Skip to main content
Scientific Reports logoLink to Scientific Reports
. 2021 Mar 25;11:6875. doi: 10.1038/s41598-021-86310-2

Hemodynamic analysis for stenosis microfluidic model of thrombosis with refined computational fluid dynamics simulation

Yunduo Charles Zhao 1,2,#, Parham Vatankhah 1,#, Tiffany Goh 1,2,3, Rhys Michelis 4, Kiarash Kyanian 1, Yingqi Zhang 1, Zhiyong Li 5, Lining Arnold Ju 1,2,3,
PMCID: PMC7994556  PMID: 33767279

Abstract

Disturbed blood flow has been increasingly recognized for its critical role in platelet aggregation and thrombosis. Microfluidics with hump shaped contractions have been developed to mimic microvascular stenosis and recapitulate the prothrombotic effect of flow disturbance. However the physical determinants of microfluidic hemodynamics are not completely defined. Here, we report a refined computational fluid dynamics (CFD) simulation approach to map the shear rate (γ) and wall shear stress (τ) distribution in the stenotic region at high accuracy. Using ultra-fine meshing with sensitivity verification, our CFD results show that the stenosis level (S) is dominant over the bulk shear rate (γ0) and contraction angle (α) in determining γ and τ distribution at stenosis. In contrast, α plays a significant role in governing the shear rate gradient (γ) distribution while it exhibits subtle effects on the peak γ. To investigate the viscosity effect, we employ a Generalized Power-Law model to simulate blood flow as a non-Newtonian fluid, showing negligible difference in the γ distribution when compared with Newtonian simulation with water medium. Together, our refined CFD method represents a comprehensive approach to examine microfluidic hemodynamics in three dimensions and guide microfabrication designs. Combining this with hematological experiments promises to advance understandings of the rheological effect in thrombosis and platelet mechanobiology.

Subject terms: Biomedical engineering, Lab-on-a-chip

Introduction

Thrombotic diseases have become the leading cause of modern mortality1,2. Thrombi, also known as blood clots, mainly consist of aggregated platelets. They become dangerous when they grow large and occlude blood vessels in the heart, brain and peripheral vascularized organs, leading to heart attack, stroke and deep vein thrombosis respectively2. While it is well known that platelet adhesion, activation, and subsequent aggregation play a central role in thrombosis, the interplay of biochemical and biomechanical factors regulating platelet thrombosis remains incompletely understood35.

Recent studies have observed the association between enhanced platelet aggregation and blood flow disturbance3,6. In vivo, blood flow within vessels exerts hemodynamic forces on both the vessel surface in the form of wall shear stress (WSS; τ), as well as blood components such as red blood cells (RBC), white blood cells and platelets in the form of shear rate (γ). τ is the tangential force on the vessel or channel wall due to friction, while γ describes the shear effect that the interstitial fluid experiences during flow7. It has been recognized that a high shear rate and its gradient may activate mechanosensitive proteins such as von Willebrand factor (VWF)8,9 and platelet glycoprotein Ib receptor1012, leading to subsequent platelet activation and aggregation7,13. Nevertheless, the relationship between the microfluidic boundary conditions (γ0, S, α,) and hemodynamic parameters of flow disturbance, such as the peak shear rate (γmax), peak WSS (τmax) and shear rate gradient (γ) at the stenotic (narrowing) region, remains incompletely defined5,14,15. Furthermore, it has been hypothesized that biomechanical platelet aggregation is not simply attributed to elevated shear and elongational force at the narrowing apex alone4, but the entire ‘shear history’ where the platelet experiences accumulated mechanical stimuli and concomitant cellular interactions15,16. Thus, it is important to map the hemodynamic distribution at both the pre- and post-stenosis regions5.

In a healthy patent blood vessel (Fig. 1A, top), the blood flow pattern is laminar and steady, yielding a constant velocity profile, γ and τ distribution. However, when a vessel is narrowed concentrically from all directions (e.g. atherosclerotic plaque formation; Fig. 1A, middle)30, or eccentrically narrowed from one direction (e.g. medical device insertion; Fig. 1A, bottom)31,32, the τ and γ0 become elevated at the stenosis with blood flow acceleration (Table 1). To mimic the pathological microvascular conditions, stenosis microfluidic models of thrombosis have been designed with a default width of Y0 = 100 µm, height of Z0 = 130 µm, α = 85° and S = 80% (Fig. 1B)13,33. It is worth noting that the flow profile within such microchannels is considered two-dimensional (2D), as the geometry and flow vary only in the x–y plane at the stenotic region (Fig. 1B)7,13,34. Previous studies have justified these stenosis microfluidic models of thrombosis which have been well characterized and validated experimentally6,7,13,16,38,40, demonstrating the shear rate gradient effects on platelet aggregation in blood flow disturbance. The advantages of such microfluidic approaches enable direct visualization of biomechanical platelet aggregation at the downstream face of stenosis (Fig. 1C). Nevertheless, the detailed rheological mechanisms, or specifically the exact τ, γ and γ thresholds that trigger such biomechanical platelet aggregation are incompletely understood. In this context, CFD simulation represents the first step in revealing the hemodynamic profile within disturbed blood flow and correlating with experimental results of thrombotic response.

Figure 1.

Figure 1

Pathological and physiological blood vessel stenoses and the mimicking microfluidic models. (A) Blood flow in a healthy vessel without stenosis (top), in a vessel with concentric stenosis due to atherosclerotic plaque (middle) and a vessel with eccentric stenosis due to medical device insertion (bottom). (B) CFD contour maps in the eccentric (top) and concentric (bottom) stenosis microfluidic channels. Note that the channel walls were colored to display the WSS distribution; the representative streamline of a platelet trajectory was colored to display the shear rate γ distribution. (C) Differential interference contrast microscopy image of biomechanical platelet aggregation in an eccentric stenosis microfluidic channel after whole blood perfusion at γ0 = 1,800 s−1, mimicking the physiological shear rate in arteries and arterioles7. To visualize platelet aggregates in better clarity, the microchannel was washed with Tyrode’s buffer after whole blood perfusion. Note that platelets aggregate at the downstream face of the stenosis. Scale bar = 20 µm. (D) ANSYS finite volume meshing scheme for the concentric stenosis model as in panel C. The entire microfluidic channel was meshed into 1,638,336 hexahedral elements for CFD analysis. The coordinate origin is located at the center of the front bottom edge. Note that for illustration purpose, a coarse mesh is shown in the figure.

Table 1.

Physiological and pathological shear rate ranges of in vivo circulating blood.

Vascular conditions Vessel types WSS τ (dyne/cm2) Bulk shear rate γ0 (s−1)
Healthy vessel Arteries 5–361719 300–80018,20
Veins 1–617,21,22 15–20018,20
Microvessels 20–8021 450–180018,20
Concentric stenosis Atherosclerotic plaque 36–6,00023,24 800–10,00018,20,25
Eccentric stenosis Occlusive thrombus 36–45017 5,000–400,00025
Coronary stent 0–4017,26 0–11,00027
Catheter insertion 0–20,00017,28  > 10,00029

CFD is the most popular computational method to model hemodynamic parameters i.e. τ, γ and γ and simulate microfluidic outcomes before experiments are done7,15. However, rarely do the CFD practices nowadays thoroughly show the mesh sensitivity verification, or systematically benchmark the effects of the microfluidic boundary conditions on the hemodynamic outcome with certain control variables i.e. γ0, S, α and fluid medium (Table 2). The caveat is that the CFD solutions may have some extents of error with coarse mesh sizes. Besides, although the non-Newtonian property of blood has significant impacts on the viscosity, its influence in the disturbed flow region remains poorly characterized. Hereby, we present an ultra-fine CFD study to map hemodynamic profiles for stenosis microfluidic models of thrombosis, and address the above concerns with mesh sensitivity verification and analytical validation.

Table 2.

Working fluids modeled and physical properties.

Working fluid Density ρ (kg m−3) Viscosity µ (Pa s)
Water 998 0.001003
Blood (Newtonian) 1,060 0.003450
Blood (non-Newtonian) 1,060 GPL model

Methods

Stenosis microfluidic models of thrombosis

Microfluidics is a miniaturized approach that has significant advantages in controling hemodynamic parameters, ligand presentation and agonist stimulation at the micro-scale13,3537. Microfluidic channels with hump-like contractions have been developed and utilized to mimic microvessel stenosis and examine the prothrombotic effect of flow disturbance through which whole blood or washed platelets are perfused (Fig. 1C)6,7,13,3841.

Numerical formulation and governing equations

The commercially available software ANSYS FLUENT version 2020 R1 is utilized to computationally simulate the flow profile. The flow was assumed as steady, laminar, and incompressible. Under these assumptions, the fluid can be described using the continuity Eq. (1) and Cauchy momentum Eq. (2) as follows:

·u=0 1
ρu·u=-P+·μu+uT 2

where u is the 3D velocity vector (m s−1), P is the pressure (Pa), μ is the dynamic viscosity (Pa s) and ρ is the density (kg m−3). This can be simplified to the well-known Navier–Stokes equation for the case of Newtonian fluids (fluids with constant dynamic viscosity):

ρu·u=-P+μ2u 3

In ANSYS FLUENT, to assess the viscosity effect we simulated 3 working mediums: water, blood as a Newtonian fluid, and blood as a non-Newtonian fluid under the Generalized Power-Law (GPL) viscosity model (Table 2).

Numerous non-Newtonian models have been used throughout literature for simulating blood flow: Carreau42, Modified Cross Law (Carreau-Yasuda)43, Power-Law (PL)44, non-Newtonian Power-Law44, Generalized Power-Law45, Casson46 and Walburn-Schneck Law47 are among the well-recognized models48. In this study, the GPL model is utilized due to its accuracy in calculating the shear rate γ48, which is calculated from:

μ=Kγn-1 4

where K is the consistency index, and n is the Power-Law index. The two parameters are calculated from45:

K=μ+δμexp-1+γaexp-bγ 5
n=n-δnexp-1+γcexp-dγ 6

where µ = 0.00345 Pa s, n = 1.0, δµ = 0.25, δn = 0.45, a = 50, b = 3, c = 50 and d = 445.

Geometrical properties, boundary conditions and Reynolds number

A schematic diagram of the microfluidic channel is illustrated in Fig. 1B. By default, the axial length is X0 = 200 µm, the width is Y0 = 100 µm and the height of the cross-section is Z0 = 130 µm. A zero-gauge pressure boundary condition is applied at the inlet of the microchannel (Fig. 1B). A no-slip boundary condition is applied to the walls. Finally, a boundary condition of the experimentally implemented flow rate Q (µL min−1) is implemented at the outlet of the flow region (Fig. 1B,C). The flow rate calculation is defined as49:

Q=0.12ADhγ0λ 7

where γ0 (s−1) is the bulk shear rate, A (µm2) is the cross-sectional area, Dh (m) is the hydraulic diameter, and, λ is the shape factor of the microfluidic cross-section49. A, Dh and λ are calculated by the following equations:

A=Y0Z0 8
Dh=2AY0+Z0 9
λ=241-0.351Y0/Z01+Y0/Z02 10

The above also allows for the calculation of the Reynolds number, a dimensionless group used to indicate whether fluid flow in a channel is characterized as laminar or turbulent. The Reynolds number for a Newtonian fluid is given by:

Re=Dhρ2QμA 11

Substituting Q from Eq. (7) into Eq. (11) yields the following expression of the Reynolds number for this particular investigation:

Re=0.12Dh2ρ2γoλμ 12

Using γ0 = 150 and 3,000 s−1 for Eq. (12), the Reynolds number for this study is found to range from 0.04 to 0.9. Unlike the turbulent state of blood flow in the stenotic region of the common carotid arteries with large diameters shown in some studies5052, a Reynolds number with such low magnitude indicates a laminar flow regime, as flow typically does not transition into a turbulent regime before the critical point of Re = 2,30053, validating our laminar flow assumption.

Furthermore, our steady flow assumption stems from the Windkessel effect which dampens the pulsatile flow changes far away from the heart, leading to steady flow in the capillary region54,55. This steady assumption is further validated by the Womersley number which is a dimensionless group to evaluate the pulsatile flow effect in relation to oscillation frequency and viscosity56.

Wo=D2ρωμ 13
ω=2πf 14

where D is the diameter of the vessel (m); f and ω are the oscillation and angular frequency (both Hz) respectively. We selected f = 2 Hz and D = 130 µm, corresponding to a heart rate of 120 beats per minute to validate our assumption at even extreme conditions. The calculated Womersley number of 0.128 is much less than the critical point of Wo = 156, validating our steady flow assumption.

Stenosis microfluidic control variables

To define the physical determinants of hemodynamic profiles, four studies were conducted with microfluidic control variables as defined in Table 3: Study (1) the bulk (wall) shear rate γ0 = 150–3000 s−1 by adjusting Q = 1.84–36.77 µL min−1 as calculated with Eq. (7) (Fig. 2); Study (2) the stenosis level S = 30—95% are assessed under γ0 = 1,000 s−1 (Fig. 3); Study (3) the contraction angle α = 30—85° are assessed under γ0 = 1,000 s−1 (Fig. 4); Study (4) benchmark the viscosity effect on water, blood as Newtonian fluid, and blood as non-Newtonian fluid (Fig. 5).

Table 3.

Control variable selection.

Control variables Test values Default values
Bulk shear rate γ0 (s−1) γ0 = 150, 600, 1000, 1500, 2000, 2500, 3000 s−1 α = 85°, S = 80%, Blood
Stenosis level S (%) S = 30, 40, 50, 60, 70, 80, 90, 95% γ0 = 1,000 s−1, α = 85°, Blood
Contraction angle α (°) α = 30, 45, 60, 75, 85° γ0 = 1,000 s−1, S = 80%, Blood
Fluid medium Water; Blood; non-Newtonian Blood; γ0 = 50–1,050 s−1 S = 80%, α = 85°

Figure 2.

Figure 2

CFD simulated shear rate, WSS and shear rate gradient plot for eccentric (AD) and concentric (EH) stenosis microfluidics under various bulk shear rates. (A and E) The cross-section WSS view of the eccentric (A) and concentric (E) stenoses microfluidic for γ0 = 2,000 s−1. Note that the WSS occurs at the middle of stenotic region. (B and F) The streamline of blood flow in the eccentric and concentric stenosis microfluidic channel for γ0 = 2,000 s−1. The representative streamline of a platelet trajectory was colored by shear rate values. Note that the peak shear rate γmax occurs at the stenosis apex. (C and G) The peak shear rate γmax and peak WSS τmax linearly correlated with the input bulk shear rate γ0 for both eccentric and concentric stenosis microfluidics respectively. (D and H) The shear rate and shear rate gradient history of a single platelet particle trajectory 30 μm from the bottom wall and 1 μm from the symmetric geometry surface at γ0 = 2000 s−1 for eccentric and concentric stenosis microfluidic channels. Note that the γmax ≈50,000 s−1 and γmax ≈5,000 µm−1 s−1 for eccentric and concentric stenosis respectively.

Figure 3.

Figure 3

CFD simulated shear rates and shear rate gradients in eccentric (AD) and concentric (EH) stenoses geometry under various stenosis level. (A and E) The streamlines of blood flow in the eccentric and concentric stenosis microfluidic channels for S = 70% (top) and 90% (bottom) respectively. The representative streamline of a platelet trajectory was colored by shear rate values. (B and F) Peak shear rate γmax exponentially correlates with the input stenosis level S for both eccentric and concentric stenoses respectively. Note that the concentric γmax is higher than that of eccentric stenosis channel. The shear rate γ (C and G) and shear rate gradient γ (D and H) distribution is plotted along a sample streamline 1 µm above the stenosis apex spanning the shear acceleration (x = −100 to 0 µm) and deceleration (x = 0–100 µm) zones. Note that γmax and γmax occur at the same location as in Fig. 2.

Figure 4.

Figure 4

CFD simulated shear rates and shear rate gradients in eccentric (AD) and concentric (EH) stenoses geometry under various contraction angles. (A and E) The streamlines of blood flow in the eccentric stenosis and concentric stenosis microfluidic channel for α = 60° (top) and 30° (bottom) respectively. The representative streamline of a platelet trajectory was colored by shear rate values. (B and F) Peak shear rate γmax slightly changes correlated to the contraction angle for both eccentric and concentric stenoses geometries. Note that the eccentric stenosis upstream shear rate is higher than that of the concentric stenosis channel. The shear rate γ (C and G) and shear rate gradient γ (D and H) distribution is plotted along a sample streamline 1 µm above stenosis apex spanning the shear acceleration (x = −100 to 0 µm) and deceleration (x = 0–100 µm) zones. Note that the shear rates in the post-stenosis zone remain unchanged with respect to α, while the upstream of the pre-stenosis area displays faster acceleration.

Figure 5.

Figure 5

Benchmarks of viscosity models and flow medium effects. (A) Relationship between bulk shear rate γ0 and viscosity for the different viscosity models. Note that the viscosity for most models first decreases then approaches to constant 0.00345 Pa s at γ0 > 500 s−1. (B) Relationship between peak shear rate and viscosity for water (blue), Newtonian blood (red) and non-Newtonian blood (green). The peak shear rates γmax are indifferent for all cases.

By default, the shear rate and velocity profile are simulated with a streamline sampled as the single platelet trajectory passing 1 µm (half of the diameter of a single platelet) above the stenosis apex, and z = 30 µm from the bottom (Fig. 1B).

In addition, we set up an ultra-fine CFD mesh with approximately 2,000,000 elements (Fig. 1D and Supplementary Fig. S1) for all the studies to achieve high accuracy.

Results

Bulk shear rate impact on WSS and shear rate distribution at stenosis

It has been well recognized that platelets activate and aggregate in response to flow disturbance5759. In microcirculation, blood flow is considered laminar and steady as the change in geometry does not transfer the flow into a turbulent regime. Here we investigated a wide range of γ0 = 150–3,000 s−1 and the effects on the τ and γ distribution using our refined CFD method. Although τmax is similar on both eccentric (Fig. 2A) and concentric (Fig. 2E) stenosis microfluidic wall with 7% difference, a higher τ distribution is observed at the ceiling of the eccentric stenosis (compare Fig. 2A vs. E), which could lead to higher probabilities of vessel damage48.

Then we compared a sample platelet trajectory streamline across the stenotic region (Fig. 2B,F), displaying 10% higher γmax in the eccentric stenosis than the concentric stenosis. Interestingly, as γ0 increases from 150 to 3,000 s−1, the τmax and γmax both increased linearly with respect to γ0 about 20.0- and 19.4-fold respectively for eccentric stenosis (Fig. 2C). Similarly, the τmax and γmax increased by 16.7- and 20.0-fold respectively for concentric stenosis (Fig. 2G). Notably, we found eccentric stenosis has larger γmax and γmax than concentric stenosis (compare γmax = 53,875 s−1, γmax = 25,701 µm−1 s−1 in Fig. 2D vs. γmax = 48,691 s−1, γmax = 23,295 µm−1 s−1 in Fig. 2H). Of note, our simulated τmax displays strong correlation to γmax. Thus, we focused on the shear rate γ as the examining parameter in the following studies.

Relation between micro-contraction geometry and hemodynamic profile at stenosis

Flow disturbance caused by vessel lumen constriction has a major prothrombotic effect via platelet mechanosensing7. Hereby, we examined γ and γ distributions as functions of S for degree (Fig. 3) and α for topology (Fig. 4) of vessel narrowing. Notably, we observed that S plays a significant role in the shear rate acceleration towards γmax (Fig. 3A,E). The γmax for eccentric and concentric geometries increased 34.6- and 41.0-fold respectively as S increased from 30 to 95% (Fig. 3C,G). Similarly, the γmax for both geometries increased 28.3- and 35.7-fold respectively (Fig. 3D,E). Notably, we identified an exponential γmax increment once the stenosis level exceeded 70% (Fig. 3B,F). Trajectory analyses demonstrated similar increases in γ and γ in the S > 70% regimes for both concentric (Fig. 3C,D) and eccentric (Fig. 3G,H) stenoses.

In contrast, α did not significantly affect the hemodynamics within the stenotic region, and subsequently had little effect on the γ and γ distribution (Fig. 4B,F). It is worth noting that with increasing α, the upstream face or pre-stenosis area displayed rapid acceleration in γ (Fig. 4C,G) and γ (Fig. 4D,H), whilst the post-stenosis downstream area was not affected. Our findings demonstrated that the shear rate profile changes most significantly when S is varied, suggesting that (1) although the development of vessel narrowing might be negligible at earlier stages of development, its hemodynamic impacts are rapidly enhanced once severe pro-occlusive conditions are reached; (2) the impact of the stenosis contraction angle is negligible due to its limited influence on γmax, however, its significant impact on the shear acceleration zone should be further investigated, especially its elongational effects acting on blood cells and plasma proteins4,9.

Hemodynamic benchmarks of viscosity models

The selection criteria for viscosity models on microfluidic simulation remain incompletely standardized. For example, it is fundamental to understand whether the hemodynamic forces experienced by washed platelets in Tyrode’s buffer is the same as those in whole blood41,60. Accurate computational modelling of blood is a complicated task due to multiplexed components in whole blood. RBCs consist a large amount, 45% of whole blood by volume or hematocrit61. Considering that RBCs have viscoelastic properties, blood exhibits shear-thinning behavior (decreasing viscosity with increasing shear rates), or non-Newtonian feature, at low shear rate γ < 1000 s−1 (Fig. 5A)62. While at high shear rates γ > 1000 s−1, blood behaves as a Newtonian fluid with a linear relationship between γ and τ, in another words nearly constant viscosity (Fig. 5A)62,63.

To benchmark the viscosity model choice, we mapped the viscosity changes with respect to γ0 by applying five different constitutive viscosity models: one Newtonian and four non-Newtonian62, where a wide range of bulk shear rates γ0 = 50–1050 s−1 were examined. All the non-Newtonian models have similar viscosity predictions at low γ0, while the viscosity of the PL model becomes inaccurate at high γ0 region. Besides, we also found negligible difference in γmax for water, blood as a Newtonian fluid, and blood as a non-Newtonian fluid (Fig. 5B). Interestingly, qualitative similarities between water and blood suggest that blood can be considered Newtonian in CFD analyses of shear rate distribution, even in low γ0 conditions (Fig. 5B).

This finding supports the feasibility of CFD practices with the water Newtonian model for microfluidic characterization as a reductionist approach with much reduced computational costs. Having said that, the water Newtonian model does not fully recapitulate the behavior of whole blood61. Viscoelastic models, such as the simplified Phan-Thien-Tanner or Gisekus61, are good alternatives to capture the blood rheology in future studies.

Discussion

Our numerical studies present a refined CFD approach to map hemodynamic parameters for both concentric and eccentric stenosis microfluidic models of thrombosis, providing unprecedented rheological insights underlying biomechanical platelet aggregation and thrombosis. The results present that (1) the stenosis level S is the major determinant of the shear rate γ and shear rate gradient γ within disturbed flow at the micro-contraction; (2) the contraction angle α plays a significant role in governing γ, while having negligible influence on γmax; (3) the shear rate γ experienced by washed platelets in Tyrode’s buffer is similar to that in whole blood; (4) water as a Newtonian fluid can be applied to microfluidic CFD characterization as a reductionist approach with reduced computational costs.

Previous experimental studies have demonstrated that in vivo aggregation is sensitive to γ at stenosis7. The follow-up studies suggested that γ0, S and α affect platelet aggregation in terms of kinetics, stability and sizes of aggregates38. The underlying platelet mechanobiology at molecular and cellular levels is exciting yet incompletely understood5. One possibility is that a γ threshold exists for VWF elongation, and its subsequent conformational activation6,13,60,64. The other working mechanism is due to the collision and compression between RBCs and platelets at the stenosis and surface of developing aggregates16,41. To investigate these mechanobiological mechanisms, future experiments are required to correlate our CFD results with molecular and cellular behaviors of biomechanical platelet aggregation in these stenotic microfluidic devices13.

Supplementary Information

Supplementary Information (475.2KB, docx)

Acknowledgements

We thank Qing Li, Chi Wu, and Jiaqiu Wang for help with the discussion, Xuanyi Chen for help with figure improvement and Sydney Manufacturing Hub, Gregg Suaning, Kevin Doan and Mark Richardson for support of our lab startup. We also thank Jacky He and Nadia Court for microfluidic design and assistance in soft lithography. This work was conducted (in part) using the Research Prototype Foundry (RPF) core research facilities at the University of Sydney (USYD) node of the Australian National Fabrication Facility (ANFF-USYD), a company established under the National Collaborative Research Infrastructure Strategy to provide nano- and micro-fabrication facilities for Australia’s researchers. This work was supported by the Australian Research Council (ARC) (DP200101970 – L.A.J.); the National Health and Medical Research Council (NHMRC) of Australia (APP2003904 – L.A.J.); NSW Cardiovascular Capacity Building Program (Early-Mid Career Researcher Grant – L.A.J.); Sydney Research Accelerator prize (SOAR – L.A.J.); Ramaciotti Foundations (2020HIG76 – L.A.J.), and the USYD Core Research Facilities User Access Scheme (RPF – L.A.J., Y.C.Z., Y.Z.); Lining Arnold Ju is an ARC DECRA fellow (DE190100609).

Author contributions

Y.C.Z., P.V. and T.G. performed the research, analyzed the data and co-wrote the paper. Z.L., R.M., K.K. and Y.Z. provided critical advice. L.A.J. supervised the study, designed the research and wrote the paper. Research activities related to this work complied with relevant ethical regulations at the Faculty of Engineering, The University of Sydney.

Data availability

The data that supports the findings of this study are available from the corresponding author upon reasonable request.

Competing interests

The authors declare no competing interests.

Footnotes

Publisher's note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

These authors contributed equally: Yunduo Charles Zhao and Parham Vatankhah.

Supplementary Information

The online version contains supplementary material available at 10.1038/s41598-021-86310-2.

References

  • 1.Virani SS, et al. Heart disease and stroke statistics-2020 update: a report from the american heart association. Circulation. 2020;141:e139–e596. doi: 10.1161/CIR.0000000000000757. [DOI] [PubMed] [Google Scholar]
  • 2.Jackson SP. Arterial thrombosis–insidious, unpredictable and deadly. Nat. Med. 2011;17:1423–1436. doi: 10.1038/nm.2515. [DOI] [PubMed] [Google Scholar]
  • 3.Jackson SP, Nesbitt WS, Westein E. Dynamics of platelet thrombus formation. J. Thromb. Haemost. 2009;7(Suppl 1):17–20. doi: 10.1111/j.1538-7836.2009.03401.x. [DOI] [PubMed] [Google Scholar]
  • 4.Rana A, Westein E, Niego B, Hagemeyer CE. Shear-dependent platelet aggregation: mechanisms and therapeutic opportunities. Front. Cardiovasc. Med. 2019;6:141. doi: 10.3389/fcvm.2019.00141. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5.Chen Y, Ju LA. Biomechanical thrombosis: the dark side of force and dawn of mechano-medicine. Stroke Vasc. Neurol. 2020;5:185–197. doi: 10.1136/svn-2019-000302. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6.Brazilek RJ, et al. Application of a strain rate gradient microfluidic device to von Willebrand's disease screening. Lab. Chip. 2017;17:2595–2608. doi: 10.1039/c7lc00498b. [DOI] [PubMed] [Google Scholar]
  • 7.Nesbitt WS, et al. A shear gradient-dependent platelet aggregation mechanism drives thrombus formation. Nat. Med. 2009;15:665–673. doi: 10.1038/nm.1955. [DOI] [PubMed] [Google Scholar]
  • 8.Schneider SW, et al. Shear-induced unfolding triggers adhesion of von Willebrand factor fibers. Proc. Natl. Acad. Sci. USA. 2007;104:7899–7903. doi: 10.1073/pnas.0608422104. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9.Fu H, et al. Flow-induced elongation of von Willebrand factor precedes tension-dependent activation. Nat. Commun. 2017;8:324. doi: 10.1038/s41467-017-00230-2. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10.Ju L, Lou J, Chen Y, Li Z, Zhu C. Force-induced unfolding of leucine-rich repeats of glycoprotein ibalpha strengthens ligand interaction. Biophys. J. 2015;109:1781–1784. doi: 10.1016/j.bpj.2015.08.050. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11.Zhang W, et al. Identification of a juxtamembrane mechanosensitive domain in the platelet mechanosensor glycoprotein Ib-IX complex. Blood. 2015;125:562–569. doi: 10.1182/blood-2014-07-589507. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12.Ju L, Chen Y, Xue L, Du X, Zhu C. Cooperative unfolding of distinctive mechanoreceptor domains transduces force into signals. Elife. 2016;5:1. doi: 10.7554/eLife.15447. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13.Chen Y, et al. An integrin alphaIIbbeta3 intermediate affinity state mediates biomechanical platelet aggregation. Nat. Mater. 2019;18:760–769. doi: 10.1038/s41563-019-0323-6. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 14.Ting LH, et al. Contractile forces in platelet aggregates under microfluidic shear gradients reflect platelet inhibition and bleeding risk. Nat. Commun. 2019;10:1204. doi: 10.1038/s41467-019-09150-9. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15.Tovar-Lopez FJ, et al. Structural and hydrodynamic simulation of an acute stenosis-dependent thrombosis model in mice. J. Biomech. 2011;44:1031–1039. doi: 10.1016/j.jbiomech.2011.02.006. [DOI] [PubMed] [Google Scholar]
  • 16.Tovar-Lopez FJ, et al. An investigation on platelet transport during thrombus formation at micro-scale stenosis. PLoS ONE. 2013;8:e74123. doi: 10.1371/journal.pone.0074123. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 17.Hong JK, et al. Evaluating medical device and material thrombosis under flow: current and emerging technologies. Biomater. Sci. 2020;8:5824–5845. doi: 10.1039/D0BM01284J. [DOI] [PubMed] [Google Scholar]
  • 18.Kroll MH, Hellums JD, McIntire LV, Schafer AI, Moake JL. Platelets and shear stress. Blood. 1996;88:1525–1541. doi: 10.1182/blood.V88.5.1525.1525. [DOI] [PubMed] [Google Scholar]
  • 19.Farag MB, et al. Review of recent results using computational fluid dynamics simulations in patients receiving mechanical assist devices for end-stage heart failure. Methodist. Debakey Cardiovasc. J. 2014;10:185–189. doi: 10.14797/mdcj-10-3-185. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 20.Nascimbene A, Neelamegham S, Frazier OH, Moake JL, Deong J. Acquired von Willebrand syndrome associated with left ventricular assist device. Blood. 2016;127:3133–3141. doi: 10.1182/blood-2015-10-636480. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 21.Ballermann BJ, Dardik A, Eng E, Liu A. Shear stress and the endothelium. Kidney Int. 1998;54:S100–S108. doi: 10.1046/j.1523-1755.1998.06720.x. [DOI] [PubMed] [Google Scholar]
  • 22.Papaioannou TG, Stefanadis C. Vascular wall shear stress: basic principles and methods. Hellen. J. Cardiol. 2005;46:9–15. [PubMed] [Google Scholar]
  • 23.Dhawan SS, et al. Shear stress and plaque development. Expert Rev. Cardiovasc. Ther. 2010;8:545–556. doi: 10.1586/erc.10.28. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 24.Xing R, et al. Temporal and spatial changes in wall shear stress during atherosclerotic plaque progression in mice. R. Soc. Open Sci. 2018;5:171447. doi: 10.1098/rsos.171447. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 25.Kim D, Bresette C, Liu Z, Ku DN. Occlusive thrombosis in arteries. APL. Bioengineering. 2019;3:041502. doi: 10.1063/1.5115554. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 26.Mongrain R, Rodés-Cabau J. Role of shear stress in atherosclerosis and restenosis after coronary stent implantation. Rev. Esp. Cardiol. 2006;59:1–4. doi: 10.1157/13083641. [DOI] [PubMed] [Google Scholar]
  • 27.Foin N, et al. Incomplete stent apposition causes high shear flow disturbances and delay in neointimal coverage as a function of strut to wall detachment distance. Circ. Cardiovasc. Interven. 2014;7:180–189. doi: 10.1161/CIRCINTERVENTIONS.113.000931. [DOI] [PubMed] [Google Scholar]
  • 28.Piper R, et al. The mechanistic causes of peripheral intravenous catheter failure based on a parametric computational study. Sci. Rep. 2018;8:3441. doi: 10.1038/s41598-018-21617-1. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 29.Lucas TC, et al. Blood flow in hemodialysis catheters: a numerical simulation and microscopic analysis of in vivo-formed fibrin. Artif. Organs. 2013;38:556–565. doi: 10.1111/aor.12243. [DOI] [PubMed] [Google Scholar]
  • 30.Yang S, et al. Computational simulation of postoperative pulmonary flow distribution in Alagille patients with peripheral pulmonary artery stenosis. Congenit. Heart Dis. 2017;13:241–250. doi: 10.1111/chd.12556. [DOI] [PubMed] [Google Scholar]
  • 31.Jin C, Liu Y. Influence of competitive flow caused by different stenosis on coronary artery bypass hemodynamics and PIV study. Mol. Cell. Biomech. 2019;16:51–52. doi: 10.32604/mcb.2019.05728. [DOI] [Google Scholar]
  • 32.Zhao Z, Mao B, Liu Y, Yang H, Chen Y. The study of the graft hemodynamics with different instant patency in coronary artery bypassing grafting. Comput. Model. Eng. Sci. 2018;116:229–245. [Google Scholar]
  • 33.Zhu C, Chen Y, Ju LA. Dynamic bonds and their roles in mechanosensing. Curr. Opin. Chem. Biol. 2019;53:88–97. doi: 10.1016/j.cbpa.2019.08.005. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 34.Zhao YC, et al. Computational fluid dynamics simulations at micro-scale stenosis for microfluidic thrombosis model characterization. Mol. Cell. Biomech. 2021;18:1–10. doi: 10.1016/j.cbpa.2019.08.005. [DOI] [Google Scholar]
  • 35.Zhou F, Chen Y, Felner EI, Zhu C, Lu H. Microfluidic auto-alignment of protein patterns for dissecting multi-receptor crosstalk in platelets. Lab. Chip. 2018;18:2966–2974. doi: 10.1039/c8lc00464a. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 36.Colace TV, Tormoen GW, McCarty OJ, Diamond SL. Microfluidics and coagulation biology. Ann. Rev. Biomed. Eng. 2013;15:283–303. doi: 10.1146/annurev-bioeng-071812-152406. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 37.Hansen CE, Lam WA. Clinical implications of single-cell microfluidic devices for hematological disorders. Anal. Chem. 2017;89:11881–11892. doi: 10.1021/acs.analchem.7b01013. [DOI] [PubMed] [Google Scholar]
  • 38.Tovar-Lopez FJ, et al. A microfluidics device to monitor platelet aggregation dynamics in response to strain rate micro-gradients in flowing blood. Lab. Chip. 2010;10:291–302. doi: 10.1039/b916757a. [DOI] [PubMed] [Google Scholar]
  • 39.Sundd P, et al. ‘Slings’ enable neutrophil rolling at high shear. Nature. 2012;488:399–403. doi: 10.1038/nature11248. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 40.Tsai M, et al. In vitro modeling of the microvascular occlusion and thrombosis that occur in hematologic diseases using microfluidic technology. J. Clin. Invest. 2012;122:408–418. doi: 10.1172/JCI58753. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 41.Ju L, et al. Compression force sensing regulates integrin alphaIIbbeta3 adhesive function on diabetic platelets. Nat. Commun. 2018;9:1087. doi: 10.1038/s41467-018-03430-6. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 42.Cho YI, Kensey KR. Effects of the non-Newtonian viscosity of blood on flows in a diseased arterial vessel. Part 1: Steady flows. Biorheology. 1991;28:241–262. doi: 10.3233/BIR-1991-283-415. [DOI] [PubMed] [Google Scholar]
  • 43.Abraham F, Behr M, Heinkenschloss M. Shape optimization in steady blood flow: a numerical study of non-Newtonian effects. Comput. Methods Biomech. Biomed. Engin. 2005;8:127–137. doi: 10.1080/10255840500180799. [DOI] [PubMed] [Google Scholar]
  • 44.Sharma K, Bhat S. Non-Newtonian rheology of leukemic blood and plasma: are n and k parameters of power law model diagnostic? Physiol. Chem. Phys. Med. NMR. 1992;24:307–312. [PubMed] [Google Scholar]
  • 45.Ballyk P, Steinman D, Ethier C. Simulation of non-Newtonian blood flow in an end-to-side anastomosis. Biorheology. 1994;31:565–586. doi: 10.3233/BIR-1994-31505. [DOI] [PubMed] [Google Scholar]
  • 46.Fung Y-C. Biomechanics: mechanical properties of living tissues. Springer; 2013. [Google Scholar]
  • 47.Walburn FJ, Schneck DJ. A constitutive equation for whole human blood. Biorheology. 1976;13:201–210. doi: 10.3233/BIR-1976-13307. [DOI] [PubMed] [Google Scholar]
  • 48.Soulis JV, et al. Non-Newtonian models for molecular viscosity and wall shear stress in a 3D reconstructed human left coronary artery. Med. Eng. Phys. 2008;30:9–19. doi: 10.1016/j.medengphy.2007.02.001. [DOI] [PubMed] [Google Scholar]
  • 49.Miller C. Predicting non-Newtonian flow behavior in ducts of unusual cross section. Ind. Eng. Chem. Fundam. 1972;11:524–528. doi: 10.1021/i160044a015. [DOI] [Google Scholar]
  • 50.Lancellotti RM, Vergara C, Valdettaro L, Bose S, Quarteroni A. Large eddy simulations for blood dynamics in realistic stenotic carotids. Int. J. Numer. Method Biomed. Eng. 2017;33:1. doi: 10.1002/cnm.2868. [DOI] [PubMed] [Google Scholar]
  • 51.Lee SE, Lee SW, Fischer PF, Bassiouny HS, Loth F. Direct numerical simulation of transitional flow in a stenosed carotid bifurcation. J. Biomech. 2008;41:2551–2561. doi: 10.1016/j.jbiomech.2008.03.038. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 52.Stroud JS, Berger SA, Saloner D. Numerical analysis of flow through a severely stenotic carotid artery bifurcation. J. Biomech. Eng. 2002;124:9–20. doi: 10.1115/1.1427042. [DOI] [PubMed] [Google Scholar]
  • 53.Fox RW, McDonald AT, Mitchell JW. Fox and McDonald's introduction to fluid mechanics. Wiley; 2020. [Google Scholar]
  • 54.Vahidkhah K, Balogh P, Bagchi P. Flow of red blood cells in stenosed microvessels. Sci. Rep. 2016;6:28194. doi: 10.1038/srep28194. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 55.Fedosov DA, Caswell B, Popel AS, Karniadakis GE. Blood flow and cell-free layer in microvessels. Microcirculation. 2010;17:615–628. doi: 10.1111/j.1549-8719.2010.00056.x. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 56.Fung Y-C. Biomechanics: motion, flow, stress, and growth. Springer; 2013. [Google Scholar]
  • 57.Wurzinger L, Opitz R, Wolf M, Schmid-Schönbein H. “Shear induced platelet activation”-A critical reappraisal. Biorheology. 1985;22:399–413. doi: 10.3233/BIR-1985-22504. [DOI] [PubMed] [Google Scholar]
  • 58.Hellums, J., Peterson, D., Stathopoulos, N., Moake, J. & Giorgio, T. in Cerebral ischemia and hemorheology 80–89 (Springer, 1987).
  • 59.O'Brien J. Shear-induced platelet aggregation. The Lancet. 1990;335:711–713. doi: 10.1016/0140-6736(90)90815-M. [DOI] [PubMed] [Google Scholar]
  • 60.Ju L, Dong JF, Cruz MA, Zhu C. The N-terminal flanking region of the A1 domain regulates the force-dependent binding of von Willebrand factor to platelet glycoprotein Ibalpha. J. Biol. Chem. 2013;288:32289–32301. doi: 10.1074/jbc.M113.504001. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 61.Campo-Deaño L, Dullens RP, Aarts DG, Pinho FT, Oliveira MS. Viscoelasticity of blood and viscoelastic blood analogues for use in polydymethylsiloxane in vitro models of the circulatory system. Biomicrofluidics. 2013;7:34102. doi: 10.1063/1.4804649. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 62.Mendieta JB, et al. The importance of blood rheology in patient-specific computational fluid dynamics simulation of stenotic carotid arteries. Biomech. Model. Mechanobiol. 2020 doi: 10.1007/s10237-019-01282-7. [DOI] [PubMed] [Google Scholar]
  • 63.Brust M, et al. Rheology of human blood plasma: viscoelastic versus Newtonian behavior. Phys. Rev. Lett. 2013;110:078305. doi: 10.1103/PhysRevLett.110.078305. [DOI] [PubMed] [Google Scholar]
  • 64.Momi S, et al. Reperfusion of cerebral artery thrombosis by the GPIb-VWF blockade with the Nanobody ALX-0081 reduces brain infarct size in guinea pigs. Blood. 2013;121:5088–5097. doi: 10.1182/blood-2012-11-464545. [DOI] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

Supplementary Information (475.2KB, docx)

Data Availability Statement

The data that supports the findings of this study are available from the corresponding author upon reasonable request.


Articles from Scientific Reports are provided here courtesy of Nature Publishing Group

RESOURCES