Skip to main content
PLOS One logoLink to PLOS One
. 2021 Apr 22;16(4):e0250455. doi: 10.1371/journal.pone.0250455

A quasi-experimental study on stethoscopes contamination with multidrug-resistant bacteria: Its role as a vehicle of transmission

Raeseok Lee 1, Su-Mi Choi 1,2,*, Sung Jin Jo 3, Songyi Han 2, Yun Jeong Park 2, Min A Choi 2, Bo Kyung Kong 4
Editor: Jose Alvarez5
PMCID: PMC8062016  PMID: 33886664

Abstract

Stethoscopes have been suggested to be a possible vector of contact transmission. However, only a few studies have focused on the prevalence of contamination by multidrug-resistant (MDR) bacteria and effectiveness of disinfection training to reduce. This study is to investigate the burden of stethoscope contamination with nosocomial pathogens and multidrug-resistant (MDR) bacteria and to analyze habit changes in disinfection of stethoscopes among healthcare workers (HCWs) before and after education and training. We performed a prospective pre and post quasi-experimental study. A total of 100 HCWs (55 doctors and 45 nurses) were recruited. HCWs were surveyed on their disinfection behavior and stethoscopes were cultured by pressing the diaphragm directly onto a blood agar plate before and after education on disinfection. Pulsed-field gel electrophoresis was performed to determine the relatedness of carbapenem-resistant Enterobacteriaceae. Most of the stethoscopes were contaminated with microorganisms before and after the intervention (97.9% and 91.5%, respectively). The contamination rate of stethoscopes with nosocomial pathogens before and after education was 20.8% and 19.2%, respectively. Stethoscope disinfection habits improved (55.1% vs 31.0%; p<0.001), and the overall bacterial loads of contamination were reduced (median colony-forming units, 15 vs 10; p = 0.019) after the intervention. However, the contamination rate by nosocomial pathogens and MDR bacteria did not decrease significantly. A carbapenemase-producing Klebsiella pneumoniae isolates from a stethoscope was closely related to isolates from the patients admitted at the same ward where the stethoscope was used. Stethoscopes were contaminated with various nosocomial pathogens including MDR bacteria and might act as a vehicle of MDR bacteria. Continuous, consistent education and training should be provided to HCWs using multifaceted approach to reduce the nosocomial transmission via stethoscopes.

Introduction

The burden of healthcare-associated infections has been steadily increasing [1]. An increasing incidence of multidrug-resistant (MDR) bacterial infection and cross-contamination with MDR bacteria is the one of the key reasons why the burden of healthcare-associated infections has increased [2]. Cross-contamination of bacteria occurs in various ways. Direct or indirect contact transmission are the most important routes of disease transmission in the hospital setting [3]. Although, hands are known to be a vector of direct contact transmission during patient care, recent studies have shown that hand hygiene alone is not enough to prevent nosocomial transmission [4]. In addition to hands, various medical devices, including blood pressure cuffs, doppler probes and even marker pens have been identified as potential vehicles of contact transmission [57].

The stethoscope is the most commonly used medical device that has a surface that is in direct contact with patients. Diaphragms of stethoscopes are known to be the second most contaminated area after the fingertips, even after single physical examination [8]. However, unlike hand hygiene, the role of the stethoscope as a vehicle of transmission has not yet been fully determined. Previous studies have shown that stethoscopes are contaminated mostly by gram-positive organisms such as coagulase-negative Staphylococcus and Staphylococcus aureus [912]. However, several studies have shown that stethoscopes may also be contaminated by gram-negative bacteria (GNB) [9, 12, 13]. A few studies have reported the proportion of resistant bacteria that were identified, and the results were limited to S. aureus [911].

The rate of stethoscope disinfection has been reported to be lower than the rate of hand hygiene [14, 15]. Only 8–35% of healthcare workers (HCWs) were disinfected their stethoscope every time, and doctors had a lower disinfection rate than nurses [11, 16]. Moreover, only 4% of HCWs performed disinfection correctly according to the standard by The US Centers for Disease Control and Prevention [14].This is partly because stethoscope disinfection is not taught or promoted to the same extent as hand hygiene. The lack of education and training is attributable to the lack of specific data on contamination by various microorganisms, and a lack of clear evidence of their role as a vehicle of transmission. The aim of this study was to investigate the burden of stethoscope contamination and the proportion of nosocomial pathogens and MDR bacteria, and to analyze habit changes in disinfection of stethoscopes, before and after education and training.

Materials and methods

Study design and questionnaire

We conducted a pre and post quasi-experimental study from November 6, 2018 to March 31, 2019 at a 450-bed university-affiliated teaching hospital in Seoul, Korea (Fig 1). Among doctors and nurses working in the general wards, intensive care units and emergency room, only those who voluntarily agreed to participate in this study were enrolled. The medical department and ward included the departments of internal medicine and pediatrics. The surgical department and ward included the departments of general surgery, neurosurgery and obstetrics and gynecology. After obtaining informed consent, a pre-intervention questionnaire was administered, and specimens were collected from stethoscopes. The staff of the department of hospital infection control conducted a training and education on stethoscope disinfection methods and practices, separately by divisions of participants, using flyers and PowerPoint slides (Fig 1). The training and education were conducted twice for each participant, one hour at a time. The post-intervention questionnaire administration and collection of specimens from stethoscopes was carried out by the same investigator, one month after the last education session. All samples for culture were collected without prior notice. Various parameters potentially associated with stethoscope contamination were included in the survey. The questionnaire included (1) demographic data such as sex, age group, specialty, work department, and length of career and; (2) stethoscope disinfection habits such as frequency, preferred methods, preferred places, and frequency of hand washing. The questionnaire was mainly created based on previous studies, but several items were added and modified after study meeting [11, 17, 18]. Furthermore, the questionnaire was pre-tested by three doctors and two nurses who did not participate in the study and finally revised before the study (S1 and S3 Files). This study was approved by the institutional review board of the Yeouido St. Mary’s Hospital (No. SC18OESI0120 and No. SC19RESI0081). All participants provided written informed consent.

Fig 1. Study flow for pre and post quasi-experimental study.

Fig 1

PFGE, Pulsed-field gel electrophoresis.

Specimen collection and culture methods

Personal stethoscopes of doctors and shared stethoscopes used by nurses in the wards were included. Culture specimens from stethoscopes were obtained by pressing the diaphragm directly onto a blood agar plate for 6 to 10 seconds by a trained investigator. The inoculated plates were incubated aerobically at 37°C for 24 hours. Colony-forming units (CFUs) were identified and counted by a single microbiologist. Bacterial identification and antimicrobial susceptibility tests were performed using the Vitek 2 automated system (bioMérieux, Marcy L’Étoile, France) following the manufacturer’s recommendations. The results of the susceptibility test were interpreted based on the guideline document M100-S28 by Clinical and Laboratory Standards Institute [19]. The modified carbapenem inactivation method was used as a phenotypic test for carbapenem-resistant Enterobacteriaceae. The BD MAX system (BD Molecular Diagnostics, Franklin Lakes, NJ, USA) with Check-Direct carbapenemase-producing Enterobacteriaceae (CPE) real-time PCR was used for the detection of carbapenemase genes according to the manufacturer’s instructions.

We defined potential nosocomial pathogens as follows: S. aureus, Enterococcus species, Pseudomonas aeruginosa, Acinetobacter species, and Enterobacteriaceae. MDR was defined as non-susceptibility to one or more agents in at least three different antibiotic classes among gram-negative organisms and methicillin-resistant staphylococci and vancomycin-resistant enterococci [20].

Pulsed-field gel electrophoresis

During the study period, we were collecting CPE clinical isolates from patients for other purposes. We performed pulsed-field gel electrophoresis (PFGE) to determine the genetic relatedness of CPE isolated from the stethoscope and patients admitted to the ward where the stethoscope was used during study period. Genomic DNA extracted from the isolates were digested by XbaI restriction endonuclease (Roche, Mannheim, Germany) and DNA fragments were separated with a running time of 20 hr at 6 V/cm at 11°C on Genemapper X (Bio-Rad, Hercules, CA) with initial and final pulse times of 0.5 sec and 30 sec, respectively. A lambda ladder (Bio-Rad) was used as a DNA size marker. Similarity coefficients were calculated from Dice coefficients. Banding patterns were analyzed with BioNumerics software version 6.0 (Applied Maths, Sint-Martens-Latem, Belgium) to make a dendrogram. Organisms with PFGE profiles with greater than 90% similarity were considered genetically related strains.

Statistical analysis

Categorical variables were described as frequencies and proportions, and continuous variables were reported using the mean and standard deviation, or the median and interquartile range. Categorical variables were compared by using the χ2-test or Fisher’s exact test. The non-parametric McNemar’s test for categorical variables and Wilcoxon signed-rank test for continuous variables were used for pre- and post-intervention comparisons. Multivariate logistic regression was performed to identify risk factors for contamination by nosocomial pathogens using all variables with a P-value of less than 0.1 in the univariate analysis and clinically important variables. Before the multivariate regression, multicollinearity was assessed using variance inflation factors. All statistical analyses were performed using SAS version 9.4 (SAS Institute, Cary, NC) with a two-sided P-value of less than 0.05 considered significant.

Results

Demographic characteristics and pre-intervention status of the participants

One hundred participants, comprising 55 doctors and 45 nurses were enrolled, of whom 59 were women. However, the proportion of women was lower among doctors (29.1%, 16/55) and higher among nurses (95.6%, 43/45). Most of the participants (88.0%) were aged less than 40 years and 84.0% had used stethoscopes for more than one year. In the pre-intervention assessment, 85.0% reported that they had been educated and had an accurate knowledge of hand hygiene, but only 22.0% reported that they had been educated about the need for stethoscope disinfection. The demographic characteristics and pre-intervention education status of the participants are shown in Table 1.

Table 1. Demographic characteristics and pre-intervention status of the participants.

Characteristics Total N = 100 (%)
Sex, woman 59 (59.0)
Age, years
    20–29 31 (31.0)
    30–39 57 (57.0)
    40–49 9 (9.0)
    ≥50 3 (3.0)
Subject
    Doctor 55 (55.0)
        Medical department 38 (69.1)
        Surgical department 7 (12.7)
        Emergency medicine 4 (7.3)
        Intern 6 (10.9)
    Nurse 45 (45.0)
        Medical ward 20 (44.5)
        Surgical ward 6 (13.3)
        Emergency room 4 (8.9)
        Intensive care unit 15 (33.3)
Career period, years
    <2 29 (29.0)
    2–5 32 (32.0)
    5–10 18 (18.0)
    ≥10 21 (21.0)
Period of stethoscope use
    <6 months 7 (7.0)
    6 months to 1 year 9 (9.0)
    ≥1 year 84 (84.0)
Educational experiences in stethoscope disinfection
    Need for disinfection 22 (22.0)
        during college education 2 (2.0)
        by hospital infection control unit 15 (15.0)
        by senior or colleague 5 (5.0)
    Methods of disinfection 16 (16.0)
        during college education 1 (1.0)
        by hospital infection control unit 12 (12.0)
        by senior or colleague 3 (3.0)
Use a personal stethoscope to auscultate patients with MDR pathogens 36 (36.0)
    Reasons of not using designated stethoscope
        Ignorance of designated stethoscope 3 (3.0)
        Hard to access 5 (5.0)
        Function is not good 19 (19.0)
Exact knowledge of hand washing 85 (85.0)
Time of hand washing
    <20 sec 61 (61.0)
    20–40 sec 34 (34.0)
    ≥40 sec 5 (5.0)

MDR, Multidrug-resistant

Stethoscope disinfection habits before and after intervention

Before the intervention, 88.0% of the participants responded that they believed that disinfection of stethoscopes would reduce contact transmission. However, only 31.0% reported that they regularly–defined as for every patient to at least once a week–disinfected their stethoscope. This rate was higher in nurses (44.4%, 20/45) than doctors (20.0%, 11/55) (odds ratio [OR], 0.313; 95% confidence interval [CI], 0.129–0.757; p = 0.008), Also, the rate of disinfection of stethoscopes at least once a week was higher in women (39.0%, 23/59) than men (19.5%, 8/41) (OR, 0.379; 95% CI, 0.149–0.965; p = 0.038). There were no significant differences according to age or length of career. Overall, the use of alcohol swabs was the preferred methods of disinfection. Twelve of the doctors, but none of the nurses, used the alcohol-based hand gel provided next to the patients’ beds (Table 2).

Table 2. Disinfection habit changes of the participants after education.

Question Habit and practices Pre-intervention Post-intervention
N = 100 (%) N = 98 (%)1
Helpful to change disinfection habit Yes 64 (65.3)
Frequency of disinfection
Every patient 10 (10.0) 20 (20.4)
At least once a week 21 (21.0) 34 (34.7)
On occasion 51 (51.0) 40 (40.8)
Not at all 18 (18.0) 4 (4.1)
Method of disinfection
Alcohol swab 75 (75.0) 78 (79.6)
Alcohol-based hand gel 12 (12.0) 19 (19.4)
Soap and water 0 (0.0) 0 (0.0)
Place of disinfection
Patient’s bed side 36 (36.0) 45 (45.9)
Nurse office 42 (42.0) 40 (40.8)
Outpatient room 9 (9.0) 11 (11.2)

1Two participants moved to another hospital after the pre-intervention samples.

After the education on disinfection, the proportion of participants who reported that they disinfected their stethoscope at least once a week significantly increased (31.0% [31/100] vs 55.1% [54/98]; p<0.001). This was consistent in men (19.5% [8/41] vs 37.5% [15/40]; p = 0.039) and women participants (39.0% [23/59] vs 67.2% [39/58]; p<0.001), and in doctors (20.0% [11/55] vs 41.5% [22/53]; p = 0.001) and nurses (44.4% [20/45] vs 71.1% [32/45]; p = 0.004). The participants’ preferred methods of disinfection did not change after the education and training, and alcohol swabs were the most frequently used method. Most of the participants reported that the education and training had been helpful and effective in changing their disinfection habits. Their self-reported changes in disinfection practices are summarized in Table 2.

Burden of contamination of stethoscopes and nosocomial pathogens identified

Four participants were excluded from the contamination analysis because they changed their stethoscope during the study period and an additional two participants were excluded from the post-intervention analysis because they moved to other hospital during the study period.

Before the intervention, 94 of 96 stethoscopes (97.9%) were contaminated with at least one microorganism, and there was a median of 15 CFUs (interquartile range: 5–36) per stethoscope. The bacterial load was significantly lower in the stethoscopes of participants who disinfected their stethoscope at least once a week than that of participants who disinfected their stethoscope less frequently or did not disinfect them at all (median CFUs, 9 vs 19; p = 0.015), nurses than doctors (median CFUs, 10 vs 20; p = 0.018) and women than men (median CFUs, 13 vs 29; p = 0.017). However, the bacterial load did not differ significantly according to the participant’s career duration, method of stethoscope disinfection, or duration of stethoscope use. Twenty-two potential nosocomial pathogens were isolated from 20 of 96 stethoscopes (20.8%) (Table 3). Three (13.6%) were MDR organisms consisting of two methicillin-resistant S. aureus and one extended-spectrum beta-lactamase producing Klebsiella pneumoniae isolate.

Table 3. Isolated nosocomial pathogens and proportion of MDR organisms.

Pre-intervention N = 96 (%) Post-intervention N = 94 (%)
Pathogens MDR Pathogens MDR
No. of stethoscope 20 (20.8) 3 (3.1) 18 (19.2) 6 (6.4)
Staphylococcus aureus 13 (13.5) 2 (2.1)1 15 (15.7) 4 (4.3)
Enterococcus species 6 (6.3) 0 (0.0) 4 (4.3) 0 (0.0)
Acinetobacter baumannii 0 (0.0) 0 (0.0) 1 (1.1) 1 (1.1)
Pseudomonas aeruginosa 0 (0.0) 0 (0.0) 0 (0.0) 0 (0.0)
Enterobacteriaceae 3 (3.1) 1 (1.2) 2 (2.1) 2 (2.1)
    Klebsiella pneumoniae 1 (1.2) 1 (1.2)2 1 (1.1) 1 (1.1)3
    Escherichia coli 0 (0.0) 0 (0.0) 1 (1.1) 1 (1.1)2
    Enterobacter cloacae 2 (2.3) 0 (0.0) 0 (0.0) 0 (0.0)

MDR, Multidrug-resistant

1Methicillin-resistant Staphylococcus aureus

2Extended-spectrum beta-lactamase producer

3Carbapenemase-producing Enterobacteriaceae

After the intervention, 86 of 94 (91.5%) stethoscopes were contaminated, and the load of contaminated bacteria was a median of 10 CFUs (interquartile range: 3–35). The overall bacterial load was significantly lower in the post-intervention period than in the pre-intervention period (median CFUs 15 vs 10; p = 0.019). However, the reduction was only observed in the stethoscopes of women and nurses, and not in the stethoscopes of men and doctors (Fig 2). The overall contamination rate was reduced but it was not statistically significant (97.9% vs 91.5%; p = 0.073) and the contamination rate by potential nosocomial pathogens and MDR bacteria did not decrease. Similar to before the intervention, 22 potential nosocomial pathogens were isolated from 18 (19.2%) of 94 stethoscopes of which 7 (31.8%) were MDR organisms (Table 3). One isolate of K. pneumoniae was found to be carbapenem-resistant. Other than bacteria, Aspergillus fumigatus, A. niger and a mold were cultured in the pre-intervention cultures and three molds were cultured in the post-intervention cultures. These fungi were not included in the statistical analysis and reporting of the colony counts.

Fig 2. Changes in colony forming units of bacteria isolated from stethoscopes during pre- and post-intervention period.

Fig 2

CFUs, colony forming units; ns, non-specific.

The risk of contamination with nosocomial pathogens was independently associated with a high bacterial load as the result of the multivariate analysis with sex, department of participants, frequency and methods of disinfection, and period of stethoscope use during both the pre- and post-intervention period (adjusted odds ratio, 1.016; 95% CI, 1.000–1.031; p = 0.049 and adjusted odds ratio, 1.033; 95% CI, 1.006–1.060; p = 0.015, respectively).

PFGE for CPE

During post-intervention period, a carbapenemase-producing K. pneumoniae was isolated from a stethoscope shared in the ward. During the period preceding the isolation of the carbapenemase-producing K. pneumoniae from the stethoscope, three isolates of carbapenemase-producing K. pneumoniae from three patients admitted to the ward in which the stethoscope was used, were available for PFGE. Two isolates were cultured from rectal swabs 6 weeks prior to the stethoscope culture and another was cultured in a blood sample 5 weeks prior to the stethoscope culture. All isolates showed the same antibiotic susceptibility and had KPC-2 in real-time PCR. K. pneumoniae isolates from the stethoscope and the patients belonged to one cluster pattern with a similarity of over 95% on PFGE (Fig 3 and S1 Fig).

Fig 3. Result of pulsed field gel electrophoresis and dendrogram of carbapenemase-producing K. pneumoniae.

Fig 3

Percentage similarities are shown above the dendrogram. ST_7W, K. pneumoniae from the stethoscope; SM 01 to 03, K. pneumoniae isolates form the patients. PFGE, Pulsed-field gel electrophoresis.

Discussion

In this study, we found most of the stethoscopes were contaminated with microorganisms. Twenty percent of the stethoscopes were contaminated with potential nosocomial pathogens, and 3–6% with MDR bacteria. After the intervention, the participants reported changes in their stethoscope disinfection habits, and overall bacterial loads of contamination were lowered. However, the rate of contamination with nosocomial pathogens and MDR bacteria did not decrease. A CPE from stethoscope showed a genetic relatedness with the isolates from patients admitted at the ward in which the stethoscope was used.

Noncritical items in the medical environment such as bed rails, bed side tables, blood pressure cuffs, monitors can be the source of healthcare-associated infections [21]. There is evidence that thorough medical environmental disinfection reduces healthcare-associated infections [22]. However, there is disagreement regarding the role of stethoscopes as a vehicle of transmission of nosocomial pathogens and MDR bacteria [9, 13, 23, 24]. According to previous studies, gram-positive bacteria were the most prevalent contaminants of stethoscopes and other medical equipment [9, 10, 25]. GNB were known to rarely contaminate stethoscopes [9, 12, 18]. Although, we did not identify all isolated microorganisms as species, GNB were not uncommon. Six of all 44 nosocomial pathogens cultured from stethoscopes were GNB, which was similar to those published by Knecht et al [13] using RNA sequencing. Methicillin-resistant S. aureus was the only microorganism reported to be resistant bacteria that contaminate stethoscopes and methicillin-resistant rate varied greatly from 0 to 42% [9, 11]. To our knowledge, the rate of MDR GNB contamination has not been reported previously. In this study, four of the six (66.7%) potential nosocomial GNB were MDR bacteria. Notably, the MDR rate was even higher in GNB than in gram-positive bacteria in both the pre- and post-intervention period. However, the absolute total number of GNB is so small, further study in the larger groups should be needed.

The US Centers for Disease Control and Prevention recommends disinfecting noncritical patient-care devices such as stethoscopes and blood pressure cuffs when they are visibly soiled and on a regular basis such as after use on each patient, once daily, or once weekly [26]. However, in actual clinical practice, less than a quarter of the participants reported that they had been educated to disinfect stethoscopes, and less than a third reported that they had disinfected their stethoscope at least once a week. Similar results have been found in previous studies using self-reported questionnaires [16, 18] and these rates might be overreported because the frequency has been only 15 to 18% in studies that used direct observation [14, 15]. To increase the frequency of stethoscope disinfection, disinfection education was done using flyers and PowerPoint presentations for doctors and nurses for 3 weeks. Post-intervention stethoscope culture was performed a month after the final education session to assess whether the participants maintained the disinfection practices that had been taught. There are opposing views on the effectiveness of education or training [23, 24]. In our study, the self-reported frequency of disinfection increased and the total CFUs of the contaminants were significantly reduced. Although not statistically significant, the proportion of stethoscopes contaminated with microorganisms decreased.

There might be several reasons for this result. First, we spent equal time and effort on education and training all participants, rather than focusing on a particular group known to have low disinfection rates. Men and doctors have previously been shown to have relatively low disinfection rates and the same results were observed in our study [10, 16]. Moreover, they showed a relatively limited improvement, even after education. Second, education session and a short training period were insufficient to change the behavior of HCWs. Third, instead of a team-based approach, only hospital infection control staff participated in education and training as educators. The increase in the self-reported disinfection rates and reduction of total bacterial loads was insufficient to reduce the opportunity of stethoscope as a transmission vehicle of pathogens. Continuous efforts are needed in the multifaceted aspect to improve HCWs’ behavior along with repeated education and training [27, 28]. Lastly, the limitation of alcohol-based disinfection methods used by most HCWs. Alcohol-based disinfection methods are widely recommended as its effectiveness and convenience [10]. However, it evaporates quickly, and residual effect can not be expected. Effect of disinfection would be reduced if disinfection is not performed after each contact with the patient.

Although there was a gap of 5 to 6 weeks, one isolate of carbapenemase-producing K. pneumoniae from a stethoscope and three clinical isolates from patients showed one cluster pattern. This suggested that stethoscope might be contaminated during patient care and demonstrated that nosocomial pathogens can survive for more than a few weeks on the inanimate surface, as is well known [29]. Because stethoscope might be a vehicle of nosocomial pathogens even MDR bacteria, we emphasize that disinfecting stethoscope is as important as hand hygiene.

This is the first study to evaluate the contamination rate of stethoscopes by GNB-MDR bacteria and to demonstrate the possibility of stethoscope as a vehicle for MDR bacteria such as CPE, but with several limitations. Since this study was conducted in a single hospital, the generalization of the results is limited. Due to the small number of study participants, a more detailed subgroup analysis could not be performed. Second, we chose the direct imprinting method as the contamination level of the diaphragm is higher than that of other parts of the stethoscope after performing physical examinations [8, 30], and direct imprinting method was known to be more efficient than collecting samples with a sterile cotton swab [9]. When we checked the blood agar plates on which the diaphragm was pressed, most of the bacterial colonies were detected only in urethane rim area. For this reason, the level of contamination might have been underestimated. However, as the peripheral rim was found to be the most heavily contaminated area of the stethoscope, the result can be expected to be close to the actual level of contamination [31]. Though our study only provided the effectiveness of education and the traditional disinfection methods, further study about the usefulness with a variety of options such as barriers, ultraviolet disinfection, and disposable stethoscope is needed [17, 32].

In conclusion, stethoscopes were contaminated with various nosocomial pathogens including MDR bacteria and could cause cross-transmission in hospital setting. Accurate and practical guidelines for disinfection of stethoscopes based on evidence are needed. Also, education and training to the HCWs must be conducted on an ongoing basis, using an intensive multi-disciplinary and team-based approach.

Supporting information

S1 Fig. Raw image of pulsed field gel electrophoresis of carbapenemase-producing K. pneumoniae.

M, lambda ladder; ST_7W, K. pneumoniae from the stethoscope; SM 01 to 03, K. pneumoniae isolates form the patients.

(PDF)

S1 File

(PDF)

S2 File

(PDF)

S3 File

(PDF)

S4 File

(PDF)

Acknowledgments

We would like to express our gratitude to Dr. Jong-Sung Lim at National Instrumentation Center for Environmental Management for his technical assistance for PFGE.

Data Availability

All data files are available from EPOPS database (https://doi.org/10.8888/EPOPS202101201MI).

Funding Statement

This study was supported by grant of the Institute of Clinical Medical Research in the Yeouido St. Mary’s Hospital, Catholic University of Korea (cooperative agreement no. YSI 2019-07) (to RL). The funding agencies had no role in the study design, the data collection or analysis, the decision to publish, or the preparation of the manuscript. URL: https://www.cmcsungmo.or.kr/en.common.main.main.sp.

References

  • 1.Blot K, Hammami N, Blot S, Vogelaers D, Lambert ML. Increasing burden of Escherichia coli, Klebsiella pneumoniae, and Enterococcus faecium in hospital-acquired bloodstream infections (2000–2014): A national dynamic cohort study. Infect Control Hosp Epidemiol. 2019; 40:705–709. 10.1017/ice.2019.59 [DOI] [PubMed] [Google Scholar]
  • 2.Weist K, Pollege K, Schulz I, Ruden H, Gastmeier P. How many nosocomial infections are associated with cross-transmission? A prospective cohort study in a surgical intensive care unit. Infect Control Hosp Epidemiol. 2002; 23:127–132. 10.1086/502021 [DOI] [PubMed] [Google Scholar]
  • 3.Siegel JD, Rhinehart E, Jackson M, Chiarello L. 2007 Guideline for Isolation Precautions: Preventing Transmission of Infectious Agents in Health Care Settings. Am J Infect Control. 2007; 35:S65–164. 10.1016/j.ajic.2007.10.007 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4.Harbarth S, Sax H, Gastmeier P. The preventable proportion of nosocomial infections: an overview of published reports. J Hosp Infect. 2003; 54:258–266; quiz 321. 10.1016/s0195-6701(03)00150-6 [DOI] [PubMed] [Google Scholar]
  • 5.Matsuo M, Oie S, Furukawa H. Contamination of blood pressure cuffs by methicillin-resistant Staphylococcus aureus and preventive measures. Ir J Med Sci. 2013; 182:707–709. 10.1007/s11845-013-0961-7 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6.Tadiparthi S, Shokrollahi K, Juma A, Croall J. Using marker pens on patients: a potential source of cross infection with MRSA. Ann R Coll Surg Engl. 2007; 89:661–664. 10.1308/003588407X209419 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7.Whitehead EJ, Thompson JF, Lewis DR. Contamination and decontamination of Doppler probes. Ann R Coll Surg Engl. 2006; 88:479–481. 10.1308/003588406X114866 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8.Longtin Y, Schneider A, Tschopp C, Renzi G, Gayet-Ageron A, Schrenzel J, et al. Contamination of stethoscopes and physicians’ hands after a physical examination. Mayo Clin Proc. 2014; 89:291–299. 10.1016/j.mayocp.2013.11.016 [DOI] [PubMed] [Google Scholar]
  • 9.Campos-Murguia A, Leon-Lara X, Munoz JM, Macias AE, Alvarez JA. Stethoscopes as potential intrahospital carriers of pathogenic microorganisms. Am J Infect Control. 2014; 42:82–83. 10.1016/j.ajic.2013.06.015 [DOI] [PubMed] [Google Scholar]
  • 10.Nunez S, Moreno A, Green K, Villar J. The stethoscope in the Emergency Department: a vector of infection? Epidemiol Infect. 2000; 124:233–237. 10.1017/s0950268800003563 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11.Tang PH, Worster A, Srigley JA, Main CL. Examination of staphylococcal stethoscope contamination in the emergency department (pilot) study (EXSSCITED pilot study). CJEM. 2011; 13:239–244. 10.2310/8000.2011.110242 [DOI] [PubMed] [Google Scholar]
  • 12.Uneke CJ, Ogbonna A, Oyibo PG, Onu CM. Bacterial contamination of stethoscopes used by health workers: public health implications. J Infect Dev Ctries. 2010; 4:436–441. 10.3855/jidc.701 [DOI] [PubMed] [Google Scholar]
  • 13.Knecht VR, McGinniss JE, Shankar HM, Clarke EL, Kelly BJ, Imai I, et al. Molecular analysis of bacterial contamination on stethoscopes in an intensive care unit. Infect Control Hosp Epidemiol. 2018:1–7. 10.1017/ice.2018.319 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 14.Boulee D, Kalra S, Haddock A, Johnson TD, Peacock WF. Contemporary stethoscope cleaning practices: What we haven’t learned in 150 years. Am J Infect Control. 2018. 10.1016/j.ajic.2018.08.005 [DOI] [PubMed] [Google Scholar]
  • 15.Vasudevan RS, Mojaver S, Chang KW, Maisel AS, Frank Peacock W, Chowdhury P. Observation of stethoscope sanitation practices in an emergency department setting. Am J Infect Control. 2018. 10.1016/j.ajic.2018.08.028 [DOI] [PubMed] [Google Scholar]
  • 16.Fafliora E, Bampalis VG, Lazarou N, Mantzouranis G, Anastassiou ED, Spiliopoulou I, et al. Bacterial contamination of medical devices in a Greek emergency department: impact of physicians’ cleaning habits. Am J Infect Control. 2014; 42:807–809. 10.1016/j.ajic.2014.03.017 [DOI] [PubMed] [Google Scholar]
  • 17.Kalra S, Reddy S. New Scope for the Stethoscope. Mayo Clin Proc Innov Qual Outcomes. 2020; 4:1–2. 10.1016/j.mayocpiqo.2020.01.001 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18.Jones JS, Hoerle D, Riekse R. Stethoscopes: a potential vector of infection? Ann Emerg Med. 1995; 26:296–299. 10.1016/s0196-0644(95)70075-7 [DOI] [PubMed] [Google Scholar]
  • 19.Clinical and Laboratory Standards Institute. Performance standards for antimicrobial susceptibility testing. Wayne, PA Clinical and Laboratory Standards Institute; 2018. [Google Scholar]
  • 20.Magiorakos AP, Srinivasan A, Carey RB, Carmeli Y, Falagas ME, Giske CG, et al. Multidrug-resistant, extensively drug-resistant and pandrug-resistant bacteria: an international expert proposal for interim standard definitions for acquired resistance. Clin Microbiol Infect. 2012; 18:268–281. 10.1111/j.1469-0691.2011.03570.x [DOI] [PubMed] [Google Scholar]
  • 21.Havill NL. Best practices in disinfection of noncritical surfaces in the health care setting: creating a bundle for success. Am J Infect Control. 2013; 41:S26–30. 10.1016/j.ajic.2012.10.028 [DOI] [PubMed] [Google Scholar]
  • 22.Goodman ER, Platt R, Bass R, Onderdonk AB, Yokoe DS, Huang SS. Impact of an environmental cleaning intervention on the presence of methicillin-resistant Staphylococcus aureus and vancomycin-resistant enterococci on surfaces in intensive care unit rooms. Infect Control Hosp Epidemiol. 2008; 29:593–599. 10.1086/588566 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 23.Holleck JL, Merchant N, Lin S, Gupta S. Can education influence stethoscope hygiene? Am J Infect Control. 2017; 45:811–812. 10.1016/j.ajic.2017.02.004 [DOI] [PubMed] [Google Scholar]
  • 24.Uneke CJ, Ndukwe CD, Nwakpu KO, Nnabu RC, Ugwuoru CD, Prasopa-Plaizier N. Stethoscope disinfection campaign in a Nigerian teaching hospital:results of a before-and-after study. J Infect Dev Ctries. 2014; 8:86–93. 10.3855/jidc.2696 [DOI] [PubMed] [Google Scholar]
  • 25.Lemmen SW, Hafner H, Zolldann D, Stanzel S, Lutticken R. Distribution of multi-resistant Gram-negative versus Gram-positive bacteria in the hospital inanimate environment. J Hosp Infect. 2004; 56:191–197. 10.1016/j.jhin.2003.12.004 [DOI] [PubMed] [Google Scholar]
  • 26.RutalaWA W, WeinstenRA, the Healthcare Infection Control Practices Advisory Committee (HICPAC). Guideline for Disinfection and Sterilization in Healthcare Facilities; 2008 [cited March, 1 2020]. Available from: https://www.cdc.gov/infectioncontrol/pdf/guidelines/disinfection-guidelines-H.pdf
  • 27.Lecat P, Cropp E, McCord G, Haller NA. Ethanol-based cleanser versus isopropyl alcohol to decontaminate stethoscopes. Am J Infect Control. 2009; 37:241–243. 10.1016/j.ajic.2008.08.006 [DOI] [PubMed] [Google Scholar]
  • 28.Raghubanshi BR, Sapkota S, Adhikari A, Dutta A, Bhattarai U, Bhandari R. Use of 90% ethanol to decontaminate stethoscopes in resource limited settings. Antimicrob Resist Infect Control. 2017; 6:68. 10.1186/s13756-017-0224-x [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 29.Kramer A, Schwebke I, Kampf G. How long do nosocomial pathogens persist on inanimate surfaces? A systematic review. BMC Infect Dis. 2006; 6:130. 10.1186/1471-2334-6-130 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 30.Tschopp C, Schneider A, Longtin Y, Renzi G, Schrenzel J, Pittet D. Predictors of Heavy Stethoscope Contamination Following a Physical Examination. Infect Control Hosp Epidemiol. 2016; 37:673–679. 10.1017/ice.2016.40 [DOI] [PubMed] [Google Scholar]
  • 31.Schmidt MG, Tuuri RE, Dharsee A, Attaway HH, Fairey SE, Borg KT, et al. Antimicrobial copper alloys decreased bacteria on stethoscope surfaces. Am J Infect Control. 2017; 45:642–647. 10.1016/j.ajic.2017.01.030 [DOI] [PubMed] [Google Scholar]
  • 32.Messina G, Burgassi S, Messina D, Montagnani V, Cevenini G. A new UV-LED device for automatic disinfection of stethoscope membranes. Am J Infect Control. 2015; 43:e61–66. 10.1016/j.ajic.2015.06.019 [DOI] [PubMed] [Google Scholar]

Decision Letter 0

Miquel Vall-llosera Camps

16 Dec 2020

PONE-D-20-26077

A quasi-experimental study on stethoscopes contamination with multidrug-resistant bacteria: Its role as a vehicle of transmission

PLOS ONE

Dear Dr. Choi,

Thank you for submitting your manuscript to PLOS ONE. After careful consideration, we feel that it has merit but does not fully meet PLOS ONE’s publication criteria as it currently stands. Therefore, we invite you to submit a revised version of the manuscript that addresses the points raised during the review process.

I would like to sincerely apologise for the delay you have incurred with your submission. It has been exceptionally difficult to secure reviewers to evaluate your study. We have now received three completed reviews; their comments are available below. Reviewer#1 and #3 have raised some concerns about the study that need to be addressed in a revision. Please provide more details on how the sample size was calculated, the sampling procedure and the use of training as an intervention.

Please submit your revised manuscript by Jan 29 2021 11:59PM. If you will need more time than this to complete your revisions, please reply to this message or contact the journal office at plosone@plos.org. When you're ready to submit your revision, log on to https://www.editorialmanager.com/pone/ and select the 'Submissions Needing Revision' folder to locate your manuscript file.

Please include the following items when submitting your revised manuscript:

  • A rebuttal letter that responds to each point raised by the academic editor and reviewer(s). You should upload this letter as a separate file labeled 'Response to Reviewers'.

  • A marked-up copy of your manuscript that highlights changes made to the original version. You should upload this as a separate file labeled 'Revised Manuscript with Track Changes'.

  • An unmarked version of your revised paper without tracked changes. You should upload this as a separate file labeled 'Manuscript'.

If you would like to make changes to your financial disclosure, please include your updated statement in your cover letter. Guidelines for resubmitting your figure files are available below the reviewer comments at the end of this letter.

If applicable, we recommend that you deposit your laboratory protocols in protocols.io to enhance the reproducibility of your results. Protocols.io assigns your protocol its own identifier (DOI) so that it can be cited independently in the future. For instructions see: http://journals.plos.org/plosone/s/submission-guidelines#loc-laboratory-protocols

We look forward to receiving your revised manuscript.

Kind regards,

Miquel Vall-llosera Camps

Senior Editor

PLOS ONE

Journal Requirements:

When submitting your revision, we need you to address these additional requirements.

1. Please ensure that your manuscript meets PLOS ONE's style requirements, including those for file naming. The PLOS ONE style templates can be found at

https://journals.plos.org/plosone/s/file?id=wjVg/PLOSOne_formatting_sample_main_body.pdf and

https://journals.plos.org/plosone/s/file?id=ba62/PLOSOne_formatting_sample_title_authors_affiliations.pdf

2. Please provide additional details regarding participant consent.

In the ethics statement in the Methods and online submission information, please state what type of consent you obtained (for instance, written or verbal, and if verbal, how it was documented and witnessed).

If your study included minors, state whether you obtained consent from parents or guardians.

If the need for consent was waived by the ethics committee, please include this information.

If you are reporting a retrospective study of medical records or archived samples, please ensure that you have discussed whether all data were fully anonymized before you accessed them and/or whether the IRB or ethics committee waived the requirement for informed consent. If patients provided informed written consent to have data from their medical records used in research, please include this information.

3. Please include additional information regarding the pre- and post-intervention questionnaire used in the study and ensure that you have provided sufficient details that others could replicate the analyses.

For instance, if you developed a questionnaire as part of this study and it is not under a copyright more restrictive than CC-BY, please include a copy, in both the original language and English, as Supporting Information.

4. We note that you have stated that you will provide repository information for your data at acceptance. Should your manuscript be accepted for publication, we will hold it until you provide the relevant accession numbers or DOIs necessary to access your data. If you wish to make changes to your Data Availability statement, please describe these changes in your cover letter and we will update your Data Availability statement to reflect the information you provide.

[Note: HTML markup is below. Please do not edit.]

Comments to the Author

1. Is the manuscript technically sound, and do the data support the conclusions?

The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented.

Reviewer #1: Yes

Reviewer #2: Yes

Reviewer #3: Partly

**********

2. Has the statistical analysis been performed appropriately and rigorously?

Reviewer #1: Yes

Reviewer #2: Yes

Reviewer #3: I Don't Know

**********

3. Have the authors made all data underlying the findings in their manuscript fully available?

The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified.

Reviewer #1: Yes

Reviewer #2: Yes

Reviewer #3: Yes

**********

4. Is the manuscript presented in an intelligible fashion and written in standard English?

PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here.

Reviewer #1: Yes

Reviewer #2: Yes

Reviewer #3: Yes

**********

5. Review Comments to the Author

Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters)

Reviewer #1: This is a well-conducted study regarding the effect of an educational program as an intervention to decrease the contamination rate of stethoscopes in use within a hospital. However, in this regard I have some observations:

1. Although the program was shown to have a positive effect on cleaning habits and cleaning frequency in stethoscopes, no effect on the contamination rate was observed. The authors focus the discussion of this point on aspects related to the training program; however, they do not consider the nature of the disinfectant used, which could also have an effect on the contamination rate. The disinfectant reported by the participants was ethyl alcohol, which has a good effect on bacteria, fungi and even viruses; however, it evaporates quickly, allowing re-contamination of the stethoscope in the time after cleaning, where the stethoscope can be in contact with the patient's environment or the patient himself.

2. The authors make the following statement in the discussion: “This is the first study to evaluate the contamination rate of stethoscopes by nosocomial pathogens and MDR bacteria and to demonstrate the possibility of stethoscope as a vehicle for MDR bacteria such as CPE, but with several limitations. " This statement is not entirely correct and in any case, they must show evidence to support such statement.

3. In table 3 the names of the isolated bacteria must be written according to the binomial system, following the nomenclature rules for publication in scientific journals.

Reviewer #2: This is an extremely well written article in a challenging area of research that has never been published before. They evaluated the potential for the stethoscope to trasmit multidrug resistant (MDR) organisms, as well as the impact of education on stethoscope hygiene. While the latter has previously been shown to have little impact (and is consistent with the authors findings), the former is uniquely important, as it clearly demonstrate the potential for the stethoscope to serve as a MDR organism vector. This paper will be frequently cited, and will serve as an important contribution to the literature. I have reviewed over 1000 articles for various journals (including PLOS ONE) in my career. I have NEVER recommended acceptance without a number of suggestions. The only suggestion I have for this manuscript is that they suggest methods that may decrease the potential for MDR transmission by the stethoscope. There are a number of stethoscope hygiene strategies on the market that include barriers, caps, UV light, and disposable stethoscopes (although these are terrible tools) that do not require health care practioner cleaning (which has been repetitively proven to be an ineffective intervention). It would be reasonable to suggest them here. Further, it is a covid world, and the potential for the stethoscope being a vector may have greater impact than when this reasearch started. Here are a few citations. Thank you for the privilege of reviewing this excellent manuscript. New Scope for the Stethoscope. Kalra S, Reddy S. Mayo Clin Proc Innov Qual Outcomes. 2020 Feb 5;4(1):1-2. Aseptic Barriers Allow a Clean Contact for Contaminated Stethoscope Diaphragms. Vasudevan R, Shin JH, Chopyk J, Peacock WF, Torriani FJ, Maisel AS, Pride DT. Mayo Clin Proc Innov Qual Outcomes. 2020 Feb 5;4(1):21-30. The stethoscope: a potential vector for COVID-19? Vasudevan RS, Bin Thani K, Aljawder D, Maisel S, Maisel AS. Eur Heart J. 2020 Sep 21;41(36):3393-3395. Comparing the auscultatory accuracy of health care professionals using three different brands of stethoscopes on a simulator. Mehmood M, Abu Grara HL, Stewart JS, Khasawneh FA. Med Devices (Auckl). 2014 Aug 14;7:273-81.

Reviewer #3: I just congratulate all the authors for this interesting study.

Line 22: please include a paragraph that describe HCWs stethoscope disinfection practice of HCWs from previously conducted studies and the challenges behind/ or the driving force/. Use the following article as an example and many others from PubMed

Ex:" Sahiledengle B. Stethoscope disinfection is rarely done in Ethiopia: What are the associatedfactors?PLoSONE.2019: 14(6):e0208365. https://doi.org/10.1371/journal.pone.0208365 "

Line 36: Does the authors included all HCWs from the specific dep't ?

Line 39: self-administered or what ? what about the validity and reliability of the data collection tool ?

line 45: "conducting a training for a month" is not clear: does the training given for 30 days? or how many days; how many hours ? when ? should be clear

Line 95: We did not see a multivariate logistic regression table in the manuscript ?

**********

6. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files.

If you choose “no”, your identity will remain anonymous but your review may still be made public.

Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy.

Reviewer #1: Yes: José A. Alvarez-Canales

Reviewer #2: Yes: W. Frank Peacock, MD, FACEP, FACC, FESC

Reviewer #3: Yes: Biniyam Sahiledengle

[NOTE: If reviewer comments were submitted as an attachment file, they will be attached to this email and accessible via the submission site. Please log into your account, locate the manuscript record, and check for the action link "View Attachments". If this link does not appear, there are no attachment files.]

While revising your submission, please upload your figure files to the Preflight Analysis and Conversion Engine (PACE) digital diagnostic tool, https://pacev2.apexcovantage.com/. PACE helps ensure that figures meet PLOS requirements. To use PACE, you must first register as a user. Registration is free. Then, login and navigate to the UPLOAD tab, where you will find detailed instructions on how to use the tool. If you encounter any issues or have any questions when using PACE, please email PLOS at figures@plos.org. Please note that Supporting Information files do not need this step.

PLoS One. 2021 Apr 22;16(4):e0250455. doi: 10.1371/journal.pone.0250455.r002

Author response to Decision Letter 0


13 Mar 2021

We would like to thank editor and all reviewers for the invaluable comments and advices that help a great deal in improving our original manuscript.

Attachment

Submitted filename: PLOS one_Response to Reviewer.docx

Decision Letter 1

Jose Alvarez

7 Apr 2021

A quasi-experimental study on stethoscopes contamination with multidrug-resistant bacteria: Its role as a vehicle of transmission

PONE-D-20-26077R1

Dear Dr. Su-Mi Choi,

We’re pleased to inform you that your manuscript has been judged scientifically suitable for publication and will be formally accepted for publication once it meets all outstanding technical requirements.

Within one week, you’ll receive an e-mail detailing the required amendments. When these have been addressed, you’ll receive a formal acceptance letter and your manuscript will be scheduled for publication.

An invoice for payment will follow shortly after the formal acceptance. To ensure an efficient process, please log into Editorial Manager at http://www.editorialmanager.com/pone/, click the 'Update My Information' link at the top of the page, and double check that your user information is up-to-date. If you have any billing related questions, please contact our Author Billing department directly at authorbilling@plos.org.

If your institution or institutions have a press office, please notify them about your upcoming paper to help maximize its impact. If they’ll be preparing press materials, please inform our press team as soon as possible -- no later than 48 hours after receiving the formal acceptance. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information, please contact onepress@plos.org.

Kind regards,

Jose Alvarez

Guest Editor

PLOS ONE

Acceptance letter

Jose Alvarez

12 Apr 2021

PONE-D-20-26077R1

A quasi-experimental study on stethoscopes contamination with multidrug-resistant bacteria: Its role as a vehicle of transmission

Dear Dr. Choi:

I'm pleased to inform you that your manuscript has been deemed suitable for publication in PLOS ONE. Congratulations! Your manuscript is now with our production department.

If your institution or institutions have a press office, please let them know about your upcoming paper now to help maximize its impact. If they'll be preparing press materials, please inform our press team within the next 48 hours. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information please contact onepress@plos.org.

If we can help with anything else, please email us at plosone@plos.org.

Thank you for submitting your work to PLOS ONE and supporting open access.

Kind regards,

PLOS ONE Editorial Office Staff

on behalf of

Dr. Jose Alvarez

Guest Editor

PLOS ONE

Associated Data

    This section collects any data citations, data availability statements, or supplementary materials included in this article.

    Supplementary Materials

    S1 Fig. Raw image of pulsed field gel electrophoresis of carbapenemase-producing K. pneumoniae.

    M, lambda ladder; ST_7W, K. pneumoniae from the stethoscope; SM 01 to 03, K. pneumoniae isolates form the patients.

    (PDF)

    S1 File

    (PDF)

    S2 File

    (PDF)

    S3 File

    (PDF)

    S4 File

    (PDF)

    Attachment

    Submitted filename: PLOS one_Response to Reviewer.docx

    Data Availability Statement

    All data files are available from EPOPS database (https://doi.org/10.8888/EPOPS202101201MI).


    Articles from PLoS ONE are provided here courtesy of PLOS

    RESOURCES