Right ventricular failure after left ventricular assist device implantation: a review of the literature

Valeria Lo Coco 1,, Maria Elena De Piero 1,2, Giulio Massimi 1,3, Giovanni Chiarini 1,4, Giuseppe M Raffa 5, Mariusz Kowalewski 1,6, Jos Maessen 1,7, Roberto Lorusso 1,7
PMCID: PMC7947472  PMID: 33717597

Abstract

Right ventricular failure (RVF) following left ventricular assist device (LVAD) implantation remains a major complication which may significantly impair patient outcome. The genesis of RVF is, however, multifactorial, and the mechanisms underlying such a condition have not been fully elucidated, making its prevention challenging and the course not always predictable. Although preoperative risks factors can be associated with RV impairment, the physiologic changes after the LV support, can still hamper the function of the RV. Current medical treatment options are limited and sometimes, patients with a severe post-LVAD RVF may be unresponsive to pharmacological therapy and require more aggressive treatment, such as temporary RV support. We retrieved 11 publications which we assessed and divided in groups based on the RV support [extracorporeal membrane oxygenation (ECMO), right ventricular assist device (RVAD), TandemHeart with ProtekDuo cannula]. The current review comprehensively summarizes the main studies of the literature with particular attention to the RV physiology and its changes after the LVAD implantation, the predictors and prognostic score as well as the different modalities of temporary mechanical cardio-circulatory support, and its effects on patient prognosis for RVF in such a setting. In addition, it provides a decision making of the pre-, intra and post-operative management in high- and moderate- risk patients.

Keywords: Right ventricular failure (RVF), left ventricular assist device (LVAD), right ventricular assist device (RVAD), mechanical cardio-circulatory support, extracorporeal membrane oxygenation (ECMO)

Introduction

Right ventricular failure (RVF) following left ventricular assist device (LVAD) placement is a complex disease including many causes and pathological events (1). A right-sided failure may complicate 10% to 40% of LVAD implantations and is associated with prolonged length of intensive care unit and hospital stay, as well as high morbidity and mortality (2-4).

In case of refractory RVF, temporary mechanical cardio-circulatory supports may be required. New percutaneous devices, specifically designed for the RV, have been introduced into the clinical practice (5,6). Although technology developments, the perfect timing of implant, the more appropriate configuration and device, as well as the identification of high-risk patients is not well clarified representing still a challenge. We present the following article in accordance with the PRISMA reporting checklist (available at http://dx.doi.org/10.21037/jtd-20-2228).

Material and methods

We systematically searched the main articles in the current literature through PubMed and MEDLINE database. Search terms included right ventricular failure-left ventricular assist device-right ventricular assist device-mechanical cardio-circulatory support-extracorporeal membrane oxygenation. Complete details, including electronic search strategy, objectives, criteria for study selection, eligibility, data collection, and assessment of study quality, were registered and published online in PROSPERO-International Prospective Register of Systematic Reviews (Registration Number: CRD42019134835).

The search was limited to human studies; no language restrictions were imposed. We made a screening of titles and abstracts identifying those records which considered patients with heart failure, treated with LVAD, and in which a proportion of the whole recruited population developed acute RVF (i.e., within 2 weeks of LVAD implantation). Furthermore, we verified that relevant data were available focusing on their quality and relevance assessment. We included studies with the following criteria: data were stratified for both RVF and No-RVF patients; data were expressed as n (%) or mean ± standard deviation and included clear denominators. Additional references were identified analyzing the bibliographies of the evaluated articles. The exclusion criteria were as follows: patients treated with medical therapy or intra-aortic balloon pump (IABP), biventricular assist device (BiVAD), neonatal/pediatric populations (i.e., 17 years old or younger) and case reports.

We retrieved 21 publications; after removal of non-pertinent reports, case reports, and studies in which the demography of the patients, the strategy and their outcome were not detailed, 12 papers remained. These studies were assessed and divided in four categories based on the strategy of right ventricle (RV) support:

  • ❖ Isolated LVAD vs. LVAD-extracorporeal membrane oxygenation (ECMO);

  • ❖ Isolated LVAD vs. LVAD-right ventricular assist device (RVAD);

  • ❖ Isolated LVAD vs. LVAD-ECMO vs. LVAD-RVAD;

  • ❖ LVAD-Tandemlife ProtekDuo.

The PRISMA flow diagram, describing the decision-making process about the management of the RVF after LVAD, is presented in Figure 1.

Figure 1.

Figure 1

Prisma flow diagram.

Changes in right ventricular function after LVAD placement

The genesis of the RVF is multifactorial and pathological events can increase RV afterload, decrease RV preload, or impair the contractility (7). Indeed, preoperative existing RV dysfunction may be aggravated by further physiologic changes related to LVAD implant (8,9). The increasing LV output could precipitate RVF through two mechanisms: volume overloading of the RV owing to an increased flow in the systemic circulation or a decrease in RV pump function caused by an interventricular septal (IVS) shift into the LV due to an exaggerated drainage (10). Thus, emptying the LV by exaggerated LVAD pump flow could interfere with RV function by compromising the contribution of the IVS to RV contraction due to the two ventricles interdependence (10). On the other hand, the reduction in LV congestion obtained by appropriate draining, may relieve the RV by reducing RV afterload, increasing the aortic and thus coronary blood flow, reversing a septal shift into the RV from a distended LV (11). The acute unloading of the LV, the increase of right-sided preload and the cardiac output leads to a septal shift thereby alter RV shape and size with modifications in the RV contractility. In addition, in the setting of an incompetent tricuspid valve, the increased RV volume and the left-ward shift of the septum may cause a tethering of valve leaflets, further deteriorating or generating hemodynamically critical tricuspid regurgitation (11,12). Among other reasons, an increased RV afterload due to pulmonary vasoconstriction and tachyarrhythmia which can double the risk of RVF have been also proposed (13,14).

Although the above-mentioned reasons are the most common explanations, the full understanding of the pathophysiology of LVAD-related RVF remains still not well understood.

Prognostic score and risk factors for RVF

In the last years, a growing interest has been observed about realizing risk models to predict RVF occurrence post-LVAD. Identifying patients at high risk for such a complication has been associated with improved management and outcomes (15,16) (Figure 2). Recent studies have highlighted a few predictors for such an adverse event: female gender, preoperative circulatory support, prior cardiac surgery and non-ischemic etiology of heart failure have all been found prognostic of RVF in this setting (17-19). The HeartMate II risk model reported CVP/pulmonary capillary wedge pressure >0.63, pre-operative mechanical ventilator support and blood urea nitrogen >39 mg/dL as independent risk factors (20). Other important predictors for RVF are the pre-LVAD implant grade of tricuspid incompetence, the geometry of the RV (right ventricle end diastolic diameter >35 mm, RV ejection fraction <30% and right atrial dimension >50 mm) in relation to pulmonary vascular resistance, elevated CVP, and evidence of liver and renal impairments (21).

Figure 2.

Figure 2

Definition of moderate and high-risk patients for right ventricular failure after left ventricular assist device and their management in the pre-, intra- and post-operative. AST, aspartate aminotransferase; BiVAD, biventricular assist device; BUN, blood urea nitrogen; CI, cardiac index; CPB, cardio-pulmonary bypass; CVP, central venous pressure; DPG, diastolic pulmonary gradient; FAC, fractional area change; LVEDD, left ventricle end-diastolic diameter; MR, mitral regurgitation; PA, pulmonary artery; PVR, peripheral vascular resistance; RA, right atrium; RV, right ventricle; RVEDD, RV end-diastolic diameter; RV EF, RV ejection fraction; RVESV, RV end-systolic volume; RVF, RV failure; RVSWI, RV stroke work index; TAPSE, tricuspid annular plane systolic excursion; TR, tricuspid regurgitation.

After LVAD implant, a moderate-severe mitral regurgitation might also be predictive of the onset or deterioration of RVF. Regarding intervention on mitral valve regurgitation as protective intervention to prevent RVF occurrence, there is no agreement about the appropriate management (11). Indeed, the disappearance or improvement of mitral valve insufficiency, after the LVAD-related LV unloading, has been reported (22). In contrast, other studies have advocated the need of mitral repair as required and protective factor for early or late RVF occurrence (23,24). This aspect, therefore, remains still controversial. Regarding tricuspid valve anomalies (annulus >40 mm, moderate or severe regurgitation), tricuspid valve repair concomitant LVAD implant has also been advocated (25,26).

The risk for post-LVAD RVF has been associated at other parameters. In the study by Atluri and colleagues (27), a preoperative high CVP was associated with a higher risk of RFV. Furthermore, Fukamachi et al. (28), demonstrated low RV stroke-work index and low pulmonary artery (PA) pressure to be significant predictors of RVF. Other studies identified low cardiac index, diastolic pulmonary gradient, elevated pulmonary vascular resistance as important risk factors (29-31).

As far as risk models for RVF are concerned, the Michigan RVF risk score and the CRITT score have been shown to be helpful about predicting the probability of developing LVAD-related RVF (27,32). The Michigan score considers four variables: vasopressor use (assigning 4 points), creatinine >2.3 mg/dL (3 points), bilirubin >2 mg/dL (2.5 points) and aspartate aminotransferase >80 IU/dL (2 points); a greater risk for RVF is associated with a higher score (32). Instead, the CRITT score is a risk score based on 5 clinical variables and 0 or 1 point is allotted for each of them, including CVP >15 mmHg, severe RV dysfunction, preoperative mechanical ventilation/intubation, severe tricuspid regurgitation, and tachycardia. A CRITT score less than 2 is highly predictive of the ability to tolerate isolated LVAD therapy (negative predictive value of 93%) (26). Recently, Loforte and collaborators developed the ALMA score model (33). The parameters considered for the ALMA risk score include destination therapy, PA pulsatility index <2, right to left ventricular end-diastolic diameter ratio >0.75, RV stroke work index <300 mmHg/mL/m2, model for end-stage liver disease excluding international normalized ratio (MELD-XI) >17. In this risk model, an isolated LVAD is recommended for score of 0 or 1, a BiVAD for patients with score of 4 or 5, whereas a score of 2 represents a grey zone and they can benefit from an LVAD with an appropriate pharmacologic and/or a temporary RVAD (33). Besides these risk factors, unfortunately, the pre-operative risk assessment cannot account for intra-operative events such as air embolism, blood transfusion etc., which may influence the hemodynamic status and consequently, the RV function.

Temporary right ventricular support

Despite the optimal medical treatment, 4–6% of the patients with RVF after LVAD implantation are unresponsive to pharmacological therapy and requires a temporary ventricular support (2,7). Several strategies have been described to support the RVF after LVAD implant. The RVF-device options include veno-arterial (V-A) ECMO, temporary RVAD (with/without oxygenator) with single lumen (Biomedicus or TandemLife) or double-lumen cannula (ProtekDuo® TandemLife), percutaneous devices (Impella RP® Abiomed), and, TandemHeart (TH-RVAD) (Figure 3) (Table 1).

Figure 3.

Figure 3

Temporary and long-term mechanical assist device for right ventricular failure after left ventricular assist device. LVAD, left ventricular assist device; OxyRVAD, right ventricular assist device with oxygenator; TH RVAD, TandemHeart RVAD; V-A ECMO, veno-arterial extracorporeal membrane oxygenation.

Table 1. Temporary mechanical cardio-circulatory support for right ventricular failure: advantages, disadvantages.

Temporary right ventricular support Features Effects Advantages Disadvantages
V-A ECMO Extracorporeal pump with membrane oxygenator; peripheral femo-femoral cannulation with inflow cannula in the RA and the outflow in femoral artery < RA pressure; < MAP; provides systemic oxygenation; indirect unloading RV; > LV afterload (non-physiological); peripheral oxygenation (non-physiological) Support for biventricular failure; hybrid and dynamic configuration Infection, leg ischemia, hemolysis, vascular and bleeding complications; differential hypoxia for impaired lung function; flow competition with LVAD (thrombosis LVAD); no specific support for RVF
RVAD with single-lumen cannula for PA cannulation Extracorporeal pump with or without membrane oxygenator; inflow cannula in the RA through the femoral vein and outflow in PA through singe-lumen cannula inserted percutaneously in IJV Direct unloading RV; physiological support Easy insertion; isolated RV support; possibility to add an oxygenator in the circuit leading to a central oxygenation (physiological)-OxyRVAD Right heart chambers or PA perforation; pulmonary insufficiency; arrhythmias; double-site cannulation (IJV and femoral vein)
RVAD with double-lumen cannula for PA cannulation Extracorporeal pump with or without membrane oxygenator; percutaneously, through IJV, inflow in the RA and outflow in the PA through the multi-fenestrated tip Direct unloading RV; physiological support Easy insertion; isolated RV support; single site cannulation and double effect (drainage and perfusion); possibility to add an oxygenator in the circuit leading to a central oxygenation (physiological)-OxyRVAD; dynamic support; mobilization patients Right heart chambers or PA perforation; pulmonary insufficiency; arrhythmias
Impella RP Intra-corporeal micro-axial flow pump; inserted in femoral vein with the inflow in the IVC and the outflow in the PA < CVP; > CI; > weaning of inotrope and vasopressor support; unloading RV; < RA pressure; < systemic venous congestion; > end-organ perfusion Easy insertion; single venous access site; small dimension of the machine support Hemolysis, vascular and bleeding complications; rare: worsened tricuspid or pulmonary valve dysfunction; no mobilization (easy dislocation of the device)
TandemHeart RVAD Extracorporeal centrifugal flow pump; inflow cannula in the RA through the left femoral vein Outflow cannula in the main PA through the right femoral vein or the right IJV > MAP; > CI; > RV stroke work; < RA pressure; < PA systolic pressure Single groin access (dual access via femoral vein), ideally suited to the needs of ambulatory LVAD patients; rapidly deployed Hemolysis, vascular and bleeding complications

CI, cardiac index; CVP, central venous pressure; IJV, internal jugular vein; IVC, inferior vena cava; LV, left ventricle; LVAD, left ventricular assist device; MAP, mean arterial pressure; PA, pulmonary artery; RA, right atrium; RV, right ventricle; RVAD, right ventricular assist device; RVF, right ventricular failure; V-A ECMO, veno-arterial extracorporeal membrane oxygenation.

One conventional approach includes peripheral V-A ECMO. It reduces LVAD preload due to RV unloading and increases the LVAD afterload with risk of reduced forward blood flow. With peripheral approach, the removal of the ECMO-related cannulas does not require a re-sternotomy and it costs less than any other mechanical support device. Common complications are bleeding, infections, and vascular complications (34). Regarding the reduced pre-load and increased afterload of the LVAD in the presence of peripheral V-A ECMO, such conditions translate into decreasing pump preload and increased pump afterload which might negatively influence the LVAD function. Therefore, in such a condition, it is important to accommodate the ECMO /LVAD interaction to avoid the competition between the two devices and the differential hypoxia phenomenon (35-37).

Another possible option implies the cannulation of the PA which may be approached with a surgical access in case of intraoperative, open-chest implant, or percutaneously via the right internal jugular vein (38). The surgical approach may be performed through a direct cannulation or via a prosthetic graft anastomosed to the main PA (“Chimney technique”). The direct PA cannulation requires open chest cannulation of the PA with cannulation of the RA with direct access or via the femoral vein, and it ensures complete unloading of the RV with an anterograde trans-pulmonary blood flow and an adequate preload for the LVAD. The associated limitations are the general anesthesia, mechanical ventilation and the necessity of re-sternotomy to withdraw the PA cannula, and, in case, also the RA cannula, which can increase bleeding and infection complications (39). Instead, the end-graft facilitated technique represents a simplified ECMO in which the venous inflow cannula is placed percutaneously into the RA through the femoral vein and the outflow cannula, through an 8 mm Dacron graft is anastomosed end-to-side to the main PA using a side clamp. The Dacron prosthesis is in front of the LVAD outflow graft and is tunneled to its percutaneous exit side at the subxiphoid level. The outflow cannula is then inserted into the Dacron graft and secured firmly to the chest wall using multiple sutures. Thus, it allows a bedside removal without a re-sternotomy. In literature, no vascular graft related problems (after months the graft appeared completely thrombosed without pseudo aneurysm, peri-graft fluid collections, pulmonary embolism or bleeding) were observed. It is associated with good LVAD pre- and post-load and with less thromboembolic events than the peripheral V-A ECMO (40,41).

Nowadays, new tools which can allow entirely percutaneous procedures have been developed. Indeed, the drainage cannula can be introduced from the femoral vein into the RA and a flexible outflow cannula can be place into the PA via the right internal jugular vein (42). An alternative fully percutaneous approach may be achieved with a single, double-lumen cannula, the PA is named ProtekDuo (TandemLife, Livanova, UK). This cannula is inserted in the right internal jugular vein to reach the main PA under fluoroscopy guidance, and allows 4 L/min flow (43); one lumen works as inflow draining the blood from the RA, while the second lumen delivers the blood into the main PA thanks to a multi-fenestrated distal tip. Such a configuration includes an oxygenator in the circuit (realizing a so called “OxyRVAD” configuration). When there is no impairment of the gas exchange, the oxygenator can be removed/avoided and a pure RVAD is obtained, enabling a prolonged support without anticoagulation, and avoiding the oxygenator-related complications (plasma leakage, systemic inflammation, coagulopathy and thrombo-embolism) (44-46). In addition, the groin-free strategy of support may allow patient awakening and mobilization, reducing thereby possible respiratory or vascular complications (44-46).

Another percutaneous option is the Impella RP (Abiomed, Denver) which is characterized by a micro-axial flow catheter with a flow of 4-5L/min. The catheter is placed into the right femoral vein and it ejects blood from the inferior vena cava into the PA, thereby bypassing the RV (47,48). This device may, therefore, represents an additional fully percutaneous approach, but consistent series and evidences are still not available (49).

The TH RVAD constitutes an additional percutaneous option for RVF. It delivers the blood through 2 venous cannulas. The inflow cannula is placed in RA via the left femoral vein, instead, the outflow cannula in PA through the right femoral vein. Another site of cannulation for the outflow is the right internal jugular vein, it can be used in case the femoral venous access is prohibitive: thrombosis, infection or inferior cava filters or in case of tall patients (distance from the femoral vein to the fifth intercostal space exceeds 58 cm) (50).

Another possible strategy, mostly in case of patient with an LVAD bridge-to-transplantation, is represented by a right VAD. In this respect, there are only small series and case reports available. The weaning rate from the right VAD was possible in the 45–75% of the patients (51). The major complication was represented bleeding and thrombosis of the VAD (52). They include the LVAD (Jarvik and HeartWare devices) (53) used in right position and the CentriMag paracorporeal VAD (54,55). For several reasons, such an approach is currently less considered for the related cost and invasiveness.

Discussion

Acute RVF is still a common complication after LVAD implantation associated with a reduced efficacy of LVAD in terms of flow and organ perfusion and a 6-month mortality reaching up to 29%. Its pathophysiology and management are complex and remain controversial (56-58). Despite optimized medical therapy, a significant number of LVAD patients still need a perioperative mechanical RV support. The outcome is different depending on the type of right assist device and the patient characteristics (Table 2). Kormos and collaborators (20), showed that patients receiving the HeartMate II with early RVF had a 20% reduction in 1-year survival compared with patients who did not have RVF. In the EUROMACS Registry, Soliman and colleagues described 2988 patients who had a LVAD, and the cumulative survival in the postoperative 24 months was higher in patients without RVF at the 6-month (79% versus 61%), 12-month (71% versus 53%), 18-month (65% versus 49%), and 24-month (58% versus 45%) follow-up, compared with patients with RVF (66).

Table 2. Comparison outcome of different right mechanical supports for right ventricular failure after left ventricular assist device.

Study Patients features Outcome
Isolated LVAD vs. LVAD-ECMO
   Shehab et al. (59) 75 isolated LVAD—mean age: 53; male: 60 (80%); etiology: DCMP 53 (71%), ischemic 15 (20%), other 7 (9%). 23 LVAD-V-A ECMO (femoral vein and PA through “Chimney technique”) —mean age: 51; male: 19 (83%); etiology: DCMP 12 (52%), ischemic 9 (39%), other 2 (9%) Isolated LVAD—survival 30-d: 75 (100%); survival to discharge: 68 (91%); 1-y survival: 63 (84%); CTx: 46 (61%); died on support: 17 (23%). LVAD-V-A ECMO—survival 30-d: 20 (87%); survival to discharge: 19 (83%); 1-y survival: 15 (65%); CTx: 11 (48%); died on support: 10 (43%)
   Scherer et al. (60) 10 LVAD-ECMO—6 pts had ECMO before LVAD for CS and after the LVAD the ECMO hasn’t been removed; 4 pts had ECMO after LVAD implantation; 8 femo-femoral ECMO, 2 subclavian-femoral ECMO; mean age: 53±10; male: 9 (90%); etiology: DCMP 7 (70%), ischemic 2 (20%), myocarditis 1 (10%); mean time support: 8±4 days Weaning ECMO: 10 (100%); mortality after weaning from ECMO: 4 (40%) (2 sepsis, 1 mesenteric ischemia, 1 GI bleeding); no death RV-related; overall survival: 6 (60%)
   Riebandt et al. (61) 122 isolated LVAD. 32 LVAD-ECMO—subclavian artery with a side graft + femoral vein percutaneously or femo-femoral ECMO percutaneously; mean age: 52±14; male: 27 (84.4%); etiology: DCMP 16 (50%), ischemic 13 (40.6%), other 3 (9.4%); in-hospital stay: 37±30 days LVAD-ECMO—weaning ECMO: 29 (90.6%); mortality on LVAD-ECMO: 3 (9.4%) (1 for sepsis, 1 for MOF, 1 for ischemic stroke); mortality 30-days: 6 (18.8%); in-hospital mortality: 8 (25%) [4 (12.5%) MOF, 2 (6.3%) sepsis, 1(4.5%) ischemic stroke, 1 (4.5%) hemorrhagic stroke]; 1-year survival: 24 (75%)
Isolated LVAD vs. LVAD-RVAD
   Kierman et al. (50) 9,580 isolated LVAD. 396 LVAD-RVAD—mean age >60 y: 161 (41.7%); male: 303 (78.5%); etiology: ischemic 189 (49.7%) Isolated LVAD—survival 30-d: 96.1%; LVAD-RVAD—survival 30-d: 73.5%
   Kormos et al. (20) 386 isolated LVAD—mean age: 51.8±13.5; male: 306 (79%); etiology: ischemic 174 (45%). 30 RVAD-LVAD—mean age: 51.0±13.3; male: 23 (77%); etiology: ischemic 15 (50%) Isolated LVAD—survival at 1-year: 78%±3%; RVAD-LVAD—survival at 1-year: 59%±9%
   Fischer et al. (62) 22 isolated LVAD—mean age: 61.9±8.9; male: 19 (86%); etiology: DCMP 10 (45%), ischemic 8 (36%), AMI 2 (9%). 22 LVAD-RVAD—mean age: 54.6±14.2; male: 19 (86%); etiology: DCMP 13 (59%), ischemic 8 (6%), AMI 1 (5%) Isolated LVAD—mean time to hospital discharge: 38.5 days; CTx: 10(46%); survival rate (%): at 1 month 90.9±6.1; at 3 months 76.5±9.2; at 6 months 71.4±9.9. LVAD-RVAD—mean duration RVAD support: 8.5 days; mean time to hospital discharge: 37 days; CTx: 10 (46%); survival rate (%): at 1 month 89.9±7.9; at 3 months 80.5±8.8; at 6 months 60.4±12
   Deschka et al. (45) 27 isolated LVAD—etiology: DCMP 14 (51.8%), ischemic 11 (40.7%), congenital 1 (3.7%), amyloidosis 1 (3.7%). 25 LVAD-RVAD—mean age: 55.4±12.4; male: 20 (80%); etiology: DCMP 10 (40%), ischemic 14 (56%), chronic rejection 1 (4%); mean duration RVAD: 11.1±7.2 days; annuloplasty tricuspid valve for IT: 15 (60%) Isolated LVAD—in-hospital survival: 19 (70.4%); 1-year survival: 66.7%. LVAD- RVAD—weaning RVAD: 23 (92%); in-hospital death on RVAD: 2 (8%) (1 MOF and 1 cerebral hemorrhage); in-hospital death after weaning RVAD: 6 (24%) (5 sepsis, 1 hemorrhagic shock and MOF); in-hospital survival: 17 (68%); follow-up at 18 months: 3/17 CTx, 5/17 deceased, 9 still on LVAD; 1-year survival: 56%
   Bhama et al. (63) 42 LVAD-RVAD—mean age: 55±13; male: 27 (64%); etiology: DCMP 18 (43%), ischemic 22 (52%), other 2 (5%) Weaning RVAD: 35 (83%); mortality <30 days or before discharge: 10 (24%); mortality>30 days or after discharge: 13 (31%)
Isolated LVAD vs. LVAD-ECMO vs. LVAD-RVAD
   Liedenfrost et al. (64) 240 isolated LVAD. 15 LVAD-RVAD—mean age: 64.6±7; etiology: ischemic 12 (80%). 12 LVAD-OxyRVAD—mean age: 45.6.6±16; etiology: ischemic 0 LVAD-RVAD—weaning RVAD: 10 (64%); 30-days mortality: 7 (47%). LVAD-OxyRVAD—weaning OxyRVAD: 10 (83%); 30-days mortality: 1 (8%)
   Noly et al. (40) 56 isolated LVAD. 10 V-A ECMO-LVAD—etiology: DCMP 2 (20%), ischemic 8 (80%); mean duration ECMO: 7.12±5.4days. 8 LVAD-OxyRVAD—etiology: DCMP 4 (50%) ischemic 4 (50%); mean duration RVAD: 9.57±3.5 days Isolated LVAD—mortality follow up: 24 (42%). V-A ECMO-LVAD—mortality on ECMO: 2 (20%) (MOF); mortality after weaning ECMO: 4 (40%) (1 suicide, 2 ischemic stroke, 1 hemorrhagic stroke); CTx: 3 (30%); mortality follow-up: 7 (70%). LVAD-OxyRVAD—mortality on OxyRVAD: 1 (12.5%); mortality after weaning RVAD: 2 (37.5%) (1 hemorrhagic stroke, 1 tamponade); CTx: 2 (25%); mortality follow-up: 4 (50%)
LVAD-Tandemlife ProtekDuo
   Ravichandran et al. (65) 12 ProtekDuo-LVAD—mean age: 56.3±8; male: 9 (76%); etiology: ischemic 5 (42%) Weaning ProtekDuo: 4 (33%); mortality on ProtekDuo: 7 (58%); duration ProtekDuo support: 10.5±6.5 days
   Schmack et al. (43) 11 ProtekDuo-LVAD—mean age: 51.6±13.1; male: 10 (90.0%); etiology: DCMP 5 (45.5%), ischemic 6 (54.5%) Weaning ProtekDuo: 10 (90.9%); survival 30-d: 8(72.7%); mortality follow-up: 4 (36.4%) (3 MOF, 1 cerebral hemorrhage); duration ProtekDuo support: 16.8±9.5 days

AMI, acute myocardial infarction; CS, cardiogenic shock; CTx, heart transplantation; DCMP, dilative cardiomyopathy; ECMO, extracorporeal membrane oxygenation; GI, gastrointestinal; IT, tricuspid insufficiency; LVAD, left ventricular assist device; MOF, multi-organ failure; Pt(s), patient(s); RV, right ventricle; RVF, right ventricular failure; RVAD, right ventricular assist device; V-A ECMO, veno-arterial ECMO.

Another factor which influences the prognosis in such a setting is the timing of RV support. Sometimes the severity of RVF is underestimated prior to or at the LVAD implant, with delayed RV support and consequent worse outcome. Bhama and associates, showed a better overall survival at 3 months in patients who received an immediate RVAD as compared to a delayed support (79% vs. 46%) (63). Additionally, a recent study reported that the early use of a temporary RVAD in case of risk factors of RVF could improve the outcome and consequently decrease the overall mortality (62).

The duration of the support seems to affect the outcomes as well. In particular, a RV support lasting for more than 7 days and a blood flow greater than 4L/min after LVAD may be associated with pulmonary bleeding complications due to the non-pulsatile flow perfusion of the lung during RVAD (67). Other studies confirmed the importance of the temporary ECMO for the right support (68). In particular, Riebandt and associates (61), showed that patients treated with ECMO have a 75% of in-hospital survival with no post-discharge mortality at 1 year. Another relevant experience was reported by Scherer and colleagues (60), who suggested that a preoperative ECMO can stabilize organ perfusion before the LVAD implantation and, in order to avoid RVF, it should not be removed for the immediate postoperative period. Furthermore, Leidenfrost and collaborators (64), reported improved metabolic parameters and a lower 30-day mortality in patients who received a RVAD after continuous-flow LVAD implantation.

In case of biventricular failure, it’s still not clear which is the perfect strategy of management. Shehab and colleagues (59), compared the outcome of BIVAD and a veno-pulmonary arterial ECMO, and observed that the ECMO is associated with a better in-hospital and late survival. A long-term dual VAD could be considered in patients with evident RVF. Those with mild signs of RVF may benefit from temporary percutaneous implantable pumps in the early perioperative period to allow for RV hemodynamic stabilization (69). A recent study showed another viable option in case of biventricular failure through the combination of an Impella CP, inserted via the axillary artery, with the ProtekDuo, connected with a TH pump allowing the patient to be awake during the support and the mobilization (70). Since its introduction in 2016, ProtekDuo cannula has been used mostly for RVF after LVAD. It provides good outcome with a high rate of weaning and RV recovery from the support. Clinical trials have to be conducted to elucidate the timing and the exact patient population (71,72). The largest experience with ProtekDuo has been reported by Ravichandran and collaborators (65). They showed how an earlier placement of the double-lumen cannula improves outcomes and reduces common complications (65).

Another percutaneous device for the RVF is the Impella RP. Hemodynamically, Impella has shown to increase the cardiac index restoring the forward perfusion, decrease the CVP, unloading of the RV, reducing systemic venous congestion and enhancing the end-organ perfusion. Its placement in LVAD patients with post-operative RVF has been associated with less necessity of inotropic or vasopressor therapy and a better outcome (survival at 30-days or discharge in 83%; all discharged patients were alive at 180 days) (48). There were no thromboembolic complications, no pulmonary embolism, and minimal cardiac structure or vascular perforation or damage. The most prevalent complication was bleeding, however, with a lower rate compared to surgical RVAD and hemolysis, while worsening of tricuspid or pulmonary valve dysfunction was infrequent. The access site to the femoral vein is however, limiting patient mobilization (47,73,74). The use of Impella is growing in the context of isolated RVF and biventricular dysfunction in patients with short- and long-term LVAD. For the biventricular support, the BIPELLA, a combination of Impella CP for the left support and Impella RP for the right one has shown promising results (49,75).

In literature, the use of TH for RVF is associated with a widely different mortality, with the lowest in the after-LVAD setting. Patients who didn’t survived to hospital discharge were older and had biventricular failure or major bleeding complications (7,76,77).

The choice of the mechanical support in such a setting is still a challenge. In the decision-making, it is important to consider the timing of RVF occurrence and the strategy of the LVAD implantation. If the RVF occurs in the operating room, valid options might be a direct RV support (RA and PA cannulation) or to consider a peripheral V-A ECMO as temporary support. An additional intra-operative approach might include a dedicated RV support with RA drainage through a femoral vein and graft-interposed PA cannulation. Instead, if a full percutaneous approach is considered, implantation of a percutaneous double-lumen ProtekDuo cannula in the PA or Impella RP through the femoral vein, are good options, particularly if the RVF occurs in the ICU or perioperatively. V-A ECMO is often the first-line of support because of the facility of implantation, the availability of the machine in most of the centers, and the lower cost compared to the percutaneous or other devices. However, as previously mentioned, the effects of reduced LVAD loading and the competitive retrograde aortic flow against the LVAD ejection, should be carefully considered.

The strategy of the LVAD (bridge-to-transplantation or destination therapy) may also influence the decision-making. In any pre-LVAD implant situation, a proper RV assessment should be applied to avoid or reduce the risk of RVF regardless the strategy of the LVAD. In any post-LVAD situations, the first approach should be always a temporary RV support. In case of no recovery of the RV after adequate waiting time and a bridge-to-transplantation objective, an attempt to speed up the waiting time to organ availability and transplanting the patient should be pursued. In case of a destination therapy and failing temporary RV support, it is doable to consider a long-term VAD (HeartWare, Heartmate) in right position, therefore, achieving a BiVAD support. Despite the available series, however, new studies are needed to define the best mechanical cardio-circulatory support for RVF, the appropriate timing of implantation, and the best candidates for each of devices.

Conclusions

RVF after LVAD is rather frequent and poses a significant challenge. Prediction is difficult and prognostic scores are variable. In case of refractory RVF, the use of temporary mechanical support may be required, providing a better survival. The percutaneous V-A ECMO is increasingly considered and with several options for post-LVAD RVF, supplying many treatments modalities and configurations. In case of RVF, a timely initiation of support provides a better survival with results comparable to those in LVAD implantation without RVF. Despite ECMO is a valuable tool, drawbacks and complications must be considered, particularly the effects on LVAD function. Finally, the development of new approaches and cannulas, such as the ProtekDuo, may lead to a minimally invasive and full percutaneous support enhancing the decision-making and management of this population. However, more studies are warranted to prove the hemodynamic and clinical effect of such attractive approaches.

Supplementary

The article’s supplementary files as

jtd-13-02-1256-rc.pdf (120.9KB, pdf)
DOI: 10.21037/jtd-20-2228
jtd-13-02-1256-coif.pdf (159.5KB, pdf)
DOI: 10.21037/jtd-20-2228

Acknowledgments

Funding: None.

Ethical Statement: The authors are accountable for all aspects of the work in ensuring that questions related to the accuracy or integrity of any part of the work are appropriately investigated and resolved.

Footnotes

Reporting Checklist: The authors have completed the PRISMA reporting checklist. Available at http://dx.doi.org/10.21037/jtd-20-2228

Conflicts of Interest: All authors have completed the ICMJE uniform disclosure form (available at http://dx.doi.org/10.21037/jtd-20-2228). RL reports other from Medtronic, other from LivaNova, other from Eurosets, outside the submitted work. MK serves as an unpaid editorial board member of Journal of Thoracic Disease from Sep 2020 to Aug 2022. The other authors have no conflicts of interest to declare.

References

  • 1.Harjola VP, Mebazaa A, Čelutkienė J, et al. Contemporary management of acute right ventricular failure: a statement from the Heart Failure Association and the Working Group on Pulmonary Circulation and Right Ventricular Function of the European Society of Cardiology. Eur J Heart Fail 2016;18:226-41. 10.1002/ejhf.478 [DOI] [PubMed] [Google Scholar]
  • 2.Dang NC, Topkara VK, Mercando M, et al. Right heart failure after left ventricular assist device implantation in patients with chronic congestive heart failure. J Heart Lung Transplant 2006;25:1-6. 10.1016/j.healun.2005.07.008 [DOI] [PubMed] [Google Scholar]
  • 3.Scherer M, Sirat AS, Moritz S. Extracorporeal membrane oxygenation as perioperative right ventricular support in patients with biventricular failure undergoing left ventricular assist device implantation. Eur J Cardiothorac Surg 2011;39:939-44. 10.1016/j.ejcts.2010.09.044 [DOI] [PubMed] [Google Scholar]
  • 4.Aissaoui N, Morshuis M, Schoenbrodt M, et al. Temporary right ventricular mechanical circulatory support for the management of right ventricular failure in critically ill patients. J Thorac Cardiovasc Surg 2013;146:186-91. 10.1016/j.jtcvs.2013.01.044 [DOI] [PubMed] [Google Scholar]
  • 5.Konstam MA, Kiernan MS, Bernstein D, et al. Evaluation and Management of Right-Sided Heart Failure: A Scientific Statement from the American Heart Association. Circulation 2018;137:e578-622. 10.1161/CIR.0000000000000560 [DOI] [PubMed] [Google Scholar]
  • 6.Mulaikal TA, Bell LH, Li B, et al. Isolated Right Ventricular Mechanical Support: Outcomes and Prognosis. ASAIO J 2018;64:e20-7. 10.1097/MAT.0000000000000597 [DOI] [PubMed] [Google Scholar]
  • 7.Kapur NK, Esposito ML, Bader Y, et al. Mechanical Circulatory Support Devices for Acute Right Ventricular Failure. Circulation 2017;136:314-26. 10.1161/CIRCULATIONAHA.116.025290 [DOI] [PubMed] [Google Scholar]
  • 8.Ali HR, Kiernan MS, Choudhary G, et al. Right Ventricular Failure Post-Implantation of Left Ventricular Assist Device: Prevalence, Pathophysiology, and Predictors. ASAIO J 2020;66:610-9. 10.1097/MAT.0000000000001088 [DOI] [PubMed] [Google Scholar]
  • 9.Cordtz J, Nilson JC, Hansen PB, et al. Right ventricular failure after implantation of a continuous-flow left ventricular assist device: Early hemodynamic predictors. Eur J Cardiothorac Surg 2014;45:847-53. 10.1093/ejcts/ezt519 [DOI] [PubMed] [Google Scholar]
  • 10.Greyson CR. Pathophysiology of right ventricular failure. Crit Care Med 2008;36:S57-65. 10.1097/01.CCM.0000296265.52518.70 [DOI] [PubMed] [Google Scholar]
  • 11.Feldman D, Pamboukian SV, Teuteberg JJ, et al. The 2013 International Society for Heart and Lung Transplantation Guidelines for mechanical circulatory support: executive summary. J Heart Lung Transplant 2013;32:157-87. 10.1016/j.healun.2012.09.013 [DOI] [PubMed] [Google Scholar]
  • 12.Krishan K, Nair A, Pinney S, et al. Liberal use of tricuspid-valve annuloplasty during left-ventricular assist device implantation. Eur J Cardiothorac Surg 2012;41:213-7. 10.1016/j.ejcts.2011.05.047 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13.Kavarana MN, Pessin-Minsley MS, Urtecho J, et al. Right ventricular dysfunction and organ failure in left ventricular assist device recipients: a continuing problem. Ann Thorac Surg 2002;73:745-50. 10.1016/S0003-4975(01)03406-3 [DOI] [PubMed] [Google Scholar]
  • 14.Brisco MA, Sundareswaran KS, Milano CA, et al. Incidence, risk, and consequences of atrial arrhythmias in patients with continuous-flow left ventricular assist devices. J Card Surg 2014;29:572-80. 10.1111/jocs.12336 [DOI] [PubMed] [Google Scholar]
  • 15.Turner KR. Right Ventricular Failure After Left Ventricular Assist Device Placement-The Beginning of the End or Just Another Challenge? J Cardiothorac Vasc Anesth 2019;33:1105-21. 10.1053/j.jvca.2018.07.047 [DOI] [PubMed] [Google Scholar]
  • 16.Raina A, Seetha Rammohan HR, Gertz ZM, et al. Postoperative right ventricular failure after left ventricular assist device placement is predicted by preoperative echocardiographic structural, hemodynamic and functional parameters. J Card Fail 2013;19:16-24. 10.1016/j.cardfail.2012.11.001 [DOI] [PubMed] [Google Scholar]
  • 17.Lampert BC, Teuteberg JJ. Right ventricular failure after left ventricular assist devices. J Heart Lung Transplant 2015;34:1123-30. 10.1016/j.healun.2015.06.015 [DOI] [PubMed] [Google Scholar]
  • 18.Ochiai Y, McCathy PM, Smedira NG, et al. Predictors of severe right ventricular failure after implantable left ventricular assist device insertion: analysis of 245 patients. Circulation 2002;106:I198-202. [PubMed] [Google Scholar]
  • 19.Fitzpatrick JR, 3rd, Friederick JR, Hiesinger W, et al. Early planned institution of biventricular mechanical circulatory support results in improved outcome compared with delayed conversion of a left ventricular assist device to a biventricular assist device. J Thorac Cardiovasc Surg 2009;137:971-7. 10.1016/j.jtcvs.2008.09.021 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 20.Kormos RL, Teuteberg JJ, Pagani FD, et al. Right ventricular failure in patients with the HeartMate II continuous-flow left ventricular assist device: incidence, risk factors, and effect on outcomes. J Thorac Cardiovasc Surg 2010;139:1316-24. 10.1016/j.jtcvs.2009.11.020 [DOI] [PubMed] [Google Scholar]
  • 21.Potapov EV, Stepanenko A, Dandel M, et al. Tricuspid incompetence and geometry of the right ventricle as predictors of right ventricular function after implantation of a left ventricular assist device. J Heart Lung Transplant 2008;27:1275-81. 10.1016/j.healun.2008.08.012 [DOI] [PubMed] [Google Scholar]
  • 22.Morgan JA, Brewer RJ, Nemeh HW, et al. Left ventricular reverse remodeling with a continuous flow left ventricular assist device measured by left ventricular end-diastolic dimensions and severity of mitral regurgitation. ASAIO J 2012;58:574-7. 10.1097/MAT.0b013e31826e4267 [DOI] [PubMed] [Google Scholar]
  • 23.Tang PC, Haft JW, Romano MA, et al. Right ventricular function and residual mitral regurgitation after left ventricular assist device implantation determines the incidence of right heart failure. J Thorac Cardiovasc Surg 2020;159:897-905.e4. 10.1016/j.jtcvs.2019.03.089 [DOI] [PubMed] [Google Scholar]
  • 24.Fukuhara S, Takeda K, Chiuzan C, et al. Concomitant mitral repair and continuous-flow left ventricular assist devices: Is it warranted? J Thorac Cardiovasc Surg 2017;154:1303-1312.e4. 10.1016/j.jtcvs.2017.04.021 [DOI] [PubMed] [Google Scholar]
  • 25.Robertson JO, Grau-Sepulveda MV, Okada S, et al. Concomitant tricuspid valve surgery during implantation of continuous-flow left ventricular assist devices: a Society of Thoracic Surgeons database analysis. J Heart Lung Transplant 2014;33:609-17. 10.1016/j.healun.2014.01.861 [DOI] [PubMed] [Google Scholar]
  • 26.Maltais S, Topilsky Y, Tchantchaleishvili V, et al. Surgical treatment of tricuspid valve insufficiency promotes early reverse remodeling in patients with axial-flow left ventricular assist devices. J Thorac Cardiovasc Surg 2012;143:1370-6. 10.1016/j.jtcvs.2011.07.014 [DOI] [PubMed] [Google Scholar]
  • 27.Atluri P, Goldstone AB, Fairman AS, et al. Predicting right ventricular failure in the modern, continuous flow left ventricular assist device era. Ann Thorac Surg 2013;96:857-63. 10.1016/j.athoracsur.2013.03.099 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 28.Fukamachi K, McCarthy PM, Smedira NG, et al. Preoperative risk factors for right ventricular failure after implantable left ventricular assist device insertion. Ann Thorac Surg 1999;68:2181-4. 10.1016/S0003-4975(99)00753-5 [DOI] [PubMed] [Google Scholar]
  • 29.Drakos SG, Janicki L, Horne BD, et al. Risk factors predictive of right ventricular failure after left ventricular assist device implantation. Am J Cardiol 2010;105:1030-5. 10.1016/j.amjcard.2009.11.026 [DOI] [PubMed] [Google Scholar]
  • 30.Alnsasra H, Asleh R, Schettle SD, et al. Diastolic Pulmonary Gradient as a Predictor of Right Ventricular Failure After Left Ventricular Assist Device Implantation. J Am Heart Assoc 2019;8:e012073. 10.1161/JAHA.119.012073 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 31.MacGowan GA, Schueler S. Right heart failure after left ventricular assist device implantation: early and late. Curr Opin Cardiol 2012;27:296-300. 10.1097/HCO.0b013e3283511e60 [DOI] [PubMed] [Google Scholar]
  • 32.Matthews JC, Koelling TM, Pagani FD, et al. The right ventricular failure risk score a pre-operative tool for assessing the risk of right ventricular failure in left ventricular assist device candidates. J Am Coll Cardiol 2008;51:2163-72. 10.1016/j.jacc.2008.03.009 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 33.Loforte A, Montalto A, Musumeci F, et al. Calculation of the ALMA Risk of Right Ventricular Failure After Left Ventricular Assist Device Implantation. ASAIO J 2018;64:e140-7. 10.1097/MAT.0000000000000800 [DOI] [PubMed] [Google Scholar]
  • 34.Lo Coco V, Lorusso R, Raffa GM, et al. Clinical complications during veno-arterial extracorporeal membrane oxygenation in post-cardiotomy and non-post-cardiotomy shock: still the Achilles’ heel. J Thorac Dis 2018;10:6993-7004. 10.21037/jtd.2018.11.103 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 35.Loforte A, Baiocchi M, Gliozzi G, et al. Percutaneous PA venting via jugular vein while on peripheral extracorporeal membrane oxygenation running: a less invasive approach to provide full biventricular unloading. Ann Cardiothorac Surg 2019;8:163-6. 10.21037/acs.2018.08.06 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 36.Loforte A, Baiocchi M, Dal Checco E, et al. Percutaneous PA Venting via Jugular Vein While on Peripheral Extracorporeal Life Support. ASAIO J 2020;66:e50-4. 10.1097/MAT.0000000000000991 [DOI] [PubMed] [Google Scholar]
  • 37.de Pommereau A, Radu C, Boukantar M, et al. Left Ventricle Unloading Through Pulmonary Artery in Patients with Venoarterial Extracorporeal Membrane Oxygenation. ASAIO J 2021;67:e49-51. 10.1097/MAT.0000000000001179 [DOI] [PubMed] [Google Scholar]
  • 38.Lorusso R, Raffa GM, Heuts S, et al. Pulmonary artery cannulation to enhance extracorporeal membrane oxygenation management in acute cardiac failure. Interact Cardiovasc Thorac Surg 2020;30:215-22. [DOI] [PubMed] [Google Scholar]
  • 39.Haneya A, Philipp A, Puehler T, et al. Temporary percutaneous right ventricular support using a centrifugal pump in patients with postoperative acute refractory right ventricular failure after left ventricular assist device implantation. Eur J Cardiothorac Surg 2012;41:219-23. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 40.Noly PE, Kirsch M, Quessard A, et al. Temporary right ventricular support following the left ventricle assist device implantation: A comparison of two techniques. Interact Cardiovasc Thorac Surg 2014;19:49-55. 10.1093/icvts/ivu072 [DOI] [PubMed] [Google Scholar]
  • 41.Lenoir M, Quessard A, N’guyen A, et al. Simplified temporary right ventricular support after implantation of a left ventricular assist device. Heart Surg Forum 2013;16:E152-4. 10.1532/HSF98.20131003 [DOI] [PubMed] [Google Scholar]
  • 42.Takayama H, Naka Y, Kodali SK, et al. A novel approach to percutaneous right ventricular mechanical support Eur J Cardiothorac Surg 2012;41:423-6. 10.1016/j.ejcts.2011.05.041 [DOI] [PubMed] [Google Scholar]
  • 43.Schmack B, Weymann A, Popov AV, et al. Concurrent left ventricular assist device (LVAD) implantation and percutaneous temporary RVAD support via Cardiac Assist Protek-Duo TandemHeart to preempt right heart failure. Med Sci Monit Basic Res 2016;22:53-7. 10.12659/MSMBR.898897 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 44.Bermudez CA, Lagazzi L, Crespo MM. Prolonged support using a percutaneous OxyRvad in a patient with end-stage lung disease, pulmonary hypertension, and right cardiac failure. ASAIO J 2016;62:e37-40. 10.1097/MAT.0000000000000343 [DOI] [PubMed] [Google Scholar]
  • 45.Deschka H, Holthaus AJ, Sindermann JR, et al. Can Perioperative Right Ventricular Support Prevent Postoperative Right Heart Failure in Patients with Biventricular Dysfunction Undergoing Left Ventricular Assist Device Implantation? J Cardiothorac Vasc Anesth 2016;30:619-26. 10.1053/j.jvca.2016.02.023 [DOI] [PubMed] [Google Scholar]
  • 46.Wang D, Zou X, Liu X, et al. OxyRVAD for total right heart and respiratory support. Conf Proc IEEE Eng Med Biol Soc 2006;2006:5390-1. 10.1109/IEMBS.2006.260849 [DOI] [PubMed] [Google Scholar]
  • 47.Morgan JA, O’Neil WW. Percutaneous right ventricular assist device support in a patient supported by an LVAD. ASAIO J 2016;62:e41-2. 10.1097/MAT.0000000000000344 [DOI] [PubMed] [Google Scholar]
  • 48.Anderson MB, Goldstein J, Milano C, et al. Benefits of novel percutaneous ventricular assist device for right heart failure, the prospective RECOVER RIGHT study of the Impella RP device. J Heart Lung Transplant 2015;34:1549-60. 10.1016/j.healun.2015.08.018 [DOI] [PubMed] [Google Scholar]
  • 49.Pieri M, Pappalardo F. Impella RP in the Treatment of Right Ventricular Failure: What We Know and Where We Go. J Cardiothorac Vasc Anesth 2018;32:2339-43. 10.1053/j.jvca.2018.06.007 [DOI] [PubMed] [Google Scholar]
  • 50.Kiernan MS, Grandin EW, Brinkley M, Jr, et al. Early Right Ventricular Assist Device Use in Patients Undergoing Continuous-Flow Left Ventricular Assist Device Implantation: Incidence and Risk Factors from the Interagency Registry for Mechanically Assisted Circulatory Support. Circ Heart Fail 2017;10:e003863. 10.1161/CIRCHEARTFAILURE.117.003863 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 51.Kapur NK, Paruchuri V, Jagannathan A, et al. Mechanical circulatory support for right ventricular failure. JACC Heart Fail 2013;1:127-34. 10.1016/j.jchf.2013.01.007 [DOI] [PubMed] [Google Scholar]
  • 52.Schürner A, Wilhelm MJ, Falk V, et al. Recurrent clotting of a continuous-flow right ventricular assist device: repeated thrombolysis with two different protocols. J Cardiothorac Vasc Anesth 2015;29:1614-7. 10.1053/j.jvca.2014.11.013 [DOI] [PubMed] [Google Scholar]
  • 53.Krabatsch T, Potapov E, Stepanenko A, et al. Biventricular circulatory support with two miniaturized implantable assist devices. Circulation 2011;124:S179-86. 10.1161/CIRCULATIONAHA.110.011502 [DOI] [PubMed] [Google Scholar]
  • 54.Griffith KE, Jenkins E, Stulak J, et al. Long-term use of the CentriMag(R) Ventricular Assist System as a right ventricular assist device: a case report. Perfusion 2012;27:65-70. 10.1177/0267659111424634 [DOI] [PubMed] [Google Scholar]
  • 55.Coromilas EJ, Takeda K, Ando M, et al. Comparison of Percutaneous and Surgical Right Ventricular Assist Device Support After Durable Left Ventricular Assist Device Insertion. J Card Fail 2019;25:105-13. 10.1016/j.cardfail.2018.12.005 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 56.John R, Lee S, Eckman P, et al. Right ventricular failure- a continuing problem in patients with left ventricular assist device support. J Cardiovasc Transl Res 2010;3:604-11. 10.1007/s12265-010-9216-4 [DOI] [PubMed] [Google Scholar]
  • 57.Patlolla B, Beygui R, Haddad F. Right ventricular failure following left ventricle assist device implantation. Curr Opin Cardiol 2013;28:223-33. 10.1097/HCO.0b013e32835dd12c [DOI] [PubMed] [Google Scholar]
  • 58.Marquez TT, D’Cunha J, John R, et al. Mechanical support for acute right ventricular failure: evolving surgical paradigms. J Thorac Cardiovasc Surg 2009;137:e30-40. 10.1016/j.jtcvs.2008.02.025 [DOI] [PubMed] [Google Scholar]
  • 59.Shehab S, Rao S, Macdonald P, et al. Outcome of venopulmonary arterial extracorporeal life support as temporary right ventricular support left ventricular assist implantation. J Thorac Cardiovasc Surg 2018;156:2143-52. 10.1016/j.jtcvs.2018.05.077 [DOI] [PubMed] [Google Scholar]
  • 60.Scherer M, Moritz A, Martens S. The use of extracorporeal membrane oxygenation in patients with therapy refractory cardiogenic shock as a bridge to implantable left ventricular assist device and perioperative right heart support. J Artif Organs 2009;12:160. 10.1007/s10047-009-0464-5 [DOI] [PubMed] [Google Scholar]
  • 61.Riebandt J, Haberl T, Wiedemann D, et al. Extracorporeal membrane oxygenation support for right ventricular failure after left ventricular assist device implantation. Eur J Cardiothorac Surg 2018;53:590-5. 10.1093/ejcts/ezx349 [DOI] [PubMed] [Google Scholar]
  • 62.Fischer Q, Kirsch M. Liberal Right Ventricular Assist Device Extracorporeal Membrane Oxygenation Support for Right Ventricular Failure after Implantable Left Ventricular Assist Device Placement. ASAIO J 2018;64:741-7. 10.1097/MAT.0000000000000735 [DOI] [PubMed] [Google Scholar]
  • 63.Bhama JK, Bamsal U, Winger DG, et al. Clinical experience with temporary right ventricular mechanical circulatory support. J Thorac Cardiovasc Surg 2018;156:1885-91. 10.1016/j.jtcvs.2018.04.094 [DOI] [PubMed] [Google Scholar]
  • 64.Leidenfrost J, Prasad S, Itoh A, et al. Right ventricular assist device with membrane oxygenator support for right ventricular failure following implantable left ventricular assist device placement. Eur J Cardiothorac Surg 2016;49:73-7. 10.1093/ejcts/ezv116 [DOI] [PubMed] [Google Scholar]
  • 65.Ravichandran AK, Baran DA, Stelling K, et al. Outcomes with the Tandem ProtekDuo Dual-Lumen Percutaneous Right Ventricular Assist Device. ASAIO J 2018;64:570-2. 10.1097/MAT.0000000000000709 [DOI] [PubMed] [Google Scholar]
  • 66.Soliman O, Akin S, Muslem R, et al. Derivation and Validation of a Novel Right-Sided Heart Failure Model After Implantation of Continuous Flow Left Ventricular Assist Devices: The EUROMACS (European Registry for Patients with Mechanical Circulatory Support) Right-Sided Heart Failure Risk Score. Circulation 2018;137:891-906. 10.1161/CIRCULATIONAHA.117.030543 [DOI] [PubMed] [Google Scholar]
  • 67.Welp H, Sindermann JR, Deschka H, et al. Pulmonary Bleeding During Right Ventricular Support After Left Ventricular Assist Device Implantation. J Cardiothorac Vasc Anesth 2016;30:627-31. 10.1053/j.jvca.2015.07.012 [DOI] [PubMed] [Google Scholar]
  • 68.Dalen M, Sartip U, Lund LH, et al. Peripheral Extracorporeal Membrane Oxygenation as Short-Term Right Ventricular Support After HeartWare Left Ventricular Assist Device Implantation. ASAIO J 2013;59:523-5. 10.1097/MAT.0b013e31829be043 [DOI] [PubMed] [Google Scholar]
  • 69.Shehab S, Hayward CS. Choosing Between Left Ventricular Assist Devices and Biventricular Assist Devices. Card Fail Rev 2019;5:19-23. 10.15420/cfr.2018.23.2 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 70.Patel NJ, Verma DR, Gopalan R, et al. Percutaneous Biventricular Mechanical Circulatory Support with Impella CP and Protek Duo Plus TandemHeart. J Invasive Cardiol 2019;31:E46. [DOI] [PubMed] [Google Scholar]
  • 71.Xu T, Trimlett S, Davies S, et al. Use of ProtekDuo in patients with acute right ventricular failure. Conference: 8th EuroELSO Congress; 10-13.04.2019 At Barcellona, Spain. Volume: Perfusion Supplement, Vol 34(15) 139-140. [Google Scholar]
  • 72.Kazui T, Tran PL, Echeverria A, et al. Minimally invasive approach for percutaneous CentriMag right ventricular assist device support using a single PROTEKDuo Cannula. J Cardiothorac Surg 2016;11:123. 10.1186/s13019-016-0515-y [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 73.Anderson M, Morris DL, Tang D, et al. Outcomes of patients with right ventricular failure requiring short-term hemodynamic support with the Impella RP device. J Heart Lung Transplant 2018;37:1448-58. 10.1016/j.healun.2018.08.001 [DOI] [PubMed] [Google Scholar]
  • 74.Cheung AW, White CW, Davis MK, et al. Short-term mechanical circulatory support for recovery from acute right ventricular failure: clinical outcomes. J Heart Lung Transplant 2014;33:794-9. 10.1016/j.healun.2014.02.028 [DOI] [PubMed] [Google Scholar]
  • 75.Pappalardo F, Scandroglio AM, Latib A. Full percutaneous biventricular support with two Impella pumps: The BI-Pella Approach ESC Heart Failure 2018. ESC Heart Fail 2018;5:368-71. 10.1002/ehf2.12274 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 76.Kapur NK, Paruchuri V, Korabathina R, et al. Effects of a percutaneous mechanical circulatory support device for medically refractory right ventricular failure. J Heart Lung Transplant 2011;30:1360-7. 10.1016/j.healun.2011.07.005 [DOI] [PubMed] [Google Scholar]
  • 77.Loor G, Khani-Hanjani A, Gonzalez-Stawinski GV. Use of RotaFlow DMAQUET for temporary right ventricular support during implantation of HeartMate II left ventricular assist device. ASAIO J 2012;58D3:275-7. [DOI] [PubMed]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

The article’s supplementary files as

jtd-13-02-1256-rc.pdf (120.9KB, pdf)
DOI: 10.21037/jtd-20-2228
jtd-13-02-1256-coif.pdf (159.5KB, pdf)
DOI: 10.21037/jtd-20-2228

Articles from Journal of Thoracic Disease are provided here courtesy of AME Publications

RESOURCES