Skip to main content
NIHPA Author Manuscripts logoLink to NIHPA Author Manuscripts
. Author manuscript; available in PMC: 2021 Mar 19.
Published in final edited form as: Int J Environ Anal Chem. 2015 Dec 21;96(1):15–37. doi: 10.1080/03067319.2015.1114104

Evaluation of portable XRF instrumentation for assessing potential environmental exposure to toxic elements

Kathryn G McIntosh 1,2,a, Diana Guimarães 1,2, Matthew J Cusack 3,b, Alexei Vershinin 3,c, ZW Chen 3, Karl Yang 1,2, Patrick J Parsons 1,2,*
PMCID: PMC7978405  NIHMSID: NIHMS794405  PMID: 33746339

Abstract

Portable instruments based on X-Ray Fluorescence Spectrometry (XRF) have the potential to assist in field-based studies provided the data produced are reliable. In this study, we evaluate the performance of two different types of XRF instrument (XOS prototype, and Thermo Niton XL3t). These two XRF analyzers were evaluated in a laboratory setting, and data were reported for 17 elements (As, Ba, Cd, Co, Cr, Cu, Fe, Hg, Mn, Ni, Pb, Se, Sn, Sr, Ti, V, and Zn). Samples analyzed (n=38) included ethnic herbal medicine products (HMP), ethnic spices (ES), and cosmetic products (CP). Comparison analyses were carried out using Inductively Coupled Plasma Optical Emission Spectrometry (ICP-OES). In general, results reported for Cd, Cu and Pb by the XOS prototype analyzer, and based on the instrument’s non-metal mode, were negatively biased (5 % to 95 %) compared to ICP-OES. In contrast, results reported for Pb, As, Cd, Cu and Zn by the Niton, based on using the soil mode, were positively biased, in some instances (Cd) by up to 4 orders of magnitude. While the sensitivity of both instruments was insufficient for reliably “quantifying” toxic elements below 15 mg/kg, XRF was still capable of positively “detecting” many elements at the low single digit mg/kg levels. However, for semi-quantification estimates of contaminants at higher levels, and with limited sample preparation, both XRF instruments were deemed fit for the purpose. This study demonstrates that modern XRF instrumentation is valuable for characterizing the elemental content of food, cosmetic, and medicinal products. The technology is particularly useful for rapidly screening large numbers of products (100s per day) in the field, and quickly identifying those that may contain potentially hazardous levels of toxic elements. Toxic elements can be confirmed by examining the raw spectrum, and the limitations of factory-based calibration are generally manageable for field-based studies.

Keywords: X-Ray Fluorescence, portable instrument, toxic metals, herbal medicine products, cosmetic products, ethnic spices, consumer products screening

1. Introduction

Public health agencies traditionally rely on an analytical laboratory to support their investigations of human exposure to heavy metals. Such laboratories are typically accredited to national or international standards, and use certified methods of analysis that require a high level of skill and training. For the major toxic elements Pb, Cd, Hg and As, approved (or official) methods typically mean those based on Inductively Coupled Plasma Mass Spectrometry (ICP-MS), ICP-Optical Emission Spectrometry (OES) and/or Atomic Absorption Spectrometry (AAS). These “destructive” methods require that the sample be usually digested prior to analysis, and are both time-consuming, relatively expensive, and required significant technical skills.

For public health investigations, the advantages of non-destructive, portable technologies are clearly obvious, especially for field-based investigations where the turnaround time and low-cost per analysis is important. For example, use of Ayurvedic medicines among immigrants from the Indian sub-continent, and the wide spread use of traditional herbal medicine products (HMP) among Chinese communities, represent a growing concern from a public health perspective. The World Health Organization (WHO) estimates that 70–80% of the population in developed countries has utilized some form of alternative or complementary treatment, while up to 80% of the population in some African and Asian countries rely on traditional medicine for primary health care [1].

In the United States, HMPs are widely available and while some may be regulated as drugs, others may be considered dietary supplements. Furthermore, the use of alternative therapies is often unsupervised by either a medical doctor or a practitioner of alternative medicine [2]. While HMPs may be considered as ‘natural’ and therefore they may be interpreted as having low risk of harm, the reality is that little is known about the chemical content of HMPs. Potential safety concerns related to the use of HMPs include undeclared ingredients such as prescription drugs, interactions with conventional medicines, and contamination with heavy metals [3,4]. Indeed, some studies have reported detecting one or more toxic metals or metalloids up to weight percent levels in some Ayurvedic products [5]. Metals may be intentionally incorporated into traditional medicines, such as in the rasa shastra practice of Ayurveda, where the “equilibrium” of elements such as Hg, Pb, and Zn is considered essential for the normal functioning of the human body. However, potential sources of contamination for other HMPs include use of agricultural pesticides (e.g., lead arsenate), growing plant raw materials in polluted soil, and poor manufacturing controls such as use of metal grinding apparatus or storage container [6,7]. HMPs and other ethnic products may constitute a significant and preventable source of exposure to toxic elements, so characterizing their elemental content is important for monitoring the human environment and in protecting public health.

Recent studies have reported on specific elements in HMPs using methods such as AAS or ICP-OES [712] while others have reported data based on XRF methods [1316]. The XRF-based approaches are generally non-destructive, rapid, and require limited sample preparation compared with AAS or ICP-OES, and they are also relatively inexpensive. In addition, the availability of portable XRF instrumentation offers the potential for rapid trace element determination in HMPs and similar products under field conditions [17,18] such as in a subject’s home or at a port of entry. Thus, XRF has the potential to facilitate rapid identification of potentially toxic products. However, correct use of portable XRF instrumentation requires some understanding of X-Ray Fluorescence Spectrometry, the sample being analyzed, the element being measured and, more importantly, the limitations of the technology. Comprehensive reviews of XRF and its application to food, biological and commercial product samples are available from Atomic Spectrometry Updates, and are published annually [19].

The development of new technologies that can be used to assess potential personal environmental exposure either at home or in the work place creates new possibilities for exposure assessment studies. Such efforts have been supported in the U.S. by the National Institutes of Health (NIH) as part of the agency’s The Genes, Environment and Health Initiative (GEI) [20]. In particular, the NIH effort is focused on exposure biology, and the need to develop new technologies for assessing environmental exposure to individuals [21].

In response to these needs, new XRF instrumentation based on doubly curved crystal (DCC) enabled monochromatic X-Ray Optic Technology (M-XRF) was developed by X-Ray Optical Systems, Inc. (XOS, East Greenbush) with NIH support. According to XOS, DCC optics have the potential to increase the sensitivity of XRF and improve performance at low levels [22]. With NIH support, a prototype M-XRF instrument, which was initially called the “Personal Environmental Analyzer” (PEA), was assembled by XOS. The prototype XOS analyzer was designed to measure a broad range of elements in liquids, gels, or solid sample types, including HMPs, personal care products, food, beverages, etc. It was envisioned that the analyzer would enable individual environmental exposure assessments to be conducted in the home, place of employment, or in a recreational setting with little or no sample preparation required.

The principal goal of this study was to evaluate the analytical performance of the XOS prototype analyzer in the laboratory, including its accuracy, precision, limits of detection, and possible interferences, and to report on the feasibility for field-based studies. The XOS prototype analyzer was compared to a commercial hand-held XRF instrument, the Niton XL3t (Thermo Scientific, Tewksbury, MA), which is also used for elemental screening of samples collected from the personal environment. Analytical performance for both instruments was assessed using HMPs, ES, and CP for up to 17 elements including: As, Ba, Cd, Cr, Co, Cu, Fe, Hg, Mn, Ni, Pb, Se, Sr, Sn, Ti, V, Zn. The XRF data, obtained using factory-based algorithms and/or specific calibration modes, were compared to independent values obtained by ICP-OES following acid digestion.

2. Experimental

2.1. Certified reference materials

A variety of certified reference materials (CRM) were analyzed on the XOS prototype analyzer and on the Niton. The CRMs were selected based on specific elements present, with a certified elemental content >1 mg/kg. An effort was made to obtain CRMs with a variety of matrices that might be encountered in the personal environment. CRMs included National Institute of Standards and Technology (NIST) Standard Reference Material (SRM) 1547 Peach Leaves (NIST, Gaithersburg, MD), Institute of Reference Materials and Metrology (IRMM, Geel, Belgium) BCR 627 Tuna Fish Tissue, National Research Council (NRC, Ottawa, Canada) DORM-2 Dogfish Muscle, SRM 612 Trace Elements in Glass and SRM 2976 Mussel Tissue. All measurements were completed in triplicate, and sample cups were removed from the instrument and repositioned between each measurement. Concerning the ICP-OES comparison, SRM 2976 Mussel Tissue and ERA Metals in Soil D069–540 (Golden, CO) were used to evaluate the recoveries for the method.

2.2. Archived laboratory samples

A broad selection of archived samples that had been collected previously for various public health investigations were used for evaluation purposes in this study. Many of the HMPs (n=22) and CP (n=12) were provided by the New York City Department of Health and Mental Hygiene’s (NYC DOHMH) Division of Environmental Health, and had been the subject of investigations by that agency. Most of the samples had been confiscated during retail inspections and/or as a result of product recalls conducted by NYC DOHMH (Dr. Paromita Hore, NYC DOHMH, personal communication). Nearly all of the products had been imported, and several had been associated with childhood lead poisoning cases in New York City. Foodstuff, including imported spices and tea samples (ES, n=4), were also analyzed. Archived samples had been stored at room temperature in their original packaging.

2.3. XOS Prototype Environmental Analyzer

An optimized prototype instrument was assembled by X-Ray Optical Systems, Inc., (XOS, East Greenbush, NY) specifically for use in this study (Fig. 1a). The prototype analyzer was developed from earlier iterations that demonstrated the feasibility of the approach. The prototype analyzer is equipped with a 50-W molybdenum (Mo) anode X-Ray tube as the excitation source, a shutter and three multi-segment DCC optics, designed to generate monochromatic excitation energies at 7, 17.5, and 31 keV during a pre-programmed measurement sequence. The diameter of the beam spot at the measurement site is ~1 mm. A silicon drift detector (SDD; Vortex-90EX, SII Nano Technology, Northridge, CA) is utilized for energy dispersive detection of fluorescent X-Rays. The XOS prototype instrument bench top unit evaluated in this study weighs ~20 kg and measures 38 (W) x 45 (D) x 45 (H) cm (Fig. 1b).

Figure 1.

Figure 1.

(a) Optical configuration of the XOS prototype analyzer (b) image of the XOS prototype instrument – computer monitor is also attached to the instrument (c) Elements measurable by the respective optics of the instrument.

For field use, the instrument is transported using a collapsible cart. It is equipped with a safety interlock and is fully shielded. Nonetheless, staff wore personal dosimeters to monitor their exposure to ionizing radiation during all XRF operations. Instrument operation and data reporting are controlled with an integrated touch screen; a separate monitor was used for laboratory-based studies. Quantitation is accomplished via XOS’ proprietary software based on fundamental parameters (FP), which reports results for more than 30 elements in parts-per-million (ppm) units, i.e., mg/kg or mg/L. User selectable parameters include substrate (metal, non-metal, or liquid) and excitation type (tri-chromatic through a combination of three separate optics and a 50-W Mo source). Although the prototype instrument was equipped with a tri-chromatic DCC excitation optic, only the 17.5 and 31 keV excitation energies were utilized for this study (the 7 keV beam was not necessary for this application). Fig. 1c shows those elements that can be fluoresced by each of the three optics. For public health purposes and this study, only the medium and high-energy optics were needed to capture the major toxic elements of interest: Pb, Cd, Hg and As. Each sample measurement was made using the “non-metal” substrate mode. The analyzer was operated with a 3.5-minute count time, which was fixed by the manufacturer for this particular prototype instrument.

A standard 32-mm diameter x 24 mm height XRF sample cup (Premier Lab Supply, Port St. Lucie, FL), was used with a 4-μm thick Ultralene window (Chemplex Industries, Palm City, FL). The polyethylene cup was filled to approximately 1/2 full, typically 2 to 6 grams of sample, equivalent to ~1 cm, which ensured infinite path thickness. The XRF cups were sealed with end-caps, and prepared samples were utilized repeatedly throughout the study. Aqueous standards were prepared by diluting 10,000 μg/mL (Inorganic Ventures, Christiansburg, VA) or 1,000 μg/mL (Spex Certiprep, Metuchen, NJ) single element standards with 18.2 MΩ.cm double-deionized water produced with a Barnstead Nanopure system (Thermo Scientific, Waltham, MA). Each aqueous standard (5 mL) was placed in an uncapped cup, which was discarded daily at the completion of analysis.

Archived samples were typically pills or tablets that were ground to a coarse powder able to pass through a 2-mm (USS#10) sieve. Powders were prepared simply by using a pestle to crush whole tablets placed inside a Whirl-Pak (Nasco, Fort Atkinson, WI), and secured within a piece of folded card stock. This simplified sample preparation approach was selected due to its feasibility for field use where cleaning a common mortar would be impractical and would likely lead to cross-contamination. Preparation for XRF analysis entailed placing about 2.75 – 5.75 g of coarse powdered sample into the aforementioned 32-mm polyethylene XRF sample cup with a 4-μm Ultralene window. Once prepared, the open end of the sample cup was sealed and the sample was used for all subsequent XRF measurements on both the XOS prototype and Niton instruments. In some cases, samples (n=5) consisted of pastes or creams, and were analyzed by placing the product directly into a sample cup (~2 cm thickness). All sample preparation was carried out in the Wadsworth Center’s Trace Elements Laboratory using protocols (gloves, disposable plastic ware, etc.) designed to avoid contaminating either the samples or the laboratory. Regardless, all samples were measured in triplicate on both the XOS prototype and the Niton, with the sample repositioned between each measurement. A silicon dioxide (SiO2, 99.999%) blank (provided by the instrument manufacturer) was measured each day of the analysis.

2.4. Niton XL3t GOLDD XRF analyzer

A commercial handheld XRF instrument, the Niton (Thermo Scientific, Tewksbury, MA) was also evaluated in this study, since it is intended for a wide variety of applications, including toy testing and alloy identification. The Niton is equipped with a 2-W Au anode tube source, and primary filters that produce excitation beams of different energies, with a spot size of 8 mm. It also uses an SDD to detect fluorescent X-Rays that are analyzed using Thermo’s proprietary software to yield quantitative results based on mass fraction, i.e., mg/kg or parts-per-million (ppm). The Niton measurements in this study were conducted using the soil substrate mode with the analyzer mounted in a SmartStand (Thermo Scientific, Tewksbury, MA) that allows for remote control via a PC, and more closely mimics a bench top instrument. The SmartStand ensures the Niton device is firmly secured, and provides for improved measurement repeatability by reducing variability, that results from variable positioning of the analyzer when used in the manual handheld mode. In addition, use of the SmartStand reduces unnecessary exposure to scattered X-Rays. As with the XOS prototype instrument, staff wore personal dosimeters to monitor their exposure to ionizing radiation during all operations with the Niton analyzer. The same samples prepared for analysis on the XOS prototype instrument, were reanalyzed on the Niton. A 2-minute measurement time was selected for use with the Niton instrument, based on guidance provided by the manufacturer. While longer count times would decrease measurement uncertainty, this must be balanced against a decrease in sample throughput.

2.5. ICP-OES instrumental analysis

For comparison purposes, samples analyzed by XRF (non-destructive) for the evaluation study were also analyzed by ICP-OES (US EPA Method 6010C) following acid digestion (destructive). Digested samples were analyzed for 17 elements using a PerkinElmer Optima 5300DV ICP-OES instrument (PerkinElmer Inc., Shelton, CT 06484). The plasma source was viewed axially for all elements (except Sr, which was viewed radially). A concentric nebulizer and cyclonic spray chamber were used to transfer samples into the plasma source. The Optima is equipped with an echelle polychromator with a solid-state segmented-array, charge-coupled-device detector. The optical system was purged with Ar gas during operation to improve sensitivity for elements with wavelengths below 190 nm, and the alignment optimized with 1 mg/L Mn solution. The analytical procedure followed EPA Method 6010C, including calibration, initial and continuous calibration verification. Calibration and spike standards were obtained from High-Purity Standards (Charleston, SC 29432); the calibration verification standards were obtained from AccuStandard (New Haven, CT 06513). Table 1 lists the elements determined, along with the analytical wavelength selected, plasma-viewing position, and other instrumental parameters. ICP-OES performance data, based on analytical recoveries for ERA Metals in Soil D069–540 (n=7), are also given in Table 1.

Table 1.

ICP-OES parameters, recovery data, and method detection limits

Element Wavelength (nm) Plasma Viewing Position Certified Value (mg/kg) ICP-OES* (mg/kg) Recovery (%) MDL§ (mg/kg)
As 188.979 Axial 109 107 98 2.3
Ba 233.527 Axial 325 300 92 0.2
Cd 214.44 Axial 110 88 80 4.9
Cr 267.716 Axial 93.4 80 86 0.3
Co 228.616 Axial 133 113 85 0.03
Cu 324.752 Axial 74.7 68 91 0.5
Fe 259.939 Axial 13100 11600 89 11
Pb 220.353 Axial 152 128 84 11
Mn 257.61 Axial 443 375 85 0.1
Hg 194.227 Axial 16.3 13 80 1.0
Ni 231.604 Axial 109 90 82 0.3
Se 196.026 Axial 207 18 87 0.8
Sr 421.552 Radial 111 98 88 0.1
Sn 189.927 Axial 135 130 96 3.1
Ti 334.94 Axial 193 215 111 0.2
V 292.402 Axial 110 100 91 0.1
Zn 213.857 Axial 299 260 87 8.4

PE Optima 5300DV operated with: 1500 W RF power, 0.55 L/min nebulizer flow rate, 0.2 L/min auxiliary flow rate, and 15 L/min plasma flow rate; signal integration times ranged from 5s to 20 s; read delay time 55 s; rinse time 55 s (n=3 replicates).

*

Recovery data are based on the analysis of ERA Metals in Soil D069–540 (n=7).

§

MDL calculated based on ISO/IUPAC guidelines: 3 SD of blank (n=10); mg/kg value determined for 50 mL sample volume and 50 mg sample mass

Prior to ICP-OES analysis, all samples were acid digested using a procedure based on EPA Method 3050B. Approximately 50 mg of tablet (or powder) was weighed into a disposable borosilicate glass test tube (Kimble Chase, Vineland, NJ), and 100 μL of a 500 μg/mL Au solution, prepared from a 10,000 μg/mL standard (Inorganic Ventures, Christiansburg, VA) was added to the tube, along with 5 mL of 50% (v/v) trace metal grade HNO3 (Fisher Scientific, Pittsburgh, PA). Addition of 1 mg/L Au, as AuCl3 is effective at preventing Hg volatilization by maintaining Hg2+ ions in solution [23]. Four heating and reflux steps of approximately 30 minutes each were completed at 90°C using a hot block. After an initial heating, 3 mL of concentrated HNO3, 2 mL of 18.2 MΩ.cm deionized water and 3 mL of 30% (v/v) H2O2 (J.T. Baker, Phillipsburg, NJ), and 3 mL of concentrated trace metal grade HCl (Fisher Scientific, Pittsburgh, PA) were added prior to successive heating steps. The samples were cooled before each addition and before filtration with a 1-μm Teflon filter (SCP Science, Champlain, NY). The samples were brought to final 50 mL volume with 18.2 MΩ.cm deionized water and stored in a polypropylene tube (SCP Science, Champlain, NY) pending multi-element analysis by ICP-OES.

Samples were separated into two batches for analysis by ICP-OES based on expected Hg content as determined by XRF. Samples with low Hg content were analyzed before those with very elevated Hg to minimize potential carryover. Diluent and rinse were prepared with 18.2 MΩ.cm deionized water and 5% (v/v) HNO3, 2% (v/v) HCl, and 1 mg/L Au. In many cases it was found necessary to analyze both undiluted and diluted samples to ensure that the wide range of analytes were determined within the calibration region established for ICP-OES. Standards analyzed in addition to the typical 5-point calibration curve indicated the linear dynamic range (LDR) reached 50 mg/L for most of the elements of interest, which reduced the overall number of dilutions required significantly. ERA Metals in Soil D069–540, a quality control (QC) material containing the 17 analytes of interest, was analyzed in each run along with an aqueous digestion spike, digestion blank, diluent blank, and sample duplicate, consistent with standard EPA approved methods of analysis. The accuracy of the ICP-OES procedure was validated against NIST SRM 2976 Mussel Tissue (data not shown).

3. Results and discussion

3.1. XRF instrumental performance

Instrumental sensitivity and the linear dynamic range for the XOS prototype were determined for Pb, As, Hg, and Cd using single-element aqueous standard solutions up to 10,000 μg/mL. The data are presented as aqueous calibration curves in Fig. 2.

Figure 2.

Figure 2.

Sensitivity and linearity for As, Cd, Hg, and Pb by XOS prototype instrument

Sensitivity was higher for As, followed by Pb and Hg, while Cd had poorer sensitivity. With the exception of Cd, which was linear up to 10,000 μg/mL (R2=0.98), the responses for As, Pb, and Hg are only linear up to ~1,000 μg/mL. The deviation from linearity is likely due to self-absorption of the fluorescence emitted by the sample at the higher concentrations, although there may be other factors too. The poorer precision observed at some elevated concentrations was unexpected and cannot be explained adequately. Despite the lack of linearity for As above 1,000 μg/mL, the predicted values for the aqueous standards as reported in μg/mL by the FP-software, showed a negative bias of –22% and –8% for the 5,000 μg/mL and 10,000 μg/mL standards respectively. It would appear that the FP-software is correcting for non-linearity, to some extent.

Instrumental detection limits (IDL) for the two XRF devices are shown in Table 2. The IDL is based on the analysis of a SiO2 blank, and is analogous to what is typically reported in other areas of analytical atomic spectrometry, where aqueous solutions are used to obtain IDL data [24]. The IDL for this study was obtained from the XRF instrument software, where values are reported as either “not detected (ND)”, “< Limit of detection (LOD)”, or “<”. According to IUPAC, the IDL may not accurately reflect performance with more complex samples, e.g., vegetation, food, soil, pharmaceuticals, etc., and they recommend a practical approach to detection limits that includes the sample matrix. A more rigorous approach is to estimate the “method” detection limit (MDL) by analyzing a complex matrix containing multiple detectable elements, and calculating the MDL using Equation 1 [25]:

MDL=3BPC Equation 1

where C is the concentration of the element and B and P are the count rates for the background and characteristic peak respectively. MDL values are an indication of analytical performance, but they are matrix specific.

Table 2.

Instrumental and detection limits for the XOS prototype analyzer and the Niton based on the analysis of an archived soil sample.

Element XOS prototype XRF Niton
IDL (mg/kg) MDL§ (mg/kg) IDL (mg/kg) MDL (mg/kg)
As 0.1 2 2.5 7
Cd 1.2 36 6.6 14
Hg 0.1 1 4.0 2
Pb 0.1 2 3.4 7
Ba 4.9 40 28 65
Co 0.2 6 18 4
Cr 0.8 7 14 14
Cu 0.1 17 10 22
Fe 0.2 45 23 64
Mn 0.4 37 37 85
Ni 0.2 2 35 5
Se 0.1 14 2.0 6
Sn 0.2 28 19 12
Sr 0.1 2 <1 3
Ti 2.3 54 25 12
V 1.4 9 17 13
Zn 0.1 34 5.1 23
§

n=9 days (triplicate within day)

n=5 days (triplicate within day)

IDL is instrument detection limit and is defined within the instrument software as “≤” and is based on the analysis of a SiO2 blank

MDL is method detection limit based on analysis of an archived soil sample. See text for details

For this study, repeated analyses of an archived soil sample were carried out over 9 days on the XOS prototype and over 5 days on the Niton, to estimate the MDL for 17 elements (Table 2).

The archived soil sample provided a convenient matrix for the MDL study and contained endogenous elements at levels that were within 1 order of magnitude of the MDL for 11 of the 17 elements studied, while 5 were within 2 orders of magnitude. The sole exception was Fe, which is a major component of most soils. A comparison between the IDL and MDL data (based on a soil matrix for the latter) show that, for the Niton, the agreement is within a factor of 4–5, whereas for the XOS prototype instrument, the IDL is more typically an order of magnitude lower than the corresponding MDL. With respect to the Niton, the observed discrepancies between some poorer IDL values, compared to better MDL estimates for some elements (e.g., Ni, Ti, and Co), is probably due to the different algorithms used to calculate these parameters. The lower MDL values for the XOS prototype instrument, when compared to the Niton, are most likely due to a combination of factors, including the use of the DCC monochromatic optic, which increases the signal-to-noise ratio, a higher power source and longer measurement time (3.5 versus 2 min), which collectively yield increased sensitivity for many elements. In several cases (Se, Ti, Sn and Cd), the MDL performance of the Niton analyzer is slightly better than for the XOS prototype. The reason for this may be related to poorer precision (noise) as the smaller beam size of the XOS prototype may be interrogating a non-homogeneous soil surface, while the Niton beam averages the signal over a larger sample area.

Accuracy and precision of the XOS prototype instrument were evaluated using three CRMs including Mussel Tissue, Tuna Fish, and Dogfish Muscle. NIST SRM 612 Trace Elements in Glass was also analyzed for informational purposes. Most of the CRMs were analyzed on a single day with the instrument operated in the non-metal substrate mode. Between-day analytical performance was assessed by analyzing SRM 2976 Mussel Tissue over 5 days (Table 3)

Table 3.

Accuracy, precision and uncertainty of the XOS prototype instrument based on analysis of certified reference materials.

CRM Matrix Element Certified (mg/kg) ±U XOS prototype (mg/kg) %RSD %Bias n Variability
BCR 627 Tuna Fish Tissue As 4.8 0.3 5 0 4 5 Within-day

DORM 2 Dogfish Muscle As 18 1.1 17 9 −7 10 Within-day
Cr 34.7 5.5 23 29 −33
Cu 2.34 0.16 2 0 −15
Fe 142 10 102 19 −28
Mn 3.66 0.34 ND - -
Hg 4.64 0.26 3 15 −40
Ni 19.4 3.1 11 22 −44
Se 1.4 0.09 2 23 29
Zn 25.6 2.3 25 28 −2

SRM 612 Glass Matrix Ba* 41 75.5 30 84 5 Within-day
Co* 35.5 1.2 43.6 5 23
Cu* 37.7 0.09 51 5 35
Au* 5 ND - -
Fe 51 2 36.4 8 −29
Mn* 39.6 0.8 24.4 8 −38
Ni 38.8 0.2 50.6 6 30
Ag 22 0.3 18.2 18 −17
Sr 78.4 0.2 62.4 2 −20
Ti* 50.1 0.8 ND - -
Pb 38.57 0.2 35.4 3 −8
K* 64 267 71 317
Rb 31.4 0.4 29.4 3 −6

SRM 2976 Mussel Tissue As 13.3 1.8 13 4 −4 5 Between-day
Pb 1.19 0.18 2 39 77
Br* 329 289 3 −12
Ca* 7600 6055 9 −20
Cl* 57000 52184 5 −8
Fe 171 4.9 157 9 −8
K* 9700 7610 7 −22
Mn* 33 28 10 −16
P* 8300 7974 9 −4
Rb* 4.14 5 16 10
S* 19000 14332 2 −25
Se 1.8 0.15 1 59 −31
Sr* 93 66 3 −29
Zn 137 13 137 6 <1
*

reference or informational value; ±U is the expanded uncertainty from the CRM certificate of analysis; PE=polyethylene

For the within-day data, performance was deemed acceptable for field-screening purposes if the absolute bias was <25% for certified values in the range of 100 to 1,000 mg/kg, and <50% for certified values <100 mg/kg. Low levels (<5 mg/kg) of As and Hg were also measured in biological CRMs with a bias of +4% and −40% respectively. Results for these CRMs may be considered indicative of what is possible for ‘real world’ samples such as Ayurvedic medicines or herbal supplements. Within-day precision was typically better than 10% for analytes meeting the acceptance criteria, except for NRC DORM-2 Dogfish Muscle, where the %RSD was inflated due to measured values approaching the IDL. For the between-day measurements, variability was <20% RSD except for Pb and Se, for which the certified mass fractions were <2 mg/kg. For the reference material SRM 2976 Mussel Tissue, a downward drift in measured As values was apparent between days 6–9, while a positive drift occurred for Pb over the same period. This is probably due to sub-optimal deconvolution by the FP software of the As Kα and Pb Lα characteristic lines that appear around 10.5 keV. For comparison in the same sample, within-day variability (n=3) was <8% relative for As and <50% relative for Pb (data not shown).

It is noted that, in general, the % bias increases at lower concentrations, such as for Pb at 1.19 mg/kg in SRM 2976, and that a high bias is not surprising for low Z elements such as K at a concentration of 64 mg/kg in SRM 612., Additionally, high Z elements that have low L energy lines also show a high bias, such as Ba at a concentration of 41 mg/kg in SRM 612. Overall, the sensitivity of the XOS instrument for a given element depends on the difference between the DCC optic (thence energy) that is used and the excitation energy of the element. In general, the smaller the difference, the better the excitation efficiency, and thus, the sensitivity.

3.2. ICP-OES method performance

The ICP-OES comparison method (EPA Method 6010C) used for the determination of metals in soil was validated at the Wadsworth Center as part of the laboratory’s requirements for accreditation under the American Association for Laboratory Accreditation (A2LA) as a PT provider for environmental testing laboratories [26]. Analytical recoveries based on repeated analysis of an archived QC soil sample containing the 17 elements of particular interest for this study were found to be within ±20% (Table 1). MDL determined according to the IUPAC/ISO Harmonized guidelines (3SD of 10 digestion blank measurements) are also shown in Table 1. Duplicate analysis of SRM 2976 Mussel Tissue for Cd, Cu, Se, and Zn were within 87–102% of the assigned value, while values for Sr and As were low (71–75%). Additional aqueous pre-digestion spikes were conducted and recoveries were within ±10% of the expected value (data not shown). Four samples (Emperor’s Tea Pill, Mahayogaraj Guggulu, Litargirio, Unlabeled powder) were digested and analyzed in duplicate, and the difference in duplicates was <15%.

3.3. Comparison of archived sample data by XRF with ICP-OES

A total of 38 archived samples, including medicinal products, cosmetics and foods, were analyzed for As, Ba, Cd, Cr, Co, Cu, Fe, Hg, Ni, Mn, Pb, Se, Sr, Sn, Ti, V, and Zn on the XOS prototype, on the Niton, and results compared to data obtained by ICP-OES. Samples analyzed were categorized as either HMP; ES; or CP. Table 4 shows the results for 5 representative HMPs originating from the Indian subcontinent, and 1 from China.

Table 4.

Results obtained by ICP-OES (n=1), and XRF (n=3, mean (SD)) for a representative number of HMPs samples; blank results indicate result <MDL by ICP-OES or <IDL by XRF methods. Units are mg/kg, except where indicated.

Name, Manufacturer Origin Indication Lakshmivilash Ras, Shree Baidyanath Ayurved Bhawan Pvt. Ltd. India Mahayogaraj Guggulu, Shree Baidyanath Ayurved Bhawan Pvt. Ltd. India Suketu, Sohum Bioenergetis Natura Health Care India
Chronic fever & cough Rheumatic pain

Method ICP-OES XOS prototype Niton ICP-OES XOS prototype Niton ICP-OES XOS prototype Niton
As 28 65 (4) 46 (15) 0.73% 0.40 (0.01)% 1.3 (0.06)% 720 2930 (463) 736 (625)
Cd 17 (3) 178 (4)
Hg 5.5% 8.9 (0.40)% 4.6 (0.06)% 4.7% 3.0 (0.09)% 2.9 (0.14)% 6.0% 17 (1.3)% 2.9 (2.5)%
Pb 20 (12) 4.5% 4.9 (0.10)% 6.8 (0.09)% 125 (136)
Ba 300 132 160 (<1) 530 (40) 12
Co 9.2 163 (<1) 6 0.6 18 (12)
Cr 38 7 (7) 15 105 (16) 43 (<1) 16
Cu 19 173 151 (7) 365 (25) 32 60 (18)
Fe 0.96% 2.3 (0.4)% 1.9 (0.03)% 2.9% 6.7 (0.2)% 13 (0.62)% 0.13% 0.81 (0.01)% 0.27 (0.21)%
Mn 86 222 (40) 179 (45) 526 749 (2) 1785 (224) 34 189 (34) 45 (<1)
Ni 43 54 (11) 23 16 (12) 573 (79) 7.9
Se 45 (<1)
Sn 6.7 19 (2) 1.1% 1.8 (0.07)% 3.8 (0.003)% 56 76 (56)
Sr 81 62 71 (6) 23 251 (212)
Ti 406 1456 (288) 852 (30) 303 877 (156) 841 (97) 36 174 (97) 130 (89)
V 7.7 22 (15) 21 91 (16) 68 (<1) 2.2
Zn 45 488 (55) 304 (24) 330 149 (40) 880 (11) 17 1019 (<1) 189 (144)
Name, Manufacturer Origin Indication Garbhapal Ras India Emperor’s Tea Pill, Kingsway Trading Inc. China Kumar Kalyan Ras Unknown
Healthy pregnancy Natural balance

Method ICP-OES XOS prototype Niton ICP-OES XOS prototype Niton ICP-OES XOS prototype Niton
As 0.05% 0.15 (0.09)% 0.72 (0.07)% 234 (19) 422 289 (160) 987 (359)
Cd 30 (<1) 122 (9) 25 17 (3) 32 (33)
Hg 5.1% 2.8 (0.60)% 2.9 (0.18)% 0.71% 0.39 (0.02)% 0.40 (0.008)% 15% 6.2 (0.70)% 6.8 (2.5)%
Pb 3.3% 3.3 (0.98)% 4.9 (0.45)% 0.49% 0.22 (0.02)% 0.44 (0.005)% 0.34% 0.17 (0.04)% 0.27 (0.11)%
Ba 11.4 147 (19) 1506 1108 (189) 177 149 (<1)
Co 0.7 213 (15) 0.7 14 17 (<1)
Cr 2.7 4.6 133 246 (173) 120 (58)
Cu 39 21 (20) 76 (8) 27 9 (0.6) 35 (5) 310 100 (39) 870 (523)
Fe 0.98% 2.1 (0.6)% 3.8 (0.3)% 593 777 (22) 1200 (14) 14% 9.6 (1.4)% 44 (17)%
Mn 110 93 (53) 234 (78) 28 37 (2) 0.11% 0.09 (0.05)% 0.38 (0.16)%
Ni 2.7 296 (120) 1.9 0.4 (0.1) 67 18 (12) 391 (46)
Se 57 (8) 4.2 2 (<1) 15 (2) 109 (6)
Sn 0.70% 1.2 (0.5)% 1.9 (0.16)% 3.5 4 (0.7) 301 0.26 (0.22)% 0.08 (0.06)%
Sr 94 78 (2) 27 18 (4) 25 (0.6) 78 0.10 (0.02)%
Ti 27 194 (167) 235 (78) 10 385 (102) 171 (<1) 0.09% 0.14 (0.11)% 0.13 (0.04)%
V 1.2 0.6 188 (51) 7.6 76 (23)
Zn 82 104 (53) 317 (67) 0.19% 0.13 (0.02)% 0.32 (0.01)% 496 248 (389) 1671 (681)

The complete dataset for all 22 HMPs analyzed by each of the 3 analytical methods for all 17 elements (n = 1122 results) can be accessed as an electronic supplementary information (ESI) files. All HMP samples were analyzed in triplicate using both XRF devices. A portion of the same original sample was acid digested and analyzed for multiple elements by routine ICP-OES, for comparison purposes. For each detectable value obtained by XRF, the SD (n=3) is shown in parentheses.

Table 5 shows representative results obtained for 4 CP and one ES that originated from the Indian subcontinent.

Table 5.

Results obtained by ICP-OES (n=1) and XRF (n=3, mean (SD)) for a representative number of CPs and ES samples; blank results indicate result <MDL by ICP-OES or <IDL by XRF methods. Units are mg/kg, except where indicated.

Type Name, Manufacturer Origin CP Personal Kajal, Blue Heaven Cosmetics Ltd. India CP Kohl Aswad, Hashmi Pakistan CP Surmi Special, Hasmi Pakistan

Method ICP-OES XOS prototype Niton ICP-OES XOS prototype Niton ICP-OES XOS prototype Niton
As 0.04 (0.04)% 8.7 (0.34)% 7.2 359 (34) 8.2 (0.11)%
Cd 10 (13) 13 1.1 (0.9) 1.1 (0.17)% 16 0.4 (0.2) 2.6 (2.4)%
Hg 3.7 2.5
Pb 25 3 (2) 20% 4.9 (2.9)% >100 (2.3)% 19% 3.0 (0.09)% >100 (0.61)%
Ba 2.5 60 192 (<1) 5.1 (0.84)% 96 12 (11)%
Co 2.6 38 (<1) 0.3 3 (<1) 2.2 21 (<1)
Cr 3.9 14 (15) 0.4 18 (<1) 1.1
Cu 3.3 3 (<1) 348 41 (20) 713 (116) 0.16% 0.02 (0.002)% 0.6 (0.04)%
Fe 2.8% 1.0 (0.9)% 6.6 (0.76)% 213 112 (22) 720 108 (9)
Mn 190 84 (99) 319 (56) 1.4 15.4
Ni 4.0 5 (6) 0.8 15 (<1) 1.8
Se 0.2 (0.1) 0.9 81 (<1)
Sn 34 (58) 3 (2) 2.7 (0.48)% 1 (<1) 6.3 (5.7)%
Sr 0.8 109 246 (<1) 168
Ti 62 185 (301) 0.4 69 (<1) 0.4 6 (<1)
V 65 (83) 0.3 67 (<1) 44 (1) 0.3
Zn 6 (<1) 0.54% 0.08 (0.01)% 2.2 (0.04)% 0.55% 0.07 (0.004)% 2.5 (0.02)%
Type Name, Manufacturer Origin CP Kajal powder Unknown ES Chili powder India

Method ICP-OES XOS prototype Niton ICP-OES XOS prototype Niton
As 263 (176) 8.7 (0.40)% 6 (2) 14 (6)
Cd 37 0.6 (0.3) 1.7 (1.2)% 1 (<1)
Hg 3.2
Pb 20% 3.7 (0.60)% >100 (4.5)% 193 218 (11) 261 (19)
Ba 162 7.9 (5.7)% 32 34 (12)
Co 0.5 4 (5) 897 (<1) 0.8 2 (<1)
Cr 0.5 47 54 (10) 84 (8)
Cu 152 7 (1) 11 12 (0.6) 17 (1)
Fe 605 68 (6) 228 295 (38) 273 (51)
Mn 7.3 0.16 (0.04)% 19 24 (3)
Ni 1.1 2 3 (<1)
Se 8 (2) 0.3 (0.1)
Sn 0.9 (0.2) 4.0 (2.8)% 20 (0.6)
Sr 193 13 13 (<1) 17 (0.5)
Ti 0.6 98 (48) 3.9 34 (10) 55 (10)
V 0.4 16 (19) 0.1 7 (1)
Zn 0.30% 385 (189) 1.5 (0.18)% 19 23 (2) 23 (3)

Again, a complete dataset with results for all 12 CPs and 4 ES samples, by each XRF device, and comparison data from ICP-OES, for all 17 elements (n = 816 results) are available as ESI files. Each of the HMP, CP and ES samples was analyzed in singlet by ICP-OES, however, the uncertainty in the ICP-OES value is estimated at approximately ±20% based on the QC (Table 3). Missing values in Tables 4 and 5, and in their associated ESI files, were either below the MDL (ICP-OES) or the XRF instrument software (XOS and/or Niton) did not report it. In some cases, the minimum reporting limits (i.e., the IDL) for the XRF software were lower than the MDL calculated according to the ISO/IUPAC guideline. Descriptive information available on each of 38 samples sample analyzed, including the manufacturer, country of origin, indication, and labeling details, are also given in Tables 4 and 5, and in the ESI files.

The elements Zn and Ti were detected by both XRF instruments in nearly all of the samples analyzed. Values for As, Cu, Hg, and Pb were also reported in many of the samples. However, based on the comparison with ICP-OES, several false positive results were evident for As when Pb was also present. This is a well-known problem in energy dispersive XRF due to overlap of the As Kα and Pb Lα lines at 10.53 keV and 10.55 keV, respectively. Both XRF instruments calculate results for As and Pb using FP-software that is designed to deconvolute the spectrum, and correct for the overlap. However, it is clear from the data reported here that the overlap correction is not always successful, and caution is warranted when As and Pb are both present. Another possibility for spectral overlap can occur from high levels of Fe, i.e., Fe sum peaks can cause a false positive for Pb in the presence of As. For Co, Ni, and Se, ICP-OES found values were very often near or below the XRF minimum reporting limits for both the XOS prototype and Niton instrument. Consequently, several false positive results for Co, Ni, and Se were evident by XRF.

In one case, the Lakshmivilash product from India (Table 4) was found to contain 300 mg/kg Ba via ICP-OES. Yet both the XOS prototype and Niton analyzer reported less than the manufacturer’s detection limit. The most likely explanation for this false negative discrepancy is not that these instruments are unable to detect Ba at 300 ppm, but rather the software cannot accurately fit the X-Ray peak above the background for this particular sample. In some cases the background may be so high that it completely obscures the peak, thereby resulting in a false negative. It is possible that use of the low energy optic in the XOS prototype instrument might have addressed this.

Discrepancies between XRF-reported data and ICP-OES data can be investigated in more detail using difference plots i.e., where the difference between the XRF and ICP-OES results are plotted as a function of increase in the latter. Fig. 3 shows difference plots for the elements Pb, As, and Hg obtained using the XOS prototype and the Niton analyzer with HMP, ES and CP samples. The elements were selected based on their public health significance. Dashed lines in Fig. 3 denote an arbitrary performance criterion of ±50% relative to ICP-OES, and might be considered acceptable for screening purposes when attempting to identify gross contamination.

Figure 3.

Figure 3.

Difference plots (XRF – ICP-OES value) for Pb, As, and Hg in powdered samples measured with the XOS prototype and the Niton; left side shows all results, right includes only low mg/kg range; XRF(n=3, error bars +/− SD), dashed lines indicate +/− 50% of reference value; HMP (herbal medicine product), ES (ethnic spice), CP (cosmetic product)

However, a negative 50% bias at lower levels could result in materials being erroneously identified as “safe” when they are contaminated, i.e., a false negative. The nature of a difference plot requires that there be two detectable results from both techniques. Thus, only detectable (ICP-OES >MDL) or reportable data (XRF) can be included in these plots. Given the levels of Pb, As and Cd found in these samples covers several orders of magnitude, the difference plots shown in Fig. 3 are arranged such that the left panels (a-c) represent the full range of contents found (up to wt.% levels), while the right hand panels (d-f) show the lower range of values detected.

It is evident that a large number of data points in Fig. 3 fall outside the ±50% dashed lines. Overall, the Niton tends to have a positive bias, while the XOS prototype instrument has a negative bias, though performance appears to be element and matrix specific. Performance for Pb on both the XOS and Niton was generally within ±50% (relative to ICP-OES) up to 800 mg/kg, while above 10% by weight, it was much poorer, with the Niton being positively biased and the XOS prototype instrument negatively biased. Most of the samples showing poor agreement were CP such as kohl, surma, or kajal, which are fine black powders often consisting of lead sulfide that is applied to the conjunctive surface of the eyelids. However, both the XOS and Niton instruments correctly identified gross Pb contamination in the CP samples.

A sample of chili powder imported from India that contained approximately 190 mg/kg Pb based on analysis by ICP-OES was reported as approximately 220 mg/kg Pb using the XOS prototype analyzer (16% bias) and approximately 260 mg/kg on the Niton (37% bias).

A second example is the Mahayogaraj Guggulu pill, an HMP manufactured in India and labeled for use with rheumatic pain. The sample was found to contain 4.5 wt.% Pb, 4.7 wt.% Hg, and 7300 mg/kg As based on analysis by ICP-OES. Both XRF instruments identified the three toxic elements, although the XOS prototype instrument had less bias (Pb 8%; Hg −36%; As 45%) compared to the Niton (Pb 51%; Hg −38%; As 78%). This is in accordance with the general trend for Hg of an absolute bias <50% for all the concentration ranges measured and with the general trend of higher absolute bias at As concentrations < 1%.

A third example is the Emperor’s Tea Pill which is an HMP imported from China and labeled for “natural balance”. Analysis by ICP-OES showed it contains 4900 mg/kg Pb and 7100 mg/kg Hg. Both XRF instruments correctly identified the presence of Pb: XOS 2200±200 mg/kg; Niton 4400±50 mg/kg, and Hg: XOS 3900±200 mg/kg; Niton 4000±80 mg/kg when the sample was prepared in a powdered form.

The samples were re-analyzed as whole tablets, using several layers of intact Emperor’s Tea Pills placed in the sample cup, to explore the effect of sample preparation on performance. With this approach, Pb and Hg were detected on the XOS (680±90 mg/kg Pb and 1010±40 mg/kg Hg) and on the Niton (3400±100 mg/kg Pb and 2700±80 mg/kg Hg). They both demonstrated much poorer accuracy than when the sample was analyzed in powdered form. Imprecision was also slightly poorer for whole tablets (up to 13% RSD) compared to powders (up to 10% RSD) for both XRF instruments. The poorer performance for whole tablets versus powdered samples was less pronounced for the Niton than the XOS prototype instrument. This may be due in part to the larger spot size of the Niton (8 mm), which is less affected by sample heterogeneity, compared with the smaller (<1 mm) spot of the XOS prototype. Nevertheless, each of the XRF instruments correctly identified very high-levels of Pb and Hg in whole tablets, suggesting that either would be adequate for rapid screening of such products without the need for sample preparation.

Fig. 4 shows difference plots for the elements Cd, Cu, Ba, Fe, Zn and Mn in HMP, ES, CP powdered samples. The dashed lines denote an arbitrary performance criterion of ±50% relative to ICP-OES. The elements were selected based on either their public health significance (Cd) or frequency of detection by XRF (Fe, Cu, Zn, Mn and Ba).

Figure 4.

Figure 4.

Difference plots (XRF – ICP-OES value) for Cd, Fe, Cu, Zn, Mn and Ba in powdered samples measured with the XOS prototype and the Niton; XRF(n=3, error bars +/− SD), dashed lines indicate +/− 50% of reference value; HMP (herbal medicine product), ES (ethnic spice), CP (cosmetic product)

In general, the Niton reports values for Cd, Cu, Fe, and Ba that are positively biased relative to ICP-OES. The XOS prototype data are less biased by comparison for Fe, Cu, and Zn. In the case of Mn, XRF performance is more complicated with several false positives errors noted for both the Niton and the XOS instruments. For the XOS prototype, this may be the result of disabling the lower energy beam at 7 keV, which may be more optimal for exciting the Mn Kα line.

For Cd, the XOS prototype instrument reported a negative bias that was proportional to the concentration range studied, while the Niton reported a large systematic positive bias that was >1,000 mg/kg. These data underline the real challenges of measuring Cd in field-type samples by portable XRF, given the low energy of the Cd Lα line (3.133 keV), its low fluorescence yield, and a high spectral background in this region. Overall, Cd detection was poor for both XRF instruments, due in part to the low concentrations (≤37 mg/kg) present in these samples, and the technical challenges in measuring Cd by XRF using these instruments. However, the performance for Cd appeared better for the XOS prototype than for the Niton instrument. For example, Cd was detected with the XOS (17±3 mg/kg) in the Emperor’s Tea Pill but was reported as <LOD=5 mg/kg by the Niton.

Of the samples analyzed by ICP-OES, 63% (24/38) contained Pb, while 89% (34/38) contained As, Cd, Hg, or Pb. Of the contaminated products, 74% (25/34) were found to contain at least 2 of these toxic elements. Thus, while lead poisoning associated with use of HMPs is most often reported in case studies and remains the primary health concern, potential toxicity associated with As, Cd, or Hg exposure should also be considered for consumers using HMPs. While numerous products either did not have labeling in English or packaging was not available, one sample was labeled as “lead-free” (defined on the product as <20 ppm), and 3 were marked with a claim of good manufacturing practice (GMP). The sample labeled as lead-free was found to contain Pb (~30 mg/kg), and each of the GMP-labeled samples contained detectable levels of Cd, Hg, or Pb.

The samples analyzed in this study can be compared to elemental contaminant limits for dietary supplements (maximum intake of 10 g/day) suggested by the U.S. Pharmacopeia; 1.5 mg/kg As, 0.5 mg/kg Cd, 1 mg/kg Pb, and 1.5 mg/kg Hg [27]. For color additives, As is typically permitted up to 3 mg/kg, Hg at 1 mg/kg, and Pb at 10–20 mg/kg dependent upon the colorant [28]. Since the MDLs of both the XOS prototype and the Niton are near or above these limits, it is unlikely that either instrument would be appropriate to determine compliance with regulatory limits. However, considerable time and cost savings, as well as increased screening rates, could be realized if either is employed as a tool to identify materials with potential toxicologically significant elemental content. Then a follow up study with accepted laboratory methods could be performed if either XRF instrument detects a potential problem.

3.4. Study limitations

Limitations exist in the sample preparation techniques used for this study. First, many of the HMPs analyzed by ICP-OES were a sub-sample of the original product that was digested in their initial (e.g., tablet) form, rather than a digestion of the actual powdered sample measured by XRF. Therefore, the comparability of XRF and ICP-OES data could potentially be influenced by within-product variability. Analysis of the original tablets by ICP-OES was carried out to permit the calculation of the elemental content per tablet, and for direct comparison to measurement of the intact tablets by XRF (data not shown). None of the samples were dried prior to analysis, because drying would be difficult under field conditions, which was the intended purpose of the XOS instrumentation. Ideally, it might be useful to analyze these samples as wet and dry to determine what effect drying would have on their reported elemental composition by XRF. However, the contribution to overall uncertainty in the XRF results has been reported to be minimal for a residual moisture content up to 20% [29]. Several of the samples provided by the NYC DOHMH for use in this validation study were previously analyzed for As, Hg, and Pb content by another laboratory, and those results were published elsewhere4. Moreover, most of the samples analyzed in this study were received as sub samples (i.e., from the same package). In some cases, samples were removed from an unopened package of the same lot, while in other cases they may have been from a different lot. HMPs are an example of products where there is a lack of homogeneity. Many are essentially “hand-made” and don’t have the uniformity characteristics usually associated with pharmaceutical grade dietary supplements and drugs. So differences between tablets within the same lot are often very large. Despite this, it is notable that our ICP-OES results for Hg are generally higher (by as much as 75%) than those reported previously.4 This discrepancy may warrant further investigations of the method used by the other laboratory, and its validation for these types of samples, as those details were not given in that report.

4. Conclusions

The performance (accuracy, precision, and robustness) of two field-portable XRF instruments, a prototype XOS instrument featuring DCC optics and a Niton XL3t analyzer, were evaluated and compared to ICP-OES under laboratory conditions, for determination of multiple elements in HMPs and other ethnic (ES and CP) products. Both instruments performed adequately for the identification of contaminated HMPs and ethnic products that may pose a significant public health threat. Neither XRF instrument was sufficiently sensitive for quantifying major toxic contaminants below 15 mg/kg with good accuracy. Nonetheless, the real advantage to this type of XRF technology is its potential for portability, use on site and real time analysis, as well as pre-screening samples prior to analysis by ICP-OES/MS. However, as designed, the XOS prototype instrument was not as portable or as easy to transport as was the Niton analyzer. Further improvements in the design of the XOS instrument are warranted for future field applications. The results of this study highlight some of the challenges that may be encountered when using portable XRF instruments in the field, including issues related to sample preparation and the limitations of calibration and quantitation based on FP models. Users should be aware of the potential for misinterpreting results, especially for those samples that contain both As and Pb. The key advantages of XRF over laboratory-based methods are that it is non-destructive, provides for rapid identification of multiple toxic elements at elevated levels, and can be used under field-based conditions by technicians with some basic training in analytical science.

Supplementary Material

Supplementary Material

Acknowledgements

Development of the XOS prototype instrument was supported by grant number R43 ES016689 from the National Institute of Environmental Health Sciences, under the Small Business Innovative Research program to X-Ray Optical Systems (East Greenbush, NY), and the Wadsworth Center. The content is the sole responsibility of the authors and does necessarily not represent the views of NIEHS or NIH. Use of trade names is for identification purposes only and does not imply an endorsement by the New York State Dept. of Health or the NIH. The authors acknowledge Dr. Paromita Hore and Munerah Ahmed of the Bureau of Environmental Disease Prevention of the New York City Department of Health and Mental Hygiene for providing the archived samples and supporting information for use in this study.

References

  • [1].World Health Organization (WHO), Traditional medicine, http://www.who.int/mediacentre/factsheets/2003/fs134/en/, accessed February 2015.
  • [2].Eisenberg DM, Davis RB, Ettner SL, Appel S, Wilkey S, Van Rompay M and Kessler RC, JAMA 280, 1569 (1998). [DOI] [PubMed] [Google Scholar]
  • [3].Dargan PI, Gawarammana IB, Archer JRH, House IM, Shaw D and Wood DM, Int. J. Environ. Health 2, 463 (2008). [Google Scholar]
  • [4].Centers for Disease Control and Prevention (CDC), Morbidity and Mortality Weekly Report 61, 641 (2012)22914225 [Google Scholar]
  • [5].Saper RB, Phillips RS, Sehgal A, Khaouri N, Davis RB, Paquin J, Thuppil V and Kales SN, JAMA, 300, 915 (2008). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [6].Ernst E, Trends Pharmacol. Sci, 23, 136 (2002). [DOI] [PubMed] [Google Scholar]
  • [7].Pakade Y, Kumari A, Singh S, Sharma R and Tewary D, B. Environ. Contam. Tox 86, 133 (2011). [DOI] [PubMed] [Google Scholar]
  • [8].Kalny P, Wyderska S, Fijalek Z and Wroczynski P, Acta Pol. Pharm. - Drug Res 69, 279 (2012). [PubMed] [Google Scholar]
  • [9].Wu J, Tan Y, Wang Y and Xu R, J. Nat. Med 65, 656 (2011). [DOI] [PubMed] [Google Scholar]
  • [10].Hina B, Rizwani GH, Shareef H and Ahmed M, Pak. J. Pharm. Sci 25, 247 (2012). [PubMed] [Google Scholar]
  • [11].Soylak M, Cihan Z and Yilmaz E, Environ. Monit. Assess 184, 3455 (2012). [DOI] [PubMed] [Google Scholar]
  • [12].Giacomino A, Abollino O, Malandrino M, Karthik M and Murugesan V, Microchem. J 99, 2 (2011). [Google Scholar]
  • [13].Saper RB, Kales SN, Paquin J, Burns MJ, Eisenberg DM, Davis RB and Phillips RS, JAMA 292, 2868 (2004). [DOI] [PubMed] [Google Scholar]
  • [14].Mahawatte P, Dissanayaka KR and Hewamanna R, J. Radioanal. Nucl. Ch 270, 657 (2006). [Google Scholar]
  • [15].Al-Omari S, X-Ray Spectrom. 40, 31 (2011). [Google Scholar]
  • [16].Lin CG, Schaider LA, Brabander DJ and Woolf AD, Pediatrics 125, 828(2010). [DOI] [PubMed] [Google Scholar]
  • [17].Murphy T, Lim S, Huong SP, Irvine K, Bayen S, Kelly BC and Wilson K, J. Health Pollution 2, 21 (2012). [Google Scholar]
  • [18].Palmer PT, Jacobs R, Baker PE, Ferguson K and Webber S, J. Agric. Food Chem 57, 2605 (2009). [DOI] [PubMed] [Google Scholar]
  • [19].West M, Ellis AT, Potts PJ, Streli C, Vanhoof C, Wobrauschek P. 2015 Atomic Spectrometry Update - a review of advances in X-ray fluorescence spectrometry and their applications. Journal of Analytical Atomic Spectrometry. 2015. doi: 10.1039/C5JA90033F. [DOI] [Google Scholar]
  • [20].National Human Genome Research Institute, http://www.genome.gov/19518663, accessed January 2015.
  • [21].National Institutes of Health (NIH), Environmental Sensors for Personal Exposure Assessment (SBIR [R43/R44]), http://grants.nih.gov/grants/guide/rfa-files/RFA-ES-07-001.html, accessed February 2015. [Google Scholar]
  • [22].Chen Z, Gibson WM, Powder Diffr. 17, 99 (2002). [Google Scholar]
  • [23].Environmental Protection Agency (EPA), Mercury Preservation Techniques 090CMB03-FS-MPT, http://www.epa.gov/esd/factsheets/mpt.pdf, accessed October 2015.
  • [24].Thompson M, Ellison SLR and Wood R, Harmonized Guidelines for Single-laboratory Validation of Methods of Analysis (IUPAC Technical Report) (2002).
  • [25].M Gibson W, W Chen Z and Li D, X-Ray Opt. Inst (2008), 17 pages. [Google Scholar]
  • [26].New York State Department of Helath (NYSDOH), Standard Operating Procedure for Metals in Water and Soil by Inductively Coupled Plasma Optical Emission Spectrometry, 2011, Doc DOH-LINC-203–11.
  • [27].US Pharmacopeia (USP), Elemental Contaminants in Dietary Supplements. Pharamcopeial Forum 38, (2012). [Google Scholar]
  • [28].Code of Federal Regulations, Title 21, Part 74 Listing of Color Additives Subject to Certification, http://www.ecfr.gov/cgi-bin/text-idx?c=ecfr&sid=1070b19eb50e562daa872cfa1755aa09&rgn=div5&view=text&node=21:1.0.1.1.27&idno=21#21:1.0.1.1.27.3.31.1, accessed February 2015.
  • [29].Environmental Protection Agency (EPA), Field Portable X-Ray Fluorescence Spectrometry for the Determination of Elemental Concentrations in Soil and Sediment 2007. Method 6200, http://www.epa.gov/osw/hazard/testmethods/sw846/pdfs/6200.pdf, accessed February 2015.

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

Supplementary Material

RESOURCES