Skip to main content
PLOS ONE logoLink to PLOS ONE
. 2021 Apr 14;16(4):e0243333. doi: 10.1371/journal.pone.0243333

Versatile and flexible microfluidic qPCR test for high-throughput SARS-CoV-2 and cellular response detection in nasopharyngeal swab samples

Julien Fassy 1,#, Caroline Lacoux 1,#, Sylvie Leroy 1,2, Latifa Noussair 3, Sylvain Hubac 4, Aurélien Degoutte 2, Georges Vassaux 5, Vianney Leclercq 6, David Rouquié 7, Charles-Hugo Marquette 2, Martin Rottman 3,8, Patrick Touron 4, Antoinette Lemoine 3, Jean-Louis Herrmann 3,8, Pascal Barbry 1, Jean-Louis Nahon 1, Laure-Emmanuelle Zaragosi 5, Bernard Mari 1,*
Editor: Tzong-Yueh Chen9
PMCID: PMC8046349  PMID: 33852580

Abstract

The emergence and quick spread of SARS-CoV-2 has pointed at a low capacity response for testing large populations in many countries, in line of material, technical and staff limitations. The traditional RT-qPCR diagnostic test remains the reference method and is by far the most widely used test. These assays are limited to a few probe sets, require large sample PCR reaction volumes, along with an expensive and time-consuming RNA extraction step. Here we describe a quantitative nanofluidic assay that overcomes some of these shortcomings, based on the BiomarkTM instrument from Fluidigm. This system offers the possibility of performing 4608 qPCR end-points in a single run, equivalent to 192 clinical samples combined with 12 pairs of primers/probe sets in duplicate, thus allowing the monitoring of SARS-CoV-2 including the detection of specific SARS-CoV-2 variants, as well as the detection other pathogens and/or host cellular responses (virus receptors, response markers, microRNAs). The 10 nL-range volume of BiomarkTM reactions is compatible with sensitive and reproducible reactions that can be easily and cost-effectively adapted to various RT-qPCR configurations and sets of primers/probe. Finally, we also evaluated the use of inactivating lysis buffers composed of various detergents in the presence or absence of proteinase K to assess the compatibility of these buffers with a direct reverse transcription enzymatic step and we propose several protocols, bypassing the need for RNA purification. We advocate that the combined utilization of an optimized processing buffer and a high-throughput real-time PCR device would contribute to improve the turn-around-time to deliver the test results to patients and increase the SARS-CoV-2 testing capacities.

Introduction

To control the pandemic and monitor virus propagation of SARS-CoV-2 in real time, extensive testing is necessary. Although alternatives are available [1, 2], viral load detection from nasopharyngeal or saliva samples is still the most appropriate method to identify SARS-CoV-2 carriers. The current diagnostic testing methods recommended by the Centers for Diseases Control (CDC) and the World Health Organization (WHO) are based on a traditional RT-qPCR assay, with validated primers [3]. However, the availability of this assay has been a major hurdle in the orderly and efficient management of the pandemic because of reagent shortages, as well as material and staff limitations. A rapid processing of the samples is also a crucial factor. To further stem the spread of coronavirus disease 2019 (COVID-19), a series of innovative approaches have been recently proposed [4]. Notably, to face material shortage and reduce processing times, two kinds of process optimization should be undertaken: (1) multiplexing sample and PCR probes, as well as (2) reducing the steps in sample preparation.

Most approved RT-qPCR assays are “one-step” kits that should be performed on standard real-time thermocyclers. When sample pooling [5] is not used, throughput is quite limited and the use of alternative systems, such as the BiomarkTM HD device from Fluidigm may overcome this limitation. The BiomarkTM HD device is a nanofluidic automated real-time PCR system that exploits the microfluidic technology through the use of dynamic arrays of integrated fluidic circuits (IFCs). Using, for example, the 192.24 Integrated Fluid Circuit (IFC), 192 samples can be processed in parallel with 24 independent sets of probes, allowing increased confidence in viral RNA detection as additional SARS-CoV-2 probes can be included. Probes to detect supplementary targets such as other RNA viruses, or host response genes can also be included. This flexibility presents a clear added value for both clinical monitoring of viral and bacterial pathogens as well as for research studies [6, 7]. Another advantage of this system resides in the low amount of reagents needed compared to classical real-time PCR machines. The reaction volume is down to the 10 nL-scale compared to the 10 μL-scale in classical qPCR, thus reducing reagent requirement, an important point in times of shortages.

The second improvement resides in the reduction of the number of steps for sample preparation. RT-qPCR detection starts with viral RNA extraction. This is time consuming for laboratories that are not equipped with high-throughput automated systems. It is also reagent consuming, and as March 2020, laboratories have suffered from major RNA extraction kit shortages. Several teams have proposed RNA extraction-free protocols. The resulting improvement in processing times was balanced by a loss in sensitivity. The comparison of these different studies is made difficult by the great variabilities of the protocols. This includes parameters such as the type of collection media, the use of additives such as detergents, heat-inactivation, or selection of a particular RT-qPCR kit [816]. In 2 of these studies, measurements on a large set of clinical samples demonstrated the effectiveness of direct RT-PCR assays with or without heat inactivation on various transport medium (VTM) [8, 9]. Srivatsan et al. proposed to collect dry swabs and elute them in Tris-EDTA (TE) to eliminate collection medium dilution and variation, and to bypass the RNA extraction step [17]. This simplified protocol was as sensitive as a conventional protocol (sample collection in VTM). They also evaluated the addition of detergents (IGEPAL, Triton X-100 and Tween-20) that facilitates virus inactivation in the elution medium. Addition of all three detergents resulted in a loss of sensitivity.

In the present study, we evaluated the use of a high-throughput real-time PCR device, the BiomarkTM HD, to increase throughput, flexibility in probe inclusion, and decrease reagent consumption, together with an optimized protocol for SARS-CoV-2 RNA detection without RNA extraction.

Materials and methods

Primers and probes for the detection of viral and cellular genes

All the DNA primers and probes used in this study are listed and described in Tables 1 and 2. The set of primers/probe used to detect the wild-type spike (S) gene was described in [18].

Table 1. List of primers/probe for virus detection.

  Target Name used in this study Name Sequence 5’ to 3’
USA CDC, N Gene N1 2019-nCoV_N1-F  GACCCCAAAATCAGCGAAAT
2019-nCoV_N1-R  TCTGGTTACTGCCAGTTGAATCTG
2019-nCoV_N1-P  ACCCCGCATTACGTTTGGTGGACC
N Gene N2 2019-nCoV_N2-F TTACAAACATTGGCCGCAAA
2019-nCoV_N2-R  GCGCGACATTCCGAAGAA
2019-nCoV_N2-P  ACAATTTGCCCCCAGCGCTTCAG
N Gene N3 2019-nCoV_N3-F ATCACATTGGCACCCGCAATCCTG
2019-nCoV_N3-R AGATTTGGACCTGCGAGCG
2019-nCoV_N3-P TTCTGACCTGAAGGCTCTGCGCG
Charité, Germany E gene E E_Sarbeco_F1 ACAGGTACGTTAATAGTTAATAGCGT
E_Sarbeco_R2 ATATTGCAGCAGTACGCACACA
E_Sarbeco_P1 ACACTAGCCATCCTTACTGCGCTTCG
China, CDC Orf1 / Rdrp gene ORF1ab ORF1ab-F  CCCTGTGGGTTTTACACTTAA
ORF1ab-R  ACGATTGTGCATCAGCTGA
ORF1ab-P  CCGTCTGCGGTATGTGGAAAGGTTATGG
Japan N Gene N NIID_2019-nCOV_N_F2  AAATTTTGGGGACCAGGAAC
NIID_2019-nCOV_N_R2  TGGCAGCTGTGTAGGTCAAC
NIID_2019-nCOV_N_P2  ATGTCGCGCATTGGCATGGA
Variants detection (this study) Wild type S gene Spike-F Spike-F TCAACTCAGGACTTGTTCTTAC
Spike-R Spike-R TGGTAGGACAGGGTTATCAAAC
WT-Spike Wild type Spike-P TGGTCCCAGAGACATGTATAGCAT
Mutant S gene Mutant spike 1 Mutant S-Probe 1 CCATTGGTCCCAGAGATAGCATGG
Mutant spike 2 Mutant S-Probe 2 CAGAGATAGCATGGAACCAAGTAA
Mutant spike 3 Mutant S-probe 3 CCCAGAGATAGCATGGAACCAAGT
Orf1ab / Wild type NSP6 gene NSP6-F NSP6-F CTAGTTGGGTGATGCGTATT
NSP6-R NSP6-R ACACAGTTCTTGCTGTCATAA
Wild type NSP6 Wild type NSP6-P AGTCTTTTAGCTTAAAACCAGACAAACTAGT
Orf1ab / NSP6 mutant gene Mutant NSP6 Mutant NSP6 AACACAGTCTTTTAGCTTCAAACTAGTATCAA

TaqMan probes are labeled at the 5’-end with the reporter molecule 6-carboxyfluorescein (FAM) and with the quencher, Blackhole Quencher 1 (BHQ1) at the 3’-end.

Table 2. List of primers/probe for cellular genes detection.

Target Name used in this study Name Sequence 5’ to 3’
RnaseP RNP CDC-RP-F  AGATTTGGACCTGCGAGCG
CDC-RP-R  GAGCGGCTGTCTCCACAAGT
CDC-RP-P TTCTGACCTGAAGGCTCTGCGCG
TMPRSS2 TMPRSS2 TMPRSS2 Forward TATAGCCTGCGGGGTCAAC
TMPRSS2 Reverse CACTCGGGGGTGATGATGG
TMPRSS2 Probe TCAAGCCGCCAGAGCAGGATCGT
ACE2 ACE2 ACE2 Forward GGCTCCTTCTCAGCCTTGTT
ACE2 Reverse GGTCTTCGGCTTCGTGGTTA
ACE2 Probe TGCTGCTCAGTCCACCATTGAGG
IL1a IL1a IL1-a Forward CATTGGCGTTTGAGTCAGCA
IL1-a Revrse CATGGAGTGGGCCATAGCTT
IL1-a Probe GTCAAGATGGCCAAAGTTCCAGACA
IL1b IL1b IL1b Forward CAGAAGTACCTGAGCTCGCC
IL1b Reverse AGATTCGTAGCTGGATGCCG
IL1b Probe CCAGGACCTGGACCTCTGCCC
CXCL8 CXCL8 CXCL8 Forward TGGACCCCAAGGAAAACTGG
CXCL8 Reverse TTTGCTTGAAGTTTCACTGGCA
CXCL8 Probe GTGCAGAGGGTTGTGGAGAAGTTT
IL6 IL6 IL6 Forward TGCAATAACCACCCCTGACC
IL6 Reverse GTGCCCATGCTACATTTGCC
IL6 Probe TGCCAGCCTGCTGACGAAGC
IFNb1 IFNb1 IFNb1 Forward AGTAGGCGACACTGTTCGTG
IFNb1 Reverse GCCTCCCATTCAATTGCCAC
IFNb1 Probe TGCTCTCCTGTTGTGCTTCTCCA
IFIT1 IFIT1 IFIT1 Forward GATCTCAGAGGAGCCTGGCTAA
IFIT1 Reverse TGATCATCACCATTTGTACTCATGG
IFIT1 Probe CAAAACCCTGCAGAACGGCTGCC

TaqMan probes are labeled at the 5’-end with the reporter molecule 6-carboxyfluorescein (FAM) and with the quencher, Blackhole Quencher 1 (BHQ1) at the 3’-end.

Positive control plasmid

Positive control plasmid containing the full length nucleocapsid coding sequence (N) was purchased at Integrated DNA Technologies (2019-nCoV_N_Positive Control; cat. no. 10006625). Plasmid was supplied at a final concentration of 200 000 copies/μL in IDTE pH = 8 buffer, and has been used to evaluate both the detection efficiency and sensitivity of the BiomarkHD System Analysis.

In vitro transcription

The synthetic transcript for the 2019-nCoV N coding sequence (N transcript) was generated using the 2019-nCoV_N_Positive Control plasmid. The template for the in vitro transcription was produced by PCR: briefly, 6 ng of plasmid was used to amplify the entire coding sequence of the N gene using: T7-For-sens (ATATAATACGACTCACTATAGGATGTCTGATAATGGACCC; T7 promoter sequence underlined) and Rev-sens (TTAGGCCTGAGTTGAGTC) as primers and the Q5-High Fidelity DNA polymerase (New England Biolabs, cat. no. M0491S). Thermal cycling conditions were: 98°C for 30 s, 40 cycles of 98°C for 10 s, 52°C for 30 s, 72°C for 1 min followed by 72°C for 2 min (SimpliAmp Thermal Cycler, ThermoFisher Scientific). The size of the PCR product was verified on a 1.5% agarose gel in 0.5X TAE and further purified using the QIAquick PCR purification kit (Qiagen, cat. 28106). Template was used in the in vitro transcription reaction using the T7 RiboMAXExpress Large Scale Production System (Promega, cat. no. P1320) according to manufacturer’s instructions. The in vitro transcript was extracted using phenol chloroform isoamyl followed by a chloroform wash and further precipitated using 300 mM sodium acetate in absolute ethanol at -20°C, over-night. Sample was centrifugated at 4°C for 30 min and washed twice with 70% cold ethanol. The RNA pellet was dried at RT for 5 min and resuspended in 100 μL water.

Clinical samples

Pulmonology department, Nice University Hospital

20 clinical samples from study participants were collected as part of the ELISpot study (ClinicalTrials.gov Identifier: NCT04418206). Participants were enrolled after signing written informed consent. Ethics committee approval was obtained from the “Comité de Protection des Personnes Sud Méditerranée V” (registration # 2020-AO1050-39) on April 22, 2020. Nasopharyngeal swabs were collected in ESwab™ (COPAN) transport medium (2 mL), stored at 4°C between sample collection and transport to the laboratory. One COVID-19 confirmed patient was collected consecutively using 2 different transport media: ESwabTM or a Tris-EDTA (TE: 10 mM Tris HCl pH 7.0, 2 mM EDTA, 20 μg/mL yeast tRNA) buffer.

Garches Hospital (AP-HP)

Purified RNA samples (n = 92) from nasopharyngeal swabs with known COVID-19 status were obtained from the Garches Hospital following an ISO 15189 certified lab protocol previously described [19] and used to validate the BiomarkTM HD protocol. Additional frozen VTMs from 55 patients with known COVID-19 status collected in 1–2 mL of saline buffer were obtained to set up the protocol. All samples were recorded for traceability on the basis of a unique barcode identifier.

Bioesterel laboratories

Samples (18 in virucide VTM, 2 in non-virucide VTM, Lingen) were stored overnight at 4°C and processed the next day. Purified RNA samples (n = 74) from COVID-19 positive nasopharyngeal swabs with known status (collected week 3, 2021, in the region from Nice to Cannes, France) were used to screen for the presence of SARS-CoV-2 variants. A generic consent form was signed by the patients, allowing the utilization of the samples for scientific research. All samples were anonymized and their COVID-19 status was recorded.

RNA extraction

Both miRNeasy Serum/Plasma Advanced Kit (Qiagen, cat. no. 217204) and QIAamp Viral RNA Mini Kit (Qiagen, cat. no. 52906) were used for Total RNA extraction from clinical samples according to manufacturer’s instructions and using the Qiacube (Qiagen) apparatus. Final elution was performed in 20 μL water and 60 μL of AVE buffer for the miRNeasy Serum/Plasma Advanced Kit and QIAamp Viral RNA Mini Kit, respectively. Total RNA (containing miRNAs) were directly used for further analysis or stored at -20°C.

Detergent treatments of clinical samples

All clinical samples were handled in a biosafety level 2 laboratory and under biological safety cabinet using the adapted biosafety personal protective and respiratory equipment, according to the recommendations of the French society of microbiology (https://www.sfm-microbiologie.org/). The following detergents were used: Triton X-100 (Sigma, cat. no. T9284-100ML), Brij™35 (Millipore, cat. No. 1.01894.0100), Tween-20 (Sigma, cat. no. P7949-500ML), Brij™O10 (Sigma, cat. no. P6136-100G) and Poly Ethylene Glyco-600 (Aldrich, cat no. 202401-250G). They were included alone or in combination to prepare a 2X detergent master mix that was added to clinical transport media samples. A 10 mg/mL Proteinase K (Sigma P4850) solution was mixed extemporaneously with the 2X detergent master mix to reach a 2 mg/mL final concentration. Heating treatments were carried out at 95°C for 5 min in a thermo-cycler apparatus (SimpliAmp Thermal Cycler, ThermoFisher Scientific). Treatment using the Quick ExtractTM DNA Extraction Solution (Lucigen, QE09050) was performed using the same volume of reagent and sample, according to the manufacturer’s instructions. Mixtures were then incubated at 95°C for 5 min or at 60°C for 10 min in a thermo-cycler. TE buffer was prepared using 10 mM Tris-HCl pH7.0, 2 mM EDTA, 20 μg/mL yeast tRNA with or without 0.5% Triton X-100.

Detection of mRNA expression protocol using the BiomarkTM HD system analysis

Reverse transcription using the Fluidigm reverse transcriptase

Extracted RNA (2 μL) or processed clinical samples were reverse transcribed using the Reverse Transcription Master Mix kit according to the manufacturer’s instructions (catalog number # PN 100–6297). Thermal cycling conditions were: 25°C for 5 min, 42°C for 30 min, 85°C for 5 min.

Reverse transcription using the Vilo SuperScript IV step

2.5 μL of processed clinical samples were used in a reverse transcription reaction using the SuperScript IV Vilo Master Mix (ThermoFisher Scientific, cat. no. 11756500) following the manufacturer’s instructions. Reactions were incubated at 25°C for 10 min, 55°C for 10 min, 85°C for 5 min.

cDNA pre-amplification step

2.5 μL of cDNA were preamplified using the Preamp Master Mix kit (Fluidigm, cat. no. PN 100–5744) according to the manufacturer’s instructions: 1 μL of Preamp Master Mix was combined with 1.25 μL of Pooled Taqman assay mix and 2.5 μL of cDNA in a 5 μL total volume reaction. The Pooled Taqman assay (180 nM) used in these reactions was prepared from an intermediate pooled Taqman assay solution (6.7 μM Forward and Reverse primers, 1.7 μM probes). Thermal cycling conditions were: 95°C for 2 min followed by 20 cycles of 95°C for 15 s, 60°C for 2 min. After each preamplification reaction, samples were diluted 1:5 by adding nuclease-free water up to 25 μL.

One step “RT-Preamplification”

One step “RT-Preamplification reactions” were performed using the Cells Direct One-Step qRT-PCR kit (Invitrogen, cat. no. 46–7200) according to a protocol adapted from the Reverse Transcription-Specific Target Amplification (Fluidigm). Briefly 5.5 μL of total RNA were mixed with 6.25 μL the Pooled Taqman assay, 0.5 μL Superscript III RT/Platinum Taq, 12.5 μL 2x Reaction mix and nuclease-free water to 25 μL total volume. Thermal cycling conditions were: 42°C for 15 min, 95°C for 2 min followed by 15 to 20 cycles of 95°C for 15 s, 60°C for 2 min. Then, samples were diluted 1:5 using nuclease-free water up to 125 μL.

Real time qPCR using BiomarkTM HD system

PCR was performed following Gene Expression Standard TaqMan Assays protocol (Fluidigm cat n° 100–6170 C1), using a 10X assays mix and a pre-sample mix prepared separately. The 10x assays mix was prepared by mixing 2 μL of combined Primer (Forward/Reverse 6.7uM, Probe 1.7 μM) and 2ul 2X Assay Loading Reagent (Fluidigm PN 100–7611) to a final volume of 4 μL (per reaction). The pre-sample mix was prepared by mixing 2 μL TaqMan Universal PCR Master Mix (2X) (Life Technologies PN 4304437) and 0.2 μL 20X GE Sample Loading Reagent (Fluidigm PN 100–7610) and 1.8 μL preamplified cDNA to a final volume of 4 μL.

Then, 3 μL of 10x assays mix and of pre-sample mix are transferred into the 192.24 IFC, loaded into the BiomarkTM IFC controller RX and transferred to the BiomarkTM HD apparatus. Thermal cycling conditions were as follows: 50°C for 120 s, 95°C for 600 s followed by 20 cycles of 95°C for 15 s, 60°C for 1min.

Detection of microRNA-expression protocol using the Biomark HD system analysis

RNAs (1 μL) were processed using the miScript 2 RT kit (QIAGEN cat n° 218161). Preamplification was performed using 5 μL of diluted cDNA using the mi Script microfluidics PreAMP kit (QIAGEN, cat n° 331455) and following the 384-plex preamplification cycling conditions (see manufacturer’s protocol). QPCR was performed following miScript microfluidics PCR kit (QIAGEN cat n° 331431) with commercially-available primers (miScript Primer assays). Six μL of assays mix and sample mix were prepared as described in the manufacturer’s protocol and 5 μL of each are loaded onto the 96.96 IFC plates (Fluidigm cat n° BMK-M-96.96). The list of microRNAs (miRNAs) tested in this study is presented in S1 Table.

Biomark HD system analysis

Exhaustive protocols are available at:

https://dx.doi.org/10.17504/protocols.io.bd3ii8ke (Two-Step qPCR Protocol)

https://dx.doi.org/10.17504/protocols.io.bnx4mfqw (One-Step qPCR Protocol)

Fluidigm HD system qPCR analysis

qPCR results were analyzed using the Real Time PCR analysis software provided by Fluidigm.

Statistical analysis

Statistical analyses were performed using GraphPad Prism software. Results are given as mean ± SEM. One-way ANOVA followed by Kruskal-Wallis test was used for multiple comparison. P-value < 0.05 was considered significant.

Results

Performance and validation of a Biomark-HD SARS-CoV-2 qPCR assay

We first set up a classical Fluidigm qPCR assay protocol that combines all primers/probe sets in a single reaction. This system requires a pre-amplification step after cDNA synthesis. This is due to the low volume of the IFC reaction chamber (9nl). As a result, and compared to the traditional RT-qPCR protocol (cDNA synthesis and qPCR), three consecutive reaction steps (RT, pre-amplification and qPCR) are required in the classical BiomarkTM -HD assay.

To evaluate the efficiency of the qPCR, we first assessed the test sensitivity using a range of synthetic viral N transcript dilutions with the US CDC primers/probe set N1. We found that the N transcript detection limit was seven transcript copies per reaction (Fig 1A). Linear regression showed a good correlation across Cq value and dilution series of N transcript (R2 = 0.9846) (Fig 1A). Similar sensitivity experiments were performed with additional primers/probes from the US CDC and by diluting the transcript in TE buffer without (S1A Fig) or with a RNA purification step (S1B Fig). Dilution of a SARS-CoV-2 positive clinical sample also indicated a large dynamic range of detection of 6 orders of magnitude for N1 and the E [20] primers/probe (Fig 1B). We then tested the combination of various primers/probe sets during the pre-amplification reaction, showing that 2 distinct mixes (mix #1: N1, E, human RNP; mix #2: ORF1ab, E, N and human RNP) gave similar data on clinical samples (Fig 1C). We finally evaluated the performance of the test on a set of 18 samples from positive patients showing a large range of Cq values. The initial reference analysis was performed using the Elitech GenefinderTM COVID-19 detection kit, comprising 3 sets of viral primers/probe targeting the same viral genes as the mix #2. We noticed excellent correlations between both methods for the detection of the 3 transcripts with R2 above 0.95 (Fig 1D), only affected by some high-range samples (i.e. Cq below 5), which displayed a saturation on BiomarkTM. Similarly, we also evaluated the performance of the test using the probe of the viral N gene on a set of 92 purified RNA samples of known status, comprising 15 positive samples, previously analyzed using the GeneFirst COVID-19 detection kit (comprising two sets of primers/probe including the N target). We plotted the correlation for the Cq obtained for the detection of the N transcript using both methods, showing a R2 of 0.97, with no false positive detected on the remaining 77 negative samples (S2 Fig).

Fig 1. Performance and validation of the Biomark-HD SARS-CoV-2 qPCR assay.

Fig 1

(A) Titration of the diluted (N) nucleocapside spike-in transcript. A serial dilution of the synthetic N transcript from 2.104 to 2 copies was performed and processed through the Biomark-HD protocol. Correlation and amplification curves of detected Cq values according to synthetic N transcript copy number obtained with N1 primers/probe are shown. (B) Amplification curves showing the range of detection of a SARS-CoV-2 positive clinical sample serial dilution with N1 and E primers/probe. (C) Typical amplification curves showing the combination of two primers/probe sets on 3 SARS-CoV-2 positive clinical samples. Mix #1: E, N1, human RNP. Mix #2: E, N, ORF1ab and human RNP. (D) Validation of the Biomark-HD protocol on a cohort of 18 biopsies from positive patients. The correlation of the Cq values obtained for the N, E and the ORF1ab genes are presented. The data presented are representative of at least two independent experiments performed in quadruplicate.

Optimization of the test

Additional tests were performed to reduce handling and reaction times by optimizing cycle numbers, elongation times and primer concentrations in the preamplification reaction (S3 Fig). In another optimization, we combined RT and pre-amplification into a one-step reaction, using the Cells Direct One-Step qRT-PCR kit (ThermoFisher Scientific). We compared this method to the classical Fluidigm RT-qPCR on a set of 18 clinical samples with a wide range of Cq values. Very similar Cq values were obtained with both methods for the cellular (RNP) and for 2 viral (E and N) primers/probe sets. The detection limit for ORF1ab gene was even improved using the Cells Direct One-Step qRT-PCR kit-based protocol (Fig 2A). Overall, this protocol offers the possibility to run 192 samples with 24 couples of primers/probe sets (using the Fluidigm 192.24 IFC) corresponding to 4608 end-points in less than 3 hours.

Fig 2. Optimization of the assay.

Fig 2

(A) Total RNA from 18 clinical samples with a wide range of SARS-CoV-2 infection were subjected to either a Two-step reaction (red circles; consecutive Reverse Transcription and Pre-amplification) or One-step reaction (blue circles; combined Reverse Transcription and Pre-amplification). Quantitative PCR reactions were performed on the Biomark-HD using cellular (RNP) and viral (N, E, ORF1ab) primers/probe sets. (B) Total RNA from 18 clinical samples with a wide range of SARS-CoV-2 infection were extracted with either the miRNeasy Advanced Serum Plasma Kit (red circles) or the Virus QIAamp Viral RNA kit (blue circles). RNAs were processed using the One step reaction. Quantitative PCR reactions were performed on the Biomark-HD using cellular (RNP) and viral (N, E, ORF1ab) primers/probe sets. The Cq presented are representative of two independent experiments performed in quadruplicate.

Alternative RNA extraction method using the miRNeasy Serum/Plasma Advanced Kit (Qiagen)

To limit the possible impact of a shortage of viral RNA extraction kits, we compared the performance of the QIAamp Viral mini kit to that of the miRNeasy Serum/Plasma Advanced Kit. A set of 18 SARS-CoV-2 positive samples with a wide range of Cq values was extracted using both protocols. The 36 resulting RNA samples were run in parallel on the BIOMARK-HD using 3 human and 3 viral primers/probe sets. Identical Cq values were obtained for the human probe (RNP) with both extraction protocols. A slight gain of signal (around 1–2 Cq) was measured after QIAamp Viral purification for the 3 viral primer/probes (E, N and ORF1ab) (Fig 2B). These data suggest that the QIAamp Viral RNA mini Kit slightly increases the efficiency of viral transcript extraction, but the miRNeasy Serum/Plasma Advanced Kit represents an acceptable alternative solution in case of shortage.

Use of the BiomarkTM HD-based protocol to analyze the host response to SARS-CoV-2 infection at the mRNA and miRNA levels

Exploiting the potential for multiplexing of the BiomarkTM HD-based technology, we included sets of human primers/probe to evaluate the expression of genes required for viral entry (ACE2, TMPRSS2), as well as inflammatory/antiviral response genes (CXCL8, IL1A, IL1B, IL6, IFNB1, IFIT1). A set of miRNAs (see S1 Table for detail) was also tested (Fig 3A). We divided the patient samples (n = 72) into groups, according to their viral load (strong, medium, weak or negative). ACE2 and TMPRSS2 were detected in all samples, regardless of the viral status, with Cq values close to those for RNP, suggesting they could represent valid human control genes to assess the quality and the presence of epithelial cells in the samples (Fig 3B and 3C). The signal obtained for the inflammation and interferon responses was more heterogeneous (Fig 3B). We particularly noticed a strong signal (in the range of 5 Cq) for IFIT1 in the 3 samples with the highest viral load (Cq values < 10 for the viral probes) (Fig 3C). Additionally, we evaluated the possibility to quantify miRNA levels in the same samples. A large number of miRNAs were detected in most of the samples (examples in Fig 3B and 3C). Overall, our results demonstrate that cellular markers can be easily quantified in nasopharyngeal swabs and may provide useful information to refine the diagnosis or prognosis of COVID-19 patients.

Fig 3. Use of the Biomark-based protocol to analyze the host response to SARS-CoV-2 infection at the mRNA and microRNA levels.

Fig 3

(A) Overview of the analysis strategy on 72 patient samples. They were divided into 4 groups, according to their viral load, from negative, weak (Cq for viral probes >20), medium (20 > Cq for viral probes >10) and strong (Cq for viral probes < 10) SARS-CoV-2 positive. (B) Typical amplification curves of the different genes (cellular, viral and micro-RNA) on three different SARS-CoV-2 patient status. (C) Modulation of cellular markers in the different groups of patients according to their SARS-CoV-2 viral load. IFIT1 expression was statistically elevated (p < 0.001) in the strong COVID-19-positive samples compared to the three other groups considered separately. The Cq presented are representative of two independent experiments performed in quadruplicate.

Detection of mutations associated with SARS-CoV-2 variants by RT-qPCR

We next assessed whether the potential of the BiomarkTM HD-based technology for multiplexing was adapted to the detection of SARS-CoV-2 variant in patient samples. For this purpose, a set of probes was designed to detect selectively SARS-CoV-2 genomes bearing the H69-V70 deletion (21765–21770) which emerged in several lineages of the virus, including the B.1.1.7/501Y.V1 variant and the 3675-GSF deletion (11288–11296) detected in all 501Y variants (B.1.1.7/501Y.V1, B.1.351/501Y.V2, P.1/501Y.V3) [21] (Fig 4). In a set of 74 COVID-19 patient samples tested in mid-January in the south of France, 11 samples showed loss of signal for wild-type spike and NSP6 probes while all the corresponding variant probes were detected, strongly suggesting the presence of the B.1.1.7 variant in these 11 samples. This experiment demonstrated that this methodology can be rapidly adapted according to SARS-CoV-2 evolution, allowing in a single assay the COVID-19 diagnosis as well as information regarding the presence of various variants of interest.

Fig 4. Use of the Biomark-based protocol to detect mutations associated with SARS-CoV-2 variants.

Fig 4

(A) Overview of the probes design used for the detection of two deletion mutations found in the Spike gene (H69-V70, top panel) and in the NSP6 gene (3675-SGF, bottom panel). Deletions are highlighted in yellow with both nucleotide at the 5’ and 3’ side shown in red and green respectively. Of note: for the H69-V70 mutation an A-to-G punctual deletion is shown in blue. Three mutant probes and one wild type probe were generated for the H69—V70 deletion (21765–21770) and one mutant and one wild type probes were generated for the 3675 –GSF deletion (11288–11296). The nucleotide numbering is based on the reference SARS-CoV-2 complete sequence (NCBI Reference Sequence: NC_045512.2). (B) Heatmap of Cq value from RT-qPCR using various combinations of primer/probes performed on 74 SARS-CoV-2 positive clinical samples. Cq scale range is shown on the right. Using the Biomark-based protocol, 11 positive samples for the mutation associated with the SARS-CoV-2 variants have been detected (on 74 positive clinical samples).

Direct qPCR detection of SARS-CoV-2 using inactivating lysis buffers-based protocols

Direct SARS-CoV-2 diagnostic methods have been described, mainly performing RT-qPCR directly on crude or heat-inactivated subject samples [810]. An additional refinement would be to improve the sensitivity and safety of these protocols by the use of RT-qPCR-compatible inactivating lysis buffers containing detergents such as Triton X-100, a standard non-ionic detergent widely used for inactivation of enveloped viruses such as SARS-CoV [22, 23]. We first compared a regular extraction-based method with a direct RT-qPCR protocol on a synthetic N transcript spike in TE buffer in the presence or the absence of 0.5% Triton X-100 and confirmed that Triton X-100 did not interfere with the efficiency of the RT-qPCR steps (S4A Fig). We then performed two consecutive nasopharyngeal swab samplings on a COVID-19 diagnosed patient, using either a regular VTM or a TE buffer containing 0.5% Triton X-100 followed or not by a heating step at 65°C for 10 min. As expected, the presence of Triton X-100 and the heating process did not affect the detection of the human RNP transcript for both RT-qPCR methods. Conversely, while the Triton X-100-lysis buffer/heating process slightly improved the signal for the virus N1 primers/probe using the regular extraction-based protocol, this treatment strongly inhibited the direct RT-qPCR method for the same cellular transcript (S4B Fig), indicating that 0.5% Triton X-100-lysis buffers are not compatible for a sensitive direct SARS-CoV-2 RT-qPCR assay.

We then tested the use of additional detergents and emulsifiers (Tween-20, Brij™-35, Brij™ O10), alone or in combination with polyethylene glycol (PEG600), in the presence or the absence of a treatment with proteinase K (PK), followed by heat inactivation. The experiment was performed on 2 clinical samples of known COVID-19 status, sampled in a commercial VTM medium, aliquoted and then treated in parallel using 16 distinct detergent conditions in the presence or absence of PK, for a total of 64 experimental conditions (Fig 5 and S5 Fig). Additional controls included the crude and heat-inactivated samples, the use of Quick ExtractTM DNA Extraction Solution (QE), which has been recently proposed as an alternative method to extraction [24, 25] as well as a regular extraction-based protocol (positive control). Three and five sets of SARS-CoV-2 (Fig 5A) and human (Fig 5B) primers/probe, respectively, were used in the same assay on a 192.24 dynamic array. In these conditions, a direct assay on a crude sample resulted in a 9 to 15-Cq increase, depending on the viral primers/probe set, when compared to the extraction method, with an even more pronounced effect when samples were heated 5 min at 95°C (Fig 5A). While the addition of 0.5% Triton X-100 alone was similar to the crude direct assay condition, the combination of Triton X-100 with the emulsifier Brij™ O10 (Oleth-10) improved the detection for all viral primers/probes, with a drop of around 5 Cq compared to Triton X-100 alone. All other detergents used alone (Tween-20, Brij™ O10, Brij™-35) gave a similar signal as 0.5% Triton X-100 alone. Of note, the addition of PK resulted in an improvement of sensitivity in the presence of various detergents including Tween-20, Brij™ O10 and Brij™-35 while addition of PEG had no beneficial effect (Fig 5 and S5 Fig).

Fig 5. Detection of viral and cellular genes using inactivating lysis buffers-based protocols.

Fig 5

Two clinical samples of known COVID-19 status (one positive and one negative) sampled in a commercial VTM medium, were aliquoted and then treated in parallel using the indicated detergent conditions (selected from S5 Fig) in the presence or absence of PK. RNA was extracted as a control. Biomark-HD RT-qPCR was performed using (A) viral (N, E, ORF1ab) and (B) cellular (ACE2, IFIT1, IL6, TMPRSS and RNP) primers/probe sets. Legend: +: 10 min at 65°C; ++: 10 min at 65°and 5 min at 95°C; TX: Triton X-100; QE: Quick ExtractTM DNA Extraction Solution. The Cq presented are representative of two independent experiments performed in quadruplicate.

Based on these results, we selected two of the best direct detergents based assays (Triton X-100/Brij™ O10 and Tween-20/PK) and compared the sensitivity of these methods with a crude direct assay and the extraction method on additional clinical samples. For all direct assays, all samples were heat-inactivated at 95°C (5 min). Moreover, we used a set of 17 clinical samples from SARS-CoV-2-diagnosed patients collected in saline solution with a wide range of Cq values. To control pH conditions and limit RNA degradation, all samples were diluted in TE buffer. We plotted the relationships between Cq obtained for the detection of the N and ORF1ab primers/probe sets for the Biomark assay and the GeneFirst COVID-19 detection kit (Fig 6). As expected, an excellent correlation was obtained when comparing the RNA extraction-based Fluidigm protocol with the reference assay (R2 = 0.8258 and R2 = 0.8208, all samples detected). Conversely to the data obtained with the commercial VTM, the best direct protocol corresponded to the crude assay (TE), approaching a very similar performance to that obtained with the extraction-based assay (R2 = 0.6447 and R2 = 0.6782, 16/17 samples detected). The R2 values dropped in a dramatic way for the two direct detergents based assays, with 11 to 14 positive samples detected out of 17. Overall, these data suggest that a direct qPCR method using saline as a VTM and a basic TE buffer followed by a 5 min inactivation step at 95°C can efficiently support detection of SARS-CoV-2.

Fig 6. Correlation of the detection of two viral genes (N, ORF1ab) obtained using a commercial COVID-19 detection kit and extraction-based or direct Biomark-HD assays.

Fig 6

We used a set of 17 clinical samples from SARS-CoV-2-diagnosed patients collected in saline solution with a wide range of Cq values. To control pH conditions and limit RNA degradation, all samples were diluted 2X in TE buffer. RNA was extracted (as a control) or samples were treated using Tween 20 and PK (Tw + PK) or Triton X-100 and Brij010 detergent solutions (Tx + Brij010). The different RT-qPCR Biomark methods were compared with the GeneFirstTM COVID-19 method as a reference. The correlation between the Cq obtained with the two methods is presented. One negative sample was processed in the three direct conditions (TE, TE plus Tw + PK and TE plus Tx + Brij010) with as results an absence of signal with the two probes (Cq >40). Double Pos: double positive.

Discussion

The present paper describes a reliable and flexible multiplex nanofluidic qPCR system-based protocol to detect SARS-CoV-2. Its versatile format makes it easily adaptable to detect mutations associated with SARS-CoV-2 variants, multiple other pathogens and/or host cellular markers. The same run allows to test in parallel viral and host RNAs, including miRNAs. We show a high concordance between this method and clinically approved traditional qPCR tests. This assay addressed some of the challenges of RT-qPCR assays, including analyzing a larger number of reactions per run, making the assay more cost-effective and less time-consuming. Further, IFCs dynamic arrays not only reduces the reaction volume from about 10 μL down to the 10 nL scale, but allows large multiplexing as well as increased parallelization throughput of qPCR reactions.

The flexibility of this platform may be used in biomarker studies aiming at predicting at diagnosis the severity (requirement or not for hospitalization/intensive care) or the length (development of so-called “long-COVID”) of the disease. Several studies have depicted an elevated innate and adaptive immune activation in severe COVID-19 patients [2628], and a differential immune phenotype in moderate versus severe disease after the second week of infection [28, 29]. The Biomark HD offers the possibility to extend the number of genes tested to biomarkers including pro-inflammatory cytokines, chemokines, interferons, tissue repair genes and miRNAs by nasopharyngeal swabs or in saliva.

Monitoring miRNAs expression levels in swabs or saliva during the course of infection might be of great interest, because of their stability in biological fluids and frequent alterations during both chronic and acute airway infections [30, 31]. For instance, small RNA profiling of in vitro SARS-CoV-2 infected cell lines resulted in an increased expression of miR-155, one of the miRNAs detected in our clinical samples (Fig 3C) [32]. This miRNA, frequently associated with virus infection, was recently shown to contribute to the development of lethal acute respiratory distress syndrome (ARDS) in H1N1 influenza A virus-infected mice, suggesting it could represent a valuable biomarker to follow the individual immune response of patients [33]. Additionally, several other miRNAs targeting ACE2 and TMPRSS2 might influence SARS-CoV-2 entry and the course of the disease, as recently suggested [34].

With the emergence of new SARS-CoV-2 variants that are potentially associated with increased transmissibility, immune escape and/or vaccine efficacy reduction [3537], it is crucial to make available a test that determines whether a patient is SARS-CoV-2-positive but also identified the variant type. We demonstrate here that the BiomarkTM HD-platform is particularly adapted to detect the presence of known variants in patient samples. Such a methodology, which is more cost-effective and flexible than sequencing, is likely to have implication in the epidemiological monitoring of the pandemic.

An important limitation of current tests is the requirement for RNA extraction that constitutes an obstacle to scale-up the capacity of testing both in term of time and cost. Several groups have explored methods to circumvent RNA extraction by performing RT-qPCR directly on crude or heat-inactivated clinical samples [810]. Overall these methods show that testing for SARS-CoV-2 infection can be performed without RNA extraction, with a limited loss in accuracy for determining negative and positive cases. While this procedure is simple and attractive, it might be improved by the addition of detergents to facilitate viral capsid lysis to release genomic RNA and also directly inactivate the virus to facilitate sample handling and safety. Several methods and commercial kits have developed approaches to lyse efficiently mammalian cells and directly perform RT-qPCR or RNA-seq libraries. However, these methods are not fully optimized for virus lysis, which requires increased concentrations of detergents. Some studies indicate that Triton X-100, widely used in virus inactivation procedures, or Tween-20 may slightly improve or at least not interfere with the RT-qPCR SARS-CoV-2 direct testing of nasopharyngeal swabs or saliva [9, 16]. However, these initial reports clearly mentioned that additional efforts were needed to optimize direct RT-qPCR assays on detergent-inactivated samples. We evaluated here the use of several detergents and emulsifiers (Triton X-100, Tween-20, Brij™-35, Brij™ O10), alone or in combination, in the presence or the absence of a treatment with PK or polyethylene glycol (PEG) followed by heat inactivation to assess their compatibility with a direct reverse transcription enzymatic step. This initial screening indicated that the addition of each of these detergents, alone, did not improve the assay compared to a direct assay performed on a crude commercial VTM sample. By contrast, the addition of PK improved the detection with all the detergents tested, in agreement with a recent study [38]. Of note, a combination of Triton X-100 with Brij™ O10, in the absence of PK, had a similar effect, by decreasing the Cqs for the 3 SARS-CoV-2 primers/probes. (Fig 4A) Brij™ O10 contains Oleth-10, a polyoxyethylene oleyl ether used in aqueous emulsions that may increase solubilization of the viral capsid or contribute to stabilization of the emulsion. When we tested the three best direct protocols on a cohort of samples collected in saline solution, we could not confirm the detection improvement observed in the screening assay using both lysis buffers (Tween-20/PK and Triton X-100/Brij™ O10, Fig 5). This discrepancy may arise from the fact that the screened samples were collected in a commercial VTM while the cohort experiment was performed on biopsies collected in saline buffer. This difference suggests a complex interaction between the various constituents of this VTM (pH, nature of the medium, presence of albumin, gelatin, anti-bacterial agents, for examples), and the different detergents and emulsifiers tested. Further developments are definitely required to fine tune specific combinations of detergents and emulsifiers with specific VTMs. In any case, our data clearly indicate that a direct qPCR method using saline as a VTM and a basic TE buffer followed by a 5 min inactivation step at 95°C shows almost identical performance to that of a classical extraction-based assay, allowing notably the identification of samples with high Cq values (Fig 5). Considering that SARS-CoV-2 remains detectable in phosphate buffer for up to a month when stored at various temperatures [39], our data supports the use of PBS, as an alternative to VTM for direct SARS-CoV-2 testing. We further confirm and refine previous studies showing that testing for SARS-CoV-2 infection can be performed with simplified protocols omitting RNA extraction steps without major loss in accuracy [8, 9]. Based on the data presented here, the simplest sample collection and preparation for direct RT-qPCR COVID-19 test would be to sample the swab into a small volume (around 0.5 mL) of TE buffer at a pH of 7.0 to limit dilution of the virus, supplemented with yeast tRNA or RNAse inhibitors to increase RNA stability. This type of buffer was recently shown to have an excellent capacity to preserve the SARS-CoV-2 signal [9, 39]. Lysis buffers containing detergents would provide the possibility to fully inactivate the virus, allowing a rapid and safe handling of the clinical samples, but our data indicate that their efficiency may vary depending on clinical sample processing and should be further optimized.

In conclusion, we propose that such a direct qPCR procedure offers an interesting option for massively scaling up SARS-CoV-2 testing. We believe that this protocol can be particularly well-adapted for pooling approaches [40], to screen asymptomatic individuals in communities at risk. It would provide an additional tool to enhance testing capacity and affordability across the world, as widely recommended by the Health Community worldwide, such as the “all-in” approach to testing recently proposed [4].

Supporting information

S1 Fig. Titration of the diluted nucleocapsid spike-in transcript.

A ten-fold serial dilution ranging from 1 to 10–6 was prepared from the stock solution of the in vitro-transcribed N gene and supplemented with 2 ng/μL of total RNA from HEK 293 Cells. Reverse Transcription was performed followed by 15 cycles of pre-Amplification and 30 cycles of qPCR. The RT-qPCR reaction was performed without (A) or with (B) a RNA purification step. Linear regression was performed by logarithmic plots of transcript copy number against Cq value. We observed a good correlation according Cq linear regression curves according to dilution for the three viral CDC primers/probe sets (N1, N2, N3). No Cq value has been detected for the E primers/probe. RNP, used as internal control, shows constant detection of Cq value, suggesting a good performance of the qPCR.

(TIF)

S2 Fig. Validation of the BiomarkTM -HD protocol on a cohort of 92 biopsies including 15 positive patients.

The correlation of the Cq values obtained for the N primers/probe (BiomarkTM -HD) and the GeneFirst COVID-19 detection kit is presented.

(TIF)

S3 Fig. Preamplification step optimization.

The elongation time used in the preamplification reaction was reduced to 1 min at 60°C (A) from 2 min at 60°C (B) using diluted total RNA from a SARS-nCov2 positive patient sample.

(TIF)

S4 Fig. Effect of Triton X-100 on extraction-based and direct RT-qPCR protocol performed on a synthetic N transcript or a clinical sample.

A. In vitro-transcribed viral N gene was added to either a Triton X-100 containing lysis buffer (Tx) or to TE buffer (TE). The samples were heated or not at 65°C for 10 min. RNA extraction was performed or not (direct) and N1 or RNP levels were determined by RT-qPCR using the BiomarkTM -HD system. B. A similar protocol as in A was used but the starting material were two consecutive sample collection from the same patient processed either in a Triton X-100 containing lysis buffer or to TE buffer.

(TIF)

S5 Fig. Heatmap of Cq values from direct RT-qPCR experiments using various combinations of detergent on clinical samples.

Samples from a positive- or negative-COVID-19 patient collected in a commercial VTM was mixed with different combinations of detergents/emulsifiers, in presence or absence of PK (2 mg/mL) and further heat at 95°C for 5 min or not. Cq values obtained in quadruplicate are presented. Tx: Triton X100; PEG: poly ethylene glycol 600; QE: Quick ExtractTM DNA Extraction Solution.

(TIF)

S1 Table. List of the micro-RNA tested using the Biomark HD system.

(DOCX)

Acknowledgments

We thank our colleagues from the IPMC and the Délégation CNRS Côte d’Azur for their help and support, especially Catherine Lecalvez, Michel Bordes, Véronique Campbell and Simon Szmidt for administrative support and gratefully acknowledge the staff from the Département de Pneumologie of the Nice Hospital, especially Jennifer Griffonet and Charlotte Maniel. We also thank the technical support of the UCA GenomiX platform and MICA imaging facility of the University Côte d’Azur. We are grateful to Bayer SAS for loan of equipment and reagents, to Hervé Groux (Immunosearch) as well as to Dr Pol-Henri Guivarch and Alexandre Romain (Agence Régionale De Santé PACA) for their support and interesting discussion. We are also grateful to the “Propagate” consortium, notably Syril D. Pettit, Emily A. Bruce, Jason W Botten and Keith R. Jerome for their support and fruitful discussion.

Data Availability

All data are provided in supplemental information.

Funding Statement

Supported by funds from the “Centre National de la Recherche Scientifique” (CNRS), the “Université Côte d’Azur”, the French “French Defence Innovation Agency – Agence de l’Innovation de Défense (“project “Safe and direct COV-2 qPCR Test”) and the Département des Alpes Maritimes (COVID-19 Health program). JF is supported by the Cancéropole PACA and CL is supported by Plan Cancer 2018 « ARN non-codants en cancérologie: du fondamental au translationnel » (number 18CN045). P.B. is recipient of a chair from ANR-19-P3IA-0002–3IA Côte d'Azur–Nice–Interdisciplinary Institute for Artificial Intelligence. The Biomark equipment was funded by Canceropole PACA and France Génomique (Commissariat aux Grands Investissements: ANR-10-INBS-6 09–03, ANR-10-INBS-09–02). The funders had no role in study design, data collection and analysis, decision to publish, or preparation of the manuscript. Bayer and LBM Bioesterel provided support in the form of salaries for authors [DR and VL], but did not have any additional role in the study design, data collection and analysis, decision to publish, or preparation of the manuscript. The specific roles of these authors are articulated in the ‘author contributions’ section. Bayer SAS also loaned equipment and reagents for the study.

References

  • 1.La Marca A, Capuzzo M, Paglia T, Roli L, Trenti T, Nelson SM. 2020. Testing for SARS-CoV-2 (COVID-19): a systematic review and clinical guide to molecular and serological in-vitro diagnostic assays. Reprod Biomed Online 41:483–499. 10.1016/j.rbmo.2020.06.001 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 2.Ghaffari A, Meurant R, Ardakani A. 2020. COVID-19 Serological Tests: How Well Do They Actually Perform? Diagnostics 10:453. 10.3390/diagnostics10070453 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3.Lieberman JA, Pepper G, Naccache SN, Huang M-L, Jerome KR, Greninger AL. 2020. Comparison of Commercially Available and Laboratory-Developed Assays for In Vitro Detection of SARS-CoV-2 in Clinical Laboratories. J Clin Microbiol 58. 10.1128/JCM.00821-20 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4.Pettit SD, Jerome KR, Rouquié D, Mari B, Barbry P, Kanda Y, et al. 2020. ‘All In’: a pragmatic framework for COVID‐19 testing and action on a global scale. EMBO Mol Med 12. 10.15252/emmm.202012634 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5.Lohse S, Pfuhl T, Berkó-Göttel B, Rissland J, Geißler T, Gärtner B, et al. 2020. Pooling of samples for testing for SARS-CoV-2 in asymptomatic people. Lancet Infect Dis 20:1231–1232. 10.1016/S1473-3099(20)30362-5 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6.Olwagen CP, Adrian P V., Madhi SA 2019. Performance of the Biomark HD real-time qPCR System (Fluidigm) for the detection of nasopharyngeal bacterial pathogens and Streptococcus pneumoniae typing. Sci Rep 9:6494. 10.1038/s41598-019-42846-y [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7.Goecke NB, Krog JS, Hjulsager CK, Skovgaard K, Harder TC, Breum SØ, et al. 2018. Subtyping of Swine Influenza Viruses Using a High-Throughput Real-Time PCR Platform. Front Cell Infect Microbiol 8. 10.3389/fcimb.2018.00165 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8.Bruce EA, Huang M-L, Perchetti GA, Tighe S, Laaguiby P, Hoffman JJ, et al. 2020. Direct RT-qPCR detection of SARS-CoV-2 RNA from patient nasopharyngeal swabs without an RNA extraction step. PLOS Biol 18:e3000896. 10.1371/journal.pbio.3000896 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9.Smyrlaki I, Ekman M, Lentini A, Rufino de Sousa N, Papanicolaou N, Vondracek M, et al. 2020. Massive and rapid COVID-19 testing is feasible by extraction-free SARS-CoV-2 RT-PCR. Nat Commun 11:4812. 10.1038/s41467-020-18611-5 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10.Beltrán-Pavez C, Alonso-Palomares LA, Valiente-Echeverría F, Gaggero A, Soto-Rifo R, Barriga GP. 2021. Accuracy of a RT-qPCR SARS-CoV-2 detection assay without prior RNA extraction. J Virol Methods 287:113969. 10.1016/j.jviromet.2020.113969 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11.Fomsgaard AS, Rosenstierne MW. 2020. An alternative workflow for molecular detection of SARS-CoV-2 –escape from the NA extraction kit-shortage, Copenhagen, Denmark, March 2020. Eurosurveillance 25:1–8. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12.Grant PR, Turner MA, Shin GY, Nastouli E, Levett LJ. 2020. Extraction-free COVID-19 (SARS-CoV-2) diagnosis by RT- PCR to increase capacity for national testing programmes during a pandemic. bioRxiv 10.1101/2020.04.06.028316. [DOI] [Google Scholar]
  • 13.Hasan MR, Mirza F, Al-Hail H, Sundararaju S, Xaba T, Iqbal M, et al. 2020. Detection of SARS-CoV-2 RNA by direct RT-qPCR on nasopharyngeal specimens without extraction of viral RNA. PLoS One 15:e0236564. 10.1371/journal.pone.0236564 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 14.Fukumoto T, Iwasaki S, Fujisawa S, Hayasaka K, Sato K, Oguri S, et al. 2020. Efficacy of a novel SARS-CoV-2 detection kit without RNA extraction and purification. Int J Infect Dis 98:16–17. 10.1016/j.ijid.2020.06.074 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15.Wei S, Kohl E, Djandji A, Morgan S, Whittier S, Mansukhani M, et al. 2020. Direct diagnostic testing of SARS-CoV-2 without the need for prior RNA extraction. medRxiv 10.1101/2020.05.28.20115220 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 16.Ranoa DRE, Holland RL, Alnaji FG, Green KJ, Wang L, Christopher B, et al. 2020. Saliva-Based Molecular Testing for SARS-CoV-2 that Bypasses RNA Extraction. bioRxiv 10.1101/2020.04.22.056283. [DOI] [Google Scholar]
  • 17.Srivatsan S, Ha PD, van Raay K, Wolf CR, McCulloch DJ, Kim AE, et al. 2020. Preliminary support for a “dry swab, extraction free” protocol for SARS-CoV-2 testing via RT-qPCR. bioRxiv 10.1101/2020.04.22.056283 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18.Zhen W, Berry GJ. 2020. Development of a New Multiplex Real-Time RT-PCR Assay for Severe Acute Respiratory Syndrome Coronavirus 2 (SARS-CoV-2) Detection. J Mol Diagnostics 22:1367–1372. 10.1016/j.jmoldx.2020.09.004 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19.Touron P, Siatka C, Pussiau A, Follot S, Fritz T, Petit M, et al. 2020. A mobile DNA laboratory for forensic science adapted to coronavirus SARS-CoV-2 diagnosis. Eur J Clin Microbiol Infect Dis 10.1007/s10096-020-03989-3 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 20.Corman VM, Landt O, Kaiser M, Molenkamp R, Meijer A, Chu DKW, et al. 2020. Detection of 2019 novel coronavirus (2019-nCoV) by real-time RT-PCR. Eurosurveillance 25. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 21.Leung K, Shum MH, Leung GM, Lam TT, Wu JT. 2021. Early transmissibility assessment of the N501Y mutant strains of SARS-CoV-2 in the United Kingdom, October to November 2020. Eurosurveillance 26. 10.2807/1560-7917.ES.2020.26.1.2002106 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 22.Darnell MER, Taylor DR. 2006. Evaluation of inactivation methods for severe acute respiratory syndrome coronavirus in noncellular blood products. Transfusion 46:1770–1777. 10.1111/j.1537-2995.2006.00976.x [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 23.Colavita F, Quartu S, Lalle E, Bordi L, Lapa D, Meschi S, et al. 2017. Evaluation of the inactivation effect of Triton X-100 on Ebola virus infectivity. J Clin Virol 86:27–30. 10.1016/j.jcv.2016.11.009 [DOI] [PubMed] [Google Scholar]
  • 24.Sentmanat M, Kouranova E, Cui X. 2020. One-step RNA extraction for RT-qPCR detection of 2019-nCoV. bioRxiv 10.1101/2020.04.02.022384. [DOI] [Google Scholar]
  • 25.Ladha A, Joung J, Abudayyeh OO, Gootenberg JS, Zhang F. 2020. A 5-min RNA preparation method for COVID-19 detection with RT-qPCR. Medrxiv 1–3. [Google Scholar]
  • 26.Blanco-Melo D, Nilsson-Payant BE, Liu W-C, Uhl S, Hoagland D, Møller R, et al. 2020. Imbalanced Host Response to SARS-CoV-2 Drives Development of COVID-19. Cell 181:1036–1045.e9. 10.1016/j.cell.2020.04.026 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 27.Zhou Z, Ren L, Zhang L, Zhong J, Xiao Y, Jia Z, et al. 2020. Heightened Innate Immune Responses in the Respiratory Tract of COVID-19 Patients. Cell Host Microbe 27:883–890.e2. 10.1016/j.chom.2020.04.017 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 28.Lucas C, Wong P, Klein J, Castro TBR, Silva J, Sundaram M, et al. 2020. Longitudinal analyses reveal immunological misfiring in severe COVID-19. Nature 584:463–469. 10.1038/s41586-020-2588-y [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 29.Hadjadj J, Yatim N, Barnabei L, Corneau A, Boussier J, Smith N, et al. 2020. Impaired type I interferon activity and inflammatory responses in severe COVID-19 patients. Science (80-) 369:718–724. 10.1126/science.abc6027 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 30.Mirzaei R, Mahdavi F, Badrzadeh F, Hosseini-Fard SR, Heidary M, Jeda AS, et al. 2021. The emerging role of microRNAs in the severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) infection. Int Immunopharmacol 90:107204. 10.1016/j.intimp.2020.107204 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 31.Cortez MA, Bueso-Ramos C, Ferdin J, Lopez-Berestein G, Sood AK, Calin GA. 2011. MicroRNAs in body fluids—the mix of hormones and biomarkers. Nat Rev Clin Oncol 8:467–477. 10.1038/nrclinonc.2011.76 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 32.Emanuel W, Kirstin M, Vedran F, Asija D, Theresa GL, Roberto A, et al. 2021. Transcriptomic profiling of SARS-CoV-2 infected human cell lines identifies HSP90 as target for COVID-19 therapy. iScience 102151. 10.1016/j.isci.2021.102151 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 33.Woods PS, Doolittle LM, Rosas LE, Nana-Sinkam SP, Tili E, Davis IC. 2020. Increased expression of microRNA-155-5p by alveolar type II cells contributes to development of lethal ARDS in H1N1 influenza A virus-infected mice. Virology 545:40–52. 10.1016/j.virol.2020.03.005 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 34.Nersisyan S, Shkurnikov M, Turchinovich A, Knyazev E, Tonevitsky A. 2020. Integrative analysis of miRNA and mRNA sequencing data reveals potential regulatory mechanisms of ACE2 and TMPRSS2. PLoS One 15:e0235987. 10.1371/journal.pone.0235987 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 35.Tegally H, Wilkinson E, Giovanetti M, Iranzadeh A, Fonseca V, Giandhari J, et al. 2020. Emergence and rapid spread of a new severe acute respiratory syndrome-related coronavirus 2 (SARS-CoV-2) lineage with multiple spike mutations in South Africa. medRxiv 2. [Google Scholar]
  • 36.Volz E, Mishra S, Chand M, Barrett JC, Johnson R, Hopkins S, et al. 2021. Transmission of SARS-CoV-2 Lineage B.1.1.7 in England: Insights from linking epidemiological and genetic data. medRxiv 2020.12.30.20249034. [Google Scholar]
  • 37.Wibmer CK, Ayres F, Hermanus T, Madzivhandila M, Kgagudi P, Lambson BE, et al. 2021. SARS-CoV-2 501Y.V2 escapes neutralization by South African COVID-19 donor plasma. bioRxiv. [DOI] [PubMed] [Google Scholar]
  • 38.Lalli MA, Langmade SJ, Chen X, Fronick CC, Sawyer CS, Burcea LC, et al. 2020. Rapid and extraction-free detection of SARS-CoV-2 from saliva by colorimetric reverse-transcription loop-mediated isothermal amplification. Clin Chem 2:1–28. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 39.Rodino KG, Espy MJ, Buckwalter SP, Walchak RC, Germer JJ, Fernholz E, et al. 2020. Evaluation of Saline, Phosphate-Buffered Saline, and Minimum Essential Medium as Potential Alternatives to Viral Transport Media for SARS-CoV-2 Testing. J Clin Microbiol 58. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 40.Perchetti GA, Sullivan K-W, Pepper G, Huang M-L, Breit N, Mathias P, et al. 2020. Pooling of SARS-CoV-2 samples to increase molecular testing throughput. J Clin Virol 131:104570. 10.1016/j.jcv.2020.104570 [DOI] [PMC free article] [PubMed] [Google Scholar]

Decision Letter 0

Tzong-Yueh Chen

22 Jan 2021

PONE-D-20-36827

Versatile and flexible microfluidic qPCR test for high-throughput SARS-CoV-2 and cellular response detection in nasopharyngeal swab samples

PLOS ONE

Dear Dr. Mari,

Thank you for submitting your manuscript to PLOS ONE. After careful consideration, we feel that it has merit but does not fully meet PLOS ONE’s publication criteria as it currently stands. Therefore, we invite you to submit a revised version of the manuscript that addresses the points raised during the review process.

Please submit your revised manuscript by Mar 07 2021 11:59PM. If you will need more time than this to complete your revisions, please reply to this message or contact the journal office at plosone@plos.org. When you're ready to submit your revision, log on to https://www.editorialmanager.com/pone/ and select the 'Submissions Needing Revision' folder to locate your manuscript file.

Please include the following items when submitting your revised manuscript:

  • A rebuttal letter that responds to each point raised by the academic editor and reviewer(s). You should upload this letter as a separate file labeled 'Response to Reviewers'.

  • A marked-up copy of your manuscript that highlights changes made to the original version. You should upload this as a separate file labeled 'Revised Manuscript with Track Changes'.

  • An unmarked version of your revised paper without tracked changes. You should upload this as a separate file labeled 'Manuscript'.

If you would like to make changes to your financial disclosure, please include your updated statement in your cover letter. Guidelines for resubmitting your figure files are available below the reviewer comments at the end of this letter.

If applicable, we recommend that you deposit your laboratory protocols in protocols.io to enhance the reproducibility of your results. Protocols.io assigns your protocol its own identifier (DOI) so that it can be cited independently in the future. For instructions see: http://journals.plos.org/plosone/s/submission-guidelines#loc-laboratory-protocols

We look forward to receiving your revised manuscript.

Kind regards,

Tzong-Yueh Chen, Ph.D.

Academic Editor

PLOS ONE

Journal Requirements:

When submitting your revision, we need you to address these additional requirements.

1. Please ensure that your manuscript meets PLOS ONE's style requirements, including those for file naming. The PLOS ONE style templates can be found at

https://journals.plos.org/plosone/s/file?id=wjVg/PLOSOne_formatting_sample_main_body.pdf and

https://journals.plos.org/plosone/s/file?id=ba62/PLOSOne_formatting_sample_title_authors_affiliations.pdf

2. Please include captions for your Supporting Information files at the end of your manuscript, and update any in-text citations to match accordingly. Please see our Supporting Information guidelines for more information: http://journals.plos.org/plosone/s/supporting-information.

3.Thank you for stating the following in the Acknowledgments Section of your manuscript:

"Supported by funds from the “Centre National de la Recherche Scientifique” (CNRS), the “Université Côte d’Azur”, the French “French Defence Innovation Agency – Agence de l’Innovation de Défense “ (project “Safe and direct COV-2 qPCR Test”) and the Département des Alpes Maritimes (COVID-19 Health program). JF is supported by the Cancéropole PACA and CL is supported by Plan Cancer 2018 « ARN non-codants en cancérologie: du fondamental au translationnel » (number 18CN045). The Biomark equipment was funded by Canceropole PACA and France Génomique (Commissariat aux Grands Investissements: ANR-10-INBS-6 09–03, ANR-10-INBS-09–02)."

We note that you have provided funding information that is not currently declared in your Funding Statement. However, funding information should not appear in the Acknowledgments section or other areas of your manuscript. We will only publish funding information present in the Funding Statement section of the online submission form.

Please remove any funding-related text from the manuscript and let us know how you would like to update your Funding Statement. Currently, your Funding Statement reads as follows:

 " The funders had no role in study design, data collection and analysis, decision to publish, or preparation of the manuscript."

Please include your amended statements within your cover letter; we will change the online submission form on your behalf.

4.Thank you for stating the following in the Competing Interests section:

"No authors have competing interests"

We note that one or more of the authors are employed by a commercial company: Bayer SAS and LBM BIOESTEREL

a) Please provide an amended Funding Statement declaring this commercial affiliation, as well as a statement regarding the Role of Funders in your study. If the funding organization did not play a role in the study design, data collection and analysis, decision to publish, or preparation of the manuscript and only provided financial support in the form of authors' salaries and/or research materials, please review your statements relating to the author contributions, and ensure you have specifically and accurately indicated the role(s) that these authors had in your study. You can update author roles in the Author Contributions section of the online submission form.

Please also include the following statement within your amended Funding Statement.

“The funder provided support in the form of salaries for authors [insert relevant initials], but did not have any additional role in the study design, data collection and analysis, decision to publish, or preparation of the manuscript. The specific roles of these authors are articulated in the ‘author contributions’ section.”

If your commercial affiliation did play a role in your study, please state and explain this role within your updated Funding Statement.

b) Please also provide an updated Competing Interests Statement declaring this commercial affiliation along with any other relevant declarations relating to employment, consultancy, patents, products in development, or marketed products, etc.  

Within your Competing Interests Statement, please confirm that this commercial affiliation does not alter your adherence to all PLOS ONE policies on sharing data and materials by including the following statement: "This does not alter our adherence to  PLOS ONE policies on sharing data and materials.” (as detailed online in our guide for authors http://journals.plos.org/plosone/s/competing-interests) . If this adherence statement is not accurate and  there are restrictions on sharing of data and/or materials, please state these. Please note that we cannot proceed with consideration of your article until this information has been declared.

Please include both an updated Funding Statement and Competing Interests Statement in your cover letter. We will change the online submission form on your behalf.

Please know it is PLOS ONE policy for corresponding authors to declare, on behalf of all authors, all potential competing interests for the purposes of transparency. PLOS defines a competing interest as anything that interferes with, or could reasonably be perceived as interfering with, the full and objective presentation, peer review, editorial decision-making, or publication of research or non-research articles submitted to one of the journals. Competing interests can be financial or non-financial, professional, or personal. Competing interests can arise in relationship to an organization or another person. Please follow this link to our website for more details on competing interests: http://journals.plos.org/plosone/s/competing-interests

Additional Editor Comments:

1. The data cannot support the conclusions. PLOS ONE is designed to communicate primary scientific research, and welcome submissions in any applied discipline that will contribute to the base of scientific knowledge. But the data of this manuscript cannot support the conclusions.

2. This manuscript needs to adhere the PLOS Data Policy. The authors need to make all methods, materials and data underlying the findings in their manuscript fully available.

3. The method must be of use to the community and must present a proven advantage over existing alternatives. If similar options already exist, the submitted manuscript must demonstrate that the new method is an improvement over existing options in some way. This requirement may be met by including a proof of principle experiment or analysis.

4. This manuscript has the statistical analysis problem.

5. The revised manuscript needs to address each of the comments of the reviewers.

[Note: HTML markup is below. Please do not edit.]

Reviewers' comments:

Reviewer's Responses to Questions

Comments to the Author

1. Is the manuscript technically sound, and do the data support the conclusions?

The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented.

Reviewer #1: Yes

Reviewer #2: Yes

**********

2. Has the statistical analysis been performed appropriately and rigorously?

Reviewer #1: Yes

Reviewer #2: No

**********

3. Have the authors made all data underlying the findings in their manuscript fully available?

The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified.

Reviewer #1: Yes

Reviewer #2: Yes

**********

4. Is the manuscript presented in an intelligible fashion and written in standard English?

PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here.

Reviewer #1: Yes

Reviewer #2: Yes

**********

5. Review Comments to the Author

Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters)

Reviewer #1: In this study, Dr. Bernard Mari and colleagues addressed a significant question regarding the two major obstacles of current COVID-19 test, the reagent shortage and tedious process of sample preparation. The authors delicately evaluated current approaches with a high-throughput platform in clinical samples. Overall, the data presented in the manuscript are of high quality with several important connections demonstrated, including accuracy and sensitivity. This manuscript should be great interest to a general audience, especially those countries are suffering from the massive COVID-19 test loading. In conclusion, I would suggest the editor directly accept this manuscript without any further revision.

Reviewer #2: The authors demonstrated a way to multiplex and bypass the RNA extraction for SARS-CoV-2 detection. The application of the IFC method is new and serve the purpose. However, the sample number of SARS-CoV-2 patients is low and damper the conclusive findings. I have several comments.

Specific comments

1. Increase the sample size of SARS-CoV-2 patients if possible.

2. Author should explain why to detect miRNA and the importance.

3. The label in figure 5 second row is quite confusing. Maybe change to VTM.

I don't understand the labeling in the figure 5 right part. DP? N? How you calculate sensitivity, by adding?

4. I won't say the 11 to 14 out of 17 is high. It is not acceptable for any applications.

5. The specificity was not evaluated in the lysis buffer and direct assay. Please revise.

6. The picture quality of the supplementary figures are very poor.

7. Please check all the labeling in text and figures and be consistent, such as SARS-CoV-2 and COVID-19.

8. Is there any conflict of interest?

**********

6. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files.

If you choose “no”, your identity will remain anonymous but your review may still be made public.

Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy.

Reviewer #1: No

Reviewer #2: No

[NOTE: If reviewer comments were submitted as an attachment file, they will be attached to this email and accessible via the submission site. Please log into your account, locate the manuscript record, and check for the action link "View Attachments". If this link does not appear, there are no attachment files.]

While revising your submission, please upload your figure files to the Preflight Analysis and Conversion Engine (PACE) digital diagnostic tool, https://pacev2.apexcovantage.com/. PACE helps ensure that figures meet PLOS requirements. To use PACE, you must first register as a user. Registration is free. Then, login and navigate to the UPLOAD tab, where you will find detailed instructions on how to use the tool. If you encounter any issues or have any questions when using PACE, please email PLOS at figures@plos.org. Please note that Supporting Information files do not need this step.

PLoS One. 2021 Apr 14;16(4):e0243333. doi: 10.1371/journal.pone.0243333.r002

Author response to Decision Letter 0


5 Mar 2021

Response to Reviewer's comments

- Reviewer #1: In this study, Dr. Bernard Mari and colleagues addressed a significant question regarding the two major obstacles of current COVID-19 test, the reagent shortage and tedious process of sample preparation. The authors delicately evaluated current approaches with a high-throughput platform in clinical samples. Overall, the data presented in the manuscript are of high quality with several important connections demonstrated, including accuracy and sensitivity. This manuscript should be great interest to a general audience, especially those countries are suffering from the massive COVID-19 test loading. In conclusion, I would suggest the editor directly accept this manuscript without any further revision.

We thank the reviewer for these very positive comments.

- Reviewer #2: The authors demonstrated a way to multiplex and bypass the RNA extraction for SARS-CoV-2 detection. The application of the IFC method is new and serve the purpose. However, the sample number of SARS-CoV-2 patients is low and damper the conclusive findings. I have several comments.

Specific comments

1. Increase the sample size of SARS-CoV-2 patients if possible.

We agree with the reviewer that increasing the size of SARS-CoV-19 patient would improve the manuscript. However, this is not possible as our access to patient samples is limited, specifically for direct detection as we receive samples collected and transported in different VTM. As a result, this paper should, in our view, be regarded as a technical report and not necessarily as a fully validated clinical study.

2. Author should explain why to detect miRNA and the importance.

It is unclear, at this stage, whether miRNA detection will be of any importance. However, we felt that being able to demonstrate that high-throughput miRNA detection is possible enriches the manuscript. We feel that this proof of principle adds value to the report and we have decided to keep this section in the revised manuscript.

3. The label in figure 5 second row is quite confusing. Maybe change to VTM. I don't understand the labeling in the figure 5 right part. DP? N? How you calculate sensitivity, by adding?

We have modified the figure 5 (Figure 6 in the revised version of the manuscript) and we feel that this new version is more reader-friendly. The reviewer is right. We have calculated the sensitivity by adding the positive samples with the different probes. However, we omitted to define the acronym “DP” (which stands for double positive). This omission may have confused the reviewer. We have amended the legend of the figure accordingly.

4. I won't say the 11 to 14 out of 17 is high. It is not acceptable for any applications.

We have removed the adjective “high”. The new sentence reads: “The R2 values dropped in a dramatic way for the two direct detergents based assays, with 11 to 14 positive samples detected out of 17” (lanes 469-470).

5. The specificity was not evaluated in the lysis buffer and direct assay. Please revise.

We performed some assays on one negative sample and no-SARS-CoV2-specific signal was obtained in conditions of the Figure 5 (Figure 6 in the revised version). This is now mentioned in the legend of figure 6 in the revised manuscript.

6. The picture quality of the supplementary figures are very poor.

We have provided a new set of supplementary figures of a higher quality.

7. Please check all the labeling in text and figures and be consistent, such as SARS-CoV-2 and COVID-19.

We have performed a few rounds of editing of the paper and are think that the text and figures are consistent.

8. Is there any conflict of interest?

We do not report any conflict of interest (see point number 3 and 4 in the responses to the Editor’s comments).

________________________________________

Attachment

Submitted filename: Response to Reviewers.pdf

Decision Letter 1

Tzong-Yueh Chen

16 Mar 2021

PONE-D-20-36827R1

Versatile and flexible microfluidic qPCR test for high-throughput SARS-CoV-2 and cellular response detection in nasopharyngeal swab samples

PLOS ONE

Dear Dr. Mari,

Thank you for submitting your manuscript to PLOS ONE. After careful consideration, we feel that it has merit but does not fully meet PLOS ONE’s publication criteria as it currently stands. Therefore, we invite you to submit a revised version of the manuscript that addresses the points raised during the review process.

1. The author should have some sentences or a paragraph to describe the rationales, the findings, and the indications of miRNA. It is not a proper way to explain by merely adding more data.

2. Please check the SARS-CoV-2 and be consistent. The manuscript still see many SARS-CoV2 throughout the manuscript.

Please submit your revised manuscript by Apr 30 2021 11:59PM. If you will need more time than this to complete your revisions, please reply to this message or contact the journal office at plosone@plos.org. When you're ready to submit your revision, log on to https://www.editorialmanager.com/pone/ and select the 'Submissions Needing Revision' folder to locate your manuscript file.

Please include the following items when submitting your revised manuscript:

  • A rebuttal letter that responds to each point raised by the academic editor and reviewer(s). You should upload this letter as a separate file labeled 'Response to Reviewers'.

  • A marked-up copy of your manuscript that highlights changes made to the original version. You should upload this as a separate file labeled 'Revised Manuscript with Track Changes'.

  • An unmarked version of your revised paper without tracked changes. You should upload this as a separate file labeled 'Manuscript'.

If you would like to make changes to your financial disclosure, please include your updated statement in your cover letter. Guidelines for resubmitting your figure files are available below the reviewer comments at the end of this letter.

If applicable, we recommend that you deposit your laboratory protocols in protocols.io to enhance the reproducibility of your results. Protocols.io assigns your protocol its own identifier (DOI) so that it can be cited independently in the future. For instructions see: http://journals.plos.org/plosone/s/submission-guidelines#loc-laboratory-protocols

We look forward to receiving your revised manuscript.

Kind regards,

Tzong-Yueh Chen, Ph.D.

Academic Editor

PLOS ONE

Journal Requirements:

Please review your reference list to ensure that it is complete and correct. If you have cited papers that have been retracted, please include the rationale for doing so in the manuscript text, or remove these references and replace them with relevant current references. Any changes to the reference list should be mentioned in the rebuttal letter that accompanies your revised manuscript. If you need to cite a retracted article, indicate the article’s retracted status in the References list and also include a citation and full reference for the retraction notice.

[Note: HTML markup is below. Please do not edit.]

Reviewers' comments:

Reviewer's Responses to Questions

Comments to the Author

1. If the authors have adequately addressed your comments raised in a previous round of review and you feel that this manuscript is now acceptable for publication, you may indicate that here to bypass the “Comments to the Author” section, enter your conflict of interest statement in the “Confidential to Editor” section, and submit your "Accept" recommendation.

Reviewer #1: All comments have been addressed

Reviewer #2: (No Response)

**********

2. Is the manuscript technically sound, and do the data support the conclusions?

The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented.

Reviewer #1: Yes

Reviewer #2: Yes

**********

3. Has the statistical analysis been performed appropriately and rigorously?

Reviewer #1: Yes

Reviewer #2: Yes

**********

4. Have the authors made all data underlying the findings in their manuscript fully available?

The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified.

Reviewer #1: Yes

Reviewer #2: Yes

**********

5. Is the manuscript presented in an intelligible fashion and written in standard English?

PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here.

Reviewer #1: Yes

Reviewer #2: Yes

**********

6. Review Comments to the Author

Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters)

Reviewer #1: (No Response)

Reviewer #2: The authors have address most of my comments. There is a few comments still need to be addressed.

1. The author should have some sentences or a paragraph to describe the rationales, the findings, and the indications of miRNA. It is not a proper way to explain by merely adding more data.

2. Please check the SARS-CoV-2 and be consistent. I still see many SARS-CoV2 throughout the manuscript.

**********

7. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files.

If you choose “no”, your identity will remain anonymous but your review may still be made public.

Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy.

Reviewer #1: No

Reviewer #2: No

[NOTE: If reviewer comments were submitted as an attachment file, they will be attached to this email and accessible via the submission site. Please log into your account, locate the manuscript record, and check for the action link "View Attachments". If this link does not appear, there are no attachment files.]

While revising your submission, please upload your figure files to the Preflight Analysis and Conversion Engine (PACE) digital diagnostic tool, https://pacev2.apexcovantage.com/. PACE helps ensure that figures meet PLOS requirements. To use PACE, you must first register as a user. Registration is free. Then, login and navigate to the UPLOAD tab, where you will find detailed instructions on how to use the tool. If you encounter any issues or have any questions when using PACE, please email PLOS at figures@plos.org. Please note that Supporting Information files do not need this step.

PLoS One. 2021 Apr 14;16(4):e0243333. doi: 10.1371/journal.pone.0243333.r004

Author response to Decision Letter 1


19 Mar 2021

Responses to the reviewers

Reviewer #2: The authors have address most of my comments. There is a few comments still need to be addressed.

1. The author should have some sentences or a paragraph to describe the rationales, the findings, and the indications of miRNA. It is not a proper way to explain by merely adding more data.

As recommended, we have added a paragraph in the discussion (lanes 523-532) describing the rationale, the findings and the indication of following miRNAs as biomarkers in nasopharyngeal swabs and cited 5 additional references to illustrate this point.

2. Please check the SARS-CoV-2 and be consistent. I still see many SARS-CoV2 throughout the manuscript.

We have corrected these last typos.

Attachment

Submitted filename: Response to Reviewers2.docx

Decision Letter 2

Tzong-Yueh Chen

23 Mar 2021

Versatile and flexible microfluidic qPCR test for high-throughput SARS-CoV-2 and cellular response detection in nasopharyngeal swab samples

PONE-D-20-36827R2

Dear Dr. Mari,

We’re pleased to inform you that your manuscript has been judged scientifically suitable for publication and will be formally accepted for publication once it meets all outstanding technical requirements.

Within one week, you’ll receive an e-mail detailing the required amendments. When these have been addressed, you’ll receive a formal acceptance letter and your manuscript will be scheduled for publication.

An invoice for payment will follow shortly after the formal acceptance. To ensure an efficient process, please log into Editorial Manager at http://www.editorialmanager.com/pone/, click the 'Update My Information' link at the top of the page, and double check that your user information is up-to-date. If you have any billing related questions, please contact our Author Billing department directly at authorbilling@plos.org.

If your institution or institutions have a press office, please notify them about your upcoming paper to help maximize its impact. If they’ll be preparing press materials, please inform our press team as soon as possible -- no later than 48 hours after receiving the formal acceptance. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information, please contact onepress@plos.org.

Kind regards,

Tzong-Yueh Chen, Ph.D.

Academic Editor

PLOS ONE

Additional Editor Comments (optional):

Reviewers' comments:

Reviewer's Responses to Questions

Comments to the Author

1. If the authors have adequately addressed your comments raised in a previous round of review and you feel that this manuscript is now acceptable for publication, you may indicate that here to bypass the “Comments to the Author” section, enter your conflict of interest statement in the “Confidential to Editor” section, and submit your "Accept" recommendation.

Reviewer #2: All comments have been addressed

**********

2. Is the manuscript technically sound, and do the data support the conclusions?

The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented.

Reviewer #2: Yes

**********

3. Has the statistical analysis been performed appropriately and rigorously?

Reviewer #2: Yes

**********

4. Have the authors made all data underlying the findings in their manuscript fully available?

The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified.

Reviewer #2: Yes

**********

5. Is the manuscript presented in an intelligible fashion and written in standard English?

PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here.

Reviewer #2: Yes

**********

6. Review Comments to the Author

Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters)

Reviewer #2: The authors have addressed all the issues.

**********

7. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files.

If you choose “no”, your identity will remain anonymous but your review may still be made public.

Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy.

Reviewer #2: No

Acceptance letter

Tzong-Yueh Chen

8 Apr 2021

PONE-D-20-36827R2

Versatile and flexible microfluidic qPCR test for high-throughput SARS-CoV-2 and cellular response detection in nasopharyngeal swab samples

Dear Dr. Mari:

I'm pleased to inform you that your manuscript has been deemed suitable for publication in PLOS ONE. Congratulations! Your manuscript is now with our production department.

If your institution or institutions have a press office, please let them know about your upcoming paper now to help maximize its impact. If they'll be preparing press materials, please inform our press team within the next 48 hours. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information please contact onepress@plos.org.

If we can help with anything else, please email us at plosone@plos.org.

Thank you for submitting your work to PLOS ONE and supporting open access.

Kind regards,

PLOS ONE Editorial Office Staff

on behalf of

Prof. Tzong-Yueh Chen

Academic Editor

PLOS ONE

Associated Data

    This section collects any data citations, data availability statements, or supplementary materials included in this article.

    Supplementary Materials

    S1 Fig. Titration of the diluted nucleocapsid spike-in transcript.

    A ten-fold serial dilution ranging from 1 to 10–6 was prepared from the stock solution of the in vitro-transcribed N gene and supplemented with 2 ng/μL of total RNA from HEK 293 Cells. Reverse Transcription was performed followed by 15 cycles of pre-Amplification and 30 cycles of qPCR. The RT-qPCR reaction was performed without (A) or with (B) a RNA purification step. Linear regression was performed by logarithmic plots of transcript copy number against Cq value. We observed a good correlation according Cq linear regression curves according to dilution for the three viral CDC primers/probe sets (N1, N2, N3). No Cq value has been detected for the E primers/probe. RNP, used as internal control, shows constant detection of Cq value, suggesting a good performance of the qPCR.

    (TIF)

    S2 Fig. Validation of the BiomarkTM -HD protocol on a cohort of 92 biopsies including 15 positive patients.

    The correlation of the Cq values obtained for the N primers/probe (BiomarkTM -HD) and the GeneFirst COVID-19 detection kit is presented.

    (TIF)

    S3 Fig. Preamplification step optimization.

    The elongation time used in the preamplification reaction was reduced to 1 min at 60°C (A) from 2 min at 60°C (B) using diluted total RNA from a SARS-nCov2 positive patient sample.

    (TIF)

    S4 Fig. Effect of Triton X-100 on extraction-based and direct RT-qPCR protocol performed on a synthetic N transcript or a clinical sample.

    A. In vitro-transcribed viral N gene was added to either a Triton X-100 containing lysis buffer (Tx) or to TE buffer (TE). The samples were heated or not at 65°C for 10 min. RNA extraction was performed or not (direct) and N1 or RNP levels were determined by RT-qPCR using the BiomarkTM -HD system. B. A similar protocol as in A was used but the starting material were two consecutive sample collection from the same patient processed either in a Triton X-100 containing lysis buffer or to TE buffer.

    (TIF)

    S5 Fig. Heatmap of Cq values from direct RT-qPCR experiments using various combinations of detergent on clinical samples.

    Samples from a positive- or negative-COVID-19 patient collected in a commercial VTM was mixed with different combinations of detergents/emulsifiers, in presence or absence of PK (2 mg/mL) and further heat at 95°C for 5 min or not. Cq values obtained in quadruplicate are presented. Tx: Triton X100; PEG: poly ethylene glycol 600; QE: Quick ExtractTM DNA Extraction Solution.

    (TIF)

    S1 Table. List of the micro-RNA tested using the Biomark HD system.

    (DOCX)

    Attachment

    Submitted filename: Response to Reviewers.pdf

    Attachment

    Submitted filename: Response to Reviewers2.docx

    Data Availability Statement

    All data are provided in supplemental information.


    Articles from PLoS ONE are provided here courtesy of PLOS

    RESOURCES