Abstract
Background
Cytokine storm-related hypercoagulation may be important in the pathogenesis of stent thrombosis in patients with SARS-CoV-2. Whether stent polymers behave differently under such conditions has never been explored.
Methods
Fluorinated polymer-nanocoated and uncoated COBRA stents (CeloNova), BioLinx-polymer-coated Resolute Onyx stents (Medtronic), and Synergy stents (Boston Scientific), which are abluminally coated with a bioabsorbable polymer, were exposed to human blood from healthy donors which was supplemented with 400 pg/mL IL-6 and 100 pg/mL TNF-α, similar to what is seen in cytokine storm caused by SARS-CoV-2. Platelet adhesion and neutrophil activation, assessed by immunofluorescence, were compared under cytokine storm and control conditions (untreated blood) (n = 4 experimental runs).
Results
Platelet adhesion values, defined as %platelet-covered area x staining intensity, were significantly lower in coated and uncoated COBRA and in Resolute Onyx than in Synergy under control conditions (1.28 × 107 ± 0.43 × 107 vs. 2.92 × 107 ± 0.49 × 107 vs. 3.57 × 107 ± 0.73 × 107 vs. 9.94 × 107 ± 0.99 × 107; p ≤0.0001). In cytokine storm, platelet adhesion values remained low in coated COBRA-PzF (1.78 × 107 ± 0.38 × 107) compared to all other devices (uncoated COBRA: 5.92 × 107 ± 0.96 × 107; Resolute Onyx: 7.27 × 107 ± 1.82 × 107; Synergy: 11.28 × 107 ± 1.08 × 107; p ≤ 0.0001). Although cytokine storm conditions significantly increased neutrophil activation in all stents, it was significantly less in coated and uncoated COBRA, and in Resolute Onyx than in Synergy.
Conclusions
Blood-biomaterials interactions may determine the thrombogenic potential of stents. Under simulated cytokine storm conditions, fluoropolymer-coated stents showed the most favorable anti-thrombogenic and anti-inflammatory properties.
Abbreviations: SARS-CoV-2, severe acute respiratory syndrome coronavirus 2; PCI, percutaneous coronary intervention; ST, stent thrombosis; PzF, poly-bis(trifluoroethoxy)phosphazene or polyzene-F; MI, myocardial infarction; STEMI, ST-elevation myocardial infarction; IL-6, interleukin 6; TNF-α, tumor necrosis factor alpha; PBS, phosphate buffer saline; MPO, myeloperoxidase; CAC, COVID-19-associated coagulopathy; CoCr, cobalt‑chromium; PtCr, platinum‑chromium; EES, everolimus-eluting stent; DES, drug-eluting stent
Keywords: Stent thrombosis, Myocardial infarction, COVID-19, Cytokine storm, Platelets, Neutrophils
Graphic abstract
Severe COVID-19 infection is associated with overproduction of cytokines by macrophages, activated T-cells, and neutrophils. Especially IL-6 and TNF-α trigger hypercoagulation, potentially playing a role in the pathogenesis of MI. PCI in this situation holds an increased risk of ST. Under in vitro simulated cytokine storm conditions, stents with fluoropolymer coatings had less platelet adhesion and neutrophil activation than uncoated and BioLinx polymer-coated stents and compared with stents abluminally coated with bioabsorbable polymers.

1. Introduction
The severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) pandemic has resulted in considerable morbidity and mortality throughout the world. 20%–30% of patients with this coronavirus disease 2019 (COVID-19) have evidence of cardiac injury defined as decline in ejection fraction or troponin I elevation, which is associated with an even higher mortality [1]. Severe SARS-CoV-2 infection is associated with overproduction of cytokines (“cytokine storm”). Notably interleukin-6 (IL-6) and tumor necrosis factor alpha (TNF-α) surge during the illness and decline during recovery [2]. Both IL-6 and TNF-α have been implicated in promoting overexpression of tissue factor in platelets and macrophages [3,4], establishing a procoagulant shift in the hemostatic balance and promoting fibrin generation in severe inflammatory states.
Cytokine storm associated with SARS-CoV-2 causes hypercoagulation with excess risk of thrombotic events [5]. Patients presenting with ST-segment elevation myocardial infarction (STEMI) and concurrent COVID-19 infection had higher rates of multi-vessel thrombosis and stent thrombosis (ST) versus non-infected patients [6,7]. Despite its clinical relevance, the performance of different stent biomaterials has never been explored in the setting of COVID-19 infection.
Here, we present cases from our stent database with and without COVID-19 infection who underwent stent implantation for STEMI and compare signs of inflammation and thrombosis in these stents. To evaluate which biomaterial has the least thrombogenicity and might be more beneficial for stent implantation during COVID-19, we simulated cytokine storm conditions in vitro and examined compared platelet and neutrophil adhesion to stents of different biomaterials.
2. Material and methods
2.1. Data availability
The data that support the findings of this study are available from the corresponding author upon reasonable request.
2.2. Selection and histological processing of human stented coronary arteries
We searched the CVPath human coronary stent database for patients with and without confirmed COVID-19 infection who underwent stent implantation within 48 h before death. Only stents of equal type were included.
For histological staining, stented arteries were processed as described previously [8]. Briefly, the stented artery segments were fixed in formalin, dehydrated in a graded series of ethanol, and embedded in methylmethacrylate polymer. Segments of 2- to 3 mm thickness were sawed from each stent, and cross-sections of 4- to 6 μm thickness were cut from each of the segments on a Leica RM2155 rotary microtome equipped with a tungsten carbide blade, mounted on slides, and stained with H&E and Movat pentachrome.
2.3. In vitro experiments
We tested the relative thromboresistance of stents coated with a fluorinated polymer (i.e. polyzene-F in COBRA-PzF (CeloNova, Carlsbad, CA)) versus the BioLinx polymer (i.e. C10, C19 and polyvinyl-pyrrolidinone polymers in Resolute Onyx (Medtronic, Minneapolis, MN)), and versus a bioabsorbable polymer (i.e. polylactic-co-glycolic acid in Synergy (Boston Scientific, Marlborough, MA)) (Table 1 ). Human blood was treated with TNF-α and IL-6 at levels consistent with what is seen in severe COVID-19 infections. Untreated blood served as control. The blood was circulated in a flow loop model, and stents were examined using confocal microscopy.
Table 1.
Stent characteristics.
| Coated COBRA-PzF | Uncoated COBRA | Resolute Onyx | Synergy | |
|---|---|---|---|---|
| Manufacturer | CeloNova | CeloNova | Medtronic | Boston Scientific |
| Stent material | L-605 cobalt‑chromium CoCr alloy | L-605 cobalt‑chromium CoCr alloy | Platinum‑iridium alloy core and cobalt‑chromium alloy shell | Platinum‑chromium PtCr alloy |
| Polymer coating | Polyzene-F [poly-bis(trifluoroethoxy) phosphazene] | None | BioLinx® (C10, C19, and polyvinyl-pyrrolidinone polymer blend) | Synchrony™ PLGA [poly(DL-lactide-co-glycolide)]; abluminally only |
| Strut thickness | 71 μm | 71 μm | 81 μm | 79 μm |
| Polymer thickness | 0.050 μm | – | 5.6 μm | 4 μm |
| Drug component | – | – | Zotarolimus | Everolimus |
2.4. Silicone tube flow loop and stent deployment
Flow loops were assembled from silicone tubing (Cole-Palmer), connected via hose barb connectors (Cole-Parmer) (Fig. 1 ). Coronary stents were deployed within the tubing based on the manufacturer's suggested nominal pressure to obtain a stent-to-tube ratio of 1.1:1.0. Coated and uncoated COBRA stents were compared with Resolute Onyx stents (n = 4 experimental runs) and with Synergy stents (n = 4 experimental runs). Each flow loop contained only one stent at a time, so one experimental run comparing 3 different types of stents (coated COBRA-PzF vs. uncoated COBRA vs. Resolute Onyx and coated COBRA-PzF vs. uncoated COBRA vs. Synergy) comprised 6 flow loops in a side-by-side experimental design (control vs. cytokine storm conditions, for each stent).
Fig. 1.
Experimental setup. A) Each flow loop was assembled from silicone tubing. B) Whole blood was circulated for 60 min using a perfusion pump. C) Enlarged view on the stents deployed in silicone tubes.
2.5. Preparation of human whole blood
Ethical approval was obtained by the Institutional Review Board at CVPath Institute. Whole blood (60 mL) was collected by venipuncture from healthy volunteers, none of whom was on anti-platelet therapy or other regular prescription. Blood was mixed with sodium citrate to obtain a final concentration of 0.32%. To simulate cytokine storm conditions, blood was supplemented with 400 pg/mL IL-6 (PeproTech) and 100 pg/mL TNF-α (Sigma Aldrich). Cytokine concentrations were chosen in accordance with the highest blood levels of severe COVID-19 cases that have been published [[9], [10], [11]], rounded up to the nearest 100. Each flow loop was filled with whole blood and connected to a perfusion pump (ISMATEC). Blood was circulated at a flow rate of 35 mL/min for 60 min at 37 °C, 5% CO2 (Fig. 1).
2.6. Sample fixation
Stented tubing pieces were washed in PBS and immersion-fixed in Zamboni's fixative (American MasterTech Scientific, Inc.) for 20 min. Stents were bisected longitudinally and placed in 15% sucrose in PBS at 4 °C overnight.
2.7. Immunofluorescence staining
One half of each stent was immunostained for adherent platelets using antibodies against CD42b (abcam) and CD61 (Immunotech). Staining against myeloperoxidase (MPO), which is released by activated neutrophils [12], was performed in the other halves of each stent (Dako). The antibodies were visualized by secondary antibodies conjugated to an Alexa Fluor® 555 fluorophore (both Invitrogen). The nuclear counterstain was DAPI (Invitrogen).
2.8. Image quantification
After immunostaining, stents were mounted ‘en face’ on glass slides and coverslipped using aqueous mounting media. For platelet immunofluorescence, the entire surface of the stent was scanned (10× objective) (Zeiss, LSM 880, Zen Black software, 2.3 SP1). Representative high-power images were taken of 3 randomly selected struts from the proximal, mid, and distal part of each stent half using a 20× objective. Exported representative 20× JPG images were imported into Nikon NIS-Elements Advanced Research analysis software. To correct for possible photobleaching and other confocal error sources, image parameters were normalized by correcting the maximum pixel intensity value present to 255, and the average background value to 0. A stock noise reduction filter was applied uniformly across all images. A general analysis tool was created in NIS-Elements which applies a binary layer to the red channel, to pixels in the intensity range 40 to 255. The general analysis calculates the total pixel area with a red channel signal above 40 and calculates the sum of the intensities-per-pixel of every pixel covered by that binary. Total strut area was calculated, and the area fraction covered by platelets or neutrophils was calculated by dividing coverage area by total strut area. This area fraction was multiplied by the summed intensity to generate an intensity per coverage area metric (Coverage Area Fraction*Sum Red Channel Intensity), referred to as “Platelet Adhesion Value” and “MPO Quantification Value”. Representative images with obvious and un-correctable imaging defects such as visible flaking, photobleaching, or out-of-plane Z axis artifacts were excluded from quantitation.
2.9. Statistical analysis
Data are expressed as means ± SEM. The data were statistically analyzed using GraphPad Prism software (version 8.4.3). Shapiro-Wilk test was used to check normality. Overall group means were compared using one-way ANOVA, followed by Tukey's multiple comparisons test, and differences between cytokine storm and control conditions in each stent were compared using Student's t-test. A value of p ≤ 0.05 was considered statistically significant.
3. Results
3.1. Human autopsy cases of DES in COVID-19 and non-COVID-19 patients who died shortly after implantation
We consulted our database for autopsy cases with and without COVID-19 who underwent stent implantation for STEMI within 48 h before death and had patent stents at the time of death. Only stents of equal type and similar implantation duration were included. Our search yielded one 3.0 × 34 mm cobalt‑chromium (CoCr) everolimus-eluting stent (EES) implanted for STEMI in a patient with severe COVID-19 infection and three CoCr-EES (3.0 × 23 mm, 3.0 × 28 mm, and 3.0 × 27 mm) that were implanted in patients without COVID-19. While we observed unusually high adhesion of inflammatory cells and platelets to the stent struts, but no ST, in the patient with confirmed COVID-19 infection (Fig. 2 a), none of the patients without COVID-19 showed signs of platelet accumulation or severe inflammation around the stent struts (Fig. 2b–c). Although limited in number, these observations suggest COVID-19 infection may induce a pro-inflammatory and pro-thrombogenic environment. Prior reports suggested an increased risk of ST in COVID-19 patients presenting with STEMI [[13], [14], [15], [16], [17], [18], [19]], suggesting COVID-19 may affect vascular responses to stenting.
Fig. 2.
Histopathology of drug-eluting stents, implanted for acute myocardial infarction within 48 h before death. A) 3.0 × 34 mm CoCr-EES stent implanted in the left anterior descending artery of an 81-year-old man who presented with COVID-19 positivity and STEMI. Subject was provided with mechanical ventilator support but developed cardiogenic shock and expired 1 day after admission. Yellow lines indicate approximate levels of histologic sections; boxed areas show fibrin-platelet thrombi around stent struts. Note the abundance of inflammatory cells around the stent struts, which is unusually high compared with stents of similar type and implantation duration in patients without COVID-19 (B–D). B) 3.0 × 23 mm CoCr-EES, implanted in the proximal right coronary artery of a 32-year-old female who was diagnosed with acute inferior wall myocardial infarction due to dissection of the proximal right coronary artery. The patient became unresponsive during procedure and could not be resuscitated. Note the absence of inflammation around struts most of which are well apposed. C) 3.0 × 28 mm CoCr-EES stent, implanted in mid left anterior descending artery of a 76-year-old woman who presented with STEMI. The patient's condition deteriorated over the next two days with acute kidney injury and pulseless electric activity arrest. Sections show uncovered struts and focal areas of plaque disruption, but thrombi and inflammatory cells are practically absent. D) 3.0 × 27 mm CoCr-EES stent placed in the proximal right coronary artery of a 58-year-old male who died shortly after stenting procedure. The section shows plaque disruption and limited medial dissection but absence of inflammation. PLAD, proximal left anterior descending; MLAD, mid left anterior descending; PRC, proximal right coronary.
In order to explore this hypothesis, we evaluated the thrombogenic and inflammatory potential of different types of stents during a simulated cytokine storm induced by COVID-19 using a novel in vitro flow loop. To simulate cytokine storm conditions, blood was supplemented with 400 pg/mL IL-6 and 100 pg/mL TNF-α. These concentrations were chosen in accordance with the highest blood levels seen in severe COVID-19 cases that have been published [[9], [10], [11]]. Each flow loop was filled with whole blood and connected to a perfusion pump (ISMATEC). Blood was circulated at a flow rate of 35 mL/min for 60 min at 37 °C, 5% CO2 (Fig. 1). We compared the relative thromboresistance of fluorinated polymer-coated stents (COBRA-PzF) versus BioLinx polymer-coated stents (Resolute Onyx) and versus an abluminally bioabsorbable polymer-coated stent (Synergy) (Table 1).
3.2. Platelet adhesion assessment
First, platelet adhesion was analyzed under control vs. cytokine storm conditions in each stent (Supplemental Table 1). While cytokine storm conditions did not enhance platelet adhesion in coated COBRA-PzF, we observed significantly greater platelet adhesion under cytokine storm conditions compared with control in uncoated COBRA stents, and a trend towards higher platelet adhesion under cytokine storm conditions in Resolute Onyx. Platelet adhesion in Synergy did not differ from control under cytokine storm conditions and was higher than for all other stent tested regardless of condition.
Next, we compared all four devices under control conditions and under cytokine storm conditions, respectively (Fig. 3 ). Under control conditions, platelet adhesion was significantly lower in coated COBRA-PzF, uncoated COBRA, and Resolute Onyx than in Synergy. No differences were seen between Resolute Onyx and coated and uncoated COBRA. Under cytokine storm conditions, platelet adhesion remained low in coated COBRA-PzF, with significantly lower platelet adhesion values than in all other stents tested. Platelet adhesion in cytokine storm was similar in uncoated COBRA and Resolute Onyx, and – as opposed to control conditions - there was no difference between Resolute Onyx and Synergy under cytokine storm conditions anymore.
Fig. 3.
Platelet adhesion. Representative images of platelet adhesion, assessed by immunodetection of surface antigens CD61 and CD42b (red channel) under control (A) and cytokine storm conditions (B), derived from confocal microscopy (10× and 20× magnifications). A + C) Under control conditions, platelet adhesion was significantly lower in coated and uncoated COBRA and Resolute Onyx than in Synergy. B + C) In cytokine storm, platelet adhesion remained significantly lower in coated COBRA-PzF compared to all other devices evaluated.
3.3. Neutrophil activation assessment
We also evaluated neutrophil activation under control and cytokine storm conditions by staining against MPO (Supplemental Table 2). Cytokine storm enhanced neutrophil activation in coated and uncoated COBRA stents, and in Resolute Onyx, albeit to a modest level (compared to its effect on Synergy). In Synergy, we observed tendencies towards enhanced neutrophil activation under cytokine storm conditions compared with control, but without statistical significance.
When directly comparing the devices, neutrophil activation was significantly higher in Synergy versus coated and uncoated COBRA stents, and versus Resolute Onyx, both under control and under cytokine storm conditions (Fig. 4 ). There was no difference between the coated and uncoated version of COBRA. Enhanced cell adhesion to Synergy as compared to coated and uncoated COBRA and Resolute Onyx stents was confirmed by SEM imaging (Fig. 5 ).
Fig. 4.
Comparison of neutrophil activation in coated and uncoated COBRA, Resolute Onyx, and Synergy. A + B) Representative images derived from confocal microscopy (20× magnification). MPO deposition was very low in coated and uncoated COBRA, and lower in Resolute Onyx than in Synergy, both under control (A) and under cytokine storm conditions (B). C) MPO quantification values were significantly higher in Synergy versus coated and uncoated COBRA, and versus Resolute Onyx, both under control conditions and under cytokine storm conditions. There was no difference in neutrophil activation between the coated and uncoated version of COBRA.
Fig. 5.
Representative scanning electron microscopy images (15×, 200×, and 600× magnifications). A) Only a moderate number of cells is attached to coated COBRA-PzF under control conditions, and cell adhesion was not enhanced in cytokine storm (B). C) While a comparatively low number of cells adhered to uncoated COBRA under control conditions, cytokine storm significantly enhanced cell adhesion (D). E) Resolute Onyx stents were covered with a layer of plasma proteins, inflammatory cells, and platelets under control conditions, and cell adhesion was not further enhanced under cytokine storm conditions (F). G) Cell adhesion to Synergy was already high under control conditions and was not further enhanced by cytokine storm (H).
4. Discussion
The purpose of this study was to assess acute thrombogenicity and inflammatory potential of different stent biomaterials under control and COVID-19-simulated cytokine storm conditions in an in vitro flow loop using human whole blood.
4.1. Effect of cytokine storm on coagulation and thrombogenicity
Elevated cytokine levels have been found in patients with COVID-19-associated pneumonia [20]. Cytokine storm was simulated by the addition of IL-6 and TNF-α to human whole blood. To simulate a worst-case scenario, the dosages of 400 pg/mL IL-6 and 100 pg/mL TNF-α were chosen in accordance with the highest blood levels of these cytokines in patients with COVID-19 that have been published [[9], [10], [11]]. A study reporting data on 389 confirmed COVID-19 patients from Wuhan, China, reported IL-6 levels up to 160 pg/mL and TNFα levels up to 31 pg/mL even in patients with mild disease, while IL-6 rose to 300 pg/mL and TNFα went up to 41 pg/mL in patients with severe disease [10]. Likewise, IL-6 levels ranged between 68 and 333 pg/mL in a single-center study reporting clinical features of the first 50 COVID-19 patients admitted to a German tertiary hospital [11]. Another study reported serum IL-6 levels up to nearly 2000 pg/mL in 54 patients with COVID-19 infection and macrophage activation syndrome or immune dysregulation, while TNFα levels ranged between 8 and 37 pg/mL even in patients with only intermediate functional state of the immune system [9].
SARS-CoV-2 infections are associated with coagulopathy (“COVID 19-associated coagulopathy”, CAC), presenting with elevation of D-dimer and fibrin/fibrinogen-degradation products [21]. While the virus itself does not appear to intrinsically trigger coagulation, CAC is thought to derive from the profound inflammatory response [5] with excess production of pro-inflammatory cytokines contributing to the activation of coagulation [22]. Especially IL-6 and TNF-α are important mediators as they upregulate tissue factor on the cell surface [3,4], which initiates the extrinsic pathway of coagulation [23]. Furthermore, TNF-α suppresses endogenous anticoagulant pathways [4]. Hence, IL-6 and TNF-α establish a procoagulant shift in the hemostatic balance.
4.2. Relationship between COVID-19 and myocardial infarction
Patients with coronary artery disease (CAD) and those with cardiovascular risk factors are at increased risk for MI during infections [24]. Elevated troponin I levels have been found in 8%–28% of patients with COVID-19 [25,26]. With surging numbers of COVID-19 patients, and given the high prevalence of CAD, reasonable numbers of patients with combined STEMI and COVID-19 can be expected. Primary PCI remains the gold standard of care for STEMI patients during the COVID-19 pandemic [27]. However, the procoagulant state of COVID-19 patients may trigger acute ST. While the event of ST in STEMI patients who have tested positive for COVID-19 has been described in a number of case reports [[13], [14], [15], [16], [17], [18], [19]], only two larger multicenter studies have been published so far. The first multicenter study, published in July 2020, included 78 patients with STEMI and COVID-19. Stent thrombosis occurred in 4 out of 19 patients treated with primary PCI (21%) [28]. Another study compared in-hospital outcomes of 1010 consecutive STEMI patients with and without COVID-19 and reported a higher incidence of stent thrombosis in COVID-19 patients (3.3% vs. 0.8%; p = 0.020) [6].
Indeed, even though the case presented did not have a ST, we observed unusually high adhesion of inflammatory cells and platelets to the struts in a stent implanted for STEMI in a patient with severe COVID-19 infection, while inflammation was practically absent around struts of the same type of stent in patients without COVID-19. We acknowledge the fact that findings in a single patient can only be hypothesis generating, however, in conjunction with the increased risk of ST in clinical studies, we assume the use of a stent that is less thrombotic and less likely causing inflammatory reactions might be beneficial in a situation of STEMI and simultaneous COVID-19 infection, but this needs to be confirmed in clinical trials.
4.3. Polymer coatings affect stent thrombogenicity, inflammation, and endothelialization
Cytokine storm triggered platelet adhesion in uncoated COBRA stents, and in Biolinx-coated stents, while it was already high under control conditions in stents abluminally coated with bioabsorbable polymer. Furthermore, we observed enhanced neutrophil activation in all stents under cytokine storm conditions. These experiments confirmed our hypothesis that COVID-19-induced cytokine storm seems to affect the interaction of stents with blood. We saw more inflammation and platelet accumulation, consistent with our clinical case.
Fluoropolymers are known to have anti-thrombotic properties. Based on a CoCr alloy, COBRA-PzF has a nano-thin coating of poly-bis(trifluoroethoxy)phosphazene (Polyzene-F or PzF) which has been shown to preferentially adsorb albumin instead of coagulation-stimulating proteins [29]. Less platelet adhesion was observed in COBRA-PzF as compared to conventional drug-eluting stents (DES) in a pig shunt study [30], and clinical studies reported low incidences of ST and spontaneous MI [[31], [32], [33], [34]]. On the other hand, hydrophilic polymer surfaces, such as BioLinx in Resolute Onyx, have been suggested to have superior biocompatibility compared with hydrophobic polymer coatings, mostly because they do not induce monocyte adhesion [35]. Consisting of an outer shell of CoCr and a platinum‑iridium inner core, Resolute Onyx showed particularly low rates of ST [36]. Bioabsorbable polymers are often cited as an alternative to biostable polymers for DES coatings. Synergy is an EES on a platinum‑chromium (PtCr)-based platform with abluminal bioabsorbable polymer coating which demonstrated accelerated vascular healing and minimal inflammation in animal studies [37] and excellent outcomes in randomized trials [38].
When comparing these devices in our study, coated COBRA-PzF showed the most favorable anti-thrombogenic and anti-inflammatory properties, both under control and under cytokine storm conditions, whereas platelet adherence and neutrophil activation were highest in Synergy. While cytokine storm conditions enhanced platelet adhesion in uncoated COBRA and in Resolute Onyx, we did not observe significant differences between control and cytokine storm conditions in coated COBRA-PzF and in Synergy. Platelet adherence in Synergy, however, was already high under control conditions, and cytokine storm might not have had the ability to further increase it. In contrast, platelet adhesion was overall low in coated COBRA-PzF, suggesting that the anti-thrombogenic properties can be attributed to its PzF-nanocoating.
The question why even the uncoated COBRA stent had less platelet and neutrophil adhesion than the bare luminal surface of Synergy may be related to differential stent platforms. COBRA stents have thinner struts than Synergy (71 μm vs. 79 μm plus the 4 μm-abluminal polymer coating), which is an important determinant of thrombogenicity [39]. Furthermore, the metal alloy is different in COBRA and Synergy. While COBRA is manufactured from a CoCr alloy, PtCr is used in Synergy. Recent data from 7045 patients suggested that Synergy was associated with a higher risk of acute ST when compared with a CoCr-based EES (1.2% vs. 0.3%, p = 0.032) [40]. Of note, 14 of 15 acute ST events among those treated with Synergy, occurred in high-risk patients.
5. Limitations
Our study has several limitations. First, this is a pure benchwork study, and we did not confirm our findings in animals. Second, although cytokine storm is considered the major determinant for thrombogenicity in SARS-CoV-2 infections, we did not perform the experiments with blood from COVID-19-infected patients. Thus, we might have missed the potential impact of other blood components potentially playing a role in CAC. In addition, although IL-6 and TNF-α are believed to be the most important players in COVID-19 induced cytokine storm [41,42], we excluded other cytokines which may have additional effects on stent thrombogenicity. Third, we compared platelet and neutrophil adhesion in a select set of stents only, all of which, however, were specifically designed to reduce inflammation and to enhance vascular healing after PCI [35,43,44]. Fourth, stents were implanted in non-endothelialized silicone tubes, missing the important pro- or anticoagulant impact of endothelial cells. Considering these important limitations, the findings of our study can only be hypothesis generating, and not practice changing.
Furthermore, the histopathology comparison between stents in COVID-19-infected and non-COVID-19-infected patients did not allow for representative statistical evaluation because of the small sample size given the limited availability of stent samples obtained from autopsy of COVID-19 cases. Nevertheless, to the best of our knowledge, histopathology data from patients with acute stent implantation during COVID-19 infection have never been published before, and our findings emphasize the clinical relevance of evaluating thromboresistance of stents in a situation of cytokine storm.
6. Conclusions
Nanocoated COBRA-PzF showed the most favorable anti-thrombogenic and anti-inflammatory properties, both under control and under cytokine storm conditions, whereas platelet adherence and neutrophil activation were highest in Synergy. We assume the anti-thrombogenic properties can be attributed to the unique Polyzene-F nanocoating, preventing platelet adhesion even in an in vitro simulated cytokine storm.
Funding
This study was funded by CeloNova Biosciences, San Antonio, Tx, USA.
CRediT authorship contribution statement
Anne Cornelissen: Conceptualization, Methodology, Investigation, Writing – original draft. Matthew Kutyna: Formal analysis, Investigation. Qi Cheng: Visualization. Yu Sato: Methodology, Writing – review & editing. Rika Kawakami: Writing – review & editing. Atsushi Sakamoto: Writing – review & editing. Kenji Kawai: Writing – review & editing. Masayuki Mori: Writing – review & editing. Raquel Fernandez: Writing – review & editing. Liang Guo: Writing – review & editing. Dario Pellegrini: Resources, Investigation. Giulio Guagliumi: Resources, Investigation. Mark Barakat: Conceptualization, Resources, Investigation, Writing – review & editing. Renu Virmani: Writing – review & editing. Aloke Finn: Supervision, Writing – original draft.
Declaration of competing interest
R.V. and A.V.F. have received institutional research support from R01 HL141425 Leducq Foundation Grant; 480 Biomedical; 4C Medical; 4Tech; Abbott; Accumedical; Amgen; Biosensors; Boston Scientific; Cardiac Implants; CeloNova; Claret Medical; Concept Medical; Cook; CSI; DuNing, Inc.; Edwards LifeSciences; Emboline; Endotronix; Envision Scientific; Lutonix/Bard; Gateway; Lifetech; Limflo; MedAlliance; Medtronic; Mercator; Merill; Microport Medical; Microvention; Mitraalign; Mitra assist; NAMSA; Nanova; Neovasc; NIPRO; Novogate; Occlutech; OrbusNeich Medical; Phenox; Profusa; Protembis; Qool; Recor; Senseonics; Shockwave; Sinomed; Spectranetics; Surmodics; Symic; Vesper; W.L. Gore; Xeltis. A.V.F. has received honoraria from Abbott Vascular; Biosensors; Boston Scientific; CeloNova; Cook Medical; CSI; Lutonix Bard; Sinomed; Terumo Corporation; and is a consultant to Amgen; Abbott Vascular; Boston Scientific; CeloNova; Cook Medical; Lutonix Bard; Sinomed. R.V. has received honoraria from Abbott Vascular; Biosensors; Boston Scientific; Cook Medical; Cordis; CSI; Lutonix Bard; Medtronic; OrbusNeich Medical; CeloNova; SINO Medical Technology; ReCore; Terumo Corporation; W. L. Gore; Spectranetics; and is a consultant to Abbott Vascular; Boston Scientific; CeloNova; Cook Medical; Cordis; CSI; Edwards Lifescience; Lutonix Bard; Medtronic; OrbusNeich Medical; ReCore; Sinomedical Technology; Spectranetics; Surmodics; Terumo Corporation; W. L. Gore; Xeltis. GG has received institutional research grants from Abbott Vascular, Boston Scientific, and Infraredx and is a consultant to Abbott Vascular and Boston Scientific. M.B. is an employee of CeloNova BioSciences. The other authors declare no competing interests.
Footnotes
Supplementary data to this article can be found online at https://doi.org/10.1016/j.carrev.2021.03.023.
Appendix A. Supplementary data
Supplementary material
References
- 1.Mitrani R.D., Dabas N., Goldberger J.J. COVID-19 cardiac injury: implications for long-term surveillance and outcomes in survivors. Heart Rhythm. 2020;17:1984–1990. doi: 10.1016/j.hrthm.2020.06.026. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2.Chen G., Wu D., Guo W., Cao Y., Huang D., Wang H., et al. Clinical and immunological features of severe and moderate coronavirus disease 2019. J Clin Invest. 2020;130:2620–2629. doi: 10.1172/JCI137244. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3.Levi M., van der Poll T. Two-way interactions between inflammation and coagulation. Trends Cardiovasc Med. 2005;15:254–259. doi: 10.1016/j.tcm.2005.07.004. [DOI] [PubMed] [Google Scholar]
- 4.Zelová H., Hošek J. TNF-α signalling and inflammation: interactions between old acquaintances. Inflamm Res. 2013;62:641–651. doi: 10.1007/s00011-013-0633-0. [DOI] [PubMed] [Google Scholar]
- 5.Connors J.M., Levy J.H. COVID-19 and its implications for thrombosis and anticoagulation. Blood. 2020;135:2033–2040. doi: 10.1182/blood.2020006000. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 6.Rodriguez-Leor O., Cid Alvarez A.B., de Prado A.P., Rossello X., Ojeda S., Serrador A., et al. In-hospital outcomes of patients with ST-segment elevation myocardial infarction and COVID-19. EuroIntervention. 2020 doi: 10.4244/EIJ-D-20-00935. Online ahead of print. [DOI] [Google Scholar]
- 7.Choudry F.A., Hamshere S.M., Rathod K.S., Akhtar M.M., Archbold R.A., Guttmann O.P., et al. High thrombus burden in patients with COVID-19 presenting with ST-segment elevation myocardial infarction. J Am Coll Cardiol. 2020;76:1168–1176. doi: 10.1016/j.jacc.2020.07.022. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8.Yazdani S.K., Sheehy A., Nakano M., Nakazawa G., Vorpahl M., Otsuka F., et al. Preclinical evaluation of second-generation everolimus- and zotarolimus-eluting coronary stents. J Invasive Cardiol. 2013;25:383–390. [PubMed] [Google Scholar]
- 9.Giamarellos-Bourboulis E.J., Netea M.G., Rovina N., Akinosoglou K., Antoniadou A., Antonakos N., et al. Complex immune dysregulation in COVID-19 patients with severe respiratory failure. Cell Host Microbe. 2020;27:992–1000.e3. doi: 10.1016/j.chom.2020.04.009. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10.Hou H., Zhang B., Huang H., Luo Y., Wu S., Tang G., et al. Using IL-2R/lymphocytes for predicting the clinical progression of patients with COVID-19. Clin Exp Immunol. 2020;201:76–84. doi: 10.1111/cei.13450. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 11.Dreher M., Kersten A., Bickenbach J., Balfanz P., Hartmann B., Cornelissen C., et al. The characteristics of 50 hospitalized COVID-19 patients with and without ARDS. Dtsch Arztebl Int. 2020;117:271–278. doi: 10.3238/arztebl.2020.0271. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12.Winterbourn C.C., Kettle A.J., Hampton M.B. Reactive oxygen species and neutrophil function. Annu Rev Biochem. 2016;85:765–792. doi: 10.1146/annurev-biochem-060815-014442. [DOI] [PubMed] [Google Scholar]
- 13.Lacour T., Semaan C., Genet T., Ivanes F. Insights for increased risk of failed fibrinolytic therapy and stent thrombosis associated with COVID-19 in ST-segment elevation myocardial infarction patients. Catheter Cardiovasc Interv. 2021;97:E241–e3. doi: 10.1002/ccd.28948. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14.Ayan M., Kovelamudi S., Al-Hawwas M. Subacute stent thrombosis in a patient with COVID-19 pneumonia. Proc (Bayl Univ Med Cent) 2020;34:175–177. doi: 10.1080/08998280.2020.1812792. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 15.Prieto-Lobato A., Ramos-Martínez R., Vallejo-Calcerrada N., Corbí-Pascual M., Córdoba-Soriano J.G. A case series of stent thrombosis during the COVID-19 pandemic. JACC Case Rep. 2020;2:1291–1296. doi: 10.1016/j.jaccas.2020.05.024. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16.Choudhary R., Kaushik A., Sharma J.B. COVID-19 pandemic and stent thrombosis in a post percutaneous coronary intervention patient-a case report highlighting the selection of P2Y12 inhibitor. Cardiovasc Diagn Ther. 2020;10:898–901. doi: 10.21037/cdt-20-485. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 17.Galeazzi G.L., Loffi M., Di Tano G., Danzi G.B. Severe COVID-19 pneumonia and very late stent thrombosis: a trigger or innocent bystander? Korean Circ J. 2020;50:632–633. doi: 10.4070/kcj.2020.0166. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 18.Hinterseer M., Zens M., Wimmer R.J., Delladio S., Lederle S., Kupatt C., et al. Acute myocardial infarction due to coronary stent thrombosis in a symptomatic COVID-19 patient. Clin Res Cardiol. 2021;110:302–306. doi: 10.1007/s00392-020-01663-4. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 19.Seif S., Ayuna A., Kumar A., Macdonald J. Massive coronary thrombosis caused primary percutaneous coronary intervention to fail in a COVID-19 patient with ST-elevation myocardial infarction. Catheter Cardiovasc Interv. 2020 doi: 10.1002/ccd.29050. Online ahead of print. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 20.McGonagle D., Sharif K., O’Regan A., Bridgewood C. The role of cytokines including interleukin-6 in COVID-19 induced pneumonia and macrophage activation syndrome-like disease. Autoimmun Rev. 2020;19:102537. doi: 10.1016/j.autrev.2020.102537. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 21.Wang D., Hu B., Hu C., Zhu F., Liu X., Zhang J., et al. Clinical characteristics of 138 hospitalized patients with 2019 novel coronavirus-infected pneumonia in Wuhan, China. Jama. 2020;323:1061–1069. doi: 10.1001/jama.2020.1585. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 22.Magro G. COVID-19: review on latest available drugs and therapies against SARS-CoV-2. Coagulation and inflammation cross-talking. Virus Res. 2020;286 doi: 10.1016/j.virusres.2020.198070. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 23.Osterud B., Bjorklid E. Tissue factor in blood cells and endothelial cells. Front Biosci (Elite Ed) 2012;vol. 4:289–299. doi: 10.2741/e376. [DOI] [PubMed] [Google Scholar]
- 24.Kwong J.C., Schwartz K.L., Campitelli M.A., Chung H., Crowcroft N.S., Karnauchow T., et al. Acute myocardial infarction after laboratory-confirmed influenza infection. N Engl J Med. 2018;378:345–353. doi: 10.1056/NEJMoa1702090. [DOI] [PubMed] [Google Scholar]
- 25.Lippi G., Lavie C.J., Sanchis-Gomar F. Cardiac troponin I in patients with coronavirus disease 2019 (COVID-19): evidence from a meta-analysis. Prog Cardiovasc Dis. 2020;63:390–391. doi: 10.1016/j.pcad.2020.03.001. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 26.Clerkin K.J., Fried J.A., Raikhelkar J., Sayer G., Griffin J.M., Masoumi A., et al. COVID-19 and cardiovascular disease. Circulation. 2020;141:1648–1655. doi: 10.1161/CIRCULATIONAHA.120.046941. [DOI] [PubMed] [Google Scholar]
- 27.Mahmud E., Dauerman H.L., Welt F.G., Messenger J.C., Rao S.V., Grines C., et al. Management of acute myocardial infarction during the COVID-19 pandemic. J Am Coll Cardiol. 2020;76:1375–1384. doi: 10.1016/j.jacc.2020.04.039. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 28.Hamadeh A., Aldujeli A., Briedis K., Tecson K.M., Sanz-Sánchez J., Al Dujeili M., et al. Characteristics and outcomes in patients presenting with COVID-19 and ST-segment elevation myocardial infarction. Am J Cardiol. 2020;131:1–6. doi: 10.1016/j.amjcard.2020.06.063. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 29.Welle A., Grunze M., Tur D. Blood compatibility of poly [bis (trifluoroethoxy) phosphazene] JAMP. 2000;4:6–10. [Google Scholar]
- 30.Jinnouchi H., Mori H., Cheng Q., Kutyna M., Torii S., Sakamoto A., et al. Thromboresistance and functional healing in the COBRA PzF stent versus competitor DES: implications for dual antiplatelet therapy. EuroIntervention. 2019;15:e342–e353. doi: 10.4244/EIJ-D-18-00740. [DOI] [PubMed] [Google Scholar]
- 31.Maillard L., Tavildari A., Barra N., Billé J., Joly P., Peycher P., et al. Immediate and 1-year follow-up with the novel nanosurface modified COBRA PzF stent. Arch Cardiovasc Dis. 2017;110:682–688. doi: 10.1016/j.acvd.2017.04.010. [DOI] [PubMed] [Google Scholar]
- 32.Maillard L., Vochelet F., Peycher P., Ayari A., Barra N., Billé J., et al. MAPT (mono antiplatelet therapy) as regular regimen after COBRA PzFTM nanocoated coronary stent (NCS) implantation. Cardiovasc Revasc Med. 2020;21:785–789. doi: 10.1016/j.carrev.2019.10.007. [DOI] [PubMed] [Google Scholar]
- 33.Maillard L., de Labriolle A., Brasselet C., Faurie B., Durel N., de Poli F., et al. Evaluation of the safety and efficacy of the Cobra PzF NanoCoated coronary stent in routine, consecutive, prospective, and high-risk patients: the e-Cobra study. Catheter Cardiovasc Interv. 2020 doi: 10.1002/ccd.29065. Online ahead of print. [DOI] [PubMed] [Google Scholar]
- 34.Cutlip D.E., Garratt K.N., Novack V., Barakat M., Meraj P., Maillard L., et al. 9-month clinical and angiographic outcomes of the COBRA polyzene-F nanocoated coronary stent system. JACC Cardiovasc Interv. 2017;10:160–167. doi: 10.1016/j.jcin.2016.10.037. [DOI] [PubMed] [Google Scholar]
- 35.Hezi-Yamit A., Sullivan C., Wong J., David L., Chen M., Cheng P., et al. Impact of polymer hydrophilicity on biocompatibility: implication for DES polymer design. J Biomed Mater Res A. 2009;90:133–141. doi: 10.1002/jbm.a.32057. [DOI] [PubMed] [Google Scholar]
- 36.Blum M., Cao D., Mehran R. Device profile of the Resolute Onyx Zotarolimus eluting coronary stent system for the treatment of coronary artery disease: overview of its safety and efficacy. Expert Rev Med Devices. 2020;17:257–265. doi: 10.1080/17434440.2020.1736037. [DOI] [PubMed] [Google Scholar]
- 37.Nakazawa G., Torii S., Ijichi T., Nagamatsu H., Ohno Y., Kurata F., et al. Comparison of vascular responses following new-generation biodegradable and durable polymer-based drug-eluting stent implantation in an atherosclerotic rabbit iliac artery model. J Am Heart Assoc. 2016;5 doi: 10.1161/JAHA.116.003803. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 38.Meredith I.T., Verheye S., Dubois C., Dens J., Farah B., Carrié D., et al. Final five-year clinical outcomes in the EVOLVE trial: a randomised evaluation of a novel bioabsorbable polymer-coated, everolimus-eluting stent. EuroIntervention. 2018;13:2047–2050. doi: 10.4244/EIJ-D-17-00529. [DOI] [PubMed] [Google Scholar]
- 39.Lu S., Ng J., Ang H., Paradies V., Wong P.E., Al-Lamee R., et al. Is there light at the end of the thin-strut tunnel?: in vitro insights on strut thickness impact on thrombogenicity in bioresorbable stents or scaffolds. JACC Cardiovasc Interv. 2018;11:714–716. doi: 10.1016/j.jcin.2018.01.237. [DOI] [PubMed] [Google Scholar]
- 40.Zanchin C., Ueki Y., Zanchin T., Häner J., Otsuka T., Stortecky S., et al. Everolimus-eluting biodegradable polymer versus everolimus-eluting durable polymer stent for coronary revascularization in routine clinical practice. JACC Cardiovasc Interv. 2019;12:1665–1675. doi: 10.1016/j.jcin.2019.04.046. [DOI] [PubMed] [Google Scholar]
- 41.Broman N., Rantasärkkä K., Feuth T., Valtonen M., Waris M., Hohenthal U., et al. IL-6 and other biomarkers as predictors of severity in COVID-19. Ann Med. 2021;53:410–412. doi: 10.1080/07853890.2020.1840621. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 42.Que Y., Hu C., Wan K., Hu P., Wang R., Luo J., et al. Cytokine release syndrome in COVID-19: a major mechanism of morbidity and mortality. Int Rev Immunol. 2021:1–14. doi: 10.1080/08830185.2021.1884248. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 43.Koppara T., Cheng Q., Yahagi K., Mori H., Sanchez O.D., Feygin J., et al. Thrombogenicity and early vascular healing response in metallic biodegradable polymer-based and fully bioabsorbable drug-eluting stents. Circ Cardiovasc Interv. 2015;8 doi: 10.1161/CIRCINTERVENTIONS.115.002427. [DOI] [PubMed] [Google Scholar]
- 44.Shreenivas S.S., Kereiakes D.J. Evolution of the SYNERGY bioresorbable polymer metallic coronary stent. Future Cardiol. 2018;14:307–317. doi: 10.2217/fca-2018-0040. [DOI] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Supplementary Materials
Supplementary material
Data Availability Statement
The data that support the findings of this study are available from the corresponding author upon reasonable request.