Abstract
Background
Although 7T functional MRI (fMRI) provides better signal‐to‐noise ratio and higher spatial resolution than 3T fMRI, geometric distortions become more challenging because fMRI is more susceptible to distortions than structural MRI. Accurate alignment of 7T fMRI to structural MRI data is critical for precise cortical surface‐based analysis.
Purpose
To quantify the effectiveness of distortion corrections of 7T fMRI data.
Study Type
Prospective.
Subjects
Fifteen healthy individuals aged 19–26 years (mean: 21.9 years).
Field Strength/Sequence
Multiband gradient‐echo echo‐planar imaging sequence at 7T; 3D T1/T2‐weighted sequences (magnetization prepared rapid acquisition with gradient echo [MPRAGE] and sampling perfection with application optimized contrast using different flip angle evolution [SPACE]) at 3T.
Assessment
fMRI data at 7T were registered to cortical surfaces reconstructed from 3T structural data acquired in the same subjects. Distortions induced by B0 inhomogeneity and gradient nonlinearity (B0 and gradient distortions) were evaluated as cortical fallout (misregistration of noncortical areas) and displacement (misregistration along gray matter).
Statistical Tests
Repeated measures analyses of variance with post‐hoc t‐tests with Bonferroni correction.
Results
The accuracy of fully corrected fMRI images based on the intensity distribution was 89.2%. Without any corrections, 9.7% of vertices in the whole surfaces were fallout and the average displacement was 0.96 mm for the rest of the vertices. B0 and gradient distortion corrections significantly reduced the fallout (to 2.1% and 8.7%) and displacement (to 0.29 mm and 0.86 mm). These corrections were effective even around regions with moderate distortions (the somatosensory and visual cortices for B0 distortion, and the anterior frontal, inferior temporal, and posterior occipital cortices for gradient distortion).
Data Conclusion
B0 distortion correction is crucial for surface‐based analysis of fine‐resolution fMRI at 7T. Gradient distortion correction should be considered when regions of interest include regions distant from the isocenter of scanners.
Evidence Level
1
Technical Efficacy Stage
1
Keywords: geometrical distortion, B0 inhomogeneity, gradient nonlinearity, ultra‐high field MRI, cortical surface‐based analysis
ULTRA‐HIGH FIELD (UHF) magnetic resonance imaging (MRI) allows us to acquire high‐resolution blood oxygenation‐level dependent (BOLD) functional images by echo‐planar imaging (EPI) 1 with better contrast 2 , 3 and signal‐to‐noise ratio (SNR) 4 , 5 compared with conventional field strength (≤3T). UHF also allows imaging with increased spatial resolution, which reduces partial volume effects, thus providing better cortical separation from the surrounding white matter, cerebrospinal fluid (CSF), and pial veins. Spatial specificity of detected brain activity in BOLD functional (f)MRI is also improved at 7T 6 and it is estimated that the full‐width at half‐maximum of the average point‐spread function is below 1 mm. These advantages have helped in developing cortical column‐ 7 and layer‐based fMRI analyses. 1 , 8 , 9 , 10 These anatomically informed analyses 11 depend on a cortical ribbon defined by high‐resolution structural images. Thus, an appropriate correction of BOLD functional image distortions is critical for precise cross‐modal registration.
Because of the higher resonance frequency, UHF‐fMRI is more susceptible to static magnetic field (B0) inhomogeneity resulting from differences in magnetic permeability of human brain tissues. Distortion induced by B0 inhomogeneity (B0 distortion) is seen around the nasal cavity and ear canals 12 , 13 in high‐resolution EPI because of the narrow acquisition bandwidth (BW). 14 Additionally, its appearance is strongly dependent on the phase‐encoding (PE) direction. Conversely, distortion induced by gradient magnetic field nonlinearity (gradient distortion) gradually increases with distance from the isocenter of an MRI scanner. These distortions result in misregistration of EPI to structural images and reduce the accuracy of surface‐based analysis in which the signals in the cortex are selectively sampled and analyzed. In particular, signal sampling at incorrect locations at the surface affects the geodesic distances between points on the cortex. These are neurobiologically more meaningful than simple 3D Euclidean distances, as the cortex is like a complexly folded sheet. 15 The distortion correction of fMRI images is, therefore, an essential preprocess for surface‐based analysis in UHF‐MRI.
B0 inhomogeneity can be estimated and corrected by traversing k‐space twice with different PE directions 14 rather than using the conventional field map method. 12 This method of correction is available as the tool topup in FSL (FMRIB [Functional MRI of the Brain] Software Library, Oxford, UK) 14 , 16 (https://fsl.fmrib.ox.ac.uk/fsl/fslwiki/topup). Gradient distortion can be corrected by a voxel displacement map calculated from a gradient coefficient file describing the characteristic of gradient coils as a spherical harmonics expansion. 17 , 18 This method of correction is available as the tool gradunwarp in FreeSurfer (https://surfer.nmr.mgh.harvard.edu/fswiki/GradUnwarp). Recently, the minimum preprocessing pipelines for the Human Connectome Project (HCP) 19 , 20 have incorporated both topup and gradunwarp, enabling systematic correction for EPI distortions. However, there has been no quantitative analysis of the effects of 7T EPI image distortion corrections on surface‐based analysis in a way that reflects how often noncortical signals are mapped and how far cortical signals are incorrectly mapped onto the surface.
The aim of this study was to quantitatively evaluate the correction of image distortions induced by B0 inhomogeneity and gradient nonlinearity on surface‐based analysis in 7T fMRI data.
Materials and Methods
We first evaluated the accuracy of fully corrected fMRI images based on the intensity distribution of fMRI image voxels within the cortical ribbon defined by the cortical surfaces reconstructed from structural images through HCP Pipelines. In the second step, we quantified the effects as cortical fallout (misregistration of EPI image voxels to the outside of the gray matter in an anatomical image) and cortical displacement (distance between the distorted and corrected EPI image voxels within the surface) by regarding the fully corrected images as ideally corrected ones that corresponded to the structure‐based surfaces and by comparing these surfaces with those replicating fMRI image distortions.
Subjects
This study was approved by the Ethical Committee of the institute and all subjects provided written informed consent. Fifteen healthy individuals aged 19–26 years (eight men and seven women; mean age = 21.9 years; standard deviation [SD] = 1.6 years) participated in the study. None of the subjects had a history of symptoms requiring neurological, psychological, or other medical care.
Data Acquisition
FUNCTIONAL DATA ACQUISITION WITH 7T MRI
Functional data were acquired with a 32‐channel phased‐array head coil on a Magnetom 7T scanner (Siemens Healthcare, Erlangen, Germany). For whole‐brain coverage at high spatial and typical temporal resolutions, the multiband (MB) gradient‐echo (GRE) EPI sequence 21 was used in conjunction with the generalized autocalibrating partially parallel acquisition (GRAPPA) technique 22 with repetition time (TR) = 2000 msec, echo time (TE) = 25 msec, flip angle (FA) = 65°, BW = 1,562 Hz/pixel, in‐plane field of view (FOV) = 192 × 192 mm2, matrix = 160 × 160, 128 slices tilted at 15–20° to the transversal plane, 1.2‐mm isotropic voxels, echo spacing (ES) = 0.74 msec, MB acceleration factor = 4, PE undersampling factor GRAPPA = 3, and an anterior‐to‐posterior (AP) PE direction. In this sequence, a single‐band (SB) GRE EPI image was also acquired using a separate SB excitation of the same slices (TR = 2000 msec; volume TR = 8000 msec), with otherwise identical parameters, including PE undersampling factor of 3. This image was used for the evaluations because of its better image quality than MB images and identical distortion to MB images.
Two spin‐echo (SE) EPI datasets with reversed PE directions were also acquired (three volumes with each PE direction) with the same geometric and ES parameters (TR = 11,900 msec, TE = 60 msec, FA = 90°, refocus FA = 180°) as the GRE EPI images. These images, which were referred to as SE field maps in the HCP, 20 were used to correct B0 distortion with FSL's topup tool using a method similar to that in Anderson et al 14 (explained in more detail below).
STRUCTURAL DATA ACQUISITION WITH 3T MRI
Image intensity is substantially inhomogeneous at 7T due to less uniform excitation (B1 inhomogeneity) and larger dielectric effects, especially in the temporal base area. As B0 and B1 inhomogeneities are less problematic at lower field strength, the structural data for subjects' cortical surface reconstructions were acquired with a 32‐channel phased‐array head coil on a Magnetom Verio 3T scanner (Siemens Healthcare). Two T1‐weighted (T1w) and two T2‐weighted (T2w) whole‐brain images were acquired at 0.8‐mm isotropic resolution in each subject to process with HCP Pipelines. 20 These images were aligned and averaged in each of the T1w and T2w images to improve the image SNR. T2w images were used not only for correcting for bias fields but also for removing veins and dura that were not adequately contrasted with gray matter in T1w images. The 3D magnetization prepared rapid acquisition with GRE (MPRAGE) sequence 23 was used for T1w images (TR = 2400 msec, TE = 2.24 msec, inversion time [TI] = 1060 msec, in‐plane FOV = 256 × 240 mm2, matrix = 320 × 300, 224 sagittal slices in a single slab, FA = 8°, BW = 210 Hz/pixel, ES = 8.1 msec, PE undersampling factor GRAPPA = 2, and an AP PE), while the 3D sampling perfection with application optimized contrast using different FA evolution (SPACE) sequence 24 was used for T2w images without a fat suppression pulse (same in‐plane FOV, matrix, slices, and PE direction as in the T1w images, TR = 3200 msec, TE = 560 msec, BW = 744 Hz/pixel, ES = 3.6 msec, PE undersampling factor GRAPPA = 2, total turbo factor = 334, and echo train length = 1,156). Similarly, two SE field maps with reversed PE directions were also acquired for readout distortion correction with FSL's FUGUE tool (TR = 7700 msec, TE = 60 msec, in‐plane FOV = 196 × 196 mm2, matrix = 98 × 98, 72 transversal slices, FA = 78°, refocus FA = 160°, BW = 1,822 Hz/pixel, and ES = 0.73 msec).
Software
HCP Pipelines 4.0.0‐alpha.5, 20 Connectome Workbench 1.2.3 or 1.3.2, FSL 5.0.10, FreeSurfer 5.3.0‐HCP, and HCP gradunwarp tool 1.0.3 running on a 64‐bit Linux Operating System (Ubuntu 16.04 LTS or CentOS 6.8 or 6.9) were used for preparation before evaluation of distortion correction such as EPI image distortion correction, registration to structural images, cortical surface reconstruction, and creation of distorted EPI images and distortion‐replicated surfaces. MatLab R2017b (MathWorks, Natick, MA) was used to write home‐made scripts for statistical tests, judgment of cortical fallout of vertices, and measurement of cortical displacements.
Preparation for Evaluation of Effects of Distortion Corrections
CREATION OF FULLY CORRECTED ECHO‐PLANAR IMAGES
We created fully corrected EPI images through HCP's fMRI volume preprocessing pipeline 20 (zone surrounded with an orange dotted line in Fig. 1). A warp field to correct gradient distortion of an SB GRE EPI image was first obtained using HCP gradunwarp tool (black W G in Fig. 1), which was part of FreeSurfer. This calculated the amount of gradient distortion based on a scanner‐specific gradient coefficient file. 18 Gradunwarp was also applied to the two SE EPI images with reversed PE directions (W G_SEAP and W G_SEPA in Fig. 1). The amount of B0 distortion was next estimated from these gradient distortion‐corrected SE EPI images using FSL's topup tool 14 (black W B0 in Fig. 1). The resulting warp fields were then sequentially applied to a raw SB GRE EPI image to fully correct the image with the intensity correction using Jacobian modulation. 14 Rigid‐body transformation was used to remove a misalignment between SE and GRE EPI images due to subject's head motion in discrete runs (black T R in Fig. 1). For registration of the corrected SB GRE EPI image to a T1w one (black T A in Fig. 1), a transform matrix was calculated by sequentially using two registration tools: FSL's BBR and FreeSurfer's BBRegister. 25 Instead of rigid‐body transform, affine transform was used for the registration to minimize interscanner gradient distortion correction (GDC) errors, 26 as we used 3T and 7T scanners for structural and EPI acquisitions, respectively.
CREATION OF DISTORTED ECHO‐PLANAR IMAGES
For the evaluation of distortion corrections, we prepared three types of distorted SB GRE EPI images for each subject: uncorrected, gradient distortion‐corrected, and B0 distortion‐corrected. A raw SB GRE EPI image was directly used as an uncorrected EPI image (green bold arrow in Fig. 1). For a gradient distortion‐corrected image, a corrected image just after applying gradunwarp to a raw image was utilized (blue bold arrow in Fig. 1). A B0 distortion‐corrected image was created by skipping GDC and applying only a warp field output by topup to a raw image (gray and red bold arrows in Fig. 1). The intensity of these partially distortion‐corrected images was corrected in a similar way to the fully corrected images described above. A transform matrix that the fMRI volume preprocessing pipeline normally outputs to register a fully corrected (undistorted) SB GRE EPI image was not used (black T A in Fig. 1) for registrations of these distorted images to the structural data because the estimated transform matrix (black T A) is influenced by the preceding distortion correction procedures. An affine transform matrix calculated for each distorted image (uncorrected, gradient distortion‐corrected, and B0 distortion‐corrected) was employed for conversion from distorted EPI image space to structural image space (green, blue, and red “T A”s in Fig. 1, respectively).
CREATION OF DISTORTION‐REPLICATED CORTICAL SURFACES AND RIBBONS
Undistorted native cortical surfaces were created from the 3T structural images through a series of the original HCP's surface reconstruction pipelines, 20 where gradient and readout distortions were corrected. Meanwhile, surfaces where EPI image distortions were replicated for each distortion condition in each subject were generated as follows. To distort the undistorted native surface, inverse warp fields, upsampled to the same resolution as the structural images (0.8 mm isotropic resolution), were first made from the warp fields for distortion corrections. A warp field for creating B0 distortion‐replicated (gradient distortion‐corrected) surfaces was generated by concatenating the inverse warp field to replicate B0 distortion, inverse affine and rigid‐body transform matrices to convert from the structural image space to the SE EPI space where the inverse warp field existed, and rigid‐body and inverse affine transform matrices to convert back to the structural image space (Fig. 2b). Then, by applying this new warp field to the undistorted native surfaces with the command surface‐apply‐field of the Connectome Workbench, surfaces replicating the B0 distortion of the EPI images were created. Similarly, a warp field for creating gradient distortion‐replicated (B0 distortion‐corrected) surfaces was generated by concatenating the inverse warp field to replicate gradient distortion, inverse affine transform matrix to convert from the structural image space to the GRE EPI space where the inverse warp field existed, and the affine transform matrix to convert back to the structural image space (Fig. 2c). Moreover, for a warp field creating surfaces where distortions of uncorrected EPI images were replicated, all of these transform matrices and inverse warp fields were concatenated (Fig. 2a). Identical procedures were applied to the undistorted cortical ribbons to generate distortion‐replicated cortical ribbons.
Evaluation of Distortion Correction Effects
CALCULATION OF ECHO‐PLANAR IMAGE DISTORTION
The absolute amounts of B0 and gradient distortions of the EPI images were calculated from the x, y, and z components of the warp field output by topup for B0 distortion correction (B0DC) and gradunwarp for GDC in each subject, respectively. The affine transform matrix for the registration of the fully corrected SB EPI image to the T1w image was applied to these calculated data to overlay onto the T1w image. The rigid‐body transform matrix to convert the SE EPI space to the GRE EPI space was also applied to the calculated data for the B0 distortion, as the warp field for the calculation was in the SE EPI space.
EVALUATION OF ACCURACY OF FULLY CORRECTED ECHO‐PLANAR IMAGES
The overall efficiency of the full distortion correction was evaluated before the surface‐based evaluations. Fully corrected EPI images were regarded as perfectly correct (corresponding to the structural images) and were thus used as a reference against other partially corrected and uncorrected ones in the surface‐based evaluation.
Voxels in the cortical ribbon defined by the structural images were first extracted from the fully corrected SB GRE EPI image in each subject to generate a normalized intensity histogram of these voxels as a probability density. Given that all these voxels are of the gray matter, the shape of the histogram should have a single peak with a normal probability density function. However, this may not be the case because of white matter and CSF contamination resulting from imperfect distortion correction, and intensity dropout, which is inevitable in EPI images. Moreover, the partial volume effect may broaden the bottom of the single peak. Therefore, considering these factors, we fitted the probability density function with four normal distribution functions for the four components: gray and white matter, CSF, and intensity dropout using MatLab's mle function. Maximum likelihood estimation was performed to estimate 12 parameters: weights, means, and SDs for these four normal distribution functions. An estimated intensity histogram F(x) as probability density for normalized intensity x was calculated according to the following equations:
where f(x, μ, σ) is a normal distribution function for the average μ and SD σ, μ g, μ w, μ f, μ d, σ g, σ w, σ f, and σ d are intensity averages and SDs of the gray and white matter, CSF, and dropout, respectively, and w g, w w, w f, and w d are weights for the same four components.
To confirm that the fitting was valid, simulated spatial distribution maps for the four components were created for each subject. Voxels in the cortical ribbon were classified into four categories based on the frequency ratio of the four components in each bin (0.1 SD). For example, when a bin is composed of 1000 voxels and the frequency ratio of the gray matter is 0.8, 800 voxels of these voxels were randomly labeled as gray matter.
CALCULATION OF CORTICAL FALLOUT AND CORTICAL AND EUCLIDEAN DISPLACEMENTS
A Euclidean displacement map was obtained by calculating spatial distances between corresponding vertices of undistorted and distorted surfaces and mapping them onto the undistorted surfaces for each distortion condition in each subject.
We needed to determine whether each vertex of the distorted “midthickness” surface, which was the intermediate surface between “white” and “pial” surfaces of FreeSurfer, was located in or out of the undistorted cortical ribbon for each distortion condition in each subject. Therefore, x, y, and z coordinate values of the vertex were first obtained in the native volume space after the “acpc alignment” step. 20 The vertex was classified as a fallout, when the ribbon did not contain the vertex (Fig. 3a). Meanwhile, for the vertex contained in the ribbon (surviving), we searched the vertex nearest to the coordinate values on the distorted surface and the vertex correspondent to the nearest vertex on the undistorted surface (Fig. 3b). A cortical distance between these vertices was measured and mapped onto the undistorted surface as a cortical displacement. A cortical fallout map was finally obtained by assigning “1”s to fallout vertices on the undistorted surface.
Surface maps of the cortical fallout and cortical and Euclidean displacements for each distortion condition from the 15 subjects were transformed to the 164 k‐CIFTI (Connectivity Informatics Technology Initiative) format 20 with the application of MSMSulc 26 and averaged over the subjects in each vertex to obtain group maps. Group‐averaged maps of the cortical thickness and curvature, which expressed whether the surface was convex or concave, were also calculated with the output data from the HCP structural pipelines. The displacement ratio was defined as the ratio between the cortical and Euclidean displacements of each vertex. The group displacement ratio map was created by computing the group cortical displacement map divided by the group Euclidean map in each distortion condition.
The fallout rate was calculated by dividing the number of fallout vertices by that of the valid ones in the 164 k CIFTI format in each subject and then averaging across subjects in each distortion condition to calculate the group fallout rate. Additionally, the average cortical displacement of each subject was obtained by averaging the values stored in all vertices, except fallout ones, of the surface map in each condition.
Statistical Tests
We applied repeated measures analysis of variance (ANOVA) to each of the 15 subjects' cortical fallout rates and displacements among the three distortion conditions. A Bonferroni correction was applied to correct for multiple comparisons; α was set at 0.0167 for these conditions in post‐hoc paired t‐tests.
Results
Echo‐Planar Image Distortions
Figure 4a shows a comparison of a T1w image (top row) with a raw SB GRE EPI image acquired with the AP PE direction (middle row) in four different slices in a single subject. A distortion‐corrected EPI image with topup is also shown in the bottom row. Mismatches between the gray matter region of the EPI image and trace lines of gray matter defined by the structural images were observed, particularly, around the frontal and occipital poles, and the orbito‐frontal cortex (OFC). The amount of distortions experienced at each position because of gradient nonlinearity and B0 inhomogeneity are shown in Fig. 4b,c, respectively. Gradient nonlinearity induced distortions over 2 mm were seen around the frontal and inferior occipital regions, although the distortions were milder than B0 inhomogeneity‐induced ones. In addition, the inferoposterior part of the cerebellum was distorted by ~4 mm. Meanwhile, the heaviest B0 distortions were observed around the OFC near the nasal cavity and the inferior temporal cortex (ITC) near the ear canal (~10 mm). Distortions over 5 mm were also seen around the occipital pole, and several‐millimeter distortions occurred in many other parts.
Accuracy of Fully Corrected Echo‐Planar Images
The black curve in Fig. 5a indicates the normalized intensity histogram of voxels in the cortical ribbon defined by the structural images as a probability density in a typical subject. Estimated histograms for the four components are also shown with yellow, blue, red, and purple curves, respectively. The simulated distribution of these four components in the ribbon is shown in Fig. 5b. Yellow voxels are prominently distributed in the ribbon with a scattering of blue and red voxels, consistent with the notion that the yellow voxels correspond to gray matter (see also Fig. 5c). As the peak density of the blue component was lower than that of the yellow component, which was in turn lower than the peak of the red component, the blue and red voxels corresponded to white matter and CSF, respectively. A substantial bump seen in the left side of the histogram (purple curve in Fig. 5a) was distributed around the frontal pole, OFC, and ITC where intensity dropout was prominent (Fig. 5c). Overall, these data indicate that the estimated components of the probability density represent the corresponding tissues of gray and white matter, CSF, and intensity dropout.
Table 1 shows the distortion correction accuracy for fully corrected EPI images in all subjects. As the average w d was 0.048, 4.8% of the voxels in the cortical ribbon lost signal in the fully corrected EPI images. The simulation (Fig. 5) showed that voxels labeled as white matter and CSF were diffusely distributed in the ribbon and occupied 1.9% (average w w = 0.019) and 8.3% (average w f = 0.083) of its volume, respectively. Altogether, the average distortion correction accuracy was 89.2 ± 2.0%.
TABLE 1.
Subject number | Gray (w g) | White (w w) | CSF (w f) | Dropout (w d) | Accuracy (%) |
---|---|---|---|---|---|
01 | 0.840 | 0.049 | 0.061 | 0.050 | 88.4 |
02 | 0.887 | 0.011 | 0.054 | 0.048 | 93.1 |
03 | 0.872 | 0.022 | 0.074 | 0.032 | 90.1 |
04 | 0.862 | 0.010 | 0.084 | 0.044 | 90.1 |
05 | 0.827 | 0.055 | 0.081 | 0.037 | 85.9 |
06 | 0.836 | 0.043 | 0.080 | 0.041 | 87.2 |
07 | 0.831 | 0.074 | 0.069 | 0.027 | 85.4 |
08 | 0.846 | 0.000 | 0.098 | 0.057 | 89.7 |
09 | 0.857 | 0.000 | 0.100 | 0.043 | 89.5 |
10 | 0.848 | 0.017 | 0.091 | 0.044 | 88.7 |
11 | 0.838 | 0.005 | 0.099 | 0.058 | 88.9 |
12 | 0.842 | 0.000 | 0.100 | 0.058 | 89.4 |
13 | 0.860 | 0.000 | 0.080 | 0.060 | 91.4 |
14 | 0.855 | 0.002 | 0.077 | 0.067 | 91.5 |
15 | 0.843 | 0.000 | 0.100 | 0.057 | 89.4 |
Average | 0.850 | 0.019 | 0.083 | 0.048 | 89.2 |
Standard deviation | 0.016 | 0.024 | 0.015 | 0.011 | 2.0 |
Group Average Effects of Distortion Corrections on Surface‐Based Analysis
Figure 6a shows the effects of uncorrected distortion on surface‐based analysis when the uncorrected SB GRE EPI images were registered to the structural images. Cortical fallout was obvious in the OFC, ITC, occipital cortex, and posterior bank of the central sulcus. In the former two cortices, cortical and Euclidean displacements and their ratio were large, while, in the latter two regions, the ratio was large in spots despite the fact that these displacements were small. The group‐averaged cortical fallout rate and displacement were 9.7 ± 2.5% and 0.96 ± 0.17 mm in the whole cortex, respectively (see also green bars in Fig. 7).
Figure 6b shows the effects of GDC. The cortical fallout was again obvious in the OFC, ITC, occipital cortex, and posterior bank of the central sulcus. The maps of Euclidean and cortical displacements and the displacement ratio were also similar to those in Fig. 6a. Cortical displacement was typically smaller than that in Fig. 6a, except in part of the OFC. Reflecting these results, the group‐averaged cortical fallout rate and displacement were significantly smaller (8.7 ± 2.4% and 0.86 ± 0.16 mm, respectively; see also blue bars in Fig. 7) than those in the uncorrected condition.
Finally, the effects of B0DC are shown in Fig. 6c. Fallout was seen in the anterior frontal cortex, ITC, and posterior occipital cortex (POC). Cortical and Euclidian displacements were relatively large in these regions, and the distance ratio was large in spots in the ITC and POC. The group‐averaged cortical fallout rate and displacement was the lowest in this case (2.1 ± 0.7% and 0.29 ± 0.07 mm, respectively; see also magenta bars in Fig. 7).
Figure 7a shows the comparisons of group‐averaged cortical fallout rates for the three distortion conditions. The repeated measures ANOVA for 15 subjects indicated statistically significant differences among the three conditions (F(2,28) = 133.95, P < 0.05). The post‐hoc paired t‐tests revealed significant differences among all pairs of conditions (P < 0.0167 with Bonferroni correction, t(14) = 5.45 between the uncorrected and GDC conditions, t(14) = 12.35 between the uncorrected and B0DC conditions, and t(14) = 11.13 between GDC and B0DC conditions). The comparisons of group‐averaged cortical displacements among the three distortion conditions are shown in Fig. 7b. Again, the repeated measures ANOVA indicated statistically significant differences among the conditions (F(2,28) = 241.71, P < 0.05). The post‐hoc paired t‐tests revealed significant differences among all pairs of conditions (P < 0.0167 with Bonferroni correction, t(14)= 9.13 between the uncorrected and GDC conditions, t(14) = 17.03 between the uncorrected and B0DC conditions, and t(14) = 14.36 between GDC and B0DC conditions).
Individual Effects of Distortion Correction on Surface‐Based Analysis and Volume Data
Figure 8a demonstrates maps of Euclidean and cortical displacements and their ratio in the right hemisphere of a typical subject when no distortion corrections were applied to the EPI image. Euclidean and cortical displacements were large in the medial frontal cortex (MFC), OFC, and ITC. Cortical fallouts were seen not only around these cortices but also around the central sulcus and part of the occipital cortex.
The cortical fallouts around the AFC, posterior bank of the central sulcus, and POC (left, middle, and right columns, respectively) are reflected in the volume data from the same typical subject in Fig. 8b. In the T1w images in the first row, gray matter voxels are located between two trace lines: pial and white surfaces. When these lines are overlaid on the corrected SB EPI images, we find gray matter voxels included between them (see images in the second row). In contrast, voxels from white matter or CSF intruded into the gray matter zones in the uncorrected SB EPI image registered to the structural image (see images in the third row). This can be more clearly seen from images in the fourth row, where distortion‐replicated cortical ribbon is overlaid on the distorted EPI images and the correct trace lines. The point F in the AFC includes signals outside the CSF (quite low voxel intensity) because of a distortion to the posterior (see the left column). We also found a cortical displacement over 5 mm in the medial area. Meanwhile, the points S and V in the somatosensory and visual cortices included signals from the CSF and white matter (higher and lower voxel intensity than gray matter), as shown in the middle and right columns, respectively.
Discussion
This study quantified the effects of geometric distortion corrections on surface‐based analysis in terms of cortical fallouts and displacements, instead of simple 3D Euclidean displacements. To this end, cortical surfaces replicating distorted EPI images were compared with those reconstructed from undistorted structural images, which were regarded as surfaces replicating fully corrected EPI images, that is, ideally corrected ones. However, since the accuracy of fully corrected EPI images was unknown, it was first evaluated based on intensity distributions of voxels within a cortical ribbon. The simulated spatial distribution map obtained by fitting the probability density function with four normal distribution functions for gray and white matter, CSF, and intensity dropout exhibited clear intensity dropout around the OFC and ITC, as expected. Some voxels labeled as white matter and CSF were randomly scattered in gray matter, while most were located on the edge. SB GRE EPI images, the least noisy EPI images, are therefore imperfect for voxel classification of brain tissues. Thus, the actual distortion correction accuracy for fully corrected EPI images may be greater than the 89.2% estimated.
Cortical fallout and displacement have allowed us to clearly delineate the impact of subtle distortions. In the comparison between undistorted and distortion‐replicated cortical surfaces, the most striking finding was that even moderate distortions can yield important cortical fallout and displacement. Hitherto, for distortion correction evaluations, distorted EPI images were simply compared to structural or corrected EPI images, or the amounts of distortions were shown or overlaid onto structural images. 13 , 20 , 27 , 28 , 29 As with the other approaches, B0 distortion was indirectly evaluated by presenting effects on functional connectivity, 30 and the effect of gradient distortion correction was assessed by evaluating voxel‐based image intensity reproducibility. 18 These studies remind us of the importance of distortion correction in brain regions where distortions are prominent, such as the OFC and ITC for B0 distortion. Our results supported these findings, showing a high cortical fallout rate and large cortical displacement around these regions. Additionally, prominent cortical fallouts were observed in the posterior bank of the central sulcus (the primary somatosensory cortex) and visual cortex, where distortions were supposed to be mild or moderate without B0DC, unlike areas near the nasal cavity and ear canals. 12 , 13 The primary somatosensory cortex is one of the thinnest cortices. 31 , 32 The visual cortex is also thin, 33 , 34 with complex gyri and sulci running throughout. 35 Due to these structural factors, small distortions in the AP direction resulting from PE can easily cause a fallout.
Across the whole brain, we observed far higher cortical fallout rate and larger cortical displacement with only GDC compared to only B0DC, indicating that B0 distortion was a major cause of distortions. The effect of GDC in our study was small. Glasser et al 20 have previously claimed that GDC is not required for standard scanners with gradients that are linear over a large FOV and where the head position is close to the isocenter. Most gradient coils have specifications of <1% linearity over a 40‐cm sphere. 27 Nevertheless, in our study gradient distortion resulted in measurable cortical fallouts and displacements in the AFC, ITC, and POC. Gradient distortion is mild but global, so that it changes the volume size of whole EPI images. Therefore, the burden of whole image distortion cumulatively arises as registration errors in the volume periphery, where the AFC, ITC, and POC are located. Further, these cortices are located in a zone where gradient distortion was high enough for measurable cortical fallouts and displacements. As for the POC, the above‐mentioned structural factors might have also affected these quantities similarly.
Our results examining cortical fallout and displacement indicate that GDC is necessary for precise surface‐based fMRI analysis. Offline GDC is especially important when we use MRI scanners where the availability of a vendor's GDC depends on the scanning sequences. In fact, the vendor's GDC was not available in the present fMRI sequence on our 7T scanner, but it was available for the structural image sequences acquired at 3T. Conversely, the importance of offline GDC applications has been less emphasized in scanners where a vendor's GDC is applicable in fMRI sequences. However, replacing the vendor's GDC with our own offline one may be important, as the vendor's GDC is applied only in the in‐plane direction in 2D EPI sequences. Although comparison of 2D and 3D GDCs is outside the scope of this study, several studies have reported or referred to the effectiveness of 3D GDC over 2D GDC in structural imaging sequences, 17 , 18 , 36 implying that offline GDC should not be confined to scanners that have a poor gradient nonlinearity due to a short or small bore and/or where the subject's head is not positioned at the isocenter.
The metrics used in our study (cortical fallout and displacement) were informative in demonstrating the impact of distortions on surface‐based analysis in functional neuroimaging. The introduction of cortical displacement highlighted a problem caused by voxels surviving on gray matter despite image distortion. The group‐averaged cortical displacement in the uncorrected condition was comparable to the length of a side of the voxel in this study (0.96 mm compared to 1.2 mm). The displacement ratio was also helpful to accentuate vertices where cortical displacement was observed. Meanwhile, because of the vertexwise metrics relying on only positional information of vertices, we did not consider partial cortical fallout resulting from voxel volume change and the partial volume effect. Moreover, a vertex was determined as a fallout when experiencing a distortion yielding displacement over half the cortical thickness in the direction normal to the surface. The present group‐averaged cortical fallout rates are, therefore, likely to be conservative. However, even if ~90% of the vertices could survive on gray matter without any distortion corrections in the conservative evaluation, fMRI data would be missampled on vertices with every surviving voxel displaced by nearly the length of the side of a voxel, on average.
Limitations
We evaluated displacements only on a cortical surface modeled as a sheet. However, as the cortex indeed has thickness, displacements caused by image distortions should also be observed in the depth direction. These displacements are much more critical for cortical layer‐based analysis, 1 , 8 , 9 , 10 which is more advanced than surface‐based analysis. 37 , 38 For future study, we need to consider the reduction of image distortions during fMRI data acquisition and methods for more precise distortion corrections. At the same time, evaluation for displacements in the depth direction will be more important.
Conclusion
This study quantified the effects of geometric distortion corrections in surface‐based analysis as cortical fallouts and displacements, rather than simple 3D Euclidean displacements, by creating cortical surfaces and ribbons replicating distortions of EPI images. Using these neurobiologically more meaningful quantities, our method of distortion correction evaluation revealed that even moderate distortion (B0 distortion in regions other than in the vicinity of the nasal cavity and ear canals and gradient distortion) can yield significant cortical fallouts and displacements in regions where the cortex is thin (eg, the primary somatosensory cortex), where the folding pattern of the cortex is complex (eg, the visual cortex), or where gradient linearity is imperfect (eg, the AFC, ITC, and POC, that is regions distant from the isocenter of an MRI scanner). This suggests that careful distortion corrections, as performed in the preprocessing pipelines, are essential for surface‐based analysis of high‐resolution fMRI data at 7T.
Acknowledgments
We thank Dr. Hans‐Peter Fautz and Dr. Tobias Kober (Siemens Healthineers) for providing the sequences for prescanning adjustment on 7T MRI.
Contract grant sponsor: Strategic Research Program for Brain Sciences from Japan Agency for Medical Research and Development (AMED); Contract grant numbers: JP18dm0107152 and JP18dm0307005Contract grant sponsor: Japan Society for the Promotion of Science (JSPS); Contract grant numbers: 16H03305 and 19 K22985.
References
- 1. De Martino F, Moerel M, Ugurbil K, Goebel R, Yacoub E, Formisano E. Frequency preference and attention effects across cortical depths in the human primary auditory cortex. Proc Natl Acad Sci U S A 2015;112:16036‐16041. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2. Ogawa S, Tank DW, Menon R, et al. Intrinsic signal changes accompanying sensory stimulation: Functional brain mapping with magnetic resonance imaging. Proc Natl Acad Sci U S A 1992;89:5951‐5955. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3. Yacoub E, Shmuel A, Pfeuffer J, et al. Imaging brain function in humans at 7 Tesla. Magn Reson Med 2001;45:588‐594. [DOI] [PubMed] [Google Scholar]
- 4. Vaughan JT, Garwood M, Collins CM, et al. 7T vs. 4T: RF power, homogeneity, and signal‐to‐noise comparison in head images. Magn Reson Med 2001;46:24‐30. [DOI] [PubMed] [Google Scholar]
- 5. Uğurbil K, Adriany G, Anderson P, et al. Ultrahigh field magnetic resonance imaging and spectroscopy. Magn Reson Imaging 2003;21:1263‐1281. [DOI] [PubMed] [Google Scholar]
- 6. Chaimow D, Yacoub E, Uğurbil K, Shmuel A. Spatial specificity of the functional MRI blood oxygenation response relative to neuronal activity. NeuroImage 2018;164:32‐47. [DOI] [PubMed] [Google Scholar]
- 7. Yacoub E, Harel N, Uğurbil K. High‐field fMRI unveils orientation columns in humans. Proc Natl Acad Sci U S A 2008;105:10607‐10612. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8. Polimeni JR, Fischl B, Greve DN, Wald LL. Laminar analysis of 7T BOLD using an imposed spatial activation pattern in human V1. NeuroImage 2010;52:1334‐1346. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9. Muckli L, De Martino F, Vizioli L, et al. Contextual feedback to superficial layers of V1. Curr Biol 2015;25:2690‐2695. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10. Kok P, Bains LJ, van Mourik T, Norris DG, de Lange FP. Selective activation of the deep layers of the human primary visual cortex by top‐down feedback. Curr Biol 2016;26:371‐376. [DOI] [PubMed] [Google Scholar]
- 11. Polimeni JR, Renvall V, Zaretskaya N, Ficshl B. Analysis strategies for high‐resolution UHF‐fMRI data. NeuroImage 2018;168:296‐320. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12. Jezzard P, Balaban RS. Correction for geometric distortion in echo planar images from B0 field variations. Magn Reson Med 1995;34:65‐73. [DOI] [PubMed] [Google Scholar]
- 13. Hutton C, Bork A, Josephs O, Deichmann R, Ashburner J, Turner R. Image distortion correction in fMRI: A quantitative evaluation. NeuroImage 2002;16:217‐240. [DOI] [PubMed] [Google Scholar]
- 14. Andersson JLR, Skare S, Ashburner J. How to correct susceptibility distortions in spin‐echo echo‐planar images: Application to diffusion tensor imaging. NeuroImage 2003;20:870‐888. [DOI] [PubMed] [Google Scholar]
- 15. Robinson EC, Jbabdi S, Glasser MF, et al. MSM: A new flexible framework for multimodal surface matching. NeuroImage 2014;100:414‐426. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16. Smith SM, Jenkinson M, Woolrich MW, et al. Advances in functional and structural MR image analysis and implementation as FSL. NeuroImage 2004;23:S208‐S219. [DOI] [PubMed] [Google Scholar]
- 17. Janke A, Zhao H, Cowin GJ, Galloway GJ, Doddrell DM. Use of spherical harmonic deconvolution methods to compensate for nonlinear gradient effects on MRI images. Magn Reson Med 2004;52:115‐122. [DOI] [PubMed] [Google Scholar]
- 18. Jovicich J, Czanner S, Greve D, et al. Reliability in multi‐site structural MRI studies: Effects of gradient non‐linearity correction on phantom and human data. NeuroImage 2006;30:436‐443. [DOI] [PubMed] [Google Scholar]
- 19. Van Essen DC, Ugurbil K, Auerbach E, et al. The human connectome project: A data acquisition perspective. NeuroImage 2012;62:2222‐2231. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 20. Glasser MF, Sotiropoulos SN, Wilson JA, et al. The minimal preprocessing pipelines for the human connectome project. NeuroImage 2013;80:105‐124. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 21. Moeller S, Yacoub E, Olman CA, et al. Multiband multislice GE‐EPI at 7 Tesla, with 16‐fold acceleration using partial parallel imaging with application to high spatial and temporal whole‐brain fMRI. Magn Reson Med 2010;63:1144‐1153. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 22. Griswold MA, Jakob PM, Heidemann RM, et al. Generalized autocalibrating partially parallel acquisitions (GRAPPA). Magn Reson Med 2002;47:1202‐1210. [DOI] [PubMed] [Google Scholar]
- 23. Mugler JP 3rd, Brookeman JR. Three‐dimensional magnetization‐prepared rapid gradient‐echo imaging (3D MP RAGE). Magn Reson Med 1990;15:152‐157. [DOI] [PubMed] [Google Scholar]
- 24. Mugler JP 3rd, Bao S, Mulkern RV, et al. Optimized single‐slab three‐dimensional spin‐echo MR imaging of the brain. Radiology 2000;216:891‐899. [DOI] [PubMed] [Google Scholar]
- 25. Greve DN, Fischl B. Accurate and robust brain image alignment using boundary‐based registration. NeuroImage 2009;48:63‐72. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 26. Glasser MF, Smith SM, Marcus DS, et al. The Human Connectome Project's neuroimaging approach. Nat Neurosci 2016;19:1175‐1187. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 27. Jezzard P, Clare S. Sources of distortion in functional MRI data. Hum Brain Mapp 1999;8:80‐85. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 28. Holland D, Kuperman JM, Dale AM. Efficient correction of inhomogeneous static magnetic field‐induced distortion in echo planar imaging. NeuroImage 2010;50:175‐183. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 29. Smith SM, Beckmann CF, Andersson J, et al. Resting‐state fMRI in the Human Connectome Project. NeuroImage 2013;80:144‐168. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 30. Togo H, Rokicki J, Yoshinaga K, et al. Effects of field‐map distortion correction on resting state functional connectivity MRI. Front Neurosci 2017;11:656. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 31. Fischl B, Dale AM. Measuring the thickness of the human cerebral cortex from magnetic resonance images. Proc Natl Acad Sci U S A 2000;97:11050‐11055. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 32. MacDonald D, Kabani N, Avis D, Evans AC. Automated 3‐D extraction of inner and outer surfaces of cerebral cortex from MRI. NeuroImage 2000;12:340‐356. [DOI] [PubMed] [Google Scholar]
- 33. Hutton C, De Vita E, Ashburner J, Deichmann R, Turner R. Voxel‐based cortical thickness measurements in MRI. NeuroImage 2008;40:1701‐1710. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 34. Glasser MF, Van Essen DC. Mapping human cortical areas in vivo based on myelin content as revealed by T1‐ and T2‐weighted MRI. J Neurosci 2011;31:11597‐11616. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 35. Rabiei H, Richard F, Coulon O, Lefèvre J. Local spectral analysis of the cerebral cortex: New gyrification indices. IEEE Trans Med Imaging 2017;36:838‐848. [DOI] [PubMed] [Google Scholar]
- 36. Wang D, Strugnell W, Cowin G, Doddrell DM, Slaughter R. Geometric distortion in clinical MRI systems part I: Evaluation using a 3D phantom. Magn Reson Imaging 2004;22:1211‐1221. [DOI] [PubMed] [Google Scholar]
- 37. Dale AM, Fischl B, Sereno MI. Cortical surface‐based analysis. I. Segmentation and surface reconstruction. NeuroImage 1999;9:179‐194. [DOI] [PubMed] [Google Scholar]
- 38. Fischl B, Sereno MI, Dale AM. Cortical surface‐based analysis. II: Inflation, flattening, and a surface‐based coordinate system. NeuroImage 1999;9:195‐207. [DOI] [PubMed] [Google Scholar]