi-Newton (QN) and truncated Newton (TN) methods in our simulations.

For each dataset, we define a mask using the convex hull of all voxels that contribute to the signal (with coil-combined image magnitude thresholded below by 0.1ymax, where ymax denotes the maximum image magnitude in the coil-combined image for the first echo time.), with a dilation of two voxels. We then computed ω within the mask, and tuned the regularization parameter β by sweeping across a range of values. All our experiments used MATLAB R2020a, with a 2.4-GHz dual-core Intel Core i7. The MATLAB code that reproduces the experiments with our efficient algorithm will be available as part of the Michigan Image Reconstruction Toolbox (MIRT) [42].

A. Brain Simulation

We first simulated a 3D brain dataset with 40 64 × 64 slices, 4 simulated coils and 3 echo times tl = 0, 2, 10 ms, with added complex Gaussian noise so that the SNR ≈ 20 dB. To generate multi-coil data, we simulated coil sensitivity maps with 4 coils based on [43] using the MIRT. We set β = 2−4 with first order regularization to achieve visual resemblance to the ground truth field map. In light of the trade-off between storage and approximation error discussed in Section III-B, we explored preconditioners using the incomplete Cholesky factorization both without tolerance (NCG-MLS-IC-0) and with a tolerance of Hmaxi×103 for each iteration i (NCG-MLS-IC).

Fig. 3 shows four selected slices, their initial field map, and the regularized estimate by our algorithm. To examine the speed of convergence, we plot the root mean square error (RMSE) ωiωtrue2Nv versus wall time in Fig. 4. The RMSE plots show that the QM and all the NCG-MLS methods converge to RMSE ≈ 5.6 Hz, though going through a slightly lower RMSE in the iterative process. Both the quasi-Newton and the truncated Newton methods converge to minimizers with higher RMSE, hence we omitted their comparison in the phantom experiment below. The plots show a significant computational gain of NCG-MLS preconditioned with the incomplete Cholesky factorization over all the other methods. We also observe that using a nonzero tolerance in the incomplete Cholesky factorization gives a slightly faster convergence than not using one, hence we adopt that choice for the NCG-MLS-IC implementations in our next experiments.

Fig. 3.

Fig. 3.

Top to bottom: selected slices of coil-combined simulation image, initial field map (in Hz), regularized field map estimate ω^, ground truth field map ωtrue, and error ω^ωtrue.

Fig. 4.

Fig. 4.

RMSE versus wall time of seven algorithms used in simulation. Every 10 iteration is marked by a dot.

B. Phantom Dataset

Our second experiment uses a Function Biomedical Informatics Research Network (FBIRN) phantom [44] with two pieces of metal staple to induce field inhomogeneity, collected on a GE MR750 3T scanner with a 32-channel Nova Head Coil receiver. This dataset has size 74 × 74 × 10 with 3 mm3 isotropic voxel size, TR = 10.5 ms, with 3 echo times tl = 0, 1, 2.3 ms. We computed coil sensitivity maps using ESPIRiT [45], and set β = 2−3 with first-order finite difference regularizaiton.

Fig. 5 shows four selected slices, their initial field map, and the regularized estimate by our algorithm. To compare convergence, we computed the root mean square difference (RMSD) ωiω2Nv to the converged ω of the QM method. The RMSD plots in Fig. 6 show that our algorithm converges much faster than the other three, reaching 0.33 Hz RMSD in 1 iteration, and 0.005 Hz RMSD in 2 iterations. Since this 3D dataset has a more realistic problem size than the simulated data, we quantify the convergence speedup by comparing the time it takes for each method to reach an RMSD below 0.5 Hz. Table II shows that our NCG-MLS algorithm with an incomplete Cholesky preconditioner provides a speedup of 15 times from NCG-MLS with a diagonal preconditioner, 18 times from that without a preconditioner, and 21 times from the quadratic majorizer implementation.

Fig. 5.

Fig. 5.

Top to bottom: selected slices of coil-combined phantom image, initial field map (in Hz), and regularized field map estimate.

Fig. 6.

Fig. 6.

RMSD versus wall time of four algorithms used in the phantom experiment. Every iteration is marked by a dot.

Table II.

Time for each method to reach an RMSD below 0.5 Hz, and their relative proportions to the time taken by NCG-MLS-IC.

QM NCG-MLS NCG-MLS-D NCG-MLS-IC
Time (s) 96 81 69 4.5
vs. IC time 21× 18× 15×

C. Cardiac Water-Fat Simulation

For water-fat imaging, we first performed a cardiac simulation based on one of the 8-echo datasets used in the ISMRM Fat-Water Toolbox [46]. Since implementations in the toolbox work only for 2D datasets, and coil combination such as [47] is often used in practice, we illustrate the flexibility of our algorithm in a 2D coil-combined case by simply setting the number of coils Nc = 1 and the coil sensitivitiy map s = 1. We also consider the multipeak model in water-fat imaging.

This dataset has size 256 × 192 with 8 echo times from 1.5 to 17.4 ms (each 2.3 ms apart). We generated ground truth field map and water and fat images using golden section search with multiresolution [22]. We used the same values {αp} and {Δf,p} as in the toolbox implementations both for simulating images with 8 echo times using the model (1) and for estimation. For comparison, we also ran the graph cut (GC) method [10] using the same cost (6) with second-order finite differences as in [10], and β = 2−7 as the regularization parameter.

Fig. 7 shows the first echo image, the initial field map ω~0 by voxel-wise estimation, and the initial ω0 after 10 CG iterations of PWLS minimization (9). Fig. 8 shows the ground truth field map, water and fat images, and the estimates and error images by the graph cut and by our algorithm. Compared with graph cut, our algorithm achieves slightly lower NRMSE on the water image (20.09% vs. 23.57%) and the fat image (20.93% vs. 23.43%), with lower final RMSE on the field map, shown in Fig. 9. To explore a combination suggested by [10], we ran 100 graph cut iterations followed by 100 optimal transfer iterations using a quadratic majorizer [9]. We used the implementation in the toolbox [46] which did not precompute rcdmnj by (4). Fig. 9 shows the graph cut RMSE curve jumps up (to 615 Hz) on its first iteration, and converges to its own minimizer. The subsequent quadratic majorizer update lowers the RMSE further, which opens a promising future direction of combining graph cut with the faster NCG-MLS-IC with precomputation of common terms. Fig. 9 also shows the truncated Newton and quasi-Newton methods again converge to their minimizers with higher RMSE. We omitted all methods with higher RMSE in the real data experiment below.

Fig. 7.

Fig. 7.

Left to right: simulated image for the 1st echo, initial field map ω~0 (in Hz) by voxel-wise estimation, and initial fieldmap ω0 by PWLS (9).

Fig. 8.

Fig. 8.

1st row: ground truth field map, water, and fat images. 2nd and 3rd row: graph cut estimates and their error images. 4th and 5th row: NCG-MLS-IC estimates and their error images.

Fig. 9.

Fig. 9.

Field map RMSE versus wall time of seven algorithms used in the water-fat simulation. Every 20 iterations is marked by a dot.

D. Ankle Water-Fat Dataset

We further illustrate our algorithm in the 3D multi-coil setting using an ankle dataset from the ISMRM Fat-Water Separation Dataset [46]. This dataset has 4 256 × 256 slices, 8 coils and 3 echo times tl = 2.2, 3, 3.8 ms, in a 3T scanner that corresponds to a single Δf ≈ 440 Hz. We chose β = 2−10 with first-order finite difference regularization to achieve visual separation of water and fat components.

Fig. 10 shows the first echo image, the initial field map ω~0 by voxel-wise estimation, the initial ω0 after 10 CG iterations of PWLS minimization (9), and the regularized estimate by our algorithm. For completeness, Fig. 10 also shows the estimated water and fat images using (7), which achieve a visual separation of the two components. However, it is worth emphasizing that our main interest is in the speed of finding a minimizer of the problem (6). In this case, since QM converged to a different local minimum than the other three methods, we computed the RMSD to ω of the NCG-MLS method (without preconditioner). The RMSD plots in Fig. 11 show a significant computational gain of our algorithm over the other algorithms.

Fig. 10.

Fig. 10.

Top to bottom: coil-combined water-fat image for the 1st echo, initial field map ω~0 (in Hz) by voxel-wise estimation, initial fieldmap ω0 by PWLS (9), regularized field map estimate, estimated water image, and estimated fat image.

Fig. 11.

Fig. 11.

Field map RMSD versus wall time of four algorithms used in the water-fat experiment. Every iteration is marked by a dot.

V. Conclusion

This paper presents an efficient algorithm for both multi-echo field map estimation and water-fat imaging problem in the 3D multi-coil MRI setting. Given the nonconvex cost function, our algorithm uses the nonlinear conjugate gradient method with a preconditioner based on an incomplete Cholesky factorization, and a monotonic step size line search based on a quadratic majorizer with optimal curvatures. This is the first work to use the incomplete Cholesky factorization as a preconditioner for multi-coil field map estimation. Experiments with simulation and real data show that our method has faster convergence than existing memory-efficient methods.

Acknowledgement

We thank Jon-Fredrik Nielsen at the University of Michigan for the multi-echo 3D multi-coil MRI phantom data.

This work was supported in part by NIH grants R01 EB023618, U01 EB026977, R21 AG061839, and NSF grant IIS 1838179.

Contributor Information

Claire Yilin Lin, Department of Mathematics, University of Michigan, Ann Arbor, MI, 48109 USA.

Jeffrey A. Fessler, Department of Electrical Engineering and Computer Science, University of Michigan, Ann Arbor, MI 48109 USA.

References

  • [1].Sekihara K, Matsui S, and Kohno H, “NMR imaging for magnets with large nonuniformities,” IEEE Trans. Med. Imag, vol. 4, no. 4, pp. 193–199, 1985. [DOI] [PubMed] [Google Scholar]
  • [2].Jezzard P and Balaban RS, “Correction for geometric distortion in echo planar images from B0 field variations,” Mag. Res. Med, vol. 34, no. 1, pp. 65–73, 1995. [DOI] [PubMed] [Google Scholar]
  • [3].Reber PJ, Wong EC, Buxton RB, and Frank LR, “Correction of off resonance-related distortion in echo-planar imaging using EPI-based field maps,” Mag. Res. Med, vol. 39, no. 2, pp. 328–330, 1998. [DOI] [PubMed] [Google Scholar]
  • [4].Noll DC, Fessler JA, Jeffrey A, and Sutton BP, “Conjugate phase MRI reconstruction with spatially variant sample density correction,” IEEE Trans. Med. Imag, vol. 24, no. 3, pp. 325–336, 2005. [DOI] [PubMed] [Google Scholar]
  • [5].Funai AK, Fessler JA, Yeo DTB, Olafsson VT, and Noll DC, “Regularized field map estimation in MRI,” IEEE Trans. Med. Imag, vol. 27, no. 10, pp. 1484–1494, 2008. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [6].Glover GH and Schneider E, “Three-point Dixon technique for true water/fat decomposition with B0 inhomogeneity correction,” Mag. Res. Med, vol. 18, no. 2, pp. 371–383, 1991. [DOI] [PubMed] [Google Scholar]
  • [7].Reeder SB, Wen Z, Yu H, Pineda AR, Gold GE, Markl M, and Pelc NJ, “Multicoil Dixon chemical species separation with an iterative least-squares estimation method,” Mag. Res. Med, vol. 51, no. 1, pp. 35–45, 2004. [DOI] [PubMed] [Google Scholar]
  • [8].Hernando D, Diego JP Haldar BP Sutton J Ma P. Kellman, and Liang Z-P, “Robust water/fat separation in the presence of large field inhomogeneities using a graph cut algorithm,” Mag. Res. Med, vol. 59, no. 3, pp. 571–580, 2008. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [9].Huh W, Fessler JA, and Samsonov AA, “Water-fat decomposition with regularized field map,” in Proc. Intl. Soc. Mag. Res. Med, 2008, p. 1382. [Google Scholar]
  • [10].Hernando D, Kellman P, Haldar JP, and Liang Z-P, “Robust water/fat separation in the presence of large field inhomogeneities using a graph cut algorithm,” Mag. Res. Med, vol. 63, no. 1, pp. 79–90, 2010. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [11].Bley TA, Wieben O, Franois CJ, Brittain JH, and Reeder SB, “Fat and water magnetic resonance imaging,” J. Mag. Res. Im, vol. 31, no. 1, pp. 4–18, 2010. [DOI] [PubMed] [Google Scholar]
  • [12].Reeder SB, Pineda AR, Wen Z, Shimakawa A, Yu H, Brittain JH, Gold GE, Beaulieu CH, and Pelc NJ, “Iterative decomposition of water and fat with echo asymmetry and least-squares estimation (IDEAL): Application with fast spin-echo imaging,” Mag. Res. Med, vol. 54, no. 3, pp. 636–644, 2005. [DOI] [PubMed] [Google Scholar]
  • [13].Yu H, Reeder SB, Shimakawa A, Brittain JH, and Pelc NJ, “Field map estimation with a region growing scheme for iterative 3-point water-fat decomposition,” Mag. Res. Med, vol. 54, no. 4, pp. 1032–1039, 2005. [DOI] [PubMed] [Google Scholar]
  • [14].Lu W and Hargreaves BA “Multiresolution field map estimation using golden section search for water-fat separation,” Mag. Res. Med, vol. 60, no. 1, pp. 236–144, 2008. [DOI] [PubMed] [Google Scholar]
  • [15].Jacob M and Sutton BP, “Algebraic decomposition of fat and water in MRI,” IEEE Trans. Med. Imag, vol. 28, no. 2, pp. 173–184, 2009. [DOI] [PubMed] [Google Scholar]
  • [16].Zhou K, Zaitsev M, and Bao S, “Reliable two-dimensional phase unwrapping method using region growing and local linear estimation,” Mag. Res. Med, vol. 62, no. 4, pp. 1085–1090, 2009. [DOI] [PubMed] [Google Scholar]
  • [17].Doneva M, Börnert P, Eggers H, Mertins A, Pauly J, and Lustig M, “Compressed sensing for chemical shift-based waterfat separation,” Mag. Res. Med, vol. 64, no. 6, pp. 1749–1759, 2010. [DOI] [PubMed] [Google Scholar]
  • [18].Windischberger C, Robinson S, Rauscher A, Barth M, and Moser E, “Robust field map generation using a triple-echo acquisition,” J. Mag. Res. Im, vol. 20, no. 4, pp. 730–734, 2004. [DOI] [PubMed] [Google Scholar]
  • [19].Schneider E and Glover G “Rapid in vivo proton shimming,” Mag. Res. Med, vol. 18, no. 2, pp. 335–347, 1991. [DOI] [PubMed] [Google Scholar]
  • [20].Irarrazabal P, Meyer CH, Nishimura DG, and Macovski A, “Inhomogeneity correction using an estimated linear field map,” Mag. Res. Med, vol. 35, no. 2, pp. 278–282, 1996. [DOI] [PubMed] [Google Scholar]
  • [21].Sharma SD, Hu HH, and Nayak KS, “Chemical shift encoded water-fat separation using parallel imaging and compressed sensing,” Mag. Res. Med, vol. 69, no. 2, pp. 456–466, 2013. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [22].Lu W and Hargreaves BA, “Multiresolution field map estimation using golden section search for water-fat separation,” Mag. Res. Med, vol. 60, no. 1, pp. 236–244, 2008. [DOI] [PubMed] [Google Scholar]
  • [23].Sharma SD, Hu HH, and Nayak KS, “Accelerated water-fat imaging using restricted subspace field map estimation and compressed sensing,” Mag. Res. Med, vol. 67, no. 3, pp. 650–659, 2012. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [24].Tsao J and Jiang Y “Hierarchical IDEAL: Fast, robust, and multiresolution separation of multiple chemical species from multiple echo times,” Mag. Res. Med, vol. 70, no. 1, pp. 155–159, 2013. [DOI] [PubMed] [Google Scholar]
  • [25].Bioucas-Dias JM and Valado G, “Phase unwrapping via graph cuts,” IEEE Trans. Imag. Proc, vol. 16, no. 3, pp. 698–709, 2007. [DOI] [PubMed] [Google Scholar]
  • [26].Diefenbach MN et al. , “Improving chemical shift encoding-based water–fat separation based on a detailed consideration of magnetic field contributions” Mag. Res. Med, vol. 80, no. 13 pp. 990–1004, 2018. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [27].Samsonov A, Liu F, Velikina Fang J. V., “Resolving estimation uncertainties of chemical shift encoded fat-water imaging using magnetization transfer effect,” Mag. Res. Med, vol. 82, no. 1, pp. 202–212, 2019. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [28].Allison MJ and Fessler JA, “Accelerated computation of regularized field map estimates,” in Proc. Intl. Soc. Mag. Res. Med, 2012, p. 0413. [Google Scholar]
  • [29].Ongie G, Shi J, and Fessler JA, “Efficient computation of regularized field map estimates in 3D,” IEEE ISBI, pp. 700–703, 2017. [Google Scholar]
  • [30].Manteuffel TA, “An incomplete factorization technique for positive definite linear systems,” Mathematics of Computation, vol. 34, no. 150, pp. 473–97, 1980. [Google Scholar]
  • [31].Lai SH, and Fang M, “A dual image approach for bias field correction in magnetic resonance imaging,” Mag. Res. Med, vol. 21, no. 2, pp. 121–125, 2003. [DOI] [PubMed] [Google Scholar]
  • [32].Reeder SB, Brittain JH, Grist TM, Yen YF, “Least-squares chemical shift separation for 13C metabolic imaging,” Mag. Res. Med, vol. 26, no. 4, pp. 1145–1152, 2007. [DOI] [PubMed] [Google Scholar]
  • [33].Bydder M, Yokoo T T, Hamilton G, Middleton MS, Chavez AD, Schwimmer JB, Lavine JE, Sirlin CB, “Relaxation effects in the quantification of fat using gradient echoimaging,” Mag. Res. Med, vol. 26, no. 3, pp. 347–359, 2008. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [34].Yu H, Shimakawa A, McKenzie CA, Brodsky EK, Brittain JH, Reeder SB, “Multiecho water-fat separation and simultaneous R2* estimation with multifrequency fat spectrum modeling,” Mag. Res. Med, vol. 60, no. 5, pp. 1122–1134, 2008. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [35].Reeder SB, Brittain JH, Grist TM, Yen YF, “Improving chemical shift encoded water-fat separation using object-based information of the magnetic field inhomogeneity,” Mag. Res. Med, vol. 73, no. 2, pp. 597–604 2015. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [36].Polak E and Ribierer G, “Note sur la convergence de méthodes de directions conjuguées,” ESAIM, vol. 3, no. R1, pp. 35–43, 1969. [Google Scholar]
  • [37].Kang N, Zhang J and Carlson ES, “Performance of ILU preconditioning techniques in simulating anisotropic diffusion in the human brain,” Future Gener. Comp. Sy, vol. 20, no. 4, pp. 687–698, 2004. [Google Scholar]
  • [38].Nocedal J and Wright S, “Efficient computation of regularized field map estimates in 3D,” Numerical optimization, 2006. [Google Scholar]
  • [39].Fessler JA and Booth SD, “Conjugate-gradient preconditioning methods for shift-variant PET image reconstruction,” IEEE Trans. Imag. Proc, vol. 8, no. 5, pp. 688–699, 1999. [DOI] [PubMed] [Google Scholar]
  • [40].Huber PJ, “Robust Statistics,” New York: Wiley, 1981. [Google Scholar]
  • [41].“Poblano Toolbox for MATLAB,” https://github.com/sandialabs/poblano_toolbox, [Online; accessed 16-Aug.-2020].
  • [42].“Michigan Image Reconstruction Toolbox (MIRT),” http://web.eecs.umich.edu/~fessler/code, [Online; accessed 20-Feb.-2020].
  • [43].Grivich MI and Jackson DP, “The magnetic field of current-carrying polygons: An application of vector field rotations,” American J. Phys, vol. 68, no. 5, pp. 469–474, 2000. [Google Scholar]
  • [44].Keator DB et al. , “The function biomedical informatics research network data repository,” Neuroimage, vol. 124, pp. 1074–1079, 2016. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [45].Uecker M, Lai P, Murphy MJ, Virtue P, Elad M, Pauly JM, Vasanawala SS, and Lustig M, “ESPIRiT an eigenvalue approach to autocalibrating parallel MRI: where SENSE meets GRAPPA,” Mag. Res. Med, vol. 71, pp. 990–1001, 2014. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [46].“Fat-Water Toolbox,” http://ismrm.org/workshops/FatWater12/data.htm, [Online; accessed 7-March-2020].
  • [47].Walsh DO, Gmitro AF and Marcellin MW, “Adaptive reconstruction of phased array MR imagery,” Mag. Res. Med, vol. 43, no. 5, pp. 682–690, 2000. [DOI] [PubMed] [Google Scholar]

RESOURCES