Skip to main content
Diabetes Technology & Therapeutics logoLink to Diabetes Technology & Therapeutics
. 2021 Mar 2;23(Suppl 1):S-27–S-34. doi: 10.1089/dia.2021.0007

Type 2 Diabetes and the Use of Real-Time Continuous Glucose Monitoring

Melanie A Jackson 1, Andrew Ahmann 2,, Viral N Shah 3
PMCID: PMC7957379  PMID: 33534631

Abstract

The role of continuous glucose monitoring (CGM) in type 1 diabetes (T1D) is well established in improving glycemic control and reducing hypoglycemia. Type 2 diabetes (T2D) is more prevalent than T1D and management of T2D is more heterogeneous, requiring treatment ranging from lifestyle modification to oral medications to intensive insulin therapy. Recent randomized controlled trials in intensively insulin-treated T2D demonstrated the efficacy and safety of real-time CGM (rtCGM) in reducing glycated hemoglobin without increasing hypoglycemia. Although evidence is limited, early studies have indicated a role for rtCGM in selected patients with non-insulin requiring T2D to improve glycemic control and/or reduce hypoglycemia. Based on literature review, we summarized current data on the use of rtCGM in T2D management and provided future research direction to generate more evidence on the utility of CGM in this population.

Keywords: Type 2 diabetes, Continuous glucose monitor, Glucose monitoring, Glycemic control, HbA1c

Introduction

Diabetes management introduces a significant burden for the patient, who is faced with a wealth of new knowledge at the time of diagnosis. This is later accompanied by ongoing lifestyle modification, multiple medications, and frequent glucose monitoring. As more data have emerged clarifying the importance of glucose monitoring, it has become apparent that treating diabetes without monitoring is comparable to taking a long trip in an automobile without the benefit of a map. Despite glucose meters becoming smaller, faster, and more accurate, while requiring less blood, adherence to self-monitoring of blood glucose (SMBG) is difficult.1,2 In addition, capillary glucose measurements do not provide glucose trends, making proactive management unfeasible.

Among the remarkable advances in diabetes technology witnessed by the diabetes community in the past decade, continuous glucose monitoring (CGM) is having the greatest impact. CGM devices have become smaller, more affordable, more accurate, and more user-friendly. With increasing CGM use, most notably in type 1 diabetes (T1D) treatment, and increased awareness of the limitations of HbA1c in the management of diabetes, new CGM-based metrics such as time in range (TIR), time in hypoglycemia, and blood glucose variability (coefficient of variation or CV) have been recommended by an international expert panel as more meaningful targets for diabetes management than HbA1c.3,4 From the mean glucose, one can also calculate an estimated HbA1c (eA1c) or the glucose management index, which can be particularly helpful in telemedicine where a laboratory HbA1c value may not be available.

The prevalence of type 2 diabetes (T2D) is higher than T1D, and over time, insulin often becomes necessary to achieve glycemic control.5–7 Individualized goals for glycemic control in T1D and T2D are similar, and include TIR greater than 70% of the time, time below range (<70 mg/dL) less than 4%, and glucose variability (CV) ≤36%4 (Table 1). Despite numerous advances in therapeutics, many of the patients with T2D are not able to achieve glycemic control.8 This may be due to relatively less use of diabetes technology in patients with T2D compared to patients with T1D.

Table 1.

Recommended Glycemic Targets for Type 2 Diabetes Patients on Continuous Glucose Monitors

Parameter T2D (%) T2D with advanced age or significant comorbidities
HbA1c <7.0 <8.0%
TIR (70–180 mg/dL) >70 >50%
% Under 70 mg/dL <4 <1%
% Under 54 mg/dL <1 To be avoided
Glucose variability (CV) ≤36 ≤33%

Adapted from Battelino et al.4

CV, coefficient of variation; T2D, type 2 diabetes; TIR, time in range.

Several studies demonstrated improvement in glycemic control and reduction in hypoglycemia with the use of CGM in patients with T1D irrespective of age, sex, educational status, or mode of insulin delivery.9–15 Evidence suggests sustained improvement in glycemic control with the use of CGM in early-onset T1D, calling for benefit in starting CGM as early as from the disease onset.16 There is a potential role of CGM in T2D as well, with patients facing many of the same obstacles to self-management as those with T1D.14,17,18 In addition, CGM has shown characteristics of various pharmacologic agents that aid in their use by differentiating postprandial effects, hypoglycemia risk, and glucose variability.19,20

In this review, we searched the literature for published real-time CGM (rtCGM) studies reporting efficacy, safety, quality of life, and lifestyle modifications in patients with T2D. We initiated our search by cross referencing “continuous glucose monitoring” and “type 2 diabetes mellitus” in PubMed. The initial 1678 references were narrowed by adding the search term “real-time” (148 items). Additional refinement included specifying articles representing controlled trials, prospective nonrandomized trials, and previous review articles on the topic, including meta-analyses. Professional guidelines as well as information gathered from professional CGM (blinded) or flash CGM were also reviewed but are cited sparingly to demonstrate principles that could apply to T2D and highlight novel opportunities to expand the use of CGM in T2D. However, the described outcomes are specific to real-time CMG reports.

This is a review of current CGM use and not a comprehensive meta-analysis. Several studies using devices no longer available or not widely used are not included. Importantly, given the high number of individuals with T2D not on insulin therapy, we looked for data that may specifically pertain to this group. Finally, we highlighted some of the limitations of currently available literature and future directions. Randomized controlled trials (RCTs) examining multiple parameters of diabetes care as well as observational studies providing an early look at the role of rtCGM in telemedicine are summarized in Table 2.

Table 2.

Real-Time Continuous Glucose Monitoring Trials in Patients with Type 2 Diabetes

Author (year)ref. Design N/Duration Treatment Change in HbA1c Hypoglycemia QOL/Behavior change Conclusion
RCTs
 Yoo et al. (2008)21 RCT
SMBG versus rtCGM
N = 65
12 weeks
About 60% on insulin
CGM to guide meds/adherence
HbA1c ↓ 1.1%
Δ = 0.7%
NR QOL NR
↓ weight and caloric intake and ↑ exercise
CGM used as a motivational intervention improved HbA1c
 Cox et al. (2020)22 RCT 1:2 ratio
Routine care versus CGM to educate on pp excursions
N = 30
2 months active
3-month follow-up
No insulin HbA1c ↓ 1.3%
Decreased medication use
NR Improved QOL, diabetes distress, and DM knowledge CGM helps behavioral modification to ↓ pp excursions
 Ehrhardt et al. (2011)23 RCT
rtCGM versus SMBG
N = 100
12 weeks
Oral agents ± basal insulin HbA1c ↓ 1.0%
Δ = 0.5%
NR QOL NR
No Δ in weight
First RCT to show benefit with prandial insulin
 Beck et al. (2017)24 RCT
rtCGM versus SMBG
Blinded CGM data in SMBG group
N = 158
24 weeks
MDI
No detailed dietary or exercise effort
HbA1c ↓ 0.8%
Δ = 0.3%
Low rate at baseline. No statistical difference No difference in QOL scales
Highly satisfied
Perceived less hassle
Best study, but low glucose variability and hypoglycemia at baseline
Telemedicine prospective and observational trials
 Majithia et al. (2020)25 Prospective
Single-arm study
N = 55
4 months
36% on insulin
Telemedicine visits with CGM
HbA1c ↓ 1.6%
TIR ↑10.2%
No change in glucose less than 70 mg/dL No QOL study
Weight ↓9 lbs
CGM facilitated glucose control in telehealth
 Dixon et al. (2020)26 Observational study N = 740
90–180 days
Telehealth with CGM support
31% on insulin
HbA1c ↓ 0.85% when starting HbA1c >7.0% NR NR Potential benefit of CGM in telehealth

TIR (70–180 mg/dL).

HbA1c, hemoglobin A1c; CGM, continuous glucose monitoring (refers to rtCGM in this study); MDI, multiple daily insulin dosing; NR, not reported; QOL, quality of life; RCT, randomized controlled trial; rtCGM, real-time CGM; SMBG, self-monitoring of blood glucose.

Effect of rtCGM on lifestyle modification

Lifestyle modification is recommended as the first step in the management of prediabetes and T2D.27 There is evidence in the literature demonstrating that rtCGM use contributes to patient education and assists with behavioral change, indicating it may be an effective teaching tool to modify lifestyle and improve glycemic control.28,29

In the first RCT using rtCGM (3 days at a time for 3 months), Yoo et al. randomized 65 patients with poorly controlled T2D on oral and/or insulin therapy to either rtCGM or SMBG. They demonstrated a 0.7% greater HbA1c reduction in the rtCGM intervention group compared to a group randomized to SMBG alone, as well as changes in weight, exercise, and postprandial glucose.21 Looking at rtCGM incorporation in diabetes education, Lee et al. sequentially enrolled T2D patients receiving initial diabetes education into groups of pattern management guided by Guardian rtCGM or standard diabetes education as a control.30 The participants were on a mix of therapies, including insulin and non-insulin medications. The rtCGM group showed improvement in classic lifestyle factors of diet, exercise, and self-management concepts. The overall self-care behavior score was significantly higher in the rtCGM group 3 and 6 months after the education was completed, showing a role for rtCGM in developing habits that are helpful for diabetes management in a standard T2D population.

Cox et al. created a model using education and lifestyle interventions to minimize glucose excursions and employed rtCGM (Dexcom Platinum G4) to educate patients on the effects of eating and activity choices previously demonstrated to reduce glycemic variability.22 They found not only an improvement in HbA1c but also a remarkable improvement in self-reported diabetes knowledge in those using rtCGM. Taylor et al. explored the educational potential of rtCGM in a pilot study where 20 obese patients with T2D were randomized to either professional CGM or rtCGM, in addition to standard lifestyle modification over 12 weeks.31 They saw reductions in HbA1c and body weight in both groups without any statistical difference between the two groups. However, authors reported a 40% greater reduction in blood glucose-lowering medication in the rtCGM group compared to the control group. In a systematic review by the same author that included 5542 participants from 11 studies (eight RCTs and three observational studies), CGM use (either professional or rtCGM) was associated with decreased body weight, decreased caloric intake, higher adherence to eating plans, and increased physical activity compared to SMBG.32

These studies demonstrate rtCGM has the potential to improve lifestyle changes and adherence to treatment in patients with T2D. However, most of the studies were of shorter duration with small sample sizes limiting the evidence of rtCGM as an adjunct to lifestyle management in patients with T2D.

Improvement in glucose control with rtCGM

In an early RCT examining CGM use by Garg et al., 91 insulin-requiring patients with diabetes (75 with T1D and 16 with T2D) were randomized to receive 3 days of rtCGM (STS, Dexcom) or masked CGM (control).33 When compared with control subjects, the CGM group spent 21% less time in hypoglycemia (<55 mg/dL), 23% less time in hyperglycemia (≥240 mg/dL), and 26% more time in the target (81–140 mg/dL) glucose range. This was the first study to demonstrate the feasibility and safety of rtCGM in insulin-treated diabetes. The data for those with T2D were not separately analyzed and thus, potential glycemic benefit in this group was not assessed. Similarly, in a small observational study of 140 patients with diabetes (109 with T1D, 24 with T2D on insulin, and 7 with T2D on non-insulin therapies), rtCGM (STS Dexcom) was associated with a significant reduction in HbA1c by 0.4% in both T1D and T2D over 3 months compared to baseline.34 Those individuals with T2D on non-insulin medications did not have a significant HbA1c reduction, probably due to small sample size. Similarly, in the GLADIS (Glucose Awareness in Diabetes Study) clinical trial, New et al. randomized 100 patients with insulin-treated diabetes (including 19 with T2D) to either rtCGM (Freestyle Navigator) or SMBG.35 With the small sample size, there was no improvement detected in HbA1c in the rtCGM group compared to SMBG.

An RCT by Ehrhardt et al. randomized 100 patients with T2D on various therapies, excluding prandial insulin to either usual care or four cycles of rtCGM (Dexcom SEVEN, 2 weeks on/1 week off) over 12 weeks.23 After 12 weeks, all subjects were followed by their primary care providers for an additional 40 weeks. There was significant improvement in HbA1c by 0.5% in the rtCGM group compared to control over 12 weeks, and the glycemic benefits seen with rtCGM use persisted over 40 weeks.36

Earlier systematic reviews and meta-analyses of CGM studies (including both rtCGM and professional CGM) suggested benefits of CGM in improving glycemic control, especially in insulin-treated patients with T2D.28,37 The DiaMonD (Multiple Daily Injections and CGM in Diabetes) trial was a larger study that randomized 158 individuals with T2D using multiple daily insulin (MDI) to either rtCGM (Dexcom G4 with 505 software) or SMBG with the goal of further evaluating the effectiveness of rtCGM in individuals with T2D.24 The study design included minimal patient contact, designed to replicate a real-world scenario. Despite this, there was significant reduction in HbA1c by 1% at 12 weeks and 0.8% at 24 weeks. The control and intervention groups both showed improvement in HbA1c at 12 weeks, but there remained a significant difference between the two, with the rtCGM group showing an additional 0.3% decrease (95% confidence interval 0.6%–0.1%, P = 0.005). Although this difference is of debatable clinical significance, there were also advantages in the proportion of subjects with a 10% reduction in HbA1c (57% vs. 35%, P = 0.016), proportion with ≥1% reduction in HbA1c (53% vs. 33%, P = 0.034), and proportion with ≥0.5% reduction HbA1c (79% vs. 51%, P = 0.002) in the rtCGM group compared to SMBG group. TIR was also higher in the rtCGM group.

A recent meta-analysis looked exclusively at RCTs specifying individuals with T2D and examined HbA1c as an outcome.38 They identified five studies that included a total of 382 patients and found a significant improvement in HbA1c (pooled mean difference of 0.25%) with the use of rtCGM compared to SMBG, providing additional evidence of HbA1c change with rtCGM use.

The role of rtCGM in telemedicine to improve HbA1c was studied by Majithia et al., who used Dexcom G6 sensors as part of a clinical model to evaluate patient response to medical interventions. HbA1c from an initial 10-day run-in period was compared to HbA1c in a 10-day period later in the study. While there was no separate control group in this study, their interventions showed an improvement of 1.6% in HbA1c, from a mean of 8.9% at baseline to 7.3% at follow-up (P < 0.001).25

Dixon et al. studied a telehealth diabetes clinic model utilizing rtCGM (Dexcom G5 or G6) in 740 participants enrolled across 21 states, with groups stratified by starting HbA1c. All groups with HbA1c >7% saw a significant change ranging from 0.2% ± 0.8% with initial HbA1c values 7.0%–7.9% to 2.3% ± 1.0% for those with baseline Hba1c >9.0% (P < 0.001), while those with HbA1c <7% maintained this level of glycemic control.26 In conclusion, there is mounting evidence that rtCGM can improve glycemic control in insulin-treated patients with T2D, in addition to other benefits.

Effect of rtCGM on hypoglycemia

Optimal glycemic control is necessary to prevent microvascular complications in patients with diabetes. Hypoglycemia, however, remains a major obstacle to achieving this. Severe hypoglycemia is a serious acute complication of diabetes that affects many more patients with T2D than we generally recognize.39–42 Patients with elevated blood glucose are not protected from hypoglycemia, as studies seeking to identify which patients have elevated risk of hypoglycemia found that blood glucose variability was a strong predictor.43,44 In a study of 108 patients with T2D on a mix of insulin and non-insulin agents, Gehlaut et al. found that 49% of individuals experienced hypoglycemia at least once during a 5-week period of professional CGM monitoring (Medtronic iPro), and that many of these episodes were asymptomatic.40 In addition, 21% of patients had blood glucose levels of 50 mg/dL or lower. Several studies have suggested that SMBG underreports hypoglycemia in patients with T2D, and it is thought that CGM would more effectively detect hypoglycemia.45–47

These studies shine light on the important and perhaps underrecognized issue of hypoglycemia in patients with T2D. The main question, however, is whether rtCGM is effective in reducing hypoglycemia in this population.

In the largest prospective trial of subjects with T2D treated with MDI, Beck et al. did not find a statistically significant reduction in hypoglycemia less than 70 mg/dL with use of rtCGM, although with a rate of only 11 min per day of hypoglycemia, the study was not powered to detect a difference.24 There was a nonsignificant reduction in hypoglycemia from a median of 11 minutes per day to 4 minutes per day in the rtCGM group, while the SMBG group experienced a median of 12 minutes per day in hypoglycemia both at baseline and at the 24-week mark.

Despite the very high prevalence of hypoglycemia in patients with T2D, there is no RCT using rtCGM with the primary objective of hypoglycemia reduction. T2D patients with hypoglycemia unawareness or with a greater percentage of hypoglycemia at baseline are likely to benefit from rtCGM; however, further dedicated studies are necessary to establish evidence of rtCGM utility in hypoglycemia prevention or reduction.

Glucose variability outcomes with rtCGM

HbA1c has been a longstanding measure of adequate glycemic control in individuals with T2D, although it has a known limitation of incompletely describing glycemic variability. There is growing evidence that glycemic variability predicts chronic complications. Cross-sectional studies have found associations between glycemic variability and microvascular complications such as diabetic retinopathy, painful diabetic neuropathy, and cardiovascular autonomic neuropathy in patients with T2D.24,48–50 Moreover, the need for cardiac revascularization in patients observed following STEMI was shown to be greater in those with higher glycemic variability.51

Although studies of rtCGM in patients with T1D have demonstrated reduced variability, it has not been adequately studied in T2D.10,52 In a clinical trial that used glucose variability as a secondary outcome, the baseline rate of variability was quite low (31%) and no significant difference was noted compared to SMBG.53 At present, it has not been clearly demonstrated that reducing glycemic variability would reduce diabetes complications or cardiovascular events.

Patient satisfaction and quality of life with rtCGM

T2D is associated with reduced health-related quality of life.54 The high work load of SMBG and predicting blood glucose levels are a frequent cause of concern. A survey-based study by Runge et al. found that wearing an rtCGM and the associated increased awareness of TIR reduced stress experienced by those with diabetes.55 Vigersky et al. found no difference in diabetes distress in their study on intermittent CGM use, although the participants in the rtCGM group were requested to perform fingersticks at the same frequency as the SMBG group, negating any difference in work load.36 In a study by Beck et al., the use of rtCGM (Dexcom G4 with 505 software) in T2D patients on MDI showed a high level of satisfaction with CGM (4.3 with a scale of 1–5) and a low “hassle” subscore.24 Other evaluations of patient-reported outcomes have found high levels of use, good satisfaction, and a better understanding of their glucose patterns, among other benefits.56,57

Quality of life remains a measure that is less frequently reported. There is a logical trend toward improved quality of life in individuals who have more health knowledge and less hassle, but currently available evidence is insufficient to fully understand the effect of rtCGM on patient satisfaction and quality of life.

Recommendation for the use of rtCGM in T2D by the professional societies

All three major U.S. diabetes professional societies have provided evidence-based guidance on the use of CGM for patients with T2D.58–61 All professional organizations have recommended CGM use in selected patients with T2D to improve glycemic control or reduce hypoglycemia, while acknowledging limited and low-quality evidence, especially non-insulin-treated T2D. A summary of their recommendations is provided in Table 3.

Table 3.

Professional Society Recommendations for Continuous Glucose Monitoring Use in the Management of Type 2 Diabetes

Professional society (reference) Recommendations
ADA55 When used properly, real-time continuous glucose monitors in conjunction with multiple daily injections and continuous subcutaneous insulin infusion [A], and other forms of insulin therapy [C] are a useful tool to lower and/or maintain A1C levels and/or reduce hypoglycemia in adults and youth with diabetes.
Use of professional CGM and/or intermittent real-time or intermittently scanned CGM can be helpful in identifying and correcting patterns of hyperglycemia and hypoglycemia, and improving A1C levels in people with diabetes on noninsulin, as well as basal insulin regimens. [C]
AACE56 CGM devices should be considered for patients with T1D and T2D who are on intensive insulin therapy to improve HbA1c levels and reduce hypoglycemia (Grade B), early reports suggest that even patients not taking insulin may benefit from CGM (Grade D).
The Endocrine Society57,58 We suggest short-term, intermittent rtCGM use in adult patients with T2DM (not on prandial insulin), who have A1c levels >7% and are willing and able to use the device. (2|⊕⊕○○)

ADA level A evidence—high-level, clear evidence from well conducted, generalizable RCT, ADA level C evidence—supportive evidence from well-conducted studies. AACE grade B evidence is intermediate level, while D means not evidence based. Endocrine society level of evidence 2|⊕⊕ means weak, low-quality evidence.

AACE, American Association of Clinical Endocrinologists; ADA, American Diabetes Association; T1D, type 1 diabetes.

Other considerations and future directions

The use of rtCGM for those with T2D will continue to evolve. The trials done thus far are small, but they add valuable information, showing a clear potential benefit of using rtCGM to more effectively educate patients and assist them in improving their diabetes management. Practitioners involved in providing CGM to patients will seek more guidance than is presently available to make educated choices about which patients will benefit most from this technology, and how to optimize its use in those individuals. Although an early estimate of cost-effectiveness has been favorable at $33,039 per quality-adjusted life year, insurers are likely to seek more outcomes data and reduced cost to apply to the widest spectrum of patients with T2D.62 However, cost-effectiveness may be demonstrated for a subset of the T2D population, particularly if they have significant hypoglycemia or high glycemic variability. So far, no study in T2D has selected patients with documented hypoglycemic unawareness or a history of severe hypoglycemic episodes. It will also be important to study patients with advanced age and with comorbidities that may increase their risk of hypoglycemia.

Telemedicine has greatly expanded during the COVID-19 pandemic, and the role of CGM in communicating patient data to providers has been crucial. This will likely remain an important tool in maximizing patient access options and improving communication with providers.25,26,56

Another developing application of rtCGM relevant to both T1D and T2D is use in the inpatient setting, where frequent monitoring with remote display may show significant advantages in nursing demands, more dynamic therapeutic adjustments, and reduced hypoglycemia.63–67 This use requires further research, including cost-benefit analysis, but holds great promise. Researchers and manufacturers have made efforts to improve the human interface in CGM technology. However, future studies should include more patient education on how to most effectively use the information provided by CGM to modify behavior, empowering patients to optimize their self-management skills.

Conclusions

The development of CGM has had a profound impact on the field of diabetes. There is real potential for rtCGM to improve glycemic outcomes and prevent diabetes complications in patients with T2D. Current evidence is promising, but the majority of trials looking at the role of CGM with T2D management have selected those on intensive insulin therapies, leaving out a large number of T2D patients. Studies have suggested a role for rtCGM as a teaching tool to improve lifestyle management, which would be helpful even in early T2D or prediabetes. These studies have been small, however, and their data often inconclusive. This exciting new field would benefit from large RCTs confirming usefulness of rtCGM for lifestyle improvement and diabetes self-management, which remain the mainstay of diabetes care.

Author Disclosure Statement

M.A.J. reports no conflict of interests. A.A. receives research support from Dexcom and is on the advisory board for Medtronic. V.N.S. reports receiving research grants through University of Colorado from Sanofi, Eli-Lilly, NovoNordisk, vTv Therapeutics, Mylan GmbH, Dexcom, and Insulet. V.N.S. had served on advisory board for Sanofi and Medscape LLC.

Funding Information

No funding was received for this article.

References

  • 1. Polonsky WH, Fisher L, Hessler D, Edelman SV: A survey of blood glucose monitoring in patients with type 2 diabetes: are recommendations from health care professionals being followed? Curr Med Res Opin 2011;27 Suppl 3:31–37 [DOI] [PubMed] [Google Scholar]
  • 2. Rossi MC, Lucisano G, Ceriello A, et al. : Real-world use of self-monitoring of blood glucose in people with type 2 diabetes: an urgent need for improvement. Acta Diabetol 2018;55:1059–1066 [DOI] [PubMed] [Google Scholar]
  • 3. American Diabetes Association: Glycemic targets: standards of medical care in diabetes-2021. Diabetes Care 2021;44:S73–S84 [DOI] [PubMed] [Google Scholar]
  • 4. Battelino T, Danne T, Bergenstal RM, et al. : Clinical targets for continuous glucose monitoring data interpretation: recommendations from the International Consensus on Time in Range. Diabetes Care 2019;42:1593–1603 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5. U.K. Prospective Diabetes Study Group: Overview of 6 years' therapy of type II diabetes: a progressive disease. Diabetes 1995;44:1249–1258 [PubMed] [Google Scholar]
  • 6. Shah VN, Moser EG, Blau A, et al. : The future of basal insulin. Diabetes Technol Ther 2013;15:727–732 [DOI] [PubMed] [Google Scholar]
  • 7. Shah RB, Patel M, Maahs DM, Shah VN: Insulin delivery methods: past, present and future. Int J Pharm Investig 2016;6:1–9 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8. Fang M: Trends in diabetes management among US adults: 1999–2016. J Gen Intern Med 2020;35:1427–1434 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9. Juvenile Diabetes Research Foundation Continuous Glucose Monitoring Study Group, Tamborlane WV, Beck RW, et al. : Continuous glucose monitoring and intensive treatment of type 1 diabetes. N Engl J Med 2008;359:1464–1476 [DOI] [PubMed] [Google Scholar]
  • 10. Beck RW, Riddlesworth T, Ruedy K, et al. : Effect of continuous glucose monitoring on glycemic control in adults with type 1 diabetes using insulin injections: the DIAMOND randomized clinical trial. JAMA 2017;317:371–378 [DOI] [PubMed] [Google Scholar]
  • 11. Lind M, Polonsky W, Hirsch IB, et al. : Continuous glucose monitoring vs conventional therapy for glycemic control in adults with type 1 diabetes treated with multiple daily insulin injections: the GOLD randomized clinical trial. JAMA 2017;317:379–387 [DOI] [PubMed] [Google Scholar]
  • 12. Pratley RE, Kanapka LG, Rickels MR, et al. : Effect of continuous glucose monitoring on hypoglycemia in older adults with type 1 diabetes: a randomized clinical trial. JAMA 2020;323:2397–2406 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13. Aleppo G, Ruedy KJ, Riddlesworth TD, et al. : REPLACE-BG: a randomized trial comparing continuous glucose monitoring with and without routine blood glucose monitoring in adults with well-controlled type 1 diabetes. Diabetes Care 2017;40:538–545 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 14. Danne T, Nimri R, Battelino T, et al. : International consensus on use of continuous glucose monitoring. Diabetes Care 2017;40:1631–1640 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15. Heinemann L, Freckmann G, Ehrmann D, et al. : Real-time continuous glucose monitoring in adults with type 1 diabetes and impaired hypoglycaemia awareness or severe hypoglycaemia treated with multiple daily insulin injections (HypoDE): a multicentre, randomised controlled trial. Lancet 2018;391:1367–1377 [DOI] [PubMed] [Google Scholar]
  • 16. Mulinacci G, Alonso GT, Snell-Bergeon JK, Shah VN: Glycemic outcomes with early initiation of continuous glucose monitoring system in recently diagnosed patients with type 1 diabetes. Diabetes Technol Ther 2019;21:6–10 [DOI] [PubMed] [Google Scholar]
  • 17. Carlson AL, Mullen DM, Bergenstal RM: Clinical use of continuous glucose monitoring in adults with type 2 diabetes. Diabetes Technol Ther 2017;19:S4–S11 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18. Vigersky R, Shrivastav M: Role of continuous glucose monitoring for type 2 in diabetes management and research. J Diabetes Complications 2017;31:280–287 [DOI] [PubMed] [Google Scholar]
  • 19. Wang L, Liu X, Yang W, et al. : Comparison of blood glucose variability between exenatide and biphasic insulin aspart 30 in Chinese participants with type 2 diabetes inadequately controlled with metformin monotherapy: a multicenter, open-label, randomized trial. Diabetes Ther 2020;11:2313–2328 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 20. FLAT-SUGAR Trial Investigators: Glucose variability in a 26-week randomized comparison of mealtime treatment with rapid-acting insulin versus GLP-1 agonist in participants with type 2 diabetes at high cardiovascular risk. Diabetes Care 2016;39:973–981 [DOI] [PubMed] [Google Scholar]
  • 21. Yoo HJ, An HG, Park SY, et al. : Use of a real time continuous glucose monitoring system as a motivational device for poorly controlled type 2 diabetes. Diabetes Res Clin Pract 2008;82:73–79 [DOI] [PubMed] [Google Scholar]
  • 22. Cox DJ, Banton T, Moncrief M, et al. : Minimizing glucose excursions (GEM) with continuous glucose monitoring in type 2 diabetes: a randomized clinical trial. J Endocr Soc 2020;4:1–7 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 23. Ehrhardt NM, Chellappa M, Walker MS, et al. : The effect of real-time continuous glucose monitoring on glycemic control in patients with type 2 diabetes mellitus. J Diabetes Sci Technol 2011;5:668–675 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 24. Beck RW, Riddlesworth TD, Ruedy K, et al. : Continuous glucose monitoring versus usual care in patients with type 2 diabetes receiving multiple daily insulin injections: a randomized trial. Ann Intern Med 2017;167:365–374 [DOI] [PubMed] [Google Scholar]
  • 25. Majithia AR, Kusiak CM, Armento Lee A, et al. : Glycemic outcomes in adults with type 2 diabetes participating in a continuous glucose monitor-driven virtual diabetes clinic: prospective trial. J Med Internet Res 2020;22:e21778. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 26. Dixon RF, Zisser H, Layne JE, et al. : A virtual type 2 diabetes clinic using continuous glucose monitoring and endocrinology visits. J Diabetes Sci Technol 2020;14:908–911 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 27. American Diabetes Association: Facilitating behavior change and well-being to improve health outcomes: standards of medical care in diabetes-2021. Diabetes Care 2021;44:S53–S72 [DOI] [PubMed] [Google Scholar]
  • 28. Ehrhardt N, Al Zaghal E: Behavior modification in prediabetes and diabetes: potential use of real-time continuous glucose monitoring. J Diabetes Sci Technol 2019;13:271–275 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 29. Maurizi AR, Pozzilli P: Do we need continuous glucose monitoring in type 2 diabetes? Diabetes Metab Res Rev 2013. [Epub ahead of print]; DOI: 10.1002/dmrr.2450 [DOI] [PubMed] [Google Scholar]
  • 30. Lee SK, Shin DH, Kim YH, Lee KS: Effect of diabetes education through pattern management on self-care and self-efficacy in patients with type 2 diabetes. Int J Environ Res Public Health 2019;16:3323. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 31. Taylor PJ, Thompson CH, Luscombe-Marsh ND, et al. : Efficacy of real-time continuous glucose monitoring to improve effects of a prescriptive lifestyle intervention in type 2 diabetes: a pilot study. Diabetes Ther 2019;10:509–522 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 32. Taylor PJ, Thompson CH, Brinkworth GD: Effectiveness and acceptability of continuous glucose monitoring for type 2 diabetes management: a narrative review. J Diabetes Investig 2018;9:713–725 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 33. Garg S, Zisser H, Schwartz S, et al. : Improvement in glycemic excursions with a transcutaneous, real-time continuous glucose sensor: a randomized controlled trial. Diabetes Care 2006;29:44–50 [DOI] [PubMed] [Google Scholar]
  • 34. Bailey TS, Zisser HC, Garg SK: Reduction in hemoglobin A1C with real-time continuous glucose monitoring: results from a 12-week observational study. Diabetes Technol Ther 2007;9:203–210 [DOI] [PubMed] [Google Scholar]
  • 35. New JP, Ajjan R, Pfeiffer AF, Freckmann G: Continuous glucose monitoring in people with diabetes: the randomized controlled Glucose Level Awareness in Diabetes Study (GLADIS). Diabet Med 2015;32:609–617 [DOI] [PubMed] [Google Scholar]
  • 36. Vigersky RA, Fonda SJ, Chellappa M, et al. : Short- and long-term effects of real-time continuous glucose monitoring in patients with type 2 diabetes. Diabetes Care 2012;35:32–38 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 37. Poolsup N, Suksomboon N, Kyaw AM: Systematic review and meta-analysis of the effectiveness of continuous glucose monitoring (CGM) on glucose control in diabetes. Diabetol Metab Syndr 2013;5:39. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 38. Janapala RN, Jayaraj JS, Fathima N, et al. : Continuous glucose monitoring versus self-monitoring of blood glucose in type 2 diabetes mellitus: a systematic review with meta-analysis. Cureus 2019;11:e5634. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 39. Lipska KJ, Yao X, Herrin J, et al. : Trends in drug utilization, glycemic control, and rates of severe hypoglycemia, 2006–2013. Diabetes Care 2017;40:468–475 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 40. Gehlaut RR, Dogbey GY, Schwartz FL, et al. : Hypoglycemia in type 2 diabetes—more common than you think: a continuous glucose monitoring study. J Diabetes Sci Technol 2015;9:999–1005 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 41. Heller SR, Peyrot M, Oates SK, Taylor AD: Hypoglycemia in patient with type 2 diabetes treated with insulin: it can happen. BMJ Open Diabetes Res Care 2020;8:e001194. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 42. Zoungas S, Patel A, Chalmers J, et al. : Severe hypoglycemia and risks of vascular events and death. N Engl J Med 2010;363:1410–1418 [DOI] [PubMed] [Google Scholar]
  • 43. Torimoto K, Okada Y, Hajime M, et al. : Risk factors of hypoglycemia in patients with type 2 diabetes mellitus: a study based on continuous glucose monitoring. Diabetes Technol Ther 2018;20:603–612 [DOI] [PubMed] [Google Scholar]
  • 44. Lipska KJ, Warton EM, Huang ES, et al. : HbA1c and risk of severe hypoglycemia in type 2 diabetes: the Diabetes and Aging Study. Diabetes Care 2013;36:3535–3542 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 45. Zick R, Petersen B, Richter M, et al. : Comparison of continuous blood glucose measurement with conventional documentation of hypoglycemia in patients with type 2 diabetes on multiple daily insulin injection therapy. Diabetes Technol Ther 2007;9:483–492 [DOI] [PubMed] [Google Scholar]
  • 46. Pazos-Couselo M, Garcia-Lopez JM, Gonzalez-Rodriguez M, et al. : High incidence of hypoglycemia in stable insulin-treated type 2 diabetes mellitus: continuous glucose monitoring vs. self-monitored blood glucose. Observational prospective study. Can J Diabetes 2015;39:428–433 [DOI] [PubMed] [Google Scholar]
  • 47. Klimontov VV, Myakina NE: Glucose variability indices predict the episodes of nocturnal hypoglycemia in elderly type 2 diabetic patients treated with insulin. Diabetes Metab Syndr 2017;11:119–124 [DOI] [PubMed] [Google Scholar]
  • 48. Guo QY, Lu B, Guo ZH, et al. : Continuous glucose monitoring defined time-in-range is associated with sudomotor dysfunction in type 2 diabetes. World J Diabetes 2020;11:489–500 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 49. Xu W, Zhu Y, Yang X, et al. : Glycemic variability is an important risk factor for cardiovascular autonomic neuropathy in newly diagnosed type 2 diabetic patients. Int J Cardiol 2016;215:263–268 [DOI] [PubMed] [Google Scholar]
  • 50. Jun JE, Jin SM, Baek J, et al. : The association between glycemic variability and diabetic cardiovascular autonomic neuropathy in patients with type 2 diabetes. Cardiovasc Diabetol 2015;14:70. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 51. Zhang JW, He LJ, Cao SJ, et al. : Effect of glycemic variability on short term prognosis in acute myocardial infarction subjects undergoing primary percutaneous coronary interventions. Diabetol Metab Syndr 2014;6:76. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 52. Olafsdottir AF, Polonsky W, Bolinder J, et al. : A randomized clinical trial of the effect of continuous glucose monitoring on nocturnal hypoglycemia, daytime hypoglycemia, glycemic variability, and hypoglycemia confidence in persons with type 1 diabetes treated with multiple daily insulin injections (GOLD-3). Diabetes Technol Ther 2018;20:274–284 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 53. Beck RW, Riddlesworth TD, Ruedy K, et al. : Continuous glucose monitoring versus usual care in patients with type 2 diabetes receiving multiple daily insulin injections: a randomized trial. Ann Intern Med 2017;167:365–374 [DOI] [PubMed] [Google Scholar]
  • 54. Trikkalinou A, Papazafiropoulou AK, Melidonis A: Type 2 diabetes and quality of life. World J Diabetes 2017;8:120–129 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 55. Runge AS, Kennedy L, Brown AS, et al. : Does time-in-range matter? Perspectives from people with diabetes on the success of current therapies and the drivers of improved outcomes. Clin Diabetes 2018;36:112–119 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 56. Bergenstal RM, Layne JE, Zisser H, et al. : Remote application and use of real-time continuous glucose monitoring by adults with type 2 diabetes in a virtual diabetes clinic. Diabetes Technol Ther 2021;23:128–132 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 57. Ida S, Kaneko R, Murata K: Utility of real-time and retrospective continuous glucose monitoring in patients with type 2 diabetes mellitus: a meta-analysis of randomized controlled trials. J Diabetes Res 2019;2019:4684815. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 58. American Diabetes Association. Diabetes Technology: Standards of medical care in diabetes-2021. Diabetes Care 2021;44:S85–S99 [DOI] [PubMed] [Google Scholar]
  • 59. Handelsman Y, Bloomgarden ZT, Grunberger G, et al. : American Association of Clinical Endocrinologists and American College of Endocrinology—Clinical Practice Guidelines for Developing a Diabetes Mellitus Comprehensive Care Plan—2015—Executive Summary. Endocr Pract 2015;21:413–437 [PubMed] [Google Scholar]
  • 60. Klonoff DC, Buckingham B, Christiansen JS, et al. : Continuous glucose monitoring: an Endocrine Society Clinical Practice Guideline. J Clin Endocrinol Metab 2011;96:2968–2979 [DOI] [PubMed] [Google Scholar]
  • 61. Fonseca VA, Grunberger G, Anhalt H, et al. : Continuous glucose monitoring: a Consensus Conference of the American Association of Clinical Endocrinologists and American College of Endocrinology. Endocr Pract 2016;22:1008–1021 [DOI] [PubMed] [Google Scholar]
  • 62. Tsuji S, Ishikawa T, Morii Y, et al. : Cost-effectiveness of a continuous glucose monitoring mobile app for patients with type 2 diabetes mellitus: analysis simulation. J Med Internet Res 2020;22:e16053. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 63. Davis GM, Galindo RJ, Migdal AL, Umpierrez GE: Diabetes technology in the inpatient setting for management of hyperglycemia. Endocrinol Metab Clin North Am 2020;49:79–93 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 64. Sadhu AR, Serrano IA, Xu J, et al. : Continuous glucose monitoring in critically ill patients with COVID-19: results of an Emergent Pilot Study. J Diabetes Sci Technol 2020;14:1065–1073 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 65. Spanakis EK, Levitt DL, Siddiqui T, et al. : The effect of continuous glucose monitoring in preventing inpatient hypoglycemia in general wards: the Glucose Telemetry System. J Diabetes Sci Technol 2018;12:20–25 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 66. Fortmann AL, Spierling Bagsic SR, Talavera L, et al. : Glucose as the fifth vital sign: a randomized controlled trial of continuous glucose monitoring in a non-ICU hospital setting. Diabetes Care 2020;43:2873–2877 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 67. Singh LG, Satyarengga M, Marcano I, et al. : Reducing inpatient hypoglycemia in the general wards using real-time continuous glucose monitoring: the Glucose Telemetry System, a randomized clinical trial. Diabetes Care 2020;43:2736–2743 [DOI] [PMC free article] [PubMed] [Google Scholar]

Articles from Diabetes Technology & Therapeutics are provided here courtesy of Mary Ann Liebert, Inc.

RESOURCES