Diagnostic performance and clinical implications for enhancing a hybrid quantitative flow ratio–FFR revascularization decision-making strategy

Joyce Peper 1,2,, Robbert W van Hamersvelt 2, Benno J W M Rensing 1, Jan-Peter van Kuijk 1, Michiel Voskuil 3, Jurriën M ten Berg 1, Jeroen Schaap 4, Johannes C Kelder 1, Diederick E Grobbee 5, Tim Leiner 2, Martin J Swaans 1
PMCID: PMC7979768  PMID: 33742077

Abstract

Invasive fractional flow reserve (FFR) adoption remains low mainly due to procedural and operator related factors as well as costs. Alternatively, quantitative flow ratio (QFR) achieves a high accuracy mainly outside the intermediate zone without the need for hyperaemia and wire-use. We aimed to determine the diagnostic performance of QFR and to evaluate a QFR–FFR hybrid strategy in which FFR is measured only in the intermediate zone. This retrospective study included 289 consecutive patients who underwent invasive coronary angiography and FFR. QFR was calculated for all vessels in which FFR was measured. The QFR–FFR hybrid approach was modelled using the intermediate zone of 0.77–0.87 in which FFR-measurements are recommended. The sensitivity, specificity, and accuracy on a per vessel-based analysis were 84.6%, 86.3% and 85.6% for QFR and 88.0%, 92.9% and 90.3% for the QFR–FFR hybrid approach. The diagnostic accuracy of QFR–FFR hybrid strategy with invasive FFR measurement was 93.4% and resulted in a 56.7% reduction in the need for FFR. QFR has a good correlation and agreement with invasive FFR. A hybrid QFR–FFR approach could extend the use of QFR and reduces the proportion of invasive FFR-measurements needed while improving accuracy.

Subject terms: Cardiology, Interventional cardiology, Coronary artery disease and stable angina

Introduction

The most frequently used reference test for the assessment of significant obstructive coronary artery disease (CAD) is invasive coronary angiography (ICA)1,2. However, the correlation between anatomically significant stenosis and physiological reduction of myocardial blood flow is weak3. Fractional flow reserve (FFR), a physiological test that serves as a proxy for myocardial blood flow, can be used in addition to ICA to assess the hemodynamical impact of a stenosis. Application of FFR has been proven cost-effective for the management of patients suffering from CAD compared to ICA-guided therapy by improving clinical outcomes and reducing stent implantations and thereby costs46. Nevertheless, it is used in less than 10% of patients due to a combination of procedural and operator related factors as well as high costs7.

Recent developments in functional coronary imaging may overcome this problem by allowing for wire- and hyperaemia-free FFR measurements for the assessment of physiological significant stenosis. Quantitative flow ratio (QFR) is a recently developed coronary-based physiological test to calculate a virtual FFR value. QFR is computed by applying (simplified) computational fluid dynamics principles on a three-dimensional model of the coronary artery constructed from two invasive angiographic views811. Flow-velocity information acquired by Thrombolysis in Myocardial Infarction (TIMI) frame counting can be added to improve the calculations. QFR was recently evaluated in a meta-analysis and was shown to have a sensitivity of 0.89 (95% CI 0.83–0.95) and a specificity of 0.89 (95% CI 0.87–0.92) to diagnose functionally significant CAD (FFR ≤ 0.80)12.

Since there are no results of (randomized) outcome studies available, a QFR-only strategy may perhaps not be feasible and safe yet. Furthermore, imperfect correlation between FFR and QFR might result in different treatment decisions based on QFR and FFR measurements, especially when close to the QFR cut-off value of 0.80 (0.75–0.85), also known as the intermediate zone. The low adoption of functional lesion assessment by invasive FFR combined with the high classification agreement between QFR and FFR outside of the intermediate zone provide an opportunity for a combined QFR–FFR strategy. A hybrid QFR–FFR strategy, in which only lesions with intermediate QFR values would require FFR measurements, may seem the most suitable clinical application of QFR. Various QFR-hybrid limits for defining the intermediate zone have been proposed based on 90–95% sensitivity–specificity, namely QFR-treat limits between 0.75 and 0.78 and QFR-defer limits between 0.85 and 0.878,10,13,14. However, these QFR–FFR hybrid limits have only been validated in the cohorts from which they were derived. To our knowledge, this is the first study that assesses the diagnostic value of a QFR–FFR hybrid strategy in an independent, real-world cohort.

Accordingly, the aim of this study was to determine (1) the diagnostic performance of QFR compared to FFR and (2) to evaluate the impact of QFR and QFR–FFR hybrid strategies on the proportion of avoided invasive FFR measurements in consecutive enrolled patients whilst matching the diagnosis of a FFR strategy.

Methods

Patient population

The present study is a multicentre study of patients who underwent CT, ICA and FFR to compare QFR versus FFR. All consecutive patients undergoing these diagnostic tests for diagnosing CAD between October 2009 and October 2017 at the St. Antonius Hospital Nieuwegein and all consecutive patients between June 2012 and July 2016 at the University Medical Centre Utrecht were retrospectively enrolled. Three-dimensional quantitative coronary angiography (3D-QCA) and QFR were performed and compared with the reference standard, invasive FFR. 517 vessels in 378 patients were screened. Patients with (1) lack of 2 optimal angiographic views; (2) overlap of other vessels with the lesion or areas around the lesion in the vessel of interest; (3) foreshortening of the target coronary artery in one or both angiographic acquisitions; (4) insufficient contrast injection; (5) location of the lesion of interest at the ostial left main coronary artery or ostial right coronary artery; (6) prior percutaneous coronary intervention (PCI) of vessel of interest (7) bypass grafts were excluded. Since this concerned a retrospective study, the Medical Ethics committees of both institutions (Medical Research Ethics Committees United [MEC-U] and Medisch Ethische Toetsingscommissie Utrecht [METC Utrecht]) and both local institutional boards approved the protocol and waived the need for informed consent. The ethical principles for medical research on human beings of the Declaration of Helsinki and the Good Clinical Practice Guidelines were followed.

ICA/FFR

Invasive coronary angiography was performed according to the standardized protocol used in both hospitals. ICA biplane views were acquired via either femoral or radial artery access. The coronary artery tree was fully examined for presence of stenosis. Invasive FFR measurements were acquired for clinical indications unrelated to this study using a pressure wire passed beyond the stenosis. Patients suffering from visually assessed intermediate stenosis, defined as a diameter reduction between 50 and 90%, or multivessel disease were subsequently assessed by measuring FFR to determine its functional severity15. The pressure gradient across the stenosis was measured during an intracoronary adenosine bolus or continuous intravenous infusion of adenosine at 140 µg/kg/min. The exact location of the wire during measurement was recorded. Vessel-based analysis was performed from which diagnostic accuracy of QFR was determined. To compare and assess the diagnostic performance of QFR, the clinical standard of FFR ≤ 0.80 indicating hemodynamically significant stenosis was applied.

QFR

3D-QCA analysis and the computation of QFR were done using dedicated QAngio software (QAngio XA 3D 1.1, Medis Medical Imaging System, Leiden, The Netherlands) by a QFR-certified observer while blinded for the FFR results and the treatment decision. The analysis was performed as previously described11. In brief, two angiographic views, at least 25° apart, were selected based on the least foreshortening of the stenosis and a minimum overlap between the main vessels and side branches. In both views, the end-diastolic frames guided by ECG (if available) were used for analysis and an anatomical landmark was indicated as reference point. The most proximal and distal points were indicated, and vessel contours were automatically extracted. Manual corrections were made if needed. A 3D reconstruction of the vessel was made for the 3D-QCA and minimum lumen diameter, reference vessel diameter, percentage diameter stenosis and lesion length were extracted. QFR calculations were performed on the anatomical information acquired by 3D-QCA in combination with modelled hyperaemic flow velocity with as input TIMI frame count analysis16.

QFR–FFR hybrid approach

The hybrid approach as proposed in this study includes an intermediate zone of QFR values around the cut-off point to define hemodynamically significant CAD for which FFR measurements need to be performed (Fig. 1). The QFR–FFR hybrid approach was modelled using the limits as proposed in the FAVOR II Europe-Japan study8. Normal QFR was defined as QFR > 0.87 (hemodynamically normal), abnormal QFR was defined as QFR < 0.77 (hemodynamically significant) and the intermediate zone was defined as QFR 0.77–0.87 (non-conclusive). In addition, multiple limits (per 0.01 QFR-units) for normal (QFR-defer) and abnormal (QFR-treat) were simulated. The proportions of potential adenosine- and wire-free (FFR-free) procedures defined as the fraction of stenosis outside the intermediate zone, were calculated.

Figure 1.

Figure 1

Study overview QFR – FFR hybrid strategy. All 50–90% stenosis or multivessel disease on invasive coronary angiography are assessed by QFR using dedicated software (QAngio XA 3D 1.1, Medis Medical Imaging System, Leiden, The Netherlands). Based on the QFR-value, lesions are categorized into hemodynamically normal (QFR > 0.87), intermediate zone (QFR between 0.77 and 0.87) or hemodynamically significant (QFR < 0.77). Only lesions in the intermediate zone are further assessed by FFR. In our population, this leads to a reduction of invasive FFR measurements of 56.7%. FFR fractional flow reserve, QFR quantitative flow ratio.

Statistical analyses

To compare and assess the diagnostic performance of both QFR and the QFR–FFR hybrid approach, the clinical standard of FFR ≤ 0.80 indicating hemodynamically significant stenosis was applied. The same threshold of ≤ 0.80 was applied for the QFR measurements. For the statistical analyses, the distribution of continuous variables was assessed using histograms and Q–Q plots. Continuous variables were expressed as means and standard deviations (SD) if normally distributed or medians and 25–75% interquartile ranges (IQR) if non-normally distributed. Categorical variables were represented as totals and percentages. Diagnostic performance on a per vessel-basis as sensitivity, specificity, positive predictive value (PPV), negative predictive value (NPV) and accuracy were calculated. The variables were calculated as proportions with a 95% confidence interval. The correlation, differences, and diagnostic performance between QFR and wire-based FFR were further assessed using Pearson correlation coefficient, Bland–Altman plots, receiver operating characteristics-curves (ROC-curves) and area under the ROC curve (AUROC-curve). QFR–FFR hybrid approaches using various cut-off points for the intermediate zone in which FFR is measured were simulated. All statistical analyses were performed using R statistical software (www.r-project.org, version 3.6.2).

Results

A total of 378 (252 patients of St. Antonius Hospital Nieuwegein and 126 patients of University Medical Centre Utrecht) patients were identified for potential inclusion in this study. Patients with a history of coronary bypass grafting or lesions in the ostia of the left main or right coronary artery (RCA) (n = 39) were excluded. 339 patients were evaluated by QFR and due to overlap, foreshortening of the stenosis, lack of contrast or lack of two angiographic projections at least 25 degrees apart, an additional 50 patients were excluded. Therefore, 289 patients and 381 vessels were included in the statistical analyses (Fig. 2). Patient characteristics and the characteristics of the analysed vessels are presented in Table 1. Approximately 73.0% of the study cohort was male, the mean age was 64.1 ± 10.4 years and 51.9% had at least three cardiovascular risk factors. On average, the FFR was 0.81 and 169 vessels (44.4%) had a significant stenosis, while the mean QFR was 0.82 and 172 (45.1%) had a significant stenosis.

Figure 2.

Figure 2

Study enrolment flow chart. FFR fractional flow reserve, LM left main coronary artery, N number of patients, QFR quantitative flow ratio, RCA right coronary artery, V number of vessels.

Table 1.

Baseline and procedural characteristics.

Baseline characteristics
N (patient) 289
Male (%) 211 (73.0)
Age (years) 64.1 (10.4)
BMI (kg/m2) 27.4 (4.1)
Previous PCI (%) 45 (15.6)
eGFR (ml/min/1.73 m2) (median (IQR)) 86.1 (70.1–96.7)
Risk factors
Hypercholesterolemia (%) 196 (67.8)
Smoking (current or past) (%) 136 (47.1)
Diabetes (%) 53 (18.3)
Family history of CAD (%) 151 (53.0)
Hypertension (%) 213 (73.7)
Risk factors ≥ 3 150 (51.9)
Procedural characteristics
N (vessel) 381
Lesions (%)
 RCA 195 (51.2)
 Left main stem 7 (1.8)
 LAD 117 (30.7)
 LCx 35 (9.2)
 Diagonal branch 13 (3.4)
 Other 14 (3.7)
Access site a. radialis (%) 238 (62.5)
Contrast (ml) [median (IQR)] 115 (85.0–130.0)
Diameter stenosis (%) 48.5 (9.4)
Lesion length (mm) 20.3 (12.7–24.1)
Minimal lumen diameter (mm) 1.35 (0.38)
Reference diameter (mm) 2.63 (0.57)
Area stenosis (%) [median (IQR)] 62.05 (52.92–71.05)
FFR 0.81 (0.08)
FFR ≤ 0.80 (%) 169 (44.4)
FFR 0.75–0.85 176 (46.2)
QFR 0.82 (0.09)
QFR ≤ 0.80 (%) 172(45.1)
QFR 0.75–0.85 177 (46.5)

Variables are reported as means (standard deviation), unless otherwise specified.

BMI body mass index, CAD coronary artery disease, eGFR estimated glomerular filtration rate, FFR fractional flow reserve, IQR inter quartile range, LAD left anterior descending, LCx left circumflex artery, OM branch; obtuse marginal branch, PCI percutaneous coronary intervention, QFR quantitative flow ratio, RCA right coronary artery.

Diagnostic performance

Sensitivity, specificity, NPV, PPV and the accuracy of QFR were 84.6%, 86.3%, 87.6%, 83.1% and 85.6%, whereas the diagnostic test results for the QFR–FFR hybrid approach improved to 88.0%, 92.9%, 86.8%, 93.6% and 90.3%, respectively (Table 2). A scatterplot of QFR and FFR showed a good correlation of 0.72 (p < 0.001). The Bland–Altman plot showed a small bias of 0.007 ± 0.058 of QFR. The area under the ROC curve of QFR was 0.89 and improved to 0.92 with the QFR–FFR hybrid approach. The accuracy plot showed a high accuracy for QFR analysis except for FFR-values around the cut-off 0.80 (Fig. 3).

Table 2.

Diagnostic test results quantitative flow ratio (QFR).

% Diameter stenosis QFR ≤ 0.80 QFR–FFR Hybrid approach (QFR: 0.77–0.87)
Estimate 95% CI Estimate 95% CI Estimate 95% CI
True positive 101 154 103
False positive 57 10 7
False negative 68 15 14
True negative 155 202 92
Sensitivity 59.8 52.0 67.2 84.6 78.4 89.3 88.0 80.9 92.7
Specificity 73.1 66.6 79.0 86.3 81.0 90.3 92.9 86.1 96.5
NPV 69.5 65.1 73.6 87.6 82.4 91.4 86.8 79.0 92.0
PPV 63.9 57.9 69.6 83.1 76.8 88.0 93.6 87.4 96.9
Accuracy 67.2 62.2 71.9 85.6 81.7 88.7 90.3 85.6 93.6
AUROC 0.72 0.89 0.92

The diagnostic performance of ICA based on diameter stenosis and QFR with fractional flow reserve (FFR) as reference standard. Both, FFR ≤ 0.80, diameter stenosis ≥ 50% QFR ≤ 0.80 and FFR ≤ 0.80—intermediate zone QFR (0.77–0.87) are used as diagnostic cut-off values.

95% CI 95% confidence interval, AUROC area under the receiver operator characteristic curve, NPV negative predictive value, PPV positive predictive value.

Figure 3.

Figure 3

The per-vessel correlation and agreement between QFR and FFR. (A) The per-vessel scatterplot shows a good correlation between QFR and FFR (r = 0.72, p < 0.001). (B) The Bland–Altman plot shows a small bias of 0.007 ± 0.058 of QFR versus FFR. (C) Per-vessels agreement by plotting agreement against the average of invasive FFR and QFR value. QFR shows a high accuracy except for FFR-values around the cut-off 0.80 (0.75–0.85). (D) The per-vessels receiver operating characteristic curves (ROC-curves) for QFR and QFR–FFR hybrid approach. The area under the curve was higher for QFR–FFR hybrid approach (AUC = 0.92) versus QFR (AUC = 0.89).

QFR–FFR hybrid strategy

The diagnostic accuracy of a QFR–FFR hybrid strategy using the pre-specified intermediate zones cut-offs of QFR < 0.77 (positive predictive value of 93.6%) and QFR > 0.87 (negative predictive value of 86.8%), on which coronary revascularization decisions could have been made without invasive FFR measurement was 90.3% (Table 2). The use of the proposed QFR–FFR hybrid strategy would have resulted in a reduction in the use of invasive FFR of 56.7% (Fig. 4). When aiming for an accuracy of QFR of 90% (intermediate zone between 0.79 and 0.83), the proportion of patients in whom invasive FFR could have safely been omitted increased to 88.7%.

Figure 4.

Figure 4

Hybrid revascularization strategy with quantitative flow ratio (QFR) and fractional flow reserve (FFR) reduces the requirement of FFR in clinical practice. (A) QFR limits for the intermediate zone on which coronary revascularization decisions can be made without invasive FFR measurement when QFR is < 0.77 (positive predictive value of 93.9%) or when QFR is > 0.87 (negative predictive value of 93.1%). When the QFR-values are between 0.77 and 0.87, invasive FFR is measured and a threshold of 0.80 is applied to guide revascularization. The dots inside the red rectangles represents the false positive and false negative QFR measurements. The dots inside the intermediate zone represent the stenosis that should be classified by FFR. (B) A QFR–FFR hybrid approach reduces the number of patients requiring invasive FFR measurements. An accuracy of QFR to FFR of 90% (intermediate zone between 0.79 and 0.83) reduces the invasive FFR requirement by 88.7%.

Discussion

In this multicentre study, we determined the diagnostic accuracy of QFR versus FFR. A high diagnostic accuracy of 85.6% was found, while having minimal bias and sufficient discrimination. This study also showed that a QFR–FFR hybrid approach improves the revascularization decision-making strategy over QFR alone. This hybrid approach has the potential to reduce the number of invasive FFR-measurements by more than half, while maintaining a high accuracy-level (accuracy of 90.3%).

Diagnostic performance

Several studies have been performed to assess the diagnostic performance of QFR. The first prospective study, FAVOR pilot, included 84 vessels in 73 patients. Significant stenoses were found in 27 vessels (32.1%)11. A good correlation (r = 0.77, p < 0.001) and agreement between FFR and QFR (mean difference = 0.001, SD 0.059) were found. Diagnostic performance measurements as sensitivity, specificity, PPV, NPV and accuracy on a per vessel basis of 74%, 91%, 80%, 88% and 86% were observed. The subsequent prospective multicentre FAVOR II studies (performed in China, Japan and Europe)8,9 also reported good correlations (r = 0.86, p < 0.001 and r = 0.83, p < 0.001 ) and small mean difference between FFR and QFR (− 0.01 SD 0.063 and 0.01 SD 0.06). The sensitivity, specificity, PPV, NPV and accuracy on a per vessel basis were 95%, 92%, 86%, 97% and 93% in the FAVOR II China and 87%, 87%, 76%, 93% and 87% in the European and Japanese cohorts. The WIFI II, a prospective multicentre sub-study of the Dan-NICAD study, included 172 patients and 255 lesions of which 28% of the lesions were functionally significant10. A correlation of r = 0.70 (p < 0.001) was observed and precision with a mean difference of 0.01 (SD 0.08). The sensitivity, specificity, PPV, NPV and accuracy on a per vessel basis were 77%, 86%, 75%, 87% and 83%. Our study further corroborates the diagnostic accuracy found in these prior studies.

QFR–FFR hybrid approach

In the FAVOR II Europe-Japan study, a QFR intermediate zone between 0.77 and 0.87 was proposed, which formed the basis for the present study8. FFR assessment could have been avoided in 64% when using the hybrid strategy. Similar results were found in the WIFI II study where an intermediate zone of 0.78–0.87 would have saved pressure wires and adenosine in 68% of the lesions10 and by Smit et al. who proposed an intermediate zone with cut-off limits 0.77–0.86 resulting in an accuracy of 93.8% and a FFR reduction of 70.1%14. In the current study, we validated the 0.77–0.87 intermediate zone in an independent cohort and found a slightly lower rate of avoided FFR (56.7%), and comparable diagnostic accuracy(93.8% accuracy in this study versus 95% in the FAVOR II study) This might be caused by the difference in proportion FFR-values around the cut-off (0.75–0.85) of 32% in the FAVOR II study versus 46.2% in the present study8,9. A second explanation could be found in the different proportion of the measured vessels, whereas the RCA was more often included in the present study (50%) as compared with the FAVOR II study (22%).

Clinical implications and perspectives

In this study we showed that the proportion of patients in whom invasive FFR measurements can be avoided and the accuracy both depend on the window of the QFR limits; the wider the intermediate zone, the higher the overall accuracy and the lower the proportion of patients in whom invasive FFR might be safely omitted (Fig. 5). An overall accuracy of 95% using an intermediate zone of 0.07 (0.74–0.86) could reduce the proportion of patient in whom FFR measurements can be omitted to 47.2%. Previously, FFR-guided revascularization has been shown to reduce stent implantations and improve long-term outcome when compared to angiography-guided strategy46. However, in routine clinical practice FFR is not used very often due to need for hyperaemia induction and a pressure-wire7. Hyperaemia might cause patient symptoms and prolongation of the procedural time and the use of a pressure-wire is costly, might lead to procedural complications and can be challenging in tortuous arteries. Other limitations of FFR are waveform artefacts or drift that impacts the accuracy of FFR17. Instantaneous wave-free ratio (iFR) overcomes the need of hyperaemia and its negative side effects, but does require invasive measurements.

Figure 5.

Figure 5

Effect of hybrid QFR–FFR strategies on the population free from FFR and the overall accuracy. Population free from invasive FFR measurements and the overall agreement depends on the size of the intermediate zone: the larger the window of QFR values, the higher the overall accuracy (upper panel). However, this decreases the percentage of patients free from invasive FFR (lower panel).

One of the benefits of QFR is that a virtual FFR can be computed in approximately 5 min (IQR 3.5–6.1) minutes compared to the average of 7 min needed to perform FFR (IQR: 5.0–10.0 min)8,9. Moreover, QFR-computation does not require induction of hyperaemia, the hazard of passing an intracoronary wire, or additional equipment, training, or cost for measuring invasive FFR. A QFR–FFR hybrid strategy could potentially increase the adoption of QFR by retaining a high accuracy while decreasing the number of invasive FFR measurements, especially whilst awaiting randomized outcome trials. One downside of the hybrid QFR–FFR approach would be that in addition to the time required to perform FFR, an additional 5 min of QFR time is needed (total time 13 min) in the case of an QFR-value within the intermediate zone.

Another application of QFR is as gatekeeper for hospitals not capable to perform FFR for referrals to hospitals where FFR and PCI can be performed. Currently these hospitals assess the severity of a lesions visually, although visual assessment alone is known to be inaccurate for the assessment of functional significant CAD14,18. It has been shown that QFR has a high diagnostic accuracy as well as a high NPV and PPV for the diagnosis of functionally significant CAD. This suggests that QFR could be safely used as gatekeeper. However, randomized studies are needed to confirm the diagnostic value of QFR for this purpose.

Limitations

This study has several limitations inherent to its retrospective character such as selection bias, which might influence the feasibility and reliability of the QFR analysis. Most limitations are related to imaging quality; a substantial number of patients (14.7%) could not be included in the analysis due to suboptimal projections and/or lack of good quality images. It can be assumed that a prospective study would have less exclusions as the clinician would be more focused on optimal image quality suitable for QFR analysis. Moreover, the QFR analysis could not be applied to ostial lesions and bypass grafts resulting in the exclusion of approximately 10% of the patients and therefore decreasing the applicability of the software. Another limitation could be found in the small differences in vessel characteristics in our study compared to previous studies. Most of target vessels in this study were the right coronary artery, whereas in most of studies reporting on QFR the most common target vessel was the left anterior descending coronary artery. Although our study is performed retrospectively, the vessel characteristics and diagnostic test results are comparable to those observed in prospective studies as the FAVOR II China, Europe-Japan and the WIFI II implicating the limited effect of the differences in vessel characteristics. More importantly, the patients and FFR-measurement included in this study were not specifically performed for study purposes and therefore reflect the clinical practice and the potential benefits of physiological FFR-measurements in addition to visual assessment.

Conclusion

In conclusion, QFR has a good correlation and agreement with invasive FFR and a high diagnostic performance. With a QFR–FFR hybrid approach, QFR can effectively reduce the number of invasive FFR-measurements with high PPV and NPV. In this way, FFR can be used for equivocal cases only. A hybrid method could potentially increase the adoption of QFR and expand the utilization of physiology-guided decision making in clinical practice while awaiting the results of clinical outcome studies.

Abbreviations

3D-QCA

Three dimensional quantitative coronary angiography

AUROC-curve

Area under the receiver operating characteristic-curve

CAD

Coronary artery disease

CT

Computed tomography

ECG

Electrocardiography

FFR

Fractional flow reserve

ICA

Invasive coronary angiography

iFR

Instantaneous wave-free ratio

IQR

Inter quartile range

LAD

Left anterior descending artery

LCx

Left circumflex artery

LM

Left main

NPV

Negative predictive value

PCI

Percutaneous coronary intervention

PPV

Positive predictive value

QFR

Quantitative flow ratio

RCA

Right coronary artery

ROC-curve

Receiver operating characteristics curve

SD

Standard deviation

TIMI

Thrombolysis in myocardial infarction

Author contributions

All authors have made substantial contribution to the study; J.P., R.W.v.H., T.L and M.J.S were involved in conception, design, analysis, and interpretation of data and writing the manuscript. All authors were involved in revising the manuscript critically for important intellectual content and gave final approval of the manuscript submitted.

Funding

The authors received no specific funding for this work.

Data availability

The (anonymized) clinical data, methods used in the analysis, and materials used to conduct the research can be requested by qualified researchers who engage in independent scientific research, and could be provided following review and approval of a research proposal. Data requests can be submitted at any time by contacting the corresponding author.

Competing interests

The authors declare no competing interests.

Footnotes

Publisher's note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

References

  • 1.Force T, et al. 2013 ESC guidelines on the management of stable coronary artery disease The Task Force on the management of stable coronary artery disease. Eur. Heart J. 2018;38:2949–3003. doi: 10.1093/eurheartj/eht296. [DOI] [PubMed] [Google Scholar]
  • 2.Fihn SD, et al. 2014 ACC/AHA/AATS/PCNA/SCAI/STS focused update of the Guideline for the diagnosis and management of patients with stable ischemic heart disease. A Report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines, an. Circulation. 2014;130:1749–1767. doi: 10.1161/CIR.0000000000000095. [DOI] [PubMed] [Google Scholar]
  • 3.Toth G, et al. Evolving concepts of angiogram: Fractional flow reserve discordances in 4000 coronary stenoses. Eur. Heart J. 2014;35:2831–2838. doi: 10.1093/eurheartj/ehu094. [DOI] [PubMed] [Google Scholar]
  • 4.Van Nunen LX, et al. Fractional flow reserve versus angiography for guidance of PCI in patients with multivessel coronary artery disease (FAME): 5-year follow-up of a randomised controlled trial. Lancet. 2015;386:1853–1860. doi: 10.1016/S0140-6736(15)00057-4. [DOI] [PubMed] [Google Scholar]
  • 5.Pijls NHJ, et al. Percutaneous coronary intervention of functionally nonsignificant stenosis. J. Am. Coll. Cardiol. 2007;49:2105–2111. doi: 10.1016/j.jacc.2007.01.087. [DOI] [PubMed] [Google Scholar]
  • 6.Pijls NHJ, et al. Fractional flow reserve versus angiography for guiding percutaneous coronary intervention in patients with multivessel coronary artery disease: 2-year follow-up of the FAME (fractional flow reserve versus angiography for multivessel evaluation) study. J. Am. Coll. Cardiol. 2010;56:177–184. doi: 10.1016/j.jacc.2010.04.012. [DOI] [PubMed] [Google Scholar]
  • 7.Dattilo PB, Prasad A, Honeycutt E, Wang TY, Messenger JC. Contemporary patterns of fractional flow reserve and intravascular ultrasound use among patients undergoing percutaneous coronary intervention in the United States insights from the National Cardiovascular Data Registry. J. Am. Coll. Cardiol. 2012;60:2235–2242. doi: 10.1016/j.jacc.2012.08.990. [DOI] [PubMed] [Google Scholar]
  • 8.Westra J, et al. Diagnostic performance of in-procedure angiography-derived quantitative flow reserve compared to pressure-derived fractional flow reserve: The FAVOR II Europe-Japan study. J. Am. Heart Assoc. 2018;7:20. doi: 10.1161/JAHA.118.009603. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9.Xu B, et al. Diagnostic accuracy of angiography-based quantitative flow ratio measurements for online assessment of coronary stenosis. J. Am. Coll. Cardiol. 2017;70:3077–3087. doi: 10.1016/j.jacc.2017.10.035. [DOI] [PubMed] [Google Scholar]
  • 10.Westra J, et al. Evaluation of coronary artery stenosis by quantitative flow ratio during invasive coronary angiography: The WIFI II study (wire-free functional imaging II) Circ. Cardiovasc. Imaging. 2018;11:e007107. doi: 10.1161/CIRCIMAGING.117.007107. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11.Tu S, et al. Diagnostic accuracy of fast computational approaches to derive fractional flow reserve from diagnostic coronary angiography. JACC Cardiovasc. Interv. 2016;9:2024–2035. doi: 10.1016/j.jcin.2016.07.013. [DOI] [PubMed] [Google Scholar]
  • 12.Collet C, et al. Diagnostic performance of angiography-derived fractional flow reserve: A systematic review and Bayesian meta-analysis. Eur. Heart J. 2018;39:3314–3321. doi: 10.1093/eurheartj/ehy445. [DOI] [PubMed] [Google Scholar]
  • 13.Yazaki K, et al. Applicability of 3-dimensional quantitative coronary angiography-derived computed fractional flow reserve for intermediate coronary stenosis. Circ. J. 2017;81:988–992. doi: 10.1253/circj.CJ-16-1261. [DOI] [PubMed] [Google Scholar]
  • 14.Smit JM, et al. Referral of patients for fractional flow reserve using quantitative flow ratio. Eur. Hear. J. Cardiovasc. Imaging. 2018;10:1–8. doi: 10.1093/ehjci/jey187. [DOI] [PubMed] [Google Scholar]
  • 15.Knuuti J, et al. 2019 ESC Guidelines for the diagnosis and management of chronic coronary syndromes. Eur. Heart J. 2019;20:1–71. doi: 10.1093/eurheartj/ehz425. [DOI] [PubMed] [Google Scholar]
  • 16.Tu S, et al. Fractional flow reserve calculation from 3-dimensional quantitative coronary angiography and TIMI frame count. JACC Cardiovasc. Interv. 2014;7:768–777. doi: 10.1016/j.jcin.2014.03.004. [DOI] [PubMed] [Google Scholar]
  • 17.Matsumura M, et al. Accuracy of fractional flow reserve measurements in clinical practice: Observations from a core laboratory analysis. JACC Cardiovasc. Interv. 2017;10:1392–1401. doi: 10.1016/j.jcin.2017.03.031. [DOI] [PubMed] [Google Scholar]
  • 18.Park S-J, et al. Visual-functional mismatch between coronary angiography and fractional flow reserve. JACC Cardiovasc. Interv. 2012;5:1029–1036. doi: 10.1016/j.jcin.2012.07.007. [DOI] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Data Availability Statement

The (anonymized) clinical data, methods used in the analysis, and materials used to conduct the research can be requested by qualified researchers who engage in independent scientific research, and could be provided following review and approval of a research proposal. Data requests can be submitted at any time by contacting the corresponding author.


Articles from Scientific Reports are provided here courtesy of Nature Publishing Group