Abstract
Introduction: Since the first wave of COVID-19 in Europe, new diagnostic tools using antigen detection and rapid molecular techniques have been developed. Our objective was to elaborate a diagnostic algorithm combining antigen rapid diagnostic tests, automated antigen dosing and rapid molecular tests and to assess its performance under routine conditions.
Methods: An analytical performance evaluation of four antigen rapid tests, one automated antigen dosing and one molecular point-of-care test was performed on samples sent to our laboratory for a SARS-CoV-2 reverse transcription PCR. We then established a diagnostic algorithm by approaching median viral loads in target populations and evaluated the limit of detection of each test using the PCR cycle threshold values. A field performance evaluation including a clinical validation and a user-friendliness assessment was then conducted on the antigen rapid tests in point-of-care settings (general practitioners and emergency rooms) for outpatients who were symptomatic for <7 days. Automated antigen dosing was trialed for the screening of asymptomatic inpatients.
Results: Our diagnostic algorithm proposed to test recently symptomatic patients using rapid antigen tests, asymptomatic patients using automated tests, and patients requiring immediate admission using molecular point-of-care tests. Accordingly, the conventional reverse transcription PCR was kept as a second line tool. In this setting, antigen rapid tests yielded an overall sensitivity of 83.3% (not significantly different between the four assays) while the use of automated antigen dosing would have spared 93.5% of asymptomatic inpatient screening PCRs.
Conclusion: Using tests not considered the “gold standard” for COVID-19 diagnosis on well-defined target populations allowed for the optimization of their intrinsic performances, widening the scale of our testing arsenal while sparing molecular resources for more seriously ill patients.
Keywords: COVID-19, SARS-CoV-2, immunoassay, diagnostic, antigen, PCR, point-of-care, NAAT
Introduction
At the time of writing (January 7, 2021), Belgium is emerging from a second wave of COVID-19 epidemic. The World Health Organization (WHO) recommended mass use of reverse transcription real-time PCR (RT-PCR) to detect active SARS-CoV-2 infections (1). However, the unprecedented high volume of samples reaching laboratories led to global scarcities of reagents and delays making prolonged containment measures less acceptable by the population (2). Since then, a new set of diagnostic tools have been developed, such as antigen detection immunoassays or molecular point-of-care tests. These tools could allow diversification of testing strategies and decrease shortages and overflows.
Thanks to their high sensitivity, ranging from 73.9 to 89.5% for high viral load samples [105-107 RNA copies/swab (3)], and their overall specificity (4, 5), antigen-detection rapid diagnostic tests have been integrated in several countries' testing strategies (6–10)1,2. Both Centers for Disease Control and Prevention (CDC) (11) WHO (12) and European Center for Disease Control and Prevention (ECDC) (13) have issued guidelines for their use. However, practical considerations are still lacking (including the best target populations). Meanwhile, several manufacturers have developed molecular point-of-care tests, most of which additionally target influenza and/or RSV (14, 15) while others offer wider respiratory syndromic panel (16).
In addition, high throughput antigen-dosing systems based on chemiluminescence enzyme immunoassay represent an interesting alternative (17). This solution, recently deployed in German airports, is a striking example of delocalized laboratory medicine (18).
Following this expansion of available diagnostic tools, a deeper reflection has come to light on the best use of these various testing solutions according to their sensitivity, their turnaround time, the context in which the result will be used (patient vs. population-centered approach), the kinetics of the epidemic and the availability of reagents and consumables (19).
All of the above may partly explain the apparent confusion we are currently witnessing in the deployment of antigen rapid diagnostic tests and/or molecular point-of-care tests in most industrialized countries, either in terms of choosing the most appropriate diagnostic tests or the target population to apply these tests to. We would like to share here the results of evaluations we performed on four antigen rapid diagnostic tests, one automated antigen dosing assay and one molecular point-of-care test for the diagnosis of COVID-19, not only from an analytical “laboratory” point-of-view but also through their field implementation during the second Belgian COVID-19 wave. Using different techniques at different levels in a multi-step, integrated, and adaptive diagnostic algorithm helped us to diversify and increase our overall testing capacity.
Methods
Population
LHUB-ULB (Laboratoire Hospitalier Universitaire de Bruxelles—Universitair Laboratorium Brussel) is a clinical laboratory serving five university hospitals (containing a capacity of around 3,000 beds) as well as a network of general practitioners in Brussels, Belgium. LHUB-ULB's service area covers 700,000 inhabitants (20). From July to September 2020, patients undergoing a SARS-CoV-2 RT-PCR were retrospectively categorized through a structured algorithm into four categories according to the information provided on the orders: symptomatic outpatients, hospital admissions (symptomatic or not), asymptomatic high-risk contacts, or mandatory screenings. The RT-PCR median CT values from these four groups were compared using the Tukey-Kramer method.
Symptomatic Cases Definition
We used the case definition provided by the Belgian national health institute (Sciensano) for COVID-19 (21). The acute apparition of one major symptom, the presence of two minor symptoms, or the aggravation of chronic respiratory symptoms without any other obvious cause was defined as a possible case (Supplementary Table 1). A confirmed case was a person with a SARS-CoV-2 positive sample.
Diagnostic Tests
Antigen Rapid Diagnostic Tests
Four lateral-flow immunoassays were evaluated: Panbio™ COVID-19 Ag Rapid Test Device (Abbott Rapid Diagnostics, Germany), BD Veritor™ SARS-CoV-2 (Becton-Dickinson and Company, USA), COVID-19 Ag Respi-Strip (Coris BioConcept, Belgium) and SARS-CoV-2 Rapid Antigen Test (SD Biosensor, Republic of Korea). Reading was performed by trained operators except for the BD Veritor™ for which an automated reader (BD Veritor™ System) was used.
An analytical performance study was performed using nasopharyngeal swabs. The swabs preserved in universal transport media (UTM) were sent to our laboratory for a SARS-CoV-2 RT-PCR, and then kept refrigerated overnight after the RT-PCR was performed. The four assays were performed at the same time by two trained operators. The amount of UTM engaged was according to the recommendations by each manufacturer for evaluation purposes but not for clinical use.
After the performance study, antigen rapid diagnostic tests were done in point-of-care settings, either a practice within our network of general practitioners, or in the emergency room of the Saint-Pierre university hospital. Each possible COVID-19 outpatient, who was within 7 days of symptoms onset, was offered an antigen rapid diagnostic test and informed that a negative result would require an additional sampling for RT-PCR as recommended at the time (21). Each antigen rapid diagnostic test sampling and test procedure was performed according manufacturer instructions (Supplementary Table 2).
The user-friendliness of each antigen rapid diagnostic test was assessed with a four-part questionnaire adapted from the Scandinavian evaluation of laboratory equipment for point-of-care testing SKUP/2008/114 evaluation (22).
Molecular Point-of-Care Test
To assess the analytical performance of the Cobas® Liat SARS-CoV-2 & Influenza A/B nucleic acid test (Roche Molecular Systems, USA), nasopharyngeal swabs, which were sent to our laboratory for a SARS-CoV-2 RT-PCR and tested positive, were kept refrigerated overnight before testing. In addition, frozen samples from February 2020 which underwent at that time a Cobas® Liat Influenza A/B & RSV RT-PCR assay were also tested.
Automated Antigen Dosing Assay
Antigen dosing was performed using the Lumipulse® G SARS-CoV-2 Ag (Fujirebio, Japan) assay, expressing the dosage in pg/mL. For biosafety consideration, a viral-deactivation step (56°C heating for 30 min) was added to the manufacturer's instructions protocol (23).
The analytical performance study was performed on UTM swabs kept refrigerated overnight after a SARS-CoV-2 RT-PCR. All available positive samples were selected. Negative samples were randomly selected to obtain a positive/negative ratio around 2:1.
In the second part of the evaluation, we evaluated the Lumipulse® performance on UTM samples sent to our laboratory for SARS-CoV-2 RT-PCR for patients who required scheduled hospital admission, COVID-19 contacts, or for healthcare workers.
Gold Standard and Statistical Analysis
Analytical performance study of antigen rapid diagnostic tests, molecular point-of-care test and automated antigen dosing were carried out on three different sets of samples.
SARS-CoV-2 RT-PCR was considered as the gold-standard. Except for some antigen rapid diagnostic tests, for which negative results were controlled by various other RT-PCR protocols, samples underwent the RealTime SARS-CoV-2 assay (Abbott Molecular, USA) on our m2000 platform. As detection of both targeted genes (RdRp and N) is performed using the same fluorophore, the CT values of this assay are observed up to 32 cycles, and not comparable with CT values of other RT-PCR assays. Consequently, only the CT values obtained using the RealTime SARS-CoV-2 assay were considered.
Statistical analyses and receiver operating characteristic (ROC) curves were performed using Analyse-it® for Microsoft Excel v3.80.
Results
Trends of CT Value in the Different Populations
LHUB-ULB performed 31,397 SARS-CoV-2 RT-PCR including 1,708 positive nasopharyngeal samples (5.4%) from 1,568 patients. 1,169 SARS-CoV-2 infected patients were categorized as follows: 580 symptomatic outpatients (49.6%), 318 admissions (27.2%), 178 contacts (15.2%), and 93 screenings (7.9%). The median CT for symptomatic outpatients (13.8/32) was significantly lower than for any other group (Figure 1). The median CT for contacts (17.4/32) was significantly lower than for admissions (20.8/32, p = 0.0044) and for screenings (23.2/32, p = 0.0002). Hence, antigen rapid diagnostic test was considered for symptomatic outpatients, automated antigen dosing for screenings and molecular point-of-care tests for hospital admissions.
Analytical Performance Studies
Antigen Rapid Diagnostic Tests
Ninety-nine UTM samples including 61 positive (CT ranging from 3.86/32 to 30.94/32) were selected. In this frame, the sensitivities of each antigen rapid diagnostic test were ranging from 36.1 to 49.2% (Table 1). The latest CT detected antigen rapid diagnostic tests was 18.06/32. No false positive result was observed.
Table 1.
Buffer dilution factor | Sensitivity (IC95) | Specificity | Last CT detected | |
---|---|---|---|---|
Panbio™ COVID-19 | 1/2 | 45.9% (34.0–58.3%) | 100% | 18.06/32 |
Coris COVID-19 Ag Respi-Strip | 1/2 | 39.3% (28.1–51.9%) | 100% | 13.31/32 |
SD Biosensor™ SARS-CoV-2 | 1/2 | 49.2% (37.1–61.4%) | 100% | 18.06/32 |
BD Veritor™ SARS-CoV-2 | 1/6 | 36.1% (25.2–48.8%) | 100% | 13.9/32 |
Molecular Point-of-Care Test
The agreement of the Cobas® Liat with the m2000 system for SARS-CoV-2 diagnostic was of 90.9% (50/55) for positive samples. CT values correlation between instruments was good (R2 = 0.931). The Cobas® Liat yielded positive results for all positive samples presenting a CT value below 27.29/32 and yielded positive results for samples with CT of up to 29.11/32. Eighteen of the 19 frozen Influenza A positive samples and 5 of the 6 frozen influenza B positive samples yielded coherent positive results. Agreements for negative samples were of 100% for each parameter.
Automated Antigen Dosing Assay
Two hundred fourteen samples were selected including 136 positive samples. ROC curve analysis yielded an area under the curve (AUC) of 0.893±0.021 (Supplementary Figure 1). The highest Youden Index was at a threshold of 13.75 pg/mL (sensitivity 67.7%, specificity 97.1%). At a threshold set at 1.32 pg/mL [similar to a previous study (17) and to the manufacturer proposed cut-off at 1.34 pg/mL (24)], sensitivity was 78.9% and specificity of 73.9%. To exclude any false positive, the threshold had to be set at 20.27 pg/mL (sensitivity 63.9%). Finally, using a CT <20/32 as a judgement criterion, the AUC of the ROC curve was 0.984 ± 0.007 (Supplementary Figure 2) with an optimal Youden index at a threshold of 20.27 pg/mL (sensitivity 87.4%, specificity 98.1%).
Elaboration of the Diagnostic Algorithm
Following these results, we elaborated the algorithm described in Figure 2: whereas the diagnosis of outpatients was mainly based on point-of-care antigen rapid diagnostic tests, the hospital algorithm combined antigen rapid diagnostic tests, molecular point-of-care tests and conventional RT-PCR in an integrative diagnostic strategy. Four clinical situations were further identified: screening of asymptomatic patients, patients requiring immediate admission, symptomatic outpatients with symptoms lasting for less or more than 5 days.
Field Performance Evaluation
Antigen Rapid Diagnostic Tests
Four hundred ninety-four symptomatic outpatients underwent an antigen rapid diagnostic test. Two hundred and nine (42.3%) were positive. Sixteen negative antigen rapid diagnostic tests were excluded due to missing RT-PCR results. Overall sensitivity was 83.3% (95% confidence interval (IC95): 78.2–87.4%—Table 2). Taken individually, each assay's sensitivity was not significantly different from the others, ranging from 78.3 to 87.7%. Only the BD Veritor™ was conducted on a sufficient number of patients to allow a meaningful comparison between the emergency room (sensitivity: 88.2%, IC95: 76.6–94.5%) and the general practitioners (sensitivity of 87.3%, IC95: 76.0–93.7%), yielding no significant difference. Sensitivity according to days since onset of symptoms (DSO), dropped significantly from 86.9% (IC95: 81.6–90.8%) for up to 4 DSO to 63.6% (IC95: 46.6–77.8%) from 5 DSO (t-test, p < 0.001). False negative antigen rapid diagnostic tests had CT ranging from 4.93/32 to 29.02/32.
Table 2.
N | Sensitivity (IC95) | False negative median CT (range) | |
---|---|---|---|
Overall | 478 | 83.3% (78.2–87.4%) | 17.60 (4.93–29.02) |
Manufacturer | |||
BD Veritor™ SARS-CoV-2 | 177 | 87.7% (80.1–92.7%) | 15.46 (4.93–18.54) |
- At the general practitioner consultation | 110 | 87.3% (76.0–93.7%) | |
- At the emergency room | 67 | 88.2% (76.6–94.5%) | |
Panbio™ COVID-19 Ag Rapid Test Device | 101 | 80.8% (68.1–89.2%) | 18.32 (10.29–23.68) |
Coris COVID-19 Ag Respi-Strip | 135 | 80.0% (69.2–87.7%) | 21.56 (15.52–29.02) |
SD Biosensor™ SARS-CoV-2 Rapid Antigen Test | 65 | 78.3% (58.1–90.3%) | 15.53 (14.92–16.15) |
DSO | |||
<5 DSO | 395 | 86.9%* (81.6–90.8%) | 18.38 (10.90–29.02) |
- 0–1 DSO | 97 | 89.1% (78.2–94.9%) | |
- 2 DSO | 118 | 90.3% (80.5–95.5%) | |
- 3 DSO | 118 | 80.3% (68.7–88.4%) | |
- 4 DSO | 62 | 89.3% (72.8–96.3%) | |
≥5 DSO | 53 | 63.6%* (46.6–77.8%) | 15.46 (4.93–27.02) |
DSO, days since symptoms onset; N, number of performed tests; IC95, 95% confidence interval.
p-value < 0.001 (Student's t-test).
The user-friendliness was satisfactory for all four antigen rapid diagnostic tests tested (Table 3). The Coris COVID-19 Ag Respi-strip had a less satisfactory rating. The main practical issue was its readiness: its “strip-in-a-tube” format was considered by operators as non-practical and leading to a potential biosafety hazard when the reading is difficult. Notably, SD Biosensor™ and Coris BioConcept did not provide any internal control in their kit. BD Veritor™ was the only kit offering nasal swabbing and automated reading.
Table 3.
Mean of N = 3 questioned operators | ||||
---|---|---|---|---|
Operation facilities | BD Veritor™ SARS-CoV-2 | Coris COVID-19 Ag Respi-strip | Panbio™ COVID-19 Ag rapid test device | SD Biosensor™ SARS-CoV-2 Rapid Antigen Test |
To prepare the test | Intermediate (1S 2I)a | Intermediateb | Intermediate (1S 2I)b | Satisfactory (2S 1I)a |
To prepare the sample | Satisfactory | Satisfactory | Satisfactory | Satisfactory |
Application of specimen | Satisfactory | Satisfactory | Satisfactory | Satisfactory |
Number of procedure step | Satisfactory | Satisfactory | Satisfactory | Satisfactory |
Test design | Satisfactory | Unsatisfactory (2U, 1I)c | Satisfactory | Satisfactory |
Reading of the result | Satisfactory | Difficultd | Satisfactory | Satisfactory |
Sources of errors | Satisfactory | Intermediated | Satisfactory | Satisfactory |
Hygiene when using the test | Satisfactory | Unsatisfactorye | Satisfactory | Satisfactory |
Size and weight of the package | Satisfactory (2S 1I)f | Satisfactory | Satisfactory | Satisfactory |
Storage conditions for tests, unopened package* | 15–30°C | 15–30°C | 15–30°C | 15–30°C |
Storage conditions for tests, opened package* | 15–30°C | 15–30°C | 15–30°C | 15–30°C |
Environmental aspects: waste handling* | Special precautions | Special precautions | Special precautions | Special precautions |
Intended users* | Health care personnel | Health care personnel | Health care personnel | Health care personnel |
Information in instruction in the insert | ||||
Preparations/Pre-analytic procedure | Satisfactory | Satisfactory | Satisfactory | Satisfactory |
Specimen collection | Satisfactory | Intermediateg | Satisfactory | Satisfactory |
Measurement procedure | Satisfactory | Satisfactory | Satisfactory | Satisfactory |
Reading of result | Satisfactory | Satisfactory | Satisfactory | Satisfactory |
Description of the sources of error | Satisfactory | Satisfactory | Satisfactory | Satisfactory |
Help for troubleshooting | Satisfactory | Satisfactory | Satisfactory | Satisfactory |
Readability/clarity of presentation | Satisfactory | Intermediateh | Satisfactory | Satisfactory (1I 2S)j |
General impression | Satisfactory | Intermediatei | Satisfactory | Satisfactory (1I 2S)j |
Measurement principle* | Satisfactory | Satisfactory | Satisfactory | Satisfactory |
Available insert in ENG + FR + NL* | Partlyk | Partlyk | Partlyl | Partlyk |
Time factors* | ||||
Required training time | <2 h | <2 h | <2 h | <2 h |
Duration of preparations/Pre-analytical time | <6 min | <6 min | <6 min | <6 min |
Duration of analysis | 10–20 min | >20 min | 10–20 min | 10–20 min |
Stability of test, unopened package | >5 months | >5 months | >5 months | >5 months |
Stability of test, opened package | >30 days or disposable | >30 days or disposable | >30 days or disposable | >30 days or disposable |
Stability of quality control material unopened | >5 months | No QC provided | >5 months | No QC provided |
Analytical quality control* | ||||
Reading of the internal quality control | Satisfactory | Unsatisfactorym | Satisfactory | Unsatisfactorym |
Usefulness of the internal quality control | Satisfactory | Unsatisfactorym | Satisfactory | Unsatisfactorym |
Objective informational items were filled by the principal investigator.
Caps of the buffer tubes difficult to manipulate.
Requires a tube rack.
The use of a strip in a closed tube with a very difficult capping was not considered practical for the operators.
Difficult reading through a closed tube although transparent.
Operators were forced to open the tubes to extract the strip in case of a doubt with the reading causing biosafety concern.
Oversized packaging compared to the number of test.
Lack of precise instruction.
Lack of clarity.
A quick reference guide would have been appreciated.
Small typo and dense content.
Only available in English.
Not available in Dutch.
Not provided.
Automated Antigen Dosing Assay
Two hundred seventy-nine patients (including 93 asymptomatic patients screened for a scheduled hospitalization) were tested. Their SARS-CoV-2 carriage status was categorized as “unlikely” if dosing below 1.32 pg/mL (n = 219, 78.5%), “possible” if dosing from 1.32 to 20.27 pg/mL (n = 46, 16.5%) and “certain” if dosing higher than 20.27 pg/mL (n = 14, 5.0%). All patients with “certain” results had a positive RT-PCR. Seven patients out of 46 (15.2%) with a “possible” result and five out of 219 (2.3%) with an “unlikely” result were tested positive according to RT-PCR, respectively (Table 4). Thus, the overall sensitivity for asymptomatic patients was of 86.7% (13/15). Hence, using this assay for the pre-admission screening of these 93 patients would have spared 87 RT-PCR (93.5%) for the cost of one missed low-positive (CT = 26.04/32).
Table 4.
Lumipulse® automated antigen detection | |||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|
Certain (>20.27 pg/mL) | Possible (1.32–20.27pg/mL) | Unlikely (≤1.32pg/mL) | |||||||||
PCR result | N | Positive | Lowest CT | N | Negative | Positive | Lowest CT | N | Negative | Positive | Lowest CT |
Overall | 279 | 14 | 3.89 | 46 | 39 | 7 | 15.7 | 219 | 214 | 5 | 22.61 |
Scheduled hospitalizations | 93 | 1 | 12.73 | 6 | 4 | 2 | 15.7 | 86 | 85 | 1 | 26.04 |
Contacts | 67 | 4 | 6.98 | 13 | 12 | 1 | 31.23 | 50 | 49 | 1 | 22.9 |
Health workers | 119 | 9 | 3.89 | 27 | 23 | 4 | 19.92 | 83 | 80 | 3 | 22.61 |
- With symptoms | 67 | 8 | 3.89 | 13 | 13 | - | - | 46 | 43 | 3 | 22.61 |
Discussion
In most industrialized countries, the large scale use of RT-PCR to detect active SARS-CoV-2 infections has shown limits in its capacity to broadly screen the population while providing timely and therefore meaningful results for optimized prevention and treatment. To fill this gap, SARS-CoV-2 antigen rapid diagnostic tests and molecular point-of-care tests are now considered as an adjunct to the RT-PCRs performed on large automated platforms (25).
Our results provide substantial evidence that no current antigen rapid diagnostic test is sensitive enough to be performed on UTM specimen (i.e., at the laboratory). During the first wave in Europe, we proposed a strategy combining antigen rapid diagnostic tests and RT-PCR, both performed in the laboratory (26). We stopped using antigen rapid diagnostic tests in the laboratory during the declining phase of the epidemic, not because of their low sensitivity [as stated by colleagues (27)], but because the proportion of samples from recently infected patient dropped, impairing these tests' usefulness (28). Regular follow-up of the positivity rate could allow adaptations of antigen rapid diagnostic test strategy as proposed by CDC (11) and ECDC (13). Here, we demonstrate the added-value of antigen rapid diagnostic tests at the point-of-care level for <5-days symptomatic outpatient thanks to their ease-of-use, rapid time-to-result, and low cost.
Our results show slightly lower sensitivity than previously reported (25). Indeed, part of the false negative results observed is likely due to variability in the adherence to protocol regarding sampling, incubation time and DSO. Sensitivity and specificity of such antigen rapid diagnostic tests strongly depend on their good execution and reading which are harder to achieve at the frontline where the expertise of personnel can vary; especially in this time of pandemic when the turn-over is higher than usual. This was confirmed by other recently published studies targeting the same population, with sensitivity ranging from 70.0 to 80.4% (29–31).
The absence of significant difference between antigen rapid diagnostic tests clinical performances highlights the need to assess their user-friendliness as a main criterion of choice. Our analysis underlined the need to consider very practical aspects such as opening caps while wearing gloves, ensuring biosafety outside a laboratory (see Figure 3) and instructions targeting non-laboratory operators, as recently discussed for low-resource settings (32). Besides, an immediate, in-person communication of a positive result likely allowed a stronger message and a better adhesion regarding quarantine, hygiene and contact-tracing than if done through virtual means, days after the consultation.
The Cobas® Liat yielded stunning performances for a 20-min triplex molecular point-of-care test compared to our RT-PCR. However, invalid results were experienced with viscous samples. The addition of a molecular point-of-care test for patients attending the emergency room and needing hospitalization, regardless of the suspicion of COVID-19, allowed a faster management of inpatients avoiding the admission of asymptomatic SARS-CoV-2 carrier in “COVID-free” units, or the admission of SARS-CoV-2-negative patients in COVID-19 units pending their RT-PCR results. Furthermore, influenza and SARS-CoV-2 co-detection allows a better surveillance at a time where the potential co-circulation of the influenza and SARS-CoV-2 is still unknown. The costs of these molecular point-of-care tests stay high and their availability low. Hence, their use should be considered by targeting the best population with regards to the reduction of global costs related to isolation, use of protective equipment and prevention of nosocomial clusters.
In the present study, the Lumipulse® G SARS-CoV-2 Ag showed an overall good analytical performance compared to RT-PCR; and more specifically, to exclude negative and low positive samples using different criteria and cut-off values than the ones proposed by the manufacturer. These cut-offs need to be adapted and chosen regarding the local epidemiology and the objectives of the screening. Our cut-off values diverged from the one proposed in a previous study (17). However, despite the fact we added a viral deactivation by heating, our results yielded a better AUC of the ROC curve. In case of limited access to RT-PCR, such technique can allow testing people who would be otherwise not tested. Its higher throughput and sensitivity than antigen rapid diagnostic tests and its faster time-to-result than RT-PCR make it an interesting intermediary tool. Its low costs and its probable good assessment of infectiousness allow a relevant periodic testing in terms of infection control. Therefore, using antigen dosing could be the best solution to repeatedly test high number of high risk contacts while sparing RT-PCR resources. However, their biosafety must be carefully considered and viral neutralization applied if needed; viscous samples may cause pipetting errors and specific interpretation algorithm should be elaborated.
Our study presents some limitations. We did not consider alternative specimens for SARS-CoV-2 detection such as saliva, the use of serology or broad molecular “syndromic” respiratory panels that could be of use in a larger diagnostic algorithms (33). The emergence of new variants should not impact the value of our algorithm due to the different targets of the assays. However a careful follow-up of their performances over time should be implemented.
Conclusion
In conclusion, our study underlines the importance of shifting our attention from a narrow focus on the sole analytical performances of the diagnostic tools available (especially when these are similar) to an integrated approach taking into account (i) practical consideration such as time-to-result, field ease-of-use, availability of reagents (ii) target populations (iii) intended use of produced results, and (iv) kinetic of the epidemic. Hence, we elaborated here a diagnostic algorithm based on these considerations to optimize the use of the newly extended arsenal of SARS-CoV-2 direct diagnostic tools, from the decentralized setting to the automated lab, to ensure clinical microbiologists enough ammunition for a reliable and meaningful COVID-19 diagnostic.
Data Availability Statement
The raw data supporting the conclusions of this article will be made available by the authors, without undue reservation.
Ethics Statement
Ethical review and approval was not required for the study on human participants in accordance with the local legislation and institutional requirements. Written informed consent from the participants' legal guardian/next of kin was not required to participate in this study in accordance with the national legislation and the institutional requirements. Written informed consent was obtained from the individual(s) for the publication of any potentially identifiable images or data included in this article.
Author Contributions
NY, CDe, MD, F-ZB, FP, MW, and CDu did the investigations. NY and MH contributed to literature review and the writing of the initial draft. NY, CM, FP, HD, ND, M-LD, and SM contributed to manuscript revision, data compilation, and figure presentation. All authors provided critical review and commentary. NY, ND, FC, MB, and MH contributed to study design, manuscript preparation, literature review, revision, and project administration. All authors contributed to the article and approved the submitted version.
Conflict of Interest
Becton Dickinson, Coris BioConcept, and Fujirebio have offered reagents for this study. The authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.
Acknowledgments
We wish to thank the personnel of the LHUB-ULB for its daily technical assistance. The authors would like to thank Olivier Vandenberg for his always appreciable support, guidance, and assistance and Alex van Belkum for his generous and always positive support. This work is dedicated to the healthcare workers, the patients, and families affected by SARS-CoV-2.
Footnotes
1Bundesministerium für Soziales Gesundheit Pflege und Konsumentenschutz. Österreichische Teststrategie SARS-CoV-2.
2Robert Koch Institut. Hinweise zur Testung von Patienten auf Infektion mit dem neuartigen Coronavirus SARS-CoV-2.
Supplementary Material
The Supplementary Material for this article can be found online at: https://www.frontiersin.org/articles/10.3389/fmed.2021.650581/full#supplementary-material
References
- 1.World Health Organization . Laboratory Testing of 2019 Novel Coronavirus (2019-nCoV) in Suspected Human Cases: Interim Guidance (2020). Available online at: https://www.who.int/publications/i/item/laboratory-testing-for-2019-novel-coronavirus-in-suspected-human-cases-20200117 (accessed August 12, 2020).
- 2.Vandenberg O, Martiny D, Rochas O, van Belkum A, Kozlakidis Z. Considerations for diagnostic COVID-19 tests. Nat. Rev. Microbiol. (2020) 19:171–83. 10.1038/s41579-020-00461-z [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3.Corman VM, Haage VC, Bleicker T, Schmidt ML, Mühlemann B, Zuchowski M, et al. Comparison of seven commercial SARS-CoV-2 rapid point-of-care antigen tests. medRxiv. (2020). 10.1101/2020.11.12.20230292 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4.World Health Organization . Antigen-Detection in the Diagnosis of SARS-CoV-2 Infection Using Rapid Immunoassays (2020). Available online at: https://www.who.int/publications/i/item/antigen-detection-in-the-diagnosis-of-sars-cov-2infection-using-rapid-immunoassays (accessed October 1, 2020).
- 5.Dinnes J, Deeks JJ, Adriano A, Berhane S, Davenport C, Dittrich S, et al. Rapid, point-of-care antigen and molecular-based tests for diagnosis of SARS-CoV-2 infection. Cochrane Database Syst. Rev. (2020) CD013705. 10.1002/14651858.CD013705 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 6.Rijskinstituut voor Volksgezondheid en Milieu . Advies antigeen(snel)testen (2020). Available online at: https://www.rijksoverheid.nl/onderwerpen/coronavirus-covid-19/documenten/rapporten/2020/10/14/advies-antigeensneltesten (accessed November 8, 2020).
- 7.Ministry of Health Labour and Welfare . Clinical Management of Patients with COVID-19 (2020). Available online at: https://www.mhlw.go.jp/content/000646531.pdf (accessed November 8, 2020).
- 8.Haute Autorité de Santé. Revue rapide sur les tests de détection antigénique du virus SARS-CoV-2 (2020). Available online at: https://www.has-sante.fr/jcms/p_3213483/fr/revue-rapide-sur-les-tests-de-detection-antigenique-du-virus-sars-cov-2 (accessed November 8, 2020).
- 9.Robert Koch Institut . Nationale Teststrategie SARS-CoV-2 (2019). Available online at: https://www.sozialministerium.at/Informationen-zum-Coronavirus/Neuartiges-Coronavirus-(2019-nCov).html (accessed November 8, 2020).
- 10.Sistema de Saúde . Estratégia Nacional de Testes para SARS-CoV-2 (2020). [Google Scholar]
- 11.US Centers for Disease Control and Prevention . Interim Guidance for Antigen Testing for SARS-CoV-2 (2020). [Google Scholar]
- 12.World Health Organization . Coronavirus Disease (COVID-19) Pandemic – Emergency Use Listing Procedure (EUL) open for in vitro diagnostics (2020). Available online at: https://www.who.int/news-room/articles-detail/coronavirus-disease-(covid-19)-pandemic-emergency-use-listing-procedure-(eul)-open-for-in-vitro-diagnostics (accessed November 9, 2020).
- 13.European Centre for Disease Prevention and Control . Options for the Use of Rapid Antigen Tests for COVID-19 in the EU/EEA and the UK (2020). Available online at: https://www.ecdc.europa.eu/en/publications-data/options-use-rapid-antigen-tests-covid-19-eueea-and-uk (accessed December 20, 2020).
- 14.Hansen G, Marino J, Wang Z-X, Beavis KG, Rodrigo J, Labog K, et al. Clinical performance of the point-of-care cobas Liat for detection of SARS-CoV-2 in 20 minutes: a multicenter study. J. Clin. Microbiol. (2020). 59:e02811–20. 10.1128/JCM.02811-20 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 15.Mostafa HH, Carroll KC, Hicken R, Berry GJ, Manji R, Smith E, et al. Multicenter evaluation of the cepheid xpert xpress SARS-CoV-2/Flu/RSV test. J Clin Microbiol. (2021) 59:e02955-20. 10.1128/JCM.02955-20 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16.Creager HM, Cabrera B, Schnaubelt A, Cox JL, Cushman-Vokoun AM, Shakir SM, et al. Clinical evaluation of the BioFire® Respiratory Panel 2.1 and detection of SARS-CoV-2. J. Clin. Virol. (2020) 129:104538. 10.1016/j.jcv.2020.104538 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 17.Hirotsu Y, Maejima M, Shibusawa M, Nagakubo Y, Hosaka K, Amemiya K, et al. Comparison of automated SARS-CoV-2 antigen test for COVID-19 infection with quantitative RT-PCR using 313 nasopharyngeal swabs, including from seven serially followed patients. Int. J. Infect. Dis. (2020) 99:397–402. 10.1016/j.ijid.2020.08.029 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 18.Dx . Centogene to Offer Fujirebio SARS-CoV-2 Antigen Test at German Airports (2020). Available online at: https://www.360dx.com/infectious-disease/centogene-offer-fujirebio-sars-cov-2-antigen-test-german-airports#.X99e_NJKhhE (accessed December 20, 2020).
- 19.Mina MJ, Parker R, Larremore DB. Rethinking Covid-19 test sensitivity — a strategy for containment. N. Engl. J. Med. (2020) 383:e120. 10.1056/NEJMp2025631 [DOI] [PubMed] [Google Scholar]
- 20.Van den Wijngaert S, Bossuyt N, Ferns B, Busson L, Serrano G, Wautier M, et al. Bigger and better? Representativeness of the Influenza A surveillance using one consolidated clinical microbiology laboratory data set as compared to the belgian sentinel network of laboratories. Front. Public. Health. (2019) 7:150. 10.3389/fpubh.2019.00150 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 21.Sciensano . Définition de cas et indications de demande d'un test et déclaration obligatoire de cas COVID-19 (2020). Available online at: https://covid-19.sciensano.be/sites/default/files/Covid19/COVID-19_Case%20definition_Testing_FR.pdf (accessed December 20, 2020).
- 22.SKUP . Report From the Evaluation SKUP/2018/114. DIAQUICK Strep A Blue Dipstick (DIALAB GmbH), A System for Measurement of Streptococcus Pyogenes Group A. Available online at: https://skup.org/GetFile.ashx?fileid=856 (accessed December 20, 2020).
- 23.Chin AWH, Chu JTS, Perera MRA, Hui KPY, Yen H-L, Chan MCW, et al. Stability of SARS-CoV-2 in different environmental conditions. Lancet Microbe. (2020) 1:e10. 10.1016/S2666-5247(20)30003-3 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 24.Fujirebio Inc. Lumipulse® G SARS-CoV-2 Ag Immunoreaction Cartridges (Package Insert, ver.1) (2020). [Google Scholar]
- 25.Mak GC, Cheng PK, Lau SS, Wong KK, Lau CS, Lam ET, et al. Evaluation of rapid antigen test for detection of SARS-CoV-2 virus. J. Clin. Virol. (2020) 129:104500. 10.1016/j.jcv.2020.104500 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 26.Mertens P, De Vos N, Martiny D, Jassoy C, Mirazimi A, Cuypers L, et al. Development and potential usefulness of the COVID-19 Ag respi-strip diagnostic assay in a pandemic context. Front. Med. (2020) 7:225. 10.3389/fmed.2020.00225 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 27.Scohy A, Anantharajah A, Bodéus M, Kabamba-Mukadi B, Verroken A, Rodriguez-Villalobos H. Low performance of rapid antigen detection test as frontline testing for COVID-19 diagnosis. J. Clin. Virol. (2020) 129:104455. 10.1016/j.jcv.2020.104455 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 28.Tom MR, Mina MJ. To interpret the SARS-CoV-2 test, consider the cycle threshold value. Clin. Infect Dis. (2020) 71:2252–4. 10.1093/cid/ciaa619 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 29.Albert E, Torres I, Bueno F, Huntley D, Molla E, Fernández-Fuentes MÁ, et al. Field evaluation of a rapid antigen test (Panbio™ COVID-19 Ag Rapid Test Device) for COVID-19 diagnosis in primary healthcare centres. Clin. Microbiol. Infect. (2020) 27:472.e7–472.e10. 10.1016/j.cmi.2020.11.004 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 30.Nalumansi A, Lutalo T, Kayiwa J, Watera C, Balinandi S, Kiconco J, et al. Field evaluation of the performance of a SARS-CoV-2 antigen rapid diagnostic test in Uganda using nasopharyngeal samples. Int. J. Infect. Dis. (2020) 104:282–6. 10.1016/j.ijid.2020.10.073 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 31.Bulilete O, Lorente P, Leiva A, Carandell E, Oliver A, Rojo E, et al. Panbio™ rapid antigen test for SARS-CoV-2 has acceptable accuracy in symptomatic patients in primary health care. J Infect. (2021) 82:391–8. 10.1016/j.jinf.2021.02.014 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 32.Jacobs J, Kühne V, Lunguya O, Affolabi D, Hardy L, Vandenberg O. Implementing COVID-19 (SARS-CoV-2) rapid diagnostic tests in Sub-Saharan Africa: a review. Front. Med. (2020) 7:557797. 10.3389/fmed.2020.557797 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 33.Sheridan C. Coronavirus testing finally gathers speed. Nat Biotechnol. (2020). [Epub ahead of print]. 10.1038/d41587-020-00021-z [DOI] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Supplementary Materials
Data Availability Statement
The raw data supporting the conclusions of this article will be made available by the authors, without undue reservation.