. 2016 Sep;37(9):1696–1699. doi: 10.3174/ajnr.A4804

Suspected Metallic Embolism following Endovascular Treatment of Intracranial Aneurysms

R Yasuda a,, M Maeda b, M Umino b, Y Nakatsuka a, Y Umeda a, N Toma a, H Sakaida a, H Suzuki a
PMCID: PMC7984696  PMID: 27102315

SUMMARY:

We describe a case series of suspected metallic embolism after coil embolization for intracranial aneurysms. Between January 2012 and December 2014, 110 intracranial aneurysms had been treated by coil embolization in our institution. In 6 cases, the postprocedural MR imaging revealed abnormal spotty lesions not detected on the preprocedural MR imaging. The lesions were also undetectable on the postprocedural CT scan. They were demonstrated as low-intensity spots on T1WI, T2WI, DWI, and T2*-weighted imaging. On DWI, they were accompanied by bright “halo,” and on T2*-weighted imaging, they showed a “blooming” effect. In 3 of the 6 cases, follow-up MR imaging was available and all the lesions remained and demonstrated no signal changes. Although histologic examination had not been performed, these neuroradiologic findings strongly supported the lesions being from metallic fragments. No specific responsible device was detected after reviewing all the devices used for the neuroendovascular treatment in the 6 cases.


Clinically asymptomatic emboli are a common complication of cerebral angiography and neuroendovascular procedures, occurring in approximately 20% of cases.1 Foreign body emboli after cerebral angiography are a much less known complication, and only a few isolated cases have been reported.24 Recently, we have noticed abnormal signals suggesting metallic emboli on the postprocedural MR imaging in several cases of coil embolization for cerebral aneurysms. In this report, we describe the particular MR imaging findings of suspected metallic emboli following neuroendovascular procedures and discuss possible sources for these findings.

Case Series

After the retrospective review of our medical data base between January 2012 and December 2014 that contains 110 intracranial aneurysms treated by coil embolization, we encountered 6 cases that showed abnormal spotty lesions on the postprocedural MR imaging. The details of the clinical features of the 6 cases are summarized in Table 1. The devices used for the treatment in the 6 cases are summarized in Table 2. In all 6 cases, the lesions were low intensity on T1WI, T2WI, DWI, and T2*-weighted images. Additionally, they were accompanied by a hyperintense “halo” on DWI and showed a “blooming” effect on T2*-weighted images. In 5 of the 6 cases (cases 1–5), a postprocedural CT scan was obtained but none of the lesions were detected. Follow-up MR images were available in 3 of the 6 cases (cases 1, 5, and 6), and in all 3 cases, the lesions remained without any signal changes during the follow-up periods (range, 6–34 months). No patients showed any symptoms associated with the lesions.

Table 1:

Patient characteristics

Case No. Patient Age (yr)/Sex Location of Aneurysm Aneurysm Size (mm) Procedure Procedural Time (min) Postprodedural Symptoms
1 68/F Left ICA 10.2 × 8.2 × 7.3 Coil embolization 170 None
2 56/M Right vertebral artery 15.6 × 7.1 × 6.5 Internal trapping 135 None
3 80/F Right ICA 12.6 × 10.3 × 11.0 Coil embolization 150 None
4 45/F Right ICA 4.9 × 4.5 × 4.3 Coil embolization 158 None
5 70/M Anterior communicating artery 9.6 × 7.1 × 9.5 Coil embolization 145 None
6 41/F Left ICA 5.6 × 4.9 × 3.8 Coil embolization 150 None

Table 2:

Devices used for treatment in each case

Devices 1 2 3 4 5 6
0.035-inch Radifocus Standarda
0.035-inch Radifocus Stiffa
0.035-inch Quick Flex Standardb
0.010–0.014-inch Tenroub
0.014-inch CHIKAIc
0.010-inch X-Pediond
0.010-inch SilverSpeedd
0.014-inch Traxcesse
Excelsior SL-10f
Headway17e
Prowler Select Plusg
Hydrogel coilse
ED coilsb
Target coilsf
Cashmereg
VFCe
Orbit Galaxyg
Deltapaqg
Deltaplushg
Matrix coilsf
Scepter XCe
HyperFormd
Enterprise VRDg
Amplatz GooseNeck Snared

Note:—○ indicates that the product was used.

a

Terumo, Tokyo, Japan.

b

Kaneka Medix, Osaka, Japan.

c

Asahi Intecc, Aichi, Japan.

d

Covidien, Irvine, California.

e

MicroVention, Tustin, California.

f

Stryker, Kalamazoo, Michigan.

g

Codman & Shurtleff, Raynham, Massachusetts.

Illustrative Cases

Case 1.

A 68-year-old woman was treated for a left ICA aneurysm (Fig 1A) with coil embolization. On the preprocedural MR imaging, no abnormal findings apart the aneurysm were identified. For the coil embolization procedure, a 5F guiding sheath was placed in the left ICA and a 1.7F microcatheter was inserted into the aneurysm by using a 0.014-inch microguidewire. The aneurysm was embolized with multiple platinum coils (Fig 1B). The MR images were obtained the next day by using a 1.5T MR imaging unit (Achieva Nova Dual; Philips Healthcare, Best, the Netherlands) with a 5-mm section thickness, and DWI revealed low-intensity spots accompanied by a hyperintense halo at the left medial frontal and left parietal regions (Fig 1C, -D). These signals were obviously different from the bright lesion in the left parietal lobe indicating thromboembolism (Fig 1E) that showed no signal changes on T2*-weighted images.

Fig 1.

Fig 1.

A left ICA aneurysm in case 1 (A) is embolized by using multiple coils (B). Postprocedural DWI shows abnormal low-intensity spots with a hyperintense halo (white arrow) in the left parietal (C) and frontal (D) lobes. These are different from a bright spot in the left parietal lobe (E), indicating thromboembolism. These abnormal spots are low intensity on T1WI (F) and T2WI (G) and show a blooming effect on T2*-weighted images (H). Follow-up MR imaging obtained 13 months after the procedure shows that all the abnormal signal spots remain without any signal changes (I–L).

While these lesions were shown as low-intensity spots on T1WI (Fig 1F) and T2WI (Fig 1G), they had a blooming effect on T2*-weighted images (Fig 1H). Head CT (Aquilion ONE CT scanner; Toshiba Medical Systems, Tokyo, Japan) obtained on the following day demonstrated no abnormal findings except a coil mass. The patient was discharged from our hospital 1 week after the coil embolization without any symptoms. Follow-up MR imaging was performed at 13 months after the procedure, and both abnormal spots remained without any signal changes (Fig 1I, -L). The patient remained asymptomatic at the last follow-up.

Case 5.

A 70-year-old man was referred to our hospital for treatment of his unruptured anterior communicating artery aneurysm (Fig 2A). Preprocedural MR imaging showed no abnormal findings apart from the aneurysm. The coil embolization was performed; a 6F guiding catheter was placed in his left ICA. After a balloon catheter was placed underneath the aneurysmal neck by using a 0.01-inch microguidewire, a 1.7F microcatheter was placed in the aneurysm by using another 0.014-inch microguidewire. During the embolization, the first coil uncoiled, and it was withdrawn by using a retrieval device. Because the hub of the microcatheter had to be cut to advance the retrieval device, another 1.7F microcatheter was used for the embolization. The aneurysm was finally embolized by using multiple platinum coils (Fig 2B). The MR imaging performed on postoperative day 1 by using a 1.5T machine (Achieva Nova Dual; Philips) with 5-mm section thickness demonstrated a low-intensity spot with a hyperintense halo on DWI in the right frontal lobe (Fig 2C). A head CT scan (Aquilion ONE; Toshiba) with 5-mm section thickness obtained on postoperative day 3 showed no abnormal findings except a coil mass (Fig 2D). The patient's postoperative course was good, and he was discharged from our hospital 1 week after the procedure without any symptoms. Follow-up MR imaging obtained 3 years after the procedure demonstrated that the lesion remained at the right frontal lobe without any signal changes (Fig 2E, -F). The patient remained asymptomatic at the last follow-up.

Fig 2.

Fig 2.

An anterior communicating artery aneurysm in case 5 (A) is embolized by using multiple coils (B). Postprocedural DWI shows an abnormal low-intensity spot with a hyperintense halo (white arrow) in the right frontal lobe (C), which is undetectable on the postprocedural CT (D). Follow-up MR imaging obtained 3 years after the procedure shows that the lesion remains without any signal changes (E and F).

Discussion

Because the new low-signal spotty lesions appeared just after the endovascular procedure and they were restricted to the vascular territory of the catheterized arteries, it is reasonable to believe they represented embolic phenomena.

Potential etiologies of the low-signal spotty lesions include air, thrombus, hemosiderin, calcium deposits, and metal. If they were air or thrombus, resolution of signal changes would be expected with time. Moreover, parenchymal ischemic changes due to an embolism could occur. Hemosiderin is also unlikely because it does not appear at this early stage just after coil embolization. Furthermore, if the lesions were calcium, they might be detected by CT. Metal results in a dark spot on MR imaging because it has no MR imaging signal. Additionally, it can cause substantial MR imaging artifacts,5 which could exhibit the hyperintense halo. Because only microscopic metal fragments are required to produce MR imaging artifacts,6 it is not surprising that the metal fragments were invisible on CT. Moreover, if microscopic metal fragments were present, it would be reasonable that patients had no embolic complications. Taking these observations into consideration, we believe that the spotty lesions found on the postoperative MR imaging in our cases were from metallic fragments.

Cerebral MR imaging findings suggesting embolic metal fragments are rare, and only 4 cases have been reported (Table 3).710 In addition to a mechanical prosthetic cardiac valve and robotic surgery, angiographic guidewires used during cardiac catheterization were suspected as an embolic source in 2 cases.9,10 To the best of our knowledge, this cases series is the first in reporting MR imaging metallic signal following neuroendovascular treatment of intracranial aneurysms. Besides the guidewires, various devices (ie, coils, balloons, and stents) composed of metal are used in neuroendovascular procedures, which can provide more frequent opportunities for metallic embolism compared with cardiac catheterization.

Table 3:

Past cases of suspected metallic embolism

Author Patient/Sex T2WI/Proton T2*-Weighted Image CT Suspected Etiology Symptom
Wingerchuk et al7 35/F Low intensity halo Low intensity blooming Normal Prosthetic cardiac valve None
Naumann et al8 66/M Low intensity halo Not available Not available Prosthetic cardiac valve Seizure
Jassal et al9 55/M Low intensity halo Not available Normal Coronary guidewire Headache
Roshal et al10 55/F Low intensity Low intensity blooming, halo Normal Coronary guidewire, robotic surgery Headache

Our study has several limitations. First and most important, it was not pathologically confirmed that the lesion represented metallic emboli. Because our patients were completely free from symptoms, it was not reasonable to perform a brain biopsy. Second, we cannot determine which devices and which parts could be a source of metallic emboli in our cases. In neuroendovascular treatment, metallic emboli can be released at any step, including delivery; deployment or withdrawal of the devices; friction between the devices; and detachment of the coils. Coil uncoiling and withdrawal of the uncoiled coil were likely responsible for metallic emboli in case 5. However, in the remainder of cases, coil embolization was completed uneventfully. Although MR imaging susceptibility artifacts vary depending on metals,11 these were not metal-specific and there is no way to differentiate various metals on MR imaging. We have reviewed all the devices used during endovascular treatment in these cases (Table 3). No specific devices except one 0.035-inch guidewire (0.035-inch Radifocus Standard; Terumo, Tokyo, Japan) were used in all cases. It is unlikely, however, that the 0.035-inch guidewire is responsible because we have routinely used the guidewire in all endovascular procedures, including diagnostic angiography. Devices for intracranial use seem more likely than the 0.035-inch guidewire. Further investigations should be done in the near future.

Finally, the exact incidence of the abnormal MR imaging signals was unclear. Our retrospective review included 110 coiling procedures for aneurysm treatment. Although periprocedural MR imaging was available in almost all cases, many cases lacked pre- and/or postoperative DWI, T2*-weighted images, or SWI, and only 29 cases had adequate imaging. Prospective registration with a specific perioperative MR imaging protocol is ongoing.

Conclusions

This is the first report on the brain MR imaging findings suggesting that metallic emboli were found after coil embolization of cerebral aneurysms, to our knowledge. Particular DWI and T2*-weighted imaging findings of low intensity accompanied by a hyperintense halo should be kept in mind for patients undergoing coil embolization.

Footnotes

Paper previously presented, in part, at: 44th Annual Meeting of the Chubu Society of Neuroendovascular Therapy, August 8, 2015; Fukui, Japan.

References

  • 1. Bendszus M, Koltzenburg M, Burger R, et al. Silent embolism in diagnostic cerebral angiography and neurointerventional procedures: a prospective study. Lancet 1999;354:1594–97 10.1016/S0140-6736(99)07083-X [DOI] [PubMed] [Google Scholar]
  • 2. Mehta RI, Mehta RI, Solis OE, et al. Hydrophilic polymer emboli: an under-recognized iatrogenic cause of ischemia and infarct. Mod Pathol 2010;23:921–30 10.1038/modpathol.2010.74 [DOI] [PubMed] [Google Scholar]
  • 3. Cruz JP, Marotta T, O'Kelly C, et al. Enhancing brain lesions after endovascular treatment of aneurysms. AJNR Am J Neuroradiol 2014;35:1954–58 10.3174/ajnr.A3976 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4. Minks D, Briley D, Schulz U, et al. Suspected cerebral foreign body granuloma following endovascular treatment of intracranial aneurysm: imaging features. Neuroradiology 2015;57:71–73 10.1007/s00234-014-1440-1 [DOI] [PubMed] [Google Scholar]
  • 5. Hargreaves BA, Worters PW, Pauly KB, et al. Metal-induced artifacts in MRI. AJR Am J Roentgenol 2011;197:547–55 10.2214/AJR.11.7364 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6. Alanen A, Bondenstam S, Komu M. Artifacts in MR imaging caused by small quantities of powdered iron. Acta Radiol 1995;36:92–95 [PubMed] [Google Scholar]
  • 7. Wingerchuk DM, Krecke KN, Fulgham JR. Multifocal brain MRI artifacts secondary to embolic metal fragments. Neurology 1997;49:1451–53 10.1212/WNL.49.5.1451 [DOI] [PubMed] [Google Scholar]
  • 8. Naumann M, Hofmann E, Toyka KV. Multifocal brain MRI hypointensities secondary to embolic metal fragments from a mechanical heart valve prosthesis: a possible source of epileptic seizures. Neurology 1998;51:1766–67 10.1212/WNL.51.6.1766 [DOI] [PubMed] [Google Scholar]
  • 9. Jassal DS, Fast MD, McGinn G. Multifocal brain MRI hypointensities secondary to cardiac catheterization. Neurology 2000;54:2023–24 10.1212/WNL.54.10.2023 [DOI] [PubMed] [Google Scholar]
  • 10. Roshal D, Snapp M, Friedman DP, et al. Iatrogenic brain and cervical cord magnetic resonance imaging susceptibility artifacts from metallic microemboli. Arch Neurol 2011;68:132–33 10.1001/archneurol.2010.336 [DOI] [PubMed] [Google Scholar]
  • 11. Matsuura H, Inoue T, Konno H, et al. Quantification of susceptibility artifacts produced on high-field magnetic resonance images by various biomaterials used for neurosurgical implants: technical note. J Neurosurg 2002;97:1472–75 10.3171/jns.2002.97.6.1472 [DOI] [PubMed] [Google Scholar]

Articles from AJNR: American Journal of Neuroradiology are provided here courtesy of American Society of Neuroradiology