Wireless Powered Encoding and Broadcasting of Frequency Modulated Detection Signals

WEI QIAN 1, XIN YU 2, CHUNQI QIAN 1,3
PMCID: PMC8023641  NIHMSID: NIHMS1680236  PMID: 33828933

Abstract

Wireless transmission of locally detected RF signals is necessary for long-term operation of batteryless and embedded transducers. To improve signal transmission efficiency over larger distances, multi-stage circuits were employed to down-convert RF signals before encoding them onto the emitted carrier wave. Such multi-stage arrangement had complicated design and high-power consumption. Here, a compact and low-power wireless modulator is introduced to directly encode input RF signals onto its oscillation carrier wave. The modulator consists of a double frequency parametric resonator that is overlaid with a single frequency passive resonator to create three resonance modes. By properly adjusting the substrate thickness between resonators, the highest resonance frequency is tuned to approximately the sum of lower two resonance frequencies, enabling efficient conversion of wireless pumping power into sustained oscillation currents. When an input RF signal is present with a certain frequency offset, the oscillation signal can be frequency modulated by the input signal to create multiple modulation sidebands separated by the offset frequency. The frequency encoded carrier wave can transmit MRI signals over larger distance separations to maintain constant image sensitivity, making the modulator useful to improve the remote detectability of miniaturized implantable and interventional devices.

Keywords: Frequency modulation, nonlinear circuits, oscillator, inductive power transmission, magnetic resonance

I. INTRODUCTION

Wireless signal transducers are particularly advantageous to improve the long-term applicability of physiological sensors [1]–[3] embedded inside confined body cavities. When an external receiver coil is close enough to the embedded transducer, passive inductive coupling is normally utilized to wirelessly relay encoded signals. For communications beyond the near field region, these embedded transducers can be wirelessly powered to convert physiological parameters into voltage signals that can modulate the amplitude [4] or frequency [5] of the emitted carrier wave using analog or digital schemes. However, these wireless transducers are mainly effective for input frequencies that are much lower than the carrier frequency. To encode input signals in the radio frequency range, RF signals are usually down converted to below the MHz range [6] before being digitized to modulate a GHz carrier. Alternatively, RF input signals can be directly digitized and encoded onto carriers waves with increasingly higher frequencies, from the millimeter [7] to the optical [8] range. These multi-stage schemes require complicated circuits with larger size and higher power consumption. Even though high performance laser diodes [9] or electro-optical modulators [10]–[14] can obviate the need for ADC conversion thus reducing power consumption, inefficient transmission of light through biological tissues mandate the use of optical fibers that are still not suitable for long-term insertion into enclosed cavities. Therefore, it will be ideal to combine down-conversion and signal encoding into a single stage, for simpler circuit design and lower power consumption. Such a compact and integrated transducer can be useful to encode locally detected ultrasound signals onto EM waves that can penetrate the skull [15], or to increase the effective coupling distance of implantable MRI detectors [16]–[18] beyond the near field region. Removal of hard-wired connections will also be useful to improve the operation flexibility of interventional detectors [19]–[21].

In this work, we are going to construct a wirelessly powered frequency modulator that can combine down conversion and signal encoding into a single stage. The core component of this compact modulator is a double-frequency parametric resonator whose second resonance mode is created by bridging the virtual voltage grounds of its first resonance mode. The double-frequency parametric resonator is coaxially overlaid with another single frequency resonator to create a third resonance mode. By properly adjusting the substrate thickness between these two resonators, the third resonance frequency is tuned to approximately the sum of two lower resonance frequencies. When activated by a pumping field near the highest resonance frequency, the coupled resonator pair can utilize its nonlinear capacitance to convert wireless pumping power into oscillation currents circulating near its two lower resonance frequencies. In the presence of an input RF signal that is deviated from the oscillation signal by an offset frequency, the oscillation signal is frequency modulated by the input signal to create multiple sidebands that are separated by the offset frequency. As a result, this frequency modulated oscillation signal can be utilized to transmit locally detected MRI signals with improved sensitivity over large distance separations, until the oscillation signal is detected by an external receiver coil that is interfaced with a conventional MRI scanner. After frequency demodulation of the received oscillation signal, MR images can be reconstructed from 2D Fourier transform to maintain constant signal-to-noise ratio even though the parametric resonator is displaced from the external detector by 11-fold of its own dimension (corresponding to a 34-dB transmission attenuation with respect to a cable-connected surface coil). The wireless modulator requires only ~ 9 dBm of pumping power on the activation antenna that is separated by 4-fold the modulator’s largest dimension, making itself particularly suitable to improve the detection sensitivity of implantable or interventional MRI detectors, whose small dimension and large penetration depth will prevent efficient signal transmission by passive coupling alone. Complementary to other types of RF transducers (Table 1), our wirelessly powered frequency modulator may also operate in conjunction with embedded perpetual sensors [22] to encode multiple types of environmental signals [23] in the radio frequency range.

TABLE 1.

Summary of different types of RF transducers.

Reference [1618] [68] [914] This work
Type Inductive coupler Digital transmitter Optical modulator Parametric oscillator
Signal Processing Method On-resonance coupling Down-Conversion & Digitize & Encode Direct AM Direct FM
Carrier/input freq ratio ~1 >938 >2x106 ~1
Power consumption Passive >77 mW DC >10 mW fiber-optic <10 mW external RF
Dynamic Range --- 86 dB 81 dB 75 dB
Surmountable Attenuation <10 dB 31 dB 33 dB 34 dB

II. OPERATION PRINCIPLE

A parametric oscillator is a nonlinear multi-frequency resonator that can convert wireless pumping power into sustained oscillation currents. As shown in Fig. 1a, the resonator can be implemented as a rectangular inductor that is serially connected to two varactors on its top and bottom edges, creating a dipole mode at resonance frequency ωdr. When the virtual voltage grounds of the dipole mode are connected by a continuous conductor in the center, the resonator will have a butterfly mode at resonance frequency ωbr (Fig. 1b), where current flows in both circuit halves are opposite. When the circuit is inductively activated by an external pumping field at approximately the sum frequency ωp = ωdr + ωbr, the pumping power is effectively converted into sustained oscillation currents in both the dipole and butterfly modes. This power conversion process is facilitated by the voltage-dependent capacitance of both varactors:

C2C20(1+λ2Vp/ϕ2) (1)

where C20 is the junction capacitance at zero bias, λ2 is a device constant describing the abruptness of charge distribution across the depletion layer, ϕ2 is the junction potential of the depletion layer, Vp is the effective pumping voltage across each diode.

FIGURE 1.

FIGURE 1.

The double frequency parametric resonator can be implemented as a rectangular resonator with identical nonlinear capacitors (C2) symmetrically distributed on its top and bottom edges. (a) When the circuit’s left and right edges are connected by a continuous conductor in their centers, the dipole mode resonance can be analyzed as two identical half circuits that are grounded in the center. Each half circuit has an effective impedance of Rd+jXd with current flows in the same directions. (b) The circuit’s butterfly mode resonance mode can be considered as two half circuits that are electrically isolated in the center. Both half circuits have current flows in opposite directions; each half circuit has effective impedance of Rb+jXb. (c) When the double frequency parametric resonator is coaxially overlaid with another single frequency passive resonator, a third resonance mode is created to allow opposite current flows in both resonators. By properly adjusting the substrate thickness (shown in grey) between the coupled resonators, the highest resonance frequency becomes approximately the sum of the two lower resonance frequencies, enabling effective conversion of pumping currents induced in the counter-rotating mode into sustained oscillation currents in the dipole and butterfly modes.

Although parametric oscillators were implemented for pressure [24] and voltage [5] sensing, those circuits were only utilized for quasi-static signal sensing. In this work, we will extend the applicability of coupled parametric resonator towards RF signal sensing. Suppose an input voltage source ζd has a frequency that is slightly deviated from the dipole mode oscillation frequency ωd by Δωd, the voltage-current relation [24] at ωd can be modified to the following approximation:

ξdejΔωdtId(Rd+jXd)+λ2ϕ2VpIbjC20ωb(ϕ22λ22|Vp|2) (2)

where Rd+jXd is the effective impedance of each half circuit at ωd (as shown in Fig. 1a). Similarly, the electromotive force ζb at ωb (the butterfly mode oscillation frequency) satisfies the following approximation:

ξbejΔωbtjλ2ϕ2VpIdC20ωd(ϕ22λ22|Vp|2)+Ib(Rb+jXb) (3)

By combining Eqns. (2) and (3), we obtain

Ib=(ξdejΔωdtId(Rd+jXd))jC20ωb(ϕ22λ22|Vp|2)λ2ϕ2Vp=ξbejΔωbt+Idλ2ϕ2VpjC20ωd(ϕ22λ22|Vp|2)(RbjXb) (4)

When there is no input e.m.f. for the butterfly mode, i.e. ξb=0, Eq. (4) is equivalated to

ξdejΔωdt(RbjXb)=Id(Rd+jXd)(RbjXb)Idλ22ϕ22|Vp|2C202ωdωb(ϕ22λ22|Vp|2)2 (5)

Equating the real parts on both sides of Eq. (5):

ξd(Rbcos(Δωdt)+Xbsin(Δωdt))=Id(RdRb+XdXb)Idλ22ϕ22|Vp|2C202ωdωb(ϕ22λ22|Vp|2)2Id0 (6)

The second line in Eq. (6) is mandated to approximate zero because circuit oscillation requires the circulating current Id to be non-negligible even though the input signal ξd is close to zero. This relation can be satisfied by gradually increasing the externally applied pumping power until the effective pumping voltage Vp across both varactors is above a certain threshold. Besides the pumping power requirement, the modulated oscillation frequency can be obtained by equating the imaginary parts on both sides of Eq. (5):

ξd(jRbsin(Δωdt)jXbcos(Δωdt))=+Id(jXdRbjRdXb) (7)

Or equivalently

ξd(Rbsin(Δωdt)Xbcos(Δωdt))=ξdRbsin(Δωdt)2(ωbωbr)Lbξdcos(Δωdt)=IdRd(XdRbXbRd1)IdRd((2(ωdωdr)LdRd)/(2(ωbωbr)LbRb)1) (8)

Eq. (8) describes the relation between the RF input voltage ξd and the circuit oscillation frequencies ωd and ωb that are slightly deviated from ωdr and ωbr (i.e. the dipole and butterfly mode resonance frequencies respectively). By plugging ωb = ωpωd into Eq. (8) and solving for ωd:

ωd=2IdLbRdωbr2IdLdRbωdr2IdLbRdωp2Lb(ωbrωp)ξdcos(Δωdt)Rbξdsin(Δωdt)2LdIdRb2LbIdRd+2Lbξdcos(Δωdt)LbRdωbrLdRbωdrLbRdωpLdRbLbRd+(ωbrωp)ξdcos(Δωdt)LdIdRb/Lb+IdRdωd0+2πkfξdcos(Δωdt) (9)

Approximation in the second line of Eq. (9), as shown at the bottom of the page, is obtained under the common condition that the resonator has high quality factor, i.e. Lb(ωbrωp) ≫ Rb, and that the oscillation induced voltage is much larger than the input electromotive force, i.e. IdRdξd. According to the third line of Eq. (9), the oscillation frequency ωd is linearly modulated by the input signal ξd around a center frequency ωd0, and the magnitude of frequency deviation can be expressed as:

2πkfξd=(ωbrωp)ξd(LdIdRb/Lb+IdRd) (10)

where kf is the modulator’s frequency sensitivity. Eq. (10) indicates that the frequency deviation is independent of the frequency offset Δωd of the input signal. By integrating the instantaneous frequency in Eq. (9) to obtain the phase, the oscillation signal at time point t can be represented as:

Idexp(jωd0t+j2πkfξd0tcos(Δωdt)dt)=Idexp(jωd0t+j2πkfξdΔωdsin(Δωdt))Idexp(jωd0t+jβsin(Δωdt))=Idn=n=Jn(β)exp(jωd0t+j2πnΔωdt) (11)

where β is the frequency deviation ratio and Jn(β) is Bessel function of the first kind. Therefore, the spectrum of frequency modulated oscillation signal is the superposition of multiple peaks at ωp0+nΔωd. Moreover, the relative intensity of the first sideband with respect to the center band is:

J1(β)J0(β)β22πkfξd2Δωd=(ωbrωp)ξd2(LdIdRb/Lb+IdRd)Δωd (12)

Besides analysis of its spectral feature in the frequency domain, the oscillation signal expressed in Eq. (11) can also be demodulated in the time domain by derivatizing its phase angle, thus retrieving the magnitude of frequency deviation for a specific input signal, as specified in Eq. (10).

In order to estimate the noise contribution to the demodulated output, it is necessary to consider both the internal noise (that is detected by the modulator) and the external noise (that resides on the external receiver coil). In the time domain, the time-dependent internal noise voltage ni(t) contributes to instantaneous frequency shift of oscillation signal according to Δfi(t) = kfni(t). Because kf is invariant to input signal offset as indicated by Eq. (10), the Fourier transform of Δfi(t) is

f^(kfni(t))=kfNi (13)

where Ni is the frequency domain spectrum of the modulator’s noise voltage that is normally frequency independent. Eq. (13) describes the “internal” noise from the modulator that will be frequency encoded onto the oscillation signal. Subsequently, this oscillation signal will be detected by the external receiver coil that is remotely coupled to the modulator (Fig. 2).

FIGURE 2.

FIGURE 2.

When activated by the pumping field (red), the parametric oscillator (orange) will encode locally detected MRI signals ξd and internal noise ni onto its oscillation signal that is detected by the external receiver coil (cyan) via remote inductive coupling, where the external noise ne also affects the retrieved spectra baseline.

For very weak coupling condition, there is also non-negligible noise contribution from the external receiver coil, whose instantaneous noise voltage ne(t) will introduce an additional phase fluctuation according to ϕe(t) = ne(t)/A where A is the oscillation voltage induced on the external receiver coil. As a result, the equivalent noise in the frequency domain can be obtained by Fourier transform of the time-domain derivative of phase fluctuation:

f^(dϕe(t)2πdt)=f^(dne(t)2πAdt)=ΔfdNeA (14)

where Δfd is the frequency offset of the input signal from the oscillation signal and Ne is the frequency domain spectrum of the noise voltage originating from the external receiver coil. Ne is normally frequency independent, but its contribution to the FM demodulated output is dependent on the frequency offset Δfd, as indicated by Eq. (14). By combining the contribution from Eqns. (13) and (14), the signal-to-noise power ratio is

PSPN=kf2|ξd|2kf2Ni2+(ΔfdNe/A)2 (15)
L=μ0π[(A+B)ln(2AB/W)Aln(A+A2+B2)Bln(B+A2+B2)(A+B)/2+2(A2+B2)+0.447W] (16)

Eq. (15) indicates that when the oscillation voltage A induced on the external receiver coil is sufficiently large, ΔfdNe/A will be much smaller than kfNi, making the noise baseline almost flat in the demodulated spectrum. But when A is small due to very large transmission attenuation, the frequency dependent noise floor can no longer be neglected.

III. MATERIALS AND METHODS

A. ELECTROMAGNETIC SIMULATION

To design a wireless modulator, the first step is to specify the dimension of a double frequency parametric resonator (blue lines in Fig. 3a) whose self-inductance is [25], (16) as shown at the bottom of the page, where A and B are the rectangle’s side lengths, W is the width of the conductor strip. For a resonator with A = 10 mm and B = 7.2 mm, its effective inductance is L = 23.2 nH. When the rectangle’s conductor gaps are filled by properly chosen varactors, its dipole mode resonance ω02 should be tuned to a frequency that is slightly above the MR frequency. Meanwhile, existence of the continuous center conductor will introduce another butterfly mode resonating at ~77% the resonance frequency of the dipole mode.

FIGURE 3.

FIGURE 3.

(a) The 3D model of a wireless modulator created by overlaying a parametric resonator (blue) with a passive coupler (green). Both resonators consist of a rectangular loop inductor split by two capacitors. The modulator is activated by an antenna (red) separated by 40 mm, which is 4-fold the modulator’s largest dimension. (b) By specifying its circuit dimension, the parametric resonator is tuned by two serial connected varactors to a frequency ω02 that is slightly above the MR frequency. Subsequently, the resonance frequency ω01 of the passive coupler and the mutual inductance M between these two resonators are also determined. (c) To tune the parametric resonator to the desired frequency of ω02, two serial capacitors of C2 values are required. Similarly, to tune the passive resonator to the desired frequency of ω01, two serial capacitors of C1 values are required. (d) Each mutual inductance of M requires the two resonators to be separated by a gap size of H that can be directly incorporated into the 3D model in (a). (e) The blue curve shows the excitation efficiency of the coupled resonator pair that is separated from the antenna by 4-fold the modulator’s own dimension. The orange curve shows the excitation efficiency of a parametric in its stand-alone configuration.

To improve the circuit’s coupling efficiency with the activation antenna, a passive coupler is coaxially overlaid with the parametric resonator to locally concentrate the pumping field. This passive coupler has identical dimension as the parametric resonator but is tuned to a higher frequency at ω01. Overlapping these two resonators will not affect the butterfly mode frequency of the parametric resonator, but will create shifted dipole mode frequencies that can be estimated from the following equation [24]:

0jωLjω012Lω+ω2M2jωLjω022L/ω (17)

where ω01 and ω02 are the resonance frequencies of the passive coupler and the parametric resonator in their stand-alone configuration. M is the mutual inductance of overlapped inductors that can be estimated from [25]:

M=2(M11+M22M13M24)=2{μ0A2π[ln(AH2+B2+1+A2H2+B2)+H2+B2AH2+B2A2+1]μ0B2π[ln(BH2+A2+1+B2H2+A2)+H2+A2BH2+A2B2+1]+μ0A2π[ln(AH+1+A2H2)+HAH2A2+1]+μ0B2π[ln(BH+1+B2H2)+HBH2B2+1]} (18)

where H is the substrate thickness between the coupled resonator pair. To evaluate the relationship between dipole mode frequencies, Eq. (17) can be rearranged as:

L2ω4L2ω2(1ω012+1ω022)+L2M2ω012ω022=0 (19)

where ω01 and ω02 are resonance frequencies of uncoupled resonators. According to Vedic theorem, Eq. (19) should have two solutions for 1/ω2 that correspond to the inverse square of the lower coupled frequency 1/ωL2 and the inverse square of the higher coupled frequency 1/ωH2:

1ωL2+1ωH2=1ω012+1ω022 (20)
1ωL2ωH2=1M2/L2ω012ω022 (21)

Proper adjustment of the mutual inductance M between coupled resonators should allow the higher coupled frequency ωH to be the sum of the lower coupled frequency ωL and the butterfly mode frequency ωbr of the parametric resonator. Meanwhile, the lower coupled frequency ωL should be equal to the MR frequency:

ωHωL=ωbr0.77ω01 (22)
ωL=ωMR (23)

By plugging Eqns. (22) and (23) into Eq. (20), the required resonance frequency ω01 of the passive coupler can be estimated as:

ω01=1/1/(ωMR+0.77ω02)2+1/ωMR21/ω022 (24)

Therefore, once the resonance frequency ω02 of the parametric resonator is known, the required resonance frequency ω01 of the passive coupler can be estimated from Eq. (24), and the required mutual inductance M between these two resonators can be estimated from Eq. (21). To incorporate these parameters into the circuit model in Fig. 3a, the required serial capacitance for both resonators can be calculated as C2=2/(ω022L) and C1=2/(ω012L), as shown in Fig. 3c. Meanwhile, the required substrate thickness H between these two resonators can be estimated from Eq. (18) as shown at the bottom of the next page, to match the corresponding mutual inductance M in Fig. 3d. When unit current is applied on the dipole antenna that is separated from the modulator by 4-fold of the modulator’s own dimension, the current Ip induced inside the parametric resonator can be numerically simulated by CST Microwave Studio (Dassault Systèmes, France). According to the blue curve in Fig. 3e, the modulator has optimal excitation efficiency of about −20 dB when the resonance frequency ω02 of the parametric resonator is tuned to 306 MHz. This level of power efficiency is 29-dB better than the power efficiency of a parametric resonator without passive coupler (orange curve in Fig. 3e). Of course, for actual circuit construction, it may not be easy for the parametric resonator to precisely resonate at 306 MHz due to component value variations. But as long as the ω02 value resides between 304 MHz and 310 MHz, the excitation efficiency will only decrease by less than 0.3 dB compared to its optimal value.

B. MODULATOR CONSTRUCTION

As shown in Fig. 4a, the parametric resonator is etched on a copper-clad polyimide as a rectangular conductor loop with a continuous conductor in its center. When A = 10 mm, B = 7.2 mm and W = 0.6 mm, the rectangular loop’s total inductance is 23.2 nH. As a result, by filling the conductor gaps on the bottom and top legs with two variable capacitors (MA2737600L, Panasonic, Japan) whose zero bias capacitance is ~22.6 pF, the parametric resonator is expected to have dipole mode resonance at 310.8 MHz. This calculated resonance frequency is in close agreement with the measured value of ω02 = 310.5 MHz on the S21 curve obtained by a pair of partially overlapped receiving coils connected to the input and output ports of a network analyzer.

FIGURE 4.

FIGURE 4.

(a) The picture of a double frequency parametric resonator made of a rectangular conductor loop with a horizontal conductor in the center. The gaps in the top and bottom conductors are filled by variable capacitors C2 (MA2737600L, Panasonic, Japan). (b) The picture of a single frequency passive resonator of the same dimension but that is connected to two 9.1-pF chip capacitors connected in series. (c) The schematic representation of overlapped resonators through a polyimide substrate with thickness H. (d) When the activation antenna provides a pumping signal near its highest resonance frequency, the induced pumping current inside the modulator is converted into sustained oscillation currents in the two lower-frequency resonance modes. An input loop placed near the modulator can inject RF signals that will frequency modulate the oscillation signal. Upon reception by the weakly-coupled antennas, the oscillation signal is mixed down to below 1 MHz and observed on an oscilloscope via a low pass filter.

To bring down the dipole mode frequency to the MR frequency at ~300 MHz, the required resonance frequency for the passive resonator can be estimated as ω01 ~ 489 MHz from Eq. (24). For this purpose, another loop inductor with the same dimension as 10 × 7.2 mm2 can be serially connected to two 9.1-pF capacitors. As a result, the passive resonator is estimated to have a resonance frequency at 489.9 MHz in its stand-alone configuration, which is in close agreement with the measured value at 489.6 MHz.

By adjusting the substrate thickness H to ~1.5 mm, the mutual inductance is estimated to be 7.9 nH. As a result, the coupled resonator pair is expected to have a lower coupled frequency at 300.2 MHz and a higher coupled frequency at 538.5 MHz, both of which are in good agreement with the measured values. Because the higher coupled frequency at 538.5 MHz is approximately the sum of the lower coupled frequency at 300.2 MHz and the butterfly resonance at 238.3 MHz, the coupled resonator can be efficiently activated by a pumping signal at 538.5 MHz to generate sustained oscillation currents at 300.2 MHz and 238.3 MHz. When the activation antenna is placed 4 cm away from the modulator, which is 4-fold the modulator’s largest dimension, the antenna only requires ~9 dBm of power to activate the modulator. This level of pumping power would produce much smaller heat compared to the 10 W/kg limit for local SAR recommended by IEC 60601-2-33.

C. ON-BENCH CHARACTERIZATION

To validate the circuit’s capability for simultaneous down-conversion and frequency encoding, an additional circular loop is used to inject input signals at 300.21 MHz that is 10 kHz above the oscillation signal. After the modulated oscillation signal reaches the receiving coil, it is down-converted by a 300.0-MHz reference signal through a mixer (ZFY-2, Minicircuits, Brooklyn, NY) before being sampled at 3.25 MHz by a digital oscilloscope via a low-pass filter (BLP-1.9+, Minicircuits, Brooklyn, NY). Using Matlab (Mathworks, Natick, MA) for post processing, the digitally sampled data is Fourier transformed to obtain the frequency domain spectrum of the carrier wave and sidebands. Alternatively, to retrieve the input signal in the time domain, the digitally sampled data is combined with its Hilbert transformation to create a complex dataset. After low pass filtering by the Hamming window, the phase angle of this complex dataset is retrieved by the angle and unwrap functions, before the baseline is linearly corrected using the detrend function. Finally, the instantaneous frequency at each time point is obtained by derivatizing the phase angle using the differential filter function (differentiatorfir).

D. MR IMAGING

To demonstrate the modulator’s capability for wireless transmission of MRI signals, the modulator is placed on top of an aqueous gel phantom containing 1%-agarose. With the activation antenna placed 4-cm away, the modulator on top of the phantom is inserted into a cylindrical volume coil (Bruker Biospin, Billerica, MA) with 70-mm inner diameter. The entire assembly is located inside a 7T horizontal bore magnet equipped with Bruker Avance III spectrometer console. Inside the scanner, the pumping signal is provided by an external RF generator (E4423, Agilent, CA) through a dipole antenna. The pumping frequency is empirically adjusted to make the oscillation frequency separated from the proton MR frequency at 7T by the image bandwidth (50 kHz in this case). To emulate weak coupling condition, the modulator is fixed at the center of a homogeneous magnet, while the volume coil is displaced from the magnet’s center. Nuclei spins are excited by the RF field produced by the volume coil and coupled through the modulator. During signal reception, the oscillation signal that is frequency modulated by MRI signals are received by the external volume coil, in a similar way as how MR signals were traditionally acquired. To retrieve MR signals from phase angles of the oscillation signal, the sampling speed should be at least the Nyquist rate (twice the image bandwidth) to obtain enough phase angles for derivatization. As shown in Fig. 5b, because the spectrometer can already perform down conversion and low-pass filtering, the angle, unwrap and detrend functions are directly applied on the received oscillation signal to obtain phase angle information before the differential filter is used to retrieve MR signals in the time domain. Fourier Transform is finally used to reconstruct 2D images. To evaluate the ratio between signal and noise voltages, the same image is acquired again whose magnitude is subtracted from the first image to obtain the noise map. The sensitivity map is calculated by dividing the average intensity of individual pixels with the standard deviation of the noise map:

SNR=(|S1|+|S2|)/2std(|S1||S2|) (25)

FIGURE 5.

FIGURE 5.

(a) In MR experiments, the modulator is placed on top of a gel phantom containing 1% agarose. The modulator, along with an activation antenna placed nearby, is inserted into a cylindrical volume coil and the entire assembly is inserted into the center of a 7T magnet. To emulate weak coupling condition, the volume coil is displaced away from the magnet’s center. For each displacement, the MR excitation power is adjusted to maintain the same effective excitation angle. During MR signal acquisition, pumping power is applied on the antenna to activate the modulator whose oscillation signal is received by the cylindrical volume coil. (b) The multi-step procedure to retrieve input RF signals from the instantaneous frequencies of the oscillation signal. For MR image reconstruction, the procedures enclosed in the dashed box are already performed by the spectrometer, so the output data can be directly processed by the five remaining steps.

For comparison, the relative SNR is obtained by dividing the SNR map obtained by the parametric oscillator with the SNR map obtained by a surface coil of the same dimension but with direct wired connection to the MRI console.

IV. RESULTS

Using the experimental setup shown in Fig. 4, we first characterized the modulator on-bench. Fig. 6 was the Fourier spectrum of the oscillation signal obtained by the receiving coil that was separated from the modulator by a distance that was 4-fold the modulator’s largest dimension. As shown in the figure, the center oscillation peak was surrounded by multiple sidebands, each of which was separated from its neighbor by 10 kHz (the frequency offset between the input RF signal and the oscillation signal). As predicted by Eq. (11), the sideband pattern demonstrated the modulator’s capability to encode the input signal at the offset frequency, without the need for dedicated down-converters. Moreover, because the height of the first sideband was 24.2-dB lower than the center band, the frequency deviation ratio was β = 0.123, according to J1(β)/J0(β) ≈ β/2 in Eq. (12). This small β corresponded to narrow band modulation, thus justifying the time-domain sampling rate at twice the message bandwidth for oscillation signal demodulation, as will be mentioned below.

FIGURE 6.

FIGURE 6.

The Fourier spectrum of the oscillation signal (obtained using the setup shown in Fig. 4). An input RF signal created multiple sidebands that were separated from each other by the offset frequency between the input signal and the oscillation signal.

To retrieve the input signal from the oscillation signal, we performed frequency demodulation using the procedures described in Fig. 5b. As shown in the frequency demodulated spectrum (Fig. 7), there was a sharp peak centered around 10 kHz whose height was approximately 320-fold larger than the noise floor. Therefore, the input signal could be decreased by another factor of 320 while still being observable. The almost flat noise floor also indicated that internal noise from the modulator dominates over external noise from the receiving antenna, even though the distance separation was 4-fold the modulator’s largest dimension.

FIGURE 7.

FIGURE 7.

The input spectrum retrieved from the oscillation signal using the multi-step procedures described in Fig. 5b. The vertical axis represents the magnitude of frequency deviation defined in Eq. (10), while the horizontal axis represents frequency offset between the input signal and the oscillation signal.

Up till now, all measurements were performed when the input signal had a power level of −60 dBm. To estimate the dynamic range of the modulator, we swept the input power and measured the peak height in the demodulated spectrum. As shown in Fig. 8, the wireless modulator had a dynamic range of 75 dB. The minimum observable input signal was −110 dBm, which was 50 dBm lower than the −60-dBm input power used previously. Fig. 8 also demonstrated the good linear relation between the input signal and the frequency deviation magnitude, thus validating the prediction of Eq. (10).

FIGURE 8.

FIGURE 8.

The linear relation between the input RF signal and the frequency deviation magnitude obtained by FM demodulation.

To evaluate the wireless modulator’s frequency response, we swept the input frequency and performed frequency demodulation using the procedures described in Fig. 5b. As shown in Fig. 9, the demodulated peak had the same magnitude over at least 100 kHz. This uniform frequency response of the demodulated peak was consistent with the offset-independent relation described in Eq. (10). Therefore, white noise detected by the modulator (as input signal) would lead to flat noise baseline upon FM demodulation, as will be demonstrated below.

FIGURE 9.

FIGURE 9.

When the input signal was swept over a range of offsets from the oscillation signal, the demodulated output maintained the same magnitude of frequency deviation.

To demonstrate the modulator’s capability for wireless transmission of MR signals, the modulator was placed on top of a 1% agarose phantom and inserted into the center of a 7T MRI scanner. To emulate weak coupling condition, the volume coil was displaced by 11 cm away from the center of magnet. For this distance separation, the modulator already laid outside the volume coil’s cylindrical edge. To estimate the magnitude of transmission attenuation, an MR image (shown in Fig. 10a) was first reconstructed from signals that were passively relayed to the volume coil, using FOV = 45 × 45 mm2, acquisition matrix = 256 × 256, bandwidth = 50 kHz, effective flip angle = 5 deg, TE = 10 ms, TR = 20 ms. Compared to a conventional surface coil of the same dimension but with wired connection to the scanner console, the passively coupled modulator had a relative SNR of only ~2%, corresponding to 50-fold (or 34-dB) transmission attenuation under 11-cm distance separation. Subsequently, we turned on pumping signal and set its power level at ~0.1 dBm below the oscillation threshold to enable stable amplification with high gain (33 dB). As shown in Fig. 10b, there was significant increase in signal intensity because the circuit could regeneratively amplify MR input signals by parametric frequency mixing processes [26]–[29]. Finally, we increased the pumping power above the oscillation threshold and recorded the oscillation signal at doubled sampling rate. To maintain the same image resolution, the horizontal field-of-view (FOV) was correspondingly doubled to confine the original image region to the right half portion of the enlarged FOV. (As shown in Fig. 10c, when the FOV was increased to 90 × 45 mm2, the acquisition matrix was increased to 512 × 256 and the sampling rate was increased to 100 kHz.) By overlapping the oscillation frequency to the center of the enlarged FOV, time-domain MR signals could be retrieved by derivatizing the phase angle. The retrieved signals were subsequently Fourier Transformed to obtain the 2D image. By comparing Figs. 9c and 9b, the image obtained by frequency modulation was 24% more sensitive than the image obtained by regenerative amplification, demonstrating the better noise immunity of FM encoding. It is worth mentioning that the reconstructed image contained mirrors of the original object in both dimensions, due to the cosine-encoding relation of the message signal as described in Eq. (9).

FIGURE 10.

FIGURE 10.

(a) When the modulator was separated from the center of volume coil by 11 cm to lay outside the volume coil’s cylindrical edge, signal acquisition by passive inductive coupling produced small observable signal. (b) When the modulator was activated by pumping signal at a power level ~0.1 dBm below its oscillation threshold, the locally detected MR signals were significantly amplified, producing a clear image with sharp intensity profile. (The intensity profile in the right most column was plotted against the yellow dashed line in MR images.) (c) When the modulator was activated by a pumping signal above its oscillation threshold, locally detected MR signals were frequency encoded onto the oscillation signal that could more effectively overcome transmission attenuation, leading to MR images with even higher SNR. All SNR values shown in the right most column were referenced against the SNR of a directly connected surface coil.

To estimate the modulator’s effective operation range, we positioned the modulator in the center of magnet to detect the gel phantom in the same location with the same static magnetic field, while displacing the volume coil away from the magnet’s center. This would enable us to systematically vary the transmission attenuation between the modulator and the volume coil without modifying their circuit dimensions. (For each displacement, the transmission attenuation was estimated as the reciprocal of relative SNR of passive inductive coupling with respect to a wire-connected surface coil. Subsequently, the excitation power on the volume coil was proportionally scaled by the transmission attenuation to maintain the same effective flip angle.) As shown in Fig. 10, both the amplifier and the modulator had similar SNR when the attenuation was smaller than 21 dB. But the modulator could maintain constant sensitivity for attenuation up to 34 dB when the modulator already laid outside the edge of the volume coil. Even though the modulator’s sensitivity would eventually drop under very large distance separations, its SNR always remained higher than the amplifier, demonstrating the advantage of FM encoding of MRI signals to overcome larger transmission attenuation.

The effectiveness of the modulator can also be evaluated by its flatter image baseline along the horizontal direction. As shown in Fig. 12a, for 7.5-cm displacement that was 7.5-fold the modulator’s largest dimension, the corresponding transmission attenuation of 25-dB could be effectively compensated by the 33-dB gain provided by regenerative amplification, leading to uneven baseline that was characteristic of a high-gain regenerative amplifier with a narrowed bandwidth. Meanwhile, the image reconstructed from the modulator had a flatter baseline (Fig. 12b) when the internal noise from the modulator dominated over external noise from the volume coil, making frequency modulated detection the preferred choice. For the 15-cm displacement that was 15-fold the modulator’s largest dimension, the transmission attenuation was much larger than what regenerative amplification could overcome, leading to a flatter image baseline (Fig. 12c) that was primarily determined by the external noise. Meanwhile, the image reconstructed from the modulator had uneven baseline, due to the frequency dependent transfer function of the external noise as described in Eq. (14). Therefore, frequency modulation would provide flatter image baseline than regenerative amplification within a tolerable range of transmission attenuation. Outside this range, observation of a tilted baseline in the reconstructed image could be practically used as an indicator of too large a transmission attenuation for the frequency modulator to completely overcome.

FIGURE 12.

FIGURE 12.

(a, c) Horizontal baselines of images acquired in the presence of regenerative amplification. (b, d) Horizontal baselines of images reconstructed from frequency modulated oscillation signal. For the 7.5-cm distance separation, the required amplifier gain to overcome transmission attenuation would make the image baseline uneven (in Fig. a) due to the amplifier’s narrowed bandwidth. Meanwhile, the baseline of the FM-demodulated image was flatter (in Fig. b) due to the modulator’s frequency invariant response to internal noise. For the 15-cm distance separation, the baseline of the amplified image was almost flat (in Fig. c), due to the predominance of external noise originating from the volume coil. Meanwhile, the baseline of the FM-demodulated image was uneven (in Fig. d), indicating too large a transmission attenuation for the frequency modulator to overcome.

V. DISCUSSION

In this work, we have fabricated a triple frequency parametric oscillator that can utilize wireless pumping power provided by an activation antenna to directly encode locally detected RF signals onto its oscillation signal as the modulated carrier wave. The modulator consists of a double frequency parametric resonator with dipole and butterfly modes to sustain circuit oscillation, as well as a coupled single frequency resonator to concentrate the local pumping field. Unlike voltage-controlled oscillators that can only encode signals much slower than circuit oscillation, the parametric oscillator can combine down conversion and frequency encoding into a single stage, thus directly encoding RF inputs as frequency modulated oscillation signals for detection over larger distance separations. As predicted by Eqns. (11) and (12), the modulated oscillation signal observed by the remotely coupled receiver coil can be visualized in the frequency domain as summation of equally spaced sidebands distributed around the center carrier (Fig. 6); and the relative height of the first sideband with respect to the center peak scales proportionally to the RF input voltage ξd. Alternatively, the input signal can be retrieved by derivatizing the modulated phase angle of the oscillation signal in the time domain (Fig. 7), leading to a constant magnitude response of decoded signals over a range of input signal offsets (Fig. 9).

Although parametric oscillators have been previously utilized for quasi-static signal sensing [5], [24], [30], this work has for the first time demonstrated the circuit’s built-in capability for simultaneous down-conversion and frequency-encoding of RF signals. Compared to on-board digitization, FM-encoded oscillation is easier to implement using a compact modulator. Because of its design simplicity, this wirelessly powered modulator is amenable to miniaturization, making itself useful for chronic operation inside confined body cavities. The modulator is a low-power device: it can be wirelessly activated by about 9-dBm of power on the activation antenna separated by a distance that is 4-fold the modulator’s largest dimension. This level of pumping power is 20 dB lower than that required for a parametric resonator in its stand-along configuration, owing to the on-resonance concentration of magnetic flux at the pumping frequency. Further reduction in the required level of pumping power would benefit from the use of loop antennas and impedance matching network that can generate larger magnetic flux per unit current [31], or benefit from the exploration of midfield power transfer schemes [32].

To demonstrate effective MRI signal transmission under weak coupling conditions, we have systematically varied the modulator’s distance separation from the external receiver coil without modifying their circuit dimensions. This can be achieved by keeping the modulator in the magnet’s center while gradually displacing the volume coil away from the center, thus emulating increasingly large transmission attenuation. For each displacement, the overall sensitivity is affected by both the “internal” noise and the “external” noise. The internal noise comes from the dissipation current in the gel phantom and the modulator itself, while the external noise originates from the volume coil. When the modulator’s oscillation current induces a voltage A that is much larger than the noise voltage on the external volume coil, the internal noise will dominate and modulate the oscillation frequency in an offset independent manner as predicted by Eq. (13). This flat frequency response will lead to flat baselines in the reconstructed image (Fig. 12b). However, when the oscillation-induced voltage A is no longer dominant over the external noise voltage from the receiver coil, the baseline of the reconstructed images will become uneven (Fig. 12d), because Fourier transform of the time-domain derivative is equivalent to frequency domain multiplication of input signal offset as described by Eq. (14). Therefore, observation of flat baseline could be practically used to evaluate the tolerable range of transmission attenuation within which constant sensitivity can be maintained. According to the distant-dependent sensitivity plot in Fig. 11, the wireless modulator can maintain constant sensitivity even when the transmission attenuation is up to 34 dB, which is 13 dB larger than what is tolerable for regenerative amplification. To flatten the image baseline for larger transmission attenuations, it is possible to incorporate pre-emphasis filtering [33] into the parametric oscillator, or to use a chain of parametric oscillators that can frequency encode relayed signals multiple times during the wireless transmission pathway [34]. More work is going on along this direction.

FIGURE 11.

FIGURE 11.

When the relative SNR of MR images was measured at different distance separations, the amplified resonator (gray) could retain constant sensitivity for transmission attenuation up to 21 dB (at 4.5 cm separation). In comparison, the frequency modulator (yellow) could retain constant sensitivity for transmission attenuation up to 34 dB (at 11 cm separation).

In Summary, we have demonstrated the simultaneous down-conversion and frequency-encoding capability of an antenna powered parametric oscillator. This oscillator can mix the input RF signal with its own oscillation signal and modulate the oscillation signal at the offset frequency. Without the need for DC power source or multi-stage signal processors, this compact circuit can simultaneously detect, encode and broadcast locally acquired MR signals over larger distance separations, using tiny amount of wireless power. This wireless modulator can be easily inserted into body cavities or placed on skin surfaces to sensitively detect focal regions of interest in close proximity. Its broadcasting capability will also enable the modulator to wirelessly communicate with any conventional detectors that are readily available on ordinary MRI scanners, thus greatly improving MRI’s applicability for any parts of body without the need for dedicated focal coils using specialized signal interfaces. Although frequency encoding of MR signals is somewhat noisier than a directly connected surface coil, the greatly improved flexibility of this wireless modulator will justify its use inside confined space or body cavities. Owing to the better noise immunity of FM encoding, this modulator can potentially be adapted to travelling wave detection [35] where the cylindrical magnet bore can be advantageously used as a waveguide for longer range signal transmission.

Acknowledgments

This work was supported in part by the National Institutes of Health under Grant R00EB016753 and Grant 1RF1NS113278.

Biographies

graphic file with name nihms-1680236-b0013.gif

WEI QIAN (Member, IEEE) received the B.S. and M.S. degrees from Wuhan University, China, in 2003 and 2006, respectively, and the Ph.D. degree from Michigan State University, East Lansing, MI, USA, in 2011. She is currently an Assistant Professor with Michigan State University. Her research interests include power electronics and energy harvesting.

graphic file with name nihms-1680236-b0014.gif

XIN YU received the B.S. degree from Wuhan University, China, in 2000, the M.S. degree from Seton Hall University, in 2002, and the Ph.D. degree from New York University, in 2007. He is currently an Assistant Professor with the Massachusetts General Hospital, focusing on multi-modal functional MRI.

graphic file with name nihms-1680236-b0015.gif

CHUNQI QIAN received the B.S. degree from Nanjing University, China, in 2002, and the Ph.D. degree from the University of California at Berkeley, Berkeley, CA, USA, in 2007. He is currently a Principle Investigator with Michigan State University, focusing on the development of wireless sensors for physiologic monitoring and medical imaging.

Footnotes

The associate editor coordinating the review of this manuscript and approving it for publication was Abdul Halim Miah.

REFERENCES

  • [1].Horton BE, Schweitzer S, DeRouin AJ, and Ong KG, “A varactor-based, inductively coupled wireless pH sensor,” IEEE Sensors J, vol. 11, no. 4, pp. 1061–1066, April. 2011. [Google Scholar]
  • [2].Drazan JF, Abdoun OT, Wassick MT, Dahle R, Beardslee L, Marcus GA, Cady NC, and Ledet EH, “Simple implantable wireless sensor platform to measure pressure and force,” Med. Eng. Phys, vol. 59, pp. 81–87, September. 2018. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [3].Riistama J, Aittokallio E, Verho J, and Lekkala J, “Totally passive wireless biopotential measurement sensor by utilizing inductively coupled resonance circuits,” Sens. Actuators A, Phys, vol. 157, no. 2, pp. 313–321, February. 2010. [Google Scholar]
  • [4].Sileo L, Martiradonna L, Arcuti P, Monti G, Tasco V, Dal Maschio M, Pruzzo G, Bozzini B, Tarricone L, and De Vittorio M, “Wireless system for biological signal recording with gallium arsenide high electron mobility transistors as sensing elements,” Microelectronic Eng, vol. 111, pp. 354–359, November. 2013. [Google Scholar]
  • [5].Qian W and Qian C, “Frequency modulated parametric oscillation for antenna powered wireless transmission of voltage sensing signals,” IEEE Trans. Biomed. Circuits Syst, vol. 13, no. 6, pp. 1783–1791, December. 2019. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [6].Wei J, Liu Z, Chai Z, Yuan J, Lian J, and Shen GX, “A realization of digital wireless transmission for MRI signals based on 802.11 b,” J. Magn. Reson, vol. 186, pp. 63–358, June. 2007. [DOI] [PubMed] [Google Scholar]
  • [7].Aggarwal K, Joshi KR, Rajavi Y, Taghivand M, Pauly JM, Poon ASY, and Scott G, “A millimeter-wave digital link for wire-less MRI,” IEEE Trans. Med. Imag, vol. 36, no. 2, pp. 574–583, February. 2017. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [8].Sporrer B, Wu L, Bettini L, Vogt C, Reber J, Marjanovic J, Burger T, Brunner DO, Pruessmann KP, Troster G, and Huang Q, “A fully integrated dual-channel on-coil CMOS receiver for array coils in 1.5–10.5 t MRI,” IEEE Trans. Biomed. Circuits Syst, vol. 11, no. 6, pp. 1245–1255, December. 2017. [DOI] [PubMed] [Google Scholar]
  • [9].Memis OG, Eryaman Y, Aytur O, and Atalar E, “Miniaturized fiber-optic transmission system for MRI signals,” Magn. Reson. Med, vol. 59, pp. 73–165, January. 2008. [DOI] [PubMed] [Google Scholar]
  • [10].Takeda K, Nagasaka K, Noguchi A, Yamazaki R, Nakamura Y, Iwase E, Taylor JM, and Usami K, “Electro-mechano-optical detection of nuclear magnetic resonance,” Optica, vol. 5, pp. 152–158, February. 2018. [Google Scholar]
  • [11].Simonsen A, Sánchez-Heredia JD, Saarinen SA, Ardenkj®r-Larsen JH, Schliesser A, and Polzik ES, “Magnetic resonance imaging with optical preamplification and detection,” Sci. Rep, vol. 9, no. 1, pp. 1–7, December. 2019. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [12].Tominaga Y, Nagasaka K, Usami K, and Takeda K, “Studies on NMR-signal up-conversion from radio-frequency to optical regimes using a lightweight nanomembrane transducer,” J. Magn. Reson, vol. 298, pp. 6–15, January. 2019. [DOI] [PubMed] [Google Scholar]
  • [13].Ilchenko VS, Savchenkov AA, Matsko AB, and Maleki L, “Whispering-gallery-mode electro-optic modulator and photonic microwave receiver,” J. Opt. Soc. Amer. B, Opt. Phys, vol. 20, pp. 333–342, February. 2003. [Google Scholar]
  • [14].Gaborit G, Dahdah J, Duvillaret L, Sablong R, and Perrier AL, “Linear electro-optic effect for nuclear magnetic resonance coil,” Opt. Sens. Detection III, vol. 9141, May 2014, Art. no. 91411J. [Google Scholar]
  • [15].Seo D, Carmena JM, Rabaey JM, Maharbiz MM, and Alon E, “Model validation of untethered, ultrasonic neural dust motes for cortical recording,” J. Neurosci. Methods, vol. 244, pp. 22–114, April. 2015. [DOI] [PubMed] [Google Scholar]
  • [16].Schnall MD, Barlow C, Subramanian VH, and Leigh JS, “Wireless implanted magnetic resonance probes for in vivo NMR,” J. Magn. Reson, vol. 68, no. 1, pp. 161–167, June. 1986. [Google Scholar]
  • [17].Wirth ED, Mareci TH, Beck BL, Fitzsimmons JR, and Reier PJ, “A comparison of an inductively coupled implanted coil with optimized surface coils for in vivo NMR imaging of the spinal cord,” Magn. Reson. Med, vol. 30, pp. 33–626, November. 1993. [DOI] [PubMed] [Google Scholar]
  • [18].Ginefri J-C, Rubin A, Tatoulian M, Woytasik M, Boumezbeur F, Djemaï B, Poirier-Quinot M, Lethimonnier F, Darrasse L, and Dufour-Gergam E, “Implanted, inductively-coupled, radiofrequency coils fabricated on flexible polymeric material: Application to in vivo rat brain MRI at 7T,” J. Magn. Reson, vol. 224, pp. 61–70, November. 2012. [DOI] [PubMed] [Google Scholar]
  • [19].Riffe MJ, Yutzy SR, Jiang Y, Twieg MD, Blumenthal CJ, Hsu DP, Pan L, Gilson WD, Sunshine JL, Flask CA, Duerk JL, Nakamoto D, Gulani V, and Griswold MA, “Device localization and dynamic scan plane selection using a wireless magnetic resonance imaging detector array,” Magn. Reson. Med, vol. 71, no. 6, pp. 2243–2249, June. 2014. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [20].Quick HH, Zenge MO, Kuehl H, Kaiser G, Aker S, Massing S, Bosk S, and Ladd ME, “Interventional magnetic resonance angiography with no strings attached: Wireless active catheter visualization,” Magn. Reson. Med, vol. 53, no. 2, pp. 446–455, February. 2005. [DOI] [PubMed] [Google Scholar]
  • [21].Jordan CD, Thorne BRH, Wadhwa A, Losey AD, Ozhinsky E, Kondapavulur S, Fratello V, Moore T, Stillson C, Yee C, Watkins RD, Scott GC, Martin AJ, Zhang X, Wilson MW, and Hetts SW, “Wireless resonant circuits printed using aerosol jet deposition for MRI catheter tracking,” IEEE Trans. Biomed. Eng, vol. 67, no. 3, pp. 876–882, March. 2020. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [22].Munir B and Dyo V, “On the impact of mobility on battery-less RF energy harvesting system performance,” Sensors, vol. 18, no. 11, p. 3597, October. 2018. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [23].Wang L, Ma J, Huang Y, Tang D, and Huang Q-A, “Passive wireless hermetic environment monitoring system for spray painting workshop,” Sensors, vol. 16, no. 8, p. 1207, August. 2016. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [24].Qian W and Qian C, “Wirelessly powered signal regeneration to improve the remote detectability of an inductive pressure sensor,” IEEE Trans. Biomed. Circuits Syst, vol. 13, no. 5, pp. 1011–1020, October. 2019. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [25].Collin RE, Foundations for Microwave Engineering. New York, NY, USA: McGraw-Hill, 1966, pp. 541–570. [Google Scholar]
  • [26].Yu X, He Y, Wang M, Merkle H, Dodd SJ, Silva AC, and Koretsky AP, “Sensory and optogenetically driven single-vessel fMRI,” Nature Methods, vol. 13, pp. 40–337, April. 2016. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [27].He Y, Wang M, Chen X, Pohmann R, Polimeni JR, Scheffler K, Rosen BR, Kleinfeld D, and Yu X, “Ultra-slow single-vessel BOLD and CBV-based fMRI spatiotemporal dynamics and their correlation with neuronal intracellular calcium signals,” Neuron, vol. 97, p. 925–939, February. 2018. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [28].He Y, Wang M, and Yu X, “High spatiotemporal vessel-specific hemodynamic mapping with multi-echo single-vessel fMRI,” J. Cerebral Blood Flow Metabolism, vol. 40, no. 10, pp. 2098–2114, November. 2019, Art. no. 271678X19886240. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [29].Wang M, He Y, Sejnowski TJ, and Yu X, “Brain-state dependent astrocytic Ca2+ signals are coupled to both positive and negative BOLD-fMRI signals,” Proc. Nat. Acad. Sci. USA, vol. 115, no. 7, pp. E1647–E1656, February. 2018. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [30].Pais-Roldán P, Takahashi K, Sobczak F, Chen Y, Zhao X, Zeng H, Jiang Y, and Yu X, “Indexing brain state-dependent pupil dynamics with simultaneous fMRI and optical fiber calcium recording,” Proc. Nat. Acad. Sci. USA, vol. 117, no. 12, pp. 6875–6882, March. 2020. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [31].Yu X, Qian C, Chen DY, Dodd SJ, and Koretsky AP, “Deciphering laminar-specific neural inputs with line-scanning fMRI,” Nature Methods, vol. 11, pp. 8–55, January. 2014. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [32].Zhu C, Fallah-Rad M, Lang M, Schellekens W, Champagne K, and Petropoulos L, “A novel multichannel wireless receive phased array coil without integrated preamplifiers for high field MR imaging applications,” in Proc. Int. Soc. Magn. Reson. Med, vol. 20, 2012, p. 2788. [Google Scholar]
  • [33].Bulumulla SB, Fiveland E, Park KJ, Foo TK, and Hardy CJ, “Inductively coupled wireless RF coil arrays,” Magn. Reson. Imag, vol. 33, pp. 7–351, April. 2015. [DOI] [PubMed] [Google Scholar]
  • [34].Porter JR, Wright SM, and Famili N, “A four-channel time domain multiplexer: A cost-effective alternative to multiple receivers,” Magn. Reson. Med, vol. 32, no. 4, pp. 499–504, October. 1994. [DOI] [PubMed] [Google Scholar]
  • [35].Porter JR and Wright SM, “A sixteen-channel multiplexing upgrade for single channel receivers,” Magn. Reson. Imag, vol. 19, pp. 16–1009, September. 2001. [DOI] [PubMed] [Google Scholar]
  • [36].He W, Qin X, Jiejing R, and Gengying L, “Four-channel magnetic resonance imaging receiver using frequency domain multiplexing,” Rev. Sci. Instrum, vol. 78, no. 1, January. 2007, Art. no. 015102. [DOI] [PubMed] [Google Scholar]
  • [37].Riffe MJ, Twieg MD, Gudino N, Blumenthal CJ, Heilman JA, and Griswold MA, “Identification and mitigation of interference sources present in SSB-based wireless MRI receiver arrays,” Magn. Reson. Med, vol. 70, pp. 86–1775, December. 2013. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [38].Riffe MJ, Yutzy SR, Jiang Y, Twieg MD, Blumenthal CJ, Hsu DP, Pan L, Gilson WD, Sunshine JL, Flask CA, Duerk JL, Nakamoto D, Gulani V, and Griswold MA, “Device localization and dynamic scan plane selection using a wireless magnetic resonance imaging detector array,” Magn. Reson. Med, vol. 71, no. 6, pp. 2243–2249, June. 2014. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [39].Darnell D, Cuthbertson J, Robb F, Song AW, and Truong T, “Integrated radio-frequency/wireless coil design for simultaneous MR image acquisition and wireless communication,” Magn. Reson. Med, vol. 81, no. 3, pp. 2176–2183, March. 2019. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [40].Rosa EB and Cohen L, “Formule and tables for the calculation of mutual and self-inductance,” Bull. Bur. Standards, vol. 5, no. 1, pp. 51–55, 1911. [Google Scholar]

RESOURCES