Abstract.
Significance: Water content plays a vital role in the normally functioning visual system; even a minor disruption in the water balance may be harmful. Today, no direct method exists for corneal hydration assessment, while it could be instrumental in early diagnosis and control of a variety of eye diseases. The use of terahertz (THz) radiation, which is highly sensitive to water content, appears to be very promising.
Aim: To find out how THz scanning parameters of corneal tissue measured by an experimental setup, specially developed for in vivo contactless estimations of corneal reflectivity coefficient (RC), are related to pathological changes in the cornea caused by B-band ultraviolet (UVB) exposure.
Approach: The setup was tested on rabbit eyes in vivo. Prior to the course of UVB irradiation and 1, 5, and 30 days after it, a series of examinations of the corneal state was made. At the same time points, corneal hydration was assessed by measuring RC.
Results: The obtained data confirmed the negative impact of UVB irradiation course on the intensity of tear production and on the corneal thickness and optical parameters. A significant (1.8 times) increase in RC on the 5th day after the irradiation course, followed by a slight decrease on the 30th day after it was revealed. The RC increase measured 5 days after the UVB irradiation course generally corresponded to the increase (by a factor of 1.3) of tear production. RC increase occurred with the corneal edema, which was manifested by corneal thickening (by 18.2% in the middle area and 17.6% in corneal periphery) and an increased volume of corneal tissue (by 17.6%).
Conclusions: Our results demonstrate that the proposed approach can be used for in vivo contactless estimation of the reflectivity of rabbit cornea in the THz range and, thereby, of cornea hydration.
Keywords: cornea, terahertz radiation, UVB, optical coherent tomography, confocal microscopy, Scheimpflug imaging, rabbit eyes
1. Introduction
Water content plays a vital role in the normally functioning visual system; even a minor disruption in the water balance (by 10%) may be harmful.1 The majority of water in an organism is bound by connective tissues, which are able to retain water due to the presence of glycosaminoglycans in the intercellular matrix (bound water). The cornea as a connective eye tissue formation normally contains a significant amount (78%) of bound water mass.2 If cornea accumulates excessive water (bulk water, causing corneal edema), its refractive properties are changed.3 Corneal edema is a characteristic consequence of many eye pathologies, such as endothelial dystrophy, eye condition after cataract surgery or glaucoma filtrating surgery, trauma, ocular toxicity, or hypoxia.4,5 Especially important is the fact that the cornea loses transparency, which leads to significant loss of vision. On the other hand, a dehydrated cornea also changes its shape and refractive ability. If loss of water persists, the cornea is subject to a dystrophic process that causes irreversible disruption of visual function, in particular after eximer laser procedures such as corneal refractive surgeries.6–8 Water content in corneal stroma directly impacts on collagen extracellular matrix and therefore on corneal transparency, i.e., on its optical characteristics.9
In view of the above, adequate control and monitoring of corneal hydration is very important for early diagnosis and control of a variety of eye diseases, stating indications for and contraindications to keratorefractive surgeries, and the choice of effective and safe plans of local medication, including tear replacement and hypotensive therapy as well as the choice of contact lens correction regimens.10–12
Currently, corneal edema or dystrophy associated with reduced hydration can only be indirectly diagnosed by measuring the thickness of the cornea as a surrogate parameter for corneal edema (by anterior segment optical coherence tomography or ultrasound pachymetry) or determining the shape of its outer surface (computer keratotopography), as well as determining the biomechanical indices.13,14 However, a change in corneal thickness or shape may not only be associated with an improper hydration, but also with other factors, therefore the existing methods are insufficiently informative in terms of assessing corneal water content.
The use of terahertz (THz) radiation for these purposes is very promising.15 At the moment, the THz range is already being used to study biological objects. In particular, the possibilities of this approach are being investigated in the diagnosis of diseases such as breast cancer, skin cancer, as well as in the analysis of wounds, etc.16,17
At present, only a few scientific groups in the world are engaged in the use of THz radiation in ophthalmology. In particular, in Ref. 18, it was demonstrated that THz radiation can be used for quantitative measurement of physiological parameters of the tear film and determination of its hydration. The approach of other authors is based mainly on the use of pulsed THz radiation sources.19 For example, in Ref. 20, for scanning the hydration of porcine corneas in the THz range, a traditional scheme consisting of a photoconductive antenna, a femtosecond pump laser, and a Schottky diode as a receiver was used for obtaining pulsed THz radiation. In Ref. 21, the proposed approach was used to obtain images of water content in corneal tissues. However, this method is strongly limited because a powerful and bulky femtosecond laser needs to be used and, therefore, is not suitable for devices intended for clinical practice.
In Ref. 22, it was shown that the transmittance and reflectivity coefficient (RC) of the extracted cornea depend on the water content. In this case, the complex part of the dielectric constant decreases monotonically with a decrease in the water content in the cornea. In their work, the authors used a THz spectrometer with an optimized scanning speed. In another study23 by the same authors, the method of time-resolved THz spectroscopy was used for in vitro studies of the dependence of the reflection and transmission coefficients of the cornea in the range of 0.1 to 1.5 THz. The results show that the reflection coefficient decreases with increasing radiation frequency, while the absorption coefficient, on the contrary, increases, and it is almost linear. A recent review24 of THz methods of non-invasive scanning of corneal hydration for the diagnosis of eye diseases at early stages indicates the growing interest of the international scientific community toward this problem, its relevance, and scientific and clinical importance.
The results of recent studies in the field of THz scanning have demonstrated the effectiveness of this technology for non-invasive determination of the level of hydration of human organs and tissues and its high diagnostic capabilities in various pathological conditions.25 Nevertheless, for the best of our knowledge, there are currently no publications on the in vivo study of the dynamics of corneal hydration with THz radiation on a well-controlled model of corneal damage, which was the aim of this research.
In our previous work, for the first time in ophthalmology, we have created a test bed for studying corneal hydration using THz scanning, in particular, we have developed a laboratory prototype of the device for monitoring the state of the corneal water balance in vivo.26,27 It is necessary to emphasize that THz radiation will interact with both the tear film and the cornea structure. The preliminary data showed that the proposed technology is effective and has the potential for widespread use in clinical ophthalmology.
For further development of THz scanning of the cornea in clinical practice, it was necessary to validate this method on models of ophthalmic pathologies, potentially associated with changes in the water balance of this tissue. As such a pathology, we chose a model of corneal damage caused by ultraviolet (UV) B-range radiation, since it was shown that such an effect leads to an increase in corneal thickness, presumably associated with edema of its stroma.28
According the literature data, UVB-induced damage of the eye surface of experimental animals (rabbits) includes loss not only of corneal epithelium, but apoptosis of keratocytes and stromal edema. These changes are accompanied by clinically and histologically manifested corneal inflammation, neutrophil infiltration, nuclear pyknosis, vacuolization of endothelium, and exudation of the anterior chamber of the eye.29,30 Presumably, these changes can be associated with corneal hydration and can impact on the THz RC.
The aim of our study is to find out how THz optical properties of corneal tissue measured by an experimental setup, especially developed for in vivo contactless estimations of cornea reflectivity in the THz range, are related to pathological changes in the cornea parameters (thickness, volume, and tear production) caused by B-band ultraviolet (UVB) exposure.
2. Experiment
The experimental procedures followed the tenets of the code of ethics of the World Medical Association (Declaration of Helsinki) and were performed according to the 8th edition “Guide for the Care and Use of Laboratory Animals” of the National Research Council and “Statement for the Use of Animals in Ophthalmic and Visual Research” of the Association for Research in Vision and Ophthalmology. The study protocols were approved by the Helmholtz National Medical Research Center of Eye Diseases Ethical Committee.
2.1. Material and Methods
The experimental corneal damage modeling was performed according the technique described by Fris et al.31 and modified by us. We irradiated 16 eyes of 8 Chinchilla rabbits with 8-W UVB lamp (EB-280C/FE, Spectroline, 312-nm wavelength, 8 W, 230 V, 50 Hz, 0.4 A). The animal’s eyes were irradiated for 4 days during 20 min with a single dose of UVB radiation () from a distance of 0.4 m. These parameters (in particular, a 20-min exposure of UVB radiation) were used because they were shown to suffice for corneal damage, including the development of corneal edema.31 The lamp was located at a normal incidence to the visual axis of the rabbit eye.
Before the course of UVB irradiation and 1, 5, and 30 days after it, tear production evaluation by Schirmer test (Fig. 1) and corneal surface state assessment using fluorescein staining were performed. The complex of examinations of rabbit eyes also included biomicroscopy of corneal epithelium and stroma using slit lamp (XCEL 255), optical coherent tomography of anterior eye segment (OCT AS, Visante TM OCT, version 2.0, model 1000, Zeiss), Scheimpflug visualization and determination of anatomic and optical parameters of anterior eye segment using optical analyzer Galilei G2 (Ziemer Ophthalmic Systems AG 6.0.2), confocal microscopy of the cornea (Confoscan 4, Nidek), and endothelial microscopy (Specular Microscope SP-1P, Topcon, Japan).
THz scanning was carried out before the course of irradiation, as well as 5 and 30 days after the procedure, and included the determination of the RC of the cornea when positioning the continuous THz radiation source and detector at a distance of 1 and 13 cm from the animal’s eye, respectively. The 5th and 30th days were chosen for the estimation of early and late changes of the parameters studied. On the 5th day, the damaging process corresponds to the acute stage of inflammation, and in a month (30th days after UVB irradiation) the inflammation will have for the most part subsided. Thus, we determined the relationship between RC and cornea parameters in these two stages of the process.
An impact ionization avalanche transit-time diode (IMPATT) diode with a radiation frequency of 95 GHz was used as a source of THz radiation (see Fig. 2). Radiation from the IMPATT diode with a power of no more than 1 mW was directed to the rabbit cornea using a quasi-optical horn. The power of the reflected scattered radiation was measured using a Golay cell. The experimental setup was preliminary tested in our previous work in the experiment of studying the dependence of the reflection coefficient of the extracted rabbit corneas on the water content in them.26 To obtain optimal detector sensitivity, the incident radiation was modulated at a frequency of 10 Hz with a mechanical obturator. The rabbits were immobilized to exclude the impact of eye movements on the reflected signal. The duration of the RC measurement was short, and usually the rabbits’ eyes did not move. If they did move, the results were excluded from consideration. The electrical signal from Golay cell was measured with a selective voltmeter. The magnitude of this signal was used to calculate the reflectivity of the cornea, which is a characteristic of the state of its water balance.
The following scanning parameters were used: the value of the THz signal reflected from a metal plate located at a distance of 40 cm from the detector was 19 to 26 mV; the bias current of the IMPATT diode was 60 mA. All experiments were performed at the same humidity to eliminate the impact of this factor on the eye hydration.
A 3D electromagnetic simulation was performed using the finite element method. Electromagnetic properties of the cornea (a relative permittivity of 6.1, a bulk conductivity of , and a dielectric loss tangent of 1.34) at the frequency of 95 GHz were taken from Refs. 18 and 32. The spatial distribution of the incident and reflected radiation field (the magnitude of ) with used positioning of the source in front of the rabbit’s eye as well as the distribution of the specific absorption rate (SAR) of radiation over the surface of the eye are shown in Fig. 3. The following equation was used to calculate the SAR: , where is the material’s conductivity (defined as ) and is the mass density of the dielectric material in mass/unit volume. The radiation reflected in the horizontal direction was detected by a Golay cell. An insignificant part of the radiation power in some standing waves occurred in the diagonal direction. The simulation results show that the penetration depth of electromagnetic radiation does not exceed the thickness of the cornea. At the same time, the incident radiation is concentrated in a rather narrow beam, so a possibility of scanning the water concentration over the surface of the cornea and in its stroma is provided.
Statistical processing included nonparametric analysis (medians, quartiles 0.25; 0.75) as well as an assessment of the difference in terms of observation using the Wilcoxon T-test for comparing the results before and after UVB exposure and Spearman’s rank correlation coefficient. The difference was significant if .
3. Results and Discussion
The results of clinical examination (biomicroscopy) of the rabbit eyes have shown that on the second day the signs of local inflammation were revealed, which included conjunctiva redness and eyelid edemas of various degrees. A day after the end of UVB course, 50% to 100% of corneal surface were covered by defects of epithelium in all experimental eyes (Fig. 4). Five days after the end of UVB irradiation course, sufficient increase of tear production was determined, which was continuously decreasing only by the end of follow-up (Table 1). Confocal microscopy results obtained in 5 days after the end of UVB course demonstrated great amount of desquamated corneal epithelial cells. Endothelial microscopy showed the safety of corneal endothelium after UVB irradiation course.
Table 1.
Schirmer test result (median [quartile 0.25; quartile 0.75]), mm | ||
---|---|---|
Before irradiation | 5 days after irradiation | 30 days after irradiation |
16.0 [14.0; 18.5] | 19.1 [17.5; 21.5]* | 14.5 [12.0; 17.5]** |
The difference as compared to the initial level is significant according to Wilcoxon signed-rank test, .
The difference as compared to the initial level is not significant according to Wilcoxon signed-rank test, .
Besides this, OCT and Sheimpflug analyzer data (Table 2) showed stromal edema, nonuniform increase of corneal thickness and volume, change of corneal curvature and optical power, as well as a decrease of anterior chamber depth as compared with values of these parameters that we obtained earlier.33
Table 2.
Parameters | Before irradiation | 5 days after |
---|---|---|
Corneal thickness in middle area, | 394.1 [175.3; 552.7] | 480.3 [262.2; 768.7]* |
Corneal thickness in the periphery, | 396.4 [172.2; 555.0] | 489.6 [254.8; 783.4]* |
Mean corneal thickness, | 423.0 [377.2; 486.5] | 470.1 [391.8; 564.2] |
Corneal tissue volume, | 20.4 [16.8; 22.6] | 24.1 [19.8; 26.9]* |
Optical power of the anterior corneal surface in middle area, D | 42.7 [41.8; 42.9] | 45.6 [42.7; 44.0]* |
Curvature radius in the middle area of anterior corneal surface, mm | 8.0 [6.9; 8.4] | 7.6 [6.1; 8.0]* |
Curvature radius in the periphery of anterior corneal surface, mm | 8.2 [7.9; 8.4] | 7.7 [7.5; 7.8] |
Refraction in the cornea apex area, D | 44.7 [40.8; 48.7] | 48.6 [43.4; 50.7]* |
Radius in the cornea apex area, mm | 7.6 [7.1; 8.1] | 7.0 [6.2; 7.3]* |
Eccentricity (asphericity) of posterior cornea surface | 1.3 [0.7; 2.1] | 4.7 [2.7; 5.2]* |
Total corneal refraction in middle area, D | 44.1 [39.9; 47.3] | 46.0 [42.4; 49.7]* |
Anterior chamber depth, mm | 3.0 [2.7; 3.1] | 2.5 [1.9; 2.8]* |
The difference as compared to the initial level is significant according to Wilcoxon signed-rank test, .
Complex of all data obtained confirms the negative impact of UVB irradiation on the intensity of tear production, and on the corneal state, that allows assessing the induced pathology as a suitable model for searching correlations between clinical changes in the tissue and the results of its THz scanning.
THz scanning of the cornea of experimental animals revealed a significant (1.8 times) increase in RC on the 5th day after the irradiation course, followed by a slight decrease on the 30th day after it (amounting to a factor of 1.7 of the initial level), and in all cases the value of this indicator remained significantly higher than the initial level (Table 3).
Table 3.
Reflected signal (median [quartile 0.25; quartile 0.75]), mV | ||
---|---|---|
Before irradiation | 5 days after irradiation | 30 days after irradiation |
0.78 [0.33; 1.1] | 1.2 [0.57; 2.4]* | 0.98 [0.50; 1.8]** |
The difference as compared to the initial level is significant according to Wilcoxon signed-rank test, .
The difference as compared to the initial level is significant according to Wilcoxon signed-rank test, .
Comparison of obtained results with above-mentioned clinical data gives an evidence to propose the significant correlations between RC changes and the parameters of the course of UVB-induced pathological process. Thus, the RC increase measured 5 days after the UVB irradiation course generally corresponded to the increase (by a factor of 1.3) of tear production (Table 1).
The RC increase (with regard to the initial RC level) 5 days after the UVB exposure course was mostly in moderate inverse relationship with the value of the initial individual parameter of Schirmer test (Spearman’s correlation coefficient ). In most cases, more expressed RC increase in 5 days was noted in case of lowest initial tear production and vice versa, in cases of high initial values of Schirmer test the increase of RC was less expressed. Obviously, normal tear production and as a consequence normal tear film better protect a cornea against edema induced by UVB damage.
Important qualitative relationships were revealed when changes of RC and anatomical and optical parameters measured using optical analyzer were compared (Tables 2 and 3). In particular, RC increase occurred with the corneal edema which was manifested by corneal thickening (by 18.2% in the middle area and 17.6% in corneal periphery) and an increased volume of corneal tissue (by 17.6%). The increase of these parameters in 5 days after UVB irradiation course corresponded to the increase of RC in this follow-up period.
At 30 days after UVB exposure, RC remained above baseline and signs of corneal edema were still detected. Although the cornea has become thinner compared to the 5-day period (413.1 in the middle area and 414.3 in the periphery) and more uniform in thickness, but small differences in thickness compared to the initial level (see Table 3) in the middle zone (5%) and on the periphery of the cornea (4.5%) remained. The corneal tissue volume (21.2 ) approached the initial value by this time, but a slight increase (by 4.5%) compared to the initial level remained.
Thus, the in vivo verification of THz scanning for the detection of the dynamics of corneal hydration on the model of controlled UVB-induced inflammation and edema of this tissue has been made. To the best of our knowledge, it is the first time when such verification is made.
4. Conclusion
Our results demonstrate that the proposed approach can be used for in vivo contactless estimation of the reflectivity of rabbit cornea in the THz range and, thereby, of cornea hydration. On the one hand, RC values correspond to the dynamics of certain indirect parameters of corneal edema induced by UVB exposure at different stages of the pathologic process and, on the other hand, an RC value is a more informative parameter because it directly represents the overall corneal hydration. The benefit of determining the reflection coefficient is explained by the fact that this parameter directly characterizes the water content in the cornea. As the changes in the water content are consistent with the qualitative changes of the thickness and volume of the cornea and tear production at certain stages of the pathological process occurring in the cornea, we conclude that the developed experimental setup is valid and informative and can justifiably be used in the overall in vivo assessment of water content in the cornea.
Acknowledgments
The reported study was funded by RFBR, Sirius University of Science and Technology, and JSC Russian Railways and Educational Fund “Talent and Success,” Project No. 20-32-51004.
Biographies
Elena N. Iomdina is the principal researcher at Helmholtz National Medical Research Center of Eye Diseases. Her main research interests include myopia pathogenesis, surgical and nonsurgical methods of myopia treatment and sclera reinforcement in progressive myopia, research in glaucoma pathogenesis and treatment, biomechanics and biochemistry of the eye (particularly the cornea and the sclera), and experimental ophthalmology. She has authored over 150 articles in peer-reviewed journals, a book on eye biomechanics, and 40 patents and inventions.
Sergey V. Seliverstov is a senior researcher in the Physics Department of Moscow Pedagogical University. He received his PhD in physics in 2014. His scientific interests include a nonequilibrium superconductivity, terahertz (THz) radiation, and applied physics. The focus of his work is on the interaction of THz radiation with nanoscale objects (primarily superconductors) and also on the applications of the THz technologies in medicine.
Kseniya O. Teplyakova received her master’s degree in physics in 2010. She is a junior researcher in the Physics Department of Moscow Pedagogical State University (MPSU), where she develops experience in applications of THz radiation in biology and medicine. She has also organized many scientific events.
Elena V. Jani is the head of the Department of Infectious and Allergic Eye Diseases at the Helmholtz National Medical Research Center for Eye Diseases. Her main research interests include etiology, pathogenesis, clinical changes, and treatment of inflammatory diseases of ocular surface including the cornea and experimental ophthalmology. She is the author of over 100 articles in peer-reviewed journals, guidelines, and 4 patents and inventions.
Viktorya V. Pozdniakova is a senior researcher at Helmholtz National Medical Research Center of Eye Diseases. Her main research interests include problems of diagnosis and treatment of infectious, allergic, autoimmune eye diseases, and experimental ophthalmology. She is the author and coauthor of more than 90 publications.
Olga N. Polyakova is a researcher at the Laboratory of Quantum Detectors (MPSU). Her scientific interests include studying the dielectric characteristics of inhomogeneous dispersed media in the THz frequency range, photonic integrated devices. She is a co-author of about 20 articles and 2 patents.
Gregory N. Goltsman is a professor and a leading researcher in the Physics Department of MPSU. He received his PhD in 1973 and his Dr. Habil. in 1985. His scientific interests include superconductivity, applied physics, and THz technology. He has extensive experience in industry, where he has contributed to the development of small innovative high-tech businesses. He is the author and coauthor of more than 200 publications.
Disclosures
Authors have no relevant financial interests in this paper.
Contributor Information
Elena N. Iomdina, Email: iomdina@yandex.ru.
Sergey V. Seliverstov, Email: seliverstovsv@mail.ru.
Kseniya O. Teplyakova, Email: ksotepl@gmail.com.
Elena V. Jani, Email: e_yani@mail.ru.
Viktorya V. Pozdniakova, Email: vikapoz@yandex.ru.
Olga N. Polyakova, Email: ariya78@mail.ru.
Gregory N. Goltsman, Email: goltsman10@mail.ru.
References
- 1.Hall D., Jackson D., International Review of Connective Tissue Research, Vol. 10, Elsevier; (2013). [Google Scholar]
- 2.Maurice D., “Clinical physiology of the cornea,” Int. Ophthalmol. Clin. 2, 561–572 (1962). 10.1097/00004397-196210000-00002 [DOI] [PubMed] [Google Scholar]
- 3.Kim Y., et al. , “Variation of corneal refractive index with hydration,” Phys. Med. Biol. 49(5), 859 (2004). 10.1088/0031-9155/49/5/015 [DOI] [PubMed] [Google Scholar]
- 4.Afshari N., “Structure and function of the external eye and cornea,” in External Disease and Cornea, Thomas J., Ed., pp. 8–9, American Academy of Ophthalmology, San Francisco: (2016). [Google Scholar]
- 5.Liu Y. C., et al. , “Cataracts,” Lancet 390(10094), 600–612 (2017). 10.1016/S0140-6736(17)30544-5 [DOI] [PubMed] [Google Scholar]
- 6.Fatt I., Weissman B., Physiology of the Eye: An Introduction to the Vegetative Functions, Butterworth-Heinemann; (2013). [Google Scholar]
- 7.VanMeter W., Lee W., Katz D., Corneal Edema. Duane’s Clinical Ophthalmology, Vol. 4, Chapter 16A, Lippincott; (2006). [Google Scholar]
- 8.Lim G., et al. , “Toxic keratopathy-related corneal dehydration after laser in situ keratomileusis,” J. Cataract Refract. Surg. 31(8), 1656–1658 (2005). 10.1016/j.jcrs.2005.01.020 [DOI] [PubMed] [Google Scholar]
- 9.Pinsky P., Cheng X., “The electrochemical basis of corneal hydration, swelling, and transparency,” in Biomech. Eye 62–79 (2018). [Google Scholar]
- 10.Kim W., Jo J., “Corneal hydration affects ablation during laser in situ keratomileusis surgery,” Cornea 20(4), 394–397 (2001). 10.1097/00003226-200105000-00011 [DOI] [PubMed] [Google Scholar]
- 11.Meek K., et al. , “Transparency, swelling and scarring in the corneal stroma,” Eye 17(8), 927–936 (2003). 10.1038/sj.eye.6700574 [DOI] [PubMed] [Google Scholar]
- 12.Hatami-Marbini H., Rahimi A., “Interrelation of hydration, collagen cross-linking treatment, and biomechanical properties of the cornea,” Curr. Eye Res. 41(5), 616–622 (2016). 10.3109/02713683.2015.1042546 [DOI] [PubMed] [Google Scholar]
- 13.Dawson D. G., Edelhauser H. F., “Chapter 9—Corneal edema,” in Ocular Disease, Levin L. A., Albert D. M., Eds., pp. 64–73, W.B. Saunders, Edinburgh: (2010). [Google Scholar]
- 14.Yu-Chi L., Lin K., Mehta J., “Evaluation of corneal oedema—tools we have and those under investigation,” Eur. Ophthalmic Rev. 13(2), 76–80 (2019). 10.17925/EOR.2019.13.2.76 [DOI] [Google Scholar]
- 15.Bennett D., et al. , “Terahertz sensing in corneal tissues,” J. Biomed. Opt. 16(5), 057003 (2011). 10.1117/1.3575168 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16.Pickwell E., Wallace V., “Biomedical applications of terahertz technology,” J. Phys. D: Appl. Phys. 39(17), R301 (2006). 10.1088/0022-3727/39/17/R01 [DOI] [Google Scholar]
- 17.Son J.-H., “Terahertz electromagnetic interactions with biological matter and their applications,” J. Appl. Phys. 105(10), 102033 (2009). 10.1063/1.3116140 [DOI] [Google Scholar]
- 18.Ozheredov I., et al. , “In vivo THz sensing of the cornea of the eye,” Laser Phys. Lett. 15(5), 055601 (2018). 10.1088/1612-202X/aaac76 [DOI] [Google Scholar]
- 19.Zaytsev K., et al. , “The progress and perspectives of terahertz technology for diagnosis of neoplasms: a review,” J. Opt. 22(1), 013001 (2019). 10.1088/2040-8986/ab4dc3 [DOI] [Google Scholar]
- 20.Singh R., et al. , “Terahertz sensing of corneal hydration,” in Annu. Int. Conf. IEEE Eng. Med. and Biol., IEEE, pp. 3021–3024 (2010). 10.1109/ICICIS.2010.5534716 [DOI] [PubMed] [Google Scholar]
- 21.Taylor Z., et al. , “THz and mm-wave sensing of corneal tissue water content: in vivo sensing and imaging results,” IEEE Trans. Terahertz Sci. Technol. 5, 184–196 (2015). 10.1109/TTHZ.2015.2392628 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 22.Liu W., et al. , “Spectroscopic measurements and terahertz imaging of the cornea using a rapid scanning terahertz time domain spectrometer,” Chin. Phys. B 25, 060702 (2016). 10.1088/1674-1056/25/6/060702 [DOI] [Google Scholar]
- 23.Liu W., et al. , “Terahertz optical properties of the cornea,” Opt. Commun. 359, 344–348 (2016). 10.1016/j.optcom.2015.09.107 [DOI] [Google Scholar]
- 24.Yawata N., et al. , “Dynamic change in natural killer cell type in the human ocular mucosa in situ as means of immune evasion by adenovirus infection,” Mucosal Immunol. 9, 159–170 (2016). 10.1038/mi.2015.47 [DOI] [PubMed] [Google Scholar]
- 25.Bajwa N., et al. , “Non-invasive terahertz imaging of tissue water content for flap viability assessment,” Biomed. Opt. Express 8(1), 460–474 (2017). 10.1364/BOE.8.000460 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 26.Iomdina E., et al. , “Study of transmittance and reflectance spectra of the cornea and the sclera in the THz frequency range,” J. Biomed. Opt. 21(9), 097002 (2016). 10.1117/1.JBO.21.9.097002 [DOI] [PubMed] [Google Scholar]
- 27.Iomdina E., et al. , “Terahertz scanning for evaluation of corneal and scleral hydration,” Sovremennye Tekhnol. v Med. 10(4), 143 (2018). 10.17691/stm2018.10.4.17 [DOI] [Google Scholar]
- 28.Cejka C., Luyckx J., Cejková J., “Central corneal thickness considered an index of corneal hydration of the UVB irradiated rabbit cornea as influenced by UVB absorber,” Physiol. Res. 61(3), 299 (2012). 10.33549/physiolres.932242 [DOI] [PubMed] [Google Scholar]
- 29.Tiulina V., et al. , “Mitochondria-targeted antioxidant SKQ1 improves corneal healing after UV-induced damage in rabbits,” FEBS Open Bio 8, 215–215 (2018). 10.1186/s12886-018-0996-7 [DOI] [Google Scholar]
- 30.Zernii E., et al. , “Mitochondria-targeted antioxidant SKQ1 protects cornea from oxidative damage induced by ultraviolet irradiation and mechanical injury,” BMC Ophthalmol. 18(1), 336 (2018). 10.1186/s12886-018-0996-7 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 31.Fris M., et al. , “The effect of single and repeated UVB radiation on rabbit cornea,” Albrecht von Graæes Arch. Ophthalmol. 244(12), 1680–1687 (2006). 10.1007/s00417-006-0297-3 [DOI] [PubMed] [Google Scholar]
- 32.Gabriel C., Compilation of the Dielectric Properties of Body Tissues at RF and Microwave Frequencies, King’s College, London: (1996). [Google Scholar]
- 33.Milash S., Iomdina E., “Possibilities of using an optical Galilei G2 Scheimpflug analyzer in experimental studies (Rus),” Russ. Ophthalmol. J. 1, 59–62 (2015). [Google Scholar]