Skip to main content
The Journal of Clinical Hypertension logoLink to The Journal of Clinical Hypertension
. 2020 Mar 9;22(4):598–604. doi: 10.1111/jch.13840

Unattended automated office blood pressure measurement: Time efficiency and barriers to implementation/utilization

John Doane 1, Michael Flynn 1, Marcus Archibald 1, Dominick Ramirez 1, Molly B Conroy 1,, Barry Stults 1
PMCID: PMC8030085  PMID: 32153104

Abstract

Unattended automated office blood pressure (BP) measurement (u‐AOBP) improves office BP measurement accuracy and reduces white‐coat BP elevation, but there are reservations about its time efficiency in primary care. We used time‐stamp methodology to measure u‐AOBP procedure times performed without a rest period in 130 patients during routine clinic visits to three primary care clinics with 2.5‐4.9 years u‐AOBP experience. We documented the clinical activities of 30 medical assistants during the u‐AOBP procedures. We also assessed MA and clinician satisfaction and knowledge about u‐AOBP performance and interpretation. Median u‐AOBP procedure time was <5 minutes, and MAs engaged in productive clinical activities during 83% of the procedures. Ninety‐three percent of MAs and 100% of clinicians in the clinics agreed that u‐AOBP is an efficient method to improve hypertension management. Barriers to effective u‐AOBP implementation and ongoing utilization included initial difficulty incorporating u‐AOBP into clinic workflow and medical staff knowledge deficiencies concerning correct u‐AOBP performance and interpretation despite prior training and experience with the procedure. Intensive u‐AOBP education and training programs are needed to facilitate effective u‐AOBP implementation into primary care. The time required to perform u‐AOBP can be utilized productively by staff.

Keywords: hypertension, unattended automated office blood pressure (u‐AOBP)

1. INTRODUCTION

Because of the 15%‐30% prevalences of white‐coat blood pressure (BP) elevation1 and masked hypertension,2 out‐of‐office BP measurement is recommended to confirm the diagnosis of hypertension for most patients using either 24‐hour ambulatory or home BP monitoring.3, 4, 5, 6, 7, 8 Still, accurate office BP measurement (OBPM) remains essential to limit the need for labor‐intensive and expensive out‐of‐office BP monitoring. Unfortunately, OBPM in routine clinical practice is often performed inaccurately due to incorrect patient preparation and manual auscultatory measurement technique.9, 10 To improve OBPM accuracy, current guidelines recommend validated, automatic oscillometric devices to replace manual auscultatory BP measurement.3, 4, 5, 6, 7 But guideline‐recommended OBPM with these devices does not eliminate white‐coat BP elevation.10, 11 In contrast, validated, automated oscillometric devices that automatically perform and average three to five BP measurements over 3‐5 minutes while the patient rests alone in the exam room or another quiet place have been demonstrated to substantially reduce white‐coat BP elevation.12, 13 This approach to OBPM, termed unattended automated office BP measurement (u‐AOBP), is recommended as the preferred OBPM technique for patients with an initial elevated observed BP measurement by the 2019 American Heart Association Scientific Statement3 and by Hypertension Canada.6

Arguments against using u‐AOBP have included a lack of data about the cardiovascular prognostic significance of u‐AOBP and a perception that u‐AOBP may not be time‐efficient and feasible in clinical practice.14, 15, 16 While recent studies demonstrate strong associations of u‐AOBP with cardiovascular events and subclinical organ damage,17, 18 few studies have investigated the practicality of u‐AOBP in primary care.19, 20 We conducted a time efficiency study in three primary care clinics to measure u‐AOBP procedure time, medical assistant (MA) clinical activities during u‐AOBP performance, and MA and clinician satisfaction with and knowledge about u‐AOBP performance and interpretation.

2. METHODS

2.1. Study setting

We studied u‐AOBP measurements performed by MAs during routine patient clinic visits to two separate academic primary care practices (University Clinics 1 and 2) and one university‐affiliated community primary care practice (Community Clinic), all at different locations, from May, 2018 through March, 2019. Characteristics of the three clinics are listed in Table 1. In Table 1, note that mean MA experience with u‐AOBP was 2.5‐4.9 years in the three clinics with a range from 0.5 to 10.5 years for individual MAs. Therefore, AOBP protocols were well in place before this study began. The study was deemed exempt by the Institutional Review Board of the University of Utah Medical Center.

Table 1.

Characteristics of participating clinics

  University clinic 1 University clinic 2 Community clinic
Patient visits/year 7100 14 200 22 000
Clinician FTE 2.75 7.0 8.0
Formal u‐AOBP Roll‐out February, 2016 February, 2016 November, 2016
u‐AOBP device(s) used

Omron HEM‐907

Welch Allyn

Connex Spot BP

Omron HEM‐907

Welch Allyn

Connex Spot BP

Welch Allyn

Connex Spot BP

Number of MAs 5 10 20
Number of MAs monitored for u‐AOBP 5 9 16
Mean (range) years measuring u‐AOBP 2.8 (1.7‐4.5) 4.9 (0.5‐10.5) 2.5 (0.8‐4.5)
Number of patients monitored for u‐AOBP 26 47 67
Observation period May‐June 2018 July‐August 2018 August 2018‐March 2019

2.2. Study design

MAs were notified at a staff meeting that an observer would follow them to evaluate general patient flow at multiple time points during entire routine clinic visits. MAs were not made aware that u‐AOBP time efficiency was the study focus. One of five trained observers not previously known to clinic staff recorded time‐flow data on an IPad tablet utilizing Excel Time stamp at the sequential time points listed in Table 2 and the Figure S1. Consecutive patients were followed until completion of at least four u‐AOBP procedures by each MA on patients with initial elevated observed BP. u‐AOBP procedure time was defined as the length of time from the MA leaving the room after AOBP device placement until re‐entering the room to document the u‐AOBP measurement.

Table 2.

Time‐stamp measures collected during patient visits

Weight: Patient steps on scale or refuses weight

Patient enters room

Vital sign measurement and completion: single observed BP, heart rate, temperature, and O2 saturation

If u‐AOBP indicated due to elevated BP, then u‐AOBP explanation and initiation of u‐AOBP

u‐AOBP procedure time:
  • MA exits room
  • MA re‐enters room

MA records u‐AOBP value: EMR, sticky note for clinician, or verbally related to clinician

Clinician enters room

Patient exits room

Observers documented the percentage of u‐AOBP procedures during which MAs spent a majority of their time performing “clinical activities” (EMR charting, clinic phone calls, clinical interactions with other patients or with other members of the care team) versus “non‐clinical activities” (non‐clinic‐related conversation or computer activities).

A questionnaire (Table S1) was designed by the study investigators to determine MA and clinician knowledge about u‐AOBP performance, perceived u‐AOBP utility and efficiency, interpretation of AOBP results, and barriers to u‐AOBP implementation. Questionnaire items were determined by face validity. After completion of the entire study, the questionnaires were distributed to the MAs and clinicians and returned anonymously.

2.3. Blood pressure measurement

Office BP measurements and u‐AOBP were performed by trained clinic MAs using one of two validated oscillometric upper arm automated devices, the Omron HEM‐907 XL device,21 or the Welch Allyn Connex Spot BP22 in University Clinics 1 and 2; the Community Clinic used only the Welch Allyn device. An initial single observed BP measurement was performed without a defined rest period using appropriate cuff size for mid‐arm circumference, seated in a chair with back supported, cuff at midsternal level, arm supported on a flat surface, feet flat on the floor, with no conversation during the measurement. If the single observed BP measurement was ≥140/90 mm Hg or ≥130/80 mm Hg, according to individual clinician preference,4, 5 u‐AOBP was performed with the same device with the technique noted above but with the patient entirely alone in the exam room resting quietly (including no cell phone activity). The Omron HEM‐907 XL device used in University Clinic 1 was programmed to perform three BP measurements at 30‐second intervals with a 0‐second time delay and in University Clinic 2 to perform three BP measurements at 1 minute intervals with a 0‐second time delay.6, 23, 24, 25, 26 In all three clinics, the Welch Allyn Connex Spot BP was programmed to perform three BP measurements at 1 minute intervals with a 0‐second time delay.

2.4. Statistical analysis

Time‐stamp measurements (minutes, seconds) were converted to a continuous variable (minutes). Descriptive statistics were examined and due to non‐normal distribution data were presented as median (interquartile range (IQR)).

3. RESULTS

3.1. u‐AOBP procedure times

Across the three clinics, 30 MAs performed 140 u‐AOBP procedures on patients with an initially elevated observed BP. Individual time‐stamp u‐AOBP procedure times from device placement until the MA re‐entered the room are recorded in Figure 1. MAs prematurely re‐entered the clinic room prior to completion of the u‐AOBP procedure on 10 occasions in University Clinic 2 and the Community clinic, responsible for the very short procedure times in Figure 2. These procedure times were excluded from subsequent calculations. Figure 2 shows median (IQR) u‐AOBP procedure times in minutes in the remaining 130 patients in the three clinics as performed with a 0‐second time delay: 3.94 (1.63) Community Clinic, 3.79 (1.09) University Clinic 2, and 3.58 (0.81) University Clinic 1. Projected procedure times incorporating a 5‐minute delay would be 8.94, 8.79, and 8.58 minutes. The narrower range for AOBP procedure time noted in University Clinic 1 may result from their prior participation in a formal implementation study of u‐AOBP20 as well as workflow issues related to the larger number of MAs and clinicians in the other two clinics.

Figure 1.

Figure 1

AOBP procedure times in three primary care clinics

Figure 2.

Figure 2

Median u‐AOBP time‐stamp procedure times with 0‐second time delay

3.2. MA productivity during AOBP procedures

MAs engaged in productive clinical activities (EMR charting, clinic phone calls, paperwork, and clinical interactions with patients or other staff members while patients were alone in the exam room) during 83% of u‐AOBP procedures including 62% in University Clinic 1, 74% in University Clinic 2, and 97% in the Community Clinic.

3.3. Questionnaire results

Twenty‐eight of the 30 MAs and 15 of the 17 clinicians completed the anonymous questionnaire on u‐AOBP (Table S1). u‐AOBP procedures were almost always completed in the clinic room with only one MA also using a treatment room as an alternate performance site. Seventy‐one percent of MAs and 66% of clinicians indicated that u‐AOBP required minimal or no effort, and 93% of MAs and all clinicians agreed that u‐AOBP efficiently improved BP control for their patients. With respect to implementation barriers, one‐third of MAs and clinicians noted some initial difficulties incorporating u‐AOBP into clinic workflow, including orientation of new MAs and clinicians to the u‐AOBP procedure, patients arriving late to clinic with u‐AOBP deferred to the end of the visit, MAs forgetting to perform indicated u‐AOBP procedures, time lost having to clear spouses and relatives from the clinic room, and “occasionally” u‐AOBP causing clinicians to run late. Clinician, MA, or nurse resistance to using u‐AOBP was noted infrequently (14%) as a barrier to implementation.

Questionnaires revealed deficiencies in staff understanding of u‐AOBP performance and interpretation. Twenty‐one percent of MAs and 7% of clinicians reported performing u‐AOBP on all, most, or some patients rather than limiting u‐AOBP to patients with initially elevated observed BP.27, 28 Eleven percent of MAs and 33% of clinicians indicated that the patient was entirely alone in the exam room on just 30%‐69% of u‐AOBP procedures.12, 29 Only 36% of MAs and 27% of clinicians were aware that our clinic protocol and u‐AOBP device settings utilized a 0‐second time delay, the rest indicating that a 1‐5 minute rest period was in place.

Many clinicians and MAs did not recognize the u‐AOBP levels that correspond to guideline‐quality observed office BP and to mean daytime ambulatory BP on a 24‐hour study. Only 60% of clinicians and 21% of MAs correctly identified that a u‐AOBP of 135/85 mm Hg corresponds to an accurately measured, observed office BP of 140/90 mm Hg, and only 47% of clinicians and 18% of MAs recognized that a u‐AOBP of 130/80 mm Hg corresponds to an accurately measured, observed office BP of 130/80 mm Hg,30 the new office hypertension threshold recommended by the 2017 American College of Cardiology/American Heart Association hypertension guideline.4 Although u‐AOBP does not entirely eliminate white‐coat BP elevation,12, 29 40% of clinicians and 14% of MAs incorrectly indicated that an elevated u‐AOBP at two or more office visits confirms the diagnosis of hypertension without the need for out‐of‐office BP monitoring. Similarly, 47% of clinicians and 7% of MAs incorrectly indicated that an elevated u‐AOBP at one office visit warranted an adjustment of therapy in treated hypertension patients without resort to out‐of‐office BP monitoring.

4. DISCUSSION

This study investigated the time efficiency and medical staff satisfaction with u‐AOBP performed without a pre‐procedure rest period in a primary care setting. One hundred thirty u‐AOBP procedures were accomplished within a median 3.58‐3.94 minutes—projected to a median 8.50‐8.94 minutes had a prior 5‐minute rest period been utilized—which compares favorably to the 8‐14 minutes required to complete two or three guideline‐quality manual BP measurements.31 MAs were objectively engaged in productive clinical activities during over 80% of the AOBP procedures, further contributing to the time efficiency of u‐AOBP. Ninety‐three percent of MAs and 100% of clinicians agreed that u‐AOBP improves hypertension control. Time efficiency and medical staff satisfaction are important since u‐AOBP is now the preferred office method recommended by multiple organizations to screen for hypertension.3, 6, 32

We did find important barriers to overcome to successfully implement u‐AOBP. One‐third of MAs and clinicians reported initial difficulty incorporating u‐AOBP into clinic workflow, not surprising as most US MAs and primary care clinicians are not familiar with u‐AOBP rationale and technique. Overcoming this barrier will require dissemination of the 2019 American Heart Scientific Association Statement3 recommending u‐AOBP as the preferred office BP measurement approach and the development of u‐AOBP educational materials; we recently posted a u‐AOBP training video.33

Despite training and more than 2 year's experience with u‐AOBP in the three clinics, we found deficiencies in medical staff knowledge about u‐AOBP performance and interpretation. Fourteen percent of MAs reported performing u‐AOBP on all or most patients rather than limiting it to patients with initially elevated observed BP; this is a time‐inefficient approach since only about 3% of patients with initial observed BP at goal will have elevated BP on subsequent measurements.27, 28 MAs and clinicians indicated that u‐AOBP was not always performed with the patient entirely alone in the exam room, and on 10 of 140 u‐AOBP measurements, MAs re‐entered the room prior to u‐AOBP completion, both of which may result in white‐coat BP elevation. Imperfect adherence to BP measurement protocols despite intensive education and monitoring has been reported previously.3, 34 Regular reinforcement of correct u‐AOBP technique is essential to minimize performance deterioration and adjust for medical staff turnover.

Many medical staff may be incorrectly interpreting u‐AOBP results. Only 60% of clinicians and 21% of MAs recognized that a u‐AOBP of 135/85 mm Hg corresponds to a guideline‐quality observed office BP of 140/90 mm Hg,6, 11, 12 and 47% of clinicians and 18% of MAs that a u‐AOBP level of 130/80 mm Hg corresponds to a guideline‐quality observed office BP of 130/80 mm Hg,30 the latter the office BP threshold considered to warrant out‐of‐office BP monitoring by the 2017 American College of Cardiology/American Heart Association hypertension guideline.4 This incorrect interpretation may result in significant under detection of patients who would benefit from out‐of‐office BP monitoring to confirm the diagnosis of hypertension. Some medical staff appear to place undue reliance on elevated u‐AOBP results as 40% of clinicians and 14% of MAs incorrectly indicated that at an elevated u‐AOBP at two or more office visits confirmed the diagnosis of hypertension. Confirmatory out‐of‐office BP monitoring is currently recommended for patients with elevated u‐AOBP.3, 25, 29

Medical staff were uncertain about the use of a rest period prior to u‐AOBP, reflecting uncertainty among current studies and recommendations.3, 6, 23, 24, 25 Recommendations that u‐AOBP may be performed with “no special period of rest”6 have been based primarily on studies with the BpTRU device (no longer available) that takes six BP measurements and averages the last five.12, 13 There are currently few studies using no rest period with u‐AOBP devices that perform and average three BP measurements.12, 13 Two studies found good concordance of three measurement u‐AOBP performed using 0‐ and 1‐minute rest periods with mean awake ambulatory BP.35, 36 A recent study of 100 hypertensive persons compared mean systolic u‐AOBP using the Omron HEM‐907 XL device, performed with and without 5 minutes rest, to mean awake ambulatory systolic BP.37 In all 100 patients, three measurement mean systolic u‐AOBP performed with and without 5 minutes of rest did not differ significantly from each other (125.7 ± 17.9 mm Hg vs 127.0 ± 18.2 mm Hg, respectively) or from systolic mean awake ambulatory BP (126.6 ± 12.9 mm Hg). Similarly, in the 37 person subset with mean systolic u‐AOBP ≥ 130 mm Hg, mean systolic u‐AOBP with and without 5 minutes of rest correlated closely (143.3 ± 14.1 mm Hg and 144.3 ± 15.3 mm Hg, respectively); however, mean awake ambulatory BP in the 37 patients was significantly lower at 135.7 ± 11.9 mm Hg, a finding at variance from recent meta‐analyses. Despite the close correlations above of mean systolic u‐AOBP values with and without 5 minutes rest, these authors recommend that if initial systolic u‐AOBP is ≥130/80 mm Hg, u‐AOBP be repeated with 5 minutes rest. This recommendation is presumably based on their other finding of an order effect in 45 persons in whom mean systolic u‐AOBP performed first with no rest was 130.0 ± 17.7 mm Hg but follow‐up systolic u‐AOBP after 5 minutes rest was 126.5 ± 16.2 mm Hg. Given the greater time efficiency using no rest period, more studies comparing three measurement u‐AOBP devices with and without 5 minutes of rest are needed.

Study strengths include using three separate primary care clinics of different sizes and at different geographic locations, a large number of MAs performing multiple u‐AOBP procedures during usual clinical activities and unaware of the study focus, time‐stamp methodology to assure accurate timing of u‐AOBP procedures, and a 90% response rate among MAs and clinicians to the anonymous questionnaire.

A study limitation is that all three clinics had been using u‐AOBP for more than 2 years, and one clinic had participated in a published study using u‐AOBP,26 so our findings may not reflect the experience and initial difficulties of clinics newly implementing u‐AOBP. Our study did not collect data about actual observed BP and u‐AOBP levels and clinician actions in response to these BP levels, so we cannot compare actual clinician behavior with their questionnaire responses.

5. CONCLUSIONS

Because it is more accurate than casual BP measurement and reduces white‐coat BP elevation, u‐AOBP has been proposed as the preferred method to screen for elevated office BP. Concerns that u‐AOBP may be impractical and time‐inefficient in routine clinical practice were not confirmed in this study which found that u‐AOBP could be accomplished in <5 minutes without a prior rest period during which time MAs engaged in productive clinical activities. u‐AOBP may be more time‐efficient to clinic workflow than performing three guideline‐quality observed BP measurements over 8‐14 minutes. MA and clinician attitudes toward u‐AOBP were positive relative to feasibility, efficiency, and contribution to BP control.

Barriers to effective implementation and ongoing utilization of u‐AOBP include knowledge deficiencies concerning performance and interpretation. Initial difficulties incorporating u‐AOBP into clinic workflow and medical staff knowledge deficiencies about u‐AOBP utilization and interpretation require the development and dissemination of effective educational materials.

CONFLICT OF INTEREST

The authors have no conflicts of interest to disclose.

AUTHOR CONTRIBUTIONS

Study concept and design: All authors. Acquisition, analysis, or interpretation of data: All authors. Drafting of the manuscript: John Doane, Barry Stults, and Molly Conroy. Critical revision of the manuscript for important intellectual content: All authors. Statistical analysis: Marcus Archibald, Molly Conroy. Obtained funding: Molly Conroy. Administrative, technical, or material support: All authors. Study supervision: Barry Stults. Final approval: All authors. Accountability for all aspects of the work: All authors.

Supporting information

Fig S1

Table S1

ACKNOWLEDGMENTS

We express our gratitude to Rachel Hinze, Chase Killpack, Alyssa Stringham, Alyssa Ngoc Ngo, and the MAs of our three clinics. We also acknowledge the assistance of Maribel Cedillo with data cleaning and analysis.

Doane J, Flynn M, Archibald M, Ramirez D, Conroy MB, Stults B. Unattended automated office blood pressure measurement: Time efficiency and barriers to implementation/utilization. J Clin Hypertens. 2020;22:598–604. 10.1111/jch.13840

Funding information

Funding was from Faculty Development Accounts in University of Utah Division of General Internal Medicine.

REFERENCES

  • 1. Gorostidi M, Vinyoles E, Banegas JR, de la Sierra A. Prevalence of white‐coat and masked hypertension in national and international registries. Hypertens Res. 2015;38:1‐7. [DOI] [PubMed] [Google Scholar]
  • 2. Schwartz JE, Burg MM, Shimbo D, et al. Clinic blood pressure underestimates ambulatory blood pressure in an untreated employer‐based U.S. population. Circulation. 2016;134:1794‐1807. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3. Muntner P, Shimbo D, Carey RM, et al. Measurement of blood pressure in humans: a scientific statement from the American heart association. Hypertension. 2019;73:e35‐e66. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4. Whelton PK, Carey RM, Aronow WS, et al. 2017 ACC/AHA/AAPA/ABC/ACPM/AGS/APhA/ASH/ASPC/NMA/PCNA. 2017 guideline for the prevention, detection, evaluation, and management of high blood pressure in adults. J Am Coll Cardiol. 2017;2018(71):2199‐2269. [Google Scholar]
  • 5. Williams B, Mancia G, Spiering W, et al. 2018 ESC/ESH guidelines for the management of arterial hypertension: the task force for the management of arterial hypertension of the European Society of Cardiology and the European Society of Hypertension. J Hypertens. 2018;36:1953‐2041. [DOI] [PubMed] [Google Scholar]
  • 6. Nerenberg KA, Zarnke KB, Leung AA, et al. Hypertension Canada’s 2018 guidelines for diagnosis, risk assessment, prevention, and treatment of hypertension in adults and children. Can J Cardiol. 2018;34(5):506‐525. [DOI] [PubMed] [Google Scholar]
  • 7. Siu AL. Screening for high blood pressure in adults: U.S. preventive services task force recommendation statement. Ann Intern Med. 2015;163:778‐786. [DOI] [PubMed] [Google Scholar]
  • 8. Piper MA, Evans CV, Burda BU, et al. Diagnostic and predictive accuracy of blood pressure screening methods with consideration of rescreening intervals: a systematic review for the U.S. preventive services task force. Ann Intern Med. 2015;162:192‐204. [DOI] [PubMed] [Google Scholar]
  • 9. Myers MG, Godwin M, Dawes M, et al. Measurement of blood pressure in the office: recognizing the problem and proposing the solution. Hypertension. 2010;55:195‐200. [DOI] [PubMed] [Google Scholar]
  • 10. Myers MG. The great myth of office BP measurement. J Hypertens. 2012;30:1894‐1898. [DOI] [PubMed] [Google Scholar]
  • 11. Myers MG. A short history of automated office blood pressure – 15 years to SPRINT. J Clin Hypertens. 2016;18:721‐724. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12. Roerecke M, Kaczorowski J, Myers MG. Comparing automated office blood pressure readings with other methods of blood pressure measurement for identifying patients with possible hypertension: a systematic review and meta‐analysis. JAMA Intern Med. 2019;179(3):351‐362. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13. Pappaccogli M, DiMonaco S, Perlo E, et al. Comparison of automated office blood pressure with office and out‐of‐office measurement techniques: a systematic review and meta‐analysis. Hypertension. 2019;73:481‐490. [DOI] [PubMed] [Google Scholar]
  • 14. Stergiou G, Kollias A, Parati G, O’Brien E. Office blood pressure measurement: the weak cornerstone of hypertension diagnosis. Hypertension. 2018;71:813‐815. [DOI] [PubMed] [Google Scholar]
  • 15. Mancia G, Kjeldsen SE. Adopting systolic blood pressure intervention trial (SPRINT)‐like office blood pressure measurements in clinical practice. J Hypertens. 2017;35:471‐472. [DOI] [PubMed] [Google Scholar]
  • 16. Asayama K, Ohkubo T. Unattended automated measurements: office and out‐of‐office blood pressures affected by medical staff and environment. Hypertension. 2019;74:1294‐1296. [DOI] [PubMed] [Google Scholar]
  • 17. Myers MG, Kaczorowski J, Dolovich L, Tu K, Paterson JM. Cardiovascular risk in hypertension in relation to achieved blood pressure using automated office blood pressure measurement. Hypertension. 2016;68:866‐872. [DOI] [PubMed] [Google Scholar]
  • 18. Salvetti M, Paini A, Aggiusti C, et al. Unattended versus attended blood pressure measurement: relationship with preclinical organ damage. Hypertension. 2019;73:736‐742. [DOI] [PubMed] [Google Scholar]
  • 19. Kaczorowski J, Myers MG, Gelfer M, et al. How do family physicians measure blood pressure in routine clinical practice? – National survey of Canadian family physicians. Can Fam Physician. 2017;63(3):e193‐e199. [PMC free article] [PubMed] [Google Scholar]
  • 20. Doane J, Buu J, Penrod MJ, et al. Measuring and managing blood pressure in a primary care setting: a pragmatic implementation study. J Am Board Fam Med. 2018;31(3):375‐388. [DOI] [PubMed] [Google Scholar]
  • 21. White WB, Anwar YA. Elevation of the overall efficacy of the Omron office digital blood pressure HEM‐907 monitor in adults. Blood Press Monit. 2001;6:107‐110. [DOI] [PubMed] [Google Scholar]
  • 22. Alpert BS. Clinical evaluation of the Welch Allyn Sure BP algorithm for automated blood pressure measurement. Blood Press Monit. 2007;12(4):215‐218. [DOI] [PubMed] [Google Scholar]
  • 23. Stergiou GS, Lin CW, Lin CM, et al. Automated device that complies with current guidelines for office blood pressure measurement: design and pilot application study of the Microlife WatchBP office device. Blood Press Monit. 2008;13:231‐235. [DOI] [PubMed] [Google Scholar]
  • 24. Stults B, Doane J, Penrod MJ, Conroy MB. Does automated office blood pressure require a 5‐minute rest period when used to screen for hypertension? J Clin Hypertens. 2019;21(1):136. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 25. Myers MG, Colella TGF. Response to: does AOBP require a 5‐minute rest period to screen for hypertension? J Clin Hypertens. 2019;21:137. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 26. Boonyasai RT, McCannon EL, Landavaso JE. Automated office‐based blood pressure measurement: an overview and guidance for implementation in primary care. Curr Hypertens Rep. 2019;21:29. [DOI] [PubMed] [Google Scholar]
  • 27. Handler J, Zhao Y, Egan BM. Impact of the number of BP measurements on blood pressure classification in U.S. adults: NHANES 1999–2008. J Clin Hypertens. 2012;14:751‐759. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 28. Burkard T, Mayr M, Winterhalder C, et al. Reliability of single office blood pressure measurements. Heart. 2018;104(14):1173‐1179. [DOI] [PubMed] [Google Scholar]
  • 29. Myers MG. A meta‐analysis that helps clarify the use of automated office blood pressure in clinical practice. J Clin Hypertens. 2019;21(4):460‐462. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 30. Myers MG, Matangi M, Kaczorowski J. Comparison of awake ambulatory blood pressure and automated office blood pressure using linear regression analysis in untreated patients in routine clinical practice. J Clin Hypertens. 2018;20:1696‐1702. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 31. Giles TD, Egan P. Pay (adequately) for what works: the economic underevaluation of office and ambulatory blood pressure recordings. J Clin Hypertens (Greenwich). 2008;10(4):257‐259. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 32. Padwal R, Campbell NRC, Shutte AE, et al. Optimizing observer performance of clinic blood pressure measurement: a position statement from the Lancet Commission on Hypertension Group. J Hypertens. 2019;37:1737‐1745. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 33. Doane J, Penrod J, Stults B. [UUHC OD] (2017, August 30). Improving the accuracy of blood pressure measurement. [Video file]. https://www.youtube.com/watch?v=VnLDJDIy0u4.
  • 34. Boonyasai RT, Carson KA, Marsteller JA, et al. A bundled quality improvement program to standardize clinical blood pressure measurement in primary care. J Clin Hypertens. 2018;20(2):324‐333. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 35. Myers MG, Valdivieso M. Evaluation of an automated sphygmomanometer for use in the office setting. Blood Press Monit. 2012;17:116‐119. [DOI] [PubMed] [Google Scholar]
  • 36. Moore MN, Schultz MG, Nelson MR, et al. Identification of the optimal protocol for automated office blood pressure measurement among patients with treated hypertension. Am J Hypertens. 2018;31(3):299‐304. [DOI] [PubMed] [Google Scholar]
  • 37. Andreadis EA, Geladari CV, Angelopoulos ET. Automated office blood pressure measurements obtained with and without preceding rest are associated with awake ambulatory blood pressure. J Clin Hypertens. 2020;22:32‐38. [DOI] [PMC free article] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

Fig S1

Table S1


Articles from The Journal of Clinical Hypertension are provided here courtesy of Wiley

RESOURCES