Abstract
The A-value used in cochlear duct length (CDL) estimation does not take malformed cochleae into consideration. The objective was to determine the A-value reported in the literature, to assess the accuracy of the A-value measurement and to evaluate a novel cochlear measurement in distinguishing malformed cochlea. High resolution Computer Tomography images in the oblique coronal plane/cochlear view of 74 human temporal bones were analyzed. The A-value and novel C-value measurement were evaluated as predictors of inner ear malformation type. The proximity of the facial nerve to the basal turn was evaluated subjectively. 26 publications report on the A-value; but they do not distinguish normal vs. malformed cochleae. The A-values of the normal cochleae compared to the cochleae with cochlear hypoplasia, incomplete partition (IP) type I, -type II, and -type III were significantly different. The A-value does not predict the C-value. The C-values of the normal cochleae compared to the cochleae with IP type I and IP type III were significantly different. The proximity of the facial nerve to the basal turn did not relate to the type of malformation. The A-value is different in normal vs. malformed cochleae. The novel C-value could be used to predict malformed anatomy, although it does not distinguish all malformation types.
Subject terms: Anatomy, Health care, Medical research
Introduction
The measurement of the cochlear duct length (CDL) was first reported by direct measurement in 1884 and many years later, the CDL was determined by indirect measurement and graphic reconstruction1; following which Escudé et al. introduced the ‘A-value’ measurement of the cochlea2 and proposed mathematical equations to estimate the CDL along the outer wall using the A-value. Alexiades et al.4. later described how the A-value could reliably be used to estimate the CDL along the basilar membrane, after modifying the equations derived by Escudé; demonstrating results that were in agreement with the outcomes of Hardy’s histological study3,4.
In the oblique-coronal plane, the basal turn of the cochlea can be viewed fully, along with the round window (RW) entrance and the three semi-circular canals of the vestibular organ. This has been termed the ‘Cochlear View’ by Xu et al.5. The ‘cochlear view’ is the best view in which to measure the A-value. The position of the facial nerve in relation to the electrode array can also be identified accurately in the ‘cochlear view’.
In the ‘cochlear view’ we have experienced that the curvature of the cochlear turn of the cochlea is highly variable6. If the beginning of the second turn of the cochlea is far away from the RW, the curvature of the second turn of the cochlea becomes very tight, affecting the insertion angle considerably. Likewise, Escudé et al. had suggested that the cochlear size influences greatly the final insertion depth of the electrode2. While the A-value has been recognized as a reliable measurement for CDL estimation for cochlear implantation, it does not account for variability in the curvature of the second turn of the cochlea. Furthermore, the A-value can only be applied to inner ears with a normal anatomy7, because malformed cochlea generally do not possess two and a half turns like the normal cochlea; the cochlear duct is typically shorter8.
This is of importance because up to 20% of the population with a sensorineural hearing loss have some degree of inner ear malformation 9, with enlarged vestibular aqueduct (EVA), incomplete partition (IP) types I, II and III, cochlear hypoplasia (CH), common cavity (CC), and cochlear aplasia (CA) being the most common 7,10. Therefore, applying the A-value measurement and CDL estimation under circumstances where the cochleae have a malformation of any kind be ill advised. In such cases, it is necessary to consider the anatomy of the cochlea and approach the estimation of the CDL individually.
Therefore, this study set out to determine the reliance on the A-value measurement in both normal and malformed cochleae. We hypothesized that the A-value is different in normal versus malformed cochleae. We also sought an alternative measurement that could be used as a landmark to distinguish between normal and malformed cochlea.
Methods
To determine the present-day reliance on the A-value measurement, a literature search was performed.
To assess the accuracy of the A-value measurement, in normal and malformed cochleae, image analyses of pre-operative high resolution Computer Tomography (HRCT) image datasets in the ‘cochlear view’ of anonymous human temporal bones shared by multiple clinics for educational purposes from the year 2011 to 2020 were analyzed retrospectively.
As an alternative landmark, the gap between the inner wall at the beginning of the cochlea to the outer wall at the beginning of the second turn of the cochlea was measured. This novel anatomical indicator was termed the C-value measurement.
Literature summary
A literature search of the key words “Cochlear duct length measurement” was performed using PubMed as the search engine. All studies that reported on the A-value taking in cadaveric human temporal bones and live human subjects were included in the analyses.
Image analyses
Seventy-four preoperative HRCT images were analyzed retrospectively using 3D slicer, version 4.10.2, freeware (https://www.slicer.org/).
The A-value was determined in the oblique coronal plane starting at the entrance of the round window and passing through the mid-modiolar section to the opposite side of the lateral wall in ‘cochlear view’ as described previously by Escudé et al.2.
The C-value was determined in the oblique coronal plane/cochlear view and along the ‘A’ value line by measuring the gap between the inner wall at the beginning of the cochlea to the outer wall at the beginning of the second turn of the cochlea.
The proximity of the facial nerve to the basal turn of the cochlea was evaluated subjectively in all 74 image datasets.
Three-dimensional (3D) segmentation of the complete inner-ear was performed using 3D slicer as described previously by Dhanasingh et al. 7,11.
Statistical analyses
The A- and C-values of the normal and malformed cochleae were compared using two sample t-tests with unequal variance in Microsoft Excel for Office 365 (Version 2020).
Regression estimates between the A- and C-values of each type of cochlea (normal vs. malformation) were determined using the data analysis tool in Microsoft Excel. A p-value < 0.05 was considered statistically significant.
Results
Literature summary
Eighty-four articles in total were retrieved from the literature search and upon reading the abstract thoroughly and reviewing the articles briefly for presence of the A-value, twenty-six articles were identified that satisfied the inclusion criteria (Fig. 1). The literature search results are shown in Table 12,4,12–35. From these twenty-six reports, a total of three-thousand-three-hundred and thirty-three cochleae were analyzed for the A-value. The A-value from this literature review ranged from a minimum of 7.9 mm to a maximum of 10.2 mm with an average value of 9.13 mm. None of the articles within the inclusion criteria reported on the A-value in malformed cochleae.
Table 1.
No | Study | N | Mean A-value | A-value range |
---|---|---|---|---|
1 | Ordonez et al.12 | 5 | 10.45 ± 0.18 | – |
2 | Khurayzi et al.13 | 256 | 8.45 | 7.5–9.4 |
3 | Nateghifard et al.14 | 10 | 8.89 ± 0.3 | – |
4 | Kuthubutheen et al.15 | 55 | 8.91 ± 0.37 | – |
5 | Nash et al.16 | 40 | 9.1 ± 0.49 | – |
6 | Zahara et al.17 | 36 | 8.75 ± 0.31 | – |
7 | Hong et al.18 | 120 | 8.55 ± 0.31 | – |
8 | Stefanescu et al.19 | 23 | 9.14 ± 0.415 | – |
9 | Schurzig et al.20 | 10 | 9.61 ± 0.54 | – |
10 | Iyaniwura et al.21 | 20 | 9.05 | 8.4–9.7 |
11 | Grover et al.22 | 124 | 8.45 | 7.7–9.2 |
12 | An et al.23 | 26 | 9.75 | 9.0–10.5 |
13 | Liu et al.24 | 102 | 8.85 | 8.1–9.6 |
14 | Rivas et al.25 | 275 | 9.22 ± 0.44 | 8.0–10.3 |
15 | Deep et al.4 | 40 | 9.5 | 8.5–10.5 |
16 | Thong et al.26 | 314 | 9.2 | 8.1–10.3 |
17 | Mosnier et al.27 | 8 | 9.3 ± 0.44 | – |
18 | Meng et al.28 | 310 | 9.3 | 8.1–10.5 |
19 | Franke-Trieger et al.29 | 10 | 9.0 | 8.3–9.67 |
20 | Van der Marel et al.30 | 671 | 9.1 | 7.3–10.9 |
21 | Avci et al.31 | 16 | 9.5 | 8.8–10.1 |
22 | Pelliccia et al.32 | 482 | 9.3 | 7.14–11.4 |
23 | Erixon et al.33 | 325 | 9.1 | 8.3–9.9 |
24 | Martinez-Monedero et al.34 | 104 | 8.5 | 6.8–10.3 |
25 | Stakovskaya et al.35 | 9 | 9.1 | 7.3–10.9 |
26 | Escudé et al.2 | 42 | 9.23 ± 0.53 | 7.9–10.8 |
3433 | 9.13 | 7.9–10.2 |
Data analyses
Of the seventy-four cochleae investigated there were: 10 normal, 3 with EVA, 25 with CH, 15 with IP type I, 11 with IP type II, and 10 with IP type III. HRCT images with the A-value and C-value, and corresponding 3D segmented images covering all the anatomical types of the inner-ear in cochlear view are shown in Fig. 2a. Images of all the seventy-four cochlear samples used are provided in Fig. 2b.
The A-values and the C-Values (mean ± standard deviation) for each group are shown in Fig. 3a,b, respectively. The A-values of the normal cochleae compared to the cochleae with CH (p < 0.001), IP type I (p = 0.049), IP type II (p = 0.038), and IP type III (p < 0.001) were significantly different. However, the A-values of the normal cochleae were not significantly different to the cochleae with EVA (p < 0.430). The C-values of the normal cochleae compared to the cochleae with IP type I (p < 0.001) and IP type III (p < 0.001) were significantly different, but there were no significant differences between the C-value and the other malformation types.
The A-values of the normal cochleae were significantly different to the combined A-values of the cochleae with a malformation (p < 0.001) (Fig. 4a). The C-values of the normal cochleae were not significantly different to the combined C-values of the cochleae with a malformation (Fig. 4b).
Table 2 shows the exact A- and C-values (mean ± standard deviation) for the normal cochleae and each type of malformation, the p-values of the differences as determined by t-test between the A- and C-values, and the p-values of the regression analyses.
Table 2.
A-value (mean ± SD) | C-value (mean ± SD) |
p-value t-test A vs C |
p-value regression A vs C |
|
---|---|---|---|---|
Normal | 9.12 ± 0.54 | 1.65 ± 0.24 | < .001 | 0.693 |
EVA | 8.82 ± 0.50 | 1.62 ± 0.41 | < .001 | 0.863 |
CH | 7.82 ± 1.04 | 1.71 ± 0.49 | < .001 | 0.104 |
IP type I | 8.67 ± 0.49 | 1.05 ± 0.20 | < .001 | 0.262 |
IP type II | 8.51 ± 0.69 | 1.52 ± 0.18 | < .001 | 0.456 |
IP type III | 7.80 ± 0.42 | 2.48 ± 0.28 | < .001 | 0.562 |
The A-value of the normal cochleae (Fig. 5a), EVA (Fig. 5b), CH (Fig. 5c), IP type I (Fig. 5d), IP type II (Fig. 5e), or IP type III (Fig. 5f) were not significant predictors of the corresponding C-value.
The proximity of the facial nerve to the cochlear basal turn of the cochlea at around 270° of angular depth, was evaluated subjectively in all 74 samples and it was determined that the proximity was highly variable, regardless of the anatomical type (Fig. 2b).
Discussion
The data collected shows that while the A-value measurement has been reported in the literature and is used to determine the CDL in clinical practice, the A-value measurement varies greatly. A-value measurements for CDL prediction apply only to normal cochleae as the mathematical equations to estimate CDL are derived taking two-and half turns of the fully formed cochlea into account. However, the malformed cochleae have defects in the full development of the cochlear turns making it difficult to apply the mathematical equations to estimate the available length of cochlear lumen for electrode placement. The C-value measurement is a novel dimension of the cochlea presented herein and it distinguishes certain types of malformed cochlea from normal cochlea.
The introduction and use of the A-value measurement to the field of cochlear implantation means that the otolaryngologist can quickly determine pre-operatively, via CDL estimation, the length of the electrode that is best suited for the cochlea. Twenty-six papers were published on the A-value measurement since 2006 of which sixteen papers were published in the last two years showing the importance given to the A-value in present-day cochlear implant therapy. However, none of the papers that reported on the use of the A-value distinguished between normal and malformed cochleae.
In the early days of cochlear implantation, patients with a malformation were initially excluded from cochlear implantation. However, these days a high percentage of patients with inner ear malformations are undergoing surgery successfully36. Despite significant benefits in sound detection, some risks and surgical challenges for these patients still exist37–41, e.g. facial nerve stimulation is one of the most common complications of CI surgery and cochlear malformation has been associated with a higher incidence of facial nerve stimulation42. It is thought that the location of the labyrinthine segment of the facial nerve in relation to the superior segment of the basal turn of the cochlea leads to the facial nerve stimulation37,43. Therefore, to avoid such risks a proper understanding of the topographic anatomy of the cochlea is essential. In particular, the use of one plane, the oblique coronal plane, or ‘cochlear view’ is advised. We have shown herein that determining the A-value in the ‘cochlear view’ the mean A-value was 9.12 ± 0.89 in normal cochleae (n = 10). This value is very much in-line with the findings in the literature that showed an average A-value of 9.13 mm measured from a sample number of 3433. However, the A-value was significantly lower in malformed cochleae (8.18 ± 0.86). Specifically, the A-value was significantly lower in CH (p < 0.001) and IP type III (p < 0.001); and it was marginally lower in IP type I (p = 0.05), -II (p = 0.04), and EVAS (p = 0.5) cochleae, compared to the normal cochleae (Fig. 3a). This indicates that the A-value is not an appropriate measurement in malformed cochleae.
If one undertakes the A-value measurement without considering the apical anatomy of the cochlea, the A-value measurement will yield a CDL with a false measurement in malformed cochleae. In clinical practice this leads to selecting the wrong electrode. For example, using a longer than necessary electrode, which increases the probability of cochlear trauma, or partial insertion. Recent reports from our clinic, indicate that electrode tip fold-over is a consequence if the full insertion of a long electrode is attempted in a malformed cochlea with a cystic apex44.
The C-value is a novel cochlear dimension introduced in this study. The A-value is significantly different to the C-value. The regression analyses showed that the A-value does not predict the C-value of the individual malformation types. The C-value measurement of normal cochleae was significantly different to the C-value measurement of IP type I (p < 0.001) and IP type III (p < 0.001) cochleae, but not significantly different to the other malformation types (Fig. 3b). Thus, the C-value could be used as a quantitative positive predictor of the malformation types IP type I and IP type III, but it does not completely distinguish all malformation types. Therefore, we recommend the 3D segmentation of the complete inner ear structures to determine malformation types and to choose the appropriate length electrode. It remains to be seen, were we not constrained by the number of HRCT images available, if the method could predict with greater accuracy the presence of malformed cochleae or different types of malformation.
Based on the HRCT images that we had available to us, we would predict that the maximum electrode coverage would be 360° for an IP type I; 450° for an IP type II, and 360° for an IP type III malformed cochlea. The CH type malformation shows great variation in the cochlear turns available and, as the data indicated, they differ significantly to the normal cochleae in terms of the A-value and on the number of turns available. However, based on the CH samples presented in this study, the ideal electrode coverage for malformation types would vary from 180° to 360°. The number of EVA malformation samples presented in this study was too small (n = 3) to enable prediction. However, EVA is thought to have the appearance of a normal cochlear part with the vestibular aqueduct being enlarged. Therefore, the electrode choice for normal anatomy cochlea could be applied to EVA, if no defect in the cochlear portion was detected.
The oblique coronal plane/cochlear view is a ‘one-stop shop’ view in which the A-value, C-value and even the proximity of the facial nerve to the basal turn of the cochlea can be measured. Our previously published data showed that patients who experience postoperative facial nerve stimulation have a significantly lower distance and bone density between the upper basal turn of the cochlea and the labyrinthine segment of the facial nerve45. However, subjective evaluation of the proximity of the facial nerve to the basal turn in the present study did not appear to have any relationship to the type of malformation.
Conclusion
In conclusion, the scientific evidence is in favor of the A-value measurement in modern cochlear implant therapy. The novel C-value measurement could prospectively be used to predict malformed anatomy, although both the A-value and the C-value do not distinguish all malformation types. Oblique coronal view/cochlear view is a complete solution in which to measure the cochlear parameters and to visualize the proximity of the facial nerve to the basal turn of the cochlea. 3D segmentation of the complete inner-ear especially in malformed cochlear anatomies should be considered as a routine in pre-operative image analysis and in our experience we find it highly useful in understanding the anatomy of the malformed cochleae.
Acknowledgements
The authors would gratefully like to acknowledge the clinics and patients from across the world that shared their anonymized pre-operative CT scans for educational purposes. We would also like to thank Masoud Zoka Assadi from MED-EL for his valuable discussion on the C-value measurement mentioned in this work and Dijana Mitrovic from MED-EL for her assistance in the 3D segmentation of the inner-ear from the pre-operative CT scans. The authors are grateful to the Deanship of Scientific Research, King Saud University for funding through Vice Deanship of Scientific Research Chairs.This was one of the research projects recommended by the Saudi Otorhinolaryngology Society.
Author contributions
T.K.: Data collection, image analyses, writing, literature summary. F.A.: Data collection, image analyses, literature summary. A.A.: Data collection, study design, writing. Y.A.: Statistical analyses, study design, critical review. Ú.D.: Statistical analyses, writing. A.D.: Data collection, image analyses, statistical analyses, literature summary, writing, critical review.
Data availability
All data generated or analyzed during this study are included in this published article.
Competing interests
The authors declare no competing interests.
Footnotes
Publisher's note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
References
- 1.Hardy M. The length of the organ of Corti in man. Am. J. Anat. 1938;62(2):291–311. doi: 10.1002/aja.1000620204. [DOI] [Google Scholar]
- 2.Escudé B, James C, Deguine O, Cochard N, Eter E, Fraysse B. The size of the cochlea and predictions of insertion depth angles for cochlear implant electrodes. Audiol. Neurootol. 2006;11(Suppl 1):27–33. doi: 10.1159/000095611. [DOI] [PubMed] [Google Scholar]
- 3.Alexiades G, Dhanasingh A, Jolly C. Method to estimate the complete and two-turn cochlear duct length. Otol. Neurotol. 2015;36(5):904–907. doi: 10.1097/MAO.0000000000000620. [DOI] [PubMed] [Google Scholar]
- 4.Deep NL, Howard BE, Holbert SO, Hoxworth JM, Barrs DM. Measurement of cochlear length using the 'A' value for cochlea basal diameter: A feasibility study. Cochlear Implants Int. 2017;18(4):226–229. doi: 10.1080/14670100.2017.1292689. [DOI] [PubMed] [Google Scholar]
- 5.Xu J, Xu SA, Cohen LT, Clark GM. Cochlear view: Postoperative radiography for cochlear implantation. Am. J. Otol. 2000;21(1):49–56. doi: 10.1016/S0196-0709(00)80075-7. [DOI] [PubMed] [Google Scholar]
- 6.Khurayzi T, Dhanasingh A, Almuhawas F, Alsanosi A. Shape of the cochlear basal turn: An indicator for an optimal Electrode-to-Modiolus proximity with precurved electrode type. Ear Nose Throat J. 2021;100(1):38–43. doi: 10.1177/0145561320920965. [DOI] [PubMed] [Google Scholar]
- 7.Dhanasingh A. Variations in the size and shape of human cochlear malformation types. Anat. Rec. (Hoboken). 2019;302(10):1792–1799. doi: 10.1002/ar.24136. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8.Zheng Y, Schachern PA, Djalilian HR, Paparella MM. Temporal bone histopathology related to cochlear implantation in congenital malformation of the bony cochlea. Otol. Neurotol. 2002;23(2):181–186. doi: 10.1097/00129492-200203000-00013. [DOI] [PubMed] [Google Scholar]
- 9.Sun B, Dai P, Zhou C. Study on 2,747 cases of inner ear malformation for its classification in patient with sensorineural hearing loss. Lin Chung Er Bi Yan HouTou Jing Wai KeZaZhi. 2015;29(1):45–47. [PubMed] [Google Scholar]
- 10.Sennaroglu L, Saatci I. A new classification for cochleovestibular malformations. Laryngoscope. 2002;112(12):2230–2241. doi: 10.1097/00005537-200212000-00019. [DOI] [PubMed] [Google Scholar]
- 11.Dhanasingh A, Dietz A, Jolly C, Roland P. Human inner-ear malformation types captured in 3D. J. Int. Adv. Otol. 2019;15(1):77–82. doi: 10.5152/iao.2019.6246. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12.Ordonez F, Sudhoff H, Todt I. Electrophysiological effects of slim straight intracochlear electrode position. J. Laryngol. Otol. 2020 doi: 10.1017/S0022215120002534. [DOI] [PubMed] [Google Scholar]
- 13.Khurayzi T, Almuhawas F, Sanosi A. Direct measurement of cochlear parameters for automatic calculation of the cochlear duct length. Ann. Saudi Med. 2020;40(3):212–218. doi: 10.5144/0256-4947.2020.218. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14.Nateghifard K, Low D, Awofala L, Srikanthan D, Kuthubutheen J, Daly M, Chan H, Irish J, Chen J, Lin V, Le TN. Cone beam CT for perioperative imaging in hearing preservation Cochlear implantation—a human cadaveric study. J. Otolaryngol. Head Neck Surg. 2019;48(1):65. doi: 10.1186/s40463-019-0388-x. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 15.Kuthubutheen J, Grewal A, Symons S, Nedzelski J, Shipp D, Lin V, Chen J. The effect of cochlear size on cochlear implantation outcomes. Erratum Biomed. Res. Int. 2020;2020:6407456. doi: 10.1155/2020/6407456. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16.Nash R, Otero S, Lavy J. Use of MRI to determine cochlear duct length in patients undergoing cochlear implantation. Cochlear Implants Int. 2019;20(2):57–61. doi: 10.1080/14670100.2018.1549186. [DOI] [PubMed] [Google Scholar]
- 17.Zahara D, Dewi RD, Aboet A, Putranto FM, Lubis ND, Ashar T. Variations in cochlear size of cochlear implant candidates. Int. Arch. Otorhinolaryngol. 2019;23(2):184–190. doi: 10.1055/s-0038-1661360. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 18.Hong Y, Liu WL, Zeng QX, Gao SL, Luo RZ. Measurement and analysis of cochlea in children with congenital sensorineural hearing loss with normal inner ear structure. Lin Chung Er Bi Yan HouTou Jing Wai KeZaZhi. 2018;32(17):1316–1318. doi: 10.13201/j.issn.1001-1781.2018.17.006. [DOI] [PubMed] [Google Scholar]
- 19.Stefanescu EH, Motoi S. Selection of the appropriate cochlear electrode array using a specifically developed research software application. J. Laryngol. Otol. 2018;132(6):544–549. doi: 10.1017/S0022215118000907. [DOI] [PubMed] [Google Scholar]
- 20.Schurzig D, Timm ME, Lexow GJ, et al. Cochlear helix and duct length identification—Evaluation of different curve fitting techniques. Cochlear Implants Int. 2018;19(5):268–283. doi: 10.1080/14670100.2018.1460025. [DOI] [PubMed] [Google Scholar]
- 21.Iyaniwura JE, Elfarnawany M, Ladak HM, Agrawal SK. An automated A-value measurement tool for accurate cochlear duct length estimation. J. Otolaryngol. Head Neck Surg. 2018;47(1):5. doi: 10.1186/s40463-018-0253-3. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 22.Grover M, Sharma S, Singh SN, Kataria T, Lakhawat RS, Sharma MP. Measuring cochlear duct length in Asian population: Worth giving a thought! Eur. Arch. Oto-Rhino-Laryngol. 2018;275(3):725–728. doi: 10.1007/s00405-018-4868-9. [DOI] [PubMed] [Google Scholar]
- 23.An SY, An CH, Lee KY, Jang JH, Choung YH, Lee SH. Diagnostic role of cone beam computed tomography for the position of straight array. ActaOtolaryngol. 2018;138(4):375–381. doi: 10.1080/00016489.2017.1404639. [DOI] [PubMed] [Google Scholar]
- 24.Liu YK, Qi CL, Tang J, Jiang ML, Du L, Li ZH, Tan SH, Tang AZ. The diagnostic value of measurement of cochlear length and height in temporal bone CT multiplanar reconstruction of inner ear malformation. ActaOtolaryngol. 2017;137(2):119–126. doi: 10.1080/00016489.2016.1221132. [DOI] [PubMed] [Google Scholar]
- 25.Rivas A, Cakir A, Hunter JB, Labadie RF, Zuniga MG, Wanna GB, Dawant BM, Noble JH. Automatic cochlear duct length estimation for selection of cochlear implant electrode arrays. Otol. Neurotol. 2017;38(3):339–346. doi: 10.1097/MAO.0000000000001329. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 26.Thong JF, Low D, Tham A, Liew C, Tan TY, Yuen HW. Cochlear duct length-one size fits all? Am. J. Otolaryngol. 2017;38(2):218–221. doi: 10.1016/j.amjoto.2017.01.015. [DOI] [PubMed] [Google Scholar]
- 27.Mosnier I, Célérier C, Bensimon JL, de Seta D, Sterkers O, Nguyen Y, Bernardeschi D. Cone beam computed tomography and histological evaluations of a straight electrode array positioning in temporal bones. ActaOtolaryngol. 2017;137(3):229–234. doi: 10.1080/00016489.2016.1227477. [DOI] [PubMed] [Google Scholar]
- 28.Meng J, Li S, Zhang F, Li Q, Qin Z. Cochlear size and shape variability and implications in cochlear implantation surgery. Otol. Neurotol. 2016;37(9):1307–1313. doi: 10.1097/MAO.0000000000001189. [DOI] [PubMed] [Google Scholar]
- 29.Franke-Trieger A, Jolly C, Darbinjan A, Zahnert T, Mürbe D. Insertion depth angles of cochlear implant arrays with varying length: a temporal bone study. Otol. Neurotol. 2014;35(1):58–63. doi: 10.1097/MAO.0000000000000211. [DOI] [PubMed] [Google Scholar]
- 30.van der Marel KS, Briaire JJ, Wolterbeek R, Snel-Bongers J, Verbist BM, Frijns JH. Diversity in cochlear morphology and its influence on cochlear implant electrode position. Ear Hear. 2014;35(1):e9–20. doi: 10.1097/01.aud.0000436256.06395.63. [DOI] [PubMed] [Google Scholar]
- 31.Avci E, Nauwelaers T, Lenarz T, Hamacher V, Kral A. Variations in microanatomy of the human cochlea. J. Comp. Neurol. 2014;522(14):3245–3261. doi: 10.1002/cne.23594. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 32.Pelliccia P, Venail F, Bonafé A, Makeieff M, Iannetti G, Bartolomeo M, Mondain M. Cochlea size variability and implications in clinical practice. ActaOtorhinolaryngol. Ital. 2014;34(1):42–49. [PMC free article] [PubMed] [Google Scholar]
- 33.Erixon E, Rask-Andersen H. How to predict cochlear length before cochlear implantation surgery. ActaOtolaryngol. 2013;133(12):1258–1265. doi: 10.3109/00016489.2013.831475. [DOI] [PubMed] [Google Scholar]
- 34.Martinez-Monedero R, Niparko JK, Aygun N. Cochlear coiling pattern and orientation differences in cochlear implant candidates. Otol. Neurotol. 2011;32(7):1086–1093. doi: 10.1097/MAO.0b013e31822a1ee2. [DOI] [PubMed] [Google Scholar]
- 35.Stakhovskaya O, Sridhar D, Bonham BH, Leake PA. Frequency map for the human cochlear spiral ganglion: implications for cochlear implants. J. Assoc. Res. Otolaryngol. 2007;8(2):220–233. doi: 10.1007/s10162-007-0076-9. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 36.Lai R, Wu W, Li W, Xie D, Liu J. An analysis of surgical management of difficulties during cochlear implant with inner ear anomalies. Lin Chung Er Bi Yan HouTou Jing Wai KeZaZhi. 2020;34(10):919–924. doi: 10.13201/j.issn.2096-7993.2020.10.012. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 37.Bigelow DC, Kay DJ, Rafter KO, Montes M, Knox GW, Yousem DM. Facial nerve stimulation from cochlear implants. Am. J. Otol. 1998;19(2):163–169. [PubMed] [Google Scholar]
- 38.House JR, 3rd, Luxford WM. Facial nerve injury in cochlear implantation. Otolaryngol. Head Neck Surg. 1993;109(6):1078–1082. doi: 10.1177/019459989310900618. [DOI] [PubMed] [Google Scholar]
- 39.Ito J, Sakota T, Kato H, Hazama M, Enomoto M. Surgical considerations regarding cochlear implantation in the congenitally malformed cochlea. Otolaryngol. Head Neck Surg. 1999;121(4):495–498. doi: 10.1016/S0194-5998(99)70244-5. [DOI] [PubMed] [Google Scholar]
- 40.Maxwell AP, Mason SM, O'Donoghue GM. Cochlear nerve aplasia: Its importance in cochlear implantation. Am. J. Otol. 1999;20(3):335–337. [PubMed] [Google Scholar]
- 41.Tucci DL, Telian SA, Zimmerman-Phillips S, Zwolan TA, Kileny PR. Cochlear implantation in patients with cochlear malformations. Arch. Otolaryngol. Head Neck Surg. 1995;121(8):833–838. doi: 10.1001/archotol.1995.01890080005001. [DOI] [PubMed] [Google Scholar]
- 42.Sefien I, Hamada S. Facial nerve stimulation as a complication of cochlear implantation. Indian J. Otolaryngol. Head Neck Surg. 2019;71(4):474–479. doi: 10.1007/s12070-019-01649-3. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 43.Kelsall DC, Shallop JK, Brammeier TG, Prenger EC. Facial nerve stimulation after Nucleus 22-channel cochlear implantation. Am. J. Otol. 1997;18(3):336–341. [PubMed] [Google Scholar]
- 44.Alsughayer L, Al-Shawi Y, Yousef M, Hagr A. Cochlear electrode array tip fold-over in incomplete partition-I—A case report. Int. J. Pediatr. Otorhinolaryngol. 2020;139:110438. doi: 10.1016/j.ijporl.2020.110438. [DOI] [PubMed] [Google Scholar]
- 45.Aljazeeri IA, Khurayzi T, Al-Amro M, Alzhrani F, Alsanosi A. Evaluation of computed tomography parameters in patients with facial nerve stimulation post-cochlear implantation. Eur. Arch. Otorhinolaryngol. 2020 doi: 10.1007/s00405-020-06486-7. [DOI] [PubMed] [Google Scholar]