Skip to main content
Medicine logoLink to Medicine
. 2021 Apr 23;100(16):e25638. doi: 10.1097/MD.0000000000025638

Comparison of corneal thickness measurements using ultrasound pachymetry, noncontact tonopachy, Pentacam HR, and Fourier-domain OCT

Jung Sub Kim a, Chang Rae Rho b, Yeon Woo Cho b, Jeongah Shin b,
Editor: Rosa E Alvarado-Villacorta
PMCID: PMC8078248  PMID: 33879743

Abstract

We retrospectively compared the central corneal thickness (CCT) obtained by ultrasound pachymetry (USP; SP-3000, Tomey Corp., Nagoya, Japan), non-contact tonopachy (TP) (NT-530P, Nidek Co., Ltd., Gamagori, Japan), Pentacam HR (OCULUS Inc., Wetzlar, Germany), and RTVue optical coherence tomography (OCT) (Optovue Inc., Fremont, CA, USA) in 78 eyes of 78 healthy subjects with myopia. Agreement between the measurement methods was evaluated using 95% confidence intervals for the limits of agreement (LoA). The mean CCT values were 546.9 ± 34.7, 548.1 ± 33.5, 559.2 ± 34.0, and 547.2 ± 34.8 μm for USP, non-contact TP, Pentacam, and RTVue, respectively. The thickest and the thinnest mean CCT values corresponded to those obtained by Pentacam HR and USP, respectively. Plots of the differences against the means showed the best agreement between USP and RTVue (LoA, 10.14–10.70 μm), while the largest discrepancy was observed between RTVue and Pentacam systems (LoA, −25.47–1.44 μm). Our data showed that CCT measurements using these 4 instruments were well correlated. However, the results from Pentacam differed significantly from those of the other instruments.

Keywords: corneal pachymetry, optical coherence tomography, Pentacam, tonopachy, ultrasound pachymetry

1. Introduction

Central corneal thickness (CCT) measurement is an integral component of ophthalmic examination. In glaucoma, a thin CCT is a significant risk factor for intraocular pressure (IOP) underestimation.[1,2] Accurate measurement of CCT is required when evaluating candidates for refractive surgery. Low residual stromal bed thickness is a risk factor for development of keratectasia after surgery.[3] Corneal thinning is known to be associated with keratoconus progression.[4] Additionally, CCT reflects the progression of Fuchs endothelial corneal dystrophy.[5] For patients with corneal problems, CCT measurements allow determination of disease status.

Several instruments are available for the measurement of CCT, each of which is based on specific principles. Contact ultrasound pachymetry (USP) is the most commonly accepted method.[6,7] In USP, a probe in contact with the anterior surface of the cornea detects the interface where the speed of sound changes. While USP is considered the gold standard in pachymetry, its measurement can be influenced by several factors. The probe should be placed perpendicular to the center of the cornea. To avoid possible corneal indentation, the operator must be trained properly and the patient must cooperate well during the examination. Moreover, this method carries additional risks such as corneal infection and erosion.[8]

Other instruments used for pachymetry are mostly automated and do not involve direct contact with the cornea. The Pentacam system (OCULUS Inc., Wetzlar, Germany) uses a rotating Scheimpflug camera to acquire cross-sectional images of the cornea. From these images, which contain up to 25,000 data points, Pentacam can calculate CCT, corneal curvature, anterior chamber angle, and anterior chamber depth.[9] RTVue (Optovue Inc., Fremont, CA, USA) is a high-speed, high-resolution optical coherence tomography (OCT) system. With the corneal anterior module option, RTVue examines the cornea and anterior segment. Non-contact tonopachy (TP) (NT-530P, Nidek, Japan) is a combination unit equipped with a non-contact tonometer and a pachymeter. The NT-530P system measures CCT with Scheimpflug camera and displays the compensated IOP, which is calculated from the measured CCT. The reliability of non-contact tonopachymeters has not been widely tested. Additionally, pachymetry evaluation is essential to confirm CCT-corrected IOP.

The purpose of this study was to compare CCT measurements obtained with USP, noncontact TP, the Pentacam system, and RTVue OCT in healthy subjects with myopia.

2. Materials and methods

2.1. Subjects

This retrospective case series study assessed 78 eyes of 78 patients who underwent preoperative examination for refractive surgery at B and Viit Eye Center between January 2017 and August 2017. To avoid inter-eye correlation, only right eye from each patient was included. This study was approved by the Institutional Review Board of Daejeon St. Mary's Hospital (DC20RISI0031) and conducted in accordance with the tenets of the Declaration of Helsinki. Patients with any of the following were excluded: history of previous ocular surgery, corneal disease, soft contact lens wear within one-week, dry eye, or ocular or systemic medication use that would likely affect corneal thickness. Four instruments were used for the measurement of corneal thickness: an NT-530P (Nidek Co., Ltd., Gamagori, Japan) for non-contact TP, an RTVue OCT (Optovue, Inc., Fremont, CA, USA) for spectral domain OCT, a Pentacam Scheimpflug imaging system (Pentacam, Oculus, Wetzlar, Germany), and an SP-3000 (Tomey Corp., Nagoya, Japan) for USP. Three non-contact measurements were followed by the application of a topical anesthetic to obtain USP data. Two examiners performed the measurements, and the mean value was used for further analysis.

2.2. CCT measurements with non-contact tonopachy

Non-contact TP measures corneal thickness using Scheimpflug imaging and calculates the compensated IOP based on the measured thickness.[10] Measurements were obtained automatically after approximately focusing the mires on the screen. CCT values were recorded for each of the 3 IOP measurements. Mean CCT values were used for the analysis.

2.3. CCT measurements with a rotating Scheimpflug camera

Pentacam HR was used to acquire Scheimpflug data.[11,12] An image was built from 50 slit-beam images using a single rotating Scheimpflug camera. CCT was recorded from the corneal thickness at the pachy apex (0, 0) displayed by Pentacam software.

2.4. CCT measurements with RTVue OCT

RTVue has a depth resolution of 5 μm in tissue for a light source working at 830 nm.[13] A corneal anterior module was added to image the anterior segment and to obtain the corneal thickness. The scans were centered on the coaxially fixating corneal light reflex observed by the central bright reflection on the OCT scan. A pachymetry map was created automatically with 3 zones: central (0–2 mm), pericentral (2–5 mm), and transitional (5–6 mm). The corneal thickness value in the central 2-mm zone was used for the analysis.

2.5. CCT measurements with ultrasound pachymetry

After applying 1 drop of 0.5% proparacaine hydrochloride, patients were asked to look at a distant, fixed point in sitting position. The USP probe was placed perpendicular to the central surface of the cornea. Three consecutive measurements were acquired, and the mean value was used for the analysis.

2.6. Statistics

All data were analyzed using SPSS for Windows software (version 19.0; Microsoft Corp., Redmond, WA, USA). Normal distributions were confirmed using the Kolmogorov-Smirnov test. CCT was compared using one-way repeated measures analysis of variance (ANOVA). Paired comparisons were further evaluated using Bonferroni adjustment. Bland–Altman plots were used to evaluate the agreement between test instruments. The limits of agreement (LOA) were calculated as the mean difference between 2 measurements ± 1.96 times the standard deviation of the difference. The R-values between the instruments were obtained using Pearson's correlation. The percentage of corneas measured within a 5-μm difference was calculated between different measurements.

3. Results

This study included 78 eyes from 78 refractive surgery candidates. Table 1 shows demographic data of the patients including their sex, mean age, median age, and spherical equivalent refraction. Table 2 displays the mean CCT with standard deviations and the CCT range of each instrument. The Pentacam and USP systems provided the thickest and the thinnest CCT values, respectively. The differences in CCT values obtained from USP compared to TP, Pentacam, and RTVue systems were 1.22 ± 5.47, 12.29 ± 5.87, and 0.28 ± 5.32 μm, respectively. Table 3 shows the mean bias, bias range, LOA, width of limits, significance level for Bonferroni analysis, R-value, and percentage of corneas measured within a 5-μm difference.

Table 1.

Demographic data.

Characteristics Value
Candidates/eyes 78/78
Sex 42% male, 58% female
Mean age, yr 25.5 ± 6.6
Median age, yr 23 (18 to 55)
Spherical equivalent refraction, Diopters −4.44 ± 2.08 (−11.625 to −0.75)

Table 2.

Mean results of ultrasound pachymetry, tonopachy, Pentacam, and RTVue.

Device CCT values (Mean ± SD) (μm) Minimum (μm) Maximum (μm)
USP 546.9 ± 34.7 469 605
TP 548.1 ± 33.5 470 607
Pentacam 559.2 ± 34.0 481 624
RTVue 547.2 ± 34.8 473 608

CCT = central corneal thickness, SD = standard deviation, TP = tonopachy, USP = ultrasound pachymetry.

Table 3.

Differences between ultrasound pachymetry, noncontact tonopachy, Pentacam, and RTVue.

Parameter TP-USP Pen-USP RTVue-USP Pen-TP RTVue-TP RTVue-Pen
Mean bias in CCT (μm) 1.22 12.29 0.28 11.08 −0.94 −12.01
SD (μm) 5.47 5.87 5.32 6.54 5.78 6.87
Maximum (μm) 15 30 18 28 14 7
Minimum (μm) −12 −2 −9 −7 −14 −26
95% CI for the mean (μm) −0.02 to 2.45 10.97 to 13.62 −0.92 to 1.48 9.60 to 12.55 −2.24 to 0.37 −13.56 to -10.47
LOA (μm) −9.50 to 11.93 0.78 to 23.81 −10.14 to 10.70 −1.75 to 23.90 −12.27 to 10.39 −25.47 to 1.44
Width of limits (μm) 21.43 23.03 20.84 25.65 22.66 26.91
95% CI for the LOA −11.61 to 14.04 −1.49 to 26.07 −12.19 to 12.75 −4.28 to 26.43 −14.5 to 12.63 −28.12 to 4.09
P (one-way ANOVA, Bonferroni) 0.316 <0.01 1.000 <0.01 0.941 <0.01
R-value 0.988 0.986 0.988 0.981 0.986 0.980
Within 5-μm difference (%) 73.1 10.3 71.8 19.2 69.2 17.9

ANOVA = analysis of variance, CCT = central corneal thickness, CI = confidence interval, LOA = limits of agreement, Pen = Pentacam, SD = standard deviation, TP = tonopachy, USP = ultrasound pachymetry.

All instruments showed a high correlation between measurements (R = 0.980–0.988). However, Bonferroni analysis showed that Pentacam CCT measurements differed significantly from those of other 3 instruments, whose values showed little or no difference (Table 3). The mean difference in CCT was largest between Pentacam and RTVue measurements. The percentage of corneas measured within a 5-μm difference was greatest, 73.1%, between USP and TP measurements. The smallest percentage was 10.3% between Pentacam and USP measurements.

The Bland–Altman plots in Figure 1 provide a graphical representation of the mean difference and LOA. The solid line represents the mean value, and the dotted lines show the agreement limits, with the smallest limit of agreement between RTVue and USP results.

Figure 1.

Figure 1

Bland–Altman plots comparing the central corneal thicknesses measured by (A) tonopachy (TP) and ultrasound pachymetry (USP), (B) Pentacam and USP, (C) RTVue and USP, (D) Pentacam and TP, (E) RTVue and TP, and (F) RTVue and Pentacam. The solid line represents the mean difference, and the dashed lines show the 95% confidence intervals for the limits of agreement.

4. Discussion

Our study investigated the agreement in CCT measurements using USP, non-contact TP, Pentacam, and RTVue OCT systems. The results revealed well-correlated CCT values. However, the measurements obtained by Pentacam differed significantly from those by other 3 instruments, while non-contact TP measurements were comparable to USP measurements.

In our study, USP, RTVue, TP, and Pentacam showed average CCT values in descending order. The difference between measurements was smallest by 0.28 μm in RTVue and USP. The differences were also small between RTVue and TP values and between TP and USP values. A comparison between TP and USP measurements showed that values from 73.1% of the eyes were within 5 μm. However, Pentacam measurements differed significantly from those of other 3 instruments. The largest gap of 12.29 μm was observed between Pentacam and USP readings; values from only 10.3% of eyes were within 5 μm. However, the variability in agreement was within the LOA. The limits were narrowest between RTVue and USP values (20.84 μm) and widest between RTVue and Pentacam values (26.91 μm).

Non-contact CCT measurement has recently gained much attention as it offers several advantages over contact methods. For example, contact with the corneal surface causes patient's discomfort, which can interfere with accurate measurements. Instilling an anesthetic eye drop can be bothersome in some patients. Additionally, probe contact can lead to corneal erosion and infection.

Among non-contact measurement methods, TP measures the corneal thickness and provides corrected IOP as opposed to other conventional tonometers. Accurate measurement of corneal thickness is a prerequisite for providing a precisely corrected IOP. Therefore, it would be important to investigate the reliability of CCT measurements in the TP.

Our results showed that TP measurements did not differ significantly from USP measurements: the average difference was 1.22 μm. Our results contradict those of earlier studies. Gonzalez-Perez et al compared non-contact TP with standard USP and found that TP underestimated CCT by 33.1 ± 33.3 μm.[14] However, their TP approach adopted a specular microscope method for pachymetric analysis.

In a separate study, Bao et al compared CCT obtained by 3 non-contact specular microscopes with that obtained by USP in 70 healthy subjects.[15] The mean CCTs measured by 3 non-contact microscopes were 513.66 ± 33.14, 529.12 ± 33.22, and 549.06 ± 40.27 μm, and the mean CCT obtained by USP was 539.01 ± 35.73 μm. Although non-contact specular microscopes and USP showed high intra-operator repeatability, the inter-device agreement was poor. Other studies have also found that non-contact specular microscopes underestimated CCT compared to USP.[16,17] Taken together, these outcomes imply the importance of scrutinizing corneal thickness measurements using non-contact pachymetric machines. In this study, TP used the Scheimpflug imaging principle to measure corneal thickness.

The accuracy of Pentacam CCT measurements, compared to USP has been challenged. While in some studies Pentacam system underestimated CCT,[18,19] in other studies Pentacam measurements were comparable to[20,21] or overestimated[2224] CCT compared to USP. In our study, Pentacam yielded overestimation of CCT. Al-Ageel et al considered the effect of ultrasound probe displacing 7 to 40-μm-thick tear film, resulting in thinner measurements.[25] They also suggested that differences in corneal hydration after local corneal anesthesia could affect differences in measurement.[25] Also, USP probe may indent the corneal epithelium.[26]

Non-contact TP using the Scheimpflug principle provided results that were similar to those obtained with USP method. However, Pentacam, which uses a rotating Scheimpflug camera, overestimated CCT. It is unclear why the measurements differed between the 2 instruments that use the same principle. One possibility could be the incorporation of a correction factor by the manufacturer; Orbscan manufacturers recommend incorporating an acoustic factor to compensate for overestimation.[27] Another explanation is that the same principle does not necessarily produce the same results. For example, a study comparing CCT measurements obtained with GALILEI, a dual Scheimpflug system, and Pentacam indicated a difference of 18.26 μm, with a thicker measurement obtained using the GALILEI system.[18]

Several studies comparing corneal thickness readings between Pentacam and RTVue systems found Pentacam measurements to be thicker than RTVue measurements by 5.16 ± 19.09, 10.52 ± 5.28, 10.9 ± 5.93, and 21.9 ± 10.6 μm.[24,2830] In our study, Pentacam reading was thicker than that of RTVue system by 12.01 ± 6.87 μm, which was approximately at middle range of previous reports.

Two recent reports compared corneal thicknesses obtained with swept-source optical coherence tomography (CASIA SS-1000; Tomey, Nagoya, Japan or CASIA2; Tomey, Nagoya, Japan) and Pentacam.[31,32] Both reports found that Pentacam CCT measurements were thicker than SS-OCT (swept-source optical coherence tomography), by 4.91 and 9.64 μm, respectively. Zhao et al insisted that SS-OCT outperformed the Scheimpflug-based corneal topographic map in measuring corneal thickness. The coefficients of variation for corneal thickness measurements were smaller with CASIA SS-1000 than with Pentacam HR.[31] Li et al emphasized that SS-OCT has the advantage of shorter scanning time and utilizing an infrared light source to minimize the effects of instrument light on pupil movement.[32]

Our study had certain limitations. First, all subjects included in this study were myopic patients who underwent examinations for refractive surgery. Patients with emmetropia or hyperopia were not included in this study. Second, we did not explore instrumental correction factors. Third, the data collection may have been flawed. Armstrong suggested that if both eyes are to be included in an evaluation, the inter-eye correlation should be assessed using the intraclass correlation coefficient.[33] Also, the author viewed that eye should be randomly selected if only one of 2 eligible eyes should be included.[33] Unfortunately, we evaluated a non-random data set, uniformly including only right eye of each patient. Murdoch et al suggested that in a condition where either eye can be equally affected, analyses made on only right, or on only left, or on a randomly selected eye are statistically equivalent.[34] Our participants were normal in terms of eye structure except for myopia, and we deemed that laterality would not influence corneal thickness in otherwise normal subjects. However, 2 concerns persist. First, only half of the available data were included in the analysis. Second, selection bias may have been in play considering that only right eye, instead of randomly selected eye, was included in this study. However, it should be mentioned that all subjects were candidates for refractive surgery with normal best-corrected visual acuity in both of their eyes, providing a rationale for our selection of right eyes.

In this study, we measured and compared CCT with 4 different instruments, and found that their readings were closely correlated in general. However, CCT measurements made using Pentacam system were significantly thicker than those using other 3 systems. The difference in average thickness between-RTVue-and-USP was minimal.

Author contributions

Conceptualization: Jung Sub Kim.

Data curation: Jung Sub Kim.

Formal analysis: Jung Sub Kim.

Investigation: Jung Sub Kim, Chang Rae Rho.

Methodology: Chang Rae Rho, Yeon Woo Cho.

Project administration: Chang Rae Rho, Jeongah Shin.

Resources: Yeon Woo Cho.

Software: Chang Rae Rho, Yeon Woo Cho, Jeongah Shin.

Supervision: Jeongah Shin.

Validation: Chang Rae Rho, Yeon Woo Cho.

Visualization: Yeon Woo Cho.

Writing – original draft: Jung Sub Kim, Chang Rae Rho.

Writing – review & editing: Jeongah Shin.

Correction

The asterisk indicating corresponding author was incorrectly placed on Dr. Jung Sub Kim in the byline. It has been corrected to Dr. Jeongah Shin.

Footnotes

Abbreviations: CCT = central corneal thickness, LoA = limits of agreement, OCT = optical coherence tomography, TP = tonopachy, USP = ultrasound pachymetry.

How to cite this article: Kim JS, Rho CR, Cho YW, Shin J. Comparison of corneal thickness measurements using ultrasound pachymetry, noncontact tonopachy, Pentacam HR, and Fourier-domain OCT. Medicine. 2021;100:16(e25638).

JSK and CRR both contributed equally to this work.

The authors have no funding and conflicts of interests to disclose.

The datasets generated during and/or analyzed during the current study are publicly available.

References

  • [1].Francis BA, Varma R, Chopra V, et al. Intraocular pressure, central corneal thickness, and prevalence of open-angle glaucoma: the Los Angeles Latino Eye Study. Am J Ophthalmol 2008;146:741–6. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [2].Herndon LW, Weizer JS, Stinnett SS. Central corneal thickness as a risk factor for advanced glaucoma damage. Arch Ophthalmol 2004;122:17–21. [DOI] [PubMed] [Google Scholar]
  • [3].Giri P, Azar DT. Risk profiles of ectasia after keratorefractive surgery. Curr Opin Ophthalmol 2017;28:337–42. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [4].Cinar Y, Cingu AK, Turkcu FM, et al. Comparison of central corneal thickness measurements with a rotating scheimpflug camera, a specular microscope, optical low-coherence reflectometry, and ultrasound pachymetry in keratoconic eyes. Semin Ophthalmol 2015;30:105–11. [DOI] [PubMed] [Google Scholar]
  • [5].Kopplin LJ, Przepyszny K, Schmotzer B, et al. Relationship of Fuchs endothelial corneal dystrophy severity to central corneal thickness. Arch Ophthalmol 2012;130:433–9. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [6].Khaja WA, Grover S, Kelmenson AT, et al. Comparison of central corneal thickness: ultrasound pachymetry versus slit-lamp optical coherence tomography, specular microscopy, and Orbscan. Clin Ophthalmol 2015;9:1065–70. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [7].Sedaghat MR, Daneshvar R, Kargozar A, et al. Comparison of central corneal thickness measurement using ultrasonic pachymetry, rotating Scheimpflug camera, and scanning-slit topography. Am J Ophthalmol 2010;150:780–9. [DOI] [PubMed] [Google Scholar]
  • [8].Scotto R, Bagnis A, Papadia M, et al. Comparison of central corneal thickness measurements using ultrasonic pachymetry, anterior segment oct and noncontact specular microscopy. J Glaucoma 2017;26:860–5. [DOI] [PubMed] [Google Scholar]
  • [9].O’Donnell C, Hartwig A, Radhakrishnan H. Comparison of central corneal thickness and anterior chamber depth measured using LenStar LS900, Pentacam, and Visante AS-OCT. Cornea 2012;31:983–8. [DOI] [PubMed] [Google Scholar]
  • [10].Garcia-Resua C, Pena-Verdeal H, Minones M, et al. Reliability of the non-contact tono-pachymeter tonopachy NT-530P in healthy eyes. Clin Exp Optom 2013;96:286–94. [DOI] [PubMed] [Google Scholar]
  • [11].Lackner B, Schmidinger G, Pieh S, et al. Repeatability and reproducibility of central corneal thickness measurement with Pentacam, orbscan, and ultrasound. Optom Vis Sci 2005;82:892–9. [DOI] [PubMed] [Google Scholar]
  • [12].Barkana Y, Gerber Y, Elbaz U, et al. Central corneal thickness measurement with the Pentacam Scheimpflug system, optical low-coherence reflectometry pachymeter, and ultrasound pachymetry. J Cataract Refract Surg 2005;31:1729–35. [DOI] [PubMed] [Google Scholar]
  • [13].Huang J, Ding X, Savini G, et al. A Comparison between Scheimpflug imaging and optical coherence tomography in measuring corneal thickness. Ophthalmology 2013;120:1951–8. [DOI] [PubMed] [Google Scholar]
  • [14].Gonzalez-Perez J, Queiruga Pineiro J, Sanchez Garcia A, et al. Comparison of central corneal thickness measured by standard ultrasound pachymetry, corneal topography, tono-pachymetry and anterior segment optical coherence tomography. Curr Eye Res 2018;43:866–72. [DOI] [PubMed] [Google Scholar]
  • [15].Bao F, Wang Q, Cheng S, et al. Comparison and evaluation of central corneal thickness using 2 new noncontact specular microscopes and conventional pachymetry devices. Cornea 2014;33:576–81. [DOI] [PubMed] [Google Scholar]
  • [16].Maloca PM, Studer HP, Ambrosio R, Jr, et al. Interdevice variability of central corneal thickness measurement. PLoS One 2018;13:e0203884. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [17].Almubrad TM, Osuagwu UL, Alabbadi I, et al. Comparison of the precision of the Topcon SP-3000P specular microscope and an ultrasound pachymeter. Clin Ophthalmol 2011;5:871–6. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [18].Jahadi Hosseini HR, Katbab A, Khalili MR, et al. Comparison of corneal thickness measurements using Galilei, HR Pentacam, and ultrasound. Cornea 2010;29:1091–5. [DOI] [PubMed] [Google Scholar]
  • [19].de Sanctis U, Missolungi A, Mutani B, et al. Reproducibility and repeatability of central corneal thickness measurement in keratoconus using the rotating Scheimpflug camera and ultrasound pachymetry. Am J Ophthalmol 2007;144:712–8. [DOI] [PubMed] [Google Scholar]
  • [20].Kim SW, Byun YJ, Kim EK, et al. Central corneal thickness measurements in unoperated eyes and eyes after PRK for myopia using Pentacam, Orbscan II, and ultrasonic pachymetry. J Refract Surg 2007;23:888–94. [DOI] [PubMed] [Google Scholar]
  • [21].Ciolino JB, Khachikian SS, Belin MW. Comparison of corneal thickness measurements by ultrasound and scheimpflug photography in eyes that have undergone laser in situ keratomileusis. Am J Ophthalmol 2008;145:75–80. [DOI] [PubMed] [Google Scholar]
  • [22].Al-Mezaine HS, Al-Amro SA, Kangave D, et al. Comparison between central corneal thickness measurements by oculus pentacam and ultrasonic pachymetry. Int Ophthalmol 2008;28:333–8. [DOI] [PubMed] [Google Scholar]
  • [23].Al-Mezaine HS, Al-Amro SA, Kangave D, et al. Comparison of central corneal thickness measurements using Pentacam and ultrasonic pachymetry in post-LASIK eyes for myopia. Eur J Ophthalmol 2010;20:852–7. [DOI] [PubMed] [Google Scholar]
  • [24].Yap TE, Archer TJ, Gobbe M, et al. Comparison of central corneal thickness between Fourier-domain OCT, very high-frequency digital ultrasound, and Scheimpflug imaging systems. J Refract Surg 2016;32:110–6. [DOI] [PubMed] [Google Scholar]
  • [25].Al-Ageel S, Al-Muammar AM. Comparison of central corneal thickness measurements by Pentacam, noncontact specular microscope, and ultrasound pachymetry in normal and post-LASIK eyes. Saudi J Ophthalmol 2009;23:181–7. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [26].Ho WC, Lam PT, Chiu TY, et al. Comparison of central corneal thickness measurement by scanning slit topography, infrared, and ultrasound pachymetry in normal and post-LASIK eyes. Int Ophthalmol 2020;40:2913–21. [DOI] [PubMed] [Google Scholar]
  • [27].Ladi JS, Shah NA. Comparison of central corneal thickness measurements with the Galilei dual Scheimpflug analyzer and ultrasound pachymetry. Indian J Ophthalmol 2010;58:385–8. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [28].Gonul S, Koktekir BE, Bakbak B, et al. Comparison of central corneal thickness measurements using optical low-coherence reflectometry, Fourier domain optical coherence tomography, and Scheimpflug camera. Arq Bras Oftalmol 2014;77:345–50. [DOI] [PubMed] [Google Scholar]
  • [29].Huang J, Pesudovs K, Yu A, et al. A comprehensive comparison of central corneal thickness measurement. Optom Vis Sci 2011;88:940–9. [DOI] [PubMed] [Google Scholar]
  • [30].Chen S, Huang J, Wen D, et al. Measurement of central corneal thickness by high-resolution Scheimpflug imaging, Fourier-domain optical coherence tomography and ultrasound pachymetry. Acta Ophthalmol 2012;90:449–55. [DOI] [PubMed] [Google Scholar]
  • [31].Zhao Y, Chen D, Savini G, et al. The precision and agreement of corneal thickness and keratometry measurements with SS-OCT versus Scheimpflug imaging. Eye Vis (Lond) 2020;7:32. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [32].Li X, Zhou Y, Young CA, et al. Comparison of a new anterior segment optical coherence tomography and Oculus Pentacam for measurement of anterior chamber depth and corneal thickness. Ann Transl Med 2020;8:857. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [33].Armstrong RA. Statistical guidelines for the analysis of data obtained from one or both eyes. Ophthalmic Physiol Opt 2013;33:07–14. [DOI] [PubMed] [Google Scholar]
  • [34].Murdoch IE, Morris SS, Cousens SN. People and eyes: statistical approaches in ophthalmology. Br J Ophthalmol 1998;82:971–3. [DOI] [PMC free article] [PubMed] [Google Scholar]

Articles from Medicine are provided here courtesy of Wolters Kluwer Health

RESOURCES