Abstract
The tubular structures and dendritic distributions of blood vessels emit anisotropic photoacoustic (PA) signals with different intensities and frequency components at different angles. Therefore, spectral analysis of PA signals from a single angle cannot accurately determine the physical characteristics of microvessels. This study investigated the feasibility of using the PA power azimuth spectrum (PA-PAS) method to evaluate microvessel structures. We mapped the acoustic power spectrum of the PA signals along the azimuth direction. Based on a frequency-domain analysis of the broadband PA signal, we calculated the spectral parameter power-weighted mean frequency (PWMF). The results demonstrate that the PA signal information of the microvessel is mainly concentrated in the direction of its width. In addition, the PWMF decreases linearly with the microvascular size. The experimental findings exhibit good agreement with the simulation results, thus demonstrating that this approach can effectively differentiate the sizes of microvessels.
Keywords: Microvascular evaluation, PA power azimuth spectrum, Angle-dependent frequency analysis, Structural direction, Size
1. Introduction
The occurrence and development of numerous diseases are closely associated with changes in the morphology and function of microvasculature, including size and density [[1], [2], [3], [4], [5], [6], [7]]. Quantitatively, these changes can provide insights into the dynamics of tumor angiogenesis, which is important for clinicians diagnosing and treating cancer.
Many well-established clinical detection modalities, such as optical coherence tomography (OCT), ultrasonography, and magnetic resonance imaging (MRI), have been adopted for vascular imaging. Owing to the strong scattering of light in biological tissues, OCT can provide high-resolution images; however, its imaging depths are limited [8,9]. Although ultrasonography can measure blood flow velocity at a satisfactory imaging depth, it is not suitable for microvessels, the diameter of which is less than 150 μm [10,11]. MRI can provide blood volume, perfusion, permeability, and other vascular information; however, its resolution is insufficient for analyzing microvessels [12,13]. At present, immunohistochemical staining of the vascular endothelium is still the conventional method for identifying microvessels in clinical practice [12]; however, it is invasive and severely painful for patients. Therefore, noninvasive diagnosis of microvascular diseases has proved extremely challenging.
Photoacoustic tomography (PAT) is a new type of noninvasive biomedical imaging modality that combines the advantages of high-contrast optical imaging with the advantages of high-resolution ultrasound imaging in deep biological tissues [9,[14], [15], [16]]. Because of the high optical contrast between the hemoglobin in blood and the surrounding tissue [17], PAT has been widely used in vascular detection [18,19]. Photoacoustic computed tomography (PACT) is based on wide-field light illumination and unfocused ultrasonic detection at multiple locations; it relies on an inverse algorithm to reconstruct the image showing light absorption distribution and can achieve penetration depths up to 70 mm in living tissue [20]. The delay-and-sum reconstruction algorithm [21,22] is a basic and widely used method in PACT. Most of the previous PACT studies on microvessels were focused on either imaging the distribution of microvessels [[23], [24], [25]] or quantifying hemoglobin oxygenation [[26], [27], [28]]. They were mainly interested in the amplitude of the time-domain photoacoustic (PA) signals. However, the imaging resolution in PACT is limited by the center frequency and bandwidth of the transducer [29]. Research has revealed that the frequency components of PA signals are related to the sizes and contents of the absorbers [30,31]. PA frequency analysis has been widely applied in examining fatty livers [32], bones [33], and blood cells [34]. Moreover, frequency analysis has proved feasible in detecting absorbers with sizes smaller than the system resolution [31,35], which is extremely advantageous for analyzing microvascular structures. To extract the main characteristics of PA signals in the frequency domain, the power spectrum of PA signals is calculated. Then the spectral parameter, namely the slope, can be obtained via linear regression fitting of the normalized power spectrum, representing the microstructure of biological tissue [33,35,36]. The slope reflects the energy proportions of the high- and low-frequency components of biological tissue, but a new quantization parameter is needed to better reflect the main frequency component of biological tissue, which represents the main size of the biological tissue. Additionally, in contrast to the point-source situation in the conventional PA theory, the tubular structures and dendritic distributions of blood vessels emit anisotropic PA signals with different intensities and frequency components at different angles. Few studies have focused on the frequency anisotropy caused by the vascular structure.
Therefore, in this study, we developed a novel method called the PA power azimuth spectrum (PA-PAS) method that combines the power spectra with the angles to evaluate microvessel structures more accurately. The PA-PAS method maps the acoustic power spectrum of the PA signals along the azimuthal direction; this reveals the physical characteristics of microvessels, including their structural direction and size. The PA spectral parameter, namely the power-weighted mean frequency (PWMF), which reflects the main frequency component of biological tissue, was calculated to quantify the sizes of the microvessels. Simulations involving k-waves and experiments using a home-made PA detection system were conducted.
2. Materials and methods
2.1. 2D numerical simulation model
To evaluate the feasibility of the PA-PAS method, we first implemented numerical simulations on models with a single microvessel in biological tissue. The PA signal from each model was simulated using the MATLAB k-wave toolbox (R2019b, MathWorks, Natick, MA) [37]. We extracted the typical vascular structural characteristics that affect the PA power spectrum for the simulation and simplified the model to a two-dimensional entity without loss of physical meaning. To simulate different sizes of blood vessel in biological tissue, the simulated microvessel were scaled to the actual microvascular size. The widths of the simulated microvessels were scaled to 100, 200, and 300 μm, respectively. The length of the simulated microvessel was scaled to 4 mm. According to the simulation setting in k-wave, the white pixels in Fig. 2(a) represent the microvessel and were designated as PA sources with the initial acoustic pressure of 1, while the black ones were designated as the other biological tissues without initial acoustic pressure. The speed of sound was set at 1500 m/s for the simulated area, and a 72-point annular acoustic sensor array around the model with an angle interval of 5° was used for receiving the PA signals, as shown in Fig. 2(a). The distance of the acoustic sensors to the center of the simulated model was set to 1 cm.
Fig. 2.
(a) Simulation model. A 72-point annular acoustic sensor array was distributed around the model with an angle interval of 5°. The distance of the acoustic sensors to the center of the simulated model was set as 1 cm. (b) Experimental setup for PA measurements. (c) Detection and spectral mapping processes used in the PA-PAS method. 1 st, 2nd, 3rd, and 19th represent the numbers indicating the acoustic sensor locations; 0, 5, 10 and 90 represent the angles; 1, 2, and 3 represent the PA spectral frequencies.
2.2. Gelatin phantom materials
The phantoms were prepared from porcine gelatin (Gelatin, Sangon Biotech, Shanghai, China) using a mixture of 8 % gelatin and 92 % water; the acoustic impedance of the phantoms is identical to that of biological tissue. The phantoms were disk-shaped with radii and thicknesses of approximately 1 and 0.8 cm, respectively. Fishing lines were used to simulate blood vessels in the experiment. The fishing lines were black and exhibited strong optical absorption in the visible spectrum. Three fishing lines with different diameters, 100, 200, and 300 μm, were cut into 5-mm segments and embedded into each phantom to simulate the single microvessel in biological tissue.
2.3. Chick embryo preparation
The feasibility of the proposed PA-PAS method for evaluating microvessels in vivo was verified using a chick embryo model, which is considered an important model for studying tumor angiogenesis [38,39]. The fertilized chicken eggs were incubated for 3–5 d and the blood vessel diameters were approximately 100–300 μm. Then, the embryos were removed from the eggshells and placed on a petri dish with plastic membrane. In order to simulate the in vivo detection of animal models, petri dishes were kept as the acoustic scattering interface to obtain PA signals with low signal-to-noise ratio (SNR). There were two groups of samples, Group-1(n = 3) and Group-2 (n = 3). The mean sizes of the blood vessels in the two groups measured microscopically (KH-7700, Hirox, Shanghai, China) were 200 and 250 μm, respectively. Fig. 1(b) presents the microscopic images and measured sizes of the blood vessels.
Fig. 1.
(a) Representative chick embryo model. Circled regions are the detection areas. (b) Microscopically measured size of detection area in chick embryo. For each detection area of the egg embryo, the vascular size was microscopically measured at eight locations at the minimum and averaged to obtain the mean size of the detection area.
2.4. System setup and PA measurements
Fig. 2(b) presents the schematic of the experimental setup. An optical parametric oscillator system pumped by an Nd:YAG laser (Phocus Mobile, OPOTEK, Carlsbad, CA) was used to provide laser pulses with a repetition rate of 10 Hz and a pulse width of 5.5 ns. At a 750-nm wavelength, the samples were illuminated by a laser beam of 1-cm diameter, which generated PA signals. These signals were received by a needle hydrophone (HNC1500, ONDA Corp., Sunnyvale, CA) with a bandwidth of 1–10 MHz, but it could cover frequencies up to 20 MHz. The energy incident on the sample was 10 mJ, which was below the safety limit specified by the American National Standards Institute. Driven by a computer-controlled stepper motor (TBR 100, Zolix, Beijing, China), the hydrophone was used to scan the samples in a circular motion with a step size of 5°; the total number of steps was I = 72. An ultrasonic coupling agent was used as the acoustic transmission medium between the samples and hydrophone. After 25-dB amplification by an amplifier (5072PR, Olympus Corp., Tokyo, Japan), the PA signals were recorded using a digital oscilloscope (HDO6000, oscilloscope, Teledyne Lecroy, USA) at a sampling rate of 2500 MHz.
2.5. PA-PAS analysis
The simulation and experimental data were processed using MATLAB. To extract the main characteristics of the PA signals in the frequency domain, the power spectra of the PA signals were computed using the Welch’s approach with a 5-μs moving Hamming window and 60 % overlapping. Then the power spectra of the PA signals were unfolded along the radius direction and the spectra acquired at all the angles were combined along the angle axis to form a PA-PAS map, as shown in Fig. 2(c). The radius axis gives the ultrasonic frequency distribution and represents the structural size of the microvessel; the angle axis represents the detection angles, showing the structural growth direction of the microvessel. Owing to the differences in the structural growth direction and size of the microvessels, each microvessel should possess a unique PA-PAS map.
In order to further quantitatively analyze the structural size of microvessel, the spectral parameter PWMF was calculated through the power spectrum of the PA signal as follows:
where is the power spectral density at each frequency. PWMF represents the main frequency component of the microvessel, reflecting its main size in the detection area. For determining the acoustic power spectra of the PA signals, the lowest spectral range of ≤1 MHz was discarded because of the poor frequency response of the hydrophone below 1 MHz. Therefore, the power spectra were analyzed in the range of 1–10 MHz, covering a sound level of approximately 25 dB.
3. Results
3.1. Simulation results of a single microvessel
Fig. 3(a) shows the waveforms of the PA signals recorded at i = 0° and 90°. Then, the power spectrum of the PA signal was analyzed, as shown in Fig. 3(b). Fig. 3(a) and (b) show that the tubular structure of the blood vessel emits anisotropic PA signals with different intensities and frequency components at different angles. Therefore, spectral analysis of PA signals from a single angle cannot accurately analyze the physical characteristics of microvessels. This frequency anisotropy caused by the vascular structure should be considered in PA spectral analysis.
Fig. 3.
(a) PA signals recorded by the 1 st (i = 0°, in the length direction of microvessel) and 19th detectors (i = 90°, in the width direction of microvessel). (b) Power spectra of PA signals in (a). diameter is the diameter of the microvessel; length is the length of the microvessel.
The PA-PAS method mapped the acoustic power spectra of the PA signals along the azimuthal direction. Fig. 4(a)-(c) show the PA-PAS maps of single microvessels of three different sizes, whose diameters are 100, 200, and 300 μm. The ultrasonic frequencies are shown along the radius axis, the detection angles are shown along the angle axis, and the amplitudes of the power spectra shown in pseudo-color. As observed in the PA-PAS maps, there are obvious PA signals along the width direction of microvessels, and the spectral amplitude along the width direction is considerably greater than that in the length direction of the microvessels. In addition, the 100-μm-diameter microvessel comprises more high-frequency components than the 200- and 300-μm-diameter microvessels. With more high-frequency components, the power spectrum of the microvessel appears more extended along the radius axis in the width direction of the microvessel. Further, to differentiate the microvascular sizes, the PWMF was calculated. As shown in Fig. 4(d), the PWMF increases from the length direction (0°) to the width direction (90°) of the microvessel. Compared to the microvessels with diameters of 200 and 300 μm, the one with a diameter of 100 μm shows higher PWMF values in the width direction; however, no difference is observed in the length direction. The simulation results show that the PA power spectra of the microvessels exhibit significant anisotropy at different angles. The PA signal in the width direction of the microvessel has a higher amplitude and more high-frequency components than that in the length direction. The PWMF values of the microvessels with different sizes differ significantly in the width direction of the microvessels but not in the length direction. Therefore, the power spectrum along the width direction of the microvessel can better characterize the microvascular size. To quantify the size of the microvessel more accurately, the detection direction of the transducer should be along the width direction of the microvessel. Then, the PWMF values in the width direction of the microvessels were extracted to characterize the microvascular sizes, as shown in Fig. 4(e). The PWMF value is linearly related to the vascular diameter, as indicated by the green dashed line. The larger the vessel diameter, the smaller the PWMF value.
Fig. 4.
The PA-PAS maps of different microvessels, whose diameters are (a) 100, (b) 200, and (c) 300 μm. (d) Quantified PWMF curve versus angle of each microvessel (the solid yellow boxes correspond to the width directions of microvessels). (e) PWMF values for different diameters in the width direction of microvessels. The PWMF values decrease linearly with the increase of the vascular diameter, as indicated by the green dashed line. (For interpretation of the references to colour in this figure legend, the reader is referred to the web version of this article).
3.2. Microvessel phantom experimental results
Fig. 5(a)-(c) are the plots of the PA signals generated by the microvessels with diameters of 100, 200, and 300 μm, respectively. There is no difference between the time-domain signals from the three differently sized microvessels. However, Fig. 5(d) shows that there are significant differences in the PA power spectra of the different microvessels. The acoustic power spectrum of the 100-μm-diameter microvessel comprises more high-frequency components than the acoustic power spectra of the 200- and 300-μm-diameter microvessels.
Fig. 5.
Representative PA signals recorded by the detector in the width direction of microvessels with diameters of (a) 100, (b) 200, and (c) 300 μm. (d) Power spectra of the PA signals in (a)–(c). The PA power spectra were normalized using the maximum amplitude for directly comparing the power spectra obtained for different vascular sizes. (For interpretation of the references to colour in this figure legend, the reader is referred to the web version of this article).
The PA power spectra at all angles are graphically displayed through PA-PAS maps in Fig. 6(d)–(f). For reference, the images obtained through PACT with the delay-and-sum reconstruction algorithm [21,22] are illustrated in Fig. 6(a)–(c). As observed, the PACT images present several ring artifacts and the PACT images of microvessel with different sizes are very similar; therefore, it is impossible to visually distinguish the differences in microvascular sizes. However, such differences can clearly be observed when using the PA-PAS method. The 100-μm vessel [Fig. 6(d)] shows more high-frequency components than the 200- [Fig. 6(e)] and 300-μm [Fig. 6(f)] vessels. Compared with the PACT method, the PA-PAS method can better demonstrate the size differences among the microvessels. The PA-PAS maps show that the PA power spectra of the microvessels have significant anisotropy at different angles. There are obvious PA signals in the width direction of the microvessels and the PA spectral amplitudes are larger than that in length direction. The PWMF values were extracted from the power spectra of the PA signals for further quantitative analysis. Fig. 6(g) plots the PWMF–angle curve. The PWMF values in the width direction are higher than those in the length direction. In addition, the smaller the microvessel size, the higher the PWMF value. The experimental results verified the correctness of the simulation results, showing that the information on the microvessels is mainly concentrated in the width direction of the microvessels. The ultrasonic frequencies in the width direction can better reflect the structural sizes of the microvessels. The angular direction corresponding to the spectral amplitude peak is considered the width direction of the microvessels. The PWMF values for signals within 30° of the peak spectral amplitude were calculated to further analyze the microvascular size quantitatively. As shown in Fig. 6(h), the experimental results are in good agreement with the simulation results showing that the PWMF value tends to decrease linearly with the microvessel size, as indicated by the green dashed line. There are statistical differences between the PWMF values obtained for different microvessel sizes. To obtain the accurate sizes of microvessels in deep tissues using the PA-PAS method, curve fitting was performed using the PWMF values. The quantitative relationship between the PWMF values and microvascular sizes was established by linear fitting of the PWMF values obtained for blood vessel sizes of 200 and 300 μm. Thereafter, the PWMF value of the 100-μm-diameter microvessel, which is difficult to detect in deep tissues, was predicted using this curve. The predicted mean PWMF for the 100-μm microvessel was 3.0 MHz and the experimentally determined mean PWMF was 3.1 MHz. Hence, the experimental results were in good agreement with the predicted values. These results indicate that the accurate size of the microvessel can be obtained using the quantitative relationship between the PWMF values and microvascular sizes.
Fig. 6.
Comparison between (a–c) PACT and (d–f) PA-PAS images. (g) PWMF–angle curves of all samples (the solid yellow boxes correspond to the width direction of the microvessels). (h) PWMF values obtained for different microvascular sizes along their width direction. The PWMF values decrease linearly with the increase in the vascular diameter, as indicated by the green dashed line. (*** p < 0.001, ** p < 0.01). (For interpretation of the references to colour in this figure legend, the reader is referred to the web version of this article).
3.3. In vivo experimental results
The PA-PAS maps of two detection areas in the two groups of chick embryos are shown in Fig. 7(a) and (b). The two detection areas display an apparent distinction. Compared to that of Group-2, the power spectrum of Group-1 appears more extended along the radius axis in the width direction of the microvessel, indicating that the diameters of the blood vessels in Group-1 are smaller than those in Group-2. The PWMF–angle curve is shown in Fig.7(c). The PWMF value at the peak position of the PA spectral amplitude, namely in the width direction of the microvessel, was extracted to quantitatively characterize the sizes of the microvessels in the detected area, as shown in Fig. 7(d). The PWMF values obtained for Group-1 are higher than those obtained for Group-2, suggesting that the average sizes of the blood vessels in Group-1 are smaller than those in Group-2. The PA-PAS results are consistent with the microscopic observations [Fig.7(e)]. We also applied Welch’s t-test to the two groups of PWMFs. The statistical results show significant differences. The proposed PA-PAS method is capable of quantifying structural changes in microvessels having sizes smaller than the system resolution.
Fig. 7.
(a) and (b) Representative PA-PAS maps of the detected microvessels in Group-1 and Group-2. (c) Quantified PWMF–angle curves of the microvessels in (a) and (b) (the solid yellow boxes correspond to the width direction of microvessels). (d) Quantified PWMF values in the width direction of the microvessels in the two groups. (e) Sizes of the blood vessels in the two groups measured microscopically. Each group contains three samples. The mean size of the blood vessels in the two groups were 200 and 250 μm, respectively. (For interpretation of the references to colour in this figure legend, the reader is referred to the web version of this article).
4. Discussion
Frequency analysis is widely used in microvascular detection because of its ability to evaluate sizes smaller than the system resolution. However, owing to the tubular structures and dendritic distributions of blood vessels, the generated PA signals are inhomogeneous and exhibit anisotropy, leading to significant differences at different receiving angles. Previous studies on the power spectra of blood vessels were mainly based on comprehensive analyses of the detection regions [40,41], ignoring the power spectral differences at different angles. Hence, in this study, we developed a new method that combines the power spectra with the angles to quantify the structural changes in microvessels more accurately. The PA-PAS method maps the acoustic power spectra of the PA signals along the azimuthal direction; this reveals the physical characteristics of the microvessels, including their structural directions and sizes. Furthermore, a new spectral parameter, namely PWMF, was extracted using the power spectra of the PA signals. The amplitude of the frequency components in the power spectrum was used as the weighted coefficient; then, the main frequency of the detection area, namely the PWMF can be obtained by weighting the frequency within a predetermined interval. As the microvascular sizes in the detection area are small, there are more high-frequency components in the PA power spectra. Thus, the high-frequency weighting coefficient increases, leading to a high PWMF value. The PWMF value decreases as the microvascular size increase. The traditional spectral parameter, namely the slope [42,43], represents the energy proportions of the high- and low-frequency components in the power spectrum, whereas the PWMF represents the main frequency component of the microvessels, reflecting the main size of the microvessels in the detection area.
The classical PACT scheme has significant advantages in determining the distribution of microvessels, but the size differences among the microvessels cannot be observed in the PACT images. In contrast, the size differences can be visually observed in the PA-PAS maps. When the vascular sizes are small, the power spectra appear more extended along the radius axis in the PA-PAS maps. Compared with the PACT method, the PA-PAS method can better distinguish microvascular sizes. Moreover, the proposed PA-PAS method has the ability of resisting low SNR than the traditional PACT method, which is very helpful for in vivo evaluation. The poor SNR in vivo detection makes it impossible to obtain clear microvascular distribution images. However, using the PA-PAS method, we can still extract the acoustic power spectrum information that can reflect the microvascular structural size at a relatively low SNR. Although, the high-frequency generated by noise increased the PWMF value in the length direction of the microvessel, resulting in the widened peak of the PWMF curve. However, the PA signal amplitude in the width direction was much larger than that in the length direction of the microvessel, which was not affected by noise; thus, the structural direction of microvessel could be identified based on the position of the maximum PA amplitude. Then the PWMF values along the structural direction could be extracted to quantify the microvascular structural sizes.
The simulation and experimental results discussed above demonstrate the feasibility of quantifying the microvascular sizes using the PA-PAS method. More experiments will be conducted to improve the robustness of the system. Additional data will be used to derive the quantitative relationship between the PWMFs and vascular sizes by linear fitting of the PWMF values obtained at different vascular diameters. Based on the fitted curve, the accurate sizes of the microvessels can be obtained. The detection areas in this study contained vessels with relatively simple dendritic distributions and no complex vascular network was involved. In future, the PA-PAS method will be used to characterize microvascular networks with multiple structural directions. In addition, the method will also be used to evaluate other physical properties of the microvessels, such as elasticity.
5. Conclusion
Considering the tubular structures and dendritic distributions of blood vessels, the PA-PAS method can characterize the microvascular structural sizes comprehensively and accurately by combining the power spectra with the detection angles. The PWMF is related to the size of the microvessels and can be used for assessing the structural changes in the microvessels that are smaller than the system resolution. The results show that the PA signal information of the microvessels is concentrated in the width direction of the microvessels. The ultrasonic frequencis in the width direction can better reflect the structural sizes of the microvessels. In addition, the PWMF values decrease linearly with the microvascular sizes. The experimental and simulation results agree well, demonstrating that the proposed approach can effectively differentiate the sizes of the microvessels.
Declaration of Competing Interest
The authors declare that there are no conflicts of interest.
Acknowledgments
This project was supported by the National Natural Science Foundation of China under grant numbers 12034015, 11827808 & 11674249, and the National Key Research and Development Program of China under grant numbers 2017YFC0111400 & 2016YFA0100800.
We would like to thank Editage (www.editage.cn) for English language editing.
Biographies

Mengjiao Zhang is a Ph.D. candidate student from Institute of Acoustics, School of Physics Science and Engineering of Tongji University. She received Bachelor degree from Ocean University of China. Her current research focuses on photoacoustic measurement of microvascular evaluation and photoacoustic imaging.

Yingna Chen is currently a Ph.D. student in the Institute of Acoustics of School of Physics Science and Engineering at Tongji University at Shanghai, China. Her research interests are in the clinical translation of photoacoustic imaging and spectrum analysis.

Weiya Xie is a Ph.D. candidate student from Institute of Acoustics, School of Physics Science and Engineering of Tongji University. She received Bachelor degree from Ocean University of China. Her current research focuses on photoacoustic measurement of bone health and photoacoustic imaging.

Shiying Wu is a Ph.D. candidate student from Institute of Acoustics, School of Physics Science and Engineering of Tongji University. She received Bachelor degree from Tongji University. During her undergraduate study, she won the scholarship for outstanding students and the second prize of mathematical modeling contest. During her postgraduate study, she won the honor of doctoral student scholarship. She current research focuses on Photoacoustic spectrum analysis and Photoacoustic imaging.

Jiangnan Liao is a master degree candidate student from Institute of Acoustics, School of Physics Science and Engineering of Tongji University. She received Bachelor degree from Tongji University. Her current research focuses on Photoacoustic spectrum analysis and Photoacoustic imaging.

Qian Cheng received the B.S. degree in physics, M.S. and PhD degree in acoustics from Tongji University, China, in 2000, 2003 and 2006, respectively. She is currently a professor at Tongji University. Since 2006, her research interests were optoacoustic phenomena, near-field acoustic imaging technique, Schlieren imaging technique and the development of the acoustic detecting instruments. Her most recent research has focused on the clinical translation of photoacoustic imaging and quantitative analysis, and in particular for tumor diagnosis and evaluation.
References
- 1.Hlatky L. Clinical application of antiangiogenic therapy: microvessel density, what it does and doesn’t tell us. Cancer Spectrum Knowl. Environ. 2002;94:883–893. doi: 10.1093/jnci/94.12.883. [DOI] [PubMed] [Google Scholar]
- 2.Gazit Y., Baish J.W., Safabakhsh N., Leunig M., Baxter L.T., Jain R.K. Fractal characteristics of tumor vascular architecture during tumor growth and regression. Microcirculation. 1997;4:395–402. doi: 10.3109/10739689709146803. [DOI] [PubMed] [Google Scholar]
- 3.Dewhirst M.W., Tso C.Y., Oliver R., Gustafson C.S., Secomb T.W., Gross J.F. Morphologic and hemodynamic comparison of tumor and healing normal tissue microvasculature. Int. J. Radiat. Oncol. 1989;17:91–99. doi: 10.1016/0360-3016(89)90375-1. [DOI] [PubMed] [Google Scholar]
- 4.Zagzag D., Brem S., Robert F. Neovascularization and tumor growth in the rabbit brain. A model for experimental studies of angiogenesis and the blood-brain barrier. Am. J. Pathol. 1988;131(2):361. [PMC free article] [PubMed] [Google Scholar]
- 5.Dennie J., Mandeville J.B., Boxerman J.L., Packard S.D., Rosen B.R., Weisskoff R.M. NMR imaging of changes in vascular morphology due to tumor angiogenesis. Magn. Reson. Med. 1998;40:793–799. doi: 10.1002/mrm.1910400602. [DOI] [PubMed] [Google Scholar]
- 6.Zama A., Tamura M., Inoue H.K. Three-dimensional observations on microvascular growth in rat glioma using a vascular casting method. J. Cancer Res. Clin. Oncol. 1991;117:396–402. doi: 10.1007/BF01612757. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 7.Harada H., Xie X., Itasaka S., Zeng L., Zhu Y., Morinibu A., Shinomiya K., Hiraoka M. Diameter of tumor blood vessels is a good parameter to estimate HIF-1-active regions in solid tumors. Biochem. Biophys. Res. Commun. 2008;373(4):533–538. doi: 10.1016/j.bbrc.2008.06.062. [DOI] [PubMed] [Google Scholar]
- 8.Zhu Q., Huang M., Chen N., Zarfos K., Jagjivan B., Kane M., Hedge P., Kurtzman S.H. Ultrasound-guided optical tomographic imaging of malignant and benign breast lesions: initial clinical results of 19 cases. Neoplasia. 2003;5:379–388. doi: 10.1016/S1476-5586(03)80040-4. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9.Zhang H.F., Maslov K., Stoica G., Wang L.V. Functional photoacoustic microscopy for high-resolution and noninvasive in vivo imaging. Nat. Biotechnol. 2006;24(7):848–851. doi: 10.1038/nbt1220. [DOI] [PubMed] [Google Scholar]
- 10.Baxter G.M., Williamson T.H. Color Doppler imaging of the eye: normal ranges, reproducibility, and observer variation. J. Ultrasound Med. 1995;14:91–96. doi: 10.7863/jum.1995.14.2.91. [DOI] [PubMed] [Google Scholar]
- 11.Wladimiroff J.W., Tonge H.M., Stewart P.A. Doppler ultrasound assessment of cerebral blood flow in the human fetus. BJOG:An International Journal of O&G. 1986;93:471–475. doi: 10.1111/j.1471-0528.1986.tb07932.x. [DOI] [PubMed] [Google Scholar]
- 12.McDonald D.M., Choyke P.L. Imaging of angiogenesis: from microscope to clinic. Nat. Med. 2003;9:713–725. doi: 10.1038/nm0603-713. [DOI] [PubMed] [Google Scholar]
- 13.Miller D., Reinig J., Volkman D. Vascular imaging with MRI: inadequacy in Takayasu’s arteritis compared with angiography. Am. J. Roentgenol. 1986;146:949–954. doi: 10.2214/ajr.146.5.949. [DOI] [PubMed] [Google Scholar]
- 14.Oraevsky A.A., Andreev V.A., Karabutov A.A., Fleming R.D., Gatalica Z., Singh H., Esenaliev R.O. Laser optoacoustic imaging of the breast: detection of cancer angiogenesis. In: Chance B., Alfano R.R., Tromberg B.J., editors. BiOS '99 International Biomedical Optics Symposium; San Jose, CA; 1999. pp. 352–363. [DOI] [Google Scholar]
- 15.Hoelen C.G.A., de Mul F.F.M., Pongers R., Dekker A. Three-dimensional photoacoustic imaging of blood vessels in tissue. Opt. Lett. 1998;23:648. doi: 10.1364/OL.23.000648. [DOI] [PubMed] [Google Scholar]
- 16.Wang L.V. Ultrasound-mediated biophotonic imaging: a review of acousto-optical tomography and photo-acoustic tomography. Dis. Markers. 2004;19:123–138. doi: 10.1155/2004/478079. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 17.Rosencwaig A. Photoacoustic spectroscopy of biological materials. Science, New Series. 1973;181:657–658. doi: 10.1126/science.181.4100.657. http://www.jstor.org/stable/1736952 [DOI] [PubMed] [Google Scholar]
- 18.Yang Z.Y., Chen J.H., Yao J.J., Lin R.Q., Meng J., Liu C.B., Yang J.H., Li X., Wang L.V., Song L. L Multi-parametric quantitative microvascular imaging with optical-resolution photoacoustic microscopy in vivo. Opt. Express. 2014;22(2):1500–1511. doi: 10.1364/OE.22.001500. [DOI] [PubMed] [Google Scholar]
- 19.Lao Y., Xing D., Yang S., Xiang L. Noninvasive photoacoustic imaging of the developing vasculature during early tumor growth. Phys. Med. Biol. 2008;53:4203–4212. doi: 10.1088/0031-9155/53/15/013. [DOI] [PubMed] [Google Scholar]
- 20.Wang L.V., Yao J. A practical guide to photoacoustic tomography in the life sciences. Nat. Methods. 2016;13:627–638. doi: 10.1038/nmeth.3925. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 21.Mucci R. A comparison of efficient beamforming algorithms, IEEE Trans. Acoust., Speech, Signal Process. 1984;32:548–558. doi: 10.1109/TASSP.1984.1164359. [DOI] [Google Scholar]
- 22.Hoelen C.A., de Mul F.M. Image reconstruction for photoacoustic scanning of tissue structures. Pure Appl. Opt. J. Eur. Opt. Soc. Part A. 2000;39:5872–5883. doi: 10.1364/ao.39.005872. [DOI] [PubMed] [Google Scholar]
- 23.Zhang E.Z., Laufer J.G., Pedley R.B., Beard P.C. In vivo high-resolution 3D photoacoustic imaging of superficial vascular anatomy. Phys. Med. Biol. 2009;54:1035–1046. doi: 10.1088/0031-9155/54/4/014. [DOI] [PubMed] [Google Scholar]
- 24.Yi-Xiong S., Wang R.K., Fan Z., Jian-Quan Y. Two-dimensional photoacoustic imaging of blood vessel networks within biological tissues. Chinese Phys. Lett. 2006;23:512–515. doi: 10.1088/0256-307X/23/2/066. [DOI] [Google Scholar]
- 25.Brecht H.-P., Su R., Fronheiser M., Ermilov S.A., Conjusteau A., Oraevsky A.A. Whole-body three-dimensional optoacoustic tomography system for small animals. J. Biomed. Opt. 2009;14 doi: 10.1117/1.3259361. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 26.Wang X., Pang Y., Ku G., Xie X., Stoica G., Wang L.V. Noninvasive laser-induced photoacoustic tomography for structural and functional in vivo imaging of the brain. Nat. Biotechnol. 2003;21:803–806. doi: 10.1038/nbt839. [DOI] [PubMed] [Google Scholar]
- 27.Lungu G., Li M.-L., Xie X., Wang L., Stoica G. In vivo imaging and characterization of hypoxia-induced neovascularization and tumor invasion. Int. J. Oncol. 2007 doi: 10.3892/ijo.30.1.45. [DOI] [PubMed] [Google Scholar]
- 28.Wang X., Xie X., Ku G., Wang L.V., Stoica G. Noninvasive imaging of hemoglobin concentration and oxygenation in the rat brain using high-resolution photoacoustic tomography. J. Biomed. Opt. 2006;11 doi: 10.1117/1.2192804. [DOI] [PubMed] [Google Scholar]
- 29.Wang L.V., Hu S. Photoacoustic Tomography: In Vivo Imaging from Organelles to Organs. Science. 2012;335:1458–1462. doi: 10.1126/science.1216210. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 30.Biswas D., Vasudevan S., Chen G.C.K., Sharma N. Quantitative photoacoustic characterization of blood clot in blood: a mechanobiological assessment through spectral information. Rev. Sci. Instrum. 2017;88 doi: 10.1063/1.4974954. [DOI] [PubMed] [Google Scholar]
- 31.Wang S., Tao C., Wang X., Liu X. Quantitative detection of stochastic microstructure in turbid media by photoacoustic spectral matching. Appl. Phys. Lett. 2013;102:114102. doi: 10.1063/1.4798272. [DOI] [Google Scholar]
- 32.Xu G., Meng Z.-X., Lin J.D., Yuan J., Carson P.L., Joshi B., Wang X. The functional pitch of an organ: quantification of tissue texture with photoacoustic Spectrum analysis. Radiology. 2014;271(1):248–254. doi: 10.1148/radiol.13130777. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 33.Feng T., Perosky J.E., Kozloff K.M., Xu G., Cheng Q., Du S., Yuan J., Deng C.X., Wang X. Characterization of bone microstructure using photoacoustic spectrum analysis. Opt. Express. 2015;23:25217. doi: 10.1364/OE.23.025217. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 34.Saha R.K., Karmakar S., Roy M. Computational investigation on the photoacoustics of malaria infected red blood cells. PLoS One. 2012;7:e51774. doi: 10.1371/journal.pone.0051774. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 35.Yang Y., Wang S., Tao C., Wang X., Liu X. Photoacoustic tomography of tissue subwavelength microstructure with a narrowband and low frequency system. Appl. Phys. Lett. 2012;101 doi: 10.1063/1.4736994. [DOI] [Google Scholar]
- 36.Xu G., Dar I.A., Tao C., Liu X., Deng C.X., Wang X. Photoacoustic spectrum analysis for microstructure characterization in biological tissue: a feasibility study. Appl. Phys. Lett. 2012;101:221102. doi: 10.1063/1.4768703. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 37.Treeby B.E., Cox B.T. k-Wave: MATLAB toolbox for the simulation and reconstruction of photoacoustic wave fields, J. Biomed. Opt. 2010;15 doi: 10.1117/1.3360308. [DOI] [PubMed] [Google Scholar]
- 38.Vergara M.N., Canto-Soler M.V. Rediscovering the chick embryo as a model to study retinal development. Neural Dev. 2012;7:22. doi: 10.1186/1749-8104-7-22. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 39.Dohle D.S., Pasa S.D., Gustmann S., Laub M., Wissler J.H., Jennissen H.P., Dünker N. Chick ex ovo culture and ex ovo CAM assay: how it really works. J. Vis. Exp. 2009;33:e1620. doi: 10.3791/1620. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 40.Fadhel M.N., Hysi E., Zalev J., Kolios M.C. Photoacoustic simulations of microvascular bleeding: spectral analysis and its application for monitoring vascular-targeted treatments. J. Biomed. Opt. 2019;24:1. doi: 10.1117/1.JBO.24.11.116001. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 41.Gao X., Tao C., Wang X., Liu X. Quantitative imaging of microvasculature in deep tissue with a spectrum-based photo-acoustic microscopy. Opt. Lett. 2015;40:970. doi: 10.1364/OL.40.000970. [DOI] [PubMed] [Google Scholar]
- 42.Xu G. Photoacoustic Spectrum analysis for microstructure characterization in biological tissue: analytical model. Ultrasound Med. Biol. 2015;41(5):1473–1480. doi: 10.1016/j.ultrasmedbio.2015.01.010. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 43.Xu G., Davis M.C., Siddiqui J., Tomlins S.A., Huang S., Kunju L.P., Wei J.T., Wang X. Quantifying Gleason scores with photoacoustic spectral analysis: feasibility study with human tissues. Biomed. Opt. Express. 2015;6:4781. doi: 10.1364/BOE.6.004781. [DOI] [PMC free article] [PubMed] [Google Scholar]