Skip to main content
NIHPA Author Manuscripts logoLink to NIHPA Author Manuscripts
. Author manuscript; available in PMC: 2022 Jul 1.
Published in final edited form as: Magn Reson Med. 2021 Feb 8;86(1):372–381. doi: 10.1002/mrm.28704

Simultaneous T1, T2, and T Relaxation Mapping of the Lower Leg Muscle with MR-Fingerprinting (MRF)

Azadeh Sharafi 1, Katherine Medina 1, Marcelo VW Zibetti 1, Smita Rao 4, Martijn A Cloos 5, Ryan Brown 1,2,3, Ravinder R Regatte 1,2,3
PMCID: PMC8005468  NIHMSID: NIHMS1662326  PMID: 33554369

Abstract

Purpose

To develop a novel MR-fingerprinting pulse sequence that is insensitive to B1+ and B0 imperfections for simultaneous T1, T2, and T relaxation mapping.

Methods

We implemented a totally-balanced-spin-lock (TB-SL) module to encode T relaxation into an existing MRF framework that encoded T1 and T2. The spin-lock module utilized two 180° pulses with compensatory phases to reduce T sensitivity to B1 and B0 inhomogeneities. We compared T measured using TB-SL MRF in Bloch simulations, model agar phantoms, and in-vivo experiments to those with a self-compensated spin-lock preparation module (SC-SL). TB-SL MRF repeatability was evaluated in maps acquired in the lower leg skeletal muscle of 12 diabetic peripheral neuropathy patients scanned two times each during visits separated by ~30 days.

Results

The phantom relaxation times measured with TB-SL and SC-SL MRF were in good agreement with reference values in regions with low B1 inhomogeneities. Compared to SC-SL, TB-SL MRF showed in experiments greater robustness against severe B1 inhomogeneities and in Bloch simulations greater robustness against B1 and B0. We measured with TB-SL MRF average of T1=950.1±28.7ms, T2=26.0±1.2ms, and T=31.7±3.2ms in skeletal muscle across patients. Bland-Altman analysis demonstrated low bias between TB-SL and SC-SL MRF and between TB-SL MRF maps acquired in two visits. The coefficient of variation was less than 3% for all measurements.

Conclusion

The proposed TB-SL MRF sequence is fast and insensitive to B1+ and B0 imperfections. It can simultaneously map T1, T2, T, and B1+ in a single scan and could potentially be used to study muscle composition.

Keywords: T, Diabetic Neuropathy, Magnetic Resonance Fingerprinting, Multiparametric mapping, Lower Leg Muscle

INTRODUCTION

Recent advancements in quantitative MRI have significantly expanded the potential to evaluate pathologic changes in muscle disorders, including composition, architecture, mechanical properties, perfusion, and function (15). Semi-quantitative methods can assess and grade the extension of compositional changes such as fatty degeneration (1). Moreover, has been shown that T1 relaxation significantly decreased with fat infiltration and increased with edema (1,6,7). T2 relaxation is sensitive to water binding (1, 810), enabling the detection of early pathologic changes in muscle integrity before deterioration (1). In subsequent pathologic stages, T2 increases with free water content, fat infiltration, and edema (11).

Moreover, a small number of studies showed that both T1 and T2 relaxations changed after exercise (9,12,13). In addition to T1 and T2, the spin-lattice relaxation time in the rotating frame, T, is shown to capture biochemical changes (1417). Virta et al. (15) showed that T is more sensitive to muscle composition than T1. The sensitivity of T to the change in muscle composition and accumulation of fibrotic tissue due to aging was shown by Wang et al. (16), while Lamminen et al. (17) showed that T relaxation time is significantly different between healthy and diseased muscle tissue.

Conventional quantitative MRI approaches can be time-consuming and generally provide only a single parameter at a time. Given the diverse physiologic information represented by T1, T2, and T, it is desirable to develop a technique to measure these parameters simultaneously. Magnetic resonance fingerprinting (MRF) is a newly developed method (18) to measure multiple MR parameters using dynamic signal patterns simultaneously.

MRF has demonstrated improved scan efficiency in several applications, including the knee (19), prostate (20), hip (21), abdomen (22,23), brain (24,25), liver (26), kidney (27), and heart (28). While most MRF implementations included the capability to measure only T1 and T2 (21), our group integrated an additional self-compensated spin-lock preparation module to encode T as well, which is particularly interesting in musculoskeletal applications. One drawback of this method was its sensitivity to B1+ and B0 imperfections that can confound T measurements. In this work, we implemented a totally balanced spin-lock module to improve robustness. To demonstrate proof-of-concept, we show results in phantom and in lower extremity skeletal muscle in a cohort of patients with diabetic peripheral neuropathy (DPN) - such patients are expected to have significant skeletal muscle deficits, including loss of strength, power, and endurance, and neurogenic muscle atrophy (2932).

METHODS

MRF-Sequence Design

Based on the method described in (21), we developed an MRF-sequence to estimate T1, T2, and T in less than 5 minutes. Similar to a previous design (21), the proposed MRF-sequence started with an adiabatic inversion pulse followed by two fast imaging with steady-state precession (FISP) segments to mainly encode T1/T2, and two fast low-angle shot (FLASH) segments, which encode T1 and B1+. Each segment contains 250 radiofrequency (RF) excitations, with a time-bandwidth product of three and a flip angle that varies from 0° to 60° in the FISP and from 0° to 20° in the FLASH segments, followed by a single radial readout whose rotation is incremented between excitations by the golden angle (33). The flip angles change in a sinusoidal fashion to smoothly vary the transient state of the magnetization. There was a delay equal to 50 repetition times (TRs) between segments to allow partial recovery of the magnetization and to enhance T1 encoding. Finally, spin-lock preparation modules with different spin-lock duration, including tsl = 2, 3.75, 7, 13, 24, 45ms, were added at the end of the train to encode T. Each of the preparation modules was followed by a FLASH segment with 125-RF excitations with a peak flip angle of 20°. The complete train of 1750 excitations constitutes one shot. We acquired four shots per slice, uniformly distributed within k-space (21).

The T preparation module starts with a 90° RF pulse along the x-axis (90x) to flip the magnetization to the transverse plane. Afterward, a continuous spin-lock pulse was applied along the y-axis (αy) to lock the transverse magnetization. The magnetization will decay during the lock time (Tsl) according to T relaxation time. Applying a second tip-up 90° RF pulse in the opposite direction of the first (90-x) will tip the magnetization back to the longitudinal direction. Crusher gradients were then applied to destroy the remaining magnetization in the transverse plane before readout. In practice, the actual spin-lock field strength and direction is affected by field inhomogeneities dictated by factors such as the excitation coil and tissue properties (34). In our application to study muscle composition, we used an in-house built 8-channel lower extremity coil (35) comprised of degenerate mode transmit/receive birdcages, which have characteristic inhomogeneous B1+ in the periphery that can affect T encoding without proper compensation.

The B1+ inhomogeneity causes a deviation of the magnetization from the transverse plane after applying the tip-down 90° pulse. As a result, the magnetization rotates with an angle (α) around the spin-lock RF pulse during Tsl, and the final longitudinal magnetizations become a complicated function of Tsl:

M(Tsl)=M0[sin(θ)eTslT1ρ+cos(α)cos(θ)eTslT2ρ] [1]

Where θ is the actual flip angle of the tip-up and tip-down RF pulses, α is the flip angle of the spin-lock pulse, and T is the relaxation time in the plane orthogonal to the spin-lock pulse. The cos(α) term causes the spatial signal modulation, which appears as banding artifacts in the acquired scans (36). Several papers (37,38) showed that the net α angle becomes zero using the rotary spin-lock pulse, and the effect of B1+ inhomogeneity will be eliminated. The B0 inhomogeneity can be compensated by inserting a 180° pulse (39). Mitrea et al. (37) further improved this approach by replacing the two spin-lock pulses with rotary self-compensated pairs. However, our experiments showed that this technique is still sensitive to the severe B1+ field inhomogeneities, such as found in our application. Hence, in this work, we applied a totally balanced T preparation module (TB-SL), in which the single 180° pulse was replaced with two rotary 180° pulses. Hence, every pulse is compensated with a correspondent pulse of the opposite phase (40). We compared our proposed MRF sequence with prior work described in (19), which uses Mitrea’s (37) self-compensated spin-lock preparation modules (SC-SL).

Model Phantoms Study

A model phantom consisting of 3%, 4%, and 8% agarose gels were scanned on a 3T MRI scanner (MAGNETOM Prisma, Siemens Healthcare GmbH, Germany) with the in-house built 8-channel lower extremity coil using both TB-SL and SC-SL MRF sequences. The imaging parameters were FOV=140×140mm2, 0.7×0.7mm2 in-plane resolution, 4.0mm slice thickness, TE/TR=3.5/7.5ms, BW=420Hz/pixel, number of slices= 4, number of shots= 4, spin-lock power fsl= 500Hz. The acquisition time was 4:44 min. The slices were acquired in a combination of sequential/interleaved fashion where two slices were interleaved at a time. There was no delay between each shot. The signal from each shot was averaged together to improve the signal to noise ratio. Moreover, the radial angle changes for each shot to reduce the undersampling artifacts. The model agar phantoms were also scanned with the MRF sequence proposed by Cloos et al. (21) for T1 and T2 comparisons and a customized turbo FLASH sequence (41) for T comparison.

In vivo Study

Our Institutional Review Board approved this Health Insurance Portability and Accountability Act study, and 12 participants diagnosed with DPN were scanned after obtaining their informed consent. The patients (age =57.5±6.4 years, 10 male, 2 female) were scanned two times each, with 31.4±10.4 days (range=21 to 60 days) between scans. Three healthy volunteers (age=30.6± 6.3 years, 1 male, 2 female) were also scanned with TB-SL and SC-SL MRF sequences in one session for comparison. The foot was enclosed in an MR compatible ergometer (56), and extensive foam padding between the calf and coil was used to reduce the motion. Four axial MRF images of the calf muscle were acquired with FOV=140×140mm2, 0.6×0.6mm2 in-plane resolution, 4.0 mm slice thickness, 224×224 matrix size, TE/TR=3.5/7.5ms, BW=420 Hz/pixel, number of shots=4, spin-lock power fsl=500Hz. The acquisition time was 4:45min.

MRF-Dictionary Construction for Multiparametric Mapping

All algorithms were implemented in MATLAB (R2019a, The MathWorks Inc., Natick, MA, USA). A dictionary of simulated MR fingerprints for possible T1, T2, T, and B1+ values was computed based on the extended phase graphs (42). The longitudinal magnetization following the T preparation module was simulated to undergo mono-exponential decay since the crusher gradients after the tip-up 90° pulse were assumed to destroy any remnant transverse magnetizations (19). Moreover, the effects of field inhomogeneities (B0 and B1) were not simulated in the dictionary.

The simulation was performed for T1:50–3000 ms, T2: 2–200ms, and T 2–200ms. The maximum excitation flip angle varied between 30° and 90° with 5% increments to simulate ±50% B1+ variations (target flip angle 60°). Afterward, the dictionary and the measured fingerprint signal were compressed using the singular value decomposition (SVD) method (22,43) and matched with each other using an iterative approach (43) to create a PD image and T1, T2, T, and B1+ maps, simultaneously.

Statistical Analysis

The mean and standard deviation for T1, T2, and T relaxation times were measured in agarose gel model phantoms and in the lower leg of DPN patients, in which we manually segmented the tibialis anterior (TA), soleus (SOL), gastrocnemius medialis (GM), and gastrocnemius lateralis (GL) muscles.

The Bland-Altman analysis was performed to assess the repeatability in the model phantoms and DPN patients. Moreover, the mean and standard deviation (SD) of each parameter was calculated for each subject across the test-retest scan, and the coefficient of variation (CV) was calculated as CV = SD/Mean to assess intra-subject repeatability.

The inter-subject repeatability was evaluated using the root mean square CV:

rmsCV=i=1nCVi2n [2]

Where CVi is the CV of an individual subject, and n (=12) is the total number of subjects.

RESULTS

The simulation results based on Bloch equations (Supplementary Figure 1) showed both TB-SL and SC-SL preparation modules provide high T accuracy when B1 and B0 are well matched to the expected values. However, TB-SL provides greater T accuracy than SC-SL when B1 is lower than expected, and B0 is offset from the center frequency. This is in agreement with our experimental results, in which the TB-SL module provided reliable T in peripheral regions in model agar phantoms and in-vivo in which B1+ was lower than expected.

Figure 1 shows improved B1+ robustness of the TB-SL spin-lock module compared to the SC-SL module in model agar gel phantoms. Specifically, T maps acquired with the TB-SL MRF sequence are largely immune to B1+ heterogeneity in the peripheries, whereas artifacts are present in maps acquired with the SC-SL MRF sequence.

Figure 1.

Figure 1.

Comparison of the SC-SL (top row) and TB-SL (bottom row) preparation modules in phantoms consisting of 3%, 4%, and 8% agarose gel. As shown by arrows, the SC-SL T map (fourth column) contains artifacts due to B1+ inhomogeneities (last column), while the artifacts are significantly reduced with the TB-SL module.

An average (Mean ± SD) of T1, T2, and T were measured in 3%, 4%, and 8% agar phantoms using the proposed TB-SL MRF technique. As shown in Table 1, the relaxation times are inversely related to agar concentration. The same trend was observed with SC-SL MRF and the reference techniques. Both TB-SL and SC-SL MRF techniques underestimated the T1 values compared to the reference values, which could be due to neglecting the magnetization transfer effect in the dictionary (48).

Table 1.

Phantom relaxation times measurements (Mean ± SD) using MRF sequence with TB and SC-SL preparation modules and comparison with conventional T1 (IR-SE), T2(SE), and T (T1ρTFL(65)) measurements.

MRF-TBSL MRF-SCSL Conventional
T1 (ms) T2(ms) T(ms) T1 (ms) T2(ms) T(ms) T1 (ms) T2(ms) T(ms)
AGAR 3% 1516.6 ± 164.6 42.0 ± 11.7 47.8 ± 10.2 1429.5 ± 141.2 46.1 ± 12.1 40.0 ± 12.1 1683.2 ± 54.1 46.1 ± 3.8 45.7 ± 1.4
AGAR 4% 1002.0 ± 125.8 26.6 ± 2.1 25.5 ± 2.8 977.8 ± 115.3 28.3 ± 2.6 24.5 ± 5.5 1656.3 ± 233.5 28.0 ± 2.7 23.2 ± 0.6
AGAR 8% 766.3 ± 44 15.9 ± 4.1 14.2 ± 2.6 746.1 ± 44.5 17.0 ± 4.0 11.1 ± 3.5 987.6 ± 64.2 17.8 ± 2.0 17.1 ± 0.4

Figure 2 shows the comparison between SC-SL and TB-SL modules in the in-vivo experiment and representative examples of the acquired signals and their respective dictionary matches in positions with low and high (periphery) B1+ inhomogeneities. In sequence with the SC-SL preparation module, the higher B1+ inhomogeneities caused a faster drop in the signal, which leads to a severe underestimation of T. We measured an average of T1=909.9±43.4, and 898.4±37.9ms, T2 =22.1±1.6, and 21.7±1.7ms, and T = 27.11±3.1, and 26.1±2.4ms across three healthy volunteers using TB-SL and SC-SL technique, respectively.

Figure 2.

Figure 2.

In-vivo comparison of the (a) SC-SL and (b) TB-SL preparation modules in the presence of B1+inhomogeneities. As shown by arrows, the T is affected using the SC-SL module, while the TB-SL-MRF sequence is robust against the inhomogeneities. (c, d) a representative example of the acquired signals using two sequences and their respective dictionary match in a position with (c) low and (d) high B1+ inhomogeneities. In sequence with the SC-SL preparation module, the higher inhomogeneities caused a faster drop in the signal, which leads to a smaller T value.

The regression and Bland-Altman plots (Figure 3) show excellent agreement between T1, T2, and T measured with SC-SL and TB-SL MRF in the part of the gastrocnemius medialis, gastrocnemius lateral, and soleus ROIs that were not affected by B1+ inhomogeneities. In the parts with high inhomogeneities, there is a drop in the estimated T1ρ, which leads to a lower overall mean value in the ROIs.

Figure 3.

Figure 3.

Regression (top row) and Bland-Altman plots (bottom row) to assess the differences in T1, T2, and T measured using SC-SL and the proposed TB-SL MRF sequence. There is a good agreement between the measured relaxation times from the two experiments.

Figure 4a demonstrates representative T1, T2, T, and B1+ maps in the lower leg of a DPN patient acquired in two separate visits. In gastrocnemius medialis, gastrocnemius lateral and soleus muscles, we measured an average of T1 = 950.1±28.7 ms, T2 =26.0±1.2 ms, and T = 31.7±3.2 across 12 DPN patients. Analysis of the Bland-Altman plots (Figure 4b) demonstrated an average difference of −6.40ms, −0.23ms, and 0.69ms between two visits for T1, T2, and T, respectively. Our in vivo study showed excellent repeatability (Figure 4c) with rmsCV smaller than 3% across all ROIs for T1, T2, and T.

Figure 4.

Figure 4.

(a) Representative PD, T1, T2, and T and B1+ maps of the calf muscle of a DPN patient acquired in two sessions one month apart. (B) Bland-Altman plots and (c) coefficient of variation in different muscle groups for T1, T2, and T measured during visits 1 (V1) and 2 (V2). The experiment showed low rmsCV (<3%) for all relaxation parameters in all ROIs.

DISCUSSION

In this work, we implemented an MRF sequence that showed the feasibility of rapid simultaneous acquisition of accurate PD image and T1, T2, B1+, and T maps of the skeletal muscle. The proposed TB-SL MRF method is fast, reproducible, and less sensitive to field inhomogeneities than SC-SL MRF. By integrating the T module into the MRF framework, we measured voxel-wise the three different relaxation times that may provide more comprehensive insight on skeletal muscle properties than T1 and T2 alone (22). The simultaneous acquisition eliminates potential misregistration due to motion between separate measurements, which can plague traditional standalone methods that may require approximately 45 minutes to separately acquire T1, T2, and T maps (41). In comparison, we used TB-SL MRF to measure T1, T2, and T in four slices in the lower extremity with 0.6×0.6×4.0 mm3 resolution in 4:45 min (1:11 min per slice).

Recently, Wyatt et al. (49) proposed a spiral MRF technique with variable flip angles and TRs for simultaneous mapping of T1, T2, and T. They used the Dixon module with a rotary spin-lock pulse, which is not robust to B0 inhomogeneities. Similar to our work, B1 variation was not considered in the T preparation module Bloch simulation. However, their use of rotary spin-lock and phase cycling made their sequence robust to B1 variations up to ±30% (49). We proposed using a balanced module with a rotary spin-lock pulse for B1+ compensation and 180° refocusing pulses to compensate for B0 inhomogeneities. Our in-vivo results showed the robustness of this technique. We were able to obtain good results, even in the sample periphery, where inhomogeneities of up to ±50% were observed. Several works proposed using adiabatic pulses for spin-locking due to their insensitivity to field inhomogeneities (5052). In contrast to conventional continuous pulses, the measured T value depends on the type of adiabatic pulse and its frequency modulation and amplitude. Moreover, T alters during the application of adiabatic spin-locking due to the frequency and amplitude modulation. Hence, the measured T is an average of its value over the adiabatic pulse duration (36). The SAR constraint also limited the use of adiabatic pulses. Recently, Ma et al. (53) proposed a 3D adiabatic T prepared ultrashort echo time technique for knee imaging. The proposed technique has lower power than continuous spin-lock pulses; however, the sequence was ~18 minutes long, limiting its use in clinical applications (53). Other studies (54,55) proposed using a constant amplitude adiabatic tip-down and tip-up pulses that simultaneously compensated for B1 and B0 inhomogeneities. In comparison with the hard pulse, the adiabatic pulse is more robust to the field inhomogeneities. However, since the spins are locked in a tilted angle instead of the transverse plane, the measured T value is different from the value measure on resonance, and the on resonance T must calculated retrospectively from the tilted angle value (54).

We used a radial readout. Compared to spiral readouts, radial trajectories are less sensitive to systematic imperfections, such as eddy current and B0 inhomogeneities (56). On the other hand, the radial pattern is less efficient in sampling the k-space and requires temporal averaging or multiple MRF train repetition (56). Block et al. (57) proposed an adaptive compensation of gradient delay where the shift in the k-space caused by the delay is estimated from a set of calibration spokes with opposite orientation. The shift is then adjusted in the k-space according to echo-pixel shifts in the gridding step (57,58). We have developed routines to calibrate out these delays in the past. However, the latest generation of scanners such as the one used here demonstrated negligible shifts for the dwell times used in this work.

Several limitations potentially affect our study. The acquisition was highly undersampled, and as a result, streaking artifacts are visible in some maps, such as T2 maps. Although the 2D TB-SL MRF total scan time is short (~5 minutes) in comparison to separate acquisition of T1, T2, and T maps, its lack of coverage (4 slices) makes volumetric measurement infeasible. One promising possibility to shorten scan time and improve coverage is to utilize a 3D-sequence with stack-of stars readout (5962). Moreover, online reconstruction is not possible due to the computationally expensive and memory inefficient dictionary matching method used in this work. We are currently optimizing the MRF sequence (e.g., flip angle train and optimal Tsl’s, etc.) based on Crammer-Rao Lower Bound (CRLB) (63) and deep learning-based reconstruction (64) instead of template matching to alleviate the problem.

Conclusion

In this work, we implemented an MRF-sequence and showed the feasibility of rapid simultaneous acquisition of accurate PD image and T1, T2, and T maps of the lower leg muscle, yielding valuable information about musculoskeletal complications. The proposed MRF method is fast, robust to B0 and B1 inhomogeneity, and repeatable, making it a candidate to assess muscle composition in pathology.

Supplementary Material

FIG S1

Supplementary Figure 1. The error of estimating T from Bloch simulation for SC-SL (a,c) and (b,d) TB-SL module in the presence of high and low field inhomogeneities. The estimation error is smaller using the TB-SL module.

Acknowledgment

The authors thank Ding Xia, M.Sc., for their assistance in data analysis.

This study was supported by NIH grants R01 DK114428, R01 DK106292, 1R01AR076328-01A1, and R01-AR068966 and was performed under the rubric of the Center of Advanced Imaging Innovation and Research (CAI2R) and NIBIB Biomedical Technology Resource Center (NIH P41 EB017183).

REFERENCES

  • 1.Paoletti M, Pichiecchio A, Piccinelli SC, et al. Advances in quantitative imaging of genetic and acquired myopathies: Clinical applications and perspectives. Front. Neurol. 2019;10:78 doi: 10.3389/fneur.2019.00078. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 2.Kumbhare DA. Advanced Skeletal Muscle MR Imaging Approaches in the Assessment of Muscular Dystrophies. Int. J. Phys. Med. Rehabil. 2014;02:1–11 doi: 10.4172/2329-9096.1000248. [DOI] [Google Scholar]
  • 3.Kim HK, Lindquist DM, Serai SD, et al. Magnetic resonance imaging of pediatric muscular disorders. Recent advances and clinical applications. Radiol. Clin. North Am. 2013;51:721–742 doi: 10.1016/j.rcl.2013.03.002. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4.Kalia V, Leung DG, Sneag DB, Del Grande F, Carrino JA. Advanced MRI Techniques for Muscle Imaging. Semin. Musculoskelet. Radiol. 2017;21:459–469 doi: 10.1055/s-0037-1604007. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5.Carlier PG, Marty B, Scheidegger O, et al. Skeletal Muscle Quantitative Nuclear Magnetic Resonance Imaging and Spectroscopy as an Outcome Measure for Clinical Trials. J. Neuromuscul. Dis. 2016;3:1–28 doi: 10.3233/JND-160145. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6.Park JH, Olsen NJ, LK Jr, et al. Use of magnetic resonance imaging and p-31 magnetic resonance spectroscopy to detect and quantify muscle dysfunction in the amyopathic and myopathic variants of dermatomyositis. Arthritis Rheum. 1995;38:68–77 doi: 10.1002/art.1780380111. [DOI] [PubMed] [Google Scholar]
  • 7.Park JH, Vansant JP, Kumar NG, et al. Dermatomyositis: Correlative MR imaging and P-31 MR spectroscopy for quantitative characterization of inflammatory disease. Radiology 1990;177:473–479 doi: 10.1148/radiology.177.2.2217788. [DOI] [PubMed] [Google Scholar]
  • 8.Wattjes MP, Kley RA, Fischer D. Neuromuscular imaging in inherited muscle diseases. Eur. Radiol. 2010;20:2447–2460 doi: 10.1007/s00330-010-1799-2. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9.Patten C, Meyer RA, Fleckenstein JL. T2 Mapping of Muscle. Semin. Musculoskelet. Radiol. 2003;7:297–305 doi: 10.1055/s-2004-815677. [DOI] [PubMed] [Google Scholar]
  • 10.Mercuri E, Pichiecchio A, Allsop J, Messina S, Pane M, Muntoni F. Muscle MRI in inherited neuromuscular disorders: Past, present, and future. J. Magn. Reson. Imaging 2007;25:433–440 doi: 10.1002/jmri.20804. [DOI] [PubMed] [Google Scholar]
  • 11.Fan RH, Does MD. Compartmental relaxation and diffusion tensor imaging measurements in vivo in λ-carrageenan-induced edema in rat skeletal muscle. NMR Biomed. 2008;21:566–573 doi: 10.1002/nbm.1226. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12.Varghese J, Scandling D, Joshi R, et al. Rapid assessment of quantitative T 1, T 2 and T 2 * in lower extremity muscles in response to maximal treadmill exercise. NMR Biomed. 2015;28:998–1008 doi: 10.1002/nbm.3332. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13.Sesto ME, Chourasia AO, Block WF, Radwin RG. Mechanical and magnetic resonance imaging changes following eccentric or concentric exertions. Clin. Biomech. 2008;23:961–968 doi: 10.1016/j.clinbiomech.2008.03.068. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 14.Wang L, Regatte RR. T MRI of human musculoskeletal system. J. Magn. Reson. Imaging 2015;41:586–600 doi: 10.1002/jmri.24677. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15.Virta A, Komu M, Lundbom N, Kormano M. T1ρ MR imaging characteristics of human anterior tibial and gastrocnemius muscles. Acad. Radiol. 1998;5:104–110 doi: 10.1016/S1076-6332(98)80130-X. [DOI] [PubMed] [Google Scholar]
  • 16.Wang P, Zhu H, Kang H, Gore JC. R1ρ dispersion and sodium imaging in human calf muscle. Magn. Reson. Imaging 2017;42:139–143 doi: 10.1016/j.mri.2017.07.020. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 17.Lamminen AE, Tanttu JI, Sepponen RE, Pihko H, Korhola OA. T(1p) dispersion imaging of diseased muscle tissue. Br. J. Radiol. 1993;66:783–787 doi: 10.1259/0007-1285-66-789-783. [DOI] [PubMed] [Google Scholar]
  • 18.Ma D, Gulani V, Seiberlich N, et al. Magnetic resonance fingerprinting. Nature 2013;495:187. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19.Sharafi A, Zibetti MVW, Chang G, Cloos M, Regatte RR. MR fingerprinting for rapid simultaneous T 1, T 2, and T 1 ρ relaxation mapping of the human articular cartilage at 3T. Magn. Reson. Med. 2020;0:1–9 doi: 10.1002/mrm.28308. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 20.Yu AC, Badve C, Ponsky LE, et al. Development of a Combined MR Fingerprinting and Diffusion Examination for Prostate Cancer. Radiology 2017;283:729–738 doi: 10.1148/radiol.2017161599. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 21.Cloos MA, Assländer J, Abbas B, et al. Rapid Radial T1 and T2 Mapping of the Hip Articular Cartilage With Magnetic Resonance Fingerprinting. J. Magn. Reson. Imaging 2019;50:810–815 doi: 10.1002/jmri.26615. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 22.Cloos MA, Knoll F, Zhao T, et al. Multiparametric imaging with heterogeneous radiofrequency fields. Nat. Commun. 2016;7:1–10 doi: 10.1038/ncomms12445. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 23.Chen Y, Jiang Y, Pahwa S, et al. MR Fingerprinting for Rapid Quantitative Abdominal Imaging. Radiology 2016;279:278–286 doi: 10.1148/radiol.2016152037. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 24.Badve C, Yu A, Rogers M, et al. Simultaneous T1 and T2 Brain Relaxometry in Asymptomatic Volunteers Using Magnetic Resonance Fingerprinting. Tomography 2015;1:136–144 doi: 10.18383/j.tom.2015.00166. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 25.Badve C, Yu A, Dastmalchian S, et al. MR fingerprinting of adult brain tumors: Initial experience. In: American Journal of Neuroradiology. Vol. 38. American Society of Neuroradiology; 2017. pp. 492–499. doi: 10.3174/ajnr.A5035. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 26.Jaubert O, Arrieta C, Cruz G, et al. Multiparametric liver tissue characterization using MR fingerprinting: Simultaneous T 1, T 2,, and fat fraction mapping. Magn. Reson. Med. 2020:mrm.28311 doi: 10.1002/mrm.28311. [DOI] [PubMed] [Google Scholar]
  • 27.Hermann I, Chacon-Caldera J, Brumer I, et al. Magnetic resonance fingerprinting for simultaneous renal T 1 and mapping in a single breath-hold. Magn. Reson. Med. 2020;83:1940–1948 doi: 10.1002/mrm.28160. [DOI] [PubMed] [Google Scholar]
  • 28.Hamilton JI, Jiang Y, Chen Y, et al. MR fingerprinting for rapid quantification of myocardial T 1, T 2, and proton spin density. Magn. Reson. Med. 2017;77:1446–1458 doi: 10.1002/mrm.26216. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 29.Tesfaye S, Boulton AJM, Dickenson AH. Mechanisms and management of diabetic painful distal symmetrical polyneuropathy. Diabetes Care 2013;36:2456–2465 doi: 10.2337/dc12-1964. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 30.Boulton AJM, Vinik AI, Arezzo JC, et al. Diabetic neuropathies: A statement by the American Diabetes Association. Diabetes Care 2005;28:956–962 doi: 10.2337/diacare.28.4.956. [DOI] [PubMed] [Google Scholar]
  • 31.Allen MD, Major B, Kimpinski K, Doherty TJ, Rice CL. Skeletal muscle morphology and contractile function in relation to muscle denervation in diabetic neuropathy. J. Appl. Physiol. 2014;116:545–552 doi: 10.1152/japplphysiol.01139.2013. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 32.Parasoglou P, Rao S, Slade JM. Declining Skeletal Muscle Function in Diabetic Peripheral Neuropathy. Clin. Ther. 2017;39:1085–1103 doi: 10.1016/j.clinthera.2017.05.001. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 33.Winkelmann S, Schaeffter T, Koehler T, Eggers H, Doessel O. An optimal radial profile order based on the golden ratio for time-resolved MRI. IEEE Trans. Med. Imaging 2007;26:68–76 doi: 10.1109/TMI.2006.885337. [DOI] [PubMed] [Google Scholar]
  • 34.Yuan J, Wang YXJ. T1rho MR imaging principle, technology, and application. In: MRI: Physics, Image Reconstruction, and Analysis. CRC Press; 2015. pp. 1–26. doi: 10.1201/b15511-20. [DOI] [Google Scholar]
  • 35.Brown R, Khegai O, Parasoglou P. Magnetic Resonance Imaging of Phosphocreatine and Determination of BOLD Kinetics in Lower Extremity Muscles using a Dual-Frequency Coil Array. Sci. Rep. 2016;6:1–8 doi: 10.1038/srep30568. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 36.Chen W Errors in quantitative T1rho imaging and the correction methods. Quant. Imaging Med. Surg. 2015;5:583–58391 doi: 10.3978/j.issn.2223-4292.2015.08.05. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 37.Mitrea BG, Krafft AJ, Song R, Loeffler RB, Hillenbrand CM. Paired self-compensated spin-lock preparation for improved T1ρ quantification. J. Magn. Reson. 2016;268:49–57 doi: 10.1016/J.JMR.2016.04.017. [DOI] [PubMed] [Google Scholar]
  • 38.Charagundla SR, Borthakur A, Leigh JS, Reddy R. Artifacts in T1ρ-weighted imaging: Correction with a self-compensating spin-locking pulse. J. Magn. Reson. 2003;162:113–121 doi: 10.1016/S1090-7807(02)00197-0. [DOI] [PubMed] [Google Scholar]
  • 39.Zeng H, Daniel G, Gochberg C, Zhao Y, Avison M, Gore JC. A composite spin-lock pulse for delta B0 and B1 insensitive T1rho measurements. In: Proceedings of the 14th Annual Meeting of ISMRM.; 2006. [Google Scholar]
  • 40.Gram M, Gensler D;, Xu A, et al. A totally balanced spin lock preparation module for accurate and artifact-free T1ρ-mapping. In: ISMRM; 27th International Society for Magnetic Resonance in Medicine Annual Meeting. Montreal; 2019. p. 1215. [Google Scholar]
  • 41.Sharafi A, Chang G, Regatte RR. Bi-component T1ρ and T2 Relaxation Mapping of Skeletal Muscle In-Vivo. Sci. Rep. 2017;7:1–9 doi: 10.1038/s41598-017-14581-9. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 42.Weigel M Extended phase graphs: Dephasing, RF pulses, and echoes - Pure and simple. J. Magn. Reson. Imaging 2015;41:266–295 doi: 10.1002/jmri.24619. [DOI] [PubMed] [Google Scholar]
  • 43.McGivney DF, Pierre E, Ma D, et al. SVD compression for magnetic resonance fingerprinting in the time domain. IEEE Trans. Med. Imaging 2014;33:2311–2322 doi: 10.1109/TMI.2014.2337321. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 44.Assländer J, Cloos MA, Knoll F, Sodickson DK, Hennig J, Lattanzi R. Low rank alternating direction method of multipliers reconstruction for MR fingerprinting. Magn. Reson. Med. 2018;79:83–96 doi: 10.1002/mrm.26639. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 45.Zhao B, Setsompop K, Adalsteinsson E, et al. Improved magnetic resonance fingerprinting reconstruction with low-rank and subspace modeling. Magn. Reson. Med. 2018;79:933–942 doi: 10.1002/mrm.26701. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 46.Doneva M, Amthor T, Koken P, Sommer K, Börnert P. Matrix completion-based reconstruction for undersampled magnetic resonance fingerprinting data. Magn. Reson. Imaging 2017;41:41–52 doi: 10.1016/j.mri.2017.02.007. [DOI] [PubMed] [Google Scholar]
  • 47.Mazor G, Weizman L, Tal A, Eldar YC. Low-rank magnetic resonance fingerprinting. Med. Phys. 2018;45:4066–4084 doi: 10.1002/mp.13078. [DOI] [PubMed] [Google Scholar]
  • 48.Hilbert T, Xia D, Block KT, et al. Magnetization transfer in magnetic resonance fingerprinting. Magn. Reson. Med. 2019:mrm.28096 doi: 10.1002/mrm.28096. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 49.Wyatt CR, Barbara TM, Guimaraes AR. T 1ρ magnetic resonance fingerprinting. NMR Biomed. 2020;33 doi: 10.1002/nbm.4284. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 50.Taheri S, Sood R. Spin-lock MRI with amplitude- and phase-modulated adiabatic waveforms: An MR simulation study. Magn. Reson. Imaging 2006;24:51–59 doi: 10.1016/j.mri.2005.10.020. [DOI] [PubMed] [Google Scholar]
  • 51.Gröhn HI, Michaeli S, Garwood M, Kauppinen RA, Gröhn OHJ. Quantitative T1ρ and adiabatic Carr-Purcell T2 magnetic resonance imaging of human occipital lobe at 4 T. Magn. Reson. Med. 2005;54:14–19 doi: 10.1002/mrm.20536. [DOI] [PubMed] [Google Scholar]
  • 52.Michaeli S, Sorce DJ, Springer CS, Ugurbil K, Garwood M. T1ρ MRI contrast in the human brain: Modulation of the longitudinal rotating frame relaxation shutter-speed during an adiabatic RF pulse. J. Magn. Reson. 2006;181:135–147 doi: 10.1016/j.jmr.2006.04.002. [DOI] [PubMed] [Google Scholar]
  • 53.Ma Y-J, Carl M, Searleman A, Lu X, Chang EY, Du J. 3D adiabatic T prepared ultrashort echo time cones sequence for whole knee imaging. Magn. Reson. Med. 2018;80:1429–1439 doi: 10.1002/mrm.27131. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 54.Chen W Artifacts correction for T1rho imaging with constant amplitude spin-lock. J. Magn. Reson. 2017;274:13–23 doi: 10.1016/j.jmr.2016.11.002. [DOI] [PubMed] [Google Scholar]
  • 55.Jiang B, Chen W. On-resonance and off-resonance continuous wave constant amplitude spin-lock and T quantification in the presence of B 1 and B 0 inhomogeneities. NMR Biomed. 2018;31:e3928 doi: 10.1002/nbm.3928. [DOI] [PubMed] [Google Scholar]
  • 56.Bipin Mehta B, Coppo S, Frances McGivney D, et al. Magnetic resonance fingerprinting: a technical review. Magn. Reson. Med. 2019;81:25–46 doi: 10.1002/mrm.27403. [DOI] [PubMed] [Google Scholar]
  • 57.Block KT, Uecker M. Simple Method for Adaptive Gradient-Delay Compensation in Radial MRI. Proc Intl Soc Magn Reson Med 2011;19:2816. [Google Scholar]
  • 58.Block KT, Chandarana H, Milla S, et al. Towards Routine Clinical Use of Radial Stack-of-Stars 3D Gradient-Echo Sequences for Reducing Motion Sensitivity. J. Korean Soc. Magn. Reson. Med. 2014;18:87 doi: 10.13104/jksmrm.2014.18.2.87. [DOI] [Google Scholar]
  • 59.Ma D, Jiang Y, Chen Y, et al. Fast 3D magnetic resonance fingerprinting for a whole-brain coverage. Magn. Reson. Med. 2018;79:2190–2197 doi: 10.1002/mrm.26886. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 60.Ma D, Jones SE, Deshmane A, et al. Development of high-resolution 3D MR fingerprinting for detection and characterization of epileptic lesions. J. Magn. Reson. Imaging 2019;49:1333–1346 doi: 10.1002/jmri.26319. [DOI] [PubMed] [Google Scholar]
  • 61.Liao C, Bilgic B, Manhard MK, et al. 3D MR fingerprinting with accelerated stack-of-spirals and hybrid sliding-window and GRAPPA reconstruction. Neuroimage 2017;162:13–22 doi: 10.1016/j.neuroimage.2017.08.030. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 62.Fujimoto K, Cloos MA, Urushibata Y, Okada T. Cortical T1 mapping with 3D MR Fingerprinting at 7T using a single transmit channel. In: Joint Annual Meeting ISMRM-ESMRMB. Paris; 2018. p. 1135. [Google Scholar]
  • 63.Zhao B, Haldar JP, Liao C, et al. Optimal experiment design for magnetic resonance fingerprinting: Cramér-rao bound meets spin dynamics. IEEE Trans. Med. Imaging 2019;38:844–861 doi: 10.1109/TMI.2018.2873704. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 64.Cohen O, Zhu B, Rosen MS. MR fingerprinting Deep RecOnstruction NEtwork (DRONE). Magn. Reson. Med. 2018;80:885–894 doi: 10.1002/mrm.27198. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 65.Sharafi A, Xia D, Chang G, Regatte RR. Biexponential T1ρ relaxation mapping of human knee cartilage in vivo at 3 T. NMR Biomed. 2017;30 doi: 10.1002/nbm.3760. [DOI] [PMC free article] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

FIG S1

Supplementary Figure 1. The error of estimating T from Bloch simulation for SC-SL (a,c) and (b,d) TB-SL module in the presence of high and low field inhomogeneities. The estimation error is smaller using the TB-SL module.

RESOURCES