Abstract
There has been a growing interest in optical neural interfaces which is driven by the need for improvements in spatial precision, real-time monitoring, and reduced invasiveness. Here, we present unique microfabrication and packaging techniques to build implantable optoelectronics with high precision and spatial complexity. Material characterization of our hybrid polymers shows minimal in vitro degradation, greater flexibility, and lowest optical loss (4.04-4.4 dB/cm at 670 nm) among other polymers reported in prior studies. We use the developed methods to build Lawrence Livermore National Laboratory’s (LLNL’s) first ultra-compact, lightweight (0.38 g), scalable and minimally invasive thin-film optoelectronic neural implant that can be used for chronic studies of brain activities. The paper concludes by summarizing the progress to date and discussing future opportunities for flexible optoelectronic interfaces in next generation clinical applications.
Keywords: Neural interfaces, flexible optoelectronics, polymer waveguides, optical interconnects
Graphical Abstract
INTRODUCTION
The long-term stability of an implantable device depends on the careful selection of interface materials, geometry and scale1. Implantable neural interfaces have undergone quite an evolution in their design, starting from millimeter-scale electrodes2 in early 20th century to microfabricated probes3-6 and multi-modal optoelectrodes of today’s age7-9. Optical tools for neural manipulation and recording have recently allowed scientists to control, map and image brain circuits with cell-type specificity at high spatial and temporal precision7,10-12. However, bidirectional optical communication between an implanted device and a tissue remains an implementation challenge. Light scatters randomly in inhomogeneous media like biological tissues, and light-guiding materials often suffer eventual degradation in vivo. New materials and advanced engineering can help to mitigate these challenges and deliver stable optical interfaces for chronic use.
Light-guiding materials in an optical implant must have high transmissivity and transparency over the visible spectrum, this allows use of a wide range of wavelengths for cell stimulation or cell imaging. Traditional silicon-based semiconductor materials are optically clear and have well understood process techniques4,13,14 but may suffer from long-term biocompatibility issues due to their mechanical stiffness15. Carbon-based materials are flexible and biocompatible15,16 but relatively less understood in the context of optical propagation. Polymer materials such as polyimide17, Parylene-C18 and SU-819 are flexible and may promote cell attachment thereby enhancing signal acquisition. However, major drawbacks of polymers include lower optical transmittance and transparency, high moisture absorbability and reduced chemical and thermal stability which limits their application.
Recently, nano-hybridization of polymers with metallic or inorganic nanoparticles has been proposed as a promising technique to overcome optical polymers’ drawbacks without reducing their inherent advantages20. These organic-inorganic hybrid polymers obtained via a sol-gel process21 are lightweight and flexible and enable a wide range of molecular designs. Unlike semiconductor material growth and deposition approaches which are carried out at higher temperatures, hybrid polymers can be spun at room temperature and cured at lower temperatures to achieve a wide range of thicknesses. Furthermore, with the aid of computational models, molecular formations can be synthesized to tune optimal physical properties like elasticity and strength. Ormocers®22 (ORganically-MOdified-CERamics) are an example of hybrid polymers which constitute characteristics useful for fabrication of flexible structures in a simple and cost-effective manner. Their UV-induced polymerization yields a cured thermosetting composite with no glass transition temperature, high chemical and thermal stability, and high optical transparency23. Ormocers® have attracted considerable attention for wafer-scale design of optical elements such as waveguides22 and Bragg gratings24 and as biocompatible tissue scaffolds25. However, to the best of our knowledge, their use in biomedical optical implants has not yet been adopted.
Lawrence Livermore National Laboratory (LLNL) has developed a series of thin-film polymer implants for research and clinical applications26-29, including the first FDA-cleared polyimide-based retinal prosthesis in the world26. In this work, we present integration of LLNL’s thin-film polymer technology with hybrid polymers to enable Polymeric Opto-Electro-Mechanical Systems (POEMS). We characterize hybrid and non-hybrid polymers (Table 1) and then use the chosen materials and developed technologies to build LLNL’s first flexible optoelectrode that can deliver guided light for neural activation. The presented work demonstrates POEMS’s potential to allow selective sensing and actuation capabilities for biomedical applications.
Table 1:
Properties of hybrid and non-hybrid polymerized polymers.
Polymer | Refractive index at 600 nm, RI |
Glass transition temperature, Tg (0 C) |
Young’s modulus, E (GPa) |
---|---|---|---|
OrmoClearFX (Hybrid) (micro resist technology, Germany) | 1.555 | N/A | <1 |
OrmoComp (Hybrid) (micro resist technology, Germany) | 1.52 | N/A | ~1 |
EpoCore (Non-hybrid) (micro resist technology, Germany) | 1.58 | 145 | 2-3 |
IP-Dip (Non-hybrid) (Nanoscribe GmbH, Germany) | 1.548 | 103 | 2-5 |
FLEXIBLE OPTOELECTRONICS: MATERIALS AND DESIGN
Material characterization
The optical performance of a material depends on its transmission and scattering properties. If a film is non-absorbing, optically homogeneous and isotropic, the transmitted ray propagates without loss of intensity. However, realistically, the transmitted ray, T, loses intensity during travel as the result of scattering from inhomogeneities within the material. The distribution of the scattered light can be represented in terms of reflected, R, back-scattered, B, and forward-scattered, F, components of the total flux using equation 130:
(1) |
We measured T of 180 μm-thick polymerized films (Figure 1a) using a UV-Vis spectrophotometer (Perkin Elmer, Houston, Texas). R, a fraction of the incident flux, I, gets reflected from the surface at the point of incidence. Fresnel's equation for normal incidence was used to determine the reflected component (equation 230), where, Refractive Index (RI) of air and polymer are denoted by n1 and n2, respectively. For example, at an air - polymer interface, 4-6% of the incident flux is reflected for polymers with n2 ranging between 1.5 to 1.6 (Table 1).
Figure 1:
Material characterization of fully polymerized, 180 μm-thick, hybrid and non-hybrid polymer films. a-b) Optical transmissivity plots show that hybrid polymers allow >90% transmissivity in 300-800 nm wavelength range while non-hybrid polymers absorb light in near-UV region, c) Change in polymer transmissivity post accelerated aging soak tests conducted for a period of one month in 1X PBS at 75 °C, equivalent to 12-month time at 37 °C in vivo. All polymers withstood soak tests with 2.92% (OrmoClearFX), 8.01% (OrmoComp), 14.07% (EpoCore), and 9.78% (IP-Dip) average drop in transmissivity over 300-800 nm wavelength range, d) Change in transmissivity post photo-exposure with a dose of 150 J/cm2 from a 360 nm UV lamp at 17 mW/cm2 light intensity. OrmoClearFX, OrmoComp and IP-Dip show less than 0.18% decrease in transmissivity as compared to 17.62% decrease for EpoCore, particularly in the 400-550 nm region.
(2) |
F and B form a very small fraction for smooth, transparent films and their summation can be calculated numerically using equations 1 and 2.
The optical transmission and appearance of the polymers can vary across the visible spectrum (Figure 1 a-b) owning to their thickness, homogeneity, and difference in light scattering processes. Ormocers® demonstrate enhanced optical transmission, transparency, and flexibility (Figure 1, Table 1). Whereas, EpoCore and IP-Dip absorb light in the near-UV spectrum and appear yellowish when viewed under white light; IP-Dip is also less flexible at the measured thickness of 180 μm, with visible cracks along the edges (Figure 1b).
Accelerated lifetime soak tests31 (Figure 1c) and photodamage tests (Figure 1d) were conducted to determine polymers’ transmission characteristics and longevity in vitro. The accelerated lifetime soak provides an initial prediction of material degradation kinetics in vivo. UV photo-exposure experiments quantify possible film degradation during waveguide operation, acknowledging the fact that polymers are damaged most by UV radiation which causes them to crack or disintegrate. The findings in Figure 1 suggest that hybrid polymers offer appreciable chemical and photo resistance with evidence of biocompatibility32, and form promising materials for further study in preparation for long-term in vivo implants33. In comparison, both EpoCore and IP-Dip show lower optical transmissivity for lower visible wavelengths and have no record of biocompatibility in the literature.
Patterning technologies
An in-depth study of polymer microfabrication has been presented extensively in the literature18,26,34. In this work, we report microfabrication and additive manufacturing methods to pattern microscale polymer optical structures on different substrates while considering material properties and surface adhesion, which are critical for long-term device performance. We use microfabrication for wafer-scale design of continuous, millimeter-long planar waveguides and three-dimensional (3D) printing for design of micro-scale non-planar optical interconnects. When combined, these technologies offer solution to interconnect challenges at the Printed Circuit Board (PCB) and Multi-Chip-Module (MCM) level, potentially replacing typical electrical interconnects with high-performance, low-cost, compact polymer-based photonic alternatives. One such envisioned device schematic is shown in Figure 2a where constituent sub-assemblies use flexible optoelectronics at the off- and on-chip interconnect level.
Figure 2:
Modular POEMS for flexible optoelectronics. a) A conceptual schematic of a POEMS interface with integrated optoelectronic transmitter and receiver subassemblies. Individual modules can be independently fabricated using microfabrication and additive manufacturing techniques and then assembled using printable interconnects or optical wirebonds. SEM images show examples of independent planar and 3D structures, b-c) Microfabricated EpoCore and OrmoClearFX planar multimode waveguides on Cytop, d) 2PP-printed IP-Dip hollow waveguide on Cytop, e) 2PP-printed optical jumper on Cytop, f) 2PP-printed cylindrical vertical waveguide on top of a 670 nm VCSEL, g) 2PP-printed cylindrical horizontal interconnect between two flat-cleaved ends of an optical fiber, h) Process flow to fabricate structures shown in b) through g).
Planar microfabrication.
Microfabrication methods of resist spin-coating, UV lithography, and curing were used to pattern photo-definable OrmoClearFX and EpoCore planar waveguides on a polymer substrate called Cytop (AGC chemicals, Japan). OrmoClearFX is an Ormocer® whereas EpoCore is an SU8-like negative resist. Cytop is an amorphous fluoropolymer that was chosen as an optical cladding material because of its low RI (RI = 1.34), transparency, insulation properties, and very low moisture absorption compared to other polymers like polyimide and parylene. The fabricated waveguides had 10 μm-thick and 10 μm-wide core and 2.5 μm-thick Cytop cladding. A characteristic rounding of OrmoClearFX waveguide edges was observed due to oxygen inhibition35 (Figure 2b), a phenomenon whereby molecular oxygen at the surface inhibits free radical polymerization leaving uncured polymer on structure surfaces and interfaces. This can be advantageously utilized to round-off waveguide edges and corners to achieve diffused illumination profiles. On the other hand, EpoCore waveguides have straighter sidewalls and can be used for applications that demand rectangular waveguide cross-section (Figure 2c). Both OrmoClearFX and EpoCore waveguides exhibit high flexibility (Figure S1) and thus attract growing interest in microfabrication of flexible optoelectronics.
3D printing.
3D waveguides and optical interconnects were additively manufactured with two-photon resists36,37, OrmoComp and IP-Dip, using a direct laser lithography system (Nanoscribe GT, Nanoscribe GmbH, Germany). OrmoComp is an Ormocer® and IP-Dip is a RI-matched negative resist specifically developed for two-photon polymerization (2PP/TPP)25,37 at 780 nm laser wavelength of the Nanoscribe. The Nanoscribe focuses ultrafast laser pulses into a volume of a photosensitive two-photon resist to initiate 2PP and a developer solution removes non-illuminated unpolymerized regions, leaving behind a 3D structure. A parametric study was conducted to capture the printable regime for OrmoComp and IP-Dip on different substrate types (Figure S2). All structures were printed using a bottom-up printing approach and were optimized for adhesion and alignment while printing within material’s thermal threshold. In addition to the commonly reported damage and writing threshold36, we also characterized a 50%-dimensional tolerance threshold (±5 μm in X, Y, and Z axes) to further characterize the dimensional integrity of printed shapes as compared to their actual programmed dimensions. The results showed that IP-Dip’s damage threshold followed previously reported trends for opaque and reflective surfaces36,37. OrmoComp exhibited wider printing windows with no boiling damage within the 50 mW-laser power and 10 mm/s writing speed on most substrate types. Power intensity in OrmoComp was also found to be lower than in IP-Dip for the same average beam power due to the RI mismatch37 between immersion objective (1.52) and OrmoComp (1.51) at 780 nm.
These printing methods were implemented to design 3D interconnects on polymers (Figure 2d-2e) and on VCSELs (SM67-3N001M, Optowell Co. Ltd., South Korea) and optical fibers (FG050LGA, Thorlabs) (Figure 2f-2g). The optical performance of printed interconnects depends on material’s degree of crosslinking which is optimized by increasing laser power and decreasing laser speed during print and using UV exposure post-print. Since 2PP printing must be carried out under carefully controlled conditions near polymerization threshold, post-print UV cure is recommended to saturate crosslinking, thus reducing structural and functional variability, i.e. adhesion, modulus, and RI, across all printed structures38.
Optical loss measurements
The direct cut-back method9 was used to evaluate propagation loss per unit length of microfabricated and 2PP printed waveguides. The propagation losses in planar 10 μm x 10 μm waveguides measured as 4.9-7.8 dB/cm (OrmoClearFX) and 7.6-11.2 dB/cm (EpoCore) (Figure 3a). The high precision optical interconnects or “optical wirebonds” were shown to enhance optical coupling between on- and off-chip components. The propagation losses in printed 50 μm diameter OrmoComp waveguides measured between 4.08-4.4 dB/cm (Figure 3b) (Figure S3, S4). IP-Dip was excluded from optical waveguide characterization due to its low optical transmission and clarity (Figure 1), high Young’s modulus, and low glass transition temperature, making it unfit for design of flexible and optically clear waveguides.
Figure 3:
Measurement cut-back method for optical loss calculation in microfabricated and additively manufactured waveguides. The measured output power for each waveguide length is plotted in a) and b) where the observed slope and y-intercept of the linear-fit corresponds to transmission loss and coupling loss, respectively. a) Optical transmission loss in 10 μm x10 μm microfabricated OrmoClearFX and EpoCore waveguides measured 4.9-7.8 dB/cm and 7.6-11.2 dB/cm, respectively, at 406 nm and 638 nm wavelengths. Fibered bench-top laser source (LDC202C, Thorlabs) was used as the light source at 1 mW optical power, b) Optical transmission loss in 50 μm-diameter 2PP-printed waveguide on a 1 mW-optical power VCSEL measured 4.4 dB/cm at 670 nm wavelength. Zoomed-in view of the printed waveguide shows precise placement of waveguide at the coupling joint and smooth waveguide walls.
TOWARDS A POEMS-BASED IMPLANTABLE NEURAL INTERFACE
Adoption of emerging flexible photonic technologies can greatly benefit the implantable device field, which has been arguably hindered by an almost singular dependence on electromagnetics. As the electrode size is reduced to attain smaller implants, electrical recording suffers from reduced detectable signals with increase in noise, further worsening the signal-to-noise ratio. The addition of optics not only reduces device footprint and inter-wire crosstalk, but also offers light as a more sensitive modality to manipulate and sense biological signals10,12,39,40.
Utilizing the materials and technologies discussed above, we developed an all-polymer POEMS neural optoelectrode that can deliver guided light for neural activation. At just 0.38 g, the volume and weight of the device (Figure 4) are almost an order of magnitude smaller than other state-of-art approaches13,14,41,42. The fabrication process leverages LLNL’s thin-film polymer technology26,27 and additionally incorporates customizable waveguide fabrication and 635 nm Edge Emitting Laser (EEL) diode assembly onto a 15 μm-thick flexible neural probe43 (Movie S1). The waveguide core has a 10 μm-by-10 μm cross-section for optical stimulation. The 2.5 μm-thick Cytop cladding prevents optical scattering loss (Figure 4g), prevents moisture absorption, and offers electrical, thermal, and chemical stability beyond the standard polyimide. The platinum/iridium (50 nm/50 nm, 20μm diameter) electrodes are arranged around the waveguide tip for electrical recording. An indium/gold eutectic bond pad for the flip-chipped EEL diode enables flux-less bonding at low bonding temperatures and no applied pressure9,44. Flip-chipped connectors are assembled directly on thin-film probe to avoid use of bulky PCBs.
Figure 4:
POEMS implantable neural optoelectrode. a) Device schematic with design components, b) Fully assembled prototype, compared to a US dime in size, c) Optoelectrode’s flexible probe shank with waveguides and recording electrodes, d) Eutectic bond pad for EEL diode placement. Zoomed-in view shows eutectic bond pad positioned next to the distal waveguide tapered end to allow easy butt-coupling during assembly (right), e) Probe shank’s tip showing waveguide emission site and Pt/Ir recording electrodes. Magnified view of waveguide front end (right), f) EEL diode flip-chipped at the eutectic pad of the thin film optoelectrode. Inset shows active diode-waveguide interface at 635nm, g) Optical scattering along the waveguide length, without and with patterned cytop cladding around the core.
The electrodes measured average baseline peak-to-peak signal of 14.5 ± 2.4 μV (mean ± std. error) and average impedance of 407 ± 50.2 kΩ (mean ± std. error) at 1 kHz (n=15, 2 devices) during in vitro characterization in saline, recording good signal-to-noise ratio. The power measured at the output of the waveguides was 80-120 μW (800-1200 mW/mm2 intensity) which is sufficiently more than 5-10 mW/mm2 optical intensity required for the target applications39. A heat transfer model was designed to study thermal behavior of the system during device operation43(Figure S5). Implantable devices should be able to deliver light deep into the brain while not causing thermal damage to the tissue or to the device sub-components. Our model results showed that POEMS optoelectrodes (thermal conductivity ~0.2 W/m.K) allow for a much slower temperature increase in the tissue compared to silicon optoelectrodes (thermal conductivity ~130 W/m.K) while operating within temperature limits of the diode. The response was found similar to the previously demonstrated GRIN lens-based assembly solution9, but with a significant reduction in device footprint, weight, and cost. This improvement is promising and will critically impact scaling up the number of shanks and diodes per shank to enable multi-color, multi-site optogenetic studies. . Future assembly modifications will focus on further reducing optical coupling losses by 3D printing the last-micron connection between assembled EEL chips and fabricated waveguides, boosting power efficiency.
DISCUSSION AND OUTLOOK
The success of devices such as cochlear and retinal implants has encouraged interest in new applications of flexible interfaces, paving the way for incorporation of novel materials and new modalities such as optical. Thermal loading via pulsed, low-power infrared light alongside electrical stimulation has been shown to reduce electrical current thresholds in cochlear implants45. Photomanipulation techniques, such as introduction of light-sensitive channels to retinal ganglion cells, may provide a functional cure for people with retinitis pigmentosa46. Waveguides with integrated photodetectors can allow simultaneous perturbation and imaging of neural activity40.
Most existing optical light delivery solutions in neurotechnology are realized using optical fibers7,10,39,47,48 which may restrict animal movement. More recently, light-emitting diodes (LEDs)8,49 and laser diodes (LDs)13,41 were integrated on silicon neural probes to provide fiber-free solutions with reduced electromagnetic interference13,14 and safer thermal design13. The search for mechanically compliable materials than silicon and glass has fueled the ongoing development of softer and flexible polymer-based interfaces. State-of-art polymer neural optoelectrodes either offer partially-flexible solutions41,50, use low optical transmission materials like SU-842,50 or are susceptible to artifacts and local heating49. In this work, we report implementation of EEL-coupled waveguides on an all-polymer neural device which is fiber-free, flexible, compact, and lightweight. Use of optically clear materials like hybrid polymers and Cytop makes the devices suitable for delivering or collecting light from the surrounding and underlying tissue while maintaining good power efficiency. An all-polymer backbone can potentially facilitate better mitigation of electromagnetic, photochemical, or photoelectrical artifacts. High optical efficiency of hybrid polymers makes them particularly well-suited for micro-optics. The attenuation in a waveguide increases with a decrease in core diameter51 which makes micro-waveguides more susceptible to optical losses. The measured transmission losses for 10 μm-core microfabricated (4.9-7.8 dB/cm at 405-635 nm) and 50 μm-core 3D-printed (4.08-4.4 dB/cm at 670 nm) hybrid polymer waveguides are amongst the lowest reported for polymer waveguides in literature. Previous work on SU-8 waveguides reported 6.4 dB/cm optical loss at 473 nm for 150 μm waveguide core42 and 25 dB/cm optical loss at 473 nm for 25 μm waveguide core50. More recent waveguide designs made from 80 μm core polycarbonate- cyclic olefin copolymer (PC-COC) core47 and 15 μm parylene core48 report 1.5 dB/cm loss at 473 nm and 5.5 dB/cm loss at 532 nm, respectively.
In summary, hybrid polymers-based POEMS show potential for high resolution and minimally invasive optical waveguiding and imaging. Further research and development can eventually enable their use in biomedical technologies such as optoelectronic brain-machine interfaces and wearables.
Supplementary Material
Novel hybrid polymers offer material advantages of both organic and inorganic materials.
Hybrid polymer-based waveguides enable efficient microscale optical systems.
Microfabricated waveguides allow minimally invasive optical waveguiding and imaging.
3D optical wirebonds can enhance optical coupling between on- and off-chip components.
Optoelectronic implants fabricated from hybrid polymers are compact, low-weight and low-loss.
ACKNOWLEDGMENTS
This work was performed under the auspices of the U.S. Department of Energy by Lawrence Livermore National Laboratory under Contract DE-AC52-07NA27344 (IM Release # LLNL-JRNL-810279). The work was supported by National Institute of Health’s National Institute on Deafness and Communication Disorders under Project NIDCD A-DC-15001-001, and Lawrence Livermore National Laboratory’s Directed Research and Development under Project 08-ERD-044. We thank Jennifer Alameda, Sarah Sahota, Cathy Reinhardt, and Jeanine Pebbles at LLNL for their assistance in task execution. We also acknowledge the support of Stanford Nanofabrication Facility for facility access for device assembly.
Footnotes
Publisher's Disclaimer: This is a PDF file of an unedited manuscript that has been accepted for publication. As a service to our customers we are providing this early version of the manuscript. The manuscript will undergo copyediting, typesetting, and review of the resulting proof before it is published in its final form. Please note that during the production process errors may be discovered which could affect the content, and all legal disclaimers that apply to the journal pertain.
REFERENCES
- 1.Wellman SM et al. A Materials Roadmap to Functional Neural Interface Design. Adv. Funct. Mater 28, 1–38 (2018). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2.Hodgkin A & Huxley A Action Potentials Recorded from Inside a Nerve Fibre. Nature 710–711 (1939). [Google Scholar]
- 3.Najafi K, Wise KD & Mochizuki T A high-yield IC-compatible multichannel Recording Array. Electron Devices IEEE Trans. 32, 1206–1211 (1985). [Google Scholar]
- 4.Nordhausen CT, Maynard EM & Normann RA Single unit recording capabilities of a 100 microelectrode array. Brain Res. 726, 129–140 (1996). [PubMed] [Google Scholar]
- 5.Seymour JP & Kipke DR Neural probe design for reduced tissue encapsulation in CNS. Biomaterials 28, 3594–3607 (2007). [DOI] [PubMed] [Google Scholar]
- 6.Jun JJ et al. Fully Integrated Silicon Probes for High-Density Recording of Neural Activity. Nature 551, 232–236 (2018). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 7.Stark E, Koos T & Buzsáki G Diode probes for spatiotemporal optical control of multiple neurons in freely moving animals. J. Neurophysiol. 108, 349–63 (2012). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8.Wu F et al. Monolithically Integrated μLEDs on Silicon Neural Probes for High-Resolution Optogenetic Studies in Behaving Animals. Neuron 88, 1136–1148 (2015). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9.Kampasi K et al. Fiberless multicolor neural optoelectrode for in vivo circuit analysis. Sci Rep, Nat. Publ. Gr 6, 30961 (2016). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10.Airan RD, Thompson KR, Fenno LE, Bernstein H & Deisseroth K Temporally precise in vivo control of intracellular signalling. Nature 458, 1025–1029 (2009). [DOI] [PubMed] [Google Scholar]
- 11.Gao R et al. Cortical column and whole-brain imaging with molecular contrast and nanoscale resolution. Science (80-. ). 363, eaau8302 (2019). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12.Eles JR, Stieger KC & Kozai TDY The temporal pattern of Intracortical Microstimulation pulses elicits distinct temporal and spatial recruitment of cortical neuropil and neurons. J. Neural Eng (2020). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13.Kampasi K et al. Dual color optogenetic control of neural populations using low-noise, multishank optoelectrodes. Microsystems Nanoeng. 4, (2018). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14.Kim K et al. Artifact-free, high-temporal-resolution in vivo opto-electrophysiology with microLED optoelectrodes. Nat. Commun 1–32 (2020). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 15.Patel PR et al. Chronic in vivo stability assessment of carbon fiber microelectrode arrays. J. Neural Eng 13, 066002 (2016). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16.Kostarelos K, Vincent M, Hebert C & Garrido J Graphene in the Design and Engineering of Next-Generation Neural Interfaces. Adv. Mater 29, (2017). [DOI] [PubMed] [Google Scholar]
- 17.de la Oliva N, Navarro X & del Valle J. Time course study of long-term biocompatibility and foreign body reaction to intraneural polyimide-based implants. J. Biomed. Mater. Res 106, (2017). [DOI] [PubMed] [Google Scholar]
- 18.Oliva N, Mueller M, Stieglitz T, Navarro X & Del Valle J On the use of Parylene C polymer as substrate for peripheral nerve electrodes. Sci. Rep (2018). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 19.Márton G et al. The neural tissue around SU-8 implants: A quantitative in vivo biocompatibility study. Mater. Sci. Eng. C 112, (2020). [DOI] [PubMed] [Google Scholar]
- 20.Muramatsu A & Miyashita T Nanohybridization of Organic-Inorganic Materials. (2009).
- 21.Ogoshi T & Chujo Y Organic–inorganic polymer hybrids prepared by the sol-gel method. Compos. Interfaces 11, (2005). [Google Scholar]
- 22.Buestrich R et al. ORMOCER®s for Optical Interconnection Technology. J. Sol-Gel Sci. Technol 20, (2001). [Google Scholar]
- 23.Haas KH & Rose K Hybrid inorganic/organic polymers with nanoscale building blocks: Precursors, processing, properties and applications. Rev. Adv. Mater. Sci 5, 47–52 (2003). [Google Scholar]
- 24.Girschikofsky M et al. Waveguide bragg gratings in ormocer®s for temperature sensing. Sensors (Switzerland) 17, 1–9 (2017). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 25.Song J, Michas C, Chen CS, White AE & Grinstaff MW From Simple to Architecturally Complex Hydrogel Scaffolds for Cell and Tissue Engineering Applications: Opportunities Presented by Two-Photon Polymerization. Adv. Healthc. Mater 9, 1–13 (2020). [DOI] [PubMed] [Google Scholar]
- 26.Pannu S et al. Artificial Retina Device Containing 240 Densely-packed Electrodes Packaged Into A Chronic, Wireless System. Invest. Ophthalmol. Vis. Sci 52, (2011). [Google Scholar]
- 27.Tooker A et al. Towards a large-scale recording system: Demonstration of polymer-based penetrating array for chronic neural recording, in 36th Annual International Conference of the IEEE Engineering in Medicine and Biology Society 6830–6833. (2014). [DOI] [PubMed] [Google Scholar]
- 28.Ivanovskaya AN et al. Electrochemical Roughening of Thin-Film Platinum for Neural Probe Arrays and Biosensing Applications. J. Electrochem. Soc 165, G3125–G3132 (2018). [Google Scholar]
- 29.Chung JE et al. High-Density, Long-Lasting, and Multi-region Electrophysiological Recordings Using Polymer Electrode Arrays. Neuron 101, 21–31.e5 (2019). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 30.Hall C Polymer Materials: An Introduction for Technologists and Scientists. (2017).
- 31.Hukins D, Mahomed A & Kukureka S Accelerated aging for testing polymeric biomaterials and medical devices. Med. Eng. Phys 30, 270–1274 (2008). [DOI] [PubMed] [Google Scholar]
- 32.Ovsianikov A, Schlie S, Ngezahayo A, Haverich A & Chichkov BN Two-photon polymerization technique for microfabrication of CAD-designed 3D scaffolds from commercially available photosensitive materials. J. Tissue Eng. Regen. Med 1, 443–449 (2007). [DOI] [PubMed] [Google Scholar]
- 33.Caldwell R et al. Characterization of Parylene-C degradation mechanisms: In vitro reactive accelerated aging model compared to multiyear in vivo implantation. Biomaterials 232, 119731 (2020). [DOI] [PubMed] [Google Scholar]
- 34.Hassler C, Boretius T & Stieglitz T Polymers for neural implants. J. Polym. Sci. Part B Polym. Phys 49, 18–33 (2011). [Google Scholar]
- 35.Chong J . Oxygen consumption during induction period of a photopolymerizing system. J. Appl. Polym. Sci 13, 241–247 (1969). [Google Scholar]
- 36.Rekštytė S, Jonavičius T & Malinauskas M Direct laser writing of microstructures on optically opaque and reflective surfaces. Opt. Lasers Eng 53, 90–97 (2014). [Google Scholar]
- 37.Saha SK, Divin C, Cuadra JA & Panas RM Effect of proximity of features on the damage threshold during submicron additive manufacturing via two-photon polymerization. J. Micro Nano-Manufacturing 5, 1–10 (2017). [Google Scholar]
- 38.Oakdale JS, Ye J, Smith WL & Biener J Post-print UV curing method for improving the mechanical properties of prototypes derived from two-photon lithography. Opt. Express 24, 27077 (2016). [DOI] [PubMed] [Google Scholar]
- 39.Mattis J et al. Principles for applying optogenetic tools derived from direct comparative analysis of microbial opsins. Nat. Methods 9, 159–172 (2011). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 40.Rickgauer J, Deisseroth K & Tank D Simultaneous cellular-resolution optical perturbation and imaging of place cell firing fields. Nat. Neurosci 17, 1816–1824 (2014). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 41.Schwaerzle M, Ringwald P, Paul O & Ruther P First dual-color optrode with bare laser diode chips diecrtly butt-coupeld to hybrid-polymer waveguides. in MEMS vol. 2 26–29 (2017). [Google Scholar]
- 42.Rubehn B, Wolff SBE, Tovote P, Löthi A & Stieglitz T A polymer-based neural microimplant for optogenetic applications: design and first in vivo study. Lab Chip 13, 579–88 (2013). [DOI] [PubMed] [Google Scholar]
- 43.Kampasi K et al. Design and microfabrication strategies for thin-film, flexible optical neural implant. in 42nd Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) (2020). [DOI] [PubMed] [Google Scholar]
- 44.Kampasi K et al. Efficient assembly of multi-color fiberless optoelectrodes with on-board light sources for neural stimulation and recording. in 38th Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) 4479–4482 (2016). [DOI] [PubMed] [Google Scholar]
- 45.Tan X et al. Auditory Neural Activity in Congenitally Deaf Mice Induced by Infrared Neural Stimulation. Sci. Rep 8, 1–14 (2018). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 46.Busskamp V, Picaud S, Sahel JA & Roska B Optogenetic therapy for retinitis pigmentosa. Gene Ther. 19, 169–175 (2012). [DOI] [PubMed] [Google Scholar]
- 47.Park S et al. One-step optogenetics with multifunctional flexible polymer fibers. Nat. Neurosci 20, 612–619 (2017). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 48.Reddy J & Chamanzar M Low-loss flexible Parylene photonic waveguides for optical implants. Opt. Lett 43, 4112–4115 (2018). [DOI] [PubMed] [Google Scholar]
- 49.Reddy JW et al. High density, double-sided, flexible optoelectronic neural probes with embedded μLEDs. Front. Genet 10, 1–15 (2019). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 50.Shin H et al. Multifunctional multi-shank neural probe for investigating and modulating long-range neural circuits in vivo. Nat. Commun 10, (2019). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 51.Ioannides N et al. Approaches to mitigate polymer-core loss in plastic optical fibers: A review. Mater. Res. Express 1, (2014). [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.