Abstract
Background
To assess the diagnostic capability of novel Bruch’s membrane opening (BMO)-based disc parameters, the BMO-minimum rim width (BMO-MRW) and the BMO-minimum rim area (BMO-MRA) in the Chinese population and compare them to the retinal nerve fiber layer (RNFL) from optical coherence tomography (OCT) and the rim area (RA) from the Heidelberg retinal tomograph-III (HRT-III).
Methods
In total, 200 eyes of 77 healthy and 123 primary open-angle glaucoma (POAG) subjects were included in this cross-sectional study. All participants underwent the visual field test and structural measurements by OCT and HRT-III. The areas under the receiver operating characteristic curves (AUCs) of different structural parameters were calculated to assess their diagnostic power and compared using the DeLong test.
Results
In populations with different characteristics, the BMO-MRW and BMO-MRA had better diagnostic power than the RA. In discriminating between all POAG subjects and healthy controls and between early-stage patients and controls, the global BMO-MRW had comparable AUCs with the RNFL, but the BMO-MRA had lower AUCs than the RNFL. In healthy subjects with macrodiscs, both the global and sectoral BMO-MRW were thinner than those in healthy subjects with normal disc size. The AUCs of BMO-MRA, BMO-MRW and RNFL in subjects with macrodiscs were comparable. Additionally, in the myopic population, the BMO-MRA and BMO-MRW had comparable AUCs with the RNFL.
Conclusions
The BMO-MRW had comparable diagnostic power with the RNFL, and compared with BMO-MRW, the BMO-MRA might have advantages in certain populations, such as macrodiscs. All OCT-derived parameters exceeded the RA in diagnostic capability.
Keywords: Primary open‐angle glaucoma, Novel Bruch’s membrane opening based disc parameters, Optical coherence tomography, Glaucoma diagnosis
Background
Glaucoma is the leading cause of irreversible blindness worldwide, and primary open-angle glaucoma (POAG) is one of the most common types of glaucoma [1]. Detection and treatment of glaucoma at an early stage can effectively slow progression and maintain visual function. Since glaucomatous damage is characterized as optic disc damage, such as rim thinning or cup excavation and corresponding retinal nerve fiber layer (RNFL) defects, quantifying structural damage is crucial for the diagnosis and monitoring of glaucoma [2].
With the development of in vivo detection of optic discs and optical coherence tomography (OCT) techniques, several structural parameters have been introduced for the early detection of POAG [3–5]. Novel Bruch’s membrane opening (BMO)-based disc parameters were proposed recently for quantifying glaucomatous damage in the optic disc[6, 7], and are supposed to accurately reflect the rim tissue in consideration of the trajectory of nerve fibers and be stable for follow-up compared to traditional rim measurements with the application of a minimum distance algorithm and individual regionalization according to the personal BMO-fovea axis [8]. Previous studies have demonstrated that the one-dimensional parameter BMO-minimum rim width (BMO-MRW) has better diagnostic capability and a stronger correlation with visual field (VF) damage than traditional disc parameters, even surpassing the RNFL in some studies [9–13]; however, the BMO-MRW values in normal subjects were found to be related to disc size, and larger discs might have a physiologically thinner BMO-MRW, resulting in inaccurate comparability between subjects with different optic disc sizes [14, 15]. Therefore, the two-dimensional parameter BMO-MRA was introduced in 2014. Gardiner et al [7] first reported a better correlation of the BMO-MRA with the RNFL and VF defects than the rim area (RA) from the Heidelberg retinal tomograph (HRT). Other studies found that the BMO-MRA had better diagnostic power than the BMO-MRW and RNFL [7, 16, 17]. Furthermore, the BMO-MRA was demonstrated to compensate for the impact of disc size [16]. Therefore, the BMO-MRA is supposed to be more advantageous than the BMO-MRW when applied to comparison of different disc sizes; however, the BMO-MRA has thus far only been investigated in only a few studies with Caucasian populations.
In our study, we aimed to compare the diagnostic capability of the RNFL, BMO-MRW, BMO-MRA from OCT and the RA from the HRT in a Chinese POAG population and we hope that the results will provide clinicians with alternative and better parameters for the early diagnosis of POAG.
Methods
This was a retrospective, cross-sectional study conducted in a Chinese population. POAG patients were enrolled consecutively from 02/2015 to 07/2019 at the Department of Ophthalmology, Peking University First Hospital, and healthy controls were recruited from physical examinations. This study followed the tenets of the Declaration of Helsinki and was approved by the Ethics Committee of Peking University First Hospital.Informed consent was waived since this was a retrospective observational study.
Inclusion criteria for POAG patients were: age > 18 years, best-corrected visual acuity (BCVA) 20/40 or better, refractive error within − 6 to + 5 diopters, astigmatism < 3 diopters, open angle on gonioscopy, reliable and repeatable VF test, typical glaucomatous optic damage and corresponding RNFL defects. The exclusion criteria included the followings: VF defects caused by other ocular or neural diseases, ocular trauma, intraocular surgery within 6 months, secondary glaucoma, and other diseases that might affect retinal and optic disc structures. If both eyes fulfilled the entry criteria, the less affected eye was chosen. For healthy controls, the inclusion criteria were: age > 18 years, BCVA ≥ 20/40, refractive error within − 6 to + 5 diopters, astigmatism < 3 diopters, open angle on gonioscopy, normal VF test and IOP ≤ 21mmHg. The exclusion criteria included the followings: family history of glaucoma, intraocular surgery within 6 months, and other diseases that might affect retinal and optic disc structures. One eye of each healthy subject was randomly selected. In addition, referencing a previous study [18], participants with refractive errors greater than − 2 diopters and typical myopic optic discs (with the presence of beta-type peripapillary atrophy, either sectorally or circumferentially) were defined as myopic subjects.
All participants underwent a comprehensive eye examination, including a slit lamp examination, BCVA test, refraction test, Goldmann applanation tonometry, gonioscopy, fundus photography, measurements of the central cornea thickness and axial length, a VF test and a structural assessment by the HRT (HRT III; Heidelberg Engineering, Heidelberg, Germany) and OCT (Spectralis, Heidelberg Engineering, Heidelberg, Germany).
The Swedish Interactive Threshold Algorithm (SITA) fast 24 − 2 pattern VF test (Humphrey field analyzer, Carl Zeiss Meditec Inc., Dublin, CA, USA) was performed on each subject. Glaucomatous VF damage was defined as a cluster of at least 3 nonedge points on the pattern deviation plot with a probability < 5 %, one of which was < 1 %, a Glaucoma Hemifield Test (GHT) outside normal limits, and p < 0.05 for the pattern standard deviation (PSD). Additionally, early-stage POAG was defined as a mean deviation (MD) of -6 dB or above. Normal VF tests were defined as GHT within normal limits and MD/PSD within 95 % confidence limits. The VF test was confirmed to be reliable at a fixation loss rate < 20 %, a false-positive rate < 15 %, and a false-negative rate < 30 %.
The total disc area and total and sectoral RA were obtained from each subject with the HRT III. An optic disc with a total disc area > 2.43 mm2 was defined as a macrodisc [16]. The margin of the optic disc was drawn manually and confirmed by two observers (R.L. and X.W.). The RA was calculated using a standard reference plane. An image was excluded when strong eye movement was noted or when the standard deviation value was > 50.
Spectralis OCT was used to measure the RNFL and BMO-based disc parameters at a wavelength of 870 nm. Before the scan, the BMO center and macular fovea were confirmed. Then, 24 angularly equidistant radial scans (BMO-MRW) and 3 annular scans (RNFL) with 3.5, 4.1, and 4.6 mm centered on the BMO center were conducted. The 3.5 mm circle was used for RNFL analysis. The internal limiting membrane (ILM) and RNFL were segmented with a segmentation algorithm and corrected manually when necessary. The BMO-MRW was defined as the minimum distance between BMO and the ILM. The BMO-MRA was calculated using Spectralis SP-X VWM (Heidelberg Engineering) freely provided by the manufacturer, and described as the minimum surface between every two adjacent BMOs and ILM, which tends to be perpendicular to the trajectories of the nerve fibers as previously described [7, 16] The BMO-MRA value was determined by summing the areas of those minimum surfaces in each sector or globally (Fig. 1). BMO was recognized automatically, then confirmed by two observers and corrected manually when wrongly marked. The BMO-MRW and BMO-MRA were divided into 6 quadrants according to the BMO-fovea axis. An image was excluded when the quality score was < 15.
Statistical analysis
All data were analyzed using SPSS (SPSS version 20, Inc., Chicago, IL, USA) and MedCalc 15.6.1 (MedCalc software, Ostend, Belgium). All continuous variables are expressed as the mean ± standard deviation. The normality of variables was verified using Shapiro-Wilk tests. Student’s t-tests and Mann-Whitney U tests were used to compare variables between groups. The results were considered statistically significant when p < 0.05. Areas under the receiver operating characteristic (ROC) curves (AUCs) and sensitivities at 90 and 95 % specificity were calculated to assess the diagnostic power. The AUCs of different parameters were compared using the DeLong test [19]. Bonferroni correction was applied during multiple comparisons and the results were considered statistically significant when p < 0.008 (0.05/6).
Results
In total, 200 subjects including 77 healthy controls and 123 POAG patients, were enrolled in this study, of which 73 early-stage glaucoma patients were included. All demographic and biometric characteristics are shown in Table 1. No significant differences in age, axial length, refractive error, central corneal thickness, BMO area or disc area were found between the controls and glaucoma subjects.
Table 1.
Controls(n = 77) | POAG(n = 123) | Early-stage POAG(n = 73) | p* | p† | |
---|---|---|---|---|---|
Age (years) | 55.2 ± 8.8 | 56.8 ± 12 | 54.5 ± 11.5 | 0.064 | 0.962 |
Gender (Male/Female) | 24/53 | 62/61 | 39/34 | 0.008 | 0.008 |
Refractive error (diopters) | -1.02 ± 1.92 | -1.34 ± 2.37 | -1.36 ± 2.31 | 0.571 | 0.654 |
Axial length (mm) | 23.81 ± 0.96 | 24.04 ± 1.39 | 24.09 ± 1.33 | 0.4 | 0.143 |
Central corneal thickness (µm) | 541.8 ± 34 | 535.8 ± 34.7 | 537.1 ± 34.6 | 0.227 | 0.397 |
Mean deviation (MD, dB) | -1.39 ± 1.03 | -6.81 ± 4.88 | -3.67 ± 1.53 | < 0.001 | < 0.001 |
Visual Field Index (VFI, %) | 98.73 ± 1.14 | 84.41 ± 14.53 | 93.1 ± 4.26 | < 0.001 | < 0.001 |
Intraocular pressure (IOP, mmHg) |
14.6 ± 2.9 | 15.3 ± 2.8 | 15.4 ± 2.8 | 0.241 | 0.219 |
BMO area | 2.36 ± 0.47 | 2.38 ± 0.46 | 2.44 ± 0.47 | 0.741 | 0.27 |
Total disc area | 2.4 ± 0.54 | 2.41 ± 0.53 | 2.46 ± 0.5 | 0.876 | 0.379 |
POAG primary open-angle glaucoma, MD mean deviation, VFI visual field index, IOP intraocular pressure, BMO Bruch’s membrane opening
p* represent for comparison between controls and POAG, p† represent for comparison between controls and early-stage POAG
The mean disc area from the HRT was 2.4 ± 0.54 mm2 and 2.41 ± 0.53 mm2 in the healthy and glaucoma groups, respectively. The optic discs in 34 controls and 53 POAG patients were defined as macrodiscs. There were no significant differences in demographic or biometric characteristics between the healthy and glaucoma subjects with macrodiscs. In healthy subjects, the global BMO-MRW in macrodiscs (256.9 ± 33.7 μm) was significantly thinner than that in subjects with a normal disc size (288.8 ± 40.7 μm, p < 0.001), and statistically significant differences in the sectoral BMO-MRW were also found (p < 0.05); however, no significant differences in either the global or sectoral BMO-MRA were observed between subjects with different disc sizes (Table 2). Similarly, in POAG patients with different disc sizes, the global BMO-MRW in macrodiscs (182.8 ± 31.8 μm) was also significantly thinner than that in subjects with a normal disc size (197.9 ± 43.6 μm, p = 0.028), while no significant difference in global BMO-MRA was found.
Table 2.
Global/Total | Infero-temporal | Supero-temporal | Temporal | Infero-nasal | Supero-nasal | Nasal | |
---|---|---|---|---|---|---|---|
Controls(disc size < 2.43 mm2) | |||||||
BMO-MRW(µm) | 288.8 ± 40.7 | 321.9 ± 53.3 | 291.6 ± 48.4 | 199.9 ± 37.5 | 356.7 ± 61.3 | 328.7 ± 62.4 | 308.9 ± 53.1 |
BMO-MRA(mm2) | 1.31 ± 0.23 | 0.17 ± 0.03 | 0.14 ± 0.03 | 0.23 ± 0.05 | 0.18 ± 0.37 | 0.16 ± 0.04 | 0.41 ± 0.08 |
RNFL(µm) | 104.4 ± 11.4 | 162 ± 20.5 | 139.1 ± 20.2 | 79.9 ± 12.2 | 119.7 ± 25.5 | 120.3 ± 28.1 | 79.8 ± 15.5 |
Rim area(mm2) | 1.37 ± 0.24 | 0.19 ± 0.05 | 0.17 ± 0.04 | 0.2 ± 0.05 | 0.22 ± 0.04 | 0.2 ± 0.05 | 0.38 ± 0.09 |
Controls(disc size > 2.43 mm2) | |||||||
BMO-MRW(µm) | 256.9 ± 33.7a | 296.9 ± 37.8a | 258.7 ± 40.6a | 180.6 ± 29.3a | 322.9 ± 46.9a | 300.6 ± 46.5a | 262.9 ± 47.6a |
BMO-MRA(mm2) | 1.36 ± 0.23 | 0.18 ± 0.03 | 0.14 ± 0.03 | 0.24 ± 0.05 | 0.19 ± 0.03 | 0.18 ± 0.03 | 0.41 ± 0.08 |
RNFL(µm) | 106.1 ± 9.3 | 164.8 ± 17.9 | 138.6 ± 21.3 | 75.7 ± 8.6 | 122.5 ± 22.1 | 131.5 ± 26.3 | 82.9 ± 10.9 |
Rim area(mm2) | 1.67 ± 0.32a | 0.21 ± 0.06 | 0.2 ± 0.05a | 0.22 ± 0.06 | 0.28 ± 0.06a | 0.25 ± 0.06a | 0.48 ± 0.13a |
POAG | |||||||
BMO-MRW(µm) | 191.4 ± 39.5b | 168.6 ± 64.9b | 180.8 ± 54.9b | 148.2 ± 33.7b | 211.9 ± 62.8b | 228.1 ± 61.7b | 218.4 ± 56.9b |
BMO-MRA(mm2) | 0.95 ± 0.19b | 0.1 ± 0.04b | 0.1 ± 0.03b | 0.18 ± 0.05b | 0.12 ± 0.03b | 0.12 ± 0.03b | 0.32 ± 0.08b |
RNFL(µm) | 75.6 ± 13.9b | 84.9 ± 36.9b | 93.6 ± 32.1b | 62.3 ± 12.4b | 85.5 ± 26.9b | 96.7 ± 29.2b | 65.6 ± 16.4b |
Rim area(mm2) | 1.21 ± 0.37b | 0.13 ± 0.07b | 0.15 ± 0.06b | 0.17 ± 0.08b | 0.2 ± 0.07b | 0.19 ± 0.06b | 0.36 ± 0.12b |
BMO-MRW Bruch;s membrane opening-minimum rim width, BMO-MRA Bruch’s membrane opening-minimum rim area, RNFL retinal nerve fiber layer, POAG primary open-angle glaucoma
arepresent for statistical difference between controls with disc size < 2.43mm2 (macrodiscs) and controls with disc size > 2.43 mm2
brepresent for statistical difference between controls and POAG patients
In distinguishing POAG subjects from healthy controls, the AUC and sensitivity at 90 % specificity were 0.931 and 82.9 % for the BMO-MRW, 0.901 and 75.6 % for the BMO-MRA, 0.954 and 87 % for the RNFL, and 0.74 and 43.1 % for the RA, respectively (Table 3). No significant difference was found between the AUC of the BMO-MRW and RNFL using the DeLong test (p = 0.176). The AUC of the BMO-MRA was comparable to that of the BMO-MRW (p = 0.072) but significantly smaller than that of the RNFL (p = 0.002). In the sectoral analysis (as shown in Table 4), the parameters in the inferotemporal quadrants had the highest AUCs among all structural parameters. The AUCs of both the BMO-MRW and BMO-MRA were comparable to that of the RNFL in most quadrants, and the BMO-MRW and BMO-MRA had higher AUCs and sensitivities at 90 and 95 % specificity than the RNFL in the inferonasal quadrants (p < 0.008). All OCT parameters had better diagnostic capability than the RA from the HRT both globally and regionally.
Table 3.
AUC | 95 % confidence intervals) | Sentivity at 90 % specificity | Sentivity at 95 % specificity | |
Global RNFL | 0.954 | (0.93,0.979) | 87 | 79.7 |
Global BMO-MRW | 0.931 | (0.896,0.965) | 82.9 | 69.1 |
Global BMO-MRA | 0.901 | (0.86.0.942) | 75.6 | 61.8 |
Global rim area | 0.74 | (0.673,0.808) | 43.1 | 35 |
DeLong- test for AUCs | ||||
AUC1 | AUC2 | p | ||
Global BMO-MRW | 0.931 | Global BMO-MRA | 0.901 | 0.072 |
Global BMO-MRW | 0.931 | Global RNFL | 0.954 | 0.176 |
Global BMO-MRW | 0.931 | Total rim area | 0.74 | < 0.001 |
Global BMO-MRA | 0.901 | Global RNFL | 0.954 | 0.002 |
Global BMO-MRA | 0.901 | Total rim area | 0.74 | < 0.001 |
Global RNFL | 0.954 | Total rim area | 0.74 | < 0.001 |
The results were considered statistically significant when p < 0.008
AUC area under the curve, BMO-MRW Bruch;s membrane opening-minimum rim width, BMO-MRA Bruch’s membrane opening-minimum rim area, RNFL retinal nerve fiber layer
Table 4.
AUC(95 % confidence intervals) | Sentivity at 90 % specificity | Sentivity at 95 % specificity | |
---|---|---|---|
BMO-MRW | |||
Inferotemporal | 0.956(0.932,0.98) | 85.4 | 81.3 |
Superotemporal | 0.908(0.869,0.947) | 76.4 | 69.9 |
temporal | 0.814(0.757,0.872) | 56.1 | 44.7 |
Inferonasal | 0.937(0.905,0.969) | 82.9 | 73.2 |
Superonasal | 0.856(0.805,0.906) | 65.9 | 60.2 |
Nasal | 0.814(0.754,0.873) | 58.5 | 53.7 |
BMO-MRA | |||
Inferotemporal | 0.93(0.897,0.964) | 82.1 | 75.6 |
Superotemporal | 0.873(0.827,0.92) | 65.9 | 59.3 |
temporal | 0.754(0.688,0.82) | 45.5 | 34.1 |
Inferonasal | 0.918(0.882,0.955) | 78 | 66.7 |
Superonasal | 0.831(0.775,0.887) | 50.4 | 42.3 |
Nasal | 0.796(0.732,0.86) | 47.2 | 27.6 |
RNFL | |||
Inferotemporal | 0.957(0.93,0.984) | 91.1 | 90.2 |
Superotemporal | 0.878(0.831,0.925) | 71.5 | 59.3 |
temporal | 0.84(0.785,0.894) | 66.7 | 47.2 |
Inferonasal | 0.849(0.797,0.901) | 64.2 | 58.5 |
Superonasal | 0.762(0.696,0.828) | 40.7 | 34.1 |
Nasal | 0.77(0.704,0.836) | 43.9 | 32.5 |
Rim area | |||
Inferotemporal | 0.804(0.743,0.864) | 58.5 | 44.7 |
Superotemporal | 0.709(0.637,0.78) | 35.8 | 31.7 |
temporal | 0.683(0.61,0.756) | 37.4 | 32.5 |
Inferonasal | 0.701(0.628,0.773) | 39 | 20.3 |
Superonasal | 0.648(0.571,0.726) | 26 | 20.3 |
Nasal | 0.642(0.564,0.721) | 28.5 | 17.1 |
AUC area under the curve, BMO-MRW Bruch;s membrane opening-minimum rim width, BMO-MRA Bruch’s membrane opening-minimum rim area, RNFL retinal nerve fiber layer
While discriminating between early-stage POAG patients and controls, all structural parameters showed lower AUCs and sensitivities than those between all glaucoma patients and controls (Table 5). Similarly, the AUC of the global BMO-MRW was comparable to that of the RNFL (p = 0.29), but the BMO-MRA had lower AUCs than the RNFL (p = 0.004). As shown in Table 6, while assessing the diagnostic capability in subjects with macrodiscs, the AUCs of the BMO-MRW and BMO-MRA increased, but no statistically significant differences were found between the AUCs of the global BMO-MRW, BMO-MRA and RNFL. All OCT parameters also surpassed the RA from the HRT in AUCs and sensitivities. Additionally, we assessed the diagnostic power in myopic subjects, including 18 healthy controls and 45 POAG patients (the demographic data were comparable between healthy and glaucoma subjects), and found that the BMO-MRW, BMO-MRA and RNFL had comparable AUCs and sensitivities, as shown in Table 7.
Table 5.
ROC | AUC | 95 % confidence intervals) | Sentivity at 90 % specificity | Sentivity at 95 % specificity |
Global RNFL | 0.933 | (0.897,0.969) | 82.2 | 71.2 |
Global BMO-MRW | 0.907 | (0.859,0.954) | 76.7 | 57.5 |
Global BMO-MRA | 0.859 | (0.801,0.917) | 65.8 | 46.6 |
Global rim area | 0.69 | (0.605,0.775) | 37 | 30.1 |
DeLong- test for AUCs | ||||
AUC1 | AUC2 | p | ||
Global BMO-MRW | 0.907 | Global BMO-MRA | 0.859 | 0.03 |
Global BMO-MRW | 0.907 | Global RNFL | 0.933 | 0.29 |
Global BMO-MRW | 0.907 | Total rim area | 0.69 | < 0.001 |
Global BMO-MRA | 0.859 | Global RNFL | 0.933 | 0.004 |
Global BMO-MRA | 0.859 | Total rim area | 0.69 | < 0.001 |
Global RNFL | 0.933 | Total rim area | 0.69 | < 0.001 |
The results were considered statistically significant when p < 0.008
AUC area under the curve, POAG primary open-angle glaucoma, BMO-MRW Bruch;s membrane opening-minimum rim width, BMO-MRA Bruch’s membrane opening-minimum rim area, RNFL retinal nerve fiber layer
Table 6.
ROC | AUC | 95 % confidence intervals) | Sentivity at 90 % specificity | Sentivity at 95 % specificity |
Global RNFL | 0.951 | (0.912,0.99) | 87 | 79.7 |
Global BMO-MRW | 0.946 | (0.9,0.993) | 82.9 | 69.1 |
Global BMO-MRA | 0.921 | (0.865,0.977) | 75.6 | 61.8 |
Global rim area | 0.768 | (0.669,0.867) | 43.1 | 35 |
DeLong- test for AUCs | ||||
AUC1 | AUC2 | p | ||
Global BMO-MRW | 0.946 | Global BMO-MRA | 0.921 | 0.091 |
Global BMO-MRW | 0.946 | Global RNFL | 0.951 | 0.835 |
Global BMO-MRW | 0.946 | Total rim area | 0.768 | < 0.001 |
Global BMO-MRA | 0.921 | Global RNFL | 0.951 | 0.212 |
Global BMO-MRA | 0.921 | Total rim area | 0.768 | < 0.001 |
Global RNFL | 0.951 | Total rim area | 0.768 | < 0.001 |
The results were considered statistically significant when p < 0.008
AUC area under the curve, BMO-MRW Bruch;s membrane opening-minimum rim width, BMO-MRA Bruch’s membrane opening-minimum rim area, RNFL retinal nerve fiber layer
Table 7.
ROC | AUC | 95 % confidence intervals) | Sentivity at 90 % specificity | Sentivity at 95 % specificity |
Global BMO-MRW | 0.945 | (0.895,0.995) | 84.4 | 82.2 |
Global RNFL | 0.943 | (0.888,0.997) | 77.8 | 77.8 |
Global BMO-MRA | 0.934 | (0.876,0.991) | 77.8 | 77.8 |
Global rim area | 0.698 | (0.564,0.831) | 53.3 | 33.3 |
DeLong- test for AUCs | ||||
AUC1 | AUC2 | p | ||
Global BMO-MRW | 0.945 | Global BMO-MRA | 0.934 | 0.648 |
Global BMO-MRW | 0.945 | Global RNFL | 0.943 | 0.939 |
Global BMO-MRW | 0.945 | Total rim area | 0.698 | < 0.001 |
Global BMO-MRA | 0.934 | Global RNFL | 0.943 | 0.773 |
Global BMO-MRA | 0.934 | Total rim area | 0.698 | < 0.001 |
Global RNFL | 0.943 | Total rim area | 0.698 | < 0.001 |
The results were considered statistically significant when p < 0.008
AUC area under the curve, BMO-MRW Bruch;s membrane opening-minimum rim width, BMO-MRA Bruch’s membrane opening-minimum rim area, RNFL retinal nerve fiber layer
Discussion
The available literature on novel BMO-based disc parameters, especially the BMO-MRA, have included mostly Caucasian populations, Therefore, in this study, the diagnostic capability of both the BMO-MRW and BMO-MRA was first assessed and compared to that of conventional parameters, the RNFL from OCT and the RA from the HRT in a Chinese population.
Consistent with most previous studies [10, 12, 17, 20], we found that the global BMO-MRW and RNFL had comparable capability in distinguishing POAG subjects from healthy subjects as well as in the detection of early-stage POAG patients (Tables 3 and 5). For the BMO-MRA, we found that the global BMO-MRA offered lower diagnostic power than the RNFL (Tables 3 and 5). However, as presented in Table 4, sectoral analysis showed that both the BMO-MRA and BMO-MRW yielded a comparable diagnostic performance with the RNFL in the supero-nasal, supero-temporal and infero-temporal quadrants (p > 0.008) and surpassed the RNFL in the infero-nasal quadrants (p < 0.008). These findings differed from previous studies: Enders et al[16] found that the BMO-MRA had better diagnostic power than the BMO-MRW and RNFL in subjects with different disc sizes, and recently, the authors demonstrated that the BMO-MRA, BMO-MRW and RNFL offered a comparable level of diagnostic capability in a large cohort [17]. Since the diagnostic capability of the BMO-MRA in previous studies varied between different cohorts, we speculate that the relatively inferior diagnostic power of the global BMO-MRA in our study in comparison with previous findings might be ascribed to the different disease stages of the glaucoma patients enrolled in the different studies. In addition, combined with global and sectoral findings, the inferior diagnostic capability of the global BMO-MRA might also be related to the algorithm: the global BMO-MRA was the sum of each sector, while the global BMO-MRW and RNFL were the average of sectoral values. Therefore, we speculate that the value and diagnostic performance of the global BMO-MRA might be influenced more by the nasal and temporal quadrants than other parameters, which resulted in relatively low diagnostic power. On the other hand, similar to previous findings, we also found that both the BMO-MRW and BMO-MRA performed significantly better than the RA from the HRT in AUCs and sensitivities both globally and regionally, indicating that novel BMO-based disc parameters might be preferred options for the assessment of optic disc damage.
Additionally, we divided the subjects according to disc size and assessed diagnostic capability in macrodiscs and found that both the BMO-MRA and BMO-MRW yielded similar performances with the RNFL in differentiating POAG subjects from controls, as shown in Table 6. However, we found that both the global and sectoral BMO-MRW in healthy subjects with macrodiscs were significantly thinner than those in healthy subjects with a normal disc size, while the BMO-MRA was comparable between subjects with different disc sizes (Table 2). These findings are in accordance with Enders’s studies [16], which demonstrated that the BMO-MRA could compensate for the influence of disc size, indicating that the BMO-MRA might represent a better tool than the BMO-MRW in the management of subjects with macrodiscs and more suitable in subjects with a wide range of disc sizes.
Myopia is a widely accepted possible risk factor for POAG, but the diagnosis of POAG in myopia with available structural parameters, such as the RNFL and traditional rim parameters, remains challenging [21, 22]. Previous studies have reported that the BMO-MRW had comparable diagnostic power and a stronger correlation with VF damage than the RNFL in myopic subjects [18, 23], showing the potential capability of the BMO-MRW in distinguishing POAG in myopic subjects, but the diagnostic power of the BMO-MRA was not investigated. Therefore, referencing Malik’s study [18], as presented in Table 7, we compared the diagnostic performance of the BMO-MRW and BMO-MRA with conventional morphological parameters in myopic participants with refraction errors greater than − 2 diopters and typical myopic optic discs. Similarly, we found that both the BMO-MRW and BMO-MRA had comparable diagnostic capability with the RNFL, further indicating the possible utility of novel BMO-based disc parameters in the myopic population.
Our study has some limitations. First, the sample size in our study was relatively small, especially the number of subjects with macrodiscs and myopia, and all patients were recruited from our clinical center, which might result in selection bias. Additionally, data from the relatively few glaucoma subjects included in our study do not represent the full spectrum of the disease. Therefore, the diagnostic performance of novel BMO-based disc parameters in the Chinese population must be verified in a larger population with a full range of disc sizes from multiple centers. Second, the BMO-MRA was calculated based on the minimum surface between BMO and the ILM in each sector and therefore might not reflect the actual minimum area where neural tissue passes through. Further modifications of the calculation might be needed for evaluation in future studies.
In conclusion, in this single-center Chinese cohort study, the BMO-MRW had comparable diagnostic power with the RNFL, and the BMO-MRA could compensate for the influence of disc size and offer similar capability with the BMO-MRW and RNFL in subjects with macrodiscs and myopia. All OCT-derived parameters exceeded the RA from the HRT in diagnostic capability. Novel BMO-based morphological parameters might serve as additional tools in the clinical management of glaucoma.
Acknowledgements
Special thanks go to Xueying Li who assisted the statistical analysis in this study.
Authors’ contributions
RSL made substantial contributions to data collection, analysis and interpretation, and drafting the article as well. XW and YHW had contributed to acquisition of data and analysis of data. YF and TT had contributed to analysis of data and design of the study. ML, YC and LK had contributed to acquisition of data. YZP designed the study and participated in revising the manuscript critically for important intellectual content. All authors have read and approved the manuscript.
Funding
No funding was received for this study.
Availability of data and materials
The datasets used and/or analysed during the current study are not publically available and can be obtained from the corresponding author on reasonable request.
Declarations
Ethics approval and consent to participate
All procedures performed in this study involving human participants were in accordance with the ethical standards of the Institutional Review Board (IRB) of Peking University First Hospital and with the 1964 Helsinki declaration and its later amendments or comparable ethical standards.As a retrospective study, informed consent was not required.
Consent for publication
Not applicable.
Competing interests
All the authors declared that they had no conflict of interest.
Footnotes
Publisher’s Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
References
- 1.Tham YC, Li X, Wong TY, Quigley HA, Aung T, Cheng CY. Global prevalence of glaucoma and projections of glaucoma burden through 2040: a systematic review and meta-analysis. Ophthalmology. 2014;121(11):2081–90. doi: 10.1016/j.ophtha.2014.05.013. [DOI] [PubMed] [Google Scholar]
- 2.Jonas JB, Aung T, Bourne RR, Bron AM, Ritch R, Panda-Jonas S. Glaucoma Lancet. 2017;390(10108):2183–93. doi: 10.1016/S0140-6736(17)31469-1. [DOI] [PubMed] [Google Scholar]
- 3.Bussel II, Wollstein G, Schuman JS. OCT for glaucoma diagnosis, screening and detection of glaucoma progression. Br J Ophthalmol. 2014;98(Suppl 2(Suppl 2):ii15–9. [DOI] [PMC free article] [PubMed]
- 4.Leung CK, Ye C, Weinreb RN, Cheung CY, Qiu Q, Liu S, et al. Retinal nerve fiber layer imaging with spectral-domain optical coherence tomography a study on diagnostic agreement with Heidelberg Retinal Tomograph. Ophthalmology. 2010;117(2):267–74. doi: 10.1016/j.ophtha.2009.06.061. [DOI] [PubMed] [Google Scholar]
- 5.Mwanza J-C, Durbin MK, Budenz DL, Sayyad FE, Chang RT, Neelakantan A, et al. Glaucoma diagnostic accuracy of ganglion cell-inner plexiform layer thickness: comparison with nerve fiber layer and optic nerve head. Ophthalmology. 2012;119(6):1151–8. doi: 10.1016/j.ophtha.2011.12.014. [DOI] [PubMed] [Google Scholar]
- 6.Reis ASC, O’Leary N, Yang H, Sharpe GP, Nicolela MT, Burgoyne CF, et al. Influence of clinically invisible, but optical coherence tomography detected, optic disc margin anatomy on neuroretinal rim evaluation. Invest Ophthalmol Vis Sci. 2012;53(4):1852–60. doi: 10.1167/iovs.11-9309. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 7.Gardiner SK, Ren R, Yang H, Fortune B, Burgoyne CF, Demirel S. A method to estimate the amount of neuroretinal rim tissue in glaucoma: comparison with current methods for measuring rim area. Am J Ophthalmol. 2014;157(3):540–9.e92. doi: 10.1016/j.ajo.2013.11.007. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8.Chauhan BC, Burgoyne CF. From clinical examination of the optic disc to clinical assessment of the optic nerve head: a paradigm change. Am J Ophthalmol. 2013;156(2):218–27.e2. doi: 10.1016/j.ajo.2013.04.016. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9.Pollet-Villard F, Chiquet C, Romanet J-P, Noel C, Aptel F. Structure-function relationships with spectral-domain optical coherence tomography retinal nerve fiber layer and optic nerve head measurements. Invest Ophthalmol Vis Sci. 2014;55(5):2953–62. doi: 10.1167/iovs.13-13482. [DOI] [PubMed] [Google Scholar]
- 10.Di Staso S, Agnifili L, Di Staso F, Climastone H, Ciancaglini M, Scuderi GL. Diagnostic capability of optic nerve head rim width and retinal nerve fiber thickness in open-angle glaucoma. Eur J Ophthalmol. 2018;28(4):459–64. doi: 10.1177/1120672117750057. [DOI] [PubMed] [Google Scholar]
- 11.Enders P, Schaub F, Adler W, Hermann MM, Dietlein TS, Cursiefen C, et al. Bruch’s membrane opening-based optical coherence tomography of the optic nerve head: a useful diagnostic tool to detect glaucoma in macrodiscs. Eye (Lond) 2018;32(2):314–23. doi: 10.1038/eye.2017.306. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12.Gmeiner JM, Schrems WA, Mardin CY, Laemmer R, Kruse FE, Schrems-Hoesl LM. Comparison of Bruch’s Membrane Opening Minimum Rim Width and Peripapillary Retinal Nerve Fiber Layer Thickness in Early Glaucoma Assessment. Invest Ophthalmol Vis Sci. 2016;57(9):Oct575–84. doi: 10.1167/iovs.15-18906. [DOI] [PubMed] [Google Scholar]
- 13.Chauhan BC, O’Leary N, AlMobarak FA, Reis ASC, Yang H, Sharpe GP, et al. Enhanced detection of open-angle glaucoma with an anatomically accurate optical coherence tomography-derived neuroretinal rim parameter. Ophthalmology. 2013;120(3):535–43. doi: 10.1016/j.ophtha.2012.09.055. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14.Enders P, Schaub F, Hermann MM, Cursiefen C, Heindl LM. Neuroretinal rim in non-glaucomatous large optic nerve heads: a comparison of confocal scanning laser tomography and spectral domain optical coherence tomography. Br J Ophthalmol. 2017;101(2):138–42. doi: 10.1136/bjophthalmol-2015-307730. [DOI] [PubMed] [Google Scholar]
- 15.Hoffmann EM, Zangwill LM, Crowston JG, Weinreb RN. Optic disk size and glaucoma. Surv Ophthalmol. 2007;52(1):32–49. doi: 10.1016/j.survophthal.2006.10.002. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16.Enders P, Adler W, Schaub F, Hermann MM, Dietlein T, Cursiefen C, et al. Novel Bruch’s Membrane Opening Minimum Rim Area Equalizes Disc Size Dependency and Offers High Diagnostic Power for Glaucoma. Invest Ophthalmol Vis Sci. 2016;57(15):6596–603. doi: 10.1167/iovs.16-20561. [DOI] [PubMed] [Google Scholar]
- 17.Enders P, Adler W, Kiessling D, Weber V, Schaub F, Hermann MM, et al. Evaluation of two-dimensional Bruch’s membrane opening minimum rim area for glaucoma diagnostics in a large patient cohort. Acta Ophthalmol. 2019;97(1):60–7. doi: 10.1111/aos.13698. [DOI] [PubMed] [Google Scholar]
- 18.Malik R, Belliveau AC, Sharpe GP, Shuba LM, Chauhan BC, Nicolela MT. Diagnostic Accuracy of Optical Coherence Tomography and Scanning Laser Tomography for Identifying Glaucoma in Myopic Eyes. Ophthalmology. 2016;123(6):1181–9. doi: 10.1016/j.ophtha.2016.01.052. [DOI] [PubMed] [Google Scholar]
- 19.Delong ER, Delong DM, Clarke-Pearson DL. Comparing the Areas under Two or More Correlated Receiver Operating Characteristic Curves: A Nonparametric Approach. Biometrics. 1988;44(3):837–45. doi: 10.2307/2531595. [DOI] [PubMed] [Google Scholar]
- 20.Enders P, Schaub F, Adler W, Nikoluk R, Hermann MM, Heindl LM. The use of Bruch’s membrane opening-based optical coherence tomography of the optic nerve head for glaucoma detection in microdiscs. Br J Ophthalmol. 2017;101(4):530–5. doi: 10.1136/bjophthalmol-2016-308957. [DOI] [PubMed] [Google Scholar]
- 21.Tan NYQ, Sng CCA, Jonas JB, Wong TY, Jansonius NM, Ang M. Glaucoma in myopia: diagnostic dilemmas. Br J Ophthalmol. 2019;103(10):1347–55. doi: 10.1136/bjophthalmol-2018-313530. [DOI] [PubMed] [Google Scholar]
- 22.Marcus MW, de Vries MM, Junoy Montolio FG, Jansonius NM. Myopia as a risk factor for open-angle glaucoma: a systematic review and meta-analysis. Ophthalmology. 2011;118(10):1989–94.e2. doi: 10.1016/j.ophtha.2011.03.012. [DOI] [PubMed] [Google Scholar]
- 23.Reznicek L, Burzer S, Laubichler A, Nasseri A, Lohmann CP, Feucht N, et al. Structure-function relationship comparison between retinal nerve fibre layer and Bruch’s membrane opening-minimum rim width in glaucoma. Int J Ophthalmol. 2017;10(10):1534–8. doi: 10.18240/ijo.2017.10.09. [DOI] [PMC free article] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Data Availability Statement
The datasets used and/or analysed during the current study are not publically available and can be obtained from the corresponding author on reasonable request.